Testing Wireless Medical Devices. Greg Kiemel



Similar documents
RF Exposure and Modular Equipment. Considerations

Wireless Technology and RF Standard in Medical Device Development

RF EXPOSURE LIMITS AND TESTING REQUIREMENTS

Evaluating Cell Phone and Personal Communications Equipment and their EMC Effects on Automotive Audio and In-Cabin Modules

How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless)

Wireless Medical Devices

Body SAR and EU. B. Derat

Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands)

Millennium Product Inc. Model: Cell Shield / Zorb

30 In Compliance October

Automated Meter Reading Frequently Asked Questions. What is AMR?

Dynamic Frequency Selection (DFS) and the 5GHz Unlicensed Band

EMR - SAR Testing & Approval Requirements for Australia

TCB RF Exposure Training

EMC & Wireless Certification

COMMON REGULATORY OBJECTIVES FOR WIRELESS LOCAL AREA NETWORK (WLAN) EQUIPMENT PART 2 SPECIFIC ASPECTS OF WLAN EQUIPMENT

TCB Workshop. Unlicensed National Information Infrastructure Devices (U-NII)/Dynamic Frequency Selection (DFS)

EMR COMPLIANCE HOW TO ENSURE YOUR NATA ACCREDITED CONTRACTOR IS ABLE TO PROVIDE A NATA ENDORSED REPORT

Dynamic Frequency Selection Requirements

Regulations and Standards for Wireless Medical Applications

ETSI EN V1.1.1 ( )

R&S R-Line Compact Test Chamber Precise measurements of radiated emissions of wireless terminals 800 MHz to 18 GHz

RF EXPOSURE PROCEDURES AND EQUIPMENT AUTHORIZATION POLICIES FOR MOBILE AND PORTABLE DEVICES

TECHNICAL SPECIFICATION FOR GSM MOBILE TERMINALS

Analysis of Immunity by RF Wireless Communication Signals

AN Application Note: FCC Regulations for ISM Band Devices: MHz. FCC Regulations for ISM Band Devices: MHz

Application Note AN-00126

Annex to the joint statement: Exposure of the general public to radiofrequency fields

RFID Design Principles

ARIB TR-T V Specific Absorption Rate (SAR) requirements and regulations in different regions

Measurement of RF Emissions from a Final Coat Electronics Corrosion Module

II. Radio equipment which transmits only under the control of electronic communications networks

OpenWay Radio Frequency FAQ

IEEE Electromagnetic Compatibility Standards (Active & Archive) Collection: VuSpec

Regulatory requirements - USA

CHAPTER 1 Introduction 1

Wireless Broadband Access

ETSI EN V1.2.1 ( )

Federal Communications Commission Office of Engineering and Technology Laboratory Division Public Draft Review

Radio Frequency Exposure Test Report

Electromagnetic Compatibility and International Regulatory Approvals

Federal Communications Commission Office of Engineering and Technology Laboratory Division

LTE, WLAN, BLUETOOTHB

Radio Frequency Wireless Technology in Medical Devices. Guidance for Industry and Food and Drug Administration Staff

EN v1.2.1 TEST REPORT FOR ag/Draft n WLAN PCI-E Mini Card MODEL NUMBER: BCM94322MC REPORT NUMBER: 07U

LTE on Shared Bands (LEONARD)

Wireless Ethernet LAN (WLAN) General a/802.11b/802.11g FAQ

Radio waves and health. Mobile communications

ETSI EN V1.2.2 ( )

CE Test Report. : Microchip Technology Inc West Chandler Blvd., Chandler, Arizona , USA

Communication Architecture for AMI and other Smart Grid/Smart City Applications. Presented By: Reji Kumar Pillai President - ISGF

Wireless Medical Telemetry Laboratory

Hot Issues in Wireless Broadband Networking

ELETRONIC COMMUNICATIONS COMMITTEE

What is ? Why are standards important?

