Comparison of spirometry and abdominal height as four-dimensional computed tomography metrics in lung

Size: px
Start display at page:

Download "Comparison of spirometry and abdominal height as four-dimensional computed tomography metrics in lung"

Transcription

1 Comparison of spirometry and abdominal height as four-dimensional computed tomography metrics in lung Wei Lu, Daniel A. Low, a Parag J. Parikh, Michelle M. Nystrom, Issam M. El Naqa, Sasha H. Wahab, Maureen Handoko, David Fooshee, and Jeffrey D. Bradley Department of Radiation Oncology, Washington University School of Medicine, St. Louis, Missouri Received 22 October 2004; revised 14 February 2005; accepted for publication 26 April 2005; published 23 June 2005 An important consideration in four-dimensional CT scanning is the selection of a breathing metric for sorting the CT data and modeling internal motion. This study compared two noninvasive breathing metrics, spirometry and abdominal height, against internal air content, used as a surrogate for internal motion. Both metrics were shown to be accurate, but the spirometry showed a stronger and more reproducible relationship than the abdominal height in the lung. The abdominal height was known to be affected by sensor placement and patient positioning while the spirometer exhibited signal drift. By combining these two, a normalization of the drift-free metric to tidal volume may be generated and the overall metric precision may be improved American Association of Physicists in Medicine. DOI: / Key words: breathing motion, spirometry, abdominal height, radiation therapy I. INTRODUCTION Internal respiratory organ motion is the principal reason that intensity modulated radiation therapy IMRT has not been widely accepted for use in lung and upper-abdomen cancer treatment. 1 4 Four-dimensional computed tomography 4D CT is one technique proposed to reduce this uncertainty by retrospectively constructing three-dimensional 3D CT images at various parts of the respiratory cycle Our group has developed a 4D CT technique for free breathing by using a 16-slice CT scanner. 8,12 For both 4D CT 5 12 and gated radiotherapy, a metric that monitors breathing motion must be used since time cannot be directly correlated to tumor motion. Various breathing metrics have been proposed, ranging from implanted marker position, 2,5,18 21 abdominal height or chest height, 6,7,9,10,15,18,22,23 spirometry, 8,12,13,16,22 26 abdominal strain gauge, 13 wraparound inductive plethysmography, 22 and temperature sensors. 13 The only commercially available system uses abdominal height Real-Time Position Management, RPM, Varian Medical Systems, Palo Alto, CA, but the relationship between internal motion and abdominal height is sensitive to sensor placement and patient positioning. 15,16,27 A spirometer measures the air flow into the lungs and may provide a useful, verifiable physiologic metric. 12,16 We have found that spirometry is an excellent quantitative metric, but its significant signal drift means that it may not be appropriate as the sole metric for 4D CT or gated radiotherapy. 12,26 This study compared spirometer-measured tidal volume and abdominal height against internal air content, a surrogate for internal motion measurements. 12 The relationship between these two metrics for an entire CT scan session was also examined. II. MATERIALS AND METHODS A. 4D CT with spirometry Brief descriptions of the 4D CT process 8 and internal air content analysis 12,28 are given as follows. Transverse slices 1.5 mm thick were acquired using a 16-slice CT scanner Sensation 16, Siemens Medical Systems, Malvern, PA operated in 12-slice mode. The scanner was operated in ciné mode couch stationary during scanning with 15 scans acquired repeatedly at each couch position for 11 s. Each scan 360 rotation required 0.5 s to acquire followed by a 0.25 s dead time. The data were acquired continuously in space, but there was usually a pause of 4 s between two neighboring couch acquisitions. This process was conducted while the patient underwent synchronized spirometry measurements. The synchronization was provided by a photoresistor connected to the X-Ray On light. 12 Both the spirometry signal and the photoresister signal were measured with approximately 100 samples per second using a customized data acquisition program written in LABVIEW National Instruments, Dallas, TX. 8,28 The spirometry was calibrated to be linear and accurate for tidal volume flow rates greater than 100 ml/s typical breathing flow rates are in excess of 200 ml/s. 28 The slow drift in the spirometry signal was corrected by retrospectively examining the spirometer-measured tidal volume v and the internal air content V for the 15 CT scans. 12 This provided a spirometer-measured tidal volume v for each CT scan. The CT scans were then sorted into variable number of bins usually bins, depending on the range of tidal volumes the patient breathed. Each bin was assigned a specific tidal volume and direction of respiration inspiration or expiration. B. Internal air content analysis The internal air content analysis provided a measure of how much air volume was within a CT volume. 29 It was 2351 Med. Phys. 32 7, July /2005/32 7 /2351/7/$ Am. Assoc. Phys. Med. 2351

