Analytical Procedures and Method Validation
|
|
|
- Wendy Miles
- 10 years ago
- Views:
Transcription
1 IPPC Discharges Monitoring Workshop Analytical Procedures and Method Validation Peter Webster Regional Chemist (EPA Cork)
2 Contents... Overview of data quality Selection of suitable Analytical Methods Choosing your method Use of test kits for regulatory reporting Fitness for purpose Method Validation Determining Performance Characteristics and suitability Demystifying detection / reporting limits Determining Measurement Uncertainty
3 Overview of EPA monitoring EPA monitor over 800 IPPC facilities and approx 150 Waste Management facilities annually Urban Wastewater Treatment plants legislation introduced in 2009 will add a further 520 WWTPs by 2012 All of these facilities have a requirement for Self-Monitoring Quality of regulatory data varies significantly between facilities hence the need for this workshop
4 Typical problems observed No documented operating procedures for test methods Limited training on test methods given to WWTP staff In-house Variations from standard/ reference methods Performance characteristics of method not determined Lack of adequate / suitable Quality Control No calibration checks or maintenance of equipment Method sensitivity not adequate to determine compliance Delays in submission of samples to contract labs No checks on competency of external contractors
5 Choosing your test method Suitable for matrix (effluent) Appropriate measurement range Free from likely interferences Ease of application Does it need specialist equipment
6 Choosing your test method contd. Is an approved / regulatory method based on CEN / ISO / AWWA APHA Standard Methods available for use? What will the results be used for? e.g. Regulatory Compliance or Process Control Is the Limit of Quantitation achievable at least 1/10 th of the regulatory ELV Use a standard method / test kit / or develop in-house?
7 Choosing your test method contd. BS ISO 17381: Water Quality Selection and Application of ready-to-use test kit methods for water analysis Useful guide to when test kits may be suitable for use BS 1427:2009 Guide to on-site test methods for the analysis of waters Provides practical guidance on the choice and limitations of test kits available for many routine water and IPPC / wastewater analysis such as COD / BOD / Nutrients etc.
8
9 A (cautionary) word on use of Test Kits Wide range of proprietary Test kits available Convenient for use by non-scientific staff No reagent preparation necessary Can be cost-effective for in-situ or even lab analyses BUT Interferences may not always be well documented Quality of instructions varies significantly May suffer from interferences due to colour / turbidity Limitations need to be appreciated and understood
10 Fit for purpose? The degree to which data produced by a measurement process enables the user to make technically and administratively correct decisions for a stated purpose IUPAC Compendium of Analytical Nomenclature the Orange book ISBN
11 Fit for purpose? What do require the method to be capable of? Performance characteristics should be adequate to meet the needs of the user / Regulator as appropriate Should be specific to the analyte of interest (interferences known) Determined by evaluating actual performance and not just that stated by the supplier / manufacturer Ref: The Fitness for Purpose of Analytical Methods EURACHEM
12
13 Main causes of unfit for purpose results Operator competence (what training provided?) Matrix interferences with your particular sample type Inappropriate test kit employed... Needs to meet your client needs Contamination (of sample and /or equipment) Inappropriate sampling protocol / sample pre-treatment (needs to be agreed with your client) Wrong range of measurement... aim for your critical limit to be % of test kit range
14 Definition of Validation Validation is the confirmation by examination and the provision of objective evidence that the particular requirements for a specific intended use are fulfilled (ISO/IEC 17025:2005 cl ) A process of evaluating method performance and demonstrating that it meets a particular requirement In essence, it is knowing what your method is capable of delivering, particularly at low concentrations
15 Method Validation.. When? When should a method be validated to verify its performance parameters are adequate for use for a particular analytical problem? New method no previous performance history Revised or established method adapted to a new problem When a review of AQC indicates an established method is changing with time When an established method is used in a different laboratory, with different analysts or with different equipment To demonstration of the equivalence between two methods, e.g. a new method vs. a standard method.
