Regulations for the Development of Medical Device Software. Master of Science Thesis ANDREAS MAGNUSSON
|
|
|
- Leslie Ross
- 10 years ago
- Views:
Transcription
1 Regulations for the Development of Medical Device Software Master of Science Thesis ANDREAS MAGNUSSON Department of Signal and Systems Division of Electrical Engineering CHALMERS UNIVERSITY OF TECHNOLOGY Gothenburg, Sweden, 2011 Report No. EX004/2012
2 Regulations for the Development of Medical Device Software ANDREAS MAGNUSSON ANDREAS MAGNUSSON 2012 Technical report no. EX004/2012 Department of Signal and Systems Chalmers University of Technology SE Göteborg Sweden Telephone +46 (0)
3 Acknowledgments I would like to thank Integrum AB for giving me this opportunity to assist them in their expansion into medical software development and for their patience with me throughout this process. I have learned a lot through this project concerning medical software development and the regulatory requirements that a medical device manufacturing and developing company faces. I would specically want to thank my supervisor, Max. J. Ortiz C., for his input, support and assistance during this project. I would also want to thank my examiner, Professor Bo Håkansson (Department of Signal and Systems at Chalmers University of Technology), for having me as his student and for his assistance and input during the project. Lastly, I would like to thank my family and friends for their continued support during the project. 3
4 Abstract Today, most medical devices contain software to some extent. As the use of software in medical devices has increased, a need to govern the development of the software used in these devices has been identied. In order to ensure that all medical devices that are released does not present any unnecessary risk to the end user, regulatory requirements has been created that govern how these medical devices should be developed. When developing software that is to be used as part of a medical device, or software that is a medical device on its own, there are specic medical software regulatory requirements that need to be considered. This thesis is about the adaptation of such regulatory requirements into a software development procedure that is to be implemented into the quality management system at Integrum AB. The result of the thesis is the development of a set of software development procedures in order to assure compliance with the regulatory requirements of IEC Medical device software - Software life cycle processes.
5 Abbreviations CDRH Center for Device and Radiological Health CFR Code of Federal Regulations CVS Concurrent Versions System EC European Commission FDA Food and Drug Administration IDE Integrated Development Environment IEC International Electrotechnical Commission ISO International Organization for Standardization MDD Medical Device Directive NCAL-Soft Natural Control of Articial Limbs Software OHMG Osseointegrated Human-Machine Gateway OPRA Osseointegrated Prostheses for the Rehabilitation of Amputees OPRA-NCAL Osseointegrated Prostheses for the Rehabilitation of Amputees with Natural Control of Articial Limbs PMA Premarket Approval SOP Standard Operating Procedure SDE Software Development Environment SVN Apache Subversive QMS Quality Management System 1
6 Contents 1 Introduction Purpose Scope Integrum AB Medical Device Regulations Background Device Regulations in the US Device Regulations in the EU Medical Software Development Background Software Validation and Verication Quality Planning Software Safety Classication Software Requirements Software Design Software Coding Software Testing Software Maintenance and Software Changes Software Life Cycle Development Models Waterfall Development Model Iterative and Incremental Development Model Agile Development Model O-the-Shelf Software Software of Unknown Pedigree Methods Literature Studies Requirements Identication Verication Problem Description Problem Description Design and Implementation The OPRA-NCAL Project OPRA-NCAL Classication According to European Union Denition of medical device Denition of accessory Medical software
7 Conclusion OPRA-NCAL Classication According to the United States of America Denition of medical device Conclusion NCAL-Soft Classication of NCAL-Soft Conclusion Evaluation of Safety Classication Reduction NCAL-Soft Intended Use Risk Assessment Software Safety Classication Software Development Environment Software Development Procedure Proposed Software Development Procedure Creating a Software Development Procedure Verication Verication of Performance Verication of Compliance Discussion NCAL-Soft Safety Classication Software Development Procedure Future Work
8 Chapter 1 Introduction Development of software for medical devices are controlled through international regulatory requirements dened by regulatory standards, such as International Organization for Standardization (ISO), or International Electrotechnical Commission (IEC). As with other devices intended for medical applications, regulatory requirements for medical software are centered around risk analysis. This is to ensure that the software does not present any risk of harming the user(s). It is also to ensure that identiable risks are properly controlled to minimize the potential harm, in the event of an accident. 1.1 Purpose The purpose of this masters thesis is to establish a medical software development procedure that complies with current medical device software regulatory requirements and implement this procedure into the quality management system at Integrum AB. 1.2 Scope The following activities make up the basis for this thesis ˆ A study of medical software development activities ˆ A study of commonly used software development models in order to properly implement development activities into a quality management system ˆ Determination of medical software safety classication based on a preliminary hazard analysis ˆ Implementation of regulatory requirements into an existing quality management system through a medical software development procedure 4
9 1.3 Integrum AB Integrum AB is a Swedish company that develops implant systems for bone anchored amputation prostheses and was founded by Dr. Rickard Brånemark in 1998 at Sahlgrenska Science Park in Gothenburg, Sweden. The implant system, developed by Integrum AB, consists of a titanium device that is surgically inserted into the bone, protruding through the skin of the amputation stump. A prosthesis can easily be attached to the anchorage by the patient, giving no direct contact with the skin. This solution improves functionality and comfort to the patient compared to the traditionally used socket prostheses. It implies reduced problems with sores, pain and pressure, as well as easier attachment and detachment, improved stability and better walking ability. Integrum's implant system is called OPRA (Osseointegrated Prostheses for the Rehabilitation of Amputees). The ability of titanium to integrate with bone is called osseointegration. In this process, bone tissue is formed around the implant causing the bone tissue and surface of the implant to anchor together. The concept of osseointegration was presented in 1977 by Per-Ingvar Brånemark, who has developed the technique primarily for prosthetic replacement of teeth, but also for orthopedic applications. 5
10 Chapter 2 Medical Device Regulations Background All devices that are used on the market have gone through some level of testing, depending on the type of device, to verify that the device is safe to use. For medical devices, these tests are stricter as a medical device can be in direct contact with a patient or be located inside a patient, such as a pacemaker. Depending on the type of medical device, there are dierent regulations which govern what tests are needed to be conducted and the level of documentation needed to prove that the device is safe to use. The regulations also govern that the device's intended use is such that it means an improvement of life quality in order to validate the potential risks inherited with the device. 2.1 Device Regulations in the US Within the United States (US), the governing agency for medical devices is Food and Drug Administration (FDA) within the Department of Health and Human Services. Within the FDA, the Center for Devices and Radiological Health (CDRH) handles the regulatory concerns for companies who manufacture, repackage, relabel and/or import medical devices which are to be sold within the US. Any regulation for medical devices is published in the Federal Register under the Code of Federal Regulations Title 21 Ÿ820. The FDA has created three regulatory classes which are based on the level of control necessary to assure the safety and eectiveness of the device, which are Class I, Class II or Class III, with Class III being for medical device with the highest risk. A medical device classied as Class I or Class II will require a premarket submission or 510(k) whilst a Class III medical device is required to get a premarket approval unless special circumstances are given, for which only a 510(k) is needed [1]. A 510(k) is a premarket submission which demonstrate that the medical is at least as safe and eective as a medical device already on the market and that is not subject to a Premarket Approval (PMA) [2]. When needing to do a 510(k), the medical device manufacturer is required to submit documentation to the FDA. The submitted documentation must prove that the device is at least as safe and eective as a similar device which is already on the market [2]. 6
11 2.2 Device Regulations in the EU In the European Union (EU) the governing organ for legislation's is the European Commission (EC). The EC publishes directives which serves as the regulatory framework, for medical devices this directive is simply called the Medical Device Directive (MDD). The MDD divides medical devices into either Class I, Class IIa, Class IIb or Class III, with Class III being for medical devices with the highest risk. In addition, there are a number of notied bodies which acts as a third party controller for medical devices. These notied bodies ensure that the medical devices that are released on the market complies with the MDD as well as other applicable directives through quality system inspections, certication and classication assessments. The notied bodies also review the technical documentation for the medical devices. It is also the notied bodies that issue the CE-marking, which is required in order to get a product onto the European market [3]. 7
12 Chapter 3 Medical Software Development Background As mentioned in the previous section, all medical devices are subjected to regulations intended to ensure safety for the operator and patient. As the use of software in medical devices has increased, the need for specic regulations for medical device software has increased. An indicator of this need was shown by an analysis performed by the FDA between 1992 and 1998 of 3140 medical device recalls. The analysis revealed that 242 (or 7.7%) of the recalls could be attributed to software failures. Out of those recalls, 192 (or 79%) were caused by software defects which had been introduced when changes were made to the software after its initial production and distribution [4]. Another example is from the mid 1980's, where coding errors in a radiation therapy device contributed to the lethal overdose of a number of patients [5]. In 2006, a new international standard was released for medical device software and has been developed by a joint working group from IEC and ISO[12]. This new standard has, since its release, become fully harmonized in both the EU and the US. 3.1 Software Validation and Verication When developing software, the software validation and software verication are the two main elements to achieve acceptance with. Where software validation is the activity of proving that the software product does what it is stated by the intended use. Software verication is what is continuously performed throughout the development process by performing testing to ensure that the software code is working properly. At the end of the software development process, given that a sucient level of software testing (verication) has been conducted, the software product can be viewed as validated through the verication eort. The process of software validation is a requirement of the quality system regulation, which was published by the FDA in the Federal Register on October 7, 1996 and took eect on June 1, The main component of software validation is what is described as the software validation plan in [4], or as the software development plan in [12], which 8
