Quality Assurance and Quality Control of Film Screen Mammography. Xuan Ho, Ph.D., R.T. (R)

Size: px
Start display at page:

Download "Quality Assurance and Quality Control of Film Screen Mammography. Xuan Ho, Ph.D., R.T. (R)"

Transcription

1 Quality Assurance and Quality Control of Film Screen Mammography Xuan Ho, Ph.D., R.T. (R)

2 Quality Assurance & Quality Control Quality Assurance Procedures Daily Weekly Monthly Quarterly Semiannual Annual

3 MAMMOGRAPHIC QC TESTS Darkroom cleanliness Daily Processor quality control Daily Mobile unit quality control Daily Screen cleanliness Weekly Viewboxes & viewing conditions Weekly Phantom images Weekly Visual checklist Monthly Repeat analysis Quarterly Analysis of fixer retention in film Quarterly Darkroom fog Semi-annually Screen-film contact Semi-annually Compression Semi-annually

4 Daily Quality Assurance Procedures Darkroom Processor

5 Daily Quality Assurance Procedures Darkroom The purpose of quality assurance for the darkroom, is to assure the cleanliness of it, the darkness (no light leaks), and the brightness of the red light. According to ACR regulations the darkroom is to cleaned each day. The darkroom needs to be free of dust and dirt. Any ventilation in to the room, can not be directly over the countertop. There should not be any storage above the countertop, and there should be no eating, drinking or smoking in this room.

6 Daily Quality Assurance Procedures Processor The purpose of quality assurance on the mammography processor, is to assure that the processor and chemistry system are working in a consistent manner. A specific box of mammography film, and cassette are dedicated to test the daily Q.A. of the processor. A sensitometer is used to expose film with different levels of densities, on the Q.A. film. A densitometer is used to read the levels of density on the film, after the film has been developed through the processor. This indicates whether the chemicals are in balance with the film.

7 Daily Quality Assurance Procedures Sensitometer

8 Daily Quality Assurance Procedures Densitometer

9 Daily Quality Assurance Procedures Processor (continued) When reading the sensitometric strip on the Q.A. film, there are three dedicated steps on the strip that are read each day. Depending on the facility, these steps can vary. There are low, medium and high density steps to be read. The low-density (LD) is subtracted from the high-density (HD), giving what is called the density difference (DD), or contrast.

10 Daily Quality Assurance Procedures Processor (continued) There is also a mid-density (MD) read, giving what is called the speed of the film. There is also one more measurement made, and that is the base-plus-fog. This is read on the clear portion of the Q.A. film.

11 Daily Quality Assurance Procedures Sensitometric Strip

12 Daily Quality Assurance Procedures Processor (continued) Corrective measures can be taken if the speed and density-difference are not with in limits. If the contrast of the film is too high, starter can be added to the developer, to bring it down. If the temperature of the processor is high, make sure the water has been turned on. If the temperature is too low, then service needs to be called, to change the parameters.

13 Daily Quality Assurance Procedures Processor (continued) If the speed of the film is off, make sure the processing time has not changed. Last, but not least, make sure the box of dedicated Q.A. film has not changed.

14 Weekly Quality Assurance Procedures Screen cleanliness Viewbox maintenance Phantom images

15 Weekly Quality Assurance Procedures Screen cleanliness is done on a weekly basis. Darkrooms processors, and manipulation of cassettes help contaminate mammography screens. This is to assure that mammographic screens are free of dust and dirt particles that may degrade image quality or mimic microcalcifications.

16 Weekly Quality Assurance Procedures To clean screens, chose a clean location. Check screens for dirt, dust, lint, pencil marks, fingernail polish and other marks under a bright light. Cleans screens using manufacturer s recommended materials and procedures. After cleaning with liquid cleaners, screens should be allowed to air dry by standing cassettes vertically, on edge, and partially open.

17 Weekly Quality Assurance Procedures Cleaning Screens (continued) Inspect the screen and cassette cover for any stray particles of dust before reloading the cassette with film. Wait at least 15 minutes or manufacturer s recommendation time after loading and before using. Be sure to document on the Q.A. log sheet that this weekly task has been performed. (Date & Tech initials)

18 Weekly Quality Assurance Procedures View Box Maintenance According to ACR regulations this is done to assure that the viewboxes and viewing conditions are optimized and maintained at an optimal level. This should be carried out weekly.

19 Weekly Quality Assurance Procedures View Box Maintenance (continued) To do so, clean view box surfaces using window cleaner and soft paper towels (magnifiers should be cleaned at the same time). Visually inspect the view boxes for uniformity and luminance. Assure that all view box masking equipment is functioning properly and easily. Visually check the room illumination levels and assure that sources of bright light are not present in the room and are not being reflected from the view box surface. Be sure to document the date and tech initials, when this procedure has been done.

20 Weekly Quality Assurance Procedures Phantom images This weekly task is performed to assure that film optical density, contrast (density difference), uniformity, and image quality due to the X-ray imaging system and film processor are maintained at optimum levels. The mammographic phantom is equivalent to a 4.2cm thick compressed breast consisting of 50% glandular and 50% adipose tissue.