16 Principal Components Selectivity and Specificity Accuracy Precision (Within batch and Between batches) Range and Linearity Detection Limit Limit of Quantitation Robustness & System suitability Uncertainty of Measurement INAB Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
17 Selectivity Selectivity and Specificity This is the ability to discriminate between the analyte or species to be determined and other materials in the test sample Specificity The ability to assess unequivocally the analyte in the presence of components that may be expected to be present in the sample such as degradation products Need to be aware of potential interferences...especially for in- house developed methods. These are usually well documented for standard methods
18 Selectivity and Specificity Selectivity and Specificity can be a real problem with gas / liquid chromatographic methods May be necessary to confirmation on more than one column or by use of different eluents Ion Chromatographic analysis for anions / cations offers good separation but there is often interference in early eluting peaks from organic compounds e.g. Acetate / Formate elute just ahead of Fluoride on carbonate eluent IC columns
19 Accuracy Accuracy = Trueness Accuracy is the closeness of agreement between a test result and the accepted reference or true value of the property being measured Can be evaluated as Bias by use of certified reference materials as controls the use of a traceable reference material or material prepared in-house the use of a standard / reference method with little or no systematic error the use of the method when participating in a proficiency testing scheme the use of spiked samples, based on blank or positive samples from assessment of routine AQC Charts
20 Precision = Repeatability Precision Combination of Within batch (Internal) and Between batches variability Influenced by changes in analyst, instrument conditions, reagents etc. Can be assessed (in simplest way) by analysis at least 10 times of known material singly or in replicate. The SD of single measurements gives an estimate of Sb whereas SD of replicate measurements gives and indication of Sw Long term assessment can derived from AQC charts ISO Accuracy (trueness and precision) of measurement methods and results
21 Accuracy vs. Precision Accuracy = Closeness to true value (error = Bias) Precision = Measure of... - Repeatability (Within Batch) - Reproducibility (Between batch) X X X X X X X X Precise but not Accurate Accurate but not Precise Perfection!! Only rarely achieved.
22 Range and Linearity All methods have an upper and lower boundary of applicability and may not be linear over all concentrations E.g. ISE electrodes exhibit non-linearity at ca 10-5 Molar Choose method whose range brackets those found in your samples (or dilute samples to fit range) E.g. COD vials come is a series of ranges. If your effluent is typically 50 there is a greater error (typically ca. 5%) when using a mg/l range tube than when using a mg/l range tube. Also check the spectrometer preset calibration at regular intervals using known standards Check using known standards at appropriate intervals
23 Detection Limits The most controversial part of any Method Validation Can be generally derived in three main ways Use of a multiple of Blank SD within a single batch Simple, Quick but dependant on one analyst set up of instrument May not be reproducible between analysts or over time From slope of calibration (assumes linear / quadratic fit) Calculation based on multiple batches in replicate Based on multiple batches in replicate Robust and generally gives more reliable estimate than above EPA recommended approach
24 Detection Limit... Definitions The lowest concentration that can be measured with reasonable statistical certainty (AOAC) The lowest concentration of analyte in a sample that can be detected, but not necessarily quantified, under the stated conditions of the test (NATA Tech, Note #3) The smallest concentration that can be determined statistically different from a blank at a specified level of confidence (typically 95%). This corresponds to the critical level. (Currie, 1988 Am. Chem. Soc) The true net concentration or amount of the analyte in the material to be analysed which will lead with probability (1-β), to the conclusion that the concentration of the analyte in the analysed material is larger than that of the blank matrix (ISO DIS ) The output signal or value above which it can be affirmed with a stated level of confidence, for example 95 %, that a sample is different from a blank sample containing no determinand of interest. (ISO :2009)
25 Detection Limit... which one? Instrumental detection limit Equates to the smallest signal above background noise that can be detected reliably. Typically 3x Signal/ Noise ratio but this is of little practical value in measurement terms Method Detection Limit (MDL) Min. concentration reportable to 99% CL that analyte > zero. Determined from analysis of a low sample concentration in given matrix. SD x student s t value or 3.14 for 7 measurements. (40 CFR Part 136, Appendix. B) MDL = Criterion of Detection (ASTM)
26 Detection Limits and LoQ (Statistical) Limit of Detection (LoD) The output signal or value above which it can be affirmed with a stated level of confidence, for example 95 %, that a sample is different from a blank sample containing no determinand of interest ISO :2009 LOD = 2 2.t (df,α = 0.05).sw For a one-sided test with 11 degrees of freedom t = and this transposes as LOD = 5.08s w Ref: ISO :2009 Water Quality Guide to analytical quality control for water analysis However Directive 2009/90/EC on WFD criteria cites LoD as only 3s w
27 Speaking softly Speaking more loudly Shouting It is often difficult to understand the concept of detection limit. The following example may help to clarify some of the concepts defined previously. (Source : Wikipedia) Suppose you are at an airport with lots of noise from jets taking off. If the person next to you speaks softly, you will probably not hear them. Their voice is less than the LOD. If they speak a bit louder, you may hear them but it is not possible to be certain of what they are saying and there is still a good chance you may not hear them. Their voice is >LOD but <LOQ. If they shout then you can understand them and take action on what they are saying and there is little chance you will not hear them. Their voice is then >LOD and >LOQ. Likewise, their voice may stay at the same loudness, but if the noise from jets is reduced allowing their voice to become >LOD you will hear them. Detection limits are thus dependent on both their voice (signal intensity ) and background noise (jet noise)
28 Limit of Quantitation / Practical Reporting Limit Practical Reporting Limit / Limit of Quantitation (PRL or LoQ) Stated multiple of the Limit of Detection, for example, 2-3 x LoD, at a concentration of the determinand that can reasonably be determined with an acceptable level of accuracy and precision. The Limit of Quantification can be calculated using an appropriate standard or sample, and may be obtained from the lowest calibration point on the calibration curve (excluding the blank). ISO :2006 Water Quality Vocabulary Part 2 Referred to (but not specifically defined) in ISO as 10s w LoQ / PRL is often assigned as lowest calibration standard in accredited laboratories as this can be shown to have a clearly defined, rather than theoretical, response.
