A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS



Similar documents
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

How To Control A Record System

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Thermo Scientific ConFlo IV Universal Interface

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

Thermo Scientific SIEVE Software for Differential Expression Analysis

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

LIFECYCLE. Thermo Scientific. Enterprise Solutions

Thermo Scientific Qtegra Intelligent Scientific Data Solution Software. Delivering quality. Driving productivity

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Bio-Rad Laboratories. QC data management solutions. Introduce Your Laboratory to a Whole New World of Unity Data Management Solutions

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

What is Eluent Generation?

FT-NIR for Online Analysis in Polyol Production

Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist

limit to percentage levels, a limited working range translates

Thermo Scientific ExactFinder Software

How To Use The Nexion 300

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Computer Specifications

MILESTONE H E L P I N G C H E M I S T S

DMA-1 Direct Mercury Analyzer

The EliA System Time for the essentials Cost efficient and flexible A boost in service for your laboratory and your clinicians

Image Lab Software for the GS-900 Densitometer

WATERS QUANTITATIVE ANALYSIS solutions

Foreign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund

Elemental Scientific. Stability. Elemental Scientific

[NUGENESIS SAMPLE MANAGEMENT ] AMPLE IMPROVING LAB EFFICIENCY, ANAGEMENT ACCELERATING BUSINESS DECISIONS. bigstock.com $69

Get the benefits of Norgren s unique range of Online services

FMA-80 Fluorescence Mercury Analyzer

QIAsymphony RGQ Protocol Sheet

aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy

Report on Government Information Requests

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Agilent 720/730 Series ICP-OES

A Complete Solution for Method Linearity in HPLC and UHPLC

Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES

Image Lab Software How to Obtain Stain-Free Gel and Blot Images. Instructions

CDC Blood Lead Reference Value, Lead Exposure, and Primary Prevention. David E. Jacobs, PhD, CIH Research Director June 2012

Reporting practices for domestic and total debt securities

How many students study abroad and where do they go?

Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System

Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA

How does a venture capitalist appraise investment opportunities?

Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry

GLOBAL DATA CENTER INVESTMENT 2013

41 T Korea, Rep T Netherlands T Japan E Bulgaria T Argentina T Czech Republic T Greece 50.

ProteinChip Energy Absorbing Molecules (EAM)

Make the invisible visible! SENSORS WITH EXCELLENT BACKGROUND SUPPRESSION

Gas Chromatography Liner Selection Guide

Agilent P940xA/C Solid State PIN Diode Switches

Global Marine Insurance Report 2008

AN2604 Application note

Global Investing 2013 Morningstar. All Rights Reserved. 3/1/2013

Diode, Schottky SMB/DO-214AA. Page <1> 24/04/08 V1.1. Dimensions Inches (Millimeters)

BT Premium Event Call and Web Rate Card

PROPOSAL GENERAL BUSINESS AND PRODUCTS LIABILITY INSURANCE

PROPOSAL GENERAL BUSINESS AND PRODUCTS LIABILITY INSURANCE

All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &

ST High voltage fast-switching NPN power transistor. Features. Applications. Description

Report on Government Information Requests

Consumer Credit Worldwide at year end 2012

Determination of PCDD/Fs in Environmental Samples using Accelerated Solvent Extraction (ASE) and GC-MS/MS

GLOBAL EDUCATION PROGRAM

CNE Progress Chart (CNE Certification Requirements and Test Numbers) (updated 18 October 2000)

How To Test For Contamination In Large Volume Water

IOOF QuantPlus. International Equities Portfolio NZD. Quarterly update

Customer Support. Superior Service Solutions for Your Laser and Laser Accessories. Superior Reliability & Performance

AT-S39 Version 1.3 Management Software for the AT-8024 and AT-8024GB Fast Ethernet Switches. Software Release Notes

UniFlux 10: hollow fiber conversion kit ( ) installation instruction

Application Note. USD 2995 (a) High Throughput Regeneration Study on MEP HyperCel Mixed-Mode Sorbent on ScreenExpert RoboColumns u

How to Efficiently Recover Merchant Fees Using ATPCO s Ticketing Fees Solution. 6 August 2013

Acrodose PL System. Enhanced Patient Safety. A New Platform for Pre-storage Pooling and Testing of Leukoreduced Whole Blood Derived Platelets

