UP-213 & UP-266 Pre-Installation Guide

Size: px
Start display at page:

Download "UP-213 & UP-266 Pre-Installation Guide"

Transcription

1 UP-213 & UP-266 Pre-Installation Guide The UP system consists of two modules, the Laser Ablation module and the Laser Power Supply. The specifications are below: Laser Ablation Module: Length 25" (64cm); Width 18" (46cm); Height (to eyepiece) 22" (56cm) Weight 130 lbs (59 kg) VAC, 3A, 50/60Hz VAC, 2A, 50/60Hz Laser Power Supply: Length 19" (48cm) Width 8.6" (22cm) Height 15" (38cm) Weight 55 lbs (25 kg) VAC, 10A, 50/60Hz VAC, 5A, 50/60Hz Laser Ablation module (with optional Trinoc assembly)

2 Front View Side View Back View Space Requirements The Laser Ablation Module requires 18 Wide x 25 Deep x 22 High (46 x 64 x 56 centimeters) of countertop space. In addition, space must be made for the computer monitor and keyboard if the system is not integrated with the mass spectrometer. The Laser Power Supply usually sits on the floor and occupies a 9 Wide x 15 Deep x 19 High (22 x 38 x 48 centimeters) footprint. The system should be placed a table or platform capable of supporting the 130 lb weight of the Laser Ablation module and/or the 55 lb Power Supply. Note: Although NWR does not recommend any particular carts or tables, customer have used the "Intermetro", one of the Metro BC Series Utility Carts with good results. These can be ordered from VWR Scientific products ( ). There are 3 sizes for the two-shelf carts (product #'s , -416, -420) and three sizes for the 3-shelf carts ( , -418, -422). For most applications, the sample flow path should be short. Therefore, the Laser Ablation System needs to be placed close to the mass spectrometer. Electrical Requirements The Laser Power Supply is shipped configured as either a 110 VAC, 10 A, 50/60 Hz or a 220 VAC, 5A, 50/60 HZ system. The Laser Ablation Module is shipped configured as either a 110 VAC, 5 A, 50/60 Hz or a 220 VAC, 3A, 50/60 HZ system. The operating voltage is specified at the time of the order. North American systems are typically configured for 110 VAV and come with standard 6 foot 3-prong power cables.

3 In either case, the system should be located within 6 feet (2 meters) of a minimum of two power outlets for the: Base Module, Laser Power Supply. Two additional outlets may be required for optional computer and computer monitor. Computer: Sharing of the MS computer or a dedicated computer can be used. With either method a video card would need to be installed in the computer. The Video Card is ordered with original system configuration and the NWR Field Service Engineer will install the PCB during the installation. All software and drivers will be supplied and installed by NWR. Note: If a computer is ordered through NWR, the computer will come preloaded with all software, drivers and video card. NWR supplies a 10 foot RS232 cable for communication between the computer and the Laser Ablation module and a 12 foot S-Video cable for video feed between the Laser Ablation module and the computer. The computer and/or the Laser Ablation module need to be located within reach of these cables. 25 foot cable options are available from NWR or can be purchased locally by the customer. Gas Lines: A carrier gas source is required to transport the ablated sample to the measurement system. Argon or Argon/Helium is normally used. It is recommended that % pure helium and argon gas be used as the cell and/or makeup gas. The supplies should be flow controlled to ensure the best analytical stability. NWR supplies 12 feet of 1/4 OD (1/8 ID) Tygon tubing to connect from the carrier gas source to the Laser Ablation module and from the Laser Ablation module to the torch. Plastic Swage TM connectors are provided to connect the Tygon tubing to the Laser Ablation module, all other supply line connectors are to be provided by the customer. NWR also supplies a torch adaptor (illustrated below). If the measurement system requires another type of adaptor, the customer should obtain the appropriate torch adaptor from the MS vendor. Glass Torch Adaptor figure 1 Glass Torch Adaptor figure 2 Most ICP systems incorporate independent mass flow control of the Argon nebuliser gas.

4 Argon Carrier Gas: When using Argon only as the carrier gas, connecting the 1/4 Tygon tubing from the Argon ( Neb ) line to the Laser Ablation module GAS IN port provides a regulated carrier gas flow to the Laser Ablation module. The customer is to supply the connection to the Argon ( Neb ) line. Do not supply argon directly from a gas bottle regulator, the gas should be precisely controlled by an MFC. The GAS OUT port is then connected to the torch using the proper torch adaptor. Helium Carrier Gas/Argon Makeup Gas: Many ICP-MS systems incorporate a second aerosol mass flow controller. This MFC must be calibrated for Helium. If this is not present, then an external gas control manifold is required for the control of the helium cell gas. The external MFC can be obtained independently or a complete MFC control module can be purchased from NWR. Using the 1/4" Tygon tubing, the Helium MFC output is connected GAS IN port on the back of the Laser Ablation module (or directly from the Helium gas supply with system with the MFC built into the Laser Ablation module). The GAS OUT port is then merged with the Argon makeup gas ( Neb gas line) using the NWR supplied Y connector, Tygon tubing and clamps. The customer is to supply the connection to the Argon ( Neb ) line. The combined gas line is then connected to the torch using the proper torch adaptor. See illustrations below. Gas Port on the back of the Laser Ablation module Helium & Argon connections

