1 APPLICATION NOTE ICP - Mass Spectrometry Authors: Madjid Hadioui, Ph.D. Université de Montréal Kevin Wilkinson, Ph.D. Université de Montréal Chady Stephan, Ph.D. PerkinElmer, Inc. Assessing the Fate of Silver Nanoparticles in Surface Water using Single Particle ICP-MS Introduction During the last decade, the production and use of engineered nanomaterials (ENMs) have experienced a drastic increase, resulting in a potential risk of their release into the environment. Therefore, the study of their impact on the environment becomes crucial. The appropriate ecological risk assessment and management of ENMs in the environment requires quantitative measurements of both exposure and effects 1 that should, ideally, be performed by in situ analysis and give physicochemical characterization. However, most analytical techniques are not suitable for environmental matrices since nanoparticle concentrations are typically very low 2. Some studies on the persistence, aggregation and dissolution of metal nanoparticles in natural freshwaters and synthetic complex waters were recently published 3-7. Historically, particle size has been measured by dispersive light scatter (DLS) and tunneling electron microscopy (TEM), while dissolved content has been measured by ultrafiltration. These common techniques have known limitations for measuring low concentrations in the presence of colloidal species in complex waters.
2 Alternatively, single particle inductively coupled plasma mass spectrometry (SP-ICP-MS) has been found to be a promising technique for detecting and characterizing metal nanoparticles at very low concentrations SP-ICP-MS is fast and efficient and can provide more information than other currently available techniques. It can lead to the determination of particle size distribution, particle number concentration, and the proportion of dissolved metal. Moreover, it can distinguish between particles of different elements. The principle of SP-ICP-MS is based on the measurement of the signal intensity produced by a single particle. The suspension of nanoparticles must be sufficiently diluted to make sure that only a single particle reaches the plasma at a time, where it is atomized and ionized, producing a signal of relatively high intensity which is measured as one pulse. If the suspension contains dissolved metal of the same element as in the particles, a constant, continuous signal of that element will be produced as a result of its homogeneous distribution. The recorded signal intensity as a function of time can be processed using a theoretical approach first developed by Duegeldre et al for natural metal colloids and then supported by other authors such as Laborda et al. 16, 17. Ag, ZnO and TiO 2 are among the most frequently studied nanoparticles 18. In the case of Ag, this is most probably due to the fact that nanosilver is among the most common nanomaterials found in consumer products and tends to release free Ag +, especially at low concentration 19. The aim of this work is to investigate the efficiency of SP-ICP-MS for the detection and characterization of metal nanoparticles in environmental waters where they can be involved in various physicochemical processes as shown by Figure 1. Dissolved silver, including released free ions (Figure 1C) and soluble complexes (Figure 1E), can easily and instantly be measured by SP-ICP-MS. These dissolved species can also be determined by ultrafiltration followed by total metal quantification using ICP-MS or AAS, but this procedure is time consuming since it requires the pre-equilibration of the membrane for at least three cycles of centrifugation, generally 20 min each 19. Aggregates (Figure 1B) and remaining stable silver nanoparticles (Figure 1G) can be counted and measured by other commonly used techniques (DLS, NTA, TEM) but SP-ICP-MS is the only method that can distinguish between nag and other colloids in surface water. Experimental A PerkinElmer NexION 350X Inductively Coupled Plasma Mass Spectrometer (ICP-MS) was used for data acquisition using the Nano Application Module within Syngistix for ICP-MS software (SP-ICP-MS instrumental parameters are given in Table 1). The sample introduction system consisted of a quartz cyclonic spray chamber, type C0.5 concentric glass nebulizer and a 2 mm bore quartz injector. Commercially available suspensions of gold and silver nanoparticles were used in this work. A NIST reference material (RM 8013) consisting of a suspension of gold nanoparticles (60 nm nominal diameter, 50 mg/l total mass concentration and stabilized in a citrate buffer) was used to determine the nebulization