1 I-1 I. PUMP OPERATION AND FLOW CALIBRATION SECTION A: PUMP OPERATION I. Purpose: To become familiar with typical air sampling pumps by: 1. adjusting and setting a sampling pump's air flow rate. 2. detecting pump failure. 3. using the pump's timing and programming mechanism. II. Readings Manuals for Pumps, particularly SKC 224-PCXR7 and MSA Elf III. Outline This exercise has three small parts that: Part 1: Develop familiarity with a personal sampling pump operation. Part 2: Demonstrate setting a timed sampling sequence. Part 3: Demonstrate setting up and wearing a sampling train. All of these parts will be completed within one laboratory period. Complete Part 1 before proceeding to parts 2 and 3. The others can be completed in any order and you may want to work at more than one exercise at time. IV. Equipment: SKC model 224-PCXR7 sampling pump, and Airchek 2000 MSA Elf sampling pump and or other pumps Low Flow Pump, such as SKC Pocket Pump or Gilian. Rotameters Small slotted screw driver Small Phillips screw driver Stop Watch Hose Clamp, tubing, and connections. Filter holder or charcoal tube holders (2) ENVH 555/2009/CS; I. Pump Operation, page I-1
2 I-2 Introduction: The sampling pump is the most commonly used industrial hygiene sampling device. It is so easy to use that the pumps are rarely used to their full technological capability. In this exercise you will work with two pumps: the SKC model 224 PCXR7, and the MSA ELF These pumps are for air sampling in the 0.5 to 4.0 liter per minute (lpm) range. With adapters air can be sampled at lower flow rates. For familiarization, your group should review the fundamentals of operation of each type of pump and its instruction pamphlet. As a minimum, learn how to turn the pump On and Off and how to change the flow rate. Additional electronic features, such as elapsed sampling time, timed operation, logging of data and flow fault notification, are available and useful, on most of the pumps, and are interesting to apply as time permits. The following sections outline necessary tasks. Part 1a: Learning Pump Basics: SKC Pump 1. With the SKC pump, turn the sampling pump ON (Figure 1 #8). 2. Press the "Start/Hold" Key (#6) and the "Flow and Battery Check Key" (#2). The Liquid Crystal Display (LCD) should indicate "battery OK". 3. Adjust the flow from 1 up to 4 and then down to 2 on the flow indicator (#18) with the flow adjust screw (#12). 4. Connect the 1/4" tubing from the intake filter housing (#14) to the top port of a rotameter. The rotameter response depends upon pressure, temperature, and humidity, so it must be calibrated to a primary standard (you will do this in lab 2). For now, adjust the flow rate so that the rotameter reads Press the "Flow and Battery Check Key" (#2) to stop the pump. A sample has not yet been collected so the timer has not started. 6. Press "Start/Hold" (#6). Allow the pump to run for at least 2 minutes and watch the LCD display as it records sampling time. 7. After two minutes partially pinch the 1/4" tubing with the hose clamp in order to engage the constant flow mechanism (you should hear the pump speed increase). Indicated flow on the rotameter should readjust after a short period. 8. Next plug the rotameter inlet and wait for the pump to fault (stop) and "Flow Fault" appear on the LCD. Record the elapsed time. The pump is designed to automatically shut down if there is flow stoppage or low voltage: Sampling time remains indicated on the LCD. 9. Start the pump again and cause a pump fault again. Then record the time. Is the time additive, or does the new sampling period start at zero? Figure out how to reset pump to start sampling time at zero. The pump's flow compensator is designed to maintain constant airflow as a sample substrate becomes partially obstructed. Table 1 (following page) illustrates the specifications for the SKC pump. You will be evaluating the flow compensator in the next laboratory. ENVH 555/2009/CS; I. Pump Operation, page I-2
3 I-3 Table 1: Allowable back pressure vs flow rate for SKC 224 PCXR7 Flow Rate (L/min) Back Pressure inches H 2 O inches H 2 O inches H 2 O inches H 2 O inches H 2 O inches H 2 O inches H 2 O Part 1b: Pump Basics for the MSA Elf Pump 1. For the Elf pump, read the instruction book pages 2-3 and 3-3 and look-over the schematic diagram on page 2-2. Following Section 3,: 2. turn the pump on and off. 3. Adjust the flow from 2 LPM to 3 LPM. How long does it take the pump to stabilize at the new flow? 4. Set the flow back to 2 LPM. Does it take the same time to stabilize? 5. Try running the pump for 3 minutes, then turn it off and on without losing the total run time. 6. Repeat the off-on procedure, but this time erasing the run time and starting a new sampling time. Part 2: Set-up of timed sampling. SKC 224-PCXR7 Pump (NB: Start Part 3 simultaneously!) In some sampling situations we would like for the pump to start at a pre-set time, or run intermittently, or shut off at the end of a specified sampling period, or any combination of these. The SKC pump has these capabilities. We will set up the pump to complete the following sampling routine: Sample period 17 minutes (total) Pump period 5 minutes (total) Delayed start 2.0 minutes While the pump LCD displays "hold", these time functions may be programmed. 