29 MDL extracted from Lachat test method for ortho-p in seawater Standard used 0.25 µm (7.8 µg/l P) Mean 0.29 µm, Sw = µm MDL applied = 0.025µM (0.8 µg/l P) LoD 5.08sw = 1.5 µg/l P PRL = probably 5 µg/l P (low std) Precision extracted from Lachat test method for ortho-p in seawater 5µM (150µg/l P), Mean 4.91 Sw = 0.072, RSD = 1.47% 2µM (60µg/l P) Mean 1.96 Sw = 0.038, RSD = 1.96% These look good but are they repeatable!
30 Response What these Limits mean in practice LoQ / PRL Limit of Linearity (LoL) IDL LoD MDL Useful measurement range = LoL - LoQ Concentration
31 Robustness / System suitability The analytical method should not be sensitive to small changes in procedures e.g. flow rates / reagents etc. Identify any variables which may have an effect of data and set up experiments using known materials to determine effects Set performance criteria for methods to achieve based on e.g. Minimum slope / Lowest standard response / Sensitivity etc. Robust methods will generally produce similar results for the same sample when used in independent laboratories. Standard methods will generally fall into this category hence their recommended use Ref: The Fitness for Purpose of Analytical Methods EURACHEM
32 Uncertainty of Measurement No analytical measurement is an absolute value... They all have an uncertainty which is influenced by environmental / physical factors! Uncertainty can be calculated from a metrology perspective taking account of all contributory factors but this is long winded and not readily applicable to chemical analysis It is much simpler to use the information you will (hopefully) already have to determine UM in the event you require it.
33 Uncertainty of measurement references The GUM Bible... An extensive reference but fairly unreadable even if you have a degree in statistics! EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Better Nordtest Report TRV 537 Handbook for calculation of Measurement Uncertainty in Environmental Laboratories Ed.2 Nordic Committee for Food Analysis (NMKL) Estimation and Expression of Measurement Uncertainty in Chemical analysis A Beginners Guide to Uncertainty of Measurement (NPL)
34 But best (in my opinion) is the new Draft Std. ISO This is a very readable document which gives clear examples of UM as it relates to laboratory data
35 Uncertainty of Measurement - demystified Assess using routine control chart precision and bias data Use information from reference materials, inter-lab performance tests, % recovery data Each input is additive Several good worked examples in ISO UoM (MU) is important from a regulatory perspective because.
