Inspection campaign summary report
|
|
|
- Steven Rice
- 9 years ago
- Views:
Transcription
1 Inspection Division Survey market inspection Unit TOPIC: Implantable defibrillation leads Inspection campaign summary report Abstract: An inspection campaign was conducted between November 2013 and June 2014, among defibrillation lead manufacturers by the Inspection Direction of ANSM. Six inspections were performed in three manufacturers and one distributor s site. The objective of this campaign was to ensure that the manufacturers take into account the regulatory requirements for CE marking in terms of manufacture and control, performances and safety demonstration of the defibrillation leads put on the French market. Inspected operators generally follow the process of putting on the market of these leads. However for one of the manufacturers, a high number of deviations was noted during the inspections of two of its sites. ANSM therefore enjoined the manufacturer to comply. Moreover, it should be noted the absence in this campaign of a major European manufacturer, because the Competent Authority of the country concerned declined the ANSM request for inspection of this manufacturer. Keywords: Inspection, medical device (MD), defibrillation lead (DL), EC marking, ethylene oxide sterilization. ANSM May Page 1 of 5
2 I. HEALTH STAKES ANSM exercises a special surveillance on medical devices at risk. It is in this context that fits the inspection campaign on the defibrillation leads (DL) as an implantable medical device (IMD). The objectives of these inspections are to verify the manufacturer s compliance to the applicable regulatory requirements for demonstrating the safety and performances of these devices and to ensure, if necessary, the implementation of the necessary corrective actions. The heart rhythm disorders are the causes of approximately 40,000 deaths per year in France, as sudden death for adult people resulting of an uncontrolled runaway of cardiac contractions. The defibrillation leads are connected to a pulse generator housing which is placed under the skin. The leads consist of sheathed electrical conductors and electrodes, they are introduced, intravenously, in the heart. Some recent models of defibrillators offer for leads subcutaneous implantation (such as housing) instead of cardiac implantation, they were not included in this campaign. II. REGULATORY REFERENCES The aim of these inspections was to assess the compliance with the provisions of the Directive 90/385/EEC of the European Parliament amended by the Directive 2007/47/EC and with the French Public Health Code and its implementing regulations concerning active implantable medical devices, particularly defibrillation leads put on the French market. III. CONDUCT OF THE CAMPAIGN OF INSPECTIONS 1. Objectives The inspection campaign on defibrillation leads was intended to check out the following points: manufacturing conditions considering the regulatory requirements; compliance and comprehensiveness of the EC mark regulatory file; regulatory obligations taken into account by the operators as responsible of the marketing including the management of claims and vigilance reports on medical devices. 2. Inspected operators The Inspection Division led the campaign between 2013 and early The inspected operators are or belong to large groups that have subsidiaries in several countries. Three foreign manufacturers (USA) and a French distributor were inspected. For North American manufacturers, inspections have taken place in two of the three North American headquarters and their three production sites in Puerto Rico. A total of six sites has been inspected during the campaign. The staff of the inspected sites is between 500 and 4300 employees. It should be noted in this summary, the absence of a major European group: BIOTRONIK, for which the German competent authority, responsible for the implementation of the regulatory requirements applicable to medical devices in its territory, has not responded favorably to the ANSM request to carry out an inspection in Germany for this campaign on the defibrillation leads. ANSM May Page 2 of 5
