Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display
|
|
|
- Helen Lloyd
- 9 years ago
- Views:
Transcription
1 Consistency of Softcopy and Hardcopy: Preliminary Experiences with the new DICOM Extensions for Image Display Marco Eichelberg a, Jörg Riesmeier a, Klaus Kleber b, Jörg Holstein b, Herman Oosterwijk c, Peter Jensch d a OFFIS e.v., D Oldenburg, Germany b Institute for MicroTherapy, D Bochum, Germany c OTech Inc., Crossroads, TX 76227, USA d Carl von Ossietzky Universität, D Oldenburg, Germany ABSTRACT The DICOM standard defines in detail how medical images can be transmitted and stored. However, there have been no precise rules on how to interpret the parameters contained in a DICOM image which deal with the image presentation. As a result, the same image frequently looks different when displayed on different workstations or printed on a film from various printers. Three new DICOM extensions attempt to close this gap by defining a comprehensive model for the display of images on softcopy and hardcopy devices: Grayscale Standard Display Function, Grayscale Softcopy Presentation State and Presentation Look Up Table. A prototype implementation of these extensions has been shown at the 1999 annual tradeshow of the Radiological Society of North America (RSNA) as part of the scientific exhibit (inforad). This demonstrated a simulated radiological workflow in which images were created, interpreted at a diagnostic workstation and later reviewed on a clinical workstation. Images could also be printed using DICOM Print. The prototype shows a proof of concept, i. e. that image integrity and consistency over a variety of display and print devices can be achieved and in addition, that the new DICOM extensions can be implemented relatively easily, without a significant performance penalty. The extensions allow to store all parameters defining how an image is displayed or printed in a separate DICOM object that can be managed with the existing DICOM database services. In particular, this satisfies the users need to view images at different locations in a consistent manner, and to document the image appearance on which a diagnosis is made in softcopy environments. Keywords: DICOM, Softcopy, Hardcopy, Display Consistency 1. INTRODUCTION The DICOM (Digital Imaging and Communications in Medicine) standard [1] has been quite successful in establishing a vendor independent platform that allows the creation of complex image management networks with different types of modalities, archives, printers and information systems. DICOM started in 1993 with basic services for image management such as transmission, archival and printing of medical images. Since then, the standard was supplemented with a multitude of enhancements defining new modalities (e. g. full-field digital X-ray) and new services (e. g. Basic Worklist Management, Modality Performed Procedure Step). One area that was never addressed by the DICOM standard in the past, however, is the problem of display consistency. Even though it might be possible to exchange a particular medical image between two display stations, there was no guarantee that both systems displayed the same image, even if all display settings (e. g. window level and width) were identical. Even worse, when the image is printed to a hardcopy printer, it is quite likely that the hardcopy looks different than the softcopy display. In installations with only a small number of different viewing station and printer types, this limitation might be acceptable because many systems can be adjusted with different kinds of correction curves to match the users preferences on image appearance. However, this requires a separate adjustment for each combination of modality and display device and, therefore, becomes an increasingly complex and time consuming task with the growing number of DICOM compliant devices in a network. Recognizing this problem, the DICOM committee has developed three supplements to the standard which address issues of display consistency for both softcopy and hardcopy displays (e. g. viewing software and printers): Grayscale Standard Display Function [2], Grayscale Softcopy Presentation State Storage [3], and Presentation Look Up Table [4].
2 The supplements on the Grayscale Standard Display Function and Presentation LUT (Look-Up Table) have been part of the standard for a while (they are included in the 1998 edition of the standard), but have not been implemented widely in the past. The supplement on Presentation States has been finalized only recently (September 1999). 2. GRAYSCALE STANDARD DISPLAY FUNCTION The human perceptual response to different levels of brightness is not linear a fact that we have to deal with in the medical imaging area, where the luminance range typically varies around cd/m 2 (candela per square meter) for conventional film-screen systems. The eye is much more sensitive to brightness changes in the bright areas of an image than in the dark areas. The DICOM Grayscale Standard Display Function (GSDF) describes this property of human vision in mathematical terms, based on Barten s model of the human visual system [5]. The GSDF uses the notion of Just Noticeable Differences (JNDs), which are defined as the luminance difference of a given target under given viewing conditions that the average human observer can just perceive, and derives a set of 1023 discrete JNDs (for a standard target and well-defined viewing conditions) that fall into the luminance range from 0.05 to about 4000 cd/m 2 which is covered by the model. The JNDs allow to determine an upper boundary for the number of shades of gray that can be visualized on a display system with a given luminance for black and white. They also allow to describe the characteristic curve of a display system that matches the sensitivity of the human eye to changes in brightness, i. e. a perceptually linearized display system. Figure 1 shows the formula from [2] which defines the Grayscale Standard Display Function as an interpolation of the luminance curve over the complete JND range (with constants a..h, k, m as defined in [2]). a + c Ln( j) + e ( Ln( j)) 2 + g ( Ln( j)) 3 + m ( Ln( j)) 4 log 10 L( j) = 1+ b Ln( j) + d ( Ln( j)) 2 + f ( Ln( j)) 3 + h ( Ln( j)) 4 + k ( Ln( j)) 5 Figure 1: DICOM Grayscale Standard Display Function L(j): Luminance as a function of the JND index The GSDF is intended as a standard curve against which different types of display devices can be calibrated, such that the calibrated image display uses the available contrast of the display device in a perceptually linear way (i. e. the difference between black and 5% gray is perceived equal to the difference between white and 95% gray). A calibration of different types of display devices (e. g. screen/film and grayscale monitor) according to the GSDF cannot yield an identical image display if the physical properties of the display devices in terms of spatial and contrast resolution differ. However, the calibration can yield a consistent image display which means that the appearance of the image to the human observer is as similar as possible given the differing properties of the display devices. The DICOM GSDF model can be applied to a wide range of display devices, including CRT (cathode ray tube) and flatpanel monitors (which are considered as a display system together with the graphics adapter to which they are attached), reflective hardcopy printers and transmissive hardcopy printers. All of these systems have in common that they allow a discrete number of shades of gray to be addressed (typically 256 for monitors and for transmissive hardcopy printers). The index number used to address the shades of gray is called Digital Driving Level (DDL) in DICOM. The calibration of a display system requires that the characteristic curve of the display system is measured, taking the reflection caused by ambient light into account. For softcopy displays, the luminance for each DDL is measured. For hardcopy printers the optical density D for each DDL is measured instead. A luminance can be computed from the optical density using the formula from [2] shown in Figure 2, where L a is the luminance contribution due to reflection of ambient light (defined to be 0 for reflective hardcopy) and L 0 is the luminance of the light box (for transmissive hardcopy) or the maximum luminance obtainable from diffuse reflection of the illumination (for reflective hardcopy). L = La + L0 10 D Figure 2: Relationship between Optical Density and Luminance Figure 3 shows the measured characteristic curve of a particular display device (the PC on the author s desk) together with the part of the DICOM GSDF which falls into the same luminance range. A correction function can be derived from the two curves in a straightforward way: For each pixel value that may occur in the image to be displayed, the corresponding luminance according to the GSDF (i. e. the luminance of a perceptually linearized display using the same luminance range as the real display system) is determined. Then a DDL of the real display system is selected such that it matches the GSDF luminance as close as possible. The resulting transformation can be implemented efficiently as a simple look-up table.
