1 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 1 of 7 Installation Manual for the SRS 300 Gas Assisted Regeneration Kit

2 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 2 of 7 TABLE OF CONTENTS SECTION 1 - INTRODUCTION... 3 SECTION 2 - INSTALLATION IN THE SRS RECYCLE MODE... 3 SECTION 3 - INSTALLATION IN THE SRS EXTERNAL WATER MODE... 4 SECTION 4 - INSTALLATION WITH ADDITIONAL DETECTORS IN SERIES... 4 SECTION 5 - RECOMMENDED APPLICATIONS... 5 SECTION 6 - OPERATION... 5 SECTION 7 - SOLVENT CONSIDERATIONS... 5 SECTION 8 - GAS FLOW CONTROL ICS and DX Systems DX-120 Systems Gas Requirements... 6 SECTION 9 - SRS 300 GAS ASSISTED REGENERATION KIT COMPONENTS... 7

3 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 3 of 7 SECTION 1 - INTRODUCTION This guide describes the Installation of the SRS 300 Gas Assisted Regeneration Kit (P/N ) for use with the Self- Regenerating Suppressor (SRS) 300. This device can be operated in Gas Assisted Recycle Mode and Gas Assisted External Water Mode. SECTION 2 - INSTALLATION IN THE SRS RECYCLE MODE 2.1 Installation Steps a. Connect the 1/8" REGEN IN line (Shortest tubing extending from the Tee) to the SRS REGEN IN port. b. Connect the line from the Mixing Tee (longer tubing extending from the Tee) to the back pressure coils. c. Start flow through the columns and suppressor (with the current on) until eluent exits the REGEN OUT port. d. Connect a gas line to the Regulator. e. Pressurize the Regulator between 5-15 psi using either Helium or Nitrogen gas. NOTE: The gas flow across the Regenerant chamber in a 4-mm SRS 300 is approximately 100 cc (ml/min) at a set pressure of psi. The gas flow across the Regenerant chamber in a 2-mm SRS 300 is approximately 100 cc (ml/min) at a set pressure of psi. ELUENT-OUT ELUENT-IN From Column SRS-300 Back Pressure Coils (40 psi) REGEN-IN To Waste REGEN-OUT From Helium/Nitrogen Gas Flow, 5-15 psi Conductivity Cell 1/4-28 Mixing Tee Check Valve Figure 1 The Gas Assisted Regeneration Mode Plumbing Diagram

4 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 4 of 7 SECTION 3 - INSTALLATION IN THE SRS EXTERNAL WATER MODE 3.1 Installation Steps a. Connect the 1/8" REGEN IN line (Shortest tubing extending from the Tee) to the SRS REGEN IN port. b. Connect the line from the Mixing Tee (the longer tubing extending from the Tee) to the line from the external water source. A pump capable of delivering 2-3 ml/min is recommended for operation in the external water mode. c. Start flow through the columns and suppressor (with the current on) until water exits the REGEN OUT port. d. Connect a gas line to the Regulator. e. Pressurize the Regulator between 5-15 psi using either Helium or Nitrogen gas. NOTE: The gas flow across the Regenerant chamber in a 4-mm SRS 300 is approximately 100 ml/min at a set pressure of psi. The gas flow across the Regenerant chamber in a 2-mm SRS 300 is approximately 100 ml/min at a set pressure of psi. Conductivity Cell To Waste ELUENT-OUT ELUENT-IN Back Pressure Coils (40 psi) SRS-300 REGEN-IN REGEN-OUT Pump To Waste Regenerant In 1/4-28 Mixing Tee (From Helium/Nitrogen Gas Flow 5-15 psi) Check Valve Figure 2 Plumbing for Gas Assisted Regeneration in the External Water Operation Mode SECTION 4 - INSTALLATION WITH ADDITIONAL DETECTORS IN SERIES The SRS 300 Gas Assisted Regeneration Kit is connected exactly as described above. Place the gas assisted apparatus directly before the REGEN IN port of the SRS 300 allowing normal operation with additional detectors.