1. Press the "Set-up" key (#7). The "Delayed Start" will display on the LCD as well as a flashing digit. The value of the flashing digit will be incremented each time the "Digit Set" (#5) is pressed. 2. The "Digit Select"(#3) is to select the flashing digit. Using these two keys, enter the desired number of minutes displayed, then press the "Mode" (#4). 3. "Sample Period" will now appear. 4. Again use the "Digit Set," "Digit Select ", and "Mode" keys to program this interval. "Pump Period" will appear on the LCD. 5. Using the same keys, program pump period. 6. Press "Start/Hold" and record delay time, start and stop times, and total duration of pump sampling. ENVH 555/2009/CS; I. Pump Operation, page I-3
4 I-4 7. Prepare a bar graph of the time sequence. When would you want to use this feature? Part 3: Mock personal sampling 1. Attach 1/4 inch tubing to pump inlet and at the other end of the tubing attach a filter holder or a charcoal tube holder. 2. Attach the sampling device holder to a person's clothing within their breathing zone and attach the pump to the person (usually their belt). 3. Turn on the pump and wear it for 10 minutes. Consider the following questions: How does it feel? Would you like to wear it for 8 hours? Can you predict instances when you, the industrial hygienist, will need to accommodate the worker by removing the pump or arranging it differently? ENVH 555/2009/CS; I. Pump Operation, page I-4
5 I-5 SECTION B: FLOW CALIBRATION The evaluation of workplace exposures requires sampling and quantitation of airborne contaminant concentrations. An essential part of any sampling is an accurate measurement of the quantity of air that was sampled. Before going on a survey of a worksite, the flow rates of the pumps and the flow measuring devices must be calibrated. The different parts of this experiment represent examples of the calibration procedures that would be a necessary part of the preparation for a survey. Emphasis is placed on the effect of atmospheric conditions. I. Purpose To become familiar with: a. Primary (e.g., bubble buret) and secondary (e.g., rotameter) calibration standards; and other flow devices such as the electronic and dry calibrators. b. Pumps and filter media used in sampling. II. Outline This experiment has three required parts which you will do in the two lab periods. You will probably have enough time to also do the optional part. You will work in teams of two or three; however, individual write-ups will be required. Part 1: Calibration of rotameter using a bubble buret. Part 2: Calibration of sampling pump using a rotameter. Part 3: Evaluating the effects of filters on pump performance. Part 4: Calibration of an electronic flow (Gilibrator) meter or Dry Cal vs. Bubble buret. III. Introduction The mainstay of industrial hygiene air sampling is the personal sampling pump. After collecting a sample, one calculates the concentration of a contaminant in air as the amount of contaminant collected divided by the volume of air drawn through the sample media (e.g., filter, adsorption tube, etc.). Concentration (mg/m 3 )= Weight Contaminant (mg)/air Volume (m 3 ) If the amount of contaminant (solid, liquid or gas) is expressed as a mass, then the resulting concentration is given as milligrams of contaminant per cubic meter of air (mg/m 3 ). If the amount of a gaseous contaminant is expressed as a volume, then the result is given as the volume of contaminant divided by the volume of air sampled (ppm). Concentration (ppm) = Volume Contaminant (L x 10 6 ) / Air Volume (L) ENVH 555/2009/CS; I. Pump Operation, page I-5
6 I-6 In either case, the volume of air sampled is determined by multiplying the sampling pump flow rate by the total time of sampling. Volume = Flow (L/min.)/Time (min.) For example, with a pump drawing 2 L/min, the total volume of a one hour sample would be 2L/min x 60 min = 120L. Thus in order to obtain an accurate result, it is necessary to accurately and precisely control and measure the flow rate of the sampling pump. In this experiment you will learn to calibrate sampling pumps, and will explore how some of the conditions such as barometric pressure or temperature can cause variations in pump performance. Calibration instruments are discussed in the following paragraphs. Pumps: Perkins, pages and Air sampling Instruments, pages The pumps that will be used in this experiment are portable sampling pumps with a range of flow rate from 0.5 to 4 liters per minute with a constant volume feature. Rotameters: Perkins, pages and ASI pages The most common air flow calibration device used in the field (and often in the laboratory) is the rotameter. The rotameter consists of a tapered vertical tube containing a float. The float rises to a height which is dependent on the flow rate and density of air moving through the tube. Rotameters are