How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters



Similar documents
Comparison between FDA QSR and ISO 13485

IVD Regulation Overview. Requirements to Assure Quality & Effectiveness

ORACLE CONSULTING GROUP

Quality System: Design Control Procedure - Appendix

Corrective and Preventive Action Background & Examples Presented by:

Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC

GxP Process Management Software. White Paper: Ten Most Common Reasons for FDA 483 Observations and Warning Letter Citations

2014 Annual FDA Medical Device Quality System Data. FDA Form 483 Observations and Warning Letter Citations

Combination Products. Presented by: Karen S. Ginsbury For: IFF March PCI Pharma

Medical Product Software Development and FDA Regulations Software Development Practices and FDA Compliance

MEDICAL DEVICE QUALITY SYSTEMS MANUAL: A SMALL ENTITY COMPLIANCE GUIDE. First Edition

Company Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature

ORACLE CONSULTING GROUP

Empowering the Quality and Regulatory Compliance Functions

The purpose of this Supplier Quality Standard is to communicate the expectations and requirements of Baxter Healthcare Corporation to its suppliers.

ISO 13485:201x What is in the new standard?

INTRODUCTION. This book offers a systematic, ten-step approach, from the decision to validate to

CORPORATE QUALITY MANUAL

SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE

Specialties Manufacturing. Talladega Castings & Machine Co., Inc. ISO 9001:2008. Quality Manual

Inspections, Compliance, Enforcement, and Criminal Investigations

Table of Contents 1. SCOPE APPLICABLE DOCUMENTS TERMS AND DEFINITIONS QUALITY MANAGEMENT SYSTEM...4-8

QUALITY POLICY MANUAL Document: Revision: E Effective Date: January 15, 2010

Surgi Manufacturing Quality Manual

Correspondence between ISO 13485:2003 and the US Quality System Regulation

Medical Device Training Program 2015

CAPA - the importance of data analysis

Documents, Records and Change Control

WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE

Medical Device Software

Quality Management System Manual

Quality Management Systems Manual

Quality Manual TABLE OF CONTENTS APPROVAL SIGNATURE PAGE AMENDMENT RECORD SCOPE EXCLUSIONS 2.1 3

Why do Project Fail? 42% 37% 27% 26% 24% 24% 0% 10% 20% 30% 40% 50% IBM Software Group Rational software. Source: AberdeenGroup, August 2006

GE Healthcare MAR

Product Liability Risk Control Checklist

MEDFORD FABRICATION CSC, INC. Quality System Manual. Date of issue: 03/25/2010 Revision : F

Q(K SVJM~jPagelIof 3

Title:: Effective GMP AUDITS for APIs and Formulation Pharma Companies By G.Sundar-Director/Consultant PharmQA Compliance solutions

Quality Assurance Manual

GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES

Quality Manual ISO9001:2008

Mapping Your Success Top Non-Conformities Requiring a CAPA Process

ALL PRODUCTS MFG & SUPPLY

PROPOSED DOCUMENT. Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change

February 5, Dear Kristin Pabst,

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

Setting Up A Complaint Handling System. Presented by Sue Jacobs Principal Consultant, QMS Consulting, Inc. Complaint Systems

Product Lifecycle Management in the Medical Device Industry. An Oracle White Paper Updated January 2008

Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702

Quality Assurance QUALITY ASSURANCE PLAN

QUALITY MANAGEMENT SYSTEM Corporate

Software Development for Medical Devices

Aligning IT investment and Business

January 12, Dear Amy Yang:

FINAL DOCUMENT. Implementation of risk management principles and activities within a Quality Management System. The Global Harmonization Task Force

April 7, Acclarent, Inc. Mr. James Patrick Garvey II Sr. Manager, Regulatory Affairs 1525-B O'Brien Drive Menlo Park, CA 94025

Harmonizing Change Control Processes Globally

Quality Agreement Template

Corrective and Preventive Actions

DEC STD ISO Quality Systems - Model for Quality Assurance in Production and Installation

How To Validate Software

CDRH Regulated Software

Making Recall Determinations

Revision Date Author Description of change Jun13 Mark Benton Removed Admin. Manager from approval

Product Liability. Risk Control

Guidance for Industry and FDA Staff Tonometers - Premarket Notification [510(k)] Submissions

