flow cytometry reinvented
|
|
|
- Elijah Beasley
- 9 years ago
- Views:
Transcription
1 FLOW CYTOMETRY flow cytometry reinvented The Attune Acoustic Focusing Cytometer Precision and at all speeds for rare events and precious samples
2 The Attune Acoustic Focusing Cytometer Precise high collection rates without compromising CVs Fast rare-event detection with dramatically shorter acquisition times Simple no-lyse, no-wash method eliminates cell loss With the Attune Acoustic Focusing Cytometer, you get both high sample input rate and high precision when analyzing precious samples or detecting rare events. You control your sample concentration, flow rate, the number of photons you detect, the length of your experiment, and more. You get the most Hydrodynamic out of your cells with dependable results. But best of all, the Attune makes it easy. focusing 60 instrument 40 Precision and at all sample throughput rates The Attune Acoustic Focusing Cytometer enables higher when you need it most. You ll be able to maintain precise alignment, even at high sample rates of up to 1 ml/min. And because the Attune 20 is not limited by physical 66.8% flow constraints, you can control the amount of time cells spend in the lasers path. This means you can use a slower rate to collect more photons when analyzing dim signals, translating to better clarity of your results and better separation (Figures 1 and 2) PE-A Hydrodynamic focusing instrument % Attune, normal mode % of Max Note: increased precision as seen by the tighter populations % PE-A BL2-A:BL2-A Figure Attune 1. Greater 0 peak separation with phosphospecific Note: antibody staining. Intracellular target detection can be challenging. Jurkat T cells were, stimulated with anti-cd3 and anti-cd28 antibodies, increased then fixed and permeabilized before staining with a PE-labeled phosphospecific antibody. normal The sample was 80run on the Attune precision and as a hydrodynamic focusing and the data compared. The plots show an overlay of the data from stimulated and unstimulated seen by samples the from each system. The data obtained with the Attune Acoustic Focusing tighter Cytometer mode have higher precision 60 (narrower peaks), resulting in greater separation and easier data interpretation. populations. % of Max % BL2-A:BL2-A
3 Conventional Conventional flow Conventional flow flow Attune Attune Attune Attune Attune Attune Conventional Conventional flow Conventional flow flow (Highest (Highest (Highest Attune Attune Attune Attune Attune Attune (Standard (Standard (Standard (High (High (High PE-Cy PE-Cy PE-Cy T T T T 1 4 T 6 4 T PE-Cy PE-Cy PE-Cy PE-Cy 3 21 PE-Cy PE-Cy (Highest (Highest (Highest (Standard (Standard (Standard (High (High (High Figure 2. Sensitivity through control of cell transit times. Fluorescent microspheres (Spherotech Rainbow 3.2 μm) were run on a high-end conventional flow and on the Attune Acoustic Focusing Cytometer using a 488 nm blue laser (top row) and a 40 nm violet laser excitation (bottom row). The conventional flow was run at its highest setting, at a sample input rate of 1 μl/min (left column). The Attune was run at 2 different settings: at its standard setting, at a 0 μl/min sample input rate (middle column), and at its highest setting, with a 0 μl/min sample input rate and 4x increase in the amount of time the particles are illuminated by the laser (right column). The Attune standard settings resulted in equal to or better than the conventional flow, while the high of the Attune instrument resulted in detection of the faintest beads not detected by the conventional flow. Rare events detection times faster The Attune Acoustic Focusing Cytometer achieves sample throughput at rates over times faster than other s up to 1,000 µl per minute, enabling rapid detection of rare events with reliable accuracy while aborting no data (Table 1, Figure 3). The Attune is designed to collect up to 20 million events per run and has adjustable collection rates of 2 1,000 µl/min. CD4R/B220-PE fluorescence Gated on live CD19 cells at collection rate of 00 ul/min Table 1. Comparison of acquisition times with a hydrodynamic focusing or the