available to measure flow rates as low as 1 ml/min and as high as 300 cfm. Any particular rotameter usually covers a range of about one order of magnitude. Their accuracy varies from ± 1% to ±10% depending on the unit. Some rotameters have multiple floats to increase their range. The float is usually a sphere or an ellipsoid; however, there are other configurations. One always reads a rotameter at the widest part of the float as a matter of standardization and in order to minimize parallax error. Bubble Burets : Perkins, pages and Air sampling Instruments, pages A bubble buret operates as a frictionless piston with carefully manufactured length and diameter. The rotameter is not a primary measuring device and thus must be calibrated against a more accurate air flow measuring device such as a bubble buret and timer. The accuracy and precision of the bubble buret depend upon easily calibrated parameters such as length, diameter and time which do not vary significantly under varying conditions of temperature and pressure. Thus the bubble buret is a primary standard. The rotameter is termed a secondary standard because its more complicated construction and physics of operation make it inherently less accurate and precise, and its readings vary with temperature, pressure and the density of the gas. Because results provided by a rotameter vary under differing conditions of temperature, pressure and density, mathematical corrections must be made if the rotameter is used under conditions which differ significantly from the conditions under which it was calibrated. It is also strongly recommended that rotameters be calibrated upon receipt from the manufacturer even if they come with the manufacturer's certificate of ENVH 555/2009/CS; I. Pump Operation, page I-6
7 I-7 calibration. In this experiment you will explore some of the conditions which can cause rotameter performance to vary, and will learn to make the necessary corrections. Other Flow Measuring Devices: Perkins, pages 406, and Air sampling Instruments, pages Additional types of equipment can be used to measure air flow. Exhibited in the lab will be mass flow meters, the Dry Cal and electronic bubble meters (Gilibrator). Mass Flow Meters are described on page 136 in Air monitoring for toxic Exposures (McDermott) IV. Equipment Each team should obtain the following equipment before beginning: a. Rotameter (1-4 l/min, w/o needle valve) j. Vacuum gage b. 1 personal sampling pump k. Stop watch c. Bubble buret l. Soap bubble solution d. Beaker m. 2 three-prong clamps e. Ring stand n. 2 ring stand clamps f. 1 side arm flask o. Tygon or rubber tubing g. 2 needle valves p. Vacuum pump h. Manometer or magnehelix q. 37 mm sampling cassettes (6) i. Dry Cal r. Gilibrator Other equipment will be provided as necessary. Record the serial numbers of the rotameter and the sampling pumps, and label these to identify them for your team. You will use these several times during the quarter. ENVH 555/2009/CS; I. Pump Operation, page I-7
8 I-8 Part 1: Calibration of Rotameter Using Bubble Buret In this part you will determine the flow rate at various rotameter readings and will plot the true flow rate (as measured with the bubble buret) vs. the rotameter reading. The resulting graph will be the calibration curve for your rotameter. You will use this curve several times during the quarter when you use the rotameter to determine air flow. Use the MSA Elf pump for this exercise. 1. Wash the buret thoroughly and then assemble the apparatus as shown in Fig Put a half inch of soap bubble solution in a beaker. Have an instructor check your set-up before you continue. Trap Pump Bubble Burette Rotameter ENVH 555/2009/CS; I. Pump Operation, page I-8 Figure Rinse and coat the inside of the buret with soap bubble solution; then drain. 3. Insure the needle valve (if present) on the rotameter is fully open. Adjust the pump air flow through its full range to determine roughly the corresponding range of rotameter settings. For example, the range might be from about 1 to about 3L/min. 4. Adjust the pump to create one a flow rate of 1.5 L/min., as measured on the rotameter. Lift the beaker to create a soap bubble at the bottom of the buret. Time the passage of the bubble through the buret. Make at least three bubble meter measurements at this rotameter reading. Record the data and determine the average flow rate. 5. Make similar bubble meter measurements at each of four other rotameter readings. You should end up with data from at least five readings representing the full range of the pump & rotameter. 6. Plot the data to create a calibration curve (rotameter reading vs. flowrate). You can use MS Excel software on the lab computer to do this. Plot this curve during lab time as it will be used in Part 2. Repeat any points that do not fall on a smooth curve. Combine the data into a single table & graph. Questions: Was a correction needed for the rotameter reading? How great is an error in a sample volume if the reading is not corrected? Is the correction constant over the range of the rotameter?