TEST REPORT FROM RFI GLOBAL SERVICES LTD

RF Exposure Procedures TCB Workshop October 2012

ECC/DEC/(04)08 ELECTRONIC COMMUNICATIONS COMMITTEE

RFID Design Principles

Wireless Broadband: Health & Safety Information

Global Wireless Device Approvals & Certification Overview Tony Henriques Principal Engineer Inspection Body, Global Approvals, WiSE EULA

Application Note AN-00125

Wireless Power Transfer

Rated Power(W) 8W 2. EG-LED W 3. EG-LED W

AKKREDITOITU TESTAUSLABORATORIO ACCREDITED TESTING LABORATORY MICROSOFT MOBILE OY TCC MICROSOFT

Espotel Test Laboratory Services. November 20, 2014

Software Defined Radio (SDR) Application Review Guide

RF safety at base station sites

Human Exposure Limits

SmartDiagnostics Application Note Wireless Interference

Bluetooth: Global Certification Requirements

Spectrum Pre-compliance for Wireless LAN Regulatory Testing

Electromagnetic Field Safety

USB 3.0* Radio Frequency Interference Impact on 2.4 GHz Wireless Devices

Wireless and Mobile Technologies for Healthcare: Ensuring Privacy, Security, and Availability

SUMMARY OF: P910077/S140 BOSTON SCIENTIFIC LATITUDE COMMUNICATOR MODELS 6460, 6280, 6290, 6250, 6498, AND Executive Summary

3 Radio Waves and Human Body 8 Specific Absorption Rate (SAR)... 8 SAR value of a mobile handset... 8 Radiation level by a mobile tower...

CE Test Report. : BL600 Series Bluetooth Low Energy Module

Radio Frequency Operations and Technology

Introduction to Clean-Slate Cellular IoT radio access solution. Robert Young (Neul) David Zhang (Huawei)

Wireless data communication in Control Systems

LTE: Technology and Health. 4G and Mobile Broadband

The U.S. FDA s Regulation and Oversight of Mobile Medical Applications

SECTION 2 TECHNICAL DESCRIPTION OF BPL SYSTEMS

EMC STANDARDS STANDARDS AND STANDARD MAKING BODIES. International. International Electrotechnical Commission (IEC)

Federal Communications Commission Office of Engineering and Technology Laboratory Division

For more information, please contact Fish s Regulatory & Government Affairs Group: Terry G. Mahn, Chair Washington, D.C. mahn@fr.

EMC Standards: Standards of good EMC engineering

Test Report. Prepared for: Technologic Systems, Inc. Model: TS-8820

RADIO TEST REPORT. EN SRD radio equipment to be used in the 1 GHz to 40 GHz frequency range. DAkkS - Registration number : D-PL

IEEE n Enterprise Class Wireless LAN?

RF Exposure Evaluation Report

Human exposure to RF

ACRS 2.0 User Manual 1

Partial GSM TEST - REPORT

Transcription:

Testing Wireless Medical Devices Greg Kiemel

Outline Wireless Trends Latest News Medical Wireless Technologies Risks Regulatory Standards Beyond EMC Compliance Considerations 2

Wireless Trends

Increasing Rate of Change It took more than 90 years for landline service to reach 100 million consumers More than 21 years for color televisions to reach 100 million consumers But less than 17 years for wireless to reach 100 million consumers. Source: CTIA The Wireless Association, Wireless Quick Facts, 2011. 4

Regulatory Approval Trends Virtually all radios sold in the U.S. require FCC certification. To keep up with the increasing number of applications, the FCC created the Telecommunications Certification Body (TCB) program in June 2000. TCBs certify wireless products under the authority of the FCC. FCC ET Docket 13-44 released 12/30/14 made TCBs the only route to certification. The FCC will no longer accept applications for certification. 5

