CCT-3300 series Conductivity (TDS) Controller
|
|
|
- Kathryn Caldwell
- 9 years ago
- Views:
Transcription
1 CCT-3300 series Conductivity (TDS) Controller Operation Manual
2 Notes before Operation 1. Notice before operation: (1) Carefully read the relevant parts of this manual before installation and operation to prevent wrong operation, measurement error and damage of instrument. (2) Improper installation and unsuitable flow speed will cause big measurement error, so please read the installation passage in detail. (3) This instrument is precise electrochemical measurement, and its installation and operation should be performed by technicians with relevant professional knowledge. 2. Maintenance items: (1) The meter s quality guarantee is one year from the date of purchasing. During this period, if the meter has quality problems, manufacturer is responsible for maintenance work for free or changes it. (2) Manufacturer offers the maintenance service for whole life of the sold meters. (3) If the damage of the meter is caused by the following reasons, it is out of the maintenance service: A The meter is burned or foundered caused by improper usage and maintenance. B The meter is refitted or misused without permit. C The meter is destroyed under the condition out of company s regulation. D The relevant damage caused by choosing the wrong type. E The cable damage and rupture caused by improper installation and usage. F The incorrect measurement of the sensor caused by disconnecting or connecting wires personally. G The inner broken wire caused by indiscreetly disassembling.
3 Catalogue I. Product General Introduction Main technical index Applications Products classification Notices of product installation Main technical index...3 II. Outline Dimension and Rear Terminals Outline dimension Rear terminals Wire connection...5 III. Electrical Connection ma wire connection diagram Electrical connection High limit and window control....7 IV. Introduction of Front Panel and Key Functions Front panel and main menu display Key board functions Introduction...9 V. Introduction of Operation Menu...9 VI. Introduction for Electrode Installation Selection for electrode constant Outline dimension of the electrode Installation method Installation and maintenance...12 VII. Failure Judgment for Meter and Electrode...13 VIII. Complete Set of Products...14 IX. Ordering Directory...14
4 I. Product General Introduction CCT-3300 series conductivity (TDS) measurement controller is a high-technology innovative product. Product performance and function updated to a new level. It is a product with high performance and price ratio for inline water quality conductivity analysis. 1.1 Main Technical Index No adjustable component inside the instrument, totally programmed automatic calibration, higher reliability; Automatic measurement range shift, digital temperature compensation. Support conductivity cell constant (0.1 cm -1, 1.0 cm -1, 5.0 cm -1, 10.0 cm -1 ), measurement range compatible with liquid from pure water to concentrated solution, named online conductivity superman. Conductivity (us/cm) or TDS (ppm) engineering unit display can be chosen in operation. present water temperature ( ) / transmitting current (4~20mA) can be check onsite. Measurement and calibration program for calculation and measurement inspection organization. Isolated, transferable 4~20mA current output, two points transmitting range random setting within full scope. Support all the two wire PLC system under transmitter model. Conductivity high limit alarm, high limit value and backlash value random setting within full measurement scope. Patent short housing and fast installation method to save more space for the project, High illumination, pure white backlight LCD display in accordance with international trend. Industrial standard wire contact, EMC electromagnetic compatible design, platform design development, serial product classification. Meeting international electronic power supply regulation, AC,DC current input which with non polarity wire connection, Our products with strict quality management, SMT (Surface Mounted Technology), AOI (automatic optical inspection (system), ICT Complete manufacturing technique management, and quality management. Electrode will be calibrated, tested, packed one by one. 1
5 1.2 Applications This series is widely used for the water treatment control and monitoring of water treatment (high salt content), concentrated solution, recirculating cooling water and etc. 1.3 Products classification Model Power Constant Frequency(Hz) Current loop modes supply selection CCT-3300E DC 24V Instrument/Transmitter CCT-3310E AC 110V 50/60 Instrument/Transmitter CCT-3320 AC 220V 50/60 Instrument/Transmitter CCT-3320E AC 220V 50/60 Instrument/Transmitter CCT-3320T AC 220V 50/60 Instrument/Transmitter Note: The model with E is the EMC enhancement type. CCT-3320T(Conductivity Monitor) without any control outputs. Display Three digits and a half Three digits and a half Three digits and a half Three digits and a half Three digits and a half 1.4 Notices of product installation i. Please choose CCT-3300E (power supply 24V DC) if the onsite with damp environment or systems with 24V power supply. ii. If the onsite voltage is unstable or the site is far away from the power station, please use wide range AC/DC voltage-stabilizing type DC 24V power supply, please choose CCT-3300E. iii. If the measurement value unstable caused by using converter system, electronic device for water treatment, UV light electronic ballast device, power supply pollution, please choose CCT-3320E iv. Under the small power generation device which the power unstable, or far away from the power supply, please use wide range AC/DC voltage-stabilizing type DC 24V power supply, please choose CCT-3300E v. UV does harm to the LCD display screen, so please do not place the meter under sunshine. 2
