BIOMEDICAL TELEMETRY: COMMUNICATION BETWEEN IMPLANTED DEVICES AND THE EXTERNAL WORLD. By Asimina Kiourti. Introduction
|
|
|
- Britney Stephens
- 10 years ago
- Views:
Transcription
1 Introduction BIOMEDICAL TELEMETRY: COMMUNICATION BETWEEN IMPLANTED DEVICES AND THE EXTERNAL WORLD By Asimina Kiourti There have been many important technological achievements in microtechnologies and microsystems in the past fifteen years. These have lead to the design of small, reliable and lowpower-consuming biomedical devices that can be implanted inside a patient s body by means of a surgical operation. Contrary to traditional, external medical devices, these implanted devices can sense data from inside the human body in real-time, offering a unique opportunity for early diagnosis and treatment of diseases. In the most complicated scenario, implanted devices communicate with the external world in terms of both powering and telemetry. Powering is the delivering of energy to the implant from the external world in order to make it work. Telemetry includes data transmission from the implanted device to an external one, and vice versa. Depending on the design, several values of implant dimensions, delivered power, data transmission and error rates can be achieved. Applications of Implanted Devices Implanted devices can act as either sensors or stimulators. Sensors measure a biosignal from inside the body and transmit this information to an external device. They can measure body temperature, blood pressure and glucose concentration, for example, and detect respiratory, cardiac and arterial wall movements, the contraction of blood vessels and cardiac pressure disorders. The information received by the external device is post-processed by monitoring units and medical experts who treat the patient accordingly. This allows several diseases, such as cancer or diabetes, to be diagnosed in their very early stages, while critical medical conditions, such as heart attacks or strokes, can be prevented. Stimulators receive information from the external world (in terms of an external unit operated by doctors) and stimulate specific nerves. Among the common applications of stimulators are microelectrodes for neural recording, used to diagnose and determine treatment for brain disorders. The electrical stimulation of retinal ganglion/bipolar cells creates visual excitation in patients who suffer from retinal diseases such as Retinitis Pigmentosa (RP) and Age-Related Macular Degeneration (AMD). Cochlear stimulators offer people who are severely hard of hearing or even profoundly deaf the hope of restoring the ability to sense sound. Meanwhile, stimulators are used in Functional Electrical Stimulation (FES) to restore function in paralyzed muscles by stimulating the nerves that control them, or the muscles themselves. Powering the Implanted Device Implanted devices need energy in order to sense or stimulate. The amount of energy required for the implant to work is small but cannot cease; if the implanted device runs out of energy, it would become useless and have to be substituted by surgical operation. Early implanted devices were interfaced with wires through the skin in order to receive energy. This way of powering soon proved to be ineffective since it restricted the movements of the patient and increased the chances of infection. Adding a battery to the implanted device is also a prohibitive solution. No matter how small the battery might be, the total size of the implant increases, limiting the possible implant locations. Moreover, battery life is limited and even rechargeable batteries have a limited number of recharge cycles before they become completely
2 useless. One might suggest energy harvesting as being the solution to the powering problem. However, while this makes use of the external environment as a source of energy (temperature, wind, water etc), these sources of energy are unavailable in the case of implanted devices. The use of implanted and external antennae to wirelessly transmit energy to the implant appears to be a suitable alternative. However, parameters such as human safety, power transfer efficiency and simplicity of electronics imply that the operational frequency must be in the frequency range between and 0 MHz. Due to this constraint, the size of the implanted antenna, which is closely related to the electromagnetic radiation frequency, becomes too large to be implanted. Inductive powering of the implanted device is the most promising solution. A pair of coils, one implanted and one placed outside the human body, form a loosely-coupled transformer, known as the biotelemetry inductive link. When appropriately driven, the external coil creates an electromagnetic field. The implanted coil, which is placed close to the external one, captures a portion of this field and current is produced. Thus, power is transferred to the implant. In order to guarantee human safety, the power transmitted from the external unit must comply with the government safety standards and must not exceed 0 mw/cm. The same biotelemetry inductive link that is used to power the implant can be used to bidirectionally transmit data between the implant and the external device. The implanted and external units along with the biotelemetry inductive link form the biomedical telemetry system. Typical Biomedical Telemetry System Figure shows a simplified block diagram of a typical biotelemetry system. At the external unit, a modulator circuit modulates the binary data ( s and 0 s) to be transmitted from the external to the implanted (internal) unit. The process modulates a property (amplitude, frequency, phase) of a high frequency carrier signal (usually a sine wave with a frequency of the order of tens of MHz) according to the binary data that is to be transmitted, and is necessary for the transference of the signal. A Power Amplifier (PA) receives and amplifies the modulated