3 100 Luminance (cd/m²) 10 1 CC GSDF 0, DDL (Digital Driving Level) Figure 3: DICOM Grayscale Standard Display Function (GSDF) versus measured Characteristic Curve (CC) DICOM does not mandate a particular implementation of the correction function, and in fact, different approaches are possible. Most graphics adapters used with high resolution / high brightness grayscale monitors for softcopy reading of digital radiography implement the correction look-up table as part of the adapter hardware. Systems using conventional color monitors and graphics boards can implement the image correction as part of the display software. It should be noted that the image correction can be implemented such that the source image data (which typically contains a contrast range of more than 8 bits for modalities like CT or CR) is directly mapped to DDLs. This eliminates a prior reduction of the source image data contrast range which could lead to the loss of shades of gray (and, thus, a loss of information caused by the software based display calibration). Figure 4: Grayscale Test Patterns (Philips DSI Pattern, SMPTE Pattern) The result of the calibration can be visualized with test patterns such as the DSI pattern from Philips Medical Systems or the SMPTE pattern from the Society of Motion Picture and Television Engineers, as shown in Figure 4. The low contrast fields (i. e. the 5% and 95% fields on the SMPTE pattern and the 2% contrast fields on the left and right of the DSI pattern) should be visible. The steps from black to white should appear equidistant and, when measured, should match the GSDF for the available contrast range of the display device.
4 3. GRAYSCALE SOFTCOPY PRESENTATION STATE Unlike conventional images which can be viewed without any additional tool (except a lightbox in an environment with appropriate lighting), digital medical images consist of binary data which must be processed and presented to the human eye on a display device such as a printer or monitor. In particular, the mapping of the raw binary image data to the display usually requires the application of image processing algorithms in order to produce a meaningful image. A well-known example is the adjustment of window level and width, but there are others, for example the mapping of vendor specific data to a standardized domain (e. g. Hounsfield units for CT) or the mapping of data acquired in optical density linear space (e. g. scanned film) to image data meaningful for softcopy display. Therefore, DICOM allows to store a number of image processing parameters in the header of each DICOM image, together with other medically relevant information about the acquisition (patient demographic data, radiation dose, contrast media, date and time of acquisition etc.) However, practice has shown that this approach has its limitations: In many PACS configurations, all modalities send their images to the archive after acquisition. The archive permanently stores the images and forwards them to the workstation scheduled for the reading of the images. Since the study is already archived, adjustments performed on the images during diagnosis (e. g. window level and width) cannot be stored unless the complete study is duplicated in the archive. Although DICOM defines fields in which parameters like Window Center (level) and Window Width can be stored, the corresponding algorithms were never precisely defined (i. e. DICOM never really defined how a Window Center and Window Width should be interpreted). Not surprisingly, this ambiguity has lead to differing implementations. It is possible today to have two different display stations which are perfectly DICOM compatible but still display the images in a different manner, even when the same image is shown with the same processing parameters. The Grayscale Softcopy Presentation State Storage Supplement [3] addresses these two problems. First of all, it precisely defines the different grayscale transformations that must be applied to the raw DICOM image data in order to yield a consistent image display. The upper part of Figure 5 shows these grayscale transformations. Rescale Slope/Intercept or Modality LUT Window/Level or VOI LUT Presentation LUT Grayscale Transformations Original Image Modality LUT Transformation Mask (Subtraction) VOI LUT Transformation Presentation LUT Transformation P-Values Shutter Transformation Image Annotation Spatial Transformation Displayed Area Annotation Display Shutter, Annotation and Spatial Transformations Figure 5: DICOM Grayscale Image Transformation Model The first step in the display chain is the Modality LUT Transformation, which maps device or vendor dependent data into a vendor-independent space. This can either be a linear mapping of the form f(x) = m*x+b where m is called rescale slope and b is called rescale intercept, or it can be a look-up table (e. g. mapping optical densities from a scanned film into a space linear to perceived intensity). The second step, mask subtraction, is used in angiographic studies, where images might be subtracted from a reference image functioning as a mask. The third step, the Value of Interest (VOI) LUT Transformation, is DICOM s notion of the window level and width adjustment. Figure 6 shows the precise (and surprisingly complex) definition of the formula which maps an arbitrary image pixel value x into the output interval [f min, f max ], depending on the Window Center c and the Window Width w, which must be 1. As an alternative to window level and width, a look-up table can also be used to specify a Value of Interest transformation.
