1 Single-Use Flow-Through Cell ph FTC-SU-HP8 SENSOR PROBES Instruction Manual

2

3 Single-Use Flow-Through Cell ph FTC-SU-HP8 Specification: ph sensor stick integrated in a single-use flow-through cell Document filename: IM_FTC-SU-HP8_dv1 All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying, recording, taping, or information storage and retrieval systems - without the written permission of the publisher. Products that are referred to in this document may be either trademarks and/or registered trademarks of the respective owners. The publisher and the author make no claim to these trademarks. While every precaution has been taken in the preparation of this document, the publisher and the author assume no responsibility for errors or omissions, or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it. In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage caused or alleged to have been caused directly or indirectly by this document. Specifications may change without prior notice. Manufacturer PreSens Precision Sensing GmbH Josef-Engert-Str Regensburg, Germany Phone Fax

4

5 Table of Contents 1 Preface Description of the Single-Use Flow-Through Cell ph Scope of Delivery Measurement Set-up Operation Unpacking the FTC-SU-HP Assembling and Connecting the FTC-SU-HP Measurement Calibration Equilibration One point adjustment Technical Data Concluding Remarks... 10

6

7 3 Preface 1 Preface You have chosen a new, innovative technology for measuring ph. Chemical optical sensors (also called optodes) have several important features: They are small. Their signal does not depend on the flow rate of the sample. They can be physically divided from the measuring system which allows a noninvasive measurement. They can be used in disposables. Therefore, they are ideally suited for the examination of small sample volumes, for highly parallelized measurements in disposables, and for biotechnological applications. A set of different minisensors, flow-through cells and integrated sensor systems is available to make sure you have the sensor which matches your application. Please feel free to contact our service team to find the best solution for your application. Your PreSens Team PLEASE READ THE FOLLOWING INSTRUCTIONS CAREFULLY BEFORE WORKING WITH THIS ITEM.

8 4 Description of the Single-Use Flow-Through Cell ph 2 Description of the Single-Use Flow- Through Cell ph The single-use flow-through cell ph (FTC-SU-HP8) comprises a miniaturized chemical optical sensor integrated in a disposable flow-through cell. The sensor is fixed to a color coded stick (sensor stick = SST; red = ph), which can be attached to flow-through cells of different size and shape (via Luer connector), according to requirements. It is delivered with a disposable T-cell made of polycarbonate, which can be easily connected to external tubing via Luer-Lock adapters. A polymer optical fiber connects the sensor inside the flow-through cell to the respective measurement device (e. g. ph-1 mini). The single-use flow-through cells can be delivered beta-irradiated (FTC-SU-HP8-S) or untreated (FTC-SU-HP8-US). Fig. 1 Single-use flow-through cell ph (FTC-SU-HP8) 2.1 Scope of Delivery The FTC-SU-HP8 is double-packed. It is delivered in a transparent plastic bag (to enable unpacking of irradiated FTCs under sterile conditions) which is again packed in a lightproof packaging (to protect the sensor material). The FTC-SU-HP8 is pre-calibrated. Fig. 2 Packed single-use flow-through cell ph. The respective polymer optical fiber has to be ordered separately.

9 5 Description of the Single-Use Flow-Through Cell ph Additionally required equipment (not supplied): Polymer optical fiber (POF); with SMA connector on one side, available lengths 1.0, 1.5, and 5.0 m (for lengths of more than 5 m, please contact our service team) Fiber optic ph transmitter, e. g. ph-1 mini (more ph transmitters can be found on PC / Notebook If your application requires a T-cell with other inner diameter or of other design, you can integrate the sensor stick in another cell as well. The flow-through cell must have a Luer connector, so the sensor stick can be properly attached! Supplier for such cells compatible with the sensor stick is e. g. Qosina (Part Numbers in the 2014 catalogue: , ). 2.2 Measurement Set-up Fig. 3 Set-up for measurement with a FTC-SU-HP8 A polymer optical fiber, which is connected to a ph transmitter, is attached to the FTC-SU- HP8. The FTC-SU-HP8 is integrated e. g. in the flow path of a perfusion reactor or the bypass of a reactor.

10 6 Operation 3 Operation 3.1 Unpacking the FTC-SU-HP8 The FTC-SU-HP8 is delivered in a lightproof packaging to ensure a long shelf-life, so do not open this packaging immediately at delivery. It is recommended to unpack the FTC just before using it.! Take care when opening the lightproof packaging with scissors, especially when handling irradiated FTCs. Cut close to the welding seam of the packaging, not to cut into the inner plastic bag. Fig. 4 Opening the lightproof outer packaging. Inner packaging: FTC in transparent plastic bag.! If your FTC-SU-HP8 is delivered irradiated open the inner plastic bag with the FTC components in a laminar flow box, or in a similar controlled environment, and wear gloves! Fig. 5 FTC-SU-HP8: the sensor stick is attached to a disposable T-cell.