signal to produce an adequate transmitting power. Figure. Simplified block diagram of a typical biotelemetry system. Powering and bi-directional data transmission take place across the biotelemetry inductive link consisting of the mutually coupled external ( L ) and implanted ( L ) coils. The degree of coupling between these two coils can be described in terms of their mutual inductance, M k LL. The coupling factor, k, shows the proportion of the external coil s ( L ) field, which is captured by the implanted coil ( L ). The external ( C ) and implanted ( C ) capacitors are respectively used to form a tuned-in-series external and a tuned-in-parallel implanted circuit. The purpose of the tuned circuits is to achieve a high voltage at the output of the implanted unit ( V o ), compared with the voltage at the input of the external unit ( V s ) (that is a high voltage gain) at the frequency of the carrier signal ( f 0 ). For example, assuming a carrier signal frequency of f MHz, the magnitude of the voltage gain of the biotelemetry inductive link as well as its dependence on k are shown in Figure. The voltage gain is maximized at the resonant frequency, which in this case is f MHz, and
3 becomes higher for better coupled coils. Figure. Dependence of the magnitude of the voltage gain on the coupling factor, k. At the internal unit, the received waveform is demodulated and the originally transmitted binary data is recovered. At the same time, the received waveform is rectified and regulated to deliver an appropriate supply voltage (and thus power) to the implant. The load resistance of the implanted circuits is represented by the block called Load in Figure. In a real system, the load is timevarying and includes both resistive and capacitive elements. However, for reasons of simplicity, the load is usually considered to be a real resistor ( R l ). The internal unit can itself send information back to the external unit. This information will be transmitted via the biotelemetry inductive link and demodulated at the receiver part of the external unit, as shown in Figure. Data Transmission from the Implant to the External Device Data transmission from the implant to the external device, known as uplink transmission, is necessary for both sensors and stimulators. Sensors measure the biosignals and transmit their measurements to the external world (e.g. medical experts, home/hospital monitoring units). Apart from stimulating specific nerves, stimulators can also send feedback to the external unit regarding the condition of the nerves or the implant itself. Different methods for uplink data transmission are available, but passive telemetry, which is performed by means of load modulation (Load Shift keying, LSK), is the most commonly used technique. Data transmission is achieved by changing the load resistance of the implant, for example by switching a second resistor loading the implant in addition to the load resistor R l. The substitution of the load resistor changes the current in the implanted device, which in turn changes the current in the external device. This change is sensed at the receiver part of the external device, shown in Figure, and is translated into the information which was originally transmitted by the implant. Data Transmission from the External Device to the Implant Data transmission from the external device to the implant, known as downlink transmission, is also, necessary for both sensors and stimulators. Sensors receive information from the external device and adjust, for instance, their rate of sensing the biosignals or performing uplink transmissions. Stimulators stimulate specific nerves with the intensity and frequency of stimulation that is defined by the external device. Amplitude Shift Keying (ASK) techniques, which vary the amplitude of the carrier signal according to the binary data to be transmitted, were the first to be used to downlink data to the implant. They were preferred because of the design simplicity of the modulator and demodulator circuits at the external and implanted devices, but the amount of power transferred to the implant and the data rates achieved at the biotelemetry inductive link were too poor. In order to achieve greater power efficiency and higher data rates, Phase Shift Keying (PSK) techniques, which vary the phase of the carrier signal according to the binary data to be 3
4 transmitted, might be used instead. Binary Phase Shift Keying (BPSK) is the simplest form of the Phase Shift Keying (PSK). In BPSK, the carrier signal can easily be modulated simply by changing its polarity by means of the binary data signal to be transmitted (a bit means that the carrier will be transmitted, while a 0 bit means that minus the carrier will be transmitted). System Design A typical biotelemetry system which downlinks BPSK modulated data to the implanted device is designed in Advanced Design System (ADS), as shown in Figure 3. The system consists of a BPSK modulator circuit as part of the external unit, a Class E Power Amplifier (PA), a biotelemetry inductive link and a BPSK demodulator circuit as part of the implanted unit. Figure 3. Design of a typical BPSK biotelemetry system in ADS. An example of the binary signal at the input of the modulator, V, which will be transmitted via the biotelemetry inductive link ( s and 0 s represented as V and -V voltage levels) and the recovered signal at the output of the implant s demodulator, V, are shown in Figure 4. In the absence of noise, the recovered binary data will be identical to the originally transmitted ones. Figure 4. Binary signal to be transmitted, V and signal at the output of the implant s demodulator, V. Probability of error In order to evaluate the effect of noise added by the biotelemetry inductive link on the transmitted signal, we now assume that Additive White Gaussian Noise (AWGN) distorts the signal before being received by the implant