Fetal Thermal Effects of Diagnostic Ultrasound

Size: px
Start display at page:

Download "Fetal Thermal Effects of Diagnostic Ultrasound"

Transcription

1 Fetal Thermal Effects of Diagnostic Ultrasound Abbreviations ALARA, as low as reasonably achievable; dga, days gestational age; FDA, Food and Drug Administration; HSP, heat shock protein; I SPTA, spatial-peak temporal-average intensity; MI, mechanical index; NTC, neural tube closure; TI, thermal index; TIB, thermal index for bone; TTO, thermal test object; WFUMB, World Federation for Ultrasound in Medicine and Biology Received April 12, 2007, from the Department of Obstetrics and Gynecology, Rush University, Chicago Illinois USA (J.S.A.); Faculty of Health Sciences, University of Sydney, Sydney, New South Wales, Australia (S.B.B.); Bath University and Royal United Hospital, Bath, England (F.A.D.); SRI International, Molecular Physics Laboratories, Menlo Park, California USA (P.D.E.); Department of Radiology, Harvard Medical School, Boston, Massachusetts USA (K.H.H.); and Center for Biomedical Physics, Temple University Medical School, Philadelphia, Pennsylvania (M.C.Z.). Revision requested April 18, Revised manuscript accepted for publication December 4, Address correspondence to Jacques S. Abramowicz, MD, Department of Obstetrics and Gynecology, Rush University Medical Center, 1635 W Congress Pkwy, Chicago, IL USA. jacques_abramowicz@rush.edu CME Article includes CME test Jacques S. Abramowicz, MD, Stanley B. Barnett, MSc, PhD, Francis A. Duck, PhD, Peter D. Edmonds, PhD, Kullervo H. Hynynen, MSc, PhD, Marvin C. Ziskin, MD Processes that can produce a biological effect with some degree of heating (ie, about 1 C above the physiologic temperature) act via a thermal mechanism. Investigations with laboratory animals have documented that pulsed ultrasound can produce elevations of temperature and damage in biological tissues in vivo, particularly in the presence of bone (intracranial temperature elevation). Acoustic outputs used to induce these adverse bioeffects are within the diagnostic range, although exposure times are usually considerably longer than in clinical practice. Conditions present in early pregnancy, such as lack of perfusion, may favor bioeffects. Thermally induced teratogenesis has been shown in many animal studies, as well as several controlled human studies; however, human studies have not shown a causal relationship between diagnostic ultrasound exposure during pregnancy and adverse biological effects to the fetus. All human epidemiologic studies, however, were conducted with commercially available devices predating 1992, that is, with acoustic outputs not exceeding a spatial-peak temporal-average intensity of 94 mw/cm 2. Current limits in the United States allow a spatial-peak temporal-average intensity of 720 mw/cm 2 for fetal applications. The synergistic effect of a raised body temperature (febrile status) and ultrasound insonation has not been examined in depth. Available evidence, experimental or epidemiologic, is insufficient to conclude that there is a causal relationship between obstetric diagnostic ultrasound exposure and obvious adverse thermal effects to the fetus. However, very subtle effects cannot be ruled out and indicate a need for further research, although research in humans may be extremely difficult to realize. Key words: bioeffects; fetus; obstetrics; safety; thermal effects; ultrasound by the American Institute of Ultrasound in Medicine J Ultrasound Med 2008; 27: /08/$3.50

2 Fetal Thermal Effects of Diagnostic Ultrasound This article analyzes thermal effects of fetal ultrasound exposure. The normal core human body temperature is generally accepted to be 37 C with a diurnal variation of ±0.5 C to 1 C, 1,2 although 36.8 C ± 0.4 C (95% confidence interval) may be closer to the actual mean for large populations. 3 During the entire gestation, the temperature of the human embryo/fetus is higher than the maternal core body temperature 4 and gradually rises until, in the final trimester (near term), it exceeds that of the mother by 0.5 C. 5 Thermally induced teratogenesis has been shown in many animal studies, as well as several controlled human studies. 6 Edwards 7 and others have shown that hyperthermia is teratogenic for numerous animal species, including the human, and suggested a 1.5 C temperature elevation above the normal value as a universal threshold. 8 An elevated maternal temperature in early gestation has been associated with an increased incidence of congenital anomalies. 9 Tolerance to increased temperature (thermotolerance) is an important aspect of thermal teratogenesis. Thermotolerance is induced by the production of heat shock proteins (HSPs), which occurs (up to a limit) during a relatively slow (10- to 15- minute) temperature increase of the whole body. 10 Diagnostic ultrasound exposures of mammalian embryos or fetuses in vivo and in vitro do not cause a whole-body temperature increase in the mother but can potentially do so in the embryo. In principle, heating with ultrasound could occur so rapidly that the protective effects of HSPs might not come into play. There are data on effects of hyperthermia and measurements of in vivo temperature induced by pulsed ultrasound but not in the human These data have been widely reviewed However, there is a serious lack of data on the effects of ultrasound while rigorously excluding other confounding factors. A number of epidemiologic studies of possible developmental effects of obstetric ultrasound were performed before 1992, when exposures of the fetus, if anything, were lower on average than they are today. The results overall were negative. Around 1992, the maximum permitted acoustic output of clinical ultrasound instruments operating in the obstetric mode was allowed to increase by a factor of almost Potentially of even greater significance, no report clearly defines the duration of actual exposure. Epidemiology, of course, cannot be expected to reveal subtle effects. Today, ultrasound is so much a part of obstetric care that it would be very difficult to design an ethically acceptable epidemiologic study. The material in this article will be presented in the following manner: Definitions, Mechanisms of Tissue Heating, Measured Temperature Rise in Human Fetal Tissue, Intracranial Temperature Elevation, Epidemiologic Data, Clinical Studies, Discussion Regarding Obstetric Issues in the Human, and Conclusions and Recommendations. Definitions bioeffect Any biological change occurring as the result of an exposure, whether physiologic or harmful. damage An irreversible adverse outcome that has occurred. embryo The conceptus from fertilization through organogenesis (late third through eighth weeks gestation). This embryonic period is the time of greatest sensitivity. fetus The conceptus from organogenesis to delivery; in common usage, the term fetus encompasses the entire gestational period. harm An adverse outcome, whether reversible or irreversible. hazard A possible source of harm. 22 risk The combination of the probability of occurrence of harm and the severity of that harm. 22 safety Freedom from unacceptable risk 22 ; absolute safety cannot be achieved. Mechanisms of Tissue Heating Two widely accepted facts are that ultrasound has the potential to elevate the temperature of the tissues being scanned, and an elevated maternal temperature, whether from illness or exposure to heat, can produce teratologic effects. 7,9,27 32 The major question is whether diagnostic ultrasound can induce a harmful temperature rise in the fetus. 16 Some believe that this temperature rise is, in fact, a major mechanism of ultrasound bioeffects. 24,32 The temperature elevation in the insonated tissue can be calculat- 542 J Ultrasound Med 2008; 27:

3 Abramowicz et al ed and estimated fairly accurately 33,34 if the field is sufficiently well characterized. A temperature change in insonated tissues depends on the balance between heat production and heat loss. Local heating is proportional to the temporalaverage intensity. Acoustic power is more appropriate for the prediction of a temperature rise within the ultrasound beam, and this fact forms the basis of the thermal index (TI). Physical characteristics of the ultrasound beam modify the temperature rise. These characteristics include ultrasound frequency, focusing (which determines the beamwidth), whether the beam is scanned, and the duration of exposure. Some properties of the exposed tissues (eg, acoustic impedances and absorption) also influence, respectively, how much ultrasound energy is reflected and how much is transformed into heat. 33 Conduction and convection influence the rate at which heat is transported locally from the sites of its generation. One other tissue property that strongly influences the amount of heat transported is local perfusion, a form of convection, which very clearly diminishes the risk, if present. Similar experimental conditions caused a 30% to 40% lower maximal temperature increase in live versus dead sheep fetuses exposed in the near field, 13 while in guinea pig fetuses exposed at the focus, the difference was approximately 10%. 14 These results were thought to be secondary to vascular perfusion in live animals. A significant cooling effect of vascular perfusion was observed only when the guinea pig fetuses reached the stage of late gestation near term, when the cerebral vessels were well developed. In the midterm, there was no significant difference when guinea pig fetal brains were exposed, alive (perfused) or postmortem (nonperfused), in the focal region of the ultrasound beam. 35 The loss of heat due to perfusion is more important in those parts of the beam close to the transducer and beyond the focus, where the beam has a large cross-sectional area. In the focal region, where the beam is narrow, the effect of perfusion (in the absence of thermally significant vessels) is much less 36 because conduction radially down the large thermal gradient is the dominant mechanism for heat transport away from the focal region. It is worth noting here that in early pregnancy (<6 weeks), there appears to be minimal maternalfetal circulation, that is, minimal fetal perfusion, which may potentially reduce heat dispersion. 37 The lack of perfusion is one reason why the spatial-peak temporal-average intensity (I SPTA ) for ophthalmic applications has been kept very low, in fact, much lower than peripheral, vascular, cardiovascular, and even obstetric scanning, despite the general increase in acoustic power that was allowed after Herman and Harris 38 have provided the rationale for the ophthalmic exposure output limit of the US Food and Drug Administration (FDA), initially 17 mw/cm 2 but increased to 50 mw/cm 2 in the 1997 revised 510(k) guidance to the industry. 39 The FDA s rationale was based on the relatively large absorption in the lens and orbital fat and the poor perfusion in the eye. The FDA initially based its findings on using (calculating) 720 mw/cm 2 but applied it to physical characteristics of the eye (no perfusion and some protein) and noted that temperature increases in the 8 C to 10 C range were obtained. The FDA then calculated that if the output was kept at or below 50 mw/cm 2, the temperature rises were all less than 1 C, hence the output limit of 50 mw/cm 2 for the eye. There are some similarities in physical characteristics between the early first-trimester embryo and the eye. Neither is perfused; they can be of similar size; and protein is present (in an increasing proportion in the fetus). The present TI values for first-trimester scanning are based on the usual perfusion rate for tissue. At about weeks 4 to 5, the gestational sac is about the size of the eye (2.5 cm in diameter), and by week 8, it is around 8 cm in diameter. There is no perfusion in very early gestation. Only at about weeks 10 to 11 does the embryonic circulation actually link up with the maternal circulation. 40 At that stage, there is a switch from previously hypoxic conditions to normo-oxic conditions and a cascading of radical scavengers (eg, superoxide dismutase) to help deal with the free radicals. 37 It appears there may be some underestimation of the actual diagnostic ultrasound-induced temperature in early gestation, mainly because of the absence of perfusion. For example, at week 5, the oxygen tension in the embryonic sac is about 10 mm Hg, which is well below normo-oxic levels 41 ; the ten- J Ultrasound Med 2008; 27:

