Automatic time gain compensation and dynamic range control in ultrasound imaging systems

Size: px
Start display at page:

Download "Automatic time gain compensation and dynamic range control in ultrasound imaging systems"

Transcription

1 Automatic time gain compensation and dynamic range control in ultrasound imaging systems Duhgoon Lee, Yong Sun Kim, and Jong Beom Ra * Dept. of EECS, KAIST, Guseong-dong, Yuseong-gu, Daejeon, Korea ABSTRACT For efficient and accurate diagnosis of ultrasound images, the appropriate time gain compensation (TGC) and dynamic range (DR) control of ultrasound echo signals are important. TGC is used for compensating the attenuation of ultrasound echo signals along the depth, and DR is for controlling the image contrast. In recent ultrasound systems, those two factors are automatically set by a system and/or manually adjusted by an operator to obtain the desired image quality on the screen. In this paper, we propose an algorithm to find the optimized parameter values for TGC and DR automatically. In TGC optimization, we determine the degree of attenuation compensation along the depth by reliably estimating the attenuation characteristic of ultrasound signals. For DR optimization, we define a novel cost function by properly using the characteristics of ultrasound images. Experimental results are obtained by applying the proposed algorithm to a real ultrasound (US) imaging system. The results prove that the proposed algorithm automatically sets values of TGC and DR in real-time so that the subjective quality of the enhanced ultrasound images may become good enough for efficient and accurate diagnosis. Keywords: ultrasound imaging, time gain compensation, dynamic range control. INTRODUCTION The ultrasound imaging system is popularly used in clinical diagnosis because the imaging is non-invasive and can be performed in real-time. However, there exist several disadvantages compared to X-CT (X-ray computed tomography) and MRI (magnetic resonance imaging) images. One of them is the presence of speckle noise in ultrasound images. The noise induces quality deterioration in the images and provides a negative impact on clinical diagnosis. Therefore, many algorithms have been continually developed to reduce the speckle noise. In consequence, the image quality for some applications is close to that of MR images. As the other disadvantage, an ultrasound imaging system needs many interactions with an operator to obtain a desired image quality. In the process of obtaining an ultrasound image, the operator actually needs to adjust several basic parameters of the system and the image quality is sensitive to their values. Among them, the parameters of TGC and DR control are to be frequently adjusted by the operator. Here, TGC is for compensating the attenuation of ultrasound echo signal along the depth and DR control is for determining a proper range to be displayed from a wide range of echo signal. Hence, it is considered meaningful to automatically find the optimized parameter values for TGC and DR in real-time without operator s interaction. In the conventional ultrasound imaging system, the time gain compensation for an ultrasound signal is done by using a compensation curve, and the curve can be adjusted by a user. An ultrasound image is composed of the arrays of hundreds of scan lines, and each scan line represents a sampled ultrasound echo signal along the depth. Several algorithms have been proposed for compensating the intensity of ultrasound signals along the depth []-[5]. These methods focus on the individual compensation of each scan line. Also, some algorithms are considered mainly at the radio frequency signal level, thereby they can be implemented only through the modification of system hardware. Dynamic range control in ultrasound systems is used to determine a range to be displayed as an 8-bit B/W image from a wide range of log-compressed echo signal. The relationship between a log-compressed ultrasound signal and the B/W ultrasound image after the DR control is usually modeled as in Fig.. Here, the range of 8 bits is mainly due to the * jbra@ee.kaist.ac.kr; phone ; fax: ; www-isl.kaist.ac.kr Medical Imaging 2006: Ultrasonic Imaging and Signal Processing edited by Stanislav Emelianov, William F. Walker, Proc. of SPIE Vol. 647, 64708, (2006) /06/$5 doi: 0.7/ Proc. of SPIE Vol

2 limitation of display system. Based on the relationship given in Fig., we can adjust the image contrast by changing the value of DR. In this case, the initial point of the DR is fixed and only the width of DR is changed. Note here that the image contrast affects the roughness of soft tissue and the edgeness near the tissue boundary, thereby strongly affects the clinical subjective quality. Intensity 2 8 US image 0 2 DR 5 Intensity Log-compressed echo signal Fig.. Relationship between a log-compressed echo signal and the corresponding ultrasound image Many algorithms for improving the ultrasound image contrast have been developed. Among them, one method tried to increase the CNR of lesions to the background by analyzing the frequency spectra [6]. However, this method is not appropriate for finding the optimal DR because it deals with only lesions rather than the overall image. As the other approach, the gray-scale image enhancement algorithm has been proposed [7]. But this method cannot find the optimal value of DR because the individual transformation of each pixel value does not yield the global transformation of DR as shown in Fig.. In this paper, we aim to develop an algorithm to automatically find the optimal parameters values for TGC and DR control in real-time without operator s interaction. The proposed algorithm is designed on the basis of a simplified model of the ultrasound imaging system without a loss of generality. Therefore, the algorithm differs from the early attempts and can be easily applied to a practical ultrasound imaging system. This paper is organized as follows. In Section 2, the proposed algorithm is described. Section 2. describes a TGC optimization scheme based on the attenuation estimation and compensation along the depth. In Section 2.2, we introduce a DR optimization scheme by using the proposed measures of edge contrast and soft tissue roughness. Experimental results are shown in Section 3. Finally, conclusions are drawn in Section PROPOSED ALGORITHM The proposed algorithm consists of two procedures, TGC and DR optimization. As an input to the algorithm, we use the log-compressed digital echo data whose intensities are not compensated along the depth. We perform the optimal TGC based on this input data. After compensating for the intensities of the input data along the depth, the DR optimization is performed. Finally the optimized parameter values for TGC and DR control are applied to the system. 2. TGC optimization It is known that ultrasound echoes attenuate exponentially with the propagation distance from the interface. Hence, logcompressed ultrasound echoes linearly decay with the distance. The degree of attenuation of echoes depends on the propagation medium. If we assume that the attenuation in an ultrasound imaging system occurs due to the one major medium in the body, the attenuation of log-compressed echoes can be modeled as a straight line. In order to alleviate undesirable errors in estimating a straight line and achieve statistically reliable TGC results, we propose the following Proc. of SPIE Vol

