A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS



Similar documents
Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.

A REVIEW ON APPLICATION OF CAD AND FEM TECHNOLOGY IN DESIGN OF TAPER DENTAL IMPLANT

IMPLANT DENTISTRY EXAM BANK

The mechanical behavior of the bone microstructure around dental implants

Clinical and Laboratory Procedures for Fixed Margin Implant Abutments

Ridge Reconstruction for Implant Placement

SCD Case Study. Treatment Considerations for Implant Rehabilitation

Three-Dimensional Analysis Using Finite Element Method of Anterior Teeth Inclination and Center of Resistance Location

LATERAL BONE EXPANSION FOR IMMEDIATE PLACEMENT OF ENDOSSEOUS DENTAL IMPLANTS

The Effect of Dental Implant Materials and Thread Profiles A Finite Element and Statistical Study

Retention of maxillary implant overdenture bars of different designs

Histologic comparison of biologic width around teeth versus implants: The effect on bone preservation

Straumann Dental Implant System. Implant Selection Guide.

Dental Implant Treatment after Improvement of Oral Environment by Orthodontic Therapy

More than a fixed rehabilitation.

Supervisors: Dr. Farhan Raza Khan

Implant rehabilitation in the edentulous jaw: the All-on-4 immediate function concept

A Study of 25 Zygomatic Dental Implants with 11 to 49 Months Follow-up After Loading

Tooth Supported Overdentures

Telescopic Denture A Treatment Modality for Minimizing the Conventional Removable Complete Denture Problems: A Case Report

Optimum Selection of the Dental Implants According to Length and Diameter Parameters by FE Method in the Anterior Position

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant

ATLANTIS ISUS the benchmark for CAD/CAM suprastructures. For all major implant systems

Quality from Switzerland

Appropriate soft tissue closure represents a critical

Photoelastic Comparison of Single Tooth Implant-Abutment-Bone of Platform Switching vs Conventional Implant Designs

Healing Abutment Selection. Perio Implant Part I. Implant Surface Characteristics. Single Tooth Restorations. Credit and Thanks for Lecture Material

Dental Implants and Esthetics

Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling

Molar Uprighting Dr. Margherita Santoro Division of Orthodontics School of Dental and Oral surgery. Consequences of tooth loss.

Placement of dental implants is a highly successful

Improving Esthetics with Sequential Treatment Planning and Implant-Retained Dentures

education Although demographic factors and growing patient awareness of the benefits of dental implants

Finite element analysis for dental prosthetic design

Treatment planning for the class 0, 1A, 1B dental arches

Dental Implant Options in Atrophic Jaws

COMPARISON OF SHORT DENTAL IMPLANTS UNDER DIFFERENT LOADING CONDITIONS AND LOCATIONS

Luigi Vito Stefanelli,D.Eng.,DDS

Replacement of the upper left central incisor with a Straumann Bone Level Implant and a Straumann Customized Ceramic Abutment

The Mandibular Two-Implant Overdenture First-Choice. Standard of Care for the Edentulous Denture Patient

GIVE YOUR PATIENTS THE FREEDOM TO EAT, SPEAK AND LAUGH AGAIN.

Overdenture Implants versus Teeth Quality of Life and Objective Therapy Evaluation

Congenital absence of mandibular second premolars

Design Optimization of a Customized Dental Implant Manufactured via Electron Beam Melting

BIOMECHANICAL ANALYSIS OF NORMAL AND IMPLANTED TOOTH USING BITING FORCE MEASUREMENT

Unless a prosthetic replacement is inserted soon

Rehabilitation of a complex case with zirconium dental implants

Oftentimes, as implant surgeons, we are

Anatomic limitations in the maxilla provide challenges

All-on-4 treatment concept with NobelSpeedy Groovy

Prosthetic treatment planning on the basis of scientific evidence.

