Neuman, M. R. Biopotential Electrodes. The Biomedical Engineering Handbook: Second Edition. Ed. Joseph D. Bronzino Boca Raton: CRC Press LLC, 2000
|
|
|
- Evan Laurence Sanders
- 10 years ago
- Views:
Transcription
1 Neuman, M. R. Biopotential Electrodes. The Biomedical Engineering Handbook: Second Edition. Ed. Joseph D. Bronzino Boca Raton: CRC Press LLC, 2000
2 48 Biopotential Electrodes Michael R. Neuman Joint Program in Biomedical Engineering, The University of Memphis and University of Tennessee, Memphis 48.1 Sensing Bioelectric Signals 48.2 Electrical Characteristics 48.3 Practical Electrodes for Biomedical Measurements Body-Surface Biopotential Electrodes Intracavity and Intratissue Electrodes Microelectrodes Electrodes Fabricated Using Microelectronic Technology 48.4 Biomedical Applications Biologic systems frequently have electric activity associated with them. This activity can be a constant dc electric field, a constant flux of charge-carrying particles or current, or a time-varying electric field or current associated with some time-dependent biologic or biochemical phenomenon. Bioelectric phenomena are associated with the distribution of ions or charged molecules in a biologic structure and the changes in this distribution resulting from specific processes. These changes can occur as a result of biochemical reactions, or they can emanate from phenomena that alter local anatomy. One can find bioelectric phenomena associated with just about every organ system in the body. Nevertheless, a large proportion of these signals are associated with phenomena that are at the present time not especially useful in clinical medicine and represent time-invariant, low-level signals that are not easily measured in practice. There are, however, several signals that are of diagnostic significance or that provide a means of electronic assessment to aid in understanding biologic systems. These signals, their usual abbreviations, and the systems they measure are listed in Table Of these, the most familiar is the electrocardiogram, a signal derived from the electric activity of the heart. This signal is widely used in diagnosing disturbances in cardiac rhythm, signal conduction through the heart, and damage due to cardiac ischemia and infarction. The electromyogram is used for diagnosing neuromuscular diseases, and the electroencephalogram is important in identifying brain dysfunction and evaluating sleep. The other signals listed in Table 48.1 are currently of lesser diagnostic significance but are, nevertheless, used for studies of the associated organ systems. Although Table 48.1 and the above discussion are concerned with bioelectric phenomena in animals and these techniques are used primarily in studying mammals, bioelectric signals also arise from plants [1]. These signals are generally steady-state or slowly changing, as opposed to the time-varying signals listed in Table An extensive literature exists on the origins of bioelectric signals, and the interested reviewer is referred to the text by Plonsey and Barr for a general overview of this area [2] Sensing Bioelectric Signals The mechanism of electric conductivity in the body involves ions as charge carriers. Thus, picking up bioelectric signals involves interacting with these ionic charge carriers and transducing ionic currents into electric currents required by wires and electronic instrumentation. This transducing function is carried out by electrodes that consist of electrical conductors in contact with the aqueous ionic solutions
3 TABLE 48.1 Bioelectric Signals Sensed by Biopotential Electrodes and Their Sources Bioelectric Signal Abbreviation Biologic Source Electrocardiogram ECG Heart as seen from body surface Cardiac electrogram Heart as seen from within Electromyogram EMG Muscle Electroencephalogram EEG Brain Electrooptigram EOG Eye dipole field Electroretinogram ERG Eye retina Action potential Nerve or muscle Electrogastrogram EGG Stomach Galvanic skin reflex GSR Skin of the body. The interaction between electrons in the electrodes and ions in the body can greatly affect the performance of these sensors and requires that specific considerations be made in their application. At the interface between an electrode and an ionic solution redox (oxidation-reduction), reactions need to occur for a charge to be transferred between the electrode and the solution. These reactions can be represented in general by the following equations: n C C + ne 1 + (48.1) m A 1A+ me (48.2) where n is the valence of cation material C, and m is the valence of anion material, A. For most electrode systems, the cations in solution and the metal of the electrodes are the same, so the atoms C are oxidized when they give up electrons and go into solution as positively charged ions. These ions are reduced when the process occurs in the reverse direction. In the case of the anion reaction, Eq. (48.2), the directions for oxidation and reduction are reversed. For best operation of the electrodes, these two reactions should be reversible, that is, it should be just as easy for them to occur in one direction as the other. The interaction between a metal in contact with a solution of its ions produces a local change in the concentration of the ions in solution near the metal surface. This causes charge neutrality not to be maintained in this region, causing the electrolyte surrounding the metal to be at a different electrical potential from the rest of the solution. Thus, a potential difference known as the half-cell potential is established between the metal and the bulk of the electrolyte. It is found that different characteristic potentials occur for different materials, and some of these potentials are summarized in Table These half-cell potentials can be important when using electrodes for low frequency or dc measurements. The relationship between electric potential and ionic concentrations or, more precisely, ionic activities is frequently considered in electrochemistry. Most commonly two ionic solutions of different activity are separated by an ion-selective semipermeable membrane that allows one type of ion to pass freely through TABLE 48.2 Half-cell Potentials for Materials and Reactions Encountered in Biopotential Measurement Metal and Reaction Half-cell Potential, V Al Al e Ni Ni e H 2 2H + + 2e (by definition) Ag + Cl AgCl + e Ag Ag + + e Au Au + + e
4 the membrane. It can be shown that an electric potential E will exist between the solutions on either side of the membrane, based upon the relative activity of the permeable ions in each of these solutions. This relationship is known as the Nernst equation RT E = nf a 1 ln a 2 (48.3) where a 1 and a 2 are the activities of the ions on either side of the membrane, R is the universal gas constant, T is the absolute temperature, n is the valence of the ions, and F is the Faraday constant. More detail on this relationship can be found in Chapter 49. When no electric current flows between an electrode and the solution of its ions or across an ionpermeable membrane, the potential observed should be the half-cell potential or the Nernst potential, respectively. If, however, there is a current, these potentials can be altered. The difference between the potential at zero current and the measured potentials while current is passing is known as the over voltage and is the result of an alteration in the charge distribution in the solution in contact with the electrodes or the ion-selective membrane. This effect is known as polarization and can result in diminished electrode performance, especially under conditions of motion. There are three basic components to the polarization over potential: the ohmic, the concentration, and the activation over potentials. Of these, the activation over potential is of greatest concern in bioelectric measurements. More details on these over potentials can be found in electrochemistry or biomedical instrumentation texts [4]. Perfectly polarizable electrodes pass a current between the electrode and