5 ARGON as carrier gas Gas Connections located on the back of the Laser Ablation module Gas In Purge Gas Out Plastic Swage TM connector provided by NWR (Qty 3) 12 feet (total) of ¼ Tygon tubing provided by NWR. Additional, if needed, provided by customer Connection of Argon (Neb) gas supply line to ¼ Tygon tubing supplied by customer. Argon MFC Gas source provided by customer Standard glass torch adaptor provided by NWR. Any other type must be obtained, by the customer, from MS vendor

6 HELIUM carrier gas & ARGON makeup gas Gas Connections located on the back of the Laser Ablation module Connection of Argon (Neb) gas supply line to ¼ Tygon tubing Gas In Purge Gas Out supplied by customer. Plastic Swage TM connector provided by NWR (Qty 3) Argon MFC Gas source provided by customer 12 feet (total) of ¼ Tygon tubing provided by NWR. Additional, if needed, provided by customer Y connector and clamps provided by NWR Connection of Helium gas supply line to ¼ Tygon tubing supplied by customer. Helium MFC gas source (or regulated Helium supply if MFC built into Laser Ablation module) provided by customer Standard glass torch adaptor provided by NWR. Any other type must be obtained, by the customer, from MS vendor

3500/93 System Display

3500/93 System Display 3500/93 System Display Bently Nevada* Asset Condition Monitoring Description The 3500/93 System Display is designed to meet the requirements of American Petroleum Institute (API) Standard 670 and provide

More information

OEM-EP Pressure Controllers

OEM-EP Pressure Controllers Pressure Controllers Typical Applications Carrier Gas Pressure Control Air over Liquid Control Mass Spectrometer Process Gas Supply Pressure Control Miniature ler The Miniature ler converts a variable

More information

PS6500 Storage Arrays Rack Mount Instructions

PS6500 Storage Arrays Rack Mount Instructions PS6500 Storage Arrays Rack Mount Instructions Part Number: R724M Rev. A01 Copyright 2010 Dell, Inc. All rights reserved. Dell is a trademark of Dell, Inc. EqualLogic is a registered trademark. All trademarks

More information

Research RF Sputtering Packages

Research RF Sputtering Packages Research RF Sputtering Packages Shown with 3" Polaris Adjustable Position Source with Tilt and Shutter 300 or 600 Watt Low Cost Packages for Those With Limited Budgets Desiring Full Capability 13.56 MHz

More information

TAC I/NETTM MR-VAV-AX. Application Specific MicroRegulator TM

TAC I/NETTM MR-VAV-AX. Application Specific MicroRegulator TM TAC I/NETTM TM TAC s TM has an extensive range ofpre-engineered VAV control sequences, which combined with a fully integrated actuator and airflow transducer, makes it a low Total Installed Cost solution

More information

[see Figure 3 below] Ferrite tile walls and ceiling RF Absorber parcel walls and ceiling Removable floor section to make fully anechoic

[see Figure 3 below] Ferrite tile walls and ceiling RF Absorber parcel walls and ceiling Removable floor section to make fully anechoic Test Room ID: R21F Description: SEMI ANECHOIC SHIELDED TEST ENCLOSURE Primary Testing Uses: FCC, CE Mark Dimensions: 9.75m x 6.10m x 5.49m (32ft. x 20ft. x 18ft.) [see Figure 3 below] Construction: Galvanized

More information

Micro GC Fusion Pre-Installation Checklist

Micro GC Fusion Pre-Installation Checklist Site Preparation Proper site preparation prior to installing is very important. Contact INFICON (www.inficon.com) with any questions regarding site preparation. NOTE: Checkboxes ( ) identify hardware that

More information

Alpha/Beta Counting Multi-Detector Systems (MDS)

Alpha/Beta Counting Multi-Detector Systems (MDS) Alpha/Beta Counting Multi-Detector Systems (MDS) Ultra High Performance Alpha/Beta Counting Systems for Maximum Laboratory Productivity The Multi-Detector System (MDS) is a multi-detector, low-background

More information

Cisco MCS 7816-I3 Unified Communications Manager Appliance

Cisco MCS 7816-I3 Unified Communications Manager Appliance Cisco MCS 7816-I3 Unified Communications Manager Appliance Cisco Unified Communications is a comprehensive IP communications system of voice, video, data, and mobility products and applications. It enables

More information

Melting Point. Electrothermal PRODUCT SPECIFICATIONS OPERATION

Melting Point. Electrothermal PRODUCT SPECIFICATIONS OPERATION M A N U A L M E L - T E M P THE LOWEST COST MELTING POINT APPARATUS FOUND ANYWHERE! Low cost unit Temperature range to 500 C Aluminum casting conducts and radiates heat uniformly to capillaries and thermometer.