11 7 Operation 3.2 Assembling and Connecting the FTC-SU-HP8 If you want to use the sensor stick (SST) with another flow-through cell than the T-cell it is delivered with, you can simply remove it from the T-cell by turning it counter-clockwise at the Luer connector. Now you can attach the sensor stick to a flow-through cell of your choice. Make sure it is attached firmly, but do not use force! Fig. 6 Connecting the sensor stick (SST) to a flow-through cell. The FTC-SU-HP8 can be integrated in the flow path of your perfusion system. According to your tube size you can use end caps (Luer-Lock adapter). Attach the tubes at the inlet and outlet of the FTC. Fig. 7 Connecting the FTC with the tubes of the perfusion system (using Luer-Lock adapter). When the FTC is connected and the flow path is closed, the polymer optical fiber can be attached. (Do not attach the polymer optical fiber before closing the flow path, if you want to avoid contamination; the fiber is not irradiated.) First loosen the red cap on the sensor stick. Then insert the polymer optical fiber in the sensor stick until stop, and fasten it with the red cap by turning the cap clockwise.

12 8 Operation Fig. 8 Loosen the red cap and insert the polymer optical fiber until stop; then fasten the red cap again. Connect the polymer optical fiber to the SMA connector on the front panel of your ph transmitter.! As the sensor stick and the flow-through cell are not glued together, please check the connection before using the FTC, and tighten it again if necessary, but without using force! Fig. 9 Tighten the connections before use! Medium can be pumped through the FTC-SU-HP8 now.

13 9 Operation 3.3 Measurement Calibration The FTC-SU-HP8 is delivered pre-calibrated. The enclosed Final Inspection Protocol contains the calibration values, which have to be entered into the software of the fiber optic ph transmitter you are using. (Please see the respective transmitter instruction manual for more detailed information on calibration.) In case of samples differing significantly from physiological buffers a multi-point calibration is recommended. Prepare at least five solutions, of similar composition like the sample and different ph, covering the range of interest. Place the sensor stick in the vessels containing the solutions of known ph or pump the solutions through the FTC one after another - and perform the multi-point calibration like described in the respective transmitter instruction manual. If the vessel is large enough to place a ph electrode, we recommend titrating directly in the vessel and following the instructions for multi-point calibration in the transmitter manual Equilibration! The sensor needs to be equilibrated before use. In order to do so you have to fill the FTC with your medium and wait for at least 15 minutes so that the sensor can equilibrate One point adjustment ph one point adjustment is advised if your media differs from physiological conditions to obtain optimal sensor performance. Ideally the starting ph of the sample is known. If the phase value is constant, follow the instructions for auto zero in the transmitter manual.

14 10 Technical Data / Concluding Remarks 4 Technical Data Specifications* Measuring range ph Response time (t90) at 37 C** < 120 sec. Resolution at ph = 7 ± 0.02 ph Accuracy at ph = 7 ± 0.05 after single point calibration at ph 7 Drift at ph = 7: < ph per day (sampling interval of 1 min.) Measurement Temperature From 5 C to 50 C Properties* Compatibility Aqueous solutions, ethanol (max. 10 % V/V), methanol (max. 10 % V/V), ph 2-10 Cross Sensitivity Reduced to ionic strength (salinity); a high concentration of small fluorescent molecules in the visible range can interfere Sterilization procedure*** Irradiation, Ethylene oxide (EtO) Calibration SST are pre-calibrated; Single point calibration is recommended Storage Stability 18 months provided the sensor is stored in the dark *provided ph sensor is used without further handling in physiological solutions **equilibrated FTC with physiological solution and sufficient flow rate (min. 10 ml/min) at 37 C ***recalibration may be required 5 Concluding Remarks Dear Customer, With this manual, we hope to provide you with an introduction to work with the single-use flow-through cell ph (FTC-SU-HP8). This manual does not claim to be complete. We are endeavored to improve and supplement this version. We are looking forward to your critical review and to any suggestions you may have. You can find the latest version at With best regards, Your PreSens Team

15 11 Technical Data / Concluding Remarks Manufacturer PreSens Precision Sensing GmbH Josef-Engert-Straße Regensburg Germany Phone Fax

Flow-Through Cell ph SENSOR PROBES. Instruction Manual

Flow-Through Cell ph SENSOR PROBES. Instruction Manual Flow-Through Cell ph SENSOR PROBES Instruction Manual Flow-Through Cell ph Specification: ph sensor integrated in a flow-through cell Document filename: IM_FTC_pH_dv1 All rights reserved. No parts of