the electrolytic solution by changing the charge distribution within the solution near the electrode. Thus, no actual current crosses the electrode-electrolyte interface. Nonpolarized electrodes, however, allow the current to pass freely across the electrode-electrolyte interface without changing the charge distribution in the electrolytic solution adjacent to the electrode. Although these types of electrodes can be described theoretically, neither can be fabricated in practice. It is possible, however, to come up with electrode structures that closely approximate their characteristics. Electrodes made from noble metals such as platinum are often highly polarizable. A charge distribution different from that of the bulk electrolytic solution is found in the solution close to the electrode surface. Such a distribution can create serious limitations when movement is present and the measurement involves low frequency or even dc signals. If the electrode moves with respect to the electrolytic solution, the charge distribution in the solution adjacent to the electrode surface will change, and this will induce a voltage change in the electrode that will appear as motion artifact in the measurement. Thus, for most biomedical measurements, nonpolarizable electrodes are preferred to those that are polarizable. The silver silver chloride electrode is one that has characteristics similar to a perfectly nonpolarizable electrode and is practical for use in many biomedical applications. The electrode (Fig. 48.1a) consists of a silver base structure that is coated with a layer of the ionic compound silver chloride. Some of the silver chloride when exposed to light is reduced to metallic silver, so a typical silver silver chloride electrode has finely divided metallic silver within a matrix of silver chloride on its surface. Since the silver chloride is relatively insoluble in aqueous solutions, this surface remains stable. Because there is minimal polarization associated with this electrode, motion artifact is reduced compared to polarizable electrodes such as the platinum electrode. Furthermore, due to the reduction in polarization, there is also a smaller effect of frequency on electrode impedance, especially at low frequencies. Silver silver chloride electrodes of this type can be fabricated by starting with a silver base and electrolytically growing the silver chloride layer on its surface [4]. Although an electrode produced in this way can be used for most biomedical measurements, it is not a robust structure, and pieces of the silver chloride film can be chipped away after repeated use of the structure. A structure with greater mechanical stability is the sintered silver silver chloride electrode in Fig. 48.1b. This electrode consists of a silver lead wire surrounded by a sintered cylinder made up of finely divided silver and silver-chloride powder pressed together.
5 FIGURE 48.1 Silver silver electrodes for biopotential measurements: (a) metallic silver with a silver chloride surface layer and (b) sintered electrode structure. The lower views show the electrodes in cross-section. In addition to its nonpolarizable behavior, the silver silver chloride electrode exhibits less electrical noise than the equivalent polarizable electrodes. This is especially true at low frequencies, and so silver silver chloride electrodes are recommended for measurements involving very low voltages for signals that are made up primarily of low frequencies. A more detailed description of silver silver chloride electrodes and methods to fabricate these devices can be found in Janz and Ives [5] and biomedical instrumentation textbooks [4] Electric Characteristics The electric characteristics of biopotential electrodes are generally nonlinear and a function of the current density at their surface. Thus, having the devices represented by linear models requires that they be operated at low potentials and currents. Under these idealized conditions, electrodes can be represented by an equivalent circuit of the form shown in Fig In this circuit R d and C d are components that represent the impedance associated with the electrode-electrolyte interface and polarization at this interface. R s is the series resistance associated with interfacial effects and the resistance of the electrode materials themselves. FIGURE 48.2 The equivalent circuit for a biopotential electrode.
6 FIGURE 48.3 An example of biopotential electrode impedance as a function of frequency. Characteristic frequencies will be somewhat different for electrode different geometries and materials. TABLE 48.3 The Effect of Electrode Properties on Electrode Impedance Property Change in Property Changes in Electrode Impedance Surface area Polarization At low frequencies Surface roughness Radius of curvature Surface contamination The battery E hc represents the half-cell potential described above. It is seen that the impedance of this electrode will be frequency dependent, as illustrated in Fig At low frequencies the impedance is dominated by the series combination of R s and R d, whereas at higher frequencies C d bypasses the effect of R d so that the impedance is now close to R s. Thus, by measuring the impedance of an electrode at high and low frequencies, it is possible to determine the component values for the equivalent circuit for that electrode. The electrical characteristics of electrodes are affected by many physical properties of these electrodes. Table 48.3 lists some of the more common physical properties of electrodes and qualitatively indicates how these can affect electrode impedance Practical Electrodes for Biomedical Measurements Many different forms of electrodes have been developed for different types of biomedical measurements. To describe each of these would go beyond the constraints of this article, but some of the more commonly used electrodes are presented in this section. The reader is referred to the monograph by Geddes for more details and a wider selection of practical electrodes [6]. Body-Surface Biopotential Electrodes This category includes electrodes that can be placed on the body surface for recording bioelectric signals. The integrity of the skin is not compromised when these electrodes are applied, and they can be used for short-term diagnostic recording such as taking a clinical electrocardiogram or long-term chronic recording such as occurs in cardiac monitoring.
7 Metal Plate Electrodes The basic metal plate electrode consists of a metallic conductor in contact with the skin with a thin layer of an electrolyte gel between the metal and the skin to establish this contact. Examples of metal plate electrodes are seen in Fig. 48.4a. Metals commonly used for this type of electrode include German silver (a nickel-silver alloy), silver, gold, and platinum. Sometimes these electrodes are made of a foil of the metal so as to be flexible, and sometimes they are produced in the form of a suction electrode (Fig. 48.4b) to make it easier to attach the electrode to the skin to make a measurement and then move it to another point to repeat the measurement. These types of electrodes are used primarily for diagnostic recordings of biopotentials such as the electrocardiogram or the electroencephalogram. Metal disk electrodes with a gold surface in a conical shape such as shown in Fig. 48.4c are frequently used for EEG recordings. The apex of the cone is open so that electrolyte gel or paste can be introduced to both make good contact between the electrode and the head and to allow this contact medium to be replaced should it dry out during its use. FIGURE 48.4 Examples of different skin electrodes: (a) metal plate electrodes, (b) suction electrode for ECG, (c) metal cup EEG electrode, (d) recessed electrode, (e) disposable electrode with electrolyte-impregnated sponge (shown in cross-section), (f) disposable hydrogel electrode (shown in cross-section), (g) thin-film electrode for use with neonates (shown in cross-section), (h) carbon-filled elastomer dry electrode.
