CHATILLON DFIS Series ACCUFORCE AF2 Series. Digital Force Gauges. Operating Instructions. CHATILLON DFIS Series Digital Force Gauge
|
|
|
- Stanley Walters
- 9 years ago
- Views:
Transcription
1 Instruction Manual NC May 2001 Issue 3 CHATILLON DFIS Series ACCUFORCE AF2 Series Digital Force Gauges Operating Instructions CHATILLON DFIS Series Digital Force Gauge AMETEK AF2 Series Shown with MHTC Mechanical Test Stand - 1 -
2 TABLE OF CONTENTS INSTRUMENT PRECAUTIONS... 4 Page 1.0 INTRODUCTION AND PRODUCT OVERVIEW PRODUCT FEATURES ELECTRICAL SAFETY RECHARGING BATTERIES PHYSICAL DIMENSIONS REVERSIBLE HOUSING ABOUT THE KEYPAD ABOUT THE DISPLAY CONNECTION TO EXTERNAL DEVICES INTERFACE CABLES SETTING UP FOR SAFE USE FRONT PANEL OPERATION NORMAL MODE Selecting Units of Measure Selecting Type of Measurement - Capturing Peaks Zeroing Transmitting Readings CONFIGURATION MODE Configuring Output Port- Analog Output Configuring Output Port- Digital Outputs Enabling the Mitutoyo Output Enabling the RS-232 Port- Numerical Data Only Enabling the RS-232 Port- Full Data Modifying Communication Parameters DATA COLLECT MODE About Data Collect Mode
3 TABLE OF CONTENTS Page 3.0 CALIBRATING THE DFIS/AF2 SERIES CALIBRATION PROCEDURE DIAGNOSTICS SOFTWARE PART NUMBER SOFTWARE REVISION LEVEL CAPACITY AND UNITS NO LOAD ZERO VALUE RS232 LOOPBACK TEST KEYPAD TEST TECHNICAL REFERENCE AND SPECIFICATIONS MODEL NUMBERS PRODUCT WARRANTY This instrument is warranted against defects in workmanship, material and design for one (1) year from the date of delivery to the extent that AMETEK will, at its sole option, repair or replace the instrument or any part thereof which is defective provided, however, that the warranty shall not apply to instruments subjected to tampering or abuse, or exposed to highly corrosive conditions. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED AND AMETEK HEREBY DISCLAIMS ALL OTHER WARRANTIES, INCLUDING, WITHOUT LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY. AMETEK SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT LIMITED TO, ANY ANTICIPATED OR LOST PROFITS. This warranty is voidable if the purchaser fails to follow any and all instructions, warnings or cautions in the instrument s operation manual. If a manufacturing defect is found, AMETEK will replace or repair the instrument or replace any defective part thereof without charge; however, AMETEK s obligation hereunder does not include the cost of transportation which must be borne by the customer. AMETEK assumes no responsibility for damage in transit, and any claims for such damage should be presented to the carrier by the purchaser
4 PRECAUTIONS Read the instruction manual completely before attempting to use the DFIS or ACCUFORCE 2 (AF2) Series. By following the instructions contained in this manual, the optimum accuracy and performance can be attained. IMPORTANT NOTE: The DFIS and ACCUFORCE 2 (AF2) gauges are identical in functionality. The procedures, specifications and operations in this manual apply to both products unless noted. Turn off the DFIS/AF2 Series before connecting or disconnecting any cables to the instrument. Do not overload the DFIS/AF2 Series! The instrument can handle an overload of up to 150% of the rated full scale. The display will read when the gauge exceeds approximately 120% of full scale. When the force applied is removed and is no longer displayed, the gauge may not read above 121% of full scale in the Tension Peak or Compression Peak modes. Therefore, you must contact your local distributor to ensure the load cell was not damaged, unless you are 100% sure the force applied did not exceed 150% of full scale. The DFIS/AF2 Series is designed for axial loading only! Applying load on an angle or eccentric loading may cause an erroneous reading. Do not use tools to attach the shaft adapter and accessories to the gauge. Use finger-tight pressure only to prevent damage to the load cell. IMPORTANT NOTE: Before using your new DFIS/AF2 Series, fully charge the rechargeable Nicad battery by plugging the supplied battery charger into the battery charger jack at the top of the gauge and the proper AC outlet. Use only the battery charger supplied with the gauge. Using other chargers can overcharge and damage the battery
5 1.0 INTRODUCTION AND PRODUCT OVERVIEW The DFIS/AF2 Series family of digital force gauges manufactured by AMETEK Test and Calibration Instruments Division. They have been designed to provide accurate compression and tension force measurements via an internal load cell. These instruments have been precisely crafted to provide years of reliable service. Each gauge comes euipped with a rechargeable battery pack, AC battery charger, Accessory Attachments, and a handy Carrying Case. 1.1 PRODUCT FEATURES The key features of the DFIS/AF2 Series are: Load capacity up to 500 lbs (250 kg, 2.5 kn) Internal sampling rate of 1000 Hz Displays Tension (T) or Compression (C) Simple keypad operation Rugged aluminum housing offers long life. Top (Front) housing is reversible Mounting arrangements for attaching to Manual and Motorized Test Stands Readings can be scaled using several engineering units: LB, KG, N Field calibration Analog and Digital outputs through common connector Can be controlled remotely via RS-232 port 1.2 ELECTRICAL SAFETY These instruments have been assessed against the essentail health and safety reuirements of the Low Voltage and the EMC Directives listed below. Based on conformity with the listed directives, the DFIS and AF2 are compliant with the following: Directive Test Standard EMC Directive (89/336/EEC) EN EMC Generic Emissions Standard EN EMC Generic Immunity Standard Low Voltage Directive (73/23/EEC) EN Safety Reuirement for Electrical Euip 1.3 RECHARGING BATTERIES A rechargeable NiCad battery powers the DFIS/AF2Series. Since NiCad batteries continue to discharge when not in use, it may be necessary to recharge the unit before use. When the battery power begins to lose its charge, the Battery Indicator will appear on the display. To recharge the battery, plug the battery charger included with the gauge into the correct voltage source and insert the charger jack into the receptacle on the end of the gauge. Charge the battery for hours to yield approximately 8-10 hours of continuous operation. The gauge can be operated continuously with the charger connected. _ WARNING - Before using the gauge, please fully charge the Nicad battery. Use only the battery charger supplied with the gauge. Using other chargers can overcharge and damage the battery
6 1.4 PHYSICAL DIMENSIONS 1.5 REVERSIBLE HOUSING The DFIS/AF2 Series is shipped shaft-down for stand-mounted applications, and the display on the front of the gauge is configured accordingly. For performing operations in which the gauge will be hand-held, it may be desirable to rotate the front of the casing so that the display and logo will be shown right side up. To do so, remove the two screws from the back of the casing, carefully rotate the front of the casing 180 degrees being careful not to pull, pinch, or twist the wires inside, and then reinstall the screws. NOTE: When removing the screws to the back casing for any reason, please take the proper ESD (Electro Static Discharge) precautions. Touching the circuit board without taking the proper ESD precautions may cause damage to the circuit board
7 1.6 ABOUT THE KEYPAD The keypad consists of four dedicated keys (PEAK, UNITS, ZERO, XMIT). The keypad is operational when the power switch, located on the side of the gauge, is in the ON position. PEAK key cycles through three different operating modes: capture of the Normal reading; Peak Compression reading; or Peak Tension reading. The selected mode is indicated in the upper left of the display screen. UNITS PEAK UNITS key allows you to select the display reading in lbs (LB), kilograms (KG) or newtons (N). ZERO key returns current display to zero reading. Depressing this key clears peak readings as well. XMIT ZERO XMIT is used to initiate data transmission to an external device, e.g. personal computer. This key is also used to display and modify the gauge s communications parameters. Figure 1.2 Keypad Layout 1.7 ABOUT THE DISPLAY The display is an LCD seven segment display that is easily read even at a distance. The display illustrates different symbols such as: battery symbol (BATT) indicating the battery is running low; a C for compression or a T for tension. The peak operational modes are also displayed (T PEAK, C PEAK). The right portion of the display indicates the force measurement units (LB, KG, N). The display updates 4 times per second. T Tension C Compression PEAK Combined with T or C to indicate peak BATT Indicates low battery condition 1. T C PEAK BATT LB KG N LB, KG, N Force measurement indicator -WARNING - Force reading accuracy may be affected by a low battery condition. BEFORE using the force gauge for the first time, fully charge the rechargeable NiCd battery. Plug the charger into the gauge s charge jack and into an AC outlet. Always charge the battery whenever the BATT indicator is highlited. A minimum charge of 8 hours is recommended
