Acceptability requirements for x-ray equipment used in health care

Size: px
Start display at page:

Download "Acceptability requirements for x-ray equipment used in health care"

Transcription

1 DECISION 17 August /310/06 Translation. Original text in Finnish. ACCEPTABILITY REQUIREMENTS FOR X-RAY EQUIPMENT USED IN HEALTH CARE Radiography and fluoroscopy equipment and computed tomography appliances Under section 30 of the Decree of the Ministry of Social Affairs and Health on the Medical Use of Radiation (no. 423 of 2000) measures causing exposure to radiation must be performed using equipment that is suited for the said purpose. The Decree also prescribes that the requirements and criteria of acceptability for specific equipment functions to be considered from the point of view of radiation safety are to be confirmed by the Radiation and Nuclear Safety Authority (STUK). This decision confirms the attached acceptability criteria, i.e. the acceptability requirements for radiography and fluoroscopy equipment and computed tomography appliances. These acceptability requirements do not apply to mammography equipment, dental x-ray imaging equipment or digital imaging sensors. This decision shall be in force as of 1 October Director General Jukka Laaksonen Director Eero Kettunen APPENDIX Acceptability requirements for x-ray equipment used in health care STUK RADIATION AND NUCLEAR SAFETY AUTHORITY Visiting address Laippatie 4, FI Helsinki, Finland Postal address PO Box 14, FI Helsinki, Finland Tel Fax

2 Appendix 1 (9) 17 August 2006 Translation. Original text in Finnish. ACCEPTABILITY REQUIREMENTS FOR X-RAY EQUIPMENT USED IN HEALTH CARE Radiography and fluoroscopy equipment and computed tomography appliances 1 Background X-ray tubes and generators Filtration of primary radiation X-ray tube voltage X-ray tube current Exposure Time Electrical charge X-ray tube radiation output Radiation beam indicators and focusing X-ray equipment operation Dose display Imaging sensors based on a film-intensifier screen combination Image quality Film cassette Film processing Automatic exposure control unit function Fluoroscopy equipment Image quality Dose rate automation in fluoroscopy Relation between radiation field and image field sizes Computed tomography equipment Image quality CT numbers Slice thickness Table movement and patient positioning... 8 Bibliography... 9 STUK RADIATION AND NUCLEAR SAFETY AUTHORITY Visiting address Laippatie 4, FI Helsinki, Finland Postal address PO Box 14, FI Helsinki, Finland Tel Fax

3 2 1. Background Health care X-ray equipment and the accessories and instruments that are associated with its use must satisfy the acceptability requirements that are specified in this decision (see also section 30 of Decree no. 423 of 2000 of the Ministry of Social Affairs and Health [9] and Guide ST 3.3 [8]). These acceptability requirements denote the minimum requirements or acceptability limits that are imposed on the performance capacity of the equipment when the equipment is used [1]. The equipment must either be repaired and its performance capacity restored to the acceptable standard or it must be decommissioned where necessary, no later than at the time when acceptability limit requirement is not met. Acceptability requirements typically concern the precision of equipment settings and the operating condition of the equipment. They are not limiting values for optimal equipment performance. Responsible parties (the party running a radiation practice) procuring new equipment, performing acceptance tests and inspecting the quality of equipment at the time of use should apply stricter requirements, which may be based, for example, on the equipment specifications or performance tolerances shown in equipment standards. The performance measurement results will depend on the conditions of measurement and may also depend on the method of measurement. The methods of measurement and test samples shown in the bibliography [references 2 7] are used in measurements of compliance with acceptability requirements unless otherwise shown in this decision. 2. X-ray tubes and generators 2.1 Filtration of primary radiation Total filtration of primary radiation must correspond to a minimum of 2.5 mm of aluminium. The total radiation filtration must be marked on the protective housing of the x-ray tube. If the additional filtration is adjustable, then the fixed filtration must be marked on the protective housing of the x-ray tube and it must be possible to verify the selected additional filtration. 2.2 X-ray tube voltage The deviation in voltage from the preset value in an X-ray tube may not exceed 10 per cent. When adjusting the voltage from one value to another the change in actual voltage must also be not less than 0.5 times and not more than 1.5 times the difference between the preset voltages. 2.3 X-ray tube current The deviation in electric current from the preset value in an X-ray tube may not exceed 20 per cent. NB. Direct measurement of X-ray tube current is not always necessary in practice, and compliance with the requirement may instead be monitored on the basis of constancy and linearity of radiation output. It may also be necessary to verify the absolute accuracy of the X-ray tube current if the result of measuring radiation output differs from the radiation output reference value determined in the equipment acceptance test by more than the action threshold imposed in the quality control programme (see Guide ST 3.3 [8]), or if the dose *) as a function of electrical charge fails to satisfy the linearity requirement stipulated in section 2.6 of this decision. *) The dose referred to in this decision may be measured as air kerma and the dose rate may be measured as air kerma rate.