6 1.5 Main technical feature Name of product CCT-3300 series of conductivity/tds Controller Display LCD display Auxiliary electrode Constant 0.10 cm cm cm cm -1 Electrode coefficient range Auxiliary electrode accuracy measurement range cm cm cm cm ~199.9µS/cm 1.0~1999µS/cm 0.05~9.99mS/cm 0.5~19.99mS/cm Measurement range conductivity TDS temperature 0.5µS/cm ~ 19.99mS/cm 0.25ppm~ 9.99 ppt 0.0~50.0 Resolution Accuracy conductivity 0.01µS/cm TDS 0.01ppm temperature 0.1 conductivity 1.5level TDS 1.5level temperature ±0.8 Temperature compensation Working Environment NTC10K temperature component Temp.0~50 Relative Humidity: 80%RH Transmitting output(4~20ma) isolated/transferable/reversible /transmitter mode/meter mode Loop resistance Ω accuracy: ±0.1mA Control Output port double contact relay output (ON/OFF) output load capacity AC 220V/AC110V/DC24V 3A Max Power consumption <2.5W Power supply CCT-3300E CCT-3310E CCT-3320/CCT-3320E CCT-3320T DC 24V AC 110V AC 220V ±10% AC 220V ±10% ±4V ±10% Frequency(Hz) /60 50/60 Installation Panel mounted,fast installation clamp Dimension 48mm 96mm 80mm(H W D) Slot dimension 44mm 92mm(H W) Weight 0.27kg 3
7 II. Outline Dimension and Rear Terminals 2.1 Outline dimension 2.2 Rear terminals CCT-3300E CCT-3310E 4
8 CCT-3320/CCT-3320E/CCT-3320T Note:CCT-3320T RELAY terminal is blank, without any connection. 2.3 Wire connection W G/B Connect white wire of electrode(white) Connect green wire of electrode(green) Y Connect yellow wire of electrode (YELLOW) R I+\I- T+\T- RELAY 24V(A)/24V (B) Connect red wire of electrode(red) Under instrument mode, power supplied by meter Under transmitter mode, power supplied by conditioning module relay contact (ON/OFF contact) DC power supply 0V/ 110V Power supply AC 110V 0V/ 220V Power supply AC 220V Protective terminal of EMC ground connection(earth) NC Empty terminal(no wire connection in-side) 5
9 III. Electrical Connection 3.1 ma wire connection diagram Connection diagram of 4~20mA signal under different modes 6
10 3.2 Electrical connection Connection diagram of ON/OFF relay 3.3 High limit and window control. 7
11 The water quality is limited within HIGH limit If the meter is used for pharmaceutical industry, food, drink, water purification and etc, when the water quality is limited at HIGH limit, the solenoid valve will switch the water flow direction to make sure the water purity. When the water quality at LOW limit, the meter will back to normal operation. The space between HIGH and LOW is backlash. Backlash will avoid the solenoid valve shock. Controlled section between HIGH and LOW When the mete is used for circulating cooling water, cooling liquid, cleaning solution, when the water quality is so bad that at HIGH, the meter will discharge the water. When the water quality is better at LOW, discharge stopped. IV. Introduction of front panel and key functions 4.1 Front panel and main menu display 8
12 4.2 Key board functions Introduction Key sign Name Function select key Add key Enter key 1.parameter setting to select thousand,hundred,ten and unit in circulate 2.radix point position setting 3.Measurement switch to display conductivity/tds 1. Adjust the value under setting status. 2. Check the temperature/ma/ conductivity (TDS) reading under measurement status. 1.Enter parameter setting under main menu 2.Save the parameters and enter next menu V. Introduction of Operation Menu Under main measuring menu, pressing automatically for three seconds and enter setting menu order Setting Menu name Introduction for function 1 Electrode constant setting radix point setting Measurement unit setting 4mA transfer v setting 20mA transfer setting Alarm hi-limit setting The sign C= blink on display screen, operating selectkey and add key to input the needed electrodeconstant value, pressing enter key to save andenter next parameter setting. The radix point sign on display screen blink, press select key to set radix point position; press enter key to save and enter next parameter setting. The sign Unit blink on display screen, press add key to select measurement unit (ppm,ppt, µs/cm or ms/cm),press enter key to save and enter next parameter setting. The sign 4mA blink on display screen, press select key and add key to input the 4mA transferable value and press enter key to save and set the radix point ( the same procedures as electrode constant setting ),press enter key to save and enter next parameter setting. The sign 20mA blink,setting the data according to 20mA and press enter key to set the radix transferable value,press enter key enter the next parameter setting the sign Hi and blink, press select key and add key to input the needed hi-limit value; press enter key to save, and then set the radix point bits, press enter key to save and enter the next parameter setting. 9
13 7 Alarm release setting The sign and Lo blink, other setting is the same as the above, pressing enter key to save and enter the next parameter setting. 8 Alarm switch Obligate function,no at present Note:CCT3320T Setting Menu don t contain item. VI. Introduction for Electrode Installation 6.1Selection for electrode constant The measurement range depends on the electrode constant, so please choose the suitable electrode constant for your applications. 6.2 Outline dimension of the electrode Outline dimension of the electrode 10