3 It was not identified marketing in France of other leads that those being the subject of this campaign. The list of operators inspected is in Appendix 1 3. Terms of the campaign The inspections were carried out by five inspectors and have been the subject of preliminary information to the operators. The inspections were lead according to a consistent and predefined methodology which includes the verification of the following items: the staff management conditions dedicated to the main activities with responsibility (organizational charts, functions forms, empowerment, delegation, training); the quality management system (documentation system, internal/external audits); the EC certificates for implantable medical devices placed on the market; the comprehensiveness of technical files, in particular on the preclinical and clinical studies, compliance with essential requirements, risks management, labeling and instructions for use of the devices; the conditions of manufacturing, control and batch release of products; the qualification of the special processes as the validation of sterilization with ethylene oxide; the production (hygiene, facilities, maintenance, storage, manufacturing, inspection, packaging; the management of non-conformities, claims, medical device vigilance as well as the process provided in case of product recalls. IV RESULTS FINDINGS This campaign has shown that the inspected manufacturers of defibrillation leads generally meet the requirements imposed by the EC marking, except one which brings almost 50% of the deviations on the overall campaign. Most operators have: made mandatory declarations and communications to ANSM, for the placing on the French market of their leads; a system of quality management sufficiently documented ; conditions of production and traceability satisfactory; management of non-conformities / claims and vigilance well suited. However, areas for improvement were identified during this campaign on: The formalization of qualification and validation processes for obtaining and maintaining sterility in particular on the following issues: implementation of procedures in particular on validation frequencies of sterilization with ethylene oxide ; existence of qualification plans for each site and for each EtO sterilization chamber ; validation of EtO re-sterilization on the MD performance and the packaging strength ; controlled atmosphere areas qualification with adequate particulate controls in terms of frequency and particle sizes. ANSM May Page 3 of 5
4 In terms of risk analysis, major deviations were on: lack of comprehensive risk analysis plan including risk reduction measures; acceptance of risks' level too high and/or an unsatisfactory approach for the calculation of the residual risk; Batch release management by the production unit. It was noted in particular the lack of independence between the production activities and batch release where the intervention of an independent qualified person is expected; Management of staff dedicated to the main activities of responsibility (organizational charts, functions forms, empowerment and delegation) must allow accurate description of the functions of a responsible person. The deviation distribution is presented in Appendix 2. An injunction was addressed to one operator following inspections of its two sites in May and June 2014 (Minnesota and Puerto Rico) involved in the production of defibrillation leads. The failures that led to this injunction are related to incomplete demonstration of products biocompatibility, to important gaps on validation processes of sterilization and desorption of ethylene oxide residues and parametric release methodology used by the company after sterilization. This injunction was published on the ANSM website. IV. CONCLUSION Defibrillation leads placed on the French market, with the exception of those covered by the injunction abovementioned are: globally in line with the regulation and with the essential requirements applicable to them; manufactured, packaged, sterilized, labeled and controlled in satisfactory conditions. Regarding the leads covered by the injunction, ANSM enjoined the manufacturer to carry out corrective actions, to comply with the regulation. ANSM May Page 4 of 5
5 Appendix 1 List of the inspected operators St JUDE MEDICAL MEDTRONIC (2 sites) BOSTON SCIENTIFIC (2 sites) SORIN CRM. Appendix 2 ANSM May Page 5 of 5
Frequently Asked Questions. Unannounced audits for manufacturers of CE-marked medical devices. 720 DM 0701-53a Rev 1 2014/10/02
Frequently Asked Questions Unannounced audits for manufacturers of CE-marked medical devices 720 DM 0701-53a Rev 1 2014/10/02 What is an unannounced audit?... 6 Are unannounced audits part of a new requirement?...
Company Quality Manual Document No. QM Rev 0. 0 John Rickey Initial Release. Controlled Copy Stamp. authorized signature
Far West Technology, Inc. ISO 9001 Quality Manual Document No.: QM Revision: 0 Issue Date: 27 August 1997 Approval Signatures President/CEO Executive Vice President Vice President/CFO Change Record Rev
Changes to Medical Device Regulations
Changes to Medical Device Regulations How these will affect distributors? IMSTA Breakfast Briefing, 24 October 2013 Niall MacAleenan Medical Device Lead, IMB. Slide 1 Content Revision of the medical device
GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES
Foreword GUIDANCE NOTES FOR MANUFACTURERS OF CLASS I MEDICAL DEVICES These guidance notes do not aim to be a definite interpretation of National Laws and/or regulations and are for guidance purpose only.