8 FIGURE 48.4 (continued)
9 Electrodes for Chronic Patient Monitoring Long-term monitoring of biopotentials such as the electrocardiogram as performed by cardiac monitors places special constraints on the electrodes used to pick up the signals. These electrodes must have a stable interface between them and the body, and frequently nonpolarizable electrodes are, therefore, the best for this application. Mechanical stability of the interface between the electrode and the skin can help to reduce motion artifact, and so there are various approaches to reduce interfacial motion between the electrode and the coupling electrolyte or the skin. Figure 48.4d is an example of one approach to reduce motion artifact by recessing the electrode in a cup of electrolytic fluid or gel. The cup is then securely fastened to the skin surface using a double-sided adhesive ring. Movement of the skin with respect to the electrode may affect the electrolyte near the skin-electrolyte interface, but the electrode-electrolyte interface can be several millimeters away from this location, since it is recessed in the cup. The fluid movement is unlikely to affect the recessed electrode-electrolyte interface as compared to what would happen if the electrode was separated from the skin by just a thin layer of electrolyte. The advantages of the recessed electrode can be realized in a simpler design that lends itself to mass production through automation. This results in low per-unit cost so that these electrodes can be considered disposable. Figure 48.4e illustrates such an electrode in cross section. The electrolyte layer now consists of an open-celled sponge saturated with a thickened (high-viscosity) electrolytic solution. The sponge serves the same function as the recess in the cup electrodes and is coupled directly to a silver silver chloride electrode. Frequently, the electrode itself is attached to a clothing snap through an insulatingadhesive disk that holds the structure against the skin. This snap serves as the point of connection to a lead wire. Many commercial versions of these electrodes in various sizes are available, including electrodes with a silver silver chloride interface or ones that use metallic silver as the electrode material. A recently developed modification of this basic monitoring electrode structure is shown in Fig. 48.4f. In this case the metal electrode is a silver foil with a surface coating of silver chloride. The foil gives the electrode increased flexibility to fit more closely over body contours. Instead of using the sponge, a hydrogel film (really a sponge on a microscopic level) saturated with an electrolytic solution and formed from materials that are very sticky is placed over the electrode surface. The opposite surface of the hydrogel layer can be attached directly to the skin, and since it is very sticky, no additional adhesive is needed. The mobility and concentration of ions in the hydrogel layer is generally lower than for the electrolytic solution used in the sponge or the cup. This results in an electrode that has a higher source impedance as compared to these other structures. An important advantage of this structure is its ability to have the electrolyte stick directly on the skin. This greatly reduces interfacial motion between the skin surface and the electrolyte, and hence there is a smaller amount of motion artifact in the signal. This type of hydrogel electrode is, therefore, especially valuable in monitoring patients who move a great deal or during exercise. Thin-film flexible electrodes such as shown in Fig. 48.4g have been used for monitoring neonates. They are basically the same as the metal plate electrodes; only the thickness of the metal in this case is less than a micrometer. These metal films need to be supported on a flexible plastic substrate such as polyester or polyimide. The advantage of using only a thin metal layer for the electrode lies in the fact that these electrodes will then become x-ray transparent. This is especially important in infants where repeated placement and removal of electrodes, so that x-rays may be taken, can cause substantial skin irritation. Electrodes that do not use artificially applied electrolyte solutions or gels and, therefore, are often referred to as dry electrodes have been used in some monitoring applications. These sensors as illustrated in Fig. 48.4h can be placed on the skin and held in position by an elastic band or tape. They are made up of a graphite or metal-filled polymer such as silicone. The conducting particles are ground into a fine powder, and this is added to the silicone elastomer before it cures so to produce a conductive material with physical properties similar to that of the elastomer. When held against the skin surface, these electrodes establish contact with the skin without the need for an electrolytic fluid or gel. In actuality such a layer is formed by sweat under the electrode surface. For this reason these electrodes tend to perform better after they have been left in place for an hour or two so that this layer forms. Some investigators have found that placing a drop of physiologic saline solution on the skin before applying
10 the electrode accelerates this process. This type of electrode has found wide application in home infant cardiorespiratory monitoring because of the ease with which it can be applied by untrained caregivers. Intracavitary and Intratissue Electrodes Electrodes can be placed within the body for biopotential measurements. These electrodes are generally smaller than skin surface electrodes and do not require special electrolytic coupling fluid, since natural body fluids serve this function. There are many different designs for these internal electrodes, and only a few examples are given in the following paragraphs. Basically these electrodes can be classified as needle electrodes, which can be used to penetrate the skin and tissue to reach the point where the measurement is to be made, or they are electrodes that can be placed in a natural cavity or surgically produced cavity in tissue. Figure 48.5 illustrates some of these internal electrodes. A catheter tip or probe electrode is placed in a naturally occurring cavity in the body such as in the gastrointestinal system. A metal tip or segment on a catheter makes up the electrode. The catheter or, in the case where there is no hollow lumen, probe, is inserted into the cavity so that the metal electrode FIGURE 48.5 Examples of different internal electrodes: (a) catheter or probe electrode, (b) needle electrode, (c) coaxial needle electrode, (d) coiled wire electrode. (Reprinted with permission from Webster JG (ed) Medical Instrumentation: Application and Design, Houghton Mifflin, Boston.)
11 makes contact with the tissue. A lead wire down the lumen of the catheter or down the center of the probe connects the electrode to the external circuitry. The basic needle electrode shown in Fig. 48.5b consists of a solid needle, usually made of stainless steel, with a sharp point. An insulating material coats the shank of the needle up to a millimeter or two of the tip so that the very tip of the needle remains exposed. When this structure is placed in tissue such as skeletal muscle, electrical signals can be picked up by the exposed tip. One can also make needle electrodes by running one or more insulated wires down the lumen of a standard hypodermic needle. The electrode as shown in Fig. 48.5c is shielded by the metal of the needle and can be used to pick up very localized signals in tissue. Fine wires can also be introduced into tissue using a hypodermic needle, which is then withdrawn. This wire can remain in tissue for acute or chronic measurements. Caldwell and Reswick have used fine coiled wire electrodes in skeletal muscle for several years without adverse effects [7]. Microelectrodes The electrodes described in the previous paragraphs have been applied to studying bioelectric signals at the organism, organ, or tissue level but not at the cellular level. To study the electric behavior of cells, electrodes that are themselves smaller than the cells being studied need to be used. Three types of electrodes have been described for this purpose: etched metal electrodes, micropipette electrodes, and metal-film-coated micropipette electrodes. The metal microelectrode is essentially a subminiature version of the needle electrode described in the previous section (Fig. 48.6a). In this case, a strong metal such as tungsten is used. One end of this wire is etched electrolytically to give tip diameters on the order of a few micrometers. The structure is insulated up to its tip, and it can be passed through the membrane of a cell to contact the cytosol. The advantage of these electrodes is that they are both small and robust and FIGURE 48.6 Microelectrodes: (a) metal, (b) micropipette, (c) thin metal film on micropipette. (Reprinted with permission from Webster JC (ed) Medical Instrumentation: Application and Design, Houghton Mifflin, Boston.)