14 6.3 Installation method Please follow the correct installation method to install the electrode strictly. The incorrect installation will cause the reading error. 1)Picture 1a the fitting is too long and the stretched parts is too short, which will lead to the dead space in electrode and incorrect measurement. please follow the picture 1b (FLOW=stretch into flow direction deeply) Installation 1a installation 1b 2)Picture 2a, the installation mode will lead to air space and incorrect measurement and instability. Please follow the picture 2b Installation 2a installation 2b 11
15 3)Picture 3, there will be air space inside the pipeline, the air will cause incorrect reading. 4)Picture 4a, the below installation will cause incorrect reading,because the water can not through the measuring hole of the electrode. 6.4 Installation and maintenance Notes: (1) The electrode should be installed in a place in the pipeline where the stream is steady and air bubbles are hard to generate. (2) No matter the conductance cell is horizontally or vertically installed, it should be deeply inserted into the moving water. (3) The conductivity signal is weak electronic signal and its collecting cable should be separately installed. When threading cable joint or connecting terminal board is used, to avoid wetting interference or breakdown of measurement unit circuit, they should not be connected to the same group of cable joint or terminal board with the power line or control line. (4) When the measurement cable needs to be lengthened, it s recommended to make an 12
16 agreement with the factory before placing an order. (5) Please keep the measuring part of electrode clean, and do not directly contact the surface by hands or contact with the oil stain objects (6) Electrode is a kind of precision components, so please do not change any part of the electrode. The accuracy will be incorrect if the electrode was destroyed by the strong acid, strong alkali, scrape from machine and etc. (7) The meter is made by precision integrated circuit and electronic components, so it needs to place in case or dry environment. (8) In order to guarantee the safety operation, pls checking after installation then switch on. VII. Failure Judgment for Meter and Electrode When the reading is incorrect or unstable, please check the meter and electrode. a.distinguish the failure source which comes from meter or electrode Firstly, remove the white wire from the wire terminal and check the conductivity reading, if the reading is 0 and stable, the meter is good. The problem can be initially identified from the electrode installation. b. Sensor installation question failure judgment. Remove the electrode from the fittings, and then use the electrode to test the water quality (user already know the conductivity reading of the water before test). If the reading is correct, so the installation is correct. If the reading with error, so the electrode failure. C. Meter mode or transmitter mode of 4~20mA judgment. Before the judgment please makes sure there is no wire connection on the terminal. ma mode Output terminal Voltage of terminal Voltage of cable Meter mode I+/I- >12V DC no Transmitter mode T+/T- no DC24V General fault inspection and trouble shooting: Symptom Possible factor Trouble shooting methods 1. No reading 2.Unstable reading A. No power supply connection B. Instrument fault A. Wire connection of CELL incorrect Incorrect electrode installation B. Unstable flow speed D. So fast flow speed in some part A. Check to the wire connection of power supply B. Maintained by professional. A. Check the wire connection of manual. B. Adjust the pipe or select another measurement point C. Use the steady water supply to eliminate the meter factor. 13
17 3.Serious error of reading 4.different ma reading between sender and receiver A. Incorrect constant Setting B. The electrode constant is changed Incorrect electrode installation A. Reviver transfer error B. Not up to 20mA C. Incorrect setting on sending D. ma transfer error A. Reset the electrode constant B. Change the electrode by new one. C. Install the electrode at where the flow s speed is suitable. D. Put the electrode deeply into the flowing water. A. Set the transfer setting again B. Loop resistance is too large, enlarge the cable. C. Set the corresponding of ma and reading again. D. Use the ammeter to check the current. VIII. Complete Set of Products Conductivity /TDS transmitting controller installation clamp) Electrode Operation menu one piece(with a pair of fast one piece (5m cable length) one piece IX. Ordering Directory CCT-3300 series including 4 models, please check the below items to choose the suitable power supply, material, fixed method for your applications Model No. Power supply Electrode Connection Electrode material constant CCT-3300E DC 24V 0.10 cm L CCT-3310E AC 110V 1.00 cm L plastic CCT-3320 AC 220V 5.00 cm -1 Thread Graphite CCT-3320E AC 220V cm -1 ferrule Graphite CCT-3320T AC 220V cm -1 Graphite NOTE: Cable length: 1m, 5m, 10m, 20m (please indicate another especial length before placing an order). The model with E is EMC enhancement type CCT-3320T(Conductivity Monitor) without any control outputs. *Without the influence on the operation, any small change or improvement on the products by the manufacturer will not be notified separately. Please make the object as the standard. 14
18
User s Manual Before using the inverter, you need to read and save the safety instructions.