Accuracy counts! SENSORS WITH ANALOG OUTPUT

Clinical and Laboratory Practices Workshops. Helping to improve laboratory skills and the quality of care in resource-limited settings

Canada GO 2535 TM World Traveller's edition Maps of North America (Canada, US, Mexico), Western and Central Europe (including Russia) CAD 349,95

Electronic Citizen Identities and Strong Authentication

MANDATORY PROVIDENT FUND SCHEMES AUTHORITY

Report on Government Information Requests

*Elevated Blood Lead Level 10 µg/dl. Objectives. Why Is Lead Poisoning Still An Issue? Childhood Lead Poisoning Prevention Program

BD241A BD241C. NPN power transistors. Features. Applications. Description. NPN transistors. Audio, general purpose switching and amplifier transistors

Brochure More information from

AN3353 Application note

[ Care and Use Manual ]

BD238. Low voltage PNP power transistor. Features. Applications. Description. Low saturation voltage PNP transistor

Brochure More information from

Agilent E6832A W-CDMA Calibration Application

Transcription:

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics, USA Application Note 43137 Key Words Blood, Research, ICP-MS, Lead Goal To develop a fully quantitative, high throughput research method for the determination of Pb levels in blood by ICP-MS. Introduction Man has used the unique properties of Lead (Pb) in a wide variety of ways for hundreds of years. For example the Romans used it for plumbing (lead is plumbum in Latin). More recently, lead was used in gasoline and paints to improve performance before it was phased out in the 1970s when the dangers associated with its use were first recognized. A series of exposure pathways allow lead to enter the body, including inhalation of particles (e.g. dust from soils exposed to gasoline and paint lead sources) as well as drinking water and (particularly for children) contaminated toys. On entering the body, lead sequesters itself into soft tissue and bone, where it can be released back to the blood. Young children and fetuses are particularly vulnerable since the blood-brain barrier is not yet fully developed and absorption rates are higher than in adults. Currently, there is no agreed safe concentration of lead in blood; however the United States Environmental Protection Agency (EPA) has defined standards for dangerous levels of lead in dust, paint and soil 1 to safeguard the population. The United States Centers for Disease Control and Prevention (CDC) states that Blood Lead Levels (BLL) >70 µg/dl (700 ng/ml) can cause serious health effects. BLL as low as 10 µg/dl (100 ng/ml) are associated with cognitive development, growth, and behavioral issues in children between the ages of 1-5 years. The CDC now designates a reference value for BLL of >5 ug/dl as elevated in children ages 1-5. The Occupational Safety & Health Administration (OSHA) also has guidelines for adult BLLs as it relates to work place exposure. All companies that expose their employees to lead must test employees at least once a year. OSHA guidelines include medical removal from the job if levels are > 60 µg/dl (600 ng/ml), and require follow up testing before returning to work. Because of these requirements, the United States performs millions of Blood Lead Level tests per year. All 50 States have BLL programs and require results to be reported back for statistical purposes. There are remaining questions about the full implications of lead exposure. Epidemiological studies may help answer these questions. A simple, fast reliable method would facilitate this research.