5 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 5 of 7 SECTION 5 - RECOMMENDED APPLICATIONS 5.1 Recycle Mode Gas Assisted Regeneration in the Recycle mode is recommended for aqueous applications requiring low noise. This mode of operation replaces the external water mode of operation that uses 5-10 cc (ml/min) of deionized water. 5.2 External Water Mode For solvent applications, Gas Assisted Regeneration in the External Water mode using a pump is recommended. The recommended deionized water flow is approximately 2-3 ml/min for most applications. This replaces the external water mode of operation that uses ml/min of deionized water. SECTION 6 - OPERATION A mixture of gas and liquid will be seen in the short tubing near the REGEN IN port. The gas and liquid should be finely mixed under the above conditions. Due to the high gas velocity, only a trace of liquid should be visible in the waste line. If discrete zones of gas followed by liquid (similar to normal SRS 300 operation) are observed in the waste line, check for high back pressure in the waste line. Replace the waste line if constriction or back pressure is observed. The waste line should be a 1/8" line that is free from bends and constrictions. The recommended SRS 300 Back Pressure Coils should remain in place after the Conductivity Cell. SECTION 7 - SOLVENT CONSIDERATIONS When operating with eluents containing solvents, the regenerant chamber may develop higher back pressure. Under this condition, increase the gas pressure to achieve higher gas flow rate to allow for complete mixing of gas and liquid.

6 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 6 of 7 SECTION 8 - GAS FLOW CONTROL 8.1 ICS-300, ICS-2500, ICS-2000, ICS-1500, ICS-1000, and DX 500, DX 600 Systems The following items are recommended for operation with DX 500 and DX 600 systems to automatically turn ON and OFF the flow of gas. AC2 Power Control Accessory P/N CAM Controlled Air Module, 115V P/N The AC2 module controls the CAM module, turning the gas ON and OFF. 8.2 DX 120 Systems The following items are recommended for operation with DX 120 systems to automatically turn ON and OFF the flow of gas. CAM Controlled Air Module, 115 V P/N Accessory Power Adapter Cable P/N The cable connects the CAM directly to the AC output on the back of the DX 120, which is activated through a solid state relay. 8.3 Gas Requirements It is important to turn off the suppressor current when the analytical pump is off; otherwise, the presence of gas flow will cause a high voltage alarm and in some instances will damage the suppressor device. In order to reduce gas waste, turn off the gas flow when the system is not operational. WARNING: Nitrogen, helium or argon gases are recommended for the above application. Compressed air is not recommended for the above application due to potential contamination. When using in-house generated gases, suitable traps to remove moisture, hydrocarbons, carbon dioxide and other particulate matter are recommended.

7 Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 7 of 7 SECTION 9 - SRS 300 GAS ASSISTED REGENERATION KIT COMPONENTS Regulator Connector to Cell Effluent Check Valve Connector to SRS REGEN-IN Port Gas Flow Mixing Tee Close up of Check Valve Figure 3 SRS-300 Gas Assisted Regeneration Kit Components Table 1 SRS-300 Gas Assisted Regeneration Kit Components Description P/N Quantity Manifold, 3-Way, 1.5 mm, ¼ Barb Fitting-Black Check Valve Regulator, 25 psi Clear Air Tubing,.062 x " Pink Air Tubing,.062 x " Assy, Tubing w/fittings-3" Long, 1/8" OD Assy, Tubing w/fittings-12" Long, 1/8" OD

Micro GC Fusion Pre-Installation Checklist

Micro GC Fusion Pre-Installation Checklist Site Preparation Proper site preparation prior to installing is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation. NOTE: Checkboxes ( ) identify hardware that

More information

GAS HANDLING SYSTEMS VACUUM INSTRUMENT CORPORATION

GAS HANDLING SYSTEMS VACUUM INSTRUMENT CORPORATION GAS HANDLING SYSTEMS VACUUM INSTRUMENT CORPORATION 2099 Ninth Avenue Ronkonkoma, NY 11779 www.vacuuminst.com Smart Charge II An advanced gas backfill and evacuation system with built-in gross leak test.

More information

Determination of Calcium and Magnesium in Brine

Determination of Calcium and Magnesium in Brine Application Note Determination of Calcium and Magnesium in Brine INTRODUCTION To prevent membrane poisoning, new membrane technology in chlor-alkali cells requires feed brine that is relatively free of

More information

HEAVY WATER LEAK DETECTOR

HEAVY WATER LEAK DETECTOR HEAVY WATER LEAK DETECTOR WEB www.overhoff.com MODEL 1925 HEAVY WATER LEAK DETECTOR (HWLD) LOW LEVEL REAL TIME TRITIUM-IN-WATER MONITOR This monitor has been designed for real time low-level detection

More information

23 The Thermal Conductivity Detector

23 The Thermal Conductivity Detector 23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations

Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations B e c k m a n C o u l t e r L i f e S c i e n c e s 2 5 0 S. K r a e m e r B o u l e

More information

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05

RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05 RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,

More information

Element D Services Plumbing

Element D Services Plumbing Medical Vacuum and Gas PART 1 - GENERAL 1.01 OVERVIEW A. This section addresses medical vacuum, waste anesthetic gas disposal, compressed air, oxygen, nitrous oxide, nitrogen and carbon dioxide systems.