4 3 2.4 Exposure Time The deviation in exposure time from the preset value may not exceed 15 per cent + 1 ms. 2.5 Electrical charge The deviation in electrical charge, i.e. the product of the X-ray tube current and the exposure time, from the preset value may not exceed 20 per cent mas. NB. Direct measurement of X-ray tube electrical charge is not always necessary in practice, and compliance with the requirement may instead be monitored on the basis of constancy and linearity of radiation output. It may also be necessary to verify the absolute accuracy of the X-ray tube electrical charge if the result of measuring radiation output differs from the radiation output reference value determined in the equipment acceptance test by more than the action threshold imposed in the quality control programme (see Guide ST 3.3 [8]), or if the dose*) as a function of electrical charge fails to satisfy the linearity requirement stipulated in section 2.6 of this decision. 2.6 X-ray tube radiation output When manual adjustment values, which correspond to the clinical use of equipment, are used, the standard deviation of dosage *) measured in the radiation beam may not exceed 10 per cent. The standard deviation will be calculated from at least five repeated measurements. If a film-screen combination and suitable phantom are used as a detector when measuring, then the deviation in x-ray image densities from the average must not exceed When using manual adjustment values the dose *) measured in the radiation beam of an X-ray appliance must be proportional to the electrical charge applied, such that K Q K Q K Q K + Q where K 1 is the dose corresponding to the electrical charge Q 1, K 2 is the dose corresponding to the electrical charge Q 2 and Q 2 < 2 Q Radiation beam indicators and focusing The radiation beam signal lights must be clearly visible under normal working illumination conditions. The deviation between the signal lights or other indicators of the radiation beam and the edges of the radiation beam at the image receptor must not exceed 1 per cent of the distance between the focus of the X-ray tube and the image receptor on any edge of the radiation field. The deviation of the central axis of the radiation beam from the centre point of the radiation beam at the image receptor as shown by the signal lights describing the radiation beam or by some other radiation beam indicator, and from the centre point of the image receptor, may not exceed 1 per cent of the distance between the focus and the image receptor. When using automatic adjustment of the radiation field or image partition automation the radiation beam must be directed onto the image receptor in a manner that is expedient and intended by the appliance manufacturer. *) The dose referred to in this decision may be measured as air kerma and the dose rate may be measured as air kerma rate.

5 4 2.8 X-ray equipment operation 3. Dose display The X-ray equipment, its accessories and its premises of use must be conducive to safe operation of the equipment. The appliance warning and indicator lights and other safety device and accessories must be undamaged and operating in the intended manner. The error of any meter or display indicating the radiation exposure of a patient from X-ray equipment may not exceed 25 per cent (50 per cent in fluoroscopy appliances). 4. Imaging sensors based on a film-screen combination 4.1 Image quality 4.2 Film cassette Image quality must satisfy the clinical requirements for X-ray examinations made with the appliance. This may be assessed in the manner specified in Guide ST 3.3 [8]. NB. No detailed technical acceptability requirements for image quality are specified in this decision, as compliance with technical requirements constitutes no direct demonstration of the clinical acceptability of an image, and because the required clinical image quality essentially depends on the medical indications that justify the X-ray examination. Control of the operating condition and performance characteristics of equipment employs a phantom suited to technical assessment of image quality and appropriately selected performance levels (e.g. resolution and contrast sensitivity) (see Guide ST 3.3 [8]). The locking mechanism of the film cassette must function reliably and the cassette must be lightproof. The compression material must be resilient and compression must be effective and even throughout the film area. Contact between the intensifying screen and film must be good. Cassettes must be classified by film speed and the speed must be clearly marked on them. The difference in film density with intensifying screens in the same speed class and images made in the same way may not exceed Film processing The darkroom and storage box for X-ray film must be lightproof. The safelight must be suitable for the type of film used, and indicator and other lights must not expose the film. Ordinary film processing in the darkroom must cause no visible change (about 0.05) in the density of pre-exposed film. The optical density of unexposed developed X-ray film must not exceed 0.3. Deviations in speed index and contrast index (figure 1) from the reference value (the value obtained in the acceptance test) in film development monitoring must not exceed 0.2. The film processor must not cause damage to X-ray films. Film processors used for developing similar films should yield similar film processing results. If this does not occur, then the film processors are to be adjusted so that at least the film speed index is the same.