Attune Acoustic Focusing Cytometer, using a blood sample from an aplastic anemia individual, each with a stop gate set on 1 million granulocyte events. Instrument (collection rate) 6 Time to acquire Relative rate 1 million compared to hydrogranulocyte events dynamic focusing Hydrodynamic focusing (high) 63 min 33 sec 4 Attune Acoustic Focusing Cytometer (200 µl/min) 13 min 20 sec 4.8x faster 3 Attune Acoustic Focusing Cytometer (00 µl/min) min 47 sec 11.0x faster Attune Acoustic Focusing Cytometer (1,000 µl/min) 3 min 13 sec 19.7x faster CD317-Alexa Fluor 488 fluorescence Figure 3. Collecting more than 1 million live cells and detecting a rare population of dendritic cells of 0.2% with mouse splenocytes. Plasmacytoid dendritic cells (pdcs) are a specialized cell population that produces large amounts of type I interferons in response to viruses and are identified using the immunophenotype CD19, B220high, CD317+. Four-color staining of mouse splenocytes included CD19 Blue, CD317 Alexa Fluor 488, CD4R/B220-PE direct conjugates, and SYTOX AADvanced Dead Cell Stain. A gate was made on live cells using SYTOX AADvanced Dead Cell Stain, followed by gating on CD19 cells. A two-parameter plot of CD4R/B220 vs. CD317 was used to identify pdcs. A collection rate of 00 µl/min was used to acquire 1.3 million total cells with a cell concentration of 7.x^7 cells/ml. Plasmacytoid dendritic cells were identified as dual B220 +/CD317+ (upper right quadrant) and constitute 0.81% of live CD19 cells, which is 0.194% of total splenocytes.
4 SSC Attune sample rate: 1,000 µl/min CD CD34 Figure 4. Identification of CD34 + cells from peripheral blood. Peripheral blood from a normal donor was stained and run on the Attune Acoustic Focusing Cytometer at a collection rate of 1,000 μl/min with a stop gate set at 00,000 total cells. A rare population of 0.04% CD34 + cells (red box) was identified within the population of cells with an acquisition time of 4 minutes, 28 seconds. In addition, the Attune provides direct cell counts without the need for expensive counting beads with volumetric analysis (Figure 4, Table 2). Table 2. Comparison of absolute cell counting with a hydrodynamic focusing or the Attune Acoustic Focusing Cytometer. Instrument Sample Flow Rate (µl/ min) Acquisition Time Percentage of CD34 + Cells CD34 + Cells/µL, Direct Measurement Attune Acoustic Focusing Cytometer 1,000 4 min, 28 sec 0.0% 0.0 High-end conventional min, 46 sec 0.02% N/A Dilute your samples, not your data quality Washing and lysis of red blood cells (RBCs) can cause significant cell loss and damage [1]. Significantly higher sample collection rates allow the Attune to deliver a no-wash, no-lyse protocol to minimize cell loss and simplify sample preparation (Figure ). This feature is particularly useful for samples that are inherently low in concentration. Dilute samples like cerebrospinal fluid (CSF), stem cells, and any sample with low cell numbers can take a long time to acquire. With the Attune, even dilute samples can be acquired quickly without compromising data. Difficult-to-collect samples like mouse blood and bone marrow, thin-needle aspirates, or any sample with low cell yield can be stained and then diluted without washing or performing RBC lysis. High rate collection makes acquisition possible you can run up to 4 ml in just four minutes. No sample loss occurs from sample preparation, and full panel testing is possible for all precious samples. Figure. Eliminating sample preparation without compromising data. Normal whole blood (0 µl) was labeled with CD4 Blue, CD3 Alexa Fluor 488, and CD4 PerCP-Cy. direct conjugates. After 1 minutes of incubation, µl of the stained whole blood was diluted into 4 ml PBS, and data were acquired on the Attune without RBC lysis or washing. A fluorescence threshold was set on the Blue dye to include only CD4-positive cells. A gate was created around the lymphocyte population in a CD4 vs. SSC plot (A) to analyze the mutually exclusive CD4 and CD8 populations (B). By diluting the sample and using a rate of 00 µl/min, all sample preparation steps can be eliminated without compromising the data.
