Correction of motion artifacts and scanning beam distortions in 3D ophthalmic optical coherence tomography imaging
|
|
|
- Bryan Owen
- 9 years ago
- Views:
Transcription
1 Correction of motion artifacts and scanning beam distortions in 3D ophthalmic optical coherence tomography imaging Robert J. Zawadzki* a, Alfred R. Fuller b, Stacey S. Choi a, David F. Wiley b,c, Bernd Hamann b,c and John S. Werner a a Vision Science and Advanced Retinal Imaging Laboratory (VSRI), Dept. of Ophthalmology & Vision Science, UC Davis, 4860 Y Street, Suite 2400, Sacramento, CA, USA 95817; b Visualization and Computer Graphics Research Group, Institute for Data Analysis and Visualization (IDAV), UC Davis, One Shields Avenue, Davis, CA 95616; c Stratovan Corporation, Woodland, CA ABSTRACT The ability to obtain true three-dimensional (3D) morphology of the retinal structures is essential for future clinical and experimental studies. It becomes especially critical if the measurements acquired with different instruments need to be compared, or precise volumetric data are needed for monitoring and treatment of retinal disease. On the other hand, it is well understood that optical coherence tomography (OCT) images are distorted by several factors. Only limited work has been performed to eliminate these problems in ophthalmic retinal imaging, perhaps because they are less evident in the more common 2D representation mode of time-domain OCT. With recent progress in imaging speed of Fourier domain - OCT (Fd-OCT) techniques, however, 3D OCT imaging is more frequently being used, thereby exposing problems that have been ignored previously. In this paper we propose possible solutions to minimize and compensate for artifacts caused by subject eye and head motion, and distortions caused by the geometry of the scanning optics. The first is corrected by cross-correlation based B-scan registration techniques; the second is corrected by incorporating the geometry of the scanning beam into custom volume rendering software. Retinal volumes of optical nerve head (ONH) and foveal regions of healthy volunteer, with and without corrections, are presented. Finally, some common factors that may lead to increased distortions of the ophthalmic OCT image such as refractive error or position of the subject s head are discussed. Keywords: ( ) Optical coherence tomography; ( ) Imaging system; ( ) Ophthalmology; ( ) Medical optics instrumentation 1. INTRODUCTION Due to the latest developments in OCT [1-2], including its Fourier-domain extension [3-5] and commercialization, it has become an increasingly important tool for in vivo imaging of the retina in clinical settings. Acquisition speed of current FD-OCT systems allows measurements of volumetric retinal structures within 2 to 10 seconds, depending on the system s speed and lateral sampling density. Therefore volumetric representation of the retina, as in other medical areas, is becoming an increasingly important tool for ophthalmic clinical diagnosis and research. In this context it is important to remember that optical coherence tomography (OCT) images are distorted by several factors including geometry of the OCT scanning beam as well as head and eye motion. Only limited work has evaluated appropriate distortion correction [6-7]. In this paper we propose methods allowing correction of both motion and scanning beam distortions resulting in corrected volume of the retina. Scanning beam distortion correction algorithm presented in this paper is a part of our custom volumetric visualization software and is designed to take into account the actual geometry of the scanning optics of the OCT system used. To make it possible we defined 5 characteristic parameters that are specific for each OCT system, therefore allowing corrections of deformations introduced by this instrument. *[email protected]; phone: ; fax: ; web: Ophthalmic Technologies XVII, edited by Fabrice Manns, Per G. Soederberg, Arthur Ho, Bruce E. Stuck, Michael Belkin, Proc. of SPIE Vol. 6426, , (2007) /07/$18 doi: / Proc. of SPIE Vol
2 The results of this correction are immediately visualized by our custom 3D volume rendering software. We think that this true three-dimensional (3D) morphology of the retina may find its application in temporal monitoring of retinal disease as well as for comparisons across OCT platforms. 2. MATERIALS AND METHODS The volumetric retinal data presented in this paper have been acquired by our clinical Fd-OCT system developed at UC Davis with collaboration with Duke University. More than 200 subjects, with healthy and diseased eyes, have been imaged with this system. This paper focuses on the correction of image distortion, thus only two representative retinal volumes acquired for the same subject over 6x6mm centered on foveal and ONH regions will be presented. Both volumes have passed our standard test of qualifying retinal volumes for volumetric reconstruction. This test includes estimating volume distortions by viewing movies showing all consecutive B-scans acquired in the volume as well as C- scans reconstructed from OCT images Experimental system The experimental Fd-OCT system used for acquiring 3D volumes is shown in Figure 1, and described in detail elsewhere [8]. For experiments described here, we used a superluminescent diode (855@75nm) from Superlum allowing 4.5 µm axial resolution of the retina (n=1.38). Measured power at the subject s eye was equal to 700 µw. With our current spectrometer design, the maximum axial range (seen after the Fourier transform) is 2.7 mm in free space and around 2 mm in the eye. I Sample Arm I x.-y scanners I Reference ArmI Achromatic Achromatic lens lens Fiber Polarization Mirror Detector Diffraction Grating Fiber Objective Faraday Isolator Polarization Controler ilight Deliveryl CCD line Fig. 1: Clinical Fd-OCT system at UC Davis. The reference arm matches the path length of the sample arm. Volumetric data presented in this paper have been acquired over 6x6 mm lateral range using 200 horizontal B-scans (500 A-scans/B-scan). Thus, volumes consist of 100,000 A-scans acquired in 5.5 s with 50 µs CCD exposure time. Software from Bioptigen, Inc. was used to acquire and display the data as well as for driving both horizontal and vertical scanners. This software permits real-time display (at the acquisition speed) of the imaged retinal structures and saving of the last acquired volume (200 B-scans). Figure 2 shows the relative lateral locations of two retinal volumes used in this paper superimposed on a fundus image. Proc. of SPIE Vol