12 can be used for neurophysiologic studies. Their principal disadvantage is the difficulty encountered in their fabrication and their high source impedance. The second and most frequently used type of microelectrode is the glass micropipette. This structure, as illustrated in Fig. 48.6b consists of a fine glass capillary drawn to a very narrow point and filled with an electrolytic solution. The point can be as narrow as a fraction of a micrometer, and the dimensions of this electrode are strongly dependent on the skill of the individual drawing the tip. The electrolytic solution in the lumen serves as the contact between the interior of the cell through which the tip has been impaled and a larger conventional electrode located in the shank of the pipette. These electrodes also suffer from high source impedances and fabrication difficulty. A combined form of these two types of electrodes can be achieved by depositing a metal film over the outside surface of a glass micropipette as shown in Fig. 48.6c. In this case, the strength and smaller dimensions of the micropipette can be used to support films of various metals that are insulated by an additional film up to a point very close to the actual tip of the electrode structure. These electrodes have been manufactured in quantity and made available as commercial products. Since they combine the features of both the metal and the micropipette electrodes, they also suffer from many of the same limitations. They do, however, have the advantage of flexibility due to the capability of being able to make films of different metals on the micropipette surface without having to worry about the strength of the metal, as would be the case if the metal were used alone. Electrodes Fabricated Using Microelectronic Technology Modern microelectronic technology can be used to fabricate many different types of electrodes for specific biomedical applications. For example, dry electrodes with high source resistances or microelectrodes with similar characteristics can be improved by incorporating a microelectronic amplifier for impedance conversion right on the electrode itself. In the case of the conventional-sized electrodes, a metal disk 5 10 mm in diameter can have a high input impedance microelectronic amplifier configured as a follower integrated into the back of the electrode so that localized processing of the high source impedance signal can produce one of lower, more practical impedance for signal transmission [8]. Single- and multiple-element electrodes can be made from thin-film or silicon technology. Mastrototaro and colleagues have demonstrated probes for measuring intramyocardial potentials using thin, patterned gold films on polyimide or oxidised molybdenum substrates [9]. When electrodes are made from pieces of micromachined silicon, it is possible to integrate an amplifier directly into the electrode [10]. Multichannel amplifiers or multiplexers can be used with multiple electrodes on the same probe. Electrodes for contact with individual nerve fibers can be fabricated using micromachined holes in a silicon chip that are just big enough to pass a single growing axon. Electrical contacts on the sides of these holes can then be used to pick up electrical activity from these nerves [11]. These examples are just a few of the many possibilities that can be realized using microelectronics and three-dimensional micromachining technology to fabricate specialized electrodes Biomedical Applications Electrodes can be used to perform a wide variety of measurements of bioelectric signals. An extensive review of this would be beyond the scope of this chapter, but some typical examples of applications are highlighted in Table The most popular application for biopotential electrodes is in obtaining the electrocardiogram for diagnostic and patient-monitoring applications. A substantial commercial market exists for various types of electrocardiographic electrodes, and many of the forms described in the previous section are available commercially. Other electrodes for measuring bioelectric potentials for application in diagnostic medicine are indicated in Table Research applications of biopotential electrodes are highly varied and specific for individual studies. Although a few examples are given in Table 48.4, the field is far too broad to be completely covered here. Biopotential electrodes are one of the most common biomedical sensors used in clinical medicine. Although their basic principle of operation is the same for most applications, they take on many forms
13 TABLE 48.4 Examples of Applications of Biopotential Electrodes Application Biopotential Type of Electrode Cardiac monitoring ECG Ag/AgCl with sponge Ag/AgCl with hydrogel Infant cardiopulmonary monitoring ECG impedance Ag/AgCl with sponge Ag/AgCl with hydrogel Thin-film Filled elastomer dry Sleep encephalography EEG Gold cups Ag/AgCl cups Active electrodes Diagnostic muscle activity EMG Needle Cardiac electrograms Electrogram Intracardiac probe Implanted telemetry of biopotentials ECG Stainless steel wire loops EMG Platinum disks Eye movement EOG Ag/AgCl with hydrogel and are used in the measurement of many types of bioelectric phenomena. They will continue to play an important role in biomedical instrumentation systems. References 1. Yoshida T, Hayashi K, Toko K The effect of anoxia on the spatial pattern of electric potential formed along the root. Ann Bot 62(5): Plonsey R, Barr RC Bioelectricity, New York, Plenum. 3. Weast RC (ed) Handbook of Chemistry and Physics, 55th ed, Boca Raton, Fla, CRC Press. 4. Webster JG (ed) Medical Instrumentation: Application and Design, Boston, Houghton Mifflin. 5. Janz GI, Ives DJG Silver silver chloride electrodes. Ann N Y Acad Sci 148: Geddes LA Electrodes and the Measurement of Bioelectric Events, New York, Wiley. 7. Caldwell CW, Reswick JB A percutaneous wire electrode for chronic research use. IEEE Trans Biomed Eng 22: Ko WH, Hynecek J Dry electrodes and electrode amplifiers. In HA Miller, Harrison DC (eds), Biomedical Electrode Technology, pp , New York, Academic Press. 9. Mastrototaro JJ, Massoud HZ, Pilkington TC, et al., Rigid and flexible thin-film microelectrode arrays for transmural cardiac recording. IEEE Trans Biomed Eng 39(3): Wise KD, Najafi K, Ji J, et al Micromachined silicon microprobes for CNS recording and stimulation. Proc Ann Conf IEEE Eng Med Biol Soc 12: Edell DJ A peripheral nerve information transducer for amputees: Long-term multichannel recordings from rabbit peripheral nerves. IEEE Trans Biomed Eng 33:203. Further Information Good overviews of biopotential electrodes are found in Geddes LA Electrodes and the Measurement of Bioelectric Events, New York, Wiley; and Ferris CD Introduction to Bioelectrodes, New York, Plenum. Even though these references are more than 20 years old, they clearly cover the field, and little has changed since these books were written. Overviews of biopotential electrodes are found in chapters of two works edited by John Webster. Chapter 5 of his textbook, Medical Instrumentation: Application and Design, covers the material of this chapter in more detail, and there is a section on Bioelectrodes in his Encyclopedia on Medical Devices and Instrumentation, published by Wiley in The journals IEEE Transactions on Biomedical Engineering and Medical and Biological Engineering and Computing are good sources of recent research on biopotential electrodes.