User s Manual Before using the inverter, you need to read and save the safety instructions. STI SERIES (STI200, STI300, STI500, STI700, STI1000) Power Frequency Pure Sine Wave Inverter The information
Digital inductive conductivity transmitter
Digital inductive conductivity transmitter Optimal solution for conductivity measurements in difficult fluids (polluted, dirty,...) PEEK/PPA version for CIP applications Large range of process connections
www.ringautomotive.co.uk Instructions: Retain these instructions for future reference SmartChargePro35 RSCPR35-12v, 2 / 8 / 16 / 35A
SmartChargePro7 RSCPR7-12v, 7A SmartChargePro10 RSCPR10-12v, 10A SmartChargePro15 RSCPR15-12v, 15A SmartChargePro25 RSCPR25-12v, 2 / 6 / 12 / 25A SmartChargePro35 RSCPR35-12v, 2 / 8 / 16 / 35A SmartChargePro50
How To Control A Power Supply With A Mini Transmitter And Switch (Power Supply)
transmitter (2 wire) / switch for continuous or On/Off Control Indication, monitoring, transmitting and continuous or On/Off control in one device Output signal 4...20 ma, 2-wire for continuous control
ELECTRONIC COUNTING SCALE. Operation Manual. l JCS- A SERIES
ELECTRONIC COUNTING SCALE Operation Manual l JCS- A SERIES CONTENTS I. Specification:... II. Operation:... III. Weight calibration:... IV. RS232 transmission confer Data format (optional) V. Regular trouble
Duct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical
PRODUCT PROFILE FEATURES
PRODUCT PROFILE Pressure Transmitters & Pneumatic Transducer Airflow Transmitter, Differential Air Pressure Transmitter and Differential & Gauge Liquid/Gas Pressure Transmitter FEATURES AIRFLOW TRANSMITTER
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
TIG INVERTER INSTRUCTION MANUAL
TIG INVERTER INSTRUCTION MANUAL Contents Warning General Description Block Diagram Main Parameters Circuit Diagram Installation and Operation Caution Maintenance Spare Parts List Troubleshooting 3 4 4
LS1024B / LS2024B/ LS3024B. Solar Charge Controller USER MANUAL
EPSOLAR LS1024B / LS2024B/ LS3024B Solar Charge Controller USER MANUAL Thank you very much for selecting our product! This manual offers important information and suggestions with respect to installation,
Operation Manual For NDJ-8S Digital Rotary Viscometer
Operation Manual For NDJ-8S Digital Rotary Viscometer 1 Contents 1. General... 2 2. Main technical data... 2 3. Working principle... 2 4. Installation... 3 5. Operation procedures... 4 6. Precautions...
Manual Ranging MultiMeter
Owner s Manual Manual Ranging MultiMeter Model 82345 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product.! Safety! Operation! Maintenance!
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Introduction. Refrigerant Leak Detecting System. Feature and Benefits
PSC European Refrigeration Controls Catalogue Catalog Section 9 Product Bulletin PD-GAS-E Refrigerant Leak Detecting System Introduction This range of refrigerant leak detecting systems is designed for
RDJ10RF/SET. Wireless room temperature controller with 24-hour time switch and LCD. Programmable, for heating systems
3 072 RDJ10RF RCR10/433 Wireless room temperature controller with 24-hour time switch and LCD Programmable, for heating systems RDJ10RF/SET Operating modes: Automatic, Comfort, Energy Saving, and Frost
OPERATING INSTRUCTIONS
PIPELINE INSPECTION COMPANY LTD. OPERATING INSTRUCTIONS Wet Sponge Holiday Detectors 670,673, and MSRB Wet Sponge Holiday Detectors Portable and In-Plant Detectors Table of Contents General Information.......................3
Features. Display. Measurements. Intelligent. Accuracy. Models. Installation DEIF A/S. Multi-instrument 4921210109D
7000/7000C/7020 Multi-instrument 4921210109D Features Measurements All 3-phase AC measurements True RMS Replaces analogue meters Demand on each phase current Accuracy U, I and F class 0.5 Other values
RISH CON - Hz. Salient Features : Application : Product Features: FREQUENCY TRANSDUCER
Application : The RISH CON - Hz transducer is used for frequency measurement. The output signal is proportional to measured frequency and is either load independent DC Current or load independent DC Voltage.
Detector transparent with Color Inserts. FAA 500 TR P Trim Ring transparent with Color Inserts. FCA 500 / FCA 500 E Detector Bases
Detector Color Detector transparent with Color Inserts FAA 500 TR W Trim Ring FAA 500 TR P Trim Ring transparent with Color Inserts FAA 500 BB Ceiling Mount Back Box FCA 500 / FCA 500 E Detector Bases
Solar Controller / Battery Charger User s Manual
Solar Controller / Battery Charger User s Manual 1 Congratulations! You have made an excellent choice by purchasing this high quality KORR PWM solar controller which has been manufactured to the highest
How To Use A Power Supply Unit (Upu)
BRAVER UPS (Uninterruptible Power System) User s Manual Safety CAUTION! This UPS utilizes voltages that may be hazardous. Do not attempt to disassemble the unit. The unit contains no user replaceable parts.
