Agilent 7800/7900 ICP-MS
|
|
|
- Alexis Oliver
- 9 years ago
- Views:
Transcription
1 Agilent 7800/7900 ICP-MS Site Preparation Checklist Agilent Technologies
2 Notices No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G Edition Rev. B, June 2015 Agilent Technologies 9-1 Takakura-cho, Hachioji-shi, Tokyo Japan Agilent Technologies, Inc Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-dod Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR (c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR (June 1987) or DFAR (b)(2) (November 1995), as applicable in any technical data. Safety Notices WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. NOTE A NOTE contains helpful information on the usage - it does not denote a hazard.
3 Contents Contents Checklist and Requirement Summary 5 Customer Responsibilities 5 Important Customer Information 6 Space, Weight and Access Requirements 7 Environmental Conditions and Site Requirements 10 Power Consumption and Configuration 12 Cooling Water Requirements 14 Gas Requirements 15 Essential Requirements for Cell Gas Installation 16 Communications 17 Laboratory Supply Requirements 18 Other Requirements 19 Important Customer Web Links 20 Appendix 21 Detailed Environmental Conditions and Site Requirements 21 Detailed Power Consumption and Configuration 23 Detailed Cooling Water Requirements and Operating Supplies 28 Detailed Gas Requirements 29 Agilent 7800/7900 ICP-MS Site Preparation Checklist 3
4 Contents This page is intentionally left blank. 4 Agilent 7800/7900 ICP-MS Site Preparation Checklist
5 Checklist and Requirement Summary Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the first key step in ensuring that your instruments and software systems operate reliably over an extended lifetime. This document is an information guide AND checklist prepared for you that outlines the supplies, consumables, space and utility requirements for your equipment for your site. Customer Responsibilities Make sure your site meets the following prior specifications before the installation date. This checklist includes information on these product G8421A and G8403A. For details, see specific sections within this checklist, including: The necessary laboratory or bench space is available. Refer to "Space, Weight and Access Requirements" on page 7. The environmental conditions for the lab as well as laboratory gases and plumbing. Refer to "Environmental Conditions and Site Requirements" on page 10. The power requirements related to the product (e.g., number & location of electrical outlets). Refer to "Power Consumption and Configuration" on page 12. The cooling water requirements for the product and installation. Refer to "Cooling Water Requirements" on page 14. The gas requirements for the product and installation. Refer to "Gas Requirements" on page 15. The required network communication configuration. Refer to "Communications" on page 17. The required items and tools necessary for the product and installation. Refer to "Laboratory Supply Requirements" on page 18. For other product specific information, refer to "Other Requirements" on page 19. Agilent Technologies service providers will not install your Agilent ICP- MS system until an adequate exhaust system is present and functioning. Refer to "Environmental Conditions and Site Requirements" on page 10. If Agilent is delivering installation and familiarization services, users of the instrument should be present throughout these services; otherwise, they will miss important operational, maintenance and safety information. Agilent 7800/7900 ICP-MS Site Preparation Checklist 5
6 Important Customer Information 1 If you have questions or problems in providing anything described as part of the Customer Responsibilities above, please contact your local Agilent or partner support/service organization for assistance prior to delivery. In addition, Agilent and/or its partners reserve the right to reschedule the installation dependent upon the readiness of your laboratory. 2 Should your site not be ready for whatever reasons, please contact Agilent as soon as possible to re- arrange any services that have been purchased. 3 Other optional services such as additional training, operational qualification (OQ) and consultation for user- specific applications may also be provided at the time of installation when ordered with the system, but should be contracted separately. 6 Agilent 7800/7900 ICP-MS Site Preparation Checklist
7 Space, Weight and Access Requirements Identify the laboratory bench space before your system arrives based on the information in Table 2. Pay special attention to the total height and total weight requirements for all system components you have ordered and avoid bench space with overhanging shelves. Also pay special attention to the total weight of the modules you have ordered to ensure your laboratory bench can support this weight. Special Notes 1 The Agilent 7800/7900 ICP- MS, its data system and accessories will be delivered to your site in large and small shipping containers. Note the size of the largest shipping container below. 2 The containers will be delivered in a large truck. You must furnish a forklift, or other suitable lifting device, and make arrangements to unload the truck and transport the containers to your site. All doorways, hallways, floors and elevators must be able to accommodate the largest, heaviest container as indicated in Table 1. Do not open any of the shipping containers unless a representative of Agilent Technologies is present. 3 At least 60 cm or 2 feet on all sides of the instrument must be kept clear for maintenance access. Table 1 Shipping Container Approximate Dimensions Instrument Description Weight Height Depth Width Kg lbs cm in cm in cm in Largest shipping container Agilent 7800/7900 ICP-MS Site Preparation Checklist 7
8 Minimum 5cm clearance required to avoid magnetic interference. Auto sampler 7900 ICP-MS PC Printer Figure 1 Table 2 Layout Example (7900 ICP-MS) Product Dimensions Product Dimensions, cm (in) Weight Height Width Depth kg (lbs) 7800/7900 ICP-MS 59.5 (23.4) * (28.7) 60 (23.6)* (221) Agilent DS402 Foreline Pump for (9.5) 16.4 (6.5) 54.1 (21.3) 35 (77) Agilent MS40+ Foreline Pump for (9) 29.7 (11.7) 41.8 (16.5) 33 (72.7) Data System (PC, Monitor, Printer) Size and weight of the data system depends on the components included. Reserve at least 100 cm (39 in) of bench space for the data system. A typical data system weight is 30 kg (66 lb). Agilent Chiller (G3292A) 57.6 (22.7) 36.8 (14.5) 70.2 (27.6) 85 (188) Agilent Heat Exchanger (G1879B) 57.2 (22.5) 38.0 (15.0) 51.4 (20.25) 42 (92) Agilent Chiller for Japan (G3159A) 71.5 (28.2) 52.0 (20.5) 68.5 (27.0) 100 (220) Agilent SPS 4 Autosampler 51.0 (20.1) 60.0 (23.6) 36.3 (14.3) 15 (33.1) Integrated Autosampler (I-AS) 22.0 (8.7) 28.0 (11.0)* (11.4) 36.0 (14.2) 5 (11) Agilent ASX-500 Series Autosampler 25.0 (9.8) 61 (24)* 3 52 (20.5) 48.2 (19) 11 (24) *1 : Maximum height is 65.5 (25.8) including the duct collar on top. *2 : Maximum depth is 72.5 (28.5) including a power connector projecting at rear. *3 : Height includes probe NOTE Approximate dimensions, refer to specific product documentation. 8 Agilent 7800/7900 ICP-MS Site Preparation Checklist
9 Top View Front View 150 Side View (Right) Rear View Cell Gas Connection Gas Connection Cooling Water Connection Power Cable Spray Chamber Electrical Connection Foreline Pump Vacuum Connection 7900_0015 Figure /7900 ICP-MS Dimensions (unit: mm) Agilent 7800/7900 ICP-MS Site Preparation Checklist 9