More information

Dionex IonPac AS19-4µm

Dionex IonPac AS19-4µm Total Thermo Scientific Dionex IonPac AS9-4µm Column Product Manual P/N: September 203 Part of Thermo Fisher Scientific Product Manual for Dionex IonPac AS9-4µm Column Dionex IonPac AG9-4µm Capillary Guard

More information

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) IonSwift MAC-100 & MAC-200 Product Manual Page 1 of 16 PRODUCT MANUAL for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC) MAC-100, 0.5 x 80mm (P/N 074702)

More information

Chromatograph Maintenance and Troubleshooting Class # 5050

Chromatograph Maintenance and Troubleshooting Class # 5050 Chromatograph Maintenance and Troubleshooting Class # 5050 Shane Hale Business Development Manager Emerson Process Management Gas Chromatographs Division 5650 Brittmoore Drive Houston, TX 77041 USA Introduction

More information

TRL-onlineTOC MONITOR

TRL-onlineTOC MONITOR TRL-onlineTOC MONITOR TRL-onlineTOC Monitor operates on well accepted HTC (high temperature combustion) technology. It is a self sufficient system that requires no external side unit to operate. It is

More information

Gas Custody Transfer Calibration

Gas Custody Transfer Calibration Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform

More information

Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes

Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes Technical Note 64 Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes INTRODUCTION The AS Autosampler is enhanced with new features to speed throughput and simplify sample

More information

105 Adopted: 27.07.95

105 Adopted: 27.07.95 105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline

More information

TTC-IIS Series Up to 11 LPM Free Flow

TTC-IIS Series Up to 11 LPM Free Flow 66 Up to 11 LPM Free Flow Typical Applications Gas Analysis Anesthesia Monitors CO 2 Monitors Patient Monitoring Wound Therapy Urinalysis Trace Detection Medical/Training Mannequins Degassing Product Specifications

More information

Heating, Ventilation, Air Conditioning and Refrigeration (HVACR)

Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) Heating, Ventilation, Air Conditioning and Refrigeration (HVACR) I. Demonstrate safety skills in typical HVACR work situations to NATE Core Installer Knowledge Areas for Technician Excellence for Safety

More information

Facility Operations Physical Safety Guidance Document

Facility Operations Physical Safety Guidance Document Facility Operations Physical Safety Guidance Document Title: Guide on Pipe Identification Standards (ASME A13.1-1996) Original Date: December 18, 2000 Section: 4.0 Facility Operations Revised Date: Number:

More information

UP-213 & UP-266 Pre-Installation Guide

UP-213 & UP-266 Pre-Installation Guide UP-213 & UP-266 Pre-Installation Guide The UP system consists of two modules, the Laser Ablation module and the Laser Power Supply. The specifications are below: Laser Ablation Module: Length 25" (64cm);

More information

Electrospray Ion Trap Mass Spectrometry. Introduction

Electrospray Ion Trap Mass Spectrometry. Introduction Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Alarm and Detection/Suppression

Alarm and Detection/Suppression Alarm and Detection/Suppression Fire Safety American Fire Technologies To provide special hazard services world wide to the industrial market through the integration of components and services Clean Agent

More information

Doc.No. NDP 192U-02. Electric Pump Controller CE-124P. Instruction Manual

Doc.No. NDP 192U-02. Electric Pump Controller CE-124P. Instruction Manual Doc.No. NDP 192U-02 Electric Pump Controller CE-124P Instruction Manual 1. Parts Name and Function 1) POWER LED Indicate that CE-124P is powered up. 2) START/STOP KEY (SELECTOR KEY) Used to start and stop

More information

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Automation of Solid Phase Extraction and Column Chromatographic Cleanup Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for

More information

NITROUS TRANSFER PUMP INSTRUCTIONS

NITROUS TRANSFER PUMP INSTRUCTIONS NITROUS TRANSFER PUMP INSTRUCTIONS SAFETY TIPS Never directly inhale nitrous oxide. When inhaled in large quantities, nitrous oxide can cause respiratory ailments or in extreme cases, death by suffocation.