RM17TE 183...528 V AC. Main. Product or component type. Product specific application. Relay monitored parameters 250 V DC 5 A DC
Characteristics multifunction control relay RM17-TE - range 183..528 V AC Complementary Reset time Maximum switching voltage Minimum switching current Maximum switching current [Us] rated supply voltage
T0118 T2118 T3118. Instruction Manual
Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity
OCSM Series. High Resolution Digital Crane Scale. User Guide
OCSM Series High Resolution Digital Crane Scale Content 1. Safety Guide... 1 2. Features... 1 3. Specifications... 2 4. Capacity... 3 5. Display & Keys... 3 Scale & Remote Keys... 3 Indicators... 4 Message...
12 Volt 30 Amp Digital Solar Charge Controller
12 Volt 30 Amp Digital Solar Charge Controller User s Manual WARNING Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions
MILLENIUM INSTALLATION MANUAL NTR 735 A. Simple Automation Control Module (MAS)
MILLENIUM Table of contents 1. INTRODUCTION 1 2. HARDWARE DESCRIPTION 2 3. INSTALLATION 5 4. CONNECTION 6 5. USER SAFETY AND PROTECTION OF EQUIPMENT 8 1. Introduction The MILLENIUM series has been designed
HEAT MONITORING AND CONTROL INSTRUMENTS FOR TRANSFORMERS MB 103 TEMPERATURE CONTROL UNIT PT 100 TEMPERATURE SENSOR
HEAT MONITORING AND CONTROL INSTRUMENTS FOR TRANSFORMERS MB 103 TEMPERATURE CONTROL UNIT PT 100 TEMPERATURE SENSOR 2 MB 103 TRANSFORMER TEMPERATURE CONTROL UNIT WITH CONTROL SIGNAL Recent approval of new
HERZ-Thermal Actuators
HERZ-Thermal Actuators Data Sheet 7708-7990, Issue 1011 Dimensions in mm 1 7710 00 1 7710 01 1 7711 18 1 7710 80 1 7710 81 1 7711 80 1 7711 81 1 7990 00 1 7980 00 1 7708 11 1 7708 10 1 7708 23 1 7709 01
Notice. Reverse engineering and disassembly are prohibited.
Notice The information in this document is subject to change without notice. Every effort has been made in the preparation of this document to ensure accuracy of the contents, but all statements, information,
Power source. Input. Contact / Open collector Voltage input (SELECTABLE) AC100-240V DC12-24V AC100-240V. Output indicator
SERIES Electronic Preset Counter Instruction Manual Please note that misuse of this device may lead to injury to the user or damage to the device. Please observe all safety precautions and warnings in
Radiant Temperature Sensor TY7321
AB-5361-U Specifications/Instructions Radiant Temperature TY7321 General TY7321 Radiant Temperature s are designed to measure infrared radiation from perimeter windows and walls and provide a proportional
Temperature Controller CD401/CD901/CD701/CD501/CD100/CD101 Instruction Manual
Ⅰ. Warn Warning of connection Temperature Controller CD401/CD901/CD701/CD501/CD100/CD101 Instruction Manual If the controller fails to operate or error occurs, the system will bring fault for it, please
12 Volt 30 Amp Digital Solar Charge Controller Installation & Operation Manual
12 Volt 30 Amp Digital Solar Charge Controller Installation & Operation Manual This 30Amp charge controller is designed to protect your 12Volt Lead-acid or Gel-cell battery from being overcharge by solar
Ground Resistance Clamp On Tester
USER MANUAL Ground Resistance Clamp On Tester MODEL 382357 Introduction Congratulations on your purchase of Extech s 382357 Ground Resistance Tester. This Clamp on device allows the user to measure ground
HCR3.2/.. Chipcard reader HOTEL SOLUTION
s 6 332 HOTEL SOLUTION Chipcard reader HCR3.2/.. Chipcard reader for hotel room access control Reads access code on chipcard Transfers access code to room controller Built-in optical display of messages
StructureScan HD Module. Installation Guide ENGLISH. www.bandg.com www.simrad-yachting.com www.lowrance.com
StructureScan HD Module Installation Guide ENGLISH www.bandg.com www.simrad-yachting.com www.lowrance.com Disclaimer As Navico is continuously improving this product, we retain the right to make changes
TYPE 65UV5-1004 CEX Integrated Flame Scanner with Internal Flame Relay and Analog Output
133-683 JUNE 28, 2006 TYPE 65UV5-1004 CEX Integrated Flame Scanner with Internal Flame Relay and Analog Output DESCRIPTION APPLICATION The Fireye 65UV5-1004 flame scanner is a microprocessor based flame
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
*.ppt 11/2/2009 12:48 PM 1
Digital Compressor Controller *.ppt 11/2/2009 12:48 PM 1 Copeland Scroll Digital Controller Simple Controller That Enables OEM s To Use Digital Scrolls Relieves OEM From Developing Special Controllers
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2. Instruction Manual
Firmware version: 1.10 Issue: 7 AUTODIALER GD30.2 Instruction Manual Firmware version: 2.0.1 Issue: 0.6 Version of the GPRS transmitters configurator: 1.3.6.3 Date of issue: 07.03.2012 TABLE OF CONTENTS
GSM Voice Auto Dialer & SMS Sender JC-999
GSM Voice Auto Dialer & SMS Sender JC-999 Main features: () Large LCD display,integrated integrated keypad, human-friendly programming; (2) Support 2 alarm receiving numbers; (3) Triggering means: N.C
Neo-Neon. LED Vision Mesh Screen Operation Manual. LED Vision Mesh Screen. Neo-Neon International Ltd He Shan Decorative Lighting Company Limited
Neo-Neon LED Vision Mesh Screen Operation Manual He Shan Decorative Lighting Company Limited LED Display Engineering Department 1 Contents Chapter1 Specification and Function Parameters 3 Chapeter2 Structure
Fire Alarm Control Panel. Family. Operating Manual
Fire Alarm Control Panel Family Operating Manual CONTENTS System Description AH-03312 System Characteristics... Fire Signal Receiving Board Description-4L... Fire Signal Receiving Board Description-8L...
