Australian Medical Devices Guidance Document Number 25. Classification of Medical Devices
|
|
|
- Leon Harrison
- 9 years ago
- Views:
Transcription
1 Australian Medical Devices Guidance Document Number 25 Classification of Medical Devices January 2005
2 DISCLAIMER This document is provided for guidance only. The classification examples provided might vary if a manufacturer assigns a different intended purpose to that used for the examples. Manufacturers and sponsors must use the Classification Rules, as contained in the Therapeutic Goods (Medical Devices) Regulations 2002, to classify their medical devices. Please refer to the Therapeutic Goods Act, 1989 as amended by the Therapeutic Goods Amendment (Medical Devices) Bill, 2002 and the Therapeutic Goods (Medical Devices) Regulations, 2002 for legislative requirements. FURTHER INFMATION The Medical Devices Information Unit of the Office of Devices, Blood and Tissues of the Therapeutic Goods Administration (TGA) can be contacted by: Telephone: Facsimile: (02) [email protected] Website: Mail: PO Box 100 Woden ACT 2606 Commonwealth of Australia
3 AMENDMENT SCHEDULE Version Number Date of Amendment Summary of Amendments 3
4 CONTENTS Disclaimer...2 Further Information...2 Amendment Schedule...3 Introduction...6 Purpose...6 Background...6 The classes of medical devices...6 The classification rules...7 Classifying a medical device...7 Deciding what classification rules apply...7 Important points to remember...8 Flowchart 1- deciding what classification rules apply...9 FLOWCHART 2 CLASSIFICATION RULES - NON INVASIVE DEVICES...10 RULE 2 NON INVASIVE MEDICAL DEVICES...11 Rule 2.1 Non-invasive medical devices general...11 Rule 2. 2 Non-invasive devices intended to channel or store blood, etc...11 Rule 2.3 Non-invasive devices intended to modify the biological or chemical composition of blood, etc...12 Rule 2.4 Non-invasive devices intended to have contact with injured skin...12 FLOWCHART 3 CLASSIFICATION RULES - INVASIVE DEVICES...13 FLOWCHART 3A CLASSIFICATION RULES - INVASIVE DEVICES...14 RULE 3 INVASIVE MEDICAL DEVICES...15 Rule 3.1 Invasive devices intended to be used to penetrate body orifices...15 Rule 3.2 Surgically invasive devices intended for transient use...15 Rule 3.3 Surgically invasive devices intended for short-term use...16 Rule 3.4 Surgically invasive devices for long-term use and implantable devices...17 FLOWCHART 4 CLASSIFICATION RULES - ACTIVE DEVICES...18 RULE 4 ACTIVE MEDICAL DEVICES...19 Rule 4.1 Active medical devices general...19 Rule 4.2 Active medical devices for therapy...19 Rule 4.3 Active medical devices for diagnosis...20 Rule 4.4 Active medical devices intended to administer or remove medicines, etc from a patient s body...20 FLOWCHART 5 CLASSIFICATION RULES Special rules...21 RULE 5 SPECIAL RULES...22 Rule 5.1 Devices incorporating a medicine...22 Rule 5.2 Devices for contraception or prevention of sexually transmitted diseases
5 Rule 5.3 Devices intended for disinfecting, cleaning, rinsing etc...22 Rule 5.4 Non-active devices intended to record X-ray diagnostic images...22 Rule 5.5 Devices containing non-viable animal tissues or derivatives...22 Rule 5.6 Devices that are blood bags...23 Rule 5.7 Active implantable medical devices...23 Rule 5.8 Medical devices for export only...23 Rule 5.9 Devices that are mammary implants...23 APPENDIX A - Definitions...24 Accessory...24 Active implantable medical device...24 Active medical device...24 Active medical device for diagnosis...24 Active medical device for therapy...24 Ancilliary action...25 Biological effect...25 Body orifice...25 Central circulatory system...25 Central nervous system...25 Chemical change...25 Direct diagnosis...26 Duration...26 Implantable medical device...26 Integral...26 Intended purpose...26 Invasive medical device...26 Measuring function...26 Medical device...27 Potentially hazardous...27 Reusable surgical instrument...27 Surgically invasive medical device...27 Vital physiological process...28 Wholly or mainly absorbed
6 INTRODUCTION This guidance document is one of a series produced to assist a wide-ranging audience to understand the regulatory system for medical devices in Australia that commenced on 4 October The system has been established by the Therapeutic Goods Act, 1989 and the Therapeutic Goods (Medical Devices) Regulations, Although each guidance document has been developed to provide information about particular aspect of the medical devices regulatory system in Australia, it is expected that a certain amount of cross-referencing to other documents in the series will be inevitable. PURPOSE The purpose of this document is to provide guidance on the classification rules for medical devices and relates to the following segments of the medical device legislation. Section 41BD of the Therapeutic Goods Act 1989 (the Act) Regulation 3.2 of the Therapeutic Goods (Medical Devices) Regulations 2002 (MD Regulations). Schedule 2 of the Therapeutic Goods (Medical Devices) Regulations Dictionary Therapeutic Goods (Medical Devices) Regulations BACKGROUND The risks associated with using medical devices can range from little or no risk to patients and users to significant potential risks inherent in the type of device. The level of premarket intervention by the regulator is proportional to the level of potential risk and is established through a classification system based on that potential risk. The classification risk is determined from:! the manufacturer s intended purpose for the medical device, and! a set of classification rules. These rules will classify medical devices into one of five classes of medical devices. THE CLASSES OF MEDICAL DEVICES The five classes of medical device based upon the classification rules are:! Class I - for low risk medical devices - includes low risk devices that are sterile and /or have a measuring function! Class IIa - for low-medium risk medical devices,! - for medium high risk medical devices,! Class III - for high risk medical devices, and! AIMD - for Active Implantable Medical Devices (are treated as Class III medical devices). 6
7 THE CLASSIFICATION RULES Schedule 2 of the MD Regulations describes the classification rules that apply to:! Non-invasive medical devices (Rule 2): General (Rule 2.1), Device to channel or store blood (Rule 2.2), Device to modify biological or chemical composition of blood (Rule 2.3), and Device to contact with injured skin (Rule 2.4).! Invasive medical devices (Rule 3): Device used to penetrate body orifices (Rule 3.1), Surgically invasive device for transient use (Rule 3.2), Surgically invasive device for short-term use (Rule 3.3), and Surgically invasive device for long-term use and implantable devices (Rule 3.4).! Active medical devices (Rule 4): General (Rule 4.1), Active device for therapy (Rule 4.2), Active device for diagnosis (Rule 4.3), and Active devices to administer or remove medicines from a patient s body (Rule 4.4).! Special rules for medical devices: Device incorporating a medicine (Rule 5.1), Device for contraception or prevention of sexually transmitted disease (Rule 5.2), Device for disinfecting, cleaning another medical device (Rule 5.3), Non-active device to record X-ray diagnostic images (Rule 5.4), Device containing animal tissues, cells or other substances, that will need to be rendered non-viable to mitigate any infectivity risk or microbial or recombinant tissues, cells or other substances (Rule 5.5), Device that is a blood bag (Rue 5.6), Active implantable device (Rule 5.7), Device for export only (Rule 5.8), and Device that is a mammary implant (Rule 5.9). CLASSIFYING A MEDICAL DEVICE DECIDING WHAT CLASSIFICATION RULES APPLY The following decisions need to be made before applying the classification rules:! Is the product a medical device? Check the definition of a medical device on page 27,! Do the special classification rules (Schedule 2, Part 5 of the MD Regulations) apply? See Flowchart 5- Special rules on page 21,! Is the device intended to be non-invasive? See Flowchart 2 on page 10, (NB- If the device is sterile or has a measuring function conformity evidence will be needed),! Is the device intended to be invasive through a body orifice? See Flowchart 3 on page 13,! Is the device intended to be surgically invasive? See Flowchart 3A on page 14, or! Is the device an active medical device? See Flowchart 4 on page 18. 7
8 IMPTANT POINTS TO REMEMBER! The manufacturer is responsible for determining the class of a device.! The classification rules are based on the manufacturer s intended purpose, taking into account the mechanism of action, and in some cases, more than one rule can apply. If this happens, the higher classification applies with the exception of medical devices for export only (Rule 5.8) which are classified as Class I.! All the relevant classification rules must be considered to determine the class of the medical device.! The classification must be consistent with the information accompanying the medical device including the label, directions for use, brochures and operating manuals.! If the intended purpose is not clear in the information, the TGA will assume an intended purpose consistent with the purpose generally accepted in clinical practice.! If the device is to be used in combination with another medical device, the classification rules must be applied separately to each device.! Accessories are classified separate to the medical device they are used with.! The duration for use must be specified for all invasive medical devices.! If the device is not for use in a specific part of the body, its classification is based on the most critical specified use.! Software intended to drive or influence the use of a medical device falls under the same classification as the medical device.! Based on the intended purpose, software may be a medical device in its own right. 8