Retrospective study on the survival rate of IBS implant

Stress distribution characteristics in dental implant influenced by its wall thickness

Resorptive Changes of Maxillary and Mandibular Bone Structures in Removable Denture Wearers

The Transition from Teeth to Implants and the Use of Post-ceramic Soldering

Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design

Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques

Implanting Brilliancy

DENTAL IMPLANTS DR JEBIN,MDS.,D.ICOI

CUSTOMIZED PROVISIONAL ABUTMENT AND PROVISIONAL RESTORATION FOR AN IMMEDIATELY-PLACED IMPLANT

How To Plan A Dental Implant With A 3D Image Based Program

Application of finite element method in dental implant research

Stress Distribution in Mandible Regulated by Bone and Dental Implant Parameters: Part I - Methodology

THE MANAGEMENT OF A COMPLEX IMPLANT CASE USING CAD-CAM TECHNOLOGY: A CLINICAL REPORT

Replacing Hopeless Retained Deciduous Teeth in Adults Utilizing Dental Implants: Concepts and Case Presentation

Lap Fillet Weld Calculations and FEA Techniques

1 The Single Tooth Implant. The Ultimate Aesthetic Challenge

Zimmer Trabecular Metal Dental Implant

ALL-CERAMIC DENTAL IMPLANT SOLUTIONS

BENDABLE VS. ANGULATED DENTAL IMPLANTS: CONSIDERATION OF ELASTIC AND PLASTIC MATERIAL PROPERTIES BASED ON EXPERIMENTAL IMPLANT MATERIAL DATA AND FEA

Implant Protected Occlusion

A Multicenter 12-Month Evaluation of Single-tooth Implants Restored 3 Weeks After 1-Stage Surgery

Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology

Osseointegrated implant connected with natural tooth using a semiprecision attachment: A clinical report

Clinical Effectiveness of Implant-Supported Removable Partial Dentures A Review of the Literature and Retrospective Case Evaluation

CAD/CAM technology supporting successful implant therapy

While the prosthetic rehabilitation of

Influence of Diameter and Length of Implant on Early Dental Implant Failure

MATERIALS AND METHODS

Dental Updates. Excerpted Article Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS

Outcomes of Placing Short Dental Implants in the Posterior Mandible: A Retrospective Study of 124 Cases

A Review of the Abutment Tooth, Part 1

Manufacturer HEAD OFFICE. OSSTEM Germany GmbH. Mergenthalerallee MANUFACTURER

CDT 2015 Code Change Summary New codes effective 1/1/2015

Restoration of the Edentulous Maxilla: The Case for the Zygomatic Implants

THE EVOLUTION OF EXTERNAL AND INTERNAL IMPLANT/ABUTMENT CONNECTIONS

IMMEDIATE CUSTOM IMPLANT PROVISIONALIZATION: A PROSTHETIC TECHNIQUE

Implant Overdentures: A New Standard of Care for Edentulous Patients Current Concepts and Techniques

Implants in your Laboratory: Abutment Design

Straumann Bone Level Tapered Implant Peer-to-peer communication

Transcription:

