A comparison of blood glucose meters in Australia



Similar documents
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care

Blood Glucose Monitoring: The Facts about Accuracy

Validation of measurement procedures

Point-of-care (POC) versus central laboratory instrumentation for monitoring oral anticoagulation

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

POCT in diagnosing and monitoring of Diabetes Mellitus. 13th EFLM Continuous Postgraduate Course, Sverre Sandberg, Noklus / EFLM

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

Symmetry in Diagnostics HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. TOSOH BIOSCIENCE

tips Insulin Pump Users 1 Early detection of insulin deprivation in continuous subcutaneous 2 Population Study of Pediatric Ketoacidosis in Sweden:

Self-Monitoring Of Blood Glucose (SMBG)

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

Blood Glucose Meter recommendations for GP practices

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)

AP Physics 1 and 2 Lab Investigations

Big Data, Socio- Psychological Theory, Algorithmic Text Analysis, and Predicting the Michigan Consumer Sentiment Index

SKUP Scandinavian evaluation of laboratory equipment for primary health care

Basic Lessons in Laboratory Quality Control

Blood Glucose Management

Understanding Diabetes

Report. European Co-ordinated Post Market Surveillance Operation under COEN authority. Blood Glucose Meters

Confidence Intervals for Cpk

Introduction 1 The system 1 The meter 2 The display 3 The mode 3 The measurement 4 Coding the meter 4 How to obtain a drop of blood 6 Application of

Quantitative Inventory Uncertainty

Name: Date: Use the following to answer questions 3-4:

STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT

Ketones & diabetes; Reduce your risk!

US Diabetic Devices Industry Research Report: KenResearch

Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

CHAPTER THREE COMMON DESCRIPTIVE STATISTICS COMMON DESCRIPTIVE STATISTICS / 13

Confidence Intervals for Spearman s Rank Correlation

Accu-Chek Inform II system. Professional glucose testing for the wireless age

Management of Diabetes

Title: Fingerstick Glucose by Accu-Chek Inform

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

Quantifying measurement error from digital instruments

Using Excel for inferential statistics

Self-Monitoring of Blood Glucose

Algorithms for Glycemic Management of Type 2 Diabetes

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

NICE insulin algorithm function. Select 'Calculate insulin dose' from web menu

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries

at The Valley Hospital (TVH) for Nursing Students/Nursing Instructors 2012

STANDARD OPERATING PROCEDURE FOR NOVA STAT STRIP CONNECTIVITY BLOOD GLUCOSE METERS FOR USE IN LEEDS NHS TRUST

To avoid potential inspection

Supplementary appendix

SMF Awareness Seminar 2014

Means, standard deviations and. and standard errors

Using Microsoft Excel to Plot and Analyze Kinetic Data

The OmniPod Insulin Management System

Terms concerned with internal quality control procedures

Take a moment Confer with your neighbour And try to solve the following word picture puzzle slides.

Forecasting in supply chains

Preface. Preface. A Healthcare professional should be contacted when Customer Service is not available.

Chapter 1 The Importance of Education in Diabetes

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

Analysis of Bayesian Dynamic Linear Models

How to get the most of your blood glucose readings

A Population Health Management Approach in the Home and Community-based Settings

Confidence Intervals for Cp

Assessment of Accuracy and Precision

Administration of Emergency Medicine

Diabetes Management in the Primary Care Setting

Nova Scotia Guidelines for Acute Coronary Syndromes (Updating the 2008 Diabetes sections of the Guidelines)

Paper KEYWORDS PROC TRANSPOSE, PROC CORR, PROC MEANS, PROC GPLOT, Macro Language, Mean, Standard Deviation, Vertical Reference.