13 species areas within the development process such as scope, approach, resources, schedules and types and extent of activities, tasks, and work items. In the following sections, I will go over some of the main components of the software development plan and what they require from a developer of medical software Quality Planning The software development plan is, as mentioned above, not the only component of the software validation. The software validation is handled through, what is called, Quality Planning and this includes plans for risk management, conguration management and problem resolution as well as the development plan. The risk management for medical software is governed by ISO Medical devices - Application of risk management to medical devices and additional regulatory requirements dened in IEC Medical device software - Software life cycle processes. These standards handles the identication and evaluation of possible hazards which are documented in a risk analysis document along with possible risk control measures. The conguration management plan should guide and control multiple parallel development activities and ensure proper communications and documentation. There should also be controls in place to ensure that all approved versions of the specications documents, source code, object code, and test suites that comprise the software system is kept updated and correct. The problem resolution procedure handles all reporting and evaluation of anomalies found during the validation or after the software product has been released as well as the resolution of identied anomalies. Both the conguration management and the problem resolution is governed by IEC Medical device software - Software life cycle processes. Finally, the software development plan handles the entire development of the software product and connects to the plans described above at dierent stages throughout the development process Software Safety Classication As with standard medical devices, software embedded in medical devices or software that are medical devices in their own right are subject to safety classication. The regulatory requirements that a software development plan must comply with is determined by the software safety classication of the software product. The classications and the supporting denitions are: ˆ Class A - No injury or damage to health is possible ˆ Class B - Non-serious injury is possible ˆ Class C - Death or serious injury is possible where serious injury is dened as an injury or illness that: ˆ directly or indirectly is life threatening, ˆ results in permanent impairment of a body function or permanent damage to a body structure, or 9
14 ˆ necessitates medical or surgical intervention to prevent permanent impairment of a body function or permanent damage to a body structure The hazardous situations are determined through a risk analysis of the software, until a risk analysis has been performed the software is to be considered as a Class C medical software. A medical software, that is composed of multiple parts, is classied based on the part with the highest potential hazard. Each separate part of the medical software, given that sucient segregation can be shown, can be classied individually so that only the high risk software parts will require the extra verication eort. The software safety classication should be performed at the start of the development and repeated as the software design and requirements change and/or are implemented. There are ways for a developer to reduce the safety classication through risk control measures, but these must be performed through hardware modications or additions. Risk control measures that are done through additional software functions or user instructions are not sucient to reduce the safety classication as they are deemed less eective and secure Software Requirements At the start of software development, there is a need to dene system requirements based on the intended use of the software. These initial system requirements may change during the development process, but they will help in creating a good framework for which the software is to be kept within. Based on the system requirements, sets of software requirements needs to be dened that de- nes how the system requirements is to be implemented into source code. The software requirements will develop as new performance or system requirements are identied, analysis of the software requirements are conducted, and as the intended use for the software becomes more detailed. All system and software requirements is to be documented in a software requirements specication document. Software requirements specication should include all the software system inputs and outputs as well as all the functions that the software system will perform. Every performance requirement for the software, such as reliability, data throughput, timing etc), should also be included. Within the software requirements, it should also be mentioned how users will interact with the system, what is classied as an error and how errors should be handled. In extension, the intended operating environment for the software system, such as hardware platform or operating system, should be stated if this is a design constraint. A software requirements traceability analysis should also be conducted to trace software requirements to (and from) system requirements and to risk analysis results. Before proceeding, the software requirements needs to be evaluated to verify that: ˆ There are no internal inconsistencies among the requirements; ˆ All of the performance requirements for the system have been spelled out; ˆ Fault tolerance, safety and security requirements are complete and correct; 10
15 ˆ Software requirements are appropriate for the system hazards; and ˆ All of the requirements are expressed in terms that are measurable or objectively veriable Software Design Once the software requirement specication has been created, the next step is to transform the requirements into an architectural design of the software. The purpose of the design specication is to describe what the software should do and how it should do it. The design specication should aim to reduce the probability of use errors due to the software design being to complex or act contrary to the user's intuitive expectations. The software requirements as well as coding guidelines and the development procedure should be included in the software design specication. It should also include information concerning any supporting software, such as operating system, drivers or other application software as well as which hardware is to be used. Systems documentation describing the context in which the program is intended to function should also be included along with a risk analysis for the software system. Additionally, information concerning error, alarm, and warning messages should be mentioned as well as what security measures (physical and logical) that have been implemented. If the software does not include alarm or warning messages, it is recommended that this be mentioned as well. Once a software design specication has been created, an evaluation is needed to verify that the design is complete, correct and possible to be maintained. A traceability analysis should also be conducted as part of the evaluation to verify that all of the software requirements have been implemented in the software design. As the software design specication is based on the software requirements specication, the design specication is subject to change during the software development as the requirements are updated Software Coding The software coding activity is where the software programmers translate the design specications into source code, or assemble previously coded software components from code libraries or from o-the-shelf software (described further in section 3.3) for use in the new software. Prior to starting the coding activity, decisions are to be made concerning the selection of which programming language to be used as well as which software development tools, such as assemblers and compilers, should be used. When selecting either which language to use or which tools to use, consideration needs be taken to the impact on the future testing and evaluation activities that will follow. The coding also needs to follow a pre-dened coding standard to ensure that all developed source code is uniform. Before implementing developed source code into the software system, a thorough level of error checking is needed to be performed to minimize the amount of residual errors. For every encountered error that is left unresolved a rationale must be documented with a risk evaluation of the error. 11
16 To aid in the error checking and tracing down errors, code comments need to provide useful and descriptive information such as expected inputs and outputs, variables referenced, expected data types and operations performed. The code comments should also be used to aid with the source code traceability analysis. The source code traceability analysis veries that all requirements has been implemented in the code and that all the modules and functions implemented in the code can be traced back to a requirement as well as to the risk analysis. The same traceability analysis should also be documented for the tests conducted to verify that the modules and functions work as intended Software Testing As touched upon in the previous section, in order to verify that the software program functions according to its intended use and design specications, testing is needed to be performed. The extent of the tests is determined largely by the software classication, as software which could cause death or serious injury if malfunctioning has to have a higher integrity than software found in an electronic thermometer. While in development, the software program will go through an iterative process which includes testing of the actual code to verify that the software behaves according to the predetermined design requirements. The testing itself consists of running the software under known conditions with documented inputs and outcomes that can be compared to their predened expectations. It is important to realize however, that software cannot be tested for every possible input as it is a complex system. In order to achieve full coverage of every possible input/output combination, the time requirement would be much to long for a sustainable development from a nancial aspect. Instead priority should be placed on ensuring that all the critical functionalities and safety related functionalities are suciently tested. Further, as mentioned in General Principals of Software Validation [4], testing of all program functionality does not mean all of the program has been tested, neither does testing all program functionality and all program code mean that the program is 100% tested. It just means that no errors were found using that set of tests, which could be due to that the testing in itself was supercial. As a software developer, it is important to create a software test plan that include the particular testing activities that are to be performed during each stage of the development process. These selected testing activities should be directed to verify specied cases based on the software product's internal structure and external specication. They should also provide a thorough examination of the software product's compliance with its functional, performance and interface denitions and requirements. The software test plan should also include justication for the level of eort represented by the selected testing activities and the corresponding acceptance criteria. Following the software testing performed during development, the software needs to be tested in the actual environment where it is intended to be used. This can be done either through simulation or actual use of the software by the intended end users. When conducting user site testing, there should be a pre-dened plan containing a summary of testing and a record of acceptance. Documentation of testing procedures, test results and test input data should also be retained 12