A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS Evandro Afonso Sartori, evandroasartori@yahoo.com.br Josué Ricardo Broilo Luiz Oscar Honorato Mariano Leandro Luis Corso, llcorso@yahoo.com.br (Times New Roman, Bold, size 10) Universidade de Caxias do Sul, Caxias do Sul, RS Brazil. Rogério José Marczak, rato@mecanica.ufrgs.br Universidade Federal do Rio Grande do Sul, Porto Algre,RS Brazil. Rosemary Sadami Shinkai, rshinkai@pucrs.br Abstract. This 3D finite element study assessed the stress distribution in mandibular overdenture supported by the 3i Locator system attached to two cylindrical or conical implants placed at the canine region. A 100 N vertical load was applied over the right first molar of the two overdenture models. The von Mises stresses were concentrated at the cervical and middle implant thirds. Cylindrical implants showed larger crestal bone area of high stresses in comparison to the conical implants. The results suggest that conical implants reduce the stresses in peri-implant crestal bone at both loading and non-loading sides. (single space line, size 10) Keywords: Dental implants; implant-supported overdenture; biomechanics; finite element analysis. le space line, size 10) 1. INTRODUCTION Conical implants have been introduced to mimic the convergence of natural dental roots. Previous studies reported that conical implants present good clinical predictability in mandibular fixed prosthesis with early loading (Kronström et al., 2003) and provide easier surgical insertion and greater primary stability than cylindrical implants (Sakoh et al., 2006). An optimal implant geometry should prevent concentration of high stresses at the bone crest to minimize marginal bone resorption, although the clinical cut-off limit is unknown. Cylindrical or conical implant shapes may modify the stress distribution at bone, implant, and prosthetic component interfaces. Most biomechanical studies on this topic have focused implant-supported fixed prostheses and single implants (Sakoh et al. 2006, Huang et al. 2007, Petrie et al. 2005), and data on overdentures supported by conical implants are scarce. Therefore, this study aims to compare the von Mises stresses generated by the application of a static vertical load in a mandibular overdenture supported by two cylindrical and conical implants. 2. MATERIALS AND METHODS Two three-dimensional solid models were built with the individual models containing cylindrical and conical (5 degrees of apical convergence) 4.1 x 11.5 implants (3i Implant Innovations, Palm Beach, Florida, USA), locator 4 mm attachment components (3i, Implant Innovations, Palm Beach, Florida, USA), edentulous mandible, complete denture, and a simulated 3mm-thick mucosa. The individual images of implants, abutments, mucosa, complete denture, and bone were obtained by laser scanning, computed tomography of the mandible (seventy 1mm-thick slices), and computeraided design using the softwares Matlab (The MathWorks, Natick, Massachusetts, USA ), Geomagic 7.0 (Raindrop, Research Triangle Park, USA), Rhynoceros 3D 3.0 (McNeel & Associates, Seattle, USA), and SolidWorks 2006 (SolidWorks Corporation, Concord, Massachusetts, USA).

The geometric models of implants and abutments were mounted at the canine region (teeth 43 and 33) to build two models: model 1 overdenture supported by two cylindrical implants, and model 2 overdenture supported by two conical implants. The finite element models were generated and analyzed in the software Ansys (Ansys Inc., Houston, USA) using three types of structural solid p-elements (10-Node Quadratic Tetrahedron, Quadratic Triangular Contact, and Quadratic Triangular Target). The grid refinement until the estimator is smaller than a prescribed bound. The mechanical elastic properties of bone, implants, attachment components, simulated mucosa, and denture are displayed in Table 1. The materials were considered homogeneous, isotropic, and linearly elastic. A perfect contact between bone and implants (100% osseointegration) was assumed. (Huang et al. 2007, Petrie et al., 2005). Boundary conditions were applied by constraining three degrees of freedom at each of the nodes located at the mandibular condyles. Table 1. Mechanical elastic properties of bone and materials used in the anisotropic models. Young s modulus (MPa) Poisson s ratio Cortical bone 13,700 0.3 Cancellous bone 1,370 0.3 Mucosa 1 0.37 Mandibular nerve 0.1 0.3 Overdenture (acrylic resin) 4,500 0.35 Implant (titanium) 135,000 0.3 Screw (titanium) 114,000 0.3 Attachment (titanium) 114,000 0.3 PTPE attachment component 19,000 0.3 The models were loaded by an axial vertical load of 100 N applied on the right first molar, using a masticatory bolus simulation modeled as a rigid semi-sphere placed over the denture tooth (Daas et. al, 2008). The von Mises stresses were qualitatively analyzed in selected areas. 3. RESULTS Lower magnitude von Mises stress areas were concentrated on the buccal anterior region of the overdentures supported by either cylindrical or conical implants (Fig.1). The stress levels on both implants resulted similar and concentrated in the cervical and middle thirds (Fig.2). Figure 1. Von Mises stress levels in overdentures supported by cylindrical and conical implants.