Case Study: A Mock-QSIT Audit of Ford & Firestone By Tim Wells

Quality Management System Manual Revision L

Mossfiel Electrical & Safety Management Pty Ltd

Simpler Implant Solutions Inc Arbutus St. Vancouver, BC V6i 3Y9. haroldbergmanl~ogmail.com Telephone Fax K080115

March 20, Dear Mr. Chen:

ISO 13485:2003 U.S. QSR (21 CFR 820) Quality Systems Manual

ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan)

Total Product Lifecycle Solutions from NSF Health Sciences Medical Devices

Guidance for Industry and FDA Staff: Current Good Manufacturing Practice Requirements for Combination Products

SQF Level 2 Proposed Preventive Controls Comparison Modules 2 & 11

Micro Plastics, Inc. Quality Manual

Cartel Electronics. AS 9100 Quality Systems Manual

Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?

Example of a food company quality

Clinical evaluation Latest development in expectations EU and USA

UNDERSTANDING THE EC DIRECTIVE 98/79/EC ON IN VITRO DIAGNOSTIC MEDICAL DEVICES

Exar Corporation Quality Manual

Guidance for Industry Quality Systems Approach to Pharmaceutical CGMP Regulations

Implementation of ANSI/AAMI/IEC Medical Device Software Lifecycle Processes.

CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES

QUALITY MANUAL ISO Quality Management System

Quality Assurance Manual for Low Level Radioactive. Waste Disposal Facility

Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables

Implementation of Risk Management in the Medical Device Industry

Software Development for Medical Devices

FINAL DOCUMENT. Global Harmonization Task Force

Section 5: Quality Systems Regulation (QSR) and Good Manufacturing Practice (GMP)

DICOM Grid, Inc. January 25, Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington Street, Suite 100 ALEXANDRIA VA 22314

Transcription:

environmental failure analysis & prevention health technology development How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters Kevin L. Ong, Ph.D., P.E. Managing Engineer, Medical Device Consultant Exponent Inc., Philadelphia, PA A leading engineering & scientific consulting firm dedicated to helping our clients solve their technical problems.

2 There s never time to do it right, but there s always time to do it over. http://curiousphotos.blogspot.com/ http://curiousphotos.blogspot.com/ http://americancommondefencereview.wordpress.com/

3 Overview Where do design controls fit in? Elements of design controls Design controls and the FDA ISO and Medical Devices Directive Managing risk against product liability claims http://www.endlessautumn.net/documentation/

4 Medical Device Development Technology and Business Forecasting and Planning Design Control Safety/Efficacy Research, Development Production, Design Cycle, Performance Testing

5 Areas of Risk Risk to safety of patients, users, handlers Business Regulatory Product liability

6 Quality System Regulation QSR, 21 CFR Part 820-1996 (eff. 1997) replaced the 1978 Good Manufacturing Practices (GMP) for medical devices Ensure that good QA practices are used for medical device design and that they are consistent with quality system requirements worldwide Applies to finished medical devices intended to be commercially distributed for human use: some exemptions for Class I devices Harmonized with efforts by Global Harmonization Task Force

21 CFR Part 820 - Quality System Regulation Subpart A General Provisions 820.1 Scope 820.3 Definitions 820.5 Quality System Subpart B Quality System Requirements 820.20 Management Responsibility 820.22 Quality audit 820.25 Personnel Subpart C Design controls 820.30 Design controls Subpart D Document controls 820.40 Document controls Subpart E Purchasing controls 820.50 Purchasing controls Subpart F Identification and traceability 820.60 Identification 820.65 Traceability Subpart G Production and process controls 820.70 Production and process controls 820.72 Inspection, measuring and test equipment 820.75 Process validation Subpart H Acceptance activities 820.80 Receiving, in-process and finished device accept. 820.86 Acceptance status Subpart I Nonconforming product 820.90 Nonconforming product Subpart J Corrective and preventative action 820.100 Corrective and preventative action Subpart K Labeling and packaging control 820.120 Device labeling 820.130 Device packaging Subpart L Handling, storage, distribution and installation 820.140 Handling 820.150 Storage 820.160 Distribution 820.170 Installation Subpart M Records 820.180 General requirements 820.181 Device master record 820.184 Device history record 820.186 Quality system record 820.198 Complaint files Subpart N Servicing 820.200 Servicing Subpart O Statistical techniques 820.250 Statistical techniques 7