2 Historically, atomic absorption (AA) has been employed for the determination of Pb in whole blood. However, with analysis times often in excess of three minutes, this technique can be quite slow, especially if complicated ashing and atomization programs are used, leading to reduced sample throughput. In addition, AA is not a multi-element approach and often utilizes matrix-modifiers, which can complicate the sample preparation and add cost to the analysis. More recently, ICP-MS has emerged as a rapid, multi-element alternative, with improved detection limits (allowing for reduced sample amounts) and sample throughput over AA. Here, we report a simple, rapid method for BLL by ICP-MS and evaluate between run consistency and sample carry-over. Sample Preparation Because of the high instrumental sensitivity of the Thermo Scientific icap Q ICP-MS, samples and reference standards could be diluted 100 times in 0.5% HNO 3 (prior to an additional 2x dilution from on-line addition of an internal standard) while still maintaining the required detection limits. An additional advantage of the relatively high dilution factor used is the improved nebulization of whole blood samples as well as reduced maintenance of the sample introduction system, both leading to longer analysis periods. Calibration was carried out using external aqueous standards in 0.5% HNO 3 with concentrations of 0.001, 0.010, 0.100, 1.00, 2.00, 5.00 and 10.0 ng/ml. The method was validated by measuring the Lyphochek Whole Blood Level Controls (Levels 1, 2 and 3 (561, 562 and 563)) (Bio-Rad Laboratories Inc.). Instrument Settings All analyses were made using an icap Qc ICP-MS. While the icap Qc is equipped with collision cell technology for the suppression of spectral interferences, it was not used in this study and the icap Qc ICP-MS was therefore operated in STD mode (Table 1). The standard icap Qc sample introduction system consisted of a PFA-ST nebulizer, quartz glass peltier-cooled cyclonic spraychamber and a 2.5 mm ID quartz injector. The 1:100 diluted samples were presented for analysis using a SC-2DX autosampler with integrated 7-port FAST valve from Elemental Scientific Inc (Omaha, NE, USA). A 1 ng/ml Bi (in 0.5% HNO 3 ) internal standard was added on-line (a further 1:1 dilution) using the 7-port FAST valve, giving a total uptake rate of 400 µl/min (an effective 200x sample dilution). Through the combination of the icap Q and the FAST sample introduction system, sample to sample analysis times (including uptake, data acquisition and washout) of < 24 seconds per sample were achieved. Table 1. icap Qc instrument parameters. Parameter Peristaltic pump speed Pump tubing Nebulizer Interface cones RF Power Cool gas flow Auxiliary gas flow Nebulizer gas flow Injector Type; ID Value 32 rpm orange/green tubing for both carrier (sample) and internal standard PFA-ST Nickel 1550 W 14 L/min 0.8 L/min 1.06 L/min Quartz; 2.5 mm Results The analytical figures of merit for five consecutive runs are listed in Table 2, and an example calibration curve is shown in Figure 1. In order to keep analysis times at a minimum, particularly important when sample sizes are limited as is often the case in clinical analyses, a dwell time of 10 ms was used with only 10 sweeps. The instrument detection limit (IDL) obtained with aqueous calibration standards prepared in 0.5% HNO 3, was ~1 ng/l (0.0001 µg/dl) and the blank equivalent concentration (BEC) was ~2 ng/l. Sensitivity is expressed as 206 Pb+ 207 Pb+ 208 Pb and is half of the normal STD mode level, due to the 1:1 on-line IS addition. The percentage recovery of the internal standard, Bi, over the course of the analysis (Run #3) is shown in Figure 2 and ranged from 94% to 103%. This indicates the analytical robustness of the method and freedom from matrix suppression.

Run # Analyte Dwell (sec) Mode Sweeps Sensitivity cps/ng/l R value BEC (ng/l) LOD (ng/l) 1 Pb 0.010 STD 10 125 0.999996 1.7 1.0 2 Pb 0.010 STD 10 140 0.999990 1.9 0.9 3 Pb 0.010 STD 10 142 0.999993 0.8 1.1 4 Pb 0.010 STD 10 121 0.999950 1.3 0.2 5 Pb 0.010 STD 10 137 0.999991 1.6 0.8 Table 2. Analytical figures of merit for five consecutive analytical runs. 3 Figure 1. Example calibration curve for Pb from 0 to 10 ng/ml. The insert shows the same curve with concentrations between 0 and 0.01 ng/ml. Figure 2. Screenshot from the Thermo Scientific Qtegra software showing the percentage recovery of the internal standard, Bi, during Run #3.

4 To determine the ability of the icap Q to accurately achieve low level BLL measurements, an additional 10x dilution was performed on the low level control (Table 3). The excellent agreement (97-105 %) with the reference values indicates that the method could easily support dilution factors in excess of the 200x used in this work. Table 3. Serial dilution results for the low level Bio-Rad control (561). Dilution Factor Pb Meas. (µg/l) D.F. Corr. (µg/dl) Certified Value (ug/dl) % Recovery 1000 0.085 8.51 8.14 105 100 0.810 8.10 8.14 100 1000 0.079 7.89 8.14 97 100 0.854 8.54 8.14 105 1000 0.083 8.28 8.14 102 100 0.794 7.94 8.14 98 Due to the clinical implications of this test, it is also important to assess the washout characteristics of the sample introduction system, and the likelihood that a false positive could be generated by inadequate washout from a previously analyzed sample. To assess this, a blank was repeatedly analyzed after five measurements of the high control standard (Figure 3). Using a FAST washout of two seconds, the lead background was reduced by >2 orders of magnitude. This equates to a BLL of ~0.001 µg/dl and is, therefore, a more accurate measure of the effective detection limit of the method described. Figure 3. Washout results from five analyses of the high (Level 3) Bio-Rad control followed by five blank measurements. Finally, an assessment of the accuracy of the proposed method was carried out by analyzing the control standards during each of the five analytical runs. The results for the BioRad Level 1-3 controls are given in Table 4.