efficiency. Suspensions of silver nanoparticles were purchased from Ted Pella Inc.: citrate coated (40 and 80 nm nominal diameter) and bare (80 nm nominal diameter) nanosilver suspensions (product numbers , and SC, respectively). Accidental or deliberate release of silver nanoparticles into surface water Groundwater Figure 1. Possible fates of silver nanoparticles in surface waters: (A) Dissolution process leading to free ions release and smaller particles; (B) Aggregation into larger particles, which may settle out of the water, depending on the aggregate size; (C, D) Adsorption of released Ag + and nag, respectively, onto other solids present in the water; (E) Formation of soluble complexes; (F) Reaction with other components in the water, which may result in precipitation; (G) nag remaining stable. 2
3 The surface water was sampled in Rivière des Prairies, Montreal, Canada and filtered with 0.2 µm filter paper prior to spiking with silver nanoparticles. Nano-Ag suspensions were added to water samples with concentrations ranging from 2.5 to 33.1 µg Ag L -1 and left to equilibrate under continuous and gentle shaking. Prior to SP-ICP-MS analysis, small aliquots of the samples were diluted to below 0.2 µg Ag L -1. Data acquisition was performed in triplicate measurements for each sample, and deionized (DI) water was analyzed between replicates to check memory effects. As shown by Figure 2, Syngistix Nano App Module generates frequency vs. size (nm) plots with an interactive focusing window, eliminating the need for any subsequent data processing using theoretical equations SP-ICP-MS data processing is based on distinguishing between the signal of dissolved metal and that of nanoparticles, counting the pulses (or events) corresponding to individual nanoparticles and converting their intensities to particle size. Indeed, the frequency of the events (pulses) provides particle number concentration, and the intensity of each pulse is proportional to the mass of analyte. The latter was converted to volume and then into size knowing the density and the geometry of the particle using the Syngistix Nano Application Module, which provides automated data treatment, so no subsequent manual data processing is required. Results Even after filtration of surface waters at 0.2 µm, Nanoparticle Tracking Analysis (NTA), using an LM14 instrument (NanoSight Ltd. Amesbury, Wiltshire, UK) with a green laser at 532 nm, showed the presence of non-negligible amounts of colloidal particles with an average diameter of ca. 110 nm. Thus, the addition of metal nanoparticles to this complex matrix will make their detection and characterization very difficult, if not impossible, with commonly used techniques (DLS, TEM, etc.). Table 1. Instrumental parameters for SP-ICP-MS data acquisition. Parameter Instrument Nebulizer Spray Chamber Torch and Injector Value NexION 350X ICP-MS Concentric Cyclonic Power (W) 1600 Plasma Gas (L/min) 18 Aux Gas (L/min) 1.2 Neb Gas (L/min) 0.97 Sample Uptake Rate (ml/min) 0.5 Sample Tubing Dwell Time (µs) 100 Sampling Time (s) 60 Quartz Torch and Quartz 2.0 mm bore injector Black/Black Figure 2. Interactive view of data acquisition with Syngistix for ICP-MS software: signal intensity as a function of acquisition time for silver nanoparticles (bare nag, 60 nm nominal diameter, total metal concentration ng/l) in surface water. 3
4 Furthermore, even the determination of the dissolved fraction that is usually performed by ultrafiltration may be inadequate because silver ions may adsorb on the surface of the colloids and, therefore, will be retained by the filtration membrane. Consequently, the proportion of dissolved metal will be underestimated. SP-ICP-MS measurements were found to be more effective and to have fewer limitations than other techniques. Indeed, the presence of other insoluble particles does not interfere with the analysis of silver nanoparticles, as the signal of Ag is recorded independently of the other constituent elements of the colloids. Figures 3-5 show the evolution of the average particle diameter and the percentage of dissolved metal over time in both pure and river water. In all cases, the average particle size of the persistent nanoparticles remains substantially constant (Figures 3a, 4a and 5a). For suspensions of particles with a nominal diameter greater than 40 nm, between 50 and 80% of the particles persist for at least five days of equilibration in pure and surface water (Figures 3b and 4b). Under the experimental conditions of this work, the coating appears to have no significant effect on the dissolution of nanoparticles over time, as both citrate-coated