Advances in Stereomicroscopy

Size: px
Start display at page:

Download "Advances in Stereomicroscopy"

Transcription

1 Advances in Stereomicroscopy H. Schnitzler*, Klaus-Peter Zimmer Leica Microsystems (Schweiz) AG, 9435 Heerbrugg, Switzerland ABSTRACT Similar to human's binocular vision, stereomicroscopes are comprised of two optical paths under a convergence angle providing a full perspective insight into the world's microstructure. The numerical aperture of stereomicroscopes has continuously increased over the years, reaching the point where the lenses of left and right perspective paths touched each other. This constraint appeared as an upper limit for the resolution of stereomicroscopes, as the resolution of a stereomicroscope was deduced from the numerical apertures of the two equally sized perspective channels. We present the optical design and advances in resolution of the world's first asymmetrical stereomicroscope, which is a technological breakthrough in many aspects of stereomicroscopes. This unique approach uses a large numerical aperture and thus an, so far, unachievable high lateral resolution in the one path, and a small aperture in the other path, which provides a high depth of field ("Fusion Optics"). This new concept is a technical challenge for the optical design of the zoom system as well as for the common main objectives. Furthermore, the new concept makes use of the particular way in which perspective information by binocular vision is formed in the human's brain. In conjunction with a research project at the University of Zurich, Leica Microsystems consolidated the functionality of this concept in to a new generation of stereomicroscopes. Keywords: Microscopy, Stereomicroscope, Binocular Vision, Zoom, Common Main Objective, Stereopsis 1. INTRODUCTION Binocular vision and stereoscopic perception are one of our primary senses. The input from our eyes is processed into three-dimensional picture of the space surrounding us. This seems to happen inconspicuously, continuously, and on a fast time scale without significant delay. As naturally and fundamental this physiological function appears in daily life, as complex are the processes involved. Recent research shows, that the retina is not just a receptor. It rather works like a piece of the brain interpreting the image information, like finding contours, motion, etc. About a dozen different kinds of interpreted information is send to the brain via the optical nerve. This and many more aspects of image processing, parallel and serial computing can be found in the physiology of human vision. In the past as well as today understanding the physiology of human s vision is of high interest in popular sciences [1] as well as for regular research [2], and it seems there is still a lot to be discovered. Insight into the small details in the microstructure of matter is another basic interest in research. The classical optical compound microscope got many partners like the electron microscope or scanning techniques like confocal microscopes. For three-dimensional observation, stereomicroscopes provide a perspective view into the world above micrometer scale, and diverse new techniques come into the field in the last decades like tomographic techniques, holography, and interferometric techniques. As different as all these techniques might appear they all have the common goal to provide better views into our nano- to micro-structures, at higher resolution, higher speed, and higher ease of use. The presentation reports about the latest advances in stereomicroscopes. Stereomicroscopes might be seen as the traditional approach to look at the micro world in three dimensions. They cannot compare with the highly dedicated systems of later developments. However, their purely optical functionality, no principal requirement of electronics or computers, and observation with bare eyes in real time make them rather universatile tabletop instruments well suitable for a large variety of applications. Two convergent optical beam path, zoom optics, long working distances, and the request of high resolution make stereomicroscopes a challenge for the optical designers. Optical Design and Engineering III, edited by Laurent Mazuray, Rolf Wartmann, Andrew Wood, Jean-Luc Tissot, Jeffrey M. Raynor, Proc. of SPIE Vol. 7100, 71000P 2008 SPIE CCC code: X/08/$18 doi: / Proc. of SPIE Vol P-1

2 INTEROCULAB The presentation introduces the optical layout of two types of stereomicroscopes, Greenough- and CMO-type. Challenges and limitations for the optical design and performance are revealed. In face with request to step further in terms of resolution as well of depth of field, it is shown how an alternative new approach was found and validated. Conclusively, the achieved advances are explained and summarized. 2. STEREOMICROSCOPES A stereomicroscope allows stereoscopic observation of micro- and macroscopic objects. The optical layout comprises two beam paths emerging from the object under a convergence angle. The two beams pass a pair of magnification changers and are finally brought to the observer s eyes with a binocular tube. 2.1 Greenough type stereomicroscopes The invention of stereomicroscopes goes back to Horatio S. Greenough in the year It was found by the combination of two compound microscopes under a convergence angle (Fig. 1). Besides of the two optical paths the image erecting is of essential importance to achieve proper depth perception. P4T BEVEL GEARS Fig. 1. Optical system of a Greenough type stereomicroscope. The drawback of Greenough s optical set up is found in the tilted focal planes of left and right channel. Proper overlap of the focal planes within the depth of field is given near the object s centre, only. Nevertheless, Greenough type stereomicroscopes allow a compact design and are of good use for many applications till today. Proc. of SPIE Vol P-2

3 2.2 Common-main-objective type stereomicroscopes One could reconcile the focal planes of left and right perspective channels with the common main objective (CMO) optical layout. A typical optical layout of a present high performance CMO stereomicroscope is sketched in Fig. 2. Observer Eyepieces Intermediate Images Image Inversion Tube Lenses Stereo Zoom Entrance Pupils stereo basis Common Main Objective IL Workirnz V I I i I J I na Distance Object Fig. 2. Optical layout of a CMO type stereomicroscope The object plane is in the focal plane of the common main objective. This creates an afocal beam path after the objective, which is also the interface to a large set of accessories offered together with CMO stereomicroscopes. A telescope type magnification changer, typically a zoom, is placed between objective and tube. The diameter of the entrance pupil D EP of the zoom is physically defined by the optical components of the zoom, and varies with the zoom position to finally achieve a good ratio between magnification and resolution. Together with the focal length f O of the common main objective one finds the numerical aperture na of the stereomicroscope as D na =. (1) 2 f EP O Proc. of SPIE Vol P-3

