SW 28 and SW 28.1 ROTORS
|
|
|
- Laurel Bryan
- 10 years ago
- Views:
Transcription
1 L5-TB-069M July 2002 TM SW 28 and SW 28.1 ROTORS Used in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2002 Beckman Coulter, Inc. Published by the Centrifuge Instrument Systems Development Center of Beckman Coulter, Inc., Palo Alto, California 94304
2 ! SAFETY NOTICE This safety notice summarizes information basic to the safe use of the rotors described in this manual. The international symbol displayed above is a reminder to the user that all safety instructions should be read and understood before operation or maintenance of this equipment is attempted. When you see the symbol on other pages throughout this publication, pay special attention to the specific safety information presented. Observance of safety precautions will also help to avoid actions that could damage or adversely affect the performance of the rotors. These rotors were developed, manufactured, and tested for safety and reliability as part of a Beckman Coulter ultracentrifuge/rotor system. Their safety or reliability cannot be assured if used in a centrifuge not of Beckman Coulter s manufacture or in a Beckman Coulter ultracentrifuge that has been modified without Beckman Coulter s approval.!!!!!!! Handle body fluids with care because they can transmit disease. No known test offers complete assurance that such fluids are free of micro-organisms. Some of the most virulent Hepatitis (B and C) viruses, HIV (I V), atypical mycobacteria, and certain systemic fungi further emphasize the need for aerosol protection. Operator error or tube failure may generate aerosols. Do not run toxic, pathogenic, or other hazardous materials in this rotor unless you take all appropriate safety precautions. Handle all infectious samples according to good laboratory practices and methods to prevent the spread of disease. Ask your laboratory safety officer to advise you about the level of containment required for your application and about the proper decontamination or sterilization procedures to follow if fluids escape from containers. Biosafe containment should be used when Risk Group II materials (as identified in the World Health Organization Laboratory Biosafety Manual) are handled; materials of a higher group require more than one level of protection. Because spills may generate aerosols, observe proper safety precautions for aerosol containment. The rotors and accessories are not designed for use with materials capable of developing flammable or explosive vapors. Do not centrifuge such materials in nor handle or store them near the centrifuge. Although rotor components and accessories made by other manufacturers may fit in the SW 28 and SW 28.1 rotors, their safety in these rotors cannot be ascertained by Beckman Coulter. Use of other manufacturers components or accessories in these rotors may void the rotor warranty and should be prohibited by your laboratory safety officer. Only the components and accessories listed in this publication should be used in this rotor. Hook all six buckets, loaded or empty, to the rotor for every run. Make sure that filled containers are loaded symmetrically into the rotor and that opposing tubes are filled to the same level with liquid of the same density. Make sure that buckets containing Quick-Seal tubes have the proper floating spacers inserted (if applicable) before installing the bucket cap. Never exceed the maximum rated speed of the rotor and labware in use. Refer to the section on RUN SPEEDS. If disassembly reveals evidence of leakage, you should assume that some fluid escaped the rotor. Apply appropriate decontamination procedures to the centrifuge and accessories as required. Do not use sharp tools on the rotor that could cause scratches in the rotor surface. Corrosion begins in scratches and may open fissures in the rotor with continued use. 2
3 SW 28 ROTOR SW 28.1 ROTOR rmin rav rmax Axis of Rotation rmin rav rmax Axis of Rotation SPECIFICATIONS SW 28 SW 28.1 Maximum speed rpm rpm Density rating at maximum speed g/ml g/ml Relative Centrifugal Field* at maximum speed at r max (161.0 mm) g at r max (171.3 mm) g at r av (118.2 mm) g at r av (122.1 mm) g at r min (75.3 mm) g at r min (72.9 mm) g k factor at maximum speed k factors at maximum speed (5 to 20% sucrose gradient; 5 C) When particle density = 1.3 g/ml When particle density = 1.5 g/ml When particle density = 1.7 g/ml Conditions requiring speed reduction see RUN SPEEDS see RUN SPEEDS Number of buckets Available tubes see Table see Table 3 Nominal tube dimensions (largest tube) mm mm Nominal tube capacity (largest tube) ml ml Nominal rotor capacity ml ml Approximate acceleration time to maximum speed (rotor fully loaded) to 5 min to 5 min Approximate deceleration time from maximum speed (rotor fully loaded) to 5 min to 5 min Weight of fully loaded rotor kg (13 lb) kg (12.7 lb) Rotor material aluminum body; aluminum body; titanium buckets titanium buckets * Relative Centrifugal Field (RCF) is the ratio of the centrifugal acceleration at a specified radius and speed (rω 2 ) to the standard acceleration of gravity (g) according to the following formula: rω RCF = g where r is the radius in millimeters, ω is the angular velocity in radians per second (2 π RPM /60), and g is the standard acceleration of gravity (9807 mm/s 2 ). After substitution: RPM RCF = 1.12 r
4 DESCRIPTION SW 28 Adapter SW 28.1 Adapter Drive Pins Rotor Stand (332400) Drive Pins Rotor Stand (332400) These Beckman Coulter rotors have been manufactured in an NSAIregistered ISO 9001 or 9002 facility for use with the appropriately classified Beckman Coulter ultracentrifuges. The SW 28 and SW 28.1, 1 both rated for rpm, are swinging bucket rotors designed to centrifuge up to six tubes each. Used in Beckman Coulter class H, R, and S preparative ultracentrifuges, these rotors develop centrifugal forces for the separation of subcellular particles and viruses in density gradients. The rotors have a common rotor body with buckets that can be used interchangeably (see ROTOR PREPARATION). Bucket and rotor body positions are numbered for operator convenience. The SW 30 and SW 30.1 rotor buckets can be used on the SW 28/ SW 28.1 rotor body as well. However, the reverse is not true. The matrix in Table 1 indicates interchangeability of rotor buckets between the SW 30 series, SW 28 series, and the older SW 27 series of rotors. Table 1. Rotor Bucket Interchangeability May be used with rotors Buckets SW 30.1 SW 30 SW 28.1 SW 28 SW 27.1 SW 27 SW 30.1/SW 30 Yes Yes Yes Yes No No SW 28.1/SW 28 No No Yes Yes No No SW 27.1/SW 27 No No No No Yes Yes The rotor body and bucket caps are made of aluminum, anodized for corrosion resistance. The buckets are made of titanium, finished with clear polyurethane paint. Each bucket and cap assembly hooks into grooves on the rotor body. Bucket and rotor body positions are numbered for operator convenience. O-rings, made of Buna N rubber, between each bucket and bucket cap maintain atmospheric pressure inside the buckets during centrifugation. Drive pins in the rotor drive hole prevent the rotor from slipping on the centrifuge drive hub during acceleration and deceleration. 1 U.S. Patent Nos. 4,102,490 and 4,190,195; Canadian Patent No. 1,120,903. 4
5 For overspeed protection, a Beckman Coulter ultracentrifuge equipped with a photoelectric detector will monitor the overspeed disk on the adapter bottom and shut down the run if a speed exceeding the maximum allowable run speed is detected. For warranty information, see the Warranty at the back of this manual. PREPARATION AND USE Specific information about the SW 28 and SW 28.1 rotors is given here. Information common to these and other rotors is contained in the manual Rotors and Tubes for Preparative Ultracentrifuges (publication LR-IM), which should be used together with this manual for complete rotor and accessory operation. NOTE Although rotor components and accessories made by other manufacturers may fit in the SW 28 and SW 28.1 rotors, their safety in these rotors cannot be ascertained by Beckman Coulter. Use of other manufacturers components or accessories in these rotors may void the rotor warranty and should be prohibited by your laboratory safety officer. Only the components and accessories listed in this publication should be used in these rotors. PRERUN SAFETY CHECK! Read the Safety Notice page at the front of this manual before using the rotor. 1. Make sure that the rotor, buckets, and caps are clean and show no sign of corrosion or cracking. 5