3 method. 2.. Vertical profiles and straight line modeling To estimate the attenuation, we need to measure the intensity variation along the depth. It is usually difficult to estimate the attenuation by using a log-compressed echo signal because of the noise. In order to overcome this problem, we may use a vertical profile which represents the average of several neighboring echo signals. To obtain the profiles, we divide an input echo image into M subsequent vertical strips. If each vertical strip consists of N scan lines, the corresponding vertical profile v k (n) can be written as vk ( n) = N N k m= u ( m, n) () where u k (m, n) denotes the intensity at (m, n) within the kth strip and the range of k is to M. As mentioned above, the attenuation function of log-compressed echoes in a vertical strip can be modeled as a straight line. And this straight line can be modeled by using the averaged vertical profile v k (n). In approximating the vertical profile to a line profile, we adopt the least squares fit technique. Fig. 2 shows one of the vertical profiles and its approximated line profile to estimate the attenuation. We can see in the graph that the approximated line profile well represents the attenuation of the vertical profile along the depth. Vertical profile Approximated line profile 96.4 Intensity Depth [pixel] Fig. 2. Average intensity profile of a vertical strip along the depth in an ultrasound image (solid line) and its approximated line profile (dashed line) 2..2 Compensation In an ultrasound image, there can be local dark regions such as a vessel and mass in the scanned area. And these dark regions may disturb a correct estimation of the attenuation line because they tend to make the slope of the approximated line profile steeper. Therefore, to avoid this problem, we exclude the line profiles with steeper slopes in averaging M line profiles. Then, the remained line profiles are averaged to obtain the attenuation line function for the ultrasound image. In the attenuation line, the attenuation value at the zero depth is converted to one for normalization. Based on the slope of the attenuation line, we perform TGC, or compensate for the intensities along the depth. In this case, we can consider that the TGC is being done by adjusting the slope of the attenuation line to zero. However, if we simply make the slope zero, the noise will be considerably amplified in the far field area. In the clinical sense, such noise Proc. of SPIE Vol

4 amplification is not desirable. Therefore, in order to prevent the unwanted noise amplification in the far field area, we need to alleviate the degree of TGC. Fig. 3 shows an estimated attenuation line a(n) and its compensated versions. In the figure, α denotes the normalized attenuation factor at the deepest point of the image and has a value in the range of 0 to. For TGC, we use a value of α which is smaller than, while the value is to be for ideal compensation as in c i (n) in the figure. Practically, TGC can be performed by multiplying the intensities of echoes by a proper gain according to the depth. However, in the ultrasound image, the multiplicative model is changed into the additive model because the input echo signals are log-compressed in the system. Therefore, the image intensity u k (m, n) can be compensated as follows; c u ( m, n) = u ( m, n) + β ( c( n) a( n)), (2) k k where a(n) denotes the estimated attenuation line and c(n) denotes the desired attenuation line after the compensation. Also, β denotes the intensity value at the zero depth of the estimated attenuation line before normalization. Note in Eq. (2) that TGC is simply done by the addition. Intensity α Ideal compensation Practical compensation Estimated attenuation line c i (n) c(n) a(n) Depth [pixel] Fig. 3. Estimated attenuation line (solid) and desired attenuation lines (dashed) 2.2 DR optimization The ultrasound imaging system displays an 8-bit B/W image which is selected from a log-compressed echo data with a wide range of intensity by using a given DR. A DR optimization algorithm is to find a value of DR that provides the best quality of 8-bit images displayed on the screen. The overall structure of the proposed DR optimization algorithm is given in Fig. 4. The proposed algorithm introduces a cost function as a quantitative measure representing the quality of the displayed image. And the algorithm tries to find the value of DR that minimizes the cost function. The cost function consists of two measures that are obtained on the basis of detected edges in a B/W image. For efficient calculation, we first obtain an edge map for the image in advance. Then, we calculate the measures only on edge regions described in the edge map Edge detection To detect edges, we adopt the coherent nonlinear anisotropic diffusion model which is known to be effective [8]. In the adopted diffusion model, a pixel locating on an edge has a large difference between its two eigenvalues. And the edge direction is determined as one of the two eigenvectors which represent the normal and tangential directions. However, a pixel in a homogeneous region where speckle noise is dominant, has a small difference between its two eigenvalues. Based on these characteristics of eigenvectors, we generate an edge map which includes the edgeness and direction at Proc. of SPIE Vol

5 each edge pixel. Start Initialize DR Edge map Convert input data into a B/W image Measure the characteristics Calculate the cost Change DR Minimum? No Stop Fig. 4. Proposed DR optimization algorithm Two proposed measures The DR directly affects the boundary contrast and the roughness of soft tissue. Hence, we define two measures, namely, edge contrast (EC) and soft tissue roughness (STR) based on the edge map. The edge contrast represents the intensity difference between two regions within a window, which are separated by the edge line centered at an edge pixel. It can be obtained by using the following equation; EC = K ( i, E ( f ( i, m + f ( i, ), A m B (3) where E denotes a set of edge pixels in an ultrasound image and K denotes the total number of edge pixels. And f(i, represents the average of intensity values of the pixels on the edge line, which passes through an edge pixel, (i,, and locates in a window centered at the edge pixel. Here, the edge line is determined by using the direction vector of the edge pixel, and the intensity value of each point on the edge line is calculated by interpolating its neighboring pixels. Also, m A and m B denote the averages of the intensity values of the two regions separated by the edge line in the window, respectively. The soft tissue roughness is then defined as a standard deviation of the intensities of non-edge pixels in the image. Since soft tissue regions in the image should be homogeneous, the soft tissue roughness becomes to represent the roughness caused by speckle noise, and can be obtained by using the equation, STR = L ( u( i, L ( i, E ( i, E u( i, ) 2. (4) Here, L denotes the total number of non-edge pixels and u(i, denotes the intensity at pixel (i,. Proc. of SPIE Vol