2 2352 Lu et al.: Comparison of spirometry and abdominal height 2352 FIG. 1. Measuring abdominal height. a Video acquisition system consisting of a digital camera pointed at the distance scale moving with the abdomen. b A sample frame taken from a video shows the template position determined by the template matching algorithm. The fiber optic cable appears bright in this sample frame indicating that the CT radiation is on. determined by the Hounsfield values in segmented aircontaining tissues lungs, trachea, bronchi. 12,28 In a previous study, we showed that the internal air content correlated with tissue motion with a residual of less than 1 mm for 12 patients and provided quantitative evaluations for the 4D CT process. 12 In this study, the internal air content was used as a surrogate for internal motion and the basis for metric comparison. C. Abdominal surface height measurement During the 4D CT procedure, an abdominal surface height measurement h was simultaneously acquired through digital video recording of a lightweight block placed on the patient s abdomen Fig. 1 a. The block was positioned between the umbilicus and the xiphoid process, based on the position of greatest tidal excursion as determined by a physician present at the simulation. We took care to tape the block tightly and straight up so that the motion was mostly in the anterior-posterior direction with negligible rotation effect. A distance scale was attached to the block surface facing the camera DCR-TRV 240, Sony, Tokyo, Japan for measurement calibration. A dark dot was marked at the center of the distance scale. It provided a template for automatically tracking of the block position on the digital movie frames Fig. 1 b. The video was synchronized to the CT scanner acquisition by a fiber optic cable connected to the same X-Ray On light as used for spirometry synchronization. Figure 1 b shows that the end of the fiber optic cable appears bright in the sample frame when the CT radiation is on. The video was downloaded from the camera and cropped to reduce the video file size with commercial video processing software PREMIERE V6.5, Adobe Systems Incorporated, San Jose, CA. The video had a sampling rate of 30 frames/s. The position of the block within each frame was automatically tracked using a two-dimensional template matching algorithm 30 implemented in MATLAB The Mathworks, Inc., Natick, MA. The template matching algorithm determines the position of a template in a target frame by finding the maximum normalized correlation between the template and the target. A rectangular region around the center dark dot in the first frame was manually selected as the template. The size of the template was approximately 2 by 2 mm. The template position determined by the algorithm was superimposed on each frame of the video Fig. 1 b. This generated a new video which was retrospectively viewed and from which the uncertainty of the template tracking was estimated for every patient to be less than 0.3 mm in both the lateral and anteroposterior directions. The anteroposterior position of the template provided the abdominal height measurement, which was calibrated from pixels into distance in millimeters using the imaged distance scale. This inexpensive in-house system provided similar surface height measurements to those provided by the RPM system with the advantage that it provided synchronization signals with the CT scanner. D. Comparison of spirometry and abdominal height as 4D CT metrics The spirometer-measured tidal volume and the abdominal height were compared against internal air content V as 4D CT metrics at each couch position 11 s. A patient-specific time offset t was assumed to exist between diaphragm motion causing abdominal motion, internal lung motion, and airflow at the mouth. The value of t between each metric and V was determined by maximizing a cross-correlation function CCF between each metric and V. The CCF evaluates the degree to which two functions are correlated and allows for the identification and estimation of t in two related signals. 12,18,31 33 On the other hand, a specific fractional error in V would correspond to the same fractional error in internal motion estimation, the fitting residual V root mean squared error of a first-order fit between each metric and V was used as a measure of the metric precision. Three 4D CT lung cancer patients and two 4D CT abdomen study patients had both spirometry and abdominal height data and were used in this study. Only couch positions that intercepted the lungs were used in this comparison. E. Comparison of spirometry and abdominal height for entire scan session The abdominal height measurement h was considered drift-free 22,26 and its relationship with the spirometermeasured tidal volume v was examined for the entire CT session s. This was done by dividing the CT session into 11 s segments based on the amount of time re-