4 2.3 Basic optical considerations The optical resolution R of a stereomicroscope is limited by diffraction of light [3]. The resolution (in units of line pairs per millimetre) translates from the numerical aperture as lp R = 3000 na. (2) mm Let the telescopic magnification of the zoom be Γ, the tube lens have focal length f TL, and the eyepiece have a visual 250 mm magnification M E =, then the total visual magnification of the instrument is f E M TOT VIS ftl = Γ M E. (3) f O Considering the resolution of human s eye, one finds a relation for the useful magnification of the instrument, in which the resolution of the instrument matches the receiver s resolution 500 VIS na < M TOT < 1000 na. (4) Great lateral resolution as described in (2) is only the half win in stereomicroscopy because the object s topographic properties are of significant interest, too. The sharpness in vertical direction in object space is commonly described by Max Berek s formula [4] for the depth of field λ 0.34 mm δ = +. (5) 2 2 na M TOT VIS na As known, both formulas resolution (2) and depth of field (5) depend on the numerical aperture but in physically conflicting directions. 2.4 Limitations of the current design of stereomicroscopes Over the past years the resolution (2) was pushed to higher and higher levels. According to Fig. 2 two identical telescopes are arranged side by side, symmetrically in the device. The plane of symmetry divides the object symmetrically into a right and a left half. The distance between the telescope axes is referred to as stereo basis. To achieve higher resolutions the entrance pupil diameters (1) of the telescopes were expanded. A limitation is set when the physical dimensions of the telescopes lenses come into touch of each other. Other options to further increase the numerical aperture are to decrease the focal length of the common main objective or to expand the stereo basis (compare Fig. 2). Shortening the focal length of the objective would decrease the working distance of the instrument. Working distance is a key figure for stereomicroscopes where accessibility and manipulation of the specimen under 3-dimensional observation in real time is a key application. Increase of the stereo basis leads to more bulky mechanical design of the stereo zoom module, as well as for the optical and mechanical design of the objectives. Beyond that, both approaches increase the convergence angle. The object side convergence angle is a figure in the physiological perception chain how the depth impression is formed. The convergence angle on the observer s side for CMO type stereomicroscopes is conventionally close to zero, which allows relaxed stereoscopic viewing. The mismatch between object-side and observer-side convergence angle causes an exaggerated depth impression for human s 3- dimensional perception. In the past, the for the sake of higher resolution, the focal length of the objectives were already brought down to the edge of what is reasonable within proper stereoscopic perception. The figures of stereo basis and/or focal length of the objectives have already experienced many generations of optimization. The margin for significant improvements on them appears to be rather exhausted. Proc. of SPIE Vol P-4

5 3. IDEAS TO ADVANCE New ideas had to be found to really achieve significant improvements. When the entrance pupils become larger than the stereo basis and if the stereo basis shall be conserved then the pupils overlap as sketched in Fig. 3. The overlapping beams could be split into two separated optical paths by use of a semitransparent mirror. Stereo Zoom..., / /, -, I, Stereo Basis fo ye1we \ina I Fig. 3. Sketch of an optical layout where the entrance pupils are wider than the stereo basis. The Stereo Basis sin(convergence Angle/2) = 2 f is kept the same as in the set up depicted in Fig. 2. The dashed line indicates O a semi transparent mirror, which splits the overlapping entrance pupils into two separated optical paths. Going into more details, this optical design turns out to become even more bulky as drawn in the sketch above: as the pupils of the stereomicroscope are defined by the zoom lenses, the entrance pupils lay somewhere in the zoom. Due to the field angle the long optical path between zoom and common main objective leads to a huge effective diameter required for the lenses of objective. The result would be very bulky and also very costly and highly demanding on the optical design of the objective. Other stereoscopic approaches are found where common optics is used for the objective and the zoom, having the perspective split after the zoom. Still, entrance pupils wider than the stereo basis still mean overlapping optical beams. Such an optical design is reasonable as long as the two perspective optical beams are not too wide and not too far out of the physical symmetry axes of the lenses. The optical correction of an off-symmetry zoom can hardly achieve the optical quality, apertures, and the zoom range as known for the classic optical design. In consequence, this set up can only achieve small convergence angles with very poor stereoscopic depth perception. Even more, the convergence angle Proc. of SPIE Vol P-5

6 changes with the zoom if the perspective split is placed after a variable focal length, which leads to a very unfavourable perception. Of course, combination of the above set ups is possible. But with respect to ease of use, handiness, and cost efficiency, it is not the primary choice, neither from the designer s nor from the customer s point of view. Having a view into the nature, we are rarely shown pupil sizes as wide (or wider) than the interpupillary distance. The pupils of owls and eagles are still in good distance to each other. It seems that Mother Nature uses the optical perception rather for distant objects and not for microscopic objects in close distance. But still the observation of nature indicates the great potential hidden in the physiology of image processing, e.g. the sensitivity on motion for the eagle. Use of existing physiological potential in the human s brain is finally the key contribution to the advances that we achieved with the new optical design for stereomicroscopes, which we titled fusion optics. 4. FUSION OPTICS The idea behind Fusion Optics is to apply differently sized entrance pupils for the left and right perspective channel. The wide entrance pupil of the one channel (might be wider than the stereo basis) provides high spatial resolution. The narrow narrow entrance pupil of the other channel simultaneously provides increased depth of field. The fused image perceived by the observer will locally extract the best resolution available from either side, leading to an stereoscopic perception superior in spatial resolution and in depth of field at the same time. An optical layout of a CMO type stereo zoom with differently sized entrance pupils is sketched in Fig. 4. Entrance Pupils Stereo Basis T fo Fig. 4. Optical layout of a CMO type stereomicroscope with differently sized entrance pupils. While the idea is explained very quickly, still many aspects need to be evaluated, validated, and confirmed. The specification of the optical design allows wide variations. 4.1 Specification Without loss of generality, but for better quantitave discussion we now assume the stereo basis to be 24 mm and the reference focal length to be f O = 80 mm, which is the same as established in current high performance stereomicroscopes. A standard outfit comprising an objective of reference focal length, a standard tube ( f TL = 160 mm), and standard 10x eyepieces is considered. Proc. of SPIE Vol P-6