6 2.2.3 Cost function An image with a high EC and a low STR is clinically desirable. And the optimal condition may be obtained by adjusting parameter DR. Hence, we propose a cost function to find the optimal value of DR in a clinical sense. The cost function is defined as a weighted sum of STR and the inverse of EC, i.e., λ J ( DR) = + STR. (5) EC + ε Here, ε is a constant of a small value that prevents the denominator of the first term of the right side in the equation from being zero, and λ is a weighting factor which will be selected empirically. The proposed cost J gets lower as EC increases and/or STR decreases. Therefore, we can expect that it has the minimum value when DR is optimized. Fig. 5 shows a plot of the cost versus the value of DR. To find a value of DR which makes the cost minimum, we adopt the downhill simplex method, which reliably guarantees the convergence within a capture range where the global minimum exists Proposed cost Value of DR [x0 5 ] Fig. 5. The cost versus the value of DR 3. EXPERIMENTAL RESULTS To test the performance of the proposed algorithm, we implement the algorithm to a practical ultrasound imaging system which includes a PC with a CPU of 2.0GHz and GB memory. The ultrasound imaging system takes the optimized parameter values of TGC and DR based on the proposed algorithm and displays the enhanced images. The processing time to determine the parameter values depends on the characteristic of the input image, but is approximately between 00ms to 300ms without software code optimization. So, the proposed algorithm can be considered fast enough for realtime processing. The variation of the processing time is mainly due to the number of iterations to minimize the cost in DR optimization. We obtain experimental results for thyroid and carotid which are most frequently scanned in the clinical diagnosis. In the experiments, values of α and λ are determined on the basis of the opinions of clinicians. For thyroid and carotid, α and λ are set to 0.95 and 450, respectively. In the experiment, images before and after parameter optimization are not scanned at the same time. Since the images are obtained by scanning a real body, a certain time interval is involved before obtaining the enhanced image. During this interval, the initial image is stored and processed to obtain the optimal parameters values of TGC and DR. Since this interval is small, we can obtain quite similar images before and after parameter optimization, which are still good enough Proc. of SPIE Vol

7 for examining the performance of the proposed algorithm. Fig. 6 shows experimental results for the thyroid. Fig. 6(a) is an initial image in which the parameters of TGC and DR are not properly adjusted. The result after TGC optimization by using the input echo signal is given in Fig. 6(b). Compared to Fig. 6(a), Fig. 6(b) shows a more homogeneous image because it is compensated along the depth. The image after the DR optimization is shown in Fig. 6(c). In this figure, we can verify that the contrast of the input image is improved and the structures look better. Finally, the image after both TGC and DR optimization is shown in Fig. 6(d). As expected, the image shows the best quality compared to the other images. Experimental results for carotid are given in Fig. 7. Fig. 7(a) shows that the image intensity rapidly attenuates along the depth and the image looks rough due to high contrast. Meanwhile, Fig. 7(b) shows the image after the TGC optimization. In this image, the intensity is compensated but the contrast is still too high. In the image after the DR optimization, however, the contrast becomes lower and the image looks smoother as shown in Fig. 7(c). Of course, we can obtain the best result as shown in Fig. 7(d) after both TGC and DR optimization. Fig. 6. Experimental results for Thyroid: (a) Initial image and the images after (b) TGC optimization, (c) DR optimization, and (d) TGC and DR optimization, respectively. Proc. of SPIE Vol

8 Fig. 7. Experimental results for Carotid: (a) Initial image and the images after (b) TGC optimization, (c) DR optimization, and (d) TGC and DR optimization, respectively. 4. CONCLUSIONS In this paper, we present an algorithm to optimize parameters TGC and DR which strongly affect the image quality in ultrasound imaging systems. In the proposed algorithm, in order to find the optimized parameters for TGC, we divide an ultrasound image into consecutive vertical strips and obtain vertical profiles from the strips. Then, we approximate every vertical profile to a straight line, and estimate the degree of attenuation along the depth by averaging the straight lines. The parameter values for TGC are obtained by using the estimated attenuation line. For DR optimization, we define two measures of edge contrast and soft tissue roughness which are the functions of DR. For the optimal value of DR, the corresponding image is expected to have high edge contrast and low soft tissue roughness. Hence, the proposed cost function is defined as a weighted sum of the soft tissue roughness and the inverse of edge contrast, and it has the minimum value when DR is optimized. This cost function enables a robust DR optimization. By applying the proposed Proc. of SPIE Vol

9 algorithm to a practical ultrasound system, we showed that it is possible to automatically optimize the parameters for TGC and DR control in real-time. We prove that the subject quality of the image with the optimized parameters is clearly improved for clinical diagnosis. REFERENCES. D. I. Hughes and F. A. Duck, Automatic attenuation compensation for ultrasonic imaging, Ultrasound Med. Biol., vol. 23, pp , W. D. Richard, A new time-gain correction method for standard B-mode ultrasound imaging, IEEE Trans. Med. Imag., vol. 8, pp , T. Herrick and A. Bryant, Automated gain control for medical diagnostic ultrasound imaging, in Proc. IEEE Circuits and Systems, A. Bryant and T. J. Herrick, Adaptive gain control and contrast improvement for medical diagnostic ultrasound B- mode imaging system using charge-couples devices, in Proc. IEEE Circuits and Systems, B. Richard and O. Charliac, A new digital adaptive time gain correction for B-mode ultrasonic imaging, in Proc. IEEE Engineering in Medicine, P. F. Stetson, F. G. Sommer, and A. Macovski, Lesion contrast enhancement in medical ultrasound imaging, IEEE Trans. Med. Imag., vol. 6, pp , August C. Munteanu and A. Rosa, Gray-scale image enhancement as an automatic process driven by evolution, IEEE Trans. Systems, Man, and Cybernetics, vol. 34, pp , April Y. S. Kim and J. B. Ra, "Improvement of ultrasound image based on wavelet transform: speckle reduction and edge enhancement," in Proc. SPIE Medical Imaging 2005, Vol. 5747, pp. 085~092, Proc. of SPIE Vol

STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI

STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI STATISTICAL ANALYSIS OF ULTRASOUND ECHO FOR SKIN LESIONS CLASSIFICATION HANNA PIOTRZKOWSKA, JERZY LITNIEWSKI, ELŻBIETA SZYMAŃSKA *, ANDRZEJ NOWICKI Institute of Fundamental Technological Research, Department

More information

Digital Camera Imaging Evaluation

Digital Camera Imaging Evaluation Digital Camera Imaging Evaluation Presenter/Author J Mazzetta, Electro Optical Industries Coauthors Dennis Caudle, Electro Optical Industries Bob Wageneck, Electro Optical Industries Contact Information

