The Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) by LC/MS/MS Using an Atmospheric Pressure Photoionization Source

Size: px
Start display at page:

Download "The Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) by LC/MS/MS Using an Atmospheric Pressure Photoionization Source"

Transcription

1 Application Note Mass Spectrometry The Analysis of Polycyclic Aromatic Hydrocarbons (PAHs) by LC/MS/MS Using an Atmospheric Pressure Photoionization Source The atmospheric pressure photoionization source, PhotoSpray ion source (Robb, Covey and Bruins 1 ) uses photons to ionize large quantities of a dopant molecule added along with the vaporized mobile phase. Analyte molecules are efficiently ionized through secondary reactions initiated by the charged dopant. This new technique was used to develop an analytical method for polycyclic aromatic hydrocarbons (PAHs), comparing the difference under reverse phase and normal phase chromatographic conditions. A total of 16 PAHs were analysed including the determination of unknown quantities in atmospheric samples from both rural and urban areas. Introduction Polycyclic aromatic hydrocarbons (PAHs) are a group of over 100 different chemicals, typically formed during incomplete combustion, which are present in a variety of environmental matrices. Airborne PAHs, usually found in aerosol particles, are implicated in heterogeneous atmospheric processing and as important indicators of gas-phase atmospheric processes. Certain carcinogenic PAHs may be carried by ultrafine particles in ambient air into the lungs, and are often the targets of rigorous environmental monitoring. Currently, environmental PAHs are most commonly analysed by HPLC or GC/EI-MS; techniques which require long, high-resolution Figure 1. PAH structures. separations, and complicated MS methods. Atmospheric pressure photoionization is a new technique that uses the energy of photons to generate ions from the vaporized eluent of an LC. This study compares the analysis of PAHs by LC/MS/MS using both reverse phase and normal phase chromatographic conditions. Materials and methods A 20 ppm stock standard solution containing 16 common PAHs (see Figure 1). Under optimized LC/MS/MS conditions for both reversed phase and normal phase chromatographic conditions, the performance, linearity and LOD levels were determined. The most abundant single positive transition for each PAH compound was used to construct an +MRM method. Concentrations from ambient air for several PAHs were determined from samples collected on high volume filters at both urban and rural sites. Standard solutions were prepared by serial dilution in acetonitrile/water and iso-octane for reversed phase and normal phase conditions, respectively. Spectrometer: API 3000 LC/MS/MS System with Analyst Software PhotoSpray Source (see Figure 2 and 3) HPLC: Perkin-Elmer 200 Series Autosampler and Micro pumps

2 Figure 3. PhotoSpray source. LC conditions: Acetonitrile/water (gradient mode) flow at 1.0 ml/min Figure 2. Schematic diagram of the PhotoSpray source Column: Hypersil PAH, 100 x 4.6 mm; 5 µm, 50:50 post column split Normal phase conditions: 100% iso-octane (isocratic mode) flow at 0.5 ml/min Column: Spherisorb Silica, 250 x 4.6 mm; 5 µm Dopant: Toluene at a flowrate of 100 µl/min. Results Figure 4. Reversed phase chromatography. The source uses a modified housing of a conventional APCI source. The corona discharge needle has been removed and replaced with a connection for the lamp gas (nitrogen) which protects the Magnesium Fluoride optical window from the vaporised mobile phase. Reversed phase PAH separation was accomplished by gradient flow (ACN/H2O) through a Hypersil PAH column, 100 x 4.6 mm; 5 µm. Individual MRM transitions were monitored simultaneously for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 4 is a 20 µl injection of a 5 ng/µl solution (100 ng on column). Figure 5. Normal phase chromatography.

