1 ICD leads under advisory What should be done? Shu Zhang, MD, PhD, FHRS, FESC Beijing Fu Wai Hospital Present by Chu-Pak Lau, MD Director, Cardiac Health Heart Centre Honorary Clinical Professor Queen Mary Hospital The University of Hong Kong
2 INTRODUCTION
3 Why do ICD leads fail and what s the consequences? Complex design with multiple cables Must tolerate high voltages Must tolerate high stress High frequency/low flex (heart/vasculature) Low frequency/high flex (shoulder) Manifestations of electrical and mechanical failure: inappropriate therapy failure to defibrillate loss of pacing -- mechanical injury to CVS structure or other leads, and/or thrombosis
4 Recent ICD leads under advisory Fidelis ICD leads (2009): FDA issued a Class I Recall of Medtronic Inc. Sprint Fidelis Defibrillator Leads, model numbers 6930, 6931, 6948 and Medtronic is advising physicians to stop implanting the leads and to return unused products to the firm. Riata ICD leads (2011): FDA is providing information and recommendations regarding safety concerns with the recalled Riata leads manufactured by St Jude Medical. These leads have an increased risk of premature insulation failure that can impact the lead s ability to function properly.
5 ICD LEAD FAILURE
6 Example of Sprint Fidelis Fracture Failure due to fracture of conductors: 90% of fractures are to pace-sense circuit, 10% to HV circuits Locations of fractures are highlighted in red above
7 Sprint Fidelis Performance Medtronic System Longevity Study*(SLS) and CareLink data 14.6% failure at 93 months, increasing 2% per year * System Longevity Study is a prospective and multi-center study to evaluate long-term performance of all Medtronic leads (29 years, over 75,000 leads evaluated proactively).
8 Example of Riata/Riata ST Inside-out Abrasion Failure due to inside out insulation defects leading to: externalized conductor cables (may be visible under fluoroscopy) short-circuiting between conductors internally Hauser, et al. Heart Rhythm May 2012; 9:
9 Riata/Riata ST Performance 2013 St Jude Medical Phase I Results
10 RISK FACTORS FOR LEAD FAILURE
11 ICD Lead Failure in 990 ICD patients in 13 years Insulation defect is the most common type of lead failure Kleemann T, et al. Circulation 2007; 115:
12 Age and gender impacting Lead Failure, e.g. Fidelis Fracture Hauser RG, et al. Circulation 2011; 123: Canadian Heart Rhythm Society Report Female gender, younger age and cardiac disease tend to have higher failure rate in Fidelis fracture (perhaps due to more active lifestyle and higher HR). Birnie DH, et al. Circulation 2012; 125:
13 Venous Access and lead failure Canadian Heart Rhythm Society Report Birnie DH, et al. Circulation 2012; 125:
14 Implant Procedure- Suture sleeve and Pocket location Subcutaneous Easy to access device/leads May predispose to erosion in thin patients Suture sleeve Submuscular More cosmetic looking Difficult to access device/leads Increase risk of adding stress or damaging connector, insulation and lead
15 Class I recall of ICD leads: A comparison of the Sprint Fidelis and Riata leads Sprint Fidelis: more likely to present with oversensing and inappropriate shocks Riata: showed a trend toward presenting with more high-voltage (shock) failure Liu J, et al. Heart Rhythm 2012; 9:
16 MANAGEMENT OF LEADS UNDER ADVISORY
17 Manufacturer s Sprint Fidelis Patient Management Recommendations The Lead Integrity Alert (LIA) provides three days advance notice prior to inappropriate therapy to 76% of patients with lead fractures. As a result, we strongly recommend that all Sprint Fidelis patients who have the ability to upgrade to Lead Integrity Alert do so promptly. Also ensure that high voltage lead impedance alerts (maximum of 100 ohms) are programmed. When a lead fracture is suspected or confirmed, immediate patient attention is strongly recommended. Physicians should inform their patients to seek medical attention without delay if they experience unexpected shocks. If a Fidelis lead fracture of any type has occurred, we recommend implanting a new high voltage lead with or without extraction of the Fidelis lead. In patients with normal device function and no manifestation of lead fracture, no action is recommended. The risk of prophylactic intervention appears to be greater than serious injury resulting from lead fracture even for pacemaker dependent patients, except in select individual patient circumstances as determined by the physician. In the event of a device change-out or upgrade procedure, with no manifestation of lead fracture, consider the patient age and lead model data above, as well as patient life expectancy, co-morbidities, ease of extraction related to implant time, patient preference, etc., for the following options: Leave a properly performing lead intact. Implant a new ICD lead without extraction of the existing lead. Carefully consider all factors before prophylactic placement of a pace-sense lead given the data in Table 1, which shows an increased risk of high voltage conductor fracture if a pace-sense conductor fracture has previously occurred. Individual patient circumstances may warrant extracting and implanting a new ICD lead. If warranted, Medtronic s Independent Physician Quality Panel recommends the lead extraction procedure
18
19 Riata TM Lead replacement and extraction recommendations