36 Interpreting exceedences Assessing whether a value exceeds a regulatory limit requires consideration of the Uncertainty of Measurement associated with the result(s). Assessment based on single measurements (most of our data) is difficult since there is no information on the range of repeatability of the actual measurement. Variation has often to be inferred from historic data from replicate analysis / QC data.. Replicate measurements are better in difficult situations. Limit value C D Small error uncertainty, Definitely exceeding at 95% confidence. Would be likely to merit some enforcement action if e.g. >1.5 x limit A B Numerically > limit and excedence is more probable than compliance but not at > 95% confidence. Would merit notification by EPA and action by licensee to reduce mean value Probably compliant but confidence < 95%. Large error uncertainty. Small probability on non-compliance. May necessitate corrective action to reduce mean value Definitely within limit at 95% confidence even with large error uncertainty. System OK
37 Determining Performance Characteristics The EPAs preferred approach to Method Validation, particularly for anyone involved in regulatory Drinking Waters analysis, is outlined in the EPA Handbook on the Implementation of the Regulations (...the Drinking Water Regs. 2007) for Water Services Authorities and Private Water Supplies... Section 5
38 Section 5 explained Based on WRc publication NS30 (M. Gardner / A. Wilson ) Blank, Low Standard (20%), High Standard (80%), Low sample, Spiked sample (+ ca. 50%) Analyse in Duplicate over 11 batches over several days with as many analysts as possible to reflect lab practices Calculate the Standard deviations and Recovery from each relevant matrix e.g. Waters, Sewage, Trade wastes etc. EPA defines LoD as 3x S w of low sample or... 5 x S w of the blank solution (not calibration blank) Recommends the use of Shewart Control Charts and rules
39 Control Charts In-house Standards / Certified Reference materials Use of Mean and/or Mean / Range charts Log Difference charts for microbiology (especially if using Lenticules)
40 What performance is acceptable? How long is a piece of string?... It really depends on the use to which the data will be put. General Rule of Thumb is aim for a Maximum Error Threshold of 20% comprising Bias (10%) and Std. Dev. (5%) Most analytical methods will meet the Bias comfortably but some will may have difficulty with the Std. Dev. Threshold is typically 30% for trace organics (10% SD) Parameter Method % Uncert Parameter Method % Uncert Parameter Method % Uncert COD B1 5.1 TOC B Silicate (300) B B2 3.2 Ammonia B Silicate (3000) 13.3 PH B Nitrite B Salinity B Conductivity B4 1.8 Phosphate 9.8 Ammonia B BOD B Ammonia 10.3 Nitrite B Alkalinity B6 1.5 Total Solids B Phosphorus B Suspended. Solids B7 7.2 TDS 3.7 Chloride B48 3.3
41 Useful References NS30 A Manual of Analytical Quality Control for the Water Industry, WRc ISO/TR 17944:2004 Water Quality Criteria for establishing the equivalence between microbiological methods. UK Drinking Water Inspectorate Guidance Note on Interpretation of Aspects of AQC ISO Water quality Guidance on analytical quality control for chemical and physicochemical water analysis Environment Agency (UK) Performance Standard for Organisations Undertaking Sampling and Chemical Testing of Water (v1.1 May 2009) Environment Agency (UK) Monitoring of Discharges to Water and Sewer (TGN M18)
42 Extract from Appendix B of EA s Performance Standard document
43 Conclusion of above data table Precision for sewage analysis is considerably poorer than the 5% target so an F-test is performed using a critical concentration... In this case 5 mg/l. An absolute target of mg/l was applied rather than an RSD of 5% (equivalent to an SD of only mg/l)
44 Summary of performance Performance characteristics for the Trade effluent sample are more variable and recovery is only just acceptable at 85%. It may be necessary to undertake some further checks on method interferences to see whether this can be improved.
45 Performance Assessment Example: Environment Agency Batch Replicate Sweage Efflt Spiked Sewage Recovery Sewage Recovery % Trade Effflt Spiked T.Efflt Recovery Trade Efflt Recovery % (+ 5 mg/l) (+15mg/l) Batch Mean Batch SD Variance Batch Mean Batch SD Variance Batch Mean Batch SD Variance Batch Mean Batch SD Variance Batch Mean Batch SD Variance Batch Mean Batch SD Variance Batch Mean Batch SD Variance Batch Mean Batch SD
46 Too complicated? Not really... It s worth the effort but very time consuming if doing it even using on Excel Simpler to purchase software to do the job e.g. mval (from NMS) htpp:// amid=358 (NMS publications page)
47 Thank you
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
American Association for Laboratory Accreditation
Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
Signal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
ANALYTICAL DETECTION LIMIT GUIDANCE. & Laboratory Guide for Determining Method Detection Limits
ANALYTICAL DETECTION LIMIT GUIDANCE & Laboratory Guide for Determining Method Detection Limits Wisconsin Department of Natural Resources Laboratory Certification Program April 1996 PUBL-TS-056-96 introduction
LAB 37 Edition 3 June 2013
LAB 37 Edition 3 June 2013 Accreditation Requirements for Sampling and Testing in Accordance with the Drinking Water Testing Specification (DWTS) CONTENTS SECTION PAGE 1 Introduction 2 2 Regulatory requirements
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters. Environment Agency July 2015 Version 2
Performance Standard for Organisations Undertaking Radioanalytical Testing of Environmental and Waste Waters Environment Agency July 2015 Version 2 Foreword We set up our Monitoring Certification Scheme
Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories
Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
ALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE
QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014 1 TABLE OF CONTENTS FOREWORD... 1 1.0 SCOPE... 1 1.5 Safety Requirements... 1 4.0 MANAGEMENT REQUIREMENTS... 1 4.1
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample
1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.
Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description
Analytical Methods for Cleaning Validation
Available online at www.scholarsresearchlibrary.com Scholars Research Library Der Pharmacia Lettre, 2011, 3 (6): 232-239 (http://scholarsresearchlibrary.com/archive.html) ISSN 0975-5071 USA CODEN: DPLEB4
integrity without compromise Environmental Monitoring
integrity without compromise Environmental Monitoring CLS is the leading provider of sampling and analysis to the food, environmental, pharmaceutical and medical device industries. Founded in 1994, CLS
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
Quality Assurance (QA) & Quality Control (QC)
Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
A commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters
Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters 1. SCOPE and APPLICATION 1.1 Gravimetric analysis is used to determine total suspended
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
Laboratory analysis for wastewater treatment, industrial monitoring, and environmental protection
Laboratory analysis for wastewater treatment, industrial monitoring, and environmental protection The man who started it all Dr. W.D. Hatfield was an innovator in wastewater treatment technologies and
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
Validating Methods using Waters Empower TM 2 Method. Validation. Manager
Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information
ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:
How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
General and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
DATA QUALITY OBJECTIVES
The Wastewater Treatment Plant Operators Guide to Biosolids Sampling Plans CHAPTER 5 DATA QUALITY OBJECTIVES Goals of the Sampling Plan ESSENTIAL ELEMENTS OF A SAMPLING PLAN Description of the Facility
STANDARD OPERATING PROCEDURES
PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
Introduction. The following definitions may help you better understand the components of the data report.
Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will
CLIENT NAME: MISC AGAT CLIENT ON, ON (403) ATTENTION TO: Chantell Cambal
CLIENT NAME: MISC AGAT CLIENT ON, ON (403) WATER ANALYSIS REVIEWED BY: Anthony Dapaah, PhD (Chem), Inorganic Lab Manager DATE REPORTED: May 19, 2015 PAGES (INCLUDING COVER): 5 VERSION*: 1 Should you require
18.6.1 Terms concerned with internal quality control procedures
18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable
Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory
Customer Application Note: 114 Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory Sébastien N. Ronkart, Ph.D.; Société wallonne des eaux, rue de la
LC-MS/MS, the new reference method for mycotoxin analysis
LC-MS/MS, the new reference method for mycotoxin analysis What is necessary for the implementation of the multi-residue mycotoxin method in an analytical laboratory? Azel Swemmer [email protected] AFMA
Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water
Online SPE-LC-APCI-MS/MS for the Determination of Steroidal ormones in Drinking Water presented by Paul Fayad, Ph.D. Candidate under the supervision of Sébastien Sauvé, Ph.D. Department of Chemistry [email protected]
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
Dissolved and precipitated oxalate
Accepted 2005 Process liquors from bleach plants Dissolved and precipitated oxalate Using Ion Chromatography 0 Introduction In bleach plants of pulp mills with a high degree of system closure, there is
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team
Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.
0(752/2*
ph Alkalinity of Water
ph Alkalinity of Water DOC316.52.93085 Based on ISO standard 9963-1:1994 ph-metric Titration 0.4 to 20 mmol/l of Total Alkalinity 1. Introduction Alkalinity of water is its acid-neutralizing capacity.
Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.
Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Quality Assurance/Quality Control in Acid Deposition Monitoring
Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial
Quantifying Uncertainty in Analytical Measurement
EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition QUAM:000.P1 EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Editors S L R Ellison
Approvals [Signature On File]
Page 1 of 7 Minimum Reporting Requirements for Analytical Data (Chemistry) for the Water Quality Planning Bureau Approvals SOP WQPBDMS-010 Prepared by: Chris Shirley, QA Officer, WQPB Michael Pipp, Environmental
Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity
For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,
IAEA-TECDOC-1350. Development and use of reference materials and quality control materials
IAEA-TECDOC-1350 Development and use of reference materials and quality control materials April 2003 The originating Section of this publication in the IAEA was: Industrial Applications and Chemistry Section
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES October 2008 Revision 08 Great Lakes Institute for Environmental Research University of Windsor Windsor, Ontario
Data Quality. Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program
Data Quality Tips for getting it, keeping it, proving it! Central Coast Ambient Monitoring Program Why do we care? Program goals are key what do you want to do with the data? Data utility increases directly
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014 This experiment explores quality assurance practices which are commonly
Medical Cannabis Laboratory Approval Program
Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
S olid Ph ase Extraction (SP E) Method:
Solid Phase Extraction of US EPA 535 for Chloroacetanilide and Acetamide Degradates in Drinking Water Samples Prior to Liquid Chromatography/ Tandem Mass Spectrometry Toni Hofhine, Horizon Technology,
EffiValidation 3.0 software basic training module
EffiValidation 3.0 software basic training module Oct 3, 2012 www.effichem.com Agenda EffiChem company EffiValidation 3.0 software : overview International standards and guidelines Software features Structure
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites
Policy on Monitoring of Stack Emissions to Air at EPA Licensed Sites 1. OVERVIEW In Ireland at present there are no mandatory minimum criteria specified in relation to the completion of stationary source
Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
DHI DANETV Test Centre Quality Manual
Purpose This quality manual has been prepared to comply with the requirements from the EU ETV pilot programme and the international standard for inspection, ISO/IEC 17020. The quality manual is integrated
M20. Technical Guidance Note (Monitoring) Quality assurance of continuous emission monitoring systems - application of EN 14181 and BS EN 13284-2
linear Technical Guidance Note (Monitoring) M20 Quality assurance of continuous emission monitoring systems - application of EN 14181 and BS EN 13284-2 Environment Agency Version 3 June 2015 Foreword This
Standard methods in water analysis
Branch General analytical laboratories; water analysis Keywords Water analysis; standard methods; ASTM; DIN; ISO; USP; EPA; SLMB; EN; SCA; titration; ion chromatography; voltammetry; branch 1; branch 2
Oasis HLB Cartridges and 96-Well Plates
CONTENTS I. INTRODUCTION II. SAMPLE PRE-TREATMENT a. Biological Samples b. Solid Samples: Soil, Whole Foods, Tissue c. Aqueous Samples: Water, Beverages d. Non-Aqueous Liquid III. SOLID PHASE EXTRACTION
Canadian Environmental Technology Verification Program Applicant Information Package CONTENTS
Canadian Environmental Technology Verification Program CONTENTS Contents... 1 Overview... 2 Performance Verification... 2 Application Information... 2 How much does verification cost?... 2 What is the
EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM. LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007)
EPA NATIONAL LEAD LABORATORY ACCREDITATION PROGRAM LABORATORY QUALITY SYSTEM REQUIREMENTS (LQSR) REVISION 3.0 (November 05, 2007) The requirements in this document apply to all lead testing laboratories
The Determination of Uncertainties in Charpy Impact Testing
Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 06 The Determination of Uncertainties in Charpy Impact Testing M.A. Lont
Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
DRAFT Changes to 25 Pa. Code Chapter 252 For Discussion Purposes Only January 21, 2015 CHAPTER 252. ENVIRONMENTAL LABORATORY ACCREDITATION
CHAPTER 252. ENVIRONMENTAL LABORATORY ACCREDITATION Subchapter A. GENERAL PROVISIONS B. APPLICATION, FEES AND SUPPORTING DOCUMENTS C. GENERAL STANDARDS FOR ACCREDITATION D. QUALITY ASSURANCE AND QUALITY
Magruder Statistics & Data Analysis
Magruder Statistics & Data Analysis Caution: There will be Equations! Based Closely On: Program Model The International Harmonized Protocol for the Proficiency Testing of Analytical Laboratories, 2006
CHAPTER ONE TABLE OF CONTENTS 1.0 INTRODUCTION... 1
CHAPTER ONE TABLE OF CONTENTS Section Page 1.0 INTRODUCTION... 1 2.0 QA PROJECT PLAN... 1 2.1 DATA QUALITY OBJECTIVES... 2 2.2 PROJECT OBJECTIVES... 2 2.3 SAMPLE COLLECTION... 3 2.4 ANALYSIS AND TESTING...
State Water Resources Control Board Environmental Laboratory Accreditation Program
State Water Resources Control Board Environmental Laboratory Accreditation Program P.O. Box 100, Floor 24 850 Marina Bay Parkway, Build. P 500 N Central Ave, Suite 500 Sacramento, CA 95812-0100 Richmond,
The Fitness for Purpose of Analytical Methods
The Fitness for Purpose of Analytical Methods EURACHEM Guide The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method Validation and Related Topics This document has been developed by