5 f(x, c, w) = f min, f max, ( x c w 1 2 x c 1 2 w 1 2 x > c w 1 2 ) (f max f min )+f min otherwise Figure 6: DICOM Formula for Window Level/Width Calculation The last DICOM grayscale transformation is called Presentation LUT Transformation. This is a rather new addition which was introduced with the 1998 edition of the standard. A Presentation LUT allows to encode non-linear image transformations for specific applications or user preferences, for example a gamma correction. A good example might be the display of orthopedic images which might need more contrast to clearly show potential fractures. The output space of the Presentation LUT is called Presentation Values (P-Values) and by definition approximately related to human perceptual response. P- Values do not depend on the characteristics of a particular display device and are used as the input for a standardized (calibrated) display device such as a workstation or printer conforming to the DICOM Grayscale Standard Display Function. In addition to a precise definition of DICOM s grayscale image transformation model, the Grayscale Softcopy Presentation State Storage Supplement creates a new DICOM Information Object Definition (IOD) called Presentation State. This object contains an extensive set of parameters defining how a particular image or set of images is to be presented (displayed) to the user. A Presentation State only contains references to the images it applies to and, therefore, does not duplicate the image data. Presentation State objects are relatively small (typically only few Kbytes) and can be stored and transmitted with a minimal resource increase. Presentation State objects fit well into the established DICOM information model (they are just a separate DICOM series within the study containing the images) and can be transmitted, stored and retrieved with the existing DICOM storage and query/retrieve services, requiring few changes in existing systems. It is possible to have a single Presentation State for a complete series of images or to have different Presentation States ( views ) of the same image (see Figure 7). Presentation State applying to multiple images Presentation State Presentation State 1 tip position Presentation State 2 Multiple Presentation States applying to a single image Figure 7: Different Mappings between Presentation States and Images A Presentation State contains all the grayscale transformations described before (all of these transformations are optional, though, and can be replaced by an identity transform). In addition, a Presentation State can contain graphical annotation and a number of spatial transformations (see lower part of Figure 5):
6 Shutter Transformation: Display shutters allow to mask unwanted parts of the image, e. g. unexposed parts of an X-ray which are displayed very brightly because of collimation and affect the visibility of relevant parts of the image. Image Annotation: Images can be annotated with graphical and textual comments displayed as overlays on the image, e. g. in order to highlight regions of interest (ROI) or document measurements on the image. Unlike overlays that are burned into the image, annotations in a Presentation State could be displayed and switched off at the user s discretion. Spatial Transformations: Images can be rotated and flipped (often required for Computed Radiography images). Images can be zoomed to a defined displayed area (rectangle) or to a specified zoom factor. It is also possible to request the display of an image at its true physical size if this size is known. Displayed Area Annotation: This annotation is not attached to the image but to the display (view port). This allows for example to display the patient demographic data (name, birth date, sex) or patient comments in the edges of the screen independent from the zoom factor or rotation applied to the image. Most DICOM image viewing applications already support similar annotation features, mostly preserved in a proprietary manner. However, DICOM Presentation States allow to archive this information in a PACS and to have it available for the full lifetime of the images, even if devices change over time. Presentation States can also be exchanged between viewing stations of different vendors supporting this DICOM service. 4. PRESENTATION LOOK UP TABLE The concept of the DICOM Grayscale Standard Display Function and the notion of P-Values as the input to a standardized display system are not limited to the softcopy domain but can also be applied to hardcopy. The DICOM Presentation LUT Supplement [4], which is already integrated into the main standard text since the 1998 edition, makes these concepts available for printing as an extension of the DICOM Print Management Service Class. This service class consists of a number of print services and optional extensions and has been available as part of the DICOM standard since its initial release in Basic Grayscale Print Management is the service that is most often used for printing grayscale images. It differs from softcopy display by the fact that the image sent to the printer is always pre-formatted. This means that most steps of the DICOM grayscale transformation model (Figure 5) must be rendered by the print client before sending the image to the printer. Although a pre-formatted image does not leave much room for ambiguity, the fact that DICOM never defined the precise relationship between image pixel values and luminance or optical density turned out to be a major problem since different vendors interpreted pixel values differently (e. g. linear to optical density or linear to human perceptual response). Most print vendors addressed this problem by implementing a large number of correction curves and configuration options that allow to adapt a printer to the characteristics of the modality from which images are printed. The correction curves also serve as a non-standard replacement for the Presentation LUT, i. e. they allow to change the image appearance in accordance with the user s preferences. The obvious disadvantage of this approach is that it requires an adaptation of the printer settings for each modality which is to be connected to the printer (and, at least in theory, even for each user with different preferences about image appearance). While this may be acceptable for installations with a limited number of modalities and printers, it becomes a significant problem in complex, heterogeneous PACS where printers need to be able to serve requests from a large number of modalities and workstations. The DICOM Presentation LUT Supplement [4] provides a solution to this problem by precisely defining how the preformatted image relates to the P-Value space and by making the parameters which are required to control hardcopy image appearance an explicit part of the communication protocol. Figure 8 shows a number of parameters which are required by a DICOM printer to control the image appearance on the hardcopy. The black arrows show extensions of the DICOM print communication protocol from [4]. The basic idea of the Presentation LUT Supplement is that a print client (modality or workstation) and a print server (printer) which are both aware of this DICOM extension can negotiate support for it during establishment of a connection (i. e. print session). The print client then either transmits a pre-formatted image in P-Values, or it sends a pre-formatted bitmap that is accompanied by a Presentation LUT which allows the print server to convert the bitmap into P-Values. The latter option is useful for modalities which only support the transmission of 8-bit image data. The print client also sends information about the lighting conditions (illumination and reflection caused by ambient light) under which the resulting hardcopy will be viewed. DICOM recommends default values of 2000 cd/m 2 for illumination and 10 cd/m 2 for reflected ambient light for transmissive hardcopies, which are typical viewing conditions for hardcopy reading. Knowing its own internal characteristic curve, which is a result of the printer s calibration procedure, and the desired maxi-