Spreadsheets and Laboratory Data Analysis: Excel 2003 Version (Excel 2007 is only slightly different)

BLOOD GLUCOSE METER. Getting Started Guide for Single Patient Use Only

Calibration and Linear Regression Analysis: A Self-Guided Tutorial

BLOOD GLUCOSE MONITORING SYSTEM USER GUIDE

How to Verify Performance Specifications

Auxiliary Variables in Mixture Modeling: 3-Step Approaches Using Mplus

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

BLOOD GLUCOSE METER / GLYCOMÈTRE. User s Manual Manuel de l'utilisateur

Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document

1/27/2013. PSY 512: Advanced Statistics for Psychological and Behavioral Research 2

What are confidence intervals and p-values?

INTRODUCTION TO CLINICAL PRACTICE AND CLINICAL SKILLS 2nd Year MEDICAL YEAR 2009/2010

British Columbia Pharmacy Association (BCPhA) Clinical Service Proposal Self-Monitoring of Blood Glucose in Type 2 Diabetes

Confidence Intervals for One Standard Deviation Using Standard Deviation

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA

Simple Start TM Diabetes Log Book

Transcription:

Diabetes Research and Clinical Practice 71 (2006) 113 118 www.elsevier.com/locate/diabres A comparison of blood glucose meters in Australia Matthew Cohen *, Erin Boyle, Carol Delaney, Jonathan Shaw International Diabetes Institute, 250 Kooyong Road, Caulfield, Vic. 3162, Australia Received 1 April 2005; accepted 26 May 2005 Available online 11 July 2005 Abstract Objective: To assess the accuracy and precision of the five currently available blood glucose meters in Australia. Design and setting: Control solutions from manufacturers were used to determine the precision for each meter. Glucose levels in capillary blood samples from 49 patients attending a diabetes clinic were measured with each meter and with a laboratory reference method. Outcome measures: The coefficient of variation was calculated to determine precision. Bias, Error Grid analysis, and Bland Altman plots were used to determine accuracy. Results: The CVs of most meters were acceptable at <5%. Bias ranged from 4.0 to 15.5% with only 1 meter satisfying the American Diabetes Association recommendation of <5% bias. Error Grid analysis showed that 94 100% of readings were clinically accurate, and that none of the differences from the reference method would lead to clinical errors. Bland Altman plots showed that for two meters the magnitude of the difference between the meter and the reference method increased with increasing glucose values, but did not change significantly with glucose level for the other 3 meters. Conclusions: Currently available blood glucose meters show acceptable precision, and any errors (with respect to a laboratory method) are highly unlikely to lead to clinical errors. However, only the CareSens meter achieved a bias of less than 5%. # 2005 Elsevier Ireland Ltd. All rights reserved. Keywords: Self blood glucose monitoring; Blood glucose meters; Accuracy 1. Introduction It is well established that optimal glycemic control reduces the long-term risk of complications in diabetes. It is currently recommended by the American Diabetes Association (ADA) that treatment of all individuals with diabetes should aim to lower blood * Corresponding author. E-mail addresses: mattcohen@idi.org.au (M. Cohen), jshaw@idi.org.au (J. Shaw). glucose to normal or near normal levels [1]. Self monitoring of blood glucose levels allows the patient to respond immediately to glucose changes with the appropriate action and is currently recommended as an important component of diabetes care [2]. Adequate glycemic control is therefore dependent on the accuracy and reliability of the self-blood glucose monitoring (SBGM) system used. Since the introduction of SBGM in 1979, there have been ongoing, competition-driven developments in both meter and strip technology, which have allowed greater accuracy 0168-8227/$ see front matter # 2005 Elsevier Ireland Ltd. All rights reserved. doi:10.1016/j.diabres.2005.05.013