and bare nag (80 nm) suspensions showed a slight decrease of particulate silver by ca. 20% during five days. In the meantime, for the same size and equilibration time, the proportion of dissolved silver was found higher in the case of citrate-coated nag. This does not necessarily mean that bare nanosilver is more stable than citrate-coated nag. In fact, the release of silver ions may be due to oxidation and/or to residual Ag + adsorbed on the surface of nag or bonded to the coating. Thus, we believe that the stability and behavior of nanoparticles in any medium will depend on the synthesis procedure. According to Figure 5b, smaller particles with a nominal diameter below 40 nm tend to dissolve in greater quantities, but caution should be taken here due to the limitation of the technique to the detection of particles under 20 nm. Figure 3. Evolution of (a) the average particle size and (b) the proportion of particulate silver as a function of equilibration time in pure (l) and surface (o) waters. Waters were spiked with 80 nm citrate-coated nag, and the total metal concentrations were equal to 72.2 and 72.4 ng L -1 in DI water and surface water, respectively. Figure 4. Evolution of (a) the average particle size and (b) the proportion of particulate silver as a function of equilibration time in pure (l) and surface (o) waters. Waters were spiked with uncoated 80 nm nag, and the total metal concentrations were equal to and ng L -1 in DI water and surface water, respectively. 4
5 Figure 5. Evolution of (a) the average particle size and (b) the proportion of particulate silver as a function of equilibration time in pure (l) and surface (o) waters. Waters were spiked with 40 nm citrate-coated nag, and the total metal concentrations were equal to and ng L -1 in DI water and surface water, respectively. Indeed, the signal intensity of such small nanoparticles is very low and overlaps with the background noise therefore, all particles smaller than ca. 15 nm will be erroneously counted as dissolved metal. In the case of bare 80 nm nag, both average particle size and percentage of particles were found higher in surface water compared to DI water (Figure 4). In the case of citrate-coated nag, no noticeable difference in behavior was found in either aqueous media. This can be explained by the fact that bare nanoparticles are more favorable to aggregation than citrate-stabilized ones. But overall, no significant aggregation was observed. Conclusion Using the Nano App Module in the Syngistix for ICP-MS software, it was possible to study the behavior of silver nanoparticles in surface water without using any subsequent manual data processing. The technique has allowed the effective and selective measurement of changing particle size, aggregation and dissolution over time at low concentration. SP-ICP-MS is practically the only suitable technique that can provide such information on the fate of metal nanoparticles at very low concentrations in environmental waters. Although this study only showed the effectiveness of the technique in the particular case of nag in river water, it is, without any doubt, applicable to other types of metal and metal oxide nanoparticles in a variety of complex matrices form wastewater, effluents to biological fluids and culture media. References 1. Klaine, S.J., et al., Nanomaterials in the environment: Behavior, fate, bioavailability, and effects. Environmental Toxicology and Chemistry, (9): p Domingos, R.F., et al., Characterizing Manufactured Nanoparticles in the Environment: Multimethod Determination of Particle Sizes. Environmental Science & Technology, (19): p Chinnapongse, S.L., R.I. MacCuspie, and V.A. Hackley, Persistence of singly dispersed silver nanoparticles in natural freshwaters, synthetic seawater, and simulated estuarine waters. Science of the Total Environment, (12): p Weinberg, H., A. Galyean, and M. Leopold, Evaluating engineered nanoparticles in natural waters. Trac-Trends in Analytical Chemistry, (1): p Van Hoecke, K., et al., Aggregation and ecotoxicity of CeO 2 nanoparticles in synthetic and natural waters with variable ph, organic matter concentration and ionic strength. Environmental Pollution, (4): p Badireddy, A.R., M.R. Wiesner, and J. Liu, Detection, Characterization, and Abundance of Engineered Nanoparticles in Complex Waters by Hyperspectral Imagery with Enhanced Darkfield Microscopy. Environmental Science & Technology, (18): p Li, X. and J.J. Lenhart, Aggregation and Dissolution of Silver Nanoparticles in Natural Surface Water. Environmental Science & Technology, (10): p Krystek, P., et al., Application of plasma spectrometry for the analysis of engineered nanoparticles in suspensions and products. Journal of Analytical Atomic Spectrometry, (9): p Tuoriniemi, J., G. Cornelis, and M. Hassellov, Size Discrimination and Detection Capabilities of Single-Particle ICPMS for Environmental Analysis of Silver Nanoparticles. Analytical Chemistry, (9): p Pace, H.E., et al., Single Particle Inductively Coupled Plasma- Mass Spectrometry: A Performance Evaluation and Method Comparison in the Determination of Nanoparticle Size. Environmental Science & Technology, (22): p