Chapter 13: Electrochemistry. Electrochemistry. The study of the interchange of chemical and electrical energy.
Chapter 13: Electrochemistry Redox Reactions Galvanic Cells Cell Potentials Cell Potentials and Equilbrium Batteries Electrolysis Electrolysis and Stoichiometry Corrosion Prevention Electrochemistry The
Figure 1. A voltaic cell Cu,Cu 2+ Ag +, Ag. gas is, by convention, assigned a reduction potential of 0.00 V.
Voltaic Cells Introduction In this lab you will first prepare a set of simple standard half-cells and then measure the voltage between the half-cells with a voltmeter. From this data you will be able to
Galvanic cell and Nernst equation
Galvanic cell and Nernst equation Galvanic cell Some times called Voltaic cell Spontaneous reaction redox reaction is used to provide a voltage and an electron flow through some electrical circuit When
Gel free Bio-potential Electrodes for Long-term Wearable Health Monitoring Applications
Volume 3, Number 2-2015 PP: 104-110 IJSE Available at www.ijse.org International Journal of Science and Engineering ISSN: 2347-2200 Gel free Bio-potential Electrodes for Long-term Wearable Health Monitoring
Discovering Electrochemical Cells
Discovering Electrochemical Cells Part I Electrolytic Cells Many important industrial processes PGCC CHM 102 Cell Construction e e power conductive medium What chemical species would be present in a vessel
Edmund Li. Where is defined as the mutual inductance between and and has the SI units of Henries (H).
INDUCTANCE MUTUAL INDUCTANCE If we consider two neighbouring closed loops and with bounding surfaces respectively then a current through will create a magnetic field which will link with as the flux passes
Chem 1721 Brief Notes: Chapter 19
Chem 1721 Brief Notes: Chapter 19 Chapter 19: Electrochemistry Consider the same redox reaction set up 2 different ways: Cu metal in a solution of AgNO 3 Cu Cu salt bridge electrically conducting wire
Indoor Environment Electromagnetic Radiation
Indoor Environment Electromagnetic Radiation Purpose: To increase the students awareness of the sources of electric and magnetic fields and the potential for adverse health effects from prolonged exposure
Electrochemistry Revised 04/29/15
INTRODUCTION TO ELECTROCHEMISTRY: CURRENT, VOLTAGE, BATTERIES, & THE NERNST EQUATION Experiment partially adapted from J. Chem. Educ., 2008, 85 (8), p 1116 Introduction Electrochemical cell In this experiment,
ELECTROCHEMICAL CELLS
1 ELECTROCHEMICAL CELLS Allessandra Volta (1745-1827) invented the electric cell in 1800 A single cell is also called a voltaic cell, galvanic cell or electrochemical cell. Volta joined several cells together
5s Solubility & Conductivity
5s Solubility & Conductivity OBJECTIVES To explore the relationship between the structures of common household substances and the kinds of solvents in which they dissolve. To demonstrate the ionic nature
4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:
4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing
Electrochemistry - ANSWERS
Electrochemistry - ANSWERS 1. Using a table of standard electrode potentials, predict if the following reactions will occur spontaneously as written. a) Al 3+ + Ni Ni 2+ + Al Al 3+ + 3e - Al E = -1.68
Theory of Heating by Induction
CHAPTER 2 Theory of Heating by Induction INDUCTION HEATING was first noted when it was found that heat was produced in transformer and motor windings, as mentioned in the Chapter Heat Treating of Metal
Laboratory Guide. Anatomy and Physiology
Laboratory Guide Anatomy and Physiology TBME04, Fall 2010 Name: Passed: Last updated 2010-08-13 Department of Biomedical Engineering Linköpings Universitet Introduction This laboratory session is intended
Subminiature Load Cell Model 8417
w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...
Rapid Prototyping and Development of Microfluidic and BioMEMS Devices
Rapid Prototyping and Development of Microfluidic and BioMEMS Devices J. Sasserath and D. Fries Intelligent Micro Patterning System Solutions, LLC St. Petersburg, Florida (T) 727-522-0334 (F) 727-522-3896
Electrophysiological Recording Techniques
Electrophysiological Recording Techniques Wen-Jun Gao, PH.D. Drexel University College of Medicine Goal of Physiological Recording To detect the communication signals between neurons in real time (μs to
Miniaturizing Flexible Circuits for use in Medical Electronics. Nate Kreutter 3M
Miniaturizing Flexible Circuits for use in Medical Electronics Nate Kreutter 3M Drivers for Medical Miniaturization Market Drivers for Increased use of Medical Electronics Aging Population Early Detection
Experiment 9 Electrochemistry I Galvanic Cell
9-1 Experiment 9 Electrochemistry I Galvanic Cell Introduction: Chemical reactions involving the transfer of electrons from one reactant to another are called oxidation-reduction reactions or redox reactions.
Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe.
Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe. Brief Overview of STM Inventors of STM The Nobel Prize in Physics 1986 Nobel
Directory chapter 04
M Directory chapter 04 D-Sub Mixed subminiature D connectors New Page D-Sub mixed connector system general information................... 04.02 Contact arrangements............................................
Solid shape molding is not desired in injection molding due to following reasons.
