Virtual CT Simulator QA 2 topics Electromechanical. Virtual Simulator CT QA. What is Virtual CT Simulator (VSim)? Original Simulators
|
|
- Loraine Shelton
- 7 years ago
- Views:
Transcription
1 Virtual CT Simulator QA 2 topics Electromechanical Virtual Simulator CT QA Virtual CT simulation software Performance and Image quality Alex Markovic, PhD Director, Medical Radiation Physics Program, Rosalind Franklin University, North Chicago, IL Lead Therapeutic Physicist, NorthShore University Health System, Evanston, IL CT operator console Alex Prasad 2 Original Simulators What is Virtual CT Simulator (VSim)? TG66:Quality assurance for computed-tomography simulators and the computed tomographysimulation process X ray tube A virtual simulator is a set of software which recreates the treatment machine and which allows import, manipulation, display, and storage of images from CT and/or other imaging modalities. Image intensifier 3 4
2 Examples of VSim software packages: Varian SomaVision GE Advantage SIM Elekta Group - CMS Focal Phillips AcQSim 3 Siemens syngo RT Oncologist Phillips AcQSim VSim Software Features DRR generation Automatic structure segmentation Interpolation of contours Manual and semi-automatic (eg. smart brush ) contouring tools Volume expansion to apply margins 4D Visualization of organ motion Isocenter placement, design of treatment fields Moving laser support 5 6 Advantages of VSim Relevant Guidance Documents No guidance document dedicated to Vsim QA Efficient workflow Doctors are usually present during the CT scan Good time to have them complete their contouring and planning work. Poor scans can be repeated Doctors work is less likely to be delayed Vsim software can be installed on doctor s PC or can be accessed through Citrix TG 2(1993): Specification and acceptance testing of CT scanners TG 53(1998) QA of treatment planning NCRP report #99: Quality Assurance for Diagnostic Imaging TG 66 (2003): QA for CT-simulators and CT simulation process TG 117 (in progress): The Use of MRI Data in Treatment Planning and Stereotactic Procedures Spatial Accuracy and Quality Control Procedures TG132 (In progress): Use of Image Registration and Data Fusion Algorithms and Techniques in Radiotherapy 7 Treatment Planning 8
3 Why Perform QA on the VSim? Ensure data to and from VSim transfers correctly Reduce possibility of treatment errors due to inaccurate transfer of data Ensure complete dataset is sent VSim Quality Management Program Acceptance testing - manufacturer procedure, if it exists Commissioning - resident physicist establishes procedures baseline values, mostly for electromechanical/imaging system components Periodic QA ensures consistency of baseline values End of warranty tests - 6 weeks prior to end of warranty 9 Post service/upgrade tests 10 Data Transfer from CT to VSim Node Setup VSim IP address Data transfer VSim AE Title CT scanner DICOM export configuration screen 11
4 Data Transfer Issues trouble in paradise Data Transfer from VSim to TPS DICOM RT Information standard to move images, etc VSim hosts file VSim hosts file does not include CT scanner IP Routers are mac layered Dicom service has stopped on VSim IT has secretly upgraded the router When in doubt reboot! Images: CTs, DRRs Structure sets: Points of interest, bolus, 3D objects Plan parameters: Tx fields, tolerance tables, Rx, patient orientation, fractionation, machine ID, accessories 14 Network Transfer Glossary SOP: service object pair (CT -> Vsim) SCP: Service Class Provider (eg VSim) SCU: Service Class User (eg CT) ftp: file transfer protocol UID: Unique identifier (transfer session ID) Data Transfer issues from VSim Eg: Beam to energy TPS does not transfer GE Advantage sim DICOM conformance statement XiO TPS DICOM conformance statement 15 16
5 Patient orientation Machine definition Movable lasers Structure segmentation DRR accuracy 4DCT SRS 17 patient orientation Ensure images are transferred to VSim and TPS correctly and they display correct orientation Implications: Reversed beam orientation Contours drawn on wrong side of patient HFS FFS HFP 18 patient orientation CT screen Incorrect patient orientation TPS screen 19 machine definition Available beam energies Position and limits of jaws Collimator rotation MLC type (52,58, , 120 HD, m3, 160 leaf) Angle convention (IEC , IEC 1217 or non-iec) Tolerance levels should be tight Implications are numerous: missing or incorrect beam parameters transferred to TPS 20
6 machine definition machine definition machine definition - verification movable lasers Mechanical system Absolute, relative position accuracy Scan phantom with radio-opaque reference point and mm markings Test system through its maximum mechanical extents Implications: incorrect isocenter markings which may not be caught until film is reviewed 1) Set up beams in VSim 2) Transfer to TPS 3) Verify Parameters: Machine/energy scale gantry angle rotations field size etc TPS 23 24