s demodulator. AWGN is the fundamental source of noise associated with real channels. It is said to be additive because it is added to the desired signal being transmitted. The term white refers to the fact that the power spectral density is flat as a function of frequency, meaning that the power of the noise signal is the same at all frequencies. Finally, the probability density function of the noise is a normal distribution, also known as Gaussian distribution. The bit rate and the carrier signal s frequency are considered to be equal to Mbps and 3.56 MHz respectively. The BPSK biotelemetry system is simulated for two values of k ( k 0.03,0. ). The simulated probability of error curves are depicted in Figure 5. The y-axis of the graph indicates the probability of error, or equivalently the bit error rate (BER). The x-axis represents the signal-tonoise ratio (SNR) per bit, defined as the E b/n 0 ratio, where E b is the energy per bit before entering the inductive link channel and N 0 is equal to the noise power (related to the AWGN s variance, σ, as N 0=σ ). 4
5 Figure 5: Simulated probability of error curves for BPSK modulated signals in case of a biotelemetry system with k=0.03, 0.. Figure indicates that at the resonant frequency (which equals the carrier signal s frequency), the biotelemetry inductive link amplifies the BPSK modulated signal at its input. As a result, due to the presence of the inductive link channel, AWGN is expected to distort an amplified version of the initial BPSK modulated signal. Moreover, Figure shows that the higher the inductive link s coupling factor, k, the more the signal will be amplified at its input. As a result, as the value of k gets higher, AWGN is added to an increasingly amplified version of the signal at the input of the inductive link. This explains why, for higher values of k, the same probability of error can be achieved for lower values of the defined E b/n 0 ratio. Conclusions Implanted devices play an increasingly important role in modern medicine, for both the diagnosis and treatment of diseases. This article has addressed the powering of the implanted devices, as well as the bi-directional communication between the implant and the external world, which can be simultaneously performed via a biotelemetry inductive link. It has also discussed BPSK modulation as a means of modulating the data to be transmitted from the external to the implanted device, concluding that this method achieves fast transmission with very few errors, which keep decreasing for better coupled coils. Even though there is considerable scientific work on the development of implanted devices that act as sensors or stimulators, further miniaturization of existing implanted devices or the design of new implanted devices that can perform more complicated tasks, such as organ monitoring, remain areas of great scientific research. The improvement of the circuit design to increase the bit rate in both directions and the power delivered to the implant (while at the same time decreasing the error rate and satisfying several security-related issues) is a very fruitful area for future research. Asimina Kiourti, 00 MSc in Technologies for Broadband Communications Department of Electronic and Electrical Engineering Acknowledgement I would like to express my gratitude to Dr Andreas Demosthenous for his supervision and guidance. I am also grateful to the Greek Foundation of Education and European Culture that funded my MSc studies at UCL during the academic year
6 References Baker, Michael and Sarpeshkar, Rahul, Feedback Analysis and Design of RF Power Links for Low-Power Bionic Systems, IEEE Transactions on Biomedical Circuits and Systems, (March 007): Vol., No., Couch, Leon, Digital and Analog Communication Systems, 5 th ed., Prentice Hall. Finkenzeller, Klaus, RFID Handbook, nd ed., 003. Ghovanloo, Maysam and Atluri, Suresh, An Integrated Full-Wave CMOS Rectifier With Built-In Telemetry for RFID and Implantable Biomedical Applications, IEEE Transactions on Circuits and Systems, (November 008): Vol. 55, No. 0, Goldsmith, Andrea, Wireless Communications, Cambridge University Press, 005. Hochmair, Erwin, System Optimization for Improved Accuracy in Transcutaneous Signal and Power Transmission, IEEE Transactions on Biomedical Engineering, (February 984): Vol. BME-3, No., Hu, Gervais and Sawan, Mohamad, High Power Efficiency Inductive Link with Full-Duplex Data Communication, 9th International Conference on Electronics, Circuits and Systems, (00): Vol., Hu, Yamu et al., A Fully Integrated Low-Power BPSK Demodulator for Implantable Medical Devices, IEEE Transactions on Circuits and Systems, (December 005): Vol. 5, No., Huang, Qiuting and Oberle, Michael, A 0.5-mW Passive Telemetry IC for Biomedical Applications, IEEE Journal of Solid-State Circuits, (July 998): Vol. 33, No. 7, Liu, Wentai et al., A Neuro-Stimulus Chip with Telemetry Unit for Retinal Prosthetic Device, IEEE Journal of Solid-State Circuits, (October 000): Vol. 35, No. 0, Parramon, Jordi et al., Asic-Based Batteryless Implantable Telemetry Microsystem for Recording Purposes, Proceedings of the 9 th International Conference, IEEE/EMBS, (Oct. 30 Nov., 997): Vol. 5, 5-8. Proakis, John and Salehi, Masoud, Communication Systems Engineering, nd ed., Prentice Hall. Razavi, Behzad, RF Microelectronics, ed. Rappaport, Prentice Hall Communications Engineering and Emerging Technologies Series. Sawan, Mohamad and Gosselin, Benoit, CMOS Circuits for Biomedical Implantable Devices, VLSI Circuits for Biomedical Applications, Ed. K. Iniewski, Artech House, Inc Sawan, Mohamad et al., Wireless Smart Implants Dedicated to Multichannel Monitoring and Microstimulation, IEEE Circuits and Systems Magazine, (005): Vol. 5, No., -39. Smith, Brian et al., An Externally Powered, Multichannel, Implantable Stimulator for Versatile Control of Paralyzed Muscle, IEEE Transactions on Biomedical Engineering, (July 987): Vol. BME- 34, No. 7, Sokal, Nathan and Sokal, Alan, Class-E A new Class of High-Efficiency Tuned Single-Ended Switching Power Amplifiers, IEEE Journal of Solid-State Circuits, (June 975): Vol. SC-0, No. 3,