6 5 3,2 2,8 2,4 Film density 2 1,6 1,2 0,8 0, Density step number Figure 1. Determination of film processing reference values from a density curve derived by exposing a film by a sensitometer. The film fog is defined at step 1 (0.2 in the graph). For a speed index reference value a step is sought where the density is closest to a value of 1.0 above the base line fog (the value shown on the graph is 1.2). The density at this step (the density value (1.1) at step 12 on the graph) expresses the value of the speed index. The contrast index is determined using two density steps, selecting the steps for which the densities are closest to values of 1.0 and 2.0 above the base line fog (steps 12 and 15 (1.1 and 2.3) on the graph). The difference in densities between these steps is the contrast index value of the film (1.2 in the situation shown on the graph). The values of the indices will always be determined in monitoring from the densities at these same density steps (12 and 15 in this example): the density at step 12 is the speed index and the difference between the densities at steps 12 and 15 is the contrast index. 4.4 Automatic exposure control unit function When imaging a phantom that optimally corresponds to real patient imaging using an automatic exposure control unit, at the various X-ray tube voltage values used in such X-ray examinations, the deviation in X-ray film densities from the average may not exceed 0.3 (checking operation of automatic exposure control unit voltage compensation). The thickness of the phantom must correspond to the thickness of body parts to be imaged in these X-ray examinations. 5. Fluoroscopy equipment The following requirements apply to fluoroscopy in addition to the requirements (in chapters 2 and 3) for the X-ray tube and generator and for the dose display. 5.1 Image quality Image quality must satisfy the clinical requirements for fluoroscopy examinations and fluoroscopically guided procedures made with the appliance. This may be assessed in the manner specified in Guide ST 3.3 [8]. NB. No detailed technical acceptability requirements for image quality are specified in this decision, as compliance with technical requirements constitutes no direct demonstration of the clinical acceptability of an image, and because the required clinical image quality essentially depends on the medical indications that justify the examination or procedure. Control of the operating condition and performance characteristics of equipment employs a phantom suited to technical assessment of image quality and appropriately selected performance levels (e.g. resolution and contrast) (see Guide ST 3.3 [8]).

7 6 5.2 Automatic dose rate control in fluoroscopy The level of dose rate, i.e. the air kerma rate regulated by the dose rate control at the surface of the image receptor entrance plate without a grid, may not exceed 0.8 µgy/s when using the most common image receptor field size (e.g. an image diameter of about 25 cm) in fluoroscopic examinations. This limiting value is a recommendation when using other image receptor field sizes. The level of dose rate must be checked, for example using the measurement system shown in figure 2. The primary radiation dose rate (air kerma rate) in normal operation under the measurement system shown in figure 3 may not exceed 100 mgy/min. If the examination or procedure requires a dose rate that is higher than this (in which case the level of the dose rate may also exceed 0.8 µgy/s), then the operation may be accepted subject to the following conditions: The dose rate to the patient s skin in the primary beam of the X-ray tube does not exceed 50 mgy/min. in normal operation. The equipment includes a switch, which the operator must activate continually in order to use the enhanced function in question. The dose rate according to the measuring system shown in figure 3 does not exceed 200 mgy/min. A continuous sound or light signal indicates use of the enhanced function to the person performing the procedure. There is an extra shield for the operator in the vicinity of the equipment corresponding in effectiveness to a layer of no less than 1.0 mm of lead. The equipment includes a DAP display (dose-area product display) or other suitable dose display. Figure 2. Arrangements for measuring the level of dose rate in fluoroscopy equipment. 1 focus of X-ray tube, 2 test phantom (20 mm of aluminium or other test phantom as necessary, e.g. 10 mm of aluminium for paediatric fluoroscopy), 3 table surface (where present), 4 cassette partition automation (where present), 5 radiation meter of an ionization chamber, 6 entrance plate, 7 front surface of image receptor (e.g. image intensifier).