More information

CATTURA TM DOWNDRAFT MODELS D49M30SB, D49M36SB, D49M48SB

CATTURA TM DOWNDRAFT MODELS D49M30SB, D49M36SB, D49M48SB SPECIFICATIONS CATTURA TM DOWNDRAFT MODELS D49M30SB, D49M36SB, D49M48SB Page 1 Powerful VERTEX TM Complete Capture Design rivals the finest island hoods. Compact design allows the choice of many cooktop

More information

Interface Adapters PS/2 Interface Adapter 1 pack 262588-B21 PS/2 Interface Adapter 8 pack 262587-B21 USB Interface Adapter 1 pack 336047-B21

Interface Adapters PS/2 Interface Adapter 1 pack 262588-B21 PS/2 Interface Adapter 8 pack 262587-B21 USB Interface Adapter 1 pack 336047-B21 Overview Models HP Server Console Switches 1X8-Port KVM 336044-B21 2X16-Port KVM 336045-B21 CAT5e Cables 3 Foot (1 m) 4 pack 263474-B21 6 Foot (2 m) 8 pack 263474-B22 12 Foot (4 m) 8 pack 263474-B23 20

More information

Storage System. Planning Checklist. EC1434-2-E1, First Edition

Storage System. Planning Checklist. EC1434-2-E1, First Edition Storage System Planning Checklist EC1434-2-E1, First Edition Proprietary Rights Notice This document contains proprietary information of Engenio Information Technologies, Inc. and Storage Technology Corporation.

More information

SIMULATION PRE-INSTALLATION GUIDELINES CONTRACTOR REQUIREMENTS FOR ROOM PREPARATION

SIMULATION PRE-INSTALLATION GUIDELINES CONTRACTOR REQUIREMENTS FOR ROOM PREPARATION SIMULATION PRE-INSTALLATION GUIDELINES CONTRACTOR REQUIREMENTS FOR ROOM PREPARATION version 1.0_042011 www.foresightsports.com TABLE OF CONTENTS INTRODUCTION 1 COMPONENTS-AT-A-GLANCE 2 SPACE DIMENSIONS

More information

CutMaster 82. 6.04 Major External Replacement Parts

CutMaster 82. 6.04 Major External Replacement Parts CutMaster 8 6.0 Major External Replacement Parts Item # Qty Description Catalog # Cover with labels 9-08 Base Enclosure Assembly 9-08 Tube, roll handle 9-0 Front Panel 9-07 Rear Panel 9-00 Art # A-080

More information

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE

OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &

More information

Model 349KLX Remote Display Digital Scale User Instructions

Model 349KLX Remote Display Digital Scale User Instructions Model 349KLX Remote Display Digital Scale User Instructions P/N UM349KLX Rev 3-080713 1 Model 349KLX Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully,

More information

Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001

Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001 Express5800/120Ed Rack Mount Kit Installation Procedures PN: 455-01607-001 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,

More information

STT -10 WIRE FEEDER. For Surface Tension Transfer TM. Process

STT -10 WIRE FEEDER. For Surface Tension Transfer TM. Process STT -10 WIRE FEEDER For Surface Tension Transfer TM Process STT -10 WIRE FEEDER ADVANTAGE LINCOLN The STT-10 does a lot more than just feed wire. This new generation microprocessor controlled wire feeder

More information

Installing Cisco TelePresence MX200 G2 - Wheel Base

Installing Cisco TelePresence MX200 G2 - Wheel Base This installation guide covers MX200 G2 with a wheel base. Dimensions Height: 1480 mm / 58.3 in. Width: 995 mm / 39.2 in. Depth: 715 mm / 28.0 in. Weight total: 33 kg / 73 lb Weight monitor: 24 kg / 53

More information

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS PORTER MASS MEASUREMENT AND CONTROL SYSTEMS PORTER: THE STANDARD OF EXCELLENCE IN MASS Porter Mass Flow products reflect nearly three decades of experience in the design and manufacture of precision instruments

More information

Vision Engravers and Routers PRE-Installation Guide. 2015 Vision Engraving & Routing Systems

Vision Engravers and Routers PRE-Installation Guide. 2015 Vision Engraving & Routing Systems Vision Engravers and Routers PRE-Installation Guide Revised: 8/19/2015 Vision Engravers and Routers PRE-Installation Guide All rights reserved. No parts of this work may be reproduced in any form or by

More information

Dell Vostro 3460/3560

Dell Vostro 3460/3560 Dell Vostro 3460/3560 Setup And Features Information About Warnings WARNING: A WARNING indicates a potential for property damage, personal injury, or death. Vostro 3460 - Front And Back View Figure 1.