7 The zoom shall be set to a maximum magnification position, resulting in a total visual magnification of 160. The diameters of the entrance pupils are assumed to be 28 mm for the right channel and 20 mm for the left channel. This describes the optical layout as it is chosen in our newly developed high performance stereomicroscope Leica M205. Tab. 1 collects the optical key performance figures for a stereomicroscope of this kind, i.e. with differently sized entrance pupils. Left perspective channel Right perspective channel Entrance pupil diameter 20 mm 28 mm Numerical aperture Spatial resolution 375 lp/mm 525 lp/mm Depth of field mm mm Luminosity 51% of right channel 100% Contribution to the convergence angle (17 ) 10 7 Diameter of exit pupil 0.39 mm 0.55 mm Tab. 1: Optical key figures for a stereo zoom with differently sized entrance pupils. The figures are evaluated for a standard outfit and the zoom set to maximum magnification. Detailed discussion in the text. The consequences of the listed values will be discussed in the following; needless to say that magnification and object field are always the same for both perspective channels throughout all configurations. The exciting feature of this optical design is how the observer merges the two images, which locally show slightly different details due to different resolution and depth of field, respectively. Extensive scientific studies were done at the University and ETH Zurich which are presented in section 4.2. The differences in luminosity between left and right channel principally could be compensated using a filter, but fortunately the human eyes provide great flexibility in terms of individual adaptation. There is no physiological coupling for the adaptation between left and right eye (different to humans eyes accommodation process). Unequal adaption between left and right eye is well possible and even rather normal. Human eyes adapt to different light situations quickly and individually, as we can easily observe in daily life, when exposed to and moving in direct sunlight and when covering one eye. Beyond that, it is even found, that the human eye is not very sensitive to differences in luminosity in the same image, e.g. vignetting of 50% to the sides of a images is hardly recognized. The overall convergence angle of the suggested stereomicroscope remains the same as what is was in a symmetric set up, because stereo basis and focal length are kept the same. If the objective were centred to the centre of the stereo channels the perspective impression would be accurately the same. Since the effectively used area is laterally shifted to the right side, it is possible to avoid too bulky objective designs if the objective is shifted the same way. Then, the viewing angles for left and right perspective in the centre of the object field are not the same. This situation is the same as if the centre of the object would not be observed from exactly above. Fortunately the human perception is well trained to look at extended objects over which the viewing angle varies significantly. The particular point of the object field where the viewing angles of left and right side are equal not perceived different from the other points. Slight displacements of the objective to the centre of the stereo basis therefore do not mean a significant change for stereoscopic perception. Proc. of SPIE Vol P-7

8 4.2 Fusion of binocular images providing locally different resolutions This section is to discuss the stereoscopic perception from a pair of perspective images with locally slightly different resolutions, as received from a stereomicroscope like in Fig. 4. The phenomenon of binocular rivalry is described in literature [2]. It is known that similar images presented to the two eyes appear as one and are processed simultaneously rather than successively. Superimposed images from the two eyes always rival by mutual inhibition (suppression theory). Only one eye s input is seen at any one time, but the eye that dominates varies from place to place and alternates over time, resulting in a mosaic of dominance and suppression. Less is known what the rules of combination in the fusion process are like. Several studies ([5], [6], [7]) showed that when one eye views a blurred version of the other eye s image, an inter-ocular blur suppression mechanism ensures a blur free percept. Moreover, Schor et al. [7] showed that inter-ocular blur suppression acts locally: When two nearly identical monocular images each contain blurred regions at different locations, the resulting binocular percept contains only non-blurred regions. In other words, the visual system is able to select, for each spatial location, the monocular signal with the highest resolution. This remarkable ability explains why presbyopic patients wearing one lens in the dominant eye for distant vision and a lens for near vision in the other (a situation known as monovision) enjoy blur-free vision at all distances [7]. To date, it was not clear whether inter-ocular blur suppression can occur for regions located at different depths. In other words, it was not known whether the human visual system can use binocular disparities to derive depth information and simultaneously suppress blur in one of the two eyes. A study at the University and ETH Zurich was performed to determine whether the human visual system is capable of extracting the highest resolution available at each location, irrespective of the eye of origin. Visual acuity at different depth planes created by binocular disparity was measured, and it was investigated whether the eye of origin of the signals influence performance. The results show that the human visual system can use the highest resolution available between the two monocular images, irrespective of eye of origin and depth plane [8]. 4.3 Optical design The specification of the first fusion optics type CMO stereomicroscope comprises a 20.5 x stereo zoom, with an entrance pupil on one side as large as 28 mm. With the standard objective of f O = 80 mm attached a numerical aperture of is achieved, which allows a so far unachieved magnification range of x. This gives a ratio of M TOT VIS /na = 914 at maximum magnification and so nicely fulfils the rules of useful magnification as given in (4). A field number of 23 mm leads to an object field from 1.44 mm to 29 mm. The optical layout is sketched in Fig. 5. J rn LH Fig. 5. Optical design of the first stereo zoom with fusion optics. Upper half of the picture shows the lens position and optical rays at minimum magnification position, lower half of the picture at maximum magnification position. Proc. of SPIE Vol P-8

9 Each zoom consists of 11 lenses in four groups with positive-negative-positive-negative refractive powers. Telescopic magnification goes from.39x to 8x. Lenses of both channels have the same optical design except for the free diameter of the first lens group. This results in the same zoom curve for both sides. The outer lens groups stay at fixed positions while both of the inner lens groups travel in opposite directions. The zoom (and the complete microscope) is apochromatically corrected over the whole range. An iris diaphragm is placed at a fixed position near the centre of the zoom. The size of the entrance pupil depends on the zoom position. The discussed physical limitation of the entrance pupils due to the spacing of the stereo basis is only a restriction near the maximum magnification position, where the high resolution is required. For this reason the idea of fusion optics was actually implemented only for the high zoom positions. In the middle and low zoom positions resolution and depth of field are sufficient, why the zoom design implements equally sized beam paths for both channels. A detailed diagram of the behaviour of the numerical apertures over the zoom range will be given in Fig. 6. The physical size of the front lens group carefully needs to be taken into consideration also at the lower magnification positions. Without further constraints the zoom might tend to pull the axial position of the entrance pupil to its interior. The optical designer must make sure that the field angles have a proper path through the front lens group without significant vignetting, also for the narrow channel. The stereomicroscope is embedded in the system of Leica s high performance stereomicroscopes offering a large variety of common main objectives, among which a 2x objective is available. Using it, a numerical aperture of 0.35 is achieved, so reaching a resolution of 1050 lp/mm. These resolutions until now were reserved for compound microscopes and have now become available under full stereoscopic vision and at a working distance of 20 mm. Proc. of SPIE Vol P-9

10 4.4 Performance of fusion optics The optical layout as described in 4.3 leads to a behaviour of the numerical apertures as shown in Fig. 6. Fig. 6. Behaviour of the numerical aperture over the whole zoom range. Solid black line shows the numerical apertures of the fusion optics designed stereo zoom, for left and right channel, respectively. Dashed gray line shows the numerical aperture for a conventional stereo zoom with equally sized entrance pupils; since maximum achievable resolution is lower than with the fusion optics layout, the magnification range ends at a lower position, too, to avoid empty magnification. Proc. of SPIE Vol P-10