6 11. Degueldre, C. and P.Y. Favarger, Colloid analysis by single particle inductively coupled plasma-mass spectroscopy: a feasibility study. Colloids and Surfaces a-physicochemical and Engineering Aspects, (1-3): p Degueldre, C., P.Y. Favarger, and S. Wold, Gold colloid analysis by inductively coupled plasma-mass spectrometry in a single particle mode. Analytica Chimica Acta, (2): p Degueldre, C. and P.Y. Favarger, Thorium colloid analysis by single particle inductively coupled plasma-mass spectrometry. Talanta, (5): p Degueldre, C., P.Y. Favarger, and C. Bitea, Zirconia colloid analysis by single particle inductively coupled plasma-mass spectrometry. Analytica Chimica Acta, (1-2): p Degueldre, C., et al., Uranium colloid analysis by single particle inductively coupled plasma-mass spectrometry. Talanta, (3): p Laborda, F., et al., Selective identification, characterization and determination of dissolved silver(i) and silver nanoparticles based on single particle detection by inductively coupled plasma mass spectrometry. Journal of Analytical Atomic Spectrometry, (7): p Laborda, F., et al., Critical considerations for the determination of nanoparticle number concentrations, size and number size distributions by single particle ICP-MS. Journal of Analytical Atomic Spectrometry, (8): p Baun, A., et al., Setting the limits for engineered nanoparticles in European surface waters - are current approaches appropriate? Journal of Environmental Monitoring, (10): p Hadioui, M., S. Leclerc, and K.J. Wilkinson, Multimethod quantification of Ag+ release from nanosilver. Talanta, (0): p PerkinElmer, Inc. 940 Winter Street Waltham, MA USA P: (800) or (+1) For a complete listing of our global offices, visit Copyright 2014, PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners _01
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media
Enhanced Organic Precursor Removals Using Aged Media Page 1 Enhanced Organic Precursor Removals Using Aged Media O b j e c t i v e s The main goals of this project was to evaluate the dissolved organic
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
How To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
Department of Environmental Engineering
Department of Environmental Engineering Master of Engineering Program in Environmental Engineering (International Program) M.Eng. (Environmental Engineering) Plan A Option 1: (1) Major courses: minimum
Sample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this document
NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight.
NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight. When your instrument is more efficient and productive, so too are your scientists and your lab. And nothing puts the flow in your
Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
Elemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
Supporting Information
Supporting Information Copyright Wiley-VCH Verlag GmbH & Co. KGaA, 69451 Weinheim, 2013 A Study of Commercial Nanoparticulate g-al 2 O 3 Catalyst Supports Yahaya Rozita, [a] Rik Brydson,* [a] Tim P. Comyn,
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
The 30-Minute Guide to ICP-MS
Technical Note The 30-Minute Guide to ICP-MS ICP-Mass Spectrometry A Worthy Member of the Inorganic Analysis Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been gaining
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
HS 1003 Part 2 HS 1003 Heavy Metals Test
HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy
Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
Chromium Speciation in Drinking Water by LC-ICP-MS
APPLICATION NOTE LC-ICP-MS Authors: Helmut Ernstberger Ken Neubauer PerkinElmer, Inc. Shelton, CT Chromium Speciation in Drinking Water by LC-ICP-MS Introduction With the growing concern of pollutants
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
NUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility
Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES
ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES HORIBA Scientific Longjumeau, France Keywords: metallurgy, non-ferrous, halogens 1 Introduction