More information

Oxygen Sensor Spots. PSt3 / PSt6. Instruction Manual SENSOR PROBES

Oxygen Sensor Spots. PSt3 / PSt6. Instruction Manual SENSOR PROBES Oxygen Sensor Spots SENSOR PROBES PSt3 / PSt6 Instruction Manual Oxygen Sensor Spots PSt3 / PSt6 Specification: Chemical optical oxygen sensor spots for integration into transparent glass or plastic vessels

More information

VisiSens AnalytiCal 2 Software

VisiSens AnalytiCal 2 Software VisiSens AnalytiCal 2 Software SYSTEMS Instruction Manual Specification: Software for the VisiSens A2 ph Imaging System Version VA2.10 Document filename: IM_VA2_dv2 All rights reserved. No parts of this

More information

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent

More information

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use

Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use GE Healthcare Instructions 28-9226-84 AD ReadyToProcess Sterile ReadyToProcess Hollow Fiber Cartridges Instructions for Use 1 Introduction... 1 2 Inspect the Cartridge... 1 3 General Description... 2 4

More information

INSTRUCTIONS 56-1190-98. Edition AC

INSTRUCTIONS 56-1190-98. Edition AC Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

Halogen Light Source with Attenuator and TTL-Shutter Installation and Operation Manual

Halogen Light Source with Attenuator and TTL-Shutter Installation and Operation Manual Halogen Light Source with Attenuator and TTL-Shutter Installation and Operation Manual For Products: HL-2000-FHSA / HL-2000-FHSA-LL / HL-2000-FHSA-HP Document: 000-10000-070-02-201504 000-00000- AMERICAS

More information

GENERAL USE AND LABORATORY ph ELECTRODES PRODUCT SPECIFICATION SHEET

GENERAL USE AND LABORATORY ph ELECTRODES PRODUCT SPECIFICATION SHEET GENERAL USE AND LABORATORY ph ELECTRODES The right sensor makes the difference. Different applications require different types of ph sensors to ensure measurement accuracy and extend instrument lifetime.

More information

Operating Manual Fast Thermal Conductivity Detector FTC110-OEM

Operating Manual Fast Thermal Conductivity Detector FTC110-OEM Operating Manual Fast Thermal Conductivity Detector FTC110-OEM Version 12_09 Messkonzept GmbH Operating Manual 2 1. Features Precise and long term stable thermal conductivity measurement High sensitivity

More information

asense CO 2 / temperature transmitter

asense CO 2 / temperature transmitter Function Description and Maintenance Instruction asense CO 2 / temperature transmitter General The IAQ-sensor product asense is used to measure indoor air carbon dioxide concentration and temperature in

More information

Instruction Manual. OXY-4 4-Channel Fiber-Optic Oxygen Meter

Instruction Manual. OXY-4 4-Channel Fiber-Optic Oxygen Meter Instruction Manual OXY-4 4-Channel Fiber-Optic Oxygen Meter Instruction Manual OXY-4 Software Version OXY4v2_11FB May 2005 Table of contents Table of Contents 1 Preface...1 2 Safety Guidelines...2 3 Description

More information

MACS GMP Cell Culture Bags

MACS GMP Cell Culture Bags Handling guidelines MACS GMP Cell Culture Bags Contents Page 1 Safety precautions 2 2 MACS GMP Cell Culture Bags 2 types and components 3 Preparing medium 3 4 Fluid transfer into bags 3 4.1 Small-volume

More information

WatchManager Data Backup

WatchManager Data Backup WatchManager Data Backup WatchManager Data Backup All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying,

More information

Synthesis_GB :38 Uhr Seite 1 C M Y CM MY CY CMY K Parallel Synthesis Probedruck

Synthesis_GB :38 Uhr Seite 1 C M Y CM MY CY CMY K Parallel Synthesis Probedruck Parallel Synthesis Technical specifications Model Solid Phase Synthesis Liquid Phase Synthesis Multiple Evaporator P/N Solid 16: 552-14200-00 Liquid 12: 552-25000-00 Multiple Evaporator 12: 552-35000-00

More information

Phase-Contrast & Fluorescence Microscopy of Adherent Cells grown on ECIS Electrodes

Phase-Contrast & Fluorescence Microscopy of Adherent Cells grown on ECIS Electrodes Phase-Contrast & Fluorescence Microscopy of Adherent Cells grown on ECIS Electrodes Prof. Joachim Wegener University of Regensburg Institut für Analytische Chemie, Chemo- und Biosensorik 93040 Regensburg

More information

E&P HYDRAULICS Remote control EPRC-01 Level System Camper

E&P HYDRAULICS Remote control EPRC-01 Level System Camper USER S GUIDE E&P HYDRAULICS Remote control EPRC-01 Level System Camper 1 Copyright 2012, E&P Hydraulics This manual is copyrighted, with all rights reserved. Under the copyright laws, this may not, in