8 1.8 CONNECTION TO EXTERNAL DEVICES The DFIS/AF2 Series can be configured to accept input from an external source, and/or to send output to an external data-recording device, such as a strip chart recorder, an RS-232 or MITUTOYO printer, or a personal computer. It can be used in combination with CHATILLON Test Stands and application software such as NEXYGEN to meet a wide variety of application reuirements. To connect the gauge to a device, attach the appropriate interface cable to the port located at the top of the gauge, and the other end to the external device. Power Adapter I/O Connector I/O Connector (View from Solder End) The DFIS/AF2 Series is supplied with a 12-pin female connector to provide both digital and analog outputs. AMETEK offers a variety of cables to handle most applications (Refer to Section 1.9). Select the cable type reuired to connect the gauge to the peripheral device using the mating connector. The pinout of the connector is shown above. Pin assignments are defined in the table below. PIN Symbol I/O Purpose Description 1 TXD O RS-232 Transmitted Data 2 RXD I RS-232 Received Data 3 GND - Ground Digital Ground 4 - Clock Mitutoyo Clock 5 I Ready Mitutoyo Ready 6 I Reuest Mitutoyo Reuest 7 O Data Mitutoyo Data 8 (Reserved) 9 (Reserved) 10 - Ground Digital Ground 11 - Analog GND Analog Ground 12 O Analog SIG Analog Output Table 1.8 I/O Connector Pinout - 8 -
9 1.9 INTERFACE CABLES Cable NC NC NC NC NC NC NC NC NC ENC0125 Description 6 ft (182 cm), Connects to a personal computer with a 25-pin RS-232 connection. 10 ft (305 cm), Connects gauge to a personal computer with a 25-pin RS-232 connection. 15 ft (450 cm), Connectsto a personal computer with a 25-pin RS-232 connection. 20 ft (610 cm), Connects to a personal computer with a 25-pin RS-232 connection. 6 ft (182 cm) Connects gauge to X-Y recorder (double banana) 5 ft ( 152 cm) Connects gauge to a Mitutoyo device, 10-pin 10 ft (305 cm) Connects gauge to a Mitutoyo device,rs ft (305 cm) Connects gauge to a Datamyte 6 ft (182 cm) Connects to CHATILLON Model TCD 8 ft (243 cm) Connects to CHATILLON Model TCM - 9 -
10 1.10 SETTING UP FOR SAFE USE The DFIS/AF2 Series should be properly setup before accurate and most of all, safe measurements can be made. Only when you have verified that you will be conducting a safe test, proceed as follows: 1. Determine what and how it will be tested and verify that the DFIS/AF2 Series can handle the test. 2. If desired, mount the gauge in the test fixture. 3. If you will be powering the instrument from the AC line, plug the charger into its jack on the instrument. Plug the charger into the correct AC voltage outlet (the reuired AC voltage and freuency are listed on the charger). 4. Otherwise, if you will be powering the instrument from batteries, fully charge the batteries before using. IMPORTANT NOTE: Before using your new DFIS/AF2 Series, fully charge the rechargeable Nicad battery by plugging the supplied battery charger into the battery charger jack at the top of the gauge and the proper AC outlet. Use only the battery charger supplied with the gauge. Using other chargers can overcharge and damage the battery. 5. Attach all necessary attachments and plug in the cables. Route the cables so they do not interfere with the test. 6. Zero the gauge by depressing the ZERO key. 7. Attach the specimen and begin to take measurements
11 2.0 FRONT PANEL OPERATION The DFIS/AF2 Series can operate in one of the three modes: Normal, Data Collect and Configuration. 2.1 NORMAL MODE The principal function of the Normal mode is to provide indications of load applied to the load cell. Instantaneous force, peak-tension and peak-compression readings can be displayed and scaled in a variety of units of measure Selecting Units of Measure Force indications can be displayed in LB, KG, or N. The gauge displays the active units of measure on the LCD display. Pressing the UNITS key will cycle between the three available measuring units. UNITS LB KG N With gauge power ON, depress the UNITS key to cycle through available force measurement units (LB, KG, N). Release the key when the desired unit is displayed Selecting Type of Measurement Capturing Peaks The gauge continuously measures instantaneous tension or compression forces (normal readings). It also captures peak-tension and peak-compression forces. The gauge displays which measurement is being indicated. Pressing the PEAK key will cycle between the available force measurement modes. C LB PEAK T PEAK LB C PEAK LB With gauge power ON, depress the PEAK key to cycle through available force modes (C, T, T PEAK, C PEAK). Release the key when the desired unit is displayed. The display will indicate C until tension is applied to the loadcell. This will cause the display to indicate tension (T)
12 2.1.3 Zeroing Zeroing allows you to exclude the weight of accessories and attachments from the indicated force readings. It also clears captured peak tension and compression readings. To zero the gauge, momentarily press the ZERO key. Only the current measurement type (i.e., NORM, T-PK, C-PK) will be zeroed. To zero and clear all measurement types press and hold the PEAK key and then press the ZERO key. Zeroing the gauge takes approximately 0.5 seconds. ZERO 0 C PEAK LB To TARE, affix the reuired adapter or fixture to the gauge. Orient the gauge so that it is positioned as it will be during testing. Depress the ZERO key to tare the gauge. To ZERO, after a test, depress the ZERO key to zero the force readings. Removing power to the gauge will also zero the force readings Transmitting Readings Pressing the XMIT key will cause the gauge to transmit a single reading (current measurement type and current units) through the output port (MITUTOYO or RS-232). If the gauge is in the NORM mode, the data value (NORM, T-PK or C-PK) displayed on the screen will be transmitted. If the gauge is in the DATA COLLECT mode, where a series of measurements are taken, the gauge will transmit an averaged value of the outputs at the loadcell. XMIT 4 C PEAK LB To TRANSMIT, depress the XMIT key to transmit the instantaneous force reading from the gauge to a personal computer or external device. The proper interface cable must be connected to the gauge and the device
13 2.2 CONFIGURATION MODE The Configuration Mode facilitates the communications setup for the gauge Configuring Output Port- Analog Output The analog output is always available at the output port without configuration. The analog output is 0 to -1.50Vdc (min) in tension mode and 0 to 1.50Vdc (min) in compression mode Configuring Output Port- Digital Output The DFIS/AF2 gauge supports three types of digital output: Modified BCD (Mitutoyo), RS232 Numerical Data Only and RS232 Full Data. When the digital output is configured for RS232 Numerical Data Only, the format of the data being transmitted to the external device is: - XXXXX^^^<CR><LF> <LF> = Line Feed <CR> = Carriage Return ^^^ = 3 spaces (Note: ^^^ are NOT shown) X X X X X = 5-character Force Value with decimal point SPACE = Compression Reading (C) - = Tension Reading (T) When the digital output is configured for RS232 Full Data, the format of the data being transmitted to the external device is: - XXXXX^YY <CR><LF> <LF> = Line Feed <CR> = Carriage Return YY = Units of Measure (2 characters) ^ = 1 space (Note: ^ is NOT shown) X X X X X = 5-character Force Value with decimal point SPACE = Compression Reading (C) - = Tension Reading (T)