20 Recommendations from HRS/ESC Heart Rhythm Society Task Force on Lead Performance * (2009) Managing Lead on Advisory Notices Periodic device monitoring (remote or in-person as appropriate) Lead revision or replacement considered based on risk-benefit analysis Reprogramming to mitigate risk from a lead malfunction Factors to considering replacement/revision of normally functioning leads subject to advisory Both patient and lead, for example: pacemaker dependence, surgical risk, rate of abnormal performance, malfunction characteristics and mechanism ESC/EHRA Pacing Guidelines 2013 Remote monitoring should be considered to provide early detection * developed in collaboration with the American College of Cardiology (ACC) and the American Heart Association (AHA)
21 MANAGEMENT OF FAILED LEADS
22 Should we extract or abandon leads (failed or on advisory)? Canadian Experience with Fidelis Revisions and Complications Significantly more complications when lead removed (complications required reoperation and minor complications). Death and cardiac arrest were similar. No randomized trials of extraction against abandonment of failed or advisory leads Every patient is different (clinical co-morbidities, anatomic variations, lead types and numbers) Parkash R, et al. Circulation 2010; 121:
23 Potential issues with abandoned leads Lower success of future lead removal Risk of failed lead removal doubles every 3 years 1 More Infection Risk of erosion from more pocket bulk 2 Risk of infection increases 2-7% at each device change 3 More venous lead burden Risk of venous thrombosis or occlusion (prevent access to new leads; right-sided implant Risk of clotting or SVC syndrome More tricuspid lead burden Risk of increased tricuspid regurgitation 4 Lead-to-lead interaction Risk of inappropriate shocks or inappropriate pace/sense behavior MRI Contraindication Risk of tissue damage or cardiac stimulation due to inadvertent heating of the capped leads or induced lead current 1. Byrd, PACE Bohm, PACE Wilkoff, HRS De Cook, PACE 2000
24 When should leads be extracted (failed or on advisory)? Transvenous Lead Extraction: Heart Rhythm Expert Consensus (2009) *Expanded from the original NASPE Indication in 2000 Category Indication Class FUNCTIONAL LEAD Due to design or failure, may pose immediate threat I NON FUNCTIONAL LEAD Risk of interference with device operation Due to design or failure poses potential future threat Functional leads not being used (e.g. ICD upgrade) Need MRI with no other imaging options for diagnosis To permit the implantation of an MRI conditional CIED system Implant would require >4 leads on one side or >5 leads through SVC Need MRI with no other imaging options for diagnosis To permit the implantation of an MRI conditional CIED system Non functional lead at time of indicated CIED procedure Lead extraction apply only in patients where the benefits of lead removal outweigh the risks when assessed based on individualized patient factors and operator specific experience and outcomes. Wilkoff, et al. Heart Rhythm 2009; 6(7): IIb IIb IIb IIb IIb IIa IIa IIb IIb
25 LEAD PERFORMANCE MONITORING
26 Failed lead analysis and post market study Returned Product and Complaints Provide insights to failure mechanism, but Significantly underestimates failure rate (failed leads are not often returned/reported) Post-market Lead Study Since 2008, FDA required manufacturers to conduct 5 year studies on new FDA-approved ICD leads (at least 1,000 patients for 5 years after implantation) Prior to 2008, most manufacturers tracked lead performance via returned product and complaints Analysis of manufacturers ICD lead database (such as over 72,000 Fidelis leads for analysis)
27 Lead Integrity Alert (LIA) LIA is an alert algorithm in Medtronic devices that addresses lead failure: Rapid oversensing trigger, Automatic NID extension and Impedance Blanck Z, et al. JCE 2011;22:
28 Lead Integrity Alert (LIA) Ellenbogen et al - LIA found 67% more ICD lead failures across manufacturers (Medtronic, St Jude Medical and Boston Scientific) compared to impedance monitoring 1 LIA now FDA approved for monitoring of non-medtronic leads Detection for lead issues Biotronik Boston Scientific St Jude Medical Medtronic Lead impedance Yes Yes Yes Yes Oversensed signals (SIC*) X X X Yes Auto NID extension post-alert X X X Yes Alert Device-based (audible or vibration) X Yes Yes Yes Wireless Yes** Yes Yes Yes Diagnostics EGM storage for NSVTs Yes Yes X Yes Oversensed signals counter (SIC*) X X X Yes Auto Pre-storage EGM activation post-alert X X X Yes Published data Performance on alert X X X 75% of patients receive at least a 3 day notification; false positive rate 1 in 1285 patient years * SIC Sensing Integrity Counter ** Some Biotronik models only 1 Ellenbogen KA, et al. Circ Arrhythm Electrophysiol 2013;6:
29 Remote monitoring of lead performance - Home Monitoring (BIOTRONIK, Berlin, Germany) - CareLink Network (Medtronic Inc., MN, USA) - Latitude Patient Management system (Boston Scientific, St. Paul, USA) - Merlin.net (St. Jude Medical, Sylmar, USA) - Smartview (Sorin CRM SAS, Clamart, France)
30 ICDs are life saving ICD leads are the weak link Summary Consider all factors before replacing or extracting an ICD lead under advisory Returned product/complain analysis underestimates failure rate and delays detection of poor performance, hence post-market lead study is now required by FDA for all new ICD leads Implant techniques may also help with ICD lead survival LIA and remote monitoring are recommended for lead monitoring sicd can be an alternative to lead failure