21 Weekly Quality Assurance Phantom images Procedures It is important to use the same cassette each time this test is done. It is also important to use the #2 photocell on the image receptor, the same view box, same technologists eyes and same technical factors, each time. This keeps all factors consistent.

22 Weekly Quality Assurance Procedures Optical disc This is a mammography phantom, with the optical disc sitting on top of it.

23

24 Weekly Quality Assurance Phantom (continued) When the designated cassette is loaded, place the phantom (with the disc on a place not obscuring the fibers, masses or specks), on the image receptor, flush with the chest wall edge of the bucky, and center it. Bring the compression paddle down, to touch the top of the phantom disc. Make an exposure and develop the film. Make sure to note the technique used.

25 Weekly Quality Assurance Phantom (continued) After the film is developed, use the sensitometer to measure the background optical density (the disc) and the density difference, by measuring right next to the disc. Subtract the optical density from the density just to the left or right of the disc. This gives the density difference.

26 Weekly Quality Assurance Phantom images (continued) Plot the mas, optical density (disc), and the DD (area measured next to the disc). Next, the technologist will then view the number of fibers, masses and specks seen on the phantom and document them on the control chart. This should be done under optimal viewing conditions.

27 Weekly Quality Assurance Phantom images (continued) When scoring the image, it is important to note that the largest whole mass, fiber or group of specks should be viewed first, and given a value of 1. If they are seen, but partially seen, they are given a value of.5. If less than half of the item is seen, then a value of 0 is given and documented. A minimum of 4 fibers, 3 speck groups and 3 masses must be seen, to pass.

28 Monthly Quality Assurance Procedures Visual checklist This is done on a monthly basis to assure that mammographic X-ray system indicator lights, displays, mechanical locks and detents are working properly and that the mechanical rigidity and stability of the equipment is optimum. To do this procedure, review all of the items listed on the visual checklist and indicate the status. Be sure to rotate the C-arm the way you would for patient imaging. Date and initial the checklist where indicated.

29 Quarterly Quality Assurance Procedures Repeat Analysis Fixer Retention Test

30 Quarterly Quality Assurance Repeat analysis Procedures This test is performed quarterly (every three months) to determine the number and cause of repeated mammograms and rejected films. Analysis of these data will help identify ways to improve efficiency and reduce costs as well as patient exposures. Start by disposing of all existing rejected films in the department.

31 Quarterly Quality Assurance Procedures Repeat analysis (continued) Take inventory of film supply as a starting point to determine total number of films consumed during this test. Start to collect all rejected films into the categories on the repeat-reject analysis sheet used for the test.

32 Quarterly Quality Assurance Procedures Repeat analysis (continued) Films may be repeated for a number of reasons including poor positioning, patient motion, being too light or dark, or artifacts. Some films may not have resulted in additional exposure to the patient, but may have been rejected. (clear films or Q.C. films) Although it is important to include wire localization films as part of the reject analysis, they should not be included in the repeat analysis because they are taken as part of the wire localization process. Tabulate the counts and determine the total number of repeated films, rejected films, and the total number exposed.

33 Quarterly Quality Assurance Procedures Repeat Analysis (continued) Determine the overall percentage of repeated/rejected films by dividing the total number of repeated films by the total number of exposed during the analysis period, and multiply by 100. Document this number on the analysis sheet. Determine the percentage of repeats in each reason for repeat category by dividing the repeats in the category by the total number of repeated films and multiply by 100. Document this number on the analysis sheet.

34 Quarterly Quality Assurance Procedures Repeat Analysis (continued) MQSA requirements for repeat analysis say the repeat/reject rates can not rate by more than 2%. If they are, corrective action must take place.

35 Quarterly Quality Assurance Procedures Repeat analysis documentation form.

36 Quarterly Quality Assurance Procedures Fixer Retention Test This test is done quarterly to determine the quantity of residual fixer (hypo) in processed film as an indicator of keeping quality. Residual hypo indicated insufficient washing and considerably degrades image stability.

37 Quarterly Quality Assurance Procedures Fixer retention kit

38 Quarterly Quality Assurance Procedures Fixer Retention test (continued) To do this test, a hypo estimator and a hypo test solution are needed. First process one sheet of unexposed film in the mammography processor. Place one drop of the hypo test solution on the emulsion (less shiny) side of the film. Allow the solution to stand for 2 minutes. Blot off the excess solution. Place the film on a sheet of white paper.

39 Quarterly Quality Assurance Procedures Fixer Retention test (continued) Compare the stain with a hypo estimator strip by looking at the radiographic film on a sheet of white paper. The comparison should be made with the estimator over the film sample to help compensate for differences in the color of the base of the film and the hypo estimator strip in its sleeve for protection. Record the results on the monthly, quarterly and semiannual checklist

40 Semiannual Quarterly Assurance Procedures Darkroom Fog test Film-screen Contact test Compression test

41 Semiannual Quarterly Assurance Procedures Darkroom fog test This test is done to assure the darkroom safelights and other light sources inside and outside of the darkroom do not fog mammographic films. To perform this test, a mammographic phantom, densitometer, mammographic film, one sheet of each type of film used for mammography in the darkroom (from a new box), an opaque card, and a watch or timer are needed.