3 Fig. 2: The OCT scanning areas of 3D volumetric measurements used in this paper superimposed on a fundus photo. To visualize and analyze our 3D OCT data, we post-process raw OCT data and save them as consecutive B-scans in tiff file format. This is represented on a figure 3. Fast Axis B-scan Slow Axis B-scan (Reconstructed frame) Slow 4 Axis k I I I I I I t r I Fast Axis B-scan (Original frame) Fig. 3: The OCT scanning pattern used in volumetric acqusition (left) and its volumetric representation with virtual cutting planes (right). 'I In our standard volumetric imaging mode Fast Axis corresponds to horizontal scanner and Slow Axis corresponds to vertical scanner. This is later used as the input for our auto registration software as well as for our volumetric visualization software that has built -in scanning beam distortions correction algorithm. Respectively this volume can be later virtually cut along and plane allowing reconstruction of the virtual Slow Axis B-scans and C-scans Motion Artifacts In Fd-OCT, head and eye motion can result in B-scans distortions as well as decreases in image intensity (due to the fringe wash out). In this paper we will concentrate only on image distortions only Head and Eye motion In general, motion artifacts are minimized with faster image acquisition speed. Conversely, the amount of motion artifact, even though it strongly depends on the imaging subjects, increases with increasing acquisition time. In the ideal case, very fast volume acquisition speed (less then 100 ms) would allow virtually motion-free imaging. This however is still not possible in current ophthalmic OCT instruments unless the number of A-scans /Volume is reduced, but this will severely limit volumetric detail. Volumetric retinal imaging thus requires a compromise between lateral sampling density and total acquisition time. As with other ophthalmic imaging instruments, use of a chin and forehead rest or bite bar with forehead rest helps to reduce head motions and therefore lower axial motion artifacts. A fixation target (point) also helps the subject to focus and reduce voluntary eye motion artifacts. Unfortunately, these methods cannot completely remove motion artifacts and additional correction is necessary. In this paper we will simplify our motion artifact model to three degrees of freedom only. Namely we will assume that retinal image can be translated axially (along the z-axis) and laterally (along x- and y-axes). Figure 4 illustrates motion artifact distortions. Proc. of SPIE Vol
4 z x y volumetric sampling grid eye motion distorted retinal data set of OCT instrument Fig. 4. Schematic of the eye and head motion caused artifact. Due to eye and head motion, the retina is sampled with a distorted sampling grid. Blue arrows indicate degrees of freedom used to correct for displacements between consecutive B-scans. Red arrow represents direction where no correction is performed Motion artifact correction The initial motion artifact correction method presented in this paper is relatively simple and is often used in the OCT community. Here only distortions between consecutive B-scans are corrected, thus no correction on a single B-scan is performed. This method is useful for correcting slow fixation drift of the eye and some head motion on a time scale longer than a single B-scan acquisition time. Thus in this method very fast eye motions are not corrected and can be still seen on single B-scans. To correct for axial and lateral distortions between consecutive B-scans, we use an autocorrelation method that shifts (alignes) stack of frames. This was implemented with ImageJ [9] with the Turboreg plugin developed in P. Thévenaz, laboratory at Biomedical Imaging Group, Swiss Federal Institute of Technology Lausanne [10]. In their algorithm each frame is used as the template with respect to which the next frame is aligned, so that the alignment proceeds by propagation. After completion of the registration process, the original stack of frames is destroyed and replaced by the result of the registration. During testing of this algorithm we found out that distortion of the stacks with translation only bring best results. Thus upon translation, a straight line on a B-scan is mapped to a straight line of identical orientation, with conservation of the distance between any pair of points. A single landmark in each image gives a complete description of a translation. Thus each B-scan stays not distorted and we get set of B-scans corrected along Z-axis (depth) and X-axis (fast lateral axis). We find this method quite successful if amount of motion artifact is moderate. To estimate results of this correction one can view virtual C-scan or slow Axis B-scan. If no disruption can be observed on both, then motion artifact correction was successful and data can be used for further image processing Elimination of the axial flattening artifact in autocorrelation methods It is important to point out that implementation of the B-scan correlation method may result is some loss of information about retinal structure along the slow-axis. Thus to retrieve this information one can use additional B-scans acquired along the slow axis (with vertical scanners scanning fast) and use this as a reference retinal profile for axial shifts of the consecutive B-scans to achieve an unflattened profile along the slow axis. Figure 5 shows a schematic of this operation. relative location of the Distorted retinal Correlation- based slow axis corrected volumetric volume and reference data set corrected retinal data set B-scan data along slow axis slow axis B-scan Fig. 5. Schematic of the slow-axis axial flattening artifact removal. Separate reference slow-axis B-scan is used to correct for axial flattening along the slow axis. Proc. of SPIE Vol