More information

EZ-48-3AB-125-35 Weather Station Mounting Tripod and Mast

EZ-48-3AB-125-35 Weather Station Mounting Tripod and Mast EZ-48-3AB-125-35 Weather Station Mounting Tripod and Mast EZ-125-35M 35 Mast Extension (Optional) EZ-125-SK Stake Kit (Optional) EZ-46-3 Tar Pad Kit (Optional) EZ-GWA Guy Wire Kit (Optional) Ambient Weather

More information

Incubator with mechanical convection

Incubator with mechanical convection BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,

More information

AUTOMATIC CHANGEOVER MANIFOLD Series 915

AUTOMATIC CHANGEOVER MANIFOLD Series 915 AUTOMATIC CHANGEOVER MANIFOLD Series 915 FEATURES Fully automatic, simple, hassle-free operation No variance in delivery pressure Designed for ultra high purity gas service Built-in audio and visual alarm

More information

NexION 300 ICP-MS. ICP-Mass Spectrometry PREPARING YOUR LAB. Preparation Checklist

NexION 300 ICP-MS. ICP-Mass Spectrometry PREPARING YOUR LAB. Preparation Checklist PREPARING YOUR LAB NexION 300 ICP-MS ICP-Mass Spectrometry Preparation Checklist Environmental conditions Electrical requirements Space requirements Exhaust ventilation Coolant requirements Argon gas requirements

More information

Dell Latitude E6320. Setup and Features Information. About Warnings. Front and Back View

Dell Latitude E6320. Setup and Features Information. About Warnings. Front and Back View Dell Latitude E6320 Setup and Features Information About Warnings WARNING: A WARNING indicates a potential for property damage, personal injury, or death. Front and Back View Figure 1. Front view 1. microphone

More information

Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry

Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry PBF T Peel Back Force Tester GPD s Peel Back Force Tester with FORCEWare software is being used to set peel back standards for the entire industry 611 Hollingsworth Street, Grand Junction, CO 81505 tel:

More information

Cisco MCS 7825-H3 Unified Communications Manager Appliance

Cisco MCS 7825-H3 Unified Communications Manager Appliance Cisco MCS 7825-H3 Unified Communications Manager Appliance Cisco Unified Communications is a comprehensive IP communications system of voice, video, data, and mobility products and applications. It enables

More information

About the BitStorm 6051 POTS Splitter

About the BitStorm 6051 POTS Splitter BitStorm 6051 Installation Instructions Document Number 6051-A2-GZ40-10 July 2002 About the BitStorm 6051 The BitStorm system enables simultaneous high-speed digital data access and analog voice service

More information

APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out

APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out APC Matrix-UPS 5kVA 208V/240V In 120/208/240V Out APC Matrix-UPS, 3750 Watts / 5000 VA,Input 208V / Output 120V, 208V, Interface Port DB-9 RS-232, SmartSlot, Extended runtime model Includes: CD with software,

More information

How To Use Bodescan For 3D Imaging Of The Human Body

How To Use Bodescan For 3D Imaging Of The Human Body «Bodescan» THE ULTIMATE 3D SCANNER FOR DATA ACQUISITION OF HUMAN BODY A COMPREHENSIVE SYSTEM INCLUDING HARDWARE AND SOFTWARE TOOLS Page 2 of 9 Bodescan HUMAN BODY3D SCANNING HARDWARE AND SOFTWARE INCLUDED

More information

XPS 8900. Views. Specifications

XPS 8900. Views. Specifications XPS 8900 Views Copyright 2015 Dell Inc. All rights reserved. This product is protected by U.S. and international copyright and intellectual property laws. Dell TM and the Dell logo are trademarks of Dell

More information

CutMaster 102. 6.04 Major External Replacement Parts

CutMaster 102. 6.04 Major External Replacement Parts CutMaster 0 6.04 Major External Replacement Parts Item # Qty Description Catalog # Cover with labels 9-08 Base Enclosure Assembly 9-086 Tube, roll handle 9-0 4 Front Panel 9-087 Rear Panel 9-00 4 Art #

More information

Displayport Extender via 2 multimode fibers LC Duplex Connector Extends Displayport Link up to 200 meters

Displayport Extender via 2 multimode fibers LC Duplex Connector Extends Displayport Link up to 200 meters Description Displayport combines the audio and video through an interface. It supports high resolution displays and multiple displays with a single cable, and increases the efficiency and bandwidth of

More information

Model PS-4001 Power Supply User Instructions

Model PS-4001 Power Supply User Instructions Model PS-4001 Power Supply User Instructions 9350-7710-000 Rev E 9/2009 PROPRIETARY NOTICE The product information and design disclosed herein were originated by and are the property of Bosch Security

More information

HP rack and power infrastructure products are designed and developed to be used specifically with HP servers and storage.

HP rack and power infrastructure products are designed and developed to be used specifically with HP servers and storage. Overview HP LCD8500 1U Rackmount Console Kit The HP LCD8500 1U Rackmount Console Kit is an ultra-dense, high-performance KVM console that combines a full 18.5-inch LCD monitor with keyboard and touch pad

More information

Dell Latitude 3440/3540

Dell Latitude 3440/3540 Dell Latitude 3440/3540 Setup And Features Information About Warnings WARNING: A WARNING indicates a potential for property damage, personal injury, or death. Latitude 3440 Front and Back View Figure 1.