7 movable lasers Scales Radiopaque markers structure delineation Bifurcating structures or disconnected structures Contouring in sagittal cuts Contour interpolating Double contours (external skin) Copy structures, expansion, Boolean operations Unclosed contours Accurate volume reporting structure delineation contour interpolation structure delineation double contours Ensure correct position TPS single contour TPS - Shifted external contour axial VSim - Before interpolation VSim -After interpolation Does the TPS allow double contours? TPS - Shifted external contour sagittal 27 28
8 structure delineation volume reporting Contour object of known dimensions in VSim Verify volume, dimensions Ensure same volume, dimensions are reported in TPS Set CT image level to ½ of object s max, set window to <50 Implication: incorrect DVH reporting VSim Incorrect windowing TPS 29 DRR accuracy Physicians rely on accurate DRRs to: delineate blocking set field size dimensions view organ projections Implications: Under or over coverage of disease Inadequate sparing of healthy tissue 30 DRR accuracy VSim acceptance testing 4DCT Scan plate between plastic slabs Radiopaque markers AP DRR on VSim workstation Acquires >10 images per couch position Accurate tracking of motion surrogate Analyze motion in three planes 31
9 VSim Acceptance testing 4DCT Test to verify that total deflection of movement is being acquired by 4DCT system Should be done in 3 axes Implications: incorrect size of MIP (maximum intensity projection) mm measure d deflection VSim Acceptance testing SRS/SBRT Image quality High contrast resolution Low contrast detectability, noise Image fusion Accuracy of fusion module MRI distortion 34 Use of VSim in SRS Image fusion - distortion Use of VSim in SRS Image fusion - distortion TG 117 (in progress): The use of MRI Data in Treatment Planning and Stereotactic Procedures Spatial Accuracy and Quality Control Procedures TG132 (In progress): Use of Image Registration and Data Fusion Algorithms and Techniques in Radiotherapy Treatment Planning MRI test phantom Implications of image distortion: - Incorrect anatomical rendering Acceptable registration MRI CT Poor registration Deformable/translational or translational only QA of MRI to test image distortion Evaluate fusion with MRI/CT phantom Fused image sets 36
10 VSim Periodic QA Questions? Low likelihood of VSim parameters changing over time Yearly: repeat acceptance tests Focus on testing after upgrades and external system changes Next Talk: CT Sim: Electromechanical, image quality Guru Prasad 39
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
More informationTotal Solutions. Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA phone 412 312 6700 800 70 NOMOS www.nomos.com
Serial Tomotherapy IGRT Total Solutions Treatment Planning Brachytherapy Imaging Best NOMOS One Best Drive, Pittsburgh, PA 15202 USA BMI.v.08.2010 Best NOMOS Your Single Source Oncology Solutions Provider
More informationDICOM: Definitions and Testing
DICOM: Definitions and Testing Bruce Curran University of Michigan (AAPM Representative, DICOM WG-7 Walter R. Bosch Advanced Technology Consortium Washington University of St. Louis Acknowledgements Sam
More informationELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS. Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic
ELECTRONIC MEDICAL RECORDS (EMR) STREAMLINE CH I PROCESS Ping Xia, Ph.D. Head of Medical Physics Department of Radiation Oncology Cleveland Clinic Disclosure I have received research grants from Siemens
More informationQuality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
More informationOur Department: structure and organization
EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled
More informationNIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning
NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,
More informationRapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
More informationkv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
More informationDICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
More informationFailure Modes and Effects Analysis (FMEA)
Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide
More informationManual VIRTUAL RADIATION ONCOLOGY CLINIC (VROC) (1.1) Radiation Oncology Training. Director User Manual (1.1)
Manual 1 VIRTUAL RADIATION ONCOLOGY CLINIC (VROC) (1.1) Radiation Oncology Training Director User Manual (1.1) Virtual Radiation Oncology Clinic Resident User manual Chapter 1 Creating the Virtual Patient
More informationKey words: treatment planning, quality assurance, 3D treatment planning
American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical
More informationINTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS
INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS HOW DOES RADIATION THERAPY WORK? Cancer is the unrestrained growth of abnormal cells within the body. Radiation therapy kills both
More informationPost Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y
Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How
More informationQuality Assurance of accelerators; the technologists responsibility
Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section
More informationProstate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021