6 rpm 64-Sector (342211) 2. Make sure that the rotor is equipped with the correct overspeed disk. If the disk is missing or damaged, replace it according to the instructions in Rotors and Tubes. 3. Check the chemical compatibilities of all materials used (refer to Appendix A in Rotors and Tubes). 4. Verify that the tubes and bottles being used are listed in Table 2 or Table 3. ROTOR PREPARATION For runs at other than room temperature, refrigerate or warm the rotor beforehand for fast equilibration. Hanger Place the rotor on the rotor stand (332400) when it is not in the centrifuge. Cap O-ring (SW 28) (SW 28.1) 1. Load the filled containers into the buckets (see page 8 for tube information). Complete loading by placing the correct floating spacers (if required) over the tubes. Bucket 2. Ensure that bucket O-rings are lightly but evenly coated with silicone vacuum grease (335148). Do not run a bucket without an O-ring, as the bucket will leak. 3. Be sure that metal threads in the bucket caps are clean and lightly but evenly lubricated with Spinkote lubricant (306812). Match bucket caps with numbered buckets and screw them down manually until tight. 4. Hook the buckets to the rotor by inserting the bucket pins into the grooves on the rotor body. Swing each bucket back and forth slightly to ensure proper installation; the buckets should move freely. Six buckets must be installed, whether loaded or empty. If fewer than six tubes are being run, they must be arranged symmetrically in the rotor (see Figure 1). Opposing tubes must be filled to the same level with liquid of the same density. 6
7 Figure 1. Arranging Tubes Symmetrically in the Rotor Buckets. Two, three, four, or six tubes can be centrifuged per run if they are arranged in the rotor as shown. All buckets must be attached to the rotor, whether loaded or empty. OPERATION For low-temperature runs, precool the rotor in the centrifuge or in a refrigerator before use especially before short runs to ensure that the rotor reaches the set temperature. A suggested precooling cycle is a minimum of 30 minutes at 2000 rpm at the required temperature. 1. To install the rotor, carefully lift it up off the rotor stand with both hands do not lift the rotor by the adapter and place it on the drive hub. Slowly turn the rotor to the right (clockwise) to make sure that the rotor is seated properly on the hub.! CAUTION Remove the zonal support band from the ultracentrifuges so equipped before operating these rotors. 2. Refer to the centrifuge instruction manual for additional operating instructions. NOTE Some gradients may degrade when run time exceeds 8 hours. 7
8 For additional information, see the following: RUN TIMES, page 12, for using k factors to adjust run durations, RUN SPEEDS, page 13, for information about speed limitations, and SELECTING CsCl GRADIENTS, page 15, for methods to avoid CsCl precipitation during centrifugation. REMOVAL AND SAMPLE RECOVERY! CAUTION If disassembly reveals evidence of leakage, you should assume that some fluid escaped the rotor. Apply appropriate decontamination procedures to the centrifuge and accessories as required. 1. Remove the rotor from the centrifuge by lifting the rotor straight up and off the drive hub. 2. Set the rotor on the rotor stand and carefully remove the buckets. While pressing the rubber tip against the adapter wall, pull the tool and adapter up and out of the cavity. Extractor Tool (354468) 3. Remove the bucket caps and use the appropriate removal tool (listed in the SUPPLY LIST) to remove the spacers and tubes. If floating spacers were used, remove them with the threaded end of the floating spacer removal tool (338765). NOTE If the conical-shaped adapters that support konical tubes are difficult to remove after centrifugation, an extractor tool (354468) is available to facilitate removal. TUBES AND ACCESSORIES The SW 28 rotor uses tubes and accessories listed in Table 2; the SW 28.1 rotor uses tubes and accessories listed in Table 3. Be sure to use only those items listed, and to observe the maximum speed limits shown. Refer to Appendix A in Rotors and Tubes for information on the chemical resistances of tube and accessory materials. 8
9 Table 2. Beckman Coulter Tubes and Accessories for the SW 28 Rotor. Use only the items listed here. Dimensions and Volume Tube Description Required Accessory Part Number Description Part Number Max Speed/ RCF/ k Factor mm 38.5 ml Ultra-Clear open-top none rpm g mm 38.5 ml polyallomer open-top none rpm g mm 32 ml thickwall polyallomer open-top (pkg/25) none rpm g mm 32 ml thickwall polycarbonate open-top (pkg/25) none rpm g mm 33 ml polyallomer Quick-Seal, bell-top Noryl* floating spacer rpm g mm 30 ml konical polyallomer open-top adapter (pkg/6) rpm g mm 25 ml konical polyallomer open-top adapter (pkg/6) rpm g mm 28 ml konical polyallomer Quick-Seal, bell-top adapter Noryl floating spacer rpm g mm 27 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g mm 15 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g mm 8.5 ml konical polyallomer Quick-Seal, bell-top adapter (pkg/6) Noryl floating spacer rpm g mm 23 ml konical polyallomer Quick-Seal, bell-top adapter Noryl floating spacer rpm g 178 *Noryl is a registered trademark of GE Plastics. 9
10 Table 3. Beckman Coulter Tubes and Accessories for the SW 28.1 Rotor. Use only the items listed here. Dimensions and Volume Tube Description Required Accessory Part Number Description Part Number Max Speed/ RCF/ k Factor mm 18 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g mm 17 ml mm 17 ml Ultra-Clear, open-top polyallomer, open-top none rpm g 275 none rpm g mm 14.5 ml konical polyallomer, open-top adapter rpm g mm 12.5 ml konical polyallomer Quick-Seal, bell-top adapter Noryl floating spacer rpm g mm 10 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g mm 8 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g mm 6.3 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g mm 4.2 ml polyallomer Quick-Seal, bell-top Noryl floating spacer rpm g 63 Temperature Limit 25 C 2 C Plastic tubes and bottles have been tested for use at temperatures between 2 and 25 C. For centrifugation at other temperatures, pretest tubes under anticipated run conditions. If plastic containers are frozen before use, make sure that they are thawed to 2 C or warmer prior to centrifugation. 10
11 Quick-Seal Tubes g-max Spacer Quick-Seal tubes must be sealed prior to centrifugation. These tubes are heat sealed and do not need caps; however, spacers are required on top of the tubes when they are loaded into the rotor buckets. Fill Quick-Seal tubes leaving a small bubble of air at the base of the neck. Do not leave large air space too much air can cause excessive tube deformation. Bell-top Tube Refer to Rotors and Tubes for detailed information on the use and care of Quick-Seal tubes. Some of the tubes listed in Tables 2 and 3 are part of the g-max system, which uses a combination of small bell-top Quick-Seal tubes and floating spacers (also called g-max spacers). This means that you can run the shorter tubes listed in the Tables in the SW 28 and SW 28.1 rotors without reduction in g force. Additional information about the g-max system is available in publication DS-709. konical Tubes Adapters Polyallomer konical tubes, used to optimize pelleting separations, have a conical tip that concentrates the pellet in the narrow end of the tube. The narrow bottom also reduces the tube s nominal volume and minimizes gradient material requirement. The konical tubes come in both open-top and Quick-Seal tube designs. Conical cavity adapters hold the tubes in the rotor buckets. Polyallomer and Ultra-Clear Open-Top Tubes Polyallomer and Ultra-Clear open-top tubes should be filled to 2 or 3 mm from the tube top for tube support. If necessary, float mineral oil (or some other low-density, immiscible liquid) on top of the tube contents to fill the tube to its maximum volume. (Do not use an oil overlay in Ultra-Clear tubes.) All opposing tubes for a run must be filled to the same level with liquid of the same density. 11
12 RUN TIMES TIME HR:MIN 03:30 The k factor of the rotor is a measure of the rotor s pelleting efficiency. (Beckman Coulter has calculated the k factors for all of its rotors at maximum speed and using full tubes.) The k factor is calculated from the formula: k = ln( r max r min ) ω (1) where ω is the angular velocity of the rotor in radians per second (ω = rpm), r max is the maximum radius, and r min is the minimum radius. After substitution: k = ( ) ln( r max rmin ) rpm 2 (2) Use the k factor in the following equation to estimate the run time t (in hours) required to pellet particles of known sedimentation coefficient s (in Svedberg units, S). t = k - s (3) Run times can be estimated for centrifugation at less than maximum speed by adjusting the k factor as follows: k adj = k actual run speed (4) Run times can also be estimated from data established in prior experiments if the k factor of the previous rotor is known. For any two rotors, a and b: t a k a = t b k b (5) where the k factors have been adjusted for the actual run speed used. 12
