Test report no. : /05 Type identification: SPECTRAN HF-2025E Test specification : comparative measurements
|
|
|
- Trevor Allen
- 10 years ago
- Views:
Transcription
1 Untertuerkheimer Str. 6-10, Saarbruecken, Germany Phone: +49 (0) Fax: SAR-Laboratory Phone: +49 (0) Test report no. : /05 Type identification: SPECTRAN HF-2025E Test specification : comparative measurements As of Page 1 of 74
2 Table of Contents 1 General Information Notes Assessment of Compliance Bewertung (deutsch) Testing laboratory Details of applicant Application details Test item Test specification(s) Technical test Summary of test results Test environment Measurement and test set-up Measurement system Broadband Electromagnetic Field Test system Test equipment list Averaging Uncertainties Definition of test distances Anechoic chamber B Additional equipment utilized Test results Comparative measurement of an electromagnetic field generated inside an anechoic chamber with broadband field probe (CW signal) at 800, 1000 and 2000 MHz Measurement of electromagnetic field at different frequencies and levels Comparative measurement of an artificially generated DECT signal with broadband field probe Comparative measurement of an artificially generated GSM and UMTS signal with broadband field probe Comparative outdoor measurement of a GSM base station with broadband field probe Comparative outdoor measurement of external Log-Per antenna with FSIQ26 spectrum analyzer using the signal of a GSM base station Comparative conducted measurement of different modulated signals with FSIQ26 spectrum analyzer Annex 1 Photo documentation of test sample and PMM Annex 2 Calibration information As of Page 2 of 74
3 1 General Information 1.1 Notes The test results of this test report relate exclusively to the test item specified in 1.5. The CETECOM ICT Services GmbH does not assume responsibility for any conclusions and generalisations drawn from the test results with regard to other specimens or samples of the type of the equipment represented by the test item. The test report may only be reproduced or published in full. Reproduction or publication of extracts from the report requires the prior written approval of the CETECOM ICT Services GmbH Assessment of Compliance The Aaronia SPECTRAN portable field meter has been tested for compliance of measurement values in comparison with industrial broadband field probes and spectrum analyzers. Especially the correct display of RMSfield strength (in connection with a logarithmic-periodic test antenna) and of artificial signal sources was checked in comparison with industrial reference devices. Special attention was directed to the following signal shapes : CW, GSM, UMTS, WLAN and DECT. The measurements were exclusively performed by skilled personnel under objective and representative conditions, yet device under test and reference device were not always operated under comparative conditions and device settings. The Spectrum Analyzer HF-2025E showed a sufficient measurement accuracy during the measurements of RMS-signal strength of pulsed and unpulsed signals as required by EMF-test specifications (see chapter 2.5 for details). It can therefore be applied for EMCE measurements of all tested signal types Bewertung (deutsch) Das tragbare Feldstärke-Messgerät Aaronia SPECTRAN wurde auf Übereinstimmung der Messwerte im Vergleich zu industriellen Breitband-Feldsonden und Spektrum-Analysatoren überprüft. Zielsetzung war die Überprüfung der korrekten Anzeige von gemessenen RMS-Feldstärkewerten (in Verbindung mit einer logarithmisch-periodischen Messantenne) und von künstlichen Signalquellen im Vergleich zu industriellen Referenzgeräten. Dabei wurde auf folgende Signalformen besonderes Augenmerk gelegt : CW, GSM, UMTS, WLAN und DECT. Die Messungen wurden ausschließlich von sachkundigen Mitarbeitern unter objektiven und repräsentativen Bedingungen durchgeführt. Gemessenes Gerät und Referenzgerät wurden nicht immer unter vergleichbaren Bedingungen und Einstellungen betrieben. Der Spectrum Analyzer HF-2025E lieferte bei den in Kapitel 2.5 (test results) beschriebenen Messungen im Rahmen der eigenen Messunsicherheit genaue Ergebnisse bei der Messungen der gemittelten Signalstärke von gepulsten und ungepulsten Signalen, wie sie in diversen EMF-Messrichtlinien gefordert sind. Er ist daher zur EMVU-Messung bei allen getesteten Signalarten einsetzbar. Test engineer: Thomas Vogler Date Name Signature Test engineer: Christophe Weiden Date Name Signature Technical responsibility for area of testing: Ralf Hoehn Date Name Signature As of Page 3 of 74
4 1.2 Testing laboratory CETECOM ICT Services GmbH Untertuerkheimer Straße 6-10, Saarbruecken Germany Telephone: Fax: Internet: [email protected] Test location, if different from CETECOM ICT Services GmbH Name: --- Street: --- Town: --- Country: --- Phone: --- Fax: Details of applicant Name: Street: Town: Country: Internet Aaronia AG Gewerbegebiet Aaronia AG Strickscheid Germany [email protected] Contact: Mr. Thorsten Chmielus Telephone: Application details Date of receipt of application: Date of receipt of test item: Start/Date of test: End of test: As of Page 4 of 74
5 1.5 Test item Description of the test item: Type identification: portable field meter and spectrum analyzer SPECTRAN HF-2025E Serial number: Manufacturer: Name: Street: Town: Country: Aaronia AG Gewerbegebiet Aaronia AG Strickscheid Germany additional information on test item : (acc. to applicant s information) frequency range : level range : measurement level uncertainty : MHz dbm +/- 4 db typical additional equipment : Logarithmic-Periodic directional antenna HyperLog 60xxx (S/N ) 1.6 Test specification(s) The device under test has not been tested according to a special test specification. As of Page 5 of 74
6 2 Technical test 2.1 Summary of test results No deviations from the technical specification(s) for the applied test cases were ascertained in the course of the tests performed. The deviations as specified in 2.5 were ascertained in the course of the tests performed. 2.2 Test environment General Environment conditions in the test area are as follows: Ambient temperature: 20 C 24 C Tissue simulating liquid: 20 C 24 C Humidity: 40% 50% 2.3 Measurement and test set-up The measurement system is described in chapter 2.4. A description of positioning and test signal control can be found in chapter 2.5 together with the test results. As of Page 6 of 74