7 Suaning, Gregg and Lovell, Nigel CMOS Neurostimulation ASIC with 00 Channels, Scaleable Output, and Bidirectional Radio-Frequency Telemetry, IEEE Transactions on Biomedical Engineering, (February 00): Vol. 48, No., Tang, Zhengnian et al., Data Transmission from an Implantable Biotelemeter by Load-Shift Keying Using Circuit Configuration Modulator, IEEE Transactions on Biomedical Engineering, (May 995): Vol. 4, No. 5, Troyk, Philip and Schwan, Martin, Closed-Loop Class E Transcutaneous Power and Data Link for Microimplants, IEEE Transactions on Biomedical Engineering, (June 99): Vol. 39, No. 6, Vaillancour, P. et al., EM Radiation Behavior upon Biological Tissues in a Radio-Frequency Power Transfer Link for a Cortical Visual Implant, Proceedings of the 9th International Conference of Engineering in Medicine and Biology Society, IEEE/EMBS, (Oct Nov. 997): Vol. 6, No. 30, Valdastri, Pietro et al., An Implantable Telemetry Platform System for In Vivo Monitoring of Physiological Parameters, IEEE Transactions on Information Technology in Biomedicine, (September 004): Vol. 8, No. 3, Wang, Guoxing et al., Design and Analysis of an Adaptive Transcutaneous Power Telemetry for Biomedical Implants, IEEE Transactions on Circuits and Systems, (October 005): Vol. 5, No. 0, Yuan, Huimei et al., Design and Implementation of Costas Loop Based on FPGA, 3rd IEEE Conference on Industrial Electronics and Applications, (3-5 June 008): Zho, Mingcui et al., A Transcutaneous Data Telemetry System Tolerant to Power Telemetry Interference, Proceedings of the 8th IEEE EMBS Annual International Conference, New York City, USA, (Aug. 30-Sept. 3, 006). 7
Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network
Recent Advances in Electrical Engineering and Electronic Devices Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network Ahmed El-Mahdy and Ahmed Walid Faculty of Information Engineering
Germanium Diode AM Radio
Germanium Diode AM Radio LAB 3 3.1 Introduction In this laboratory exercise you will build a germanium diode based AM (Medium Wave) radio. Earliest radios used simple diode detector circuits. The diodes
a leap ahead in analog
EMV Contactless Payment Systems based on AS3911 Overview and System Simulations Giuliano Manzi, PhD Mannheim, May 23 25, 2012 CST EUROPEAN USER CONFERENCE 2012 a leap ahead in analog OUTLINE AS3911 OVERVIEW
NATIONAL SUN YAT-SEN UNIVERSITY
NATIONAL SUN YAT-SEN UNIVERSITY Department of Electrical Engineering (Master s Degree, Doctoral Program Course, International Master's Program in Electric Power Engineering) Course Structure Course Structures
How To Understand The Power Of An Freddi Tag (Rfid) System
Radio Frequency Identification Done by: Haitham Habli. Table of contents Definition of RFID. Do they need license? RFID vs other identification systems. Classification of RFID systems. Emerge of passive
Computer Aided Design of Home Medical Alert System
Computer Aided Design of Home Medical Alert System Submitted to The Engineering Honors Committee 119 Hitchcock Hall College of Engineering The Ohio State University Columbus, Ohio 43210 By Pei Chen Kan
Impedance Matching and Matching Networks. Valentin Todorow, December, 2009
Impedance Matching and Matching Networks Valentin Todorow, December, 2009 RF for Plasma Processing - Definition of RF What is RF? The IEEE Standard Dictionary of Electrical and Electronics Terms defines
Wireless Medical Telemetry Laboratory
Wireless Medical Telemetry Laboratory 0 Introduction The development of wireless medical telemetry has become an increasingly popular application in recent years. As the elderly population continues to
Application Note Noise Frequently Asked Questions
: What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random
ISSCC 2006 / SESSION 2 / BIOMEDICAL SYSTEMS / 2.5
2.5 Minimally Invasive Retinal Prosthesis Luke Theogarajan 1, John Wyatt 1, Joseph Rizzo 2,3, Bill Drohan 1,3, Mariana Markova 1, Shawn Kelly 1,3, Greg Swider 1,3, Milan Raj 4, Douglas Shire 3,5, Marcus
Special Topics in Security and Privacy of Medical Information. Reminders. Medical device security. Sujata Garera
Special Topics in Security and Privacy of Medical Information Sujata Garera Reminders Assignment due today Project part 1 due on next Tuesday Assignment 2 will be online today evening 2nd Discussion session
2.45 GHz Power and Data Transmission for a Low-Power Autonomous Sensors Platform
9.4.45 GHz Power and Data Transmission for a Low-Power Autonomous Sensors Platform Stefano Gregori 1, Yunlei Li 1, Huijuan Li 1, Jin Liu 1, Franco Maloberti 1, 1 Department of Electrical Engineering, University
MODULATION Systems (part 1)
Technologies and Services on Digital Broadcasting (8) MODULATION Systems (part ) "Technologies and Services of Digital Broadcasting" (in Japanese, ISBN4-339-62-2) is published by CORONA publishing co.,
Changes PN532_Breakout board
Changes PN532_Breakout board Document: Changes PN532_Breakout board Department / Faculty : TechnoCentrum - Radboud University Nijmegen Contact: René Habraken Date: 17 May 2011 Doc. Version: 1.0 Contents
Experiment # (4) AM Demodulator
Islamic University of Gaza Faculty of Engineering Electrical Department Experiment # (4) AM Demodulator Communications Engineering I (Lab.) Prepared by: Eng. Omar A. Qarmout Eng. Mohammed K. Abu Foul Experiment
Implementation of Digital Signal Processing: Some Background on GFSK Modulation
Implementation of Digital Signal Processing: Some Background on GFSK Modulation Sabih H. Gerez University of Twente, Department of Electrical Engineering [email protected] Version 4 (February 7, 2013)
Diode Applications. by Kenneth A. Kuhn Sept. 1, 2008. This note illustrates some common applications of diodes.