SYNERGISTIC APPLICATION OF ADVANCED PRIMARY AND SECONDARY WASTEWATER TREATMENT SYSTEMS
SYNERGISTIC APPLICATION OF ADVANCED PRIMARY AND SECONDARY WASTEWATER TREATMENT SYSTEMS Published in Water and Waste Digest membrane issue, November 2008 Miroslav Colic; Chief Scientist, Clean Water Technology
REMOVAL OF PHOSPHATE FROM WASTEWATER USING LOW-COST ADSORBENTS
International Journal of Engineering Inventions ISSN: 2278-7461, www.ijeijournal.com Volume 1, Issue 7 (October2012) PP: 44-50 REMOVAL OF PHOSPHATE FROM WASTEWATER USING LOW-COST ADSORBENTS Dr. C.R.Ramakrishnaiah
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
Safe Nano Design Molecule Manufacturing Market
August 14 16, 212 College of Nanoscale Science & Engineering (CNSE) of the University at Albany National Institute for Occupational Safety and Health (NIOSH) Prevention through Design Program Safe Nano
Release of Nanosilver and its Behavior in Urban Wastewater
Release of Nanosilver and its Behavior in Urban Wastewater Michael Burkhardt HSR University of Applied Sciences Institute of Environmental and Process Engineering (UMTEC) Rapperswil, Switzerland Rome,
Removing Heavy Metals from Wastewater
Removing Heavy Metals from Wastewater Engineering Research Center Report David M. Ayres Allen P. Davis Paul M. Gietka August 1994 1 2 Removing Heavy Metals From Wastewater Introduction This manual provides
CHAPTER 2 EXPERIMENTAL. g/mol, Sigma-Aldrich, Germany. 2.1.2 Magnesium acetate tetrahydrate (C 4 H 6 MgO. 4 4H 2 O), assay 99.0%,
CHAPTER 2 EXPERIMENTAL 2.1 Chemicals and Equipments 2.1.1 Zinc naphthenate (2(C 11 H 7 O 2 ). Zn), assay
Determination of calcium by Standardized EDTA Solution
Determination of calcium by Standardized EDTA Solution Introduction The classic method of determining calcium and other suitable cations is titration with a standardized solution of ethylenediaminetetraacetic
2015 2016 Ph.D. Science & Engineering Positions. Wilmington, DE. Analytical Sciences Wilmington, DE. Sciences Wilmington, DE.
2015 2016 Ph.D. Science & Engineering Positions 1) Synthetic Organic Chemistry Crop Protection Newark, DE 2) Polymer Chemistry Performance Polymers Various 3) Polymer Engineering Performance Polymers Various
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems
Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES FIMS 100 and 400 Flow Injection Mercury Systems Delivering the Performance You Need The determination and monitoring of sub-ppb concentrations
Electron microscopy as an analytical tool for detection and characterization of inorganic nanoparticles in food. EU Framework Programme 7-Nanolyse
Electron microscopy as an analytical tool for detection and characterization of inorganic nanoparticles in food EU Framework Programme 7-Nanolyse Nanostructures in food 1. Natural e.g. milk proteins 2.
Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide
chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo
Factors Affecting Precipitation of Calcium Carbonate
Factors Affecting Precipitation of Calcium Carbonate John A. Wojtowicz Chemcon Laboratory tests with clear solutions showed that precipitation of calcium carbonate does not occur in the ph range 7.5 to
All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &
105 6.1 CHARACTERIZATION OF TABLETS All the prepared formulations were subjected for following evaluation parameters and obtained results were showed in Tables 6.3 & 6.4. 6.1.1 Description (Size, Shape,
Chapter 13 - Solutions
Chapter 13 - Solutions 13-1 Types of Mixtures I. Solutions A. Soluble 1. Capable of being dissolved B. Solution 1. A homogeneous mixture of two or more substances in a single phase C. Solvent 1. The dissolving
Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
Human serum albumin (HSA) nanoparticles stabilized with. intermolecular disulfide bonds. Supporting Information
Human serum albumin (HSA) nanoparticles stabilized with intermolecular disulfide bonds Wentan Wang, Yanbin Huang*, Shufang Zhao, Ting Shao and Yi Cheng* Department of Chemical Engineering, Tsinghua University,
Phosphate Recovery from Municipal Wastewater through Crystallization of Calcium Phosphate
Phosphate Recovery from Municipal Wastewater through Crystallization of Calcium Phosphate A. Alamdari, Ph.D. * and S. Rohani, M.Sc. * Department of Chemical Engineering, School of Engineering, Shiraz University,
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
A. Types of Mixtures:
I. MIXTURES: SOLUTIONS 1) mixture = a blend of two or more kinds of matter, each of which retains its own identity and properties a) homogeneous mixture = a mixture that is uniform in composition throughout