13 For more information on k factors, see Use of k Factor for Estimating Run Times from Previously Established Run Conditions (publication DS-719). RUN SPEEDS SPEED RPM/RCF RPM The centrifugal force at a given radius in a rotor is a function of speed. Comparisons of forces between different rotors are made by comparing the rotors relative centrifugal fields (RCF). When rotational speed is selected so that identical samples are subjected to the same RCF in two different rotors, the samples are subjected to the same force. The RCF at a number of rotor speeds is provided in Table 4. Do not select run speeds in excess of rpm. In addition, speeds must be reduced under the following circumstances: 1. If nonprecipitating solutions more dense than 1.2 g/ml are centrifuged, reduce the maximum allowable run speed according to the following equation: reduced maximum speed = ( rpm) 1.2 g/ml ρ (6) where ρ is the density of the tube contents. This speed reduction will protect the rotor from excessive stresses due to the added tube load. 2. Further speed limits must be imposed when CsCl or other selfforming-gradient salts are centrifuged, as equation (6) does not predict concentration limits/speeds that are required to prevent precipitation of salt crystals. Solid CsCl has a density of 4 g/ml, and if precipitated during centrifugation may cause rotor failure. Figures 2 through 5, together with the description and examples below, show how to reduce run speeds when using CsCl gradients. 13
14 Table 4. Relative Centrifugal Fields for the SW 28 and SW 28.1 Rotors. Entries in this table are calculated from the formula RCF = 1.12r (RPM/1000) 2 and are then rounded to three significant digits. SW 28 Rotor SW 28.1 Rotor Rotor Speed (rpm) Relative Centrifugal Field ( g) At r max (161 mm) At r av (118.2 mm) At r min (75.3 mm) Rotor Spee d (rpm) Relative Centrifugal Field ( g) At r max (171.3 mm) At r av (122.1 mm) At r min (72.9 mm) Relative Centrifugal Fields, SW 28 Rotor RCF (x g) r max r av r min Speed (rpm) Relative Centrifugal Fields, SW 28.1 Rotor RCF (x g) r max r av r min Speed (rpm) 14
15 SELECTING CsCl GRADIENTS Rotor speed is used to control the slope of a CsCl density gradient, and must be limited to prevent CsCl precipitation during centrifugation. Speed and density combinations that intersect on or below the curves in Figure 2 (for the SW 28 rotor) and in Figure 4 (for the SW 28.1 rotor) ensure that CsCl will not precipitate during centrifugation in these rotors. Curves are provided at two temperatures: 20 C (black curves) and 4 C (gray curves). Curves in Figures 2 through 5 are provided up to the maximum speed of the rotor. NOTE The curves in Figures 2 through 5 are for solutions of CsCl salt dissolved in distilled water only. If other salts are present in significant concentrations, the overall CsCl concentration may need to be reduced. The reference curves shown in Figures 3 and 5 show gradient distribution at equilibrium. Each curve in Figure 3 is within the density limits allowed for the SW 28 rotor; each curve in Figure 5 is within the density limits allowed for the SW 28.1 rotor. Each curve was generated for a single run speed using the maximum allowable homogeneous CsCl densities (one for each fill level) that avoid precipitation at that speed. (The gradients in Figures 3 and 5 can be generated from step or linear gradients, or from homogeneous solutions. But the total amount of CsCl in solution must be equivalent to a homogeneous solution corresponding to the concentrations specified in Figures 3 and 5.) Figures 3 and 5 can also be used to approximate the banding positions of sample particles. Curves not shown may be interpolated. ADJUSTING FILL VOLUMES Buffer Gradient Figures 2 through 5 show that several fill volumes are possible in a tube. If a thinwall tube is partially filled with gradient solution, float mineral oil (or some other low-density, immiscible liquid) on top of the tube contents to fill the tube to its maximum volume. (Do not use an oil overlay in Ultra-Clear tubes.) Note that for a given CsCl density, as the fill level decreases the maximum allowable speed increases. Partial filling may be desirable when there is little sample or when you wish to shorten the run time. 15
16 1.90 full 3/4 1/2 1/4 1/ /4 1/2 3/4 1/2 full 3/4 full 1.70 Homogeneous CsCl Solution (g/ml) SW 28 ROTOR = 20 C = 4 C Rotor Speed (K rpm) Figure 2. Precipitation Curves for the SW 28 Rotor. Using speed and density combinations that intersect on or below the solid curves ensures that CsCl will not precipitate during centrifugation. Tube fill volumes are indicated on the curves. The dashed lines are a representation of equation (6) and are shown here to illustrate the inability of that equation to prevent CsCl precipitation. 16
17 rpm rpm Density (g/ml) rpm rpm rpm rpm 1.2 SW 28 ROTOR rpm rpm 3/4 filled tube 1/2 filled tube /4 filled tube Each square on the grid represents 1.26 mm by 0.01 g/ml. = 20 C = 4 C r min r max Distance from Axis of Rotation (mm) Figure 3. CsCl Gradients at Equilibrium for the SW 28 Rotor. Centrifugation of homogeneous CsCl solutions at maximum allowable speeds (from Figure 2) results in gradients presented here. 17
18 1.90 1/4 full 3/4 1/2 1/4 1/ /4 1/2 3/4 full 3/4 full 1.70 Homogeneous CsCl Solution (g/ml) SW 28.1 ROTOR = 20 C = 4 C Rotor Speed (K rpm) Figure 4. Precipitation Curves for the SW 28.1 Rotor. Using speed and density combinations that intersect on or below the solid curves ensures that CsCl will not precipitate during centrifugation. Tube fill volumes are indicated on the curves. The dashed line are a representation of equation (6) and are shown here to illustrate the inability of that equation to prevent CsCl precipitation. 18
19 rpm rpm Density (g/ml) rpm rpm rpm rpm rpm rpm 3/4 filled tube 1/2 filled tube 1/4 filled tube SW 28.1 ROTOR Each square on the grid represents 1.26 mm by 0.01 g/ml. = 20 C = 4 C r min Distance from Axis of Rotation (mm) r max Figure 5. CsCl Gradients at Equilibrium for the SW 28.1 Rotor. Centrifugation of homogeneous CsCl solutions at maximum allowable speeds (from Figure 4) results in gradients presented here. 19
20 For example, in the SW 28 rotor, a quarter-filled tube of 1.67-g/mL homogeneous CsCl solution at 4 C may be centrifuged at rpm (see Figure 2). The segment of the rpm curve (Figure 3) from the quarter-filled line to r max (the tube bottom) represents this gradient. The same solution in a half-filled tube may be centrifuged no faster than rpm, and rpm in a three-quarter-filled tube. A tube full of the 1.67-g/mL CsCl solution may be centrifuged no faster than rpm. Curves not shown in the figures may be interpolated. TYPICAL EXAMPLES FOR DETERMINING RUN PARAMETERS Floating Components At Speed Bands Pelleted Material Example A: Starting with a homogeneous CsCl solution density of 1.33 g/ml and approximate particle buoyant densities of 1.30 and 1.35 g/ml, at 20 C, where will particles band at equilibrium in the SW 28 rotor? r min r max 1. In Figure 2, find the curve that corresponds to the required run temperature (20 C) and fill volume (one-half full). The maximum allowable rotor speed is determined from the point where this curve intersects the homogeneous CsCl density ( rpm). At Rest in Rotor Pathlength 2. In Figure 3, sketch a horizontal line corresponding to each particle s buoyant density. 3. Mark the point in Figure 3 where each particle density intersects the curve corresponding to the selected run speed and temperature. 4. Particles will band at these locations across the tube diameter at equilibrium during centrifugation. At Rest Outside Rotor In this example, particles will band about 145 and 151 mm from the axis of rotation, about 6 mm of centerband-to-centerband separation. To determine interband volume in milliliters, use the following equation: V = π r 2 h (7) where r is the tube radius in centimeters and h is the interband separation in centimeters. 20