42 Semiannual Quality Assurance Procedures Darkroom fog test continued First assure that all safelight filters are those specified by the film manufacturer, and do not appear faded or cracked. Also assure that the wattage and type of the light bulbs and the distance from the safelights to the film-handling surface are those specified by the film manufacturer. Turn out all lights in the darkroom and wait 5 minutes for the eyes to adjust. Look for obvious light leaks around doors, around passbox and processor, and in the ceiling. Light leaks may be quite directional. Be careful in moving around a darkroom without safelight illumination.

43 Semiannual Quality Assurance Procedures Darkroom Fog Test (continued) In total darkness, load a new sheet of film into the cassette, and X-ray a phantom (like normal procedures). In total darkness, remove the exposed film from the cassette and place it on the countertop, emulsion side up. Cover one half of the film with an opaque sheet, set perpendicular to the chest wall edge of the film. Be sure the opaque card is in contact with the film, so no light leaks.

44 Semiannual Quality Assurance Procedures Darkroom fog test (continued) Turn on the safelight and let the film and opaque card lie on the countertop for 2 minutes. Process the film. Using the densitometer, measure the the density of the unfogged portion of the image. Then measure the density of the fogged portion of the image. To determine the amount of fog, subtract the unfogged value from the fogged value. Record this optical density difference on the monthly, quarterly, semiannual checklist. MQSA standards indicate that darkroom fog shall not exceed 0.05.

45 Semiannual Quality Assurance Procedures Film-screen contact test This test is done semiannually to assure that optimum contact is maintained between the screen and film in each cassette. In order to do this test a copper screen, 40 mesh (40 wires per inch), mammographic film of appropriate size, a densitometer, a sensitometer, screens and cassettes to be tested. First, all the cassettes to be tested must be cleaned. Afterwards, load film and let the air out of the cassettes by waiting at least 15 minutes before proceeding.

46 Semiannual Quality Assurance Procedures Film-screen contact test tool

47 Semiannual Quality Assurance Procedures Film-screen contact test (continued) Place the cassette to be tested on top of the imaging receptor, with the edge of the cassette flush with the chest-wall edge of the image receptor. Place the copper screen on top of the cassette. Place the compression paddle as high towards the X-ray tube as possible.

48 Semiannual Quality Assurance Procedures Film-screen contact test (continued) Select a manual technique (25 to 28kVp) that will provide a film density between 0.70 and 0.80 when measured in the image over the mesh near the chest-wall edge.

49 Semiannual Quality Assurance Procedures Film-screen contact test tool on top of the image receptor, with the compression paddle all the way up.

50 Semiannual Quality Assurance Procedures Film-screen contact test (continued) Expose and process the film. Repeat this process for each cassette. View the films on a viewbox at a distance of at least 3 feet. Look for areas of poor contact, i.e. darker areas in the mesh image. The cassettes that pass can be placed into clinical use.

51 Semiannual Quality Assurance Procedures Film-screen contact test (continued) Select those cassettes that did not pass and repeat the above process of cleaning etc. Those that do not pass, can not be used. MQSA requires a 40 mesh copper screen. All cassettes being used, need to be tested every six months.

52 Semiannual Quality Assurance Procedures Compression test This test is performed semiannually to assure that the mammography system can provide adequate compression in the manual and powered mode and that the equipment does not allow too much compression to be applied. Adequate compression is essential for high-quality mammography. Compression reduces the thickness of tissue that must be penetrated by radiation, thereby reducing scattered radiation and increasing contrast, while reducing radiation exposure to the breast. For this test a flat analog type bathroom scale and several towels or a sponge are needed.

53 Semiannual Quality Assurance Procedures

54 Semiannual Quality Assurance Procedures Compression test (continued) To do this test in power mode, place a towel on the cassette holder then place the bathroom scale on the towel with the dial and read-out positioned for easy reading. Locate the center of the scale directly under the compression device. Place more towels, or sponges on top of the scale for protection. Using the initial power drive, activate the compression device and allow it to operate until it stops automatically. Read and record the compression force on the monthly, quarterly and semiannual checklist.

55 Semiannual Quality Assurance Procedures Compression test (continued) Release the compression device. For manual mode, move the compression device downward until it stops, manually. Read and record the compression force on the monthly, quarterly and semiannual checklist. Release the compression device. MQSA requirements state that a compression force of at least 111 newtons (25 pounds) shall be provided. The maximum compression force for the initial power drive must be between 111 newtons (25 pounds) and 200 newtons (45 pounds). If the compression test fails, the mammography unit shall not be used until the problem has been corrected.