PLASTICS PART DESIGN and MOULDABILITY Injection molding is popular manufacturing method because of its high-speed production capability. Performance of plastics part is limited by its properties which
Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator
Development of New Inkjet Head Applying MEMS Technology and Thin Film Actuator Kenji MAWATARI, Koich SAMESHIMA, Mitsuyoshi MIYAI, Shinya MATSUDA Abstract We developed a new inkjet head by applying MEMS
Resistor Theory and Technology
Resistor Theory and Technology Felix Zandman Chairman, Scientific Director, and CEO, Vishay Intertechnology, Inc. Paul-Rene Simon Consultant Joseph Szwarc Chief Engineer, Vishay Israel Ltd SciTECH PUBLISHING;
An equivalent circuit of a loop antenna.
3.2.1. Circuit Modeling: Loop Impedance A loop antenna can be represented by a lumped circuit when its dimension is small with respect to a wavelength. In this representation, the circuit parameters (generally
Chemistry 122 Mines, Spring 2014
Chemistry 122 Mines, Spring 2014 Answer Key, Problem Set 9 1. 18.44(c) (Also indicate the sign on each electrode, and show the flow of ions in the salt bridge.); 2. 18.46 (do this for all cells in 18.44
Type of Chemical Bonds
Type of Chemical Bonds Covalent bond Polar Covalent bond Ionic bond Hydrogen bond Metallic bond Van der Waals bonds. Covalent Bonds Covalent bond: bond in which one or more pairs of electrons are shared
Bending, Forming and Flexing Printed Circuits
Bending, Forming and Flexing Printed Circuits John Coonrod Rogers Corporation Introduction: In the printed circuit board industry there are generally two main types of circuit boards; there are rigid printed
An Introduction to Electrochemical Impedance Measurement
An Introduction to Electrochemical Impedance Measurement Technical Report No. 6 An Introduction to Electrochemical Impedance Measurement N D Cogger and N J Evans Technical Report No. 6 Part No.: BTR006
A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer
Presented at IMAPS 2002 Denver, Colorado September 5, 2002 (Best of Session Award) A Novel Flex Circuit Area-Array Interconnect System for a Catheter-Based Ultrasound Transducer Jeff Strole*, Scott Corbett*,
The most recent advances in glass encapsulation of microelectronic devices offer important
Improvements in Glass Encapsulation Technology Offer Significant Advantages for Implantable Medical Devices By Frédéric Mauron Senior VP, Biomedical Engineering Director, Business & Administration Valtronic
Chapter 1. Introduction of Electrochemical Concepts
Chapter 1. Introduction of Electrochemical Concepts Electrochemistry concerned with the interrelation of electrical and chemical effects. Reactions involving the reactant the electron. Chemical changes
Good Boards = Results
Section 2: Printed Circuit Board Fabrication & Solderability Good Boards = Results Board fabrication is one aspect of the electronics production industry that SMT assembly engineers often know little about.
5.111 Principles of Chemical Science
MIT OpenCourseWare http://ocw.mit.edu 5.111 Principles of Chemical Science Fall 2008 For information about citing these materials or our Terms of Use, visit: http://ocw.mit.edu/terms. 26.1 5.111 Lecture
Fundamentals of Mass Flow Control
Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
Objectives. Capacitors 262 CHAPTER 5 ENERGY
Objectives Describe a capacitor. Explain how a capacitor stores energy. Define capacitance. Calculate the electrical energy stored in a capacitor. Describe an inductor. Explain how an inductor stores energy.
Direction of current flow through conductor. Fig. 1 Magnetic field generated by current flow
ingle Conductor Cable Copper heathed Cable heath Currents ingle conductor cables present certain application considerations that do not arise in multiconductor cable installations. These considerations
Sample preparation for X-ray fluorescence analysis
Technical articles Sample preparation for X-ray fluorescence analysis III. Pressed and loose powder methods Gakuto Takahashi* 1. Introduction There are two main sample preparation techniques for measurement
The DC Motor/Generator Commutation Mystery. Commutation and Brushes. DC Machine Basics
The DC Motor/Generator Commutation Mystery One small, yet vital piece of the DC electric motor puzzle is the carbon brush. Using the correct carbon brush is a key component for outstanding motor life,
Quantum Computing for Beginners: Building Qubits
Quantum Computing for Beginners: Building Qubits Suzanne Gildert Condensed Matter Physics Research (Quantum Devices Group) University of Birmingham 28/03/2007 Overview of this presentation What is a Qubit?
Electrochemical Impedance Spectroscopy (EIS): A Powerful and Cost- Effective Tool for Fuel Cell Diagnostics
Electrochemical Impedance Spectroscopy (EIS): A Powerful and Cost- Effective Tool for Fuel Cell Diagnostics Electrochemical Impedance Spectroscopy (EIS) is a powerful diagnostic tool that you can use to
RECITATION NOTES FOR EXPERIMENT # 5 A&B THIN LAYER CHROMATOGRAPHY
RECITATION NOTES FOR EXPERIMENT # 5 A&B THIN LAYER CHROMATOGRAPHY Have your lab textbook available for quick reference to specific pages, indicated in red. BASIC PRINCIPLES OF CHROMATOGRAPHY Chromatography
Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law.