9 FLOWCHART 1- DECIDING WHAT CLASSIFICATION RULES APPLY Is this a medical device? Yes No No need to proceed further Do the special rules under Part 5, Schedule 2 apply? Yes Apply classification rule 5 No Is this a non invasive device? Yes Apply classification rule 2 No Is this an invasive device? No Yes Apply classification rule 3 Is this an active device not covered by Rule 2, 3 or 5? Yes Apply classification rule 4 9
10 FLOWCHART 2 CLASSIFICATION RULES - NON INVASIVE DEVICES These rules are described in Schedule 2, Part 2 of the Medical Devices Regulations Rule 2.1 General Rule Either do not touch patient or contact only intact skin. Rule 2.2 Channelling or storing for eventual administration Rule 2.3 Modify biological or chemical composition of blood, body liquids intended for transfusion Rule 2.4 In contact with injured skin (mechanical barrier absorb exudates) Class I Class IIa Only filtration, centrifugation or exchange of gas or heat Class IIa Class I Intended for wounds that breach dermis and heal only by secondary intent Intended to manage micro environment of wound + others Class IIa 10
11 RULE 2 NON INVASIVE MEDICAL DEVICES Rule 2.1 Non-invasive medical devices general This rule applies to all medical devices that are not covered by a specific rule, devices that contact intact skin and devices that do not touch the patient. Rule 2.1 A non-invasive device is Class I, unless the device is classified at a higher level under another rule in Schedule 2 of the MD Regulations. Devices used to collect body liquid where a return flow is unlikely eg urine collection bottles, ostomy pouches, wound drainage collection bottles and incontinence pads. Devices used to immobilise body parts and/or to apply force or compression eg non-sterile dressings, plaster bandages, cervical collars and gravity traction devices or compression hosiery. Rule 2. 2 Non-invasive devices intended to channel or store blood, etc Devices covered under this rule may be indirectly invasive. Typically, they channel or store substances that will be eventually delivered into the body. Rule 2.2 A non-invasive device used to channel or store blood or body liquids that are to be infused, administered or introduced into a patient. Class IIa A non-invasive device to store an organ, parts of an organ or body tissue that is to be later introduced into a patient. Class IIa A non-invasive device to channel or store a liquid or gas that is to be infused, administered or introduced into a patient and may be connected to an active medical device classified as Class IIa or higher. Class IIa. Devices intended to be used to channel active drug delivery systems. Eg intravenous tubing, gastrostomy tubing, anaesthesia breathing circuits and pressure indicator and syringes for infusion pumps. Devices to channel blood. Eg blood transfusion sets. Devices to temporarily store and transport of organs for transplant or for long term storage of biological substances and tissues such as corneas, sperm and human embryos. Oxygen tubing & masks, anaesthetic tubing & breathing circuits and syringes & tubing for infusion pumps. 11
12 Rule 2.3 Non-invasive devices intended to modify the biological or chemical composition of blood, etc Devices in this category must be considered separately from those in Rule 2.1, as they are indirectly invasive. The devices treat or modify substances that will be delivered into the body. Rule 2.3 A non-invasive device to modify the biological or chemical composition of blood, other body liquids, or other liquids to be infused in the patient.. A non-invasive device to be used in treatment consisting of filtration, centrifugation or exchanges of gas or heat. Class IIa. Devices intended to remove undesirable substances out of the blood by exchange of solutes such as hemodyalizers. Auto transfusion systems. Devices used to separate cells such as gradient medium for sperm. Particulate filtration of blood in an extracorporeal circulation system, centrifugation of blood for transfusion or autotransfusion, removal of carbon dioxide from the blood and/or adding oxygen, and warming or cooling blood in the extracorporeal circulatory system. Rule 2.4 Non-invasive devices intended to have contact with injured skin This rule covers wound dressings without consideration of the wound depth. The technology associated with these devices is well understood and they do not result in any great hazard to the patient. Rule 2.4 A non-invasive device to be used in contact with injured skin (including a device the principal intention of which is to manage the microenvironment of a wound). Class IIa A non-invasive device to be used as a mechanical barrier or for compression; or for absorption of exudates. Class I. A non-invasive device to be used for wounds that have breached the dermis and the wounds can only heal by secondary intent.. Assists healing by controlling the level of moisture and regulating the humidity, temperature, levels of oxygen, other gases and ph values of the wound environment, or by influencing the process by other physical means. Eg adhesives for topical use, polymer film dressings, hydrogel dressings and non-medicated impregnated gauze dressings. Absorbent pads, island dressings, cotton wool, wound strips and gauze dressings to act as a barrier or absorb exudates from the wound. (NB If the device is sterile conformity evidence is required) Intended for severe wounds that have extensively breached the dermis, and healing is by secondary intent (by granulation from the base of the wound). Eg dressings for chronic extensive ulcerated wounds, severe burn, severe decubitus wounds, or dressings providing a temporary skin substitute. 12
13 FLOWCHART 3 CLASSIFICATION RULES - INVASIVE DEVICES These rules are described in Schedule 2, Part 3 of the Medical Devices Regulations Rule 3.1 Invasive in body orifice or stoma (not surgical) Rule 3.2 Surgically invasive transient use Transient use Class I Short term use Class IIa If only in oral cavity, ear canal or nasal cavity Long term use If only in oral cavity, ear canal or nasal cavity Connected to an active medical device of Class IIa or higher Class IIa Diagnose/control defect of heart/central circulatory system Class III Class IIa Supply energy/ionising radiation Class I Class IIa Reusable surgical instrument Biological effect mainly absorbed Class I System to administer medicines potentially hazardous 13
14 FLOWCHART 3A CLASSIFICATION RULES - INVASIVE DEVICES These rules are described in Schedule 2, Part 3 of the Medical Devices Regulations Rule 3.3 Surgically invasive Short term use Rule 3.4 Surgically invasive long term use and implantable devices Class IIa Supply energy/ionising radiation Undergo chemical change in the body or administer medicines (not in teeth) Specifically intended to monitor/correct defect of heart or central circulatory system by direct contact Class III For use in direct contact with central nervous system To be placed in teeth Class IIa Used in direct contact with heart or central circulatory/nervous system Biological effect mainly absorbed Class III Undergo chemical change in body - or administer medicines (NOT in teeth) Class III Class III Class III Class III Biological effect mainly absorbed Class III 14
15 RULE 3 INVASIVE MEDICAL DEVICES Rule 3.1 Invasive devices intended to be used to penetrate body orifices This rule covers devices that enter the body through existing body orifices (eg ear mouth, nose, eye etc) and surgically created stomas. Devices covered by this rule tend to be for diagnostic and therapeutic use in particular specialities (ENT, ophthalmology, dentistry, proctology, urology and gynaecology). Rule 3.1 Invasive devices that are: - not connected to an active medical device, and are for transient use. Class I; Handheld dental mirrors, dental impression materials, exam gloves and prostatic balloon dilation catheters. - for short-term use. Class IIa; - for use in the oral cavity as far as the pharynx, in an ear canal to the ear drum, or in a nasal cavity. Class I; and - for long-term use.. Invasive devices for use in the oral cavity as far as the pharynx or in an ear canal to the ear drum, or in a nasal cavity and is not liable to be absorbed by the mucous membrane. Class IIa. Invasive device to be connected to an active medical device that is classified as Class IIa or higher. Class IIa. Contact lenses, urinary catheters, tracheal tubes, stents, vaginal pessaries and perineal reduction devices. Dressing for nose bleeds, dentures removable by the patient. Long term urinary catheters, artificial eye and urethral stents. Orthodontic wire, fixed dental prostheses, fissures sealants. Tracheostomy tubes connected to a ventilator, powered nasal irrigators, nasopharyngeal airways, heat and moisture exchangers and suction catheters or tubes for stomach drainage. Rule 3.2 Surgically invasive devices intended for transient use This rule covers devices that are used to create a conduit through the skin (needles, cannulae, etc), surgical instruments (scalpels, saws, etc.) and various types of catheters, suckers, etc. Rule 3.2 Surgically invasive device for transient use. Suture needles, hypodermic needles and syringes, Class IIa. Surgically invasive device for transient use to diagnose, monitor, control or correct a defect of the heart, or central circulatory system through direct contact. Class III. A reusable surgical instrument. Class I. suckers, surgical swabs and surgical gloves. Cardiovascular catheters (eg. angioplasty balloon catheters), including related guidewires and coronary artery probes. Scissors, artery forceps, tissue forceps, tissue clamps, excavators, osteotomes and chisels. 15