56 Annual Quality Assurance Procedures Physicist s survey Inspection by federally certified or state-certified inspectors Qualification standards

57 Annual Quality Assurance Procedures Physicist s survey is done once per year. A licensed physicist is required to test the mammography equipment at a minimum of once per calendar year. The physicist records and prepares a report of the findings, according to MQSA regulations. A copy of this document is kept with all of the quality assurance/control records, for inspection.

58 Annual Quality Assurance Procedures Inspection by federally certified or statecertified inspectors: This is done once per calendar year. The inspector reviews all records of Q.A. tests done through out the year, and certification and continuing education of personnel (mammographers and radiologists). If any of the tests are not in compliance with MQSA requirements, the facility will lose it s license to perform mammography procedures. If any personnel are not current with continuing education or licensure, they can not perform mammography until corrected.

59 Annual Quality Assurance Procedures Qualification standards MQSA specifies that a lead interpreting physician be identified by the mammography facility to have the general responsibilities of ensuring that all MQSA-required activities are met. This individual will most likely be the lead mammography radiologist. All radiologists must be compliant in their training in mammography and continuing education, in order to interpret mammograms.

60 Annual Quality Assurance Procedures Mammographers According to the ARRT, a licensed mammographer must complete a minimum of 12 continuing education units on a biennium basis. For the state of California, the continuing education requirement is 15 ceu s per every three years. Each state varies with their requirements. In order to stay current in mammography, a technologist must perform 100 mammograms each year, according to the ARRT.

61 Annual Quality Assurance Procedures Medical physicists The medical physicist must under go extensive education and licensure. He must be current in his education and licensure in order to be compliant with MQSA requirements.

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality

More information

Annual MQSA Inspection Questions NMQAAC April 19, 2004

Annual MQSA Inspection Questions NMQAAC April 19, 2004 1 1 1 1 1 1 1 1 0 1 0 1 0 1 0 1 Annual MQSA Inspection Questions NMQAAC April 1, 00 1.0 Inspection Information 1.1 Name and Address 1. Equipment Registration.0 Facility Inspection Download.0 Facility Inspections.0

More information

MIN-R 2000 Plus Film

MIN-R 2000 Plus Film Technical Data Sheet Issued 2014-03 MIN-R 2000 Plus Film MIN-R 2000 Plus Film is a medium speed, dual coated, orthochromatic medical x-ray film for mammographic use with single green-emitting intensifying

More information

Quality Control of Full Field Digital Mammography Units

Quality Control of Full Field Digital Mammography Units Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of

More information

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes. IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641

More information

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).

More information

QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 3 RADIOGRAPHIC OR FLUOROSCOPIC MACHINES

QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 3 RADIOGRAPHIC OR FLUOROSCOPIC MACHINES CRCPD Publication 01-6 $20.00 QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 3 RADIOGRAPHIC OR FLUOROSCOPIC MACHINES July 2001 Published by Conference of Radiation Control Program Directors,

More information

Quality Control and Maintenance Programs

Quality Control and Maintenance Programs Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment

More information

QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 1 DENTAL FACILITIES

QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 1 DENTAL FACILITIES CRCPD Publication 01-4 $20.00 QUALITY CONTROL RECOMMENDATIONS FOR DIAGNOSTIC RADIOGRAPHY VOLUME 1 DENTAL FACILITIES July 2001 Published by Conference of Radiation Control Program Directors, Inc. This page

More information

Aspire FCRm/CRm Fujifilm CR for Mammography

Aspire FCRm/CRm Fujifilm CR for Mammography : May 2013 Fujifilm Medical Systems U.S.A., Inc. 2 CONTENTS FOREWORD... 5 Introduction to Measurements... 6 S value... 7 Overview of Testing... 8 Section A Getting Started... 9 0 Set-up & Baseline Images...10

More information

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis

More information

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,

More information

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ

More information

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.

More information

KODAK Dental Radiography Series. Quality Assurance in Dental Radiography DENTAL PRODUCTS

KODAK Dental Radiography Series. Quality Assurance in Dental Radiography DENTAL PRODUCTS KODAK Dental Radiography Series Quality Assurance in Dental Radiography DENTAL PRODUCTS Quality Assurance in Dental Radiography A complete quality assurance program is easy to establish and maintain. Quality

More information

The Mammography Quality Standards Act Final Regulations Document #1

The Mammography Quality Standards Act Final Regulations Document #1 Compliance Guidance The Mammography Quality Standards Act Final Regulations Document #1 Document issued on: March 19, 1999 This document supersedes document Draft Compliance Guidance August 27, 1998 U.S.