More information

Jitter Measurements in Serial Data Signals

Jitter Measurements in Serial Data Signals Jitter Measurements in Serial Data Signals Michael Schnecker, Product Manager LeCroy Corporation Introduction The increasing speed of serial data transmission systems places greater importance on measuring

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

THREE DIMENSIONAL REPRESENTATION OF AMINO ACID CHARAC- TERISTICS

THREE DIMENSIONAL REPRESENTATION OF AMINO ACID CHARAC- TERISTICS THREE DIMENSIONAL REPRESENTATION OF AMINO ACID CHARAC- TERISTICS O.U. Sezerman 1, R. Islamaj 2, E. Alpaydin 2 1 Laborotory of Computational Biology, Sabancı University, Istanbul, Turkey. 2 Computer Engineering

More information

Combining an Alternating Sequential Filter (ASF) and Curvelet for Denoising Coronal MRI Images

Combining an Alternating Sequential Filter (ASF) and Curvelet for Denoising Coronal MRI Images Contemporary Engineering Sciences, Vol. 5, 2012, no. 2, 85-90 Combining an Alternating Sequential Filter (ASF) and Curvelet for Denoising Coronal MRI Images Mohamed Ali HAMDI Ecole Nationale d Ingénieur

More information

Ultrasonic Echosignal Applied to Human Skin Lesions Characterization

Ultrasonic Echosignal Applied to Human Skin Lesions Characterization ARCHIVES OF ACOUSTICS Vol.37,No.1, pp.103 108(2012) Copyright c 2012byPAN IPPT DOI: 10.2478/v10168-012-0014-7 Ultrasonic Echosignal Applied to Human Skin Lesions Characterization HannaPIOTRZKOWSKA (1),JerzyLITNIEWSKI

More information

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL

More information

Level Set Evolution Without Re-initialization: A New Variational Formulation

Level Set Evolution Without Re-initialization: A New Variational Formulation Level Set Evolution Without Re-initialization: A New Variational Formulation Chunming Li 1, Chenyang Xu 2, Changfeng Gui 3, and Martin D. Fox 1 1 Department of Electrical and 2 Department of Imaging 3

More information

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt.

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt. Medical Image Processing on the GPU Past, Present and Future Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt.edu Outline Motivation why do we need GPUs? Past - how was GPU programming

More information

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Routine Ultrasound Equipment Tests for Quality Assurance. Contents Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI,

More information

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number 1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression

More information

2. MATERIALS AND METHODS

2. MATERIALS AND METHODS Difficulties of T1 brain MRI segmentation techniques M S. Atkins *a, K. Siu a, B. Law a, J. Orchard a, W. Rosenbaum a a School of Computing Science, Simon Fraser University ABSTRACT This paper looks at

More information

Detection and Restoration of Vertical Non-linear Scratches in Digitized Film Sequences

Detection and Restoration of Vertical Non-linear Scratches in Digitized Film Sequences Detection and Restoration of Vertical Non-linear Scratches in Digitized Film Sequences Byoung-moon You 1, Kyung-tack Jung 2, Sang-kook Kim 2, and Doo-sung Hwang 3 1 L&Y Vision Technologies, Inc., Daejeon,

More information

How To Fix Out Of Focus And Blur Images With A Dynamic Template Matching Algorithm

How To Fix Out Of Focus And Blur Images With A Dynamic Template Matching Algorithm IJSTE - International Journal of Science Technology & Engineering Volume 1 Issue 10 April 2015 ISSN (online): 2349-784X Image Estimation Algorithm for Out of Focus and Blur Images to Retrieve the Barcode

More information

High-definition 3D Image Processing Technology for Ultrasound Diagnostic Scanners

High-definition 3D Image Processing Technology for Ultrasound Diagnostic Scanners Hitachi Review Vol. 63 (204), No. 9 584 Featured Articles High-definition 3D Image Processing Technology for Ultrasound Diagnostic Scanners Realistic 3D Fetal Imaging Masahiro Ogino Takuma Shibahara, Ph.D.

More information

Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006

Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006 Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 13, 2006 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain

More information

PERFORMANCE ANALYSIS OF HIGH RESOLUTION IMAGES USING INTERPOLATION TECHNIQUES IN MULTIMEDIA COMMUNICATION SYSTEM

PERFORMANCE ANALYSIS OF HIGH RESOLUTION IMAGES USING INTERPOLATION TECHNIQUES IN MULTIMEDIA COMMUNICATION SYSTEM PERFORMANCE ANALYSIS OF HIGH RESOLUTION IMAGES USING INTERPOLATION TECHNIQUES IN MULTIMEDIA COMMUNICATION SYSTEM Apurva Sinha 1, Mukesh kumar 2, A.K. Jaiswal 3, Rohini Saxena 4 Department of Electronics

More information

Simultaneous Gamma Correction and Registration in the Frequency Domain

Simultaneous Gamma Correction and Registration in the Frequency Domain Simultaneous Gamma Correction and Registration in the Frequency Domain Alexander Wong a28wong@uwaterloo.ca William Bishop wdbishop@uwaterloo.ca Department of Electrical and Computer Engineering University

More information

Assessment of Camera Phone Distortion and Implications for Watermarking

Assessment of Camera Phone Distortion and Implications for Watermarking Assessment of Camera Phone Distortion and Implications for Watermarking Aparna Gurijala, Alastair Reed and Eric Evans Digimarc Corporation, 9405 SW Gemini Drive, Beaverton, OR 97008, USA 1. INTRODUCTION

More information

Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997

Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997 Physics 623 Transistor Characteristics and Single Transistor Amplifier Sept. 8, 1997 1 Purpose To measure and understand the common emitter transistor characteristic curves. To use the base current gain

More information

Chapter 2 The Research on Fault Diagnosis of Building Electrical System Based on RBF Neural Network

Chapter 2 The Research on Fault Diagnosis of Building Electrical System Based on RBF Neural Network Chapter 2 The Research on Fault Diagnosis of Building Electrical System Based on RBF Neural Network Qian Wu, Yahui Wang, Long Zhang and Li Shen Abstract Building electrical system fault diagnosis is the