8 Significance Of Design Controls FDA found that approximately 44% of the quality problems that led to voluntary recall actions during this 6-year period were attributed to errors or deficiencies that were designed into particular devices and may have been prevented by adequate design controls. Federal Register, October 7, 1996 21 CFR Parts 808, 812, and 820 Medical Devices; Current Good Manufacturing Practice (CGMP); Final Rule

9 2010 Warning Letters FDA issued 89 Warning Letters to medical device companies for QS/GMP deficiencies QS subsystem Corrective and preventive actions Production & process controls Design controls Management controls Document controls Source: CDRH http://www.fda.gov/downloads/aboutfda/centersoffices/cdrh/cdrhtransparency/ucm256354.pdf # warning letters with cite 81 91 69 78 49 55 43 48 33 37 % warning letters with cite

10 Most Frequent Quality Systems Warning Letter Cites 2010 21 CFR 820.198(a) 42 21 CFR 820.100(a) 30 21 CFR 820.75(a) 28 21 CFR 820.22 26 21 CFR 820.184 21 21 CFR 820.20(c) 19 21 CFR 820.30(i) 18 21 CFR 820.90(a) 18 21 CFR 820.30(g) 17 21 CFR 820.50 (general) 17 Source: CDRH http://www.fda.gov/downloads/aboutfda/centersoffices/cdrh/cdrhtransparency/ucm256354.pdf Design changes Design validation

11 Design Control Warning Letter Cites 2010 21 CFR 820.30(i) - Design Changes 18 21 CFR 820.30(g) - Design Validation 17 21 CFR 820.30(a) - General 14 21 CFR 820.30(j) - Design History File 6 21 CFR 820.30(c) - Design Input 4 21 CFR 820.30(f) - Design Verification 4 21 CFR 820.30(e) - Design Review 3 21 CFR 820.30(d) - Design Output 3 21 CFR 820.30 2 21 CFR 820.30(b) - Design & Development Planning 1

12 FDA Design Control Guidance Assist manufacturers in understanding quality system requirements concerning design control Make systematic assessment of the design an integral part of development Increase the likelihood that the design transferred to production will translate into a device that is appropriate for its intended use Applies to the design of medical devices and associated manufacturing processes; new designs as well as modifications or improvements to existing device designs

13 Application of Design Controls User Needs Review Design Input Design Process Validation Intended use is fulfilled Verification Requirements are fulfilled Design Output Medical Device

14 Elements of Design Controls CFR 820.30 Establishment, maintenance, and documentation of: Design and Development Planning Design Input (Requirements) Design Output (Specifications) Design Review Design Verification (Meets specifications) Design Validation (Meets clinical needs) Design Transfer (Moves from design to manufacturing) Design Changes Design History File

15 Medical Device Classification Class I - General Controls Minimal potential harm to the user, e.g., elastic bandages, examination gloves Prohibit adulteration and misbranding; allow for bans and restriction of sale and distribution; notification of risk and of repair, replacement or refund; registration of manufacturers and listing of devices Most are exempt from the premarket notification and/or GMP regulation Class II General Controls with Special Controls Those for which general controls alone are insufficient to assure safety and effectiveness, e.g. infusion pumps, surgical drapes Special controls may include special labeling requirements, mandatory performance standards, patient registries, postmarket surveillance Few Class II devices are exempt from the premarket notification Class III - General Controls and Pre-Market Approval Those for which insufficient information exists to assure safety and effectiveness solely through the general or special controls, e.g., implantable pacemaker, artificial disc replacement Pre-Market Approval (PMA) application scientific review Clinical data may be necessary - IDE/HDE

16 Design Control Scope 820.30(a) Design control requirements apply to: All Class II and Class III devices The following class I devices: Devices automated with computer software Tracheobronchial suction catheters Surgeon's gloves Protective restraints Manual radionuclide applicator system Radionuclide teletherapy source

17 Design & Development Planning - 820.30(b) Product concept is usually outlined in a set of marketing requirements and include the wish list for device features and functionality Is there a market opportunity for the device? Where will the new device be used and who will use it? How will the new device be used? How long will the new device be used? Once the project moves forward, a project plan is drawn up to guide and control the design and development of the device Implementation of design controls is initiated