9 I-9 Part 2: Calibration of Sampling Pump Using a Rotameter In this part you will use the rotameter which you have just calibrated to calibrate a sampling pump for flow rate. This is a common field procedure. Use the SKC pump for this exercise. 1. Assemble the apparatus as shown in Fig Pump Rotameter Figure Adjust the pump air flow through its full range to determine roughly the corresponding range of rotameter settings. For example, the range might be from about 1 to about 3L/min. 3. Choose four flow rates in this range and record them in your lab book. For example, you might choose 1.5, 1.7, 2.0 and 2.5 liter per minute. 4. Refer to your rotameter calibration curve and adjust the pump to create one of the corrected flow rates (as measured by the rotameter) that you have chosen. Record the reading on the rotameter which is integral to the pump. 5. Repeat steps 4 and 5 for all other flow rates which you have chosen. For each step record both the setting on your calibrated rotameter, and the flow reading on the rotameter that is integral to the pump. How accurate is the rotameter on the pump? Which is more accurate--the rotameter on the pump or the secondary standard? Considering the primary standard (bubble burette) and the secondary standard (calibrated rotameter), which is best in accuracy or convenience for routine calibrations of the sampling pump? ENVH 555/2009/CS; I. Pump Operation, page I-9
10 I-10 Part 3: Evaluating the Effects of Filters on Pump Performance In this part you will evaluate the effects of filter resistance on the pump performance for the SKC pump. 1. Load four 37 mm cassettes with the following filters: a. Glass fiber used for sampling/analysis of SVOCs b. Metricel VM (MCE, 5.0 um pore size) used for metals sampling. c. Millipore AA (MCE, 0.8 um pore size) --used for asbestos sampling. d. Nucleopore filter (polycarbonate) used for electron microscopy samples. 2. Obtain one or two "dirty" filter cassettes from the instructors. 3. Assemble the apparatus as in Fig. 2.3 (initially without the filter cassette in line). Rotameter Filter Cassette Figure 2.3 Vacuum Gauge Pump 4. Without any filter cassette in line, refer to your calibration curve and adjust the pump to draw 1L/min. Record the pressure (vacuum) reading, and the rotometer readings for your calibrated rotatmter and the pump rotameter. 5. Without changing the pump setting, place each of the 6 filter cassettes in line, one at a time. For each cassette note the pressure (vacuum), the (upstream) rotameter reading and the pump rotameter reading. 6. After completing Step 5 for each cassette, place each filter in line, one at a time, and adjust the pump to draw in 2.8 L/min. (Is this possible?). Note the pressure (vacuum) for each cassette. Questions: a. Does a filter or a loaded filter change the flow measured on the pump? b. Does a filter or a loaded filter change the flow measured upstream of the filter cassette? c. Would the effects you note for a loaded filter alter the measured or actual sample volume for a filter sample? ENVH 555/2009/CS; I. Pump Operation, page I-10
11 I-11 Part 4: Comparison of a Bubble Buret and Electronic Bubble Meter and/or DryCal Attach the soap bubble buret to electronic bubble meter (Gillibrator) or the DryCal and pump in series, as in Diagram 2.1. Select two flow settings with replicate readings, and compare results of the two devices. Compare the use of the electronic bubble meter with that of the soap bubble meter from Part 1 for the calibration of other flow measuring devices. Which would you prefer for field calibration and why? Should the electronic bubble meters or the Dry Cal require calibration? The SKC UltraFlo can be substituted for the Gilibrator. LAB WRITE UP, EXPERIMENT ONE: A full scale write up is not required for this experiment. Instead, please answer the questions below and include the data requested. Section 1: Part 1a: When the SKC pump flow-faulted and was then re-started, was the sampling time additive? Part 2: Provide a bar graph of timed sampling sequence with time on the x-axis and air flow on the y-axis. Are the intervals equal across the entire duration? Section 2: Part One - Rotameter Calibration: Tabulate the raw data you collected in this section, and include your calibration curve of bubble burette flow vs rotameter flow. Where you have taken multiple measurements under the same conditions, use the data to calculate standard deviations and coefficients of variation and place appropriate error bars on your calibration graph. a. What error variable(s) is (are) measured? b. Was a correction needed to adjust the rotameter reading to the true flow rate as measured by the bubble buret? c. How great is an error in a sample volume if the reading is not corrected? d. Is the correction constant over the range of the rotameter? Part 2 - Pump Calibration: Tabulate the raw data you collected in this section. a. How accurate is the rotameter on the pump (compared to the primary standard)? b. Which is more accurate--the rotameter on the pump or the secondary standard? c. Considering the primary standard (bubble burette) and the secondary standard (calibrated rotameter), which is best in accuracy or convenience for routine calibrations of the sampling pump. Is there any difference? Part 3 - Filter Effects: Tabulate the raw data you collected in this section. ENVH 555/2009/CS; I. Pump Operation, page I-11
12 I-12 Besides discussing the effects of each filter, if you observe a noticeable effect for specific filters make a graph of pump flow rate (i.e., pump performance) as a function of pressure drop across the filter (i.e., filter resistance). a. Does a filter or a loaded filter change the flow measured on the pump? b. Does a filter or a loaded filter change the flow measured upstream of the filter cassette? c. Would the effects you note for a loaded filter alter the measured or actual sample volume for a filter sample? Part 4- Electronic meters: Compare the ease of use and accuracy of the Dry Cal and Gilibrator (UltraFlo) with the soap bubble buret. a. Which would you prefer for field calibration and why? b. Should the electronic bubble meters or the Dry Cal require calibration? References Air Sampling Instruments for evaluation of atmospheric contaminanats. Susanne V. Herring (Tech. Ed.) 7 th ed., ACGIH, Cincinnati, OH. McDermott, H.J., Air sampling for toxic exposures, 2 nd Ed. John Wiley and sons, Hoboken, NJ Perkins, J. Modern Industrial Hygiene, Van Nostrand Reinhold, NY(1997). ENVH 555/2009/CS; I. Pump Operation, page I-12
Determining Equivalent Weight by Copper Electrolysis
Purpose The purpose of this experiment is to determine the equivalent mass of copper based on change in the mass of a copper electrode and the volume of hydrogen gas generated during an electrolysis reaction.