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry
Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
Paper No. 4071 APPLICATION OF EQCM TO THE STUDY OF CO2 CORROSION
Paper No. 471 APPLICATION OF EQCM TO THE STUDY OF CO2 CORROSION Yang Yang, Bruce Brown and Srdjan Nešić Institute for Corrosion and Multiphase Technology, Department of Chemical and Biomolecular Engineering
Thème Analytical Chemistry ECUE composant l UE
Thème Analytical Chemistry ECUE composant l UE UE 7.1 Performing analyses in the environment and complex systems UE 7.2 Heterogeneity and complexity UE 7.3 Processes in aquatic systems and at geo-interfaces
STUDENT BACKGROUND READING FOR EXPERIMENT C: COLORIMETRIC GOLD NANOSENSOR
STUDENT BACKGROUND READING FOR EXPERIMENT C: In this experiment you will synthesize and test a plasmonic colorimetric nanosensor made of nanoparticles of gold. Here we provide you with some background
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic Absorption Spectrometry
application Note Atomic Absorption Authors Jijun Yao Renkang Yang Jianmin Chen PerkinElmer, Inc. Shelton, CT 06484 USA Accurate Determination of Lead in Different Dairy Products by Graphite Furnace Atomic
Start of introducing ICP MS method for analyzing water samples for Strontium. November 27, 2014 Tokyo Electric Power Company
Start of introducing ICP MS method for analyzing water samples for Strontium November 27, 2014 Tokyo Electric Power Company 1. Current Strontium analysis and introduction of various methods Strontium analysis
Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
Simplicity. Reliability. Performance
Simplicity Reliability Performance ICP Without Compromise At Teledyne Leeman Labs atomic spectroscopy is our business, our only business. We are deeply committed to providing you with technically superior
Environmental Engineering Professors Cal (Chair of Department), Richardson Associate Professor Huang Adjunct Faculty Brady, Hendrickx
Environmental Engineering Professors Cal (Chair of Department), Richardson Associate Professor Huang Adjunct Faculty Brady, Hendrickx Degrees Offered: B.S. in Environmental Engineering; M.S. in Environmental
Contract No: Disclaimer:
Contract No: This document was prepared in conjunction with work accomplished under Contract No. DE-AC09-09SR22505 with the U.S. Department of Energy (DOE) National Nuclear Security Administration (NA).
IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS
IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS INTRODUCTION Each REACH registrant is required to file its
Purification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
ENE 806, Project Report 3 CHEMICAL PRECIPITATION: WATER SOFTENING. Grégoire Seyrig Wenqian Shan
ENE 806, Project Report 3 CHEMICAL PRECIPITATION: WATER SOFTENING Grégoire Seyrig Wenqian Shan College of Engineering, Michigan State University Spring 2007 ABSTRACT The groundwater with high level initial
Application Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
Mass Spectrometry Signal Calibration for Protein Quantitation
Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of
Advanced Treatment of Hazardous Wastes(1) Advanced Treatment of Hazardous Wastes(2) Advanced Environmental Chemistry. Design of Solid Waste Landfill
Course Description (전체 개설 교과목 개요) Advanced Treatment of Hazardous Wastes(1) This course is concerned with the management of hazardous materials and wastes in depth. We will deal with the physico-chemical
ICP - Mass Spectrometry APPLICATION NOTE. Authors. Introduction. Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA
APPLICATION NOTE ICP - Mass Spectrometry Authors Lee Davidowski Ewa Pruszkowski PerkinElmer, Inc. Shelton, CT USA Benefits of the NexION 300X/350X ICP-MS Coupled with the prepfast In-line Auto-dilution/calibrationSystem
Prentice Hall. Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition. High School. High School
Prentice Hall Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition High School C O R R E L A T E D T O High School C-1.1 Apply established rules for significant digits,
Silver Nanoparticles: No Threat to the Environment. George J. Maass, Ph.D. (Colloidal Science Laboratories)
Abstract: Silver Nanoparticles: No Threat to the Environment George J. Maass, Ph.D. (Colloidal Science Laboratories) Silver is an effective germ fighter and silver nanoparticles are widely recognized as
Particle Surface Area
Particle Surface Area TM Introducing the Acorn Area TM The Acorn Area TM is a revolutionary instrument designed to measure the surface area of nanoparticles dispersed in a liquid. This patented nuclear