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765
IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS: 61-763 61-765 The Service Information provides the following information: Precautions and safety information Specifications Performance test procedure Calibration
Drayton Digistat +2RF/+3RF
/+3RF Programmable Room Thermostat Wireless Model: RF700/22090 Model: RF701/22092 Power Supply: Battery - Thermostat Mains - Digistat SCR Invensys Controls Europe Customer Service Tel: 0845 130 5522 Customer
CW-5000/ 5200,1'8675,$/ &+,//(5 86(5 0$18$/
CW-5000/ 5200 Contents Cautions ---------------------------------------------------------------- 3 Parts introduction --------------------------------------------------- 4 Installation ------------------------------------------------------------
Manual. Simrad StructureScan LSS-1 Sonar Module. English
Manual Simrad StructureScan LSS-1 Sonar Module English www.simrad-yachting.com A brand by Navico - Leader in Marine Electronics Disclaimer As Navico is continuously improving this product, we retain the
Electronic timer CT-AHD.12 OFF-delayed with 1 c/o (SPDT) contact
Data sheet Electronic timer CT-AHD.12 OFF-delayed with 1 c/o (SPDT) contact The CT-AHD.12 is an electronic time relay with OFF-delay. It is from the CT-D range. With their MDRC profile and a width of only
PC INTERFACE MODULES FOR TOLL-PLAZA CLASSIFICATION SENSORS
PC INTERFACE MODULES FOR TOLL-PLAZA CLASSIFICATION SENSORS FRANCE Head office USA 4 Rue du Bois Chêne le loup Parc d Activité de Brabois 54 500 VANDOEUVRE LES NANCY (33) 0383442413, Fax (33) 0383443797
Sanitary and CIP Toroidal Sensor
Instruction Sheet PN 51A-225/rev.I December 2010 Model 225 Sanitary and CIP Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS
AutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
FIREDEX 2200. Conventional Fire Panels
68 Flexible, high specification system Choice of 1, 2, 4 or 8 zones Simple one-shot auto-reset user test facility Approved to EN54 Maintenance free poly switch circuit protection, with auto reset Class
EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM
EVC40 EMERGENCY VOICE COMMUNICATION SYSTEM INSTALLATION MANUAL Protec Fire Detection PLC, Protec House, Churchill Way, Nelson, Lancashire, BB9 6RT. Telephone: +44 (0) 1282 717171 Fax: +44 (0) 1282 717273
TMS TANK MANAGEMENT SYSTEM
TMS TANK MANAGEMENT SYSTEM Page 1 of 9 Operating Instructions GENERAL The Tank Management System is a bespoke design to control, monitor and accommodate efficient storage and dispensing of TMS. FUNCTIONS
RAINCONTROL LCD Special
RAINCONTROL LCD Special edition 1.1 01/2007 - code 5880 1/16 RAINCONTROL LCD SPECIAL... 3 RAINCONTROL SPECIAL LCD PANEL PRESENTATION... 4 AUTO... 5 LOW AND HIGH THRESHOLD SETTING... 6 LOW THRESHOLD...
AUTODIALLER / QUICKDIALLER - SA132
AUTODIALLER / QUICKDIALLER - SA132 INSTRUCTION LEAFLET ENGLISH www.thermomax-group.com CONTENTS 1 SETUP AT A GLANCE... 2 2 FOREWORD....... 3 3 INSTALLATION...... 4 4 KEYPAD AND INDICATORS...... 5 SETTING
Nokeval. User's Manual Models 301 / 302 / 305
Nokeval User's Manual Models 301 / 302 / 305 No 16042010 1 Contents Contents... 2 General description... 3 Disabling front panel keys... 4 Front panel buttons and indicators... 5 Configuration mode...