7 mum and minimum density for the print, which can also be transmitted as part of the DICOM print protocol, the print server can compute a correction curve which calibrates the hardcopy to the DICOM Grayscale Standard Display Function for the defined viewing conditions. Modality Printer Hardware image characteristic curve presentation LUT illumination User Preference max/min density DICOM Printer reflection Viewing Conditions Figure 8: Parameters affecting the Rendering of an Image in a DICOM Printer This approach combines the advantages of a standard display function with the flexibility required to adapt hardcopy image appearance to user preferences. Since modalities and printers supporting the Presentation LUT Supplement use a welldefined interface for image transmission, P-Values, there is no need to adapt each printer to each modality any more. A calibration of the printers against the GSDF makes sure that hardcopies from different printers become directly comparable (consistent), of course within the physical limitations of the printer. Since DICOM proposes that both softcopy and hardcopy devices should be calibrated against the GSDF, this even allows to make softcopy display and hardcopy image appearance directly comparable. On the other hand, the possibility to use Presentation look-up tables gives the flexibility needed to adapt image appearance to user (or vendor) preferences, independent from the printer. Finally, Presentation States allow to carry the concept of user preference over to the softcopy world. 5. EXPERIENCES In order to validate the feasibility of the new services described above and to demonstrate its potential to the radiological society, the NEMA Committee for the Advancement of DICOM commissioned a prototype implementation of a DICOM image viewer supporting Presentation States, GSDF calibration and DICOM printing with Presentation LUT support. This prototype has been shown at RSNA inforad 1999, together with a number of prototype implementations of DICOM printers supporting Presentation LUT, as the DICOM Softcopy and Hardcopy Consistency Demonstration. Figure 9 shows the concept of the demonstration. Modality Diagnostic Workstation Clinical Review Station image presentation image presentation Printer (Laser Camera) Printer (Paper) Figure 9: Concept of RSNA inforad 1999 Demonstration
8 The exhibit demonstrated a simulated radiological workflow in which images were created by a modality and transmitted to a diagnostic workstation (with 2 2.5k high resolution grayscale monitor) where a Presentation State was created for the image. Both image and Presentation State were then sent to a review workstation (PC with conventional color monitor) and displayed. Finally, images could be printed on a laser camera or a paper printer. A number of printers from different vendors and softcopy displays using different technologies (monochrome CRTs with hardware support for display calibration, conventional color CRTs, flat-panel color LCD) allowed to compare the display consistency between display devices with very different characteristics. The prototype DICOM viewer which was installed on the simulated modalities as well as on the diagnostic and review workstations is an extension of the software shown for the first time at the European Congress of Radiology 1999 [6]. It is implemented in a combination of Java and C++ (which allows easy porting of the software to different operating systems such as Microsoft Windows, Sun Solaris or Linux) and freely available as Open Source software on the Internet [7,8]. Figure 10 shows a screenshot of the demonstration software. Figure 10: DICOM Softcopy Presentation State Demonstration Software Altogether the demonstration included seven independent implementations of a DICOM GSDF calibration (four printers and three softcopy display systems including the demonstration software mentioned before) which can be regarded as a true proof of concept for the applicability of the GSDF to different types of display devices. An image displayed on different types of monitors and printed with different printers does not really look the same because the physical properties of these devices (resolution, brightness, color temperature) are quite different, however, the visual impression confirms that the image appearance is consistent. Most healthcare professionals visiting the demonstration acknowledged that the possibility to calibrate both hardcopy and softcopy displays to a common standard solves a problem which is a major issue in many institutions. Many of them were also not aware that the concept of calibration is not bound to specialized hardware and can be implemented with rather moderate expenditure like a portable photometer and a software-based correction. This fact might be of importance wherever images are viewed on standard consumer devices (e. g. a radiologist reading an emergency case at home on a standard PC). 6. DISCUSSION Since human visual perception depends on the lighting conditions under which images are viewed, measuring and controlling these conditions is important if one wants to guarantee a consistent image appearance. In particular the reflection caused by ambient light on a monitor or screen film must be measured and taken into account for the calibration the conventional suction cup type of photometer do not allow to measure this factor. Changing lighting conditions require an
9 adjustment of the image correction tables implementing the display calibration. Whereas this can be automated with dedicated hardware (an ambient light sensor permanently tracking changing lighting conditions and adjusting the correction tables), it is an open question how this could be handled in a way which does not become a burden to the user on systems where such hardware is not available. It is also debatable whether it can be guaranteed that printed films are always viewed under the same lighting conditions, in particular if films are moved from the radiology department to the operating theatre or the clinical ward. However, even if a hardcopy is only calibrated for average viewing conditions, this is still a significant step forward from the current state where no consistency at all is the normal case and not the exception. The calibration of softcopy displays requires that the monitors are carefully adjusted first and measured under stable conditions (e. g. full operating temperature). One particular problem with standard consumer quality monitors is the fact that the brightness and contrast settings, which affect the calibration, can be easily changed because the control is usually located on the front side of the monitor and cannot be locked. There is also a number of other factors which affect image display quality on softcopy displays that are outside the scope of the DICOM GSDF, such as spatial resolution, image geometry, image stability, artifacts and color temperature. However, there are other standards such as the German DIN standard on acceptance testing of image display devices [9] which address these issues. Finally, image consistency by means of calibration requires organizational measures. Responsibilities for the calibration procedure need to be assigned and the intervals for the calibration must be defined (the vendors recommendations on how often a calibration should be performed vary significantly). Test patterns allowing to easily check whether the display looks right (e. g. whether the 5% and 95% steps in the SMPTE pattern are visible) should be available to the medical users, for example as a screen saver or as part of the login dialog. This also implies that users need to be trained what to look for (on the test patterns) and what to avoid (changing the monitor s contrast and brightness). The software developed for the RSNA inforad 1999 demonstration shows that DICOM Presentation States can be integrated with an image viewer application in a way that does not require the user to know the details about the underlying concepts. For the user, Presentation States only manifest themselves as the possibility to store, retrieve and select not only images but also snapshots of the viewer settings defining how the image is displayed. Implementing this added value for the user, however, is everything but easy because the Presentation State definition contains many details that need to be addressed. In order to prevent partial implementations of the standard from defeating the concept of exchanging the visual presentation of an image between different viewers, DICOM requires that all mandatory parts of a Presentation State (which means most parts) must be supported by a Presentation State aware viewer if it chooses to accept a Presentation State at all: A display device acting as an SCP of this SOP Class shall make all mandatory presentation attributes available for application to the referenced images at the discretion of the display device user, for all Image Storage SOP Classes defined in the Conformance Statement for which the Grayscale Softcopy Presentation State Storage SOP Class is supported. [3] Another complication is caused by the fact that the linkage between image and Presentation State is only contained in the Presentation State object and not in the image. This means that, in order to find out which Presentation States are available for a particular image, a viewer must search and download all Presentation States that are part of the same study. Finally, there are a number of limitations in the current Presentation State object which need to be addressed by further extensions of the standard: Presentation States do not provide information about how to display cine-based multi-frame images (e. g. angiography) and they do not support color images. There is also no support yet for hanging protocols (i. e. screen layouts). These features might be defined in a future Advanced Presentation State object, but are not yet available. Finally, Presentation States, or the graphical annotations contained in them, are not associated with any semantics. This means that it is possible to draw the graphical representation of a scale into a Presentation State as the result of a on-screen measurement, but it is not possible to store the information that the lines and characters comprising this graphical element define a measurement. It is also not possible to mark a number of frames from a series as key frames. Such features will be available with the advent of DICOM Structured Reporting (SR), and a tight integration of SR with Presentation States is, therefore, more than desirable. 7. CONCLUSION The DICOM Grayscale Standard Display Function defines a standard for the calibration of softcopy and hardcopy display devices. It allows to implement a consistent image display over a wide variety of display devices with different characteris-