TCB Application Trend 40000 Approvals 30000 20000 10000 0 1999 2000 2001 2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2014 Year FCC TCB 6 Source: https://apps.fcc.gov/oetcf/eas/reports/genericsearch.cfm

Application Trends Jan 1, 2014 Dec 31, 2014 TCBs issued 42,669 Approvals, FCC 425 7

Wireless Data Standards Usable Range Miles Feet GSM, GPRS, Edge, HSPA 802.11ad 802.15.1 (Bluetooth) WWAN 4G LTE WLAN 802.11a/b/g/n WPAN 802.15.3 (UWB) WBAN 802.15.6 MBAN MedRadio Inductive Telemetry 802.15.4 (Zigbee) 802.16e (WiMax) 802.11ac 8

Evolution of Applications / Platforms TV, Video Computer Convergence of Technologies Medical Wireless Telecommunications Internet 9

Multiple Radios in the Same Host Device Patient Monitor with Wi-Fi, Bluetooth, and WWAN Radios 4G Cellphone with Wi-Fi, Bluetooth, NFC, and GPS Notebook PC or Tablet with Cellular modem, Wi-Fi, and Bluetooth 10

New Technologies Spectrum Efficiency is the driving force more users require more throughput while occupying the same finite chunk of frequency spectrum Biggest Breakthroughs Smart Antenna Systems MIMO (Multiple Input and Multiple Output) Adaptive Beam forming OFDM (Orthogonal frequency-division multiplexing) Cognitive Radio Listen Before Talk (LBT) Dynamic Frequency Selection (DFS) Adaptive Frequency Hopping (e.g. Bluetooth devices) 11

The New Wireless Medical World 12

Wireless Proliferation into Medical Why? Convenience: don t want to go to the doctor Speed: want and need fast service Separation: can keep people or equipment out of the sterile environment Appearance: nobody wants to look sick being hooked up to wires or tubes 13

Wireless Proliferation into Medical Where - Point of care Hospital Clinic Home Mobile 14

Latest News

FCC Report and Order (FCC 14-208) FCC will no longer accept applications for certification. All grants will be issued by TCBs. The Permit But Ask procedure will be codified. Clarification of TCB responsibilities for performing market surveillance. Specify steps for addressing instances of deficient TCB performance, including sanctions and rescinding TCB authority. Modify the rules to reference new standards (ISO/IEC 17065) used to accredit TCBs Require labs that test for certification or DoC to be accredited. [ Testing laboratories located in non MRA countries are not FCC recognized as accredited under the current procedures. ] Codify a procedure through which the Commission currently recognizes new laboratory accreditation bodies; Update references to industry measurement procedures in the Commission s rules (e.g. ANSI C63.4-2014 and ANSI C63.10-2013); and Provide greater flexibility to enable the FCC s OET to address minor technical issues that may be raised when updating to the latest versions of industry 16 standards that are referenced in Parts 2, 5, 15, and 18 of the Commission s rules.

FCC Report and Order (FCC 14-30) New rules for U-NII devices (e.g. 802.11an) became effective on June 2, 2014. Transition period, see FCC KDB 926956: New devices can be approved until June 1, 2015 under either the old or new rules. Starting June 2, 2015, new approvals must be to the new rules. Starting June 2, 2016, permissive changes must be to the new rules. Also products previously certified under the old rules can t be marketed unless they meet the new rules (C2PC application required). Indoor-only restriction removed from the 5.2 GHz band. Also increased the output power in that band. Extended the upper frequency edge of the 5.8 GHz band to match FCC 15.247. They are now both 5725 5850 MHz Harmonized the rules between 15.247 and 15.407 for DTS device operating in the 5.8 GHz band. U-NII device software must be secured to prevent its modification New DFS radar waveform and new test requirements. 17