Konformitätsbewertung 3.9 B 17. Guidance for Notified Bodies auditing suppliers to medical device manufacturers
Antworten und Beschlüsse des EK-Med Konformitätsbewertung 3.9 B 17 Guidance for Notified Bodies auditing suppliers to medical device manufacturers Herkunft Notified Body Operations Group Quellen NBOG BPG
Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702
Quality Management System MANUAL SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Doc. No. G5500 Rev. 9.1 Status : APPROVED Effective: 12/11/2014 Page 2 of 23 Quality Manual Table of Contents
Quality Management System MANUAL. SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702
Quality Management System MANUAL SDIX, LLC Headquarters: 111 Pencader Drive Newark, Delaware 19702 Doc. No. G5500 Rev. 9.2 Status : APPROVED Effective: 9/25/2015 Page 2 of 23 Quality Manual Table of Contents
Revision Date Author Description of change. 10 07Jun13 Mark Benton Removed Admin. Manager from approval
Page 2 of 15 Document Revision History Revision Date Author Description of change 10 07Jun13 Mark Benton Removed Admin. Manager from approval 12Feb13 Mark Benton 08 01Oct12 Mark Benton 07 8/30/2012 Refer
Working Party on Control of Medicines and Inspections. Final Version of Annex 16 to the EU Guide to Good Manufacturing Practice
Version 8 (final) EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single market, regulatory environment, industries under vertical legislation Pharmaceuticals and cosmetics Brussels, July 2001 S\common\legal-legislation\75-319nd81-851\91-356\eudralexvol4\Annex
Medical Devices. Notified Bodies and the CE certification Process for Medical Devices. European Surgical Robotics Demonstration Day
Medical Devices Notified Bodies and the CE certification Process for Medical Devices European Surgical Robotics Demonstration Day 27 th March 2014, Leuven, Belgium 1 Bart Mersseman Product Manager Medical
Quality Management System Manual
Quality Management System Manual Assurance ISO / AS Manual Quality Management System EXCEEDING ALL EXPECTATIONS Since our inception in 1965, Swiss-Tech has supplied the medical, aerospace, hydraulic, electronic
DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL
L 218/82 EN Official Journal of the European Union 13.8.2008 DECISIONS ADOPTED JOINTLY BY THE EUROPEAN PARLIAMENT AND THE COUNCIL DECISION No 768/2008/EC OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of
ASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION
PHARMACEUTICAL INSPECTION CONVENTION PHARMACEUTICAL INSPECTION CO-OPERATION SCHEME PI 038-1 26 March 2012 AIDE-MEMOIRE ASSESSMENT OF QUALITY RISK MANAGEMENT IMPLEMENTATION PIC/S March 2012 Reproduction
CE Marking: Your Key to Entering the European Market
CE Marking: Your Key to Entering the European Market N. Gerard Zapiain U.S. Department of Commerce International Trade Administration Office of Microelectronics, Medical Equipment and Instrumentation I.
How To Inspect A Blood Bank
Site visit inspection report on compliance with HTA minimum standards Belfast Cord Blood Bank HTA licensing number 11077 Licensed for the procurement, processing, testing, storage, distribution and import/export
GUIDELINES FOR IMPORT AND MANUFACTURE OF MEDICAL DEVICES
GUIDELINES FOR IMPORT AND MANUFACTURE OF MEDICAL DEVICES The Ministry of Health and F.W. under Gazette notification S.O. 1468 (E) dated 6/10/2005 declared the following sterile devices to be considered
COMMISSION DIRECTIVE 2003/94/EC
L 262/22 COMMISSION DIRECTIVE 2003/94/EC of 8 October 2003 laying down the principles and guidelines of good manufacturing practice in respect of medicinal products for human use and investigational medicinal
Inspections, Compliance, Enforcement, and Criminal Investigations
Page 1 of 7 Home Inspections, Compliance, Enforcement, and Criminal Investigations Compliance Actions and Activities Warning Letters 2014 Inspections, Compliance, Enforcement, and Criminal Investigations
Med-Info. Introduction to Korean Medical Device regulations. TÜV SÜD Product Service GmbH
Med-Info International expert information for the Medical Device industry Introduction to Korean Medical Device regulations The Ministry of Health and Welfare (MHW) is the healthcare agency having overall
Gamma Sterilisation Validation according to ISO 11137 - Sterilising dose -
Synergy makes sence Gamma Sterilisation Validation according to ISO 11137 - Sterilising dose - MG-FSI72-105 Last revision: March 2011 5, Chemin du Catupolan - 69120 Vaulx en Velin - France - Tel. 33 (0)4