31 Thank you
Managing Patients With RIATA Leads
Managing Patients With RIATA Leads Bruce L. Wilkoff M.D. Cleveland Clinic Medical Advisory Boards: Medtronic, St. Jude Medical, Spectranetics Professor of Medicine Cleveland Clinic Lerner College of Medicine
More informationDEVICE RECALLS: The Era of Regulation and Outcome Metrics: Optimizing Benefits and Managing Risks
DEVICE RECALLS: The Era of Regulation and Outcome Metrics: Optimizing Benefits and Managing Risks Kenneth A. Ellenbogen, MD Kontos Professor & Chairman Virginia Commonwealth University School of Medicine
More informationMinimizing and Managing Lead Complications. Eric Buch MD Assistant Professor of Medicine Director, Specialized Program for Atrial Fibrillation
Minimizing and Managing Lead Complications Eric Buch MD Assistant Professor of Medicine Director, Specialized Program for Atrial Fibrillation Topics History of lead management Surveillance of implanted
More informationSW012. Installation software for the Right Ventricular Lead Integrity Alert feature. Reference Manual
Installation software for the Right Ventricular Lead Integrity Alert feature Reference Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. The following list
More informationSt. Jude Medical Credit Suisse s 2009 Healthcare Conference. Eric S. Fain, M.D. President, Cardiac Rhythm Management Division November 12, 2009
St. Jude Medical Credit Suisse s 2009 Healthcare Conference Eric S. Fain, M.D. President, Cardiac Rhythm Management Division November 12, 2009 Forward-Looking Statements The following presentation contains
More informationEstablishing a Remote Monitoring Program. Martha Ferrara, FNP
Establishing a Remote Monitoring Program Martha Ferrara, FNP Establishing a Remote Monitoring Program What is Remote Monitoring? Martha Ferrara, FNP, CCDS November 2012 CIED Timeline: Cardiovascular Implantable
More informationChristopher M. Wright, MD, MBA Pioneer Cardiovascular Consultants Tempe, Arizona
Christopher M. Wright, MD, MBA Pioneer Cardiovascular Consultants Tempe, Arizona Areas to be covered Historical, current, and future treatments for various cardiovascular disease: Atherosclerosis (Coronary
More informationDisclosures. Anesthesia and Lead Extractions. Lead Extractions: Objectives. Lead Removal Techniques. None
Anesthesia and Lead Extractions Disclosures None Bryan Ahlgren DO Assistant Professor University of Colorado Dept of Anesthesiology Objectives Define lead extraction procedures and why Anesthesiologists
More informationCardiac Rhythm Device Summit:
Division of Cardiovascular Diseases Cardiac Rhythm Device Summit: Implantation, Management, and Follow Up June 17-19, 2016 Swissotel Chicago Course Directors: LEARN. ADVANCE. CONNECT. celinks.mayo.edu/cardiacdevice2016
More informationCareLink Network. Clinic Setup Guide WORKFLOWS. FAQs
INTRODUCTION CareLink Network Clinic Setup Guide All patient and clinical data displayed on the screenshots are fictitious and for demonstration purposes only. Introduction INTRODUCTION Transforming cardiac
More informationHow To Improve Health Care For Remote Workers
CRM Devices and Telemonitoring Where the industry stands today Annette Brüls VP CRDM Marketing CareLink Status worldwide More than 450.000000 patients in > 4000 clinics > 30 countries 9 years of experience
More informationSome of the slides and the video were supplied by a device manufacturer
Financial Disclosure The Ins and Outs of Implantable Lead Extraction: The Value of Implementing a Complication Prevention and Management Initiative Some of the slides and the video were supplied by a device
More informationSTANDARDS FOR IMPLANTATION AND FOLLOW-UP OF CARDIAC RHYTHM MANAGEMENT DEVICES IN ADULTS January 2013
1. INTRODUCTION STANDARDS FOR IMPLANTATION AND FOLLOW-UP OF CARDIAC RHYTHM MANAGEMENT DEVICES IN ADULTS January 2013 This document replaces the previous Heart Rhythm UK documents Clinical Guidance by Consensus
More informationExamining the Sprint Fidelis Effect on Medicare Costs. H. Dennis Tolley, PhD, ASA. April 2010
Examining the Sprint Fidelis Effect on Medicare Costs by H. Dennis Tolley, PhD, ASA April 2010 The Sprint Fidelis Effect on Medicare Costs H. Dennis Tolley, PhD, ASA SUMMARY This white paper presents an
More informationWorldwide Experience with S-ICD: Early results from the EFFORTLESS Registry
Worldwide Experience with S-ICD: Early results from the EFFORTLESS Registry Results published in European Heart Journal (March 2014) Pier D. Lambiase, Craig Barr, Dominic A.M.J. Theuns, Reinoud Knops,
More informationINSERTABLE CARDIAC MONITORING SYSTEM. UNLOCK the ANSWER. Your heart and long-term monitoring
INSERTABLE CARDIAC MONITORING SYSTEM UNLOCK the ANSWER Your heart and long-term monitoring UNLOCK the ANSWER Irregular heartbeats can be related to a variety of conditions, including unexplained fainting,
More informationManagement of malfunctioning and recalled pacemaker and defibrillator leads: results of the European Heart Rhythm Association survey
Europace (2014) 16, 1674 1678 doi:10.1093/europace/euu302 EP WIRE Management of malfunctioning and recalled pacemaker and defibrillator leads: results of the European Heart Rhythm Association survey Maria
More informationCRM Product Performance Report 2014 Q4 Edition
CRM Performance Report 2014 Q4 Edition 1 Advancing Science for Life. For almost forty years, meaningful innovation at Boston Scientific Rhythm Management has helped patients live healthier, longer lives.