11 The resolution achieved with a stereomicroscope having differently sized pupils is shown in Fig. 7. The right channel has the highest peak resolution in the focal plane of the objective, while the left channel has a lower peak but higher depth of field and thus a higher resolution out of the focal plane. For comparison, the dashed line shows the resolution for a stereomicroscope with symmetrically sized entrance pupils. Resolution [lp/mm],/3oo z [mm] Fig. 7. Resolution in different levels z in object space at maximum magnification position. Solid lines show the right and left channel, respectively. Thick solid line is the perceived resolution after fusion of both channels. Dashed gray line is the achievable performance with a stereomicroscope of equally sized entrance pupils. The perceived resolution with the fusion optics layout is superior in all z-planes. 5. CONCLUSION In this presentation the limitation of conventional stereomicroscopes was shown, which prevents these microscopes to achieve higher resolution. An idea how to overcome this limitation was explained and validated. The fusion optics called optical layout supplies the observer with slightly differently resolved stereoscopic images. The physiological fusion process finally lets the observer perceive a 3-dimensional image with increased resolution and increased depth of field at the same time. The optical design was presented and the performance described. This new approach combined with a 20.5x zoom and a new series of CMO objectives now allows so far unachieved resolutions up to 1050 lp/mm under full stereoscopic observation at a free working distance of 20 mm. Proc. of SPIE Vol P-11

12 REFERENCES [1] [2] [3] [4] [5] [6] [7] [8] Spektrum der Wissenschaft 41 05/08 Ian P. Howard, Brian J. Rogers, Binocular Vision and Stereopsis, Oxford Psychology Series No. 29, Born & Wolf, Principles of Optics, 6th Ed., Cambridge University Press, 694 (1980) Max Berek, Grundlagen der Tiefenwahrnehmung im Mikroskop, Marburger Sitzungsberichte 61, (1927) Collins, M.J. and A. Goode, Interocular blur suppression and monovision. Acta Ophthalmol (Copenh), 72(3), (1994) Johannsdottir, K.R. and L.B. Stelmach, Monovision: a review of the scientific literature. Optom Vis Sci, 78(9), (2001) Schor, C., L. Landsman, and P. Erickson, Ocular dominance and the interocular suppression of blur in monovision, Am J Optom Physiol Opt, 64(10), (1987) C. Schulthess, H. Schnitzler, D.C. Kiper, Spatial visual acuity at different depth planes: The role of eye of origin, Institute of neuroinformatics, University and ETH Zurich, to be published (2008) Proc. of SPIE Vol P-12

Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.

Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away. Revision problem Chapter 18 problem 37 page 612 Suppose you point a pinhole camera at a 15m tall tree that is 75m away. 1 Optical Instruments Thin lens equation Refractive power Cameras The human eye Combining

More information

Light and Sound. Pupil Booklet

Light and Sound. Pupil Booklet Duncanrig Secondary School East Kilbride S2 Physics Elective Light and Sound Name: Pupil Booklet Class: SCN 3-11a - By exploring the refraction of light when passed through different materials, lenses

More information

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM

More information

Endoscope Optics. Chapter 8. 8.1 Introduction

Endoscope Optics. Chapter 8. 8.1 Introduction Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize

More information

Care and Use of the Compound Microscope

Care and Use of the Compound Microscope Revised Fall 2011 Care and Use of the Compound Microscope Objectives After completing this lab students should be able to 1. properly clean and carry a compound and dissecting microscope. 2. focus a specimen

More information

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus

More information

2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec.

2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec. Physics for Scientists and Engineers, 4e (Giancoli) Chapter 33 Lenses and Optical Instruments 33.1 Conceptual Questions 1) State how to draw the three rays for finding the image position due to a thin

More information

Introduction to 3D Imaging

Introduction to 3D Imaging Chapter 5 Introduction to 3D Imaging 5.1 3D Basics We all remember pairs of cardboard glasses with blue and red plastic lenses used to watch a horror movie. This is what most people still think of when

More information

Lenses and Apertures of A TEM

Lenses and Apertures of A TEM Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics

More information

Information. From the LowVision Specialists. Guidelines for the fitting of telescopic systems

Information. From the LowVision Specialists. Guidelines for the fitting of telescopic systems Information From the LowVision Specialists Guidelines for the fitting of telescopic systems About a successful fitting Eye care professionals dispensing telescopic spectacles must ensure they have successfully

More information

Lecture 17. Image formation Ray tracing Calculation. Lenses Convex Concave. Mirrors Convex Concave. Optical instruments

Lecture 17. Image formation Ray tracing Calculation. Lenses Convex Concave. Mirrors Convex Concave. Optical instruments Lecture 17. Image formation Ray tracing Calculation Lenses Convex Concave Mirrors Convex Concave Optical instruments Image formation Laws of refraction and reflection can be used to explain how lenses

More information

SLIT LAMP BQ 900 Sophisticated microscopy

SLIT LAMP BQ 900 Sophisticated microscopy SLIT LAMP BQ 900 Sophisticated microscopy Tradition and Innovation Since 1858, visionary thinking and a fascination with technology have guided us to develop innovative products of outstanding reliability:

More information

First let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between

First let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between Optical Differences Between Telescopes and Microscopes Robert R. Pavlis, Girard, Kansas USA icroscopes and telescopes are optical instruments that are designed to permit observation of objects and details

More information

4. CAMERA ADJUSTMENTS

4. CAMERA ADJUSTMENTS 4. CAMERA ADJUSTMENTS Only by the possibility of displacing lens and rear standard all advantages of a view camera are fully utilized. These displacements serve for control of perspective, positioning

More information

Optical laser beam scanner lens relay system

Optical laser beam scanner lens relay system 1. Introduction Optical laser beam scanner lens relay system Laser beam scanning is used most often by far in confocal microscopes. There are many ways by which a laser beam can be scanned across the back

More information

Magnification Devices

Magnification Devices LOW VISION AIDS Optical Characteristics of the Low Vision Patient The definition of visual loss includes two components and limited resolving power or acuity, a blur that can't be eliminated with a simple

More information

Measuring. User Manual

Measuring. User Manual 0 1 2 3 4 5 6 7 8 9 10 11 Measuring User Manual Accessories for measuring tasks Stage micrometer (1) for calibration Graticules with various measuring pitches (2) in mm and inches Graticule with mesh (3)