5 Cell cycle analysis using the Attune Acoustic Focusing Cytometer Cell cycle analysis is just one example of where it is critical to precisely detect differences in fluorescence intensity between multiple cell populations. With the Attune Acoustic Focusing Cytometer, minimal variation in results is seen regardless of sample throughput rate. You no longer need to sacrifice throughput for (Figure 6). Minimal variation, even at high sample rates (Figures 6 and 7) Less variability in results (see changes in %S phase from Table 1) No sacrifice of for speed %CV C-Low (12 µl/min) C-Med (3 µl/min) C-High (60 µl/min) Attune (2 µl/min) Attune (0 µl/min) 60 µl/min G O /G 1 CV 1,000 µl/min Attune (200 µl/min) Attune (00 µl/min) Attune (1,000 µl/min) Figure 7. Percent CV of G 0 / G 1 at different flow rates on the Attune and a competitor (C) at different sample rates. Note the minimal change in variability (%CV) for the Attune Acoustic Focusing Cytometer, even at a high sample rate. A High-End Hydrodynamic Focusing Cytometer Low - 12 µl/min B Attune Acoustic Focusing Cytometer 2 µl/min Five other cell types were compared to a high-end hydrodynamic focusing, including ModFit analysis in each case. The sample throughput rate of 2uL/min was used for the analysis (Figure 8). C Medium - 3 µl/min D 200 µl/min Ramos B Cells A Hydrodynamic System G0G1 CV = 3.3% B Attune Acoustic G0G1 CV = 3.11% E High - 60 µl/min F 1,000 µl/min KG-1 a Myeloblast Cells C Hydrodynamic System G0G1 CV =.3% D Attune Acoustic G0G1 CV = 3.8% HL60 Promyeloblast Cells E Hydrodynamic System G0G1 CV = 6.% F Attune Acoustic G0G1 CV = 2.14% Figure 6. Minimal data variation at high sample rates with the Attune Acoustic Focusing Cytometer. Jurkat cells were fixed and stained with propidium iodide, treated with RNase, and analyzed at a concentration of 1 x 6 cells/ml on a high-end instrument that uses hydrodynamic focusing, and on the Attune Acoustic Focusing Cytometer at different sample rates. The left peak in all graphs reflects cells in G 0 /G 1 phase, while the right peak reflects cells in G 2 /M phase. Note that as sample rates increase on the instrument that uses hydrodynamic focusing, the width of the G 0 / G 1 and G 2 /M peaks increase, whereas for the Attune the peaks are relatively stable, even at the highest sample rate of 1,000 µl/min. U266 Myeloma Cells G Hydrodynamic System H Attune Acoustic G0G1 CV = 7.17% G0G1 CV = 4.41% ST486 B Cells I Hydrodynamic System J Attune Acoustic G0G1 CV =.84% G0G1 CV = 2.81% Figure 8. Attune Acoustic Focusing Cytometer exhibits greater precision. Data collected on the Attune exhibits lower coefficients of variation (%CV) and more distinct cell populations when compared to a high-end hydrodynamic focusing flow.
6 What is acoustic focusing cytometry? The Attune Acoustic Focusing Cytometer is the first that uses ultrasonic waves (over 2 MHz, similar to those used in medical imaging), rather than hydrodynamic forces, to position cells into a single focused line along the central axis of a capillary (Figure 9). Acoustic focusing is largely independent of the sample input rate, enabling cells to be tightly focused at the point of laser interrogation regardless of the sample-to- ratio (Figure ). This, in turn, allows slowed cell velocity to collect more photons for high-precision analysis at unprecedented volumetric sample throughput. The Attune accomplishes all this without high velocity or high volumetric fluid, which can damage cells. In addition, volumetric syringe pumps enable absolute cell counting without beads minimizing cost and sample preparation time. In contrast, s that use hydrodynamic focusing maintain the same sample speed at all flow rates, causing cells to lose focus as the sample core widens to increase flow rate. A. Acoustic focusing = better precision Acoustically focused sample stream Laser (cross-section) Acoustic focusing = OFF Sample is unfocused Acoustic focusing = ON Sample is focused Figure 9. Acoustic focusing in action. Fluorescent microspheres were applied to the capillary system of an acoustic focusing. Beads flow through randomly without any acoustic focusing (left). With the application of acoustic focusing, the beads are focused into a single line (right). No more tradeoffs Hydrodynamic focusing s have many limitations. As a result, users often have to sacrifice: Throughput for Features for ease of use Performance for price Power for footprint B. Traditional hydrodynamic focusing: compromised of data quality Hydrodynamic core device Laser (cross-section) Acoustic focusing technology keeps cells within a confined focal point, so these tradeoffs are not required. Figure. Acoustic focusing vs. traditional hydrodynamic focusing. (A) In acoustic focusing, cells remain in tight alignment even at higher sample rates. With this tight alignment, cells pass through the laser beam at its optimal focal point, resulting in less signal variation and improved data quality. (B) In traditional hydrodynamic focusing, increasing the sample rate results in widening of the sample core stream. The speed at which cells pass through the laser is not changed, and is determined by the speed of the fluid flow. Cells are distributed throughout the sample core stream because of reduced differential pressure between sample stream and stream, resulting in reduced cell focusing. Cells are not in tight alignment as they pass through the laser beam, resulting in increased signal variation and compromised data quality.