14 Enabling MITUTOYO Digital Output To enable the MITUTOYO digital output, turn the gauge power OFF. UNITS XMIT H - NOTICE - You must power down the gauge to configure the digital output and communication specifications for the gauge. To enable the MITUTOYO output, with gauge power OFF, DEPRESS and HOLD the XMIT key while simultaneously turning the gauge power switch to the ON position. Release the XMIT key after approximately 2 seconds. The gauge will display the current communication parameters for the gauge. The parameters will be displayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity. - NOTICE - Your gauge is shipped with the following factory settings: Baud Rate = 9600 Units = 0 1 (Transmit with Units) Word Length/Stop Bits = 7-2 (7 bit word/2 stop bits Parity = -P (No Parity) To select the MITUTOYO output, change the BAUD RATE to 000H by depressing the UNITS key until 000H is displayed. Once 000H is displayed (000H will flash), depress the XMIT key to accept and proceed to the next communication parameter. If you do not need to select a new parameter, the display will first flash 0000, then display the normal operating display to begin testing
15 Enabling RS232 Digital Output- Numerical Data Only To enable the RS232 digital output with numerical data only, turn the gauge power OFF. UNITS XMIT NOTICE - You must power down the gauge to configure the digital output and communication specifications for the gauge. To enable the RS232 NUMERICAL DATA ONLY output, with gauge power OFF, DEPRESS and HOLD the XMIT key while simultaneously turning the gauge power switch to the ON position. Release the XMIT key after approximately 2 seconds. The gauge will display the current communication parameters for the gauge. The parameters will be displayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity. - NOTICE - Your gauge is shipped with the following factory settings: Baud Rate = 9600 Units = 0 1 (Transmit with Units) Word Length/Stop Bits = 7-2 (7 bit word/2 stop bits Parity = -P (No Parity) To select the RS232 NUMERICAL DATA ONLY output, change the UNITS to 0 2 by depressing the UNITS key until 0 2 is displayed. Once 0 2 is displayed (0 2 will flash), depress the XMIT key to accept and proceed to the next communication parameter. If you do not need to select a new parameter, the display will first flash 0000, then display the normal operating display to begin testing
16 Enabling RS232 Digital Output- Full Data (Default) To enable the RS232 digital output with full data, turn the gauge power OFF. Full data means that UNITS will be included with your force value during transmission to an external device. UNITS XMIT NOTICE - You must power down the gauge to configure the digital output and communication specifications for the gauge. To enable the RS232 FULL DATA output, with gauge power OFF, DEPRESS and HOLD the XMIT key while simultaneously turning the gauge power switch to the ON position. Release the XMIT key after approximately 2 seconds. The gauge will display the current communication parameters for the gauge. The parameters will be displayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity. - NOTICE - Your gauge is shipped with the following factory settings: Baud Rate = 9600 Units = 0 1 (Transmit with Units) Word Length/Stop Bits = 7-2 (7 bit word/2 stop bits Parity = -P (No Parity) To select the RS232 FULL DATA output, change the UNITS to 0 1 by depressing the UNITS key until 0 1 is displayed. Once 0 1 is displayed (0 1 will flash), depress the XMIT key to accept and proceed to the next communication parameter. If you do not need to select a new parameter, the display will first flash 0000, then display the normal operating display to begin testing
17 Modifying Communication Parameters You may change the communication parameters for the force gauge by seuencing through the communication setup display. UNITS XMIT NOTICE - You must power down the gauge to configure the digital output and communication specifications for the gauge. To modify any communication parameter, with gauge power OFF, DEPRESS and HOLD the XMIT key while simultaneously turning the gauge power switch to the ON position. Release the XMIT key after approximately 2 seconds. The gauge will display the current communication parameters for the gauge. The parameters will be displayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity. - NOTICE - Your gauge is shipped with the following factory settings: Baud Rate = 9600 Units = 0 1 (Transmit with Units) Word Length/Stop Bits = 7-2 (7 bit word/2 stop bits Parity = -P (No Parity) To change any communication parameter, gain access to a parameter by depressing the XMIT key. BAUD RATE will be the first parameter displayed, followed by UNITS, WORD LENGTH, STOP BITS and PARITY in this order respectively. To change a parameter, once you have seuenced to the parameter, depress the UNITS key until you display the option desired for that specific parameter. To enable the parameter, depress the XMIT key. The selected parameter will flash. You must then depress the XMIT key to go to the next parameter. You must seuence through each parameter regardless of whether or not you will be making a change. After the PARITY parameter, the display will flash 0000 and then display the normal operating display
18 2.3 DATA COLLECT MODE The DFIS/AF2Series can accept and execute commands through the RS-232 serial port. The command set is tailored to make it easy to configure and operate the instrument under computer control About Data Collect Mode Data transmitted by the gauge in Data Collect mode is a filtered average of 45 instantaneous force readings. Digital filtering is employed to eliminate false readings resulting from noise and vibration. The external device, which the gauge is communicating to, e.g. personal computer, must be configured to transmit and ASCII character F to the gauge. This signals the gauge to go into, or out of DATA COLLECT mode. When the external device sends an ASCII character? or X to the gauge, the gauge will always respond even if the instantaneous reading is a zero value. This permits real time data collection to begin at a force reading or zero value. The ASCII character? or X is a reuest to the gauge to transmit data. The gauge will return to Normal mode when it receives and ASCII character F. The table on the next page shows the available ASCII characters and their functions. The diagram below will illustrate how the gauge operates based on these ASCII characters. - NOTICE - Each ASCII character transmission will advances the gauge one step, similar to a keypad entry. PC ASCII? or X Data Collect Mode ASCII F D ASCII P ASCII P N Normal Mode ASCII P ASCII P D Data Collect Mode ASCII F C Compression Peak Mode ASCII P ASCII P T Tension Peak Mode ASCII F D Data Collect Mode Figure 2.3 Data Collect Mode and ASCII Command Relationship
19 The DFIS/AF2Series can accept and execute commands through the RS-232 serial port. The command set is tailored to make it easy to configure and operate the instrument under computer control. The string commands are sent as ASCII characters. The following string commands are recognized: Command Response Description Legend: ^ is space (20 hex); 9 is a number 0 through 9; ± is a space or a minus; <CR> is Carriage Return (0D hex); <LF> is Line Feed (0A hex). ^^^^^^^^^^<CR><LF> On power-up, this line is sent. A F P R S U X or? Z Unit^=^lb^<CR><LF> Unit^=^kg^<CR><LF> Unit^=^N^^<CR><LF> ^N-MODE^^^<CR><LF> TP-MODE^^^<CR><LF> CP-MODE^^^<CR><LF> DC-MODE^^^<CR><LF> ±99.99^lb<CR><LF> ±99.99^kg<CR><LF> ±999.9^N^<CR><LF> ERROR^^^^^<CR><LF> Sends currently selected units to external device. Toggles between Data Collect and the Normal and Peak Modes Steps through Normal and Peak Modes: Normal, Tension Peak, Compression Peak Resets the gauge: zeroes all modes. Sends currently selected mode: Normal, Tension Peak, or Data Collect Steps through Units in the following order: lb, kg, N Sends data on display with the position of the decimal place the same as on display. If in Data Collect Mode, sends the Data Collect filter instead. If Transmit Units is set to transmit without units, lb, kg, and N are replaced by ^^. Response during force overload. Zeroes the currently selected mode: Normal, Tension Peak or Compression Peak