Prostate IMRT: Promises and Problems Chandra Burman, Ph.D. Memorial Sloan-Kettering Cancer Center, New York, NY 10021 Introduction Prostate is one of the treatment sites that is well suited for IMRT. For
More informationQuality Assurance in Stereotactic. Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationMedical Device Correction
Medical Device Correction Field Safety Notice Re.: Digital Linear Accelerators of type Primus TM, ONCOR TM and ARTISTE TM running Control Console software version 12.0.25 or 13.0.65 and syngo RT Therapist
More informationInformatics Systems Overview. R Dahl*, M Herman Mayo Foundation Rochester, MN
Welcome! Informatics Systems Overview R Dahl*, M Herman Mayo Foundation Rochester, MN Objectives Understand basic network infrastructure for both LAN s s and WAN s Understand interfaces to both hospital
More informationM D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
More informationDepartment of Radiation Oncology
Department of Radiation Oncology Welcome to Radiation Oncology at Emory Clinic Every member of Emory Clinic Department of Radiation Oncology strives to provide the highest quality of care for you as our
More informationTHE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
More informationAnalysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients
대한방사선종양학회지 2005;23(3):169~175 Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients Department of Radiation Oncology, College of Medicine, Ewha Womans University
More informationImplementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
More informationChapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY W. PARKER, H. PATROCINIO Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 7.1. INTRODUCTION
More informationPrecise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care
Precise Treatment System Clinically Flexible Digital Linear Accelerator Personalized radiotherapy solutions for everyday treatment care Can you offer personalized cancer care? As more and more people require
More informationmultimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images
multimodality image processing workstation Visualizing your SPECT-CT-PET-MRI images InterView FUSION InterView FUSION is a new visualization and evaluation software developed by Mediso built on state of
More informationAccreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
More informationPerformance evaluation and quality assurance of Varian enhanced dynamic wedges
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department
More informationChapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy
Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review
More informationQuality Reports With. PlanIQ. *Export only. 510(k) pending. Patent pending. Your Most Valuable QA and Dosimetry Tools
Quality Reports With PlanIQ * *Export only. 510(k) pending. Patent pending. Your Most Valuable QA and Dosimetry Tools Objective Plan Assessment? How good are my treatment plans? Quality assurance of radiation
More informationTarget Volumes for Anal Carcinoma For RTOG 0529
Target Volumes for Anal Carcinoma For RTOG 0529 Robert Myerson, M.D. Ph.D.*, Lisa Kachnic, M.D.**, Jacqueline Esthappan, Ph.D.*,Parag Parikh M.D.*William Straube M.S.*, John Willins, Ph.D.** *Washington
More informationDiagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
More informationHead and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD
Good Morning Head and Neck Treatment Planning: A Comparative Review of Static Field IMRT Rapid Arc Tomotherapy HD Barbara Agrimson, BS RT(T)(R), CMD Steve Rhodes, BS RT(T), CMD Disclaimer This presentation
More informationFirst Three Years After Project Proton Therapy Facility:
39. Project Related Utilization and Financial Information. Proton Therapy Facility: Unit of Service Delivery Projected Utilization FY2017 FY2018 FY2019 Estimated Patients 448 670 808 Average Revenue Per
More informationThinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
More informationRadiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
More informationDICOM- an overview with an emphasis on Therapy
DICOM- an overview with an emphasis on Therapy R. Alfredo C. Siochi, PhD Outline a. What is DICOM? i. Standard ii. Applications b. Making the connection i. TCP/IP + DICOM ii. Configuring Clients and Servers
More informationMLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline
Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50
More informationPurchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage
More informationHDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014
HDR Brachytherapy 1: Overview of QA Bruce Libby Department of Radiation Oncology University of Virginia Health System Disclosures: Nondisclosure agreement with Varian Brachytherapy Shareholder- Varian,
More informationIGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
More informationQA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
More information1. Provide clinical training in radiation oncology physics within a structured clinical environment.
Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible
More informationTASK INVENTORY ARRT BOARD APPROVED: JULY 2015 IMPLEMENTATION DATE: JANUARY 2017
Radiation Therapy Certification and registration requirements for radiation therapy are based, in part, on the results of a comprehensive practice analysis conducted by The American Registry of Radiologic
More informationTISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
More informationHITACHI Proton Beam Therapy System
HITACHI Proton Beam Therapy System Masumi Umezawa, M.S. Hitachi, Ltd., Hitachi Research Laboratory HITACHI Proton Beam Therapy System Contents 1. Overview of HITACHI 2. HITACHI Proton Beam Therapy and
More informationPhysics and Dosimetry of SBRT: Simulation and Localization. Moyed Miften, PhD University of Colorado
Physics and Dosimetry of SBRT: Simulation and Localization Moyed Miften, PhD University of Colorado Team Kelly Stuhr Bodo Reitz Quentin Diot Tracey Schefter Laurie Gaspar Brian Kavanagh Rad Onc Staff at
More information4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
More informationClinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION
Clinical Rotation 3: PHYS 705 Fall 2015 (Aug. 25, 2015 to Feb. 25, 2016) COURSE INFORMATION Days: Monday-Friday Times: Full Time Location: One of the participating cancer clinics Program Director: George
More informationCHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY
Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY WILLIAM PARKER Department of Medical Physics McGill
More informationDaily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT
Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Wen Li, Andrew Vassil, Lama Mossolly, Qingyang Shang, Ping Xia Department of Radiation Oncology Why
More informationRadiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements
PRIMARY CERTIFICATION DIDACTIC AND CLINICAL COMPETENCY REQUIREMENTS EFFECTIVE JANUARY 2014 Radiation Therapy 1. Introduction Candidates for certification and registration are required to meet the Professional
More informationA PACS-Aware DICOM Image Object
Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry
More informationData Flow and Management in Radiation Therapy
Data Flow and Management in Radiation Therapy Issues in Data Flow and Data Management in Radiation Oncology DICOM- an Overview with an Emphasis on Therapy - R. Siochi Issues in Data Flow Management in
More informationTom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015
INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design
More informationMOC Exam Preparation Therapy
MOC Exam Preparation Therapy R. Alfredo C. Siochi, PhD., DABR Overview MOC Exam Requirement Available Guidance What I did What I wished I did 1 MOC Exam Requirement Rolling Window of 10 years Must pass
More informationThe Medical Physicist as Computer Geek: Custom Software in the Clinic
The Medical Physicist as Computer Geek: Custom Software in the Clinic Brian Tonner, Ph.D. Moffitt Cancer Center Tampa, Florida Florida AAPM Fall Meeting Sarasota Florida, November 2015 NO CONFLICTS OF
More informationGated Radiotherapy for Lung Cancer
Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego sbjiang@ucsd.edu radonc.ucsd.edu/research/cart Two Types of Gating Internal gating
More informationDIFFERENT FROM THE REST FOCUSED ON THE FUTURE. A guide to understanding TomoTherapy treatments
DIFFERENT FROM THE REST FOCUSED ON THE FUTURE A guide to understanding TomoTherapy treatments Advanced treatment planning Integrated, daily CTrue imaging 360º treatment delivery A platform for truly personalized
More informationPath Finder. CMA Engineering Designing & Consulting Inc. Miami, Florida
Path Finder CMA Engineering Designing & Consulting Inc. Miami, Florida Table of Contents I. General Purpose II. III. Contents of the Path Finder Box Path Finder I. General Purpose This QA Tool for beam
More informationVIRTUAL SIMULATION FOR RADIATIOTHERAPY TREATMENT USING CT MEDICAL DATA
VIRTUAL SIMULATION FOR RADIATIOTHERAPY TREATMENT USING CT MEDICAL DATA Dr. Stelios Zimeras (zimste@aegean.gr) University of the Aegean, Department of Statistics and Assurance Sciences, 83200 Karlovasi
More informationVirtual Reality Training for Radiotherapy becomes a Reality
Virtual Reality Training for Radiotherapy becomes a Reality R PHILLIPS a,1, J W WARD a, L PAGE a, C GRAU b, A BOJEN b, J HALL c, K NIELSEN b, V NORDENTOFT b, A W BEAVIS d a Department of Computer Science,
More informationOncentra Brachy Comprehensive treatment planning for brachytherapy. Time to focus on what counts
Oncentra Brachy Comprehensive treatment planning for brachytherapy Time to focus on what counts Time to focus on what counts In contemporary brachytherapy, the medical physicist needs to process an increasing
More informationSpecifications Varian Treatment Streamlined Treatment Delivery Management Application. Specifications
Specifications Varian Treatment Streamlined Treatment Delivery Management Application i Varian Treatment Streamlined Treatment Delivery Management Application Specifications Specifications Varian Treatment
More informationAssessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
More informationWHERE IN THE WORLD JILL LIPOTI?
WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain
More informationRadiosurgery 8/16/2011. Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago
Intracranial Stereotactic Radiosurgery (SRS) and Stereotactic Radiotherapy (SRT) Kamil M. Yenice, PhD University of Chicago Radiosurgery The use of radiation as a surgical tool Small volumes of tissues
More informationIAEA-TECDOC-1540. Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems April 2007 IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems
More informationNEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM
NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM Purpose: To provide licensees and registrants with a
More informationInnovative RT SBRT. The variables with REQ in superscript are required.