Ground Resistance Clamp-On Tester
User s Manual Ground Resistance Clamp-On Tester MODEL 382357 Warranty EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants the basic instrument to be free of defects in parts and workmanship for one
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
1. SAFETY INFORMATION
RS-232 Sound Level Meter 72-860A INSTRUCTION MANUAL www.tenma.com 1. SAFETY INFORMATION Read the following safety information carefully before attempting to operate or service the meter. Use the meter
Voltage Regulator SPAU 341 C. Product Guide
Issued: July 1998 Status: Updated Version: D/25.04.2006 Data subject to change without notice Features Comprehensive voltage regulation for power transformers with on-load tapchangers in distribution substations
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &
REGULINE 600VA / 1000VA REGULATOR USER MANUAL
REGULINE 600VA / 1000VA REGULATOR USER MANUAL TUNÇMATİK REGULINE SERIES AUTOMATIC VOLTAGE REGULATOR Models: REGULINE 600VA / REGULINE 1000VA Principle of Operation Automatic Voltage Regulators regulate
Delta Transportation Automation Solution. www.deltaww.com
Delta Transportation Automation Solution www.deltaww.com Smarter. Greener. Together. Who We Are Delta is a market leader that has won wide recognition as a: Global power and thermal solution provider for
SOLAR POWER - THE NEW WAY TO RV!
IMPORTANT! - YOU'RE SOLAR READY SOLAR POWER - THE NEW WAY TO RV! Your RV is PRE-WIRED for SOLAR POWER! Benefits of Solar Zamp Solar Portable Solar Kit Maximize Battery Life Electrical Independence Green,
Operating Instructions
Operating Instructions Dear customer, Congratulations on your decision to purchase the Smart-Switch from our range! This innovative product is the world s first electronic multi-function switch system,
SYSTEM 4C. C R H Electronics Design
SYSTEM 4C C R H Electronics Design SYSTEM 4C All in one modular 4 axis CNC drive board By C R Harding Specifications Main PCB & Input PCB Available with up to 4 Axis X, Y, Z, A outputs. Independent 25
Digital input modules
8 172 TX-I/O Digital input modules TXM1.8D TXM1.16D Two fully compatible versions: TXM1.8D: 8 inputs, each with a three-color LED (green, yellow or red) TXM1.16D: As TXM1.8X, but 16 inputs, each with a
FT-100 Fire Training Fog Machine User Manual English
FT-100 Fire Training Fog Machine User Manual English 2015 Antari Lighting and Effects Ltd. User Manual - English Safety Information Please read the following safety information carefully before operating
FPV MONITOR. User Manual
FPV MONITOR User Manual Important Safety Instructions: Please read User Guide before using this product. Please keep User Guide for future reference. Please read the cautions to prevent possible danger
RM35UA13MW. range 15..600 V. Main. Product or component type. Relay monitored parameters
Characteristics multifunction voltage control relay RM35-U - range 15..600 V Complementary Reset time Maximum switching voltage [Us] rated supply voltage Supply voltage limits Power consumption in W Main
EPSOLAR LS0512R / LS0524R. Solar Light Controller INSTRUCTION MANUAL. Please read this manual carefully before using the product!
EPSOLAR LS0512R / LS0524R Solar Light Controller INSTRUCTION MANUAL Please read this manual carefully before using the product! LandStar LS0512R / LS0524R Solar Light Controller Nominal system voltage
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
User Guide. Model 380260 Insulation Tester / Megohmmeter
User Guide Model 380260 Insulation Tester / Megohmmeter Introduction Congratulations on your purchase of Extech s Insulation Tester/Megohmmeter. The Model 380260 provides three test ranges plus continuity
Power consumption (kw)
OS Series Series description LIT OS systems are applied to wastewater disinfection. The models of this series were developed on base of high-efficiency low-pressure amalgam lamps and have all components
COLOR TFT LCD MONITOR. User Manual
COLOR TFT LCD MONITOR User Manual GENERAL INFORMATION Thank you for choosing our TFT LCD(liquid crystal display) monitor. This product employs integrate circuits, low power consumption, and no radiation
INSTRUCTION MANUAL OVERVIEW. ---------- Remote Meter: MT-5 Remote meter (Model MT-5) is available to connect with solar controller Tracer MPPT series.
INSTRUCTION MANUAL ---------- Remote Meter: MT-5 Remote meter (Model MT-5) is available to connect with solar controller Tracer MPPT series. OVERVIEW The digital Remote Meter displays system operating
Hygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
INTELLIGENT CONTROL MODULE SA-2000-II HARDWARE MANUAL. Access Technologies International, Inc.