More information

Elemental analyzer multi EA 5000

Elemental analyzer multi EA 5000 Elemental analyzer multi EA 5000 Analytik Jena AG Konrad-Zuse-Str. 1 07745 Jena/Germany www.analytik-jena.com info@analytik-jena.com 1/6 multi EA 5000 The multi EA 5000 is a powerful elemental analyzer

More information

SERVICE & OPERATING MANUAL ELECTRONIC LEAK DETECTOR 032-044-000 032-045-000 WARNING 12-32 VDC

SERVICE & OPERATING MANUAL ELECTRONIC LEAK DETECTOR 032-044-000 032-045-000 WARNING 12-32 VDC SERVICE & OPERATING MANUAL ELECTRONIC LEAK DETECTOR 032-044-000 032-045-000 } 12-32 VDC Specifications 032-044-000 and 032-045-000 WARNING Hazardous voltage. This Leak Detector is an electrical device.

More information

Thermal Conductivity Detector

Thermal Conductivity Detector Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)

More information

VC Series Zone Valves BALANCED 2-WAY AND 3-WAY HYDRONIC VALVES

VC Series Zone Valves BALANCED 2-WAY AND 3-WAY HYDRONIC VALVES VC Series Zone Valves BALANCED 2-WAY AND 3-WAY HYDRONIC VALVES PRODUCT DATA Design VC Series zone valves consist of: 2-way or 3-way valve housing available with various connection ends Spindle and cartridge

More information

Flow Solution Parts & Accessories

Flow Solution Parts & Accessories Flow Solution Parts & Accessories Assemblies Part # Instrument U/M Price Debubbler Debubbler, Acrylic A000172 All ea $140.00 Distillation Coil Heater Spool A515017 FS III/IV ea $495.00 Digestion Coil UV

More information

ORGANIC SAMPLE PREPARATION

ORGANIC SAMPLE PREPARATION ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,

More information

The Breather Box. High Performance Grade-D Breathing Air Filtration system. BB50-cO

The Breather Box. High Performance Grade-D Breathing Air Filtration system. BB50-cO The Breather Box High Performance Grade-D Breathing Air Filtration system Specifications Filter change Indicators Incoming Air Plug 1/2 Industrial Interchange 2 Automatic Drains with Discharge tubes 3

More information

The Electron Capture Detector (ECD)

The Electron Capture Detector (ECD) Theory of operation The ECD is based on the phenomenon that electronegative species can react with thermal electrons present to form negatively charged ions. The loss of such electrons is related to the

More information

MIC-WKT and MIC-WKT-IR

MIC-WKT and MIC-WKT-IR MIC-WKT and MIC-WKT-IR Installation Manual Bosch Security Systems EN Installation and Operation Manual MIC-WKTI and MIC-WKT-IR Installation Manual EN 2 MIC-WKT and MIC-WKT-IR Washer Pump Drive Card Kits

More information

Qualification Guide. BPEC 601/2514/7 - Level 1 Diploma in Plumbing Foundation. BPEC 600/9432/1 - Level 2 Diploma in Plumbing Foundation

Qualification Guide. BPEC 601/2514/7 - Level 1 Diploma in Plumbing Foundation. BPEC 600/9432/1 - Level 2 Diploma in Plumbing Foundation Qualification Guide BPEC 601/2514/7 - Level 1 Diploma in Plumbing Foundation BPEC 600/9432/1 - Level 2 Diploma in Plumbing Foundation BPEC 600/5270/3 - Level 2 NVQ Diploma in Plumbing and Heating BPEC

More information

Geocontrol PRO INSTALLATION AND OPERATION MANUAL. Rev. 1 03/18/04 Part # 11150263

Geocontrol PRO INSTALLATION AND OPERATION MANUAL. Rev. 1 03/18/04 Part # 11150263 Geocontrol PRO INSTALLATION AND OPERATION MANUAL Rev. 1 03/18/04 Part # 11150263 TABLE OF CONTENTS Chapter 1: System Description... p. 05 Function and Theory...p. 05 Chapter 2: System Installation... p.

More information

PortaSens ll G AS DETECTOR

PortaSens ll G AS DETECTOR PortaSens ll G AS DETECTOR ONE DETECTOR... FOR MANY GASES! Sensors can be exchanged quickly and easily without the need for calibration. Sensors are maintained in ready condition in the optional sensor

More information

TRL-onlineTOC Monitor is designed for monitoring carbon content in

TRL-onlineTOC Monitor is designed for monitoring carbon content in TRL-onlineTOC Online TOC Monitor TRL-onlineTOC Monitor is designed for monitoring carbon content in Industrial process water To control and monitor different processes where water is used for heating or

More information

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC

FRACTION COLLECTOR. Microcomputer Controlled CHF122SC FRACTION COLLECTOR Microcomputer Controlled CHF122SC From open column to HPLC liquid chromatography and general liquid samplings, the Super Fraction Collector, model CHF122SC has a wide range of uses.