10 tics. The GSDF can be implemented in hardware or also rather efficiently in software. It only requires the use of a photometer (for monitors) or densitometer (for printers) to measure the characteristic curve of the display device during calibration. Calibrating a display device cannot improve its physical properties but makes sure that the available contrast range is used in an optimal manner. Moreover, the continuous control over the device's life-time allows to detect and potentially compensate for failures or degradation of the display quality early in time. A regular check of all output devices is a critical part of the quality control in clinical environments. It should be noted, however, that the DICOM GSDF is only applicable to the display of monochrome images and cannot be used for a color calibration as it might be required in pathology or dermatology. At the time being the DICOM standard does not include or recommend a procedure for color calibration. The possibility to store and distribute the presentation of an image including annotations between softcopy devices from different vendors in a standard manner is new to the PACS world. The same applies to the possibility to offer multiple alternative views of the same image (by means of multiple Presentation States). The well-defined image processing pipeline paves the way for a consistent image display on calibrated monitors in conjunction with the DICOM Grayscale Standard Display Function. Several application fields for Presentation States come to mind, for example: Documentation of softcopy reading: When images are read from plain film, the film itself precisely documents on which information the radiologist based the diagnosis. However, digital images can be viewed in different ways, e. g. with different window level and width, magnification factor, etc. Presentation States allow to accurately document the diagnostic process in a softcopy environment. Image transfer to clinical departments: Presentation States permit to make sure that images are sent out to the clinical departments with reasonable viewing defaults (window level and width adjusted, image orientation corrected etc.) Annotations allow to point out important details on an image. Teleradiology: Since Presentation States are very small, they are well suited for teleradiology applications in which images are transferred in advance and only Presentation States (e. g. together with reports) are exchanged online. The current definition of Presentation States is still relatively basic. Nevertheless, based on our feedback during the RSNA, Presentation States seem to be a useful extension of the set of DICOM services. It will be interesting to see at which time frame they will start to be supported in commercial products. 8. REFERENCES 1. NEMA Standards Publication PS 3.x, Digital Imaging and Communications in Medicine (DICOM), National Electrical Manufacturers Association, 2101 L Street, N. W., Washington, D. C , NEMA Standards Publication PS 3.14, Digital Imaging and Communications in Medicine (DICOM), Part 14: Grayscale Standard Display Function, National Electrical Manufacturers Association, 2101 L Street, N. W., Washington, D. C , DICOM Standards Committee, Working Group 11 Display: Digital Imaging and Communications in Medicine (DICOM), Supplement 33: Grayscale Softcopy Presentation State Storage, Final Text, September DICOM Standards Committee, Working Group 6 Printer Ad Hoc: Digital Imaging and Communications in Medicine (DICOM), Supplement 22: Presentation Look Up Table (LUT), Final Text, January Barten, PGJ, Physical model for the Contrast Sensitivity of the human eye, Proceedings SPIE 1666, (1992). 6. Eichelberg M, Riesmeier J, Jensch P: Grayscale Softcopy Presentation States a new DICOM Service for Documenting Image Appearance in a Softcopy Environment, in: P. Rehak, H. Hutten: Medical & Biological Engineering & Computing, Vol. 37, Supplement 2, pp (1999). 7. OFFIS DICOM Project, 8. Institute for MicroTherapy, 9. DIN : Image quality assurance in X-ray diagnosis Part 57: Acceptance testing for image display devices, Normenausschuß Radiologie im DIN Deutsches Institut für Normung e.v.
User Manual. DICOMscope 3.6.0 2004-01-09. Software developed by: M. Eichelberg 1, K. Kleber 2, J. Riesmeier 1, A. Schröter 2, A.
User Manual DICOMscope 3.6.0 2004-01-09 Software developed by: M. Eichelberg 1, K. Kleber 2, J. Riesmeier 1, A. Schröter 2, A.Thiel 1 Organizational support: J. Holstein 2, H. Oosterwijk 3 1 2 3 Kuratorium
Technical Paper DENTAL MONITOR CALIBRATION
Technical Paper DENTAL MONITOR CALIBRATION A REPORT ON DENTAL IMAGE PRESENTATION By Tom Schulte Dental radiographs and oral photographs are often both viewed on the same dental workstation monitor. The
A PACS-Aware DICOM Image Object
Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry
Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices
IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists
Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2
Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for
Medical imaging monitors specification guidelines
Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Understanding Digital Modalities: System Integration and Use
Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI [email protected] Dicom Digital Imaging and Communications in Medicine Part 3: Information
Workstation, Server Hosted
Workstation, Server Hosted With so many PACS out there, Why just make another? So, we didn't. Distribute, Burn, Print, Import, Archive The Internet DICOM Eye Web Web Viewer PACStoGo DICOM Distribution
DICOM Conformance Statement. DICOMscope 3.5. Software developed by: M. Eichelberg 1, K. Kleber 2, J. Riesmeier 1, A. Schröter 2, A.
DICOM Conformance Statement DICOMscope 3.5 Software developed by: M. Eichelberg 1, K. Kleber 2, J. Riesmeier 1, A. Schröter 2, A. Thiel 1 Organizational support: J. Holstein 2, H. Oosterwijk 3 1) Kuratorium
Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors
Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4
MQSA Quality Control Manual for Monochrome Displays for Mammography
MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography
Version 8 DICOM Conformance Statement. Version 3.04, September 2014
Version 8 DICOM Conformance Statement Version 3.04, September 2014 1 Conformance Statement Overview The application described in this Conformance Statement VEPRO EMR Manager PACS is a collection of processes
Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function
PS 3.14-2011 Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn,
Coronis 5MP Mammo. The standard of care for digital mammography
Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment
PARCA Certified PACS Interface Analyst (CPIA) Requirements
PARCA Certified PACS Interface Analyst (CPIA) Requirements Copyright notice: Copyright 2005 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved.
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS
COST EFFECTIVE FLAT PANEL DIGITAL RADIOGRAPHY UPGRADE SOLUTIONS DRive is a digital imaging DR hardware & Software solution designed for General Radiography of anatomy. It is intended to replace film/screen
Nio. Industry-standard diagnostic display systems
Nio Industry-standard diagnostic display systems Diagnostic confidence in grayscale With the Nio diagnostic display system, Barco brings dependable diagnostic imaging to its true potential. Presenting
DICOM Conformance Statement For Diagnostic Review Workstation Software Version 5-x MAN-00546
DICOM Conformance Statement For Diagnostic Review Workstation Software Version 5-x MAN-00546 SecurView DX DICOM Conformance Statement Part Number MAN-00546 Revision 001 For Software version 5-X Service
Supports screening, diagnostic, and multimodality workflows
GE Healthcare Universal Viewer Breast Imaging Product Data Sheet Supports screening, diagnostic, and multimodality workflows Introduction Centricity PACS with Universal Viewer 1 puts clinical insight within
DryView CHROMA Imaging System
DryView CHROMA Imaging System VERSATILE PRINTING ON FILM OR PAPER One of the most trusted names in medical printing now offers a highly versatile printer that empowers physicians and other healthcare professionals
Calibration Best Practices
Calibration Best Practices for Manufacturers SpectraCal, Inc. 17544 Midvale Avenue N., Suite 100 Shoreline, WA 98133 (206) 420-7514 [email protected] http://studio.spectracal.com Calibration Best Practices
Extracting, Storing And Viewing The Data From Dicom Files
Extracting, Storing And Viewing The Data From Dicom Files L. Stanescu, D.D Burdescu, A. Ion, A. Caldare, E. Georgescu University of Kraiova, Romania Faculty of Control Computers and Electronics www.software.ucv.ro/en.