114 M. Cohen et al. / Diabetes Research and Clinical Practice 71 (2006) 113 118 and reliability of results, as well as improved patient acceptance. While independent studies exist assessing various models of glucose meters to reference values [3 5] assessment of the accuracy and precision of the latest models available in Australia (Accu-Chek Go, Accu-Chek Advantage (Roche), Optium (Abbott) CareSens (i-sens) and GlucoMen PC (Menarini)) is limited to the manufacturers in house evaluations. There is a need for a direct and independent comparison to be made in order to alleviate any uncertainty regarding performance of these individual meters. 2. Aim To provide an independent assessment of the accuracy and reliability of the latest models of blood glucose meters available in Australia. 3. Patients and methods The latest models of blood glucose meters available in Australia in September 2004 were studied. The five latest available meters were selected: (1) Accu-Chek Go (Roche), which has the same technology as Accu- Chek Active which it is replacing; (2) Accu-Chek Advantage (Roche); (3) CareSens (i-sens a new entrant in Australia); (4) GlucoMen PC (Menarini) and (5) Optium (Abbott). The GlucoCare (Diabetes Australia NSW) was not available outside the state of New South Wales and the distributors declined requests to submit a meter for evaluation. One new meter of each type was randomly selected from stock. Prior to the collection of samples, and after the calibration procedure for new vials of strips, quality control measurements were performed using an aqueous solution supplied by the manufacturer for each meter with the exception of GlucoMen PC, for which a control solution is apparently not available in Australia. Precision was measured by calculating the coefficient of variation (CV) using 25 samples from control solutions. Each meter measured within the recommended ranges for the control solutions on each occasion. Accuracy was measured using blood samples obtained from 49 patients with diabetes attending the International Diabetes Institute for their usual clinic visit. Capillary blood was obtained using a sterile lancet fixed in a spring-loaded device, and was applied directly from the fingertip to the reagent strip. Successive drops of blood from the same site were used for simultaneous analysis first with the reference method and then with each meter in random order. The reference method used was the YSI Glucose Analyser (Yellow Springs Instruments, Ohio, USA) using capillary whole blood collected in a hinge cap vial. All samples were collected and tested by the same operator, a pathology nurse with many years experience using various glucose meters. To compare the performance of each meter to the reference method, several analyses were undertaken. The traditional Error Grid analysis [6] was performed to determine the clinical significance of the differences between the meter and reference value. However, it has been suggested that this can be misleading, and that Bland Altman charts are preferable [7]. In this analysis, the mean blood glucose obtained from the reference and meter measurements is plotted against the difference between the two results (meter minus reference) with a regression line and 95% limits of agreement calculated. The width of the limits of agreement indicates the variability of the difference between the meter and reference methods over the range of glucose values. A significant Pearson s correlation coefficient (r) indicates the degree to which the difference changes with the magnitude of the measurement. Paired reference and glucose meter readings were used to calculate regression equations for each glucose meter. Bias was also calculated for each meter, as the mean of the difference between the reference and test meter as a percentage of the reference value. 4. Results Table 1 shows that the CVs of the meters, using the manufacturers test solutions, ranged between 2.8 and 5.5%, and were close to the figures given by the manufacturers, except for the Accu-Chek Go. Table 2 shows that all the meters read higher than the reference device, with only the CareSens meter meeting the ADA recommendation of <5% bias.

M. Cohen et al. / Diabetes Research and Clinical Practice 71 (2006) 113 118 115 Table 1 Coefficient of variation (CV) of glucose meters, in rank order and compared to product information. Meter CV% (low control) CV% (high control) Product information YSI (reference) 0.99 NA CareSens 2.83 2.5 3.4 Accu-Chek Advantage 3.64 2.62 <3.1 Optium 4.36 3.71 2.9 5.1 Accu-Chek Go 5.51 3.26 1.6 GlucoMen PC NA NA 2.79 4.46 The Error Grid analysis (see Fig. 1) showed that all measurements from all meters lay in zones A and B. GlucoMen had three measurements (6.1%) and Accu- Chek Advantage had one measurement (2%) in zone Table 2 Comparison of blood glucose concentrations in 49 patients obtained by glucose meters and reference method (YSI) Reference/glucose meter Glucose Bias (%) Regression concentration equation (mmol/l). Mean (S.D.) YSI (reference) 8.933 (3.886) Optium 9.533 (3.91) 6.7 y = 0.99x + 0.67 Accu-Chek Go 9.619 (4.16) 7.7 y = 1.06x + 0.12 CareSens 9.288 (3.70) 4.0 y = 0.93x + 0.95 Accu-Chek Advantage 9.510 (4.06) 6.5 y = 1.03x + 0.29 GlucoMen 10.317 (4.52) 15.5 y = 1.15x + 0.03 y, glucose meter; x, YSI (reference). Regression equations were calculated from paired values obtained from the reference method and each glucose meter. Fig. 1. Error Grid analysis.