More information

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT

CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers

More information

Plumbing Code in Commercial Kitchens

Plumbing Code in Commercial Kitchens Plumbing Code in Commercial Kitchens City of Concord Code Administration 225-8580 Dan Clark Plumbing Inspector Tedd Evans, Chief Building Inspector Code References from the 2009 edition of the International

More information

INSTALLATION INSTRUCTIONS Fan Coil Replacement Coil Kit EBX & EBXX

INSTALLATION INSTRUCTIONS Fan Coil Replacement Coil Kit EBX & EBXX Fan Coil Replacement Coil Kit EBX & EBXX These instructions must be read and understood completely before attempting installation. These instructions covers the installation of replacement coil kit into

More information

Validation of process gas systems

Validation of process gas systems Validation of Process Gas Systems By Jeff Hargroves Alza Corporation Validation of process gas systems involve documenting the expected system behavior, and verifying that the system performs as expected.

More information

Electronic Diesel Control EDC 16

Electronic Diesel Control EDC 16 Service. Self-Study Programme 304 Electronic Diesel Control EDC 16 Design and Function The new EDC 16 engine management system from Bosch has its debut in the V10-TDI- and R5-TDI-engines. Increasing demands

More information

Rheodyne Manual Sample Injector Valves

Rheodyne Manual Sample Injector Valves Rheodyne Manual Sample Injector Valves Rotor Seal Shaft End Rotor Seal Groove Stator Face Stator Stator Face Seal Rotor Seal Grooves Position A Rotor Seal Grooves Position B Position C Flow paths showing

More information

WAYNE-DRESSER HARMONY SERIES DISPENSER RETROFIT for HEALY SYSTEMS, INC. MODEL VP1000 VAPOR RECOVERY ASSIST SYSTEM OUTLINE

WAYNE-DRESSER HARMONY SERIES DISPENSER RETROFIT for HEALY SYSTEMS, INC. MODEL VP1000 VAPOR RECOVERY ASSIST SYSTEM OUTLINE Page 98 WAYNE-DRESSER HARMONY SERIES DISPENSER RETROFIT for HEALY SYSTEMS, INC. MODEL VP1000 VAPOR RECOVERY ASSIST SYSTEM (KIT Z078) OUTLINE Notice: USE THIS PROCEDURE IF CONVERTING A BALANCE VAPOR RECOVERY

More information

Ultrasound Condition Monitoring

Ultrasound Condition Monitoring Ultrasound Condition Monitoring Whitepaper Alan Bandes UE Systems, Inc. Abstract: Instruments based on airborne/structure borne ultrasound technology offer many opportunities for reducing energy waste

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

Paper Chromatography: Separation and Identification of Five Metal Cations

Paper Chromatography: Separation and Identification of Five Metal Cations Paper Chromatography: Separation and Identification of Five Metal Cations Objectives Known and unknown solutions of the metal ions Ag +, Fe 3+, Co 2+, Cu 2+ and Hg 2+ will be analyzed using paper chromatography.

More information

Self Purging Liquid Trap (SPLT ) User s Manual

Self Purging Liquid Trap (SPLT ) User s Manual Self Purging Liquid Trap (SPLT ) User s Manual High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High

More information

ChromQuest 5.0 Chromatography Data System

ChromQuest 5.0 Chromatography Data System ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks

More information

columns IonPac CS18 Cation-Exchange Column

columns IonPac CS18 Cation-Exchange Column columns IonPac CS Cation-Exchange Column The IonPac CS is a cation-exchange column of moderate capacity and hydrophobicity, recommended for isocratic and gradient separations of polar amines such as alkanolamines,

More information

Fully Automated Distillation of Petroleum Products including Biofuels OptiDist - the Optimal Solution

Fully Automated Distillation of Petroleum Products including Biofuels OptiDist - the Optimal Solution Fully Automated Distillation of Petroleum Products including Biofuels OptiDist - the Optimal Solution The evaporating characteristics of hydrocarbons have an important effect on their performance. The

More information

(a) Method 1 Sample and Velocity Traverses for Stationary Sources.

(a) Method 1 Sample and Velocity Traverses for Stationary Sources. While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the

More information

Micro Motion 3098 Gas Specific Gravity Meter

Micro Motion 3098 Gas Specific Gravity Meter Product Data Sheet PS-001161, Rev. D April 2013 Micro Motion 3098 Gas Specific Gravity Meter Micro Motion density and concentration meters are built to tackle the most demanding process and fiscal applications.