More informationUpdate on Indications, Techniques and Controversies for Lead Extraction
Update on Indications, Techniques and Controversies for Lead Extraction W Boston 100 Stuart Street, Boston, MA Course Directors Avi Fischer, MD Laurence M. Epstein, MD Brigham and Women s Hospital Sponsored
More informationWhat Are Arrhythmias?
What Are Arrhythmias? Many people have questions about what the word arrhythmia means, and arrhythmias can be a difficult subject to understand. The text below should give you a better understanding of
More informationIntroduction. Planned surgical procedures
Guidelines for the perioperative management of patients with implantable pacemakers or implantable cardioverter defibrillators, where the use of surgical diathermy/electrocautery is anticipated. Introduction
More informationMRI Case Study MRI Cervical Spine
MRI Case Study MRI Cervical Spine Jewish Hospital Louisville, KY The Revo MRI SureScan pacing system is MR Conditional designed to allow patients to undergo MRI under the specified conditions for use.
More informationStarting an Extraction Program Getting your ducks in a row
Starting an Extraction Program Getting your ducks in a row Samir Saba, MD, FHRS Director, Cardiac Electrophysiology University of Pittsburgh Medical Center Research Grants Disclosures Medtronic Inc. Boston
More informationUnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program:
UnitedHealthcare, UnitedHealthcare of the River Valley and Neighborhood Health Partnership Cardiology Notification and Prior Authorization Program: Electrophysiology Implant Code Classification Table The
More informationEpimed Would Like To Congratulate The 20th Annual Budapest Conference The New Shape of Pain Relief SmalleST. ThiNNeST. CoNTouRed design. Precision Novi is the world s smallest, thinnest 16 contact primary
More informationREMOTE CARE ALERT MANAGEMENT REFERENCE GUIDE Managing Merlin.net
REMOTE CARE ALERT MANAGEMENT REFERENCE GUIDE Managing Merlin.net Patient Care Network (PCN) Alerts Merlin.net PCN offers a powerful range of alert capabilities to give your practice the control, flexibility
More informationADVISING THE PROFESSIONALS: THE ROLE OF MHRA
Safeguarding public health ADVISING THE PROFESSIONALS: THE ROLE OF DR SUSANNE LUDGATE Clinical Director Medicines and Healthcare products Regulatory Agency, UK Name Date WHAT IS? Medicines and Healthcare
More informationTABLE OF CONTENTS. 1 From Cleveland Clinic, Cleveland, Ohio, 2 Fondazione Cardiocentro Ticino, Lugano, Switzerland, 3 Thorax
HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): Description of Techniques, Indications, Personnel, Frequency and Ethical Considerations Developed in
More informationAtrial Fibrillation and Cardiac Device Therapy RAKESH LATCHAMSETTY, MD DIVISION OF ELECTROPHYSIOLOGY UNIVERSITY OF MICHIGAN HOSPITAL ANN ARBOR, MI
Atrial Fibrillation and Cardiac Device Therapy RAKESH LATCHAMSETTY, MD DIVISION OF ELECTROPHYSIOLOGY UNIVERSITY OF MICHIGAN HOSPITAL ANN ARBOR, MI Outline Atrial Fibrillation What is it? What are the associated
More informationPreamble. Europace (2008) 10, 707 725 doi:10.1093/europace/eun122
Europace (2008) 10, 707 725 doi:10.1093/europace/eun122 HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): Description of Techniques, Indications, Personnel,
More informationWhat Can I Do about Atrial Fibrillation (AF)?