3 2353 Lu et al.: Comparison of spirometry and abdominal height 2353 quired to acquire 15 CT scans from a single couch position and correlating h to v in every segment. When examining the relationship of these two metrics in each segment, t between h and v and a baseline drift in tidal volume were observed. We adopted an approach to simultaneously determine t and spirometry drift by maximizing the correlation between h and v for each segment: 1 t was determined by shifting v in time to maximize its correlation with h; 2 a linear spirometer drift correction was applied to v and the slope of the drift was adjusted to again maximize the correlation between v and h. The mean ratio of drift-corrected tidal volume change to the abdominal height change dv/dh, and the mean time offset t h-v were tabulated. III. RESULTS A. Comparison of spirometry and abdominal height as 4D CT metrics Figure 2 a shows the spirometer-measured tidal volume, abdominal height, and internal air content at the times corresponding to the 15 successive CT scans for one couch position. Figure 2 b shows the continuous tidal volume and abdominal height for the same couch position 11 s as shown in Fig. 2 a. The abdominal height changed by about 10 mm for the three breaths shown, while the tidal volume changed by about 500 ml. Because of unintended abdomen motion caused by involuntary abdomen muscle motion and cardiac motion, the abdominal trace has more small vibrations than the tidal volume trace. Similar vibrations were noticeable in surface displacement traces abdominal or chest measured by others. 3,34,35 Figure 3 shows the results of linearly fitting each metric to the internal air content for this couch position. The fitting residual was 2.8 ml for tidal volume Fig. 3 a and 4.3 ml for abdominal height Fig. 3 b, suggesting a greater precision of the tidal volume. The slopes of the fitted lines dv/dv and dv/dh represent the ratio of internal air content change to metric change. Figure 4 shows summary plots of these measurements as a function of couch positions for this lung cancer patient couch positions 5 18 are intercepting the lungs. The slopes were normalized to a common range of 0 1 for comparison. The two metrics display the same trend in the slopes Fig. 4 a. This indicates that both metrics have equivalent accuracy in determining changes in internal air content. For most couch positions within the lung, spirometry showed a better precision smaller fitting residual than abdominal height Fig. 4 b. This was true for all five patients. The mean time offset, fitting residual, and correlation between each metric and the internal air content are listed in Table I. We have shown that the mean fitting residuals between v and V v-v for the lung cancer patients corresponded to 3% 8% uncertainties in the overall tidal volumes. 12 For all patients, the spirometer-measured tidal volume had a better mean precision than the abdominal height. The correlation between the tidal volume and internal air content was slightly greater than or equal to that between the abdominal height and internal air content. With regard to FIG. 2. Breathing samples of spirometer-measured tidal volume, abdominal height, and internal air content vs scan index a and vs time b for a couch position in the middle of lung and intercepting the diaphragm. t, for all patients but patient 3, the abdominal height change led the internal air content change. Patient 3 showed a different phase relationship: the internal air content change led the abdominal height change. This was possibly due to the patient s breathing being driven more by the chest than for the other four patients. Further investigation showed that this patient also had the smallest peak-to-peak change in abdominal height 6 mm. The internal air content change led the spirometry change for all patients. The time offset between the spirometry and internal air content was smaller than that between the abdominal height and internal air content for every patient. B. Relationship between spirometry and abdominal height for entire scan session Figure 5 shows the relationship between the abdominal height and tidal volume for an entire scan session 350 s.

4 2354 Lu et al.: Comparison of spirometry and abdominal height 2354 FIG. 3. Linearly fitting spirometer-measured tidal volume a and abdominal height b to internal air content. Dashed lines represent ± fitting residuals. The values of dv/ dh were relatively unchanged during the session with a mean of ml/mm and a standard deviation s.d. of 3.11 ml/mm Fig. 5 a. The residual h-v for a first-order fit between v and h was relatively small with a mean of 19 ml or 3.25% of the overall tidal volume Fig. 5 b. The correlation between h and v r h-v was high throughout the session Fig. 5 c. The time offset t h-v plot shows that the abdominal height change always led the spirometry change with a mean of 0.12 s and a s.d. of 0.02 s Fig. 5 d. These measurements were tabulated for all five patients Table II. The first three lung cancer patients show less variation in dv/dh than the last two abdomen study patients, which might be the result of longer overall scan session times for abdomen study patients 600 s. The time offsets were relatively unchanged for every patient s.d s except for patient 4 s.d. =0.05 s, for whom the time offset increased significantly near the end of the session possibly due to patient relaxation. IV. DISCUSSION AND CONCLUSION The results showed that both metrics were accurate with respect to internal air content. The metric precision varied with the location of interest but spirometry was generally a better metric than the abdominal height for a location within the lung. The mean precision of spirometry within the lung FIG. 4. Summary plots for a lung cancer patient. a Normalized ratios of internal air content to tidal volume dv/dv and to abdominal height dv/dh. b Fitting residuals between tidal volume and internal air content v-v, and between abdominal height and internal air content h-v. was better for all patients, suggesting a stronger and more reproducible relationship of spirometry with internal air content and tissue motion. The time offset between the internal air content and the spirometry was smaller than that between the internal air content and the abdominal height. A recent fluoroscopic study by Hoisak et al. reached the same conclusions for 11 patients with data acquired over extended periods and over multiple days. 23 It has been reported that breathing patterns change when one breathes through a spirometer. 36,37 Gilbert et al. 36 reported that the tidal volume increased by 29±21% for 14 subjects in one study, while Askanazi et al. 37 found increases of 15.5% for 28 subjects. However, when evaluating metrics for 4D CT or gated radiation therapy, it is the relationship between internal motions and the breathing metrics that are the principal consideration. The fact that the tidal volume changes when a spirometer is employed does not necessarily change the relationship between tidal volume and internal motion. This relationship will need verification, but if the use of spirometry significantly affects this relationship, a spirometer can be employed during treatment. Though the internal air content functioned well as a surrogate for tissue motion, and the internal air content change