More information

CHROMOCTANE O N LINE ANALYZER

CHROMOCTANE O N LINE ANALYZER GECIL Process GT instruments CHROMOCTANE O N LINE ANALYZER ASTM D6733 DHA and RON and MON on line GC for Reformer unit Automatic Chromatographic system for analysis of naphta feeds and reforming gasolines

More information

HOW TO SIZE A PRESSURE REGULATOR

HOW TO SIZE A PRESSURE REGULATOR CHOOSING THE CORRECT REGULATOR HOW TO SIZE A PRESSURE REGULATOR In order to choose the correct regulator the following information is needed: - Function: Pressure reducing or backpressure control? - Pressure:

More information

Signature and ISX CM870 Fuel System

Signature and ISX CM870 Fuel System Signature and ISX CM870 Fuel System Cummins Ontario Training Center HPI-TP Fuel System Heavy Duty High Pressure Injection - Time Pressure Fuel System The fuel system developed for the Signature and ISX

More information

Agilent 7697A Headspace Sampler Data Sheet

Agilent 7697A Headspace Sampler Data Sheet Agilent 7697A Headspace Sampler Data Sheet Overview The Agilent 7697A Headspace Sampler is part of Agilent s line of gas chromtography products. The 7697A based on the architecture of the market-leading

More information

Single sub-base valve

Single sub-base valve solenoid valves Mini-valves 10 mm Electrical connection... Mounting... Port sizes... Rated temperature... Fluid... Working pressure... Nominal flow... Pneumatic connections Voltage... Power... NC poppet

More information

Line to Refrigerator Ice/Water Dispenser

Line to Refrigerator Ice/Water Dispenser Standard 18 Line to Refrigerator Ice/Water Dispenser Pump Module Bottled Water How The System Works The FLOJET Bottled Water Dispensing System was designed to pump water from a commercially available 5-gallon

More information

Compressed Air Filters

Compressed Air Filters Compressed Air Filters Superior Filtration for Increased Productivity What happens when you don t use an air filter? Kaeser Gives You the Air Quality You Require Ambient air contains contaminants that

More information

ENVIRONMENTAL STANDARDS FOR PETROLEUM OIL REFINERIES EMISSIONS. Standards for emissions from furnaces, boilers and Power Plant

ENVIRONMENTAL STANDARDS FOR PETROLEUM OIL REFINERIES EMISSIONS. Standards for emissions from furnaces, boilers and Power Plant ENVIRONMENTAL STANDARDS FOR PETROLEUM OIL REFINERIES EMISSIONS Standards for emissions from furnaces, boilers and Power Plant S. Parameter Limiting concentration in mg/nm 3, unless stated Existing refineries

More information

DETERMINING THE MOLAR MASS OF CARBON DIOXIDE

DETERMINING THE MOLAR MASS OF CARBON DIOXIDE DETERMINING THE MOLAR MASS OF CARBON DIOXIDE PURPOSE: The goal of the experiment is to determine the molar mass of carbon dioxide and compare the experimentally determined value to the theoretical value.

More information

Hazards associated with the gas system and how to mitigate them

Hazards associated with the gas system and how to mitigate them Hazards associated with the gas system and how to mitigate them Do not enter the gas shack if the rotating red light at the entrance is on. This warns of an oxygen deficiency. In General the Gas Mixing

More information

PBX Series Quick Fit Connector Bimetallic Steam Traps

PBX Series Quick Fit Connector Bimetallic Steam Traps 6262100/6 IM-P626-01 ST Issue 6 PBX Series Quick Fit Connector Bimetallic Steam Traps Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Commissioning

More information

UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE 3211-4211 ABSORPTION OF CARBON DIOXIDE INTO WATER

UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE 3211-4211 ABSORPTION OF CARBON DIOXIDE INTO WATER UNIVERSITY OF MINNESOTA DULUTH DEPARTMENT OF CHEMICAL ENGINEERING ChE 3211-4211 ABSORPTION OF CARBON DIOXIDE INTO WATER OBJECTIVE The objective of this experiment is to determine the equilibrium line,

More information

el-720 Carbon Dioxide Gas Leak Detector User s Guide

el-720 Carbon Dioxide Gas Leak Detector User s Guide el-720 Carbon Dioxide Gas Leak Detector User s Guide Congratulations on the purchase of the AccuTools el-720 Carbon Dioxide Gas Leak Detector, the most technologically advanced instrument of its kind.