7 2.4 Measurement system Broadband Electromagnetic Field Test system A state of the art Broadband Electromagnetic Field Test system was used. The probes of the system are fitted with three sensors which measure the field strength of the X, Y and Z plane directions separately. The field strength is calculated by the instrument's processor by summing the squares of the three measured values. The frequency range 5 Hz to 40 GHz is covered. Depending from the used probe type E and H or E-field only is detectable. EHP50B 5 Hz to 100 khz E and H-Field EP khz to 3 GHz E-Field EP GHz to 40 GHz E-Field H-Field measurements from 100 khz to 40 GHz have not been performed due to limitations of the measurement equipment Test equipment list No Manufacturer Device Type Serial number 1 PMM Electric and Magnetic PMM J10945 Field Meter 2 PMM Electric Field Probe EP J khz - 3 GHz 3 PMM Electric Field Probe EP J MHz - 40 GHz 4 PMM Optical Repeater OR J PMM Electric and Magnetic EHP-50B 241WM30404 Field Analyser 5 Hz 100 khz 6 Frankonia 10 m EMC-Chamber Chamber F n.a. Last calibration date of PMM equipment : May 24, 2004 As of Page 7 of 74
8 2.4.3 Averaging For time efficient testing the average function last 32 samples was used. With some spot checks was shown, that caused by the time structure of the measured responses, the results did not change with 6- minutes averaging Uncertainties The probe uncertainties stated by the manufacturer are considered to be the main relevant and dominant issues. Probe type EHP 50B EP 330 EP 408 Expanded Uncertainties k=2 in % Electrical Magn. Flux Field Density (B) Strength in µt in V/m Frequency Range Hz 100 khz Hz khz 3 GHz MHz MHz 300 MHz MHz 18 GHz GHz 40 GHz As of Page 8 of 74
9 2.4.5 Definition of test distances Due to the mechanical concept of the used probe a distance between DUT surface and electrical centre of the probe antennas remains. This distance needs to be taken into account when measuring EMF at small spatial resolution. Probe type EHP 50B Magn. Field Distance in mm max. 4 cm Electrical Field distance in mm max. 4cm EP 330 EP max. 2 cm - - max. 1.5 cm - As of Page 9 of 74
10 2.4.6 Anechoic chamber B As of Page 10 of 74
11 No. Instrument/Ancillary Manufacturer Type Serial-No. Internal identification Radiated immunity in chamber B 1. Fully anechoic chamber B Frankonia 2. Controll computer Tecline DT6/400H1300P II FW Software R&S EMS-K a 4. Position control unit Deisel HD /322/ Antenna positioner (Horn) Deisel MA /308 Bj a 6. Antenna positioner (Log. Per.) Deisel MA /309 Bj b 7. Stacked double Log.-Per Antenna ( MHz) Schwarzbeck STLP9128E 9128E Horn antenna GHz Amplifire research AT4002 ar Isotropic Field Probe HI-6005 Holaday Directional coupler unit R&S DCU / Amplifier MHz Amplifire research 250L Amplifier 25MHz-1GHz Ampl. Res. 100W1000M Amplifier GHz Bonn. BLMA Data Processing / Interface Unit EMCO Signal generator MHz R&S SMH / Signal generator MHz R&S SMHU / Relay matrix R&S PSN / Power meter R&S URV / power sensor, insertion unit A R&S URV5-Z / power sensor, insertion unit B R&S URV5-Z / power sensor, insertion unit A R&S URV5-Z power sensor, insertion unit B R&S URV5-Z Bus extender National Instruments GPIB Observation equipment, audio rack 1 in chamber B 24. Connection field R&S / Relay matrix Electronics 4874B ohne Corporation 26. Bus Extender National Instruments GPIB Control computer Software ICT 29. band pass B&K Measurement amplifier B&K Optical fibre microphone system Sennheiser keine 32. Artificial mouth with AF B&K transformer 33. sound calibrator CR511F CYRRUS empty 35. empty Additional equipment utilized Device Type Serial number last calibration Signal Analyzer Rhode & Schwarz FSIQ / Vector Signal Generator SMJ100A n/a ESG Vector Signal Agilent E4438C MY n/a Generator Base station simulator CMU n/a Wideband horn EMCO n/a n/a : calibration not necessary for test cases in chapter 2.5. A regular in-house verification of test equipment is performed annually. For all relevant calibration information see annex 2. As of Page 11 of 74
12 2.5 Test results The following test cases were applied during the comparative measurements of the Aaronia HF-2025E. 1. Comparative measurement of an electromagnetic field generated inside an anechoic chamber with broadband field probe (CW signal) at 800, 1000 and 2000 MHz. 2. Measurement of the same field at different frequencies. 3. Comparative measurement of an artificially generated GSM and UMTS signal with broadband field probe. 4. Comparative measurement of an artificially generated DECT signal with broadband field probe. 5. Comparative outdoor measurement of a GSM base station with broadband field probe. 6. Comparative outdoor measurement of external Log-Per antenna with FSIQ26 spectrum analyzer using the signal of a GSM base station. 7. Comparative conducted measurement of different modulated signals with FSIQ26 spectrum analyzer. The target of the above described tests were : - proving the ability to measure electromagnetic fields in RMS mode of pulsed and unpulsed signals - showing usability of external logarithmic-periodic antenna - checking level accuracy (conducted) As of Page 12 of 74
13 2.5.1 Comparative measurement of an electromagnetic field generated inside an anechoic chamber with broadband field probe (CW signal) at 800, 1000 and 2000 MHz. The test was performed with the following settings : Aaronia SPECTRAN HF-2025E PMM 8053 Center frequency : transmit frequency Span : 20 MHz Filter : 3 MHz Sample time : 50 ms Max hold mode : on wide band probe EP MHz RMS : last 32 samples Test results : generated E-field / V/m frequency / MHz level PMM / V/m level SPECTRAN / V/m Test signal : CW within a frequency range of MHz. Field strength deviations of up to 1.5 V/m to target values are within tolerances of generated field. For good comparability of test results during these parallel measurements it was necessary to position the PMM probe close to the logarithmic-periodic antenna. Note : 6 V/m already exceed nominal maximum measurement level of SPECTRAN without additional attenuator. As of Page 13 of 74
14 test position of log-per antenna As of Page 14 of 74