Adoption by GCP Inspectors Working Group for consultation 14 June 2011. End of consultation (deadline for comments) 15 February 2012
10 December 2013 EMA/INS/GCP/600788/2011 Compliance and Inspection Reflection paper on the use of interactive response technologies (interactive voice/web response systems) in clinical trials, with particular
Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR)
Update: Proposed Medical Device Regulation (MDR) & IVD Regulation (IVDR) Paul Brooks Healthcare Solutions Disclaimer The new regulations are not finalized and subject to change http://medicaldevices.bsigroup.com/en-us/resources/whitepapers-and-articles/
QUALITY MANAGEMENT SYSTEM REQUIREMENTS General Requirements. Documentation Requirements. General. Quality Manual. Control of Documents
Chapter j 38 Self Assessment 729 QUALITY MANAGEMENT SYSTEM REQUIREMENTS General Requirements 1. Establishing and implementing a documented quality management system 2. Implementing a documented quality
Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System
NBOG s Best Practice Guide applicable for AIMDD, MDD, and IVDD 2014-3 Guidance for manufacturers and Notified Bodies on reporting of Design Changes and Changes of the Quality System 1 Introduction The
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical Devices) Regulations, 1994
Guide for Custom-Made Dental Device Manufacturers on Compliance with European Communities (Medical SUR-G0014-1 12 NOVEMBER 2013 This guide does not purport to be an interpretation of law and/or regulations
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC
Guidance on the In Vitro Diagnostic Medical Devices Directive 98/79/EC August 2013 Contents 1 Introduction...3 2 Scope of the directive...3 2.1 What is an in vitro diagnostic medical device?... 3 2.2 Specimen
SUPPLIER QUALITY MANAGEMENT SYSTEM QUESTIONNAIRE
Company Name Street Address City, State, Zip code Phone Number Fax Company Website Email Address ORGANIZATION NAME PHONE NUMBER EMAIL ADDRESS President/CEO General Manager Engineering Manager Production
WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE
WHITEPAPER: SOFTWARE APPS AS MEDICAL DEVICES THE REGULATORY LANDSCAPE White paper produced by Maetrics For more information, please contact global sales +1 610 458 9312 +1 877 623 8742 [email protected]
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES
GUIDANCE NOTE FOR MANUFACTURERS OF CUSTOM-MADE MEDICAL DEVICES Foreword This guidance document is informative and advisory and has no legal authority. Individual national enforcement authorities are bound
New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01)
Safeguarding public health New Guidelines on Good Distribution Practice of Medicinal Products for Human Use (2013/C 68/01) Tony Orme, Senior GDP Inspector Inspection, Enforcement and Standards Division
Surgi Manufacturing Quality Manual
Surgi Manufacturing Page 1 of 18 Approvals: QA: Eng. Mgt. : A Date: 18Aug98 1. Introduction... 4 1.1 Scope... 4 1.2 Purpose... 4 1.3 Authority... 4 1.4 Issue of the Manual... 4 1.5 Amendments... 4 1.6
AVNET Logistics & EM Americas. Quality Manual
AVNET Logistics & EM Americas Quality Manual Avnet is committed to ensuring customer satisfaction while meeting all customer and applicable legal, statutory and regulatory requirements. This is accomplished
GUIDELINES ON MEDICAL DEVICES
EUROPEAN COMMISSION DIRECTORATE GENERAL for HEALTH and CONSUMERS Consumer Affairs Health technology and Cosmetics MEDDEV 2.12/2 rev2 January 2012 GUIDELINES ON MEDICAL DEVICES POST MARKET CLINICAL FOLLOW-UP
Business Management System Manual. Context, Scope and Responsibilities
Business Management System BMS Manual Page 1 of 11 Business Management System Manual Context, Scope and Responsibilities ISO 9001:2015 BMS.0100 R1 MAS Solutions LLC 29810 FM 1093 Suite C Fulshear, TX 77441
DEVICE RECALLS: The Era of Regulation and Outcome Metrics: Optimizing Benefits and Managing Risks
DEVICE RECALLS: The Era of Regulation and Outcome Metrics: Optimizing Benefits and Managing Risks Kenneth A. Ellenbogen, MD Kontos Professor & Chairman Virginia Commonwealth University School of Medicine
GUIDELINES FOR FOOD IMPORT CONTROL SYSTEMS
GUIDELINES FOR FOOD IMPORT CONTROL SYSTEMS SECTION 1 SCOPE CAC/GL 47-2003 1. This document provides a framework for the development and operation of an import control system to protect consumers and facilitate
ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan)