More information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in

More information

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS Products Include: 60 gpm eductors Style 3060, 3061, 3062 & 3070 95 gpm eductors Style

More information

SciTech, sro Praha Ceník Jour Research tisk 3.1.2000

SciTech, sro Praha Ceník Jour Research tisk 3.1.2000 Ceny bez DPH, balného a poštovného. Jour Research pricelist 08/99 Part Number Description cena CZK 0001-0737 Wash-Pro Upgrade Kit 7098,00 0001-5530 Gas Flow Meter 22932,00 0001-5531 Silicon Connection

More information

Agilent Veriflow 500 Electronic Flowmeter. Flow Range: 1 500 sccm. Operation Guide. Agilent Technologies

Agilent Veriflow 500 Electronic Flowmeter. Flow Range: 1 500 sccm. Operation Guide. Agilent Technologies Agilent Veriflow 500 Electronic Flowmeter Flow Range: 1 500 sccm Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by

More information

Flow Injection Analysis

Flow Injection Analysis Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since

More information

DIGITAL ALARM II FOR HOSPITALS AND LABORATORIES INSTALLATION AND OPERATING INSTRUCTIONS 52 635.0

DIGITAL ALARM II FOR HOSPITALS AND LABORATORIES INSTALLATION AND OPERATING INSTRUCTIONS 52 635.0 Form No. 74-00-4001 S168-195-001 Revision E DIGITAL ALARM II FOR HOSPITALS AND LABORATORIES INSTALLATION AND OPERATING INSTRUCTIONS 52 635.0 INTRODUCTION Allied Healthcare Products, Inc. s Digital Alarm

More information

LIPE- Institut National des Sciences Appliquées 31077 Toulouse, France Contact: adamou@enstimac.fr

LIPE- Institut National des Sciences Appliquées 31077 Toulouse, France Contact: adamou@enstimac.fr Characterization of the ionic liquid bmimpf 6 in supercritical conditions Abla ADAMOU a, Jean-Jacques LETOURNEAU a, Elisabeth RODIER a, René DAVID a, Pascal GUIRAUD b a LGPSD- UMR CNRS 2392, Ecole des

More information

HVAC Processes. Lecture 7

HVAC Processes. Lecture 7 HVAC Processes Lecture 7 Targets of Lecture General understanding about HVAC systems: Typical HVAC processes Air handling units, fan coil units, exhaust fans Typical plumbing systems Transfer pumps, sump

More information

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases

EuroFID Total Hydrocarbon Analyzer. Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases Product information EuroFID Total Hydrocarbon Analyzer Precise Determination of Total Hydrocarbons in Air for Corrosive as well as Condensing Gases Proven Analyzer Technology Measuring total hydrocarbons

More information

Powers Controls TH 192 S Single Temperature Room Thermostat

Powers Controls TH 192 S Single Temperature Room Thermostat Technical Instructions Document No. 155-065P25 TH 192-1 Powers Controls TH 192 S Single Temperature Room Thermostat 50 60 70 80 TH0355R1 70 80 POWERS Description The TH 192 S thermostats are proportional

More information

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter Thermal Mass meter and Temperature Transmitter Description The MTI10 insertion thermal mass flowmeter and the MTL10 in-line thermal mass flowmeter provide accurate mass flow measurement of clean, dry gases

More information

B150 Series Commercial Brewer. Direct Plumb Kit

B150 Series Commercial Brewer. Direct Plumb Kit B150 Series Commercial Brewer Welcome Keurig Authorized Dealer! Direct Plumb Kit This manual provides: Installation of the Direct Plumb Kit (DPK) that allows you to connect your customer s B150 brewer

More information

Installing a Probe. Typical Installation

Installing a Probe. Typical Installation for hazardous locations. Prior to any installation in a Classified Hazardous Location, verify installation methods by the Control Drawing referenced on the product s name tag 2. The Feed Through Assembly

More information

Thermal flow sensors TA10-ZG8c and TA10-ZG9c for exact and stable, long-term measuring of lower flow velocities (Laminar Flow)

Thermal flow sensors TA10-ZG8c and TA10-ZG9c for exact and stable, long-term measuring of lower flow velocities (Laminar Flow) Thermal flow sensors TA10-ZG8c and TA10-ZG9c for exact and stable, long-term measuring of lower flow velocities (Laminar Flow) Drawing 8c Drawing 9c U391_TA10C_DoP_e_150421 www.hoentzsch.com 1 of 5 Measured

More information

THE HUMIDITY/MOISTURE HANDBOOK

THE HUMIDITY/MOISTURE HANDBOOK THE HUMIDITY/MOISTURE HANDBOOK Table of Contents Introduction... 3 Relative Humidity... 3 Partial Pressure... 4 Saturation Pressure (Ps)... 5 Other Absolute Moisture Scales... 8 % Moisture by Volume (%M