21 Example B: Knowing particle buoyant densities (for example, 1.55 and 1.50 g/ml), how do you achieve good separation in the SW 28 rotor. 1. In Figure 3, sketch in a horizontal line corresponding to each particle s buoyant density. 2. Select the curve at the desired temperature (4 C) and tube volume (full) that gives the best particle separation. 3. Note the run speed along the selected curve ( rpm). 4. From Figure 2, select the maximum homogeneous CsCl density (in this case, 1.56 g/ml) that corresponds to the temperature and run speed established above. These parameters will provide the particle-banding pattern selected in Step 2. In this example, particles will band about 110 and 122 mm from the axis of rotation (about 12 mm apart). USE OF A CsCl CUSHION Some separations incorporate the use of cushions of CsCl. A common example is the isolation of total RNA. In this example, one-fourth of the total tube volume is filled with a cushion of 5.7 M CsCl (1.71 g/ml). A solution containing a tissue homogenate in a guanidinium thiocynate buffer is layered over the CsCl solution. Maximum run speeds must take into account the increased density of the solution as well as the use of the CsCl precipitation curves. NOTE Run speeds obtained using average densities are approximate. Example C (using SW 28 Rotor): 1. Parameters Cushion: 1.71 g/ml density CsCl (5.7 M) Cushion volume: 9.6 ml ( 1 /4 total volume) Overlay: 1.2 g/ml density homogenate/buffer Overlay Volume: 28.9 ml ( 3 /4 total volume) Average Density: 1.33 g/ml Temperature: 20 C 21
22 2. First, use the square root deration formula: RPM = g/ml = (0.95) = g/ml (8) 3. Next, use the CsCl curves for quarter-filled tubes (at 20 C) in Figure 3 to determine that the maximum run speed for the SW 28 with a quarter volume of 1.71 g/ml CsCl is rpm. 4. Choosing the lower of the two speeds gives a maximum run speed of rpm. Example D: 1. All parameters are as listed in Example C with the exception of temperature. Temperature: 4 C 2. As in Example C, the square root deration curve gives a maximum speed of rpm. 3. The CsCl curves for quarter-filled tubes at 4 C in Figure 3 show that the maximum run speed for the SW 28 with a quarter volume of 1.71 g/ml CsCl is rpm. 4. Choosing the lower of the two speeds gives a maximum run speed of rpm. CARE AND MAINTENANCE MAINTENANCE Bucket Cap O-ring (SW 28, ) (SW 28.1, ) NOTE Do not use sharp tools on the rotor that could cause scratches in the rotor surface. Corrosion begins in scratches and may open fissures in the rotor with continued use. Frequently check the bucket O-rings for signs of wear. Replace O-rings every 6 months, or whenever worn or damaged. Keep the O-rings lightly coated with silicone vacuum grease (335148). 22
23 Before every run, lubricate the bucket cap threads with a thin, even coat of Spinkote lubricant (306812). Refer to Appendix A of Rotors and Tubes for the chemical compatibility of rotor and accessory materials. Your Beckman Coulter representative provides contact with the Field Rotor Inspection Program and the rotor repair center. CLEANING Wash the rotor and rotor components immediately if salts or other corrosive materials are used or if spillage has occurred. Do not allow corrosive materials to dry on the rotor. Under normal use, wash the rotor frequently (at least weekly, or after each run if runs are infrequent) to prevent buildup of residues. Rotor Cleaning Kit (339558) 1. Wash the rotor buckets, O-rings, and caps in a mild detergent, such as Beckman Solution 555, that won t damage the rotor. The Rotor Cleaning Kit (339558) contains two plastic-coated brushes and two quarts of Solution 555 (339555) for use with rotors and accessories. Dilute the detergent 10 to 1 with water. 2. Wash the rotor body with a sponge or cloth dampened with a mild detergent, such as Solution 555, diluted 10 to 1 with water. 3. Rinse the cleaned rotor and components with distilled water. 4. Air-dry the rotor and buckets upside down. Do not use acetone to dry the rotor. Clean metal threads frequently to prevent buildup of residues and ensure adequate closure. Use a brush and concentrated Solution 555. Rinse and dry thoroughly, then lubricate lightly but evenly with Spinkote to coat all threads. 23
24 DECONTAMINATION If the rotor or other components become contaminated with toxic, radioactive, or pathogenic substances, follow appropriate decontamination procedures as outlined by your laboratory safety officer. Refer to the chemical resistances list in Appendix A of Rotors and Tubes to select solutions that will not damage the rotor and accessory materials. STERILIZATION AND DISINFECTION 121 C The rotor and all rotor components can be autoclaved at 121 C for up to one hour. Remove the caps from the rotor buckets and place the rotor, caps, and spacers in the autoclave upside down. Ethanol (70%) 2 or hydrogen peroxide (6%) may be used on all rotor components, including those made of plastic. Bleach (sodium hypochlorite) may be used, but may cause discoloration of anodized surfaces. Use the minimum immersion time for each solution, per laboratory standards. While Beckman Coulter has tested these methods and found that they do not damage components, no guarantee of sterility or disinfection is expressed or implied. When sterilization or disinfection is a concern, consult your laboratory safety officer. Refer to publication IN-192 (included with each box of tubes) for tube sterilization and disinfection procedures. Quick-Seal, Ultra Clear and thinwall open-top tubes are disposable and should be discarded after a single use. STORAGE When it is not in use, store the rotor and buckets in a dry environment (not in the instrument). Remove the bucket caps to allow air circulation so that moisture will not collect in the buckets. 2 Flammability hazard. Do not use in or near operating centrifuges. 24
25 RETURNING A ROTOR RGA Before returning a rotor or accessory for any reason, prior permission (a Returned Goods Authorization form) must be obtained from Beckman Coulter, Inc. This RGA form may be obtained from your local Beckman Coulter sales office, and should contain the following information: serial number of rotor, history of use (approximate frequency of use), reason for the return, original purchase order number, billing number, and shipping number, if possible, name and phone number of the person to be notified upon receipt of the rotor or accessory at the factory, and name and phone number of the person to be notified about repair costs, etc. To protect our personnel, it is the customer s responsibility to ensure that the parts are free from pathogens and/or radioactivity. Sterilization and decontamination must be done before returning the parts. Smaller items (such as tubes, bottles, etc.) should be enclosed in a sealed plastic bag. All parts must be accompanied by a note, plainly visible on the outside of the box or bag, stating that they are safe to handle and that they are not contaminated with pathogens or radioactivity. Failure to attach this notification will result in return or disposal of the items without review of the reported problem. Use the address label printed on the RGA form when mailing the rotor and/or accessories to: Beckman Coulter, Inc Page Mill Road Palo Alto, CA Attention: Returned Goods Customers located outside the United States should contact their local Beckman Coulter office. 25
26 SUPPLY LIST NOTE To receive copies of referenced publications, contact Beckman Coulter, Inc., Technical Publications Department, 1050 Page Mill Road, Palo Alto, CA 94304, U.S.A. (Telephone ; Fax ). Contact Beckman Coulter Sales ( in the United States; worldwide offices are listed on the back cover of this manual) or see the Beckman Coulter Ultracentrifuge Rotors, Tubes & Accessories catalog (BR-8101, available at for detailed information on ordering parts and supplies. For your convenience, a partial list is given below. REPLACEMENT ROTOR PARTS SW 28 rotor assembly SW 28 buckets (set of 6, with caps and O-rings) SW 28 bucket O-ring SW 28.1 rotor assembly SW 28.1 buckets (set of 6, with caps and O-rings) SW 28.1 bucket O-ring Rotor stand Overspeed disk ( rpm) OTHER Tubes and accessories see Tables 2 and 3 Bucket holder rack Quick-Seal Cordless Tube Topper Kit, 60 Hz Quick-Seal Cordless Tube Topper Kit, 50 Hz (Europe) Quick-Seal Cordless Tube Topper Kit, 50 Hz (Great Britain) Quick-Seal Cordless Tube Topper Kit, 50 Hz (Australia) Quick-Seal Cordless Tube Topper Kit, 50 Hz (Canada) Tube racks for the Tube Topper for 16-mm diameter tubes for 38-mm diameter tubes Floating spacer removal tool Tube removal tool (Quick-Seal tubes) Extractor tool (konical tube adapters) Silicone vacuum grease (1 oz) Spinkote lubricant (2 oz) Rotor Cleaning Kit Beckman Solution 555 (1 qt) Rotor cleaning brush