by Kenneth A. Kuhn Sept. 1, 2008 This note illustrates some common applications of diodes. Power supply applications A common application for diodes is converting AC to DC. Although half-wave rectification
Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements. Application Note 1304-6
Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements Application Note 1304-6 Abstract Time domain measurements are only as accurate as the trigger signal used to acquire them. Often
ISSCC 2003 / SESSION 13 / 40Gb/s COMMUNICATION ICS / PAPER 13.7
ISSCC 2003 / SESSION 13 / 40Gb/s COMMUNICATION ICS / PAPER 13.7 13.7 A 40Gb/s Clock and Data Recovery Circuit in 0.18µm CMOS Technology Jri Lee, Behzad Razavi University of California, Los Angeles, CA
Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99
Engineering Sciences 151 Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99 WAVE PROPAGATION II: HIGH FREQUENCY SLOTTED LINE AND REFLECTOMETER MEASUREMENTS OBJECTIVES: To build greater
Design of Bidirectional Coupling Circuit for Broadband Power-Line Communications
Journal of Electromagnetic Analysis and Applications, 2012, 4, 162-166 http://dx.doi.org/10.4236/jemaa.2012.44021 Published Online April 2012 (http://www.scirp.org/journal/jemaa) Design of Bidirectional
IFI5481: RF Circuits, Theory and Design
IFI5481: RF Circuits, Theory and Design Lecturer: Prof. Tor A. Fjeldly, UiO og NTNU/UNIK [[email protected]] Assistant: Malihe Zarre Dooghabadi [[email protected]] Syllabus: Lectured material and examples,
BSEE Degree Plan Bachelor of Science in Electrical Engineering: 2015-16
BSEE Degree Plan Bachelor of Science in Electrical Engineering: 2015-16 Freshman Year ENG 1003 Composition I 3 ENG 1013 Composition II 3 ENGR 1402 Concepts of Engineering 2 PHYS 2034 University Physics
INTRODUCTION TO COMMUNICATION SYSTEMS AND TRANSMISSION MEDIA
COMM.ENG INTRODUCTION TO COMMUNICATION SYSTEMS AND TRANSMISSION MEDIA 9/6/2014 LECTURES 1 Objectives To give a background on Communication system components and channels (media) A distinction between analogue
Department of Electrical and Computer Engineering Ben-Gurion University of the Negev. LAB 1 - Introduction to USRP
Department of Electrical and Computer Engineering Ben-Gurion University of the Negev LAB 1 - Introduction to USRP - 1-1 Introduction In this lab you will use software reconfigurable RF hardware from National
Maximizing Receiver Dynamic Range for Spectrum Monitoring
Home Maximizing Receiver Dynamic Range for Spectrum Monitoring Brian Avenell, National Instruments Corp., Austin, TX October 15, 2012 As consumers continue to demand more data wirelessly through mobile
PHASE ESTIMATION ALGORITHM FOR FREQUENCY HOPPED BINARY PSK AND DPSK WAVEFORMS WITH SMALL NUMBER OF REFERENCE SYMBOLS
PHASE ESTIMATION ALGORITHM FOR FREQUENCY HOPPED BINARY PSK AND DPSK WAVEFORMS WITH SMALL NUM OF REFERENCE SYMBOLS Benjamin R. Wiederholt The MITRE Corporation Bedford, MA and Mario A. Blanco The MITRE
Fruit Quality Monitoring System Using the Smart RFID Tag
Fruit Quality Monitoring System Using the Smart RFID Tag Nghia Truong Van 1,1, Lin Sen 1, Kyung Kwon Jung 2, Joo Woong Kim 1 and Ki Hwan Eom 1 1 Department of Electronics and Electrical Engineering Dongguk
Propagation Channel Emulator ECP_V3
Navigation simulators Propagation Channel Emulator ECP_V3 1 Product Description The ECP (Propagation Channel Emulator V3) synthesizes the principal phenomena of propagation occurring on RF signal links
Development of a Radio Frequency Identification Based Attendance Management Application With A Pictorial Database Framework
International Journal of Research in Information Technology (IJRIT) www.ijrit.com ISSN 2001-5569 Development of a Radio Frequency Identification Based Attendance Management Application With A Pictorial
GaAs Switch ICs for Cellular Phone Antenna Impedance Matching
GaAs Switch ICs for Cellular Phone Antenna Impedance Matching IWATA Naotaka, FUJITA Masanori Abstract Recently cellular phones have been advancing toward multi-band and multi-mode phones and many of them
T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p
Data Transmission Concepts and terminology Transmission terminology Transmission from transmitter to receiver goes over some transmission medium using electromagnetic waves Guided media. Waves are guided
Understanding SMD Power Inductors. Application Note. July 2011
Understanding SMD Power Inductors July 2011 Application Note Power inductors play an important role in voltage conversion applications by yielding lower core losses. They are also used to store energy,
Design of a Wireless Medical Monitoring System * Chavabathina Lavanya 1 G.Manikumar 2
Design of a Wireless Medical Monitoring System * Chavabathina Lavanya 1 G.Manikumar 2 1 PG Student (M. Tech), Dept. of ECE, Chirala Engineering College, Chirala., A.P, India. 2 Assistant Professor, Dept.