ME 315 - Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS
ME 315 - Heat Transfer Laboratory Nomenclature Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS A heat exchange area, m 2 C max maximum specific heat rate, J/(s
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
Physical Properties of a Pure Substance, Water
Physical Properties of a Pure Substance, Water The chemical and physical properties of a substance characterize it as a unique substance, and the determination of these properties can often allow one to
UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE 3211-4211 ABSORPTION OF CARBON DIOXIDE INTO WATER
UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE 3211-4211 ABSORPTION OF CARBON DIOXIDE INTO WATER OBJECTIVE The objective of this experiment is to determine the equilibrium line,
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids
Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids Dr. J. M. Meyers Dr. D. G. Fletcher Dr. Y. Dubief 1. Introduction Last lab you investigated flow loss in a pipe due to the roughness
THE LABORATORY NOTEBOOK
THE LABORATORY NOTEBOOK In scientific work keeping a permanent record of all raw data, observations, calculations, et cetera obtained during an experiment is important. Therefore, a student must become
Measurement and Calibration
Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding
Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P2.6.3.2. Thermodynamic cycle Heat pump
Heat Thermodynamic cycle Heat pump LD Physics Leaflets P2.6.3.2 Investigating the function of the expansion valve of the heat pump Objects of the experiment g To study the operational components of the
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Balzers Sputter Coater SCD 050
Balzers Sputter Coater SCD 050 The SCD 050 is a bench top, sputter deposition system designed for thin films on substrates up to 6 inches. Morphology and thickness is user controlled using power, pressure,
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid
Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid Introduction Many metals react with acids to form hydrogen gas. In this experiment, you will use the reactions
SOLUBILITY, IONIC STRENGTH AND ACTIVITY COEFFICIENTS
SOLUBILITY, IONIC STRENGTH AND ACTIVITY COEFFICIENTS References: 1. See `References to Experiments' for text references.. W. C. Wise and C. W. Davies, J. Chem. Soc., 73 (1938), "The Conductivity of Calcium
Calibration of Volumetric Glassware
CHEM 311L Quantitative Analysis Laboratory Revision 2.3 Calibration of Volumetric Glassware In this laboratory exercise, we will calibrate the three types of glassware typically used by an analytical chemist;
To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.
Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2
DENSITY MEASURING SYSTEMS
SHIBAYAMA SCIENTIFIC CO., LTD. DENSITY MEASURING SYSTEMS Density Gradient Tube Method Direct Reading Type A / Type B Comply with the standards below: JIS K-0061-1992, K-7112 1980 ASTM D1505 Type A /Left
Chemistry 119: Experiment 7. Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets
Chemistry 119: Experiment 7 Potentiometric Titration of Ascorbic Acid in Vitamin C Tablets Vitamin C is another name for ascorbic acid (C 6 H 8 O 6, see below ), a weak acid that can be determined by titration
EXPERIMENT 15: Ideal Gas Law: Molecular Weight of a Vapor
EXPERIMENT 15: Ideal Gas Law: Molecular Weight of a Vapor Purpose: In this experiment you will use the ideal gas law to calculate the molecular weight of a volatile liquid compound by measuring the mass,
KNC Model 3666 Automatic Pressure Calibration System
KNC Model 3666 Automatic Pressure Calibration System Product Description The Model 3666 Automatic Pressure Calibration (APC) System by King Nutronics Corporation is a portable secondary standards laboratory
18 Conductometric Titration
Lab Activity 18 CONDUCTOMETRIC TITRATION LAB ACTIVITY 18 Conductometric Titration Background Titration is the a method of determining the concentration of an unknown solution (the analyte) by reacting
Standard Operating Procedure
Standard Operating Procedure Equipment / Procedure: Filename: Assessor s name & date: Reviewer s name & date: Cosmed K4 2012.1201SOP_cosmed_K4 Richard Metcalfe, December 2012 Tom Nightingale, December
Experiment 3 Pipe Friction
EML 316L Experiment 3 Pipe Friction Laboratory Manual Mechanical and Materials Engineering Department College of Engineering FLORIDA INTERNATIONAL UNIVERSITY Nomenclature Symbol Description Unit A cross-sectional
(S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26)
Chem 367-2/ Physical Adsorption 63 PHYSICAL ADSORPTION (S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26) The sample used in this experiment needs to be conditioned for at least two hours before the experiment;
Stack Sampling Stack sampling or source sampling
Stack Monitoring Stack Sampling Stack sampling or source sampling may be defined as a method of collecting representative samples of pollutant laden air/gases at the place of origin of pollutants to determine
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer
Titration of a Diprotic Acid: Identifying an Unknown Computer 25 A diprotic acid is an acid that yields two H + ions per acid molecule. Examples of diprotic acids are sulfuric acid, H 2 SO 4, and carbonic
Physics 41, Winter 1998 Lab 1 - The Current Balance. Theory
Physics 41, Winter 1998 Lab 1 - The Current Balance Theory Consider a point at a perpendicular distance d from a long straight wire carrying a current I as shown in figure 1. If the wire is very long compared
COMMON LABORATORY APPARATUS
COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen
Triple Point Experiment
Equipment List Rotary vane vacuum pump 2-stage, 2 to 7 CFM. For example, Edwards 2M2 (2 CFM), Edwards RV5 (3.5 CFM), Edwards E2M8 (6.7 CFM) or equivalent. Bell jar Nalgene polycarbonate plastic, 5-5/8
The University of Toledo Soil Mechanics Laboratory
The University of Toledo Soil Mechanics Laboratory Permeability Testing - 1 Constant and Falling Head Tests Introduction In 1856 the French engineer Henri D arcy demonstrated by experiment that it is possible
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
NITRIC OXIDE and NITROGEN DIOXIDE 6014
NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm
12 Volt 30 Amp Digital Solar Charge Controller
12 Volt 30 Amp Digital Solar Charge Controller User s Manual WARNING Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions
IDEAL AND NON-IDEAL GASES
2/2016 ideal gas 1/8 IDEAL AND NON-IDEAL GASES PURPOSE: To measure how the pressure of a low-density gas varies with temperature, to determine the absolute zero of temperature by making a linear fit to
Physical Chemistry Laboratory I CHEM 445 Experiment 6 Vapor Pressure of a Pure Liquid (Revised, 01/09/06)
1 Physical Chemistry Laboratory I CHEM 445 Experiment 6 Vapor Pressure of a Pure Liquid (Revised, 01/09/06) The vapor pressure of a pure liquid is an intensive property of the compound. That is, the vapor
QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
The Molar Mass of a Gas
The Molar Mass of a Gas Goals The purpose of this experiment is to determine the number of grams per mole of a gas by measuring the pressure, volume, temperature, and mass of a sample. Terms to Know Molar
OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
Apr 17, 2000 LAB MANUAL 1302.0. 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified)
Apr 17, 2000 LAB MANUAL 1302.0 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified) 1302.1 SCOPE This method describes a procedure for the quantitative determination of the distribution
15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.
S HIFT INTO NEUTRAL 15. Acid-Base Titration Shift into Neutral Student Instruction Sheet Challenge Discover the concentration of an unknown acid solution using acid-base titration. Equipment and Materials
Pressure -Temperature Relationship in Gases. Evaluation copy. Figure 1. 125 ml Erlenmeyer flask. Vernier computer interface
Pressure -Temperature Relationship in Gases Computer 7 Gases are made up of molecules that are in constant motion and exert pressure when they collide with the walls of their container. The velocity and
CONSTANT HEAD AND FALLING HEAD PERMEABILITY TEST
CONSTANT HEAD AND FALLING HEAD PERMEABILITY TEST 1 Permeability is a measure of the ease in which water can flow through a soil volume. It is one of the most important geotechnical parameters. However,
SELECTION, APPLICATION AND MAINTENANCE
DIESEL PROTECTION SYSTEMS Automatic Diesel Engine Shut Down System for Safe Area Applications SELECTION, APPLICATION AND MAINTENANCE Series 300 Series 310 SYSTEM DESCRIPTION Suitable for attended engine
Operating Instructions & Maintenance Guidelines for Precision Vaporizers
Operating Instructions & Maintenance Guidelines for Precision Vaporizers This page intentionally blank. Examination and Preparation for Use We want you to be completely satisfied with your vaporizer. Please
Volkswagen Jetta, Golf, GTI 1999, 2000 2.8 Liter VR6 2V Engine Mechanical, Engine Code(s): AFP 17 Engine-Lubrication (Page GR-17)
17 Engine-Lubrication (Page GR-17) Lubrication system components, removing and installing Oil filter housing, disassembling and assembling Oil pan, removing and installing Oil pressure and oil pressure
Physics Lab Report Guidelines
Physics Lab Report Guidelines Summary The following is an outline of the requirements for a physics lab report. A. Experimental Description 1. Provide a statement of the physical theory or principle observed
1. The Determination of Boiling Point
1. The Determination of Boiling Point Objective In this experiment, you will first check your thermometer for errors by determining the temperature of two stable equilibrium systems. You will then use
LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS
LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS 1. Define an accumulator and explain its function A hydraulic accumulator is a device that stores the potential energy of an incompressible fluid
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
Acid Base Titrations
Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually
ENZYME ACTION: TESTING CATALASE ACTIVITY
ENZYME ACTION: TESTING CATALASE ACTIVITY LAB ENZ 1.CALC From Biology with Calculators, Vernier Software & Technology, 2000 INTRODUCTION Many organisms can decompose hydrogen peroxide (H 2 O 2 ) enzymatically.