4 Fetal Thermal Effects of Diagnostic Ultrasound sion in the eye is about 15 mm Hg. These values are surrogate indicators of the absence of perfusion. Only when the embryonic circulation links up with the maternal circulation does this value change and substantially increase. 42,43 The perfusion issue is in addition to modifications of tissue temperatures due to ambient maternal and fetal temperatures. The questions of tolerance development and the cumulative effect have not been fully evaluated. Furthermore, motions (even very small) of the examiner s hand as well as the patient s breathing and body movements (in the case of obstetric ultrasound, both the mother and the fetus) tend to spread the region being heated. However, for spectral (pulsed) Doppler studies, it is necessary to have the transducer as steady as possible. This is because, in general, blood vessels are small in comparison to the general organ or body size being scanned with B-mode imaging, and hand movements while performing Doppler studies will have more undesired effects on the resulting image. The intensity (I SPTA ) and acoustic power associated with Doppler ultrasound are the highest of all the general-use categories; hence, the associated TI values and potential for heating are also the highest. Ziskin 44 reported that among 15,973 Doppler ultrasound examinations, the average duration was 27 minutes (and the longest 4 hours). It is important to note at this point that no actual damage attributable to heat generation by ultrasound has been described in the human medical literature. Measured Temperature Rise in Human Fetal Tissue Those wishing to measure the temperature rise in the gravid human uterus in vivo during ultrasound scanning must contend with almost insurmountable problems, both from the difficulty in managing a controlled experiment and from ethical considerations. A recently funded project to carry out such experimental studies in the United Kingdom has not yet been reported (G.R. ter Harr, PhD, DSc, communication, 2006). In vitro studies must therefore be used. In one study, Doody et al 45 reported the temperature increase generated in vitro at the surface of samples of human fetal vertebrae ranging in age from 14 to 39 weeks. The exposure conditions (3.5 MHz, a weakly focused beam, and acoustic power of 50 mw) compared well with estimated in situ exposure conditions toward the upper end of pulsed Doppler ultrasound use. For a fully ossified vertebra from a 39-week fetus, a temperature rise of 1.8 C after 5 minutes of exposure was reported. For vertebrae from less mature fetuses, lower temperature rises were observed, which were related to the lower degree of ossification. For the 14-week vertebra studied, a temperature rise of no more than 0.6 C was reported. Another experimental approach uses tissuemimicking materials in the form of thermal test objects (TTOs). 46 Thermal test objects have been used to explore the validity of the TI as a predictor of a temperature rise in a significant study carried out by the National Physical Laboratory (Middlesex, England). Atkins and Duck 47 have reported results from a more limited study using a soft tissue TTO and two pulsed Doppler beams, one of them in the color Doppler mode. When surface heating effects were ignored, the maximum temperature rises in the TTO were 2.8 C after 10 minutes of exposure in the pulsed Doppler mode (3 MHz and 290 mw/cm 2 ) and 0.4 C in the Doppler imaging mode. When compensation was included for the attenuation of overlying tissue (the measurements being made using water coupling), the temperature rises were reduced to 0.95 C and 0.21 C, respectively. Higher temperatures were reported when a bone TTO was used. Use of such a phantom for precise fetal exposure calculations would need specific modification to the case of the fetal absorption coefficient, which is presently unknown. 48 The potential for nonlinear propagation effects to enhance heating has been explored both experimentally and theoretically by, among others, Cahill et al. 49 A condition was used that approximated an obstetric examination, using assumptions of a fixed attenuation near the source, transmission through a low-attenuation liquid, and a tissue-equivalent target. Both the model and experiments found that the temperature rise may be enhanced by up to 50% when compared to linear propagation calculations, but that for higher pulse amplitudes, the temperature rise may be reduced because of increased dissipation of energy by the liquid. Current clini- 544 J Ultrasound Med 2008; 27:

5 Abramowicz et al cal scanners can, under some conditions, operate with a displayed TI of 6.0 or higher. If one misinterprets the TI as accurately representing the maximum possible temperature rise in vivo, one may predict a worst-case temperature rise of 6 C or a rise from an average 37 C to 43 C in vivo. Temperature increases similar to this have been shown in vitro with commercial systems. 50 A recent publication described TI values remaining very low during routine obstetric examinations but occasionally reaching 1.5 for very brief periods. 51 Intracranial Temperature Elevation The extraordinary growth in the use of ultrasonic imaging as a diagnostic tool in medicine has led to the development of a wide range of specialized procedures and the evolution of sophisticated and powerful modern equipment. Improvements in resolution and image quality and in gray scale definition have been particularly important in obstetrics. However, it is important to realize that some developments of novel technology have been accompanied by increased acoustic outputs, 52 such that modern equipment is capable of emitting beams with acoustic intensities sufficient to produce measurable effects in biological tissue. Pulsed Doppler sonography, spectral flow analysis, and Doppler color flow imaging techniques offer the potential to increase diagnostic effectiveness and may be attractive for applications in early pregnancy or embryonic sonography. During ultrasound examinations, acoustic energy is absorbed by tissue and converted into heat. The temperature rise depends on a number of factors, including tissue properties and the ultrasound exposure conditions (see above). In simple ultrasound scanning, using gray scale B- mode, modest energy levels are applied in a constantly moving beam, so that there is negligible opportunity for tissue heating to occur. On the other hand, Doppler techniques present the highest risk of inducing thermally mediated biological effects. This follows from the use of higher pulse repetition rates and longer pulse lengths than are used in gray scale imaging and from the use of a stationary beam in spectral Doppler measurement of flow. The skull bone, with its high acoustic absorption coefficient, has a significant effect on ultrasound interactions and potential bioeffects. Bone attenuates the amount of energy that penetrates into deep brain tissue; meanwhile, the absorbed acoustic energy is rapidly transformed into heat in the bone and the nearby soft tissue of the cerebral cortex. The subject of ultrasound-induced heating and its biological consequences has been examined in detail by groups of international experts in two symposia sponsored by the World Federation for Ultrasound in Medicine and Biology (WFUMB). The first of these symposia identified the pulsed Doppler mode as being capable of producing biologically significant temperature increases in certain unperfused tissue. 53 Two WFUMB publications 18,53 have highlighted the significance of thermal bioeffects as a potential risk factor in diagnostic sonography. Since these WFUMB symposia took place, there have been several publications showing that diagnostic pulsed ultrasound can produce substantial temperature increases in live perfused fetal brains when insonated within the uterus. 11,13,54,55 These findings will have an impact on the development of relevant safety guidelines. Meanwhile, the FDA Center for Devices and Radiological Health has relaxed its regulatory limit in the United States, such that the embryo or fetus may be exposed to substantially increased acoustic intensity (nearly 8 times compared to pre-1992 limits), provided the equipment incorporates a real-time output display. The display is intended to provide an indicator of the risk of producing bioeffects for each examination in each mode of operation or whenever the equipment output controls are altered. The rationale for the output display is to alert the user to potential bioeffects. This option of selfregulation increasingly places the responsibility on the user/clinician to maximize the benefit of ultrasound examinations while minimizing the risk. The FDA-approved output display standard 56 describes a TI as an indicator of the risk of producing thermally mediated biological effects and a mechanical index (MI) as an indicator of the risk of producing mechanically induced effects. The concept of using an index value seems simple. In practice, it often can be misleading because the calculations are based on J Ultrasound Med 2008; 27:

6 Fetal Thermal Effects of Diagnostic Ultrasound various assumptions about average properties of biological tissues and ultrasound parameters as the modeled beam propagates through homogeneous tissue, whereas actual tissues are heterogeneous. Some studies have shown limitations of the TI as an estimator of a temperature rise and have shown that, in certain circumstances, the index can underestimate the potential for a temperature increase. 55,57 59 An example of this limitation is described below. There is ample published scientific evidence to show that exposure to diagnostic ultrasound can produce significant temperature increases in the fetal brain near bone. 13,36,60,61 The critical questions that need to be addressed are whether there is evidence that: 1. The extent of the ultrasound-induced temperature rise is sufficient to create a hazard; 2. There are neurophysiologic effects or responses to clinically relevant exposures; 3. There are unexplained epidemiologic effects; and 4. There is a significant risk to embryonic and fetal development. Ultimately, these data are used to establish conclusions about thresholds for temperature-mediated biological effects. The following text describes some fundamental research that has evaluated the extent of ultrasound-induced heating in the brain and further studies on the influence of changes in exposure conditions and physiologic conditions on the extent of heating. Experimental Evidence The data reviewed here are derived from a series of published collaborative studies in which one of the authors (S.B.B.) participated in different laboratories in Australasia. To facilitate comparisons of data, the same ultrasound exposure equipment was used throughout. The ultrasound exposure parameters are given together with reference to the original articles in Table 1. In all studies, the ultrasound beam was stationary so that a single-point tissue target was constantly insonated with pulsed ultrasound. This situation represents the worst case for spectral pulsed Doppler examinations. Other studies that used handheld thermocouples and ultrasound probes are not considered in this evaluation. There are variations in the animal models used, and some studies involved insonation in the near field to explore the effects of tissue heating over a larger area of tissue using a wide beam. Temperature Increase and Fetal Age Ultrasound-induced temperature elevations in fetal brains of time-mated guinea pigs were measured during in vitro/ex utero insonation. The greatest temperature rise occurred in brain tissue close to bone and correlated with both gestational age and progression in bone development. 11 The skull bone becomes progressively thicker and denser (measured by mineral content) from 30 to 60 days gestational age (dga). Normal gestation for guinea pigs is 66 to 68 days. After 120 seconds of insonation with an I SPTA of 2.9 W/cm 2, the mean maximum temperature increases were 1.2 C, 1.6 C, 3.7 C, and 5.2 C for 30-, 40-, 50-, and 60-dga fetuses, respectively (Figure 1). The temporal-average intensity is about 4 times the in situ intensity currently permitted by the FDA for diagnostic use. The mean peak temperature elevation of greater than 5 C in 60-dga fetuses was subsequently confirmed in an independent study using identical ultrasound conditions. 54 The ultrasound exposure was also found to pro- Table 1. Exposure Parameters for Different Studies Reference Frequency, MHz Power, mw I SPTA, W/cm 2 Beamwidth, cm Duration, s Mean ΔT, C Animal Model Bosward et al Guinea pig fetus Horder et al Guinea pig fetus Guinea pig fetus Horder et al 14, Guinea pig fetus Horder et al Guinea pig fetus Duggan et al Sheep fetus Duggan et al Neonatal pig The site of temperature measurement varied between studies. Intensity was measured in a free field in water, not in situ. 546 J Ultrasound Med 2008; 27:

7 Abramowicz et al duce an increase of 2.6 C in the midbrain when the bony cranium was removed from 60-dga fetuses. There was no significant difference in the results obtained for formalin-fixed or fresh tissue. Most of the heating (80% of the mean maximum temperature increase) occurred within 40 seconds. The rate of heating is relevant to the safety of pulsed Doppler examinations in which the dwell time may be an important factor. The results show that the amount of the ultrasoundinduced temperature increase that occurs in the fetal brain near bone is directly related to the amount of mineralization deposited in the bone, that is, the fetal age. The implications of this finding are that the worst-case heating will occur later in pregnancy (from the second trimester). Therefore, ultrasound-induced heating is not as intense during embryonic development where there is no bone. Fetal Brain Insonated In Utero Pulsed Doppler ultrasound examination of the fetal cerebral circulation may cause potentially harmful temperature elevations in brain tissue immediately beneath the insonated segment of the skull. Temperature increases were measured at various depths in the brain of living thirdtrimester (57- to 61-dga) fetal guinea pigs during in utero exposure to a narrow focused beam of unscanned pulsed ultrasound at an I SPTA of 2.8 W/cm Mean temperature increases of 4.9 C close to the parietal bone and 1.2 C in the midbrain were recorded after 120 seconds. Cerebral Figure 1. Mean temperature elevations in fixed guinea pig fetal brain tissue during 120-second exposures to 260-mW pulsed ultrasound. A single thermocouple was placed under the parietal bone at a distance of 6.0 cm from the transducer (after Bosward et al 11 ). blood perfusion had a negligible cooling effect on ultrasound-induced heating in these guinea pig fetuses (Figure 2). When this study was repeated in older nearterm (62- to 66-dga) guinea pig fetuses, a statistically significant but small (12%) difference was found in the mean peak intracranial temperature increase for live (perfused) versus dead (nonperfused) fetuses exposed in utero. 14 This study was designed primarily to test the effectiveness of the thermal index for bone (TIB) as an estimator of the intracranial ultrasound-induced temperature increase. Even though the temperature was measured on the inside of the cranium (ie, not the outer surface of bone proximal to the transducer, as defined by the TIB), and the temperature was reduced by cerebral blood flow, the TIB was found to underestimate the value for the temperature increase. The ratio of the measured temperature to that estimated by the TIB was 1.3. In subsequent continuing studies, a larger difference between the TIB and the measured temperature increase was found for the outer surface of the fetal skull. 55 Effect of Cerebral Perfusion on Heating A study with anesthetized neonatal pigs measured the effect of variations in cerebral blood flow (normal, double, and absent flow) on ultrasonic heating of the cerebral cortex (Table 2). 62 Cerebral blood flow was measured using the radiolabeled microsphere technique. The study Figure 2. Mean temperature measured in in utero 57- to 61-dga fetal guinea pig brains, live/perfused (dashed line) and dead/unperfused (solid line), as a function of depth below the parietal bone during exposure to 240-mW pulsed ultrasound focused on the outer surface of the parietal bone. The water path is 6.6 cm; the 6 db beamwidth is 0.26 cm (after Horder et al 35 ). J Ultrasound Med 2008; 27:

8 Fetal Thermal Effects of Diagnostic Ultrasound Table 2. Results of Ultrasound-Induced Temperature Increase in Brain Tissue Under Different Experimental Conditions Temperature Increase in Brain of Temperature Increase in Brain of Live Animals After Insonation, C Dead Animals After Insonation, C Reference 30 s 60 s 90 s 120 s 30 s 60 s 90 s 120 s Bosward et al 11 NR NR NR NR Horder et al 54 NR NR NR NR Horder et al 14 (>5 mm depth) Horder et al 61 * Duggan et al Duggan et al 62 NR NR 1.5 NR NR NR 1.5 NR NR indicates not reported. *Temperature measured at depth in the brain at the sphenoid bone (Refer to Figure 4). also compared the temperature increases achieved for a narrow beam with those achieved for a wide beam. Pulsed ultrasound was applied to the neonatal pig brain, insonating with an I SPTA of 1.4 W/cm 2 at the 0.3-cm-diameter beam focus or with the brain positioned in the 1.6-cmdiameter near field of the beam, where the intensity was 3.6 W/cm 2. In all exposures to the focused narrow beam, there was no difference in the heating curves for normal, increased, and no cerebral blood flow, for which mean peak temperature increases of 1.5 C were recorded after 90 seconds (Figure 2). However, when insonated with the wide beam, the mean peak temperature increases in the cerebral cortex were significantly different for the 3 rates of blood flow: mean temperature increases of 1.9 C (normal flow), 1.7 C (increased flow), and 2.4 C (no flow) were recorded. There was a 20% reduction in the mean peak temperature rise after 90 seconds in the healthy live neonate when compared with the dead nonperfused neonate (Figure 3). However, it was found that an increase of 300% above the normal flow rate produced only a marginally (9%) further reduction in the mean maximum temperature increase. In another study on the possible cooling effect of cerebral perfusion, 13 perfusion in the living versus the dead sheep fetuses was found to result in a 40% reduction in the mean peak temperature increase when insonation was with a similar wide beam. Ultrasound was applied to the conscious nonanesthetized near-term fetal lamb, and the temperature was measured at a depth of approximately 2 mm in the cerebral cortex under the parietal bone. The results of these studies comparing heating in wide and narrow beams support the theory underlying the bioheat equation 24,53 that perfusion affects heating by broad beams to a greater extent than that of narrow beams. The large temperature gradient in narrow, focused beams allows optimum heat dissipation so that any further cooling effects of vascular perfusion are minimal. Effect of Ultrasound-Induced Heating on Brain Electrophysiologic Mechanisms The temperature increase was measured deep in brain tissue adjacent to the sphenoid bone in fetal guinea pigs during in utero exposure to pulsed ultrasound at an intensity (I SPTA ) of 2.8 W/cm 2. This is clinically relevant for Doppler flow studies on middle cerebral vessels. Sensitive important homeostatic hormone- and growth-regulating structures, such as the hypothalamus and the pituitary, lie in the region commonly targeted Figure 3. Heating curves from the brain of a neonatal pig during ultrasound exposure with a wide beam (after Duggan et al 62 ). 548 J Ultrasound Med 2008; 27:

9 Abramowicz et al during examination of the cerebral vessels. After 120 seconds of exposure, a mean temperature increase of 1.5 C was measured at the sphenoid bone (Figure 4). This temperature is readily conducted to the adjacent hypothalamus, which regulates the body temperature by adjusting the heart rate. However, fetal electrocardiographic measurement throughout these exposures showed that ultrasound-induced heating of the hypothalamic region by 1.5 C during a brief exposure (total time, 120 seconds) did not significantly alter the normal mean fetal heart rate of 250 beats per minute. 61 Another study found no difference in the electrocortical activity in the brains of conscious nonanesthetized fetal sheep during intrauterine exposure to low-intensity (0.3 W/cm 2 ) pulsed ultrasound. 63 Ultrasound was delivered by a wide beam in the near field of a transducer that was surgically implanted onto the fetal skull. The study was designed to provide empirical data from a reliable animal physiologic model to investigate previous reports of increased fetal activity after examinations with diagnostic ultrasound. The temporal-average intensity applied was more comparable with that of B-mode scanning than current pulsed Doppler exposures. The results gave no evidence of the fetal electrophysiologic response to low-intensity ultrasound exposures. Figure 4. Mean temperatures measured in utero at the sphenoid bone in 62- to 66-dga fetal guinea pig perfused and unperfused brains during exposure to 240-mW pulsed ultrasound focused on the sphenoid bone. The axial distance is 6 cm through a water path from the transducer to the maternal abdominal surface or 7 cm to the sphenoid bone (after Horder et al 61 ). The current published WFUMB policy statements on thermal effects of pulsed Doppler ultrasound resulted from expert interpretation of bioeffects data at the time of the last WFUMB safety symposium in April Since then, a number of publications have shown that a significant temperature increase is produced in living perfused brain tissue. Furthermore, the extent of heating is not significantly reduced by vascular perfusion during exposure to clinically relevant narrow-focused beams. Epidemiologic Data For detailed analysis of epidemiology literature, as it relates to possible fetal bioeffects of ultrasound, see a separate article related to the American Institute of Ultrasound in Medicine epidemiology statement. 64 Health, Safety, and Risk: Why Should There Be Concern at All? It is certain that diagnostic ultrasound has had an enormous beneficial effect on patient health and welfare. It is a noninvasive procedure, widely available, and with excellent resolving capabilities. The topic of safety, however, is not so easily addressed. Safety is another way of discussing risk. We know that great benefit has been derived from the clinical use of diagnostic ultrasound, but there is uncertainty about its risk. This uncertainty arises primarily from the fact that there has been (1) no clinical evidence of any bioeffects or side effects from exposure to diagnostic ultrasound and (2) uncertainty as to the relevance of (a) theoretical insights about the insonating conditions leading to the occurrence of heating and nonthermal mechanisms of action and (b) reports of bioeffects from in vitro and in vivo chemical and nonhuman biological systems apparently relevant to the topic of safety. If there is a risk, it is, most likely, very subtle and therefore not so easily detected, even after 4 decades of use. It is important to bear in mind that throughout this 4-decade period, there has been a steady increase in acoustic outputs of diagnostic ultrasound devices (allegedly to gain better diagnostic information). There are major problems in designing appropriate (particularly human) studies. J Ultrasound Med 2008; 27:

10 Fetal Thermal Effects of Diagnostic Ultrasound One major question to resolve is whether a temperature threshold exists for inducing birth defects. This is a legitimate and relevant question because there is little information on this topic. This information would be useful to know from both basic science and clinical points of view. One can use the rat to illustrate the various difficulties involved with designing adequate studies. The sample size needed to detect various increments in background rates of major malformations in rat fetuses is the first parameter to consider. Rats have a background rate of 1 malformed fetus per 250 rats (ie, the rate of occurrence of malformations in the rat is [0.4%] 65 ). A 2-proportions power analysis indicates the numbers of individuals (ie, fetuses) and litters needed to determine with reasonable statistical power normal background rates for major malformations in the rat. It is obvious that a larger effect requires a smaller sample size than a smaller effect. For instance, detection of a doubling of the background rate of malformations would require 11,370 individuals, or approximately 1138 litters. At present, data on hyperthermia-induced defects in rats are meager because the usual experiments have involved 5 to 15 litters per regimen. 66,67 These sample sizes are simply far too small to gain any insights about whether temperature thresholds exist. However, exposure conditions have not been extended to determine the presence or absence of a threshold. It is unknown whether the threshold would diminish or disappear for greatly extended durations and different bioeffect end points. Although no relevant data exist, there are many opinions about whether a temperature threshold exists. Edwards 7 suggests somewhere between 1.5 C and 2.5 C above normal physiologic levels; Kimmel et al 67 suggest 2 C above normal physiologic levels; WFUMB 18 suggests 1.5 C; the National Council on Radiation Protection and Measurements 6 suggests 1.0 C; and the International Radiation Protection Association/International Non-Ionizing Radiation Committee 68 suggests 0.5 C. However, Carstensen and Gates 69 state that hyperthermia-induced damage scales linearly with time and exponentially with temperature. This latter position is quite analogous to the empirically determined thermal dose concept of Sapareto and Dewey 70 for hyperthermia-induced killing of human cells. This implies an Arrhenius relationship between thermal exposure and the rate of induction of an adverse biological effect 71,72 and, therefore, that there is no threshold for biological change (namely, enzymatic reaction rates). However, the magnitude of the induced effect at low doses may be too small to be detectable in even very large sample sizes, or the biological changes might not, in fact, lead to adverse effects. The issue is even more complicated when examining the above aspects in humans. This is precisely the domain of epidemiology. Suppose one wants to attempt to determine whether obstetric diagnostic ultrasound induces a temperature increment sufficient to cause birth defects in humans. Analyses will be limited to the first trimester because several embryologic changes (neural tube closure [NTC] in particular) have been shown in animal systems to be very heat-sensitive stages. We know that (1) the background rate for major malformations in humans is about 3%, and (2) the occurrence of diagnostic ultrasound scans during early pregnancy is increasing because of in vitro fertilization procedures and other medical reasons. The rationale for selecting a heat-sensitive stage of embryogenesis for studying this problem is that if a threshold is determined for a sensitive stage, then one can be certain that, if one remains below that threshold level when scanning during that stage as well as all other stages of pregnancy, there is reasonable assurance that diagnostic ultrasound will not induce a thermal effect on the embryo or fetus. 24 A simplistic statistical assessment with the usual criteria (80% power; α =.05), with a basis of a 3% background rate of major malformations in humans, will show that for detecting doubling, tripling, quadrupling, and factor-of-10 increases above the background rate, the total sample sizes (divided equally between control and exposed ) are 1498, 490, 268, and 58, respectively. These numbers are considerably smaller than those associated with the same anticipated increments in malformations in animal model systems in which the anticipated rate of background occurrences is approximately an order of magnitude smaller than that of humans. The problem of epidemiologically studying whether diagnostic ultrasound-induced temper- 550 J Ultrasound Med 2008; 27:

11 Abramowicz et al ature increments cause an increase in birth defects is potentially complicated by the fact that some stages of pregnancy are more sensitive to hyperthermia-induced birth defects than others. 73 In general, NTC is recognized as a heatsensitive stage. Thus, from a retrospective approach, it would be necessary to have access to patient records of the dating of the pregnancy and the timing of the diagnostic ultrasound application as well as its thermal dose, as represented by the type and duration of exposure and the temporal-average intensity. One would also want to ensure that all patients were nutritionally comparable. It would be necessary to ensure that patients taking vitamin supplementation were not mixed in with patients not taking such supplementation. For a prospective study, it would be necessary to have the same information, except that the scans would take place given medical reasons for conducting them during NTC. An additional complication is that nearly all pregnancies are scanned with diagnostic ultrasound. Thus, a control group may not be available. It may be possible to devise a situation in which the fetal outcome results of firsttrimester (NTC) scans are compared against those of subsequent diagnostic ultrasound scans. A problem with nearly all diagnostic ultrasound scans from an epidemiologic basis is the total absence of exposimetry information. Knowledge of information available about the scan (eg, temporal-average intensity, acoustic frequency, pulse length, and pulse repetition frequency, as presented in the owner s manual) would be very useful. 74 There appear to be some reasons for concern that modern diagnostic ultrasound devices are capable of inducing temperature increments in soft tissue known to be teratogenic in animals and that such diagnostic ultrasound exposures are occurring at times when the mammalian embryo is known to be sensitive to thermal insult. There are two major mechanisms by which bioeffects are known to occur: thermal and nonthermal. We do not expect to see nonthermal effects arising in the embryo or fetus simply because there appear to be no stabilized gas bodies in those structures. Mechanical phenomena may become very relevant if procedures arise by which the embryo or fetus is treated with echo contrast agents to enhance diagnostic capability. 75 It therefore seems more likely that a diagnostic ultrasound-induced thermal mechanism could be operating because the physics (sufficient acoustic output) and the biology (thermally sensitive stages) of obstetric settings are conjoined. Furthermore, this situation may worsen because acoustic outputs of instruments are rising 76 and will be exacerbated if the present regulated limitations on allowable SPTA acoustic emissions of diagnostic ultrasound devices are further relaxed or eliminated. 77,78 Clinical Studies Several biological end points have been analyzed in the human fetus/neonate: intrauterine growth restriction (previously named growth retardation) and low birth weight, delayed speech, dyslexia, mental development or behavioral issues, and, more recently, non right-handedness. Occasional studies report an association between diagnostic ultrasound and some specific abnormalities such as lower birth weight, 79 delayed speech, 80 dyslexia, 81 and non righthandedness. 82,83 With the exception of low birth weight (also shown in monkeys 84 ), these findings have never been duplicated, and findings from most studies have been negative for any association. For further details, please see American Institute of Ultrasound in Medicine epidemiology statement. 64 Recently, a follow-up study to the famous low-birth-weight study 79 was published with up to 8 years follow-up of children who were insonated in utero, and no long-term differences were shown with controls in growth and measures of development. 85 A major problem with formulating hypotheses from animal data is that in vivo effects of hyperthermia seen in animals result from exposure situations that would never be encountered by the human embryo or fetus during ultrasound examination. It is all to do with the volume of exposed developing tissue. While there exist reliable data on teratogenic effects of hyperthermia in animals (mostly rodents), these data are based on whole-body exposure of both the embryo and mother in situations in which the elevated temperature is consistently maintained for relatively J Ultrasound Med 2008; 27:

12 Fetal Thermal Effects of Diagnostic Ultrasound long periods compared with that likely to occur in a diagnostic ultrasound examination. In the human, insonation will generally be of limited parts of the fetus and the mother. However, while this is true in the second and third trimesters, early in pregnancy (up to week 10), the entire embryo/fetus is insonated. The size of the embryo is approximately 1 to 2 mm at 5 weeks, 5 to 6 mm at 6 weeks, 10 mm (1 cm) at around 6 to 7 weeks, and about 2 cm at 8 weeks and reaches 3 cm at 10 weeks (when it is now called a fetus). These sizes are well within the range of the beamwidth at a depth of 10 to 20 cm. With a whole-body temperature increase, the animal mother is stressed and compromised physiologically. Blood pressure, heart rate, systemic corticosteroid levels, stress hormone levels, and stress protein levels all would increase when the maternal core temperature is elevated greater than 5 C in typical hyperthermia experiments. We know from radio frequency experiments with nonhuman primates that only a 1 C increase in core temperature results in significant behavioral (and, presumably, physiologic) changes This result forms the basis of international safety standards for radio frequency radiation. Reports from human studies show that a similar modest increase in body temperature (<1 C) elicits physiologic responses, including sweating and an increased heart rate. 89 The National Institute for Occupational Safety and Health 90 and the American Conference of Governmental Industrial Hygienists 91 recommend an upper threshold limit value of a rise in body temperature of 1 C. This was endorsed by the World Health Organization. 92 In an ultrasound examination, the maternal physiologic state is not perturbed by environmental heating, so it is unlikely that induced systemic effects would have an impact on fetal development (that is, beyond any stress-induced effects that the clinical examination might induce). What About Synergistic Effects of Ultrasound and Raised Temperature? In studies that were designed specifically to examine the direct effects of the combination of ultrasound mechanical interaction with bulk heating, some interesting results emerged. The collaborative research joined the expertise of growing whole embryos in culture with a custom-built and calibrated ultrasound exposure system. 93 Whole rat embryos were grown in culture in the laboratory in a carefully controlled environment. This experimental model was well established and had been previously used in published studies on teratogenic effects of hyperthermia. Bioeffects end points and thresholds for hyperthermia had already been well established for the rat embryo when heated at a specifically sensitive stage in gestation: exactly 9.5 dga. In work that set out to examine the effects of combining hyperthermia and ultrasound exposure (culture vessel heated and pulsed ultrasound applied), a synergistic effect was implicated in the response of delayed somite development and enhanced HSP synthesis occurring with an increase in the whole-embryo body temperature (through bulk heating of surrounding culture media) of only 1.5 C above the normal culture temperature for rat embryos. The investigators found various effects at lower hyperthermic exposure levels (temperature increase duration of application) when ultrasound was applied (as opposed to whole-body hyperthermia alone). For example, previous work with this embryo assay system showed that a temperature elevation to 40 C for 40 minutes alone had no detectable effect on embryonic growth. 10 In contrast, adding ultrasound to modest hyperthermia produced a significant reduction in the neural protein level and an altered head-to-body surface area ratio. Exposure to ultrasound for 15 minutes at 40 C caused significant reduction in the growth of the head compared with that of control embryos. 94 The exact mechanism was not identified. Such an exposure situation (embryos [or oocytes] being pushed around by an ultrasound beam in a fluid environment) may occur in in vitro fertilization. 95,96 Considering the potential ramifications of perturbed embryonic development and the extreme financial cost factor in in vitro fertilization treatments, it seems that this topic should be carefully explored. Clinical Relevance in the Human There are problems in relating laboratory animal studies to patients. First and foremost, one cannot necessarily extrapolate from animal studies 552 J Ultrasound Med 2008; 27:

13 Abramowicz et al to humans or even from one animal species to another because differences in outcome may, in fact, be due to strain differences. 97 Animal studies are usually performed in an attempt to obtain some effects, and the values of exposure parameters are usually pushed to some extreme not applicable to human scanning, such as an I SPTA of W/cm 2, 98 or even 1500 W/cm On the other hand, the lack of findings is just that. It does not prove safety. Subtle behavioral or psychological changes would be extremely difficult to show. Even more obvious anomalies might not be directly due to one etiology but could be secondary to the presence of multiple confounding factors. In addition, even when a procedure might carry some yet unidentified risk, one has to balance the risks versus benefits. This is why most national and international medical organizations adhere to the principle of performing ultrasound only when clinically indicated and then use the ALARA (as low as reasonably achievable) principle for output power. It is important to note that if there is a threshold, and exposure is kept below this threshold, then there is no need to apply the ALARA principle. It is beyond the scope of this article to describe the clinical indications for ultrasound use, 100 but when they are present, the advantages greatly outweigh the potential risks. An important point to remember is that the indices, TI and MI, are estimators of relative risk based on assumed simple physical models. They are most meaningful for tracking changes of operating conditions of equipment when the operator is heeding ALARA and comparing displayed index values with those displayed immediately beforehand. The TI was not intended to indicate a temperature increase in degrees Celsius in numerous selected clinical situations, although many interested parties continue to criticize it for not doing so accurately. In fact, these indices may be underestimating the actual in situ acoustic pressure in clinical practice. 101 This reservation is substantiated for the MI. 102 The presence of harmonic frequencies (sound waves generated at exact multiples of the original frequency) in particular may cause overestimation or underestimation of the TI. 103 As previously mentioned, if one misconstrues the TI as a direct temperature indicator, the underestimation can be as high as 1.4 C and 1.8 C for B-mode and color imaging, respectively, and up to 5.8 C for pulsed Doppler imaging when bone is insonated through a liquid layer with low attenuation. 59 This needs to be kept in mind in obstetric ultrasound with studies of fetal intracranial vasculature. This question of applicability of the indices, particularly in regard to the TI, is a major concern. 32,104 The exposure variables are one part of the equation. The second component is the insonated tissue. This also poses somewhat of a quandary because the sensitivity of the target tissues is not precisely known and thus demands awareness by the operator, at least, of the acoustic output and exposure time. 105 The issue of a long path through fluid (such as a full bladder in first-trimester scanning or polyhydramnios) has to be considered, too, because estimation of temperature elevation by invoking the TI may be an underestimation by a factor of 2 or even more. 106 Heating at the transducer surface tends to be ignored, but clinicians have the experience of some patients reporting a warm feeling at the point of contact with their skin. This is particularly relevant in intracavitary scanning, such as transvaginal ultrasound. Discussion Regarding Obstetric Issues in the Human While above-mentioned studies present no positive information (which leads to the natural conclusion that there is no risk), there are a few issues that appear to be pertinent to obstetric diagnostic ultrasound that seem not to have been discussed. Some have been addressed above, to some extent, but will now be detailed as they relate to actual clinical reality. 1. For most of the first trimester, the embryo is hypoxic, 37,41 and the only physical mechanism for heat dissipation is diffusion. Perfusion is not a pertinent factor. The embryonic heartbeat begins about week 6, and with that begins the development of an embryonic circulatory system. By week 11, the placenta is reasonably organized, and the rudiments of a perfusion-based circulatory system are in place. For early diagnostic ultrasound scans, however, it is well to be J Ultrasound Med 2008; 27:

14 Fetal Thermal Effects of Diagnostic Ultrasound aware that the embryo s main heat dissipation mechanism is diffusion, not perfusion. This has an effect on anticipated diagnostic ultrasound-induced heat removal and may mislead an operator who interprets the TI as a precise indicator of tissue temperature because the TI assumes some level of perfusion. Thus, the embryo would be expected to be a little warmer than expected because of the absence of perfusion. However, a mitigating effect is that embryonic soft tissues may be less absorbing than adult soft tissues, which would then cause the embryo to be less warm than expected. 2. Later in pregnancy (third trimester), it becomes apparent that the placental system is not a fully efficient mechanism for perfusive dissipation of heat because the fetus becomes (naturally) warmer than the mother by about 0.5 C Whenever a pregnant woman visits her obstetrician, and the woman is febrile, the obstetrician will often scan the fetus just to be sure that everything is OK, thereby increasing the thermal load to the fetus. These points are important because the Arrhenius relationship for hyperthermiainduced effects from a chemical rate perspective is linear with time but exponential with temperature. For example, a thermal effect seen at one temperature might require only one tenth the exposure time at a 2 C higher temperature. This would not be indicated by the displayed value of the TI, and because, as previously stated, there can be differences amounting to a factor of 0.5, 2, or even 6 in the TI, the problem is compounded even more. 6 To our knowledge, the present literature on obstetric scanning ignores some basic obstetric facts such as lack of perfusion in the first few weeks of pregnancy, as mentioned above, which would lead to higher temperature change, ΔT, values than predicted. It should be noted however that, conversely, differences in protein and fat content between fetus and adult might change the ΔT in the opposite direction. In addition, there are some further problems with the epidemiologic literature in terms of obstetric diagnostic ultrasound scanning. When an effect is reported, considerable effort is made to scrutinize the protocol and verify whether there may be some deficiency that renders the report questionable. When no effect is reported by the investigation, the outcome is generally accepted with no or minimal criticism. In this respect, two reports, one by Ziskin 44 and the other by the Canadian Health Directorate, 107 are useful to consider. The Ziskin report 44 was about an international survey of clinical users and entailed about 25,000 obstetric examinations with a notation that there were no reported bad outcomes in terms of the obstetric diagnostic ultrasound exposure by any of the 68 respondents in more than 121,000 examinations. However, because the background rate for major deleterious birth defects is somewhere between 3% and 5%, there should have been (minimally) 750 major anomalous fetuses in the Ziskin report 44 on the basis of the background rate alone. None was reported. Admittedly, it was not the goal of the survey to determine the background rate of anomalies in the general population, but the report can be misread from that standpoint. The only way one could reasonably sort out whether there is an effect would be to compare what was observed with what was expected. Similarly, the Canadian Health Directorate report 107 was also a survey and entailed 1,200,000 obstetric examinations among 111 respondents. It found no bad outcomes due to ultrasound exposure. In that report, there should have been at least 3000 major birth defects on the basis of the background rate of occurrence alone. Again, none was reported. Clearly, the background rate of anomalies must have been filtered out. This is what Bello 108 also expressed in an article asking whether harmful effects of ultrasound may have been missed because the wrong time frame reference is usually used. He describes two possible factors for such errors. Changes in the fetus are extremely rapid, and if one uses a term pregnancy (264 days [37 38 weeks])-to-life expectancy of 60 years (22680 days) ratio, then 7 in utero days are comparable to about 601 ex utero days. Therefore, a much shorter time interval (perhaps even 1 day) should be used to group fetuses to evaluate effects. Furthermore, Bello 108 also discusses a dilution error, similar to what was exposed above. Assuming an event has a background rate of 10% in the general population but 554 J Ultrasound Med 2008; 27:

15 Abramowicz et al occurs in 100% of fetuses exposed on day 30, if a large number (eg, 2000) of fetuses exposed on that particular day are examined, the incidence will be 100%, that is, 90% increase over the control population. However, if we assume 2000 fetuses exposed per day for 12 weeks, this represents 168,000 scans, and only 11.1% will be affected (all 2000 scanned on day 30 and 10% of all others, due to background incidence, or 18,800), an increase of only 1.1% over the background rate. A further confounding factor, which may decrease the number of reported malformations, is that when such a fetal malformation is diagnosed in utero, there is often an option for elective abortion. Elective abortions (of malformed fetuses) lower the birth defect rates because most hospitals do not include electively aborted malformed fetuses in their birth defects registries. In summary, performing research on the risk of fetal malformations posed by thermal effects of ultrasound scanning in early pregnancy is extremely difficult for the following reasons: 1. If ultrasound-induced heating produced localized pinpoint damage in the beam focal zone, it would not be detectable; 2. Numbers in clinical studies are insufficient for statistical purposes unless all data are merged into a meta-analysis; 3. There is an absence of sham-exposed controls (virtually every pregnant woman is exposed at least once); 4. Numbers in animal experiments to detect low-level (in contrast to gross developmental) effects are too small; and 5. Without a hypothesis derived from experimental animal work, it is almost impossible to know where to look, and any finding would be seen as questionable. There is, in addition, a whole ethics aspect (eg, withholding scanning to obtain a control group) that is beyond the scope of this article. Conclusions and Recommendations Conclusions 1. Acoustic output from diagnostic ultrasound devices is sufficient to cause temperature elevations in fetal tissue. In general, temperature elevations become progressively greater from B-mode to color Doppler to spectral Doppler applications. For identical exposure conditions, the temperature rise near bone increases with ossification development throughout gestation. Although, in general, an adverse fetal outcome is possible at any time during gestation, most severe and detectable effects of thermal exposure in animals have been observed during the period of organogenesis. For identical exposure conditions, the potential for thermal bioeffects increases with the dwell time during examination. 2. The TI, a quantity related to the calculated or estimated temperature rise under certain defined assumptions, correlates, albeit not perfectly, with temperature elevation. There are 2 TIs relevant for fetal exposure: the TI for soft tissues and the TIB. 3. Ultrasound exposures that elevate fetal temperature by 4 C above normal for 5 minutes or more have the potential to induce severe developmental defects. Thermally induced congenital anomalies have been observed in a large variety of animal species. In current clinical practice, using commercially available equipment, it is unlikely that such thermal exposure would occur at a specific fetal anatomic site. 4. The epidemiology database consists of exposures occurring before 1992, after which time diagnostic equipment capable of higher acoustic output became commercially available for fetal use. No congenital anomalies have been attributed to diagnostic ultrasound in humans. 5. Transducer self-heating is a significant component of the temperature rise of tissues close to the transducer. This may be of significance in transvaginal scanning, but no data for the fetal temperature rise are available. Recommendations 1. The TI should not be interpreted as an actual degrees Celsius temperature rise in the region of interest. Its use should be limited to a relative indication of the maximum temperature rise. J Ultrasound Med 2008; 27:

16 Fetal Thermal Effects of Diagnostic Ultrasound 2. Although the TI is not ideal, it should be used to assess the potential thermal risk in conjunction with the dwell time. 3. Given the fact that maternal fever early in pregnancy is known to be associated with fetal anomalies, one should be aware that ultrasound examination in a febrile patient might further increase the local temperature. 4. If it is ever clinically indicated in the first trimester, spectral Doppler examination of the fetus should be used with caution. 5. Given the present state and methods of epidemiologic research, detection of subtle effects, should they take place, is unlikely. To detect such effects, studies with sufficient statistical power to detect low-incidence occurrences would be desirable. Because of ethical, practical, and financial issues, such studies are hard to perform. Therefore, it is necessary to rely on laboratory experimentation for advancing our knowledge of ultrasonically induced bioeffects and for our understanding of the underlying mechanisms. 6. Output display indices represent important patient information and should be documented as part of the permanent record of the examination to enable future studies. References 1. Mellette HC, Hutt BK, Askovitz SI, Horvath SM. Diurnal variations in body temperatures. J Appl Physiol 1951; 3: Hardy JD. Physiology of temperature regulation. Physiol Rev 1961; 41: Mackowiak PA, Wasserman SS, Levine MM. A critical appraisal of 98.6 degrees F, the upper limit of the normal body temperature, and other legacies of Carl Reinhold August Wunderlich. JAMA 1992; 268: Asakura H. Fetal and neonatal thermoregulation. J Nippon Med Sch 2004; 71: Macaulay JH, Randall NR, Bond K, Steer PJ. Continuous monitoring of fetal temperature by noninvasive probe and its relationship to maternal temperature, fetal heart rate, and cord arterial oxygen and ph. Obstet Gynecol 1992; 79: National Council on Radiation Protection and Measurements. Exposure Criteria for Diagnostic Ultrasound, II: Criteria Based on All Known Mechanisms. Bethesda, MD: National Council on Radiation Protection and Measurements; Report Edwards MJ. Hyperthermia and birth defects [editorial]. Cornell Vet 1993; 83: Edwards MJ. Hyperthermia as a teratogen: a review of experimental studies and their clinical significance. Teratog Carcinog Mutagen 1986; 6: Shaw GM, Todoroff K, Velie EM, Lammer EJ. Maternal illness, including fever, and medication use as risk factors for neural tube defects. Teratology 1998; 57: Walsh DA, Klein NW, Hightower LE, Edwards MJ. Heat shock and thermotolerance during early rat embryo development. Teratology 1987; 36: Bosward KL, Barnett SB, Wood AKW, Edwards MJ, Kossoff G. Heating of the guinea-pig fetal brain during exposure to pulsed ultrasound. Ultrasound Med Biol 1993; 19: Tarantal AF, Hendricx AG, O Brien WD. Evaluation of the bioeffects of prenatal ultrasound exposure in the cynomolgus macaque (Macaca fascicularis), III: developmental and hematologic studies. Teratology 1993; 47: Duggan PM, Liggins GC, Barnett SB. Ultrasonic heating of the brain of the fetal sheep in utero. Ultrasound Med Biol 1995; 21: Horder MH, Barnett SB, Vella GJ, Edwards MJ, Wood AKW. In vivo heating of the guinea-pig fetal brain by pulsed ultrasound and estimates of thermal index. Ultrasound Med Biol 1998; 24: Ziskin MC, Petitti DB. Epidemiology of human exposure to ultrasound: a critical review. Ultrasound Med Biol 1988; 14: Miller MW, Ziskin MC. Biological consequences of hyperthermia. Ultrasound Med Biol 1989; 15: Barnett SB, Rott HD, ter Haar GR, Ziskin MC, Maeda K. The sensitivity of biological tissue to ultrasound. Ultrasound Med Biol 1997; 23: World Federation for Ultrasound in Medicine and Biology (WFUMB). Barnett SB (ed) WFUMB Symposium on Safety of Ultrasound in Medicine: Conclusions and Recommendations on Thermal and Non-Thermal Mechanisms for Biological Effects of Ultrasound. Ultrasound Med Biol 1998; 24(suppl I): Edwards MJ, Saunders RD, Shiota K. Effects of heat on embryos and fetuses. Int J Hyperthermia 2003; 19: Edwards MJ. Review: Hyperthermia and fever during pregnancy. Birth Defects Res A Clin Mol Teratol 2006; 76: Miller MW, Brayman AA, Abramowicz JS. Obstetric ultrasound: a biophysical consideration of patient safety. The rules have changed. Am J Obstet Gynecol 1998; 179: International Standards Organization. Standard 14971: Medical Devices Application of Risk Management to Medical Devices. Geneva, Switzerland: International Standards Organization; J Ultrasound Med 2008; 27:

17 Abramowicz et al 23. Abraham V, Ziskin MC, Heyner S. Temperature elevation in the rat fetus due to ultrasound exposure. Ultrasound Med Biol 1989; 15: National Council on Radiation Protection and Measurements. Exposure Criteria for Diagnostic Ultrasound, Part I: Criteria Based on Thermal Mechanisms. Bethesda, MD: National Council on Radiation Protection and Measurements; Report Duck FA, Starritt HC. A study of the heating capabilities of diagnostic ultrasound. Ultrasound Med Biol 1994; 20: Nyborg WL. Safety of medical diagnostic ultrasound. Semin Ultrasound CT MR 2002; 23: Chance PF, Smith DW. Hyperthermia and meningomyelocele and anencephaly. Lancet 1978; 1: Layde PM, Edmonds LD, Erickson JD. Maternal fever and neural tube defects. Teratology 1980; 21: Milunsky A, Ulcickas M, Rothman KJ, Willett W, Jick SS, Juck H. Maternal heat exposure and neural tube defects. JAMA 1992; 268: Lynberg MC, Khoury MJ, Lu X, Cocian T. Maternal flu, fever, and risk of neural tube defects: a population-based case-control study. Am J Epidemiol 1994; 140: Werler MM, Louik C, Mitchell AA. Heat source exposure and neural tube defect risk. Am J Epidemiol 1995; 142: Miller MW, Nyborg WL, Dewey WC, Edwards MJ, Abramowicz JS, Brayman AA. Diagnostic ultrasound during pregnancy; effects of new standards on tissue heating and their relevance of hyperthermia teratogenicity. Int J Hyperthermia 2002; 18: Nyborg WL, Steele RB. Temperature elevation in a beam of ultrasound. Ultrasound Med Biol 1983; 9: Nyborg WL, O Brien WD Jr. An alternative simple formula for temperature estimates. J Ultrasound Med 1989; 8: Horder MH, Barnett SB, Vella GJ, Edwards MJ, Wood AKW. Ultrasound-induced temperature increase in guinea-pig fetal brain in utero: third-trimester gestation. Ultrasound Med Biol 1998; 24: Vella GJ, Humphrey VF, Duck FA, Barnett SB. Ultrasoundinduced heating in a foetal skull bone phantom and its dependence on beam width and perfusion. Ultrasound Med Biol 2003; 29: Jauniaux F, Guldis B, Burton GJ. The human first trimester gestational sac limits rather than facilitates oxygen transfer to the foetus: a review. Placenta 2003; 24(suppl A):S86 S Herman BA, Harris GR. Models and regulatory considerations for transient temperature rise during diagnostic ultrasound pulses. Ultrasound Med Biol 2002; 28: US Food and Drug Administration. Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers. Rockville, MD: US Food and Drug Administration, Center for Devices and Radiological Health; Makikallio K, Tekay A, Joupila P. Yolk sac and umbilicoplacental hemodynamics during early human embryonic development. Ultrasound Obstet Gynecol 1999; 14: Kingdom JC, Kaufmann P. Oxygen and placental vascular development. Adv Exp Med Biol 1999; 474: Burton GJ, Jauniaux E. Maternal vascularisation of the human placenta: does the embryo develop in a hypoxic environment? Gynecol Obstet Fertil 2001; 29: Jauniaux F, Johns J, Burton GJ. The role of ultrasound imaging in diagnosing and investigating early pregnancy failure. Ultrasound Obstet Gynecol 2005; 25: Ziskin MC. Intrauterine effects of ultrasound: human epidemiology. Teratology 1999; 59: Doody C, Porter H, Duck FA, Humphrey VF. In vitro heating of human fetal vertebra by pulsed diagnostic ultrasound, Ultrasound Med Biol 1999; 25: Shaw A, Pay NM, Preston RC, Bond AD. Proposed standard thermal test object for medical ultrasound. Ultrasound Med Biol 1999; 25: Atkins TJ, Duck FA. Heating caused by selected pulsed Doppler and physiotherapy ultrasound beams measured using thermal test objects. Eur J Ultrasound 2003; 16: Bly SH, Vlakovich S, Mabee PR, Hussey RG. Computed estimates of maximum temperature elevations in fetal tissues during transabdominal pulsed Doppler examinations. Ultrasound Med Biol 1992; 18: Cahill MD, Humphrey VF, Doody C, Duck FA. The effect of nonlinear propagation on heating of tissue: numerical modelling and experimental measurement. In: Proceedings of the IEEE Ultrasonics Symposium, Munich, Germany. Piscataway, NJ: Institute of Electrical and Electronics Engineers; 2002: Shaw A, Pay NM, Preston RC. Assessment of the Likely Thermal Index Values for Pulsed Doppler Ultrasonic Equipment Stages 2 and 3: Experimental Assessment of Scanner/Transducer Combinations. London, England: National Physical Laboratory; Report CMAM Sheiner E, Freeman J, Abramowicz JS. Acoustic output as measured by mechanical and thermal indices during routine obstetric ultrasound examinations. J Ultrasound Med 2005; 24: Whittingham TA, The acoustic output of diagnostic machines. In: ter Haar G, Duck FA (eds). The Safe Use of Ultrasound in Medical Diagnosis. London, England: British Medical Ultrasound Society/British Institute of Radiology; 2000: World Federation for Ultrasound in Medicine and Biology (WFUMB). Barnett SB, Kossoff G (eds). WFUMB Symposium on Safety and Standardization in Medical Ultrasound: Issues and Recommendations Regarding Thermal Mechanisms for Biological Effects of Ultrasound. Ultrasound Med Biol 1992; 18(special issue): J Ultrasound Med 2008; 27:

18 Fetal Thermal Effects of Diagnostic Ultrasound 54. Horder MM, Barnett SB, Vella GJ, Edwards MJ. Ultrasoundinduced temperature increase in the guinea-pig fetal brain in vitro. Ultrasound Med Biol 1998; 24: Barnett SB, Vella GJ, Horder MM. Validation of the thermal index for ultrasound-induced heating of the fetal brain. In: Proceedings of the 43rd Annual AIUM Convention. J Ultrasound Med 1999; 18(suppl):S American Institute of Ultrasound in Medicine, National Electrical Manufacturers Association. Standard for Real- Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment. Laurel, MD: American Institute of Ultrasound in Medicine; Rosslyn, VA: National Electrical Manufacturers Association; Duck FA, Starritt HC, Aindow JD, Perkins MA, Hawkins AJ. The output of pulse-echo ultrasound equipment: a survey of powers, pressures and intensities. Br J Radiol 1985; 58: Jago J, Henderson J, Whittingham TA, Willson K. How reliable are manufacturer s reported acoustic output data? Ultrasound Med Biol 1995; 21: Jago JR, Henderson J, Whittingham TA, Mitchell G. A comparison of AIUM/NEMA thermal indices with calculated temperature rises for a single third-trimester pregnancy tissue model. Ultrasound Med Biol 1999; 4: Carstensen EL, Child SZ, Norton S, Nyborg WL. Ultrasonic heating of the skull. J Acoust Soc Am 1990; 87: Horder MM, Barnett SB, Edwards MJ. Effects of pulsed ultrasound on sphenoid bone temperature and the heart rate in guinea-pig foetuses. Early Hum Dev 1998; 52: Duggan PM, Murcott MF, McPhee AJ, Barnett SB. The influence of variations in blood flow on pulsed Doppler ultrasonic heating of the cerebral cortex of the neonatal pig. Ultrasound Med Biol 2000; 26: Duggan PM, Liggins GC, Barnett SB. Pulsed ultrasound and electrocortical activity in fetal sheep. Early Hum Dev 1993; 35: Abramowicz JS, Fowlkes JB, Skelly AC, Stratmeyer ME, Ziskin MC. Conclusions regarding epidemiology for obstetric ultrasound. J Ultrasound Med 2008; 27: Palmer AK. Use of mammalian models in teratology. Prog Clin Biol Res 1985; 163A: Germain MA, Webster WS, Edwards MJ. Hyperthermia as a teratogen: parameters determining hyperthermiainduced head defects in the rat. Teratology 1985; 31: Kimmel CA, Cuff JM, Kimmel GL, et al. Skeletal development following heat exposure in the rat. Teratology 1993; 47: International Non-Ionizing Radiation Committee of the International Radiation Protection Association. Protection of the patient undergoing a magnetic resonance examination. Health Phys 1991; 61: Carstensen EL, Gates AH. The effects of pulsed ultrasound on the fetus. J Ultrasound Med 1984; 3: Sapareto SA, Dewey WC. Thermal dose determination in cancer therapy. Int J Radiat Oncol Biol Phys 1984; 10: Dewey WC. Arrhenius relationships from the molecule and cell to the clinic. Int J Hyperthermia 1994; 10: Miller MW, Miller RK, Battaglia LF, et al. the delta-t thermal dose concept, 1: in vivo teratogenesis. J Therm Biol 2004; 29; Rodier PM. Critical periods for morphological assessment. Congenit Anom (Kyoto) 1992; 32:S55 S Edmonds PD, Abramowicz JS, Carson PL, Carstensen EL, Sandstrom KL. Guidelines for Journal of Ultrasound in Medicine authors and reviewers on measurements and reporting of acoustic output and exposure. J Ultrasound Med 2005; 24: Abramowicz JS. Ultrasound contrast media in obstetrics and gynecology: is it time yet? J Ultrasound Med 2005; 24: Duck FA, Henderson J. Acoustic output of modern ultrasound equipment: is it increasing? In: Barnett SB, Kossoff G (eds). Safety of Diagnostic Ultrasound. New York, NY: Parthenon Publishing Group; 1998: Kremkau FW, Merritt CR, Carson PL, et al. The American Institute of Ultrasound in Medicine and Society of Radiologists in Ultrasound: Future directions in diagnostic US. Radiology 1998; 209: O Brien WD, Miller DL. Diagnostic ultrasound should be performed without upper intensity limits [point-counterpoint]. Med Phys 2001; 28: Newnham JP, Evans SF, Michael CA. Effects of frequent ultrasound during pregnancy: a randomized controlled trial. Lancet 1993; 342: Campbell JD, Elford RW, Brant RF. Case-control study of prenatal ultrasonography exposure in children with delayed speech. Can Med Assoc J 1993; 149: Stark CR, Orleans M, Haverkamp AD, Murphy J. Short- and long-term risks after exposure to diagnostic ultrasound in utero. Obstet Gynecol 1984; 63: Kieler H, Axelsson O, Haglund B, Nilsson S, Salvesen KA. Routine ultrasound screening in pregnancy and the children s subsequent handedness. Early Hum Dev 1998; 50: Kieler H, Cnattingus S, Halund B, et al. Sinistrality: a sideeffect of prenatal sonography a comparative study of young men. Epidemiology 2001; 12: Tarantal AF, Hendricx AG. Evaluation of the bioeffects of prenatal ultrasound exposure in the cynomolgus macaque (Macaca fascicularis), II: growth and behavior during the first year. Teratology 1989; 39: Newnham JP, Doherty DA, Kendall GE, Zubrick SR, Landau LL, Stanley FJ. Effects of repeated prenatal ultrasound examinations on childhood outcome up to 8 years of age: 558 J Ultrasound Med 2008; 27:

19 Abramowicz et al follow-up of a randomised controlled trial. Lancet 2004; 364: D Andrea, JA, de Lorge, JO. Behavioral effects of electromagnetic fields. In: Gandhi OP (ed). Biological Effects and Medical Applications of Electromagnetic Energy. Englewood Cliffs, NJ: Prentice-Hall 1990: D Andrea JA Thomas A, Hatcher DJ. Rhesus monkey behavior during exposure to high-peak-power 5.62-GHz microwave pulses. Bioelectromagnetics 1994; 15: de Lorge JO. Operant behavior and colonic temperature of Macaca mulatta exposed to radio frequency fields at and above resonant frequencies. Bioelectromagnetics 1984; 5: Abart J, Brinkler G, Irlbacher W, Grebmeier J. Temperature and heart rate changes in MRI at SAR levels of up to 3 W/kg. In: Abstracts of the International Society for Magnetic Resonance in Medicine. Berkeley, CA: International Society for Magnetic Resonance in Medicine; 1989: National Institute for Occupational Safety and Health. Criteria for a Recommended Standard for Occupational Exposure to Hot Environments. Rockville, Maryland, US Department of Health, Education, and Welfare; Report NIOSH-TR American Conference of Governmental Industrial Hygienists. Threshold Limit Values for Chemical Substances and Physical Agents in the Work Environment With Intended Changes for Cincinnati, OH: American Conference of Governmental Industrial Hygienists; World Health Organization. Radiofrequency and Microwaves: World Health Organization Environmental Health Criteria 16. Geneva, Switzerland: World Health Organization; Barnett SB, Walsh DA, Angles JA. Novel approach to evaluate the interaction of pulsed ultrasound with embryonic development. Ultrasonics 1990; 28: Angles JM, Walsh DA, Li K, Barnett SB, Edwards MJ. Effects of pulsed ultrasound and temperature on the development of rat embryos in culture. Teratology 1990; 42: Testart J, Thebault A, Souderes E, Frydman R. Premature ovulation after ovarian ultrasonography. Br J Obstet Gynecol 1982; 89: Iannaccone PM, Van Gorder M, Madsen EL, Martin AO, Garton R. The role of preimplantation sonographic exposure. J Ultrasound Med 1991; 10: Tarantal AF, O Brien WD. Discussion of ultrasonic safety related to obstetrics. In: Sabbagha RE (ed). Diagnostic Ultrasound Applied to Obstetrics and Gynecology. 3rd ed. Philadelphia, PA: JB Lippincott Co; 1994: Sikov MR. Effect of ultrasound on development, part II: studies in mammalian species and overview. J Ultrasound Med 1986; 5: Jensh RP, Lewis PA, Poczobutt MT, et al. Effects of prenatal ultrasound exposure on adult offspring behavior in the Wistar rat. Proc Soc Exp Biol Med 1995; 210: National Institutes of Health. Consensus Development Conference Statement: Diagnostic Ultrasound Imaging in Pregnancy. Vol 5, No 1. Washington, DC: Government Printing Office; National Institutes of Health publication Carstensen EL, Dalecki D, Gracewski SM, Christopher T. Nonlinear propagation and the output indices. J Ultrasound Med 1999; 18: Christopher T. Computing the mechanical index. J Ultrasound Med 1999; 18: Duck FA. Estimating in situ exposure in the presence of acoustic nonlinerarity. J Ultrasound Med 1999; 18: Abramowicz JS. Ultrasound in obstetrics and gynecology. Is this hot technology too hot [editorial]? J Ultrasound Med 2002; 21: Horder MM, Barnett SB, Edwards MJ. Diagnostic ultrasound in veterinary practice: how safe is it? Aust Vet J 1996; 73: Ramnarine KV, Nassiri KD, Pearce JM, Joseph AEA, Patel RH, Varma TR. Estimation of in situ ultrasound exposure during obstetric examinations. Ultrasound Med Biol 1993; 19: Environmental Health Directorate. Canada-Wide Survey of Nonionizing Radiation Emitting Medical Devices, Part II: Ultrasound Devices. Ottawa, Ontario, Canada: Environmental Health Directorate, Health Protection Branch; Report 80-EHD Bello SO. How we may be missing some harmful effects of ultrasound: a hypothesis. Med Hypotheses 2006; 67: J Ultrasound Med 2008; 27:

BIOMEDICAL ULTRASOUND

BIOMEDICAL ULTRASOUND BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic

More information

History of the American Institute of Ultrasound in Medicine s Efforts to Keep Ultrasound Safe

History of the American Institute of Ultrasound in Medicine s Efforts to Keep Ultrasound Safe Commemoration History of the American Institute of Ultrasound in Medicine s Efforts to Keep Ultrasound Safe Wesley L. Nyborg, PhD Physics Department University of Vermont Burlington, Vermont USA Abbreviations

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality

Hunting Bats. Diagnostic Ultrasound. Ultrasound Real-time modality Diagnostik Ultrasound Basic physics, image reconstruction and signal processing Per Åke Olofsson Dpt of Biomedical Engineering, Malmö University Hospital, Sweden Ultrasound Real-time modality 17-WEEK FETAL

More information

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim- QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound

More information

Doppler Ultrasound in the Management of Fetal Growth Restriction Chukwuma I. Onyeije, M.D. Atlanta Perinatal Associates

Doppler Ultrasound in the Management of Fetal Growth Restriction Chukwuma I. Onyeije, M.D. Atlanta Perinatal Associates Doppler Ultrasound in the Management of Fetal Growth Restriction Chukwuma I. Onyeije, M.D. Atlanta Perinatal Associates 1 For your convenience a copy of this lecture is available for review and download

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

Rate Equations and Detailed Balance

Rate Equations and Detailed Balance Rate Equations and Detailed Balance Initial question: Last time we mentioned astrophysical masers. Why can they exist spontaneously? Could there be astrophysical lasers, i.e., ones that emit in the optical?

More information

Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University

Principles of Medical Ultrasound. Pai-Chi Li Department of Electrical Engineering National Taiwan University Principles of Medical Ultrasound Pai-Chi Li Department of Electrical Engineering National Taiwan University What is Medical Ultrasound? Prevention: actions taken to avoid diseases. Diagnosis: the process

More information

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen.

An abdominal ultrasound produces a picture of the organs and other structures in the upper abdomen. Scan for mobile link. Ultrasound - Abdomen Ultrasound imaging of the abdomen uses sound waves to produce pictures of the structures within the upper abdomen. It is used to help diagnose pain or distention

More information

Thermal Mass Availability for Cooling Data Centers during Power Shutdown

Thermal Mass Availability for Cooling Data Centers during Power Shutdown 2010 American Society of Heating, Refrigerating and Air-Conditioning Engineers, Inc. (www.ashrae.org). Published in ASHRAE Transactions (2010, vol 116, part 2). For personal use only. Additional reproduction,

More information

FXA 2008. UNIT G485 Module 4 5.4.3 Ultrasound. Candidates should be able to :

FXA 2008. UNIT G485 Module 4 5.4.3 Ultrasound. Candidates should be able to : 1 Candidates should be able to : ULTRASOUND Describe the properties of ultrasound. ULTRASOUND is any sound wave having a frequency greater than the upper frequency limit of human hearing (20 khz). Describe

More information

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects. RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer

More information

EVALUATION OF GEOTHERMAL ENERGY AS HEAT SOURCE OF DISTRICT HEATING SYSTEMS IN TIANJIN, CHINA

EVALUATION OF GEOTHERMAL ENERGY AS HEAT SOURCE OF DISTRICT HEATING SYSTEMS IN TIANJIN, CHINA EVALUATION OF GEOTHERMAL ENERGY AS HEAT SOURCE OF DISTRICT HEATING SYSTEMS IN TIANJIN, CHINA Jingyu Zhang, Xiaoti Jiang, Jun Zhou, and Jiangxiong Song Tianjin University, North China Municipal Engineering

More information

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode

THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode THE IMPOSSIBLE DOSE HOW CAN SOMETHING SIMPLE BE SO COMPLEX? Lars Hode Swedish Laser-Medical Society The dose is the most important parameter in laser phototherapy. At a first glance, the dose seem very

More information

Comments on the Draft Report by the California Council on Science and Technology Health Impacts of Radio Frequency from Smart Meters

Comments on the Draft Report by the California Council on Science and Technology Health Impacts of Radio Frequency from Smart Meters Comments on the Draft Report by the California Council on Science and Technology Health Impacts of Radio Frequency from Smart Meters by Daniel Hirsch 1 31 January 2011 Abstract The draft report by the

More information

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION Richard A. Sweet, M.D. Louisville Orthopaedic Clinic Louisville, KY TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION INTRODUCTION Total hip replacement surgery (THR) has

More information

ELECTROMAGNETIC FIELDS AND PUBLIC HEALTH HEALTH AND SAFETY GUIDELINES #1

ELECTROMAGNETIC FIELDS AND PUBLIC HEALTH HEALTH AND SAFETY GUIDELINES #1 SINGAPORE, 16 AUGUST 2001 HEALTH SCIENCES AUTHORITY PRESS RELEASE ELECTROMAGNETIC FIELDS AND PUBLIC HEALTH HEALTH AND SAFETY GUIDELINES #1 THE EMF HEALTH ISSUE 1 Over the past years a large number of scientific

More information

Guidance For Research Involving Human Embryonic Stem Cells, Germ Cells, And Cells Obtained From Cord Blood

Guidance For Research Involving Human Embryonic Stem Cells, Germ Cells, And Cells Obtained From Cord Blood Guidance For Research Involving Human Embryonic Stem Cells, Germ Cells, And Cells Obtained From Cord Blood Supreme Council of Health Department of Research Guidance Regarding Research Involving Human Embryonic

More information

Ultrasound Dose Calculations

Ultrasound Dose Calculations Ultrasound Dose Calculations One of the advantages of ultrasound therapy remains the reasonably broad range of trails from which effective treatment doses can be established. In principle, there is no

More information

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole

Applications of Doppler Ultrasound in Fetal Growth Assessment. David Cole Applications of Doppler Ultrasound in Fetal Growth Assessment David Cole Aims The aim of this presentation is to consider the use of Doppler ultrasound to investigate and monitor those pregnancies at risk

More information

Chapter 4 Physiological Therapeutics. 3 Therapeutic Ultrasound

Chapter 4 Physiological Therapeutics. 3 Therapeutic Ultrasound Chapter 4 Physiological Therapeutics 3 Therapeutic Ultrasound Therapeutic Ultrasound PHYSIOLOGIC EFFECTS OF ULTRASOUND 1. Increased collagen extensibility 5. Increased nerve conduction velocities 2. Increased

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

What the experts say: The consensus of scientific opinion

What the experts say: The consensus of scientific opinion What the experts say: The consensus of scientific opinion AMTA relies on the expert judgment of independent public health authorities, such as the World Health Organization, for assessments of safety and

More information

Science 10-Biology Activity 15 The Development of the Human Embryo

Science 10-Biology Activity 15 The Development of the Human Embryo Science 10-Biology Activity 15 The Development of the Human Embryo 10 Name Due Date Show Me Hand In Correct and Hand In Again By NOTE: This worksheet is based on material from pages 379-380 in Science

More information

How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless)

How To Know If You Are Safe To Use An Antenna (Wired) Or Wireless (Wireless) 1 2 The range of RF spans 3 KHz (3000 Hz) to 300 GHz (300 million Hz) Frequencies of RF devices range from the low frequency AM broadcasts (80 MHz) to higher frequency mobile phones (1900 MHz) smart meters

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology

The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology The Effects of Ultrasonic Sound Generated by Ultrasonic Cleaning Systems on Human Hearing and Physiology Questions about the effects of ultrasonic energy on hearing and other human physiology arise from

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

How to calibrate an RTD or Platinum Resistance Thermometer (PRT) How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of

More information

2. Incidence, prevalence and duration of breastfeeding

2. Incidence, prevalence and duration of breastfeeding 2. Incidence, prevalence and duration of breastfeeding Key Findings Mothers in the UK are breastfeeding their babies for longer with one in three mothers still breastfeeding at six months in 2010 compared

More information

Statistical estimation using confidence intervals

Statistical estimation using confidence intervals 0894PP_ch06 15/3/02 11:02 am Page 135 6 Statistical estimation using confidence intervals In Chapter 2, the concept of the central nature and variability of data and the methods by which these two phenomena

More information

High-quality/High-resolution Digital Ultrasound Diagnostic Scanner

High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Hitachi Review Vol. 52 (2003), No. 4 207 High-quality/High-resolution Digital Ultrasound Diagnostic Scanner Tsuyoshi Otake Toshihiko Kawano Takashi Sugiyama Tsuyoshi Mitake Shinichiro Umemura, Ph. D. OVERVIEW:

More information

Statement of the Chief Medical Health Officer

Statement of the Chief Medical Health Officer Statement of the Chief Medical Health Officer June, 2011 Health Concerns About Cellular Phone Transmission Antennae and Base Stations In 2005, in response to community concerns and after reviewing the

More information

Scanning Acoustic Microscopy Training

Scanning Acoustic Microscopy Training Scanning Acoustic Microscopy Training This presentation and images are copyrighted by Sonix, Inc. They may not be copied, reproduced, modified, published, uploaded, posted, transmitted, or distributed

More information

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS Selection and use of Ultrasonic Ceramic Transducers The purpose of this application note is to aid the user in the selection and application of the Ultrasonic

More information

Patient Handbook on Stem Cell Therapies

Patient Handbook on Stem Cell Therapies Patient Handbook on Stem Cell Therapies Appendix I of the Guidelines for the Clinical Translation of Stem Cells www.isscr.org 2008, International Society for Stem Cell Research 2 Introduction We have all

More information

Why is prematurity a concern?

Why is prematurity a concern? Prematurity What is prematurity? A baby born before 37 weeks of pregnancy is considered premature. Approximately 12% of all babies are born prematurely. Terms that refer to premature babies are preterm

More information

Medical Ultrasound Safety

Medical Ultrasound Safety Medical Ultrasound Safety Second Edition www.aium.org Medical Ultrasound Safety Second Edition Part 1: Bioeffects and Biophysics Part 2: Prudent Use Part 3: Implementing ALARA www.aium.org 2009 by the

More information

Conductivity Sensor Calibrations to Meet Water Industry Requirements

Conductivity Sensor Calibrations to Meet Water Industry Requirements Conductivity Sensor Calibrations to Meet Water Industry Requirements Victor M. Braga Technical Service/Training Manager Mettler-Toledo Thornton Inc. 36 Middlesex Turnpike Bedford, Massachusetts 01730 Phone

More information

Human Exposure Limits

Human Exposure Limits Human Exposure Limits Session 3 0 Version December 2014 Learning objectives In this session we will: Learn about the international exposure limits for workers and the public Learn about methods for assessing

More information

Temperature Increase in the. Human Eye When Subjected to a. Laser Source

Temperature Increase in the. Human Eye When Subjected to a. Laser Source Validation of a Model to Predict Temperature Increase in the Human Eye When Subjected to a Laser Source Corinna Sue Thompson, Antonio Campo, PhD., University of Vermont, Undergraduate Research Endeavors

More information

Artificial insemination with donor sperm

Artificial insemination with donor sperm Artificial insemination with donor sperm Ref. 123 / 2009 Reproductive Medicine Unit Servicio de Medicina de la Reproducción Gran Vía Carlos III 71-75 08028 Barcelona Tel. (+34) 93 227 47 00 Fax. (+34)

More information

Antonio Rampoldi Struttura Complessa di Radiologia Interventistica Azienda Ospedale Niguarda Ca Granda, Milano. Terminology

Antonio Rampoldi Struttura Complessa di Radiologia Interventistica Azienda Ospedale Niguarda Ca Granda, Milano. Terminology Biophysical principles and clinical applications of MRgFUS Alberto Torresin Struttura Complessa di Fisica Sanitaria Azienda, Milano Università degli Studi di Milano Dip. di Fisica Antonio Rampoldi Struttura

More information

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany ABSTRACT: PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany Today, in-line inspection tools are used routinely

More information

Power Dissipation Considerations in High Precision Vishay Sfernice Thin Film Chips Resistors and Arrays (P, PRA etc.) (High Temperature Applications)

Power Dissipation Considerations in High Precision Vishay Sfernice Thin Film Chips Resistors and Arrays (P, PRA etc.) (High Temperature Applications) VISHAY SFERNICE Resistive Products Application Note ABSTRACT On our thin film chips resistors and arrays the main path for the heat, more than 90 %, is conduction through the body of the component, the

More information

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power

Describing Sound Waves. Period. Frequency. Parameters used to completely characterize a sound wave. Chapter 3. Period Frequency Amplitude Power Parameters used to completely characterize a sound wave Describing Sound Waves Chapter 3 Period Frequency Amplitude Power Intensity Speed Wave Length Period Defined as the time it take one wave vibrate

More information

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,

More information

Lecture Notes Module 1

Lecture Notes Module 1 Lecture Notes Module 1 Study Populations A study population is a clearly defined collection of people, animals, plants, or objects. In psychological research, a study population usually consists of a specific

More information

Simplifying Oncology Care

Simplifying Oncology Care Simplifying Oncology Care March 2009 Research Philips Research focuses on ultrasound-mediated drug delivery Research Philips Research focuses on ultrasound-mediated drug delivery Philips Research and its

More information

Airways Resistance and Airflow through the Tracheobronchial Tree

Airways Resistance and Airflow through the Tracheobronchial Tree Airways Resistance and Airflow through the Tracheobronchial Tree Lecturer: Sally Osborne, Ph.D. Department of Cellular & Physiological Sciences Email: sosborne@interchange.ubc.ca Useful links: www.sallyosborne.com

More information

Wireless Broadband: Health & Safety Information

Wireless Broadband: Health & Safety Information Wireless Broadband: Health & Safety Information Introduction The increasing use of mobile phones and other wireless technology has been accompanied by public debate about possible adverse effects on health.

More information

II. DISTRIBUTIONS distribution normal distribution. standard scores

II. DISTRIBUTIONS distribution normal distribution. standard scores Appendix D Basic Measurement And Statistics The following information was developed by Steven Rothke, PhD, Department of Psychology, Rehabilitation Institute of Chicago (RIC) and expanded by Mary F. Schmidt,

More information

Effects of Cell Phone Radiation on the Head. BEE 4530 Computer-Aided Engineering: Applications to Biomedical Processes

Effects of Cell Phone Radiation on the Head. BEE 4530 Computer-Aided Engineering: Applications to Biomedical Processes Effects of Cell Phone Radiation on the Head BEE 4530 Computer-Aided Engineering: Applications to Biomedical Processes Group 3 Angela Cai Youjin Cho Mytien Nguyen Praveen Polamraju Table of Contents I.

More information

NONCLINICAL EVALUATION FOR ANTICANCER PHARMACEUTICALS

NONCLINICAL EVALUATION FOR ANTICANCER PHARMACEUTICALS INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE NONCLINICAL EVALUATION FOR ANTICANCER PHARMACEUTICALS

More information

Date Issued: Page No.: Document No.: August 2002 1 SOP-Safety-01. 3.0 Working in Hot Environments

Date Issued: Page No.: Document No.: August 2002 1 SOP-Safety-01. 3.0 Working in Hot Environments August 2002 1 SOP-Safety-01 1. Introduction: The Environmental Health & Safety Standard Operating Procedure for working in hot environments was developed by the Department of Environmental Health & Safety

More information

Cell phones and brain tumors: A review including the long-term epidemiologic data. Surgical Neurology September 2009; 72; pp.

Cell phones and brain tumors: A review including the long-term epidemiologic data. Surgical Neurology September 2009; 72; pp. Cell phones and brain tumors: A review including the long-term epidemiologic data 1 Surgical Neurology September 2009; 72; pp. 205 215 Vini G. Khurana, PhD, FRACS, Charles Teo, MBBS, FRACS, Michael Kundi,

More information

Fact Sheet for Health Care Providers: Interpreting Results from the Aptima Zika Virus Assay. June 17, 2016

Fact Sheet for Health Care Providers: Interpreting Results from the Aptima Zika Virus Assay. June 17, 2016 Dear Health Care Provider: Fact Sheet for Health Care Providers: Interpreting Results from the Aptima Zika Virus Assay June 17, 2016 The U.S. Food and Drug Administration (FDA) has issued an Emergency

More information

99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, 99.42 cm

99.37, 99.38, 99.38, 99.39, 99.39, 99.39, 99.39, 99.40, 99.41, 99.42 cm Error Analysis and the Gaussian Distribution In experimental science theory lives or dies based on the results of experimental evidence and thus the analysis of this evidence is a critical part of the

More information

X-Rays Benefits and Risks. Techniques that use x-rays

X-Rays Benefits and Risks. Techniques that use x-rays X-Rays Benefits and Risks X-rays are a form of electromagnetic radiation, just like light waves and radiowaves. Because X-rays have higher energy than light waves, they can pass through the body. X-rays

More information

Neural Tube Defects - NTDs

Neural Tube Defects - NTDs Neural Tube Defects - NTDs Introduction Neural tube defects are also known as NTDs. They happen when the spine and brain do not fully develop while the fetus is forming in the uterus. Worldwide, there

More information

Charts of fetal size: limb bones

Charts of fetal size: limb bones BJOG: an International Journal of Obstetrics and Gynaecology August 2002, Vol. 109, pp. 919 929 Charts of fetal size: limb bones Lyn S. Chitty a, *, Douglas G. Altman b Objective To construct new size

More information

North Bergen School District Benchmarks

North Bergen School District Benchmarks Grade: 10,11, and 12 Subject: Anatomy and Physiology First Marking Period Define anatomy and physiology, and describe various subspecialties of each discipline. Describe the five basic functions of living

More information

Epidemiology 521. Epidemiology of Maternal and Child Health Problems. Winter / Spring, 2010

Epidemiology 521. Epidemiology of Maternal and Child Health Problems. Winter / Spring, 2010 Extended MPH Degree Program School of Public Health Department of Epidemiology University of Washington Epidemiology 521 Epidemiology of Maternal and Child Health Problems Winter / Spring, 2010 Instructor:

More information

Sample Size and Power in Clinical Trials

Sample Size and Power in Clinical Trials Sample Size and Power in Clinical Trials Version 1.0 May 011 1. Power of a Test. Factors affecting Power 3. Required Sample Size RELATED ISSUES 1. Effect Size. Test Statistics 3. Variation 4. Significance

More information

Zhao et al. 2.2 Experimental Results in Winter Season The analysis given below was based on the data collected from Nov. 2003 to Mar. 15, 2004.

Zhao et al. 2.2 Experimental Results in Winter Season The analysis given below was based on the data collected from Nov. 2003 to Mar. 15, 2004. Proceedings World Geothermal Congress 2005 Antalya, Turkey, 24-29 April 2005 A Case Study of Ground Source Heat Pump System in China Jun Zhao, Chuanshan Dai, Xinguo Li, Qiang Zhu and Lixin Li College of

More information

Dissolved Gas Analysis Guide for Transformers Filled with Beta Fluid

Dissolved Gas Analysis Guide for Transformers Filled with Beta Fluid DSI Ventures, Inc. PHONE: (903) 526-7577 FAX: (903) 526-0021 www.dsiventures.com Dissolved Gas Analysis Guide for Transformers Filled with Beta Fluid Introduction Analysis of dissolved gases in transformer

More information

EFFECTS OF ELECTROMAGNETIC RADIATION FROM MOBILE PHONES AND CASE STUDIES OF COMPUTATIONAL MODELS FOR CALCULATING SAR DISTRIBUTION IN HUMAN BODY Settapong Malisuwan National Broadcasting and Telecommunications

More information

Animal Systems: The Musculoskeletal System

Animal Systems: The Musculoskeletal System Animal Systems: The Musculoskeletal System Tissues, Organs, and Systems of Living Things Cells, Cell Division, and Animal Systems and Plant Systems Cell Specialization Human Systems The Digestive The Circulatory

More information

Genetics and Pregnancy Loss

Genetics and Pregnancy Loss Genetics and Pregnancy Loss Dorothy Warburton Genetics and Development (in Pediatrics) Columbia University, New York Estimates of Pregnancy Loss from Conception 1000 fertilized eggs (27% are lost) 728

More information

Reproductive System & Development: Practice Questions #1

Reproductive System & Development: Practice Questions #1 Reproductive System & Development: Practice Questions #1 1. Which two glands in the diagram produce gametes? A. glands A and B B. glands B and E C. glands C and F D. glands E and F 2. Base your answer

More information

In today s world of wireless communications

In today s world of wireless communications TUTORIAL RF CERAMIC CHIP CAPACITORS IN HIGH RF POWER APPLICATIONS Reprinted with permission of MICROWAVE JOURNAL from the April 2000 issue. 2000 Horizon House Publications, Inc. In today s world of wireless

More information

X-ray (Radiography) - Abdomen

X-ray (Radiography) - Abdomen Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,

More information

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications Janet E. Semmens Sonoscan, Inc. 2149 E. Pratt Boulevard Elk Grove Village, IL 60007 USA Phone: (847)

More information

Accurate Air Flow Measurement in Electronics Cooling

Accurate Air Flow Measurement in Electronics Cooling Accurate Air Flow Measurement in Electronics Cooling Joachim Preiss, Raouf Ismail Cambridge AccuSense, Inc. E-mail: info@degreec.com Air is the most commonly used medium to remove heat from electronics

More information

Copyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley.

Copyright 2008 Pearson Education, Inc., publishing as Pearson Addison-Wesley. Chapter 20. Traveling Waves You may not realize it, but you are surrounded by waves. The waviness of a water wave is readily apparent, from the ripples on a pond to ocean waves large enough to surf. It

More information

arxiv:astro-ph/0509450 v1 15 Sep 2005

arxiv:astro-ph/0509450 v1 15 Sep 2005 arxiv:astro-ph/0509450 v1 15 Sep 2005 TESTING THERMO-ACOUSTIC SOUND GENERATION IN WATER WITH PROTON AND LASER BEAMS K. GRAF, G. ANTON, J. HÖSSL, A. KAPPES, T. KARG, U. KATZ, R. LAHMANN, C. NAUMANN, K.

More information

Advanced Measurements of Microwave Oven Leakage

Advanced Measurements of Microwave Oven Leakage Conference paper Advanced Measurements of Microwave Oven Leakage M. Bangay [1] C. Zombolas [2] 1. Australian Radiation Protection and Nuclear Safety Agency 2. EMC Technologies Pty Ltd Abstract Reports

More information

Birth Defects Monitoring in Japan -Possible Effects of Environmental Endocrine Disrupters-

Birth Defects Monitoring in Japan -Possible Effects of Environmental Endocrine Disrupters- Birth Defects Monitoring in Japan -Possible Effects of Environmental Endocrine Disrupters- Fumiki Hirahara Yokohama City University School of Medicine Thank you, Paul. It is my great pleasure to be here

More information

2007 WHO Research Agenda for Extremely Low Frequency Fields

2007 WHO Research Agenda for Extremely Low Frequency Fields 2007 WHO Research Agenda for Extremely Low Frequency Fields Introduction In 1997, the WHO International EMF Project developed a Research Agenda in order to facilitate and coordinate research worldwide

More information

Certified in Public Health (CPH) Exam CONTENT OUTLINE

Certified in Public Health (CPH) Exam CONTENT OUTLINE NATIONAL BOARD OF PUBLIC HEALTH EXAMINERS Certified in Public Health (CPH) Exam CONTENT OUTLINE April 2014 INTRODUCTION This document was prepared by the National Board of Public Health Examiners for the

More information

Human Skin Temperature Response to Absorbed Thermal Power

Human Skin Temperature Response to Absorbed Thermal Power Human Skin Temperature Response to Absorbed Thermal Power Jim M. Griffith 1, Ardella Hamilton, George Long, Amir Mujezinovic, DeWayne Warren, and Kamal Vij W. L. Gore & Associates, Inc., Phoenix, AZ 85044

More information

Evaluation and Follow-up of Fetal Hydronephrosis

Evaluation and Follow-up of Fetal Hydronephrosis Evaluation and Follow-up of Fetal Hydronephrosis Deborah M. Feldman, MD, Marvalyn DeCambre, MD, Erin Kong, Adam Borgida, MD, Mujgan Jamil, MBBS, Patrick McKenna, MD, James F. X. Egan, MD Objective. To

More information

Orphan drugs: rising to the challenge to ensure a better future for 30 million patients in Europe

Orphan drugs: rising to the challenge to ensure a better future for 30 million patients in Europe www.eurordis.org S T A T E M E N T Orphan drugs: rising to the challenge to ensure a better future for 30 million patients in Europe October 2009 Rare Diseases Europe Summary Nine years after the implementation

More information

Principle of Thermal Imaging

Principle of Thermal Imaging Section 8 All materials, which are above 0 degrees Kelvin (-273 degrees C), emit infrared energy. The infrared energy emitted from the measured object is converted into an electrical signal by the imaging

More information

Imaging Markers of Brain Network Dysfunction in Multiple Sclerosis

Imaging Markers of Brain Network Dysfunction in Multiple Sclerosis Faculty of Medicine & Health Sciences School of Medicine Radiological Sciences Research Group The University of Nottinham University Park Nottingham NG7 2RD t: +44 (0)115 823 0018 f: +44 (0)115 823 0004

More information

What Alcohol Does to the Body. Chapter 25 Lesson 2

What Alcohol Does to the Body. Chapter 25 Lesson 2 What Alcohol Does to the Body Chapter 25 Lesson 2 Short-Term Effects of Drinking The short-term term effects of alcohol on the body depend on several factors including: amount of alcohol consumed, gender,

More information

CLINICAL TRIALS SHOULD YOU PARTICIPATE? by Gwen L. Nichols, MD

CLINICAL TRIALS SHOULD YOU PARTICIPATE? by Gwen L. Nichols, MD CLINICAL TRIALS SHOULD YOU PARTICIPATE? by Gwen L. Nichols, MD Gwen L. Nichols, M.D., is currently the Oncology Site Head of the Roche Translational Clinical Research Center at Hoffman- LaRoche. In this

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

15. Liquid-in-Glass Thermometry

15. Liquid-in-Glass Thermometry 123 15. Liquid-in-Glass Thermometry Even though the liquid-in-glass thermometer is used much less frequently today than formerly, it is still a very commonly used device. Although it is normally an instrument

More information

Rationale for replacing IVIG with Intralipid (IL) for immunological pregnancy loss

Rationale for replacing IVIG with Intralipid (IL) for immunological pregnancy loss Rationale for replacing IVIG with Intralipid (IL) for immunological pregnancy loss Recurrent Pregnancy Loss The reason that an embryo may not implant successfully is either because there is something intrinsically

More information

Ultrasound. Sound waves

Ultrasound. Sound waves Ultrasound Basic Idea Send waves into body which are reflected at the interfaces between tissue Return time of the waves tells us of the depth of the reflecting surface History First practical application,

More information

General Thoughts on Generator Set - Acoustic Solutions & Noise Control

General Thoughts on Generator Set - Acoustic Solutions & Noise Control t: 023 81 290160 www.excelpowerltd.co.uk f: 023 81 290260 info@excelpowerltd.co.uk General Thoughts on Generator Set - Acoustic Solutions & Noise Control Written By Steve Delaney A.M.I.O.A. of TAS Ltd

More information

Ann.wellhouse@TouchStoneScience.net 1. Enzyme Function

Ann.wellhouse@TouchStoneScience.net 1. Enzyme Function Ann.wellhouse@TouchStoneScience.net 1 Enzyme Function National Science Standards Science as Inquiry: Content Standard A: As a result of activities in grades 9-12, all students should develop: Abilities

More information

Human Growth and Reproduction

Human Growth and Reproduction Human Growth and Reproduction Sperm reach ovum and cluster around it Only one sperm is allowed to penetrate egg When the sperm penetrates the egg, the egg immediately releases a chemical creating a hard

More information

digital fever alert ear thermometer suitable from birth instruction manual

digital fever alert ear thermometer suitable from birth instruction manual digital fever alert ear thermometer suitable from birth instruction manual digital fever alert ear thermometer instruction manual Please read these instructions carefully before using the thermometer and

More information

DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling

DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE E. Günther, S. Hiebler, H. Mehling Bavarian Center for Applied Energy Research (ZAE Bayern) Walther-Meißner-Str.

More information

(Refer Slide Time 1:13 min)

(Refer Slide Time 1:13 min) Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras Module - 1 Lecture - 1 Introduction to the Study of Mechanical

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

Understanding Currency

Understanding Currency Understanding Currency Overlay July 2010 PREPARED BY Gregory J. Leonberger, FSA Director of Research Abstract As portfolios have expanded to include international investments, investors must be aware of

More information