More information

Electro-pneumatic control valves

Electro-pneumatic control valves Electropneumatic control valves 7 Miniature solenoid valves for alternating current Conform to the Low Voltage Directive or mounting on subbase or footprint in accordance with CNOMO recoendation E 066120N

More information

Fiber Optic Selector Guide for Analog & Digital Data Links, Contact Closures & Multiplexers

Fiber Optic Selector Guide for Analog & Digital Data Links, Contact Closures & Multiplexers Fiber Optic Selector Guide for Analog & Digital Data Links, Contact Closures & Multiplexers Analog Data Links Multi-Channel Contact Closures Multiplexers Digital Data Links - Contact Closures Power Supply

More information

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Chapter 28: High-Performance Liquid Chromatography (HPLC) Chapter 28: High-Performance Liquid Chromatography (HPLC) Scope Instrumentation eluants, injectors, columns Modes of HPLC Partition chromatography Adsorption chromatography Ion chromatography Size exclusion

More information

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown) Service Manual Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

More information

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas Paper #46 Heidi C. Hayes Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 Diane J.

More information

Fundamentals of Gas Chromatography

Fundamentals of Gas Chromatography Application Note Oil & Gas Fundamentals of Gas Chromatography Figure 1 - The Function Components of a Gas Chromatograph Process Line Sample System Chromatograph Oven GC Controller Probe Sample Return Slip

More information

RGC-IR Remote Gas Calibrator for IR400

RGC-IR Remote Gas Calibrator for IR400 Remote Gas Calibrator for IR400 The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by

More information

GAS LEAKAGE DETECTION & AUTO ON-OFF GAS SYSTEM

GAS LEAKAGE DETECTION & AUTO ON-OFF GAS SYSTEM GAS LEAKAGE DETECTION & AUTO ON-OFF GAS SYSTEM This system helps you to upgrade your safety standards, comply statutory requirements on environmental commitments and most important and basic function being

More information

moehwald Bosch Group

moehwald Bosch Group moehwald Bosch Group Division Testing Technology for Fuel Cells Moehwald GmbH Michelinstraße 21 Postfach 14 56 66424 Homburg, Germany Tel.: +49 (0) 68 41 / 707-0 Fax: +49 (0) 68 41 / 707-183 www.moehwald.de

More information

In-Line Filters and Coalescers with Automatic Drains

In-Line Filters and Coalescers with Automatic Drains Instructions/Parts In-Line Filters and Coalescers with Automatic Drains 310611G EN Air Filter with Automatic Drain Part No. 22601, Series B Maintenance To maintain maximum filtering efficiency and to avoid

More information

Evaluate, Clean, and Tune Guidance

Evaluate, Clean, and Tune Guidance Evaluate, Clean, and Tune Guidance The Evaluate, Clean and Tune (ECT) process serves three essential purposes in the Weatherization Assistance Program (WAP). The first is to evaluate the existing system

More information

Please also see 10 golden rules for correct Saniflo installation on www.saniflo.co.uk

Please also see 10 golden rules for correct Saniflo installation on www.saniflo.co.uk 1. Description Sanicom is a lifting pump for wastewater used in commercial applications (restaurants, hairdressing salons, the catering industry etc.). The unit operates automatically, has a high performance

More information

BTU ANALYSIS USING A GAS CHROMATOGRAPH

BTU ANALYSIS USING A GAS CHROMATOGRAPH BTU ANALYSIS USING A GAS CHROMATOGRAPH By Joseph E. Landes Southern Petroleum Laboratories, Inc. Technical Director Hydrocarbon and Field Services 8820 Interchange Drive Houston, Texas 77054 Introduction

More information

Electricity. Investigating spontaneous gas discharge in air as a function of pressure. LD Physics Leaflets P3.9.2.1. 0210-Sel

Electricity. Investigating spontaneous gas discharge in air as a function of pressure. LD Physics Leaflets P3.9.2.1. 0210-Sel Electricity Electrical conduction in gases Gas discharge at reduced pressure LD Physics Leaflets P3.9.2.1 Investigating spontaneous gas discharge in air as a function of pressure Objects of the experiments

More information

Elektromotoren und Gerätebau Barleben GmbH

Elektromotoren und Gerätebau Barleben GmbH Elektromotoren und Gerätebau Barleben GmbH OPERATING INSTRUCTIONS Transformer Protection Relay (Buchholz principle) Elektromotoren und Gerätebau Barleben GmbH Table of contents Page 1 Safety instructions

More information