More information

Fundamentals of Mass Flow Control

Fundamentals of Mass Flow Control Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of

More information

Models SFP33 & SFP32 Pressure Transmitters

Models SFP33 & SFP32 Pressure Transmitters SFP33 Flow-Through Pressure Transmitter SFP32 Single-Ended Pressure Transmitter DS-PR-SFP-eng April, 2016 Data Sheet Models SFP33 & SFP32 Pressure Transmitters Pressure Brooks Seal Free Pressure Measurement

More information

Dissolved Oxygen Sensors

Dissolved Oxygen Sensors Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit

More information

MobileCollect RF Channel Sniffer Program

MobileCollect RF Channel Sniffer Program All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying, recording, taping, or information storage and retrieval

More information

How single-use connections advance aseptic processing: Increased process flexibility and reliability, reduced costs

How single-use connections advance aseptic processing: Increased process flexibility and reliability, reduced costs WHITE PAPER 7004 How single-use connections advance aseptic processing: Increased process flexibility and reliability, reduced costs By John Boehm Business Unit Manager Colder Products Company Today s

More information

HiTrap Heparin HP, 1 ml and 5 ml

HiTrap Heparin HP, 1 ml and 5 ml Instructions 71-7004-00 AU HiTrap affinity columns HiTrap Heparin HP, 1 ml and 5 ml HiTrap Heparin HP is a prepacked ready to use, column for preparative affinity chromatography. The special design of

More information

Denver Instrument Company. BASIC ph Meter. Operation Manual 300408.1 Rev. G

Denver Instrument Company. BASIC ph Meter. Operation Manual 300408.1 Rev. G Denver Instrument Company BASIC Meter Operation Manual 300408.1 Rev. G BASIC Meter Quick Reference 1. Connect power cable to meter power jack and to AC power source. 2. Connect a glass electrode to the

More information

Groundwater Sampling for Helium / Noble Gases

Groundwater Sampling for Helium / Noble Gases Institute of Environmental Physics University of Heidelberg Im Neuenheimer Feld 229 D-69120 Heidelberg http://www.iup.uni-heidelberg.de Prof. W. Aeschbach-Hertig phone: +49 (0)6221 54 6331 fax: +49 (0)6221

More information

Trace Gas Exchange Measurements with Standard Infrared Analyzers

Trace Gas Exchange Measurements with Standard Infrared Analyzers Practical Environmental Measurement Methods Trace Gas Exchange Measurements with Standard Infrared Analyzers Last change of document: February 23, 2007 Supervisor: Charles Robert Room no: S 4381 ph: 4352

More information

450TOC Analyzer. Portable, Real-time, Continuous Total Organic Carbon Measurement. 450TOC Analyzer

450TOC Analyzer. Portable, Real-time, Continuous Total Organic Carbon Measurement. 450TOC Analyzer 450 Analyzer 450 Analyzer Fast, Continuous Measurement Compliant with USP, Ch P, EP and JP USB Data Logging and Printer Output Robust Portable Design Portable, Real-time, Continuous Total Organic Carbon

More information

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual

ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual ENrich SEC 70 ENrich SEC 650 High-Resolution Size Exclusion Columns Instruction Manual Catalog numbers 780-1070 780-1650 Please read these instructions before you use ENrich SEC high-resolution size exclusion

More information

ENDURANCE Conductivity Sensors

ENDURANCE Conductivity Sensors Instruction Manual 400 and 400VP LIQ_MAN_ABR_400_400VP December 2013 ENDURANCE Conductivity Sensors For additional information, please refer visit our website at www.rosemountanalytical.com. CAUTION Sensor/Process

More information

REDOX AND REFERENCE ELECTRODE USER MANUAL

REDOX AND REFERENCE ELECTRODE USER MANUAL REDOX AND REFERENCE ELECTRODE USER MANUAL 3 Redox and reference electrode manual Copyright 2012 Unisense A/S Version October 2012 REDOX AND REFERENCE ELECTRODE MANUAL UNISENSE A/S TABLE OF CONTENTS WARRANTY

More information

Installation and User Guide. UpIR Out-of-Compartment Diffuse Reflectance Accessory

Installation and User Guide. UpIR Out-of-Compartment Diffuse Reflectance Accessory Installation and User Guide UpIR Out-of-Compartment Diffuse Reflectance Accessory The information in this publication is provided for reference only. All information contained in this publication is believed

More information

Subminiature Load Cell Model 8417

Subminiature Load Cell Model 8417 w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...