The variables with REQ in superscript are required. The variables with a are single-select variables; only one answer can be selected. The variables with a are multi-select variables; multiple answers
More informationMoving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
More informationPhysics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationUsers Group Meeting 2011 Northeastern Society of Orthodontists 89th Annual Meeting November 11-14, 2010 Hilton Montréal Bonaventure Montréal, Quebec, Canada 14 Lecture Schedule Saturday continued w Doctor
More informationOptimization of image quality through online motion correction
Advanced imaging technologies a revolution in diagnosis and treatment Optimization of image quality through online motion correction Oline Vinter Olesen December 10, 2010 Medical Solutions, Siemens Institute
More informationGE Healthcare. pet/ct for simulation. Precision in motion.
GE Healthcare pet/ct for simulation. Precision in motion. Patient profile H: 5 10 W: 370 lbs BMI: 53 Each cancer patient is just as unique as his or her cancer. Approaches to cancer treatment are changing
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationMIM USER GUIDELINES FOR TROG QUALITY ASSURANCE DOSE REVIEW
MIM USER GUIDELINES FOR TROG QUALITY ASSURANCE DOSE REVIEW Version 3 May 2015 Contents 1 Welcome... 3 2 TROG QA Support... 3 3 Opening MIM... 3 3.1 Access to CQMS... 3 3.2 Access to MIM via Citrix... 4
More informationCTVision System. Accurate Imaging for Precise Treatment. www.siemens.com/medical
CTVision System Accurate Imaging for Precise Treatment www.siemens.com/medical Adapt in Real Time A radiation therapy treatment plan is a snapshot in time. But as time moves on, changes in patient anatomy
More informationLasers. Lasers. Lasers. Lasers for. for Patient. Lasers. Lasers. Alignment - Room
for for for for for for for for for for for for Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient Patient Alignment Alignment Alignment Alignment Alignment Alignment
More informationChapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy
Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN 92-0-107304-6):
More informationIntensity-Modulated Radiation Therapy (IMRT)
Scan for mobile link. Intensity-Modulated Radiation Therapy (IMRT) Intensity-modulated radiotherapy (IMRT) uses linear accelerators to safely and painlessly deliver precise radiation doses to a tumor while
More informationIn room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers
In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der
More informationLiberate Your Image Data
Liberate Your Image Data Does Adherence to the DICOM Standard Guarantee Interoperability? 2009-2014 Mach7 Technologies, Inc. Page 1 of 9 Table of Contents Abstract...3 The DICOM Standard Does Not Guarantee
More informationA Three-Dimensional Correlation Method for Registration of Medical Images in Radiology
A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami
More informationAutomated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging
Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging Shaun S. Gleason, Ph.D. 1, James Goddard, Ph.D. 1, Michael Paulus, Ph.D. 1, Ryan Kerekes, B.S.
More informationRADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
More informationRIS. Complete -Complete exam on RIS *** [! - 9 ] ***
PACS And The Technologist Gen Rad/CR - Workflow 1 GEN RAD/CR - Workflow CR Processing RIS Arrival - Patient Arrived Order Request Prints *** [! - 1 ] *** Verify -Verify Exam and Resource (Correct if needed)
More informationGE Healthcare. Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together
GE Healthcare Centricity PACS and PACS-IW with Universal Viewer* Where it all comes together The healthcare industry is going through an unprecedented period of change with providers being called upon
More information3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationQuality Assurance for Treatment Planning Systems
Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline
More informationA Thesis. entitled. Retrofitted Micro-MLC SRS System. George H. Hancock
A Thesis entitled Development of a Comprehensive Linac-based Quality Assurance Program for a Retrofitted Micro-MLC SRS System. by George H. Hancock Submitted to the Graduate Faculty as partial fulfillment
More informationPage: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
More informationRADIOTHERAPY Giovanna Mantello
F The impact on RADIOTHERAPY Giovanna Mantello AOU Osp. Riuniti Ancona gio@mobilia.it RADIOTHERAPY OF PULMONARY LESIONS INTENSITY MODULATION HIGH GRADIENT DOSE DOSE ESCALATION IPOFRACTIONATION STEREO BODY
More informationDENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
More informationARIA ONCOLOGY INFORMATION SYSTEM RADIATION ONCOLOGY
ARIA ONCOLOGY INFORMATION SYSTEM RADIATION ONCOLOGY The ARIA oncology information system is a powerful information and image management solution designed to support the specific clinical needs of oncology
More informationTwelve. Figure 12.1: 3D Curved MPR Viewer Window
Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from
More information