16 A surgically invasive device for transient use to supply ionising radiation. A surgically invasive device for transient use to have a biological effect. A surgically invasive device for transient use to be wholly, or mostly, absorbed by the body. A surgically invasive device for transient use to administer medicine via a delivery system, and the administration is potentially hazardous to the patient.. Catheters containing or incorporating radioactive radioisotopes where the isotope is not intended to be released into the body. Fibrin based tissue glues. Resorbable haemostat materials. Devices for repeated self-application where the dose and the medicine are critical, eg. personal insulin injectors (commonly referred to as pens). Rule 3.3 Surgically invasive devices intended for short-term use This rule covers devices used in the context of surgery or post-operative care (eg clamps and drains), infusion devices (cannulae and needles) and catheters of various types. Rule 3.3 Surgically invasive device for short term use. Class IIa. A surgically invasive device for short term use to supply ionising radiation, or A surgically invasive device for short term use device to undergo a chemical change in a patient s body (except a device intended to be placed in the teeth), or A surgically invasive device for short term use to administer medicine.. A surgically invasive device for short term use to be specifically used to diagnose, monitor, control or correct a defect of the heart, or central circulatory system, through direct contact with these parts of the body; A surgically invasive device for short term use to be used in direct contact with the central nervous system; or A surgically invasive device for short term use to have biological effect; or be wholly, or mostly, absorbed by a patient s body. Class III A surgically invasive device for short term use that is intended by the manufacturer to be placed in the teeth and to undergo a Clamps, infusion cannulae, skin closure devices or temporary filling materials, some surgical retractors eg chest retractors for cardiac surgery. Bradytherapy devices. Biological tissue adhesives. An intravenous cannula. Cardiovascular catheters, cardiac output probes and temporary pacemaker leads. Thoracic catheters intended to drain the heart, including the pericardium and a carotid artery shunt. Neurological catheters, cortical electrodes and connonoid paddles. Absorbable sutures and bone anchors. Dental Adhesives used for root canal therapy. 16
17 chemical change in the body is Class IIa. For this clause a medical device to be placed in the teeth includes a device that is intended to penetrate a tooth but does not enter the gum or bone beyond the tooth. Rule 3.4 Surgically invasive devices for long-term use and implantable devices Devices covered by this rule include implants used in orthopaedic, dental, ophthalmic and cardiovascular fields. In addition, soft tissue implants used in plastic surgery are covered by this rule. Rule 3.4 A surgically invasive device for long-term use and implantable devices. A surgically invasive device for long-term use to be placed in the teeth. Class IIa. A surgically invasive device for long-term use to be used in direct contact with the heart, the central circulatory system or the central nervous system. Implantable joint replacements, shunts, stents, nails, plates and screws, intra-ocular lenses, infusion ports, peripheral vascular grafts, non-absorbable sutures, bone cements and maxillo-facial implants. Bridges, crowns, dental filling materials and pins, dental alloys, ceramics and polymers. Prosthetic heart valves, aneurysm clips, vascular prostheses, spinal stents, vascular stents, CNS electrodes and cardiovascular sutures. A surgically invasive device for long-term use to have a biological effect. A surgically invasive device for long-term use to be wholly, or mostly, absorbed by a patient s body. A surgically invasive device for long-term use to undergo a chemical change in the patient s body (except a device that is to be placed in the teeth); or A surgically invasive device for long-term use to administer medicine. Class III A surgically invasive device for long-term use that is intended by the manufacturer to be placed in the teeth and to undergo a chemical change in -the body is Class IIa. For this clause a medical device to be placed in the teeth includes a device that is intended to penetrate a tooth but does not enter the gum or bone beyond the tooth. Permanent vena cava filters. Absorbable sutures, bioactive adhesives and implants through the attachment of surface coatings such as phosphorylcholine. Rechargeable non-active drug delivery systems. 17
18 FLOWCHART 4 CLASSIFICATION RULES - ACTIVE DEVICES These rules are described in Schedule 2, Part 4 of the Medical Devices Regulations Rule 4.1 All active medical devices not covered by Rules 2, 3 or 5 Rule 4.2 Active medical device for therapy to administer or exchange energy Rule 4.3 Active device for diagnosis. May supply energy, for imaging purpose monitor vital physiological process Rule 4.4 Active device to administer /remove medicines & substances to or from the body Class I Class IIa Class IIa Class IIa Administer or exchange energy in a potentially hazardous way Intended to control or monitor or influence directly a active medical device When used to monitor vital processes where variations could result in immediate danger If administration /removal is hazardous to patient 18
19 RULE 4 ACTIVE MEDICAL DEVICES Rule 4.1 Active medical devices general This rule applies to active medical devices that are not covered by a specific rule. Rule 4.1 An active device is Class I, unless Examination lights, surgical microscopes, diagnostic the device is classified at a higher devices for thermography, active devices for recording, level under another rule in Schedule processing or viewing of diagnostic images and dental 2 of the MD Regulations. curing lights. Rule 4.2 Active medical devices for therapy This rule covers devices that are electrical equipment used in surgery, devices used in specialised treatments and stimulation devices. Rule 4.2 An active device for therapy to administer energy to a patient, or exchange energy to or from a patient. Class IIa. An active device to administer or exchange energy in a potentially hazardous way, having regard to the nature, density and site of application of the energy.. An active device to control or monitor, or directly influence the performance of an active medical device for therapy of the kind mentioned in the 2nd box above.. Electrical, magnetic and electromagnetic energy- muscle stimulators, external bone growth stimulators, TENS devices and electrical acupuncture. Thermal energy- warming blankets except for unconscious patients, cryosurgery equipment and heat exchangers. Mechanical energy- powered dermatomes, drills and dental hand pieces. Light- phototherapy for skin treatment and for neonatal care Sound- hearing aids. Kinetic Energy- lung ventilators. Thermal Energy- infant incubators, warming blankets for unconscious patients, blood warmers, heat exchangers used in intensive care. Electrical Energy- high frequency electrosurgical generators, electrocautery, external defibrillators, electroconvulsive therapy equipment. Coherent Light- surgical lasers. Ultrasound- lithotriptors, physiotherapy ultrasound devices. Ionising Radiation - radioactive sources for after-loading therapy, therapeutic cyclotrons, linear accelerators, therapeutic X-ray sources. External feedback systems for active therapeutic devices, afterloading control devices. 19