7 Ideal for both standard and specialized research applications Most standard applications for flow cytometry have been tested on the Attune, including: Live/dead cell discrimination Cell cycle analysis Cell proliferation assays Basic phenotyping (up to 6 colors) Rare-event detection Apoptosis Phagocytosis Detection of phosphoproteinsphagocytosis Other intracellular markers Many standard cell types have been tested, including: Jurkat cells HL60 promyoblast cells U266 myeloma cells, Mouse splenocytes Mouse blood HeLa human cervical carcinoma cells Bovine pulmonary artery epithelial (BPAE) cells 3T3 mouse embryo fibroblast cells Chinese hamster ovary (CHO) cells Human mesenchymal stem cells (hmsc) E. coli, S. aureus Human platelets Human whole blood Data have also been collected in specialized research areas such as oceanography and microbiology. Cells up to 0 μm in diameter have been run without clogging our 200 μm flow cell, with a minimum particle size specification of 1 μm; however, researchers are constantly pushing these limits with the Attune. Service and support The Attune Acoustic Focusing Cytometer is backed by Life Technologies worldwide technical support and service programs. We are focused on delivering personalized service from the time our sales representative walks through the door and throughout the life of your Attune instrument. The Attune is fully supported for one year with our extensive service plan, which includes: Comprehensive training Application and assay support Worldwide technical service Preventive maintenance Extended service plans are also available. For more information about available service plans, go to 1. Gratama JW, Menéndez P, Kraan J, Orfao A (2000) Loss of CD34(+) hematopoietic progenitor cells due to washing can be reduced by the use of fixative-free erythrocyte lysing reagents. J Immunol Methods 239:13 23.
8 Attune Acoustic Focusing Cytometer Software that performs to your specifications User-defined analysis, flexible setup Attune Cytometric Software is designed to provide powerful user-defined analysis using an intuitive interface for simplified experimental analysis (Figure 11). Templates can be built around specific applications and saved for consistent experimental design. Compensation is automated or user-defined and can be set up using a compensation guide. Utilities such as quick-save, drag-and-drop, and copy-and-paste provide rapid manipulation with commonly used functions. Experiments can be easily set up with automated settings that can be completely customized and saved for future experiments. No software licensing fees The Attune Cytometric Software can be downloaded without licensing fees. The software can be added to any desktop or laptop computer at your institution without any additional costs. Results can easily be analyzed at your convenience at your own computer, allowing the next user to run experiments on the Attune. Instrument controls Instrument status Control sample flow rate Select level Run startup/shutdown protocol System performance tracking by date Visualization tools Plot overlay or side-by-side graph comparison Standard or custom gate shapes Histograms, dot plots, density plots Customize statistics Copy and paste plots into reports (Microsoft Word or PowerPoint software) Analysis setup Save protocol templates Automated or user-defined compensation Raw data capture Data display in linear, log, or lin/lg scales Set universal compensation across multiple data sets File format compatibility Save data and images in PDF format for reports Data format in standard FCS 3.0 for compatibility with third-party analysis tools (i.e., FlowJo, ModFit LT, FCS Express software programs) Statistics in.csv format
9 Instrument performance tracking The Attune Cytometric Software provides automated baseline calculation and performance test functions with minimal user interactions. Performance tracking is facilitated through the reports function in the software. Instrument performance tracking is critical for collecting and analyzing accurate experimental data. The Attune Performance Tracking Beads are designed for use with the Attune Cytometric Software to automatically characterize, track, and report performance measurements of the Attune Acoustic Focusing Cytometer (Figure 12). The beads are used to define a baseline and conduct daily measurements of the. Vials contain enough beads for 0 measurements, and specific information for each lot is downloaded to the Attune Cytometric Software prior to use. Performance tracking for the Attune includes a comprehensive set of procedures to monitor the daily performance of the instrument. The performance tracking process involves: Running performance tracking beads Monitoring changes in the coefficient of variation and in the associated PMT voltage Tracking linearity of instrument performance Evaluating the