10 Environmental Conditions and Site Requirements Operating your instrument within the recommended temperature ranges insures optimum instrument performance and lifetime. Special Notes 1 Performance can be affected by sources of heat and cold (e.g., direct sunlight, heating/cooling from air conditioning outlets, drafts and/or vibrations). The site s ambient temperature conditions must be stable for optimum performance. 2 The customer is responsible for supplying the ductwork between the instrument and the lab extraction system. WARNING User safety requires that the exhaust gases from the plasma and vacuum systems be vented externally to the building and not recirculated by the environmental control system. Health hazards include chemical toxicity of solvents, samples, and foreline pump fluid vapor. Table 3 Temperature and Humidity Specifications Parameter Specification Temperature C (59-86 F) < 2 C/h change and total change should be < 5 C Humidity 20-80% Altitude Up to 2,000m Atmosphere Non Condensing; Non Corrosive Table 4 Heat Output Product Heat Dissipated (W) Heat Absorbed (W) 7800/7900 ICP-MS 2,900 Foreline Pump 500 PC/Monitor 430 Heat Exchanger (50/60Hz) 2,000 (Maximum) 1,600 (2,000 maximum) Agilent Chiller (G3292A) 50/60Hz 3,200 (Maximum) 1,600 (2,000 maximum) Extraction Vent 1,600 (2,500 maximum) NOTE A maximum of 4,500W is removed from the ICP-MS mainframe via the cooling water and extraction duct. Approximate values. Refer to the specific product information for details. The ambient temperature around the heat exchanger must not exceed 30 C for normal ICP-MS operation. 10 Agilent 7800/7900 ICP-MS Site Preparation Checklist
11 Table 5 Exhaust Venting Requirements Product Port Diameter (mm) Exhaust Flow 7800/7900 ICP-MS * m 3 /min m/s >5 >4.7 <7 <6.6 SPS 4 Cover Kit Port for ducting with 50 mm ID m 3 /min m/s * 2 * 1 Exhaust flow must be continuous as long as the plasma is ON. Exhaust flow must be stable: maximum fluctuation of ±5% of target flow. * 2 When used with samples, standards or rinse solution(s) that release corrosive fumes in any concentration. Whenever such liquids are present, the atmospheric contents of the sampler cover must be extracted continuously at a rate of liters per minute ( m 3 /min). Corrosive samples and standards should be removed from the autosampler once analysis is completed. NOTE Flexible ducting must be used for easy removal during instrument maintenance. The back pressure of the ICP-MS is approximately 40Pa. Outside of the Building Fan Damper Flexible Duct Exhaust Port Exhaust Flow > 5 m 3 /min <7 m 3 /min 150 mm 79CEHB0185 Figure 3 Exhaust Duct Connection For more information, refer to "Detailed Environmental Conditions and Site Requirements" on page 21. Agilent 7800/7900 ICP-MS Site Preparation Checklist 11
12 Power Consumption and Configuration Special Notes 1 Power for the 7800/7900 ICP- MS is supplied in single- phase, Split phase, 3 phase star 4 wired, High- leg delta or 208 Wye configuration (see Table 12). 2 If a computer system is supplied with your instrument, be sure to account for those electrical outlets. 3 The entire ICP- MS system should share an isolated, noise- free electrical ground. This system ground should be electrically separate from the ground for the rest of the building, i.e. back to the main ground of the facility. (See Figure 4) Main Isolation Panel L1 L2 Unstable Ground G L1 L2 Other Circuits in Lab Isolated Ground Circuit Breaker G L1 L2 7800/7900 ICP-MS Main Building Ground G L2 PC/PC Monitor for MassHunter Workstation 7900_0011 Figure 4 Isolated Ground WARNING Do not use extension cords with Agilent Technologies equipment. They cannot provide enough power to the system and can be a safety hazard. If the desired location of equipment does not permit its standard power cord to reach an power receptacle, your electrician should install additional outlets. Otherwise, you should relocate the equipment closer to existing power receptacles. 12 Agilent 7800/7900 ICP-MS Site Preparation Checklist
13 Table 6 Power Requirements Instrument Description Line Voltage & Frequency (V, Hz) Current Rating (amps) Maximum Power Consumption (W) 7800/7900 ICP-MS Vac, 50/60Hz 24 PC/Monitor/Printer Vac Vac 15 8 Agilent Chiller (G3292A)* Vac, 60Hz 240Vac, 50Hz Agilent Chiller (G3159A)*2 200Vac, 50/60Hz 20 Agilent Heat Exchanger (G1879B)*3 SPS 4 Autosampler Integrated Autosampler (I-AS) Vac, 50Hz Vac, 50/60Hz V, 47/63 Hz, 1.5 A input * Vac, 50/60Hz Vac, 50/60Hz ASX-500 Series Autosampler Vac <1 40 *1: The chiller will ship with the power cable for the country of installation. *2: The chiller will ship with the power cable for Japan. *3: The heat exchanger will ship with the power cable for the country of installation. *4: This is the input rating for the AC adapter. AC adapter output/autosampler input: 24 VDC, 2.5 A Power Receptacle Power Plug 1 L1 200V L2 7800/7900 ICP-MS Circuit Breaker 30A (rating) Safety Ground Figure 5 Table 7 Main Supply - Instruments Connection without Neutral Power Plugs and Cords Country Supplied Plug Prepared Outlet U.S., all Americas (North, Central and South American countries), Japan, Korea and Taiwan NEMA L6-30P with 4.3 m power cord NEMA L6-30R Others IEC with 4.3 m power cord IEC If this is not compatible with your power receptacle, it is your responsibility to connect the ICP- MS power cord safely. This can be done via an industrial standard locking plug and socket, or directly into a switched distribution panel. For more information, refer to "Detailed Power Consumption and Configuration" on page 23. Agilent 7800/7900 ICP-MS Site Preparation Checklist 13
14 Cooling Water Requirements Special Notes 1 The preferred cooling system is the Agilent Chiller (G3292A), or Heat Exchanger (G1879B), filled with Poly- Clear Fluid (G ). 2 If you are not using the preferred system, and another type of water re- circulator is used, the reservoir should be filled with distilled water having a conductance in the range of 50 to 150 µs. 3 Distilled water will keep the system clean. Do not use tap water as it will contaminate the system and do not use deionized water as it will corrode the system. Table 8 Cooling Water Requirements Cooling Water Parameter Heat to be absorbed Flow Rate Temperature Conductivity * Connections Water Filter Specification 1,600W approximately > 5L/min (1.32 us gallons/min) Inlet pressure kPa (33-58psi) C at water inlet of ICP-MS µS/cm at the chiller reservoir Hose 1 10m long, ID=12mm (7/16inch) Male PT 1/2inch fittings on the 7800/7900 ICP-MS end <100 microns particle size * When not using the Chiller or Heat Exchanger in conjunction with Poly-Clear Fluid. CAUTION Continued operation with inadequate cooling can cause automatic shutdown of the ICP-MS and risk potential degradation in the performance of your instrument. Using tap water can cause contamination to the drinking water line from ICP-MS. The temperature of water exiting the system should be maintained at less than 40 C. The outlet water line must be open to atmosphere to minimize flow variations. For more information, refer to "Detailed Cooling Water Requirements and Operating Supplies" on page Agilent 7800/7900 ICP-MS Site Preparation Checklist
15 Gas Requirements Special Notes 1 The Argon gas regulators should be within 5 meters of the ICP- MS. It is strongly recommended that the cell gas cylinder, regulator with pressure gauge and shutoff valve are within 3 m of the ICP- MS. Argon Gas Table 9 Argon Gas Requirements Compressed gas Purity Typical Working Pressure kpa (psi) Consumption (L/min) Argon 99.99% (73-102) 20 Option Gas (If required for application) Oxygen 20% Argon 80% % (1.5-14) 1.0 Cell Gas Table 10 Cell Gas Requirements Compressed gas Purity Typical Working Pressure kpa (psi) Max Consumption (ml/min) Helium % ( ) 12 Hydrogen * % ( ) 10 Xenon * % ( ) 1 NH 3 /He 10%/90% * % * ( ) 10 *1: Hydrogen gas is an option with the ICP-MS. *2: 3rd Cell Gas Line is an option. *3: Both the NH3 and He gases used to prepare the gas mix must be % pure. For NH3/He Gas, Agilent recommends use of CONCOA Regulators and Protocol Station. To find a Concoa representative in your area please visit Concoa on the web at NOTE Regulator Connection Standards vary by country and region. Agilent 7800/7900 ICP-MS Site Preparation Checklist 15