8 7 a) Over coutch tube Focus Measurement chamber Minimum distance 30 cm Table surface b) Under coutch tube Measurement chamber Table surface Minimum distance Focus c) C or U arm Front surface of image receptor entrance plate 30 cm Measurement chamber Focus Figure 3. Primary radiation dose rate measurement geometry of fluoroscopy equipment.

9 8 5.3 Relation between radiation field and image field sizes The ratio of sizes of the radiation field and the image field (surface area of the image receptor front panel) must not exceed It is considered good practice for the edges of the radiation beam limiter to show on the monitor image. 6. Computed tomography equipment The following requirements apply to computed tomography equipment in addition to the requirements (in chapters 2 and 3) for the X-ray tube and generator and for the dose display. 6.1 Image quality 6.2 CT numbers Image quality must satisfy the clinical requirements for computed tomography examinations made with the appliance. This may be assessed in the manner specified in Guide ST 3.3 [8]. NB. No detailed technical acceptability requirements for image quality are specified in this decision, as compliance with technical requirements constitutes no direct demonstration of the clinical acceptability of an image, and because the required clinical image quality essentially depends on the medical indications that justify the examination or procedure. Control of the operating condition and performance characteristics of equipment employs a phantom suited to technical assessment of image quality and appropriately selected performance levels (e.g. resolution and contrast) (see Guide ST 3.3 [8]). The CT numbers at the reference point of the image field must not deviate from the manufacturer s notified values by more than 10 HU when measured in the manner specified by the manufacturer. 6.3 Slice thickness The true slice thickness (detector sensitivity profile half-width value) in axial imaging must not deviate from the preset value by more than 1 mm when the preset slice thickness is greater than 2 mm 50 per cent, when the preset slice thickness is 1-2 mm 0.5 mm when the preset slice thickness is less than 1 mm. 6.4 Table movement and patient positioning When the table moves a distance of 20 cm the true movement of the table must not deviate by more than 2 mm from the value shown on the table movement display. The indicated starting point of CT imaging may not deviate from the true starting point by more than 3 mm.

10 9 Bibliography [1] Criteria for acceptability of radiological (including radiotherapy) and nuclear medicine installations. Radiation Protection 91, European Commission [2] EN (IEC) Medical electrical equipment Part 1: General requirements for safety. 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment. [3] EN (IEC) Medical electrical equipment Part 2: Particular requirements for the safety of highvoltage generators of diagnostic X-ray generator. [4] IEC Medical electrical equipment Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis. [5] IEC Medical electrical equipment Part 2: Particular requirements for the safety of X-ray equipment for interventional procedures. [6] IEC Medical electrical equipment Part 2: Particular requirements for the safety of X-ray equipment for computed tomography. [7] IEC Medical electrical equipment Part 2: Particular requirements for the safety of X-ray equipment for computed tomography, Amendment 1. [8] Röntgentutkimukset terveydenhuollossa, Ohje ST 3.3, STUK (X-ray examinations in health care. Guide ST 3.3. Radiation and Nuclear Safety Authority, 20 March 2006.) [9] Sosiaali- ja terveysministeriön asetus säteilyn lääketieteellisestä käytöstä, 423/2000. (Decree of the Ministry of Social Affairs and Health on the medical use of radiation, no. 423 of 2000.)