20 Rule 4.3 Active medical devices for diagnosis This rules covers devices that are used in ultrasound diagnosis and capture of physiological signals and devices used in therapeutic and diagnostic radiology. Rule 4.3 A device to supply energy that will be absorbed by a patient s body (except a devices that illuminate the patient s body in the visible spectrum). A device to be used to image in vivo distribution of radiopharmaceuticals in patients. Devices to allow direct diagnosis or monitoring of vital physiological processes of a patient (except a device mentioned in the 1 st paragraph in the box below. Class IIa. A device to monitor vital physiological parameters of a patient, and the nature of variations monitored could result in immediate danger to the patient. A device to emit ionising radiation and to be used for diagnostic or therapeutic interventional radiology, or A device to control, monitor, or directly influence the performance of a device in the previous paragraph. Magnetic resonance equipment, pulp testers, evoked response stimulators, diagnostic ultrasound. Gamma cameras, positron emission tomography and single photon emission computer tomography. Electrocardiographs, electroencephalographs, cardioscopes with or without pacing pulse indicators and electronic thermometers. Intensive care monitoring systems, biological sensors, blood gas analysers used in open-heart surgery, cardioscopes and apnea monitors, including those in home care. Diagnostic X-ray sources. Rule 4.4 Active medical devices intended to administer or remove medicines, etc from a patient s body This rule covers drug delivery systems and anaesthesia equipment. Rule 4.4 An active device to administer or remove medicine, body liquids or other substances. Class IIa An active device to administer or remove medicine, body liquids or other substances in a way that is potentially hazardous to the patient, having regard to the substances, the part of the body concerned, and the characteristics of the device.. Suction equipment, feeding pumps Jet injectors for vaccination. Infusion pumps, ventilators, anaesthesia machines, anaesthetic vaporisers, dialysis equipment, blood pumps for heart-lung machines, hyperbaric chambers, pressure regulators for medical gases, medical gas mixers, moisture exchangers in breathing circuits. Nebulisers where the failure to deliver the appropriate dosage form could be hazardous. 20
21 FLOWCHART 5 CLASSIFICATION RULES SPECIAL RULES These rules are described in Schedule 2, Part 5 of the Medical Devices Regulations Rule 5.1 Device incorporating a medicine and has an ancillary action on the body Class III Rule 5.3 Specific for disinfecting, cleaning, rinsing or hydrating contact lenses Rule 5.5 Devices containing non viable human or animal tissue (not for intact skin) Class III Rule 5.7 AIMD Class AIMD Rule 5.2 Device for contraception or preventing sexually transmitted diseases Device for disinfecting other devices other than by physical action Device containing cells or substances of bacterial or recombinant origin Rule 5.8 Export only Class I Class III If implantable or long term invasive Rule 5.4 Non active devices to record X-ray diagnostic images Rule 5.6 Blood bags Rule 5.9 Mammary implants Class III Class IIa Class III 21
22 RULE 5 SPECIAL RULES Rule 5.1 Devices incorporating a medicine This rule covers medical devices that incorporate a medicinal substance including stable derivatives of human blood and blood plasma that assists the function of the device. Clause 5.1 A device incorporating a substance that if used separately would be a medicine and has an ancillary action on the body. Class III Antibiotic bone cements, condoms with spermicide, heparin coated catheters. Dressings incorporating an antimicrobial agent where the purpose of such an agent is to provide ancillary action on the wound. Rule 5.2 Devices for contraception or prevention of sexually transmitted diseases Some devices covered by this rule may perform both functions, eg. condoms. Rule 5.2 A device for contraception or the Condoms, contraceptive diaphragms. prevention of sexually transmitted diseases.. An implantable or invasive device for Contraceptive intrauterine devices (IUDs) and long-term use. Class III. surgically implanted contraceptive devices. Rule 5.3 Devices intended for disinfecting, cleaning, rinsing etc This rule covers various contact lens fluids and substances or equipment to disinfect another medical device. It does not cover devices that clean by a physical action only. Rule 5.3 A device specifically for disinfecting, cleaning, Contact lens solutions, comfort solutions. rinsing or hydrating contact lenses. A device specifically for disinfecting another medical device. Disinfectants for haemodialysis devices or endoscopes, sterilisers to sterilise medical devices and washer disinfectors. Rule 5.4 Non-active devices intended to record X-ray diagnostic images A non-active medical device to record X-ray diagnostic images such as X-ray films, photostimulable phosphor plates are Class IIa. Rule 5.5 Devices containing non-viable animal tissues or derivatives This rule covers devices that contain or are made of animal tissues that have been rendered nonviable or derivatives from such tissues also being non-viable. Rendering non-viable refers primarily to the inactivation of any infective agent that has been identified in a risk analysis as being potentially transmissible. 22
23 Rule 5.5 Devices containing animal tissues or derivatives that have been rendered nonviable and not intended to come into contact with intact skin only. Class III. A devices that contains tissues, cells or substances of bacterial or recombinant origin or a combination of tissues, cells or substances of animal, bacterial or recombinant origin. Wherever these devices contain material of bacterial or recombinant origin, they are Class III, even if they only come into contact with intact skin. Biological heart valves, porcine xenograft dressings, catgut sutures, implants and dressings made from collagen. Intra-ocular fluids, meniscul joint fluid replacement, anti-adhesion barriers, tissue fillers based on hyaluronic acid derived from bacterial fermentation processes. Rule 5.6 Devices that are blood bags Rule 5.6 A device that is a blood bag. If the blood bags have a function greater than storing purposes and include systems for preservation other than anti-coagulants then other rules (eg. rule 5.1) may apply. Blood bags (including those containing or coated with an anticoagulant). Rule 5.7 Active implantable medical devices An active implantable medical device is classified as Class AIMD. Rule 5.7 An implantable accessory to an active Electrode leads associated with pacemakers, implantable device. defibrillators and nerve stimulators. An active device to control, monitor, or directly influence the performance of an active implantable device. Class III Clinician s programming device for pacemakers, patient control device for nerve stimulation devices. Rule 5.8 Medical devices for export only A device that is intended by the manufacturer for export only is classified as Class I. Rule 5.9 Devices that are mammary implants A device that is a mammary implant is classified as Class III. 23
24 APPENDIX A - DEFINITIONS Accessory (reference subsection 3(1) of the Act) An article that its manufacturer specifically intended to be used together with a medical device to enable the device to be used as the manufacturer of the device intended. Active implantable medical device (reference- Dictionary in the MD Regulations) An active medical device (other than an implantable medical device) that is intended by the manufacturer: (a) either: (i) to be, by surgical or medical intervention, introduced wholly, or partially, into the body of a human being, or (ii) to be, by medical intervention, introduced into a natural orifice in the body of a human being, and (b) to remain in place after the procedure. : Pacemaker, nerve stimulator, artificial cochlear and implantable bradytherapy seeds. Active medical device (reference- Dictionary in the MD Regulations) A medical device that is intended by the manufacturer: (a) To depend for its operation on a source of electrical energy or other source of energy (other than a source of energy generated directly by a human being or gravity), and (b) To act by converting this energy; but does not include a medical device that is intended by the manufacturer to transmit energy, a substance, or any other element, between an active medical device and a human being without any significant change in the energy, substance, or other element being transmitted. : surgical diathermy, surgical lasers, air driven surgical drills and saws, powered traction systems and examination lights. Active medical device for diagnosis (reference- Dictionary in the MD Regulations) An active medical device that is intended by the manufacturer to be used on a human being, either alone or in combination with another medical device, to supply information for the purpose of detecting, diagnosing monitoring or treating physiological conditions, states or health, illness or congenital deformities. Example: Ultra sound device Active medical device for therapy (reference- Dictionary in the MD Regulations) An active medical device that is intended by the manufacturer to be used on a human being, either alone or in combination with another medical device, to support, modify, replace or restore biological functions or structures for the purpose of treating or alleviating an illness, injury or handicap. : therapeutic laser, infant incubator, lung ventilator, deep heat therapy lamp and an electrical muscle stimulator. 24