16 Essential Requirements for Cell Gas Installation Special Notes 1 The cell gas pathway from the source to the mass spectrometer must be as short and as clean as possible to ensure correct operation of the ORS. 2 Only dedicated cylinders of % pure gas are to be used. 3 House gas supplies are not allowed, and sharing of gas supplies with other instruments (except another Agilent ORS (single quad or triple quad) ICP- MS) is not allowed. 4 The Agilent ancillary parts detailed below should be used to ensure the highest quality supply of He and H 2 cell gas. 5 Only the Agilent stainless steel tubing (Part Number G should be used to connect the He, H2, (or other cell gases) cell gas cylinders to the Agilent ICP-MS, and the length of the tubing must be less than 3m. 6 No other valves, fittings or connections are to be inserted along the length of the cell gas supply lines. NOTE on use of Hydrogen Generator: Use of a H 2 generator for cell gas supplies is not recommended. If a H 2 generator is used it must be fitted with an effective moisture trap and it must be correctly operated and maintained to ensure that the gas supply reaching the ICP- MS is completely free of moisture and contaminants. Failure to ensure that cell gases are clean and dry will compromise performance and increase maintenance, and may cause damage to the instrument. Agilent Part Number Description CP17976 * 1 Filter x 2, single position base plate (for He and/or H2) CP7988 CP7981 Single-position base plate Wall mount bracket Tube cutter G /8" Stainless steel tubing, 6 m *1: Do not use for NH3/He 7 Consult your local supplier for regulators detailed below. Description Quantity Note Two stage regulator, must be fitted with stainless steel diaphragms, kPa (14-28 psi) For use with: He Two stage regulator, must be fitted with stainless steel diaphragms, kPa (14-28 psi) For use with: H2 1 These regulators must be ordered in the country of use. Gas cylinder - regulator connections vary throughout the world and are not compatible. 1 For more information, refer to "Detailed Gas Requirements" on page Agilent 7800/7900 ICP-MS Site Preparation Checklist
17 Communications Special Notes 1 A network connection to the company LAN is not required for instrument operation. If you need to connect the ICP-MS computer to your company LAN, we only recommend connection via a switching HUB. A small 10/100 6 port switch can be obtained from many sources at low cost. 2 When connecting via a switching hub, your IT department must reserve two fixed (static) IP addresses that share the same subnet mask and are dedicated to the ICP- MS. 3 The Agilent ICP- MS system does not support the use of a computer with two network interface cards. It has been observed that the use of two network cards frequently results in crosstalk between the cards. This crosstalk can cause communication failures. LAN CROSS-OVER Cable Must have an IP Address IP Address: (Default) Subnet mask: IP Address: (Default) Subnet mask: CEHB018 Figure 6 Basic Connection Corporate LAN LAN Cable LAN Cable Must have an IP Address Switching HUB Must have an IP Address (IP address consistent with Corporate LAN) 79CEHB0183 Figure 7 Connection via a Switching HUB Agilent 7800/7900 ICP-MS Site Preparation Checklist 17
18 Laboratory Supply Requirements Special Notes 1 The following items are commonly used with the ICP- MS system. Refer to Table 11 below, and make sure you obtain the necessary supplies. A set of clean hand tools dedicated for maintenance of ultra- clean parts such as the ion lens, interface, etc. Reagent- grade acetone, methanol, or isopropanol for cleaning. Proper storage, handling, and disposal of these chemicals is required for personal and environmental safety. Laboratory glassware detergent and deionized water for general cleaning of system components. WARNING Chemical solvents should be considered hazardous and must be handled with care. Contact your chemical supplier for additional solvent handling and safety information. An auxiliary work space and fume hood are needed for maintenance procedures such as interface, ion lens, etc. cleaning. Table 11 Recommended Tools, Equipment and Supplies Item Ball Driver Beakers Sample Bottles Cloths Cotton Applicators Gloves Gloves Magnifier Ultrasonic Bath Pliers Screwdrivers Wrenches Volt/Ohm Meter Description Metric, Hex set 50mL, 250mL, 600mL Polypropylene Clean Lint Free Lint Free Chemical Resistant for Inspection of Interface Cones at least 1 Quart Capacity Needle Nose Torx, Phillips & Flat head Various Metric High Impedance > 10M ohm 18 Agilent 7800/7900 ICP-MS Site Preparation Checklist
19 Other Requirements Special Notes 1 Customer must create Microsoft Account when the Excel 2013 package arrives. 2 If necessary, insert tie down brackets into mounting holes, create corresponding holes in the table, and secure the brackets with bolts. NOTE Seismic brackets are not supplied by Agilent. Prepare these items separately. Front Front 3mm 10mm 30mm Bracket (example) 7900_0394 Figure 8 Holes for Seismic Tie Down Agilent 7800/7900 ICP-MS Site Preparation Checklist 19
20 Important Customer Web Links For additional information about our solutions, please visit our web site at US/Pages/HomePage.aspx Need to get information on your product? Literature Library - Need to know more? Customer Education Need technical support, FAQs? Need supplies? 20 Agilent 7800/7900 ICP-MS Site Preparation Checklist
21 Appendix Detailed Environmental Conditions and Site Requirements Environment conditioning considerations include temperature, humidity, altitude, atmosphere, airborne dust and exhaust venting. Temperature and Humidity Be sure to add the additional heat output of other optional equipment and system peripherals installed in the lab. Agilent Technologies recommends you set up the water recirculator outside the laboratory; if possible in a well ventilated room. Additional allowances should be made for other heat sources, such as heat from adjacent rooms. Example: Air Conditioning Requirements: Case 1: Heat Exchanger in Lab. Heat Dissipated to the room is ,900-1,600-1, ,000 = 2,630 Watts Case 2: Heat Exchanger external to Lab. Heat Dissipated to the room is ,900-1,600-1, = 630 Watts Case 3: Chiller in Lab. Heat Dissipated to the room is ,900-1,600-1, ,200 = 3,830 Watts Case 4: Chiller external to Lab. Heat Dissipated to the room is ,900-1,600-1, = 630 Watts Exhaust Venting The ICP- MS can produce ozone and high temperatures. A ventilation duct must be connected to the exhaust port of the instrument during operation for removal of waste products. NOTE Exhaust gas venting must comply with all local environmental codes. NOTE Agilent Technologies engineer will not install your ICP-MS system until an adequate exhaust system is present and functioning. Agilent 7800/7900 ICP-MS Site Preparation Checklist 21
22 Your exhaust system must have the capacity to maintain negative pressure with 6m 3 /min = 5.7m/sec of gas flow. Inadequate exhaust capacity can degrade the performance of your ICP- MS or cause system failures. Refer to the following example for calculating the exhaust flow in m 3 /min from the flow meter reading (m/min). <Example> meter reading 3.142r 2 = meter reading ( ) 2 = meter reading Note: r = Duct radius r 2 = Area of Duct = Safety information for installation Installation category based on IEC61010:II The Installation category implies the regulation for impulse withstand voltage. It is also called the Over voltage category. II applies to electrical equipment. Pollution level based on IEC61010:2 Pollution level describes the degree to which a solid, liquid or gas which deteriorates dielectric strength is adhering. 2 applies to a normal indoor atmosphere. Electrical Power (Voltage, Frequency, Amperage, Phase) VAC, 50/60Hz, 30A, single phase Supply voltage fluctuations are not exceed 10% of the nominal supply voltage. 22 Agilent 7800/7900 ICP-MS Site Preparation Checklist