15 test position of log-per antenna (rear view), with observation camera at the left hand side and horizontally polarized field generating Schwarzbeck antenna in background. As of Page 15 of 74
16 SPECTRAN HF-2025E As of Page 16 of 74
17 Aaronia HyperLOG log-per antenna and PMM broadband field-probe As of Page 17 of 74
18 Measurement position of wide band field probe during parallel measurements with SPECTRAN As of Page 18 of 74
19 position of both handsets for monitoring As of Page 19 of 74
20 example screenshot : readout values at 1 V/m and 800 MHz (1.100 and 1.06 V/m) As of Page 20 of 74
21 example screenshot : readout values at 3 V/m and 1 GHz (3.024 and 3.10 V/m) As of Page 21 of 74
22 example screenshot : readout values at 6 V/m and 2 GHz (6.866 and 7.18 V/m) As of Page 22 of 74
23 2.5.2 Measurement of electromagnetic field at different frequencies and levels. These measurements have been performed under the same conditions as in chapter without the PMM 8053 field probe in place because measurement accuracy between PMM and SPECTRAN showed no difference. Note : Differing measurement values are caused by a slightly different position of the Log-Per antenna. The electromagnetic field applied during this test has been calibrated for standard immunity tests and therefore spatial differences are allowed. The following test results using levels of 1, 3 and 6 V/m, as well as some random levels has been performed to show constant measurement behaviour of the SPECTRAN used together with the Log-Per antenna. The center frequency of SPECTRAN was adjusted before each frequency change. Frequenz / MHz Level V/m Readout value SPECTRAN in Max Hold mode / V/m , , , , , , , ,5 3, , , , ,5 2, , , , , , , , , , , , , ,5 4, ,22 Test signal : CW within a frequency range of MHz. Field strength deviations of up to 1.5 V/m to target values are within tolerances of generated field. As of Page 23 of 74
24 2.5.3 Comparative measurement of an artificially generated GSM and UMTS signal with broadband field probe. test description : An EMCO wideband horn was fed with a GSM or UMTS signal generated by a CMU base station simulator. The electromagnetic field in front of the horn was measured simultaneously both with the PMM broadband probe and the SPECTRAN HF-2025E at a distance of 2 meters and at a small distance between each other. This has been a check under normal user conditions (no shielded chamber, unknown field structure, possible reflections) The test was performed with the following settings : at 2140 MHz (WCDMA): Aaronia SPECTRAN HF-2025E PMM 8053 Hotkey UMTS : Center frequency : 2140 MHz wide band probe EP MHz Span : 60 MHz RMS : last 32 samples Filter : 3 MHz Sample time : 50 ms Max hold mode : on at 900 MHz (GSM) Aaronia SPECTRAN HF-2025E PMM 8053 Hotkey GSM 900 : Center frequency : 940 MHz wide band probe EP MHz Span : 40 MHz RMS : last 32 samples Filter : 3 MHz Sample time : 100 ms Max hold mode : on Test result : frequency / MHz level PMM / V/m level SPECTRAN / V/m GSM 900 MHz band 2.34 (RMS) UMTS 2100 MHz band 4.36 (RMS) As of Page 24 of 74
25 EMCO 1-15 GHz wideband horn (vertically polarized) As of Page 25 of 74
26 R&S CMU 200 base station simulator for generating GSM and WCDMA signals As of Page 26 of 74
27 measured values in GSM900 MHz band (Log-per antenna needs to have same polarization as transmitted signal to measure maximum values) As of Page 27 of 74
28 measured values in UMTS band As of Page 28 of 74
29 2.5.4 Comparative measurement of an artificially generated DECT signal with broadband field probe. test description : An EMCO wideband horn was fed with a DECT signal generated by the E4438C vector signal generator. The electromagnetic field in front of the horn was measured simultaneously both with the PMM broadband probe and the SPECTRAN at a distance of 2 meters and at a small distance between each other. This has been a check under normal user conditions (no shielded chamber, unknown field structure, possible reflections) The test was performed with the following settings : at 1890 MHz (DECT): Aaronia SPECTRAN HF-2025E PMM 8053 Hotkey DECT : Center frequency : 1890 MHz wide band probe EP MHz Span : 20 MHz RMS : last 32 samples Filter : 3 MHz Sample time : 50 ms Max hold mode : on Pulse mode on : Test result : frequency / MHz level PMM / V/m level SPECTRAN / V/m DECT 1900 MHz band 0.81 (RMS) Remark : The test was performed with all 12 PP DECT channels activated to have direct comparison between PEP RMS-Value of Sprectran and averaged RMS of PMM As of Page 29 of 74
30 EMCO 1-15 GHz wideband horn (vertically polarized) and measured field strength As of Page 30 of 74
31 E4478C signal generator for generating DECT signal with all PP channels active As of Page 31 of 74
32 2.5.5 Comparative outdoor measurement of a GSM base station with broadband field probe. The test was performed with the following settings : Aaronia SPECTRAN HF-2025E PMM 8053 Hotkey GSM 900 : Center frequency : 940 MHz wide band probe EP MHz Span : 40 MHz RMS : last 32 samples Filter : 3 MHz Sample time : 100 ms Max hold : on Test result : frequency / MHz level PMM / V/m level SPECTRAN / V/m GSM 900 MHz band 0.52 (RMS) The PMM 8053 measures both 900 MHz and 1800 MHz at the same time. Chapter shows that GSM 900 is the dominating signal so that influences of GSM 1800 on the measurement result of the broadband probe are small. As of Page 32 of 74
33 GSM base station 300 meters away from CETECOM ICT Services GmbH. As of Page 33 of 74
34 Displayed values of both SPECTRAN and PMM 8053 at detected position with maximum signal. As of Page 34 of 74
35 2.5.6 Comparative outdoor measurement of external Log-Per antenna with FSIQ26 spectrum analyzer using the signal of a GSM base station. The test was performed with the following settings : at 900 MHz : Aaronia SPECTRAN HF-2025E Hotkey GSM 900 : Center frequency : 940 MHz Span : 40 MHz Filter : 3 MHz Sample time : 100 ms Display mode : max hold FSIQ26 Center frequency : 940 MHz Span : 40 MHz Filter : 3 MHz RMS detector : on Display mode : max hold at 1800 MHz : Aaronia SPECTRAN HF-2025E Hotkey GSM 1800 : Center frequency : 1840 MHz Span : 80 MHz Filter : 3 MHz Sample time : 100 ms Display mode : max hold FSIQ26 Center frequency : 1840 MHz Span : 80 MHz Filter : 3 MHz RMS detector : on Display mode : max hold note : cable attenuation not corrected with both devices Test result : frequency / MHz peak level FSIQ26 / dbm peak level SPECTRAN / dbm GSM 900 MHz band (no decimals available) GSM 1800 MHz band (no decimals available) As of Page 35 of 74