Conductivity of strong and weak electrolytes TEC Related concepts Kohlrausch s law, equivalent conductivity, temperature dependence of conductivity, Ostwald s dilution law. Principle It is possible to
Micro Power Generators. Sung Park Kelvin Yuk ECS 203
Micro Power Generators Sung Park Kelvin Yuk ECS 203 Overview Why Micro Power Generators are becoming important Types of Micro Power Generators Power Generators Reviewed Ambient Vibrational energy Radiant
ENS 07 Paris, France, 3-4 December 2007
ENS 7 Paris, France, 3-4 December 7 FRICTION DRIVE SIMULATION OF A SURFACE ACOUSTIC WAVE MOTOR BY NANO VIBRATION Minoru Kuribayashi Kurosawa, Takashi Shigematsu Tokyou Institute of Technology, Yokohama
WIRE, TERMINAL AND CONNECTOR REPAIR CONDUCTORS
CONDUCTORS Conductors are needed to complete the path for electrical current to flow from the power source to the working devices and back to the power source. Special wiring is needed for battery cables
Biomedical applications of polypyrrole microactuators: from single-cell clinic to microrobots
Biomedical applications of polypyrrole microactuators: from single-cell clinic to microrobots Edwin Jager, Charlotte Immerstrand, Karl-Eric Magnusson, Olle Inganäs and Ingemar Lundström Linköping University
Supercapacitors. Advantages Power density Recycle ability Environmentally friendly Safe Light weight
Supercapacitors Supercapacitors also called ultracapacitors and electric double layer capacitors (EDLC) are capacitors with capacitance values greater than any other capacitor type available today. Capacitance
Why silicon MEMS? MEMS@KTH. Silicon is a strong material... Photolithography. Micromachining. Dicing and packaging
Why silicon MEMS? MEMS@KTH Small Identical Large volumes (low cost per unit) School of Electrical Engineering Royal Institute of Technology Silicon is a strong material... Photolithography 10 µm thick
Chapter 7: Polarization
Chapter 7: Polarization Joaquín Bernal Méndez Group 4 1 Index Introduction Polarization Vector The Electric Displacement Vector Constitutive Laws: Linear Dielectrics Energy in Dielectric Systems Forces
Fabrication of Complex Circuit Using Electrochemical Micromachining on Printed Circuit Board (PCB)
5 th International & 26 th All India Manufacturing Technology, Design and Research Conference (AIMTDR 2014) December 12 th 14 th, 2014, IIT Guwahati, Assam, India Fabrication of Complex Circuit Using Electrochemical
Basic Designs Of Flex-Rigid Printed Circuit Boards
PCBFABRICATION Basic Designs Of Flex-Rigid Printed Circuit Boards Flex-rigid boards allow integrated interconnection between several rigid boards. This technology helps to reduce the number of soldered
How to Build a Printed Circuit Board. Advanced Circuits Inc 2004
How to Build a Printed Circuit Board 1 This presentation is a work in progress. As methods and processes change it will be updated accordingly. It is intended only as an introduction to the production
Microvitae présente 4 communications à l'e-mrs, 14-17 mai 2012, Strasbourg
Microvitae présente 4 communications à l'e-mrs, 14-17 mai 2012, Strasbourg Lundi 14 mai 2012 09:30 In vivo cortical and hippocampal activity recordings by using high-resolution plastic electrode arrays
Balancing the Electrical and Mechanical Requirements of Flexible Circuits. Mark Finstad, Applications Engineering Manager, Minco
Balancing the Electrical and Mechanical Requirements of Flexible Circuits Mark Finstad, Applications Engineering Manager, Minco Table of Contents Abstract...............................................................................................
ONE OF THE SMALLEST SNAP-ACTION SWITCHES IN THE WORLD. FEATURES Superminiature type, light-weight snap action switch PC board terminal type (0.
ONE OF THE SMALLEST SNAP-ACTION SWITCHES IN THE WORLD SWITCHES FEATURES Superminiature type, light-weight snap action switch PC board terminal type (0.2g) Solder terminal type with mounting holes (0.3g)
Use of Carbon Nanoparticles for the Flexible Circuits Industry
Use of Carbon Nanoparticles for the Flexible Circuits Industry Ying (Judy) Ding, Rich Retallick MacDermid, Inc. Waterbury, Connecticut Abstract FPC (Flexible Printed Circuit) has been growing tremendously
Flex Circuit Design and Manufacture.
Flex Circuit Design and Manufacture. Hawarden Industrial Park, Manor Lane, Deeside, Flintshire, CH5 3QZ Tel 01244 520510 Fax 01244 520721 [email protected] www.merlincircuit.co.uk Flex Circuit
How compact discs are made
How compact discs are made Explained by a layman for the laymen By Kevin McCormick For Science project at the Mountain View Los Altos High School Abstract As the major media for music distribution for
Electrochemistry Voltaic Cells
Electrochemistry Voltaic Cells Many chemical reactions can be classified as oxidation-reduction or redox reactions. In these reactions one species loses electrons or is oxidized while another species gains
Electric Field Mapping Lab 3. Precautions
HB 09-25-07 Electric Field Mapping Lab 3 1 Electric Field Mapping Lab 3 Equipment mapping board, U-probe, resistive boards, templates, dc voltmeter (431B), 4 long leads, 16 V dc for wall strip Reading
Prentice Hall. Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition. High School. High School
Prentice Hall Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition High School C O R R E L A T E D T O High School C-1.1 Apply established rules for significant digits,
Flexible Mehrlagen-Schaltungen in Dünnschichttechnik:
Flexible Mehrlagen-Schaltungen in Dünnschichttechnik: Technologie-Plattform für Intelligente Implantate A. Kaiser, S. Löffler, K. Rueß, P. Matej, C. Herbort, B. Holl, G. Bauböck Cicor Advanced Microelectronics
Galvanic Cells and the Nernst Equation
Exercise 7 Page 1 Illinois Central College CHEMISTRY 132 Laboratory Section: Galvanic Cells and the Nernst Equation Name: Equipment Voltage probe wires 0.1 M solutions of Pb(NO 3, Fe(NO 3 ) 3, and KNO
Broadband Slotted Coaxial Broadcast Antenna Technology
Broadband Slotted Coaxial Broadcast Antenna Technology Summary Slotted coaxial antennas have many advantages over traditional broadband panel antennas including much smaller size and wind load, higher
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Development of Ultra-Multilayer. printed circuit board
Development of Ultra-Multilayer Printed Circuit Board Yasuyuki Shinbo Using the multilayer and characteristic impedance control technologies cultivated in the field of telecommunications, OKI Printed Circuits
Implant materials. Learning outcomes. Implant materials in trauma. How to use this handout? Functions of implants. Types of materials
Implant materials How to use this handout? The left column is the information as given during the lecture. The column at the right gives you space to make personal notes. Learning outcomes At the end of
Chapter 6. Current and Resistance
6 6 6-0 Chapter 6 Current and Resistance 6.1 Electric Current... 6-2 6.1.1 Current Density... 6-2 6.2 Ohm s Law... 6-5 6.3 Summary... 6-8 6.4 Solved Problems... 6-9 6.4.1 Resistivity of a Cable... 6-9
Flexible Printed Circuits Design Guide
www.tech-etch.com/flex Flexible Printed Circuits Design Guide Multilayer SMT Assembly Selective Plating of Gold & Tin-Lead Fine Line Microvias Cantilevered & Windowed Leads 1 MATERIALS CONDUCTOR Copper
White Paper. Recommendations for Installing Flash LEDs on Flex Circuits. By Shereen Lim. Abstract. What is a Flex Circuit?