23 Detailed Power Consumption and Configuration General Power Configuration Power for the ICP- MS is supplied in single- phase, Split phase, 3 phase star 4 wired, High- leg delta or 208 Wye configuration (see Table 12). Correct grounding for the 208 Wye configuration must be verified by an electrician. The neutral wire cannot be used for safety grounding. The ground wire (green or green/yellow) should carry zero current except for ground- fault current or static electric discharge. The entire system should share an isolated, noise- free electrical ground. Table 12 Electrical Configurations Configuration Measurement Nominal voltage Single phase Split phase Line(L) to Neutral Line(L) to Ground Ground to Neutral Line(L1) to Line(L2) Line(L1) to Ground Line(L2) to Ground 3 phase star 4 wired Line to Neutral (Phase A to Neutral) Line to Ground (Phase A to Ground) Ground to Neutral High-leg Delta (3 phase delta) Line to Neutral (Phase B to Neutral) Line to Ground Ground to Neutral (Phase B to Ground) 208 Wye Line to Line (Phase A to Phase B) Line to Ground (Phase A to Ground) 220Vac 220Vac <0.5Vrms 240Vac 120Vac 120Vac 240Vac 240Vac <0.5Vrms 207Vac 207Vac <0.5Vrms 208Vac 120Vac H1 L 220Vac H2 Neutral Safety Ground 7700_0001 Figure 9 Single Phase Configuration Agilent 7800/7900 ICP-MS Site Preparation Checklist 23
24 H1 L1 120Vac L0 240Vac H2 L2 120Vac Safety Ground 7700_0002 Figure 10 Split Phase Configuration 240Vac Neutral Phase A 240Vac 415Vac 415Vac 240Vac Phase C 415Vac Phase B Safety Ground 7700_0003 Figure 11 3 Phase Star 4 Wired Configuration A Phase A 120Vac 240Vac Neutral Phase C 120Vac 240Vac B C Phase B High Leg 207Vac 240Vac Safety Ground 7700_0004 Figure 12 High-Leg Delta Configuration A 120Vac 208Vac 208Vac Phase A 120Vac Neutral Phase C B C 120Vac 208Vac Phase B Safety Ground 7700_0005 Figure Wye Configuration 24 Agilent 7800/7900 ICP-MS Site Preparation Checklist
25 Power Plugs and Cords Power Receptacle (NEMA L6-30R) 200~240Vac Active (L1, black wire) Neutral (L2, white wire)* Isolated Ground (green with yellow stripe) Power Plug (NEMA L6-30P) Power Plug (IEC 60309) Power Cord 4.3m (14.1ft) 7900_0008 Figure 14 Power Receptacle for Vac, 50/60Hz (NEMA L6-30R) Power Receptacle (IEC 60309) 200~240Vac Active (L1, black wire) Neutral (L2, white wire)* Isolated Ground (green with yellow stripe) Power Plug (IEC 60309) Power Plug (IEC 60309) Power Cord 4.3m (14.1ft) 7900_0009 Figure 15 Power Receptacle for Vac, 50/60Hz (IEC 60309) Agilent 7800/7900 ICP-MS Site Preparation Checklist 25
26 Power Conditioner/Uninterruptible Power Supply (UPS) If the power supplied is outside the limits specified, a power conditioner may be required. Power conditioners aid in filtering impulses caused by lightning strikes, line spikes, oscillatory transients and electrical noise impulses. It is the customer s responsibility to install the power conditioner and supply any additional equipment, circuit breakers, switches, etc., before the Agilent Technologies engineer arrives on site. It is your responsibility to comply with all local and national electrical and safety codes. Check with your electrical department! Data system components and accessories have power cords with plugs depending on the voltage and power cord option ordered. Power cord lengths for the data system components and accessories are 2.5m. NOTE The ICP-MS has a start up rush current of 150A for 15 milliseconds. 26 Agilent 7800/7900 ICP-MS Site Preparation Checklist
27 General Power Requirements Table 6 lists the power requirements for the ICP- MS and associated equipment. Extra power capacity for the future growth of your laboratory should be considered now. Power requirements and considerations include the following. Each product listed in Table 6 requires a dedicated circuit. The ICP- MS mainframe, PC/PC monitor, water recirculator, etc. should each have a separate circuit breaker. Power must meet the stability and transient specifications listed in Table 6. We recommend your site power specialist use a line monitor to check power stability. If your line power is unstable, you may need to install a line conditioner. Separate convenience outlets should be provided for building maintenance and other appliances. Convenience outlets must be on circuits separate from the ICP- MS system and must share the normal building distribution ground, not the ICP- MS system ground. In some geographical areas it may be advisable to install lightning protection for personnel and equipment. Electromagnetic interference (EMI) generated by NMRs, radio transmitters, and microwave links, may interfere with system performance. Protect the system from static electricity by observing humidity and temperature requirements. Minimize the presence of non- conductive products such as carpets and vinyl floor tiles. Emergency- off push buttons that will disconnect power to the ventilation system and all electric equipment in the room except overhead lighting are recommended. A minimum of 4 power receptacles are required for data system installation; 6 power receptacles are recommended. Surge protection on the data system circuit is also recommended. CAUTION There is a power receptacle on the rear of the ICP-MS, it is dedicated for the Foreline Pump. DO NOT USE FOR ANY OTHER AUXILIARY EQUIPMENT. NOTE 1 Verify that the voltage available on site is adequate for all the equipment ordered. 2 Approximate values. Refer to the specific product specification. Data systems typically require at least 4 outlets and a 15A circuit with surge suppression. Agilent 7800/7900 ICP-MS Site Preparation Checklist 27
28 Detailed Cooling Water Requirements and Operating Supplies Water Quality In the event that the chiller or heat exchanger with Poly-Clear Fluid is not used, and another type of water re- circulator is used, the reservoir should be filled with distilled water having a conductance in the range of µs. Distilled water will keep the system clean. When the Chiller or heat exchanger in combination with Poly- Clear Fluid is not being used, refer to Table 13 for the water quality standards and recommendations. Table 13 Water Quality Standards and Recommendations Inorganic Chemicals * µs/cm (Compensated at 25 C) Permissible (ppm) Calcium < Chloride 250 <25 Copper Iron 0.3 <0.1 Lead Magnesium < Manganese 0.05 <0.03 Nitrates/Nitrites 10 as N 0 Potassium < Silicate 25 <0.1 Sodium < Sulfate 250 <50 Hardness 17 <0.05 Total Dissolved Solid Other Parameters ph Conductivity * 50 * Desirable (ppm) Unfavorably high total ionized solids (TIS) can accelerate the rate of galvanic corrosion. These contaminants can function as electrolytes which increase the potential for galvanic cell corrosion and lead to localized corrosion such as pitting which can be observed at the studs and on the outside surface of cooling coils of the chiller. Eventually, the pitting will become so extensive that the coil will leak refrigerant into the water reservoir. 28 Agilent 7800/7900 ICP-MS Site Preparation Checklist