116 M. Cohen et al. / Diabetes Research and Clinical Practice 71 (2006) 113 118 B. All other values for these meters and all values for the other meters were in zone A. The Bland Altman analyses (see Fig. 2) showed a significant correlation between the absolute difference (between the meter and the reference method) and the mean of the two methods for the Accu-Chek Go and GlucoMen PC meters, while for the other meters the difference between the meter values and the reference method was the same at lower and higher glucose values. Most meter readings were higher than the reference device, with the mean differences being 0.6 mmol/l for the Optium, 0.7 mmol/l for the Accu- Chek Go, 0.4 mmol/l for the CareSens, 0.6 mmol/l for the Accu-Chek Advantage and 1.4 mmol/l for the GlucoMen. All meters tested were very compact and portable. Obtaining a BG measurement was straightforward for all meters. All models now use electrochemical technology, apart from Accu-Chek Go, which uses photo-reflectance technology. Sample volume requirements have decreased since our previous analysis [3], with CareSens having the smallest blood sample volume requirement (0.5 ml). Accu-Chek Go and CareSens had the fastest testing time of 5 s. The Optium is the only meter that also performs blood testing of ketones, a recommended part of sick day management in Type 1 diabetes. 5. Discussion Based primarily on the results of the DCCT, the ADA recommends that individuals with diabetes aim at achieving and maintaining blood glucose levels as close as is safely possible to the normal range and SBGM is recommended to assist in achieving this goal. Glucose meters are designed specifically for patient use in determining glucose concentrations in capillary blood. As the accuracy and precision of glucose meters vary, the ADA also recommends calibration and validation of glucose meters. Fig. 2. Bland Altman analysis: solid line represents the regression line; dotted lines show 95% limits of agreement.

M. Cohen et al. / Diabetes Research and Clinical Practice 71 (2006) 113 118 117 Precision and accuracy are generally greater with the newer generation models of blood glucose monitors compared to older devices [8,9]. This is reflected in the tightening of the ADA recommendations regarding acceptable limits of deviation from laboratory reference methods [10,11]. Recent studies have assessed various models of glucose meters in relation to reference values. However, despite significant improvements, some meters still fail to satisfy the current recommendations of <5% deviation from reference methods [4,8,12,13]. All meters examined in this study yielded mean glucose concentrations slightly higher than the reference (YSI) method. The current study found the CareSens and Accu-Chek Advantage meters to be the two most precise meters. The error measured as bias was smallest for the CareSens (4.0%) and Accu- Chek Advantage (6.5%) and largest for GlucoMen (15.5%), with the CareSens meter being the only one which met the ADA recommendation of <5% deviation from the reference method. The Bland Altman analysis showed that the difference between the meter and the reference method was constant across all glucose levels for all meters except the Accu-Chek Go and the Gluco- Men. The small difference between the YSI and the meters is probably due to the test strips being calibrated to plasma rather than whole blood glucose values. (The formula used by each manufacturer is unknown, but the conversion factor usually used is 1.11.) Nevertheless, based on Error Grid analysis results, we have shown adequate clinical accuracy of currently available meters in the hands of an experienced pathology nurse. All measurements from all meters lay in zones A (clinical accuracy) or B (no or benign treatment differences). GlucoMen had three measurements (6.1%) and Accu-Chek Advantage had one measurement (2%) in zone B, while all readings from the other meters were in zone A. As the interaction between user and device can be a large source of variability [2],thisispotentially a limitation of the current study, which takes no account of patient/user error. We did not study alternate sites e.g., forearm, that may be used with some meters, but other studies have shown in some circumstances, changes in blood glucose at these sites may lag behind the changes at the finger tip. 6. Operating features The latest meters are an improvement on previous models. They are quicker and easier to both learn and teach, and this represents a significant savings in staff and patient time. The convenience of using Accu-Chek Advantage, Accu-Chek Go and CareSens which all have a 5 s test may also encourage more frequent testing. Many clinicians and patients use the computerdownloading feature of these meters. This obviates the need for patients to manually record the results in a diary. It also facilitates the analysis of blood glucose levels, for example, revealing diurnal patterns that assist with altering insulin regimens. However, this feature requires the patient to maintain the correct time and date in the meter on first use, and when changing batteries. Although the actual operation of all meters is simple, the resetting of the date and time is often beyond the ability of many less technically minded people. This is particularly difficult for the optium meter, which only has a single button for all the control features. 7. Conclusion The information provided in this study indicates good precision and accuracy for all the meters studied, and should provide confidence in their role in achieving optimal glycemic control through SBGM. Acknowledgements Thanks to our patients for donating additional drops of blood. Statement of competing interests: The manufacturers of CareSens, i-sens, provided an unconditional grant to perform the study and some distributors supplied strips and control solutions. MC has been reimbursed for consultancies and conducting presentations for Abbott and Roche diagnostics, but not since 2003. The study was conducted independently of commercial organizations, and none were involved in the study design, data analysis and interpretation, or in the writing or decision to publish the article.