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
Interfacing Fuji CR in a DICOM Network Environment
Interfacing Fuji CR in a DICOM Network Environment Wayne T. DeJarnette, Ph.D., Valmore Albrecht, RT, Alan Orth, RT, Jon Sayer, M.S. DeJarnette Research Systems, Inc. 401 Washington Avenue Towson, MD 21204
DICOM 3.0 Conformance Statement
Version 2.7 DICOM 3.0 Conformance Statement 1 1. Conformance Statement Overview SOP Classes Query and Retrieve Study Root Query/Retrieve Information Model Find Study Root Query/Retrieve Information Model
Guidelines On Teleradiology In Malaysia
1 The following are Teleradiology Standards prepared by our College (MRS) and submitted to the Ministry of Health. The Ministry of Health has not endorsed these standards yet. However, we hope they will
Picture Archiving and Communication systems
Picture Archiving and Communication systems (PACS) By: Somayeh nabiuni 1 Major limitations of the conventional radiology practice Wasted time - implying diagnostic results many not be obtained in a timely
HDI 4000 Ultrasound System
HDI 4000 Ultrasound System DICOM Conformance Statement Rev 1.0 Philips Medical Systems HDI 4000 DICOM Conformance Rev. 1.0 12/17/2001 Philips Medical Systems Table of Contents 0. Introduction...3 0.1 DICOM
Tools for DICOM Implementation
DICOM INTERNATIONAL CONFERENCE & SEMINAR Oct 9-11, 2010 Rio de Janeiro, Brazil Tools for DICOM Implementation David Clunie CoreLab Partners, Inc. Outline Tools for DICOM implementation Toolkits and sample/reference
HP Medical Archive Solutions DICOM Conformance Statement. January 2007 (Third Edition) Part Number T4434-96003
HP Medical Archive Solutions DICOM Conformance Statement January 2007 (Third Edition) Part Number Copyright 2007, 2007 Hewlett-Packard Development Company, L.P. Confidential computer software. Valid license
Monitor Quality Control Solutions
Monitor Quality Control Solutions Managing the quality of monitors used every day leads to the improvement of the quality of medical care itself. Detecting changes in a monitor s display quality Controlling
MiPACS Storage Server Conformance Statement Version 1.02.0.000
MiPACS Storage Server Conformance Statement Version 1.02.0.000 December 2009 Version 1.00.4.001 Conformance Statement Document Information Issued by: Medicor Imaging 1927 S. Tryon St., Suite 200 Charlotte,
CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0. DICOM Conformance Statement
CARESTREAM PACS Suite (Client, Server, and CD Direct) Version 11.0 DICOM Conformance Statement Publication # 8G8913 22 April 2009 Table of Contents Table of Contents...2 1 Introduction...5 1.1 Abbreviations
Converting the DICOM Presentation State to AIM Version 3.0
Converting the DICOM Presentation State to AIM Version 3.0 Development Team(s) Development Technical Documentation Program Management Pattanasak Mongkolwat Pattanasak Mongkolwat Peter Yan Vladimir Kleper
CAR Standards for Teleradiology
CAR Standards for Teleradiology The standards of the Canadian Association of Radiologists (CAR) are not rules, but are guidelines that attempt to define principles of practice that should generally produce
Overview. Raster Graphics and Color. Overview. Display Hardware. Liquid Crystal Display (LCD) Cathode Ray Tube (CRT)
Raster Graphics and Color Greg Humphreys CS445: Intro Graphics University of Virginia, Fall 2004 Color models Color models Display Hardware Video display devices Cathode Ray Tube (CRT) Liquid Crystal Display
DICOM Conformance Statement
DICOM Conformance Statement CARESTREAM Vue PACS v11.3.2.4 Lesion Management and Bookmark Applications Part # 6K4188 2012-07-02 PAGE 1 of 142 Table of Contents 1 INTRODUCTION... 4 1.1 ABBREVIATIONS AND
Introducing MIPAV. In this chapter...
1 Introducing MIPAV In this chapter... Platform independence on page 44 Supported image types on page 45 Visualization of images on page 45 Extensibility with Java plug-ins on page 47 Sampling of MIPAV
GE PACS Conformance Statement for DICOM v3.0
g GE Medical Systems Technical Publications IIS P/N 4361668 Revision 01 GE PACS Conformance Statement for DICOM v3.0 Copyright 1998 By General Electric Company g GE Medical Systems GE Medical Systems,
Digital Print Manufacturing: Color Management Workflows and Roles. Ann McCarthy Xerox Innovation Group ICC Steering Committee
Digital Print Manufacturing: Color Management Workflows and Roles Ann McCarthy Xerox Innovation Group ICC Steering Committee ICC Color Management Workflows Digital Smart Factory Forum 24 June, 2003 What
MammoView 1.5. DICOM Conformance Statement 1.0. medigration GmbH All rights reserved
All rights reserved medigration GmbH, Schuhstr. 30, 91052 Erlangen Tel: +49 (0) 9131/69087-40, Fax: +49 (0) 9131/69087-50 www.medigration.de Copyright. All rights reserved. Table of Contens 1 Introduction...
A full-featured online DICOM archive built on the same SQL platform as our powerful enterprise grade PACS database. Enterprise Grade DICOM Engine
DICOM A full-featured online DICOM archive built on the same SQL platform as our powerful enterprise grade PACS database. SQL Database Enterprise Grade DICOM Engine Flexible Archiving Solutions Internet
Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function
PS 3.14-2004 Digital Imaging and Communications in Medicine (DICOM) Part 14: Grayscale Standard Display Function Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn,
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
PARCA Certified PACS System Analyst (CPSA2014) Requirements
PARCA Certified PACS System Analyst (CPSA2014) Requirements Copy right notice: Copyright 2014 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved.