25 Ancilliary action (reference European Commissions Guidelines relating to Medical Device Directives MEDDEV 2.4/1 rev 8 Part 2: Guidelines for the Classification of Medical Devices - July 2001 (MEDDEV 2.4/1) The action of the medicine is subordinate/subsidiary to that of the device. Biological effect (reference MEDDEV 2.4/1) A material is considered to have a biological effect if it actively and intentionally induces, alters or prevents a response from the tissues that is mediated by specific reactions at a molecular level. Such a device may be described as bioactive. Body orifice (reference- Dictionary in the MD Regulations) A natural opening or a permanent artificial opening, in a human being s body, and includes the external surface of the eyeball. : Includes stomas such as a colostomy, gastrostomy and tracheostomy. Central circulatory system (reference- Dictionary in the MD Regulations) The system in a human being comprising the following vessels:! Arteriae pulmonales (pulmonary artery)! Aorta ascendens (ascending aorta)! Arteriae coronariae (coronary artery)! Arteria carotis communis (common carotid artery)! Arteria carotis externa (external carotid artery)! Arteria carotis interna (internal carotid artery)! Arteria cerebrates (cerebella arteries)! Truncus brachicephalicits (brachiocephalic trunk)! Venae cordis (cardiac veins)! Venae pulmonales (pulmonary vein)! Vena cava superior (superior vena cava)! Vena cava inferior (inferior vena cava)! Arcus aorta (aortic arch)! Thoracica aorta (thoracic aorta)! Abdominalis aorta (abdominal aorta)! Iliaca communis (common iliac arteries and veins). Note: The thoracic aorta, abdominal aorta and the common iliac are not considered part of the definition of the central circulatory system in the EU Medical Devices Directive - this will result in differences between Australia and the EU in classifying devices used for these vessels. Central nervous system (reference- Dictionary in the MD Regulations) The system in a human being comprising the brain, meninges and spinal cord. Chemical change (reference MEDDEV 2.4/1) Chemical change does not apply to products such as bone cements where the chemical change takes place during the placement and does not continue in long term. 25
26 Direct diagnosis (reference MEDDEV 2.4/1) Continuous surveillance by direct measurement. Duration (reference Classification Rules Part 1 (1.1) Schedule 2 of the MD Regulations) Transient - intended to be used continuously for less than 60 minutes. Short term - intended to be used continuously for at least 60 minutes but not more than 30 days. Long term - intended for continuous use for more than 30 days. Implantable medical device (reference- Dictionary in the MD Regulations) A medical device (other than an active implantable medical device) that is intended by the manufacturer: (a) to be, by surgical intervention, wholly introduced into the body of a human being, and to remain in place after the procedure, or (b) to replace, by surgical intervention, an epithelial surface,or the surface of an eye, of a human being, and to remain in place after the procedure, or (c) to be, by surgical intervention, partially introduced into the body of a human being, and to remain in place for at least 30 days after the procedure. : cardiovascular stent, bone plates and screws, intra-ocular lees, internal/external fixation systems, resorbable anti-adhesion barriers and heart valve prosthesis. Integral (reference MEDDEV 2.4/1) A device and the ancillary medicine form a single physical unit. Intended purpose (reference- Dictionary in the MD Regulations) The purpose for which the manufacturer of the device intends it to be used, as stated in: (a) the information provided with the device, or (b) the instructions for use of the device, or (c) any advertising material applying to the device, and (d) the mechanism of action of the device. Note: The mode of action of a device must be consistent with the intended purpose of the device. Where not stated explicitly it will be assumed that the device will be used in accordance with reasonably foreseeable Australian medical practice. Invasive medical device (reference- Dictionary in the MD Regulations) A medical device that is intended by the manufacturer to be used, in whole or in part to penetrate the body of a human being through a body orifice or through the surface of the body. : wound drains, colonoscope, gastroscope, artificial larynx, urinary catheter, hypodermic needle, naso-gastric tube, and a feeding tube. Measuring function (reference- Regulation 1.4 of the MD Regulations) A measuring function is one intended by the manufacturer to measure quantitatively, a physiological or anatomical parameter, or a quantity, or a qualifiable characteristic, of energy or substances delivered to or removed from the body. Is displayed in Australian units of measurement or other units of measurement approved by the Secretary, of the particular device, or 26
27 is compared to at least one point of reference indicated in such units, and must be accurate to enable the device to achieve its intended purpose. Medical device (reference- Section 41BD of the Act) A medical device is: (a) any instrument, apparatus, appliance, material or other article (whether used alone or in combination, and including the software necessary for its proper application) intended, by the person under whose name it is or is to be supplied, to be used for human beings for the purpose of one or more of the following: (i) diagnosis, prevention, monitoring, treatment or alleviation of disease; (ii) diagnosis, monitoring, treatment, alleviation of or compensation for an injury or handicap; (iii) investigation, replacement or modification of the anatomy or of a physiological process; (iv) control of conception; and that does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but that may be assisted in its function by such means; or (c) an accessory to such an instrument, apparatus, appliance, material or other article. Potentially hazardous (reference MEDDEV 2.4/1) The concept of potentially hazardous is dependent on the type of technology involved and the intended application of the device to the patient and not on the measures adopted by the manufacturer in view of good design management. A device is potentially hazardous if the basic risk of the device before deliberate measures have been taken to minimise risk by inherent design or protection mechanisms, is considered intolerable. For example all devices intended to emit ionizing radiation, all lung ventilators and lithotriptors should be in a higher risk class. Reusable surgical instrument (reference- Dictionary in the MD Regulations) A medical device that is intended by the manufacturer: (a) to be used surgically, without being connected to an active medical device, for cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or any other similar procedure, and (b) to be used after the appropriate procedures specified by the manufacturer in the instructions for use have been carried out. Surgically invasive medical device (reference- Dictionary in the MD Regulations) (a) An invasive medical device that is intended by the manufacturer to be used with the aid, or in the context of, a surgical operation, and (b) A medical device that is intended by the manufacturer to be used to penetrate the body of a human being in any way other than through a body orifice. : scalpel blades, biopsy devices, surgical instruments, skin stapling devices, staples, clips and clamps for blood vessels tissues or organs, coronary angioplasty balloon catheters. 27
28 Vital physiological process (reference MEDDEV 2.4/1) Includes respiration, heart rate, cerebral functions, blood gases, blood pressure and body temperature. Medical devices intended to be used for continuous surveillance of vital physiological processes in anaesthesia, intensive care or emergency care are in a higher risk class, whilst medical devices intended to be used to obtain readings of vital physiological signals in routine check ups and in self-monitoring are in a lower risk class. Wholly or mainly absorbed Refers to the degradation of a material within the body and the metabolic elimination of the resulting degradation products from the body. 28
Classification Rules for Medical Devices a
Classification Rules for Medical Devices a The actual classification of each device depends on the precise claims made by the manufacturer and on its intended use. While the provision of examples in the
Classification Rules for Medical Devices
Technical Reference: TR-003 Date of Issue: 4 Jan. 2010 Classification Rules for Medical Devices Technical Reference: TR-003 Department of Health The G overnment of the Hong K ong S pec ial Adminis trative
CANADIAN RULE BASED CLASSIFICATION SYSTEM (non-ivdd) Life Sciences British Columbia NRC-Industry Research Assistance Program
CANADIAN RULE BASED CLASSIFICATION SYSTEM (non-ivdd) Life Sciences British Columbia NRC-Industry Research Assistance Program Health Canada Regulations on Medical Devices Vancouver, B.C. October 29, 2007
MEDICAL DEVICES: Guidance document - Classification of medical devices
EUROPEAN COMMISSION DG HEALTH AND CONSUMER Directorate B, Unit B2 Cosmetics and medical devices MEDICAL DEVICES: Guidance document - Classification of medical devices GUIDELINES RELATING TO THE APPLICATION
Med-Info. Council Directive 93/42/EEC on Medical Devices. TÜV SÜD Product Service GmbH
Med-Info International expert information for the Medical Device industry Council Directive 93/42/E on Medical Devices Practice-oriented summary of the most important aspects and requirements contained
Medical Devices: CE Marking Step-by Step
Medical Devices: CE Marking Step-by Step Sylvia Mohr Standards Specialist U.S. Mission to the European Union 1 Introduction CE marking in 8 Steps Overview intended purpose - technical documentation classification
CE-Marked Product List (MDD)
CE-Marked Product List (MDD) On March 21st the revised Directive 93/42/EEC concerning medical devices which includes the amendment M5 Directive 2007/47/EC of the European Parliament and of the Council
EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below.