20 3.0 CALIBRATING THE DFIS/AF2 SERIES The DFIS/AF2 Series may be field calibrated, however, for optimum performance, it is recommended that your gauge be factory calibrated at least every other year or more freuently depending on your freuency of use. Field calibration is recommended every 6 to 12 months. A 5-point calibration is reuired where five different calibrated weights are applied to the load shaft at increments of 20%, 40%, 60%, 80% and 100% of capacity. - WARNING - The DFIS/AF2 Series should always be calibrated to full capacity. A 500 lb capacity gauge should be calibrated at a full scale reading of 500 lbs. If you have a 500 lb capacity gauge, yet you use the gauge between 0 and 100 lbs, YOU CANNOT calibrate the gauge as thought the full load is 100 lbs. You must have a way to mount the load cell vertically oriented before applying tension and compression force. The DFIS/AF2 Series should be calibrated using deadweights representing 100% of the rated full scale with 20% increments. For example, calibrating a 10 lb. gauge, the deadweights must be in 20% increments for a total of 10 lbs. at correct gravity. You must have a safe method of axially applying the deadweights to the load cell in both the tension and compression directions. - WARNING - You must depress ZERO at the completion of the calibration procedure to record the calibration data into the instrument s electronics. Data is NOT SAVED until the ZERO key is depressed. If, at any time during the calibration procedure that incorrect data is being displayed, simultaneously depress PEAK and ZERO keys to restart the calibration procedure. 3.1 CALIBRATION PROCEDURE This section leads you step-by-step through the calibration process for a DFIS/AF2Series: STEP 1 Gauge Pre-Set With gauge power OFF, PRESS and HOLD the UNITS and PEAK key. LB Switch gauge power to ON. Select the units to be used during the calibration. Depress the UNITS key to scroll through the available units (LB, KG, N). Press ZERO to select the desired unit. STEP 2 Select the desired capacity that will be calibrated (2, 10, 50, 100, 200 or 500). Depress the UNITS key to scroll through the available LB capacities. Press ZERO to select the desired capacity
21 STEP 3 Test Loadcell and A/D Converter C The display indicates the loadcell reading as a percent of the A/D span. Display the % A/D Mode by depressing the UNITS key. The % A/D Mode is displayed and should be below 6.0. Press ZERO. 1.6 C 7.1 Note: If the % A/D Mode is greater than 6.0, the loadcell may be damaged. The gauge should be returned to the factory or the loadcell should be replaced. - WARNING - Once the gauge has displayed -00- place the gauge flat onto its back side. DO NOT place the gauge in a vertical position and DO NOT mount the gauge to a test fixture. The gauge must be positioned on its back to remove any effects of gravity prior to the calibration. STEP 4 The display should indicate -00-, which indicates zero offset of the loadcell and that the loadcell is ready to be measured. Make sure the gauge is laying flat on its back side. DO NOT place the gauge in a vertical position while zero offset is being measured C 0.00 LB Depress the ZERO key to prepare the gauge for calibration. Allow a 5 minute warm up time before proceeding to Step 5. STEP 5 You may now position the gauge onto a test fixture (test stand) so that it is in COMPRESSION mode. You do not need to zero or tare the gauge. STEP 6 Add FULL LOAD to the gauge and exercise the loadcell. Remove the full load weight and repeat this process 2 additional times (Exercise the loadcell with full load a total of 3 times). C LB After applying and removing full load for the third time, depress the ZERO key. The display will prompt you to add the full load weight. C PEAK LB Note: A blinking number represents an incorrect weight or a problem with the load cell or gauge electronics. The gauge should be returned for factory service. If PEAK is displayed, the weight is NOT steady. Wait for the weight to stabilize and for the PEAK indicator to disappear before going to the next step. Once stable, depress the ZERO key
22 STEP 7 The gauge will display the next value of weight to be applied to the loadcell (80% capacity). Add 80% of full load to the gauge. C LB The display should indicate the same value of the calibrate load weight. The display reading should not blink. Depress the ZERO key. Note: A blinking number represents an incorrect weight or a problem with the load cell or gauge electronics. The gauge should be returned for factory service. STEPS 8, 9, 10 The gauge will continue to display the next value of weight to be applied to the loadcell. Repeat the process until all 5-points have been successfully calibrated in COMPRESSION mode. STEP 11 Remove the gauge from the test fixture and from Compression mode and re-orient the gauge into TENSION mode. Note: The display will indicate TENSION mode by a blinking T, blinking Units or blinking decimal point. A blinking C indicates a COMPRESSION verify mode. Repeat STEPS 6-10 with the gauge in TENSION mode. STEP 12 After the gauge has been calibrated in both Compression and Tension modes, depress the ZERO key to exit the calibration process and record calibration data. The gauge is now calibrated. - WARNING - You must depress ZERO at the completion of the calibration procedure to record the calibration data into the instrument s electronics. Data is NOT SAVED until the ZERO key is depressed. If, at any time during the calibration procedure that incorrect data is being displayed, simultaneously depress PEAK and ZERO keys to restart the calibration procedure
23 4.0 DIAGNOSTICS In the Self Test operation, the DFIS/AF2 Series performs a series of self-diagnostic tests. diagnostics, the gauge will display... and then the following information: During self ZERO... - NOTICE - You must power down the gauge to perform self diagnostic tests. To enable the SELF DIAGNOSTIC, with gauge power OFF, DEPRESS and HOLD the ZERO key while simultaneously turning the gauge power switch to the ON position. 4.1 SOFTWARE PART NUMBER ZERO 0877 To display the SOFTWARE PART NUMBER used with the gauge, depress the ZERO key. A four digit number will appear representing the software part number. 4.2 SOFTWARE REVISION LEVEL ZERO To display the SOFTWARE REVISION NUMBER used with the gauge, depress the ZERO key. The gauge will display the revision number, e.g. Revision
24 4.3 CAPACITY AND UNITS ZERO 10 LB To display the CAPACITY and UNITS used with the gauge, depress the ZERO key. The gauge will display the capacity and units used in the previous and most current calibration. - WARNING - If no capacity or calibration data is found in the gauge s EPROM, the display with blink -EL. This indicates that the gauge must be recalibrated. -EL 4.4 NO LOAD ZERO VALUE ZERO 0.0 C To display the NO LOAD ZERO, depress the ZERO key. The gauge will display the no load zero measured during the previous and most current calibration. The display will indicate the no load zero reading in % of A/D range). - WARNING - You must place the gauge on its back side in a horizontal position. ZERO Place the gauge on its back side in a horizontal position, depress the ZERO key. The gauge will display a blinking -00- then display the applied force in % of A/D range until ZERO is depressed again. Press ZERO to display the current no load value
25 ZERO Depress the ZERO key and the gauge will compare the current and stored no load offsets (zeros). If the gauge displays -0-, the gauge is within specification. If another value is displayed, the current offset has shifted too far from the stored offset and recalibration is reuired. The difference between the current offset and tolerance is displayed as a % of A/D range. 4.5 RS232 LOOPBACK TEST ZERO L00P Depress the ZERO key and the gauge will transmit an ASCII character U (See Section X.X Data Collect Mode) back to itself via an RS232 loopback connector (Not included). Depress ZERO again to exit the loopback test. With a loopback connector (Txc jumpered to Rxd); depress XMIT to send a continuous U. The gauge will display a count of the number of transmission errors. The display will blink. Depress XMIT and the loopback test will toggle ON (Blinking) and OFF (steady). 4.6 KEYPAD TEST ZERO 1 Depress the ZERO key and the gauge will perform a KEYPAD test. The gauge will display the number of each key that was depressed: UNITS = 1, PEAK = 2, XMIT = 3, and ZERO = 4. Turn power OFF to exit diagnostics mode