Experiment 446.1 SURFACE TENSION OF LIQUIDS. Experiment 1, page 1 Version of June 17, 2016
Experiment 1, page 1 Version of June 17, 2016 Experiment 446.1 SURFACE TENSION OF LIQUIDS Theory To create a surface requires work that changes the Gibbs energy, G, of a thermodynamic system. dg = SdT
Carbon Dioxide and an Argon + Nitrogen Mixture. Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10
Carbon Dioxide and an Argon + Nitrogen Mixture Measurement of C p /C v for Argon, Nitrogen, Stephen Lucas 05/11/10 Measurement of C p /C v for Argon, Nitrogen, Carbon Dioxide and an Argon + Nitrogen Mixture
Experiment 4 The Relationship of Density and Molarity of an Aqueous Salt Solution
Experiment 4 The Relationship of Density and Molarity of an Aqueous Salt Solution Purpose: The purpose of this experiment is to investigate the relationship between the concentration of an aqueous salt
In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
ph: Measurement and Uses
ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic
RITTER Multiple Sonic Nozzle Calibration System
RITTER Multiple Sonic Nozzle Calibration System »Wouldn t it be great to eliminate doubtful measurement results by using proven measurement technology? Providing the most precise results ensures and increases
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
Calibration of Volumetric Glassware
Chemistry 119: Experiment 2 Calibration of Volumetric Glassware For making accurate measurements in analytical procedures, next in importance to the balance is volumetric equipment. In this section volumetric
Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.
ASGMT / Averaging Pitot Tube Flow Measurement Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. Averaging Pitot Tube Meters Introduction
Acid Dissociation Constants and the Titration of a Weak Acid
Acid Dissociation Constants and the Titration of a Weak Acid One of the most important applications of equilibria is the chemistry of acids and bases. The Brønsted-Lowry acid-base theory defines an acid
Electricity. Confirming Coulomb s law. LD Physics Leaflets P3.1.2.2. 0909-Wie. Electrostatics Coulomb s law
Electricity Electrostatics Coulomb s law LD Physics Leaflets Confirming Coulomb s law P3... Measuring with the force sensor and newton meter Objects of the experiments Measuring the force between two charged
PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE
Test Procedure for PARTICLE SIZE ANALYSIS OF SOILS TxDOT Designation: Tex-110-E Effective Date: August 1999 1. SCOPE 1.1 This method covers the quantitative determination of the distribution of particle
Experiment 1: Measurement and Density
Experiment 1: Measurement and Density Chemistry 140 Learning Objectives Become familiar with laboratory equipment and glassware Begin to see the link between measurement and chemical knowledge Begin to
Gravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
Online Infrared Flue Gas Analyzer Gasboard 3000
Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume
6 Crankcase Ventilation
6 Crankcase Ventilation Introduction... 6-3 Crankcase Ventilation Manager (CVM)... 6-5 Introduction... 6-6 CVM280B..................................... 6-7 CVM280T... 6-10 CVM424B....................................
MFP14, MFP14S and MFP14SS Automatic Pumps
Page 1 of 6 TI-P136-02 ST Issue 12 Cert. No. LRQ 0963008 ISO 9001, S and SS utomatic Pumps Description The Spirax Sarco automatic pump is a displacement receiver operated by steam or compressed air. It
Vacuum. How It Relates to Refrigeration And Air Conditioning Service
Vacuum How It Relates to Refrigeration And Air Conditioning Service Moisture In A Refrigeration System Visible Moisture Water Droplets Uncommon, but it can occur Invisible Moisture Water Vapor Found in
Molar Mass of Butane
Cautions Butane is toxic and flammable. No OPEN Flames should be used in this experiment. Purpose The purpose of this experiment is to determine the molar mass of butane using Dalton s Law of Partial Pressures
Activity P13: Buoyant Force (Force Sensor)
July 21 Buoyant Force 1 Activity P13: Buoyant Force (Force Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Archimedes Principle P13 Buoyant Force.DS P18 Buoyant Force P18_BUOY.SWS
DOE-HDBK-1122-99. Module 2.18 Air Sampling Equipment. Instructor s Guide
Course Title: Radiological Control Technician Module Title: Air Sampling Equipment Module Number: 2.18 Objectives: 2.18.01 Identify the factors that affect the operator's selection of a portable air sampler.
STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE
Page 1 of 9 PURPOSE Steam is the most commonly used heating medium for maintaining process temperatures. Compared to other heating media, steam contains a significant amount of heat energy, and this heat
DMS 680. Universal length measuring system gauge inspection for»iso 9000«requirements
DMS 680 Universal length measuring system gauge inspection for»iso 9000«requirements Universal length measuring system DMS 680 Wide range of applications For periodic inspection of gauges, reference gauges,
AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *
AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory * Paper prepared for the 4 th International Symposium on Fluid Flow
Laboratory Exercise: Calibration of a Thermometer
CHEM 109 Introduction themistry Revision 3.1 Laboratory Exercise: Calibration of a Thermometer In this exercise we will calibrate a stem-type thermometer and then use it to correctly measure the Air temperature
Air Eliminators and Combination Air Eliminators Strainers
Description Air Eliminators and Combination Air Eliminator Strainers are designed to provide separation, elimination and prevention of air in piping systems for a variety of installations and conditions.