27 ULTRACENTRIFUGE ROTOR WARRANTY All Beckman Coulter ultracentrifuge Fixed Angle, Vertical Tube, Near Vertical Tube, Swinging Bucket, and Airfuge rotors are warranted against defects in materials or workmanship for the time periods indicated below, subject to the Warranty Conditions stated below. Preparative Ultracentrifuge Rotors years No Proration Analytical Ultracentrifuge Rotors years No Proration ML and TL Series Ultracentrifuge Rotors years No Proration Airfuge Ultracentrifuge Rotors year No Proration For Zonal, Continuous Flow, Component Test, and Rock Core ultracentrifuge rotors, see separate warranty. Warranty Conditions (as applicable) 1) This warranty is valid for the time periods indicated above from the date of shipment to the original Buyer by Beckman Coulter or an authorized Beckman Coulter representative. 2) This warranty extends only to the original Buyer and may not be assigned or extended to a third person without written consent of Beckman Coulter. 3) This warranty covers the Beckman Coulter Centrifuge Systems only (including but not limited to the centrifuge, rotor, and accessories) and Beckman Coulter shall not be liable for damage to or loss of the user s sample, non-beckman Coulter tubes, adapters, or other rotor contents. 4) This warranty is void if the Beckman Coulter Centrifuge System is determined by Beckman Coulter to have been operated or maintained in a manner contrary to the instructions in the operator s manual(s) for the Beckman Coulter Centrifuge System components in use. This includes but is not limited to operator misuse, abuse, or negligence regarding indicated maintenance procedures, centrifuge and rotor classification requirements, proper speed reduction for the high density of certain fluids, tubes, and tube caps, speed reduction for precipitating gradient materials, and speed reduction for high-temperature operation. 5) Rotor bucket sets purchased concurrently with or subsequent to the purchase of a Swinging Bucket Rotor are warranted only for a term co-extensive with that of the rotor for which the bucket sets are purchased. 6) This warranty does not cover the failure of a Beckman Coulter rotor in a centrifuge not of Beckman Coulter manufacture, or if the rotor is used in a Beckman Coulter centrifuge that has been modified without the written permission of Beckman Coulter, or is used with carriers, buckets, belts, or other devices not of Beckman Coulter manufacture. 7) Rotor parts subject to wear, including but not limited to rotor O-rings, VTi, NVT, TLV, MLN, and TLN rotor tube cavity plugs and gaskets, tubing, tools, optical overspeed disks, bearings, seals, and lubrication are excluded from this warranty and should be frequently inspected and replaced if they become worn or damaged. 8) Keeping a rotor log is not mandatory, but may be desirable for maintenance of good laboratory practices. Repair and Replacement Policies 1) If a Beckman Coulter rotor is determined by Beckman Coulter to be defective, Beckman Coulter will repair or replace it, subject to the Warranty Conditions. A replacement rotor will be warranted for the time remaining on the original rotor s warranty. 2) If a Beckman Coulter centrifuge is damaged due to a failure of a rotor covered by this warranty, Beckman Coulter will supply free of charge (i) all centrifuge parts required for repair (except the drive unit, which will be replaced at the then current price less a credit determined by the total number of revolutions or years completed, provided that such a unit was manufactured or rebuilt by Beckman Coulter), and (ii) if the centrifuge is currently covered by a Beckman Coulter warranty or Full Service Agreement, all labor necessary for repair of the centrifuge. 3) If a Beckman Coulter rotor covered by this warranty is damaged due to a malfunction of a Beckman Coulter ultracentrifuge covered by an Ultracentrifuge System Service Agreement, Beckman Coulter will repair or replace the rotor free of charge. 4) If a Beckman Coulter rotor covered by this warranty is damaged due to a failure of a Beckman Coulter tube, bottle, tube cap, spacer, or adapter, covered under the Conditions of this Warranty, Beckman Coulter will repair or replace the rotor and repair the instrument as per the conditions in policy point (2) above, and the replacement policy. 5) Damage to a Beckman Coulter rotor or instrument due to the failure or malfunction of a non-beckman Coulter tube, bottle, tube cap, spacer, or adapter is not covered under this warranty, although Beckman Coulter will assist in seeking compensation under the manufacturer s warranty. Disclaimer IT IS EXPRESSLY AGREED THAT THE ABOVE WARRANTY SHALL BE IN LIEU OF ALL WARRANTIES OF FITNESS AND OF THE WARRANTY OF MERCHANTABILITY AND BECKMAN COULTER, INC. SHALL HAVE NO LIABILITY FOR SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY KIND WHATSOEVER ARISING OUT OF THE MANUFAC- TURE, USE, SALE, HANDLING, REPAIR, MAINTENANCE, OR REPLACEMENT OF THE PRODUCT. Factory Rotor Inspection Service Beckman Coulter, Inc., will provide free mechanical and metallurgical inspection in Palo Alto, California, USA, of any Beckman Coulter rotor at the request of the user. (Shipping charges to Beckman Coulter are the responsibility of the user.) Rotors will be inspected in the user s laboratory if the centrifuge in which they are used is covered by an appropriate Beckman Coulter Service Agreement. Contact your local Beckman Coulter office for details of service coverage or cost. Before shipping, contact the nearest Beckman Coulter Sales and Service office and request a Returned Goods Authorization (RGA) form and packaging instructions. Please include the complete rotor assembly, with buckets, lid, handle, tube cavity caps, etc. A SIGNED STATEMENT THAT THE ROTOR AND ACCESSO- RIES ARE NON-RADIOACTIVE, NON-PATHOGENIC, NON- TOXIC, AND OTHERWISE SAFE TO SHIP AND HANDLE IS REQUIRED.
28 Beckman Coulter Worldwide Life Science Research Division Offices AUSTRALIA Beckman Coulter Australia Pty Ltd Unit D, 24 College St. Gladesville, NSW 2111 Australia Telephone: (61) or toll free: Fax: (61) [email protected] CANADA Beckman Coulter (Canada) Inc Mississauga Road, Suite 600 Mississauga, Ontario Canada L5N 7Y2 Telephone: (905) Fax: (905) CHINA Beckman Coulter Inc. Beijing Representative Office Unit 2005A, , East Ocean Center Jian Guomenwai Avenue Beijing China Telephone: (86) Fax: (86) , EASTERN EUROPE/ MIDDLE EAST/NORTH AFRICA Beckman Coulter International S.A. 22, Rue Juste-Olivier Case Postale CH-1260 Nyon, Switzerland Telephone: (41) Fax: (41) FRANCE Beckman Coulter France S.A. Paris Nord II, 33/66 rue des Vanesses B.P Villepinte, France ROISSY CDG Cedex Telephone: Fax: [email protected] GERMANY Beckman Coulter GmbH Siemensstrasse 1 D Unterschleissheim-Lohhof Germany Telephone: (89) Fax: (89) [email protected] TM HONG KONG Beckman Coulter Hong Kong Ltd. 12th Floor, Oxford House 979 King s Road Taikoo Place, Hong Kong Telephone: (852) , Fax: (852) , ITALY Beckman Coulter S.p.a. Centro Direzionale Lombardo Palazzo F/1, Via Roma Cassina de Pecchi Milano, Italy Telephone: Fax: JAPAN Beckman Coulter K.K. Toranomon 37 Mori Bldg , Toranomon Minato-ku, Tokyo Japan Telephone: Fax: MEXICO Beckman Coulter de Mexico S.A. DE C.V. Avenida Popocatépetl #396 Colonia Gral. Pedro Maria Anaya Codigo Postal Maxico, D.F. / Mexico City Telephone: Fax: NETHERLANDS Beckman Coulter Nederland B.V. Nijverheidsweg RP-Mijdrecht Postbus AA Mijdrecht The Netherlands Telephone: Fax: SINGAPORE Beckman Coulter Singapore Pte. Ltd. 116 Changi Road Unit #03-01/02 Singapore Telephone: (65) Fax: (65) SOUTH AFRICA/SUB-SAHARAN AFRICA Beckman Coulter Stand 1A Primegro Park Tonetti Street 1685 Halfway House Johannesburg Republic of South Africa Telephone: (27) /5 Fax: (27) [email protected] SPAIN Beckman Coulter España S.A. C/ Caleruega, Madrid, Spain Telephone: (34) Fax: (34) [email protected] SWEDEN Beckman Coulter AB Archimedesvaegen 7 Box SE Bromma Sweden Telephone: (0) Telefax: (0) SWITZERLAND Beckman Coulter International S.A. 22, Rue Juste-Olivier Case Postale CH-1260 Nyon Switzerland Telephone: Fax: TAIWAN Beckman Coulter Taiwan Inc. Taiwan Branch 8th Floor 216 Tun Hwa South Road, Section 2 Taipei 106, Taiwan Republic of China Telephone: (886) Fax: (886) TURKEY Beckman Coulter Ltd. E-5 Yanyol Faith Cad Soganlik Kartal Istanbul Turkey Telephone: Fax: UNITED KINGDOM Beckman Coulter United Kingdom Ltd Oakley Court Kingsmead Business Park London Road High Wycombe Buckinghamshire HP11 1JU England, U.K. Telephone: Fax: [email protected] Authorized dealers in other countries. Beckman Coulter, Inc N. Harbor Boulevard, Box 3100 Fullerton, California Sales: Service: Internet: Telex: Fax: Printed on recycled paper 2002 Beckman Coulter, Inc. I.S. EN ISO 9001
SW 32 Ti and SW 32.1 Ti Rotors
LXL-TB-015B January 2004 TM SW 32 Ti and SW 32.1 Ti Rotors Used in Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2004 Beckman Coulter, Inc. Printed in the U.S.A. ! SAFETY NOTICE This safety
L5-TB-061P March 2003. Type 70.1 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges. 2003 Beckman Coulter, Inc.