More information

Our data centres have been awarded with ISO 27001:2005 standard for security management and ISO 9001:2008 standard for business quality management.

Our data centres have been awarded with ISO 27001:2005 standard for security management and ISO 9001:2008 standard for business quality management. AIMS is Malaysia and South East Asia s leading carrier neutral data centre operator and managed services provider. We provide international class data storage and ancillary services, augmented by an unrivaled

More information

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters Description The Tricor Mass Flow Meters measure simultaneously mass

More information

Replacement Procedure for E4982A LCR Meter REPLACEMENT PROCEDURE FOR E4982A LCR METER

Replacement Procedure for E4982A LCR Meter REPLACEMENT PROCEDURE FOR E4982A LCR METER Replacement Procedure for E498A LCR Meter REPLACEMENT PROCEDURE FOR E498A LCR METER REVISION HISTORY REVISION 1 DESCRIPTION Initial Release DATE May 01 REVISION HISTORY TOOL AND FIXTURE LIST PART NO. DESCRIPTION

More information

System Accessories Pricing. Page 0 Effective 7/02

System Accessories Pricing. Page 0 Effective 7/02 System Accessories Pricing Page 0 Effective 7/02 Firetrace Tubing 200005 Firetrace Detection Tubing (FDT) (Per Foot) (Available in continuous lengths up to 333 ft.) Accessories 400001 400004 400061 400060

More information

Lighting Controls ! WARNING RISK OF ELECTRIC SHOCK. Installation Instructions DESCRIPTION

Lighting Controls ! WARNING RISK OF ELECTRIC SHOCK. Installation Instructions DESCRIPTION GE Lighting Installation Instructions Lighting Controls Centralized Lighting Control Panel Interior Catalog Number CLCINTxx DESCRIPTION The Centralized Lighting Control System is a small network of relay

More information

Dell Precision Workstation T3600/T5600/ T7600

Dell Precision Workstation T3600/T5600/ T7600 Dell Precision Workstation T3600/T5600/ T7600 Setup and Features Information About Warnings WARNING: A WARNING indicates a potential for property damage, personal injury, or death. Dell Precision Workstation

More information

8-Port PoE Network Switch

8-Port PoE Network Switch Page 1 of 12 8-Port PoE Network Switch User s Guide Document Number 91I00-1 Rev A August, 2009 FCC Warning: Page 2 of 12 This devices has been tested and found to comply with the regulations for Class

More information

Natural gas chromatograph LNG gas quality testing

Natural gas chromatograph LNG gas quality testing Application description AD/2102760-EN Natural gas chromatograph LNG gas quality testing Introduction The portable NGC8206 is a natural gas chromatograph designed for installation in a vehicle or mounted

More information

Alpha System ADAPTABLE KEWAUNEE LABORATORY FURNITURE. encouraging new discovery

Alpha System ADAPTABLE KEWAUNEE LABORATORY FURNITURE. encouraging new discovery Alpha System ADAPTABLE LABORATORY FURNITURE KEWAUNEE encouraging new discovery Alpha 6 partition modules with Kemresin worktops. Minneapolis, Minnesota Alpha 6 partition modules with Alpha overhead service

More information

3500/25 Enhanced Keyphasor*Module

3500/25 Enhanced Keyphasor*Module 3500/25 Enhanced Keyphasor*Module Bently Nevada* Asset Condition Monitoring Description The 3500/25 Enhanced Keyphasor Module is a half-height, two-channel module used to provide Keyphasor signals to the

More information

Video conferencing expanding conferencing experiences

Video conferencing expanding conferencing experiences Video conferencing expanding conferencing experiences Stylish mounting solutions for video conferencing Video conferencing furniture Video conferencing is developing fast. Vogel s is capitalising on this

More information

Automation & Sanitization REPLACEMENT PARTS

Automation & Sanitization REPLACEMENT PARTS & Sanitization PROLOGIC GLX-PCB-PRO Main PCB, ProLogic GLX-PL-LOC-P-4 Display, Local, ProLogic, P-4 GLX-PL-LOC-PS4 Display, Local, ProLogic, PS-4 GLX-PL-LOC-PS8 Display, Local, ProLogic, PS-8 GLX-PL-LOC-PS16

More information

NEW APC Smart-UPS. Advanced line interactive power protection for servers and network equipment. The world s most popular network and server UPS.