36 Aaronia Log-Per antenna connected to SPECTRAN with GSM base station in background As of Page 36 of 74
37 Aaronia Log-per antenna connected to FSIQ26 and detected multi-carrier GSM signal in higher resolution at highest detected level to prove existence of a real GSM signal. Ref Lvl 0 dbm 0 Marker 1 [T1] dbm MHz RBW 500 khz VBW 500 khz SWT 5 ms RF Att 30 db Unit dbm A MAX 1MA Center 950 MHz Date: 24.APR :54:56 6 MHz/ Span 60 MHz As of Page 37 of 74
38 comparative test result with identical antenna position at GSM 900 (not maximum!) (no display photo of FSIQ26 available due to poor readability in outdoor conditions.) Ref Lvl -10 dbm -10 Marker 1 [T1] dbm MHz RBW 3 MHz VBW 10 MHz SWT 5 ms RF Att 20 db Unit dbm A MAX 1RM Center 940 MHz Date: 24.APR :46:04 4 MHz/ Span 40 MHz As of Page 38 of 74
39 comparative test result with identical antenna position at GSM 1800 (not maximum) (no display photo of FSIQ26 available due to poor readability in outdoor conditions.) As of Page 39 of 74
40 2.5.7 Comparative conducted measurement of different modulated signals with FSIQ26 spectrum analyzer. Device settings at 900 MHz : Aaronia SPECTRAN HF-2025E Hotkey GSM 900 : Center frequency : 940 MHz Span : 40 MHz Filter : 3 MHz Sample time : 100 ms Display mode : max hold at 1800 MHz : Aaronia SPECTRAN HF-2025E Hotkey GSM 1800 : Center frequency : 1840 MHz Span : 80 MHz Filter : 3 MHz Sample time : 100 ms Display mode : max hold at 1890 MHz (DECT): Aaronia SPECTRAN HF-2025E Hotkey DECT : Center frequency : 1890 MHz Span : 20 MHz Filter : 3 MHz Sample time : 50 ms Pulse mode : on Display mode : max hold at 2140 MHz (WCDMA): Aaronia SPECTRAN HF-2025E Hotkey UMTS : Center frequency : 2140 MHz Span : 60 MHz Filter : 3 MHz Sample time : 50 ms Display mode : max hold FSIQ26 Center frequency : 940 MHz Span : 40 MHz Filter : 3 MHz RMS detector : on Display mode : max hold FSIQ26 Center frequency : 1840 MHz Span : 80 MHz Filter : 2 MHz RMS detector : on Display mode : max hold FSIQ26 Center frequency : 1890 MHz Span : 20 MHz Filter : 3 MHz / 0.5 MHz RMS detector : on Display mode : max hold FSIQ26 Center frequency : 2140 MHz Span : 60 MHz Filter : 3 MHz RMS detector : on Display mode : max hold As of Page 40 of 74
41 at 2450 MHz (WLAN): Aaronia SPECTRAN HF-2025E Center frequency : 2450 MHz Span : 80 MHz Filter : 3 MHz Sample time : 100 ms Display mode : max hold FSIQ26 Center frequency : 2450 MHz Span : 80 MHz Filter : 3 MHz RMS detector : on Display mode : max hold Remark : Filter bandwidth of FSIQ26 was set to the most appropriate values during the measurements. Changes of different settings (RBW, span etc.) shown on the picture had no influence on the measurement results. FSIQ was used with 10 db attenuation at signal input. Summary conducted power measurements of average channel power : note : cable attenuation included in measurement values. modulation setting output power peak level FSIQ26 /dbm DECT 1 timeslot -20 dbm DECT 1 timeslot 0 dbm GSM timeslot, -20 dbm unframed GSM timeslot, -20 dbm unframed WCDMA 3.84 Mcps -20 dbm (UMTS) WLAN b CCK, framed -20 dbm Remark : SPECTRAN does not display decimals in dbm mode. peak level SPECTRAN /dbm As of Page 41 of 74
42 test set-up As of Page 42 of 74
43 DECT setting (downlink only) As of Page 43 of 74
44 DECT setting (uplink switched on additionally) As of Page 44 of 74
45 DECT peak value SPECTRAN As of Page 45 of 74
46 DECT peak value FSIQ 26 (with 20 MHz span) As of Page 46 of 74
47 DECT peak value FSIQ 26 with 1.6 MHz span (channel bandwith) : no difference in peak dbm value As of Page 47 of 74
48 DECT timeslot overview (1 uplink and 1 downlink) shown on SPECTRAN LCS software As of Page 48 of 74
49 DECT setting and measured peak value of SPECTRAN (output power 0 dbm) As of Page 49 of 74
50 DECT peak value FSIQ 26 (output power 0 dbm) As of Page 50 of 74
51 GSM 900 setting As of Page 51 of 74
52 GSM peak value SPECTRAN As of Page 52 of 74
53 GSM peak value FSIQ 26 As of Page 53 of 74
54 GSM1800 setting As of Page 54 of 74
55 GSM1800 peak value SPECTRAN Alias effects in spectrum display had no influence on measured peak level accuracy. As of Page 55 of 74
56 GSM1800 peak value FSIQ 26 As of Page 56 of 74
57 WCDMA setting As of Page 57 of 74
58 WCDMA peak value SPECTRAN Alias effects in spectrum display had no influence on measured peak level accuracy. As of Page 58 of 74
59 WCDMA peak value FSIQ 26 As of Page 59 of 74
60 WLAN setting As of Page 60 of 74
61 WLAN peak value SPECTRAN (the measurement has been performed in the 2.4 GHz band, the picture has been taken before a full display update so that a frequency marker of the previous measurement at 1800 MHz appears.) As of Page 61 of 74
62 WLAN peak value FSIQ 26 As of Page 62 of 74
63 Annex 1 Photo documentation of test sample and PMM 8053 Picture no. 1 DUT - front view As of Page 63 of 74
64 Picture no. 2 DUT - rear view As of Page 64 of 74
65 Picture no. 3 DUT - label As of Page 65 of 74
66 Picture no. 4 log-periodic antenna ( top view) As of Page 66 of 74
67 Picture no. 5 log-periodic antenna (underside view) As of Page 67 of 74
68 Picture no. 6 log-periodic antenna - serial number As of Page 68 of 74
69 Picture no. 7 calibration label As of Page 69 of 74
70 Picture no. 8 calibration certificate of log-periodic antenna As of Page 70 of 74
71 Picture no. 9 test equipment : field meter, E-field probes EP330 and EP408, optical repeater, probe EHP50B (only EP330 was used for the measurements described in chapter 2.5) As of Page 71 of 74
72 Annex 2 Calibration information As of Page 72 of 74
73 As of Page 73 of 74
74 Detailed calibration information is available on request. As of Page 74 of 74
RADIO TEST REPORT. EN 300 440-2 SRD radio equipment to be used in the 1 GHz to 40 GHz frequency range. DAkkS - Registration number : D-PL-12092-01-01
RADIO TEST REPORT SRD radio equipment to be used in the 1 GHz to 40 GHz frequency range Report Reference No.... : G0M-1206-2043-TEU440G-V01 Testing Laboratory... : Eurofins Product Service GmbH Address...