5 In the actual implementation of this method one needs to locate the reference slow-axis frame on the virtually reconstructed slow axis B-scans (from the volume). Then, the axial position of the A-scans from the reconstructed B- scans should be correlated with the A-scan on the reference frame. As a result, the set of axial displacements of each A- scan on the reconstructed B-scan is found. This information can be used to shift axially fast axis B-scans allowing removal of the flattening artifact Scanning system distortion As with ultrasound imaging, distortions from scanning in ophthalmic OCT data acquisition (in polar coordinates), need to be taken into account for undistorted data display. Because two separate scanners are used in volumetric image acquisition, two separate image distortion corrections in two coordinate systems must be performed, first along the slow axis planes and then along the fast axis. The correction methodology presented in this paper is similar to the one proposed by Podelanu et al. [7] Geometry of the scanning system and the eye To model and later correct distortion created by the OCT scanning system, one needs to use a model of the OCT scanning head and the imaged eye. Similarly to most commercial and laboratory OCT ophthalmic systems, we used an OCT sample/scanning head consisting of a fiber collimator, two scanners (for vertical and horizontal scanning) and two lenses that image scanners onto the eye s pupil. Additional correction may be required if trial lenses are used to correct a subject s refractive error (defocus). Figure 6 presents a schematic of a standard OCT scanning system. Ey L 1 Vertical - scanner Collimator L 2 Horizontal - scanner f 1 f 1 f 2 f 2 Eye L 1 L 2 Collimator Fig. 6. Schematic of the OCT scanning head geometry. Top: 3D view simulated in ZEMAX; Bottom: 2D view of the elements; L 1 - Lens 1; L 2 - Lens 2; f 1 - focal length of Lens 1; f 2 - focal length of Lens 2; In this standard configuration, rays are scanned from the central pivot point originating in the eye pupil. The exact position of this pivot point depends on the location of the eye with respect to the lens, L1. Due to the separation between vertical and horizontal scanners, their pivot points are separated as well, creating references for two perpendicular polar coordinate systems, one for vertical planes and one for horizontal planes. Figure 7 is a schematic of OCT scanning geometry and sampling grid density on the OCT B-scans defined by angular steps of the scanners ( Φ) and depth sampling ( r) for both horizontal and vertical planes. Proc. of SPIE Vol
6 Φ V Φ H L V,H Φ V,H r L V L H Fig. 7. Scanning geometry and sampling grid of the OCT retinal system; Left: schematic of the ray s path inside subject s eye for vertical scanning plane. Center: schematic of the ray s path inside subject s eye for horizontal scanning plane. L V,H distance between pivot point and retina for vertical (V) and horizontal (H) scanning pattern. Right: 2D sampling grid of the OCT system with the axial sampling ( r) and radial sampling ( Φ) for vertical (V) and horizontal (H) scanning pattern. The separation between vertical and horizontal scanners in our OCT scanning head transfer into a 4 mm separation between vertical plane and horizontal plane pivot positions. Driver signals of our scanning system have been corrected for separation of vertical and horizontal pivot points to produce undistorted lateral patterns on the retina Correction of the scanning system distortions To properly display OCT scans on a computer screen (that relies on evenly sampled data) we need to resample our data from polar to Cartesian coordinates. In our method new equidistant points are used with intensity found by linear interpolation based on the neighbor s intensity value. We treat our original data set a three dimensional matrix S ijk with the i encoding number of the A-scan along fast axis, j encoding number of A-scan along slow axis and k encoding depth location of the point. S = S, Θ, Θ ) (1) ijk ( r ijk H ijk V ijk where: i Fast axis (X); I - size j Slow Axis (Y); J size k Depth (Z); K size The next step uses the scanning system geometry to find locations for each point in the matrix S. This is performed in two steps. First we find points located on the-fast-axis plane and then resample it to equidistant coordinate system. S ijk Fast Axis plane ijk H ijk J j=0 Interpolation = S( r, Θ ) S (2) xz Fast Axis plane where: J rijk = LH + k r j= 0 Θ H ijk I = + i Θ H 2 j=0 J Proc. of SPIE Vol