26 5.0 TECHNICAL REFERENCE & SPECIFICATIONS Accuracy DFIS Series: AF2 Series: +0.15% of Full Scale +1 LSC +0.2% of Full Scale +1 LSC Tare Capacity 10% in order to utilize Full Scale (in lb Mode); can tare more than 10%, however user may not have use of gauge s Full Scale capacity during testing Deflection for loadcell Safe Overload Gauge will display when the force applied exceeds approximately 120% of the gauge s capacity Overload Maximum overload is 150%. Load Cell deformation may occur when overload exceeds 150%. Contact your local AMETEK Representative if you experienced an overload beyond 150%. Display Update 8 updates per second in Normal or Peak Mode Sampling Rate 1,000 samples per second Battery Life 8-10 hours of continuous operation between charges Temperature Range 40 to 110 o F (5 to 45 o C) Temperature Stability 0.06% per o F Warranty 1 year Analog Output (Min) to 1.50Vdc (Min) at Full Scale Digital Output RS-232 and MITUTOYO
27 6.0 MODEL NUMBERS CHATILLON Brand Model Capacity AMETEK Brand Model Capacity DFIS2 2 lb x lb 1 kg x kg 10 N x 0.01 N AF lb x lb 1 kg x kg 10 N x 0.01 N DFIS10 10 lb x 0.01 lb 5 kg x kg 50 N x 0.05 N AF lb x 0.01 lb 5 kg x kg 50 N x 0.05 N DFIS50 50 lb x 0.05 lb 25 kg x 0.02 kg 250 N x 0.2 N AF lb x 0.05 lb 25 kg x 0.02 kg 250 N x 0.2 N DFIS lb x 0.1 lb 50 kg x 0.05 kg 500 N x 0.5 N AF lb x 0.1 lb 50 kg x 0.05 kg 500 N x 0.5 N DFIS lb x 0.2 lb 100 kg x 0.1 kg 1000 N x 1 N AF lb x 0.2 lb 100 kg x 0.1 kg 1000 N x 1 N DFIS lb x 0.5 lb 250 kg x 0.2 kg 2500 N x 2 N AF lb x 0.5 lb 250 kg x 0.2 kg 2500 N x 2 N
28 TEST AND CALIBRATION INSTRUMENTS Internet Addresses: Americas Europe Asia Pacific AMETEK Test and Calibration Instruments Division 8600 Somerset Drive Largo, Florida USA AMETEK Lloyd Instruments Forum House 12 Barnes Wallis Road Segensworth East Fareham AMETEK Lloyd Instruments No. 7 Sherwood Place Alexander Heights 6064 Perth Australia Tel Hampshire PO 15 5TT Tel Tel (USA only) Tel +44 (0) Fax [email protected] Fax +44 (0) Fax AMETEK Singapore Pvt. Ltd. AMETEK Lloyd Instruments SA 3 Avenue des Coudriers Zone d Activite de l Observatoire 7810 Montigny-Le-Bretonneux 10 Ang Mo Kio Street 65 #05-12 TECHPOINT Singapore France Tel Tel Fax Fax AMETEK and CHATILLON are registered trademarks of AMETEK, Inc. LLOYD INSTRUMENTS is an AMETEK Inc. trademark Information within this document is subject to change without notice. AMETEK Precision Instruments Europe GmbH Rudolf-Diesel-Strasse 16 D-40670, Meerbusch Germany Tel Fax ISO 9001 Manufacturer Pub No. NC Issued 05/01 Copyright 2001, by AMETEK, Inc Printed in U.S.A.
CHATILLON DG Series Mechanical Force Gauge. Operating Instructions
Instruction Manual NC002737 Feb 2003 Issue 1 CHATILLON DG Series Mechanical Force Gauge Operating Instructions The DG Series mechanical force gauge is ideal for a wide range of force testing applications.
SPECIFICATIONS. +0.3% of Full Scale +1 LSC. LCD, 4-1/2 characters. RS232. Mitutoyo, 12-Pin Hirose AC Adapter, NiCd Battery Pack. and Battery Charger
DFGS-R-ND SERIES Digital Force Gauge The CHATILLON DFGS-R-ND Seriescomes without a loadcell. Instead, you specify one or many different REMOTE Series loadcells to use with the gauge. These loadcells simply
Part No. NC002739 April 2010. DMG Series Mechanical Dynamometers. User Manual. DMG Series User Manual 1
Part No. NC002739 April 2010 DMG Series Mechanical Dynamometers User Manual DMG Series User Manual 1 WARRANTY This instrument is warranted against defects in workmanship, material and design for one (1)
Part No. NC002737 May 2012. DG Series Mechanical Force Gauges. User Manual. DG Series User Manual 1
Part No. NC002737 May 2012 DG Series Mechanical Force Gauges User Manual DG Series User Manual 1 WARRANTY This instrument is warranted against defects in workmanship, material and design for two (2) years
XP Series Balances Operation Manual
XP Series Balances Operation Manual North and South America: 6542 Fig Street Arvada, Colorado 84 1-8-321-1135 Tel: 33-431-7255 Fax: 33-423-4831 U.K. and Ireland: Denver House, Sovereign Way Trafalgar Business
Instruction Manual FL
3. Operation Overview The most common used features (such as displaying force, peak hold, zero and changing of displayed units) can all be done by pressing a single dedicated key identified on the front
BW-500 BED WEIGHER USER MANUAL
BW-500 BED WEIGHER USER MANUAL 0 1 TABLE OF CONTENTS SPECIFICATION... 3 Equipment... 4 INSTALLING WEIGH BEAM... 5 PREPARATION FOR MEASUREMENT... 6 DP3800 Indicator... 8 POWER SUPPLY... 8 PANEL... 9 KEY
User Manual. Humidity-Temperature Chart Recorder. Model RH520
User Manual Humidity-Temperature Chart Recorder Model RH520 Introduction Congratulations on your purchase of the Extech RH520 Temperature + Humidity Chart Recorder. The RH520 measures and displays Temperature,
User's Guide. Integrating Sound Level Datalogger. Model 407780. Introduction
User's Guide 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 407780 Product Page Integrating Sound Level Datalogger
When any of the following symbols appear, read the associated information carefully. Symbol Meaning Description
Samba OPLC SM35-J-R20 Installation Guide The Unitronics SM35-J-R20 offers the following onboard I/Os: 12 Digital Inputs, configurable via wiring to include: 1 HSC/Shaft-encoder Input, 2 Analog inputs (only
Model 5511 Filler Controller User s Manual Version 1.1 October 2011
Thompson Scale Company WEIGHING SYSTEMS & PACKAGING MACHINERY CONTROLS 2758 Bingle Road Houston, Texas 77055 Phone: 713/932-9071 Fax: 713/932-9379 www.thompsonscale.com Model 5511 Filler Controller User
Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF
Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator
TCP/IP MODULE CA-ETHR-A INSTALLATION MANUAL
TCP/IP MODULE CA-ETHR-A INSTALLATION MANUAL w w w. c d v g r o u p. c o m CA-ETHR-A: TCP/IP Module Installation Manual Page Table of Contents Introduction...5 Hardware Components... 6 Technical Specifications...
Back-UPS Pro 1300/1500 Installation and Operation
Back-UPS Pro 1300/1500 Installation and Operation Inventory Safety Do not install the Back-UPS in direct sunlight, in excessive heat, humidity, or in contact with fluids. Connect the battery bu059a bu058a
ATS Overhead Table Shelf System INSTRUCTION MANUAL
ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System Instruction Manual Warranty Newport Corporation warrants this product to be free of defects in material and workmanship
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
ValuPrompt Computer Prompter Instruction Manual
ValuPrompt Computer Prompter Instruction Manual For Technical Information Please Call Voice: (604) 985-2250 Toll Free: 877-TEKSKIL (835-7545) Toll Free Fax: 877-576-8361 Or Visit Our Website tekskil.com
PPM Users Manual Signature Software 01-12-00
PPM Users Manual Signature Software 0-2-00 PPM User Manual /8/02 Software Versions: 0.0.27 Contents. Introduction 2 2. Parameters 3 2. Overload Limit...4 2.2 Relative Upper Limit...4 2.3 Relative Lower
Keep it Simple Timing
Keep it Simple Timing Support... 1 Introduction... 2 Turn On and Go... 3 Start Clock for Orienteering... 3 Pre Start Clock for Orienteering... 3 Real Time / Finish Clock... 3 Timer Clock... 4 Configuring
HIGH PRECISION BALANCE MWP OWNER S MANUAL. MWP-2004-12 ver.1.0
HIGH PRECISION BALANCE MWP OWNER S MANUAL MWP-2004-12 ver.1.0 INTRODUCTION------------------------------------------------1 1. INSTALLATION---------------------------------------------------2 Unpacking---------------------------------------------------------------2
T7560A,B,C Digital Wall Module
T7560A,B,C Digital Wall Module HONEYWELL EXCEL 5000 OPEN SYSTEM BEFORE INSTALLATION All wiring must comply with local electrical codes and ordinances or as specified on installation wiring diagrams. Digital
BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER. Installation, Operation, and Maintenance Instructions. Manual No.
BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER Installation, Operation, and Maintenance Instructions Manual No. M06-512 SPECIALISTS IN THE MEASUREMENT AND CONTROL OF VISCOSITY BROOKFIELD
OCSM Series. High Resolution Digital Crane Scale. User Guide
OCSM Series High Resolution Digital Crane Scale Content 1. Safety Guide... 1 2. Features... 1 3. Specifications... 2 4. Capacity... 3 5. Display & Keys... 3 Scale & Remote Keys... 3 Indicators... 4 Message...
Thermistor Thermometer MODEL NO. 93210-00
Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:
OPERATING INSTRUCTIONS Model ST-888 DTMF ANI/ENI Display Decoder
P R O D U C T G R O U P OPERATING INSTRUCTIONS Model ST-888 DTMF ANI/ENI Display Decoder Manual # 600-0901 November 30, 1999 Rev. D - 99068 DESCRIPTION The ST-888 Mobilecall Display Decoder is a desktop
Advantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
4 Port VGA Video Audio Switch with RS232 control
4 Port VGA Video Audio Switch with RS232 control VS410RVGAA *actual product may vary from photos DE: Bedienungsanleitung - de.startech.com FR: Guide de l'utilisateur - fr.startech.com ES: Guía del usuario
OPLC Installation Guide
Samba OPLC SM43-J-T20/ SM35-J-T20 OPLC Installation Guide 12 Digital Inputs, include 1 HSC/Shaft-encoder Input, 2 Analog inputs (only when the digital inputs are set to pnp), 8 Relay Outputs 12 Digital
Programming and Using the Courier V.Everything Modem for Remote Operation of DDF6000
Programming and Using the Courier V.Everything Modem for Remote Operation of DDF6000 1.0 Introduction A Technical Application Note from Doppler System July 5, 1999 Version 3.x of the DDF6000, running version
2X1 DVI Switch (DS-21R) 4X1 DVI Switch (DS-41R) USER MANUAL. DTronics Inc.
2X1 DVI Switch (DS-21R) 4X1 DVI Switch (DS-41R) USER MANUAL DTronics Inc. 40 State Highway 94, Suite L2A McAfee, NJ 07428 201-362-1606 Fax: 781-207-0351 UNPACKING Each DVI Link DVI Switch package includes
Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection
w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force
HP UPS R1500 Generation 3
HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications
RS232C < - > RS485 CONVERTER S MANUAL. Model: LD15U. Phone: 91-79-4002 4896 / 97 / 98 (M) 0-98253-50221 www.interfaceproducts.info
RS232C < - > RS485 CONVERTER S MANUAL Model: LD15U INTRODUCTION Milestone s model LD-15U is a RS232 to RS 485 converter is designed for highspeed data transmission between computer system and or peripherals
TABLE 1: Wiring Terminals. Connect to... 1C 1H 2C 2H 1H1C 2H1C 2H2C 3H2C
Installation TURN OFF POWER TO THE SYSTEM AT THE MAIN POWER PANEL TO AVOID ELECTRICAL SHOCK. Installation should be carried out by an electrician or a qualified technician. 1.1 Find a Location for the
Technical Manual. For use with Caller ID signaling types: Belcore 202, British Telecom, & ETSI
Technical Manual For use with Caller ID signaling types: Belcore 202, British Telecom, & ETSI Caller ID.com WHOZZ CALLING? POS 2 Caller ID Monitoring Unit Technical Manual For use with Caller ID signaling
Instruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
Quick Installation. A Series of Intelligent Bar Code Reader with NeuroFuzzy Decoding. Quick Installation
Quick Installation A Series of Intelligent Bar Code Reader with NeuroFuzzy Decoding This chapter intends to get your new FuzzyScan scanner working with your existing system within minutes. General instructions
Single Station Remote Alarm
ADI 5106G Certified ISO 9001:2000 Single Station Remote Alarm 529 5106-01-120 529 5106-01-220 INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Controls Corporation
EMBEDDED ACCESS CONTROL Hardware Installation Guide
EMBEDDED ACCESS CONTROL Hardware Installation Guide Lenel goentry Hardware Installation Guide, product version 1.00. This guide is item number DOC- ENHW-ENU, revision 1.003, April 2009 Copyright 2009 Lenel
SATEX SA 230. - Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND
- Manual SATEX SA 230 PRECISION BALANCE Weegtechniek holland b.v. Patroonsweg 23-27 3892 DA Zeewolde HOLLAND Tel. + 31 36 522 20 30 Fax. + 31 36 522 20 60 Email [email protected] Website www.satex.nl page
HDMI Switch USER MANUAL VS481A
HDMI Switch USER MANUAL VS481A FCC Information This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
Master Time Clock MTC-200 MTC-400 MTC-600. Users Manual
Master Time Clock MTC-200 MTC-400 MTC-600 Users Manual Toll Free (888)713-0373 Phone (972)987-4408 FAX (877)720-9291 www.midwest-time.com [email protected] TABLE OF CONTENTS TOPIC PAGE GENERAL DESCRIPTION
AC-115 Compact Networked Single Door Controller. Installation and User Manual
AC-115 Compact Networked Single Controller Installation and User Manual December 2007 Table of Contents Table of Contents 1. Introduction...5 1.1 Key Features... 6 1.2 Technical Specifications... 7 2.
Model SRMD Setra Remote Monitoring Display
Model SRMD Setra Remote Monitoring Display 1.0 GENERAL INFORMATION Thank you for purchasing the Setra Remote Monitoring Display (SRMD). The SRMD is a digital panel meter with a bright 1 LED display for
F2400 FOM II Series Fiber Optic Modem Technical Manual
F2400 FOM II Series Fiber Optic Modem Technical Manual T1 Revision B Copyright April 2003 VERSITRON, Inc. 83 Albe Drive / Suite C Newark, DE 19702 www.versitron.com A030430283T PROPRIETARY DATA All data
Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS
+ Installation and Operation Manual Back-UPS BX800CI-AS/BX1100CI-AS Inventory Safety and General Information bu001c This unit is intended for indoor use only. Do not operate this unit in direct sunlight,
STAGE SETTER-8. User Instructions. Elation Professional
STAGE SETTER-8 User Instructions Elation Professional A Division of the American DJ Group of Companies 4295 Charter Street Los Angeles, CA 90058 USA Tel: 323-582-2650 Fax: 323-582-2610 Web: www.elationlighting.com
USB to RS-422/485 Serial Adapter (ID-SC0911-S1/SC0A11-S1) User s Manual
USB to RS-422/485 Serial Adapter (ID-SC0911-S1/SC0A11-S1) User s Manual 1 Copyright Statement No part of this publication may be reproduced in any form by any means without the prior written permission.