detector and background over time Automatic setting of laser delay Instrument performance is affected by the lasers, optics, and fluidics of the instrument. For optimal instrument performance, it is critical to follow the manufacturer s recommended schedules for maintenance. Figure 11. Example of Attune Cytometric Software. The instrument control panel on the left side of the screen is used for experiment setup and control of laser selection, optics, and fluidics (flow rates). The main frame of this screen shows data analysis plots, while the right panel shows sub-file organization. Figure 12. Example of Levey-Jennings plot showing Attune performance over time. This report is generated using data from performance tracking beads.
10 NOTES
11
12 Ordering information Description Quantity Part number Attune Acoustic Focusing Cytometer (includes computer, monitor, startup solutions, installation, and warranty) Attune Focusing Fluids, 1X solution, 1 L Attune Focusing Fluids, X solution, 1 L Attune Focusing Fluids, 1X solution, 6 x 1 L Attune Wash Solution Attune X Shutdown Solution Attune Performance Tracking Beads ( x 6 beads/ml) Life Technologies offers a breadth of products DNA RNA PROTEIN CELL CULTURE INSTRUMENTS For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use. LIMITED USE LABEL LICENSE: No right to resell this product or any of its components is conveyed expressly, by implication, or by estoppel. For information on obtaining additional rights, please contact [email protected] or Out Licensing, Life Technologies, 791 Van Allen Way, Carlsbad, California Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation or their respective owners. Printed in the USA. CO
Boundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes
Immunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
CyAn : 11 Parameter Desktop Flow Cytometer
CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
Cell Cycle Tutorial. Contents
Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6
Cell Cycle Phase Determination Kit
Cell Cycle Phase Determination Kit Item No. 10009349 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 3 Safety
CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles
CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry About ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
Standardization, Calibration and Quality Control
Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic
Uses of Flow Cytometry
Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
LIVE/DEAD Fixable Dead Cell Stain Kits
USER GUIDE LIVE/DEAD Fixable Dead Cell Stain Kits Pub. No. MAN0002416 (MP34955) Rev. A.0 Table 1. Contents and storage Material Amount Storage Stability Individual Kits: Blue, violet, aqua, yellow-, green,
Introduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples
Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and
Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
APPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor
Notes on titering antibodies
Home Events This Week Bulletin Boards Lab Personnel Lab Policies Protocols MolBio Tetramers Proteins Flow Cytometry Immunoassays Instruments Computers Lab Publications Scientiific Literature Science on
These particles have something in common
These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative
Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
FlowSight. Flow cytometry with vision
FlowSight Flow cytometry with vision Flow cytometry with vision Introducing FlowSight Capable: Sensitive and flexible for every need Intuitive: Easy-to-use, with imagery for every cell Affordable: Designed
Introduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T
Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep
use. 3,5 Apoptosis Viability Titration Assays
Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,
Absolute counting using flow cytometry. Bright Count Microspheres
Absolute counting using flow cytometry Single-platform system for absolute counts I PRODUCT MANUAL [REF] IQP-391 [REF] IQP-392 s 50 tests s 100 tests [IVD] In Vitro Diagnostic medical device IQ Products
Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo ([email protected]) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
a Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper Violet-Excited nim-da Allows Efficient and Reproducible Cell Cycle Analysis on the Gallios Flow Cytometer Authors: Valdez, Ben 1. Carr, Karen 2. Norman, John
How to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping
CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
Annexin V Conjugates for Apoptosis Detection
Annexin V Conjugates for Apoptosis Detection Table 1. Spectral characteristics and storage information. Catalog no. Material Annexin V conjugate Amount Ex/Em (nm)* Storage Stability A23202 Alexa Fluor
MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