7 Figure 8 shows a graphical representation of this process on a single fast-axis plane. This operation is repeated for each of the fast axis B-scan planes. Fig. 8. Schematic of the resampling OCT data from radial to Cartesian coordinates. Left: orientation of the original (black) and resampled (grey) Field of View, Right: locations of the data points on the radial coordinates (black) and Cartesian coordinates (interpolated points that will be displayed) The next step on this already corrected along fast-axis data set is to perform a similar operation along the slow-axis plane and then the data are again resampled into equidistant Cartesian coordinate system. This operation is performed on each slow-axis B-scan plane S ijk Slow Axis plane = S xz Fast Axis plane ijk V ijk I i=0 Interpolation ( r, Θ ) S (3) xyz where: I rijk = LV + k z i= 0 Θ H ijk J = + j Θ V 2 i=0 I Thus, by simply knowing five specific parameters of the OCT system ( r; Φ V ; Φ H ; L V ; L H ), one can have a nondistorted representation of the data set in a Cartesian coordinate system I xyz. The figure below shows a schematic of our scanning distortion correction algorithm. Fast axis distorsions correction Slow axis distorsions correction S ijk S xz Fast axis plane S xyz Fig. 9. Schematic of the scanning beam distortion correction. Left: original data set. Center: data set corrected for Horizontal scanner distortion; Right: data set corrected for Vertical scanner distortion. 3. RESULTS We tested our method for reducing image distortions using two data sets acquired on the same subject s fovea and optic nerve head (ONH). Proc. of SPIE Vol
8 3.1. Imaging artifacts Before presenting corrected results, we want to point out that the axial and lateral positioning of the eye pupil in the front of the imaging lens plays a critical role in achieving good correction of the volumetric data. Any displacement of the eye along x- or y-axes will result in tilting of the retinal structures on the OCT scans. Axial displacement of the eye will result in the bending of the retinal structures seen with OCT. - 8 mm axial shift Center + 8 mm axial shift Fig. 10. Visualization of the retinal distortions caused by axial displacement of the pupil with respect to the scanning beam pivot. Top: schematic of the eye position; Bottom corresponding B-scans acquired over 12 mm. Left column represents 8 mm axial displacement toward the imaging lens; central row represents reference displacement used in our scanning beam distortion algorithm. Right: 8 mm axial displacement backwards from the imaging lens Correct positioning of the eye s pupil is, therefore, critical for the correction method presented in this paper as it assumes central location of the pupil. This is very important because the distance between two imaging lenses is very often used to correct for subject refractive error (defocus). Thus, if one wants to keep the eye at the correct position, alteration of reference arm length is necessary. Alternatively, changes in the scanners displacement would be necessary to keep proper lateral dimensions of the scanning area on the retina. If the pupil is located off axis, a more complex distortion correction model is necessary to take into account and correct for pupil s lateral distortions Motion artifact correction To illustrate the effect of eye/head motion and scanning beam distortion on the volumetric data we visualized those data without any corrections. Figure 11 shows data as acquired. Note that the axial motion artifact manifesting itself in wavy structures seen on the virtual C-scans. H Fig.11. Visualization of raw volume acquired over ONH region (left) and fovea (right). Proc. of SPIE Vol
9 Figure 12 presents the same data after motion artifact correction. One can see that most of the wavy structures previously seen on the C-scans are now removed. Fig. 12. Visualization of Volume acquired over ONH region (left) and fovea (right) corrected for lateral and axial motion artifacts. 3.3 Scanning beam distortion correction In the next step, motion artifact-corrected data are used as an input for our two step scanning beam distortion algorithm. As noted, we used five parameters to perform this transformation. The axial spacing of the retinal data is r = mm; the angular separation of the A-scans on vertical planes is equal to Φ V = rad/line; the angular separation of the A-scans on horizontal planes is equal to Φ H = rad/line; distance between vertical pivot pint and top of the data set is equal L V = 16.5 mm, respectively; distance between horizontal pivot point and top of the data set is equal L H = 20.5 mm. XZ - Axis ZY - Axis V V V Fig. 13. Visualization of corrected volumes (along fast- and slow-axis planes) acquired over ONH region Proc. of SPIE Vol
10 XZ - Axis ZY - Axis 7 Fig. 14. Visualization of corrected volumes (along fast- and slow-axis planes) acquired over fovea. Characteristic circular shapes seen on the reconstructed C-scans indicate the spherical shape of the retina after implementation of the scanning beam distortion correction. 4. CONCLUSIONS A method to correct for eye and head motion and scanning beam distortion artifacts is presented. Corrected volumes show spherically shaped retinae. These results demonstrate our method in qualitative correction of retinal volumes. In order to have precisely corrected data, more accurate measurements of the correction parameters are needed. Our method could be verified with independent data acquired with other imaging techniques free of artifacts present on OCT scans. A high-resolution magnetic resonant imaging system will be used to test our method for scanning beam distortion correction. ACKNOLEDGEMENTS We gratefully acknowledge the contributions of Joseph A. Izatt, Mingtao Zhao, Bradley A. Bower and Marinko Sarunic from the Department of Biomedical Engineering, Duke University Durham, NC. The help of Bioptigen Inc. for providing OCT data acquisition software is appreciated. This research was supported by the National Eye Institute (EY ) REFERENCES 1. D. Huang, E. A. Swanson, C.P. lin, J.S. Schuman, W.g. Stinson, W. Chang, M.R. Flotte, K. Gregory, C.A. Puliafito, Optical Coherence Tomography, Science 254, (1991) 2. A. F. Fercher, C.K. Hitzenberger, G. Kamp, Y. Elzaiat, Measurment of intraocular distances by backscattering spectral interferometry, Opt. Commun. 117, (1995) 3. M. Wojtkowski, T. Bajraszewski, P. Targowski and A. Kowalczyk Real time in vivo imaging by high-speed spectral optical coherence tomography, Opt. Lett. 28, (2003) Proc. of SPIE Vol