17 along with documentation proving that the software and supporting hardware is installed and congured as specied in accompanying documents of the software. It should also be ensured that all software system components are tested and that these are of the correct version. The user site testing plan should specify the full range of operating conditions for which the software is to be tested under as well as specify a duration suciently long as to allow possible conditions or events to occur that could trigger detection of previously undetected bugs. There should also be an evaluation performed on the end users ability to understand and correctly use the software Software Maintenance and Software Changes Following the software testing activity, changes are commonly needed to be made to the software due to identied errors. As mentioned previously in chapter 3.1.1, the activity of software changes is handled by the conguration management plan. As part of the conguration management, all potential software changes needs to be evaluated to determine if the change is necessary and how the change will aect the software as well as the organization. Once a decision has been made, the software change is either implemented into the software code or the rationale for why the change was not implemented is documented as part of the risk management documentation. Software maintenance, which handles the post-release activity of software development, is also part of the conguration management. Software maintenance primarily handles any received feedback, either from end users or internal, and performing evaluation eorts to analyze the feedback to determine the response to the feedback. If required, changes to the released software will be performed and depending on the nature of the changes, it is classied as either corrective, perfective or adaptive maintenance. Changes made to correct errors and faults in the software are corrective maintenance. Changes made to the software to improve the performance, maintainability, or other attributes of the software system are perfective maintenance. Software changes to make the software system usable in a changed environment are adaptive maintenance. When changes are made to a software system, either during initial development or during post release maintenance, sucient regression analysis and testing should be conducted to demonstrate that portions of the software not involved in the change were not adversely impacted. 3.2 Software Life Cycle Development Models The activities that make up the software development plan can be used dierently depending on the approach of the developer. These dierent approaches are commonly expressed as development models or life cycle models. This section will briey cover some of the more common software development models though there are several more development models available, such as Evolutionary Systems Development and Spiral Development, but these are not the focus of this report. 13
18 3.2.1 Waterfall Development Model The Waterfall development model is credited to Dr. Winston W. Royce due to his article Managing the Development of Large Software Systems [7] and originates from an earlier development model called the Stagewise model. The Waterfall development model uses a linear approach, where each activity is only performed once and a project may not proceed into the next activity until the current activity has been completed. Figure 3.1 shows a graphical illustration of a basic Waterfall development model. Although it does not represent the nished development model proposed by Dr. Royce, it gives a good indication to how the Waterfall model is being implemented. Figure 3.1: Graphical illustration of the Waterfall model [7] For a software development, this kind of model is not optimal, as the requirements (both system and software) can change during the development process. The software design can also change which means that there is a direct need for a more iterative approach to ensure that both the requirements fully describe the software design but also that the software design fully describes the requirements. The nished model described in Managing the Development of Large Software Systems [7], includes feedback loops between the dierent phases to correct initial errors as well as a Do it twice methodology Iterative and Incremental Development Model The development model described by Dr. Royce is closer to how the Iterative and Incremental Development (IID) model is designed[8]. At the start of an IID project there is a need to subdivide the project into binary deliverables. This is done by performing a high level analysis of the project with the output of slices, or increments. Each slice should meet a number of criterias, such as; 14
19 ˆ implement a use case, or part of a use case, ˆ represent features, ˆ be executable and demonstrable, and ˆ have completion and acceptance criteria. as described in Iterative and Incremental Development [8]. Once the project has been subdivided into slices, each slice is developed in isolation from the other slices until all slices have been completed. Figure 3.2 shows a graphical illustration of the Iterative and Incremental development model, where the project is iteratively improved through testing, re-planning, re-analyzes and evaluations until the criterias have been met for the program to be deployed. The Initial Planning shown in the gure represents the initial subdivision of the project and other, company specic, activities (such as nancial analyses). The iterative process represents the development of each slice. Figure 3.2: Graphical illustration of the Iterative and Incremental model [14] However, this development model does invite the developer to stay too long in the iterative process to reach a perfect program. But releasing software without any bugs is near impossible which would lead to the software staying in development for too long and delivery dates being missed. As such, it is important to recognize that not all bugs has to be xed for the software to be released and not to set the completion criterias to high. A well conducted risk analysis during the development will help identify which bugs are required to be xed and which that can be left for a later software update Agile Development Model The Agile development model is a continuation of the IID model and the most commonly used Agile development model is called SCRUM [9]. Due to this, I will here describe SCRUM as an example of Agile development. Similar to how the IID divided the project into slices, the Scrum model uses Sprints which commonly last for 2-4 weeks. During these sprints, the development is carried out like the model shown in gure 3.2. Each day within 15
20 these sprints also follow this model, leading to nestled iterative development, which is depicted in gure 3.3. Figure 3.3: Graphical illustration of SCRUM [13] The daily iteration starts with a SCRUM meeting where the team of developers go over what each of them will be doing during the day and bring up if there are any blockers that prevent them from performing their assigned tasks. This ensures that everyone within the developing team knows what the rest are doing and resources can be allocated to quickly deal with blockers to minimize possible delays in development. The product backlog contains the requirements that should be implemented into the nal product whilst the sprint backlog contain the dierent functionalities or tasks that should be completed during that sprint to achieve a working increment of the nal software product. This gives the project leader a complete overview of the development and makes it easier to get a good sense of the progress. 3.3 O-the-Shelf Software Manufacturers of medical devices that use software can choose to buy commercially available or open-source software instead of developing the needed software themselves. The software, which is not developed by the manufacturer and for which the manufacturer can not provide a complete software life cycle control, is then classied as O-the-shelf software or OTS software. Using OTS software allows the medical device manufacturer to focus on developing their device without having to spend resources on software already existing on the market. However, the manufacturer needs to ensure that the OTS software will not introduce any risks towards the patient. It is also important to state that although the manufacturer has not developed the software used, the responsibility to ensure that the software's performance is maintained safe and eective remains with the developer [11]. As a general guidance on how much documentation the manufacturer needs to provide when using OTS software the schematic shown in gure 3.4 can be 16
21 used, which takes into account the level of concern presented by the OTS software. Level of concern is a measurement of the severity of potential hazardous situation that the OTS software is implemented to prevent, or if the OTS software in itself can cause a potential hazardous situation Software of Unknown Pedigree As an extension to the OTS software, Software of Unknown Pedigree (sometimes Software of Unknown Provenance), or SOUP, is used to referred to software obtained from a third party and for which documentation is dicult to obtain. As documentation is essential when submitting a PMA to the FDA or to the EC, this would make all devices using third party software illegible for market distribution. Instead, by classifying a third party software as an SOUP, the documentation requirements decreases to enclose motivation to the use of the software, explaining the origin as well as presenting what documentation is available. Further, the risk analysis of the software being developed should include the risks associated with the use of the SOUP in regards to lack of documentation prior to testing the SOUP within the software device. As mentioned in Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices[6], the responsibility for adequate testing of the device and for providing appropriate documentation of software test plans and results remains with the developer issuing the PMA. 17
22 Figure 3.4: OTS Software Decision Schematic [11] 18
23 Chapter 4 Methods This chapter describes the methods used during the master thesis project. 4.1 Literature Studies First a literature study was conducted to gain a better understanding for software development in general and medical software development specically. A focus was placed on obtaining guidance documents from the FDA and the EC as acceptance from these organizations is one of the main goals. The literature study also included gaining a basic understanding of the software development models described in chapter 3.2. Lastly, the literature study involved identifying current medical software regulatory standards to which compliance would be required. 4.2 Requirements Identication Through the literature study, needed activities were identied that would make up the medical software development procedure. The identied regulatory standard, IEC Medical device software, were used identify the dierent processes required to be established to provide a framework for the procedure that were to be created. The FDA guidance document to software validation were used to provide a foundation of software development practices to the framework. Lastly, existing procedures in Integrum's quality management system were reviewed to determine if these could used for the medical software development procedure as well as to identify if there were any restrictions within the quality management system that needed to be taken into consideration. 4.3 Verication Verication of the medical software development procedure was needed to ensure that the procedure meet the regulatory requirements. The procedure also needed to be understandable to the software development team and that the activities described in the procedure were free of redundancies. 19