3 Normal phase PAH separation was accomplished with 100% iso-octane through a Spherisorb Silica column, 250 x 4.6 mm; 5 µm. Individual MRM transitions were monitored for 16 common PAHs (see Figure 1 for corresponding PAH). Figure 5 is a 20 µl injection of a 5 ng/µl solution (100 ng on column). The observed response for PAHs under normal phase conditions show an increase on the order of 10 to 70 times of that observed under reversed phase conditions (Figure 6). Figure 6. Response comparison for normal phase and reversed phase conditions. The post-column addition of dopant flow can have an effect on the overall response. This will vary from compound to compound. As shown in Figure 7, the response of naphthalene is increased by a factor of 7, while the response of benzo(a)pyrene is relatively unchanged. Under normal phase conditions, 5000 pg/µl fell out of the calibration range for most PAHs, except Indeno(1,2,3,c,d) pyrene and Dibenz(a,h)anthracene. Reversed phase conditions could be calibrated as high as 5000 pg/µl (Figure 8). Figure 7. Effect of dopant on response. Ambient air samples were collected on high volume filters(subsequently solvent extracted, dried to 0.2 ml) from three locations, rural, urban and a traffic tunnel. Sample chromatogramsare shown in Figure 9, and calculated ambient concentrations are shown in Table 2. Figure 8. PAH calibrations under normal phase conditions.

4 Conclusions The new PhotoSpray source provides a sensitive ionization technique for analysis of polycyclic aromatic hydrocarbons by LC/MS/MS. Normal phase conditions provide an increase in sensitivity by a factor of at least 10 and as much as 70 over reversed phase conditions. The APPI LC/MS/MS normal phase method provides adequate sensitivity to be useful in the analysis of atmospheric aerosol samples for PAHs. Table 1. Limits of detection for 16 common PAHs under normal phase and reversed phase conditions. References 1 Robb, D. B., Covey, T. R., Bruins, A. P., Anal. Chem. 2000, 72, Authors Gary Impey, Byron Kieser, Jean-François Alary, Applied Biosystems/MDS SCIEX Table 2. Ambient concentrations at three sites: rural, urban, and traffic tunnel. *Not shown in Figure 9. Acknowledgements The authors wish to thank Satoshi Irei (York University, Toronto, Canada) for supplying PAH standards and the ambient air samples for analysis.

5 iscience. To better understand the complex interaction of biological systems, life scientists are developing revolutionary approaches to discovery that unite technology, informatics, and traditional laboratory research. In partnership with our customers, Applied Biosystems provides the innovative products, services, and knowledge resources that make this new, Integrated Science possible. Worldwide Sales Offices Applied Biosystems vast distribution and service network, composed of highly trained support and applications personnel, reaches 150 countries on six continents. For international office locations, please call the division headquarters or refer to our Web site at Applera is committed to providing the world s leading technology and information for life scientists. Applera Corporation consists of the Applied Biosystems and Celera Genomics businesses. Applied Biosystems/MDS SCIEX is a joint venture between Applera Corporation and MDS Inc. Headquarters 850 Lincoln Centre Drive Foster City, CA USA Phone: Toll Free: Fax: For Research Use Only. Not for use in diagnostic procedures. Analyst, Applied Biosystems, and QSTAR are registered trademarks and AB (Design), iscience and iscience (Design), API 2000, API 3000, API 4000, and Applera are trademarks of Applera Corporation or its subsidiaries in the US and/or certain other countries. Figure 9. Comparison of ambient air samples at three sites: rural, urban and traffic tunnel. PhotoSpray is a trademark and Q TRAP is a registered trademark of Applied Biosystems/MDS Sciex, a joint venture between Applera Corporation and MDS, Inc. All other trademarks are the sole property of their respective owners Applied Biosystems. All Rights Reserved. Information subject to change without notice. Printed in the USA, 10/04, P+s, Publication 114AP22-01

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION

Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,

More information

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,

More information

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined

API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation

More information

Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS

Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS Application note Explosives by Direct Injection LC/MS/MS Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS API 5000 LC/MS/MS System

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.

Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories. Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine

More information

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:

More information

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification

Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment

More information

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water Chen Jing, 1 Dai Zhenyu, 1 Xu Qun, 1 Liang Lina, 1 and Jeffrey Rohrer 2 1 Thermo Fisher Scientific, Shanghai, People s

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird

More information

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process Save Hours in Sample Preparation Reduce the time to prepare and inject samples

More information

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs

More information

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers

More information

LC-MS/MS for Chromatographers

LC-MS/MS for Chromatographers LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,

More information

High Sensitivity MS Determination of Carryover in a New Autosampler Design

High Sensitivity MS Determination of Carryover in a New Autosampler Design High Sensitivity MS Determination of Carryover in a New Autosampler Design 1 William Hedgepeth*, Sue Steinike Shimadzu Scientific Instruments, Columbia, MD USA Mauro Aiello, Robert Ellis, Andre Schreiber,