5 2355 Lu et al.: Comparison of spirometry and abdominal height 2355 TABLE I. Comparison of tidal volume v and abdominal height h against internal air content V as 4D CT metrics within the lung. Between v and V Between h and V Patient Site Time Fitting Time Fitting offset a residual offset a residual s ml Correlation s ml Correlation 1 Lung Lung Lung Abdomen Abdomen a A positive time offset indicates that the first variable is leading the second variable. FIG. 5. Relationship between abdominal height h and tidal volume v for the entire scan session. a Ratio of change in v to change in h dv/dh. b Fitting residual between h and v h-v. c Correlation between h and v r h-v. d Time offset between h and v t h-v. can be monitored either locally or globally in the lung, actual tumor motion would be better for evaluating breathing metric in radiotherapy. Reliably tracking tumor motion in 3D is ongoing research. Fluoroscopically tracking implanted markers 2,3,5,18 21 may yield the most reliable 3D tumor position, however, this technique is invasive and has other limitations. 19,21 A recent study showed that tumors in the lung can be tracked with a biplane digital radiography unit, while implanted markers were still needed for tumors in the liver and esophagus. 18 Fluoroscopically tracking the superior-inferior diaphragm motion has been conducted but is limited by the fact that only one-dimensional diaphragm motion was measured. 15,17,34,35 The development of respiration-correlated 4D CT techniques may provide a noninvasive way for tracking and modeling 3D tumor/tissue motion. For the five patients in this study, there was a high correlation between the spirometry and abdominal height 0.98, and their relationship dv/dh and time offset was relatively unchanged for the entire CT session s under free breathing. Zhang et al. 26 reported a stable correlation between spirometry and chest wall height for healthy volunteers during normal breathing cycles. Because of the large variation exhibited in breathing patterns, it is important to establish and verify the dv/dh relationship for each scanning session and each treatment session. We have so far assumed that this relationship would change only slightly during each treatment session min, in which the patient is kept at a fixed position and undergoes quiet respiration. Our results, acquired during CT simulation, have shown that TABLE II. Relationship between the tidal volume v and abdomen height h for the entire scan session s. Values are mean s.d. for all 11 s segments. Patient Site Time offset a s dv/dh ml/mm Fitting residual % Correlation 1 Lung Lung Lung Abdomen Abdomen a A positive time offset indicates that the first variable is leading the second variable.