ISO 9001: 2008 Construction Quality Management System Sample - Selected pages (not a complete plan) Part 1: Project-Specific Quality Plan Part 2: Company Quality Manual Part 3: Submittal Forms Part 4:
Med-Info. Malaysia Medical Device Regulations. TÜV SÜD Product Service GmbH. International expert information for the medical device industry
Med-Info International expert information for the medical device industry Malaysia Medical Device Regulations Passed in 2012, the Medical Device Act (Act 737) and the Medical Device Authority Act 2012
DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL
29.3.2014 Official Journal of the European Union L 96/149 DIRECTIVE 2014/32/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 26 February 2014 on the harmonisation of the laws of the Member States relating
QUALITY OPERATING PROCEDURE QOP-00-02 QUALITY ASSURANCE SYSTEM MANUAL
PAGE 1 OF 40 QUALITY Note: The revision level of this document and any referenced documents must be verified for latest issue before use QOP-00-02 TITLE QUALITY ASSURANCE SYSTEM MANUAL APPROVALS DATE REVISIONS
TG 47-01. TRANSITIONAL GUIDELINES FOR ISO/IEC 17021-1:2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES
TRANSITIONAL GUIDELINES FOR ISO/IEC 17021-1:2015, ISO 9001:2015 and ISO 14001:2015 CERTIFICATION BODIES Approved By: Senior Manager: Mpho Phaloane Created By: Field Manager: John Ndalamo Date of Approval:
Reflection paper on the Use of Interactive Response Technologies (Interactive Voice/Web Response Systems) in Clinical Trials
1 2 3 5 August 2011 EMA/INS/GCP/600788/2011 Compliance and Inspection 4 5 6 7 Reflection paper on the Use of Interactive Response Technologies (Interactive Voice/Web Response Systems) in Clinical Trials
Specialties Manufacturing. Talladega Castings & Machine Co., Inc. ISO 9001:2008. Quality Manual
Specialties Manufacturing Talladega Castings & Machine Co., Inc. ISO 9001:2008 This document is the property of TMS and may not be reproduced, wholly, or in part, without the express consent of TMS. Rev.
Product Liability Risk Control Checklist
Product Liability Risk Control Checklist Appendix C1 This material will help you evaluate your products liability exposure. It is designed to enable you to focus on areas that need to be addressed to safeguard
The Risk Prevention Management System of the European Commission's staff
The Risk Prevention Management System of the European Commission's staff Calvo Alonso-Cortés, Antonio Directorate-General "Human Resources and Security" 200, Rue de la Loi, B-1049- Brussels, Belgium Phone:
How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters
environmental failure analysis & prevention health technology development How to Use the Design Process to Manage Risk: Elements of Design Controls and Why It Matters Kevin L. Ong, Ph.D., P.E. Managing
Qualified Persons in the Pharmaceutical Industry Code of Practice 2009, updated August 2015
Qualified Persons in the Pharmaceutical Industry Code of Practice 2009, updated August 2015 *QP Code of Practice 2008 updated Aug15 Page 1 of 13 Code of Practice for Qualified Persons 1. INTRODUCTION 2.
LinkTech LLC 3000 Bunsen Ave. #A Ventura, CA 93003
D LT-422-001 Page 1 of 12 QUALITY MANUAL LinkTech LLC 3000 Bunsen Ave. #A Ventura, CA 93003 Based on ISO 9001:2008 D LT-422-001 Page 2 of 12 INTRODUCTION: This Quality Manual establishes the quality system
PROPOSED DOCUMENT. Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change
AHWP/WG3/P001:2013 PROPOSED DOCUMENT Title: Quality management system Medical devices Nonconformity Grading System for Regulatory Purposes and Information Ex-change Author: AHWP Work Group 3 Date: 13 November
20-21 May 2015, Lisbon, Portugal. Highlights. Speakers:
Speakers: Dr Afshin Hosseiny Chair of the ECA Expert Working Group on GDP, Tabriz Consulting Comply with the new EU GDP Guideline Dr Martin Egger Pharmaserv Dr Daniel Müller GMP/GDP Inspector All participants
ISO 9001:2008 Audit Checklist
g GE Power & Water ISO 9001:2008 Audit Checklist Organization Auditor Date Page 1 Std. 4.1 General s a. Are processes identified b. Sequence & interaction of processes determined? c. Criteria for operation
Hong Kong Joint Venture Agreements
Hong Kong Exchanges and Clearing Limited and The Stock Exchange of Hong Kong Limited take no responsibility for the contents of this announcement, make no representation as to its accuracy or completeness,
RECALLS in EUROPE. Past, present, near & further future. Gert Bos BSI Medcon May 2012