Additional Device Information 9529 Reveal XT Insertable Cardiac Monitor The Reveal XT Insertable Cardiac Monitor is an implantable patientactivated and automatically activated monitoring system that records
More informationTreatment of Cardiac Device Infections
Treatment of Cardiac Device Infections Peter Ammann 2 2 40 yrs old patient Pocket infection after pacemaker change. Staph. aureus growing in blood cultures. Pacemaker dependent. Echo: no vegetations on
More informationInspection campaign summary report
Inspection Division Survey market inspection Unit TOPIC: Implantable defibrillation leads Inspection campaign summary report Abstract: An inspection campaign was conducted between November 2013 and June
More informationGuidelines for Perioperative Management of Pacemakers and Defibrillators
Guidelines for Perioperative Management of Pacemakers and Defibrillators Developed By: Deborah Wolbrette, MD, Medical Director, Electrophysiology Lab, Dept of Cardiology Kane High, MD, Department of Anesthesiology
More informationIMAGEREADY MR CONDITIONAL DEFIBRILLATION SYSTEM
MRI TECHNICAL GUIDE IMAGEREADY MR CONDITIONAL DEFIBRILLATION SYSTEM REF D000, D002, D010, D012, D020, D022, D044, D046, D050, D052, D140, D142, D150, D152, D174, D176, G058, G148, G158, G179, 0265, 0266,
More information2015 Subcutaneous Implantable Defibrillator (the S-ICD TM System) Coding Guide Rhythm Management
2015 Subcutaneous Implantable Defibrillator (the S-ICD TM System) Coding Guide Rhythm Management Contents Physician Coding... 2 Hospital Outpatient Coding and Payment... 3 Ambulatory Surgery Center (ASC)
More informationTelecardiology Technical Innovations and Challegenes in Clinical Practice
Telecardiology Technical Innovations and Challegenes in Clinical Practice Axel Müller Clinic of Internal Medicine I (Head of Department: Prof. Dr. med. J. Schweizer) Klinikum Chemnitz ggmbh June 24, 2015
More informationREVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual
REVEAL LINQ LNQ11 Insertable Cardiac Monitor MRI procedural information MRI Technical Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. The following list
More informationONC Project Drives Results with High- Value, Consumer- Friendly Data. [Font: Verdana Size 30]
ONC Project Drives Results with High- Value, Consumer- Friendly Data [Font: Verdana Size 30] Implantable Cardioverter-Defibrillators: High Value Data Why implantable cardioverter-defibrillators (ICDs)?
More informationLiving with Your Pacemaker
Brief Statement Additional Device Information An implantable pacemaker system relieves symptoms of heart rhythm disturbances. They do this by restoring normal heart rates. A normal heart rate provides
More information1547-5271/$ -see front matter 2009 Heart Rhythm Society. All rights reserved. doi:10.1016/j.hrthm.2009.05.020
Transvenous Lead Extraction: Heart Rhythm Society Expert Consensus on Facilities, Training, Indications, and Patient Management This document was endorsed by the American Heart Association (AHA). Bruce
More informationRome, Italy December 4-7, 2012 Rome Cavalieri TIMETABLE
Rome, Italy December 4-7, 2012 Rome Cavalieri Monday, December 3 ROOM Terrazza Monte Mario PATIENT MANAGEMENT CONCEPTS AND CONTROVERSIES IN ELECTROPHYSIOLOGY BIOTRONIK International Fellows Program (IFP)
More informationObjectives. Preoperative Cardiac Risk Stratification for Noncardiac Surgery. History
Preoperative Cardiac Risk Stratification for Noncardiac Surgery Kimberly Boddicker, MD FACC Essentia Health Heart and Vascular Center 27 th Heart and Vascular Conference May 13, 2011 Objectives Summarize
More informationBasics of Pacing. Ruth Hickling, RN-BSN Tasha Conley, RN-BSN
Basics of Pacing Ruth Hickling, RN-BSN Tasha Conley, RN-BSN The Cardiac Conduction System Cardiac Conduction System Review Normal Conduction Conduction QRS QRS Complex Complex RR PP ST ST segment segment
More informationCatheter Ablation. A Guided Approach for Treating Atrial Arrhythmias
Catheter Ablation A Guided Approach for Treating Atrial Arrhythmias A P A T I E N T H A N D B O O K This brochure will provide an overview of atrial arrhythmias (heart rhythm problems affecting the upper
More informationClinical trials for medical devices: FDA and the IDE process
Clinical trials for medical devices: FDA and the IDE process Owen Faris, Ph.D. Deputy Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health,
More informationOverview. Total Joint Replacement in the U.S. KP National Total Joint Registry EMR Tools and Outcome Assessment: A Model for Vascular Surgery?
KP National Total Joint Registry EMR Tools and Outcome Assessment: A Model for Vascular Surgery? Liz Paxton Director of Surgical Outcomes and Analysis Overview KP Total Joint Replacement Registry Background
More informationLATITUDE NXT 4.0 Quick Reference Guide
LATITUDE NXT 4.0 Quick Reference Guide LATITUDE NXT is compatible with Boston Scientific s most advanced pacemaker and defibrillator families 1, and now also with the EMBLEM sub-cutaneous defibrillator.
More informationCCAD Training Manual. Cardiac Rhythm Management (CRM)
CCAD Training Manual Cardiac Rhythm Management (CRM) Version 1.0 A D Cunningham 19/3/2008 Introduction This manual is intended to assist users of the Notes Client version of the CCAD Cardiac Rhythm Management
More informationHOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3
3 How to Use the LIFEPAK 1000 Defibrillator HOW TO USE THE LIFEPAK 1000 DEFIBRILLATOR 3 This section provides an overview of information and instructions for using the LIFEPAK 1000 defibrillator. Modes
More informationOEM MAXNIBP Frequently Asked Questions
Frequently Asked Questions Why does the monitor sometimes inflate the BP cuff, then shortly thereafter reinflate the cuff? How will I know if the monitor is experiencing motion artifact during a measurement?