11 4. R. Leitgeb, C.K. Hitzenberger, and A.F. Fercher Performance of fourier domain vs. time domain optical coherence tomography, Opt. Express 11, (2003) 5. N.A. Nassif, B. Cense, B.H. Park, M.C. Pierce, S.H. Yun, B.E. Bouma, G.J. Tearney, t.c. Chen, J,F. de Boer In vivo high-resolution video-rate spectral-domain optical coherence tomography of the human retina and optic nerve, Opt. Express 12, (2004) 6. V. Westphal, A.M. Rollins and S. Radhakrishnan J.A. Izatt Correction of geometric and refractive image distorsions in optical coherence tomography applying Fermat s principle, Opt. Express 10, (2002) 7. A. Podoleanu, I. Charalambous, L. Plesea, A. Dogariu and R. Rosen Correction of distortions in optical coherence tomography imaging of the eye, Phys. Med. Biol. 49, (2004) 8. R. Zawadzki, S. Jones, S. Olivier, M. Zhao, B. Bower, J. Izatt, S. Choi, S. Laut, and J. Werner, "Adaptiveoptics optical coherence tomography for high-resolution and high-speed 3D retinal in vivo imaging," Opt. Express 13, (2005) Proc. of SPIE Vol
The Nurse/Technician Role Within the Emerging Ophthalmic Technology - OCTs/B-Scan
The Nurse/Technician Role Within the Emerging Ophthalmic Technology - OCTs/B-Scan Margie V. Wilson, COMT Chief Clinical Supervisor UCSD Shiley Eye Center Thanks, Carol! Unfortunately. I have not financial
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus
Understanding astigmatism Spring 2003
MAS450/854 Understanding astigmatism Spring 2003 March 9th 2003 Introduction Spherical lens with no astigmatism Crossed cylindrical lenses with astigmatism Horizontal focus Vertical focus Plane of sharpest
Optic Nerve Imaging. Management of Glaucoma
Optic Nerve Imaging in the Diagnosis and Management of Glaucoma Scott D. Smith, MD, MPH Associate Professor of Clinical Ophthalmology Director, Glaucoma Service Columbia University, New York, NY Glaucoma
Shape Measurement of a Sewer Pipe. Using a Mobile Robot with Computer Vision
International Journal of Advanced Robotic Systems ARTICLE Shape Measurement of a Sewer Pipe Using a Mobile Robot with Computer Vision Regular Paper Kikuhito Kawasue 1,* and Takayuki Komatsu 1 1 Department
DICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
Optical Coherence Tomography OCT. 3D Imaging in Medical Technology and Quality Control
Optical Coherence Tomography OCT 3D Imaging in Medical Technology and Quality Control SLN Seminar, EPMT Lausanne, 26. May. 2011 Ch. Meier www.optolab.ch 1 / 27 SLN/EPMT, Lausanne, 26.5.2011 Ch. Meier Outline
Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD
2008 AGI-Information Management Consultants May be used for personal purporses only or by libraries associated to dandelon.com network. Optical Metrology Third Edition Kjell J. Gasvik Spectra Vision AS,
Anamorphic imaging with three mirrors: a survey
Anamorphic imaging with three mirrors: a survey Joseph M. Howard Optics Branch (Code 551), NASA Goddard Space Flight Center, Greenbelt, MD 20771 Ph: 301-286-0690 Fax: 301-286-0204 [email protected]
Polarization of Light
Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
PHYS 39a Lab 3: Microscope Optics
PHYS 39a Lab 3: Microscope Optics Trevor Kafka December 15, 2014 Abstract In this lab task, we sought to use critical illumination and Köhler illumination techniques to view the image of a 1000 lines-per-inch
WAVELENGTH OF LIGHT - DIFFRACTION GRATING
PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant
Experimental and modeling studies of imaging with curvilinear electronic eye cameras
Experimental and modeling studies of imaging with curvilinear electronic eye cameras Viktor Malyarchuk, 1 Inhwa Jung, 1 John A. Rogers, 1,* Gunchul Shin, 2 and Jeong Sook Ha 2 1 Department of Materials
How To Understand The Current Of An Optic Nerve Fiber Layer
A Practical Guide for Journal Interpretation of Current of Optical Glaucoma Coherence Practice, Tomography January-April Retinal 2009;3(1):9-13 Nerve Fiber Layer Measurement A Practical Guide for Interpretation
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
High-resolution imaging of the human retina with a Fourier deconvolution technique
D. Catlin and C. Dainty Vol. 19, No. 8/August 2002/J. Opt. Soc. Am. A 1515 High-resolution imaging of the human retina with a Fourier deconvolution technique David Catlin and Christopher Dainty The Blackett
Modern Classical Optics
Modern Classical Optics GEOFFREY BROOKER Department of Physics University of Oxford OXPORD UNIVERSITY PRESS Contents 1 Electromagnetism and basic optics 1 1.1 Introduction 1 1.2 The Maxwell equations 1
Michalina Góra, Jan Marczak, Antoni Rycyk, Piotr Targowski APPLICATION OF OPTICAL COHERENCE TOMOGRAPHY TO MONITORING OF LASER ABLATION OF VARNISH
Michalina Góra, Jan Marczak, Antoni Rycyk, Piotr Targowski APPLICATION OF OPTICAL COHERENCE TOMOGRAPHY TO MONITORING OF LASER ABLATION OF VARNISH Institute of Physics, Nicolaus Copernicus University, Toruń,
SafetyScan Lens. User Manual
SafetyScan Lens User Manual Contents Introduction 2 Lens selection 3 Beam quality parameter Using the tables Use with lasers whose M 2 is 1.2 4 Use with lasers whose M 2 is 3.0 5 Use with lasers whose
DETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY. Rainer Boerret, Dominik Wiedemann, Andreas Kelm
URN (Paper): urn:nbn:de:gbv:ilm1-2014iwk-199:0 58 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 08 12 September 2014 URN: urn:nbn:de:gbv:ilm1-2014iwk:3 DETECTION OF SUBSURFACE DAMAGE
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
Scanning Acoustic Microscopy Training
Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
Optical Coherence Tomography OCT
Optical Coherence Tomography OCT Optische 3D Tomographie SLN Seminar 2. Dez. 2010 1 / 37 SLN Seminar, Buchs 2.12.2010 Ch. Meier Outline 1. Introduction 2. History and technological Basics Time Domain TD-OCT,
Nederland België / Belgique
Cronus Brochure 2 Introduction Cronus: The first spectrometer and colorimeter in one The Cronus is world s first spectrocolorimeter combining a high en VIS spectrometer with a high speed XYZ colorimeter.