Radio Equipment Directive (RED) 2014/53/EU Published in the OJEU on April 16, 2014 will be applicable June 13, 2016. No one can claim compliance with this new directive until then. However, if equipment meets the requirements of the R&TTE Directive before June 13, 2016, compliance with the RED is not mandatory until June 13, 2017. For new equipment placed on the EU market after June 12, 2016 must meet the requirements of the RED. Sound and TV broadcast receivers will be included in the RED Telecom Terminal Equipment will be excluded Radio determination will be included, e.g. RFID, movement detection. Radio equipment operating below 9 khz will be included ISM equipment (generates RF to perform work, not telecommunications) will be excluded. Role of a Notified Body is still to assess compliance at the request of the manufacturer. This is optional if harmonized standards have been applied in full. Notified Bodies will issue an EU-Type Examination Certificate that documents their assessment of the manufacturer s supplied Technical Documentation (see Annex V of the RED) 18

Changes to EN 300 328 and EN 301 893 EN 300 328 V1.8.1 and EN 301 893 V1.7.1 required starting Jan 1, 2015. Even products that have been selling for years under the previous version must now meet the latest version if they are still being shipped into the EU. Products already in the EU prior to Jan 1, 2015 do NOT need to be recalled or retested. The main differences between the old and new versions of the standards are significant and can be summarized as follows: Adequate spectrum sharing. The efficiency of the various sharing mechanisms must be tested per the procedures of the harmonized standards. Previous versions permitted just an attestation from the manufacturer. An output power test methodology that would address a wide-range of technologies. This new approach is completely different from what was included in V1.7.1 of the standard. A power spectral density test methodology that would permit a higher output power. Many devices are not limited by the 100mW EIRP limit, but instead by the 10mW/MHz power spectral density (PSD) limit. EN 300 328 V1.8.1 has very different and much more complex method of measuring power spectral density. Harmonize test methods with other ETSI standards operating in the same bands. EN 300 328 V1.9.1 published in the OJ on April 17, 2015. Mandatory starting Nov. 30, 2016. Contains minor fixes to previous version. 19

Medical Wireless Technology

Medical Wireless Technology Wearable, wireless devices capable of offering real time data 2012 14 million 2016 171 million (estimated) $6 billion market minimum 21 DIGI FIT LIFE, Wearble, Wireless Technology to Generate Minimum $6 Billion by 2016, August 2012, retrieved from http://digifitlife.com/wearable-wireless-technology-to-generate-minimum-6-billion-by-2016/#.vl_xwkff-so

Medical Wireless Technology Inductive Typically below 200 khz Very short range, low level Inductive charging ISM/SRD (Industrial, Scientific & Medical / Short range Devices) Frequencies 13.56 MHz (ISM/SRD), 27.12 MHz (ISM), 40.68 MHz (ISM), 433.92 MHz (ISM), 865.0 MHz (SRD), 915.0 MHz (ISM), 2450 MHz (ISM, SRD), 5.8 GHz (ISM, SRD) Crowded Medical Device Radiocommunication Service (MedRadio) or MICS (Medical Implant Communication Service) 401-406 MHz Special implanted devices Short range - approximately 3-5meters 22

Medical Wireless Technology A Sub-set of MedRadio, the FCC also permits operation of Medical Micropower Network (MMN) devices In November 2011, the FCC adopted rules for the operation of ultra-low power wideband networks in the 413-419 MHz, 426-432 MHz, 438-444 MHz, 451-457 MHz, and 2360-2400 MHz bands: http://www.fcc.gov/document/medical-micro-power-networks MMN s are used to restore functions to paralyzed limbs. Implanted transmitters in the body take the place of damaged nerves, restoring sensation and mobility. Multiple MMNs may be present within a patient. Implants may only communicate with the programmer /controller for their MMN, and not with each other. 23

Medical Wireless Technology MBAN (Medical Body Area Networks) 5/2012 - FCC Designates 2360-2400 MHz band License-by-rule users will not have to apply for/receive individual station licenses Sharing 2360-2390 w/aeronautical Mobile Telemetry (AMT) licenses. Use in that band restricted to indoor health care facilities & requires registration w/mban coordinator. FCC still needs to select. Rest of band no registration, use anywhere (mobile, home, health care facilities) Low power, short range http://www.fcc.gov/document/medical-body-area-networks-first-report-andorder/ 24