L5-TB-061P March 2003 Type 70.1 Ti Rotor Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2003 Beckman Coulter, Inc. ! SAFETY NOTICE This safety notice summarizes information basic
INSTRUCTIONS FOR USING THE A-110 ROTOR In the Beckman Coulter Airfuge Ultracentrifuge
AF-TB-029BA April 2014 INSTRUCTIONS FOR USING THE A-110 ROTOR In the Beckman Coulter Airfuge Ultracentrifuge SPECIFICATIONS 18 Maximum speed (±5000 rpm)...................... 110 000 rpm Air pressure at
Type 90 Ti Rotor. Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges. LXL-TB-005M August 2007
LXL-TB-005M August 2007 Type 90 Ti Rotor Used In Beckman Coulter Class H, R, and S Preparative Ultracentrifuges 2007 Beckman Coulter, Inc. Printed in the U.S.A. !!!!!!!! SAFETY NOTICE This safety notice
Care and Handling of Rotors
Care and Handling of Rotors Today s Objectives Help you to use centrifuge systems safely Ensure maximum life of your investment Avoid unbudgeted lab expenses Maximize experiment integrity Table of Contents
BR-8101K. Rotors, Tubes & Accessories ULTRACENTRIFUGE
BR-80K Rotors, Tubes & Accessories ULTRACENTRIFUGE Since the introduction of the first commercial ultracentrifuge in 949 the classic Beckman Model L Beckman Coulter has been at the forefront of centrifuge
Thermo Scientific Rotor Care Guide. Protect your rotor investment
Thermo Scientific Rotor Care Guide Protect your rotor investment Thermo Scientific Rotor Care Guide Thermo Scientific rotors meet the high standards we set for all of our laboratory products. The precision-engineered
Using k-factor to Compare Rotor Efficiency
Using k-factor to Compare Rotor Efficiency Most researchers refer to the maximum speed of a rotor to compare efficiency. This can be misleading, however, since other factors such as the geometry of the
Thermo Scientific Rotor Care Guide. Protect your rotor investment
Thermo Scientific Rotor Care Guide Protect your rotor investment pg 2 pg 4 pg 6 pg 8 Routine Evaluation and Care of Your Rotor Rotor Maintenance Rotor Inspection Services Rotor Life Management Thermo Scientific
ROTORS AND TUBES. For Beckman Coulter Preparative Ultracentrifuges. User s Manual
LR-IM-20 TM ROTORS AND TUBES For Beckman Coulter Preparative Ultracentrifuges User s Manual Published by the Centrifuge Instrument Systems Development Center of Beckman Coulter, Inc., Palo Alto, California
ROTORS, TUBES, BOTTLES AND CAPS FOR Hitachi Preparative Ultracentrifuge
INSTRUCTION MANU ROTORS, TUBES, BOTTLES AND CAPS FOR Hitachi Preparative Ultracentrifuge This manual contains instruction for operation of Angle rotor, Neo-angle rotor, Swing rotor and Vertical rotor.
ROTOR LOADER OWNER S MANUAL
ROTOR LOADER OWNER S MANUAL ROTOR LOADER OWNER S MANUAL WARNING IMPORTANT SAFETY INSTRUCTIONS AND GUIDELINES. Misuse of paintball equipment may cause serious injury or death. QUICK SET-UP GUIDE BATTERY
Featuring fast rotor exchanges with AUTO-LOCK
Thermo Scientific Sorvall T1 Centrifuge Series Versatility, convenience and value for routine applications Featuring fast rotor exchanges with AUTO-LOCK Versatility, convenience and value for routine applications
Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.
Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam
CappController Pipette Controller. Operation Manual Version: 1.2. www.capp.dk
CappController Pipette Controller Operation Manual Version: 1.2 www.capp.dk skal vises her. Table of Content 1. Safety Precautions... 1 2. General Description... 2 3. Getting Started... 4 3.1. Recharging
Autoclave Safety. Autoclaves are sterilizers using high pressure and high temperature steam. The potential safety risks for the operators are:
Autoclave Safety Purpose: Sterilization refers to the complete killing of all living organisms, including spores. Common sterilization techniques include the application of wet heat, dry heat, chemicals,
Industrial Robotics Solutions
Industrial Robotics Solutions Complete Solutions for Discovery Research Genomics Proteomics Cell Analysis Centrifugation Lab Tools Particle Characterization Bioseparation Lab Automation BR-10552A The Anatomy
Destiny Shower Enclosure Installation Instructions
Destiny Shower Enclosure Installation Instructions IMPORTANT Please read carefully the following instructions before installing your shower enclosure. If you have any questions on this shower enclosure
ATS Overhead Table Shelf System INSTRUCTION MANUAL
ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System Instruction Manual Warranty Newport Corporation warrants this product to be free of defects in material and workmanship
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support
Dehumidifier Users manual. For Models: DH45S DH65S
Dehumidifier Users manual For Models: DH45S DH65S 950-0062-revD Jan. 9 2007 FORWARD The appearance of the units that you purchase might be slightly different from the ones described in the Manual, but
81000 Series Aluminum Gate Valve Installation and Repair Manual
High Vacuum Valves Leader in Quality and Value 81000 Series Aluminum Gate Valve Installation and Repair Manual Dear Customer, Thank you for choosing HVA as your vacuum valve supplier. Your valve will give
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW
Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer
Standard laryngoscope blade assemblies Directions for use
1 2014 Welch Allyn, Inc. MM 721358 Ver. C Standard laryngoscope blade assemblies Directions for use Intended use About this document The rigid laryngoscope is a device used to examine and visualize a patient
3 WATT LED SPOTLIGHT Model No. SLM - 3801
3 WATT LED SPOTLIGHT Model No. SLM - 3801 OWNER'S MANUAL Customer Service Tel: 1-800-268-3319 Superex Canada Ltd, Toronto,M2H 3B8 Made in China Table of Contents A). Important Safety Instructions B). Charging
London Farm Kitchen Sinks
INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE London Farm Kitchen Sinks 24" Sink model number 35120 30" Sink model number 35130 36" Sink model number 35140 Certified to comply with ANSI A112.19.3.M CAN/C.S.A
EZ-PAGE Electrophoresis System USER MANUAL
EZ-PAGE Electrophoresis System USER MANUAL Table of Contents Safety Information.. 2 Product Description... 2 Product Contents..... 3 Specifications & Storage Conditions.. 3 Product Use..... 3 Getting Started
Retractable Screen. Replacement Screen Instructions
Retractable Screen Patent Pending Replacement Screen Instructions For RTS, RET, & RTM Products ATTENTION! Retractable screens are not intended to provide security or provide for the retention of objects,
PALLET JACK - 2.5 TON
PALLET JACK - 2.5 TON 39939 SET UP AND OPERATING INSTRUCTIONS Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result in serious injury.