EMKI COMPETENCE in MD EMKI has competence in conformity assessment - on the basis of EU directives 93 /42/EEC - for companies dealing with product ranges listed below. Annex II.: EC DECLARATION OF CONFORMITY
Regulatory Strategy for Drug-Device Combinations By John Worroll
THE MEDICAL DEVICE CONSULTING, AUDITING & TRAINING EXPERTS Regulatory Strategy for Drug-Device Combinations By John Worroll Introduction Drug-device combinations have always been a challenge to the regulatory
X-Plain Subclavian Inserted Central Catheter (SICC Line) Reference Summary
X-Plain Subclavian Inserted Central Catheter (SICC Line) Reference Summary Introduction A Subclavian Inserted Central Catheter, or subclavian line, is a long thin hollow tube inserted in a vein under the
DENTAL ASSISTING CATEGORIES
DENTAL ASSISTING CATEGORIES EFFECTIVE JANUARY 1, 2010 Starting January 1, 2010, the dental assisting scope of practice will include new duties and two new specialty permits in orthodontics and dental sedation
CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow
CE Marking Changing & Evolving Requirements September 2013 2013 BSI Healthcare Roadshow Existing Directives 2 Directive 93/42/EEC 1. Check Scope of Medical Device Directive (Article 1) 2. Determine Device
Australian Medical Devices Guidance Document Number XX. Custom Made Medical Devices
Australian Medical Devices Guidance Document Number XX Custom Made Medical Devices December 2008 Custom made medical devices V1.1 Page 1 of 29 Disclaimer This document is provided for guidance only. Please
Procedure Code A4215 22 Insulin pen needles A4230 Infusion set for external insulin pump, non needle cannula type A4231 Infusion set for external
A4215 22 Insulin pen needles A4230 Infusion set for external insulin pump, non needle cannula type A4231 Infusion set for external insulin pump, needle type A4232 Syringe with needle for external insulin
Healthcare services requiring prior authorisation
Annex 2 Healthcare requiring prior authorisation This list does not include organ transplants and does also not apply to long-term care and the primary purpose of which is to support people in need of
ORGAN SYSTEMS OF THE BODY
ORGAN SYSTEMS OF THE BODY DEFINITIONS AND CONCEPTS A. Organ a structure made up of two or more kinds of tissues organized in such a way that they can together perform a more complex function that can any
Why is prematurity a concern?
Prematurity What is prematurity? A baby born before 37 weeks of pregnancy is considered premature. Approximately 12% of all babies are born prematurely. Terms that refer to premature babies are preterm
NAPCS Product List for NAICS 62121 (US, Mex): Offices of Dentists
NAPCS List for NAICS 62121 (US, Mex): Offices of Dentists 62121 1 Services of dentists Providing dental medical attention by means of consultations, preventive services, and surgical and non-surgical interventions.
UW MEDICINE PATIENT EDUCATION. Aortic Stenosis. What is heart valve disease? What is aortic stenosis?
UW MEDICINE PATIENT EDUCATION Aortic Stenosis Causes, symptoms, diagnosis, and treatment This handout describes aortic stenosis, a narrowing of the aortic valve in your heart. It also explains how this
Enables MDA Medical Teams to categorize victims in mass casualty scenarios, in order to be able to triage and treat casualties
MDA Disposable ALS + BLS Medical Ambulance Equipment Prices shown in CDN. Funds Items Description Picture Mass Casualty ID tag 1000 units = $350 Enables MDA Medical Teams to categorize victims in mass
Paramedic Program Anatomy and Physiology Study Guide
Paramedic Program Anatomy and Physiology Study Guide Define the terms anatomy and physiology. List and discuss in order of increasing complexity, the body from the cell to the whole organism. Define the
Clinical Guideline for: Aseptic Technique
Clinical Guideline for: Technique Summary This guideline provides the principles of, Non Touch, and Clean Techniques to be implemented in the hospital environment. Key Points The essential elements of
Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules?
Medical Device Directive 2007/47/EC What is New? Are we moving towards Drug Rules? CEMO Congress, Paris, February 3rd, 2011 Dr. Martine Dehlinger-Kremer Vice President Global Regulatory Affairs Agenda
INFORMATION ON CUSTOMS CLEARANCES FOR TRADERS FOR CLASS A MEDICAL DEVICES EXEMPTED FROM PRODUCT REGISTRATION
INFORMATION ON CUSTOMS CLEARANCES FOR TRADERS FOR CLASS A MEDICAL DEVICES EXEMPTED FROM PRODUCT REGISTRATION The HS Codes required for Customs clearances, such as through the TradeNet System, are managed
X-Plain Trigeminal Neuralgia Reference Summary
X-Plain Trigeminal Neuralgia Reference Summary Introduction Trigeminal neuralgia is a condition that affects about 40,000 patients in the US every year. Its treatment mostly involves the usage of oral
Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service.
Joint Working Group to produce guidance on delivering an Endovascular Aneurysm Repair (EVAR) Service. Royal College of Radiologists British Society of Interventional Radiology The Vascular Society of Great
Urinary Diversion: Ileovesicostomy/Ileal Loop/Colon Loop
Urinary Diversion: Ileovesicostomy/Ileal Loop/Colon Loop Why do I need this surgery? A urinary diversion is a surgical procedure that is performed to allow urine to safely pass from the kidneys into a
CITY OF COLORADO SPRINGS TAX GUIDE MEDICAL EXEMPTIONS
CITY OF COLORADO SPRINGS TAX GUIDE MEDICAL EXEMPTIONS The City of Colorado Springs Sales and Use Tax Ordinance contains a medical exemption and definitions that, when combined, provide a variety of exemptions
Disease/Illness GUIDE TO ASBESTOS LUNG CANCER. What Is Asbestos Lung Cancer? www.simpsonmillar.co.uk Telephone 0844 858 3200
GUIDE TO ASBESTOS LUNG CANCER What Is Asbestos Lung Cancer? Like tobacco smoking, exposure to asbestos can result in the development of lung cancer. Similarly, the risk of developing asbestos induced lung
Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau [email protected]
Software Regulated as a Medical Device Sarah Chandler A/Head, Regulatory and Scientific Section Medical Devices Bureau [email protected] Therapeutic Products Directorate Director General S. Sharma
GUIDANCE DOCUMENT Guidance on the Risk-based Classification System for Non- In Vitro Diagnostic Devices (non-ivdds)
GUIDANCE DOCUMENT for Non- In Vitro Diagnostic Devices (non-ivdds) Published by authority of the Minister of Health Date Adopted 2015/04/23 Effective Date 2015/06/12 Health Products and Food Branch Our
BIOL 1108 Vertebrate Anatomy Lab
BIOL 1108 Vertebrate Anatomy Lab This lab explores major organs associated with the circulatory, excretory, and nervous systems of mammals. Circulatory System Vertebrates are among the organisms that have
Medical Surgical Nursing Skills List
Medical Surgical Nursing Skills List Read each of the required clinical skills for a Registered Nurse working on a typical acute medical-surgical unit. Write the number that corresponds to the level of
Levels of Critical Care for Adult Patients
LEVELS OF CARE 1 Levels of Critical Care for Adult Patients STANDARDS AND GUIDELINES LEVELS OF CARE 2 Intensive Care Society 2009 All rights reserved. No reproduction, copy or transmission of this publication
Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
Minimum Standards for Transport of Critically Ill Patients
COLLEGE OF INTENSIVE CARE MEDICINE OF AUSTRALIA AND NEW ZEALAND ABN 16134 292 103 AND AUSTRALIAN AND NEW ZEALAND COLLEGE OF ANAESTHETISTS ABN 82055 042 852 AND AUSTRALASIAN COLLEGE FOR EMERGENCY MEDICINE
A PATIENT S GUIDE TO CARDIAC CATHETERIZATION
A PATIENT S GUIDE TO CARDIAC CATHETERIZATION The science of medicine. The compassion to heal. This teaching booklet is designed to introduce you to cardiac catheterization. In the following pages, we will
College of Applied Medical Sciences\ Department of Nursing
2 nd Edition 2014/2015 College of Applied Medical Sciences\ Department of Nursing CAMS/ Department of Nursing/ Internship Training Logbook 2 nd Edition 1 INTERNSHIP TRAINING LOGBOOK Nurse Intern Name:
IAF Mandatory Document for the Application of ISO/IEC 17021 in Medical Device Quality Management Systems (ISO 13485)
IAF MD 9:2011. International Accreditation Forum, Inc.(IAF) IAF Mandatory Document for the Medical Device Quality Management Systems () Issue 1, Version 2 (IAF MD 9:2011) IAF MD9:2011 International Accreditation
2002 No. 618 CONSUMER PROTECTION. The Medical Devices Regulations 2002
STATUTORY INSTRUMENTS 2002 No. 618 CONSUMER PROTECTION The Medical Devices Regulations 2002 Made - - - - - 20th May 2002 Laid before Parliament 21st May 2002 Coming into force - - 13th June 2002 ARRANGEMENT
Inferior Vena Cava filter and removal
Inferior Vena Cava filter and removal What is Inferior Vena Cava Filter Placement and Removal? An inferior vena cava filter placement procedure involves an interventional radiologist (a specialist doctor)
AUSTRALIAN REGULATORY GUIDELINES FOR MEDICAL DEVICES
AUSTRALIAN REGULATORY GUIDELINES FOR MEDICAL DEVICES Version 1.0 April 2010 Version history Version Description of change Author Effective Date 1.0 Initial publication on TGA website ODBT 28 April 2010
GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES
EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.14/1 rev. 1 January
2015 SUMMER STUDENT NURSE EXTERNSHIP PROGRAM SKILLS CHECK LIST
2015 SUMMER STUDENT NURSE EXTERNSHIP PROGRAM SKILLS CHECK LIST STUDENT NURSE EXTERNNAME SCHOOL OF NURSING STUDENT AGREEMENT: I request the Clinical Skills Check list be released to (hospital/agency). I
TABLE OF PERMITTED DUTIES (Rev. 1/1/10)
TBLE OF PERMITTE UTIES (Rev. 1/1/10) Following is a table of duties which ental ssistants (), Registered ental ssistants (R), Registered ental ssistants in Extended Functions (REF), Orthodontic ssistants
LPN / LVN SKILL CHECKLIST
LPN / LVN SKILL CHECKLIST Name: When completing this ckecklist, please indicate your level of proficiency in each area according to the scale below. Place a check mark in box which best describes your
Fields of Education. Last updated August 2011
Fields of Education Last updated August 2011 Monash University is required to report to the Department of Education, Employment and Workplace Relations (DEEWR) the number of higher degree by research (HDR)
EUROPEAN COMMISSION DG ENTERPRISE and INDUSTRY Directorate F, Unit F3 Cosmetics and medical devices. MEDDEV 2. 1/3 rev 3
EUROPEAN COMMISSION DG ENTERPRISE and INDUSTRY Directorate F, Unit F3 Cosmetics and medical devices MEDICAL DEVICES: Guidance document - Borderline products, drug-delivery products and medical devices
ADA Standards Committee on Dental Products National Standards Status of Projects. Standard # Title of Standard WG Status Activity
ANSI/ADA 1-2003 (R2013) Alloy for Dental Amalgam 1.2 AS 0 ADA 6-1987 Dental Mercury 1.2 AS 0 ANSI/ADA 15-2008 (R2013) Artificial Teeth for Dental Prostheses 2.11 AP 0 ANSI/ADA 17-1983 (R2014) Denture Base
Bachelor s degree in Nursing
Tbilisi State Medical University Faculty of Physical Medicine and Rehabilitation The first level of academic higher education Bachelor s degree in Nursing Tbilisi 2012 Name of program Nursing Name of qualification
Patient Care Technician (Bridge Program)
Patient Care Technician (Bridge Program) Introduction: At Prestige Medical Solutions we are fully vested in helping students succeed. Our vision is to be a great place to learn, where people are inspired
IMPLANT CONSENT FORM WHAT ARE DENTAL IMPLANTS?
IMPLANT CONSENT FORM WHAT ARE DENTAL IMPLANTS? Dental implants are a very successful and accepted treatment option to replace lost or missing teeth. A dental implant is essentially an artificial tooth
F.E.E.A. FONDATION EUROPEENNE D'ENSEIGNEMENT EN ANESTHESIOLOGIE FOUNDATION FOR EUROPEAN EDUCATION IN ANAESTHESIOLOGY
créée sous le Patronage de l'union Européenne Detailed plan of the program of six courses 1. RESPIRATORY 1. ESPIRATORY AND THORAX 1.1 Physics and principles of measurement 1.1.1 Physical laws 1.1.2 Vaporizers
Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.
Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery. Technique Guide Instruments and implants approved by the AO Foundation Table of Contents Introduction Scout Vessel Guard 2
Functions of Blood System. Blood Cells
Functions of Blood System Transport: to and from tissue cells Nutrients to cells: amino acids, glucose, vitamins, minerals, lipids (as lipoproteins). Oxygen: by red blood corpuscles (oxyhaemoglobin - 4
INTRAPERITONEAL HYPERTHERMIC CHEMOTHERAPY (IPHC) FOR PERITONEAL CARCINOMATOSIS AND MALIGNANT ASCITES. INFORMATION FOR PATIENTS AND FAMILY MEMBERS
INTRAPERITONEAL HYPERTHERMIC CHEMOTHERAPY (IPHC) FOR PERITONEAL CARCINOMATOSIS AND MALIGNANT ASCITES. INFORMATION FOR PATIENTS AND FAMILY MEMBERS Description of Treatment A major difficulty in treating
Name Class Date Laboratory Investigation 24A Chapter 24A: Human Skin
Name Class Date Laboratory Investigation 24A Chapter 24A: Human Skin Human Anatomy & Physiology: Integumentary System You may refer to pages 386-394 in your textbook for a general discussion of the integumentary
PERIPHERAL STEM CELL TRANSPLANT INTRODUCTION
PERIPHERAL STEM CELL TRANSPLANT INTRODUCTION This booklet was designed to help you and the important people in your life understand the treatment of high dose chemotherapy with stem cell support: a procedure
Gastrointestinal Bleeding
Gastrointestinal Bleeding Introduction Gastrointestinal bleeding is a symptom of many diseases rather than a disease itself. A number of different conditions can cause gastrointestinal bleeding. Some causes
North Bergen School District Benchmarks
Grade: 10,11, and 12 Subject: Anatomy and Physiology First Marking Period Define anatomy and physiology, and describe various subspecialties of each discipline. Describe the five basic functions of living
Provide chairside support during the extraction of teeth and minor oral surgery
About this workforce competence This workforce competence is intended for those who provide close support during the extraction of erupted teeth, the extraction of unerupted teeth or roots, and bone removal.
41 Assisting with Minor Surgery
Learning Outcomes 41.1 Define the medical assistant s role in minor surgical procedures. 41-2 CHAPTER 41 Assisting with Minor Surgery 41.2 Describe types of wounds and explain how they heal. 41.3 Describe
Prerequisites: None. Course Description: Studies principles and procedures essential to the basic nursing care of patients.