More information

Product Range Overview

Product Range Overview Product Range Overview Stereo Optical Inspection Microscopes Non-Contact Measurement Systems Laboratory Microscopes Metallurgical Microscopes FM 557119 Vision Engineering Ltd has been certificated for

More information

9/16 Optics 1 /11 GEOMETRIC OPTICS

9/16 Optics 1 /11 GEOMETRIC OPTICS 9/6 Optics / GEOMETRIC OPTICS PURPOSE: To review the basics of geometric optics and to observe the function of some simple and compound optical devices. APPARATUS: Optical bench, lenses, mirror, target

More information

Anamorphic imaging with three mirrors: a survey

Anamorphic imaging with three mirrors: a survey Anamorphic imaging with three mirrors: a survey Joseph M. Howard Optics Branch (Code 551), NASA Goddard Space Flight Center, Greenbelt, MD 20771 Ph: 301-286-0690 Fax: 301-286-0204 Joseph.M.Howard@nasa.gov

More information

Rodenstock Photo Optics

Rodenstock Photo Optics Apo-Sironar-S Apo-Macro-Sironar Apo-Grandagon Grandagon-N Accessories: Center filters Accessories: Focus-Mount Lenses for Analog Professional Photography Even in the age of digital photography, the professional

More information

Light and its effects

Light and its effects Light and its effects Light and the speed of light Shadows Shadow films Pinhole camera (1) Pinhole camera (2) Reflection of light Image in a plane mirror An image in a plane mirror is: (i) the same size

More information

Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012

Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012 Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012 Figure 1.1 illustrates the parts of an upright compound microscope and indicates the terminology that I use in these notes. Figure 1.1. Parts of a Compound

More information

ReLEx smile Minimally invasive vision correction Information for patients

ReLEx smile Minimally invasive vision correction Information for patients ReLEx smile Minimally invasive vision correction Information for patients Seeing is living Our eyes are our most important sensory organ. The human brain obtains over 80 % of its information via the sense

More information

The Kaps SOM -Series. Stereo Examination Microscopes for ENT Specialists. The variable space-saving system for every ENT practice

The Kaps SOM -Series. Stereo Examination Microscopes for ENT Specialists. The variable space-saving system for every ENT practice The Kaps SOM -Series Stereo Examination Microscopes for ENT Specialists The variable space-saving system for every ENT practice The Kaps SOM -Series The economical modern version for ENT practice and minor

More information

LEICA TRI-ELMAR-M 28 35 50 mm f/4 ASPH. 1

LEICA TRI-ELMAR-M 28 35 50 mm f/4 ASPH. 1 LEICA TRI-ELMAR-M 28 35 5 mm f/4 ASPH. 1 This lens combines three of the focal lengths that are most popular among Leica M photographers. The appropriate bright-line frame appears in the viewfinder when

More information

Understanding astigmatism Spring 2003

Understanding astigmatism Spring 2003 MAS450/854 Understanding astigmatism Spring 2003 March 9th 2003 Introduction Spherical lens with no astigmatism Crossed cylindrical lenses with astigmatism Horizontal focus Vertical focus Plane of sharpest

More information

Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information

Rediscover quality of life thanks to vision correction with technology from Carl Zeiss. Patient Information Rediscover quality of life thanks to vision correction with technology from Carl Zeiss Patient Information 5 2 It was really w Vision defects: Light that goes astray For clear vision the eyes, cornea and

More information

6 Space Perception and Binocular Vision

6 Space Perception and Binocular Vision Space Perception and Binocular Vision Space Perception and Binocular Vision space perception monocular cues to 3D space binocular vision and stereopsis combining depth cues monocular/pictorial cues cues

More information

Imaging techniques with refractive beam shaping optics

Imaging techniques with refractive beam shaping optics Imaging techniques with refractive beam shaping optics Alexander Laskin, Vadim Laskin AdlOptica GmbH, Rudower Chaussee 29, 12489 Berlin, Germany ABSTRACT Applying of the refractive beam shapers in real

More information

Forensic Science: The Basics. Microscopy

Forensic Science: The Basics. Microscopy Forensic Science: The Basics Microscopy Chapter 6 Jay A. Siegel,Ph.D. Power point presentation by Greg Galardi, Peru State College, Peru Nebraska Presentation by Greg Galardi, Peru State College CRC Press,

More information

Rodenstock Photo Optics

Rodenstock Photo Optics Rogonar Rogonar-S Rodagon Apo-Rodagon N Rodagon-WA Apo-Rodagon-D Accessories: Modular-Focus Lenses for Enlarging, CCD Photos and Video To reproduce analog photographs as pictures on paper requires two

More information

XXIX ENFMC - Annals of Optics 2006. 3D shadowgram projection using a simple diffractive screen José J. Lunazzi, Noemí I. Rivera Rodríguez.

XXIX ENFMC - Annals of Optics 2006. 3D shadowgram projection using a simple diffractive screen José J. Lunazzi, Noemí I. Rivera Rodríguez. 3D shadowgram projection using a simple diffractive screen José J. Lunazzi, Noemí I. Rivera Rodríguez. Universidade Estadual de Campinas - Instituto de Física lunazzi@ifi.unicamp.br, nrivera@ifi.unicamp.br

More information

waves rays Consider rays of light from an object being reflected by a plane mirror (the rays are diverging): mirror object

waves rays Consider rays of light from an object being reflected by a plane mirror (the rays are diverging): mirror object PHYS1000 Optics 1 Optics Light and its interaction with lenses and mirrors. We assume that we can ignore the wave properties of light. waves rays We represent the light as rays, and ignore diffraction.