More information

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter Description The coriolis mass flow meters measure simultaneously mass

More information

COMPACT CORIOLIS MASS FLOW METER/CONTROLLER FOR LOW FLOWS

COMPACT CORIOLIS MASS FLOW METER/CONTROLLER FOR LOW FLOWS COMPACT CORIOLIS MASS FLOW METER/CONTROLLER FOR LOW FLOWS W. Jouwsma Bronkhorst Cori-Tech B.V. Nijverheidsstraat -, 71 AK Ruurlo, The Netherlands Phone: +31 573 59; Fax: +31 9375; Internet: www.bronkhorst-cori-tech.com

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

Duct Humidity & Temperature Transmitter

Duct Humidity & Temperature Transmitter SDC-HT Duct Humidity & Transmitter Features Replaceable sensor element Humidity and temperature measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable

More information

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and

More information

TROUBLESHOOTING GUIDE HPLC

TROUBLESHOOTING GUIDE HPLC TROUBLESHOOTING GUIDE HPLC Before starting any troubleshooting, whether it is related to instruments or columns, it is essential that safe laboratory practices be observed. The chemical and physical properties

More information

User information Fill level indicator for rain water storage tanks

User information Fill level indicator for rain water storage tanks User information Fill level indicator for rain water storage tanks We congratulate you on the purchase of our fill level measuring device. You have purchased a high-quality product built to the highest

More information

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force

More information

LABORATORY: TEMPERATURE AND RESPIRATION OF TOADS AND MAMMALS

LABORATORY: TEMPERATURE AND RESPIRATION OF TOADS AND MAMMALS LABORATORY: TEMPERATURE AND RESPIRATION OF TOADS AND MAMMALS Aerobic and Anaerobic Cellular Respiration In most cells, metabolism is supported by oxidation of carbon fuels, including lipid, carbohydrate

More information

Installation Guide for Fiber Taps

Installation Guide for Fiber Taps Installation Guide for Fiber Taps 800-0146-001 CPUBTPLCSLMU Rev. E, 5/12 Contents Introduction....1 Key Features....2 About This Guide....3 Unpacking and Inspection....4 Product Diagram....4 Connecting

More information

CALIBRATING TOROIDAL CONDUCTIVITY SENSORS

CALIBRATING TOROIDAL CONDUCTIVITY SENSORS Application Data Sheet ADS 43-025 October 2010 Theory CALIBRATING TOROIDAL CONDUCTIVITY SENSORS BACKGROUND A toroidal conductivity sensor consists of two wire wound toroids encased in a plastic body. When

More information

Instruction and Maintenance Manual

Instruction and Maintenance Manual Instruction and Maintenance Manual GRYF OXY FS 01/1000/1 B Contact GRYF HB, spol. s r.o. Cechova 314 Havlickuv Brod 580 01 tel.: +420 569 426 627 fax: +420 569 426 627 Czech Republic www.gryf.eu Technical

More information

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com

More information

August 2012 Web Application for Enprise Job Costing. 1.0.2 1.2.3863 SAP 2007A, SAP 2007A SP01, SAP 8.8 All regions

August 2012 Web Application for Enprise Job Costing. 1.0.2 1.2.3863 SAP 2007A, SAP 2007A SP01, SAP 8.8 All regions August 2012 Document Version: Product Version: SAP Version: Localization: 1.0.2 1.2.3863 SAP 2007A, SAP 2007A SP01, SAP 8.8 All regions 2012 Enprise Solutions All rights reserved. No parts of this work

More information

Linear Position Sensors

Linear Position Sensors McLaren Technology Centre Chertsey Road Woking Surrey GU21 4YH UK McLaren Electronics Inc. Suite 115 Bostick Building 9801 West Kincey Avenue Huntersville NC 28078 USA Copyright McLaren Electronics 2012

More information

Electroblotting Systems

Electroblotting Systems user manual Electroblotting Systems Complete Electroblotting System for 10 10 cm Mini Gels, Cat. No. E4103 Complete Electroblotting System for 20 20 cm Standard Gels, Cat. No. E4203 um DNTV100-EBK-IM/Rev.

More information

DEAE Sephacel. Sephacel. GE Healthcare

DEAE Sephacel. Sephacel. GE Healthcare GE Healthcare Instructions 71 7100 00 AC Ion Exchange DEAE Sephacel DEAE Sephacel is a weak anion exchanger based on beaded cellulose. The ion exchange group is diethylaminoethyl, which remains charged

More information

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration

More information

Chapter 17 Operation Instruction of Soap Bubble Meter

Chapter 17 Operation Instruction of Soap Bubble Meter Chapter 17 Operation Instruction of Soap Bubble Meter PRINCIPLE OF OPERATION The APM 520A Flow Calibrator is designed for use as a primary standard for calibrating flow meter in the low or medium flow

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Volumetric two-component methanol based Karl Fischer Titration

Volumetric two-component methanol based Karl Fischer Titration Volumetric two-component methanol based Karl Fischer Titration DOC316.52.93098 1 to 100 mg of water in sample used This document applies to volumetric Karl Fischer titrations determined with two-component

More information

BUSINESS. Elektronisches Türschloss. Installation and User Manual. www.burg-waechter.de. Ingeniously safe. Ingeniously simple.

