Care and Handling of Rotors



Similar documents
Thermo Scientific Rotor Care Guide. Protect your rotor investment

Thermo Scientific Rotor Care Guide. Protect your rotor investment

INSTRUCTIONS FOR USING THE A-110 ROTOR In the Beckman Coulter Airfuge Ultracentrifuge

ROTORS, TUBES, BOTTLES AND CAPS FOR Hitachi Preparative Ultracentrifuge

A. Work on this project shall consist of, but is not limited to, the following:

SW 28 and SW 28.1 ROTORS

TX-1000 Rotor For Thermo Scientific General Purpose Centrifuges

BR-8101K. Rotors, Tubes & Accessories ULTRACENTRIFUGE

VERTICAL / FIXED BEAM CLAMP I-BEAM ANCHOR ADJUSTABLE FROM 4-14 Model # HD26249

SW 32 Ti and SW 32.1 Ti Rotors

L5-TB-061P March Type 70.1 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges Beckman Coulter, Inc.

3. Maintenance of Laboratory Equipment

orlando OWNER S MANUAL

SPROCKETS 1 - TECHNICAL SPECIFICATIONS 2 - COMPATIBILITY WARNING!

Spectrafuge 24D Microcentrifuge

Fleetguard CS41018 SpiraTec Centrifuge Rotor Installation Instructions

Type 90 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges. LXL-TB-005M August 2007

INSPECTION, MAINTENANCE AND RECHARGE SERVICE MANUAL P/N 16303

WARNING: Important: Keep these instructions for future reference. Gate Instructions (1) Assembly and Installation Instructions

Profiles for floors of same height Cerfix Projoint

Archived. Gloves should be changed frequently during the analysis.

BRAKE DRUM AND ROTOR SERVICE INFORMATION

Step-by-step instructions:

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Standard laryngoscope blade assemblies Directions for use

DM 0412 Clinical Centrifuge

Super Slider ADJUSTABLE FROM Model #HD26229

SAFE HEALTHY CHILDREN A Health and Safety Manual for Childcare Providers

Using k-factor to Compare Rotor Efficiency

San josé OWNER S MANUAL

WARNING! WARNING! ENGLISH 18 ENGLISH

Maintenance Instructions

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

Trouble shooting guide

Operations and Instruction Manual Might Swivel - Part # Concrete and Steel Anchorage Connector ANSI Z ,000 lbs / 44kn

Featuring fast rotor exchanges with AUTO-LOCK

Sales and Support in the US and CA by: NuAire, Inc Fernbrook Lane Plymouth, MN

Operating, Installation, and Maintenance Instructions

User s Manual XXL Care Lift, Electrically Elevating Shower-, Commode- & Care Lift with Tilt-in-Space Seat Unit

Depending on which elastic support you have you can secure the V2 power unit to the following seat bracket tubes: EXTERNAL DIAMETER 27.

Dehumidifier Users manual. For Models: DH45S DH65S

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

This instruction is valid for all ACD pump models shown on page 2

Care and use guide. Contents. a handy solution made easy. page

World-Class Oil Sampling It is Possible

Installation Instructions / Warranty

The Drink-Aide The NEW look of independence

Cleaning & Sanitisation

Active and Passive Sash Replacement

2006 HEADSHOK Service Video #1

PREVENTATIVE MAINTENANCE

Repair of Hyd-ro-ac Actuators

CONGRATULATIONS! TABLE OF CONTENTS. 1 Safety Warnings 2 About Your Wheels 3 Setting Up Your Wheels 7 Care and Cleaning 8 Warranty

Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:

DNA Analyst Training Laboratory Training Manual Protocol 2.02 Clean Technique

Training on Standard Operating Procedures for Health Care Waste Management Swaziland 12 May, 2011

PART 17 SAFETY/RESCUE HARNESS JACKET TYPE M 500/1 BRIGADE ORDER TECHNICAL NO. 13 C O N T E N T S. Page No PURPOSE STRATEGIC AIMS AND OBJECTIVES

1000-LB. TRAILER JACK OWNER S MANUAL

Destiny Shower Enclosure Installation Instructions

HYDRAULIC TABLE CART 500-LB.