Revised: Fall 2015 22TPNE 142 22TNursing Skills II Prerequisites: None Course Description: Studies principles and procedures essential to the basic nursing care of patients. Semester Credits: 3 Lecture
Tibial Intramedullary Nailing
Tibial Intramedullary Nailing Turnberg Building Orthopaedics 0161 206 4898 All Rights Reserved 2015. Document for issue as handout. Procedure The tibia is the long shin bone in the lower leg. It is a weight
Nursing college, Second stage Microbiology Dr.Nada Khazal K. Hendi L14: Hospital acquired infection, nosocomial infection
L14: Hospital acquired infection, nosocomial infection Definition A hospital acquired infection, also called a nosocomial infection, is an infection that first appears between 48 hours and four days after
Objective of This Lecture
Component 2: The Culture of Health Care Unit 3: Health Care Settings The Places Where Care Is Delivered Lecture 5 This material was developed by Oregon Health & Science University, funded by the Department
Injury Law Center OTHER INJURIES
Injury Law Center Note: This information is provided to give you a basic understanding of the injury. It is not intended as medical advice. You should consult a qualified medical provider. OTHER INJURIES
2.3. The management in each HCF shall be responsible for ensuring good waste management practices in their premises.
1. PURPOSE Health-care activities lead to production of medical waste that may lead to adverse health effects. Most of this waste is not more dangerous than regular household waste. However, some types
Vaxcel PICCs Valved and Non-Valved. A Patient s Guide
Vaxcel PICCs Valved and Non-Valved A Patient s Guide Information about your Vaxcel PICC is available by calling the Navilyst Medical Vascular Access Information Line 800.513.6876 Vaxcel Peripherally Inserted
Medical devices adverse incidents reported to MHRA 2011 to 2013
Medical devices adverse incidents reported to MHRA 2011 to 2013 This summary of medical device adverse incident data replaces the annual reports that we have previously published. Further changes are planned
Certified Nursing Assistant Essential curriculum- Maryland Board of Nursing
Certified Nursing Assistant Essential curriculum- Maryland Board of Nursing Module I Orientation/Introduction A. Identify general information pertaining to the nursing assistant course B. List requirements
Anatomy and Physiology: Understanding the Importance of CPR
Anatomy and Physiology: Understanding the Importance of CPR Overview This document gives you more information about the body s structure (anatomy) and function (physiology). This information will help
Total knee replacement
Patient Information to be retained by patient What is a total knee replacement? In a total knee replacement the cartilage surfaces of the thigh bone (femur) and leg bone (tibia) are replaced. The cartilage
Guidelines for Specialized Health Care Procedures (Revision 2004)
Guidelines for Specialized Health Care Procedures (Revision 2004) Vickie H. Southall, MSN, RN Family, Community, and Mental Health Systems Department School of Nursing University of Virginia for the Virginia
Practical class 3 THE HEART
Practical class 3 THE HEART OBJECTIVES By the time you have completed this assignment and any necessary further reading or study you should be able to:- 1. Describe the fibrous pericardium and serous pericardium,
DENTAL ASSISTING CATEGORIES EFFECTIVE JANUARY 1, 2010
DENTAL ASSISTING CATEGORIES EFFECTIVE JANUARY 1, 2010 Starting January 1, 2010, dental assisting scope of practice includes new duties and two new add-on specialty permits in orthodontics and dental sedation,
Virginia Office of Emergency Medical Services Scope of Practice - Procedures for EMS Personnel
Specific tasks in this document shall refer to the Virginia Education Standards. AIRWAY TECHNIQUES Airway Adjuncts Airway Maneuvers Alternate Airway Devices Cricothyrotomy Obstructed Airway Clearance Intubation
Oxford Centre for Respiratory Medicine Bronchial-Artery Embolisation Information for patients
Oxford Centre for Respiratory Medicine Bronchial-Artery Embolisation Information for patients This leaflet tells you about the bronchial-artery embolisation procedure. It explains what is involved and
Ohio State Dental Board Permissible Duties of Dental Hygienists and Dental Assistants
Ohio State Dental Board Permissible Duties of Dental Hygienists and Dental Assistants Pursuant to Ohio Administrative Code Section 4715-3-01 (C) which defines the following dental personnel as: Licensed
Investigating the Human Body On-site student activities: Years 7-8 Investigating the Human Body On-site student activities Years 7 8
Investigating the Human Body On-site student activities Years 7 8 Student activity (and record) sheets have been developed with alternative themes for students to use as guides and focus material during
Are any artificial parts used in the ACE Malone surgery?
ACE Malone (Antegrade Continence Enema) What is the ACE Malone? The Antegrade Continence Enema (ACE) is a type of surgery designed for the child who has chronic bowel problems with bouts of constipation,
CONSENT FOR MRI SCAN WITH INTRAVENOUS CONTRAST
CONSENT FOR MRI SCAN WITH INTRAVENOUS CONTRAST Your Doctor has asked that an MRI scan be performed using intravenous contrast. This contrast may give the radiologist additional information that may not
Oxygen - update April 2009 OXG
PRESENTATION Oxygen (O 2 ) is a gas provided in compressed form in a cylinder. It is also available in liquid form, in a system adapted for ambulance use. It is fed via a regulator and flow meter to the
2161-1 - Page 1. Name: 1) Choose the disease that is most closely related to the given phrase. Questions 10 and 11 refer to the following:
Name: 2161-1 - Page 1 1) Choose the disease that is most closely related to the given phrase. a disease of the bone marrow characterized by uncontrolled production of white blood cells A) meningitis B)
PROPOSED AMENDMENTS NURSES (REGISTERED) AND NURSE PRACTITIONERS REGULATION. Health Professions Act
Health Professions Act NURSES (REGISTERED) AND NURSE PRACTITIONERS REGULATION PROPOSED AMENDMENTS TO B.C. REG. 284/2008 abc abc text to be removed text to be added Contents 1 Definitions 2 College name
Ureteral Stenting and Nephrostomy
Scan for mobile link. Ureteral Stenting and Nephrostomy Ureteral stenting and nephrostomy help restore urine flow through blocked ureters and return the kidney to normal function. Ureters are long, narrow
Health Science Career Field Allied Health and Nursing Pathway (JM)
Health Science Career Field Allied Health and Nursing Pathway (JM) ODE Courses Possible Sinclair Courses CTAG Courses for approved programs Health Science and Technology 1 st course in the Career Field
Academy of Health Professions Foundations of Medicine and Health Science
Length of course 1 semester Overview Academy of Health Professions Foundations of Medicine and Health Science This course is designed to introduce students to the variety of health career opportunities
Medical device registration
Medical device registration Medical devices need pre-market approval All medical devices entering the Chinese market are required to obtain pre-market approval from the Chinese State Food and Drug Administration
Dentalworkers JOB DESCRIPTIONS Great Team Members make your Office function!
Dentalworkers JOB DESCRIPTIONS Great Team Members make your Office function! Dental Assistant Registered Dental Assistant with Expanded Function: RDAEF Sterilization Assistant Dental Hygienist General
Treating your abdominal aortic aneurysm by open repair (surgery)
Patient information Abdominal aortic aneurysm open surgery Treating your abdominal aortic aneurysm by open repair (surgery) Introduction This leaflet tells you about open repair of abdominal aortic aneurysm,
Catheter Embolization and YOU
Catheter Embolization and YOU What is catheter embolization? Embolization therapy is a minimally invasive (non-surgical) treatment that occludes or blocks one or more blood vessels or vascular channels
Oxygenation. Chapter 21. Anatomy and Physiology of Breathing. Anatomy and Physiology of Breathing*
Oxygenation Chapter 21 Anatomy and Physiology of Breathing Inspiration ~ breathing in Expiration ~ breathing out Ventilation ~ Movement of air in & out of the lungs Respiration ~ exchange of O2 & carbon
Biol 111 Comparative & Human Anatomy Lab 9: Circulatory System of the Cat Spring 2014
Biol 111 Comparative & Human Anatomy Lab 9: Circulatory System of the Cat Spring 2014 Philip J. Bergmann Lab Objectives 1. To learn how blood flows through a dual circuit circulation with lungs. 2. To
TRANSPORT OF CRITICALLY ILL PATIENTS
TRANSPORT OF CRITICALLY ILL PATIENTS Introduction Inter-hospital and intra-hospital transport of critically ill patients places the patient at risk of adverse events and increased morbidity and mortality.
Small cell lung cancer
Small cell lung cancer Small cell lung cancer is a disease in which malignant (cancer) cells form in the tissues of the lung. The lungs are a pair of cone-shaped breathing organs that are found within
Reavis High School Anatomy and Physiology Curriculum Snapshot
Reavis High School Anatomy and Physiology Curriculum Snapshot Unit 1: Introduction to the Human Body 10 days As part of this unit, students will define anatomy, physiology, and pathology. They will identify
An introduction to the regulation of apps and wearables as medical devices
An introduction to the regulation of apps and wearables as medical devices Introduction Phones are increasingly equipped with a range of devices capable of being used as sensors, including gyroscopes,
Chetek-Weyerhaeuser High School
Chetek-Weyerhaeuser High School Anatomy and Physiology Units and Anatomy and Physiology A Unit 1 Introduction to Human Anatomy and Physiology (6 days) Essential Question: How do the systems of the human