6 2356 Lu et al.: Comparison of spirometry and abdominal height 2356 this assumption is valid for 6 11 min. If this assumption holds, we hypothesize that it will be possible to acquire both metrics for several breathing cycles and normalize the abdominal height into tidal volume for each patient just before the treatment. 16,25,26 After that, the tidal volume measurement would not be necessary unless removing the spirometer is found to significantly alter the relationship of internal motion and abdomen height. This would be advantageous for patients who did not want to undergo spirometry during the entire process for which the tidal volume was required. Finally, one should be cautious when using any external breathing metrics spirometry and abdominal height to infer internal tumor or tissue motion. 3,23 At the least the following issues need to be addressed. 1 Is there a high correlation between the external metric and internal motion? 2 Is this relationship reproducible during the time of consideration? 3 Is there a mechanism to determine and account for the time offsets between the external metric and internal motion? Nevertheless, external breathing metrics have been demonstrated useful in predicting the tumor motion and thus in reducing the uncertainties caused by respiratory motion for many patients. 6 10,12,13,15,16,18,22 26 A more thorough verification of their relationships with tumor motions by using realtime, direct tumor or indirect implanted marker imaging is under development in our institution. ACKNOWLEDGMENT This work was supported in part by National Institute of Health Grant No. R01 CA a Author to whom correspondence should be addressed. Electronic mail: low@wustl.edu 1 A. M. Allen, K. M. Siracuse, J. A. Hayman, and J. M. Balter, Evaluation of the influence of breathing on the movement and modeling of lung tumors, Int. J. Radiat. Oncol., Biol., Phys. 58, D. P. Gierga, G. T. Chen, J. H. Kung, M. Betke, J. Lombardi, and C. G. Willett, Quantification of respiration-induced abdominal tumor motion and its impact on IMRT dose distributions, Int. J. Radiat. Oncol., Biol., Phys. 58, C. Ozhasoglu and M. J. Murphy, Issues in respiratory motion compensation during external-beam radiotherapy, Int. J. Radiat. Oncol., Biol., Phys. 52, C. X. Yu, D. A. Jaffray, and J. W. Wong, The effects of intra-fraction organ motion on the delivery of dynamic intensity modulation, Phys. Med. Biol. 43, H. Shirato, S. Shimizu, K. Kitamura, T. Nishioka, K. Kagei, S. Hashimoto, H. Aoyama, T. Kunieda, N. Shinohara, H. Dosaka-Akita, and K. Miyasaka, Four-dimensional treatment planning and fluoroscopic realtime tumor tracking radiotherapy for moving tumor, Int. J. Radiat. Oncol., Biol., Phys. 48, S. S. Vedam, P. J. Keall, V. R. Kini, H. Mostafavi, H. P. Shukla, and R. Mohan, Acquiring a four-dimensional computed tomography dataset using an external respiratory signal, Phys. Med. Biol. 48, T. Pan, T. Y. Lee, E. Rietzel, and G. T. Chen, 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT, Med. Phys. 31, D. A. Low, M. Nystrom, E. Kalinin, P. Parikh, J. F. Dempsey, J. D. Bradley, S. Mutic, S. H. Wahab, T. Islam, G. Christensen, D. G. Politte, and B. R. Whiting, A method for the reconstruction of four-dimensional synchronized CT scans acquired during free breathing, Med. Phys. 30, P. J. Keall, G. Starkschall, H. Shukla, K. M. Forster, V. Ortiz, C. W. Stevens, S. S. Vedam, R. George, T. Guerrero, and R. Mohan, Acquiring 4D thoracic CT scans using a multislice helical method, Phys. Med. Biol. 49, P. Keall, 4-dimensional computed tomography imaging and treatment planning, Semin. Radiat. Oncol. 14, G. S. Mageras, A. Pevsner, E. D. Yorke, K. E. Rosenzweig, E. C. Ford, A. Hertanto, S. M. Larson, D. M. Lovelock, Y. E. Erdi, S. A. Nehmeh, J. L. Humm, and C. C. Ling, Measurement of lung tumor motion using respiration-correlated CT, Int. J. Radiat. Oncol., Biol., Phys. 60, W. Lu, P. J. Parikh, I. M. ElNaqa, M. M. Nystrom, J. P. Hubenschmidt, S. H. Wahab, S. Mutic, A. K. Singh, G. E. Christensen, J. D. Bradley, and D. A. Low, Quantitation of the four-dimensional computed tomography process for lung cancer patients, Med. Phys. in press. 13 H. D. Kubo and B. C. Hill, Respiration gated radiotherapy treatment: A technical study, Phys. Med. Biol. 41, P. J. Keall, V. R. Kini, S. S. Vedam, and R. Mohan, Potential radiotherapy improvements with respiratory gating, Australas. Phys. Eng. Sci. Med. 25, S. Minohara, T. Kanai, M. Endo, K. Noda, and M. Kanazawa, Respiratory gated irradiation system for heavy-ion radiotherapy, Int. J. Radiat. Oncol., Biol., Phys. 47, T. Zhang, H. Keller, M. J. O Brien, T. R. Mackie, and B. Paliwal, Application of the spirometer in respiratory gated radiotherapy, Med. Phys. 30, S. S. Vedam, P. J. Keall, V. R. Kini, and R. Mohan, Determining parameters for respiration-gated radiotherapy, Med. Phys. 28, Y. Tsunashima, T. Sakae, Y. Shioyama, K. Kagei, T. Terunuma, A. Nohtomi, and Y. Akine, Correlation between the respiratory waveform measured using a respiratory sensor and 3D tumor motion in gated radiotherapy, Int. J. Radiat. Oncol., Biol., Phys. 60, Y. Seppenwoolde, H. Shirato, K. Kitamura, S. Shimizu, M. van Herk, J. V. Lebesque, and K. Miyasaka, Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy, Int. J. Radiat. Oncol., Biol., Phys. 53, G. C. Sharp, S. B. Jiang, S. Shimizu, and H. Shirato, Prediction of respiratory tumour motion for real-time image-guided radiotherapy, Phys. Med. Biol. 49, G. C. Sharp, D. Gierga, and S. B. Jiang, Reliable real-time tracking in fluoroscopy for image guided radiotherapy, AAPM 46th Annual Meeting, Pittsburgh, PA, H. D. Kubo, P. M. Len, S. Minohara, and H. Mostafavi, Breathingsynchronized radiotherapy program at the University of California Davis Cancer Center, Med. Phys. 27, J. Hoisak, K. Sixel, R. Tirona, P. Cheung, and J.-P. Pignol, Correlation of lung tumour motion with surrogates of respiration, AAPM 46th Annual Meeting, Pittsburgh, PA, T. Zhang, R. Jeraj, H. Keller, W. Lu, G. H. Olivera, T. R. McNutt, T. R. Mackie, and B. Paliwal, Treatment plan optimization incorporating respiratory motion, Med. Phys. 31, T. Zhang, H. Keller, R. Jeraj, R. Manon, J. Welsh, R. Patel, J. Fenwick, M. Mehta, T. R. Mackie, and B. Paliwal, Breathing synchronized delivery - A new technique for radiation treatment of the targets with respiratory motion, Int. J. Radiat. Oncol., Biol., Phys. 57, S T. Zhang, H. Keller, R. Jeraj, M. J. O Brien, K. Sheng, T. R. Mackie, and B. Paliwal, Lung motion tracking with a combined spirometer-laser sensor system, AAPM 44th Annual Meeting, Montreal, D. Gierga, G. Sharp, J. Brewer, M. Betke, C. Willett, and G. T. Chen, Correlation between external and internal markers for abdominal tumors: Implications for respiratory gating, Int. J. Radiat. Oncol., Biol., Phys. 57, S D. A. Low, P. J. Parikh, I. M. ElNaqa, M. M. Nystrom, W. Lu, J. P. Hubenschmidt, S. H. Wahab, S. Mutic, A. K. Singh, G. Christensen, and J. D. Bradley, Quantitative 4-D CT using a multislice CT scanner, Proceedings of the 14th International Conference on the Use of Computers in Radiation Therapy, Seoul, Korea, E. A. Hoffman and E. L. Ritman, Effect of body orientation on regional lung expansion in dog and sloth, J. Appl. Physiol. 59, R. C. Gonzalez and R. E. Woods, Digital Image Processing Addison Wesley, Reading, MA, 1992.

7 2357 Lu et al.: Comparison of spirometry and abdominal height R. F. W. Coates, Modern Communication Systems Macmillan, London, R. N. Bracewell, The Fourier Transform and its Applications McGraw- Hill, New York, J. D. Hoisak, K. E. Sixel, R. Tirona, P. C. Cheung, and J. P. Pignol, Correlation of lung tumor motion with external surrogate indicators of respiration, Int. J. Radiat. Oncol., Biol., Phys. 60, Q. S. Chen, M. S. Weinhous, F. C. Deibel, J. P. Ciezki, and R. M. Macklis, Fluoroscopic study of tumor motion due to breathing: Facilitating precise radiation therapy for lung cancer patients, Med. Phys. 28, G. S. Mageras, E. Yorke, K. Rosenzweig, L. Braban, E. Keatley, E. Ford, S. A. Leibel, and C. C. Ling, Fluoroscopic evaluation of diaphragmatic motion reduction with a respiratory gated radiotherapy system, J. Appl. Clin. Med. Phys. 2, R. Gilbert, J. H. Auchincloss, Jr., J. Brodsky, and W. Boden, Changes in tidal volume, frequency, and ventilation induced by their measurement, J. Appl. Physiol. 33, J. Askanazi, P. A. Silverberg, R. J. Foster, A. I. Hyman, J. Milic-Emili, and J. M. Kinney, Effects of respiratory apparatus on breathing pattern, J. Appl. Physiol.: Respir., Environ. Exercise Physiol. 48,

Gated Radiotherapy for Lung Cancer

Gated Radiotherapy for Lung Cancer Gated Radiotherapy for Lung Cancer Steve B. Jiang, Ph.D. Depart Of Radiation Oncology University of California San Diego sbjiang@ucsd.edu radonc.ucsd.edu/research/cart Two Types of Gating Internal gating

More information

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan

Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan An Brief Introduction to 4D CT Scanning Steve B. Jiang, Ph.D. Dept of Radiation Oncology Univ of California San Diego Outline Problems with free breathing 3D scanning What is 4D CT? How does it work? Acknowledgements

More information

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D. 4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?

More information

GEOMETRIC ACCURACY OF A REAL-TIME TARGET TRACKING SYSTEM WITH DYNAMIC MULTILEAF COLLIMATOR TRACKING SYSTEM

GEOMETRIC ACCURACY OF A REAL-TIME TARGET TRACKING SYSTEM WITH DYNAMIC MULTILEAF COLLIMATOR TRACKING SYSTEM doi:10.1016/j.ijrobp.2006.04.038 Int. J. Radiation Oncology Biol. Phys., Vol. 65, No. 5, pp. 1579 1584, 2006 Copyright 2006 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/06/$ see front

More information

Real-time Position Management System

Real-time Position Management System Real-time Position Management System COMPREHENSIVE SYSTEM FOR TOTAL MOTION MANAGEMENT RPM gates the imaging and treatment beams on and off at predetermined intervals based upon the patient s breathing

More information

Tracking Errors and Repair

Tracking Errors and Repair INSTITUTE OF PHYSICS PUBLISHING Phys. Med. Biol. 49 (2004) 5347 5356 PHYSICS IN MEDICINE AND BIOLOGY PII: S0031-9155(04)85063-2 Tracking errors in a prototype real-time tumour tracking system Gregory C

More information

EVALUATION OF LINEAR ACCELERATOR GATING WITH REAL-TIME ELECTROMAGNETIC TRACKING

EVALUATION OF LINEAR ACCELERATOR GATING WITH REAL-TIME ELECTROMAGNETIC TRACKING doi:10.1016/j.ijrobp.2009.01.034 Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 3, pp. 920 927, 2009 Copyright Ó 2009 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/09/$ see front

More information

The Management of Respiratory Motion in Radiation Oncology

The Management of Respiratory Motion in Radiation Oncology AAPM REPORT NO. 91 The Management of Respiratory Motion in Radiation Oncology Report of AAPM Task Group 76 July 2006 DISCLAIMER: This publication is based on sources and information believed to be reliable,

More information

Real-Time DMRT Delivery For Mobile and Deforming Targets

Real-Time DMRT Delivery For Mobile and Deforming Targets Real-time DMLC IMRT delivery for mobile and deforming targets Lech Papiez and Dharanipathy Rangaraj Department of Radiation Oncology, Indiana University School of Medicine, Indianapolis, Indiana 46202

More information

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients

Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients 대한방사선종양학회지 2005;23(3):169~175 Analysis of Set-up Errors during CT-scan, Simulation, and Treatment Process in Breast Cancer Patients Department of Radiation Oncology, College of Medicine, Ewha Womans University

More information

Radiation Protection in Radiotherapy

Radiation Protection in Radiotherapy Radiation Protection in Radiotherapy Albert Lisbona Medical Physics Department CLCC Nantes Atlantique 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Radiation therapy The lecture is oriented to

More information

A Patient s Guide to the Calypso System for Breast Cancer Treatment

A Patient s Guide to the Calypso System for Breast Cancer Treatment A Patient s Guide to the Calypso System for Breast Cancer Treatment Contents Introduction How Radiation Works To Treat Cancer Radiation Treatment for Breast Cancer Special Concerns for Treating the Left

More information

4DCT radiotherapy for NSCLC: a review of planning methods

4DCT radiotherapy for NSCLC: a review of planning methods Journal of Radiotherapy in Practice Journal of Radiotherapy in Practice (2014) Page 1 of 10 & Cambridge University Press 2014 doi:10.1017/s1460396914000041 Literature Review 4DCT radiotherapy for NSCLC:

More information

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,

More information

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.

IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment. DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering

More information

ONLINE FIRST. http://journal.publications.chestnet.org/ Online First articles are not copyedited prior to posting.

ONLINE FIRST. http://journal.publications.chestnet.org/ Online First articles are not copyedited prior to posting. Page 1 of 25 ONLINE FIRST This is an Online First, unedited version of this article. The final, edited version will appear in a numbered issue of CHEST and may contain substantive changes. We encourage

More information

Validation of 3D motion tracking of pulmonary lesions using CT fluoroscopy images for robotically assisted lung biopsy

Validation of 3D motion tracking of pulmonary lesions using CT fluoroscopy images for robotically assisted lung biopsy Validation of 3D motion tracking of pulmonary lesions using CT fluoroscopy images for robotically assisted lung biopsy Sheng Xu* a, Gabor Fichtinger a, Russell H. Taylor a, Kevin Cleary b a Engineering

More information

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.

Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source

More information

AK IMRT Würzburg 26./27.03.2009. Roland Kramer, Conmedica GmbH, Schriesheim

AK IMRT Würzburg 26./27.03.2009. Roland Kramer, Conmedica GmbH, Schriesheim AK IMRT Würzburg 26./27.03.2009 Roland Kramer, Conmedica GmbH, Schriesheim 2 AK IMRT Würzburg 26./27.03.2009 Roland Kramer, Conmedica GmbH, Schriesheim OVERVIEW 3 Company Background AlignRT GateCT and

More information

STABILITY OF MARKERS USED FOR REAL-TIME TUMOR TRACKING AFTER PERCUTANEOUS INTRAPULMONARY PLACEMENT

STABILITY OF MARKERS USED FOR REAL-TIME TUMOR TRACKING AFTER PERCUTANEOUS INTRAPULMONARY PLACEMENT doi:10.1016/j.ijrobp.2010.12.026 Int. J. Radiation Oncology Biol. Phys., Vol. 81, No. 3, pp. e75 e81, 2011 Copyright Ó 2011 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ - see front

More information

Intensity-Modulated Radiation Therapy (IMRT)

Intensity-Modulated Radiation Therapy (IMRT) Scan for mobile link. Intensity-Modulated Radiation Therapy (IMRT) Intensity-modulated radiotherapy (IMRT) uses linear accelerators to safely and painlessly deliver precise radiation doses to a tumor while

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

Deformable Registration for Image-Guided Radiation Therapy

Deformable Registration for Image-Guided Radiation Therapy Symposium Multimodality Imaging for Radiotherapy : State of the art, needs and perspectives 1/33 Deformable Registration for Image-Guided Radiation Therapy David Sarrut «Radiation, images, oncology» team

More information

QA of intensity-modulated beams using dynamic MLC log files

QA of intensity-modulated beams using dynamic MLC log files 36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,

More information

Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate

Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate Corporate Medical Policy Intensity-Modulated Radiation Therapy (IMRT) of the Prostate File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_prostate

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

FOR IMMEDIATE RELEASE

FOR IMMEDIATE RELEASE FOR IMMEDIATE RELEASE The PROBEAT-RT Proton Beam Therapy System Incorporating Real-time Tumor-tracking is Approved for Commercial Manufacturing and Sales under Japan s Pharmaceutical Affairs Law Treatments

More information

Multi-slice Helical CT Scanning of the Chest

Multi-slice Helical CT Scanning of the Chest Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.