Medical Wireless Technology MBAN (Medical Body Area Networks) Multiple body sensors for monitoring & control Facebook for your organs removes the tether 25

Medical Wireless Technology WMTS (Wireless Medical Telemetry Service instituted in 2000) Freqs: 608-614 MHz, 1395-1400 MHz, 1429-1432 MHz Long range Safe band, proprietary, very common use for wireless patient monitoring Limited to use within health care facilities. 26

Wireless Technology Wi-Fi connectivity of monitors 92% Wi-Fi penetration into large (500+ employees) healthcare facilities now & 100% by 2016 17% of the world s population uses Wi-Fi Wi-Fi Alliance working on WLAN best practices & Risk Mgmnt, plus personal home healthcare architecture 27

Wireless Technology Bluetooth low power, secure, interoperability 2 billion BT products ship in 2012 Bluetooth 4.0 is capable of Basic Rate and Enhance Data Rate operation (BR/ EDR) as well as Low Energy (LE), or LE only. Bluetooth BR and EDR modes operate on 79 hopping channels that are tested at: 2402, 2441, and 2480 MHz Bluetooth LE Advertising mode utilizes only three channels: 2402, 2426, and 2480 MHz Bluetooth LE Data mode has 37 hopping channels: 2404, 2442, and 2478 MHz 28

Wireless Technology RFID equipment or people tags Zigbee some but mostly industrial devices Cell bands mobile communication/ apps 29

Risks

Are Wireless Medical Devices Safe and Effective? Interference could result in injury or death Failure to provide critical status or drug injection information Problem Reports (FDA) Feb 2006 cell phone interfered w/infusion pump causing higher prescribed rate June 2006 cell phone caused infusion pump to stop Sept 2008 electrocautery device caused a loss of therapeutic effect on an implant Other studies ICD reactions to RFID Legal Perspective http://www.fr.com/files/uploads/attachments/finalregulatorywhitepaperwirelessmedicaltechnologies.pdf 31

RF Wireless Coexistence Due to the increasing number of wireless devices, the risk of interference with wireless medical devices is also increasing Critical data transmitted by medical devices could be delayed or blocked by other nearby wireless devices. Current EMC standards are inadequate. In-band interference is specifically excluded. Manufacturers cannot claim the safe operation of their wireless medical device based solely on EMC test results. Test methods are under development to assess the risk of interference. (ANSI, Univ of Oklahoma) 32

Current Guidance on Coexistence Co-existence is discussed in the FDA Guidance for Wireless Technology in Medical Devices. Draft issued Jan 2007, became formal guidance August 2013. http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidancedocuments/ucm077210.htm Also contains guidance on risk management, design & development, verification, validation, labeling, etc. Environmental Considerations Radio Channel Characteristics, Polarization, Co-channel & Adjacent Channel Interference, Distance from the interfering source Testing should be done in both an anechoic chamber and in situ, with each wireless medical device separately exposed to one or more wireless networks. http://www.medicalelectronicsdesign.com/article/wireless-medical-device-coexistence 33

Regulatory

US FDA Medical Regulatory Regulates all medical devices & general public health and safety http://www.fda.gov/medicaldevices/deviceregulationandguidance/guidanc edocuments/ucm077210.htm FCC Wireless and RF Exposure Spectrum management Standards TCB route for approvals short turnaround 35

EU Regulatory Medical & Wireless - Directives Radio (RTTE) / EMC (EMCD) RTTE repealed and RED begins June 13, 2016 Old EMCD repealed and New EMCD begins April 20, 2016 Medical Devices (MDD, AIMDD) Safety (Electrical, Mechanical, RF) Self declaration 36