29 As an example, raw water in the United States averages 171ppm (of NaCl). The recommended level for use in a water system is between 0.5 to 5.0ppm (of NaCl). NOTE Initially fill the tank with distilled water. Do not use tap water as the total ionized solids level may be too high. Do not use deionized water as it will corrode the system. Detailed Gas Requirements The Argon gas needed for installation will be supplied by the customer. It is recommended that there be a shutoff valve and a regulator with a pressure gauge as shown in Figure 16, within 5m of the ICP-MS. The customer will supply and prepare any gas piping from a remote gas storage area. This must be at least 1/4 inch in diameter and should be made of stainless steel. Gas cylinders or liquid Argon may be used. Typically a cylinder of Argon lasts 8 hours. A 260L tank of liquid Argon lasts for approximately one month of daily (8 hours) operation. WARNING Strictly adhere to all local and national regulations and guidelines for the proper storage, handling and transport of all gases. Compressed gas tanks must be handled with care. The contents of the cylinders also may be hazardous, depending on the gases you choose to use. Contact your gas supplier for cylinder handling, storage, transport and safety information for the gases that you will be using. Figure 16 NPT Fitting and 1/4 Inch Nut Connectors at the Pressure Regulator Agilent 7800/7900 ICP-MS Site Preparation Checklist 29
30 Example: Regulator with Pressure Gauges and ShutOff Valve Argon Gas Option Gas (if required) Argon Gas Option Gas 375 Note: The length of line after the regulator should be 5m or shorter. Do not use the supplied nylon tubing before the shutoff valve. 240 Note: Fittings and the Filter used for External Connection included in shipping kit _0016 Figure 17 Example Gas Connection and Connection Location (all units in mm) 30 Agilent 7800/7900 ICP-MS Site Preparation Checklist
31 Argon Gas Purity Contamination of instrumentation and poor ignition of plasma can be caused by low grade Argon, an acceptable specification for Argon purity is listed in Table 14. Table 14 Argon Purity Contamination Level O 2 Oxygen <2ppm N 2 Nitrogen <10ppm CO Carbon Oxide <0.5ppm CO 2 Carbon Dioxide <0.5ppm CH 4 Methane <0.5ppm H 2 O Water <5ppm Requirements for Argon Gas Supply Installation Agilent Technologies will provide nylon tubing, 1/4 inch (6.4mm) O.D., 5m long. This is used for supplying gases to the ICP- MS from the laboratory shutoff valve. NOTE The length of line after the regulator should be within 5m. Do not use the supplied nylon tubing before the shutoff valve. NOTE Fittings and the filter used for external connection included in shipping kit. Customers will provide connection from the gas supply to a shutoff valve. This tubing should be 1/4 inch (6.4mm) Electro Polished stainless steel or stainless steel that has been cleaned. The customer should provide the pressure regulator for the Argon gas cylinder. All pressure regulators should be stainless steel two- stage for ultra high purity; Input: Pa (0-3500psig) Output: Pa (0-150psig) When ordering, consider the size of the output tubing, 1/4 inch (6.4mm), and also the Compressed Gas Association (CGA) number. Your gas supplier can advise, supply and install all items for a successful installation. NOTE Please ensure that the type of connector used at the outlet side of the gas-pressure regulator conforms to applicable national requirements. Agilent 7800/7900 ICP-MS Site Preparation Checklist 31
32 WARNING Compressed gas tanks must be handled with care. The contents of the cylinders also may be hazardous depending on the gases you choose to use. Please contact your gas supplier for cylinder handling and safety information for the gases that you will be using. Cell Gas Requirement It is strongly recommended that the cell gas cylinder, regulator with pressure gauge and shutoff valve are within 3m of the ICP- MS. The ICP- MS will be shipped with 1/8 inch Swagelok fitting for Helium (optoins #110 and #111 will be shipped with 1/8 inch Swagelok fittings for Helium and Hydrogen cell gas connections) on the rear of the instrument. Example: Regulator with Pressure Gauges and ShutOff Valve Helium Gas Hydrogen Gas Helium Cell Gas Hydrogen Cell Gas 3rd Cell Gas (Option) _0017 Figure 18 Example of Cell Gas Connection and Connection Location (all units in mm) 32 Agilent 7800/7900 ICP-MS Site Preparation Checklist
33 Gas Purity Problem Atmospheric Oxygen, moisture and volatile organics entering the cell gas lines will result in poor/unpredictable ORS interference removal, and severe signal drift or changes in sensitivity due to contamination of the octopole (which will then have to be replaced). The common causes are: Use of existing He and H 2 lab supplies, instead of using dedicated cylinders for the ICP- MS Dirty, or previously used piping to connect the cylinder to the ICP-MS (or use of Cu instead of stainless steel) Use of gas cylinder regulators with plastic diaphragms (use only stainless steel diaphragm regulators designed for use with high purity gas supplies) Use of a Hydrogen generator (we don't recommend the use of these as the moisture content of the H 2 is too high) Use of cell gas of less than % quality Sharing of cell gas lines with other instrumentation The fitting of a gas purifier offers further protection for the ORS - for example if a cylinder of contaminated gas is connected, the gas purifier assures peak performance (spectral interference removal) of the ORS cell. Follow the instructions below to ensure your system is operating at peak performance. Essential Requirements for Cell Gas Supply Installation NOTE The following regulators are available for North America Regulator for Helium Cell Gas Regulator for Hydrogen Cell Gas Regulator for Ammonia Cell Gas Regulator for Argon Gas Customers will supply all gases that are needed for installation. Customers will provide connection from the gas supply to the instrument. This tubing should be 1/8 inch (3.2mm) Electro Polished stainless steel or Bright Anneal stainless steel or stainless steel that has been cleaned. The ICP- MS requires high purity cell gases for best performance. Stainless Steel Tubing (Part# G ) must be used for the cell gas lines; other materials (such as copper) are not satisfactory. Agilent does not guarantee the ICP- MS performance when using alternative materials for the cell gas line tubing. The customer must supply pressure regulators for the cell gas cylinders. Stainless steel two stage regulators are recommended for ultra high purity, corrosive or toxic gas applications; Agilent 7800/7900 ICP-MS Site Preparation Checklist 33
34 Input: Pa (0-2000psig) Output: Pa (0-15psig), or Pa (0-30psig) When ordering, consider the size of the output tubing, 1/8 inch (3.2mm), and also the Compressed Gas Association (CGA) number. NOTE Please ensure that the type of connector used at the outlet side of the gas-pressure regulator conforms to applicable national requirements. 34 Agilent 7800/7900 ICP-MS Site Preparation Checklist
35
36 In this Book This manual will help you prepare your facility for the arrival of your new Agilent 7800/7900 Inductively Coupled Plasma Mass Spectrometer (ICP- MS). The specifications in this manual ensure consistent, reliable and safe installation of your ICP- MS system. Agilent Technologies, Inc Rev. B, June 2015 *G * G Agilent Technologies
Micro GC Fusion Pre-Installation Checklist
Site Preparation Proper site preparation prior to installing is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation. NOTE: Checkboxes ( ) identify hardware that
NexION 300 ICP-MS. ICP-Mass Spectrometry PREPARING YOUR LAB. Preparation Checklist
PREPARING YOUR LAB NexION 300 ICP-MS ICP-Mass Spectrometry Preparation Checklist Environmental conditions Electrical requirements Space requirements Exhaust ventilation Coolant requirements Argon gas requirements
Lava. Precision Solutions. Site Planning and Installation Guide
Lava Precision Solutions Site Planning and Installation Guide TABLE OF CONTENTS Table of Contents Introduction...........................................................................................