Below is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
Science is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
Standard Operating Procedure
1.0 Purpose: 1.1 The characterisation of of main leukocyte subsets in peripheral blood cells from mice by flow cytometry. Reliable values of frequencies of leukocyte clusters are very much dependent on
INSIDE THE BLACK BOX
FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben
BD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
ArC Amine Reactive Compensation Bead Kit
ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component
Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835
Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that
Real-time PCR: Understanding C t
APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence
EdU Flow Cytometry Kit. User Manual
User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg
Attune Acoustic Focusing Cytometer
USER GUIDE Attune Acoustic Focusing Cytometer USER GUIDE Publication Part Number 4453328 Rev. F Revision Date June 2012 Copyright 2012, Life Technologies Corporation. All rights reserved. For Research
Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1
78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction
Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size
Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences [email protected] What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements
Easy to use. Compact. Flexible. Sorting Chip
SH800Z Cell Sorter The SH800Z is a compact, affordable, and easy to use cell sorter developed using the state- of-the art technologies from Sony. An array of automated mechanisms makes it an ideal sorter
Instrument Characterization and Performance Tracking for Digital Flow Cytometers
Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic
UCHC Guide to CellQuest Pro. FACS Calibur instruments
UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn
CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection
CyTOF2 Mass cytometry system Unveil new cell types and function with high-parameter protein detection DISCOVER MORE. IMAGINE MORE. MASS CYTOMETRY. THE FUTURE OF CYTOMETRY TODAY. Mass cytometry resolves
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences
FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data Monisha Sundarrajan, PhD BD Biosciences 23 13795 00 Agenda Introduction Overview of BD CBA bead based immunoassays FCAP
Becton-Dickinson FACSCalibur Flow Cytometer Instructions
Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If
HIGH PERFORMANCE FLOW CYTOMETER FOR EVERYONE
POWERFUL. INTUITIVE. CUSTOMIZABLE. HIGH PERFORMANCE FLOW CYTOMETER FOR EVERYONE ABOUT ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose
Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:
UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a
FACSCanto RUO Special Order QUICK REFERENCE GUIDE
FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW
Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer
Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer Purpose This procedure gives instruction on minimal residual disease (MRD) detection in patients with
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)
Three Rules for Compensation Controls
From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for
Application Note. Cell Cycle and DNA Content Analysis Using the BD Cycletest Assay on the BD FACSVerse System. Summary
Cell Cycle and DNA Content Analysis Using the BD Cycletest Assay on the BD FACSVerse System Yibing Wang, Catherine McIntyre, and Dev Mittar BD Biosciences, San Jose Contents 1 Summary 2 Introduction 3
EDF Extended Depth of Field
EDF Extended Depth of Field An upgrade for the ImageStream system Think outside the dot. Break the classic depth of field barrier Improve precision Enhance discrimination Simplify analysis Increase resolution
PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.
PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based
BD FACSVerse. Simply brilliant
BD FACSVerse Simply brilliant BD FACSVerse: Simply Brilliant The BD FACSVerse flow cytometer was engineered from the ground up to offer remarkable performance, flexibility, and ease of operation for research
Navios Quick Reference
Navios Quick Reference Purpose: The purpose of this Quick Reference is to provide a simple step by step outline of the information needed to perform various tasks on the system. We begin with basic tasks
1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8
Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms
Let the BD FACSAria II accompany your best performance ever
Let the BD FACSAria II accompany your best performance ever Meet the new BD FACSAria II high-speed, fixed-alignment cuvette cell sorter. The first generation BD FACSAria system brought the complex world
Flow cytometry basics fluidics, optics, electronics...
Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence
Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility
Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection
LSR II Users Guide Stanford Shared FACS Facility 06/10/13
LSR II Users Guide Stanford Shared FACS Facility 06/10/13 A configuration sheet is located on the Website in the Instrument Overview area Do I need to run beads when I collect data using Diva Digital?
Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood
Application Note Primary Cell Analysis Enumeration and Viability of Nucleated Cells from Bone Marrow, Cord Blood, and Mobilized Peripheral Blood Cell Viability Introduction: Bone Marrow, Cord Blood, Stem
Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
BD CellQuest Pro Software Analysis Tutorial
BD CellQuest Pro Analysis Tutorial This tutorial guides you through an analysis example using BD CellQuest Pro software. If you are already familiar with BD CellQuest Pro software on Mac OS 9, refer to
Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers
An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,
NyONE - Cell imaging in a bird s eye view 4. NyONE...resolution matters! 8. Features & benefits 10. Fluorescence excitation channels 12
Envisions confirmed Content NyONE - Cell imaging in a bird s eye view 4 From cells to numbers 6 NyONE...resolution matters! 8 Features & benefits 10 Fluorescence excitation channels 12 Technical specifications
Analyzer Experiment setup guide for LSRII and Canto s
Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,
Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit
2 Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit Presented by Ellen Meinelt, MS MLS(ASCP) CM, Technical Applications Specialist and Calin Yuan, Product Course Developer,
Attune Acoustic Focusing Cytometer. Training Guide
Attune Acoustic Focusing Cytometer Training Guide For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use. Information in this document is subject to change without notice.
BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin
BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to
Cytell Cell Imaging System
Cytell Cell Imaging System Simplify your cell analysis and start changing tomorrow today www.gelifesciences.com/cytell The Cytell Cell Imaging System is an intuitive cell analysis instrument that streamlines
Attune NxT Acoustic Focusing Cytometer
USER GUIDE Attune NxT Acoustic Focusing Cytometer Catalog Number A24858, A24859, A24860, A24861, A24862, A24863, A24864 Publication Number 100024235 Revision B.0 For Research Use Only. Not for use in diagnostic
guava Flow Cytometry easycyte Systems
Data Sheet Data Sheet guava Flow Cytometry easycyte Systems Compact, Intuitive, and Affordable The guava easycyte flow cytometry systems are uncomplicated instruments that deliver complex cell analysis
UCHC Guide to FACS DiVa. LSR II instruments
UCHC Guide to FACS DiVa LSR II instruments Start Up: Turn on the instrument using the Green power button. Fluidics pump power, LSR II-B & C LSR II-B and LSRII-C-> Turn on the fluidics pump and ensure that
Compensation in Flow Cytometry
Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome
CytoSelect 48-Well Cell Adhesion Assay (ECM Array, Fluorometric Format)
Product Manual CytoSelect 48-Well Cell Adhesion Assay (ECM Array, Fluorometric Format) Catalog Number CBA-071 CBA-071-5 48 assays 5 x 48 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
protocol handbook 3D cell culture mimsys G hydrogel
handbook 3D cell culture mimsys G hydrogel supporting real discovery handbook Index 01 Cell encapsulation in hydrogels 02 Cell viability by MTS assay 03 Live/Dead assay to assess cell viability 04 Fluorescent
Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction
Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number
CFSE Cell Division Assay Kit
CFSE Cell Division Assay Kit Item No. 10009853 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions
LOAD & GO and make flow cytometry simplicity itself. the full spectrum of Flow Cytometry
LOAD & GO and make flow cytometry simplicity itself the full spectrum of Flow Cytometry Introducing Aquios CL...the first true LOAD & GO flow cytometer What would it mean for your laboratory s productivity
COMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract
arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents
Annexin V-EGFP Apoptosis Detection Kit
ab14153 Annexin V-EGFP Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in various samples This product is for research use only and is not intended
Introduction to Flow Cytometry:
Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning
Zecotek S Light Projection Network Marketing
White Paper Zecotek Visible Fiber Laser Platform Enabling the future of laser technology Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology company