Typical Wireless Test Requirements Output Power Effective Radiated Power Frequency Stability Occupied Bandwidth Emissions Mask Power Spectral Density Spurious Emissions Immunity 37

Effective Radiated Power MedRadio 403 405 MHz Maximum EIRP: 25uW = 85.2 dbuv/m @ 3 meters measured over reference ground plane. Implants must be configured to transmit in a human torso simulator placed 1.5 m above the ground plane. Dimensions of the fixture, and the properties of the tissue substitute material are defined in 95.639(f)(2). The properties of the tissue substitute material should be verified on the day of test, and the temperature of the material recorded at the start and conclusion of the test. 38

Torso Simulator Cylindrical Plexiglas container with a size of 30 cm by 76 cm with a sidewall thickness of 0.635 cm. A mounting grid for the implant and associated leads must be provided inside the container that permits the implant to be positioned vertically and horizontally The implant must be mounted 6 cm from the sidewall and centered vertically within the container. 39

Radiated Emissions Test Setup EUT Vertical EUT Horizontal 40

Radiated Emissions Test Setup 41

Tissue Substitute Material Torso simulator must be completely filled with a tissue substitute material whose dielectric and conductivity properties match those of human muscle tissue at 403.5 MHz Simple saline solutions do not meet the above criteria. A formula for a suitable tissue substitute material is defined in the paper Simulated Biological Materials for Electromagnetic Radiation Absorption Studies by G. Hartsgrove, A. Kraszewski, and A. Surowiec as published in Bioelectromagnetics 8:29 36 (1987). All emissions measurements are made with the tissue material at a nominal temperature of 20 25 C 42

Standards

Standards Inductive (communications) US EU FCC Part 15.209 / 207 (tested in air) ERC/REC 70-03 ETSI EN 302 195-2 / -1 (tested in saline) ETSI EN 301 489-31 / -1 44

Standards Inductive (charging portion only) US FCC Part 18 RF Exposure (KDB 680106) EU CISPR 11 IEC 60601-1-2 45

Standards ISM/SRD US EU 15.209 / 15.109 / 15.107 15.247 2.4-2.4835 GHz Band 15.249 902-928 MHz ERC/REC 70-03 EN ETSI 300 220-2/-1 / 300 440 /300 328 EN ETS 301 489-3 / -1 46

Standards MedRadio/MICS US EU Part 95I / 2 15.109 / 107 ERC/REC 70-03 EN ETSI 301 839-2 / -1 EN ETS 301 489-27 / -1 Other Torso simulator 47

MBAN US EU Part 95I Standards ERC/REC 70-03 Unknown we are to assume Part of MICS but could also be Wi-Fi stds or some combo EN ETSI 301 839-2 / -1 EN ETS 301 489-27 / -1 48

Wi-Fi/BT US EU Standards Part 15.407 / 15.247 15.207 ERC/REC 70-03 EN ETSI 300 328 EN ETS 301 489-17 / -1 49

Medical Device Standards FDA IEC 601-1-2 EMC of medical electrical equipment & systems IEC 60601-2-X specific types of equipment FDA Guidance Higher immunity levels might be required & have been requested 50

Medical Device Standards Other AAMI PC69 EMC for Active Implantable Medical Devices EN 45502-2 (Active Implantable Medical Devices) ISO 14708-2 (implantable cardiac pacemakers) EN & ETSI reqs ISO Standards (wheelchairs, pulse oximeters, neurostimulators) 51

Beyond EMC

Special Requirements RF Exposure / RF Safety RF Exposure is the effect of emissions from transmitters on the quality of the human environment. On August 1, 1996, the Commission adopted the NCRP's recommended Maximum Permissible Exposure limits for field strength and power density for the transmitters operating at frequencies of 300 khz to 100 GHz. FCC and Industry Canada adopted the specific absorption rate (SAR) limits for devices operating within close proximity to the body as specified within the ANSI/IEEE C95.1-1992 guidelines. More info: http://transition.fcc.gov/oet/rfsafety/ 53