118 M. Cohen et al. / Diabetes Research and Clinical Practice 71 (2006) 113 118 References [1] D.E. Goldstein, R.R. Little, R.A. Lorenz, et al. Tests of glycemia in diabetes (technical review), Diabetes Care 18 (1995) 896 909. [2] American Diabetes Association, Clinical practice recommendations 2000. Position statement: tests of glycemia in diabetes, Diabetes Care 23 (2000) S80 S82. [3] L. Engel, C. Delaney, M. Cohen, Blood glucose meters: an independent had-to-head comparison, Pract. Diabetes Int. 15 (1998) 15 18. [4] B. Solnica, J.W. Naskalski, J. Sieradzki, Analytical performance of glucometers used for routine glucose self monitoring of diabetic patients, Clin. Chem. Acta 331 (2003) 29 35. [5] M.O. Ajala, O.O. Oladipo, O. Fasanmade, et al. Laboratory assessment of three glucometers, Afr. J. Med. Sci. 32 (2003) 279 282. [6] W.L. Clarke, D. Cox, L.A. Gonder-Fredrick, et al. Evaluating clinical accuracy of systems for self-monitoring of blood glucose, Diabetes Care 10 (1987) 622 628. [7] J.M. Bland, D.G. Altman, Comparing methods of measurement: why plotting difference against standard method is misleading, Lancet 346 (1995) 1086 1087. [8] R. Weitgasser, B. Gappmayer, M. Pichler, Newer portable glucose meters analytical improvement compared with previous generation devices, J. Clin.Chem. 45 (1999) 1821 1825. [9] P. Bohme, M. Floriot, M.-A. Sirveaux, et al. Evolution of analytical performance in portable glucose meters in the last decade, Diabetes Care 26 (2003) 1170 1175. [10] American Diabetes Association, Consensus statement on self monitoring of blood glucose, Diabetes Care 10 (1987) 95 99. [11] American Diabetes Association, Self monitoring of blood glucose consensus statement, Diabetes Care 19 (Suppl. 1) (1996) S62 S66. [12] A. Brunner, M. Ellmerer, G. Sendlhofer, et al. Validation of home blood glucose meters with respect to clinical and analytical approaches, Diabetes Care 21 (1998) 585 590. [13] R.N. Johnson, J.R. Baker, Error detection and measurement in glucose monitors, Clin. Chem. Acta 307 (2001) 61 67.