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)

More information

USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY

USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY GUIDE ST 5.3 / 4 MAY 2007 USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY 1 Ge n e r a l 3 2 Safety licences and exemption from licensing 3 3 Safety licence-exempt use of radiation 3

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)

More information

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

RADIATION SAFETY OF LASER DISPLAYS AND SHOWS

RADIATION SAFETY OF LASER DISPLAYS AND SHOWS GUIDE ST 9.4 / 30 April 2015 RADIATION SAFETY OF LASER DISPLAYS AND SHOWS 1 General 3 2 The responsible party is liable for the safety of the laser show 3 3 Equipment safety 4 3.1 Laser equipment must

More information

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information

More information

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recommendations for a technical quality control program for diagnostic X-ray equipment Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

MALAYSIAN STANDARD CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) Copyright 2007

MALAYSIAN STANDARD CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) Copyright 2007 MALAYSIAN STANDARD MS 838:2007 CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) ICS: 13.280 Descriptors: code of practice, medical diagnosis, radiation protection, x-ray

More information

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and

More information

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC

More information

WARNING SIGNS FOR RADIATION SOURCES

WARNING SIGNS FOR RADIATION SOURCES WARNING SIGNS FOR RADIATION SOURCES 1 General 3 2 All radiation sources must be marked 3 3 Warning signs for ionising radiation 3 4 Warning signs for non-ionising radiation 4 4.1 Magnetic fields 4 4.2

More information

SECURITY ARRANGEMENTS OF RADIATION SOURCES

SECURITY ARRANGEMENTS OF RADIATION SOURCES SECURITY ARRANGEMENTS OF RADIATION SOURCES 1 General 3 2 The party running the radiation practice shall be responsible for security arrangements 3 3 Places of use and storages of radiation sources shall

More information

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding

More information

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General

More information

Patient Exposure Doses During Diagnostic Radiography

Patient Exposure Doses During Diagnostic Radiography Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University

More information

RADIATION SAFETY STANDARD

RADIATION SAFETY STANDARD Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession

More information

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 CHAPTER 227. RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT, X-RAY GAUGING EQUIPMENT, ELECTRON MICROSCOPES AND X-RAY CALIBRATION SYSTEMS Sec. 227.1.

More information

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Measuring Size and Distance in Digital Mammography

Measuring Size and Distance in Digital Mammography On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

RADIATION PROTECTION AND EXPOSURE MONITORING OF NUCLEAR FACILITY WORKERS

RADIATION PROTECTION AND EXPOSURE MONITORING OF NUCLEAR FACILITY WORKERS RADIATION PROTECTION AND EXPOSURE MONITORING OF NUCLEAR FACILITY WORKERS 1 Introduction 5 2 Scope of application 5 3 Occupational radiation protection 6 4 Operation of the radiation protection organisation

More information

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT

More information

X-Mind. Instinct for perfection

X-Mind. Instinct for perfection X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in

More information

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Radiation Safety Characteristics of the NOMAD Portable X-ray System Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality

More information

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come

More information

X-Rays Benefits and Risks. Techniques that use x-rays

X-Rays Benefits and Risks. Techniques that use x-rays X-Rays Benefits and Risks X-rays are a form of electromagnetic radiation, just like light waves and radiowaves. Because X-rays have higher energy than light waves, they can pass through the body. X-rays

More information

DENTAL Cone beam 3D X-RAY SYSTEM with

DENTAL Cone beam 3D X-RAY SYSTEM with VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic

More information

NUCLEAR POWER PLANT PRESSURE EQUIPMENT

NUCLEAR POWER PLANT PRESSURE EQUIPMENT GUIDE YVL 3.8 / 22 SEPTEMBER 2003 NUCLEAR POWER PLANT PRESSURE EQUIPMENT In-service inspection with non-destructive testing methods 1 GENERAL 5 2 DEFINITIONS 5 3 REQUIREMENTS 7 4 IN-SERVICE INSPECTION

More information

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

Required RS Training Info

Required RS Training Info C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness

More information

Checklist for submission of application form for Registration

Checklist for submission of application form for Registration Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status

More information

Staff Doses & Practical Radiation Protection in DEXA

Staff Doses & Practical Radiation Protection in DEXA Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

LASER-PROTECTION INFORMATION

LASER-PROTECTION INFORMATION LASER-PROTECTION INFORMATION µspeed-systems Version V2.0 Creation date: 18.07.2012 page 1/8 LASER PROTECTION The µspeed sensor system is a laser-based measuring system. When using a µspeed, it is necessary