18 Design & Development Planning - 820.30(b) Establish and maintain plans that: Describe or reference design and development activities Define responsibility for implementation Identify or describe interfaces with different groups or activities Review, update and approve plans as design and development evolves

19 Design & Development Planning - 820.30(b) Elements include: Major tasks Deliverables Responsible individuals/units, including interface with suppliers, contractors Resources Project timeline, milestones Major reviews and decision points Review teams Design documentation Planning method: Flow diagram or spreadsheet Gantt chart

20 Design Input - 820.30(c) Design input means the physical and performance requirements of a device that are used as a basis for device design Establish and maintain procedures to assure that device design requirements: Are appropriate Address the intended use of the device Include in procedures a mechanism for addressing requirements that are: Incomplete Ambiguous Conflicting

21 Design Input - 820.30(c) Starting point for product design Usually fall into three categories: Functional requirements specify what the device does, focusing on operational capabilities and processing of inputs and resultant outputs. Performance requirements specify how much or how well the device must perform, addressing speed, strength, response times, accuracy, limits of operation, etc. Also includes device reliability and safety requirements Interface requirements specify characteristics critical to compatibility with external systems; specifically, those characteristics which are mandated by external systems and outside the control of the developers. An example is important in every case is the user and/or patient interface

22 Types of Design Inputs Intended Use User Needs Physical/chemical characteristics Performance characteristics Safety Reliability Environmental limits User interfaces Regulatory requirements Labeling Human factors Maintenance Compatibility with other devices Sterilization Energy source Toxicity and biocompatibility

23 Sources of Design Inputs Customer input through focus groups, surveys, etc. Comparison testing of competitor product for specific performance criteria or other benchmarking activities Review of similar product histories to include production data for scrap, rework, testing Review of complaints, MDRs, FDA Enforcement Reports, recall of similar products or earlier generations of the product Performance requirements stipulated by the FDA, voluntary standards, or other regulatory agencies Literature review Input from cross-functional teams (R&D, Quality, Regulatory, Marketing, Manufacturing, etc.)

24 Design Output - 820.30(d) Design output means the results of a design effort at each design phase and at the end of the total design effort. The finished design output is the basis for the device master record. The total finished design output consists of the device, its packaging and labeling, and the device master record Establish and maintain procedures for Defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements Containing or referencing acceptance criteria Ensuring design outputs essential for the proper functioning of the device are identified

25 Design Output - 820.30(d) Design output are the design specifications which should meet design input requirements, as confirmed during design verification and validation and ensured during design review The output includes the device, its labeling and packaging, associated specifications and drawings, and production and quality assurance specifications and procedures

26 Types of Design Output Production specifications and other descriptive materials that define and characterize the design Production specifications used to procure components, fabricate, inspect, install, maintain and service the device Other materials that are necessary for establishing conformance, but are not used in device production: Biocompatibility test results Bio-burden test results

27 Design Review - 820.30(e) Documented, comprehensive, systematic examination of a design to evaluate adequacy of requirements, capability of design to meet requirements, and to identify problems Establish and maintain procedures for design reviews Plan and conduct formal documented design reviews of the design results at appropriate stages Include at each design review Representatives of all functions concerned Independent reviewer for the stage being reviewed Any specialists needed

28 Considerations for Design Review Formal process to evaluate design, resolve concerns, and implementation of corrective actions Required for phase approval Formal design review prior to release for manufacture and distribution Responsibility and authority for decision-making Determine frequency of review and design phases for review Controlling documentation and record keeping Competency of reviewers in terms of background and experience

29 Design Verification 820.30(f) vs. Design Validation 820.30(g) Design Verification Did I make the product right? i.e. specified requirements have been fulfilled Design Validation Did I make the right product? i.e. device specifications conform with user needs and intended use(s)

30 Design Verification 820.30(f) Establish and maintain procedures for verifying the device design Confirm design output meets design input requirements Types of verification activities include: Inspections Tests Analyses

31 Examples of Design Verification Activities Evaluation of physical characteristics Mechanical/chemical/electrical testing Failure modes and effects analysis Package integrity testing Biocompatibility testing Sterility testing Comparison against predicate device with successful history

32 Design Validation 820.30(g) Establish and maintain procedures for validating the device design Perform design validation Conform to defined user needs and intended uses Under actual or simulated use conditions Under defined operating conditions On initial production units, lots, or batches or their equivalents Includes software validation and risk analysis, where appropriate