EN 301 489-17 v1.2.1 TEST REPORT FOR. 802.11ag/Draft 802.11n WLAN PCI-E Mini Card MODEL NUMBER: BCM94322MC REPORT NUMBER: 07U11529-5
EN 301 489-17 v1.2.1 TEST REPORT FOR 802.11ag/Draft 802.11n WLAN PCI-E Mini Card MODEL NUMBER: BCM94322MC REPORT NUMBER: 07U11529-5 ISSUE DATE: JANUARY 29, 2008 Prepared for BROADCOM CORPORATION 190 MATHILDA
Fast and Accurate Test of Mobile Phone Boards
Products: R&S FSP Fast and Accurate Test of Mobile Phone Boards Short test times in conjunction with accurate and repeatable measurement results are essential when testing and calibrating mobile phones
Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet
Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers Data Sheet Specifications Specifications are only valid for the stated operating frequency, and apply over 0 C to +55 C unless otherwise
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity
Evaluating Cell Phone and Personal Communications Equipment and their EMC Effects on Automotive Audio and In-Cabin Modules
Evaluating Cell Phone and Personal Communications Equipment and their EMC Effects on Automotive Audio and In-Cabin Modules Craig W. Fanning Elite Electronic Engineering, Inc. 1516 Centre Circle Downers
Measurement of RF Emissions from a Final Coat Electronics Corrosion Module
Engineering Test Report No. 37802-02 Rev. A Measurement of RF Emissions from a Final Coat Electronics Corrosion Module For : Canadian Auto Preservation, Inc. 390 Bradwick Drive Concord, Ontario CANADA
R&S ZNBT8 Vector Network Analyzer Specifications
ZNBT8_dat-sw_en_3606-9727-22_v0200_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S ZNBT8 Vector Network Analyzer Specifications 20.05.2014 08:39:42 CONTENTS Definitions... 3 Measurement range... 4
Continuous, remote monitoring and logging of electromagnetic fields
MULTI-BAND EMF AREA MONITOR AMB-8057 Continuous, remote monitoring and logging of electromagnetic fields Discrimination of GSM and UMTS contributions to total emf values Magnetic fields monitoring from
AN003. Evaluation of Anechoic Chambers For EMC Measurements Using A CNE. Overview. CNE III 9kHz to 2GHz Conducted and radiated output
Evaluation of Anechoic Chambers For EMC Measurements Using A CNE Overview Anechoic chambers are used for EMC testing primarily for radiated emissions (RE) and radiated immunity (RI) in the frequency range
R&S ZNC Vector Network Analyzer Specifications
ZNC3_dat-sw_en_5214-5610-22_v0300_cover.indd 1 Data Sheet 03.00 Test & Measurement R&S ZNC Vector Network Analyzer Specifications 04.09.2012 13:39:47 CONTENTS Definitions... 3 Measurement range... 4 Measurement
Report Of. Shielding Effectiveness Test For. SafeSleeve Radiation Shielding Technology. Test Date(s): June 07 June 09, 2015
Report Of Shielding Effectiveness Test For SafeSleeve Radiation Shielding Technology Test Date(s): June 07 June 09, 2015 Issue Date: June 10, 2015 UST Project No: 15-0137 Total Number of Pages Contained
AN1200.04. Application Note: FCC Regulations for ISM Band Devices: 902-928 MHz. FCC Regulations for ISM Band Devices: 902-928 MHz
AN1200.04 Application Note: FCC Regulations for ISM Band Devices: Copyright Semtech 2006 1 of 15 www.semtech.com 1 Table of Contents 1 Table of Contents...2 1.1 Index of Figures...2 1.2 Index of Tables...2
How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless)
1 2 The range of RF spans 3 KHz (3000 Hz) to 300 GHz (300 million Hz) Frequencies of RF devices range from the low frequency AM broadcasts (80 MHz) to higher frequency mobile phones (1900 MHz) smart meters
Test result: SHENZHEN TIMEWAY TECHNOLOGY CONSULTING CO LTD. ISO/IEC17025Accredited Lab. Report No: FCC 1106180 File reference No: 2011-07-18
Report No: FCC 1106180 File reference No: 2011-07-18 ISO/IEC17025Accredited Lab. Applicant: FUJIAN NAN AN BAOFENG ELECTRONICS CO., LTD Product: Model No: Trademark: Two Way Radio UV-3R BAOFENG Test Standards:
Electromagnetic Compatibility Test Report Test results of Floww equipment
Electromagnetic Compatibility Test Report Test results of Floww equipment Reference number Status test report Brand Model number : 10C00357RPT01.doc : Final : Floww : mobilefloww screenfloww pocketfloww
MATRIX TECHNICAL NOTES
200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR
HERMES: Human Exposure and Radiation Monitoring of Electromagnetic Sources. www.hermes-program.gr
HERMES Program: The Greek Experience of an Electromagnetic Radiation Monitoring System A. Manassas, A. Boursianis, T. Samaras and J. N. Sahalos HERMES: Human Exposure and Radiation Monitoring of Electromagnetic
RF EXPOSURE LIMITS AND TESTING REQUIREMENTS
RF EXPOSURE LIMITS AND TESTING REQUIREMENTS Jay Moulton Vice President March 12, 2013 1 OVERVIEW Specific Absorption Rate (SAR) and Maximum Permissible Exposure (MPE) Standards and Limits Evaluation Methods
Continuous, remote monitoring and logging of electromagnetic fields
MULTI-BAND EMF AREA MONITOR AMB-8057-03/G Continuous, remote monitoring and logging of electromagnetic fields Discrimination of GSM and UMTS contributions to total EMF values Electric field monitoring
The tradescantia micronucleus bioassay:
Investigation of the mutagenicity of electromagnetic fields using the tradescantia micronucleus bioassay H. Lehmann 1, C. Pickl 2, M. Urech 3 1 Environment & Electromagnetic Compatibility, Swisscom Innovations,
Application Note Noise Frequently Asked Questions
: What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random
Testing of 10 GHz Instantaneous Bandwidth RF Spectrum Monitoring at Idaho National Labs
Testing of 10 GHz Instantaneous Bandwidth RF Spectrum Monitoring at Idaho National Labs Scott H. Bekker, Aaron S. Traxinger, Colton R. Stiffler, Alex J. Woidtke, Michael D. Chase, Wm. Randall Babbitt,
R&S R-Line Compact Test Chamber Precise measurements of radiated emissions of wireless terminals 800 MHz to 18 GHz
Test & Measurement Data Sheet 02.00 R&S R-Line Compact Test Chamber Precise measurements of radiated emissions of wireless terminals 800 MHz to 18 GHz R&S R-Line Compact Test Chamber At a glance The R-Line
R&S ZVA Vector Network Analyzer Specifications
www.atecorp.com 8-44-ATEC (2832) R&S ZVA Vector Network Analyzer Specifications ZVA_dat-sw_en_523-568-22_v_cover.indd Data Sheet. E stablished 98 Advanced Test Equipment Rentals Test & Measurement 4.5.22
ETSI EN 300 328-1 V1.2.2 (2000-07)
EN 300 328-1 V1.2.2 (2000-07) European Standard (Telecommunications series) Electromagnetic compatibility and Radio spectrum Matters (ERM); Wideband Transmission systems; data transmission equipment operating
TEST REPORT EN 55014-2 (1997) +A1 (2001)
Page 1 of 23 TEST REPORT EN 55014-2 (1997) +A1 (2001) Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus Part 2: Immunity Report Reference No....
Impedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
Real-life Applications of ICNIRP Guidelines to Various Human EMF Exposure Issues
Technical Forum Overview and Latest Development of Standards in Human Exposure to Electromagnetic Fields (EMF) Real-life Applications of ICNIRP Guidelines to Various Human EMF Exposure Issues Dr. Brian
Microwave Exposure from Mobile Phones and Base Stations in Sweden
Microwave Exposure from Mobile Phones and Base Stations in Sweden Prof. Dr. Yngve Hamnerius and Thomas Uddmar Chalmers University of Technology, Department of Electromagnetics, SE 41 96 Goteborg, Sweden,
SAMSUNG GT-S5369 (mobile phone) Serial number: IMEI 358858043716587. Mr. Antoine SAMAKH and Mr. Mathieu SAMAKH 92100 FRANCE
R052-DAS-12-104374-7/A Ed. 1 This report cancels and replaces the test report R052-DAS-12-104374-7/A Ed. 0 SAR TEST REPORT According to the standard: EN 62209-1: 2006 Equipment under test: Antenna patch
Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics
Jeff Thomas Tom Holmes Terri Hightower Learn RF Spectrum Analysis Basics Learning Objectives Name the major measurement strengths of a swept-tuned spectrum analyzer Explain the importance of frequency
Selective measurement of high frequency electromagnetic
Selective Radiation Meter SRM-3006 Selective measurement of high frequency electromagnetic fields Complete, easy to use test system, consisting of a base unit and measuring antennas, for non-directional
Partial GSM TEST - REPORT
EUROFINS PRODUCT SERVICE GMBH Partial GSM TEST - REPORT EN 301 511 GSM/GPRS/GPS tracking unit FOX-LT-IP Test report no.: G0M20904-2307-T-51 Eurofins Product Service GmbH Storkower Str. 38c, 15526 Reichenwalde,
Rated Power(W) 8W 2. EG-LED0840-01 8W 3. EG-LED1027-01 10W
14713221 001 Seite 2 von 37 Page 2 of 37 Model List: No Model Rated Voltage(V) 1. EG-LED0827-01 Rated Power(W) 8W 2. EG-LED0840-01 8W 3. EG-LED1027-01 10W 4. EG-LED1040-01 AC 100-240V, 10W 5. EG-LED1027-02
Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform
Application Note Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform By: Richard Komar Introduction With the rapid development of wireless technologies, it has become
Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands)
Issue 3 June 2009 Spectrum Management and Telecommunications Radio Standards Specification Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands) Aussi disponible
SRM-3006 Selective Radiation Meter
SRM-3006 Selective Radiation Meter Complete Solution for Selective Measurement of RF and Microwave Electromagnetic Fields Isotropic and Single-Axis Measurements from 9 khz to 6 GHz Excellent Immunity for
Measurement, analysis, and monitoring of RF signals
NRA-2500, NRA-3000 and NRA-6000 Narda Remote Spectrum Analyzer Measurement, analysis, and monitoring of RF signals 19" rack mountable Spectrum Analyzer for remote controlled measurements and analysis of
AM/FM/ϕM Measurement Demodulator FS-K7
Data sheet Version 02.00 AM/FM/ϕM Measurement Demodulator FS-K7 July 2005 for the Analyzers FSQ/FSU/FSP and the Test Receivers ESCI/ESPI AM/FM/ϕM demodulator for measuring analog modulation parameters
TEST REPORT FROM RFI GLOBAL SERVICES LTD
FROM RFI GLOBAL SERVICES LTD Partial Test of: BTM411 To: EN 300 328 V1.7.1 (2006-10) Test Report Serial No: RFI/RPT1/RP76018JD01A This Test Report Is Issued Under The Authority Of Brian Watson, Operations
Minimum requirements for DVB-T receiving antennas for portable indoor and portable outdoor reception
Deutsche TV Platform Minimum requirements for DVB-T receiving antennas for portable indoor and portable outdoor reception compiled by Working Group: DVB-T launch (a working group of the Deutsche TV Platform)
Measurement of Adjacent Channel Leakage Power on 3GPP W-CDMA Signals with the FSP
Products: Spectrum Analyzer FSP Measurement of Adjacent Channel Leakage Power on 3GPP W-CDMA Signals with the FSP This application note explains the concept of Adjacent Channel Leakage Ratio (ACLR) measurement
FCC Part 15C Compliance Test Report
Test Report no.: Cph_FCC_732_5.doc Date of Report: 14-8-27 Number of pages: 42 Customer s Contact person: Ole Soerensen Testing laboratory: Customer: Nokia Corporation FCC listing no.: IC recognition no.:
Selected Radio Frequency Exposure Limits
ENVIRONMENT, SAFETY & HEALTH DIVISION Chapter 50: Non-ionizing Radiation Selected Radio Frequency Exposure Limits Product ID: 94 Revision ID: 1736 Date published: 30 June 2015 Date effective: 30 June 2015
Analysis of Immunity by RF Wireless Communication Signals
64 PIERS Proceedings, Guangzhou, China, August 25 28, 2014 Analysis of Immunity by RF Wireless Communication Signals Hongsik Keum 1, Jungyu Yang 2, and Heung-Gyoon Ryu 3 1 EletroMagneticwave Technology
Broadband over Power Line (BPL) Test Procedures and Equipment Authorization
Broadband over Power Line (BPL) Test Procedures and Equipment Authorization Andy Leimer Equipment Authorization Branch Federal Communications Commission Office of Engineering and Technology Laboratory
Handheld Spectrum Analyzer
9101 Handheld Spectrum Analyzer The 9101 Handheld Spectrum Analyzer provides RF engineers with the excellent performance of a workbench analyzer in a handheld form, at a competitive price. Highlights Covering
2398 9 khz to 2.7 GHz Spectrum Analyzer
Spectrum Analyzers 2398 9 khz to 2.7 GHz Spectrum Analyzer A breakthrough in high performance spectrum analysis, combining cost effectiveness and portability in a new lightweight instrument 9 khz to 2.7
MYRICOM, INC. TEST REPORT FOR THE SERVER RACK, 10G-PCIE2-8B2-2S & 10G-PCIE2-8B2-2QP FCC PART 15 SUBPART B SECTIONS 15.107 AND 15.109 CLASS A TESTING
TESTING CERT #803.01, 803.02, 803.05, 803.06 MYRICOM, INC. TEST REPORT FOR THE SERVER RACK, 10G-PCIE2-8B2-2S & 10G-PCIE2-8B2-2QP FCC PART 15 SUBPART B SECTIONS 15.107 AND 15.109 CLASS A TESTING DATE OF
SAR Test Report: Type Number FAD-3022012-BV; FCC ID PY7FD022012; IC 4170B-FD022012
1 (34) Report issued by an Accredited Laboratory Rapport utfärdad av ett ackrediterat laboratorium This laboratory is accredited to SS-EN ISO/IEC 17025 (SWEDAC accreditation no. 1761). SAR Test Report:
A Guide to Calibrating Your Spectrum Analyzer
A Guide to Calibrating Your Application Note Introduction As a technician or engineer who works with electronics, you rely on your spectrum analyzer to verify that the devices you design, manufacture,
TCB Workshop. Unlicensed National Information Infrastructure Devices (U-NII)/Dynamic Frequency Selection (DFS)
1 TCB Workshop Unlicensed National Information Infrastructure Devices (U-NII)/Dynamic Frequency Selection (DFS) Andrew Leimer Office of Engineering and Technology/Equipment Authorization Branch FCC Laboratory
September 26, 2000 IMST GmbH Carl-Friedrich-Gauß-Str. 2 D-47475 Kamp-Lintfort
TTI-P-G 158 Report Dosimetric Assessment of the Mobile Ericsson T18z with an without the PAM System Technologies devices According to the American FCC Requirements September 26, 2000 IMST GmbH Carl-Friedrich-Gauß-Str.