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
Chemistry B11 Chapter 6 Solutions and Colloids
Chemistry B11 Chapter 6 Solutions and Colloids Solutions: solutions have some properties: 1. The distribution of particles in a solution is uniform. Every part of the solution has exactly the same composition
Opiates in Urine by SAMHSA GC/MS
application Note Gas Chromatography/ Mass Spectrometry Author Timothy D. Ruppel PerkinElmer, Inc. Shelton, CT 06484 USA Opiates in Urine by SAMHSA GC/MS Introduction The United States Department of Health
TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature
ProteoMass Peptide & Protein MALDI-MS Calibration Kit Catalog Number MSCAL1 Store at Room Temperature TECHNICAL BULLETIN Description This kit provides a range of standard peptides and proteins for the
Supporting Information
Supporting Information Simple and Rapid Synthesis of Ultrathin Gold Nanowires, Their Self-Assembly and Application in Surface-Enhanced Raman Scattering Huajun Feng, a Yanmei Yang, a Yumeng You, b Gongping
Treatment and Management of Oil Sands Tailings Water for Detoxification and Decontamination
Treatment and Management of Oil Sands Tailings Water for Detoxification and Decontamination Mohamed Gamal El-Din, Ph.D., P.Eng. Professor NSERC Senior IRC in Oil Sands Tailings Water Treatment Helmholtz
Silica Over-Saturation, Precipitation, Prevention and Remediation In Hot Water Systems Edited By Dave Peairs, Cal Water, Technical Director
Silica Over-Saturation, Precipitation, Prevention and Remediation In Hot Water Systems Edited By Dave Peairs, Cal Water, Technical Director Background Silica scaling becomes a problem when any hot water
Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)
Application Note: 52020 Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Timothy O. Deschaines, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words Array Automation
DNA NANOWIRES USING NANOPARTICLES ECG653 Project Report submitted by GOPI KRISHNA.ARI,[email protected],Fall-2008
DNA NANOWIRES USING NANOPARTICLES ECG653 Project Report submitted by GOPI KRISHNA.ARI,[email protected],Fall-2008 INTRODUCTION: Deoxyribonucleic acid (DNA) has been a key building block in nanotechnology
SUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
2. Deposition process
Properties of optical thin films produced by reactive low voltage ion plating (RLVIP) Antje Hallbauer Thin Film Technology Institute of Ion Physics & Applied Physics University of Innsbruck Investigations
Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
Identity, grouping and characterisation of silver based biocidal active substances including nano-silver
Identity, grouping and characterisation of silver based biocidal active substances including nano-silver Ulrike Frank Swedish Chemical Agency February 2012 Identity, grouping and characterisation of silver
Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.
T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
X Ray Flourescence (XRF)
X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose
LIST OF EQUIPMENT IN CHEMISTRY LABORATORY
LIST OF EQUIPMENT IN CHEMISTRY LABORATORY 1 Adana Science and Technology University-Bioengineering Department INSTRUMENTS: Analytical Balance (RADWAG AS220 C/2) 3 Analytical Balance (Denver Instrument)...4
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
6 Reactions in Aqueous Solutions
6 Reactions in Aqueous Solutions Water is by far the most common medium in which chemical reactions occur naturally. It is not hard to see this: 70% of our body mass is water and about 70% of the surface
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray
Nanoparticle Deposition on Packaging Materials by the Liquid Flame Spray Hannu Teisala a, Mikko Tuominen a, Mikko Aromaa b, Jyrki M. Mäkelä b, Milena Stepien c, Jarkko J. Saarinen c, Martti Toivakka c
Method 202 Dry Impinger Method for Determining Condensable Particulate Emissions from Stationary Sources. By: Paul Ottenstein
Air Quality Issues in the Rocky Mountain Region PM2.5 Source Testing Method 201A Determination PM 10 and PM 2.5 Emissions from Stationary Sources (Constant Sampling Rate Procedure) Method 202 Dry Impinger
SODIUM CARBOXYMETHYL CELLULOSE
SODIUM CARBOXYMETHYL CELLULOSE Prepared at the 28th JECFA (1984), published in FNP 31/2 (1984) and in FNP 52 (1992). Metals and arsenic specifications revised at the 55 th JECFA (2000). An ADI not specified
Physical & Chemical Properties. Properties
Physical & Chemical Properties Properties Carbon black can be broadly defined as very fine particulate aggregates of carbon possessing an amorphous quasi-graphitic molecular structure. The most significant