BUSINESS. Elektronisches Türschloss. Installation and User Manual. www.burg-waechter.de. Ingeniously safe. Ingeniously simple. Elektronisches Türschloss BUSINESS Set 5011 PINCODE Set 5012 FINGERSCAN Set 5013 E-KEY Ingeniously safe. Ingeniously simple. BA/MA TSE Set 5000 dp/fsc 04/2012 Installation and User Manual BURG-WÄCHTER

More information

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration Relative Humidity Measurement and Calibration The widest range of humidity products from a pioneer of RH sensor development The best in Relative Humidity technology to optimize your process Whether you

More information

External Devices Handling

External Devices Handling External Devices Handling Data transfer from digital tachograph PC NET SERVICE Stocerz Sp. J. Other devices - Tutorial - Pomoc All rights reserved. No parts of this work may be reproduced in any form or

More information

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm 1. How to prepare straws for pronuclear stages and embryos (use straws with plug, 66.5 mm) Straws are

More information

AIRMoN, ph and Conductivity Field Measurement

AIRMoN, ph and Conductivity Field Measurement AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity

More information

General Purpose ph/orp Sensors

General Purpose ph/orp Sensors General Purpose ph/orp Sensors Product Data Sheet LIQ_PDS_3900 / Rev. F December 2014 General Purpose ph/orp Sensors SMART preamplifier-stores calibration and other data. Extended sensor life provided

More information

Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate

Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate 1 919 1919P01 1919P03 QBM65/C, QBM75-1U/C Differential Pressure Sensors for air and non-corrosive gases, with calibration certificate QBM751-1/C QBM65/C QBM75/C Highly accurate measurements with calibration

More information

Related topics: Application Note 27 Data Analysis of Tube Formation Assays.

Related topics: Application Note 27 Data Analysis of Tube Formation Assays. Tube Formation Assays in µ-slide Angiogenesis Related topics: Application Note 27 Data Analysis of Tube Formation Assays. Contents 1. General Information... 1 2. Material... 2 3. Work Flow Overview...

More information

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev. Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 6 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

Oxygen Measurement Systems

Oxygen Measurement Systems INGOLD Leading Process Analytics Oxygen Measurement Systems InPro 6000 Sensor Series Improved product quality Enhanced process safety Ease of operation Full regulatory compliance In-line Oxygen Measurements

More information

Loop Calibration and Maintenance

Loop Calibration and Maintenance Loop Calibration and Maintenance Application Note Introduction Process instrumentation requires periodic calibration and maintenance to ensure that it is operating correctly. This application note contains

More information

I N S T R U C T I O N D A T A

I N S T R U C T I O N D A T A I N S T R U C T I O N D A T A 107660-200 107660-300 107660-400 107660-500 107660- RFL CMI CONVERTER E1 TX TX/RX E1 RX PWR ELECT FO FO RX FO TX Boonton, New Jersey USA Figure 1. Typical, Front Panel view

More information

SenTix 950 SenTix 980

SenTix 950 SenTix 980 SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

TROUBLESHOOTING FLOW CHART

TROUBLESHOOTING FLOW CHART TROUBLESHOOTING FLOW CHART Revised and updated on January 16, 2003, by Tony Leger and Automation Lab Technology. FACScan(FC)\FACSort(FST) FACScalibur (FCB) Fluidics/Pneumatics troubleshooting guide DCM

More information

Installation Guide for Fiber Taps

Installation Guide for Fiber Taps Installation Guide for Fiber Taps 800-0146-001 CPUBTPLCSLMU Rev. C, 08/10 Contents Introduction..................................................... 1 Key Features.....................................................

More information

Metrology Society of Australia MSA Test Method Calibration of Pressure Calibrators, Indicators and Transducers

Metrology Society of Australia MSA Test Method Calibration of Pressure Calibrators, Indicators and Transducers Metrology Society of Australia MSA Test Method 1-2008 Calibration of Pressure Calibrators, Indicators and Transducers Revision No. 5 June 2008 Contents: 1. Introduction 2. Scope 3. Desired outcomes of

More information

ph meter Care and use guide

ph meter Care and use guide ph meter Care and use guide www.getbluelab.com Low battery indicator Bluelab ph Meter The battery operated Bluelab ph Meter measures ph levels by using the Bluelab ph Probe connected to the electronic