Printlink5-ID_IV. DICOM 3.0 Conformance Statement PRINT MANAGEMENT SYSTEM CODE NO. 1118. Manufacturer:
PRINT MANAGEMENT SYSTEM Printlink5-ID_IV DICOM 3.0 Conformance Statement CODE NO. 1118 Manufacturer: 2970 Ishikawa-machi, Hachioji-shi Tokyo 192-8505 Japan DICOM Conformance Statement Important Notes Konica
Technical Publication. DICOM Conformance Statement. DICOM Proxy 2.0. Document Revision 3. October 20, 2010. 2010 Copyright Brainlab AG
Technical Publication DICOM Conformance Statement Document Revision 3 October 20, 2010 2010 Copyright Brainlab AG 1 Conformance Statement Overview This is a conformance statement for the Brainlab software
Dx MM DICOM 3.0 Conformance Statement
Dx MM DICOM 3.0 Conformance Statement NAME FONCTION VISA AUTHOR Alain Battistini Senior Engineer CONFORM TO ORIGINAL REVIEW Jean-François Hoh Senior Engineer CONFORM TO ORIGINAL APPROVAL Pascal Gilbert-Jeantet
Philips Medical Systems DICOM Conformance Statement
Philips Medical Systems DICOM Conformance Statement DSI Pro 1.0 Document Number 4522 164 07221 20 February 2001 Copyright Philips Medical Systems Nederland B.V. 2001 Page ii DICOM Conformance Statement
Available online at www.sciencedirect.com. ScienceDirect. Procedia Computer Science 37 (2014 ) 16 23
Available online at www.sciencedirect.com ScienceDirect Procedia Computer Science 37 (2014 ) 16 23 The 5th International Conference on Emerging Ubiquitous Systems and Pervasive Networks (EUSPN-2014) HIS
OPERATING INSTRUCTIONS Nikon TiE Deconvolution Microscope CCRB 1-220G
OPERATING INSTRUCTIONS Nikon TiE Deconvolution Microscope CCRB 1-220G Conventions Through these notes bold font refers to a software button and italics refer to a hardware switch. The software makes extensive
DICOM Correction Proposal
DICOM Correction Proposal STATUS Assigned Date of Last Update 20165/01/08 Person Assigned Submitter Name Harry Solomon Harry Solomon Submission Date 2015/09/21 Correction Number CP-1550 Log Summary: Refactor
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Monitor Qualit y Control Solutions
Monitor Qualit y Control Solutions Controlling the Quality of Monitors Managing the quality of monitors used every day leads to the improvement of the quality of medical care itself. Q1 A1 Do all monitors
Standard-Compliant Streaming of Images in Electronic Health Records
WHITE PAPER Standard-Compliant Streaming of Images in Electronic Health Records Combining JPIP streaming and WADO within the XDS-I framework 03.09 Copyright 2010 Aware, Inc. All Rights Reserved. No part
Mammography as you ve never seen it before
CARESTREAM Mammography Workstation and PACS Mammography as you ve never seen it before Whether you re planning a move to digital mammography or you ve already transitioned to digital, it s important to
GelAnalyzer 2010 User s manual. Contents
GelAnalyzer 2010 User s manual Contents 1. Starting GelAnalyzer... 2 2. The main window... 2 3. Create a new analysis... 2 4. The image window... 3 5. Lanes... 3 5.1 Detect lanes automatically... 3 5.2
ACR AAPM SIIM TECHNICAL STANDARD FOR ELECTRONIC PRACTICE OF MEDICAL IMAGING
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project
Software Tools for DICOM Media Exchange in Clinical Research: the SWABIK Project Software Tools for DICOM Media Exchange in Clinical Research: the SWABIK Project O. ElGazzar 1, M. Onken 1, M. Eichelberg
Digital Imaging and Communications in Medicine (DICOM) Part 1: Introduction and Overview
Digital Imaging and Communications in Medicine (DICOM) Part 1: Introduction and Overview Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn, Virginia 22209 USA Copyright
DICOM Digital Imaging and Communications in Medicine
DICOM Digital Imaging and Communications in Medicine INF5081 / Vårsemester 2007 Wolfgang Leister Reidar Øksnevad Pedro Cruz G. April 2007 PACS PACS = Picture Archive and Communication System Archive for
Head Office #5 Sugar Creek Road Piedmont, Missouri 63957. Tel: 877-838-9273, option 2, sub-option 8 Fax: 573-223-3551 [email protected].
Head Office #5 Sugar Creek Road Piedmont, Missouri 63957 Tel: 877-838-9273, option 2, sub-option 8 Fax: 573-223-3551 [email protected]. DragonflyPACS Contact Information Dragonfly Specialized Support
REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) TERMINOLOGY NEEDED TO UNDERSTAND MONITOR PERFORMANCE
REQUIREMENTS FOR MEDICAL IMAGING MONITORS (part I) By Ken Compton and Herman Oosterwijk The question of what monitor to use for diagnostic purposes with digital medical images has been a recurring theme
DICOM Conformance Statement FORUM
DICOM Conformance Statement FORUM Version 3.1 Carl Zeiss Meditec AG Goeschwitzerstraße 51-52 07745 Jena Germany www.meditec.zeiss.com Document: DICOM Conformance Statement_FORUM_3.1.doc Page 1 of 25 1
Digital Imaging and Communications in Medicine (DICOM) Part 4: Service Class Specifications
Digital Imaging and Communications in Medicine (DICOM) Part 4: Service Class Specifications Published by National Electrical Manufacturers Association 1300 N. 17th Street Rosslyn, Virginia 22209 USA Copyright
DICOM Conformance Statement
Philips Medical Systems DICOM Conformance Statement Integris Systems with High Speed DICOM Image Interface MCV 2974 MCV 3621 Cardiac DICOM XA MCV 3761 Vascular DICOM XA MCV 3031 DICOM RIS Interface and
DICOM Conformance Statement
This printed copy is not subject to any change control XXXXXXXXXXXXxxXXXXXXXXXXXXXXXXXXX MR 2013-03-14 Siemens AG 2013 All rights reserved Siemens AG Healthcare Sector, Henkestr. 127, D-91050 Erlangen,
AGFA HEALTHCARE DICOM Conformance Statement
Page 1 of 79 AGFA HEALTHCARE DICOM Conformance Statement IMPAX 6.5 Solution Document No. 001310 Revision: 1.1 When printed, this is NOT a controlled document. Page 2 of 79 Document Information Service-related
Hologic Physician s Viewer 7.0 DICOM Conformance Statement
Hologic Physician s Viewer 7.0 DICOM Conformance Statement MAN-02386 REV 001 Page 1 of 32 1. Conformance Statement Overview This document is the DICOM Conformance Statement for Hologic Physician s Viewer,
High Dynamic Range Video The Future of TV Viewing Experience
The Future of TV Viewing Experience - White Paper - www.keepixo.com Introduction The video industry has always worked to improve the TV viewing experience. More than a decade ago, the transition from SD
Microsoft Amalga Hospital Information System (HIS)
m Microsoft Amalga Hospital Information System (HIS) > Manage all hospital functions with one integrated solution PG 0 Our Vision: To improve health around the world For more than a decade, Microsoft has
Managing DICOM Image Metadata with Desktop Operating Systems Native User Interface
Managing DICOM Image Metadata with Desktop Operating Systems Native User Interface Chia-Chi Teng, Member, IEEE Abstract Picture Archiving and Communication System (PACS) is commonly used in the hospital
DICOM Conformance Statement
DICOM Conformance Statement Page 1 of 14 1 Introduction... 3 2 Quick Summary... 4 3 Storing Studies... 6 3.1 Ideal Store scenario...6 3.2 Non-ideal scenarios...6 3.2.1 Accession number...7 3.2.2 No indication
DICOM 3.0 Conformance Statement
Picture Archiving and Communication System DICOM 3.0 Conformance Statement CODE NO.1245 Manufacturer: 1 Sakura-machi, Hino-shi Tokyo 191-8511, Japan Important Notes Konica Minolta Medical & Graphic, Inc.