More information

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Presented: August 3, 2011 Daniel Kassiday United States Food and Drug Administration Center for Devices and Radiological

More information

KS-70. Catheterization Table GENERAL

KS-70. Catheterization Table GENERAL PD505-009A Catheterization Table KS-70 GENERAL This Catheterization Table KS-70, which is combined with X-ray high voltage generator, X-ray tube assembly, X-ray image system (Image Intensifier, Flat Panel

More information

The Nuclear Safety Council, in exercise of the functions conferred to

The Nuclear Safety Council, in exercise of the functions conferred to Nuclear Safety Council (BOE no. 43, of 19/02/2008) The Nuclear Safety Council s INSTRUCTION IS-17, of 30 th January, on the recognition of training courses and programmes for personnel that manage the

More information

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact

More information

Understanding Digital Modalities: System Integration and Use

Understanding Digital Modalities: System Integration and Use Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information

More information

Tube Control Measurement, Sorting Modular System for Glass Tube

Tube Control Measurement, Sorting Modular System for Glass Tube Tube Control Measurement, Sorting Modular System for Glass Tube Tube Control is a modular designed system of settled instruments and modules. It comprises measuring instruments for the tube dimensions,

More information

Diagnostic Exposure Tracking in the Medical Record

Diagnostic Exposure Tracking in the Medical Record Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand

More information

3rd International Symposium of the System of Radiological Protection Seoul, Korea, 20-22 October 2015

3rd International Symposium of the System of Radiological Protection Seoul, Korea, 20-22 October 2015 3rd International Symposium of the System of Radiological Protection Seoul, Korea, 20-22 October 2015 Dr. Pedro Ortiz López Member of ICRP Committee 3 on Radiological Protection in Medicine 2 1895 Discovery

More information

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION

CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination

More information

Safety and Radiation Protection Office Working with X-Ray Equipment

Safety and Radiation Protection Office Working with X-Ray Equipment Safety and Radiation Protection Office Working with X-Ray Equipment Radiation Safety Rules May 2016 Radiation Safety Rules X-Ray Equipment General Rules for the protection of persons exposed to ionising

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

COLLEGE ON MEDICAL PHYSICS

COLLEGE ON MEDICAL PHYSICS COLLEGE ON MEDICAL PHYSICS (30 August - 22 September 2004) Miramare - Trieste, Italy Directors: A. Benini, P. Sprawls and S. Tabakov Local Organizer: L. Bertocchi FINAL PROGRAMME (all lectures were held

More information

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number

More information

Application Note. ipix A Gamma imager to support various applications

Application Note. ipix A Gamma imager to support various applications Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose

More information

ELECTRICAL AND I&C EQUIPMENT OF A NUCLEAR FACILITY

ELECTRICAL AND I&C EQUIPMENT OF A NUCLEAR FACILITY ELECTRICAL AND I&C EQUIPMENT OF A NUCLEAR FACILITY 1 Introduction 5 2 Scope of application 6 3 Requirement specification, selection, and procurement of electrical and I&C equipment and cables 7 3.1 General

More information

APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS

APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS SAFETY REPORTS SERIES No. 39 APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

X-RAY REGULATORY GUIDE

X-RAY REGULATORY GUIDE Minnesota Department of Health Radiation Control, X-ray Unit Protecting, maintaining and improving the health of all Minnesotans by promoting radiation safety through guidance and collaboration with the

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

RADIATION PROTECTION PROGRAM (RPP) GUIDANCE

RADIATION PROTECTION PROGRAM (RPP) GUIDANCE RADIATION PROTECTION PROGRAM (RPP) GUIDANCE The Colorado Rules and Regulations Pertaining to Radiation Control, (6 CCR 1007-1) are the regulations developed by the Colorado Department of Public Health

More information

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY

ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY ESTIMATING POPULATION DOSES FROM MEDICAL RADIOLOGY B. Wall, (Co-ordinator) 1, D. Hart 1, H. Mol 2, A. Lecluyse 2, A. Aroua 3, P. Trueb 3, J. Griebel 4, E. Nekolla 4, P. Gron 5, H. Waltenburg 5, H. Beauvais-March