More information50 years of CRM Device Therapy Past, Present and Future. Richard Sutton Professor of Cardiology Imperial College, London, UK
50 years of CRM Device Therapy Past, Present and Future Richard Sutton Professor of Cardiology Imperial College, London, UK The Past First implant 1958 Sweden Dual Chamber pacing (VAT) 1962 US Transvenous
More informationHeart Rhythm UK. Standards for Implantation and Follow-up of Cardiac Rhythm Management Devices HRUK, February 2011
Heart Rhythm UK Standards for Implantation and Follow-up of Cardiac Rhythm Management Devices HRUK, February 2011 1) Introduction 2) Definitions 3) Treatment indications 4) Requirements for performing
More informationMichael Hartman, CRNA, DNP, MSN, BSN, BA
Michael Hartman, CRNA, DNP, MSN, BSN, BA Intrinsic Conduction Automaticity-cardiac cells ability to spontaneously depolarize and initiate impulse Depolarization-cardiac cells go to a+ intracellular and
More information19.109 AUTOMATED EXTERNAL DEFIBRILLATOR
19.109 AUTOMATED EXTERNAL DEFIBRILLATOR References: American Heart Association Healthcare Provider (CPR/AED) Ohio Revised Code Chapter 2305.235 Immunity of Person Involved with Providing Automated External
More informationImplantable Cardioverter Defibrillator Treatment: Benefits and Pitfalls in the Currently Indicated Population. Carel Jan Willem Borleffs
Implantable Cardioverter Defibrillator Treatment: Benefits and Pitfalls in the Currently Indicated Population Carel Jan Willem Borleffs The studies described in this thesis were performed at the Department
More informationRemote Monitoring of Cardiac Implantable Electrical Devices (CIEDs)
Remote Monitoring of Cardiac Implantable Electrical Devices (CIEDs) Changing the face of enhanced self-management and improved coordinated healthcare K Fan, CKC Tsui, KL Au, RTC Ng, CYS Chung, KW Lai,
More informationManagement of Pacing Wires After Cardiac Surgery
Management of Pacing Wires After Cardiac Surgery David E. Lizotte, Jr. PA C, MPAS, FAPACVS President, Association of Physician Assistants in Cardiovascular Surgery Conflicts: None Indications 2008 Journal
More informationLes Implantables actifs: présent et futur. Pierre-Olivier Sémon
Les Implantables actifs: présent et futur Pierre-Olivier Sémon Présent Solutions for Restoring Health Product Families Pacemakers Bradycardia Indications Pacemakers Arrêt sinusal Bloc AV Fonctionnement
More informationSporadic or short episodes of paroxysmal atrial fibrillation - still a need for antithrombotic therapy?
Sporadic or short episodes of paroxysmal atrial fibrillation - still a need for antithrombotic therapy? Carina Blomström Lundqvist Dept Cardiology, Uppsala University, Sweden Patterns of AF Terminates
More informationPacemaker Counter and Histogram Operation and Interpretation. Percentage of total count. Figure 2. Detailed ALTRUA Paced and Sensed Event Counters
A Closer Look SUMMARY This article provides information regarding the operation and interpretation of Event Counters and Histograms available in ALTRUA, INSIGNIA, PULSAR MAX II, and DISCOVERY II pacemakers.
More information2015 CPT Codes for Cardiac Device Monitoring
2015 CPT Codes for Cardiac Device Monitoring CPT 1 copyright 2014. American Medical Association. All rights reserved. CPT is a registered trademark of the American Medical Association. CPT Code IMPLANTABLE
More informationMedtronic CareLink Network :: PACE article: Remote monitoring of ICDs: A prospective analysis
Remote Monitoring of Implantable Cardioverter Defibrillators: A Prospective Analysis Reprinted with permission from Pacing and Clinical Electrophysiology (PACE) Schoenfeld MH, Compton SJ, Mead RH, et al.
More informationXXXXX File No. 108655-001 Petitioner v. Issued and entered this 28 th day of June 2010 by Ken Ross Commissioner ORDER I PROCEDURAL BACKGROUND
STATE OF MICHIGAN DEPARTMENT OF ENERGY, LABOR & ECONOMIC GROWTH OFFICE OF FINANCIAL AND INSURANCE REGULATION Before the Commissioner of Financial and Insurance Regulation In the matter of XXXXX File No.
More informationINFORMED CONSENT FOR SLEEVE GASTRECTOMY
INFORMED CONSENT FOR SLEEVE GASTRECTOMY This informed-consent document has been prepared to help inform you about your Sleeve Gastrectomy including the risks and benefits, as well as alternative treatments.