Exposure Categories Fixed Mobile Portable 54

Exposure Limits FCC RF exposure limits are based on ANSI/IEEE C95.1-1992 and NCRP Report No. 86, but not identical to these standards Maximum Permissible Exposure (MPE) limits are applicable to fixed and mobile transmitters. Used >= 20cm from the head or torso Frequency dependent free-space power density and/or field strength limits Based on whole-body averaged plane wave equivalent exposure conditions Specific Absorption Rate (SAR) are applicable to portable transmitters. Used < 20cm from the head or torso One gram averaged energy absorption limits for the US and Canada, 10g limits for the rest of the world Based on partial-body or localized near-field exposure conditions 55

Specific Absorption Rate (SAR) 56

SAR Test System 57

SAR Test System Radio Held Underneath Body Phantom 58

SAR Test System Radio Held Next to Head Phantom 59

RF Exposure Requirement for MedRadio Implants 95.1221 RF exposure: Applications for equipment authorization of devices operating under this section must demonstrate compliance with these requirements using either finite difference time domain (FDTD) computational modeling or laboratory measurement techniques. FCC KDB 447498 D01 v05v401, Section 4.2.4: When the aggregate of the maximum power available at the antenna port and radiating structures of an implanted transmitter, under all operating circumstances, is 1.0 mw, SAR test exclusion may be applied. 60

Special Requirements Listen Before Talk (LBT) Required for MedRadio operation in the 401 406 MHz band. EN 301 839-1 Clause 10 best test procedure Except in the case of a Medical Implant Event, communications sessions shall be initiated via LBT (either by programmer or implant). Before a communications session is initiated the requirements as stated specifically in clauses 10.1-10.6 of EN 301 839-1 shall be met. 61

Compliance Considerations

Research Objectives for your Target Market Spectrum Allocation Frequency Bands Output power Data vs. Voice Modulation type Licensed vs. Unlicensed Specifications and Test Methods EMC and RF Safety Equipment Authorization process System or modular approval Self Declaration or submittal to spectrum authority Labeling and User Info 63

Write a Test Plan IEC 60601-1-2, 4th edition "Prior to the start of formal testing, a detailed test plan shall be provided to the test laboratory. Compliance Objective Certification, Class II permissive change, or audit. Target markets U.S., Canada, EU, etc. Brief product description List of standards including version and year. Any applicable interpretations or procedures from spectrum authorities. Complete List of operating modes including modulation types, data rates, power levels, and antennas. Detailed equipment configuration for each test Operating instructions 64

Make EMC a Design Consideration The focus of EMC design: grounding, filtering, component selection, PCB layout, and shielding. Single solutions don t exist; but rather a combination of suppression techniques are required. Don t wait until the end of the design cycle to test. One test result is worth one thousand expert opinions Wernher Von Braun 65

Common Design Pitfalls Poor antenna matching Lack of shielding Design objectives or omissions that conflict with regulatory requirements Output power versus operating band RF Exposure Listen Before Talk, Duty Cycle, or Dynamic Frequency Selection (DFS) 66

Get Expert Testing & Approval Services Partner with a domestic EMC laboratory that is accredited for foreign specifications Permits testing in-country to obtain global approvals. For example Japan and Korea wireless can be tested locally. Easy access to someone who is familiar with the requirements of your target market Confirm that your EMC laboratory has experience in medical EMC How many MICS, MedRadio, WMTS, Wi-Fi, and Bluetooth grants have they issued? Ask if your EMC laboratory is equipped for the latest medical EMC test requirements Radiated Immunity to 6 GHz, 10V/m with uniform field Radiated Immunity in the 4th edition "IMMUNITY to proximity fields from RF wireless communications" Radiated Emissions testing at 10 meters 67

Thank You! 68