24 Chapter 5 Problem Description This chapter describes the task assigned to be performed and the dierent considerations that were taken into account. 5.1 Problem Description The task was to establish a medical software development procedure to be implemented into the quality management system at Integrum AB. Prior to this master thesis project, there were no procedures or experience on medical software development at Integrum AB. As such, the master thesis project also came to involve the selection of supportive software development environment, safety classication of the medical device software and classication of the dierent medical device components connected to the medical software. This thesis is concerning the development of the Software Development, Software Validation and Software Classication Guidelines as well as the selection of the software development environment as shown in gure
25 Figure 5.1: Problem Description 21
26 Chapter 6 Design and Implementation This chapter describes the design and implementation process performed in order to complete the assigned task. In the chapter, the dierent components that make up the project for which the medical software development procedure is to be used are considered and described in how they aect the overall project at Integrum AB. 6.1 The OPRA-NCAL Project The OPRA-NCAL project at Integrum AB concerns the development of a robotic prosthesis that use the muscle and nerve signals of the patient to control the articial limb. For this purpose, a medical software is needed to be developed which requires the establishment of quality management procedures for handling the development process in order to pass existing regulatory requirements. Figure 6.1 shows how the OPRA-NCAL project is divided into its dierent parts. The Osseointegrated Human-Machine Gateway, OHMG, is the mechanical xture that is implanted into the patient's bone, the Articial Limb Controller, ALC, is the electrical component that controls the robotic prosthetic and lastly the NCAL-Soft which is the software that is used both by the ALC Figure 6.1: OPRA-NCAL Breakdown 22
27 as well as used on a PC for training and tting purposes. As the OPRA-NCAL project consists of dierent parts, one of the initial steps that needed to be performed was to investigate the regulatory classication of these parts according to current EU and US directives. It had already been previously been established that the complete OPRA- NCAL system is a medical device but, at the same time, the OPRA-NCAL is comprised by three parts which potentially can be classied separately with only the main part being a medical device and the other two as accessories. The benets for doing this would be a decrease in documentation, verication and overall resources required to drive the OPRA-NCAL project through development, the regulatory classications and onto the market. The medical software classication for the NCAL-Soft project was also investigated in order to determine the optimal safety classication for the NCAL-Soft. This was due to a higher safety classication would require more resources from Integrum AB OPRA-NCAL Classication According to European Union Denition of medical device Within the EU a medical device is dened as follows from the medical device directive (MDD) 93/42/EEC: a 'medical device' means any instrument, apparatus, appliance, software, material or other article, whether used alone or in combination, together with any accessories, including the software intended by its manufacturer to be used specically for diagnostic and/or therapeutic purposes and necessary for its proper application, intended by the manufacturer to be used for human beings for the purpose of: - diagnosis, prevention, monitoring, treatment or alleviation of disease, - diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap, - investigation, replacement or modication of the anatomy or of a physiological process, - control of conception, and which does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means; Denition of accessory The MDD denes a medical device accessory as: an `accessory' means an article which whilst not being a device is intended specically by its manufacturer to be used together with a device to enable it to be used in accordance with the use of the device intended by the manufacturer of the device. 23
28 Medical software In the denition of active medical device, the following amendment was implemented in 2010 through the directive 2007/47/EC: Stand alone software is considered to be an active medical device. Further amendments concerning medical software was in regards to requirements for medical devices connected to or equipped with an energy source, where the following addition was made: For devices which incorporate software or which are medical software in themselves, the software must be validated according to the state of the art taking into account the principles of development lifecycle, risk management, validation and verication Conclusion In accordance with the MDD denitions above, it would be possible to label the main part of medical device product as `medical device' while labeling any attachment parts as `accessory' to that medical device as long as the attachment on its own does not fulll the requirements for a 'medical device'. Table 5.1 shows the devices and the dierent cases which would classify a device as a 'medical device'. It should also be noted that only the device on its own is taken into concideration and the functional capabilities of the device. Device name Case 1 Case 2 Case 3 Case 4 OHMG No No Yes No ALC No No No No Table 6.1: Medical device classication comparison. Case 1: Diagnosis, prevention, monitoring, treatment or alleviation of disease; Case 2: Diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap; Case 3: Investigation, replacement or modication of the anatomy or of a physical process; Case 4: Control of conception As seen, the OHMG fullls one of the medical device classications due to being implanted and thus modifying the anatomy of the patient. The ALC however does not fulll any of the cases on its own, and can be classied as an 'accessory' to the medical device, which would be the OHMG. This means that the OPRA-NCAL medical device can be broken up into a medical device component (the OHMG) and an accessory (the ALC). NCAL-Soft is, in accordance with the 2007/47/EC directive, an active medical device and is therefore not included in table 6.1. It should nally be mentioned that, while the OHMG, ALC and NCAL-Soft all treat a handicap when combined, they do not fulll this case when isolated OPRA-NCAL Classication According to the United States of America Denition of medical device The FDA's denition of a medical device is as follows: 24
29 an instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other similar or related article, including a component part, or accessory which is: - recognized in the ocial National Formulary, or the United States Pharmacopoeia, or any supplement to them, - intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, in man or other animals, or - intended to aect the structure or any function of the body of man or other animals, and which does not achieve any of its primary intended purposes through chemical action within or on the body of man or other animals and which is not dependent upon being metabolized for the achievement of any of its primary intended purposes Conclusion This would mean that if a medical device consists of multiple parts, all those parts would be labeled as medical devices due to that the product they form is a medical device. For the OPRA-NCAL product, both the OHMG and the ALC would be labeled as medical devices as they together form the actual medical device product NCAL-Soft The NCAL-Soft is the software component to the OPRA-NCAL project and is separated into two parts. One part to be installed on a PC for training and tting purposes and another intended to be installed into the ALC component of the OPRA-NCAL medical device. NCAL-Soft is described further in chapter 6.2. In accordance with IEC Medical device software - Software life cycle processes, all medical software being developed for use in a medical device or to be used as a medical device in its own right must be given a safety classication based on the potential hazardous situation that could arise from a software failure Classication of NCAL-Soft As the two parts of NCAL-Soft are installed in separate devices and does not interface with each other, they can be given an individual safety classication as mentioned in chapter ALC Following a preliminary hazard analysis of the ALC component for the NCAL-Soft, the worst potential risk associated with a hazardous situation would result in bruising. Other identied potential risks would result in a delay of prosthetic movement or no movement at all. However, there are possibilities to reduce the safety classication through hardware implemented risk control measures. The eects of these hardware changes would need to be analyzed to determine the impact it would have on the 25
30 performance of the medical device before a decision can be made to implement them or not. PC For the PC component of the NCAL-Soft the only concern would be during the sessions where the patient is training the ALC software through the PC, but this should not give rise to any risks unless the patient deliberately act against the user instructions Conclusion Based on the preliminary risk analysis it is determined that the ALC component of the NCAL-Soft will have a Class B safety classication while the PC component of the NCAL-Soft will have a Class A safety classication Evaluation of Safety Classication Reduction The gain of a reduced safety classication for Integrum AB would be the reduction in documentation and verication eort. This reduction could save resources and possibly result in a faster project closure. In order to reduce the safety classication of a medical device software, as described in chapter 3.1.2, risk control measures have to be implemented through hardware changes or adding additional hardware to the medical device. So when considering a reduction of safety classication, it is needed to take the impact of hardware changes on the medical device performance into consideration. One of the identied risks would be accidental bruising due to the prosthesis performing an unintentional action, such as closing of prosthetic hand while holding another person's hand. A possible hardware solution to this problem would be to change the prosthetic motor to a less powerful motor, but it would need to be veried that the intended use can still be met. 6.2 NCAL-Soft The NCAL-Soft is a training and tting software designed for the OPRA-NCAL medical device currently being developed by Integrum AB. The project leader is Max J. Ortiz C., who also is the project leader for the OPRA-NCAL project. Here, there dierent tasks concerning the development of NCAL-Soft are described Intended Use Dening a medical device's intended use is a critical part and is one of core aspects taken into consideration when performing the risk assessment. The intended use should clearly dene what the product is designed to do as well as any restrictions in how it is to be used. One of the rst tasks was to dene the intended use for NCAL-Soft and the ALC. Due to condentiality, these denitions can not be presented in this thesis. 26