NEW APC Smart-UPS. Advanced line interactive power protection for servers and network equipment. The world s most popular network and server UPS. NEW APC Smart-UPS Advanced line interactive power protection for servers and network equipment. > The world s most popular network and server UPS. Award-winning Smart-UPS protects critical data by supplying

More information

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter Description The coriolis mass flow meters measure simultaneously mass

More information

Easy Installation All five standard monitors fit into shallow mounting areas, requiring about 3" of depth. The XT1502, SXT1811, and SXT2011 models hav

Easy Installation All five standard monitors fit into shallow mounting areas, requiring about 3 of depth. The XT1502, SXT1811, and SXT2011 models hav Flat Panel Industrial Monitors FEATURES 10.4 to 20.1 bright, easy-to-read color TFT flat panel display with screen resolution from VGA to SXGA Large viewing area and viewing angles Operator input through

More information

Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems

Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES FIMS 100 and 400 Flow Injection Mercury Systems Delivering the Performance You Need The determination and monitoring of sub-ppb concentrations

More information

SMART Board 8055i and 8055i-SMP

SMART Board 8055i and 8055i-SMP SMART Board 8055i and 8055i-SMP Interactive flat panel Installation guide Product registration If you register your SMART product, we ll notify you of new features and software upgrades. Register online

More information

SPID Azimuth Antenna rotator Type: BIG-RAK/HR

SPID Azimuth Antenna rotator Type: BIG-RAK/HR WWW.RFHAMDESIGN.COM SPID Azimuth Antenna rotator Type: BIG-RAK/HR SPID BIG-RAK/HR MD-02 Rotor Controller Type BIG-RAK/HR is a 1 engine rotor system (Azimuth) and delivered including: BIG-RAK/HR rotor 0.2

More information

Nevco, Inc. Installation Manual

Nevco, Inc. Installation Manual Nevco, Inc. Installation Manual Serial No. Model GLH-5 Size: Each module is 18 L x 19 ¾ H x 4 D (.46 x.50 x.10 meters) Approximate hanging weight: 11 lbs each (5 kg) Power requirements 120V,.4 Amps each

More information

SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS

SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS ADVANTAGE PLUS SITE REQUIREMENTS SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS 50096-927 Rev A For additional information contact your Medivators Pg 1 of 18 SITE REQUIREMENTS The purpose of this section

More information

Pipe Cutting and Beveling Clamshells

Pipe Cutting and Beveling Clamshells Pipe Cutting and Beveling Clamshells Who We Are One Company, Total Support, Complete Solutions For more than a century, Hydratight has provided world-class bolted joint solutions and continues to set international

More information

Lower Level (Level L) Exhibit Halls A, B & C

Lower Level (Level L) Exhibit Halls A, B & C Floor Plans Wal t e r E. Was h i n g ton Con v e n t i on C e n t e r Lower Level (Level L) Exhibit Halls A, B & C Lower Level (Level L) Lower Level (Level L) Halls A, B & C Hall A 151,000 square feet

More information

CYMAQ Marking System AS200 + DS7050 N-14

CYMAQ Marking System AS200 + DS7050 N-14 CYMAQ Marking System AS200 + DS7050 N-14 Serial Number: 13716 CYMAQ www.cymaq.com INDEX 1. PRESENTATION AND GENERAL POINTS 1.1. General Points 1.2. Identification plate 2. MACHINE CHARACTERISTICS 2.1.

More information

Basic Specifications. Electrical. EPSON Perfection 3170. Color EPSON MatrixCCD TM line sensor. device Effective pixels

Basic Specifications. Electrical. EPSON Perfection 3170. Color EPSON MatrixCCD TM line sensor. device Effective pixels Scanner Parts Document table Power cord Document cover Transparency unit (TPU) under lid Web Transparency unit (TPU) connector TPU indicator light e-mail Carriage and lamp (in the home position) USB port

More information

NEW APC Smart-UPS. Advanced line interactive power protection for servers and network equipment. The world s most popular network and server UPS.

NEW APC Smart-UPS. Advanced line interactive power protection for servers and network equipment. The world s most popular network and server UPS. NEW APC Smart-UPS Advanced line interactive power protection for servers and network equipment. > The world s most popular network and server UPS. Award-winning Smart-UPS protects critical data by supplying

More information

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide

R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide R&S ZVA-Z75, -Z110, -Z140, -Z170, -Z220, -Z325, -Z500 Converters Quick Start Guide (=7ÔWÌ) 1307.7039.62 06 Test & Measurement Quick Start Guide This Quick Start Guide describes the following converter

More information

VSO MAX HP High Flow Proportional Valve

VSO MAX HP High Flow Proportional Valve VSO MAX HP High Flow Proportional Valve Typical Applications Ventilators Insufflators Anesthesia Delivery & Monitors Pressure & Flow Control Mass Flow Control Product Specifications Physical Properties

More information

CRITICAL POWER MONITORING CPM20/CPM21 DETAILS

CRITICAL POWER MONITORING CPM20/CPM21 DETAILS CRITICAL POWER MONITORING CPM20/CPM21 DETAILS The Starline Critical Power Monitor (CPM) is a distributed data acquisition system that enables Current and Power monitoring in busway systems. Each phase

More information

AIS 3000 APC AIS 3000 15kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass

AIS 3000 APC AIS 3000 15kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass AIS 3000 APC AIS 3000 15kVA 208V w/2 Batt. Module Exp. To 4, Start-Up 5X8, Internal Maintenance Bypass APC AIS 3000, 12 kw / 15 kva,input 208V 3PH / Output 120V, 208V, 208V 3PH, Interface Port DB-9 RS-232,