METHOD OF TEST FOR DETERMINATION OF PERMEABILITY OF GRANULAR SOILS
Laboratory Testing Manual Date: 99 06 21 Page 1 of 7 METHOD OF TEST FOR DETERMINATION OF PERMEABILITY OF GRANULAR SOILS 1. SCOPE 1.1 This method covers the determination of the coefficient of permeability
DISSOLVED OXYGEN METER Model : DO-5509
DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex
Determination of a Chemical Formula
1 Determination of a Chemical Formula Introduction Molar Ratios Elements combine in fixed ratios to form compounds. For example, consider the compound TiCl 4 (titanium chloride). Each molecule of TiCl
Specifications - Installation and Operating Instructions Ø5 [127.00] 17/64 [6.91] 1-59/64 [48.89]
Bulletin P-DM-1100 Series DM-1000 DigiMag Digital Differential Pressure Gage Specifications - Installation and Operating Instructions PRESSURE CONNECTIONS 21/64 [8.59] LONG Ø5 [127.00] 4-31/64 [113.89]
RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer
RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer EN FR IT DE ES CONTENTS / EN Title CONTENTS Page 1. SAFETY INFORMATION...1 2. INTRODUCTION...2 3. SPECIFICATIONS...3 4. PARTS & CONTROLS...4
Laboratory Manual for Air Quality Sampling and Analysis
Laboratory Manual for Air Quality Sampling and Analysis Department of Civil Engineering Indian Institute of Technology Delhi New Delhi-110016 MONITORING OF RESPIRABLE PARTICULATE MATTER 1. Objective Sampling
Chem 131A: Absorbance of Riboflavin
Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the
GA/7 Potentiometric Titration
Rev. 7/99 7-1 INTRODUCTION GA/7 Potentiometric Titration The potentiometric titration is a useful means of characterizing an acid. The ph of a solution is measured as a function of the amount of titrant
NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS
NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS APPLICATION NOTE ITI-089 I. Introduction The technology used in personal sampling pumps has not changed much over the past 20 years. II. There
Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Chemicals Needed:
Reaction of Magnesium with Hydrochloric Acid (Gas Laws) Your Name: Date: Partner(s) Names: Objectives: React magnesium metal with hydrochloric acid, collecting the hydrogen over water. Calculate the grams
Kompressoren. Adsorption Dryer AD
Kompressoren Adsorption Dryer AD Adsorption Drying: why? Modern industries require compressed air that is increasingly filtered with low dew point and condensate. Today, the equipment is more sophisticated
Isolation of Caffeine from Tea
Isolation of Caffeine from Tea Introduction A number of interesting, biologically active compounds have been isolated from plants. Isolating some of these natural products, as they are called, can require
VAPOR PRESSURE AS A FUNCTION OF TEMPERATURE. This laboratory covers material presented in section 11.8 of the 9 th Ed. of the Chang text.
VAPOR PRESSURE AS A FUNCTION OF TEMPERATURE Objectives: (1) Observe and measure the change in the vapor pressure (dependent variable) as a function of temperature (independent variable). (2) Analyze the
FLUID FLOW Introduction General Description
FLUID FLOW Introduction Fluid flow is an important part of many processes, including transporting materials from one point to another, mixing of materials, and chemical reactions. In this experiment, you
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
Unit 24: Applications of Pneumatics and Hydraulics
Unit 24: Applications of Pneumatics and Hydraulics Unit code: J/601/1496 QCF level: 4 Credit value: 15 OUTCOME 2 TUTORIAL 2 HYDRAULIC AND PNEUMATIC CYLINDERS The material needed for outcome 2 is very extensive
Pre-Lab Notebook Content: Your notebook should include the title, date, purpose, procedure; data tables.
Determination of Molar Mass by Freezing Point Depression M. Burkart & M. Kim Experimental Notes: Students work in pairs. Safety: Goggles and closed shoes must be worn. Dispose of all chemical in the plastic
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
WIKA INSTRUMENT CORPORATION
WIKA INSTRUMENT CORPORATION Instruction Manual GAUGE PRESSURE INDICATOR Series 1500 Series 1000 Series 300 WIKA Instrument Corporation 1000 Wiegand Boulevard Lawrenceville, GA 30043 1-888-945-2872 http://www.wika.com
A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion
A Determination of g, the Acceleration Due to Gravity, from Newton's Laws of Motion Objective In the experiment you will determine the cart acceleration, a, and the friction force, f, experimentally for
Solids, Volatile Dissolved and Fixed Dissolved
, 8277 Solids, Volatile Dissolved and Fixed Dissolved Gravimetric Method 1 Scope and Application: For wastewater. 1 Adapted from Standard Methods for the Examination of Water and Wastewater DOC316.53.001206
ThermoSorb Desiccant Dryers
ThermoSorb Desiccant Dryers Sierra Why dry compressed air? Contamination reduces efficiency The air we breathe contains contamination in the form of water vapour and airborne particles. During the compression
The University of Toledo Soil Mechanics Laboratory
1 Grain Size Distribution Sieve Analysis The University of Toledo Soil Mechanics Laboratory Introduction The grain size distribution is a representation of the approximate distribution of soil grain sizes