NITROUS TRANSFER PUMP INSTRUCTIONS
NITROUS TRANSFER PUMP INSTRUCTIONS SAFETY TIPS Never directly inhale nitrous oxide. When inhaled in large quantities, nitrous oxide can cause respiratory ailments or in extreme cases, death by suffocation.
Pre-installation Manual. CONCEPT Processor 305 DW / 405 DW / 505 DW 505 DW XP
Pre-installation Manual CONCEPT Processor 305 DW / 405 DW / 505 DW 505 DW XP Pre-installation Manual CONCEPT Processor 305 DW / 405 DW / 505 DW / 505 DW XP T11336 Edition AA, April 2013 This book has
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
Vision Engravers and Routers PRE-Installation Guide. 2015 Vision Engraving & Routing Systems
Vision Engravers and Routers PRE-Installation Guide Revised: 8/19/2015 Vision Engravers and Routers PRE-Installation Guide All rights reserved. No parts of this work may be reproduced in any form or by
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
USE &CARE GUIDE. Remote Faucet Pump System. See Important Safeguards on page 2
Remote Faucet Pump System USE &CARE GUIDE See Important Safeguards on page 2 An exclamation point within an equilateral triangle is intended to alert user to the presence of important operating and maintenance
Element D Services Plumbing
Medical Vacuum and Gas PART 1 - GENERAL 1.01 OVERVIEW A. This section addresses medical vacuum, waste anesthetic gas disposal, compressed air, oxygen, nitrous oxide, nitrogen and carbon dioxide systems.
BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS
BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS PLEASE READ COMPLETE INSTRUCTIONS BEFORE YOU BEGIN LEAVE INSTALLATION INSTRUCTIONS AND USER'S GUIDE WITH OWNER ALL ELECTRIC WIRING AND PLUMBING MUST BE DONE
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter
Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases
VSO MAX HP High Flow Proportional Valve
VSO MAX HP High Flow Proportional Valve Typical Applications Ventilators Insufflators Anesthesia Delivery & Monitors Pressure & Flow Control Mass Flow Control Product Specifications Physical Properties
Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance
Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam
SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS
ADVANTAGE PLUS SITE REQUIREMENTS SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS 50096-927 Rev A For additional information contact your Medivators Pg 1 of 18 SITE REQUIREMENTS The purpose of this section
Keysight N4877A Clock Data Recovery and Demultiplexer 1:2
Keysight N4877A Clock Data Recovery and Demultiplexer 1:2 Getting Started Getting Started with the Keysight N4877A Clock Data Recovery and Demultiplexer 1:2 You only need a few minutes to get started
HP UPS R1500 Generation 3
HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications
USER S MANUAL HSC-24A
AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL
AXIS T81B22 DC 30W Midspan
INSTALLATION GUIDE AXIS T81B22 DC 30W Midspan ENGLISH About this Document This document includes instructions for installing AXIS T81B22 on your network. Previous experience of networking will be beneficial
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MM14CCS. Voltage rating: 115V~60Hz Power rating : 1400W
Portable Air Conditioner OWNER S MANUAL Read these instructions before use Model: MM14CCS Customer Support : 1-800-474-2147 Voltage rating: 115V~60Hz Power rating : 1400W For product inquiries or support
Agilent P940xA/C Solid State PIN Diode Switches
Agilent P940xA/C Solid State PIN Diode Switches Operating and Service Manual Agilent Technologies Notices Agilent Technologies, Inc. 2007 No part of this manual may be reproduced in any form or by any
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR
USER INSTRUCTIONS FOR GET PORTABLE 12k BTU AIR CONDITIONER MODEL No. GPACU12HR CONTENTS Introduction Safety Notes Identification of parts Installation instructions Operation instructions Maintenance Troubleshooting
SunMaxx Solar Filling Station Operating Instructions
SunMaxx Solar Filling Operating Instructions Content 1. Declaration of conformity... 2 2. Introduction... 2 3. Transportation and unpacking... 4 4. Mounting and commissioning... 5 5. End of operation...
Heating, Ventilation, Air Conditioning and Refrigeration (HVACR)
Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) I. Demonstrate safety skills in typical HVACR work situations to NATE Core Installer Knowledge Areas for Technician Excellence for Safety
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
INSTALLATION INSTRUCTIONS Fan Coil Replacement Coil Kit EBX & EBXX
Fan Coil Replacement Coil Kit EBX & EBXX These instructions must be read and understood completely before attempting installation. These instructions covers the installation of replacement coil kit into
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MN12CES / MN10CESWW
Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MN12CES / MN10CESWW Voltage rating: 120V~60Hz Power rating : 1100W (MN12CES) Power rating : 900W (MN10CESWW) Customer
Dehumidifier Users manual. For Models: DH45S DH65S
Dehumidifier Users manual For Models: DH45S DH65S 950-0062-revD Jan. 9 2007 FORWARD The appearance of the units that you purchase might be slightly different from the ones described in the Manual, but
Portable Air Conditioner. OWNER S MANUAL Read these instructions before use. Model: MF08CESWW. Voltage rating: 115V~60Hz Power rating : 800W
MODE ALARM Portable Air Conditioner OWNER S MANUAL Read these instructions before use 8 Model: MF08CESWW Voltage rating: 115V~60Hz Power rating : 800W Customer Support : 1-800-474-2147 For product inquiries
NAVIEN AMERICA INC. 17855 FITCH AVENUE
Navien America Inc. Specification Details for the Navien models: CC-240 CC-240A For Potable Water Heating and Space Heating NAVIEN AMERICA INC. 17855 FITCH AVENUE IRVINE, CALIFORNIA, USA 92614 Toll Free:
FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters
FMA 3100/3100ST/3300/3300ST Series Thermal Mass Flow Sensors and Meters M-4270/0707, pg. 2 of 29 READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY
Meaco 30L and Meaco 40L dehumidifier instruction manual
Meaco 30L and Meaco 40L dehumidifier instruction manual Please read this instruction manual before using the dehumidifier and keep safe for future reference SAFETY INSTRUCTIONS PLEASE READ ALL INSTRUCTIONS
Split-type Air-Conditioner INSTALLATION MANUAL CONTENTS FOR INSTALLER MXZ-3A30NA MXZ-4A36NA ATTENTION. English. Français. Español
Split-type Air-Conditioner MXZ-3A30NA MXZ-4A36NA INSTALLATION MANUAL Refer to the installation manual of each indoor unit for indoor unit installation. English Français Español ATTENTION This manual mentions
Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF
Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator
APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out
APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out APC Matrix-UPS, 3750 Watts / 5000 VA,Input 208V / Output 120V, 208V, Interface Port DB-9 RS-232, SmartSlot, Extended runtime model Includes: CD with software,
MACBlower Model Number: MAC40R MAC60R MAC80R MAC100R. MAC120R MAC150R MAC 200R Serial # www.macblowers.com
MACBlower Model Number: MAC40R MAC60R MAC80R MAC100R MAC120R MAC150R MAC 200R Serial # Fuji Clean USA 41-2 Greenwood Road Brunswick, Maine 04011 207-406-2729 www.macblowers.com MACBlowers The Intelligent
OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE
OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:
Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS
Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS Contents Page Ratings and Specifications..................... 2 Installation Requirements......................
NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10000E, AC-10000H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS BEFORE USE GENERAL SAFETY INSTRUCTIONS: ALWAYS OPERATE THE UNIT IN AN UPRIGHT POSITION AND PLACE
Outdoor 33.6W Dual Port Passive Power-over-Ethernet Midspan For External Security Cameras and Wireless Access Points
Outdoor 33.6W Dual Port Passive Power-over-Ethernet Midspan For External Security Cameras and Wireless Access Points Features SELV Compliant No Detection Passive Injector Gigabit Compatible Full Protection
Oil and Coolant Circulating Heating System. Model - OCSM
Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine
OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608
Electric Water Heater OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608 If your water Heater is Damaged or you have questions
FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases
FLR 1000 & 1000-BR Series Flow Sensors & Meters For Gases READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY RESULT IN INJURY TO YOU OR DAMAGE TO
VOYAGER 570G. 744A Sprayer Control
VOYAGER 570G 744A Sprayer Control U S E R M A N U A L U S E R M A N U A L Table of Contents CHAPTER 1 - INTRODUCTION...1 SYSTEM CONFIGURATIONS...1 KIT CONTENTS...3 CONTROL HOUSING ASSEMBLY...5 CHAPTER
QuickSpecs. HP Fixed Cord PDUs & Extension Bars (Zero-U/1U Modular PDUs) Overview
Overview Fixed Cord PDUs and Extension Bars have a unique architecture designed specifically for 1U fixed rail servers (HP ProLiant DL360, HP ProLiant DL320, HP ProLiant DL140). This new design consists
Index 11.1. Page. Pumping Systems...11.2-11.4 Intensifiers...11.5 Gas Boosters...11.6-11.7 High Pressure Generators...11.8-11.9
Pumping Systems, Intensifiers, Gas Boosters and High Pressure Generators High Pressure Equipment Company produces a number of components and systems for general industrial, elevated pressure applications.
Ceiling Mount Air Handler Manual
www.surna.com 303.993.5271 Ceiling Mount Air Handler Manual Models: CMAH12, CMAH18, CMAH24, CMAH30, CMAH36, CMAH48, CMAH60 Revised: September 2014 Table of Contents Warranty Information 4 Limited Warranty
Match. GE Digital Energy. Uninterruptible Power Supply 500-1500 VA. Technology for the Digital World. Match UPS. GE Digital Energy.
Match Uninterruptible Power Supply 500-1500 VA Manufactured by: General Electric Company Telephone +41 (0)91 / 850 51 51 CH 6595 Riazzino (Locarno) Fax +41 (0)91 / 850 51 44 Switzerland Website www.gedigitalenergy.com
33.6W Power over Ethernet Waterproof Adapter PoE Plus Single Port Injector for Outdoor Application
33.6W Power over Ethernet Waterproof Adapter PoE Plus Single Port Injector for Outdoor Application Features Compliant with the IEEE802.3at Standard -40 to +60 C Temperature Range Diagnostic LEDs Full Protection
Portable Evaporative Air Cooler. OWNER S MANUAL Read and save these instructions before use. Model: CL30XC
Portable Evaporative Air Cooler OWNER S MANUAL Read and save these instructions before use Model: CL30XC Power rating: 250 Watts Voltage rating: 230 Volt, 50Hz Made in P.R.C. QUICK START GUIDE Fill with
Quick Installation Guide 24-port PoE switch with 2 copper Gigabit ports and 2 Gigabit SFP ports (af Version 15.4W)
Quick Installation Guide 24-port PoE switch with 2 copper Gigabit ports and 2 Gigabit SFP ports (af Version 15.4W) Table of Contents Introduction.. Power Over Ethernet (PoE) & Features.... Unpacking and
Your safety and the safety of others are very important.
NATURAL GAS TO PROPANE CONVERSION KIT 090 INSTALLATION INSTRUCTIONS FOR ALTITUDES 0 -,00 FT. ONLY PROPANE CONVERSION KIT SAFETY... INSTALLATION REQUIREMENTS... Tools and Parts... LP Gas Requirements...
OPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
Cooktop Low-Profile Ventilation Hoods
INSTALLATION GUIDE Cooktop Low-Profile Ventilation Hoods Contents Wolf Cooktop Low-Profile Ventilation Hoods........ 3 Cooktop Low-Profile Hood Specifications.......... 4 Cooktop Low-Profile Hood Installation............
REFRIGERATED TYPE COMPRESSED AIR DRYERS INSTRUCTION MANUAL. For Sales & Service please contact:
7610.478.16 8/99 INSTRUCTION MANUAL Models 8005, 8010, 8015 For Sales & Service please contact: CENTRAIR Air Systems & Supplies Phone: 705-722-5747 Fax: 705-722-5458 Email: [email protected] Website: www.centrair.ca
USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920
USER MANUAL Stand Alone Power Supply PSQ 2909 / PSQ 3909 / PSQ 4909 PSQ 2920 / PSQ 3920 / PSQ 4920 [This page intentionally left blank] Warning for Your Protection 1. Read these instructions. 2. Keep these
CG-1171-20 PTFE Diaphragm Liquid Transfer Pump
3800 N. Mill Road Vineland, NJ 08360 Phone: 800.843.1794 Fax: 800.922.4361 CG-1171-20 PTFE Diaphragm Liquid Transfer Pump This is the safety alert symbol. It is used to alert you to potential personal
SYSTEM 1E LIQUID CHEMICAL STERILANT PROCESSING SYSTEM
SYSTEM 1E LIQUID CHEMICAL STERILANT PROCESSING SYSTEM APPLICATION For rapid, safe, standardized low-temperature liquid chemical sterilization of heat-sensitive critical and semi-critical devices. DESCRIPTION
Line to Refrigerator Ice/Water Dispenser
Standard 18 Line to Refrigerator Ice/Water Dispenser Pump Module Bottled Water How The System Works The FLOJET Bottled Water Dispensing System was designed to pump water from a commercially available 5-gallon
BroadBand PowerShield. User Manual
BroadBand PowerShield User Manual 990-0375G 12/2006 Chapter 1 General Information The PowerShield provides a power source for broadband telephony and other DC applications. Safety This Safety Guide contains
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
Power Max Level 2 Charging Station en Installation and Operating Instructions This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the
Powers Controls TH 192 S Single Temperature Room Thermostat
Technical Instructions Document No. 155-065P25 TH 192-1 Powers Controls TH 192 S Single Temperature Room Thermostat 50 60 70 80 TH0355R1 70 80 POWERS Description The TH 192 S thermostats are proportional
Seaward Products OWNER S MANUAL WATER HEATERS. Serial Number:
Seaward Products WATER HEATERS OWNER S MANUAL Serial Number: IMPORTANT SAFETY INSTRUCTIONS WARNING When using electrical appliances, basic safety precautions to reduce the risk of fire, electrical shock,
Powers Controls TH 192 HC Heating/Cooling Room Thermostat
Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional
How To Use A Power Supply Unit (Upu)
BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.