More information

Spetec. Laminar Flow Box. Portable Clean Room Technology

Spetec. Laminar Flow Box. Portable Clean Room Technology Spetec Laminar Flow Box Portable Clean Room Technology 1 Fundamentals Portable Clean Room Technology Laminar Flow The basis for the development of the clean room technology was first established in the

More information

Back-UPS Pro 1300/1500 Installation and Operation

Back-UPS Pro 1300/1500 Installation and Operation Back-UPS Pro 1300/1500 Installation and Operation Inventory Safety Do not install the Back-UPS in direct sunlight, in excessive heat, humidity, or in contact with fluids. Connect the battery bu059a bu058a

More information

Cisco 7816-I5 Media Convergence Server

Cisco 7816-I5 Media Convergence Server Cisco 7816-I5 Media Convergence Server Cisco Unified Communications Solutions unify voice, video, data, and mobile applications on fixed and mobile networks, enabling easy collaboration every time from

More information

QuickSpecs. Models HP Server Console Switches

QuickSpecs. Models HP Server Console Switches Overview are key components for rack mount implementations (a console represents one keyboard, video monitor, and mouse (KVM) attached to a server). The family of allows multiple servers in a rack to be

More information

Installation Guide. 5 Series 15 W. Undercounter/Freestanding Nugget Ice Machine

Installation Guide. 5 Series 15 W. Undercounter/Freestanding Nugget Ice Machine Installation Guide 5 Series 15 W. Undercounter/Freestanding Nugget Ice Machine Table of Contents Warnings & Important Safety Instructions 3 Dimensions (Professional) 5 Dimensions (Custom Panel) 6 Specifications

More information

CONTENTS. PAGE Chapter 1 General Information 1. Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13

CONTENTS. PAGE Chapter 1 General Information 1. Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13 CONTENTS PAGE Chapter 1 General Information 1 Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13 Chapter 3 Alarm/Excitation Options 15 Wiring 17 Chapter

More information

RHE12. Hazardous Area Coriolis Mass Flow Transmitter. Features. Applications. Benefits

RHE12. Hazardous Area Coriolis Mass Flow Transmitter. Features. Applications. Benefits RHE12 Hazardous Area Coriolis Mass Flow Transmitter Features Field mounting Compact, pressure safe housing ATEX and CSA approvals for installation in hazardous areas 24 VDC power supply Configurable analog

More information

KNC Model 3666 Automatic Pressure Calibration System

KNC Model 3666 Automatic Pressure Calibration System KNC Model 3666 Automatic Pressure Calibration System Product Description The Model 3666 Automatic Pressure Calibration (APC) System by King Nutronics Corporation is a portable secondary standards laboratory

More information

HDO700 P FIBRE OPTIC TRANSMITTER

HDO700 P FIBRE OPTIC TRANSMITTER Timo Rantanen 18.1.2011 1(5) HDO700 P FIBRE OPTIC TRANSMITTER HDO700 P is a high performance, extremely linear externally modulated 1550 nm transmitter for HDO fibre optic CATV link. This transmitter type

More information

Dell Vostro 3350/3450/3550/3750

Dell Vostro 3350/3450/3550/3750 Dell Vostro 3350/3450/3550/3750 Setup And Features Information About Warnings WARNING: A WARNING indicates a potential for property damage, personal injury, or death. Front and Back View Vostro 3350 Figure

More information

Computer Gateway Specification and Technical Data

Computer Gateway Specification and Technical Data L Computer Gateway Specification and Technical Data CG03-400 9/92 detergant coffee chocolate Page 2 TDC 3000 Computer Gateway Specification and Technical Data Introduction This publication defines the

More information

Flow Injection Analysis

Flow Injection Analysis Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since

More information

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002

AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 First Edition MONITORING AND LABORATORY DIVISION August

More information

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES TSI and TSI logo are registered trademarks of TSI Incorporated. SmartTune is a trademark of TSI Incorporated. THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES IFA 300 Constant Temperature Anemometry

More information

A reasonable, new furnace (mid efficiency) has the following electrical requirements:

A reasonable, new furnace (mid efficiency) has the following electrical requirements: June 2007 Power Backup Systems for Your Home. As promised from a previous safety meeting, a couple of options to protect you home in the event of a power failure. These options are for the most serious,

More information

Electric Vehicle Charging Station Products EVoReel Charge Cable Management Solutions

Electric Vehicle Charging Station Products EVoReel Charge Cable Management Solutions Self-Retracts 30ft Cable Length Electric Vehicle Charging Station Products EVoReel Safe, Convenient & Clean EV Charging EVoReel products are engineered to provide improved safety and convenience for electric

More information

Table Of Contents 6 6 6 7 7 8 8 9-11 12 12-13 14 15 16 17 18 19 20-21 22 22 23 24 24 24