Spectrafuge 24D Microcentrifuge
Spectrafuge 24D Microcentrifuge User Manual C2400 C2400-100V C2400-230V Lit M00013 Rev 2 July 2012 Safety Precautions NEVER use the centrifuge in any manner not specified in these instructions. NEVER operate
STANDARD OPERATING PROCEDURES Safe Autoclave Operations
Department of Environmental Health & Safety Biosafety & Biosecurity STANDARD OPERATING PROCEDURES Safe Autoclave Operations The purpose of this document is to provide standard operating procedures for
Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function
Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless
Biological Safety Program
Risk Management & Safety Main Office, Wyoming Hall Phone: (307) 766-3277 Fax: (307)766-6116 Regulated Materials Management Center Phone: (307)766-3696 Fax: (307)766-3699 Web: www.uwyo.edu/ehs Email: [email protected]
How To Use The Visifault Visual Fault Locator
VisiFault Visual Fault Locator Instruction Sheet The VisiFault Visual Fault Locator (VFL) is a visible light source that helps you trace optical fibers, check fiber continuity, and find faults such as
TWIN EAGLES ALL WEATHER ELECTRIC RADIANT HEATER
TWIN EAGLES, INC. Defining the Art of Outdoor Cooking TWIN EAGLES ALL WEATHER ELECTRIC RADIANT HEATER OWNER S MANUAL MODEL TEEH-1512 TEEH-2524 TEEH-3524 NEED ASSISTANCE? PLEASE CALL: (562) 802-3488 This
MODEL G300 BRAKE BLEEDER
MODEL G300 BRAKE BLEEDER Installation, Operation & Repair Parts Information Branick Industries, Inc. 4245 Main Avenue P.O. Box 1937 Fargo, North Dakota 58103 REV060616 P/N: 81-0035G 1 THIS PAGE INTENTIONALLY
Warranty. www.tissot.ch 1/5. Warranty
Warranty Warranty Your TISSOT watch is warranted by TISSOT S.A. for a period of twenty-four (24) months from the date of purchase under the terms and conditions of this warranty. The international TISSOT
Important Information
Single Door Quadrant Instruction Manual ESQ Important Information Toughened glass is completely safe for use in our shower enclosures and bath screens; providing our products are installed according to
Hydraulic Transmission Jacks Operating Instructions & Parts Manual
Blackhawk Automotive is a Licensed Trade Mark Made by SFA Companies, Kansas City, MO Hydraulic Transmission Jacks Operating Instructions & Parts Manual Model BH7011 BH7210 Capacity 1/2 Ton 1 Ton SFA Companies
Lenntech. 1000 psi End port pressure vessels 2.5. User s Guide to: Phoenix Vessel Technology Limited. Model number: 1503
1 User s Guide to: Phoenix Vessel Technology Limited 1000 psi End port pressure vessels 2.5. Model number: 1503 Lenntech [email protected] Tel. +31-152-610-900 www.lenntech.com Fax. +31-152-616-289 Phoenix
USER MANUAL. Bottom Loading Bottled Water Dispenser SAVE THIS MANUAL FOR FUTURE USE. Model # 900172
Model # 900172: Page 1 USER MANUAL Bottom Loading Bottled Water Dispenser Model # 900172 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING
Important Operating Instructions and Warranty Information On Your New Electronic AMSEC Safe
Important Operating Instructions and Warranty Information On Your New Electronic AMSEC Safe MODELS: ES712, ES914, ES149, ES1814, ES813, ES916, ES1014, ES2014, ES412, WES149, WES2114 Read Contents Carefully
Scotchgard Multi-Layer Protective Film 1004
Scotchgard Multi-Layer Protective Film 1004 3 Technical Bulletin September, 2008 Scotchgard Multi-Layer Protective Film 1004 Application Kit (3M Part No. 70-0710-4666-1) includes: (clockwise) Heavy Duty
The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.
i800 AC Current Clamp Instruction Sheet Introduction The i800 AC Current Clamp, the Clamp, has been designed for use with multimeters, recorders, power analyzers, safety testers, etc., for accurate non-intrusive
How To Service A Paper Shredder
Paper shredder 20-2HS 80C POWERSHRED OPERATING INSTRUCTIONS NOTICE D'UTILISATION MANUAL DEL PROPIETARIO BETRIEBSANLEITUNG 20 80 20C 80C 20-2HS Edition 0/0.70.999.0 Edition 0/0 20-2HS 80C Paper shredder
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W
MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries
Sealant. 30" Model No. 35020. D: 17-7/8" TO 18-3/8" (454 to 466mm) W: 29-3/4" TO 30-1/4" (756 to 768mm) Under Countertop Installation
INSTALLATION INSTRUCTIONS CARE AND MAINTENANCE London Kitchen Sinks 24" Sink model number 35030 30" Sink model number 35020 36" Sink model number 35040 Thank you for selecting Porcher. To ensure this product
Warranties All Products. Revised: 8/1/15 7 Pages
Warranties All Products Revised: 8/1/15 7 Pages GENERAL DISCLAIMER In addition to all claims listed on each of the following individual WARRANTY pages, the following GENERAL DISCLAIMERS apply. 1. The purchaser
Ozone Generator Manual
ENMET Corporation PO Box 979 Ann Arbor, MI 48106-0979 Ozone Generator Manual 80003-118 MCN-282, 05/14/04 Table of Contents 1.0 INTRODUCTION... 1 1.1 Unpack...1 1.2 Check Order...1 1.3 Serial Numbers...1
IWAKI MAGNETIC DRIVE PUMPS MDM. Patent & Pat.Pend. JAPAN / U.S.A. / EU / CHINA / TAIWAN
IWAKI MAGNETIC DRIVE PUMPS Patent & Pat.Pend. JAPAN / U.S.A. / EU / CHINA / TAIWAN Magnetic drive process pump resistant to dry run damage The Series of Magnetic drive process pumps have wetted parts made
Agilent P940xA/C Solid State PIN Diode Switches
Agilent P940xA/C Solid State PIN Diode Switches Operating and Service Manual Agilent Technologies Notices Agilent Technologies, Inc. 2007 No part of this manual may be reproduced in any form or by any
OAK WORKBENCH WITH 2 DRAWERS
OAK WORKBENCH WITH 2 DRAWERS Model 93991 ASSEMBLY Instructions Visit our website at: http://www.harborfreight.com Read this material before using this product. Failure to do so can result in serious injury.
Super Slider ADJUSTABLE FROM 3.5-12 Model #HD26229
Super Slider ADJUSTABLE FROM 3.5-12 Model #HD26229 IMPORTANT!!! ALL PERSONS USING THIS EQUIPMENT MUST READ AND UNDERSTAND ALL INSTRUCTIONS. FAILURE TO DO SO MAY RESULT IN SERIOUS INJURY OR DEATH. USERS
ECM-D70T / ECM-D70T1.5
Elliptical Ceiling Dual Mount for 37 to 50 Flat Panels INSTALLATION INSTRUCTIONS CREATING POSITIVE CUSTOMER EXPERIENCES 9531-041-001-01 Contents ECM-D70T / ECM-D70T1.5 Installation Tools... 3 Parts List...
Instructions for Use. A-dec 500 Assistant s Instrumentation. Model 551
Instructions for Use A-dec 500 Assistant s Instrumentation Model 551 A-DEC 500 A SSISTANT S INSTRUMENTATION (MODEL 551) I NSTRUCTIONS FOR USE Copyright 2010 A-dec Inc. All rights reserved. A-dec Inc.
7. Chemical Waste Disposal Procedures
7. Chemical Waste Disposal Procedures Substances that Can and Cannot be Disposed of via Sink Drains or in Regular Solid Refuse Containers 1. It is prohibited to pour down a sink or floor drain, or place
OPERATIONS AND MAINTENANCE INFORMATION FOR: THE CANPLAS BACKWATER VALVE THE CHEMTROL BALL SHUT OFF VALVE SUMP PUMPS AND WASTEWATER EJECTOR PUMPS
OPERATIONS AND MAINTENANCE INFORMATION FOR: THE CANPLAS BACKWATER VALVE THE CHEMTROL BALL SHUT OFF VALVE SUMP PUMPS AND WASTEWATER EJECTOR PUMPS THE LOUISVILLE AND JEFFERSON COUNTY METROPOLITAN SEWER DISTRICT
SINGLE RANGE DS METERS Operation Instructions
SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax
Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version
instrumentation for fluids Electromagnetic sensor Series FLOMID FX Instructions Manual! Conforms with the Pressure Equipment Directive 97/23/EC. 0830 This equipment is considered as being a pressure accessory
Use & Care. of your Iron. All about the TA B L E O F C O N T E N T S. USA and Canada 1-888-845-7330
All about the Use & Care of your Iron Important Safeguards... 2 Iron Safety Instructions... 3 Parts and Features...4 TA B L E O F C O N T E N T S Using Your Iron...6 Cleaning and Maintenance... 8 Warranty...9
For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).
User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,
Multi Tube Rotator. Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11
Multi Tube Rotator Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11 Table of Contents Safety Information...3 Alert Signals...3 Description...4 Specifications...5 Power Requirements...5 Shaker Motion...5
UltraClean PCR Clean-Up Kit
UltraClean PCR Clean-Up Kit Catalog No. Quantity 12500-50 50 Preps 12500-100 100 Preps 12500-250 250 Preps Instruction Manual Please recycle Version: 02212013 1 Table of Contents Introduction... 3 Protocol
LEAD CRYSTAL. User Manual. Valve-regulated lead-crystal batteries Energy storage Cells
Engineering Production Sales LEAD CRYSTAL Valve-regulated lead-crystal batteries Energy storage Cells User Manual www.axcom-battery-technology.de [email protected] Chapter 1: 1. Introduction
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
Laptop Power Adapter for home/office User Manual
Laptop Power Adapter for home/office User Manual Introduction Congratulations and thank you for purchasing the Laptop Power Adapter for home/office from Belkin. This new Adapter provides a single source
Deluxe Bottle Warmer
Deluxe Bottle Warmer IMPORTANT SAFEGUARDS This product is for household use only. When using electrical appliances, basic safety precautions should always be followed including the following: Read all
LM135-LM235-LM335. Precision temperature sensors. Features. Description
Precision temperature sensors Features Directly calibrated in K 1 C initial accuracy Operates from 450µA to 5mA Less than 1Ω dynamic impedance TO-92 (Plastic package) Description The LM135, LM235, LM335
MS-800B Mobile Video Studio
MS-800B Mobile Video Studio Instruction Manual http:// www.datavideo-tek.com Rev: 141206 Contents Warnings and Precautions ----------------------------------------------------------------------------------3
Scott & White Institutional Biosafety Committee Compliance Program Biohazardous Material Spill Clean-Up Procedure Policy #IBC.002
I. Purpose Biohazardous material usage on the Scott & White campus is regulated by the Scott & White Institutional Biosafety Committee (IBC). Those investigators choosing to perform research with biohazardous
Unpacking, Installation, and Customization
Unpacking, Installation, and Customization NetShelter WX Enclosure AR100 AR100HD Contents Unpacking... 1 How to Unpack the Enclosure............................... 1 Please Recycle..........................................