ArtisanLink Staining System is an automated special stains slide

PHOENIX TAPWARE. Gloss. Shower/Wall Mixer RRP: $ Product Code: Available Finishes: GS780. Chrome MATCH WITH...

D-Ring. Fixed Arm. 1.2 LIMITATIONS: The following application limitations must be recognized and considered before using this product:

Installation Instructions / Warranty

2006 JUDY SERVICE GUIDE

TABLE OF CONTENTS CABINET SAFETY

Horizontal Lifeline. SideWalk Rope. User Instructions. 3M Occupational Health and Environmental Safety. Fall Protection Equipment

Cleaning guide for senior living facilities

English. Symbols used to mark instructions...3. Congratulations...5 Getting the best results...5. Warnings...6 Operating Procedure...

Pallet Jack. OWNER S MANUAL Model MH1230. Important Safety Instructions Assembly Instructions Parts and Hardware Identification

STABITHERM INSTALLATION & SERVICE BOOKLET

Operator s Instructions

User Instruction Manual Vertical Lifelines and Vertical Lifeline Subsystems

IMPORTANT INFORMATION - PLEASE READ. Table of Contents. Introduction. General Information

ARVINMERITOR UNITIZED FRONT WHEEL HUB INSPECTION AND MAINTENANCE

Carnegie Mellon University s Policy and Procedures for Recombinant and Synthetic Nucleic Acid Materials Spills

Infection Control in the Dental Practice through Proper Sterilization

Protocol for Disinfection of Cell Culture and Tissue Culture in Media:

Get Cleaning... User Guide Vax Careline: (UK) (ROI) Multifunction. vax.co.uk Series. What s your Vax model number?

Effective Use of Autoclaves

Assembly Instructions

KEEP FOR FUTURE REFERENCE MRTALP VERTICAL LIFT BAR READ ALL INSTRUCTIONS AND WARNINGS BEFORE USING THIS PRODUCT

STANDARD OPERATING PROCEDURES Safe Autoclave Operations

Centurion. Scientific Limited Making Science Look Good. Quality Profile

North Carolina State University Emergency Facilities Closure Checklist- Part I

T29-8x50 Rotor for Sorvall LYNX Superspeed Centrifuges

OPERATIONS AND MAINTENANCE INFORMATION FOR: THE CANPLAS BACKWATER VALVE THE CHEMTROL BALL SHUT OFF VALVE SUMP PUMPS AND WASTEWATER EJECTOR PUMPS

Biosafety Guidelines. 5. Pipetting Mechanical pipetting devices must be available and used. Mouth pipetting is prohibited.

Owner s Guide and Installation Manual. Vancouver Model Name , Model No. English Español

COASTLINE QUADRANT ENCLOSURE. Installation / Maintenance Instructions Rev

AKRAPOVIC SLIP-ON EXHAUST SYSTEM for the HONDA CB1000R

Fundamental Autoclave Techniques

FAUCET SPECIFICATIONS. Vessel Bathroom Faucet Model VG03023 MODEL VG03023

Scott & White Institutional Biosafety Committee Compliance Program Biohazardous Material Spill Clean-Up Procedure Policy #IBC.002

Installation Instructions / Warranty

Transcription:

Care and Handling of Rotors

Today s Objectives Help you to use centrifuge systems safely Ensure maximum life of your investment Avoid unbudgeted lab expenses Maximize experiment integrity

Table of Contents Proper Rotor Use Proper Use of Tools Rotor Care and Maintenance After-Sales Support Out-of-Warranty Rotors/Rotor Retirement What s in it for you?