More information

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp

More information

American College of Radiology Mammography Accreditation Program

American College of Radiology Mammography Accreditation Program American College of Radiology Mammography Accreditation Program Objectives Overview of the ACR Mammography Accreditation Program Interrelationship between mammography accreditation, FDA certification and

More information

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D. Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start eric.berns@gmail.com Overview and compare QC tests

More information

Update on ACR Digital Mammography QC Manual

Update on ACR Digital Mammography QC Manual Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual

More information

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION

More information

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes

Radiographic Image Production. Radiographic Image Production. Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Principles of Imaging Science I (RAD 119) Film, Screens, and Cassettes Radiographic Image Production X-ray photons emitted from the x-ray tube interact with the body, exit the patient (exit beam) and interact

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Digital Breast Tomosynthesis QC Requirements

Digital Breast Tomosynthesis QC Requirements Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist

More information

Update on MQSA and ACR Mammography Accreditation

Update on MQSA and ACR Mammography Accreditation MQSA - Who s Who Update on MQSA and ACR Mammography Accreditation The Law: Mammography Quality Standards Act (MQSA) The Regulator: US Food and Drug Administration (FDA) Pamela L. Platt, BSRT(R)(M)(CV)

More information

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015)

The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) The American College of Radiology Stereotactic Breast Biopsy Accreditation Program: Frequently Asked Questions (Updated: August 24, 2015) Table of Contents Stereotactic Breast Biopsy Accreditation Program

More information

MQSA Quality Control Manual for Monochrome Displays for Mammography

MQSA Quality Control Manual for Monochrome Displays for Mammography MQSA Quality Control Manual for Monochrome Displays for Mammography Ver1.0 Applicable Models LMD-DM50, LMD-DM30 Sony Corporation - 1 - MQSA Quality Control Manual for Monochrome Displays for Mammography

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)

Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition) Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition) New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral

Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come

More information

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,

More information

2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014

2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014 2014 Meeting of Mid-Atlantic States Radiation Control Programs March 25, 2014 Interpreting Physicians Final Requirements for Interpreting Physicians Meet Initial Qualifications License Certification OR

More information

QUALITY ASSURANCE IN DENTAL RADIOGRAPHY. Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000

QUALITY ASSURANCE IN DENTAL RADIOGRAPHY. Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000 QUALITY ASSURANCE IN DENTAL RADIOGRAPHY Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000 Quality Assurance (QA) Ensure consistently adequate diagnostic information whilst keeping doses

More information

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection

Mammography. For 35+ years, screenfilm has been the gold standard for breast cancer detection Digital Mammography Upright Stereo Mammography For 35+ years, screenfilm has been the gold standard for breast cancer detection IN THE BEGINNING Mammography technology has come a long way since the first

More information

X-ray 100KV. Speed as compared to that of type IX100 used as a standard 100 under each exposure conditions.

X-ray 100KV. Speed as compared to that of type IX100 used as a standard 100 under each exposure conditions. DATA SHEET INDUSTRIAL RADIOGRAPHIC SYSTEMS RADIOGRAPHIC FILM IX20, IX25, IX50, IX80, IX, IX, IX29 AND IX59 Fujifilm industrial X-ray films feature revolutionary new film technology. The combination of

More information

The Mammography Quality Standards Act Final Regulations Document #4; Final Guidance for Industry and FDA

The Mammography Quality Standards Act Final Regulations Document #4; Final Guidance for Industry and FDA The Mammography Quality Standards Act Final Regulations Document #4; Final Guidance for ndustry and FDA Document issued on: [use release date of FR Notice] This document supersedes Draft Compliance Guidance:

More information

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding

More information

Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH)

Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH) Analysis of Causes for Reject X-Ray Films as a Quality Assurance Element in Diagnostic Radiology in Jos University Teaching Hospital(JUTH) By S.O ANIKOH Dept. of Physics, University of Jos, Jos. HABILA

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

SAFELIGHT FILTERS AND DARKROOM LAMPS

SAFELIGHT FILTERS AND DARKROOM LAMPS TECHINCAL INFORMATION SAFELIGHT FILTERS AND DARKROOM LAMPS A RANGE OF HIGH QUALITY FILTERS AND LAMPS FOR DARKROOM ILLUMINATION DESCRIPTION ILFORD manufacture a range of high quality Safelight Colour Use

More information

Elements of required physical infrastructures: space, schielding, and patient flow..

Elements of required physical infrastructures: space, schielding, and patient flow.. Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto

More information

KODAK PLUS-X AEROGRAPHIC Film 2402

KODAK PLUS-X AEROGRAPHIC Film 2402 AS-45 KODAK PLUS-X AEROGRAPHIC Film 2402 KODAK PLUS-X AEROGRAPHIC Film 2402 is a panchromatic, negative camera film with medium speed, medium-high contrast, and extended red sensitivity for the reduction

More information

(1) Interpreting physicians. All physicians interpreting mammograms shall meet the following qualifications:

(1) Interpreting physicians. All physicians interpreting mammograms shall meet the following qualifications: DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration 21 CFR Part 900 Quality Mammography Standards Subpart B Quality Standards and Certification Sec. 900.12 Quality standards. (a) Personnel.