More information

(979) 703-1925

(979) 703-1925 Ý» ² Û²» ¹ Ì ²» Û» ³»² Ù«¼» www.fuelcellstore.com sales@fuelcellstore.com (979) 703-1925 All rights reserved. No part of this Experiment Guidemay be reproduced, stored in a data retrieval system or transmitted

More information

Fiber Optic Dual Switch FOS-2x2-TTL. Installation and Operation Manual Document Number 000-10000-040-02-1209

Fiber Optic Dual Switch FOS-2x2-TTL. Installation and Operation Manual Document Number 000-10000-040-02-1209 Fiber Optic Dual Switch FOS-2x2-TTL Installation and Operation Manual Document Number 000-10000-040-02-1209 Offices: Ocean Optics, Inc. World Headquarters 830 Douglas Ave., Dunedin, FL, USA 34698 Phone

More information

Adapter Catalog MOBILE METAL ANALYZER. SPECTRO isort and SPECTROTEST

Adapter Catalog MOBILE METAL ANALYZER. SPECTRO isort and SPECTROTEST Adapter Catalog MOBILE METAL ANALYZER SPECTRO isort and SPECTROTEST Testing Adapters for Mobile Metal Analyzers A full range of metal analyzer products for onsite metal analysis tasks SPECTRO isort SPECTROTEST

More information

Polymer Coated Fiber Cable (PCF)

Polymer Coated Fiber Cable (PCF) Polymer Coated Fiber Cable (PCF) Panduit has introduced a Polymer Coated Fiber (PCF) to their fiber cable offering available in 50µm and 62.5µm core diameters. Along with this cable having a stronger durability

More information

Trace Dissolved Oxygen Sensor

Trace Dissolved Oxygen Sensor Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION

More information

MKS Baratron Type 122A Absolute Pressure Gauge P1 Rev E, 6/96

MKS Baratron Type 122A Absolute Pressure Gauge P1 Rev E, 6/96 MKS Baratron Type 122A Absolute Pressure Gauge 110448-P1 Rev E, 6/96 Copyright 1996 by MKS Instruments, Inc. All rights reserved. No part of this work may be reproduced or transmitted in any form or by

More information

Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN1 2015-10. www.sks-online.com www.sks-webshop.com.

Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN1 2015-10. www.sks-online.com www.sks-webshop.com. Instruction Manual Alfa Laval SB Membrane Sample Valve ESE02963-EN1 2015-10 Original manual Table of contents The information herein is correct at the time of issue but may be subject to change without

More information

VisiSens 2D Mapping of Oxygen, ph & CO 2

VisiSens 2D Mapping of Oxygen, ph & CO 2 O ² ph CO ² VisiSens 2D Mapping of Oygen, ph & C 2D read-out of luminescent sensor foils Contactless direct sensing or through transparent walls Visualize spatial and temporal analyte gradients Numerous

More information

NEULOG HEART RATE AND PULSE LOGGER SENSOR GUIDE

NEULOG HEART RATE AND PULSE LOGGER SENSOR GUIDE NeuLog heart rate and pulse logger sensor NUL-208 The NeuLog pulse sensor can be used for any science experiment which requires a human heart rate or pulse measurement such as in the fields of Biology,

More information

DISSOLVED OXYGEN PROBE

DISSOLVED OXYGEN PROBE 51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe

More information

Process Automation. Sensors and Systems: Level Sensors Level Switches Overfill Prevention Pressure Sensors Temperature Sensors

Process Automation. Sensors and Systems:  Level Sensors Level Switches Overfill Prevention Pressure Sensors Temperature Sensors Process Automation Accurate Level Sensors Level Switches Overfill Prevention Pressure Sensors Temperature Sensors Flexible Sensors and Systems: www.fafnir.com Reliable Process Auromation Edition 08.13

More information

PRODUCT INFORMATION...

PRODUCT INFORMATION... VIROMER RED In vitro plasmid DNA and mrna Standard Transfection PRODUCT INFORMATION... 2 GENERAL... 2 RED OR YELLOW?... 3 PROTOCOL GUIDELINES... 4 GENERAL REMARKS... 4 CELL CULTURE AND PLATING... 4 FORWARD/REVERSE

More information

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters

More information

Calpro EasyExtract. 1: Intended use. 2: Materials

Calpro EasyExtract. 1: Intended use. 2: Materials 1 Calpro Easy 1: Intended use Calpro Easy is intended for extraction of stool samples for measurement of Calprotectin in humans. This device has been validated in combination with CalproLab Calprotectin

More information

MEDICAL DEVICE GUIDANCE

MEDICAL DEVICE GUIDANCE May 2014 MEDICAL DEVICE GUIDANCE GN-23: Guidance on Labelling for Medical Devices PREFACE This document is intended to provide general guidance. Although we have tried to ensure that the information contained