E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE
E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter
EMC countermeasures for High-Speed Differential Interfaces
TDK EMC Technology Practice Section EMC countermeasures for High-Speed Differential Interfaces How Do Common Mode Filters Suppress EMI in Differential Transmission Circuits? TDK Corporation Application
DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz
Author: Don LaFontaine Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Abstract Making accurate voltage and current noise measurements on op amps in
CHAPTER 1 Introduction 1
Contents CHAPTER 1 Introduction 1 CHAPTER 2 Short-Range Communications Systems 3 2.1 Radio-Frequency Spectrum and Propagation 3 2.1.1 Theory of Electromagnetism and Maxwell s Equations 3 2.1.2 RF Propagation
A Wireless Medical Monitoring Over a Heterogeneous Sensor Network
A Wireless Medical Monitoring Over a Heterogeneous Sensor Network Mehmet R. Yuce, Peng Choong Ng, Chin K. Lee, Jamil Y. Khan, and Wentai Liu Abstract This paper presents a heterogeneous sensor network
REMOTE KEYLESS ENTRY SYSTEM RECEIVER DESIGN
INTRODUCTION: REMOTE KEYLESS ENTRY SYSTEM RECEIVER DESIGN Remote keyless entry (RKE) has captivated automobile buyers, as evidenced by the popularity of RKE on new automobiles and as an aftermarket item.
An equivalent circuit of a loop antenna.
3.2.1. Circuit Modeling: Loop Impedance A loop antenna can be represented by a lumped circuit when its dimension is small with respect to a wavelength. In this representation, the circuit parameters (generally
A Highly Efficient Power Management System for Charging Mobile Phones using RF Energy Harvesting
A Highly Efficient Power Management System for Charging Mobile Phones using RF Energy Harvesting Ajay Sivaramakrishnan, Kailarajan Jeyaprakash Jegadishkumar 2 B.E Electronics & Communication, II year,
Experiment #11: LRC Circuit (Power Amplifier, Voltage Sensor)
Experiment #11: LRC Circuit (Power Amplifier, Voltage Sensor) Concept: circuits Time: 30 m SW Interface: 750 Windows file: RLC.SWS EQUIPMENT NEEDED Science Workshop Interface Power Amplifier (2) Voltage
application note The Measurement and Suppression of RF Interference from TDMA Phones GSM Cellular Introduction
APPLICATION NOTE AN-3 The Measurement and Suppression of RF Interference from TDMA Phones Introduction While the recent explosion of mobile telephone usage has improved the quality of life for many in
Curriculum and Concept Module Development in RF Engineering
Introduction Curriculum and Concept Module Development in RF Engineering The increasing number of applications students see that require wireless and other tetherless network solutions has resulted in
A 1-GSPS CMOS Flash A/D Converter for System-on-Chip Applications
A -GSPS CMOS Flash A/D Converter for System-on-Chip Applications Jincheol Yoo, Kyusun Choi, and Ali Tangel Department of Computer Science & Department of Computer & Engineering Communications Engineering
ELECTRON SPIN RESONANCE Last Revised: July 2007
QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron
EB215E Printed Circuit Board Layout for Improved Electromagnetic Compatibility
Printed Circuit Board Layout for Improved Author: Eilhard Haseloff Date: 10.07.96 Rev: * The electromagnetic compatibility (EMC) of electronic circuits and systems has recently become of increasing significance.
Features. Applications. Transmitter. Receiver. General Description MINIATURE MODULE. QM MODULATION OPTIMAL RANGE 1000m
Features MINIATURE MODULE QM MODULATION OPTIMAL RANGE 1000m 433.05 434.79 ISM BAND 34 CHANNELS AVAILABLE SINGLE SUPPLY VOLTAGE Applications IN VEHICLE TELEMETRY SYSTEMS WIRELESS NETWORKING DOMESTIC AND
AppNote 404 EM MICROELECTRONIC - MARIN SA. EM4095 Application Note RFID. Title: Product Family: TABLE OF CONTENT. Application Note 404
EM MICROELECTRONIC - MARIN SA AppNote 0 Title: Product Family: Application Note 0 EM095 Application Note RFID Part Number: EM095 Keywords: RFID Transceiver, Reader Chip, EM095 Date: 5 September 00 TABLE
Pre-Compliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions
PreCompliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions D. Schneider 1*, S. Tenbohlen 1, W. Köhler 1 1 Institute of Power Transmission and
Digital Modulation. David Tipper. Department of Information Science and Telecommunications University of Pittsburgh. Typical Communication System
Digital Modulation David Tipper Associate Professor Department of Information Science and Telecommunications University of Pittsburgh http://www.tele.pitt.edu/tipper.html Typical Communication System Source
Divvela.Santhosh Raghava Rao [1],Sreevardhan cheerla [2]
Signal Strength Enhancement Using Cellular Repeater On Three Frequency Bands For Low Signal Coverage Areas (GSM900, GSM 1800/DCS, 3G) Divvela.Santhosh Raghava Rao [1],Sreevardhan cheerla [2] [1] B.tech
FPGAs in Next Generation Wireless Networks
FPGAs in Next Generation Wireless Networks March 2010 Lattice Semiconductor 5555 Northeast Moore Ct. Hillsboro, Oregon 97124 USA Telephone: (503) 268-8000 www.latticesemi.com 1 FPGAs in Next Generation
AVX EMI SOLUTIONS Ron Demcko, Fellow of AVX Corporation Chris Mello, Principal Engineer, AVX Corporation Brian Ward, Business Manager, AVX Corporation
AVX EMI SOLUTIONS Ron Demcko, Fellow of AVX Corporation Chris Mello, Principal Engineer, AVX Corporation Brian Ward, Business Manager, AVX Corporation Abstract EMC compatibility is becoming a key design
How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless)
1 2 The range of RF spans 3 KHz (3000 Hz) to 300 GHz (300 million Hz) Frequencies of RF devices range from the low frequency AM broadcasts (80 MHz) to higher frequency mobile phones (1900 MHz) smart meters
RF Measurements Using a Modular Digitizer
RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.