More information

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2

Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2 Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2 1 OpAns LLC, Durham, NC, USA; 2 Ionics Mass Spectrometry

More information

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen. Mark Crawford Single Presenter Name Abstract With the recent decline in hive bee populations this year the investigation of acaricides

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex

More information

Supplementary Materials and Methods (Metabolomics analysis)

Supplementary Materials and Methods (Metabolomics analysis) Supplementary Materials and Methods (Metabolomics analysis) Metabolite extraction and mass spectrometry analysis. 12 Vldlr-/- knock out and 12 wild type (WT) retinas were separately frozen at -80 ºC in

More information

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,

More information

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

Creating Standard Curves with Genomic DNA or Plasmid DNA Templates for Use in Quantitative PCR

Creating Standard Curves with Genomic DNA or Plasmid DNA Templates for Use in Quantitative PCR Creating Standard Curves with Genomic DNA or Plasmid DNA Templates for Use in Quantitative PCR Overview Genomic DNA (gdna) and plasmids containing cloned target sequences are commonly used as standards

More information

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines

More information

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische

More information

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10

More information

How To Test For Contamination In Large Volume Water

How To Test For Contamination In Large Volume Water Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

Waters Core Chromatography Training (2 Days)

Waters Core Chromatography Training (2 Days) 2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.

More information

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites Carmen T. Santasania and David S. Bell SUPELCO, 595 North Harrison Road, Bellefonte, PA 16823 T404045 GUS Abstract

More information

Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report

Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report PRODUCT BULLETIN Cliquid ChemoView 3.0 Software for Routine Screening and Quantitation Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report KEY FEATURES Secure user login that

More information

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants Metabolomic Profiling of Accurate Mass LC- Data to Identify Unexpected Environmental Pollutants André Schreiber 1, David Cox 1, Nadia Pace 1, Christopher Borton 2 1 AB SCIEX, Concord, ntario, Canada; 2

More information

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental

More information

WATERS QUANTITATIVE ANALYSIS solutions

WATERS QUANTITATIVE ANALYSIS solutions More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

The Use of Hydrogen Carrier Gas for GC/MS

The Use of Hydrogen Carrier Gas for GC/MS Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors

More information

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests

Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Application Note: 386 Advantages of the LTQ Orbitrap for Protein Identification in Complex Digests Rosa Viner, Terry Zhang, Scott Peterman, and Vlad Zabrouskov, Thermo Fisher Scientific, San Jose, CA,

More information

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates

In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics

More information

Purification of reaction mixtures using flash chromatography.

Purification of reaction mixtures using flash chromatography. Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?

More information

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative

More information

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK Background Unilabs YBS are a bioanalytical CRO based in York (Uk) Copenhagen

More information

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application

More information

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection Derek Hillbeck, Thermo Fisher Scientific, Runcorn, UK Application Note 2999 Key

More information

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler

More information

METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS

METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS 1.0 SCOPE AND APPLICATION 1.1 Method 8310 is used to determine the concentration of certain polynuclear aromatic hydrocarbons (PAH) in ground water and wastes.

More information

Chemical analysis service, Turner s Green Technology Group

Chemical analysis service, Turner s Green Technology Group Chemical analysis service, Turner s Green Technology Group Hourly costs, Academic collaborations leading to co-publications: 627 SEK/hr (excl. VAT) Hourly costs, Industry/Academic collaborations, no publications:

More information

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs

AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF

More information

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access

More information

Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants

Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants Version 2004-01 / August 2004 (HGRS-JW-04-25s) HGRS-CTS/MT J. Waltisberg SUMMARY ( ) shall not

More information

SERVICE PLANS. More Productivity. Less Worry. SERVICE PLANS

SERVICE PLANS. More Productivity. Less Worry. SERVICE PLANS SERVICE PLANS More Productivity. Less Worry. SERVICE PLANS Flexible service plans to help you maximize the value of your technology investment. Let us help you improve your laboratory productivity and

More information

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis

Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** mariadellarcopassaro@gmail.com

More information

Overview. Purpose. Methods. Results

Overview. Purpose. Methods. Results A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.