33 Examples of Design Validation Activities Clinical studies Consumer preference testing Bench testing under simulated use conditions Review of labeling and packaging Environmental testing (e.g. stress during shipping and installation) Testing that distinguishes between customers, users, and patients

34 Design Transfer 820.30(h) Establish and maintain procedures to ensure that the device design is correctly translated into production specifications Design moves from R&D to manufacturing Specifications ensure that manufactured devices are repeatedly and reliably produced within product and process capabilities

35 Types of Production Specifications Assembly drawings Component procurement specifications Workmanship standards Manufacturing instructions Inspection and test specifications Computer-aided manufacturing programming files Manufacturing jigs/fixtures

36 Design Transfer: Elements Product risk documentation shows what failure modes, components, parts, etc. should be monitored and trended Process validations indicate critical control points for manufacturing and design

37 Design Changes - 820.30(i) Establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation Establish design change control procedures, determine the risk presented by the change and implement design control requirements to the extent appropriate for that risk

38 Considerations for Document Control Enumeration of design documents, and tracking their status and revision history Maintain master list or document index to define product and/or process Maintain history of document revisions Periodically inventory controlled documents to confirm that contents are up to date Develop procedure for removal and deletion of obsolete documents

39 Considerations for Change Control Document decisions for change requests identifies possible design matter which the requester believes should be corrected Document change order following acceptance of change request identifies relevant document(s) and revision of document content May also include assignment to designers to further evaluate the problem and develop corrective action Review and assess impact of design change on design input requirements and intended uses

40 Design Change Review The effect of the changes on products already in commercial or clinical use Whether prior design verification or validation results are impacted Changes are implemented in such a manner that the original problem is resolved and no new problems are created; or if new problems are created, they are also tracked to resolution

41 Design History File - 820.30(j) Establish and maintain a design history file for each type of device Include in the DHF or reference records necessary to demonstrate that the design was developed in accordance with the design plan and CFR 820 requirements Design history file (DHF) means a compilation of records which describes the design history of a finished device

42 Design History File Tells the story of a device or device family s development from beginning to end Contains the documentation necessary to assure changes to design or process for manufacture of the device do not adversely impact the device design and lead to improvement in design and process Maintains knowledge base after design decisions have been made, employees leave, company reorganizes, or subcontractors unavailable etc.

43 Elements of Design History File Design and development plan Design input documents Risk analyses documents Design output Pre-production design change control records Engineering/laboratory notebooks V&V protocols and results Design review records Design transfer records Copies of controlled documents used during the design process, e.g. SOPs etc. Initial DMR

44 Benefits of Design Controls Improve the product development process by producing higher quality devices Reduce liability exposure Reduce design related product recalls Shorten development time Recognize problems earlier and make corrections Lower manufacturing costs Meet customer expectations (user and patient needs)

45 Other Quality System Requirements Medical Devices Directive (MDD 93/42/EEC): developed in 1993; lists requirements regarding medical device design within EU Revised in 2007 (2007/47/EC) Compliance became mandatory in March, 2010

46 Medical Devices Directive All devices require clinical data Of the actual device or a similar device Scientific literature of a similar device Human factors considered a key factor

47 ISO Standards on Quality ISO 13485 (Medical devices - Quality management systems - Requirements for regulatory purposes): standard that recommends procedures in order to have a quality system that complies with Medical Devices Directive (MDD 93/42/EEC) Based on ISO 9001, but excludes some of the requirements that are not appropriate for regulatory requirements

48 ISO 13485 Similar elements to 820.30 ISO 13485 Section 7.3 Design and development Design and development planning Design and development inputs Design and development outputs Design and development review Design and development verification Design and development validation Control of design and development changes

49 ISO 14971 Provide framework within which experience, insight and judgment are applied systematically to manage the risks Risk management is the systematic application of Management policies Procedures and practices for analyzing, evaluating, and controlling risk

50 ISO 14971 Specifies process through which the manufacturer can identify hazards associated with a medical device, estimate and evaluate the risks associated with these hazards, control these risks, and monitor the effectiveness of that control