31 6.2.2 Risk Assessment The risk assessment of medical software must, like any other medical device, comply with the ISO 14971:2007 standard for risk management for medical devices. The major dierence, however, between performing a risk assessment on a software medical device contra that of a classic medical device, is that the probability of a risk or hazard occurring can not be estimated in the same statistical way due to software's unpredictability. As such, any foreseeable risk or hazard has a 100% chance of occurring. As previously stated, the outcome of the risk assessment of NCAL-Soft was a worst case risk of bruising Software Safety Classication As covered previous in chapter 3.1.2, software safety classication needs to be performed at the start of the software development. With NCAL-Soft marking the rst software development project at Integrum AB, there were no existing procedures for how the software safety classication should be performed. As such, a classication strategy document for how the software classication should be performed in compliance with the IEC standard needed to be created. The safety classication for the NCAL-Soft medical software was generated at the same time and documented as NCAL-Soft Classication. The safety classication strategy includes the identication of the dierent software entities that comprise the software system and how they interface. It also includes performing risk analysis on the identied entities and the specication of risk control measures. The software safety classication strategy is documented as Software Classication Guidelines procedure which is to be implemented into the quality management system at Integrum AB and used to determine the software safety classication of any medical software project that Integrum AB initiates Software Development Environment When developing a software, the choice of a suitable software development environment (SDE) needs to be considered. For Integrum, the requirements on a SDE was based on the compliance requirements for the IEC 62304, features such as traceability (risk assessment to code and design requirements to code), revision control, activity planning and document sharing between the developers. Dierent SDE's that could be of use to Integrum in the development of NCAL-Soft needed to be identied and evaluated. The rst possible solution was from Mortice Kerns Systems which solution, MKS Integrity, oers full SDE solutions including integrated risk management and quality management but also includes features that were not desirable for the NCAL-Soft project. The cost for the license was another factor which played a part in looking for a dierent solution. A second solution was found in IBM Rational SDE which, like MKS Integrity, includes integrated risk management, quality management as well as the possibility of creating a full traceability between implemented software requirements, risk analysis and actual source code. The biggest perk for IBM 27
32 SDE Traceability Revision control Integrated RM Integrated QM License MKS Yes Yes Yes Yes - Integrity IBM Yes Yes Yes Yes Free Rational rt:collabs Yes Yes No No - Table 6.2: SDE Feature Comparison Rational is that it is free as long as there are 10 or less developers using it, making it a very attractive choice for small companies, the downside of that is that there is no IBM support included in the free license. A third solution found is the rt:collabs SDE which includes the possibility of having traceability between source code and risk analysis or software requirements through the use of tags and comments, increasing the manual responsibility. It is designed to use a SVN client for revision control and has a web based user-interface that is built on TracWiki [15]. Table 6.2 shows a breakdown comparison of the three possible solutions described above. The decision of SDE ended up being between IBM Rational and rt:collabs. Both could oer similar features, with IBM having some additional features over rt:collabs as well as being free. The trade-o would be that the rt:collabs license would include support whilst Integrum would be without support with the IBM solution. The rt:collabs platform was ultimately selected due to its high level of con- guration, that it can be used as a platform for more than just software development and that support would be available. Following the choice of rt:collabs came the need of identifying a suitable SVN client. A number of dierent SVN clients were identied, such as Visual Studio SVN and Tortoise SVN. Tortoise SVN was selected as it works well with any le type and does not require a specic programming environment. 6.3 Software Development Procedure As the NCAL-Soft marks the rst software product being developed by Integrum AB, there was a need to create a software development procedure to be implemented into Integrum's quality system which would ensure that the development is conducted in accordance with current regulatory requirements for medical software. The rst aspect to consider was which of the software development models described in chapter 3.2 would t Integrum AB as an organization, its development team and their quality system. The development model would have to work with a very small team and require as little additional resources from Integrum AB as an organization also if it has a similar structure to the existing development procedure in place at Integrum AB for their non-software products would make it easier to implement but that is of lesser importance. 28
33 The main task of this thesis project was the development of a set of software development procedures to assure compliance with the regulatory requirements of IEC Medical device software - Software life cycle processes. This was done by initially performing a quick review of the quality management system in place and compare the implemented standard operating procedures, SOP, present in it, with the requirements within the standard. From there, it could determined if new SOP's were needed to be created, or if changes to present SOP's would be sucient Proposed Software Development Procedure The software development procedure described in this section is that of an iterative development model. This is due to that software development is an iterative process but also due to that it is the basis for the proposed software development procedure for Integrum AB. It is also easy to verify that this development procedure follows the general principals of software validation by comparing the activities described in Chapter 3.1 with the processes described below. The software development would start with an initial planning process, as shown in gure 3.2 on page 15, before the actual development starts. The development outline is shown below with a brief description of each of the processes. ˆ Initial Planning ˆ Requirements ˆ Implementation ˆ Testing ˆ Evaluation ˆ Release Initial Planning The initial planning deals with all the dierent tasks that are needed to be performed before starting a project. It covers initial risk analysis, initial software designs and establishing software system requirements. But also nancial estimates and other factors that aect the company. Requirements As mentioned in Chapter 3.1.3, the requirements activity should identify and dene the various software and system requirements and ensure that any additional requirements that are identied during the development process is properly dened for a successful implementation. Implementation This activity is where all the requirements (system and software), as well as any identied risk control measures, are implemented into actual source code. 29
34 Testing The testing activity is the verication eort to ensure that the source code fullls its intended function, and that any new code or functions implemented into the software does not cause conicts. It is through this verication eort that the software will be validated for completeness at the end of the project. Evaluation The activity of evaluation is where the test results are analyzed and decisions are made on how the software architectural design shall be rened, if additional risk control measures shall be implemented and if an implemented functionality has reached the acceptance criteria. During this activity, the software development plan is updated to include the results of the iteration ending as well as add appropriate documentation concerning the iteration that is to start. Release Once the evaluation activity shows that all requirements and risk control measures have successfully been implemented and that the software has reached the acceptance criteria, it will enter the release activity. Documentation shall be created identifying any known software anomalies that has not been corrected along with a motivation for why this has not been done. An evaluation eort shall also be performed to ensure that all created documentation supports that the software has been completed Creating a Software Development Procedure After deciding on software development model to use, the next step was to create the necessary operating procedures for software development based on the requirements set by the IEC standard. The rst step was to identify the procedures required to be established, followed by dening what these procedures should cover in terms of clauses of the IEC standard. This resulted in the following list of procedures that needed to be dened and developed: ˆ Software Development Process ˆ Software Design Process ˆ Software Integration and Testing Process ˆ Software Conguration Management Process ˆ Software Problem Resolution Process ˆ Software Maintenance Process ˆ Software Release Process ˆ Software Risk Management Process 30
35 Activity: Process: Initial Planning - Requirements Software Development, Software Design Design Software Design Implementation Software Integration and Testing Testing Software Integration and Testing Evaluation Software Conguration Management, Software Problem Resolution Release Software Release Table 6.3: Connection between development activity and development processes Comparing this list of procedures with the Iterative and Incremental development model shown on page on page 15, each process is connected to a corresponding activity. This connection is shown in table 6.3. As can be seen in this table, the risk management process and the maintenance process does not have a dedicated development model activity but exist outside of the illustrated model shown in 3.2. It is also shown that the initial planning activity is not covered by the IEC standard, but it is described in the General Principals of Software Validation [4]. Once these processes had been identied, the work of creating and rening them to ensure that they were clear in what is required to be done and how this is to be done began. As part of this process, the software design, software maintenance and software release process were merged into one single process labeled software design and release process. Further, the requirements activity was integrated into a separate document labeled IEC Implementation that serves as a guiding document to how the IEC regulatory requirements should be implemented into the rt:collabs TracWiki platform that is used by Integrum AB. At this point, there were several documents describing the dierent processes involved in medical software development but there were no clear step-by-step procedure that described how these processes could easily be used by the programmers at Integrum AB. Bearing that in mind, an eort was made in rewriting the existing processes, listed in table 6.3, into a step-by-step procedure called Software Development. The re-write also included the addition of references to the IEC Implementation document in the cases where a more detailed description would be needed, i.e. for the denition of requirements. As Integrum AB already has an existing risk management procedure, there was a concern of how the software risk management would best be implemented into the quality management system. After going over the process with the quality manager for Integrum AB, it was decided that a new separate procedure should be created which would cover the software risk management as well as the software release and software maintenance. This new procedure was labeled Software Validation. The nal software project development process is described by the owchart shown in gure 6.2. The Software Development procedure is described in detail by gure 6.3. It should be noted that the last step in the owchart is the same step described in 6.2 and does not indicate a need to repeat the activities described in the 31
36 Figure 6.2: Software Development Procedure Flowchart Software Product Release process dened in the Software Validation procedure. The integration plan described in gure 6.3 refers to the process that is described by gure 6.4. This owchart describes the integration testing process that all source code is required to go through before being implemented into the software product and the testing of the software system before being submitted to the Software Product Release process dened in the Software Validation procedure. 32