USER MANUAL. Model 6103 3U ForeFront Alarm Card SALES OFFICE (301) 975-1000 TECHNICAL SUPPORT (301) 975-1007. An ISO-9001 Certified Company
USER MANUAL Model 6103 3U ForeFront Alarm Card An ISO-9001 Certified Company Part# 07M6103 Doc# 123011U Rev. A Revised 7/15/03 SALES OFFICE (301) 975-1000 TECHNICAL SUPPORT (301) 975-1007 CONTENTS 1.0
Digital Weight Indicator Setup / Operation Manual
Digital Weight Indicator Setup / Operation Manual Revision 1.0 July 8, 2009 i TABLE OF CONTENTS Page INTRODUCTION... 1 FCC NOTE... 1 INSTALLATION... 1 PREPARATION... 1 CONNECTIONS... 2 CONNECTING THE WEIGH
ABB Drives. User s Manual. Pulse Encoder Interface Module RTAC-01
ABB Drives User s Manual Pulse Encoder Interface Module RTAC-0 Pulse Encoder Interface Module RTAC-0 User s Manual 3AFE 64486853 REV A EN EFFECTIVE:.5.00 00 ABB Oy. All Rights Reserved. Safety instructions
EZmoto V2. Product description Rev. 6 10/01/2014. EZmoto V2 Product description Rev.6 10/01/2014
EZmoto V2 Product description Rev. 6 10/01/2014 1 Contents 1. Overview... 3 2. Hardware Interface Description... 3 2.1 Main features of the EZmoto... 3 2.2 Hardware block diagram... 4 2.3 Internal Hardware
AGS. Owner's Manual. Xantrex Automatic Generator Start Control System
AGS Owner's Manual Xantrex Automatic Generator Start Control System TABLE OF CONTENTS INTRODUCTION...3 Main Features...3 Safety Summary...3 THINGS YOU SHOULD KNOW...4 THEORY OF OPERATION...5 System...5
USER MANUAL. MODEL 460 Series G.703 Coax to Twisted Pair Adapter (Balun)
USER MANUAL MODEL 460 Series G.703 Coax to Twisted Pair Adapter (Balun) An ISO-9001 Certified Company Part #07M460-E Doc. #019021UE Revised 02/23/00 SALES OFFICE (301) 975-1000 TECHNICAL SUPPORT (301)
Copyrights 2010 Victron Energy B.V. All Rights Reserved
Manual EN 1 Copyrights 2010 Victron Energy B.V. All Rights Reserved This publication or parts thereof may not be reproduced in any form, by any method, for any purpose. For conditions of use and permission
GENERATOR START CONTROL MODULE - MINI (2 Wire to 3 Wire)
FEATURES & APPLICATIONS Inexpensive 2 wire to 3 wire start controller for electric start high speed gas generators. Optimized for use with Outback Invertors. Supports three types of 3 wire generator control
User Manual. English. APC Smart-UPS SC. 1000/1500 VA 110/120/230 Vac. 2U Tower/Rack-Mount Uninterruptible Power Supply
User Manual English APC Smart-UPS SC 1000/1500 VA 110/120/230 Vac 2U Tower/Rack-Mount Uninterruptible Power Supply 990-1851C, 10/2005 Introduction The APC Uninterruptible Power Supply (UPS) is designed
JDSU PathTrak. HCU200 Install User s Guide
JDSU PathTrak HCU200 Install User s Guide Notice Copyright System Requirements The information contained in this document is the sole property of JDSU and may not be reproduced, copied, or communicated
NortechCommander Software Operating Manual MAN-00004 R6
NortechCommander Software Operating Manual MAN-00004 R6 If the equipment described herein bears the symbol, the said equipment complies with the applicable European Union Directive and Standards mentioned
Troubleshooting Tips Lifestyle SA-2 & SA-3 Amplifier. Troubleshooting Tips
Troubleshooting Tips Lifestyle SA-2 & SA-3 Amplifier Refer to the Lifestyle SA-2 & SA-3 Amplifier service manuals, part number 271720 for schematics, PCB layouts and parts lists. Preventative Repair Measures
How To Power A Power Control On An Ip40 (Ipl) With A Power Supply (Iplug) With An Ip20 Controller (Iphones) With Power Control (Power Control) With No Antenna) With The Ip20 (Power)
MODEL NUMBER: ISC910-1-0-GB-XX ISC911-5-0-GB-XX IXP20 CONTROLLER SPECIFICATIONS Working Environment Plastic Housing... Power ImproX IXP20 Controller INSTALLATION MANUAL Designed to work in an indoor (dry)
Digital Indicator. Revision 1.4 September 24, 2004. 1999-2004 Transcell Technology, Inc. Contents subject to change without notice.
TI-1600 Series Digital Indicator Setup / Operation Manual Revision 1.4 September 24, 2004 1999-2004 Transcell Technology, Inc. Contents subject to change without notice. Transcell Technology, Inc. 975
user s manual Battery Case model #: SPB3200 Battery Case Charger for Samsung Galaxy S 4
user s manual model #: SPB3200 Charger for Samsung Galaxy S 4 What s Included Unpack the battery case and make sure all accessories are put aside so they will not be lost. hello. USB to Micro USB Cable
RouteFinder SOHO. Quick Start Guide. SOHO Security Appliance. EDGE Models RF825-E, RF825-E-AP CDMA Models RF825-C-Nx, RF825-C-Nx-AP
RouteFinder SOHO SOHO Security Appliance EDGE Models RF825-E, RF825-E-AP CDMA Models RF825-C-Nx, RF825-C-Nx-AP Quick Start Guide RouteFinder RF825 Series Quick Start Guide RouteFinder SOHO Security Appliance
User Guide. HDMI Active Cable Extender. DVI-7370c
User Guide HDMI Active Cable Extender DVI-7370c TABLE OF CONTENTS SECTION PAGE PRODUCT SAFETY...1 PRODUCT LIABILITY...1 1.0 INTRODUCTION...2 2.0 SPECIFICATIONS...3 3.0 PACKAGE CONTENTS...4 4.0 CONNECTING
User Manual. Interface Adapter IA1 / Computer Interface EFOY 600 / EFOY 900 / EFOY 1200 / EFOY 1600
User Manual Interface Adapter IA1 / Computer Interface EFOY 600 / EFOY 900 / EFOY 1200 / EFOY 1600 2 1. Introduction 1. Introduction 1.1 Introduction Thank you for choosing an EFOY product. Please read
CelluLine CGW-TS GSM Cellular Gateway. Installation and Programming Manual
CelluLine CGW-TS GSM Cellular Gateway Installation and Programming Manual CelluLine CGW-TS GSM Cellular Gateway Installation and Programming Manual CGWTS-M001A Version 1, Release 1, December 2004 NOTICE
MCR-VDC-UI-B-DC. Voltage Transducer for DC Voltages. INTERFACE Data Sheet 100260_en_01. 1 Description
Voltage Transducer for DC Voltages TERFACE Data Sheet 000_en_0 PHOENIX CONTACT - /00 Description The MCR voltage transducer measures DC voltages in several signal ranges from 0... ±0 V DC to 0... ±0 V
Dear Customer, User Memo: Please visit us on facebook or twitter! Thank you for purchasing this product.
Owner s Manual 1 Dear Customer, Thank you for purchasing this product. For optimum performance and safety, please read these instructions carefully. User Memo: Date of purchase: Dealer name: Dealer address:
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &
INSTRUCTION MANUAL. FOR TECHNICAL ASSISTANCE: www.atlasrr.com [email protected] VERSION 3.2 ITEM #330
LIMITED ONE-YEAR WARRANTY Atlas Model Railroad Company, Inc. warrants that this Commander will be free from defects in material and workmanship for a period of one year from the date of purchase. If this
WEA-Base. User manual for load cell transmitters. UK WEA-Base User manual for load cell transmitters Version 3.2 UK
WEA-Base User manual for load cell transmitters 1 Contents 1. Technical data... 3 2. Assembly... 4 2.1 Power supply... 4 2.2 Load cells... 4 2.3 RS-485... 4 2.4 Relays... 5 2.5 Digital input... 5 2.6 Analogue
How To Use A Ci 2001A/B (I2) With A Power Supply (I3)
Weighing Indicator CI-2001A/B OWNER S MANUAL CONTENTS PRECAUTIONS 4 INTRODUCTION 6 THE FEATURES OF CI-2001A/B 6 TECHNICAL SPECIFICATION 7 DIMENSIONS 8 FRONT PANEL(CI-2001A) 9 FRONT PANEL(CI-2001B) 11 REAR
TI-500E Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual
TI-500E Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual Revision 1.0 975 Deerfield Parkway Buffalo Grove, IL 60089 January 28, 2008 Tel (847) 419-9180 Fax (847) 419-1515 http://www.transcell.net
OPERATION MANUAL MARSDEN-200 PLEASE TAKE THE TIME TO READ THESE INSTRUCTIONS BEFORE STARTING TO USE THE SCALES
OPERATION MANUAL MARSDEN-200 PLEASE TAKE THE TIME TO READ THESE INSTRUCTIONS BEFORE STARTING TO USE THE SCALES 1 TABLE OF CONTENTS POWER SUPPLY... 3 SPECIFICATIONS... 3 MAINATENANCE AND GENERAL CARE........