Centrifugation around since 1948 The First Ultracentrifuge Spinco Model L

There is a wide variety of centrifugation systems

Correct Rotor Use

Correct Rotor Use depends on Several Factors Use rotors only as part of approved systems Properly balance loads Understand sample density Correct attachment of rotor lids and buckets Proper use of tubes, adapters, accessories, and tools Consult your Operators Manual

Use rotors only with Approved Systems Instruments and rotors are tested as a System (IEC-10-10-2) Only approved Rotor-Instrument Systems are proven to be safe Using non-approved components can result in - Instrument or rotor damage - Unbudgeted expense - Personal injury

The energy of an operating centrifuge is enormous... Situations such as this emphasize the importance of following centrifuge safety and rotor use and care procedures

Correctly Balance Rotor Loads At 1,000,000 x g, 1 gram weighs 1,000 kg Which Equals...The Weight of a Mini Cooper

For Ultracentrifuges... Balance opposing sample loads within 0.5 grams for fixed angle rotors Balance opposing sample loads within 0.2 grams for swinging bucket rotors Consult your Operators Manual

For High Performance and Benchtop Centrifuges Balance specification is instrument / rotor dependent Some rotors offer imbalance correction - 50 to 100 grams! Consult your Operators Manual

Maximum Sample Density for Rotors Is... 1.2 g/ml for most rotors 1.7 g/ml for VTi and NVT rotors Consult your Operators Manual for information on precipitating gradients

Any Guesses??

How About This One??

Proper Rotor Handling SW Rotors - Buckets: Inspect for proper attachment - Handling: Hold SW rotors correctly - Installation: Properly attach rotor to Spindle Fixed Angle Rotors - Lid: Tighten correctly - Installation: Properly attach rotor to Spindle

Proper Handling of SW rotors INCORRECT Improper handling can lead to - Mishooked buckets - Disturbed gradients

Proper Handling of SW rotors CORRECT Hold rotor with both hands

Proper Handling of SW rotors Inspect Buckets Prior to the Run Mis-hooked buckets are the leading cause of rotor mishaps in Ultracentrifuges

Proper Handling of SW rotors Some rotors, such as the SW-32Ti, have simplified the bucket attachment process

Lid Attachment: Optima Series

Rotor Attachment: Optima Series Place the rotor straight down onto the spindle The rotor is self locking

Rotor Attachment: TL Series For TL series rotors with buttons Place the rotor on the spindle

Rotor Attachment: TL Series Push the button to lock the rotor onto the spindle

Rotor Attachment: TL Series Check to ensure proper attachment Rotors without locking buttons are self-locking

Lid Attachment: Bio-safe Rotors For rotors with dual knobs: First, turn the lower knob clockwise to attach the lid to the rotor When the rotor is placed in the centrifuge, tighten the center knob clockwise

Rotor Attachment: For rotors with single knobs: When the rotor is placed in the centrifuge, tighten the knob clockwise The single knob locks the rotor to the spindle and attaches the lid to the rotor

Rotor Attachment: Turn the locking handle clockwise to attach the rotor to the centrifuge spindle

Rotor Preparation: Vertical Rotors Consult the rotor manual for each vertical rotor for correct use of tools

Tubes, Bottles and Adapters A vast selection of tubes, bottles, adapters and other accessories are available for use in your centrifuge Consult the rotor manual for correct use of tubes, bottles, and adapters Improper labware use is a leading cause of rotor mishaps

Tubes, Bottles and Adapters Use only the tubes, bottles, and adapters listed in the rotor manual or Rotors and Tubes guides Third-party tubes and bottles may not handle the high g-forces of your centrifuges Be aware of proper filling levels and cap usage instructions

Proper Use of Tools

Proper Tools Consult your rotor manual for guidance on proper use of tools for rotors and tubes

Improper Tools Improper tool use can damage parts and cause a safety hazard for centrifuge users

Rotor Care and Maintenance

Proper Cleaning Utensils BCI Rotor Cleaning Kit

Cleaning procedure Wash rotor frequently at least weekly Remove O-rings before washing Use mild detergent, such as Solution 555, and soft brushes (both are available in the BCI Rotor Cleaning Kit) Thoroughly rinse with distilled water Air-dry upside down Re-lubricate O-rings with vacuum grease Re-lubricate metal threads with Spinkote Consult your Operators Manual