37 Figure 6.3: Software Development Method Flowchart 33
38 Figure 6.4: Software Integration Testing Flowchart 34
39 Chapter 7 Verication This chapter describes the verication eort that was performed to ensure that the medical software development procedure works as intended and fulll the regulatory requirements set by IEC Medical device software. The verication was performed in two dierent ways, one through the actual use of the procedure and another by verifying that each required regulatory clause was met. 7.1 Verication of Performance This verication was initially conducted by running a dummy project through the procedure to catch any inconsistencies or redundancies. When all identied redundancies and inconsistencies had been corrected, the procedure was handed to the medical software development team for live testing. Although the use of the software development procedure was highly limited during this test, new issues that needed to be resolved were still identied. 7.2 Verication of Compliance This verication stage was performed at the end of the project and consisted of going through each regulatory clause and verify that the clause had been met. This was done over the course of several meeting between myself and the medical software development team. It was completed up to clause 7 of the IEC standard, Risk Management Process. 35
40 Chapter 8 Discussion 8.1 NCAL-Soft Safety Classication The NCAL-Soft safety classication is based on a preliminary risk analysis that will be needed to be repeated once a more detailed design of the NCAL-Soft project has been established. It is also of importance that a risk analysis is performed continuously throughout the development process to ensure that no potential hazardous situations are missed. Also, it was determined that Integrum AB will not be developing any medical software that would be classied as a Class C medical software given their current projects. However, in the event that Integrum AB start developing medical software that will actively stimulate biological tissue this needs to be evaluated. 8.2 Software Development Procedure The proposed software development process created for Integrum AB is highly based on the IID model described in section but adapted to t the small development team at Integrum AB and the TracWiki development platform. It consists of a software development procedure, a software validation procedure and IEC Implementation guidance document. The software development procedure provides a step-by-step method for how a medical software development project should be performed for compliance with IEC Medical software standard and includes owchart representations of the software project development process and integration testing process. The software validation procedure provides instructions for performing validation prior to releasing the software product to the market, post-release activities for handling feedback and product maintenance as well as software risk management. The IEC Implementation guidance document provides more detailed explanations for why dierent activities needs to be performed by referencing directly to the IEC regulatory clause that activity complies with. It also oers additional information concerning denition of requirements and software design documentation. As previously mentioned, it was evaluated that Integrum AB will not develop any Class C medical software. As such, the software development procedure is 36
41 restricted to only comply with the regulatory requirements for Class A and Class B medical software. In the event that Integrum AB will be developing Class C medical software, additions will be needed to be made to the software development procedure to include the regulatory requirements for Class C. 8.3 Future Work As with most development work, there is always room for further work and improvements. The verication of compliance needs to be completed by verifying that remaining clauses are met. Further, the software development procedures need to be used in a real development instead of simulated development projects. Although the software development procedures were used as part of the NCAL- Soft project, it was not used to a wide extent and the use could not give any qualitative feedback on the procedure's performance. I also believe that Integrum AB would benet from looking into obtaining guidance documents concerning software requirement specication and software system requirement specication. Though it is not required by the medical software standard, I believe it will aid Integrum AB in their expansion into a medical software development company. Example of these guidance documents is the IEEE STD for software requirement specication and the IEEE STD for software system requirement specication. Although I believe these documents would result in changes needed to be made to the software development procedure. They should also result in better documented requirements which will lead to a smoother software development, as well as decrease the FDA processing. 37
42 Bibliography [1] Device Classication. Food and Drug Administration [cited ]; Available from: Overview/ClassifyYourDevice/ucm htm [2] Premarket Notications (510k). Food and Drug Administration [cited ]; Available from: HowtoMarketYourDevice/PremarketSubmissions/PremarketNotication510k/ default.htm [3] Guide to the implementation of directives based on the New Approach and the Global Approach, European Commission, 2000 [4] General Principles of Software Validation; Final Guidance for Industry and FDA Sta, Center for Devices and Radiological Health, Food and Drug Administration, January 2002 [5] A Formal Methods-based verication approach to medical device software analysis. EE Times [cited ]; Available from: [6] Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices, Oce of Device Evaluation, Center for Devices and Radiological Health, Food and Drug Administration, May 2005 [7] Royce, Winston, Managing the Development of Large Software Systems, Proceedings of IEEE WESCON 26:1-9, August 1970 [8] Martin, Robert, Iterative and Incremental Development, Engineering Notebook Column, April 1999 [9] Sutherland, Je, Agile Development: Lessons Learned From the First SCRUM, Cutter Agile Project Management Advisory Service. Executive Update, Vol. 5, No. 20, October 2004 [10] Boehm, Barry, A Spiral Model of Software Development and Enhancement, TRW Defense Systems Group, May 1988 [11] Guidance for Industry, FDA Reviewers and Compliance on O-the-Shelf Software Use in Medical Devices, Ocer of Device Evaluation, Center for 38
43 Devices and Radiological Health, Food and Drug Administration, September 1999 [12] IEC 62304, Medical device software - Software life cycle processes, International Electrotechnical Commission, May 2006 [13] Scrum (development). Wikipedia [cited ]; Available from: [14] Iterative and incremental development. Wikipedia [cited ]; [15] TracWiki - The Trac Project. Trac [cited ]; Available from: 39
Medical Software Development. International standards requirements and practice
Medical Software Development International standards requirements and practice Food and Drug Administration What? A public health agency Why? Protect American consumers How? By enforcing the Federal Food,
How To Write Software
1 Medical Device Software - Software Life Cycle Processes IEC 62304 2 Credits John F. Murray Software Compliance Expert U.S. Food and Drug Administration Marcie R. Williams Medical Device Fellow Ph.D.
Implementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes.
Implementation of ANSI/AAMI/IEC 62304 Medical Device Software Lifecycle Processes.. www.pharmout.net Page 1 of 15 Version-02 1. Scope 1.1. Purpose This paper reviews the implementation of the ANSI/AAMI/IEC
CDRH Regulated Software
CDRH Regulated Software An Introduction John F. Murray Jr. CDRH Software Compliance Expert CDRH Regulates Software in the following areas Medical Devices Automation of Production Systems Automation of
WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE
WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 [email protected]
Introduction into IEC 62304 Software life cycle for medical devices
Introduction into IEC 62304 Software life cycle for medical devices Christoph Gerber 4. September 2008 SPIQ 9/5/2008 1 Agenda Current Picture Regulatory requirements for medical device software IEC 62304
Medical Device Software
Medical Device Software Bakul Patel Senior Policy Advisor 1 Overview Medical devices and software Oversight principles and Current approach Trends, Challenges and opportunities Addressing challenges 2
Medical Product Software Development and FDA Regulations Software Development Practices and FDA Compliance
Medical Product Development and FDA Regulations IEEE Orange County Computer Society March 27, 2006 Carl R. Wyrwa Medical Product Development and FDA Regulations Introduction Regulated FDA Overview Medical
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound
Final Document. Software as a Medical Device (SaMD): Key Definitions. Date: 9 December 2013. Despina Spanou, IMDRF Chair. IMDRF/SaMD WG/N10FINAL:2013
Final Document Title: Authoring Group: Software as a Medical Device (SaMD): Key Definitions IMDRF SaMD Working Group Date: 9 December 2013 Despina Spanou, IMDRF Chair This document was produced by the
How To Know If A Mobile App Is A Medical Device
The Regulation of Medical Device Apps Prepared for West of England Academic Health Science Network and University of Bristol June 2014 June 2014 1 Table of Contents 1 Purpose...3 2 Scope...3 3 The Regulation
Document issued on: May 11, 2005
Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005 This document supersedes Guidance for the
Software development process
OpenStax-CNX module: m14619 1 Software development process Trung Hung VO This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 2.0 Abstract A software development
Medical Certification: Bringing genomic microcores to clinical use OI- VF- WP- 011
Medical Certification: Bringing genomic microcores to clinical use VoltedFlow GmbH Industriestrasse 23, 6055, Alpnach Dorf, Switzerland 1 Summary VoltedFlow has developed solutions to vastly speed up DNA
2015. All rights reserved.
DOCUMENT: Future AAMI/IEC 62304:2006/AMD1, 18-August-2015 Final Draft International Standard for Vote, Amendment 1 to IEC 62304: Medical device software Software life cycle processes. Public Review Draft
US & CANADA: REGULATION AND GUIDELINES ON MEDICAL SOFTWARE AND APPS OR
US & CANADA: REGULATION AND GUIDELINES ON MEDICAL SOFTWARE AND APPS OR A MEDICAL DEVICE IS A MEDICAL DEVICE IS A MEDICAL DEVICE AHWP Medical SW Workshop Taipei, Taiwan November 3, 2012 John G. Abbott,
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical SUR-G0014-1 12 NOVEMBER 2013 This guide does not purport to be an interpretation of law and/or regulations
Correspondence between ISO 13485:2003 and the US Quality System Regulation
Correspondence between ISO 13485:2003 and the US Quality System Regulation Correspondence between ISO 13485:2003 and the US Quality System Regulation 1 Scope 1.1 General This International Standard specifies
What is a medical device? Medical Devices: Roadmap to Market. Kathryn Klaus, Esq.
Medical Devices: Roadmap to Market Kathryn Klaus, Esq. The last installment of Regulatory 360 discussed the FDA organization in general where it came from and a broad overview of how it operates, as well
Adapting Agile Software Development to Regulated Industry. Paul Buckley Section 706 Section Event June 16, 2015
Adapting Agile Software Development to Regulated Industry Paul Buckley Section 706 Section Event June 16, 2015 Agenda FDA s expectations for Software Development What is Agile development? Aligning Agile
Medical Device Software Do You Understand How Software is Regulated?
Medical Device Software Do You Understand How Software is Regulated? By Gregory Martin Agenda Relevant directives, standards, and guidance documents recommended to develop, maintain, and validate medical
CE Marking: Your Key to Entering the European Market
CE Marking: Your Key to Entering the European Market N. Gerard Zapiain U.S. Department of Commerce International Trade Administration Office of Microelectronics, Medical Equipment and Instrumentation I.