More information

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION 35'th Annual Precise Time and Time Interval (PTTI) Systems and Applications Meeting San Diego, December 2-4, 2003 A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION Józef Kalisz and Ryszard Szplet

More information

High-quality/High-resolution Digital Ultrasound Diagnostic Scanner

High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Hitachi Review Vol. 52 (2003), No. 4 207 High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Tsuyoshi Otake Toshihiko Kawano Takashi Sugiyama Tsuyoshi Mitake Shinichiro Umemura, Ph. D. OVERVIEW:

More information

A wave lab inside a coaxial cable

A wave lab inside a coaxial cable INSTITUTE OF PHYSICS PUBLISHING Eur. J. Phys. 25 (2004) 581 591 EUROPEAN JOURNAL OF PHYSICS PII: S0143-0807(04)76273-X A wave lab inside a coaxial cable JoãoMSerra,MiguelCBrito,JMaiaAlves and A M Vallera

More information

Maximum Likelihood Estimation of ADC Parameters from Sine Wave Test Data. László Balogh, Balázs Fodor, Attila Sárhegyi, and István Kollár

Maximum Likelihood Estimation of ADC Parameters from Sine Wave Test Data. László Balogh, Balázs Fodor, Attila Sárhegyi, and István Kollár Maximum Lielihood Estimation of ADC Parameters from Sine Wave Test Data László Balogh, Balázs Fodor, Attila Sárhegyi, and István Kollár Dept. of Measurement and Information Systems Budapest University

More information

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1)

X X X a) perfect linear correlation b) no correlation c) positive correlation (r = 1) (r = 0) (0 < r < 1) CORRELATION AND REGRESSION / 47 CHAPTER EIGHT CORRELATION AND REGRESSION Correlation and regression are statistical methods that are commonly used in the medical literature to compare two or more variables.

More information

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi Diffusione e perfusione in risonanza magnetica E. Pagani, M. Filippi DW-MRI DIFFUSION-WEIGHTED MRI Principles Diffusion results from a microspic random motion known as Brownian motion THE RANDOM WALK How

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information

ISTITUTO NAZIONALE DI FISICA NUCLEARE

ISTITUTO NAZIONALE DI FISICA NUCLEARE ISTITUTO NAZIONALE DI FISICA NUCLEARE Sezione di Trieste INFN/TC-2/7 1 aprile 22 A novel scheme for the integrated voltage divider of Silicon Drift Detectors P. Burger 1, C. Piemonte 2, A. Rashevsky 3,

More information

Understanding CIC Compensation Filters

Understanding CIC Compensation Filters Understanding CIC Compensation Filters April 2007, ver. 1.0 Application Note 455 Introduction f The cascaded integrator-comb (CIC) filter is a class of hardware-efficient linear phase finite impulse response

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University

More information

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few

More information

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment Fluid Structure Interaction VI 3 Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment J. Hengstler & J. Dual Department of Mechanical and Process

More information

Low-resolution Character Recognition by Video-based Super-resolution

Low-resolution Character Recognition by Video-based Super-resolution 2009 10th International Conference on Document Analysis and Recognition Low-resolution Character Recognition by Video-based Super-resolution Ataru Ohkura 1, Daisuke Deguchi 1, Tomokazu Takahashi 2, Ichiro

More information

A Reliability Point and Kalman Filter-based Vehicle Tracking Technique

A Reliability Point and Kalman Filter-based Vehicle Tracking Technique A Reliability Point and Kalman Filter-based Vehicle Tracing Technique Soo Siang Teoh and Thomas Bräunl Abstract This paper introduces a technique for tracing the movement of vehicles in consecutive video

More information

Implementation of Modified Booth Algorithm (Radix 4) and its Comparison with Booth Algorithm (Radix-2)

Implementation of Modified Booth Algorithm (Radix 4) and its Comparison with Booth Algorithm (Radix-2) Advance in Electronic and Electric Engineering. ISSN 2231-1297, Volume 3, Number 6 (2013), pp. 683-690 Research India Publications http://www.ripublication.com/aeee.htm Implementation of Modified Booth

More information

Measurement, Modeling and Simulation of Power Line Channel for Indoor High-speed Data Communications

Measurement, Modeling and Simulation of Power Line Channel for Indoor High-speed Data Communications Measurement, Modeling and Simulation of Power Line Channel for Indoor High-speed Data Communications Jong-ho Lee, Ji-hoon Park', Hyun-Suk Lee, Gi-Won Leett and Seong-cheol Kim School of Electrical and

More information

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ. Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ., Raleigh, NC One vital step is to choose a transfer lens matched to your

More information

Super-resolution method based on edge feature for high resolution imaging

Super-resolution method based on edge feature for high resolution imaging Science Journal of Circuits, Systems and Signal Processing 2014; 3(6-1): 24-29 Published online December 26, 2014 (http://www.sciencepublishinggroup.com/j/cssp) doi: 10.11648/j.cssp.s.2014030601.14 ISSN:

More information

Chapter 29 Scale-Free Network Topologies with Clustering Similar to Online Social Networks

Chapter 29 Scale-Free Network Topologies with Clustering Similar to Online Social Networks Chapter 29 Scale-Free Network Topologies with Clustering Similar to Online Social Networks Imre Varga Abstract In this paper I propose a novel method to model real online social networks where the growing

More information

A New Programmable RF System for System-on-Chip Applications

A New Programmable RF System for System-on-Chip Applications Vol. 6, o., April, 011 A ew Programmable RF System for System-on-Chip Applications Jee-Youl Ryu 1, Sung-Woo Kim 1, Jung-Hun Lee 1, Seung-Hun Park 1, and Deock-Ho Ha 1 1 Dept. of Information and Communications

More information

REAL TIME TRAFFIC LIGHT CONTROL USING IMAGE PROCESSING

REAL TIME TRAFFIC LIGHT CONTROL USING IMAGE PROCESSING REAL TIME TRAFFIC LIGHT CONTROL USING IMAGE PROCESSING Ms.PALLAVI CHOUDEKAR Ajay Kumar Garg Engineering College, Department of electrical and electronics Ms.SAYANTI BANERJEE Ajay Kumar Garg Engineering