Case Study Competition 2013. Be an engineer of the future! Innovating cars using the latest instrumentation!
Case Study Competition 2013 Be an engineer of the future! Innovating cars using the latest instrumentation! The scenario You are engineers working on a project team that is tasked with the development
RF EXPOSURE LIMITS AND TESTING REQUIREMENTS
RF EXPOSURE LIMITS AND TESTING REQUIREMENTS Jay Moulton Vice President March 12, 2013 1 OVERVIEW Specific Absorption Rate (SAR) and Maximum Permissible Exposure (MPE) Standards and Limits Evaluation Methods
Optimizing IP3 and ACPR Measurements
Optimizing IP3 and ACPR Measurements Table of Contents 1. Overview... 2 2. Theory of Intermodulation Distortion... 2 3. Optimizing IP3 Measurements... 4 4. Theory of Adjacent Channel Power Ratio... 9 5.
How To Understand The Theory Of Time Division Duplexing
Multiple Access Techniques Dr. Francis LAU Dr. Francis CM Lau, Associate Professor, EIE, PolyU Content Introduction Frequency Division Multiple Access Time Division Multiple Access Code Division Multiple
Security & Chip Card ICs SLE 44R35S / Mifare
Security & Chip Card ICs SLE 44R35S / Mifare Intelligent 1 Kbyte EEPROM with Interface for Contactless Transmission, Security Logic and Anticollision according to the MIFARE -System Short Product Info
A Faster, More Accurate Approach for System-Level Performance Verification of a Wireless RFIC Design
A Faster, More Accurate Approach for System-Level Performance Verification of a Wireless RFIC Design By Nebabie Kebebew and Rich Davis, Cadence Wireless RFIC designs are growing more complex, increasing
Broadband Push-Pull Power Amplifier Design at Microwave Frequencies
Broadband Push-Pull Power Amplifier Design at Microwave Frequencies Robert Smith and Prof. Steve Cripps Centre for High Frequency Engineering, Cardiff University [email protected] A broadband, high
RFID Design Principles
RFID Design Principles Second Edition Harvey Lehpamer ARTECH HOUSE BOSTON LONDON artechhouse.com Contents CHAPTER 1 Introduction CHAPTER 2 Short-Range Communications Systems 2.1 Radio-Frequency Spectrum
Radio Transmission Performance of EPCglobal Gen-2 RFID System
Radio Transmission Performance of EPCglobal Gen-2 RFID System Manar Mohaisen, HeeSeok Yoon, and KyungHi Chang The Graduate School of Information Technology & Telecommunications INHA University Incheon,
Application Note SAW-Components
Application Note SAW-Components Principles of SAWR-stabilized oscillators and transmitters. App: Note #1 This application note describes the physical principle of SAW-stabilized oscillator. Oscillator
Implementing Digital Wireless Systems. And an FCC update
Implementing Digital Wireless Systems And an FCC update Spectrum Repacking Here We Go Again: The FCC is reallocating 600 MHz Frequencies for Wireless Mics 30-45 MHz (8-m HF) 174-250 MHz (VHF) 450-960 MHz
Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.
Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Abstract. I have been asked to put together a detailed article on a SWR meter. In this article I will deal
The most recent advances in glass encapsulation of microelectronic devices offer important
Improvements in Glass Encapsulation Technology Offer Significant Advantages for Implantable Medical Devices By Frédéric Mauron Senior VP, Biomedical Engineering Director, Business & Administration Valtronic
DT3: RF On/Off Remote Control Technology. Rodney Singleton Joe Larsen Luis Garcia Rafael Ocampo Mike Moulton Eric Hatch
DT3: RF On/Off Remote Control Technology Rodney Singleton Joe Larsen Luis Garcia Rafael Ocampo Mike Moulton Eric Hatch Agenda Radio Frequency Overview Frequency Selection Signals Methods Modulation Methods
Technical Article. NFiC: a new, economical way to make a device NFC-compliant. Prashant Dekate
Technical NFiC: a new, economical way to make a device NFC-compliant Prashant Dekate NFiC: a new, economical way to make a device NFC-compliant Prashant Dekate The installed base of devices with Near Field
ISSCC 2003 / SESSION 10 / HIGH SPEED BUILDING BLOCKS / PAPER 10.5
ISSCC 2003 / SESSION 10 / HIGH SPEED BUILDING BLOCKS / PAPER 10.5 10.5 Broadband ESD Protection Circuits in CMOS Technology Sherif Galal, Behzad Razavi Electrical Engineering Department, University of
Lecture 24. Inductance and Switching Power Supplies (how your solar charger voltage converter works)
Lecture 24 Inductance and Switching Power Supplies (how your solar charger voltage converter works) Copyright 2014 by Mark Horowitz 1 Roadmap: How Does This Work? 2 Processor Board 3 More Detailed Roadmap
Adaptive Equalization of binary encoded signals Using LMS Algorithm
SSRG International Journal of Electronics and Communication Engineering (SSRG-IJECE) volume issue7 Sep Adaptive Equalization of binary encoded signals Using LMS Algorithm Dr.K.Nagi Reddy Professor of ECE,NBKR
Analysis of Immunity by RF Wireless Communication Signals
64 PIERS Proceedings, Guangzhou, China, August 25 28, 2014 Analysis of Immunity by RF Wireless Communication Signals Hongsik Keum 1, Jungyu Yang 2, and Heung-Gyoon Ryu 3 1 EletroMagneticwave Technology