MOLECULAR DRAG PUMPS MDP SERIES
MOLECULAR DRAG PUMPS MDP SERIES MOLECULAR DRAG PUMPS MDP SERIES If you need: to replace a rotary vane pump with a dry pump for an application at 10-3 a turbopump that can operate up to 10 a pump that can
KGI MEDICAL WASTE MANAGEMENT PLAN
1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:
KR25i HIGH SPEED, HIGH CAPACITY REFRIGERATED CENTRIFUGE AUTO-LOCK. CQCS Traceability System VIDEOset Display!
KR25i HIGH SPEED, HIGH CAPACITY REFRIGERATED CENTRIFUGE AUTO-LOCK CQCS Traceability System VIDEOset Display! A CENTRIFUGE FOR ALL YOUR HIGH SPEED APPLICATIONS CLINICAL BIOLOGY, RESEARCH, INDUSTRY The KR25i
OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE
OSEH Occupational Safety & Environmental Health OCCUPATIONAL SAFETY AND ENVIRONMENTAL HEALTH GUIDELINE Subject: Biohazardous (Medical) Waste Disposal Date: 08/19/09 Revision: 03 Page: 1 of 7 TABLE OF Section
3. Maintenance of Laboratory Equipment
- 52 - Mainten: of Lab Equipment / National Guidelines SLCM SLCM National Guidelines / Mainte: of Lab Equipment - 53-3. Maintenance of Laboratory Equipment Introduction Equipment maintenance is one important
THE UNIVERSITY OF NEWCASTLE- SCHOOL of BIOMEDICAL SCIENCES
Page: 1 of 7 1. Purpose: 1.1. To describe the procedures to be used when dealing with chemical or microbiological spills. 2. Equipment: 2.1. Spill Kit 2.2. Miscellaneous items as listed 3. Materials: 3.1.
POSEIDON 2-29, 2-25 & 2-22 POSEIDON 2-29, 2-25 & 2-22 XT
POSEION 2-29, 2-25 & 2-22 POSEION 2-29, 2-25 & 2-22 XT Repair Manual Index A. Safety precautions 3 B. Technical data 4 C. Structure 5-6. Service / Repair 7-23 E. Tools 24 F. Function 25-26 G. Electric
Electric Fryers Instruction Manual Models: 8047D, 8048D, 8049D, 8050D, 8051D 8066, 8068, 8068FL, 8073, 8073BF and 8075
Part No. 89047 Electric Fryers Instruction Manual Models: 8047D, 8048D, 8049D, 8050D, 8051D 8066, 8068, 8068FL, 8073, 8073BF and 8075 Cincinnati, OH 45241-4807 USA ELECTRIC FRYER SAFETY PRECAUTIONS Installation
LUCCI AIRFUSION QUEST II CEILING FAN
LUCCI AIRFUSION QUEST II CEILING FAN WITH IR REMOTE INSTALLATION OPERATION MAINTENANCE WARRANTY INFORMATION CAUTION READ INSTRUCTIONS CAREFULLY FOR SAFE INSTALLATION AND FAN OPERATION. V1.0 QUEST II IR
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
M128 SERVICE MANUAL- M128 MODEL #S-11 MANUAL CAN OPENER
SERVICE MANUAL- M128 MODEL #S-11 MANUAL CAN OPENER 1 Model S-11 Can Opener Assembly Procedure The Model S-11 can opener will be assembled according to the following procedure: I. Model S-11 Handle and
Installation Instructions / Warranty
Installation Instructions / Warranty Axor Citterio 39384XX1 39387XX1 Axor Citterio 39385XX1 39386X1 Axor Massaud 18743001 18747001 Axor Starck X 10385XX1 10386XX1 Axor Starck 10375XX1 10379XX1 Axor Terrano
VERTICAL / FIXED BEAM CLAMP I-BEAM ANCHOR ADJUSTABLE FROM 4-14 Model # HD26249
VERTICAL / FIXED BEAM CLAMP I-BEAM ANCHOR ADJUSTABLE FROM 4-14 Model # HD26249 IMPORTANT!!! ALL PERSONS USING THIS EQUIPMENT MUST READ AND UNDERSTAND ALL INSTRUCTIONS. FAILURE TO DO SO MAY RESULT IN SERIOUS
Sun Vinyl New Construction Windows Homeowner's Warranty
Sun Windows Sun Vinyl New Construction Windows Homeowner's Warranty Vinyl Sun Windows, Inc. warrants for a period of 2 years that the quality of workmanship and materials of vinyl items shall be free from
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR Installation and Operation Manual 1 ENGLISH Safety Instructions 2 Features 2-3 Specifications 3 Installation 4-5 Wiring Diagrams 6-7 Warranty 8
MFT Series INTRODUCTION UNPACKING FERRULES
MFT Series Multiconductor Feedthroughs INSTRUCTION SHEET M0152/0103 Shop online at: omega.com e-mail: [email protected] For latest product manuals: omegamanual.info INTRODUCTION The OMEGA MFT multiconductor
National Optical & Scientific Instrument Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instrument Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR STEREOSCOPIC MICROSCOPES MODEL NUMBERS 400 400TL
World-Class Oil Sampling It is Possible
www.lubecontrol.com.au World-Class Oil Sampling It is Possible Lube Control Pty Ltd Providing lubrication solutions! Unit 5/53 Norfolk Rd, Marion, SA 5043 t: (08) 8298 5563 f: (08) 8298 6253 e: [email protected]
User s Guide. Oreck Air Purifier with HEPA Filtration. Important! Read this manual carefully, and keep for future reference.
User s Guide Oreck Air Purifier with HEPA Filtration Important! Read this manual carefully, and keep for future reference. Enjoy Congratulations on your purchase of the Oreck Air Purifier with HEPA Filtration.
Wafer Check Valve ASAHI AV VALVES. Contents. User s Manual. (1) Be sure to read the following description of our product warranty 1
Serial No. H-V066-E-2 Wafer Check Valve User s Manual Contents (1) Be sure to read the following description of our product warranty 1 (2) General operating instructions 2 (3) General instructions for
Bottom Loading Water Dispenser
Bottom Loading Water Dispenser Model # 601000 TO REDUCE THE RISK OF INJURY AND PROPERTY DAMAGE, USER MUST READ THIS MANUAL BEFORE ASSEMBLING, INSTALLING & OPERATING DISPENSER. SAVE THIS MANUAL FOR FUTURE
Administrative Procedure
Administrative Procedure Number: 410 Effective 1/30/98 Supercedes: 1 of Subject: MEDICAL WASTE MANAGEMENT PLAN 1.0. PURPOSE: The purpose of the California State University Los Angeles Medical Waste Management
Propane Conversion Kit PROPANE CONVERSION KIT INSTALLATION MANUAL
Propane Conversion Kit Supersedes: 145.25-IOM2 (708) Form 145.25-IOM2 (908) PROPANE CONVERSION KIT INSTALLATION MANUAL "LPKIT " - PROPANE CONVERSION KIT Kits are available for field conversion from natural
This instruction is valid for all ACD pump models shown on page 2
Screw pumps ACD Maintenance and Service Instruction This instruction is valid for all ACD pump models shown on page 2 Contents Page List of components 2 Exploded view/ordering code 3 Service intervals
How To Size A Stenner Pump
Metering Pumps Manufactured Since 1957 PCM Series Pump Control Module INSTALLATION AND MAINTENANCE MANUAL Table of Contents WarrantyandServicePolicy...3 Safety Information...............................................