More information

MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY

MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY Adapted from Foundations of Biology I; Lab 6 Introduction to Microscopy Dr. John Robertson, Westminster College Biology Department,

More information

Optical Illusions Essay Angela Wall EMAT 6690

Optical Illusions Essay Angela Wall EMAT 6690 Optical Illusions Essay Angela Wall EMAT 6690! Optical illusions are images that are visually perceived differently than how they actually appear in reality. These images can be very entertaining, but

More information

Handy Pinhole Camera (Latin Camera Obscura) to observe the transit of Venus, eclipses and other phenomena occurring on the Sun

Handy Pinhole Camera (Latin Camera Obscura) to observe the transit of Venus, eclipses and other phenomena occurring on the Sun Lech Mankiewicz Centre for Theoretical Physics, Polish Academy of Sciences, Warsaw, Global Intelligent Robotic Telescopes Network GLORIA http://www.gloria-project.eu/ Paweł Rudawy Astronomical Institute,

More information

Space Perception and Binocular Vision

Space Perception and Binocular Vision Space Perception and Binocular Vision Space Perception Monocular Cues to Three-Dimensional Space Binocular Vision and Stereopsis Combining Depth Cues 9/30/2008 1 Introduction to Space Perception Realism:

More information

THE COMPOUND MICROSCOPE

THE COMPOUND MICROSCOPE THE COMPOUND MICROSCOPE In microbiology, the microscope plays an important role in allowing us to see tiny objects that are normally invisible to the naked eye. It is essential for students to learn how

More information

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,

More information

Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images

Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images FORMATEX 2007 A. Méndez-Vilas and J. Díaz (Eds.) Software-based three dimensional reconstructions and enhancements of focal depth in microphotographic images Jörg Piper Clinic Meduna, Department for Internal

More information

1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002

1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical

More information

PHYS 39a Lab 3: Microscope Optics

PHYS 39a Lab 3: Microscope Optics PHYS 39a Lab 3: Microscope Optics Trevor Kafka December 15, 2014 Abstract In this lab task, we sought to use critical illumination and Köhler illumination techniques to view the image of a 1000 lines-per-inch

More information

Apo-Nikkor and Process-Nikkor large format lenses by Daniel W. Fromm

Apo-Nikkor and Process-Nikkor large format lenses by Daniel W. Fromm Apo-Nikkor and Process-Nikkor large format lenses by Daniel W. Fromm 1 Apo-Nikkors, according to Nikon Nikon made two families of Apo-Nikkor lenses, both intended to be used for, e.g., photoengraving.

More information

Chapter 17: Light and Image Formation

Chapter 17: Light and Image Formation Chapter 17: Light and Image Formation 1. When light enters a medium with a higher index of refraction it is A. absorbed. B. bent away from the normal. C. bent towards from the normal. D. continues in the

More information

Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012

Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 Chapter 13 Confocal Laser Scanning Microscopy C. Robert Bagnell, Jr., Ph.D., 2012 You are sitting at your microscope working at high magnification trying to sort out the three-dimensional compartmentalization

More information

PDF Created with deskpdf PDF Writer - Trial :: http://www.docudesk.com

PDF Created with deskpdf PDF Writer - Trial :: http://www.docudesk.com CCTV Lens Calculator For a quick 1/3" CCD Camera you can work out the lens required using this simple method: Distance from object multiplied by 4.8, divided by horizontal or vertical area equals the lens

More information

Lenses and Telescopes

Lenses and Telescopes A. Using single lenses to form images Lenses and Telescopes The simplest variety of telescope uses a single lens. The image is formed at the focus of the telescope, which is simply the focal plane of the

More information

EXPERIMENT 6 OPTICS: FOCAL LENGTH OF A LENS

EXPERIMENT 6 OPTICS: FOCAL LENGTH OF A LENS EXPERIMENT 6 OPTICS: FOCAL LENGTH OF A LENS The following website should be accessed before coming to class. Text reference: pp189-196 Optics Bench a) For convenience of discussion we assume that the light

More information

Technical Drawing Specifications Resource A guide to support VCE Visual Communication Design study design 2013-17

Technical Drawing Specifications Resource A guide to support VCE Visual Communication Design study design 2013-17 A guide to support VCE Visual Communication Design study design 2013-17 1 Contents INTRODUCTION The Australian Standards (AS) Key knowledge and skills THREE-DIMENSIONAL DRAWING PARALINE DRAWING Isometric

More information

The Limits of Human Vision

The Limits of Human Vision The Limits of Human Vision Michael F. Deering Sun Microsystems ABSTRACT A model of the perception s of the human visual system is presented, resulting in an estimate of approximately 15 million variable

More information

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie marc.lavoie@cavehill.uwi.edu MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment

More information

A new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31

A new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31 A new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31 Phase contrast Relief contrast Incorporation of advanced UIS2 optics ensures the highest level of clarity for cell checking applications.

More information

Leica MZ125. Modular high-performance stereomicroscope with 12.5:1 zoom

Leica MZ125. Modular high-performance stereomicroscope with 12.5:1 zoom Leica MZ125 Modular high-performance stereomicroscope with 12.5:1 zoom The vision Leading-edge technology for a new era In research and technology the bar has been raised. Advances in science and industry

More information

British Journal of Science 43 September 2012, Vol. 6 (2) Study of the Objective Focal Properties for Asymmetrical Double Polepiece Magnetic Lens

British Journal of Science 43 September 2012, Vol. 6 (2) Study of the Objective Focal Properties for Asymmetrical Double Polepiece Magnetic Lens British Journal of Science 43 Study of the Objective Focal Properties for Asymmetrical Double Polepiece Magnetic Lens Talib M. Abbass 1 and Ban A. Nasser 2 1 Department of Physics, College of Education

More information

PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES

PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES British Made Portable Microscopes with Image Erection PCB Through - Hole Scope Depthscope Brinell Hardness Impression Microscope Cathode Ray Tube Inspector Portable Video Scope Metallurgical Microscope

More information

Alignement of a ring cavity laser

Alignement of a ring cavity laser Alignement of a ring cavity laser 1 Introduction This manual describes a procedure to align the cavity of our Ti:Sapphire ring laser and its injection with an Argon-Ion pump laser beam. The setup is shown

More information

EXPERIMENT #1: MICROSCOPY

EXPERIMENT #1: MICROSCOPY EXPERIMENT #1: MICROSCOPY Brightfield Compound Light Microscope The light microscope is an important tool in the study of microorganisms. The compound light microscope uses visible light to directly illuminate

More information

Recommended alignment procedure for Shaper 6_6 / Focal- Shaper 9

Recommended alignment procedure for Shaper 6_6 / Focal- Shaper 9 Recommended alignment procedure for Shaper 6_6 / Focal- Shaper 9 The described below procedure presents an example of alignment of beam shapers Shaper and Focal- Shaper (F- Shaper) with using the standard

More information

Color holographic 3D display unit with aperture field division

Color holographic 3D display unit with aperture field division Color holographic 3D display unit with aperture field division Weronika Zaperty, Tomasz Kozacki, Malgorzata Kujawinska, Grzegorz Finke Photonics Engineering Division, Faculty of Mechatronics Warsaw University