More information

814.49. Ordinance on the Protection of Audiences from Exposure to Hazardous Sound Levels and Laser Beams. (Sound Levels and Laser Ordinance, SLO)

814.49. Ordinance on the Protection of Audiences from Exposure to Hazardous Sound Levels and Laser Beams. (Sound Levels and Laser Ordinance, SLO) English is not an official language of the Swiss Confederation. This translation is provided for information purposes only and has no legal force. Ordinance on the Protection of Audiences from Exposure

More information

The Exposure Index and Its Standardization

The Exposure Index and Its Standardization The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good

More information

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition) Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition) New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp

More information

7.1 General 5 7.2 Events resulting in pressure increase 5

7.1 General 5 7.2 Events resulting in pressure increase 5 GUIDE YVL 2.4 / 24 Ma r ch 2006 Primary and secondary circuit pressure control at a nuclear power plant 1 Ge n e r a l 3 2 General design requirements 3 3 Pressure regulation 4 4 Overpressure protection

More information

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes. Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists

More information

Personal Dosimetry in Dental Radiology

Personal Dosimetry in Dental Radiology Personal Dosimetry in Dental Radiology Position Statement Personal Dosimetry in Dental Radiology Radiological Protection Institute of Ireland May 2011 3 Clonskeagh Square, Dublin 14 Background The RPII

More information

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA N. Zarrinpanjeh a, F. Dadrassjavan b, H. Fattahi c * a Islamic Azad University of Qazvin - nzarrin@qiau.ac.ir

More information

INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer

INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer X-rays were discovered by W.C. Roentgen because of their ability to cause fluorescence. X-ray photons cannot be seen. The image produced

More information

SOURCES AND EFFECTS OF IONIZING RADIATION

SOURCES AND EFFECTS OF IONIZING RADIATION SOURCES AND EFFECTS OF IONIZING RADIATION United Nations Scientific Committee on the Effects of Atomic Radiation UNSCEAR 2008 Report to the General Assembly with Scientific Annexes VOLUME I UNITED NATIONS

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

portable x-ray survey report

portable x-ray survey report department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2

More information

TOF FUNDAMENTALS TUTORIAL

TOF FUNDAMENTALS TUTORIAL TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This

More information

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581 X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers

More information

L-LAS-TB-CL serie. laser light curtains for inline measuring tasks

L-LAS-TB-CL serie. laser light curtains for inline measuring tasks L-LAS-TB-CL serie laser light curtains for inline measuring tasks Maximum distance 2.000 mm Measurement range 6 up to 98 mm Resolution up to 4 µm up to 1,5 khz scan rate 0-10 V or 4-20 ma analogue output

More information

Checklist for the Qualification of Digital Detector Array Systems

Checklist for the Qualification of Digital Detector Array Systems Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction

More information

Quality Assurance of Radiotherapy Equipment

Quality Assurance of Radiotherapy Equipment Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria

More information

Prepublication Requirements

Prepublication Requirements Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals

More information

ABLOY DOOR AUTOMATICS Easy access for all types of swing doors

ABLOY DOOR AUTOMATICS Easy access for all types of swing doors ABLOY DOOR AUTOMATICS Easy access for all types of swing doors Automatic door operators bring additional convenience in situations where the doors opening manually would be inconvenient or laborious. Recommended

More information

Guide to Gate Safety Legislation & the Machinery Directive. In Association with:

Guide to Gate Safety Legislation & the Machinery Directive. In Association with: Guide to Gate Safety Legislation & the Machinery Directive In Association with: Tel: 0191 267 4922 info@primeautomation.co.uk Tel: 0191 267 4922 Email: info@primeautomation.co.uki Guide to Gate Safety

More information

7üVNORD. TÜV NORD Röntgentechnik

7üVNORD. TÜV NORD Röntgentechnik .. TÜV NORD Röntgentechnik Niedersachsen Nordrhein-Westfalen Hamburg Schleswig-Holstein Baden-Württemberg Mecklenburg-Vorpommern Bremen Sachsen-Anhalt Berlin Brandenburg Sachsen' Thüringen' Bayern' Hessen

More information