ETSI EN 302 774 V1.2.1 (2012-02)
EN 302 774 V1.2.1 (2012-02) Harmonized European Standard Broadband Wireless Access Systems (BWA) in the 3 400 MHz to 3 800 MHz frequency band; Base Stations; Harmonized EN covering the essential requirements
Users Manual FA106 / FA306 version 1.0. Field Analyzer. FA106 and FA306. Users Manual
Field Analyzer FA106 and FA306 Users Manual 1 Contents Contents...2 Preface...3 Chapter 1 Basic Functions...4 1.1 Turning on/off...4 1.2 Choosing the required measuring mode...4 1.3 Picture of front panel...5
HP 70950B OPTICAL SPECTRUM ANALYZER
HP 71450B, 71451B, and 71452B Optical Spectrum Analyzers Technical Specifications Spectral Measurements from 600 to 1700 nm HP 70950B OPTICAL SPECTRUM ANALYZER OPTICAL INPUT The HP 71450B, 71451B, and
Standard: EN 61000-4-2 :1995, EN 61000-4-3 :1996, ENV 50204 :1993, & EN 61000-4-4 :1995 Solid State Energy Meter
Standard: EN 61000-4-2 :1995, EN 61000-4-3 :1996, ENV 50204 :1993, & EN 61000-4-4 :1995 Model: Solid State Energy Meter Prepared for: Analog Devices, Inc. 804 Woburn Street Wilmington, MA 01887 Date of
NI USB-5681 RF Power Meter Specifications
NI USB-568 RF Power Meter Specifications General This document lists specifications for the NI USB-568 RF power meter. Minimum or maximum specifications are warranted under the following conditions: hour
Keysight Technologies N9320B RF Spectrum Analyzer
Keysight Technologies N9320B RF Spectrum Analyzer 9 khz to 3.0 GHz Data Sheet Definitions and Conditions The spectrum analyzer will meet its specifications when: It is within its calibration cycle It has
Draft ETSI EN 301 783-1 V1.2.1 (2009-07)
Draft EN 301 783-1 V1.2.1 (2009-07) European Standard (Telecommunications series) Electromagnetic compatibility and Radio spectrum Matters (ERM); Land Mobile Service; Commercially available amateur radio
Keysight N9320B RF Spectrum Analyzer
Keysight N9320B RF Spectrum Analyzer 9 khz to 3.0 GHz Data Sheet 02 Keysight N9320B RF Spectrum Analyzer - Data Sheet Definitions and Conditions Specifications describe the performance of parameters covered
EMC TEST REPORT For. Shenzhen Homelux Security Equipment Co., Ltd. Auto Dialer. Model No.: HX-GD20
EMC TEST REPORT For Shenzhen Homelux Security Equipment Co., Ltd Auto Dialer Model No.: HX-GD20 Prepared for : Shenzhen Homelux Security Equipment Co., Ltd Address : 4 th Floor, Building 5, Jiademaluan
CETECOM ICT Services GmbH Untertürkheimer Straße 6-10, D-66117 Saarbrücken Phone: +49 (0) 681-598-0 Fax:-9075
Untertürkheimer Straße 6-10, D-66117 Saarbrücken Phone: +49 (0) 681-598-0 Fax:-9075 CETECOM ICT Services GmbH Untertürkheimer Str. 6 10 66117 Saarbrücken Germany Test report no. : 1-0510-01-01/08 Type
Product Information S N O. Portable VIP protection CCTV & Alarm System 2
Product Information S N O Portable VIP protection CCTV & Alarm System 2 G O V E R N M E N T A L S E C U R I T Y S O L U T I VIP KIT Rapid Deployment VIP Protection Kit The VIP KIT has been designed to
1975-2004 GOOD WILL Instrument Co., Ltd. All rights reserved.
Introduction to 3GHz Spectrum Analyzer of Instek Front side Features Rear side 1Frequency Adjustment 2Ref. Input 3Ref. Out 4Ext. Trigger 5GPIB 6RS232 11 12 13 7USB 8VGA Output 9Ear Phone DC Input 11 Battery
Pre-Compliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions
PreCompliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions D. Schneider 1*, S. Tenbohlen 1, W. Köhler 1 1 Institute of Power Transmission and
Human Exposure Limits
Human Exposure Limits Session 3 0 Version December 2014 Learning objectives In this session we will: Learn about the international exposure limits for workers and the public Learn about methods for assessing
SHENZHEN GEMBIRD ELECTRONICS LTD EMC REPORT
Shenzhen BST Technology Co., Ltd. SHENZHEN GEMBIRD ELECTRONICS LTD EMC REPORT Prepared For : Product Name: Model : Prepared By : SHENZHEN GEMBIRD ELECTRONICS LTD 5th Floor, Building B, Shifeng Industry
FCC PART 15 SUBPART C EMI MEASUREMENT AND TEST REPORT
FCC PART 15 SUBPART C EMI MEASUREMENT AND TEST REPORT For AeroComm, Inc. 10981 Eicher Drive Lenexa, KS 66219 2003-05-22 This Report Concerns: Original Report Equipment Type: 900MHz Transceiver Test Engineer:
EMC TEST REPORT For. Webcam. Model Number: CAM67U
EMC TEST REPORT For Webcam Model Number: CAM67U Prepared for : Gembird Europe B.V. Wittevrouwen 56, 1358CD Almere, The Netherlands. Phone: +31-36-5211588 Prepared By : Shenzhen Meihua Electronic Technology
R&S ZVL Vector Network Analyzer Specifications
R&S ZVL Vector Network Analyzer Specifications Test & Measurement Data Sheet 08.02 CONTENTS Specifications... 3 Measurement range... 3 Measurement speed... 3 Measurement accuracy... 4 Effective system
Next Generation Test Chambers. Bluetest AB
Next Generation Test Chambers Bluetest Background Founded in 2001 by Professor Kildal at Chalmers University, Gothenburg, Sweden The vision was to perform tests: - Fast - Easy - Cost effectively 2006:
Minimum requirements for DVB-T2 receiving antennas
Minimum requirements for DVB-T2 receiving antennas for portable indoor and portable outdoor reception based on Minimum requirements for DVB-T receiving antennas for portable indoor and portable outdoor
The Salzburg Model: A Precautionary Strategy for Siting of Base Stations
The Salzburg Model: A Precautionary Strategy for Siting of Base Stations Dr. Gerd Oberfeld, Dr. Christoph König Federal State of Salzburg, Public Health Department, Environmental Health P.O. Box 527, A-5010
Network analyzer and spectrum analyzer two in one
R&S ZVL Vector Network Analyzer Network analyzer and spectrum analyzer two in one The R&S ZVL is the lightest and smallest vector network analyzer in its class. On top of this, it can be used as a full-featured
EMC STANDARDS STANDARDS AND STANDARD MAKING BODIES. International. International Electrotechnical Commission (IEC) http://www.iec.