More information

Physics 116. Nov 4, 2011. Session 22 Review: ray optics. R. J. Wilkes Email: ph116@u.washington.edu

Physics 116. Nov 4, 2011. Session 22 Review: ray optics. R. J. Wilkes Email: ph116@u.washington.edu Physics 116 Session 22 Review: ray optics Nov 4, 2011 R. J. Wilkes Email: ph116@u.washington.edu ! Exam 2 is Monday!! All multiple choice, similar to HW problems, same format as Exam 1!!! Announcements

More information

The Olympus stereology system. The Computer Assisted Stereological Toolbox

The Olympus stereology system. The Computer Assisted Stereological Toolbox The Olympus stereology system The Computer Assisted Stereological Toolbox CAST is a Computer Assisted Stereological Toolbox for PCs running Microsoft Windows TM. CAST is an interactive, user-friendly,

More information

ZEISS Compact Prime and Zoom lenses Flexibility and performance in a winning combination.

ZEISS Compact Prime and Zoom lenses Flexibility and performance in a winning combination. ZEISS Compact Prime and Zoom lenses Flexibility and performance in a winning combination. // FLEXIBILITY MADE BY ZEISS The moment you shout action! and feel your story come together. This is the moment

More information

Microscope Lab Introduction to the Microscope Lab Activity

Microscope Lab Introduction to the Microscope Lab Activity Microscope Lab Introduction to the Microscope Lab Activity Wendy Kim 3B 24 Sep 2010 http://www.mainsgate.com/spacebio/modules/gs_resource/ CellDivisionMetaphase.jpeg 1 Introduction Microscope is a tool

More information

Phoroptor VRx Digital Refraction System

Phoroptor VRx Digital Refraction System Drive practice efficiency and accelerate your exam process. Phoroptor VRx Digital Refraction System easily adapts to your practice, allowing you to rapidly perform all of the functions of a manual Phoroptor,

More information

Flat-Field IR Mega-Pixel Lens

Flat-Field IR Mega-Pixel Lens Flat-Field Mega-Pixel Lens Series Flat-Field Mega-Pixel Lens Flat-Field Mega-Pixel Lens 20.ver.02 E Specifications and Lineup Full MP Image Model Imager Size Mount Focal Length Aperture Range Zoom Ratio

More information

MRC High Resolution. MR-compatible digital HD video camera. User manual

MRC High Resolution. MR-compatible digital HD video camera. User manual MRC High Resolution MR-compatible digital HD video camera User manual page 1 of 12 Contents 1. Intended use...2 2. System components...3 3. Video camera and lens...4 4. Interface...4 5. Installation...5

More information

MT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL

MT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL Introduction With your purchase of an MT-30/MT-90 series type microscope you have chosen for a quality product. The MT-30/MT-90 series type microscopes are developed for use at schools and laboratories.

More information

Solution Derivations for Capa #14

Solution Derivations for Capa #14 Solution Derivations for Capa #4 ) An image of the moon is focused onto a screen using a converging lens of focal length (f = 34.8 cm). The diameter of the moon is 3.48 0 6 m, and its mean distance from

More information

Binocular Vision and The Perception of Depth

Binocular Vision and The Perception of Depth Binocular Vision and The Perception of Depth Visual Perception How one visually interprets a scene 4 forms of perception to be studied: Depth Color Temporal Motion Depth Perception How does one determine

More information

Bio 321 Lightmicroscopy Electronmicrosopy Image Processing

Bio 321 Lightmicroscopy Electronmicrosopy Image Processing Bio 321 Lightmicroscopy Electronmicrosopy Image Processing Urs Ziegler Center for Microscopy and Image Analysis Light microscopy (Confocal Laser Scanning Microscopy) Light microscopy (Confocal Laser Scanning

More information

Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998)

Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998) Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, 1849-1853, 1 April 1998) Geoff Andersen and R. J. Knize Laser and Optics Research Center

More information

Science In Action 8 Unit C - Light and Optical Systems. 1.1 The Challenge of light

Science In Action 8 Unit C - Light and Optical Systems. 1.1 The Challenge of light 1.1 The Challenge of light 1. Pythagoras' thoughts about light were proven wrong because it was impossible to see A. the light beams B. dark objects C. in the dark D. shiny objects 2. Sir Isaac Newton

More information

3D Printing LESSON PLAN PHYSICS 8,11: OPTICS

3D Printing LESSON PLAN PHYSICS 8,11: OPTICS INVESTIGATE RATIONALE Optics is commonly taught through the use of commercial optics kits that usually include a basic set of 2-4 geometric lenses (such as double convex or double concave). These lenses

More information

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY 3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY Prepared by Duanjie Li, PhD & Andrea Novitsky 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard

More information

FIFTH GRADE TECHNOLOGY

FIFTH GRADE TECHNOLOGY FIFTH GRADE TECHNOLOGY 3 WEEKS LESSON PLANS AND ACTIVITIES SCIENCE AND MATH OVERVIEW OF FIFTH GRADE SCIENCE AND MATH WEEK 1. PRE: Interpreting data from a graph. LAB: Estimating data and comparing results

More information

7.2. Focusing devices: Unit 7.2. context. Lenses and curved mirrors. Lenses. The language of optics

7.2. Focusing devices: Unit 7.2. context. Lenses and curved mirrors. Lenses. The language of optics context 7.2 Unit 7.2 ocusing devices: Lenses and curved mirrors Light rays often need to be controlled and ed to produce s in optical instruments such as microscopes, cameras and binoculars, and to change

More information

How an electronic shutter works in a CMOS camera. First, let s review how shutters work in film cameras.