51 Quality System Inspection Technique (QSIT) Guidance for auditing medical device manufacturers against QS Regulation (21 CFR Part 820) and related regulations Focuses on four major subsystems that are the basic foundation of the QS to evaluate and states the purpose and importance of each subsystem http://www.fda.gov/downloads/iceci/inspections/i nspectionguides/ucm085938.pdf

52 What are the four main subsystems? Design Controls Material Controls Corrective & Preventive Actions Management Records, Documents, & Change Controls Production & Process Controls Equipment & Facility Controls

53 What does FDA look for in the Design Control Subsystem? Design procedures and plan established? Design inputs or requirements for device identified? Design outputs or specifications for device identified? Design verification conducted? Design validation conducted? Software validation completed? Risk analysis carried out? Design reviews conducted? Design transfer to manufacturing completed successfully?

54 Warning Letters - Examples do not have design control procedures no documentation of design validation or design change controls no documentation of risk analysis or structural (b)(4) testing of the embedded software

55 Warning Letters - Examples unable to demonstrate that the validation for the redesign project was validated using initial production units the verification study for the coating which is applied to the was conducted after the validation study was completed failed to demonstrate bench testing of the [X] had been completed, prior to their installation into [Y] for design validation studies failed to establish acceptance specifications prior to conducting verification and validation studies

56 Warning Letters - Examples Design validation was not executed or documented in order to ensure that the devices conform to defined user needs and intended uses A design and development plan that describes or references the design and development activities and define responsibility for implementation was not established or maintained A design history file was not established or maintained by your firm

57 Warning Letters - Examples Design Input for [X] contains no inputs for the [Y] which are to be used with the [X] Because you did not include [Y] inputs in your [X] design verification/validation, your [X] Risk Analysis/Failure Modes and Criticality Analysis, document (b)(4) is incomplete in that you failed to identify and evaluate potential hazards involved with use of the [Y]

58 Product Liability/Negligence Claims Design Defects Foreseeable risks? Could have been reduced or avoided? Omission of reasonable, safer, and technologically feasible alternative design? Risks outweigh benefits?

59 Key Documents to Consider During Litigation DHF Regulatory documents (510(k), memo to file, PMA) DFMEA PFMEA Specifications Lot/device history records Manufacturing procedures Design drawings CAPA reports MDR/complaint files IFU Surgical brochures Marketing brochures/documents Audit reports

60 Impact on Litigation Design changes Design controls for line extensions Memo to file justification Questions to consider: After a design change or line extension, are there additional risks for the device? Is additional design verification or validation needed? Are the specifications clinically relevant? How were they developed? Are the specifications consistent? Were the inputs translated from marketing inputs into engineering inputs? Were finished products considered? Was validation completed before verification? Was there compliance with own requirements/procedures?

61 Final Thoughts The design control regulations provide a comprehensive and systematic framework for device development The design control process is iterative Design control regulations are a regulatory requirement, but when used properly serve as good business practice

Design control does not end with the transfer of a design to production. Design control applies to all changes to the device or manufacturing process design, including those occurring long after a device has been introduced to the market. This includes evolutionary changes such as performance enhancements as well as revolutionary changes such as corrective actions resulting from the analysis of failed product. The changes are part of a continuous, ongoing effort to design and develop a device that meets the needs of the user and/or patient. Thus, the design control process is revisited many times during the life of a product. 62 Design Control Guidance for Medical Device Manufacturers; March 11, 1997; FDA, CDRH

63 Reading Materials FDA Guidance Document (http://www.fda.gov/downloads/medicaldevices/deviceregulationandguidance/ GuidanceDocuments/ucm070642.pdf) Medical Devices Directive (http://eurlex.europa.eu/lexuriserv/lexuriserv.do?uri=oj:l:2007:247:0021:0055:en:pdf) ISO 13485 (http://www.iso.org/iso/catalogue_detail?csnumber=36786) ISO 14971 (http://www.iso.org/iso/catalogue_detail.htm?csnumber=38193) Quality System Inspection Techniques (http://www.fda.gov/downloads/iceci/inspections/inspectionguides/ucm08593 8.pdf) AdvaMed Points to consider when preparing for an FDA inspection under the QSIT design controls subsystem (http://www.compliance-alliance.com/wpcontent/uploads/2010/05/points-design.pdf)

64 Thank you! Questions? Kevin Ong, Ph.D., P.E. kong@exponent.com 215-594-8874