Table 4. BLL (µg/dl) measured as the sum of the stable Pb isotopes 206, 207 and 208 for the Bio-Rad Control Levels 1-3. The % RSD for each measurement is also shown. 5 Run # Analyte Level 1 % RSD Level 2 % RSD Level 3 % RSD Range (AA) Pb 6.28-9.42 20.4-30.6 34.2-51.3 Value (AA) Pb 7.85 25.5 42.8 Value (ID-ICP-MS) Pb 8.14 25.2 43.2 Run 1 1a Pb 7.98 4.8 25.2 1.9 41.9 1.9 Run 2 Pb 2a Pb 8.22 1.3 25.7 2.0 42.2 1.6 2b Pb 8.16 2.6 25.5 1.2 42.1 0.8 2c Pb 8.09 2.7 25.7 1.7 41.8 2.1 Mean ± SD 8.16 ± 0.065 25.6 ± 0.12 42.0 ± 0.21 % RSD 0.80 0.45 0.50 Run 3 3a Pb 8.10 1.4 25.7 0.4 43.5 0.1 3b Pb 8.32 1.2 25.5 1.6 43.0 2.0 3c Pb 8.16 0.7 25.1 1.1 42.4 1.3 Mean ± STD 8.19 ± 0.11 25.4 ± 0.31 43.0 ± 0.55 % RSD 1.4 1.2 1.3 Run 4 4a Pb 8.54 3.8 25.8 1.3 40.6 3.1 4b Pb 8.40 0.6 26.5 2.9 41.2 1.9 4c Pb 8.43 2.9 26.0 0.8 41.3 2.5 4d Pb 8.29 3.3 26.1 3.0 41.5 0.9 Mean ± STD 8.42 ± 0.10 26.1 ± 0.29 41.2 ± 0.39 % RSD 1.2 1.1 0.94 Run 5 5a Pb 7.94 0.6 25.0 1.1 42.5 1.3 5b Pb 8.15 2.8 25.9 1.3 42.4 2.6 5c Pb 8.19 1.3 25.1 1.0 41.6 2.3 5d Pb 8.25 2.2 25.0 1.5 43.0 2.4 5e Pb 8.07 1.2 24.5 2.2 42.5 1.4 Mean ± STD 8.12 ± 0.12 25.1 ± 0.50 42.4 ± 0.50 % RSD 1.5 2.0 1.2 Run 1 5 Mean ± STD 8.21 ± 0.16 25.5 ± 0.51 42.1 ± 0.75 % RSD 2.0 2.0 1.8

Conclusions A fully quantitative method for the determination of blood Pb by ICP-MS has been described. Calibration using aqueous calibration standards is possible and the accuracy was assessed by the measurement of control samples. The proposed method, with an instrumental detection limit of 0.0001 µg/dl (and effective method detection limit of 0.001 µg/dl due to sample carry-over) can be used for the analysis of real samples with dilution factors of 1000 as demonstrated by recoveries of the low level control. Due to the high sensitivity of the Thermo Scientific icap Q ICP-MS, this increased dilution factor would further decrease system maintenance, enabling high volume clinical laboratories to process samples at a rate of 2.5 per minute. Application Note 43137 References 1. United States Federal Register, Part III, Environmental Protection Agency, 40 CFR Part 745; Lead: Identification of Dangerous Levels of Lead; Final Rule. Disclaimer The method presented is intended for research purposes only. www.thermoscientific.com 2014 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Lyphochek and Bio-Rad are trademarks of Bio-Rad Laboratories, Inc All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, (Bremen) GmbH Management System Registered is ISO 9001:2008 Certified. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 AN43137_E 06/14C Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Spain +34 914 845 965 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752