Performance of Quasi-Constant Envelope Phase Modulation through Nonlinear Radio Channels
Performance of Quasi-Constant Envelope Phase Modulation through Nonlinear Radio Channels Qi Lu, Qingchong Liu Electrical and Systems Engineering Department Oakland University Rochester, MI 48309 USA E-mail:
Abstract. Cycle Domain Simulator for Phase-Locked Loops
Abstract Cycle Domain Simulator for Phase-Locked Loops Norman James December 1999 As computers become faster and more complex, clock synthesis becomes critical. Due to the relatively slower bus clocks
Laboratory 4: Feedback and Compensation
Laboratory 4: Feedback and Compensation To be performed during Week 9 (Oct. 20-24) and Week 10 (Oct. 27-31) Due Week 11 (Nov. 3-7) 1 Pre-Lab This Pre-Lab should be completed before attending your regular
Product Datasheet P1110 915 MHz RF Powerharvester Receiver
DESCRIPTION The Powercast P1110 Powerharvester receiver is an RF energy harvesting device that converts RF to DC. Housed in a compact SMD package, the P1110 receiver provides RF energy harvesting and power
Radiated emission measurement of a cell phone processor module using TEM cell
, pp.48-53 http://dx.doi.org/10.14257/astl.2013.28.09 Radiated emission measurement of a cell phone processor module using TEM cell Fayu Wan 1,2*, Qi Liu 3, Jian Shen 3, Jin Wang 3 and Nigel Linge 4 1
Intelligent Traffic Control and Patient Monitoring For Efficient Ambulance Service
Intelligent Traffic Control and Patient Monitoring For Efficient Ambulance Service Suneesh S 1, Muhammed Mansoor P K 2, Bibin Varghees 3, Nong Taku 4 1234 Under Graduate Students, M.A. College of Engineering,
DRM compatible RF Tuner Unit DRT1
FEATURES DRM compatible RF Tuner Unit DRT1 High- Performance RF Tuner Frequency Range: 10 KHz to 30 MHz Input ICP3: +13,5dBm, typ. Noise Figure @ full gain: 14dB, typ. Receiver Factor: -0,5dB, typ. Input
Visual System Simulator White Paper
Visual System Simulator White Paper UNDERSTANDING AND CORRECTLY PREDICTING CRITICAL METRICS FOR WIRELESS RF LINKS Understanding and correctly predicting cellular, radar, or satellite RF link performance
The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT
The Effect of Network Cabling on Bit Error Rate Performance By Paul Kish NORDX/CDT Table of Contents Introduction... 2 Probability of Causing Errors... 3 Noise Sources Contributing to Errors... 4 Bit Error
Voice services over Adaptive Multi-user Orthogonal Sub channels An Insight
TEC Voice services over Adaptive Multi-user Orthogonal Sub channels An Insight HP 4/15/2013 A powerful software upgrade leverages quaternary modulation and MIMO techniques to improve network efficiency
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER W. Li, J. Vandewege Department of Information Technology (INTEC) University of Gent, St.Pietersnieuwstaat 41, B-9000, Gent, Belgium Abstract: Precision
Sigma- Delta Modulator Simulation and Analysis using MatLab
Computer and Information Science; Vol. 5, No. 5; 2012 ISSN 1913-8989 E-ISSN 1913-8997 Published by Canadian Center of Science and Education Sigma- Delta Modulator Simulation and Analysis using MatLab Thuneibat
LM1596 LM1496 Balanced Modulator-Demodulator
LM1596 LM1496 Balanced Modulator-Demodulator General Description The LM1596 LM1496 are doubled balanced modulator-demodulators which produce an output voltage proportional to the product of an input (signal)
MIMO Antenna Systems in WinProp
MIMO Antenna Systems in WinProp AWE Communications GmbH Otto-Lilienthal-Str. 36 D-71034 Böblingen [email protected] Issue Date Changes V1.0 Nov. 2010 First version of document V2.0 Feb. 2011
EMC STANDARDS STANDARDS AND STANDARD MAKING BODIES. International. International Electrotechnical Commission (IEC) http://www.iec.
EMC STANDARDS The EMC standards that a particular electronic product must meet depend on the product application (commercial or military) and the country in which the product is to be used. These EMC regulatory
FREQUENCY RESPONSE ANALYZERS
FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road
Simplifying System Design Using the CS4350 PLL DAC
Simplifying System Design Using the CS4350 PLL 1. INTRODUCTION Typical Digital to Analog Converters (s) require a high-speed Master Clock to clock their digital filters and modulators, as well as some
Satellite Telemetry, Tracking and Control Subsystems
Satellite Telemetry, Tracking and Control Subsystems Col John E. Keesee 1 Overview The telemetry, tracking and control subsystem provides vital communication to and from the spacecraft TT&C is the only
A Study of the Design of Wireless Medical Sensor Network based u- Healthcare System
, pp.91-96 http://dx.doi.org/10.14257/ijbsbt.2014.6.3.11 A Study of the Design of Wireless Medical Sensor Network based u- Healthcare System Ronnie D. Caytiles and Sungwon Park 1* 1 Hannam University 133
Charged cable event. 1 Goal of the ongoing investigation. 2 Energy sources for the CDE. Content
Charged cable event David Pommerenke, [email protected], 916 785 4550 Last update: Feb.23, 2001 Content Goal Energy sources, which may lead to CDE. Complexity of the different discharge modes. Possible