RAY TRACING UNIFIED FIELD TRACING
RAY TRACING Start to investigate the performance of your optical system using 3D ray distributions, dot diagrams of ray positions and directions, and optical path length. GEOMETRIC FIELD TRACING Switch
Grazing incidence wavefront sensing and verification of X-ray optics performance
Grazing incidence wavefront sensing and verification of X-ray optics performance Timo T. Saha, Scott Rohrbach, and William W. Zhang, NASA Goddard Space Flight Center, Greenbelt, Md 20771 Evaluation of
5-Axis Test-Piece Influence of Machining Position
5-Axis Test-Piece Influence of Machining Position Michael Gebhardt, Wolfgang Knapp, Konrad Wegener Institute of Machine Tools and Manufacturing (IWF), Swiss Federal Institute of Technology (ETH), Zurich,
Encoded Phased Array Bridge Pin Inspection
Encoded Phased Array Bridge Pin Inspection James S. Doyle Baker Testing Services, Inc. 22 Reservoir Park Dr. Rockland, MA 02370 (781) 871-4458; fax (781) 871-0123; e-mail [email protected] Product
Opto-Mechanical I/F for ANSYS
Abstract Opto-Mechanical I/F for ANSYS Victor Genberg, Keith Doyle, Gregory Michels Sigmadyne, Inc., 803 West Ave, Rochester, NY 14611 [email protected] Thermal and structural output from ANSYS is
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement
MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters
Application Report: Running µshape TM on a VF-20 Interferometer
: Running µshape TM on a VF-20 Interferometer General This report describes how a fiber interferometer from Arden Photonics Ltd was used together with the µshape TM Generic software package. The VF-20
Comparison to other anterior segment imaging devices
found helpful in a number of areas of ophthalmology. However, in the field of anterior segment tumors its usefulness is confined to a subset of tumors that are confined to the iris, hypo pigmented, and
Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected].
Medical Image Processing on the GPU Past, Present and Future Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected] Outline Motivation why do we need GPUs? Past - how was GPU programming
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAYS AND TUNNEL LININGS. HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAY AND ROAD TUNNEL LININGS.
HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAYS AND TUNNEL LININGS. HIGH-PERFORMANCE INSPECTION VEHICLE FOR RAILWAY AND ROAD TUNNEL LININGS. The vehicle developed by Euroconsult and Pavemetrics and described
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami
The Limits of Human Vision
The Limits of Human Vision Michael F. Deering Sun Microsystems ABSTRACT A model of the perception s of the human visual system is presented, resulting in an estimate of approximately 15 million variable
Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda
Slit Skiascopic-guided Ablation Using the Nidek Laser Scott MacRae, MD; Masanao Fujieda ABSTRACT PURPOSE: To present the approach of using a scanning slit refractometer (the ARK 10000) in conjunction with
X-ray thin-film measurement techniques
Technical articles X-ray thin-film measurement techniques II. Out-of-plane diffraction measurements Toru Mitsunaga* 1. Introduction A thin-film sample is two-dimensionally formed on the surface of a substrate,
Renishaw 2008. apply innovation TM. Calibrating 5-axis machines to improve part accuracy. 5Align
Calibrating 5-axis machines to improve part accuracy 5Align Productive Process Pyramid TM Understanding and tracking machine behaviour Process verification Thermal compensation In-cycle process control
Introduction to Optics
Second Edition Introduction to Optics FRANK L. PEDROTTI, S.J. Marquette University Milwaukee, Wisconsin Vatican Radio, Rome LENO S. PEDROTTI Center for Occupational Research and Development Waco, Texas
A System for Capturing High Resolution Images
A System for Capturing High Resolution Images G.Voyatzis, G.Angelopoulos, A.Bors and I.Pitas Department of Informatics University of Thessaloniki BOX 451, 54006 Thessaloniki GREECE e-mail: [email protected]
How To Fuse A Point Cloud With A Laser And Image Data From A Pointcloud
REAL TIME 3D FUSION OF IMAGERY AND MOBILE LIDAR Paul Mrstik, Vice President Technology Kresimir Kusevic, R&D Engineer Terrapoint Inc. 140-1 Antares Dr. Ottawa, Ontario K2E 8C4 Canada [email protected]
Trends in Ophthalmology a Market Perspective. Matthias Karl, Michael Kempe
Trends in Ophthalmology a Market Perspective Matthias Karl, Michael Kempe 30.04.2013 Agenda 1 2 3 4 Role of Optical Technologies in Ophthalmology Demands of the Market - Trends in the Application Implications
Color holographic 3D display unit with aperture field division
Color holographic 3D display unit with aperture field division Weronika Zaperty, Tomasz Kozacki, Malgorzata Kujawinska, Grzegorz Finke Photonics Engineering Division, Faculty of Mechatronics Warsaw University
Lenses and Apertures of A TEM
Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics
Integration of a passive micro-mechanical infrared sensor package with a commercial smartphone camera system
1 Integration of a passive micro-mechanical infrared sensor package with a commercial smartphone camera system Nathan Eigenfeld Abstract This report presents an integration plan for a passive micro-mechanical