Regulatory Considerations for Medical Device Software. Medical Device Software
Medtec Ireland 2015 Wireless Medical Devices Regulatory Considerations for Medical Device Software Kenneth L. Block, RAC October 7, 2015 Galway, Ireland Offices: Dallas, Texas (12 employees) Tokyo, Japan
How To Validate Software
General Principles of Software Validation; Final Guidance for Industry and FDA Staff Document issued on: January 11, 2002 This document supersedes the draft document, "General Principles of Software Validation,
Information Sheet Guidance For IRBs, Clinical Investigators, and Sponsors
Information Sheet Guidance For IRBs, Clinical Investigators, and Sponsors Frequently Asked Questions About Medical Devices Additional copies are available from: Office of Good Clinical Practice Office
Usability of Medical Applications Ved Line Kagenow Svenstrup, [email protected]
Usability of Medical Applications Ved Line Kagenow Svenstrup, [email protected] What is usability? The user, rather than the system, at the center of the process. Risk of operating errors that can cause injury
Effective Software Verification for Medical Devices
STERLINGTECH AND KLOCWORK WHITE PAPER NOVEMBER 2009 Effective Software Verification for Medical Devices Achieving compliance and meeting productivity goals with static analysis In addition to producing
GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES
Foreword GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES These guidance notes do not aim to be a definite interpretation of National Laws and/or regulations and are for guidance purpose only.
Medical Device Software Standards for Safety and Regulatory Compliance
Medical Device Software Standards for Safety and Regulatory Compliance Sherman Eagles +1 612-865-0107 [email protected] www.softwarecpr.com Assuring safe software SAFE All hazards have been addressed
Combination Products Regulation in the United States
Combination Products Regulation in the United States Presenter: Scott Sardeson RAC US/EU 3M Health Care St. Paul, MN USA 1 Presentation Outline Combination products Definitions and Regulations Jurisdiction
General Wellness: Policy for Low Risk Devices. Draft Guidance for Industry and Food and Drug Administration Staff
General Wellness: Policy for Low Risk Devices Draft Guidance for Industry and Food and Drug Administration Staff DRAFT GUIDANCE This guidance document is being distributed for comment purposes only. Document
Safeguarding public health The Regulation of Software as a Medical Device
Safeguarding public health The Regulation of Software as a Medical Device Dr Head of New and Emerging Technologies Medicines and Healthcare Products Regulatory Agency (MHRA) Crown Copyright 2013 About
SYSTEMS AND SOFTWARE REQUIREMENTS SPECIFICATION (SSRS) TEMPLATE. Version A.4, January 2014 FOREWORD DOCUMENT CONVENTIONS
SYSTEMS AND SOFTWARE REQUIREMENTS SPECIFICATION (SSRS) TEMPLATE Version A.4, January 2014 FOREWORD This document was written to provide software development projects with a template for generating a System
Development of a Process Assessment Model for Medical Device Software Development
Development of a Process Assessment Model for Medical Device Software Development Marion Lepmets, Paul Clarke, Fergal McCaffery, Anita Finnegan, Alec Dorling Regulated Software Research Centre, Dundalk
CDRH Regulated Software Looking back, looking forward
CDRH Regulated Software Looking back, looking forward Medical Device Software Compliance Expert US Food & Drug Administration at the Regulatory Affairs Professional Society Indianapolis, Indiana Goal of
Regulation and Risk Management of Combination Products
Regulation and Risk Management of Combination Products Katherine Ulman Associate Scientist Global Regulatory Compliance Manager Dow Corning Healthcare Jim Curtis Senior Specialist, Healthcare Applications
Medical Devices: CE Marking Step-by Step
Medical Devices: CE Marking Step-by Step Sylvia Mohr Standards Specialist U.S. Mission to the European Union 1 Introduction CE marking in 8 Steps Overview intended purpose - technical documentation classification
Regulation of Mobile Medical Apps
Regulation of Mobile Medical Apps May 30, 2014 Copyright 2014 Software Quality Consulting Inc. Slide 1 Speaker Bio Steven R. Rakitin has over 35 years experience as a software engineer and 25 years in
Navigating the New EU Rules for Medical Device Software
Navigating the New EU Rules for Medical Device Software Mathias Klümper and Erik Vollebregt examine how European Directive 2007/47/EC will affect requirements for medical device software. Directive 2007/47/EC
Considerations for using the Web for Medical Device Applications
Considerations for using the Web for Medical Device Applications MEDS, San Diego August 23 rd, 2012 Daniel Sterling, President Who is Sterling? Your Partner in Medical Device Development What we do: o
Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH
Med-Info International expert information for the Medical Device industry Council Directive 93/42/E on Medical Devices Practice-oriented summary of the most important aspects and requirements contained
Health Informatics Application of clinical risk management to the manufacture of health software Formerly ISO/TS 29321:2008(E) DSCN14/2009
Health Informatics Application of clinical risk management to the manufacture of health software (formerly ISO/TS 29321:2008(E)) DSCN14/2009 Programme Sub-Prog / Project NPFIT Clinical Safety Document
2014 Annual Report on Inspections of Establishments
2014 Annual Report on Inspections of Establishments Table of Contents Introduction... 2 Data Collection and Definitions... 3 Section 510(h)(6)(A)(i) Number of Domestic and Foreign Establishments Registered
Risk Assessment for Medical Devices. Linda Braddon, Ph.D. Bring your medical device to market faster 1
Risk Assessment for Medical Devices Linda Braddon, Ph.D. Bring your medical device to market faster 1 My Perspective Work with start up medical device companies Goal: Making great ideas into profitable
The Shifting Sands of Medical Software Regulation
The Shifting Sands of Medical Software Regulation Suzanne O Shea Ralph Hall September 10, 2014 What Software is Regulated by FDA? FDA regulates medical devices. FDA regulates software that meets the definition
Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC
Overview of Medical Device Controls in the US By Nandini Murthy, MS, RAC 18 controls are a regulatory requirement for medical devices. In the US, compliance with the design controls section of 21 Code
The New Paradigm for Medical Device Safety. Addressing the Requirements of IEC 60601-1 Edition 3.1
The New Paradigm for Medical Device Safety Addressing the Requirements of IEC 60601-1 Edition 3.1 Medical devices play a vital role in the diagnosis and treatment of most health-related conditions, and
NOTICE TO MANUFACTURERS OF RADIOTHERAPY MEDICAL DEVICES
French health products safety agency Medical Devices Evaluation Directorate Market Surveillance Department Matter handled by Pascal Di Donato Email: [email protected] August 2007 NOTICE
Quality Assurance Manual
Quality Assurance Manual Revised April 30, 2014 2004, 2005, 2006, 2007, 2008, 2009, 2010, 2011, 2012, 2013, 2014 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
Summary of: P890003/S285, P090013/S112, P010031/S397, P010015/S215, P980016/S436, P980035/S343. Background and Summary
Summary of: P890003/S285, P090013/S112, P010031/S397, P010015/S215, P980016/S436, P980035/S343 Background and Summary This PMA Supplement is to obtain approval for updates to the Medtronic MyCareLink Patient
Clinical Risk Management: Agile Development Implementation Guidance
Document filename: Directorate / Programme Document Reference NPFIT-FNT-TO-TOCLNSA-1306.02 CRM Agile Development Implementation Guidance v1.0 Solution Design Standards and Assurance Project Clinical Risk
General Principles of Software Validation; Final Guidance for Industry and FDA Staff
General Principles of Software Validation; Final Guidance for Industry and FDA Staff Document issued on: January 11, 2002 This document supersedes the draft document, "General Principles of Software Validation,
Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau [email protected]
Software Regulated as a Medical Device Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau [email protected] Therapeutic Products Directorate Director General S. Sharma
CENTER FOR CONNECTED HEALTH POLICY
CENTER FOR CONNECTED HEALTH POLICY The Center for Connected Health Policy (CCHP) is a public interest nonprofit organization that develops and advances telehealth policy solutions to promote improvements
(OJ L 169, 12.7.1993, p. 1)
1993L0042 EN 11.10.2007 005.001 1 This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents B COUNCIL DIRECTIVE 93/42/EEC of 14 June 1993 concerning
An introduction to the regulation of apps and wearables as medical devices
An introduction to the regulation of apps and wearables as medical devices Introduction Phones are increasingly equipped with a range of devices capable of being used as sensors, including gyroscopes,
MEDICAL DEVICES INTERIM REGULATION
Issued by the Saudi Food and Drug Authority Board of Directors decree number 1-8-1429 and dated 27 December 2008 The Interim was published in Umm Al-QURA Journal year 86 Issue No:4249 dated in 17 April
615, GSB, University of Alberta, famr,sundari,hoover,[email protected]. Abstract
Software Process Improvement Model for a Small Organization: An Experience Report Amr Kamel, Sundari Voruganti, H. James Hoover and Paul G. Sorenson Dept. of Computing Science, 615, GSB, University of
The FDA Perspective on Human Factors in Medical Device Software Development
The FDA Perspective on in Medical Device Development Molly Follette Story, PhD FDA /CDRH / ODE 2012 IQPC Design for Medical Devices Europe Munich, Germany Februar y1, 2012 Overview Guidance for FDA premarket
Off-The-Shelf Software Use in Medical Devices
Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices Document issued on: September 9, 1999 This document supersedes document, Guidance on Off-the-Shelf Software
Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device
Guidance Notes for Applicants of the Certificate for Clinical Trial on Medical Device List of Contents Page 1. Introduction 1.1 The Medical Device Administrative Control System and the 3 proposed legislation
How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters
environmental failure analysis & prevention health technology development How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters Kevin L. Ong, Ph.D., P.E. Managing
Developing a Mobile Medical App? How to determine if it is a medical device and get it cleared by the US FDA
Developing a Mobile Medical App? How to determine if it is a medical device and get it cleared by the US FDA In this presentation: App stats: Explosive growth Examples already cleared by the US FDA Is
RoCKIn@Work Innovation in Industrial Mobile Manipulation
Robot Competitions Kick Innovation in Cognitive Systems and Robotics FP7-ICT-601012 RoCKIn@Work Innovation in Industrial Mobile Manipulation I: RoCKIn@Work in a Nutshell Revision: Date: Editors: Contributors:
New trends in medical software safety: Are you up to date? 5 October 2015, USA
TÜV SÜD New trends in medical software safety: Are you up to date? 5 October 2015, USA Dr. Peter Havel, Senior Vice President, Medical & Health Services (MHS) TÜV SÜD Product Service [email protected] The
National Decontamination Guidance on Loan Medical Devices (Reusable): Roles & Responsibilities GUID 5002
National Decontamination Guidance on Loan Medical Devices (Reusable): Roles & Responsibilities GUID 5002 July 2015 Contents Page 1.0 Executive summary... 3 2.0 Background... 4 3.0 Objective... 5 4.0 Scope...