Series 6000 Torque measured metal bellow coupling
Properties Free of float metal bellow coupling with integrated torque measurement Non-contact measurement system, high robustness High torsional stiffness Limited torque of inertia Performance Measurement
Sierra Dual 24 Volt Brushless DC Motor Controller Product Specification
Sierra Dual 24 Volt Brushless DC Motor Controller Product Specification Assembly 025A0215 600A0942 Rev. A May 14, 2012 025A0215 Brushless DC Motor Controller Page 1 Revision History ECN # Date Rev Description
USB/VGA Cat 5 UTP Console Extender
USB/VGA Cat 5 UTP Console Extender SV565UTPU SV565UTPUGB SV565UTPUEU Instruction Manual FCC Compliance Statement This equipment has been tested and found to comply with the limits for a Class B digital
INSTALLER S & OWNER S MANUAL
INSTALLER S & OWNER S MANUAL HVAC INSTALLER: PLEASE LEAVE MANUAL FOR HOMEOWNER DEH 3000 DEH 3000 Part No. 4028539 Dehumidifier & Ventilation System Controller P.O. Box 8680 Madison, WI 53708 TOLL-FREE
PRODUCTIVITY THROUGH INNOVATION 600 CONTROL DIRECT DRIVE TECHNICAL/OPERATION MANUAL
Rev. D PRODUCTIVITY THROUGH INNOVATION 600 CONTROL DIRECT DRIVE TECHNICAL/OPERATION MANUAL 10 BORIGHT AVENUE, KENILWORTH NEW JERSEY 07033 TELEPHONE: 800-524-0273 FAX: 908-686-9317 TABLE OF CONTENTS Page
Periodic Verification System Model 775PVS
Periodic Verification System Model 775PVS Instruction Manual Contents 1 Description... 1 775PVS System... 1 Model 775 Fieldmeter... 2 Model 775C Charger... 2 Model 775P Plate Assembly... 3 2 Operation...
8-Port HDMI Switch USER MANUAL VS0801H
8-Port HDMI Switch USER MANUAL VS0801H FCC Information This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
NEXUS FX04 and FX04-1 Servo Motor Installation. Fireye DESCRIPTION SAFETY INFORMATION. FX04-1 Shown
NEX-3004 AUGUST 4, 2015 Fireye NEXUS FX04 and FX04-1 Servo Motor Installation FX04-1 Shown DESCRIPTION This bulletin describes the installation and operating instructions for the FX04 and FX04-1, 3 lb.-ft./
CASTON. Crane Scale. OWNER S MANUAL (BT Ver.) 9007-CT3-BT33-0 2006. 07
Crane Scale CASTON OWNER S MANUAL (BT Ver.) 9007-CT3-BT33-0 2006. 07 CONTENTS PRECAUTIONS 4 OVERALL VIEW 7 SPECIFICATIONS 7 DISPLAY & KEY FUNCTIONS 8 BATTERY USAGE 9 USE OF BATTERY CHARGER 9 FUNCTIONS
1. SAFETY INFORMATION
RS-232 Sound Level Meter 72-860A INSTRUCTION MANUAL www.tenma.com 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter
GAUGEMASTER PRODIGY EXPRESS
GAUGEMASTER PRODIGY EXPRESS DCC01 USER MANUAL Version 1.1 2011 T A B L E O F C O N T E N T S 1 Getting Started Introduction Specifications and Features Quick Start Connecting to Your Layout Running a Loco
USER MANUAL MODEL 2300M SERIES. OpticLink Ethernet Media Converters (2300M, 2310M, 2311M, & 2312M)
USER MANUAL MODEL 2300M SERIES OpticLink Ethernet Media Converters (2300M, 2310M, 2311M, & 2312M) An ISO-9001 Certified Company Part# 07M2300M Doc# 077321U Rev. A Revised 4/14/03 SALES OFFICE (301) 975-1000
Operation and Installation Manual
Operation and Installation Manual RCM-10 Remote Control Monitor and RSM-10 Remote Status Monitor for the CNA-100 & CNA-200 Automations Revision 1.1 9/98 WARRANTY INFORMATION The RCM-10 Remote Control
NX-591NE-GSM NetworX GSM 3G HSPA+ Module Installation Sheet
NX-591NE-GSM NetworX GSM 3G HSPA+ Module Installation Sheet The NX-591NE-GSM is a microprocessor-controlled GSM interface module used to connect the NetworX series of control panels to GSM cellular networks
HP ProDisplay P19A LED Backlit Monitor. User Guide
HP ProDisplay P19A LED Backlit Monitor User Guide 2013 Hewlett-Packard Development Company, L.P. The only warranties for HP products and services are set forth in the express warranty statements accompanying
PERF10 Rubidium Atomic Clock
Owner s Manual Audio Stanford Research Systems Revision 1.0 February, 2011 2 Certification Stanford Research Systems certifies that this product met its published specifications at the time of shipment.
M3 Single-Slot Cradle. User Guide. Cradle Model: A-1CR-U002-040-R00, A-1CR-U00D-040-R00, A-1CR-US02-110-R00, A-1CR-U0ED-110-R00
M3 Single-Slot Cradle User Guide Cradle Model: A-1CR-U002-040-R00, A-1CR-U00D-040-R00, A-1CR-US02-110-R00, A-1CR-U0ED-110-R00 Compatible Device: M3 RED, GREEN and SKY Version 2.3 Released in March 2013
FTW 175 VERIZON Wireless Monitoring System Reference Manual Part Number 7911751VERIZON
FTW 175 VERIZON Wireless Monitoring System Reference Manual Part Number 7911751VERIZON SERIAL NUMBER Flash Technology, 332 Nichol Mill Lane, Franklin, TN 37067 (615) 261-2000 Front Matter Abstract This
M O D U L E - 7 E Model CS-Caliprompter Operator s Manual
O P E R A T I O N S A N D P R O C E D U R E S F O R C S - C A L I P R O M P T E R A N D A C C E S S O R I E S Model CS-Caliprompter Visual screen display allows easy step-by-step operation On-site programming
581 Rapid Transfer Device (RTD) Instruction Manual. Photo of product, with a.5 point black border
581 Rapid Transfer Device (RTD) Instruction Manual Photo of product, with a.5 point black border Part #60-9003-085 Copyright 1995, 2002. All rights reserved, Teledyne Isco, Inc. Revision J, April 20, 2011
Technical data. General specifications. Indicators/operating means. Electrical specifications Operating voltage U B Power consumption P 0 Interface
Release date: 06-0- 09: Date of issue: 06-0- 009_eng.xml Model Number Single head system Features Parameterization interface for the application-specific adjustment of the sensor setting via the service
Duct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
MINIMAT-EC-Servo Screwdriver Spindles
Screwdriving technology Automation Air motors Air tools Screwdriver Spindles electric MINIMAT-EC-Servo Screwdriver Spindles Maximum Flexibility and process Control Straight Spindle Form - Torque range
The Analyst RS422/RS232 Tester. With. VTR, Monitor, and Data Logging Option (LOG2) User Manual
12843 Foothill Blvd., Suite D Sylmar, CA 91342 818 898 3380 voice 818 898 3360 fax www.dnfcontrolscom The Analyst RS422/RS232 Tester With VTR, Monitor, and Data Logging Option (LOG2) User Manual Manual
Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001
Express5800/120Ed Rack Mount Kit Installation Procedures PN: 455-01607-001 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,