More information

Screw Cap Micro Tubes

Screw Cap Micro Tubes Screw Cap Micro Tubes For transport, storage and sample preparation Reliable quality and versatility Sicherheit Uncompromising... quality and purity Sicherheit... Over 30 years ago, Sarstedt introduced

More information

User's Guide bintec R4100 / R4300 GRE. Copyright July 17, 2006 Funkwerk Enterprise Communications GmbH Version 1.0

User's Guide bintec R4100 / R4300 GRE. Copyright July 17, 2006 Funkwerk Enterprise Communications GmbH Version 1.0 User's Guide bintec R4100 / R4300 GRE Copyright July 17, 2006 Funkwerk Enterprise Communications GmbH Version 1.0 Purpose Liability Trademarks Copyright Guidelines and standards How to reach Funkwerk Enterprise

More information

EasyTrack 2 Accessories List

EasyTrack 2 Accessories List EasyTrack 2 Accessories List DATA LOGGER ET6061 EasyTrack2 Data Logger (6 Channel) Measurement Range -150 to 500 C (-238 F to 932 F) Accuracy ±0.5 C (0.9 F). Type K thermocouple. 6000 readings per channel.

More information

Measuring Process Temperature In Small Diameter Lines

Measuring Process Temperature In Small Diameter Lines Measuring Process Temperature In Small Diameter Lines Burns Engineering Inc., 10201 Bren Road East, Minnetonka, Minnesota 55343 Preface In today s high-performance process systems, measuring the process

More information

Techneau, 08.October Redesigned monitoring station based on UV/Vis spectrometry

Techneau, 08.October Redesigned monitoring station based on UV/Vis spectrometry Techneau, 08.October 2007 Redesigned monitoring station based on UV/Vis spectrometry Techneau, 08.October 2007 Redesigned monitoring station based on UV/Vis spectrometry 2006 TECHNEAU TECHNEAU is an Integrated

More information

BioResearch. RAFT 3D Cell Culture Kit Protocol

BioResearch. RAFT 3D Cell Culture Kit Protocol BioResearch RAFT 3D Cell Culture Kit Protocol RAFT 3D Cell Culture Kit This protocol describes in detail how to produce RAFT 3D Cell Cultures in 96-well plates, 24-well plates and 24-well inserts. It is

More information

User Manual. Model P771 ph Meter

User Manual. Model P771 ph Meter User Manual Model P771 ph Meter 1 Table of Contents 1 INTRODUCTION... 1 2 SAFETY SUMMARY... 2 3 COMPLIANCE STATEMENTS... 3 4 PRODUCT CONTENTS AND INSPECTION... 3 5 DEVICE DESCRIPTION... 4 6 OPERATING INSTRUCTIONS...

More information

Manual Flow Meter PCE-423

Manual Flow Meter PCE-423 PCE Americas Inc. 711 Commerce Way Suite 8 Jupiter FL-33458 USA From outside US: +1 Tel: (561) 320-9162 Fax: (561) 320-9176 info@pce-americas.com PCE Instruments UK Ltd. Units 12/13 Southpoint Business

More information

EP Series Touch User Guide

EP Series Touch User Guide EP Series Touch User Guide Copyright 6 Oct 2011 by Planar Systems, Inc. All rights reserved. Contents of this publication may not be reproduced in any form without permission of Planar Systems, Inc. Trademark

More information

Designed for multi-injector endurance testing, the

Designed for multi-injector endurance testing, the Designed for multi-injector endurance testing, the ETB Test Bench can optionally provide shot-to-shot mass measurement. The injectors are precisely subject to variable controls and hydraulic pressures

More information

WHITE PAPER: Relative Humidity Sensor Behavior and Care By Steve Cavlier

WHITE PAPER: Relative Humidity Sensor Behavior and Care By Steve Cavlier WHITE PAPER: Relative Humidity Sensor Behavior and Care By Steve Cavlier 1. What is Relative Humidity? Air, in our normal environment always holds humidity. The number of water molecules in the air can

More information

WHITE PAPER: Relative Humidity Sensor Behavior and Care By Steve Cavlier. 1. What is Relative Humidity?

WHITE PAPER: Relative Humidity Sensor Behavior and Care By Steve Cavlier. 1. What is Relative Humidity? WHITE PAPER: Relative Humidity Sensor Behavior and Care By Steve Cavlier 1. What is Relative Humidity? Air, in our normal environment always holds humidity. The number of water molecules in the air can

More information

HD402T Pressure Transmitters

HD402T Pressure Transmitters HD402T Pressure Transmitters Sensor with high accuracy and stability Measurement of pressures relative to the atmosphere or differential pressures Dual analogue output: current and voltage Versions with

More information