Software Development for Medical Devices
Overcoming the Challenges of Compliance, Quality and Cost An MKS White Paper Introduction Software is fast becoming the differentiator for manufacturers of medical devices. The rewards available from software
Poten&al Impact of FDA Regula&on of EMRs. October 27, 2010
Poten&al Impact of FDA Regula&on of EMRs October 27, 2010 Agenda The case for regula&ng Impact on manufacturers Impact on providers Recommenda&ons and best prac&ces 2 A Medical Device Is an instrument,
Triteq Software Services, Products & Skills
Triteq Software Services, Products & Skills Software Management and Development Services Development of bespoke Software Development Processes Development Process Documentations Process Configuration Management
Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?
Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? CEMO Congress, Paris, February 3rd, 2011 Dr. Martine Dehlinger-Kremer Vice President Global Regulatory Affairs Agenda
a Medical Device Privacy Consortium White Paper
a Medical Device Privacy Consortium White Paper Introduction The Medical Device Privacy Consortium (MDPC) is a group of leading companies addressing health privacy and security issues affecting the medical
codebeamer INTLAND SOFTWARE codebeamer Medical ALM Solution is built for IEC62304 compliance and provides a wealth of medical development knowledge
codebeamer Medical ALM Solution is built for INTLAND Traceability matrix Medical wiki Risk management IEC 62304 compliance codebeamer INTLAND codebeamer Medical ALM Solution is built for Medical Device
IBM Rational systems and software solutions for the medical device industry
IBM Software August 2011 IBM Rational systems and software solutions for the medical device industry Improve processes, manage IEC 61508 and IEC 62304 standards, develop quality products Highlights Manage
Process Methodology. Wegmans Deli Kiosk. for. Version 1.0. Prepared by DELI-cious Developers. Rochester Institute of Technology
Process Methodology for Wegmans Deli Kiosk Version 1.0 Prepared by DELI-cious Developers Rochester Institute of Technology September 15, 2013 1 Table of Contents 1. Process... 3 1.1 Choice... 3 1.2 Description...
2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002
STATUTORY INSTRUMENTS 2002 No. 618 CONSUMER PROTECTION The Medical Devices Regulations 2002 Made - - - - - 20th May 2002 Laid before Parliament 21st May 2002 Coming into force - - 13th June 2002 ARRANGEMENT
[DOCKET NO.96N-0002] DRAFT
[DOCKET NO.96N-0002] DRAFT DRAFT DOCUMENT CONCERNING THE REGULATION OF PLACENTAL/UMBILICAL CORD BLOOD STEM CELL PRODUCTS INTENDED FOR TRANSPLANTATION OR FURTHER MANUFACTURE INTO INJECTABLE PRODUCTS DECEMBER,
(Refer Slide Time: 01:52)
Software Engineering Prof. N. L. Sarda Computer Science & Engineering Indian Institute of Technology, Bombay Lecture - 2 Introduction to Software Engineering Challenges, Process Models etc (Part 2) This
Off-The-Shelf Software Use in Medical Devices
Guidance for Industry, FDA Reviewers and Compliance on Off-The-Shelf Software Use in Medical Devices Document issued on: September 9, 1999 This document supersedes document, Guidance on Off-the-Shelf Software
Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO
Development and Validation of In Vitro Diagnostic Tests YC Lee, Ph.D. CEO 1 Validation of In Vitro Diagnostic Tests Validated d Diagnostic Test should: Provides test results that identify if positive i
University of Paderborn Software Engineering Group II-25. Dr. Holger Giese. University of Paderborn Software Engineering Group. External facilities
II.2 Life Cycle and Safety Safety Life Cycle: The necessary activities involving safety-related systems, occurring during a period of time that starts at the concept phase of a project and finishes when
Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day
Medical Devices Notified Bodies and the CE certification Process for Medical Devices European Surgical Robotics Demonstration Day 27 th March 2014, Leuven, Belgium 1 Bart Mersseman Product Manager Medical
IVD Regulation Overview. Requirements to Assure Quality & Effectiveness
IVD Regulation Overview Requirements to Assure Quality & Effectiveness CLIAC Jan. 2002 Statutory and Regulatory Requirements Statute: Food, Drug, and Cosmetic Act Food and Drugs Act of 1906 Food and Drug
PREP Course #27: Medical Device Clinical Trial Management
PREP Course #27: Medical Device Clinical Trial Management Presented by: Evelyn Huang Jeffrey Revello Office of Research Compliance North Shore-LIJ Health System CME Disclosure Statement The North Shore
Medical Device Reporting (MDR) 21 CFR Part 803
Medical Device Reporting (MDR) 21 CFR Part 803 1 Objectives Review applicable sections of 21 CFR 803 and 21 CFR 820 Review and explain MDR reporting requirements Review FDA-483 observation examples 2 Medical
SECTION 2 PROGRAMMING & DEVELOPMENT
Page 1 SECTION 2 PROGRAMMING & DEVELOPMENT DEVELOPMENT METHODOLOGY THE WATERFALL APPROACH The Waterfall model of software development is a top-down, sequential approach to the design, development, testing
Enabling Continuous Delivery by Leveraging the Deployment Pipeline
Enabling Continuous Delivery by Leveraging the Deployment Pipeline Jason Carter Principal (972) 689-6402 [email protected] Pariveda Solutions, Inc. Dallas,TX Table of Contents Matching
ORACLE CONSULTING GROUP
ORACLE CONSULTING GROUP An Official United States Agent Firm for Foreign Establishments CONSULTING MEMORANDUM: DEALING WITH A MEDICAL DEVICE IN THE U.S. 5398 Golder Ranch Rd., Ste. 1 Tucson, Arizona 85739
Software Development Life Cycle (SDLC)
Software Development Life Cycle (SDLC) Supriyo Bhattacharjee MOF Capability Maturity Model (CMM) A bench-mark for measuring the maturity of an organization s software process CMM defines 5 levels of process
Design Controls: Are They Worth the Effort?
Design Controls: Are They Worth the Effort? Compliance-Alliance conducted a survey to measure the effects of FDA s design control regulation on the industry. Most respondents believe the controls have
Understanding Medical Device Regulation for mhealth A Guide for Mobile Operators
for mhealth A Guide for Mobile Operators 1 Foreword Mobile technologies will make a significant contribution to addressing the enormous challenges of healthcare provision worldwide. Early efforts in mobile
Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System
NBOG s Best Practice Guide applicable for AIMDD, MDD, and IVDD 2014-3 Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System 1 Introduction The
The Role of Automation Systems in Management of Change
The Role of Automation Systems in Management of Change Similar to changing lanes in an automobile in a winter storm, with change enters risk. Everyone has most likely experienced that feeling of changing
Demystifying the European Machinery Directive and SEMI Requirements for the Industrial Automation and Semiconductor Markets
Demystifying the European Machinery Directive and SEMI Requirements for the Industrial Automation and Semiconductor Markets An Overview of Current Changes and Directions By Andras Szende Senior Engineer