More informationSientra Silicone Gel Breast Implants Quick Facts About Breast Augmentation And Reconstruction
Sientra Silicone Gel Breast Implants Quick Facts About Breast Augmentation And Reconstruction About This Brochure This brochure is intended to provide you with a high level overview of the facts about
More informationLong-term results of high vs. normal impedance ventricular leads on actual (Real-Life) pacemaker generator longevity
Europace (2009) 11, 200 205 doi:10.1093/europace/eun328 CLINICAL RESEARCH Leads and Lead Extraction Long-term results of high vs. normal impedance ventricular leads on actual (Real-Life) pacemaker generator
More informationHong Kong Joint Venture Agreements
Hong Kong Exchanges and Clearing Limited and The Stock Exchange of Hong Kong Limited take no responsibility for the contents of this announcement, make no representation as to its accuracy or completeness,
More informationDr. Sven Bode 24. Treffpunkt Medizintechnik 1. Juli 2010. Telemetrische Herzüberwachung Eine Dekade BIOTRONIK Home Monitoring
Dr. Sven Bode 24. Treffpunkt Medizintechnik 1. Juli 2010 Telemetrische Herzüberwachung Eine Dekade BIOTRONIK Home Monitoring RF GPRS GPRS https 0 Agenda Introduction to active implants and Home Monitoring
More informationVaginal Mesh: The FDA Decision and Repurcussions. Roger Dmochowski MD, FACS Dept of Urology Vanderbilt University Medical Center Nashville, TN
Vaginal Mesh: The FDA Decision and Repurcussions Roger Dmochowski MD, FACS Dept of Urology Vanderbilt University Medical Center Nashville, TN 1 ANATOMY FUNCTION 2 Mesh vs No Mesh Outcomes Sivaslioglu 2007
More informationNew CPT Codes for Cardiac Device Monitoring SIDE-BY-SIDE COMPARISON 2008-2009
New CPT Codes for Cardiac Device Monitoring SIDE-BY-SIDE COMPARISON 2008-2009 Effective January 1, 2009 NEW CPT 1 CODES FOR CARDIAC DEVICE MONITORING Twenty-three new CPT codes for Cardiac Device Monitoring
More informationDiagnostic Scoring System for LQTS
Medical Coverage Policy Genetic Testing: Congenital Long QT Syndrome Device/Equipment Drug Medical Surgery Test Other Effective Date: 2/15/2011 Policy Last Updated: 2/21/2012 Prospective review is recommended/required.
More informationSpecial Topics in Security and Privacy of Medical Information. Reminders. Last lecture: Recap. Sujata Garera. Project part 1 submission
Special Topics in Security and Privacy of Medical Information Sujata Garera Reminders Project part 1 submission Assignment 2 is online Last lecture: Recap Medical Telemetry Infrastructure Devices capturing
More informationCorporate Medical Policy
Corporate Medical Policy File Name: Origination: Last CAP Review: Next CAP Review: Last Review: spinal_cord_stimulation 3/1980 10/2015 10/2016 10/2015 Description of Procedure or Service Spinal cord stimulation
More informationRecalls of Cardiac Implants in the Last Decade: What Lessons Can We Learn?
RESEARCH ARTICLE Recalls of Cardiac Implants in the Last Decade: What Lessons Can We Learn? Shixuan Zhang 1, Christine Kriza 1,2, Sandra Schaller 1, Peter L. Kolominsky-Rabas 1,2 *, National Leading-Edge
More informationNASPE Training Requirements for Cardiac Implantable Electronic Devices: Selection, Implantation, and Follow-Up
NASPE Training Requirements for Cardiac Implantable Electronic Devices: Selection, Implantation, and Follow-Up DAVID L. HAYES, GERALD V. NACCARELLI, SEYMOUR FURMAN, VICTOR PARSONNET, DWIGHT REYNOLDS, NORA
More informationUpdated Cardiac Resynchronization Therapy Guidelines
The Ohio State University Heart and Vascular Center Updated Cardiac Resynchronization Therapy Guidelines William T. Abraham, MD, FACP, FACC, FAHA, FESC Professor of Medicine, Physiology, and Cell Biology
More informationScreening for Critical Congenital Heart Disease in the Apparently Healthy Newborn
Screening for Critical Congenital Heart Disease in the Apparently Healthy Newborn A presentation of Texas Pulse Oximetry Project: A Joint Educational Initiative of The University of Texas Health Science
More informationMinimally Invasive Mitral Valve Surgery
Minimally Invasive Mitral Valve Surgery Stanford Health Care offers leading, superior options in cardiac surgery, including the latest techniques and research for Minimally Invasive Cardiac surgery. Advanced
More informationJUL 2 2008. Ms. Kendra Basler Regulatory Affairs Associate Abbott Vascular Cardiac Therapies 3200 Lakeside Drive Santa Clara, CA 95054-2807
DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service JUL 2 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Kendra Basler Regulatory Affairs Associate Abbott Vascular
More informationFACTS about. MENTOR MemoryGel. Silicone Gel Filled Breast Implants
FACTS about MENTOR MemoryGel Silicone Gel Filled Breast Implants Are you considering breast implant surgery but not certain which type of implant to choose? You re not alone. Science-based information
More informationMRI SureScan Pacing System
MRI SureScan Pacing System Step-by-step Instructions For A Successful MRI Scan Timothy S. E. Albert, MD Medical Director Advanced Diagnostic Imaging Center Monterey, CA Lisa L. Hungate, BS, RN clinical
More informationSetting up and managing a remote management service