Cone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound
High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard
Analyzing LASIK Optical Data Using Zernike Functions
MATLAB Digest Analyzing LASIK Optical Data Using Zernike Functions By Paul Fricker Researchers in fields as diverse as optometry, astronomy, and photonics face a common challenge: how to accurately measure
TAGARNO AS Sandøvej 4 8700 Horsens Denmark Tel: +45 7625 1111 Mail: [email protected]
8 TAGARNO AS Sandøvej 4 8700 Horsens Denmark Tel: +45 7625 1111 Mail: [email protected] TAGARNO 2 Quick Image Capture Split Screen Function Magnification up to 320x Easy Height Adjustment SD-card ESD Security
Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm
Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and
Introduction to acoustic imaging
Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3
Geometric Camera Parameters
Geometric Camera Parameters What assumptions have we made so far? -All equations we have derived for far are written in the camera reference frames. -These equations are valid only when: () all distances
Realization of a UV fisheye hyperspectral camera
Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical
THREE DIMENSIONAL GEOMETRY
Chapter 8 THREE DIMENSIONAL GEOMETRY 8.1 Introduction In this chapter we present a vector algebra approach to three dimensional geometry. The aim is to present standard properties of lines and planes,
MRC High Resolution. MR-compatible digital HD video camera. User manual
MRC High Resolution MR-compatible digital HD video camera User manual page 1 of 12 Contents 1. Intended use...2 2. System components...3 3. Video camera and lens...4 4. Interface...4 5. Installation...5
Adaptive Optics Phoropters
Adaptive Optics Phoropters Scot S. Olivier Adaptive Optics Group Leader Physics and Advanced Technologies Lawrence Livermore National Laboratory Associate Director NSF Center for Adaptive Optics Adaptive
Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated testing of square and round bars
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Phased-Array ROWA-SPA: High-performance testing machine for combined, 100-percent automated
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors Rudolph J. Guttosch Foveon, Inc. Santa Clara, CA Abstract The reproduction
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Optical Design using Fresnel Lenses
Optical Design using Fresnel Lenses Basic principles and some practical examples Arthur Davis and Frank Kühnlenz Reflexite Optical Solutions Business Abstract The fresnel lens can be used in a wide variety
How Reed Switches are used with a Permanent Magnet
How Reed Switches are used with a Permanent Magnet Using Reed Switches in a sensing environment, one generally uses a magnet for actuation. It is important to understand this interaction clearly for proper
Dev eloping a General Postprocessor for Multi-Axis CNC Milling Centers
57 Dev eloping a General Postprocessor for Multi-Axis CNC Milling Centers Mihir Adivarekar 1 and Frank Liou 1 1 Missouri University of Science and Technology, [email protected] ABSTRACT Most of the current
Experiment 3 Lenses and Images
Experiment 3 Lenses and Images Who shall teach thee, unless it be thine own eyes? Euripides (480?-406? BC) OBJECTIVES To examine the nature and location of images formed by es. THEORY Lenses are frequently
A Simple Fiber Optic displacement Sensor for Measurement of Light Intensity with the Displacement
A Simple Fiber Optic displacement Sensor for Measurement of Light Intensity with the Displacement Trilochan patra Assistant professor, Department of Electronics and Communication Engineering, Techno India
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Wavelength stabilized high-power diode laser modules
Wavelength stabilized high-power diode laser modules Bernd Köhler *, Thomas Brand, Matthias Haag, Jens Biesenbach DILAS Diodenlaser GmbH, Galileo-Galilei-Str. 10, 55129 Mainz-Hechtsheim, Germany ABSTRACT
COMP175: Computer Graphics. Lecture 1 Introduction and Display Technologies
COMP175: Computer Graphics Lecture 1 Introduction and Display Technologies Course mechanics Number: COMP 175-01, Fall 2009 Meetings: TR 1:30-2:45pm Instructor: Sara Su ([email protected]) TA: Matt Menke
Intuitive Navigation in an Enormous Virtual Environment
/ International Conference on Artificial Reality and Tele-Existence 98 Intuitive Navigation in an Enormous Virtual Environment Yoshifumi Kitamura Shinji Fukatsu Toshihiro Masaki Fumio Kishino Graduate
Three-dimensional vision using structured light applied to quality control in production line
Three-dimensional vision using structured light applied to quality control in production line L.-S. Bieri and J. Jacot Ecole Polytechnique Federale de Lausanne, STI-IPR-LPM, Lausanne, Switzerland ABSTRACT
Integrated sensors for robotic laser welding
Proceedings of the Third International WLT-Conference on Lasers in Manufacturing 2005,Munich, June 2005 Integrated sensors for robotic laser welding D. Iakovou *, R.G.K.M Aarts, J. Meijer University of
Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging.
Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging. Ira O. Wygant a, Mustafa Karaman b, Ömer Oralkana, and Butrus T. Khuri-Yakub
Geometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The
A DATA ACQUISITION, PROCESSING AND STORAGE SYSTEM FOR AN OPHTHALMIC INSTRUMENT: FOTOBIOFTAL-1
Journal of Optoelectronics and Advanced Materials Vol. 7, No. 3, June 2005, p. 1651-1655 A DATA ACQUISITION, PROCESSING AND STORAGE SYSTEM FOR AN OPHTHALMIC INSTRUMENT: FOTOBIOFTAL-1 S. Miclos, M. Mustata,
p h a s e d - a r r a y t e c h n o l o g i e s w w w. m 2 m - n d t. c o m M2M today c o n t a c t @ m 2 m - n d t. c o m
M2M today Overview M2M was created in November 2003 after a 3-year R&D program aiming for the development of an innovative ultrasonic phased array device M2M develops, produces and sells Ultrasonic systems
Recommended alignment procedure for Shaper 6_6 / Focal- Shaper 9
Recommended alignment procedure for Shaper 6_6 / Focal- Shaper 9 The described below procedure presents an example of alignment of beam shapers Shaper and Focal- Shaper (F- Shaper) with using the standard
Laser expander design of highly efficient Blu-ray disc pickup head
Laser expander design of highly efficient Blu-ray disc pickup head Wen-Shing Sun, 1,* Kun-Di Liu, 1 Jui-Wen Pan, 1 Chuen-Lin Tien, 2 and Min-Sheng Hsieh 1 1 Department of Optics and Photonics, National
32-Channel Head Coil Imaging at 3T
32-Channel Head Coil Imaging at 3T Thomas Benner Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA
BIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
Projection Center Calibration for a Co-located Projector Camera System
Projection Center Calibration for a Co-located Camera System Toshiyuki Amano Department of Computer and Communication Science Faculty of Systems Engineering, Wakayama University Sakaedani 930, Wakayama,
Microlenses immersed in nematic liquid crystal with electrically. controllable focal length
Microlenses immersed in nematic liquid crystal with electrically controllable focal length L.G.Commander, S.E. Day, C.H. Chia and D.R.Selviah Dept of Electronic and Electrical Engineering, University College
A Simple Guide To Understanding 3D Scanning Technologies
A Simple Guide To Understanding 3D Scanning Technologies First Edition www.lmi3d.com Table of Contents Introduction At LMI Technologies, solving complex problems in a simple way is the philosophy that
OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction
American Academy of Ophthalmology Annual Meeting Orlando, Florida, 21-25 October 2011 OCT-based IOL power calculation for eyes with previous myopic and hyperopic laser vision correction Weeks Professor
A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA
A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - [email protected]
CASE HISTORY #2. APPLICATION: Piping Movement Survey using Permalign Laser Measurement System
CASE HISTORY #2 APPLICATION: Piping Movement Survey using Permalign Laser Measurement System EQUIPMENT: Dresser-Clark Hot Gas Expander (Turbine), 60-inch Inlet Flange HISTORY: Piping support modifications
TABLE OF CONTENTS. INTRODUCTION... 5 Advance Concrete... 5 Where to find information?... 6 INSTALLATION... 7 STARTING ADVANCE CONCRETE...
Starting Guide TABLE OF CONTENTS INTRODUCTION... 5 Advance Concrete... 5 Where to find information?... 6 INSTALLATION... 7 STARTING ADVANCE CONCRETE... 7 ADVANCE CONCRETE USER INTERFACE... 7 Other important
Fast Varifocal Lenses Based on KTa 1-x Nb x O 3 (KTN) Single Crystals
: Crystal Technologies Fast Varifocal Lenses Based on KTa 1-x Nb x O 3 () Single Crystals Tadayuki Imai, Shogo Yagi, Seiji Toyoda, and Masahiro Sasaura Abstract NTT Photonics Laboratories has recently
High-Performance Wavelength-Locked Diode Lasers
Copyright 29 Society of Photo-Optical Instrumentation Engineers. This paper was published in the proceedings of the SPIE Photonics West 29, Vol. 7198-38 (29), High-Power Diode Laser Technology and High-Performance
Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging
Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging Shaun S. Gleason, Ph.D. 1, James Goddard, Ph.D. 1, Michael Paulus, Ph.D. 1, Ryan Kerekes, B.S.
Twelve. Figure 12.1: 3D Curved MPR Viewer Window
Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from
Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images
FORMATEX 2007 A. Méndez-Vilas and J. Díaz (Eds.) Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images Jörg Piper Clinic Meduna, Department for Internal
Wir schaffen Wissen heute für morgen
Diffractive optics for photon beam diagnostics at hard XFELs Wir schaffen Wissen heute für morgen PSI: SLAC: ESRF: SOLEIL: APS: SACLA: EuroXFEL C. David, S. Rutishauser, P. Karvinen, Y. Kayser, U. Flechsig,
http://dx.doi.org/10.1117/12.906346
Stephanie Fullerton ; Keith Bennett ; Eiji Toda and Teruo Takahashi "Camera simulation engine enables efficient system optimization for super-resolution imaging", Proc. SPIE 8228, Single Molecule Spectroscopy