More information

A STUDY OF ECHO IN VOIP SYSTEMS AND SYNCHRONOUS CONVERGENCE OF

A STUDY OF ECHO IN VOIP SYSTEMS AND SYNCHRONOUS CONVERGENCE OF A STUDY OF ECHO IN VOIP SYSTEMS AND SYNCHRONOUS CONVERGENCE OF THE µ-law PNLMS ALGORITHM Laura Mintandjian and Patrick A. Naylor 2 TSS Departement, Nortel Parc d activites de Chateaufort, 78 Chateaufort-France

More information

Encoded Phased Array Bridge Pin Inspection

Encoded Phased Array Bridge Pin Inspection Encoded Phased Array Bridge Pin Inspection James S. Doyle Baker Testing Services, Inc. 22 Reservoir Park Dr. Rockland, MA 02370 (781) 871-4458; fax (781) 871-0123; e-mail jdoyle@bakertesting.com Product

More information

Measuring Line Edge Roughness: Fluctuations in Uncertainty

Measuring Line Edge Roughness: Fluctuations in Uncertainty Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as

More information

Force/position control of a robotic system for transcranial magnetic stimulation

Force/position control of a robotic system for transcranial magnetic stimulation Force/position control of a robotic system for transcranial magnetic stimulation W.N. Wan Zakaria School of Mechanical and System Engineering Newcastle University Abstract To develop a force control scheme

More information

Using Linear Fractal Interpolation Functions to Compress Video. The paper in this appendix was presented at the Fractals in Engineering '94

Using Linear Fractal Interpolation Functions to Compress Video. The paper in this appendix was presented at the Fractals in Engineering '94 Appendix F Using Linear Fractal Interpolation Functions to Compress Video Images The paper in this appendix was presented at the Fractals in Engineering '94 Conference which was held in the École Polytechnic,

More information

INTERFERENCE OF SOUND WAVES

INTERFERENCE OF SOUND WAVES 1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.

More information

Non-Data Aided Carrier Offset Compensation for SDR Implementation

Non-Data Aided Carrier Offset Compensation for SDR Implementation Non-Data Aided Carrier Offset Compensation for SDR Implementation Anders Riis Jensen 1, Niels Terp Kjeldgaard Jørgensen 1 Kim Laugesen 1, Yannick Le Moullec 1,2 1 Department of Electronic Systems, 2 Center

More information

Tracking Moving Objects In Video Sequences Yiwei Wang, Robert E. Van Dyck, and John F. Doherty Department of Electrical Engineering The Pennsylvania State University University Park, PA16802 Abstract{Object

More information

Visualization of large data sets using MDS combined with LVQ.

Visualization of large data sets using MDS combined with LVQ. Visualization of large data sets using MDS combined with LVQ. Antoine Naud and Włodzisław Duch Department of Informatics, Nicholas Copernicus University, Grudziądzka 5, 87-100 Toruń, Poland. www.phys.uni.torun.pl/kmk

More information

HSI BASED COLOUR IMAGE EQUALIZATION USING ITERATIVE n th ROOT AND n th POWER

HSI BASED COLOUR IMAGE EQUALIZATION USING ITERATIVE n th ROOT AND n th POWER HSI BASED COLOUR IMAGE EQUALIZATION USING ITERATIVE n th ROOT AND n th POWER Gholamreza Anbarjafari icv Group, IMS Lab, Institute of Technology, University of Tartu, Tartu 50411, Estonia sjafari@ut.ee

More information

2.500 Threshold. 2.000 1000e - 001. Threshold. Exponential phase. Cycle Number

2.500 Threshold. 2.000 1000e - 001. Threshold. Exponential phase. Cycle Number application note Real-Time PCR: Understanding C T Real-Time PCR: Understanding C T 4.500 3.500 1000e + 001 4.000 3.000 1000e + 000 3.500 2.500 Threshold 3.000 2.000 1000e - 001 Rn 2500 Rn 1500 Rn 2000

More information

Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network

Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network Recent Advances in Electrical Engineering and Electronic Devices Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network Ahmed El-Mahdy and Ahmed Walid Faculty of Information Engineering

More information

Application Note Noise Frequently Asked Questions

Application Note Noise Frequently Asked Questions : What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random

More information

DYNAMIC DOMAIN CLASSIFICATION FOR FRACTAL IMAGE COMPRESSION

DYNAMIC DOMAIN CLASSIFICATION FOR FRACTAL IMAGE COMPRESSION DYNAMIC DOMAIN CLASSIFICATION FOR FRACTAL IMAGE COMPRESSION K. Revathy 1 & M. Jayamohan 2 Department of Computer Science, University of Kerala, Thiruvananthapuram, Kerala, India 1 revathysrp@gmail.com

More information

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,

More information

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of

More information

Microcontroller-based experiments for a control systems course in electrical engineering technology

Microcontroller-based experiments for a control systems course in electrical engineering technology Microcontroller-based experiments for a control systems course in electrical engineering technology Albert Lozano-Nieto Penn State University, Wilkes-Barre Campus, Lehman, PA, USA E-mail: AXL17@psu.edu

More information

EDUMECH Mechatronic Instructional Systems. Ball on Beam System

EDUMECH Mechatronic Instructional Systems. Ball on Beam System EDUMECH Mechatronic Instructional Systems Ball on Beam System Product of Shandor Motion Systems Written by Robert Hirsch Ph.D. 998-9 All Rights Reserved. 999 Shandor Motion Systems, Ball on Beam Instructional

More information

Analysis of Wing Leading Edge Data. Upender K. Kaul Intelligent Systems Division NASA Ames Research Center Moffett Field, CA 94035

Analysis of Wing Leading Edge Data. Upender K. Kaul Intelligent Systems Division NASA Ames Research Center Moffett Field, CA 94035 Analysis of Wing Leading Edge Data Upender K. Kaul Intelligent Systems Division NASA Ames Research Center Moffett Field, CA 94035 Abstract Some selected segments of the STS114 ascent and on-orbit data

More information

Introduction to acoustic imaging

Introduction to acoustic imaging Introduction to acoustic imaging Contents 1 Propagation of acoustic waves 3 1.1 Wave types.......................................... 3 1.2 Mathematical formulation.................................. 4 1.3