EMC STANDARDS The EMC standards that a particular electronic product must meet depend on the product application (commercial or military) and the country in which the product is to be used. These EMC regulatory
Tektronix RSA306 USB Spectrum Analyzer
Tektronix RSA306 USB Spectrum Analyzer Simple Demos The Demo of the RSA306 is easy. Even you do not have signal generators, devices under test, or demo boards, using the whip antenna provided in box, you
Body SAR and EU. B. Derat
Body SAR and EU B. Derat Overview / Outline New Approach: Directive / Standards Current applicable standard IEC/EN 62209-2:2010 Tablet case study Future of EU body SAR testing Maintenance of IEC 62209-2
SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES
SAR ASSOCIATED WITH THE USE OF HANDS-FREE MOBILE TELEPHONES S.J. Porter, M.H. Capstick, F. Faraci, I.D. Flintoft and A.C. Marvin Department of Electronics, University of York Heslington, York, YO10 5DD,
Test Report. Prepared for: Technologic Systems, Inc. Model: TS-8820
Test Report Prepared for: Technologic Systems, Inc. Model: TS-8820 Description: Single Board Computer with Analog and Digital Inputs and Outputs, Relays, and RS-232/485 Ports To IEC 61000-6-1 (2005-03)
Test Report #: 1981-3 Date: April 8, 2005
Test Report #: 1981-3 Date: April 8, 2005 CERTIFICATE #2316.01 Issued To: Bill Burks American Power Conversion 85 Rangeway Road North Billerica, MA 01862 USA 978-670-2440 Product Name/Description Model
Agilent E4401B, E4402B, E4404B, E4405B, and E4407B ESA-E Series Spectrum Analyzers
Agilent, E4402B, E4404B, E4405B, and E4407B ESA-E Series Spectrum Analyzers Technical Specifications All specifications apply over 0 C to + 55 C unless otherwise noted. The analyzer will meet its specifications
FCC Test Report. FCC Part 22, 24 / RSS 132, 133 FOR: GSM Cellular Telephone with Bluetooth and WiFi. Model #: A1203
FCC Test Report FCC Part 22, 24 / RSS 132, 133 FOR: GSM Cellular Telephone with Bluetooth and WiFi Model #: A1203 Apple Inc. 1 Infinite Loop Mail Stop26A Cupertino, California 95014 U.S.A FCC ID: BCGA1203
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative
Frequency selective monitoring and logging of environmental electromagnetic fields
FREQUENCY SELECTIVE EMF AREA MONITOR AMS-8060 Frequency selective monitoring and logging of environmental electromagnetic fields Up to 20 fully programmable frequency bands Real built-in spectrum analyser
The field strength measurement and SAR experience related to human exposure in 110 MHz to 40 GHz
The field strength measurement and SAR experience related to human exposure in 110 MHz to 40 GHz J. Klima, R. Ščehovič Department of Physics, Faculty of Natural Sciences, University of Mathias Bel, Tajovského
GSM/EDGE Output RF Spectrum on the V93000 Joe Kelly and Max Seminario, Verigy
GSM/EDGE Output RF Spectrum on the V93000 Joe Kelly and Max Seminario, Verigy Introduction A key transmitter measurement for GSM and EDGE is the Output RF Spectrum, or ORFS. The basis of this measurement
Raptor RXi Ultra-fast scanning Countersurveillance Receiver
data sheet RAPTOR RXi DATA SHEET A MATTER A MATTER OF OF NATIONAL SECURITY Raptor RXi Ultra-fast scanning Countersurveillance Receiver Features + Scans 26GHz in less than 4 seconds at 3kHz resolution (approx.
TELECOMMUNICATION FREQUENCY BAND SPECTRUM OCCUPANCY IN KAMPALA UGANDA
TELECOMMUNICATION FREQUENCY BAND SPECTRUM OCCUPANCY IN KAMPALA UGANDA Gertrude Ayugi 1, Akisophel Kisolo 2, Tumps.W. Ireeta 3 1 PhD student Department of Physics, Makerere University Kampala, Uganda. P.O
Cond-IS RF Conducted Immunity System
Cond-IS RF Conducted Immunity System About RF Conducted Immunity In the global contest of EMC testing for residential and industrial EUTs (Equipments Under Test) RF Conducted Immunity compliance verification
Antenna Measurements with the Network Analyzer. Presented by Ernie Jackson RF/uW Applications Engineer Agilent Technologies
Antenna Measurements with the Network Analyzer Presented by Ernie Jackson RF/uW Applications Engineer Agilent Technologies Purpose During this presentation you will: Learn about interface requirements
RX-AM4SF Receiver. Pin-out. Connections
RX-AM4SF Receiver The super-heterodyne receiver RX-AM4SF can provide a RSSI output indicating the amplitude of the received signal: this output can be used to create a field-strength meter capable to indicate
Radio Frequency Exposure Test Report
Radio Frequency Exposure EN 62311 January 2008 Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0Hz 300GHz) (IEC 62311:2007, modified)
Scalar Network Analysis with the HP 8590 Series Spectrum Analyzers Product Overview
This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products or services now available through Agilent. It
EMC Expert System for Architecture Design
EMC Expert System for Architecture Design EMC Expert System for Architecture Design Marcel van Doorn [email protected] Philips Electromagnetics Competence Center High Tech Campus 26, 5656 AE
EUROPEAN ETS 300 718 TELECOMMUNICATION March 1997 STANDARD
EUROPEAN ETS 300 718 TELECOMMUNICATION March 1997 STANDARD Source: ETSI TC-RES Reference: DE/RES-08-0501 ICS: 33.020 Key words: Radio, safety, SAR, testing Radio Equipment and Systems (RES); Avalanche
RF and Microwave Accessories. CD-ROM Catalog. Find the right component for your Rohde & Schwarz test & measurement equipment
RF and Microwave Accessories CD-ROM Catalog Find the right component for your Rohde & Schwarz test & measurement equipment Product group Typical applications Adapters Interchanging of various connector
Making Spectrum Measurements with Rohde & Schwarz Network Analyzers
Making Spectrum Measurements with Rohde & Schwarz Network Analyzers Application Note Products: R&S ZVA R&S ZVB R&S ZVT R&S ZNB This application note describes how to configure a Rohde & Schwarz Network
SAR TEST REPORT. According to the standard: EN 62209-1: 2006
R052-DAS-12-104374-2/A Ed. 1 This report cancels and replaces the test report R052-DAS-12-104374-2/A Ed. 0 SAR TEST REPORT According to the standard: EN 62209-1: 2006 Equipment under test: Antenna patch