How an electronic shutter works in a CMOS camera. First, let s review how shutters work in film cameras. How an electronic shutter works in a CMOS camera I have been asked many times how an electronic shutter works in a CMOS camera and how it affects the camera s performance. Here s a description of the way

More information

Firearms & Tool Marks Comparison Microscope. Discovery. Leeds Forensic Systems. Your Forensic Imaging Source

Firearms & Tool Marks Comparison Microscope. Discovery. Leeds Forensic Systems. Your Forensic Imaging Source Firearms & Tool Marks Comparison Microscope Leeds Forensic Systems Your Forensic Imaging Source Leeds Discovery The Leeds Discovery Firearms & Tool Marks Comparison Microscope is an innovative microscope

More information

Master Anamorphic T1.9/35 mm

Master Anamorphic T1.9/35 mm T1.9/35 mm backgrounds and a smooth, cinematic look, the 35 Close Focus (2) 0.75 m / 2 6 Magnification Ratio (3) H: 1:32.3, V: 1:16.1 Weight (kg) 2.6 Weight (lbs) 5.7 Entrance Pupil (7) (mm) -179 Entrance

More information

Sound absorption and acoustic surface impedance

Sound absorption and acoustic surface impedance Sound absorption and acoustic surface impedance CHRISTER HEED SD2165 Stockholm October 2008 Marcus Wallenberg Laboratoriet för Ljud- och Vibrationsforskning Sound absorption and acoustic surface impedance

More information

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope. MICROSCOPY I. OBJECTIVES To demonstrate skill in the proper utilization of a light microscope. To demonstrate skill in the use of ocular and stage micrometers for measurements of cell size. To recognize

More information

STARLINN. Novacon Pioneers a New Telescope Profile. Here we have an absolutely new product all over the world! A new way of seeing!

STARLINN. Novacon Pioneers a New Telescope Profile. Here we have an absolutely new product all over the world! A new way of seeing! STARLINN Novacon Pioneers a New Telescope Profile Here we have an absolutely new product all over the world! A new way of seeing! The first radically new design approach... since the telescope invention

More information

Contents. Introduction 2. Features and Benefits 3. Freeform Design Range 10. Custom 14/16/18 13. Custom U 15. Custom V 17.

Contents. Introduction 2. Features and Benefits 3. Freeform Design Range 10. Custom 14/16/18 13. Custom U 15. Custom V 17. Contents page Introduction 2 Features and Benefits 3 Freeform Design Range Custom 14/16/18 13 Custom U 1 Custom V 17 Custom Office 19 Custom Driver Custom Relax 21 Custom Junior 22 Custom Blended Bifocal

More information

ALMA Newsletter. ALMA In-depth. How Will ALMA Make Images? April 2010

ALMA Newsletter. ALMA In-depth. How Will ALMA Make Images? April 2010 How Will ALMA Make Images? Credit: ALMA (ESO / NAOJ / NRAO), Cynthia Collao (ALMA). The invention of the optical telescope by Galileo 400 years ago marked the beginning of modern astronomy. Galileo used

More information

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and

More information

Rec. ITU-R F.699-5 1 RECOMMENDATION ITU-R F.699-5 *

Rec. ITU-R F.699-5 1 RECOMMENDATION ITU-R F.699-5 * Rec. ITU-R F.699-5 1 RECOMMENATION ITU-R F.699-5 * REFERENCE RAIATION PATTERNS FOR LINE-OF-SIGHT RAIO-RELAY SYSTEM ANTENNAS FOR USE IN COORINATION STUIES AN INTERFERENCE ASSESSMENT IN THE FREQUENCY RANGE

More information

Spatial location in 360 of reference points over an object by using stereo vision

Spatial location in 360 of reference points over an object by using stereo vision EDUCATION Revista Mexicana de Física E 59 (2013) 23 27 JANUARY JUNE 2013 Spatial location in 360 of reference points over an object by using stereo vision V. H. Flores a, A. Martínez a, J. A. Rayas a,

More information

The Visual Cortex 0 http://www.tutis.ca/neuromd/index.htm 20 February 2013

The Visual Cortex 0 http://www.tutis.ca/neuromd/index.htm 20 February 2013 T he Visual Cortex 0 Chapter contents Contents Chapter 2... 0 T he Visual Cortex... 0 Chapter Contents... 1 Introduction... 2 Optic Chiasm... 2 Where do the eye's ganglion cells project to?... 3 To where

More information

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant INTRODUCTION Except for unusual situations, a cataract operation is indicated only when you cannot function satisfactorily

More information

National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104

National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODELS 156, 156-S, 157 COMPOUND BIOLOGICAL MICROSCOPES

More information

Protocol for Microscope Calibration

Protocol for Microscope Calibration Protocol for Microscope Calibration A properly calibrated system is essential for successful and efficient software use. The following are step by step instructions on how to calibrate the hardware using

More information

RAY OPTICS II 7.1 INTRODUCTION

RAY OPTICS II 7.1 INTRODUCTION 7 RAY OPTICS II 7.1 INTRODUCTION This chapter presents a discussion of more complicated issues in ray optics that builds on and extends the ideas presented in the last chapter (which you must read first!)

More information

National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104

National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODELS 106, 106-L 107, 107-L 108, 108-L 109-L ELEMENTARY

More information

picture real beauty Capture real beauty with the one you love

picture real beauty Capture real beauty with the one you love picture real beauty Capture real beauty with the one you love Lina Dove invites you to experience photography for yourself and capture images of beauty that live around you. To help, we ve taken the principles

More information

How To Use 3D On A Computer Or Tv

How To Use 3D On A Computer Or Tv Display technologies seminar 3D Displays What is 3D? 3D History 3D production 3D Displays 3D now Agenda 13 May 2010 Metropolia University of Applied Sciences Lauri Virkki What is 3D? Eyes are average 64

More information

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

WAVELENGTH OF LIGHT - DIFFRACTION GRATING PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant

More information

Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications

Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications Screw thread image generated by WLI Steep PSS angles WLI color imaging Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications 3D optical microscopy is a mainstay metrology

More information

OPERATING INSTRUCTIONS XJL 101/101A REFLECTED LIGHT METALLURGICAL MICROSCOPE

OPERATING INSTRUCTIONS XJL 101/101A REFLECTED LIGHT METALLURGICAL MICROSCOPE OPERATING INSTRUCTIONS XJL 0/0A REFLECTED LIGHT METALLURGICAL MICROSCOPE M.R.C.LTD. OFFICES: HAHYSTADRUT 84, HOLON 58394 P.O.B. 684, TEL-AVIV 606, ISRAEL TEL: 97-3-559305,97-3-55955 FAX: 97-3-559459 www.mrclab.com

More information

Hunting Ghosts. For the development of imaging optical STRAY LIGHT ANALYSIS IN IMAGING OPTICS

Hunting Ghosts. For the development of imaging optical STRAY LIGHT ANALYSIS IN IMAGING OPTICS Virtual prototype of the camera lens defined in [3]. Besides the lenses we model only those mechanical parts that potentially contribute the most to stray light Hunting Ghosts STRAY LIGHT ANALYSIS IN IMAGING

More information