RECALLS in EUROPE Past, present, near & further future Gert Bos BSI Medcon May 2012 A C P D Vigilance, vigilant: Continually watchful or on guard; always prepared for possible danger Content 2 Current
How companies leverage quality and quality certifications to achieve competitive advantage
How companies leverage quality and quality certifications to achieve competitive advantage Eize de Boer Systems & Services Certification International Business Development Manager Pharma Supply Chain for
Pre-market Information - Class III and IV
Pre-market Information - Class III and IV Life Sciences Council British Columbia NRC Industry Research Assistance Program Vancouver, October 29, 2007 Mary-Jane Bell, Ph.D. Medical Devices Bureau (MDB)
Safeguarding public health The Regulation of Software as a Medical Device
Safeguarding public health The Regulation of Software as a Medical Device Dr Head of New and Emerging Technologies Medicines and Healthcare Products Regulatory Agency (MHRA) Crown Copyright 2013 About
BASE CONSTRUCTION INC
BASE CONSTRUCTION INC 14252 CULVER DRIVE A-630 IRVINE,CA 92604-0326 OFFICE:949 387-3471 Shop 714 994 9563 CELL:949 735-5292 FAX:949 786-8753 QUALITY ASSURANCE and QUALITY CONTROL MANUAL 1 TABLE OF CONTENTS
Self-Audit Checklist
Page 1 Company Name: Date of audit: Date of last audit performed: Name of person performing self-audit: Signature: Name of person responsible for quality system: Signature: Number of non-compliances: Page
Manual for ITC's clients, 2013. Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC
Manual for ITC's clients, 2013 Conformity assessment of ACTIVE IMPLANTABLE MEDICAL DEVICES pursuant to Council Directive 90/385/EEC Institute for Testing and Certification, inc., Czech Republic, EU 1.
FMC Technologies Measurement Solutions Inc.
Table of Contents 1 Revision History... 4 2 Company... 5 2.1 Introduction... 5 2.2 Quality Policy... 6 2.3 Quality Organization... 6 2.4 Company Processes... 7 3 Scope and Exclusions... 8 3.1 Scope...
GOOD DOCUMENTATION AND QUALITY MANAGEMENT PRINCIPLES. Vimal Sachdeva Technical Officer (Inspector), WHO Prequalification of Medicines Programme
GOOD DOCUMENTATION AND QUALITY MANAGEMENT PRINCIPLES Vimal Sachdeva Technical Officer (Inspector), WHO Prequalification of Medicines Programme Contents 1. Why good documentation is essential? 2. What constitutes
Canada Organic Regime Quality Management System Manual
Canada Organic Regime Quality Management System Manual September 11, 2007 Page 1 COR Quality Management System 1.0 COO QUALITY MANAGEMENT SYSTEM MANUAL... 4 1.2 Overview of the Canada Organic Regime (COR)...
Standard Practice for Quality Control Systems for Nondestructive Testing Agencies 1
Designation: E 1212 99 AMERICAN SOCIETY FOR TESTING AND MATERIALS 100 Barr Harbor Dr., West Conshohocken, PA 19428 Reprinted from the Annual Book of ASTM Standards. Copyright ASTM Standard Practice for
PUBLICATIONS. Introduction
Page: 1 of 10 Introduction 1. The requirements of this document supplement the Factory Production Control requirements given in Table 1 of PN111 by giving the specific requirements for Factory Production
Engineering Procurement Construction Quality Plan
Engineering Procurement Construction Quality Plan Index 1 Introduction... 4 1.1 Project Background... 4 1.2 Document Purpose... 4 1.3 Change Control... 4 1.4 Contract... 4 1.5 Quality system... 4 1.6 Distribution...
ANNEX F. Organic agriculture (Art. 11) ARTICLE 1. Objectives
ANNEX F Organic agriculture (Art. 11) ARTICLE 1 Objectives Without prejudice to their obligations with regard to products not originating in the Member States or other legislative provisions in force,
DEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation
Digital Internal Use Only DEC STD 017-2 - ISO 9002 - Quality Systems - Model for Quality Assurance in Production and Installation DOCUMENT IDENTIFIER: A-DS-EL00017-02-0000 Rev A, ABSTRACT: This standard
Guidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC. GNB-CPD position paper from SG17 - EN 1090-1:2009
GNB-CPD SG17 Guidance from the Group of Notified Bodies for the Construction Products Directive 89/106/EEC NB-CPD/SG17/09/069 Issued: 7 September 2009 APPROVED GUIDANCE GNB-CPD position paper from SG17
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM
RTP s NUCLEAR QUALITY ASSURANCE PROGRAM RTP operates under one quality program, whether you purchase products that are commercial grade, nuclear safety-related or industrial safety compliant (IEC 61508).