More information

Chapter 10. Key Ideas Correlation, Correlation Coefficient (r),

Chapter 10. Key Ideas Correlation, Correlation Coefficient (r), Chapter 0 Key Ideas Correlation, Correlation Coefficient (r), Section 0-: Overview We have already explored the basics of describing single variable data sets. However, when two quantitative variables

More information

NOVEL PRIORITISED EGPRS MEDIUM ACCESS REGIME FOR REDUCED FILE TRANSFER DELAY DURING CONGESTED PERIODS

NOVEL PRIORITISED EGPRS MEDIUM ACCESS REGIME FOR REDUCED FILE TRANSFER DELAY DURING CONGESTED PERIODS NOVEL PRIORITISED EGPRS MEDIUM ACCESS REGIME FOR REDUCED FILE TRANSFER DELAY DURING CONGESTED PERIODS D. Todinca, P. Perry and J. Murphy Dublin City University, Ireland ABSTRACT The goal of this paper

More information

DVB-T BER MEASUREMENTS IN THE PRESENCE OF ADJACENT CHANNEL AND CO-CHANNEL ANALOGUE TELEVISION INTERFERENCE

DVB-T BER MEASUREMENTS IN THE PRESENCE OF ADJACENT CHANNEL AND CO-CHANNEL ANALOGUE TELEVISION INTERFERENCE DVB-T MEASUREMENTS IN THE PRESENCE OF ADJACENT CHANNEL AND CO-CHANNEL ANALOGUE TELEVISION INTERFERENCE M. Mª Vélez (jtpveelm@bi.ehu.es), P. Angueira, D. de la Vega, A. Arrinda, J. L. Ordiales UNIVERSITY

More information

Automated Process for Generating Digitised Maps through GPS Data Compression

Automated Process for Generating Digitised Maps through GPS Data Compression Automated Process for Generating Digitised Maps through GPS Data Compression Stewart Worrall and Eduardo Nebot University of Sydney, Australia {s.worrall, e.nebot}@acfr.usyd.edu.au Abstract This paper

More information

Efficient Data Recovery scheme in PTS-Based OFDM systems with MATRIX Formulation

Efficient Data Recovery scheme in PTS-Based OFDM systems with MATRIX Formulation Efficient Data Recovery scheme in PTS-Based OFDM systems with MATRIX Formulation Sunil Karthick.M PG Scholar Department of ECE Kongu Engineering College Perundurau-638052 Venkatachalam.S Assistant Professor

More information

BMOA: Binary Magnetic Optimization Algorithm

BMOA: Binary Magnetic Optimization Algorithm International Journal of Machine Learning and Computing Vol. 2 No. 3 June 22 BMOA: Binary Magnetic Optimization Algorithm SeyedAli Mirjalili and Siti Zaiton Mohd Hashim Abstract Recently the behavior of

More information

STATISTICS AND DATA ANALYSIS IN GEOLOGY, 3rd ed. Clarificationof zonationprocedure described onpp. 238-239

STATISTICS AND DATA ANALYSIS IN GEOLOGY, 3rd ed. Clarificationof zonationprocedure described onpp. 238-239 STATISTICS AND DATA ANALYSIS IN GEOLOGY, 3rd ed. by John C. Davis Clarificationof zonationprocedure described onpp. 38-39 Because the notation used in this section (Eqs. 4.8 through 4.84) is inconsistent

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

SOFTWARE FOR GENERATION OF SPECTRUM COMPATIBLE TIME HISTORY

SOFTWARE FOR GENERATION OF SPECTRUM COMPATIBLE TIME HISTORY 3 th World Conference on Earthquake Engineering Vancouver, B.C., Canada August -6, 24 Paper No. 296 SOFTWARE FOR GENERATION OF SPECTRUM COMPATIBLE TIME HISTORY ASHOK KUMAR SUMMARY One of the important

More information

Real-time PCR: Understanding C t

Real-time PCR: Understanding C t APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence

More information

Face detection is a process of localizing and extracting the face region from the

Face detection is a process of localizing and extracting the face region from the Chapter 4 FACE NORMALIZATION 4.1 INTRODUCTION Face detection is a process of localizing and extracting the face region from the background. The detected face varies in rotation, brightness, size, etc.

More information

Machine Learning for Medical Image Analysis. A. Criminisi & the InnerEye team @ MSRC

Machine Learning for Medical Image Analysis. A. Criminisi & the InnerEye team @ MSRC Machine Learning for Medical Image Analysis A. Criminisi & the InnerEye team @ MSRC Medical image analysis the goal Automatic, semantic analysis and quantification of what observed in medical scans Brain

More information

Current control of piezoelectric actuators with power loss compensation

Current control of piezoelectric actuators with power loss compensation Tampere University of Technology Author(s) Title Citation onkanen, Pekka; Kallio, Pasi; Koivo, Heikki Current control of piezoelectric actuators with power loss compensation onkanen, Pekka; Kallio, Pasi;

More information

Lossless Medical Image Compression using Redundancy Analysis

Lossless Medical Image Compression using Redundancy Analysis 5 Lossless Medical Image Compression using Redundancy Analysis Se-Kee Kil, Jong-Shill Lee, Dong-Fan Shen, Je-Goon Ryu, Eung-Hyuk Lee, Hong-Ki Min, Seung-Hong Hong Dept. of Electronic Eng., Inha University,

More information

How To Test A Robot Platform And Its Components

How To Test A Robot Platform And Its Components An Automated Test Method for Robot Platform and Its Components Jae-Hee Lim 1, Suk-Hoon Song 1, Jung-Rye Son 1, Tae-Yong Kuc 2, Hong-Seong Park 3, Hong-Seok Kim 4 1,2 School of Information and Communication,

More information

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough

More information

Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging.

Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging. Beamforming and hardware design for a multichannel front-end integrated circuit for real-time 3D catheter-based ultrasonic imaging. Ira O. Wygant a, Mustafa Karaman b, Ömer Oralkana, and Butrus T. Khuri-Yakub

More information

POTENTIAL OF STATE-FEEDBACK CONTROL FOR MACHINE TOOLS DRIVES

POTENTIAL OF STATE-FEEDBACK CONTROL FOR MACHINE TOOLS DRIVES POTENTIAL OF STATE-FEEDBACK CONTROL FOR MACHINE TOOLS DRIVES L. Novotny 1, P. Strakos 1, J. Vesely 1, A. Dietmair 2 1 Research Center of Manufacturing Technology, CTU in Prague, Czech Republic 2 SW, Universität

More information

2695 P a g e. IV Semester M.Tech (DCN) SJCIT Chickballapur Karnataka India

2695 P a g e. IV Semester M.Tech (DCN) SJCIT Chickballapur Karnataka India Integrity Preservation and Privacy Protection for Digital Medical Images M.Krishna Rani Dr.S.Bhargavi IV Semester M.Tech (DCN) SJCIT Chickballapur Karnataka India Abstract- In medical treatments, the integrity

More information

Nano Meter Stepping Drive of Surface Acoustic Wave Motor

Nano Meter Stepping Drive of Surface Acoustic Wave Motor Proc. of 1st IEEE Conf. on Nanotechnology, Oct. 28-3, pp. 495-5, (21) Maui, Hawaii Nano Meter Stepping Drive of Surface Acoustic Wave Motor Takashi Shigematsu*, Minoru Kuribayashi Kurosawa*, and Katsuhiko

More information

How To Filter Spam Image From A Picture By Color Or Color

How To Filter Spam Image From A Picture By Color Or Color Image Content-Based Email Spam Image Filtering Jianyi Wang and Kazuki Katagishi Abstract With the population of Internet around the world, email has become one of the main methods of communication among

More information

Segmentation and Automatic Descreening of Scanned Documents

Segmentation and Automatic Descreening of Scanned Documents Segmentation and Automatic Descreening of Scanned Documents Alejandro Jaimes a, Frederick Mintzer b, A. Ravishankar Rao b and Gerhard Thompson b a Columbia University b IBM T.J. Watson Research Center

More information

Segmentation of Breast Tumor from Mammographic Images Using Histogram Peak Slicing Threshold

Segmentation of Breast Tumor from Mammographic Images Using Histogram Peak Slicing Threshold International Journal of Computer Sciences and Engineering Open Access Research Paper Volume-4, Special Issue-1 E-ISSN: 2347-2693 Segmentation of Breast Tumor from Mammographic Images Using Histogram Peak

More information

ECE 533 Project Report Ashish Dhawan Aditi R. Ganesan

ECE 533 Project Report Ashish Dhawan Aditi R. Ganesan Handwritten Signature Verification ECE 533 Project Report by Ashish Dhawan Aditi R. Ganesan Contents 1. Abstract 3. 2. Introduction 4. 3. Approach 6. 4. Pre-processing 8. 5. Feature Extraction 9. 6. Verification

More information

Multimodal Biometric Recognition Security System

Multimodal Biometric Recognition Security System Multimodal Biometric Recognition Security System Anju.M.I, G.Sheeba, G.Sivakami, Monica.J, Savithri.M Department of ECE, New Prince Shri Bhavani College of Engg. & Tech., Chennai, India ABSTRACT: Security

More information

An Ecient Dynamic Load Balancing using the Dimension Exchange. Ju-wook Jang. of balancing load among processors, most of the realworld

An Ecient Dynamic Load Balancing using the Dimension Exchange. Ju-wook Jang. of balancing load among processors, most of the realworld An Ecient Dynamic Load Balancing using the Dimension Exchange Method for Balancing of Quantized Loads on Hypercube Multiprocessors * Hwakyung Rim Dept. of Computer Science Seoul Korea 11-74 ackyung@arqlab1.sogang.ac.kr

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

High Quality Image Magnification using Cross-Scale Self-Similarity

High Quality Image Magnification using Cross-Scale Self-Similarity High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg

More information

INTRODUCTION. The principles of which are to:

INTRODUCTION. The principles of which are to: Taking the Pain Out of Chromatographic Peak Integration Shaun Quinn, 1 Peter Sauter, 1 Andreas Brunner, 1 Shawn Anderson, 2 Fraser McLeod 1 1 Dionex Corporation, Germering, Germany; 2 Dionex Corporation,

More information

Design and Simulation of Soft Switched Converter Fed DC Servo Drive

Design and Simulation of Soft Switched Converter Fed DC Servo Drive International Journal of Soft Computing and Engineering (IJSCE) ISSN: 2231-237, Volume-1, Issue-5, November 211 Design and Simulation of Soft Switched Converter Fed DC Servo Drive Bal Mukund Sharma, A.

More information

Influence of Load Balancing on Quality of Real Time Data Transmission*

Influence of Load Balancing on Quality of Real Time Data Transmission* SERBIAN JOURNAL OF ELECTRICAL ENGINEERING Vol. 6, No. 3, December 2009, 515-524 UDK: 004.738.2 Influence of Load Balancing on Quality of Real Time Data Transmission* Nataša Maksić 1,a, Petar Knežević 2,

More information

Instruction Manual Service Program ULTRA-PROG-IR

Instruction Manual Service Program ULTRA-PROG-IR Instruction Manual Service Program ULTRA-PROG-IR Parameterizing Software for Ultrasonic Sensors with Infrared Interface Contents 1 Installation of the Software ULTRA-PROG-IR... 4 1.1 System Requirements...

More information

Medical imaging monitors specification guidelines

Medical imaging monitors specification guidelines Medical imaging monitors specification guidelines Document details Contact for enquiries and proposed changes If you have any questions regarding this document or if you have a suggestion for improvements,

More information

Step Response of RC Circuits

Step Response of RC Circuits Step Response of RC Circuits 1. OBJECTIVES...2 2. REFERENCE...2 3. CIRCUITS...2 4. COMPONENTS AND SPECIFICATIONS...3 QUANTITY...3 DESCRIPTION...3 COMMENTS...3 5. DISCUSSION...3 5.1 SOURCE RESISTANCE...3

More information

SIGMOD RWE Review Towards Proximity Pattern Mining in Large Graphs

SIGMOD RWE Review Towards Proximity Pattern Mining in Large Graphs SIGMOD RWE Review Towards Proximity Pattern Mining in Large Graphs Fabian Hueske, TU Berlin June 26, 21 1 Review This document is a review report on the paper Towards Proximity Pattern Mining in Large

More information