FOXAIR 60/50 ALL MODELS
FOXAIR 60/50 ALL MODELS OPERATION AND SERVICE MANUAL UNIT WEIGHT - ALL MODELS APPROX. 300 POUNDS FOXTRONICS, INC. 38 L x 28 W x 24 H Love Field-Dallas TX MADE IN THE USA 3448 West Mockingbird Lane Dallas,
The Breather Box. High Performance Grade-D Breathing Air Filtration system. BB50-cO
The Breather Box High Performance Grade-D Breathing Air Filtration system Specifications Filter change Indicators Incoming Air Plug 1/2 Industrial Interchange 2 Automatic Drains with Discharge tubes 3
Industrial-Grade UPS System Heavy-duty power protection for harsh industrial environments
Industrial-Grade UPS System Heavy-duty power protection for harsh industrial environments Model #: SM2000RMX Heavy-duty, industrial-grade metal housings Shock and vibration rated to MIL-STD-810D 19 inch
CARING FOR YOUR WATER HEATER
http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html Water Heater Inspections CARING FOR YOUR WATER HEATER Venting System (Direct Vent Only) The venting system should be inspected
User s Manual AURORA 1.2K/2.2K
User s Manual AURORA 1.2K/2.2K Uninterruptible Power System Safety CAUTION This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user serviceable
Introduction. Safety Guidelines. Part Numbers. Operation and Maintenance Manual. Operation and Maintenance Manual
Operation and Maintenance Manual Introduction Read all instructions thoroughly. Installation of the OilTector must comply with all Federal, State and Local Codes, Regulations and Practices. The OilTector
HEAVY DUTY STORAGE GAS
Multi-Fin flue technology Flue damper saves energy Electronic controls HEAVY DUTY STORAGE GAS Dependability The Rheem heavy duty gas range is the work horse of the industry having proved itself over many
STATFlow. Direct-to-Drain System. For use with STATIM 2000 and STATIM 5000 Autoclaves. Installation Guide and Operator s Manual
STATFlow Direct-to-Drain System For use with STATIM 2000 and STATIM 5000 Autoclaves Installation Guide and Operator s Manual STATFlow Installation Guide and Operator s Manual 95-113193 Rev 2.0. Copyright
Please also see 10 golden rules for correct Saniflo installation on www.saniflo.co.uk
1. Description Sanicom is a lifting pump for wastewater used in commercial applications (restaurants, hairdressing salons, the catering industry etc.). The unit operates automatically, has a high performance
Drayton Digistat +2RF/+3RF
/+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer
Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains
Drayton Programmable Room Thermostat 7 Day (5-2 Day) Models: 22083 / 22087 Power Supply: Battery / Mains Invensys Controls Europe Technical Helpline: +44 (0) 845 130 7722 www.draytoncontrols.co.uk Installation
36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve
Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION
Select Radiators Installation Guide
Select Radiators Installation Guide Table of Contents Informational Symbols...3 Before You Begin...4 Select Rough-In... 5 Connection Installation...6 Optional Piping Arrangements...7 Conventional Wall
Agilent N9911X Economical Waveguide Calibration Components
Agilent N9911X Economical Waveguide Calibration Components Configuration Guide Agilent Technologies Notices Agilent Technologies, Inc. 2013 No part of this manual may be reproduced in any form or by any
Uninterruptible Power Supply
96-01101 / rev. 2e / 2-2-12 Uninterruptible Power Supply EXCEPTIONAL SUPPORT & PROTECTION Uninterruptible Power Supply with Energy Saver design that is optimized to address the needs of A/V systems Features
ARCO Electric Products Installation and Maintenance Manual Low Voltage Automatic Power Factor Correction Capacitor Systems 2013
ARCO Electric Products Installation and Maintenance Manual Low Voltage Automatic Power Factor Correction Capacitor Systems 2013 READ CAREFULLY These instructions are intended to cover good practices in
WHOLESALE DIRECT INC. 5620 WEST 65TH STREET
Auto Pump V 200 100 0 100 Max Rating Vehicle mount compressor ensures truck air brake system is properly pressurized for immediate dispatch from station Pressure switch regulated operation automatically
National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook
National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook Trainee:... Training Provider:... Year:... Trainee Contact no:... Competency Based Skill Training Logbook Introduction
6 Crankcase Ventilation
6 Crankcase Ventilation Introduction... 6-3 Crankcase Ventilation Manager (CVM)... 6-5 Introduction... 6-6 CVM280B..................................... 6-7 CVM280T... 6-10 CVM424B....................................
INSTALLATION AND OPERATING INSTRUCTIONS For Model GL1 Gate Locks
Securitron Magnalock Corp. www.securitron.com ASSA ABLOY, the global leader Tel 800.624.5625 [email protected] in door opening solutions INSTALLATION AND OPERATING INSTRUCTIONS For Model GL1 Gate
Utility Distribution Systems
Utility Distribution Systems 6/2012 A0011037 1 WARRANTY This equipment is warranted to be free from defects in materials and workmanship, under normal use and service, for a period of 12 months from date
I nstallation. M100Q Series Proportional Actuator with R81Q Controller Board for Thermistor Sensor Applications. Tools Needed.
FANs 268.1, 1628.3 Installation Bulletin M100Q Issue Date 1099 M100Q Series Proportional Actuator with R81Q Controller Board for Thermistor Sensor Applications I nstallation Parts Included M110QGA-1 and
Portable Evaporative Air Cooler for Outdoor, Indoor & Commercial Use. OWNER S MANUAL Read and save these instructions before use.
OFF COOL 0 3 2 1 OFF Portable Evaporative Air Cooler for Outdoor, Indoor & Commercial Use ON SPEED ON SWING OWNER S MANUAL Read and save these instructions before use Model: CO60PM Power rating: 220 Watts
Scitex FB550 and FB750 Printers. Site Preparation Checklist
Scitex FB550 and FB750 Printers Site Preparation Checklist 1 2015 Hewlett-Packard Development Company, L.P. The information contained herein is subject to change without notice. The only warranties for
Frequently Asked Questions Generators
Frequently Asked Questions Generators Fuel Related How much fuel do the generators use during normal operation? At exercise? The generator fuel consumption is given in each of the product knowledge books
AUTOMATIC CHANGEOVER MANIFOLD Series 915
AUTOMATIC CHANGEOVER MANIFOLD Series 915 FEATURES Fully automatic, simple, hassle-free operation No variance in delivery pressure Designed for ultra high purity gas service Built-in audio and visual alarm
Use and Care Manual. Model CPA12KH AIR CONDITIONER
Use and Care Manual Model CPA12KH AIR CONDITIONER Introduction Thank you for choosing this air conditioner to provide you and your family with all of the "Home Comfort" requirements for your home, cottage
EPM3. Phase Sequence and Motor Rotation Tester. Users Manual
EPM3 Phase Sequence and Motor Rotation Tester Users Manual 1 L1 L2 L3 2 3 A B C CAT 600V 3-PHASE TESTER 5 TEST M1 M2 BATT M3 EPM3 MOTOR ROTATION TESTER EPM3 MOTOR ROTATION DETERMINED WHILE FACING MOTOR
Instructions & Safety Information Models A220-20D and A220-20L Version 2
Quick 220 Voltage Converting Power Supply Instructions & Safety Information Models A220-20D and A220-20L Version 2 Quick 220 Systems LLC PO Box 47489 Phoenix, Arizona 85068-7489 800-347-0394 602-938-6057
NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL
NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Esenboğa Yolu 22 km Akyurt - ANKARA / TURKEY Tel : (00 90 312)
P R O D U C T S P E C I F I C A T I O N MSA Ultima X Series Sensor/Transmitter Specification
P R O D U C T S P E C I F I C A T I O N MSA Ultima X Series Sensor/Transmitter Specification 1.0 This specification details the attributes and operating characteristics of the MSA Ultima X Series sensors/transmitters.