US EPA REGION III QUALITY MANAGEMENT PLAN REVIEW CHECKLIST
US EPA REGION III Organization: EPA Organization: EPA Program: Contact: EPA Contact: EPA Contract/Grant/IAG Number: Phone Number: Phone Number: Reviewer: Date Reviewed: Reviewer Organization: Check New
MEAT GRADING AND CERTIFICATION BRANCH QUALITY MANUAL
United States Department of Agriculture Marketing and Regulatory Programs Agricultural Marketing Service Livestock and Seed Program Meat Grading & Certification Branch MGC Instruction 800 May 9, 2005 Page
Reserve Bank of Fiji Insurance Supervision Policy Statement No. 8 MINIMUM REQUIREMENTS FOR RISK MANAGEMENT FRAMEWORKS OF LICENSED INSURERS IN FIJI
Reserve Bank of Fiji Insurance Supervision Policy Statement No. 8 NOTICE TO INSURANCE COMPANIES LICENSED UNDER THE INSURANCE ACT 1998 MINIMUM REQUIREMENTS FOR RISK MANAGEMENT FRAMEWORKS OF LICENSED INSURERS
Keywords: inspection; nuclear material control and accountancy (NMC&A); quality management system
French Domestic Safeguards inspections with regard to quality management system of the operators, in the field of nuclear material control and accountancy Julie LASNEL-PAYAN, Flavien LEMOINE, Bruno AUTRUSSON,
General Regulations. Part Ⅱ - Quality Management System Rules ENGLISH VERSION 5.0
General Regulations Part Ⅱ - Quality Management System Rules ENGLISH VERSION 5.0 VALID FROM: 1 JULY 2015 OBLIGATORY FROM: 1 JULY 2016 TABLE OF CONTENTS 1. LEGALITY, ADMINISTRATION AND STRUCTURE... 3 1.1
Correspondence between ISO 13485:2003 and the US Quality System Regulation
Correspondence between ISO 13485:2003 and the US Quality System Regulation Correspondence between ISO 13485:2003 and the US Quality System Regulation 1 Scope 1.1 General This International Standard specifies
CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES
1.0 SCOPE CQR-1 CONTRACTOR QUALITY REQUIREMENTS for CONSTRUCTION SERVICES This document establishes the minimum quality program requirements for a contractor providing equipment, material, and construction
23. The quality management system
23. The quality management system Version 2.0 On this page: Mandatory requirements: Extracts from the HFE Act Extracts from licence conditions HFEA guidance: Definition of the quality management system
NEW AND GLOBAL + OLD APPROACH PRODUCT LEGISLATION: A. Standard questionnaire to be filled for each sector individually:
IV. NEW AND GLOBAL + OLD APPROACH PRODUCT LEGISLATION: A. Standard questionnaire to be filled for each sector individually: Sector: Noise emission in the environment by equipment for use outdoors Directive
Schweppes Australia Head Office Level 5, 111 Cecil Street South Melbourne Victoria 3205. www.schweppesaustralia.com.au
Schweppes Australia Head Office Level 5, 111 Cecil Street South Melbourne Victoria 3205 www.schweppesaustralia.com.au Quality Management Systems 1. Quality Management Systems develop, implement, verify
Table of Contents 1. SCOPE... 3 2. APPLICABLE DOCUMENTS... 4 3. TERMS AND DEFINITIONS... 4 4. QUALITY MANAGEMENT SYSTEM...4-8
Table of Contents 1. SCOPE... 3 2. APPLICABLE DOCUMENTS... 4 3. TERMS AND DEFINITIONS... 4 4. QUALITY MANAGEMENT SYSTEM...4-8 5. MANAGEMENT RESPONSIBILITY...8-9 6. RESOURCE MANAGEMENT... 10 7. PRODUCT
HACCP Compliance Audits: The Role of the Regulator
HACCP Compliance Audits: The Role of the Regulator Ahmed Al Kaabi HACCP for Catering Project Team HACCP Inspector Training Coordinator ADFCA Policy A planned process of transition from inspection to audit.
Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables
(1987) Guidelines for the Acceptance of Manufacturer's Quality Assurance Systems for Welding Consumables 1. General 1.1 Introduction 1.1.1 The present guidelines are to serve as a supplement to the IACS
Comparative analysis between the possible regulatory approaches to GMP compliance TITOLO PRESENTAZIONE
Comparative analysis between the possible regulatory approaches to GMP compliance TITOLO PRESENTAZIONE Dr. Fulvio CARLOTTI, GNOSIS SpA, Corporate QA Director September 26, 2014 Scope of GMP GMP compliance
SUPPLIER Form #: 007 AUDIT QUESTIONNAIRE
1.0 BACKGROUND INFORMATION Company Name: Company Address: Phone Number: Fax Number: Number of years in business: Number of employees: Annual Sales in Dollars: Size of Facility in Square Feet: Name of Senior
GUIDE TO IMPLEMENTING A REGULATORY FOOD SAFETY AUDITOR SYSTEM
GUIDE TO IMPLEMENTING A REGULATORY FOOD SAFETY AUDITOR SYSTEM FEBRUARY 2016 2 Contents Introduction... 4 Scope and objectives... 5 Scope... 5 Objectives... 5 Responsibilities... 5 The role of the licensee
Policy Briefing. Frequently Asked Questions on the Construction Products Regulation and CE marking. December 2012. Background
Frequently Asked Questions on the Construction Products Regulation and CE marking December 2012 Background The following list of frequently asked questions has been produced by an MPA Working Group and
ORACLE CONSULTING GROUP
ORACLE CONSULTING GROUP 9 Golder Ranch Rd., Ste. 1 Tucson, Arizona 9 Web Site: E-mail: 20-2-0 20-2-0 (FAX) CONSULTING MEMORANDUM QUALITY SYSTEM INSPECTION TECHNIQUE
QUALITY MANAGEMENT SYSTEM Corporate
Page 1 of 12 4 Quality Management System 4.1 General Requirements The Peerless Pump Quality Management System shall include: Documented statements of a quality policy and of quality objectives; A quality
CE mark now! Mandatory CE marking of construction products 1 July 2013
CE mark now! Mandatory CE marking of construction products 1 July 2013 Mandatory CE marking of construction products 1 July 2013 From 1 July 2013, construction products covered by a harmonised standard
End of consultation (deadline for comments) 14 October 2009. Adoption by Committee for advanced therapies 15 October 2010
15 October 2010 EMA/CAT/418458/2008/corr. Committee for advanced therapies (CAT) Procedural advice on the certification of quality and nonclinical data for small and medium sized enterprises developing
Registration of Class B Medical Devices
Registration of Class B Medical Devices 1 INTRODUCTION Medical devices are classified based on a rule based risk classification system into four risk classes Class A to D with Class A being the lowest
NABL NATIONAL ACCREDITATION
NABL 160 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDE for PREPARING A QUALITY MANUAL ISSUE NO. : 05 AMENDMENT NO : 00 ISSUE DATE: 27.06.2012 AMENDMENT DATE: -- Amendment
THE CONSTRUCTION PRODUCTS DIRECTIVE (COUNCIL DIRECTIVE 89/106/EEC)
THE CONSTRUCTION PRODUCTS DIRECTIVE (COUNCIL DIRECTIVE 89/106/EEC) Council Directive 89/106/EEC of 21 December 1988 on the approximation of laws, regulations and administrative provisions of the Member
Quality Management System Manual
Effective Date: 03/08/2011 Page: 1 of 17 Quality Management System Manual Thomas C. West Eric Weagle Stephen Oliver President ISO Management General Manager Representative Effective Date: 03/08/2011 Page:
2015 CANADIAN ELECTRICAL CODE UPDATE TRAINING PROVIDER PROGRAM Guidelines
2015 CANADIAN ELECTRICAL CODE UPDATE TRAINING PROVIDER PROGRAM Guidelines Under this program, CSA Group has developed a training program that provides detailed instruction on all major aspects of the CE
Technical Manufacturing Corporation (TMC) Quality Manual
Technical Manufacturing Corporation (TMC) Quality Manual 15 Centennial Drive Peabody MA 01960 Revision: 19 Pages: 2 of 15 Date Issued: 03/13/2013 1. Table of Contents 1. TABLE OF CONTENTS... 2 2. ABOUT