More information

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization

More information

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 Version 3.1 Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2 MQSA Quality Control Manual for Monochrome Monitors for Mammography Foreword Purpose of this Document This document is for

More information

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised

VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised Page 1 of 6 instructions VC 104+ Rigid Grade / VC 104+ Rigid Grade Imagine / (for the rest of this document referred to as VC 104) can be applied to new and used trucks and trailers with painted rigid

More information

Introduction to Color Process Monitoring For Minilabs

Introduction to Color Process Monitoring For Minilabs Introduction to Color Process Monitoring For Minilabs KODAK Publication No. Z-99M Eastman Kodak Company, 2011 1 Contents WHY SHOULD YOU MONITOR YOUR PROCESS?... 3 HOW DOES A COLOR PROCESS WORK?... 3 WHAT

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Recommendations for a technical quality control program for diagnostic X-ray equipment

Recommendations for a technical quality control program for diagnostic X-ray equipment Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer

More information

Checklist for submission of application form for Registration

Checklist for submission of application form for Registration Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status

More information

Scanners and How to Use Them

Scanners and How to Use Them Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types

More information

Agfa Materials Printed Circuit Boards. High Resolution Film. HRR & HRO Technical Information

Agfa Materials Printed Circuit Boards. High Resolution Film. HRR & HRO Technical Information ab Agfa Materials Printed Circuit Boards High Resolution Film HRR & HRO Technical Information April 27 HRR AND HRO Technical Information HRR is a red-sensitive line film. It is especially designed for

More information

Basics to Using the View Camera

Basics to Using the View Camera Walker Evans at work Using the View Camera Basics to Using the View Camera Because of the large-scale nature of the view camera, it is necessary to follow basic steps in setting-up the camera in order

More information

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)

More information

Color Filters and Light

Color Filters and Light activity 4 Color Filters and Light OBJECTIVES Students add to their understanding of subtractive color mixing by investigating the effect of filters on the color of light. The students shine white light

More information

PROCESSING THE RADIOGRAPH. Neill Serman. Sept. 2000 W + P. Chapter 6

PROCESSING THE RADIOGRAPH. Neill Serman. Sept. 2000 W + P. Chapter 6 1 PROCESSING THE RADIOGRAPH. Neill Serman. Sept. 2000 W + P. Chapter 6 THE STAGES IN THE PRODUCTION OF THE RADIOGRAPH Many dentists who utilize manual processing, intentionally over expose patients and

More information

MQSA, FDA Regulations, and Inspections

MQSA, FDA Regulations, and Inspections The American College of Radiology Mammography Accreditation Program: Frequently Asked Questions (Updated: December 15, 2015) Table of Contents MQSA, FDA Regulations, and Inspections... 2 ACR Mammography

More information

Darkroom Design for Amateur Photographers

Darkroom Design for Amateur Photographers TECHNICAL DATA / REFERENCE March 2005 AK-3 Darkroom Design for Amateur Photographers When you take a pleasing photograph, develop the film, and make a print, there is a lot of satisfaction in knowing the

More information

Module Connectors and Cable Specifications

Module Connectors and Cable Specifications Module Connectors, page 1 Cables and Adapters, page 2 Cleaning the Fiber-Optic Connectors, page 6 Module Connectors RJ-45 Connector The RJ-45 connector is used to connect a Category 3, Category 5, Category

More information

ABHES BULLETIN. Subject: Final Radiologic Technology Standards Effective January 1, 2013

ABHES BULLETIN. Subject: Final Radiologic Technology Standards Effective January 1, 2013 ACCREDITING BUREAU OF HEALTH EDUCATION SCHOOLS 7777 Leesburg Pike, Suite 314 N. Falls Church, Virginia 22043 Tel. 703/917.9503 Fax 703/917.4109 E-Mail: info@abhes.org ABHES BULLETIN To: From: ABHES-Accredited

More information

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Standard Test Method for Classification of Film Systems for Industrial Radiography 1 Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately

More information

PREVENTIVE MAINTENANCE OF FIBER OPTIC CABLES AND OPTICS

PREVENTIVE MAINTENANCE OF FIBER OPTIC CABLES AND OPTICS PREVENTIVE MAINTENANCE OF FIBER OPTIC CABLES AND OPTICS The end of a fiber optic cable and the inner surface of an optical module lens constitute optical surfaces that should be properly cleaned and maintained

More information

IMAGING & THERAPEUTIC TECHNOLOGY

IMAGING & THERAPEUTIC TECHNOLOGY IMAGING & THERAPEUTIC TECHNOLOGY Quality Assurance in Mammography: Artifact Analysis 1 Jacquelyn P. Hogge, MD Carolyn H. Palmer, RTR(M) Caroline C. Muller, RTR(M) Sherrill T. Little, MD Deborah C. Smith,

More information

ISO 400/27 BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE

ISO 400/27 BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE TECHNICAL INFORMATION o ISO 400/27 BLACK AND WHITE PROFESSIONAL FILM FOR HIGH PRINT QUALITY AND FLEXIBILITY IN USE ILFORD HP5 Plus is a fast black and white film. It is ideal for action, available light

More information

Wooden Shipping Crate is Available to Protect Your UV Light Source. 11 X 14 8-1" diameter tubes 120 VAC - 2.2 Amps - 265 Watts