Setting up and managing a remote management service Stuart Allen Principal Cardiac Physiologist Manchester Heart Centre stuart.allen@cmft.nhs.uk My history Southampton Melbourne Manchester Increased use
More informationIl loop recorder esterno. Roberto Maggi
Il loop recorder esterno Roberto Maggi ECG monitoring and syncope In-hospital monitoring Holter Monitoring Event recorder External loop recorder Remote (at home) telemetry Implantable loop recorder External
More informationBreast Implants: Local Complications and Adverse Outcomes
Breast Implants: Local Complications and Adverse Outcomes This booklet highlights the most common problems associated with silicone gel-filled and saline-filled breast implants: those that occur in the
More informationFY2015 Final Hospital Inpatient Rule Summary
FY2015 Final Hospital Inpatient Rule Summary Interventional Cardiology (IC) Peripheral Interventions (PI) Rhythm Management (RM) On August 4, 2014, the Centers for Medicare & Medicaid Services (CMS) released
More information2010 CPT Codes for Cardiac Device Monitoring
2010 CPT Codes for Cardiac Device Monitoring CPT copyright 2009. American Medical Association. All rights reserved. CPT is a registered trademark of the American Medical Association. CPT Code IMPLANTABLE
More informationACLS Defibrillation Protocols With the ZOLL Rectilinear Biphasic Waveform AHA/ERC Guidelines 2005
ACLS Defibrillation Protocols With the ZOLL Rectilinear Biphasic Waveform AHA/ERC Guidelines 2005 Introduction: The purpose of this document is to outline the equivalent biphasic protocols specific to
More informationSecurity and Privacy of Wireless Implantable Medical Devices
Security and Privacy of Wireless Implantable Medical Devices Security Forum 2013 Hagenberg, 17.04.2013 Dipl.-Ing. Dr. Gregor Koenig Outlook Overview Device Hacking Safety & Utility Goals Security & Privacy
More informationConnecting Cardiac Device Data and the EHR
Connecting Cardiac Device Data and the EHR Introduction In an age of complex information needs, healthcare providers often use a staggering variety of applications to capture a patient s complete clinical
More informationComplications Associated with Pacemaker and ICD Generator Replacements and Upgrade Procedures: Results from the REPLACE Registry
Complications Associated with Pacemaker and ICD Generator Replacements and Upgrade Procedures: Results from the REPLACE Registry Jeanne E. Poole, MD; Marye J. Gleva, MD; Theofanie Mela, MD; Mina K. Chung,
More informationPower of Attorney for Health Care For
Power of Attorney for Health Care For Name: Date of Birth: Address: Telephone: This document is on file at Copies of this document have been given to my health care agent(s) and: 1. 2. 3. 4. 5. Courtesy
More informationHow should we treat atrial fibrillation in heart failure
Advances in Cardiac Arrhhythmias and Great Innovations in Cardiology Torino, 23/24 Ottobre 2015 How should we treat atrial fibrillation in heart failure Matteo Anselmino Dipartimento Scienze Mediche Città
More informationCLINICIAN MANUAL. LATITUDE NXT Patient Management System
CLINICIAN MANUAL LATITUDE NXT Patient Management System CAUTION: Federal (USA) law restricts this device to distribution and use by or on the lawful order of a physician. Table of Contents INTRODUCTION:
More informationSuffering from varicose veins? Patient Information. ELVeS Radial Minimally invasive laser therapy of venous insufficiency
Suffering from varicose veins? Patient Information ELVeS Radial Minimally invasive laser therapy of venous insufficiency Do you suffer from heavy legs or visible veins? This makes diseases of the veins
More information2/21/2016. Prolapse Surgery after Transvaginal Mesh: The Evolving Landscape. Disclosures. Objectives. No Relevant Disclosures
Prolapse Surgery after Transvaginal Mesh: The Evolving Landscape David R. Ellington, MD, FACOG Assistant Professor Division of Urogynecology and Pelvic Reconstructive Surgery Disclosures No Relevant Disclosures
More informationAdvantages of Right-Sided Implantation of Permanent Cardiac Pacemakers Performed by a Surgeon
Advantages of Right-Sided Implantation of Permanent Cardiac Pacemakers Performed by a Surgeon 7 Davor Horvat General Hospital in Karlovac, Croatia 1. Introduction A pacemaker is a medical device which
More informationCORONERS COURT NEW SOUTH WALES
CORONERS COURT NEW SOUTH WALES Inquest: Inquest into the death of Kerryn Foale Hearing dates: 19 & 20 August 2015 Date of findings: 26 September 2015 Place of findings: State Coroner s Court, Glebe Findings
More informationSpinal Cord Stimulation (SCS) Therapy: Fact Sheet
Spinal Cord Stimulation (SCS) Therapy: Fact Sheet What is SCS Therapy? Spinal cord stimulation (SCS) may be a life-changing 1 surgical option for patients to control their chronic neuropathic pain and
More informationInformation Sheet Guidance For IRBs, Clinical Investigators, and Sponsors
Information Sheet Guidance For IRBs, Clinical Investigators, and Sponsors Frequently Asked Questions About Medical Devices Additional copies are available from: Office of Good Clinical Practice Office
More information