DICOM Conformance Statement
Date: 2010-10-7 DICOM Conformance Statement Stratus OCT Instrument and Review Software Version 7.0 Carl Zeiss Meditec, Inc. 5160 Hacienda Drive Dublin, CA 94568 USA www.meditec.zeiss.com Document: DICOM_Conformance_Statement_Stratus_7.0.doc
Image Acquisition and Diagnostics
en examion aqs software examion aqs software Image Acquisition and Diagnostics Professional Image Acquisition and Diagnostic Workstation examion aqs software is the universal image acquisition software
DR Retrofit Kit 1417 WL Configuration Integrated PACS
DR Retrofit Kit 1417 WL Configuration Integrated PACS Description: The DR Retrofit Kit 1417WL - Configuration Integrated PACS makes the use of CR Cassettes/Imaging Plates or the use of analogue x-ray films,
1 Introduction. 1.1 PACS Introduction
1 Introduction 1.1 PACS Introduction A PACS system displays, archives and communicates medical digital images. It provides an electronic alternative to using conventional, analogue methods of acquiring
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
A Proposal for OpenEXR Color Management
A Proposal for OpenEXR Color Management Florian Kainz, Industrial Light & Magic Revision 5, 08/05/2004 Abstract We propose a practical color management scheme for the OpenEXR image file format as used
DICOM 3.0 CONFORMANCE STATEMENT
DICOM 3.0 CONFORMANCE STATEMENT JETStream Workspace Version 1.0 9705-0067, Rev A October, 004 Copyright Statement Philips Medical Systems has taken care to ensure the accuracy of this document. However,
DigitizingStation. DICOM V3.0 Conformance Statement
DigitizingStation DICOM V3.0 Conformance Statement Copyright 2004 by image diagnost GmbH Page 1 of 17 Image Diagnost International GmbH Reproduction of this document without prior written approval is prohibited.
Dome CXtra User s Guide
Dome CXtra User s Guide 2015 NDS Surgical Imaging, LLC. All rights reserved. This document contains proprietary information of NDS Surgical Imaging. It is exclusive property of NDS Surgical Imaging. It
Workshop: Defining the Medical Imaging Requirements for a Health Center. Teleradiology and Networking
Workshop: Defining the Medical Imaging Requirements for a Health Center April 17 2011 E-health and Telemedicine ehealth is the use, in the health sector, of digital data - transmitted, stored and retrieved
Digital Imaging and Communications in Medicine (DICOM) Supplement 23: Structured Reporting Storage SOP Classes
Page 1 4 Digital Imaging and Communications in Medicine (DICOM) Supplement 23: Structured Reporting Storage SOP Classes 8 12 16 20 Status: Letter Ballot Text, 29 October, 1999 24 DICOM Standards Committee
DICOM Conformance Statement. Version: 1.0
DICOM Conformance Statement Version: 1.0 Table of Contents 1 Preface...4 2 Overview...5 3 Introduction...5 3.1 Revision History...5 3.2 Audience...5 3.3 Remarks...6 3.4 Definitions, Terms and Abbreviations...6
3D Viewer. user's manual 10017352_2
EN 3D Viewer user's manual 10017352_2 TABLE OF CONTENTS 1 SYSTEM REQUIREMENTS...1 2 STARTING PLANMECA 3D VIEWER...2 3 PLANMECA 3D VIEWER INTRODUCTION...3 3.1 Menu Toolbar... 4 4 EXPLORER...6 4.1 3D Volume
PARCA Certified PACS Associate (CPAS) Requirements
PARCA Certified PACS Associate (CPAS) Requirements Copyright notice: Copyright 2005 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved. This publication
Nuclear PACS: Problems, Solutions
Nuclear PACS: Problems, Solutions by Jerold W. Wallis, MD Introduction Reprinted with permission: Wallis JW. Nuclear PACS: Problems, Solutions. Decisions in Imaging Economics. 2004;6;21-25. There is increasing
ANALYSIS OF THE COMPRESSION RATIO AND QUALITY IN MEDICAL IMAGES
ISSN 392 24X INFORMATION TECHNOLOGY AND CONTROL, 2006, Vol.35, No.4 ANALYSIS OF THE COMPRESSION RATIO AND QUALITY IN MEDICAL IMAGES Darius Mateika, Romanas Martavičius Department of Electronic Systems,
PARCA Certified PACS System Analyst (CPSA) Requirements
PARCA Certified PACS System Analyst (CPSA) Requirements Copyright notice: Copyright 2005 PACS Administrators in Radiology Certification Association (PARCA). All rights reserved. All rights reserved. This
Philips HDR technology White paper
Philips HDR technology White paper Philips International B.V. Version 1.0 2015-08-21 1. Introduction Current video formats are still based upon the display capabilities of CRT monitors. They cannot capture