Figure 2. Von Mises stress levels in the cylindrical and the conical implants placed at the teeth 43 and 33 sites (loading and non-loading sides, respectively). Figure 2 shows that high stress levels on the implants occur on localized areas, due to stress concentration or contact. Since the stress analysis of the implants is not the objective of the present work, no effort was made to study in better detail these areas. On the other hand, Fig. 3 shows the stress distribution in peri-implant crestal bone at the loading side (tooth 43 site). The stress magnitude was similar for both types of implant but the area subjected to higher stress was larger for the cylindrical implant. At the non-loading side, the cylindrical implant also generated a larger area of high stress in comparison to the conical implant (Fig.4). 4. DISCUSSION This finite element analysis work presented the stress distribution in a mandibular overdenture supported by the 3i Locator system attached to two types of implants placed at the canine region. It is worth to note that oversimplifications traditionally used in these models were avoided here, so that the numerical models included a detailed representation of the implants threads. The stresses in both cylindrical and conical implants were concentrated at the cervical and middle thirds. The implants placed at the loading side showed larger area of stress than the contralateral implants, but the stress magnitude was similar. In relation to the mandibular bone, the anterior surface had larger stress concentration at the distal area of both types of implants on the loading side than that of the contralateral side. The stress magnitude on the bone surface was similar but the area of higher stress distribution was larger for the cylindrical implant at the loading side. In a longitudinal section of the bone at the tooth 43 site, the higher stresses were concentrated at the cervical region of both implant shapes corresponding to the cortical bone. In relation to the bone at the tooth 33 region, the stresses were higher at the mesial side of the cylindrical implant compared to the conical implant. The numerical results for stress showed that the use of two conical implants reduced the area of higher von Mises stresses in the peri-implant crestal bone at the loading and non-loading sides of the implant-supported overdenture. Nevertheless, implants and denture did not show relevant variations of stress magnitude and distribution, indicating that the small geometric differences in the shape of the implants do not affect significantly the nominal stress levels. This occurred because the conical implant used has a slight apical convergence. Certainly, it is expected that conical implants with greater apical convergence show higher crestal strain in single implants (Petrie et al., 2005). The present study assumed various simplifications to simplify the numerical model and reduce the computational cost of the analysis, and therefore one cannot extrapolate the main conclusions drew from the stress analysis to other situations. Further studies can incorporate better definition of the contact areas (osseointegration), analyze the stress generated by different load incidences, dynamic loadings, functionally graded properties, and non-linear material behavior. Nonetheless, even within these limitations, the results obtained in the present work suggest that conical implants may offer biomechanical advantages with no detectable risk to support mandibular overdentures in comparison with the standard, more widely used cylindrical implants.

The methodology used is, however, very simple and can be used by implant manufacturers to design/analyze other conical implants and attachment systems with little effort. Figure 3. Von Mises stress levels in the crestal bone surrounding the cylindrical and the conical implants at the loading side.

Figure 3. Von Mises stress levels in the crestal bone surrounding the cylindrical and the conical implants at the loading side.

5. CONCLUSIONS Within the limitations of this study, the results suggest that conical implants may provide reduction of the area of stress in the mandibular bone at the loading side of the overdenture. At the contralateral side the stress was similar for both implants. 6. REFERENCES Kronström M, Widbom T, Löfquist LE, Henningson C, Widbom C, Lundberg T. Early functional loading of conical Brånemark implants in the edentulous mandible: a 12-month follow-up clinical report. J Prosthet Dent 2003;89:335-340. Sakoh J, Wahlmann U, Stender E, Nat R, Al-Nawas B, Wagner W. Primary stability of a conical implant and a hybrid, cylindric screw-type implant in vitro. Int J Oral Maxillofac Implants 2006;21:560-566. Huang HL, Chang CH, Hsu JT, Fallgatter AM, Ko CC. Comparison of implant body designs and threaded designs of dental implants: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants 2007;22:551-562. Petrie CS, Williams JL. Comparative evaluation of implant designs: influence of diameter, length, and taper on strains in the alveolar crest. A three-dimensional finite-element analysis. Clin Oral Implants Res 2005;16:486-494. Daas M, Dubois G, Bonnet AS, Lipinski P, Rignon-Bret C.A complete finite element model of a mandibular implantretained overdenture with two implants: Comparison between rigid and resilient attachment configurations. Med Eng Phys 2008;30:218-225. 7. RESPONSIBILITY NOTICE The author(s) is (are) the only responsible for the printed material included in this paper.