Wooden Shipping Crate is Available to Protect Your UV Light Source. 11 X 14 8-1 diameter tubes 120 VAC - 2.2 Amps - 265 Watts UV Light Source General Information & Frequently Asked Questions Delivery Times Usually, you will have a unit sitting on your front door step within two weeks anywhere in the contiguous United States for

More information

Detector transparent with Color Inserts. FAA 500 TR P Trim Ring transparent with Color Inserts. FCA 500 / FCA 500 E Detector Bases

Detector transparent with Color Inserts. FAA 500 TR P Trim Ring transparent with Color Inserts. FCA 500 / FCA 500 E Detector Bases Detector Color Detector transparent with Color Inserts FAA 500 TR W Trim Ring FAA 500 TR P Trim Ring transparent with Color Inserts FAA 500 BB Ceiling Mount Back Box FCA 500 / FCA 500 E Detector Bases

More information

TECHNISCHE FICHE - TECHNISCHE FICHE - Chat Board: Classic

TECHNISCHE FICHE - TECHNISCHE FICHE - Chat Board: Classic - TECHNISCHE FICHE - Chat Board: Classic Classic is a magnetic glass- and writing board for use in modern offices and private homes. CHATBOARD Classic is Danish Design - made in Denmark. Classic is a glass

More information

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes. Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists

More information

Guide to Care and Maintenance

Guide to Care and Maintenance Village of Mount Prospect Guide to Care and Maintenance of Your Apartment Sponsored ed by: y: Village of Mount Prospect Envir nvironmental Health Division with the assistance of The Concerened ened Owners

More information

Auditing a Printed Circuit Board Fabrication Facility Greg Caswell

Auditing a Printed Circuit Board Fabrication Facility Greg Caswell Auditing a Printed Circuit Board Fabrication Facility Greg Caswell Introduction DfR is often requested to audit the PCB fabrication process of a customer s supplier. Understanding the process variations

More information

INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer

INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer INTENSIFYING SCREENS, CASSETTES AND SCREEN FILMS N. Serman & S. Singer X-rays were discovered by W.C. Roentgen because of their ability to cause fluorescence. X-ray photons cannot be seen. The image produced

More information

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

Lamp and Sleeve Replacement

Lamp and Sleeve Replacement Lamp and Sleeve Replacement G, H, J & K G Plus, H Plus, J Plus & K Plus Pro0, Pro0, Pro30 & Pro50 Always shut-off water flow and release water pressure before servicing. Electrical hazard. Be sure to unplug

More information

1.4 Metallic Impurities Iron must not exceed 150 ppm and copper shall not exceed 6 ppm when tested according to TAPPI T 266.

1.4 Metallic Impurities Iron must not exceed 150 ppm and copper shall not exceed 6 ppm when tested according to TAPPI T 266. Library of Congress Preservation Directorate Specification Number 100 132 09 Specifications for 3 x 10ʺ Paper Enclosures For the Storage of Strip Photographic Negatives This specification is provided as

More information

College of Radiology, Academy of Medicine of Malaysia Recommendations & Guidelines for Quality Assurance in MAMMOGRAPHY

College of Radiology, Academy of Medicine of Malaysia Recommendations & Guidelines for Quality Assurance in MAMMOGRAPHY College of Radiology, Academy of Medicine of Malaysia Recommendations & Guidelines for Quality Assurance in MAMMOGRAPHY 15 July 2008 These recommendations provide guidance to radiology and allied professionals

More information

KODAK PANATOMIC-X AEROGRAPHIC II Film 2412 KODAK PANATOMIC-X AERECON II Film 3412

KODAK PANATOMIC-X AEROGRAPHIC II Film 2412 KODAK PANATOMIC-X AERECON II Film 3412 AS-112 KODAK PANATOMIC-X AERECON II Film 3412 and KODAK PANATOMIC-X AERECON II Film 3412 are panchromatic, black-and-white aerial camera films with extended red sensitivity and intermediate speed. Both

More information

Chapter 15. Creating the Digital Image. Objectives. Key Terms

Chapter 15. Creating the Digital Image. Objectives. Key Terms Chapter 15 Creating the Digital Image Objectives Describe how a digital image is created and processed. Sequence the steps of image production and evaluation of the digital image. Identify the components

More information

X-RAY REGULATORY GUIDE

X-RAY REGULATORY GUIDE Minnesota Department of Health Radiation Control, X-ray Unit Protecting, maintaining and improving the health of all Minnesotans by promoting radiation safety through guidance and collaboration with the

More information

Workshop Self Inspection Checklist WORK01

Workshop Self Inspection Checklist WORK01 Workshop Self Inspection Checklist WORK01 General Mechanical Workshops Location Assessor Date This core checklist is provided by the Health and Safety Department as a template, which provides a basic framework

More information

Cleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9

Cleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9 Chapter 9 By the end of this chapter, you will be able to: l Define the terms cleaning, decontamination, disinfection l List the factors that affect disinfection l Describe the aims of a cleaning schedule

More information

Mammography Applications for MAMMOMAT 1000/3000 Nova

Mammography Applications for MAMMOMAT 1000/3000 Nova Mammography Applications for MAMMOMAT 1000/3000 Nova Introduction This booklet is intended as an application handbook for use with Siemens MAMMOMAT 1000 and 3000 Nova. The booklet contains advice on positioning,

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment

More information

SOLAR PV SYSTEM MAINTENANCE GUIDE

SOLAR PV SYSTEM MAINTENANCE GUIDE SOLAR PV SYSTEM MAINTENANCE GUIDE GUYANA HINTERLANDS STAND-ALONE SOLAR PV INSTALLATIONS IMPROVING HEALTH FACILITY INFRASTRUCTURE (IHFI) GUYANA CONTRACT NO. EPP-I-00-03-00008-00, TASK ORDER 07 APRIL 2013

More information

Routine Ultrasound Equipment Tests for Quality Assurance. Contents

Routine Ultrasound Equipment Tests for Quality Assurance. Contents Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI,

More information

How to Process and Print Black-and-White Film

How to Process and Print Black-and-White Film TECHNICAL DATA / BLACK-AND-WHITE FILM June 2005 AJ-3 How to Process and Print Black-and-White Film Processing and printing your own films can be rewarding and fun. With a little practice, you ll find it

More information

FOREWORD. The Plastics Pipe Institute. This Technical Note, TN-30, was first issued in January 2006 and was revised and republished in September 2013.

FOREWORD. The Plastics Pipe Institute. This Technical Note, TN-30, was first issued in January 2006 and was revised and republished in September 2013. Requirements for the Use of Rework Materials in Manufacturing of Polyethylene Gas Pipe TN-30/2013 105 Decker Court, Suite 825, Irving, TX 75062 P: 469-499-1044 F: 469-499-1063 www.plasticpipe.org FOREWORD

More information

Radiation safety in dental radiography

Radiation safety in dental radiography Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at

More information

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY

QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY AAPM REPORT NO. 74 QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY Report of Task Group #12 Diagnostic X-ray Imaging Committee Members S. Jeff Shepard, Chairman Pei-Jan Paul Lin, Co-Chairman John M. Boone Dianna

More information

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS guidelines for quality control testing for digital (cr DR) mammography THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS The Royal Australian and New Zealand College of Radiologists Eligibility

More information

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial

Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Mammography: Dosimetry ACMP Annual Meeting Virginia Beach Saturday, May 2, 2009 Lawrence N. Rothenberg, Ph.D. Department of Medical Physics Memorial Sloan-Kettering Cancer Center New York, NY rothenbl@mskcc.org

More information

OPERATOR MANUAL for the Kodak X-Omat 1000, 1000A, and 1000J PROCESSORS

OPERATOR MANUAL for the Kodak X-Omat 1000, 1000A, and 1000J PROCESSORS {OperatorsManual}{Production}{HealthImaging} Publication No. 9B8942 November 1997 Supersedes July 1997 HEALTH IMAGING OPERATOR MANUAL for the Kodak X-Omat 1000, 1000A, and 1000J PROCESSORS Eastman Kodak

More information

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE) SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE) We greatly appreciate your purchase of this Tamron lens. The Tamron SP AF 300mm F/2.8 LD [IF] is a fast telephoto lens developed for Canon AF single-reflex

More information

Department of Veterans Affairs VA HANDBOOK 5005/77 STAFFING

Department of Veterans Affairs VA HANDBOOK 5005/77 STAFFING Department of Veterans Affairs VA HANDBOOK 5005/77 Washington, DC 20420 Transmittal Sheet June 26, 2014 STAFFING 1. REASON FOR ISSUE: To revise the Department of Veterans Affairs (VA) qualification standard

More information

X-ray Imaging Systems

X-ray Imaging Systems Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures

More information

Radiology Services Policies and Procedures

Radiology Services Policies and Procedures Radiology Services Policies and Procedures November 2011 Table of Contents - Alphabetical Please Note! Clicking on a title in the Table of Contents will take you to that document. To find a particular

More information

FUJICOLOR CRYSTAL ARCHIVE DIGITAL PAPER TYPE DPII

FUJICOLOR CRYSTAL ARCHIVE DIGITAL PAPER TYPE DPII AF3-212E COLOR NEGATIVE PAPERS FUJICOLOR CRYSTAL ARCHIVE DIGITAL PAPER TYPE DPII 1. FEATURES AND USES FUJICOLOR CRYSTAL ARCHIVE DIGTAL PAPER TYPE DPII is a silver halide color paper designed exclusively

More information

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors

Health. NYC Quality Assurance Guidelines for. Primary Diagnostic Monitors Health NYC Quality Assurance Guidelines for Primary Diagnostic Monitors June 2015 Content I. Overview... 1 II. Definitions... 1 III. Acceptance Testing Requirements... 4 IV. Bi-weekly PDM Testing... 4

More information