Proper Brushes Improper bushes can scratch and damage rotor tube cavities

Inspect Periodically Inspect rotor body and tube cavities - Pitting, rough spots, cracks or damage - White deposits may indicate stress corrosion - Discoloration Lubricate rotor and components as indicated in rotor manual Inspect O-rings for cuts, abrasions, or flattened areas; replace if damaged Inspect and replace overspeed disk if damaged Contact Field Service if you have questions Consult your Operators Manual

Stress Corrosion Over time, the combination of tensile load and environment creates stress corrosion, which significantly reduces the service life of the metal, specifically aluminum alloys.

CORROSION can destroy aluminum rotors

NO CORROSIVE ATTACK LOAD = 6000 lbs LOAD = 26,700 N AREA = 1.00 sq in AREA = 6.45 x 10-4 m 2 STRESS = 6000 lbs/sq in STRESS = 4.14 x 10 7 N/m 2 STRESS = 41.4 MPa

CORROSIVE ATTACK LOAD = 6000 lbs LOAD = 26,700 N AREA = 0.60 sq in AREA = 3.87 x 10-4 m 2 STRESS = 10,000 lbs/sq in STRESS = 6.90 x 10 7 N/m 2 STRESS = 69.0 MPa

Points to Remember Anodized aluminum is corrosion resistant, not corrosion proof Although titanium and carbon composite rotors are highly corrosion resistant, these rotors may have anodized aluminum components Consult chemical compatibility charts

Proper Lubrication is Important Periodically lubricate O- rings with vacuum grease Lubricate threaded portions of rotor with Spinkote Lubrication maintains vacuum sealing and enhances smooth operation of components Consult your Operators Manual

Overspeed Disk Replacement

Overspeed Disk Replacement

Overspeed Disk Replacement

Overspeed Disk Replacement

Periodically Inspect Rotor Hub Pins

Decontamination Radioactive material - Use a cleaning agent that will not harm anodized aluminum Toxic or Pathogenic contamination - Follow appropriate procedures outlined by your Laboratory Safety Officer

Sterilization Follow procedures outlined by your Laboratory Safety Officer - Most metal components can be autoclaved up to 1 hour at 121 C (up to 30 minutes for most composite rotors; see manual) - Cold methods 70% ethanol, 6% hydrogen peroxide can be used Bleach may discolor anodized surfaces

Rotor Storage Store rotors upside-down to ensure that condensation does not collect in tube cavities Store rotors in a dry environment Remove all tube adapters from tube cavities before storage

Run Logging Run Logging is not required for warranty Using a logbook can help lab managers track rotor and instrument usage patterns

Rotor Care Review Use proper cleaning solutions Use proper cleaning brushes Use proper storage techniques Read the rotor instruction manual Ask your Field Engineer

Out-of-Warranty Rotors and Rotor Retirement

Rotor Life and Rotor Warranty Rotor life is not related to warranty period Rule-of-thumb rotor lifetimes - Optima Ultracentrifuge rotors FA and VTi rotors:12 years SW rotors: 10 years - High Performance Centrifuge rotors JA rotors: 15 years When in doubt, consult with Service to determine safe rotor usage

After Sales Support

Field Rotor Inspection Program Check if available as part of your service agreement Ensures a long service life for your rotors Increases lab safety Minimizes lab down time Minimizes repair and replacement expenses

Rotor Inspection by Factory-Trained Inspectors The inspector will Examine your rotors with non-destructive techniques Indicate possible repairs or modifications to run protocols Recommend rotor retirement based on condition, run cycles, or age Advise on care and handling techniques

What s In It For You? Helps you use centrifuge systems safely Ensures maximum life of your investment Avoids unbudgeted lab expenses Helps assure experiment integrity Helps insure against loss of valuable sample and lab preparation time

Any Questions?

Thank You!