SA-2000-II HARDWARE MANUAL Access Technologies International, Inc. 1 Table of Contents 1. Introduction 3 2. Features 3 3. Specification 4 4. Identifying Supplied Parts 4 5. Panel Description 5 6. Connection
NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL
NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. OT 020 BENCH TOP STEAM STERILIZER USER S MANUAL NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. Esenboğa Yolu 22 km Akyurt - ANKARA / TURKEY Tel : (00 90 312)
PIEZOMETER AIRFLOW MEASURING RING
PIEZOMETER AIRFLOW MEASURING RING M.K. Plastics is pleased to introduce The Piezometer Airflow Measuring Ring, now available as an accessory for our steel Axijet fans. The system consists of a Piezometer
IN-OUT Thermometer with Cable Free Sensor and Clock
IN-OUT Thermometer with Cable Free Sensor and Clock MODEL: RAR232 USER'S MANUAL INTRODUCTION Congratulations on your purchase of the RAR232 In-Out Thermometer with 433MHz cable free sensor and calendar
Streaming Potential System (SPT1000, SPD1000)
Streaming Potential System (SPT1000, SPD1000) User's Manual Sentrol Systems, Inc. 3949 Cotswold Dr. SW, Lilburn, GA 30047-2371 Tel: 770-564-1541, Fax: 770-564-8605 www.sentrolsystems.com Since the streaming
How to read this guide
How to read this guide The following shows the symbols used in this Quick start guide with descriptions and examples. Symbol Description Example P oint Reference Caution [ ] This symbol explains information
Programming Logic controllers
Programming Logic controllers Programmable Logic Controller (PLC) is a microprocessor based system that uses programmable memory to store instructions and implement functions such as logic, sequencing,
ACS-30-EU-EMDR-10-MOD
Multi-application heat tracing Control and Monitoring in commercial and residential buildings Roof & Gutter de-icing sensor module Technical Information Approvals Module IP rating Ambient operating temperature
Temperature & Humidity SMS Alert Controller
Temperature & Humidity Alert Controller METERS 3 simple steps starting the unit: Insert the SIM card Plug in the sensors connectors Connect the AC power cord. Specifications: AC 90~260V Auto Select Internal
GSM Alarm System User Manual
GSM Alarm System User Manual For a better understanding of this product, please read this user manual thoroughly before using it. Quick Guider After getting this alarm system, you need to do the following
Multifunction devices
devices devices devices, electronic Type Page LED multifunction displays pulse, frequency, time (DC) Codix 524 240 pulse, frequency, time (AC+DC) Codix 544 243 LCD multifunction preset counters 1 or 2
PSC-A5 Solar Charge Controller Manual
( 5A, 12V/24V Output, Double Mode ) 1. Introduction: This controller is suitable for 12V and 24V solar photovoltaic system. The charge and discharge current is 5A. The maximum solar panel for 12V system
01-3 0000-00 6810-20 AIR CONDITIONING SYSTEM 1. FFH SPECIFICATION AIR CONDITIONING SYSTEM RODIUS 2004.09
0000-00 01-3 6810-20 1. FFH SPECIFICATION 01-4 0000-00 2. SYSTEM LAYOUT AND COMPONENTS 0000-00 01-5 01-6 0000-00 3. FFH GENERAL INFORMATION The system is to increase the coolant temperature quickly by
Doc.No. NDP 192U-02. Electric Pump Controller CE-124P. Instruction Manual
Doc.No. NDP 192U-02 Electric Pump Controller CE-124P Instruction Manual 1. Parts Name and Function 1) POWER LED Indicate that CE-124P is powered up. 2) START/STOP KEY (SELECTOR KEY) Used to start and stop
BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER. Installation, Operation, and Maintenance Instructions. Manual No.
BROOKFIELD MODEL VTE250 ELECTRIC VISCOSEL PROCESS VISCOMETER Installation, Operation, and Maintenance Instructions Manual No. M06-512 SPECIALISTS IN THE MEASUREMENT AND CONTROL OF VISCOSITY BROOKFIELD
The Foundation User Manual LED Battery Slim Par Fixture (RGBA, DMX 7Ch/4CH)
The Foundation User Manual LED Battery Slim Par Fixture (RGBA, DMX 7Ch/4CH) Page 1 I. Security: 1. Make sure the user voltage cannot higher or lower than the regulating voltage. 2. Please use proper and
PSU-U-24V/AC-6A/10/FTA-TR-MC PSAC 0624
PSU-U-24V/AC-6A/10/FTA-TR-MC PSAC 0624 v.1.0 AC/AC power supply unit Edition: 1 from 24 th October 2008 Supercedes edition: ---------- 10062009 CONTENTS: 1. Technical description. 1.1. General description
Automation System TROVIS 6400 TROVIS 6493 Compact Controller
Automation System TROVIS 6400 TROVIS 6493 Compact Controller For panel mounting (front frame 48 x 96 mm/1.89 x 3.78 inch) Application Digital controller to automate industrial and process plants for general
Match. GE Digital Energy. Uninterruptible Power Supply 500-1500 VA. Technology for the Digital World. Match UPS. GE Digital Energy.
Match Uninterruptible Power Supply 500-1500 VA Manufactured by: General Electric Company Telephone +41 (0)91 / 850 51 51 CH 6595 Riazzino (Locarno) Fax +41 (0)91 / 850 51 44 Switzerland Website www.gedigitalenergy.com