Recommendations for Installing Flash LEDs on Circuits By Shereen Lim White Paper Abstract For the mobile market some PCB assemblies have been converted to flex circuit assemblies, in part because flex
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
Silicon-On-Glass MEMS. Design. Handbook
Silicon-On-Glass MEMS Design Handbook A Process Module for a Multi-User Service Program A Michigan Nanofabrication Facility process at the University of Michigan March 2007 TABLE OF CONTENTS Chapter 1...
CELL POTENTIAL, E. Terms Used for Galvanic Cells. Uses of E o Values CELL POTENTIAL, E. Galvanic Cell. Organize halfreactions
Electrons move from anode to cathode in the wire. Anions & cations move thru the salt bridge. Terms Used for Galvanic Cells Galvanic Cell We can calculate the potential of a Galvanic cell using one of
A METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS
Uludağ Üniversitesi Mühendislik-Mimarlık Fakültesi Dergisi, Cilt 9, Sayı, 24 A METHOD OF PRECISE CALIBRATION FOR PIEZOELECTRICAL ACTUATORS Timur CANEL * Yüksel BEKTÖRE ** Abstract: Piezoelectrical actuators
Name Electrochemical Cells Practice Exam Date:
Name Electrochemical Cells Practice Exam Date: 1. Which energy change occurs in an operating voltaic cell? 1) chemical to electrical 2) electrical to chemical 3) chemical to nuclear 4) nuclear to chemical
Development of High-Speed High-Precision Cooling Plate
Hironori Akiba Satoshi Fukuhara Ken-ichi Bandou Hidetoshi Fukuda As the thinning of semiconductor device progresses more remarkably than before, uniformity within silicon wafer comes to be strongly required
Question Bank Electrolysis
Question Bank Electrolysis 1. (a) What do you understand by the terms (i) electrolytes (ii) non-electrolytes? (b) Arrange electrolytes and non-electrolytes from the following substances (i) sugar solution
still different technique of assessment.
D E N S I T Y A N A L Y S I S Micromeritics gas pycnometers are being used worldwide to obtain material density measurements. These measurements are necessary in the manufacture of many of today s products.
NetShape - MIM. Metal Injection Molding Design Guide. NetShape Technologies - MIM Phone: 440-248-5456 31005 Solon Road FAX: 440-248-5807
Metal Injection Molding Design Guide NetShape Technologies - MIM Phone: 440-248-5456 31005 Solon Road FAX: 440-248-5807 Solon, OH 44139 [email protected] 1 Frequently Asked Questions Page What
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS Joseph J. Stupak Jr, Oersted Technology Tualatin, Oregon (reprinted from IMCSD 24th Annual Proceedings 1995) ABSTRACT The
Metal Migration on Electric Circuit Boards
Three Bond Technical News Issued December 1, 1986 16 Metal Migration on Electric Circuit Boards 1. Introduction Today, many mechanical parts such as those used on electronic circuit boards are made to
Part B 2. Allow a total of 15 credits for this part. The student must answer all questions in this part.
Part B 2 Allow a total of 15 credits for this part. The student must answer all questions in this part. 51 [1] Allow 1 credit for 3 Mg(s) N 2 (g) Mg 3 N 2 (s). Allow credit even if the coefficient 1 is
Wearable Sensors for medical applications
Wearable Sensors for medical applications A. M. Ionescu, A. Bazigos Ecole Polytechnique Fédérale Lausanne A.M. Ionescu, A. Bazigos - CATRENE Wkshp, Feb. 4, 2014 1 Outline Acknowledgement: working group
Introduction to Electricity & Magnetism. Dr Lisa Jardine-Wright Cavendish Laboratory
Introduction to Electricity & Magnetism Dr Lisa Jardine-Wright Cavendish Laboratory Examples of uses of electricity Christmas lights Cars Electronic devices Human body Electricity? Electricity is the presence
PIEZOELECTRIC FILMS TECHNICAL INFORMATION
PIEZOELECTRIC FILMS TECHNICAL INFORMATION 1 Table of Contents 1. PIEZOELECTRIC AND PYROELECTRIC EFFECTS 3 2. PIEZOELECTRIC FILMS 3 3. CHARACTERISTICS PROPERTIES OF PIEZOELECTRIC FILMS 3 4. PROPERTIES OF
Are you aware. Outline
Electrical Safety of Medical Equipment Hasan Al-Nashash School of Engineering American University of Sharjah (AUS) 1 Are you aware Electrocutions are the 5th leading cause of accidental death in the U.S.
Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope
International Journal of Arts and Sciences 3(1): 18-26 (2009) CD-ROM. ISSN: 1944-6934 InternationalJournal.org Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe Bedri Onur Kucukyildirim,
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion
EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1 Introduction This lab is designed for you to discover the properties of electrochemical cells. It requires little previous knowledge of electrochemical
www.axon-cable.com Power distribution systems
www.axon-cable.com Power distribution systems Power distribution systems As vital components for electrical power distribution in telecommunication satellites and landbased weapon systems, the bus and
EXPERIMENT #9 CORROSION OF METALS
EXPERIMENT #9 CORROSION OF METALS Objective The objective of this experiment is to measure the corrosion rate of two different metals and to show the effectiveness of the use of inhibitors to protect metals
Trident Connector Family Offers the Versatility and Reliability Required in Industrial & Transportation Applications
Trident Connector Family Offers the Versatility and Reliability Required in Industrial & Transportation Applications Industrial and transportation applications require robust connections with superior
How to measure absolute pressure using piezoresistive sensing elements
In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article
Solar Photovoltaic (PV) Cells
Solar Photovoltaic (PV) Cells A supplement topic to: Mi ti l S Micro-optical Sensors - A MEMS for electric power generation Science of Silicon PV Cells Scientific base for solar PV electric power generation
BIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
Chapter 10 Liquids & Solids
1 Chapter 10 Liquids & Solids * 10.1 Polar Covalent Bonds & Dipole Moments - van der Waals constant for water (a = 5.28 L 2 atm/mol 2 ) vs O 2 (a = 1.36 L 2 atm/mol 2 ) -- water is polar (draw diagram)
Applications and Benefits of Multi-Walled Carbon Nanotubes (MWCNT)
I Applications and Benefits of Multi-Walled Carbon Nanotubes (MWCNT) Table of Content 1 Introduction...1 2 Improved Properties...1 3 Potential Applications...1 3.1 Current / short-term applications...3
Encoders for Linear Motors in the Electronics Industry
Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy
