Development of a Diffusive Sampler for Vapor-Phase Hydrogen Peroxide (VHP)
|
|
- Gwen Shelton
- 7 years ago
- Views:
Transcription
1 Development of a Diffusive Sampler for Vapor-Phase Hydrogen Peroxide (VHP) C.R. (Gus) Manning, PhD, CIH Assay Technology, Inc. Livermore, CA
2 Hydrogen Peroxide (VHP = Vapor Phase Hydrogen Peroxide) H 2 O 2 = HO-OH Consumer Pharmaceutical Antimicrobial Traditional Uses: Industrial Oxidizing Agent Aerospace Propellant (rocket fuel)
3 H 2 O 2 Safety Concerns (pre-2000) Traditional Uses of H 2 O 2 Used Low concentrations Well-contained Small number of exposed employees
4 H 2 O 2 Safety Concerns (post-2000) New Application! Medical Sterilizing Agent Sterilizing Agent with lower risk profile than Ethylene Oxide Glutaraldehyde Formaldehyde
5 Hydrogen Peroxide Common substance with unusual properties I I have a bottle in the bathroom. How bad can it be?
6 H 2 O 2 Safety Concerns (new) Medical Sterilization performed in 1000 s of hospitals, surgical centers, and medical mfg facilities Sterilization at high vapor levels (ca. 500 ppm) > 25,000 employees exposed REL = TLV = PEL = 1.0 ppm
7 Pre-Existing VHP Sampling Methods
8 Area Sampling Ion-Mobility Spectrometer Interference by methanol, bleach, SO 2 Cost = $50,000 Infra Red Spectrometry Interference by Water FTIR can avoid water interference Cost = $65,000 Honeywell TLD-1 1 (Chem Key) Paper tape reading device Colorimetric for H 2 O 2 Measures 0.1-3ppm Cost = $4,500
9 Personal Sampling in breathing zone OSHA ID-006 & ID126SG reaction TiOSO 4 + H 2 O 2 H 2 TiO 3 SO 4 yellow Air from Breathing Zone Sampling Pump Impinger Solution TiOSO 4 in dil. H 2 SO 4
10 OSHA ID-006 H 2 TiO 3 SO 4 - yellow solution absorption maximum at 410 nm analyze by UV Spectrophotometry H2O2 Calibration Curve Optical Absorbance y = x R 2 = H2O2 Found (microgram)
11 OSHA ID-125SG H 2 TiO 3 SO 4 - yellow solution highly oxidized species analyze by Pulse Polarography 10 Current (µamp) applied Volts vs SCE
12 Goals of Study Design Personal Monitoring Badge Cost-effective & convenient Generate VHP (vapor phase H 2 O 2 ) x PEL or higher PEL = 1 ppm Compare Badges versus Impingers Controlled Exposure Chamber
13 Development Approach Incorporate TiOSO 4 reagent in a solid phase Fiberglass Cellulose Silica Gel Incorporate solid phase into a Personal Monitoring Badge
14 Prototype H 2 O 2 Sampler Sampler Body (with snap cap) Sampling Grid (low SR) Sampling Grid (high SR) Spacer Wafer (treated with TiOSO4)
15 Optimal H 2 O 2 Sampler Side 1 of Sampler Body (with Sampling Grid) Collector Strip (treated with TiOSO4) Side 2 of Sampler Body (with Sampling Grid)
16 Badge and Impinger (Hydrogen Peroxide)
17 Generating H 2 O 2 Vapor for Testing (issues) Not available in gaseous form Available as a 3-50% 3 solution in water Strongly attracted to water
18 Hydrogen Peroxide Interactions H 2 O 2 b.p. 150 o C decomposition H 2 O + ½ O 2 b.p. 100 o C 50% H 2 O 2 Solution in water boils at 114 o C Heating
19 H 2 O 2 Generation Humidity Effects %RH 85%RH 50% RH 100% Efficiency H2O2 Concn (ppm) H2O2 Added (mg/min)
20 VHP Vapor Pressure is Limited at High Humidity Saturated VHP Concn % Relative Humidity
21 Vaporization Methods for Liquids Equilibration with Heated Liquid Air or N2 stream passes over heated liquid Liquid Injection through Heated Needle directly into test air stream Liquid Injection into Heated Chamber chamber located in test air stream
22 Test Air Passes Over Heated Liquid Test Air with Contaminant Concentration Test Air from Miler-Nelson Generator Liquid to be Vaporized (heated) Temp Sensor Heating Block Feedback - + Applied Voltage Highest Concentration Attainment (> 1,000 ppm) Liquid components separate by volatility (boiling point)
23 Heated Needle Vaporization Liquid to be Vaporized (metered from syringe pump) Applied Voltage Injection Needle Feedback Temp Sensor Resistance Wire (NiCr) Septum Test Air with Contaminant Concentration Test Air from Miler-Nelson Generator Vaporization Chamber Vaporizes 100% of liquid (does not separates components) Lowest Concentration Attainment (1-50 ppm)
24 Heated Chamber Vaporization Liquid to be Vaporized (metered from syringe pump) Injection Needle Test Air with Contaminant Concentration Septum Test Air from Miler-Nelson Generator Heating Mantle Vaporization Chamber (3-neck flask) Temp Sensor - + Applied Voltage Higher Concentration Attainment than Heated Syringe (to 500 ppm) Does not separate liquid components (if careful)
25 Accuracy of VHP Challenge Challenge Level vs Testing Time 250 VHP Challenge Level (ppm) Elapsed Time (hr)
26 Hydrogen Peroxide Evaluation of Personal Samplers
27 MNR Atmosphere Generator Concept of Side-by by-side Sampler Comparison Chemical Vapor Generator Air Intake ChemDisk ChemDisk Sampling Tube Sampling Tubes Sampling Pump ChemDisk Diffusive Samplers Exhaust Fan
28 Evaluating Air Samplers (schematic)
29 H 2 O 2 Analysis Analysis (low levels) OSHA ID 006 Impinger Sampling Rate = 1 L/min; Sampling Time = 40 min Optical Abs (410 nm) VHP Concn (ppm)
30 H 2 O 2 Analysis Personal Monitoring Badge Sampling Rate = 105 ml/min; Sampling Time = 4 hour Optical Abs (410 nm) VHP Concn (ppm)
31 Summary Vapor Phase Hydrogen Peroxide (VHP) was generated by injecting 50% aqueous H 2 O 2 solution into flowing system. Challenge Levels from ppm Flow Rates from L/min Low Levels of Vapor Phase Hydrogen Peroxide were monitored side-by by-side using OSHA ID 006 and a Personal Badge using similar chemistry. Operating range ppm
32 Conclusion Diffusive Samplers (Personal Monitoring Badges) based on TiOSO 4 chemistry perform similarly to the OSHA ID 006 method, but with increased convenience. Eliminate need for Glass Impingers Operating range ppm
33 Finis
Generation of Vapor-Phase Hydrogen Peroxide Test Atmospheres for the Evaluation of Respirator Cartridges and Air Samplers
Generation of Vapor-Phase Hydrogen Peroxide Test Atmospheres for the Evaluation of Respirator Cartridges and Air Samplers C.R. (Gus) Manning, PhD, CIH Hydrogen Peroxide Common substance with unusual properties
More informationNITRIC OXIDE and NITROGEN DIOXIDE 6014
NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm
More informationAir Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike License. Your use of this material constitutes acceptance of that license and the conditions of use of materials on this
More informationSTERILIZATION AND DISINFECTION
STERILIZATION AND DISINFECTION Importance of hand washing shown by Semmelweis STERILIZATION A physical or chemical process that destroys or eliminates all forms of microbial life, including spores. A satisfactory
More informationCase Study: Successful Decontamination of an HVAC System s Ductwork Using Chlorine Dioxide Gas
Canadian Biosafety Symposium Date: June 1, 2009 Place: Delta Halifax, Halifax Nova Scotia Case Study: Successful Decontamination of an HVAC System s Ductwork Using Chlorine Dioxide Gas Mark A. Czarneski
More informationATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
More informationOnline Infrared Flue Gas Analyzer Gasboard 3000
Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume
More informationApplication Requirement
Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality
More informationSpectrophotometer - Milton Roy Spectronic 21D or equivalent.
3M COMPANY OCCUPATIONAL HEALTH AND ENVIRONMENTAL SAFETY DIVISION DETERMINATION OF FORMALDEHYDE VAPORS IN AIR USING 3M 3721 FORMALDEHYDE MONITORS May, 2002 SCOPE This procedure covers the method of collecting
More informationThe photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
More informationWHITE PAPER April 18, 2016
WHITE PAPER April 18, 2016 Nitrogen Dioxide Biodecontamination: A New, Effective and Cost-saving Option for Biodecontaminating Syringe Tubs Prior to the Filling Line Prepared by: David Opie, PhD Maura
More informationGraphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
More informationCalculations and OELs
Calculations and Dr. Peter Bellin, PhD, CIH EOH 466A The Occupational Environment Definitions Particulates: dust, fume, mist Vapors Gases Particle size considerations Inspirable Particulate Mass (IPM)
More informationFlow Injection Analysis
Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since
More informationMATERIAL SAFETY DATA SHEET
Material Safety Data Sheet:- DIHYDROGEN MONOXIDE Page No. 2 of 5. - - 1/7/2008. Date: January 7, 2008 Replaces MSDS Dated: MATERIAL SAFETY DATA SHEET Manufacturer(s) NFPA Designation 704 GENERIC (UBIQUITOUS)
More informationSterilization methods and equipment Lab 1-2
Sterilization methods and equipment Lab 1-2 PHT 434 Sterilization Sterilization a process that by which all viable M.O are removed or destroyed, based on a probability function. Sterilization concept Sterilization
More informationCartridge Change Schedule
3 Cartridge Change Schedule How to establish a Cartridge Change Schedule Exposure Assessment for Respirator Selection and Chemical Cartridge Changeout Schedule An exposure assessment is a process used
More informationHands on an infrared gas analyzer (IRGA) to measure photosynthesis
University of Toledo - Dept of Envir. Sci., EEES 6660/8960: Biophysical Processes of Ecosystems, Spring 2010 Hands on an infrared gas analyzer (IRGA) to measure photosynthesis Burkhard Wilske for Jiquan
More informationPPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Introduction The amount of dissolved oxygen in process water is continually gaining importance in many industries as a critical parameter
More informationPOLYDIMETHYLSILOXANE
POLYDIMETHYLSILOXANE Prepared at the 37th JECFA (1990), published in FNP 52 (1992) superseding specifications prepared at the 29th JECFA (1985), published in FNP 34 (1986). Metals and arsenic specifications
More informationDetermination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC
Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC Abstract: High performance liquid chromatography (HPLC) was used to determine the identity and concentrations of food dyes
More informationSENSORS FOR AIR QUALITY MONITORING MARCEL ZEVENBERGEN SR. RESEARCHER GAS AND ION SENSORS
SENSORS FOR AIR QUALITY MONITORING MARCEL ZEVENBERGEN SR. RESEARCHER GAS AND ION SENSORS Health and Lifestyle as Application Driver IMPROVE ILLNESS MANAGEMENT ENABLE HEALTH MANAGEMENT ( What if these systems
More information3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
More informationEmergency Telephone (800) 551-7176
MATERIAL SAFETY DATA SHEET SECTION 1 - PRODUCT & COMPANY IDENTIFICATION 1501 East Woodfield Road Suite 300 East Schaumburg, IL 60173 (847) 969-2700 Product Identifier: Chemical Name: Category: Mesna Detoxifying
More informationMulti-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
More information2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY
2nd/3rd Year Physical Chemistry Practical Course, Oxford University 2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY (4 points) Outline Spectrometry is widely used to monitor the progress
More informationDOW CORNING CORPORATION Material Safety Data Sheet
Page: 1 of 7 1. IDENTIFICATION OF THE SUBSTANCE AND OF THE COMPANY Dow Corning Corporation South Saginaw Road Midland, Michigan 48686 24 Hour Emergency Telephone: Customer Service: Product Disposal Information:
More informationKINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)
Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:
More informationCHEM 105 HOUR EXAM III 28-OCT-99. = -163 kj/mole determine H f 0 for Ni(CO) 4 (g) = -260 kj/mole determine H f 0 for Cr(CO) 6 (g)
CHEM 15 HOUR EXAM III 28-OCT-99 NAME (please print) 1. a. given: Ni (s) + 4 CO (g) = Ni(CO) 4 (g) H Rxn = -163 k/mole determine H f for Ni(CO) 4 (g) b. given: Cr (s) + 6 CO (g) = Cr(CO) 6 (g) H Rxn = -26
More informationDecontamination and Waste Management www.biosecurity.sandia.gov
Decontamination and Waste Management www.biosecurity.sandia.gov SAND No. 2006-3684C Sandia is a multiprogram laboratory operated by Sandia Corporation, a Lockheed Martin Company, for the United States
More informationSDS. GHS Safety Data Sheet. Alpha Associates, Inc. Alpha Style VR-R Plus PRODUCT AND COMPANY IDENTIFICATION. Manufacturer HAZARDS IDENTIFICATION
1 Page 1 of 5 PRODUCT AND COMPANY IDENTIFICATION Manufacturer 145 Lehigh Avenue Lakewood, NJ 871 Contact: Phone: Fax: Email: Web: James Palmer 732-634-57 732-634-143 jpalmer@alphainc.com www.alphainc.com
More informationHYDROGEN PEROXIDE SOLUTION 3%
MSDS Number: H4070 * * * * * Effective Date: 02/15/08 * * * * * Supercedes: 04/22/05 HYDROGEN PEROXIDE SOLUTION 3% 1. Product Identification Synonyms: Hydrogen Dioxide Solution, 3%; Hydrogen Peroxide Topical
More information22 APR 2010 <MSDS_ID>300088</MSDS_ID> MATERIAL SAFETY DATA SHEET. <Product> PHENOLPHTHALEIN INDICATOR SOLUTION # HR - 2497. </Product> Part#: N/A
22 APR 2010 300088 MATERIAL SAFETY DATA SHEET PHENOLPHTHALEIN INDICATOR SOLUTION # HR - 2497 Part#: N/A SECTION I SUPPLIER INFORMATION Common Name : PHENOLPHTHALEIN
More informationPREPARATION FOR CHEMISTRY LAB: COMBUSTION
1 Name: Lab Instructor: PREPARATION FOR CHEMISTRY LAB: COMBUSTION 1. What is a hydrocarbon? 2. What products form in the complete combustion of a hydrocarbon? 3. Combustion is an exothermic reaction. What
More informationNitrous Oxide: Hazards and Proper Use
1 Nitrous Oxide: Hazards and Proper Use James Kaznosky Manager, Env. & Occ. Safety Environmental Health and Safety What is Nitrous Oxide? 2 Nitrous oxide (N20) is nonflammable, colorless gas with pleasant,
More information(a) Method 1 Sample and Velocity Traverses for Stationary Sources.
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
More informationNANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For
NANYANG TECHNOLOGICAL UNIVERSITY School of Civil and Environmental Engineering CV2701: Laboratory 2A Laboratory Manual For Experiment Lab 2A-6(ENV) Wastewater Quality Analysis Location : Environment Laboratory,
More information1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
More information8001782 Owner s Manual
8001782 Digital Infrared Thermometer Owner s Manual Introduction This instrument is a portable, easy to use compact-size digital thermometer with laser sighting designed for one hand operation. The meter
More informationLUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)
October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for
More informationSelective Catalytic Reduction (SCR) and Diesel Exhaust Fluid (DEF) Training Module
Selective Catalytic Reduction (SCR) and Diesel Exhaust Fluid (DEF) Training Module DEF SCR Training Module Welcome to the Cummins Filtration DEF SCR training module. DEF & SCR systems are key to Cummins
More informationChemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
More informationA Report of Air Sampling for Mercury Vapor and Particulates During Fluorescent Tubes Disposal Operation Maywood, Illinois Survey date: 9/28/2000
A Report of Air Sampling for Mercury Vapor and Particulates During Fluorescent Tubes Disposal Operation Maywood, Illinois Survey date: 9/28/2000 Prepared for: Air Cycle Corporation 2000 South 25 th Avenue,
More informationSpectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
More informationUsing Carbon Dioxide as a Tracer Gas to Measure Air Change Rate in a Single Zone
Using Carbon Dioxide as a Tracer Gas to Measure Air Change Rate in a Single Zone C. Simmons, F. Boelter, G. Crawford, Boelter & Yates, Inc., Park Ridge, IL Presented by: Catherine E. Simmons, CIH Situation
More informationA Visual Method for the Detection of Arsenic 0 to 500 µg/l By Dan Kroll R&D Chemist. Hach Co., Loveland Colorado
Introduction: Arsenic is a common contaminant of ground water that has been found to adversely effect human health at levels as low as 10 µg/l. The current USEPA maximum contaminant level (MCL) for arsenic
More informationSingle-Use Flow-Through Cell ph FTC-SU-HP8
Single-Use Flow-Through Cell ph FTC-SU-HP8 SENSOR PROBES Instruction Manual Single-Use Flow-Through Cell ph FTC-SU-HP8 Specification: ph sensor stick integrated in a single-use flow-through cell Document
More information"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH
"ADOPTED STANDARDS FOR THE REGULATION OF MEDICAL WASTE" IN HEALTH CARE FACILITIES LICENSED BY THE MISSISSIPPI STATE DEPARTMENT OF HEALTH REGULATED MEDICAL WASTE "Infectious medical wastes" includes solid
More informationThe Properties of Water (Instruction Sheet)
The Properties of Water (Instruction Sheet) Property : High Polarity Activity #1 Surface Tension: PILE IT ON. Materials: 1 DRY penny, 1 eye dropper, water. 1. Make sure the penny is dry. 2. Begin by estimating
More informationExposure and Control of Ten High Priority MSHA Contaminants in Metal Mines. Presented by: Shannon E. Newton, MPH, CIH
Exposure and Control of Ten High Priority MSHA Contaminants in Metal Mines Presented by: Shannon E. Newton, MPH, CIH Topics What are these hazards? How can they harm you? Why do we care? How do we tell
More informationMATERIAL SAFETY DATA SHEET
Section I: IDENTIFICATION 1 OF 2 Date Prepared: 02/15/02 Date Revised*: 09/21/06 Product Name: Gram Stain, Lugol s Iodine Item Number: 400312, 400330, 400350, 400352, 400314, 400351, 400356, 400312B 400312A
More informationLecture 35: Atmosphere in Furnaces
Lecture 35: Atmosphere in Furnaces Contents: Selection of atmosphere: Gases and their behavior: Prepared atmospheres Protective atmospheres applications Atmosphere volume requirements Atmosphere sensors
More informationEMERGENCY PHONE: 1-800-364-3577 or (651) 737-6501 (24 hours)
Material Safety Data Sheet Copyright, 2012, 3M Company All rights reserved. Copying and/or downloading of this information for the purpose of properly utilizing 3M products is allowed provided that: (1)
More informationA Beer s Law Experiment
A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining
More informationUSING OPTICAL EMISSION SPECTROSCOPY TO IMPROVE EQUIPMENT UPTIME FOR AN AL2O3 ALD PROCESS *
USING OPTICAL EMISSION SPECTROSCOPY TO IMPROVE EQUIPMENT UPTIME FOR AN AL2O3 ALD PROCESS * JOHN LOO Samsung Austin Semiconductor 12100 Samsung Blvd. Austin, Texas 78754 * Presented at the AVS 5 th International
More informationMeasuring Manganese Concentration Using Spectrophotometry
Measuring Manganese Concentration Using Spectrophotometry Objectives To use spectroscopy to determine the amount of Manganese is an unknown sample. Scenario Your have just joined a "Green Team" at SMC
More informationGuidance on safe use of Autoclaves
Safety Office Reviewed: Dec 2013 Reviewed: DD Month Year Guidance on safe use of Autoclaves 1. Hazards & Uses There are several different types of hazard associated with the use of autoclaves. The main
More informationCorrosion of Copper in Water
Corrosion of Copper in Water Digby D. Macdonald Center for Electrochemical Science and Technology Department of Materials Science and Engineering Pennsylvania State University 209 Steidle Bldg University
More informationGas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
More information************************************************************************ *** SECTION 1 PRODUCT IDENTIFICATION
MATERIAL SAFETY DATA SHEET Pure Tech Emergency Phone Numbers: P.O. Box 1950 CHEMTREC: 800/424-9300 Brewster, New York 10509-8950 POISON CENTER: 800/562-8236 845/279-0900 Date of Last Revision: 03-21-97
More informationINGREDIENTS CAS No. CONTENTS(%) OSHA ACGIH
M A T E R I A L S A F E T Y D A T A S H E E T page 1/ 4 S - 2 0 0 E * Updated : October 28, 2010 MSDS No. : MS- 00163201-03 SECTION 1 IDENTIFICATION MANUFACTURER TAIYO INK MFG. CO., LTD. TEL 81-493- 61-2832
More informationLecture 11. Etching Techniques Reading: Chapter 11. ECE 6450 - Dr. Alan Doolittle
Lecture 11 Etching Techniques Reading: Chapter 11 Etching Techniques Characterized by: 1.) Etch rate (A/minute) 2.) Selectivity: S=etch rate material 1 / etch rate material 2 is said to have a selectivity
More informationCHAPTER 7 THE DEHYDRATION AND SWEETENING OF NATURAL GAS
CHAPTER 7 THE DEHYDRATION AND SWEETENING OF NATURAL GAS Natural gases either from natural production or storage reservoirs contain water, which condense and form solid gas hydrates to block pipeline flow
More informationBET Surface Area Analysis of Nanoparticles
OpenStax-CNX module: m38278 1 BET Surface Area Analysis of Nanoparticles Nina Hwang Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0
More informationORGANIC LABORATORY TECHNIQUES 10 10.1. NEVER distill the distillation flask to dryness as there is a risk of explosion and fire.
ORGANIC LABORATORY TECHNIQUES 10 10.1 DISTILLATION NEVER distill the distillation flask to dryness as there is a risk of explosion and fire. The most common methods of distillation are simple distillation
More informationCOMPOSITION / INFORMATION ON INGREDIENTS
MATERIAL SAFETY DATA SHEET DAMTITE WATERPROOFER- WHITE 01071, 01211, 01451 DAMTITE POWDER FOUNDATION WATERPROOFER GRAY 02451 SECTION I. COMPANY & PRODUCT INFORMATION Product Identification Manufacturer
More informationSewage and Wastewater Odor Control Dr. Giancarlo Riva, Ozono Elettronica Internazionale, Muggio, Italy
Introduction Sewage and Wastewater Odor Control Dr. Giancarlo Riva, Ozono Elettronica Internazionale, Muggio, Italy Sewage and industrial plants located near residential areas can be subject to political
More informationDOW CORNING CORPORATION Material Safety Data Sheet
Page: 1 of 7 1. PRODUCT AND COMPANY IDENTIFICATION Dow Corning Corporation South Saginaw Road Midland, Michigan 48686 24 Hour Emergency Telephone: Customer Service: Product Disposal Information: CHEMTREC:
More informationUTILIZATION of PLASMA ACTIVATED WATER in Biotechnology, Pharmacology and Medicine. JSC TECHNOSYSTEM-ECO Moscow, Russia April, 2009
UTILIZATION of PLASMA ACTIVATED WATER in Biotechnology, Pharmacology and Medicine JSC TECHNOSYSTEM-ECO Moscow, Russia April, 2009 METHOD of WATER ACTIVATION with PLASMA of GAS DISCHARGE ANODE VACUUM WATER
More informationExperiment 10. Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado
1 Radioactive Decay of 220 Rn and 232 Th Physics 2150 Experiment No. 10 University of Colorado Introduction Some radioactive isotopes formed billions of years ago have half- lives so long that they are
More informationCONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
More informationTable of Contents 1. INTRODUCTION... 1
Pa mphl e t73 At mo s p h e r i cmo n i t o r i n g Eq u i p me n tf o rch l o r i n e Ed i t i o n7 J une2003 Table of Contents 1. INTRODUCTION... 1 1.1 PURPOSE... 1 1.2 RESPONSIBLE CARE... 1 1.3 BACKGROUND...
More informationEXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
More informationthe importance of ph in water purification the importance of ph in water purification Water Purification Systems UK
the importance of ph in water purification Water Purification Systems UK What is ph? The ph scale has been in use for over a century, ever since it was first defined in 1909 by the Danish chemist, Søren
More informationCompact ph Measuring Instruments
testo 205, 206, 230 Compact ph Measuring Instruments With innovative probe engineering ph C The new ph measuring instruments with innovative probe engineering Measurement of the ph value plays an important
More informationGlutaraldehyde Policy
Virginia Commonwealth University Office of Environmental Health & Safety Chemical/Biological Safety Section Glutaraldehyde Policy (revised 10/13/09) Table of Contents (Sections) 1. Purpose 2. Scope 3.
More information3M MATERIAL SAFETY DATA SHEET NO. 4004, 4008, 4026 AND 4032 DOUBLE COATED URETHANE FOAM TAPE 07/29/2002
Material Safety Data Sheet Copyright, 2002, Minnesota Mining and Manufacturing Company. All rights reserved. Copying and/or downloading of this information for the purpose of properly utilizing 3M products
More informationLACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e
1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER
More informationHolcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants
Holcim EMR List of EN Standards and VDI Guidelines usable for Discontinuous Measurements in Cement Plants Version 2004-01 / August 2004 (HGRS-JW-04-25s) HGRS-CTS/MT J. Waltisberg SUMMARY ( ) shall not
More informationANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION
Chem 306 Section (Circle) M Tu W Th Name Partners Date ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION Materials: prepared acetylsalicylic acid (aspirin), stockroom samples
More informationFundamental Autoclave Techniques
Fundamental Autoclave Techniques Version 3, September 2010 Veronika Tatarinoff GSBmE University of NSW Brandy Nelson University of Kentucky The following information is directed at aseptic/sterile surgical,
More informationCAS NO. A. 7664-93-9 >94.19 % WT >94.19 % VOL SARA 313 NO
SECTION 1: PRODUCT AND COMPANY IDENTIFICATION PRODUCT NAME: FLOWEASY DRAIN CLEANER SYNONYMS: N/A PRODUCT CODES: HTL-FE16, HTL-FE20, HTL-FE32, HTL-FE64, HTL-FE128 MANUFACTURER: Hi-Tec Laboratories DIVISION:
More informationThe Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
More information105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
More informationHM1500LF Relative Humidity Module
Small size product Product free from Lead, Cr(6+), Cd and Hg Humidity calibrated within +/-2% @55%RH Typical 1 to 3.6 Volt DC output for 0 to 100% RH at 5Vdc supply Ratiometric to voltage supply from 4.75Vdc
More informationACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
More informationLaboratory 5: Properties of Enzymes
Laboratory 5: Properties of Enzymes Technical Objectives 1. Accurately measure and transfer solutions with pipettes 2. Use a Spectrophotometer to study enzyme action. 3. Properly graph a set of data. Knowledge
More informationTHE BASICS OF STERILIZATION
THE BASICS OF STERILIZATION Objectives: State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application,
More informationSteam Sterilization Cycles for Lab Applications
Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain
More informationWaters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
More informationSECTION 1 - PRODUCT IDENTIFICATION
Product Identification Physical Data Hazardous Components Fire and Explosion Hazard Data Reactivity Data Health Hazard Information Spill or Leak Procedures Handling Information Special Protection Information
More informationMATERIAL SAFETY DATA SHEET PAGE 1
1254 Chestnut Street Newton, MA 02464-1418 Tel: (617) 965-5511 Fax: (617) 965-5818 MATERIAL SAFETY DATA SHEET PAGE 1 SECTION 1. CHEMICAL IDENTIFICATION -- - CHEMICAL NAME: TRADE NAME: PRODUCT #: Organic
More informationDew-Point Measurement Solutions
Dew-Point Measurement Solutions www.michell.com Dew-Point Measurement Solutions Special Features No downtime Michell Instruments offers a unique sensor exchange programme, which means that you will never
More informationHow Sensors Work. How Oxygen, Electrochemical Toxic, and Metal Oxide Semiconductor Sensors Work *
How Oxygen, Electrochemical Toxic, and Metal Oxide Semiconductor Sensors Work * 1. Oxygen sensor detection principle Most portable or survey instruments used for workplace evaluation of oxygen concentrations
More informationIn this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
More informationPhysical Properties of a Pure Substance, Water
Physical Properties of a Pure Substance, Water The chemical and physical properties of a substance characterize it as a unique substance, and the determination of these properties can often allow one to
More informationEmergency Phone Numbers: P.O. Box 1950 CHEMTREC: 800/424-9300 Brewster, New York 10509-8950 POISON CENTER: 800/562-8236 845/279-0900
MATERIAL SAFETY DATA SHEET Pure Tech Emergency Phone Numbers: P.O. Box 1950 CHEMTREC: 800/424-9300 Brewster, New York 10509-8950 POISON CENTER: 800/562-8236 845/279-0900 Date of Last Revision: 10/16/96
More informationChapter 12 - Liquids and Solids
Chapter 12 - Liquids and Solids 12-1 Liquids I. Properties of Liquids and the Kinetic Molecular Theory A. Fluids 1. Substances that can flow and therefore take the shape of their container B. Relative
More informationLab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives.
1 Lab 2 Biochemistry Learning Objectives The lab has the following learning objectives. Investigate the role of double bonding in fatty acids, through models. Developing a calibration curve for a Benedict
More informationCALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)
CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) I. Introduction Calibration experiments or procedures are fairly common in laboratory work which involves any type of instrumentation. Calibration
More information