Table Of Contents 6 6 6 7 7 8 8 9-11 12 12-13 14 15 16 17 18 19 20-21 22 22 23 24 24 24 USER S GUIDE Table Of Contents Package Contents Accessories: Sold Separately Product Specifications Intended Use Compatibility ResMed Device Warning Battery Pack Run Times Power Backup Usage Getting Started

More information

Video conferencing. solutions. Communication was never that easy,

Video conferencing. solutions. Communication was never that easy, 35 Video conferencing solutions Communication was never that easy, with the Audipack video conferencing solutions. Furnitures, camera drawers and camera lift systems, for every situation the most perfect

More information

Airflow Sensors Line Guide

Airflow Sensors Line Guide Airflow Sensors Line Guide Go with the flow of engineering leadership. All airflow sensors operate on heat transfer flow and differential pressure. But Honeywell Sensing and Control (S&C) offers advanced

More information

MEASURING & CONTROL UNITS FOR PUBLIC SWIMMING POOLS

MEASURING & CONTROL UNITS FOR PUBLIC SWIMMING POOLS MEASURING & CONTROL UNITS FOR PUBLIC SWIMMING POOLS Measuring and Controlling Equipment ensuring required water quality in public pools. This is fully automatic measuring and controlling equipment with

More information

RESOURCES. Telehealth. Provincial Health Services Authority. Telehealth & General Meeting Room Design Requirements

RESOURCES. Telehealth. Provincial Health Services Authority. Telehealth & General Meeting Room Design Requirements RESOURCES Telehealth Provincial Health Services Authority Telehealth & General Meeting Room Design Requirements Prepared by IMITS, LMC Telehealth and C&W Media Services December, 2013 Revised, August 2014

More information

HP UPS R1500 Generation 3

HP UPS R1500 Generation 3 HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications

More information

Everything you need for protection scheme testing

Everything you need for protection scheme testing Power System Simulator for Testing Protection Relays and Schemes Everything you need for protection scheme testing The is the only instrument with the high power, flexibility and software to perform full

More information

QuickSpecs. Overview. HP Power Distribution Rack. HP Power Distribution Rack. DA - 12859 North America Version 5 August 22, 2008 Page 1

QuickSpecs. Overview. HP Power Distribution Rack. HP Power Distribution Rack. DA - 12859 North America Version 5 August 22, 2008 Page 1 Overview It's getting harder to manage a data center efficiently, with rising utility bills and ever-increasing demands due to processor power and server density increases. Growing user requests mean heavier

More information

Electrical Connection. Provide Proper Electrical Supply. Electrical Requirements. Connecting using the supplied cord and plug

Electrical Connection. Provide Proper Electrical Supply. Electrical Requirements. Connecting using the supplied cord and plug Electrical Connection When installed the range must be electrically grounded in accordance with local codes or; in the absence of local codes with the National Electrical Code ANSI/NFPA 70, latest edition.

More information

AXIS PS12 Outdoor Power Supply. Installation Guide

AXIS PS12 Outdoor Power Supply. Installation Guide AXIS PS12 Outdoor Power Supply Installation Guide AXIS PS12 Outdoor Power Supply Connect the AXIS PS12 Outdoor Power Supply This guide provides instructions for connecting and preparing the AXIS PS12

More information

Fiber Optic Distributed Temperature Sensor (B-DTS)

Fiber Optic Distributed Temperature Sensor (B-DTS) Fiber Optic Distributed Temperature Sensor (B-DTS) Low-cost Brillouin BOTDA scattering version For more information about our strain and temperature sensor system and related products, please visit www.ozoptics.com

More information

06 / 2003. Optional components. MAXDATA PLATINUM Server-Rack PLATINUM. [Product Guide] www.maxdata.de

06 / 2003. Optional components. MAXDATA PLATINUM Server-Rack PLATINUM. [Product Guide] www.maxdata.de D Positioning Key-Features Specifications Optional components MAXDATA PLATINUM Server-Rack PLATINUM [Product Guide] www.maxdata.de Positioning Description Server-Racks with 41U, 37U and 23U Highlights

More information

RiOs 3, 5, 8 Water Purification Systems

RiOs 3, 5, 8 Water Purification Systems RiOs 3, 5, 8 Water Purification Systems Reliable and affordable pure water solutions Merck Millipore is a division of Reliable and affordable pure water solutions Your water purification needs Reliable,

More information

Specifications 17/17R

Specifications 17/17R 17/17R Specifications 2013 Dell Inc. Trademarks used in this text: Dell, the DELL logo, and Inspiron are trademarks of Dell Inc. Intel, Pentium, and Celeron are registered trademarks and Core is a trademark

More information

Area is a measure of how much space is occupied by a figure. 1cm 1cm

Area is a measure of how much space is occupied by a figure. 1cm 1cm Area Area is a measure of how much space is occupied by a figure. Area is measured in square units. For example, one square centimeter (cm ) is 1cm wide and 1cm tall. 1cm 1cm A figure s area is the number

More information