More information

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM

DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural

More information

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:

More information

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification

MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can

More information

CORESTA RECOMMENDED METHOD N 72

CORESTA RECOMMENDED METHOD N 72 CORESTA RECOMMENDED METHOD N 72 DETERMINATION OF TOBACCO-SPECIFIC NITROSAMINES IN SMOKELESS TOBACCO PRODUCTS BY LC-MS/MS (February 216). INTRODUCTION The CORESTA Smokeless Tobacco Sub-Group studied various

More information

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France

Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS Mikaël LEVI Shimadzu France Welcome Seminar Barcelona 2013 Interest in Steroids assay by LC-MS/MS Steroids measurement is essential

More information

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory

More information

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Automation of Solid Phase Extraction and Column Chromatographic Cleanup Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for

More information

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas

More information

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring

The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,

More information

Air Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University

Air Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this

More information

Analyzing Small Molecules by EI and GC-MS. July 2014

Analyzing Small Molecules by EI and GC-MS. July 2014 Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,

More information

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities ICMS and LCMS Determination of Ionic Liquids, Counterions, and Impurities Leo Wang, Detlef Jensen, and William Schnute Dionex Corporation, Sunnyvale, CA, USA; Dionex Corporation, Olten, Switzerland INTRODUCTION

More information

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide chromatography Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide The following guide provides considerations, tips and general guidelines for developing SPE methods using the Thermo

More information

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588). User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,

More information

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil. March 2012 8 1 of 17 1.0 SCOPE 1.1 This procedure describes the preparation and analysis of solid matrices and wipe samples. 1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.

More information

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in uman Serum Technical verview Clinical Research Authors Anthony Macherone, Ph.D. Agilent Technologies,

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection

Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT

More information

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate

More information

Calculation of Molar Masses. Molar Mass. Solutions. Solutions

Calculation of Molar Masses. Molar Mass. Solutions. Solutions Molar Mass Molar mass = Mass in grams of one mole of any element, numerically equal to its atomic weight Molar mass of molecules can be determined from the chemical formula and molar masses of elements

More information

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System

Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS

More information

Replacing TaqMan SNP Genotyping Assays that Fail Applied Biosystems Manufacturing Quality Control. Begin

Replacing TaqMan SNP Genotyping Assays that Fail Applied Biosystems Manufacturing Quality Control. Begin User Bulletin TaqMan SNP Genotyping Assays May 2008 SUBJECT: Replacing TaqMan SNP Genotyping Assays that Fail Applied Biosystems Manufacturing Quality Control In This Bulletin Overview This user bulletin

More information

Smart Monitoring Service

Smart Monitoring Service SMART SERVICES Smart Monitoring Service Increase productivity with real-time remote instrument monitoring. Life in the lab just got easier with remote instrument monitoring. When one of your instruments

More information

How To Develop A Recipe Card

How To Develop A Recipe Card TSNA Method Development Dr Michael Intorp CORESTA Cooperation Center for Scientific Research Relative to Tobacco FDA Center for Tobacco Products Scientific Workshop Tobacco Product Analysis 30 31 July

More information

HRMS in Clinical Research: from Targeted Quantification to Metabolomics

HRMS in Clinical Research: from Targeted Quantification to Metabolomics A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier

More information

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water Sjaak de Koning 1, Mark van Lieshout 2, Hans-Gerd Janssen 2 and Wil van Egmond 1 1 ATAS, De Run 4334, PO Box

More information

HPLC Basics, Equipment, and Troubleshooting

HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting HPLC Basics, Equipment, and Troubleshooting explains chromatography in practical terms from the ground up HPLC Basics, Equipment, and Troubleshooting HPLC Basics,

More information

Dionex IonPac AS19-4µm

Dionex IonPac AS19-4µm Total Thermo Scientific Dionex IonPac AS9-4µm Column Product Manual P/N: September 203 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS9-4µm Column Dionex IonPac AG9-4µm Capillary Guard

More information

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform

Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform Charles Yang, Leo Wang, Dipankar Ghosh, Jonathan Beck,

More information

A Complete Solution for Method Linearity in HPLC and UHPLC

A Complete Solution for Method Linearity in HPLC and UHPLC Now sold under the Thermo Scientific brand A Complete Solution for Method Linearity in HPLC and UHPLC Frank Steiner, Fraser McLeod, Tobias Fehrenbach, and Andreas Brunner Dionex Corporation, Germering,

More information

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Analysis of Vegetable Oils by High Performance Liquid Chromatography Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information