More information

Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT

Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Daily IGRT with CT-on-Rails Can Safely Reduce Planning Margin for Prostate Cancer: Implication for SBRT Wen Li, Andrew Vassil, Lama Mossolly, Qingyang Shang, Ping Xia Department of Radiation Oncology Why

More information

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How

More information

intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis 11/2009 5/2016 5/2017 5/2016

intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis 11/2009 5/2016 5/2017 5/2016 Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Abdomen File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_abdomen_and_pelvis

More information

ELECTROMAGNETIC-GUIDED DYNAMIC MULTILEAF COLLIMATOR TRACKING ENABLES MOTION MANAGEMENT FOR INTENSITY-MODULATED ARC THERAPY

ELECTROMAGNETIC-GUIDED DYNAMIC MULTILEAF COLLIMATOR TRACKING ENABLES MOTION MANAGEMENT FOR INTENSITY-MODULATED ARC THERAPY doi:10.1016/j.ijrobp.2010.03.011 Int. J. Radiation Oncology Biol. Phys., Vol. 79, No. 1, pp. 312 320, 2011 Copyright Ó 2011 Elsevier Inc. Printed in the USA. All rights reserved 0360-3016/$ see front matter

More information

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning

Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,

More information

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS

INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS INTENSITY MODULATED RADIATION THERAPY (IMRT) FOR PROSTATE CANCER PATIENTS HOW DOES RADIATION THERAPY WORK? Cancer is the unrestrained growth of abnormal cells within the body. Radiation therapy kills both

More information

Tutorial for Tracker and Supporting Software By David Chandler

Tutorial for Tracker and Supporting Software By David Chandler Tutorial for Tracker and Supporting Software By David Chandler I use a number of free, open source programs to do video analysis. 1. Avidemux, to exerpt the video clip, read the video properties, and save

More information

PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops

PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops YOU SHOULD READ THE FOLLOWING MATERIAL BEFORE Tuesday March 30 Interpretation of PFTs Learning Objectives 1. Specify the indications

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning.

The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. The validation of a tumour control probability (TCP) model for supporting radiotherapy treatment planning. KAROLINA KOKUREWICZ S U P E R V I S O R S : D I N O J A R O S Z Y N S K I, G I U S E P P E S C

More information

Airways Resistance and Airflow through the Tracheobronchial Tree

Airways Resistance and Airflow through the Tracheobronchial Tree Airways Resistance and Airflow through the Tracheobronchial Tree Lecturer: Sally Osborne, Ph.D. Department of Cellular & Physiological Sciences Email: sosborne@interchange.ubc.ca Useful links: www.sallyosborne.com

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of Head and Neck File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_head_and_neck

More information

Baylor Radiosurgery Center

Baylor Radiosurgery Center Radiosurgery Center Baylor Radiosurgery Center Sophisticated Radiosurgery for both Brain and Body University Medical Center at Dallas Radiosurgery Center 3500 Gaston Avenue Hoblitzelle Hospital, First

More information

GOM Optical Measuring Techniques. Deformation Systems and Applications

GOM Optical Measuring Techniques. Deformation Systems and Applications GOM Optical Measuring Techniques Deformation Systems and Applications ARGUS Forming Analysis ARGUS Deformation analysis in sheet metal and forming industry Forming Characteristics of Sheet Metals Material

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard

More information

Failure Modes and Effects Analysis (FMEA)

Failure Modes and Effects Analysis (FMEA) Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide

More information

Basic techniques of pulmonary physical therapy (I) 100/04/24

Basic techniques of pulmonary physical therapy (I) 100/04/24 Basic techniques of pulmonary physical therapy (I) 100/04/24 Evaluation of breathing function Chart review History Chest X ray Blood test Observation/palpation Chest mobility Shape of chest wall Accessory

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

HITACHI Proton Beam Therapy System

HITACHI Proton Beam Therapy System HITACHI Proton Beam Therapy System Masumi Umezawa, M.S. Hitachi, Ltd., Hitachi Research Laboratory HITACHI Proton Beam Therapy System Contents 1. Overview of HITACHI 2. HITACHI Proton Beam Therapy and

More information

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical

More information

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems

A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department

More information

Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of the Chest

Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of the Chest Corporate Medical Policy Intensity Modulated Radiation Therapy (IMRT) of the Chest File Name: Origination: Last CAP Review: Next CAP Review: Last Review: intensity_modulated_radiation_therapy_imrt_of_the_chest

More information

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,

More information

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.

Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT

More information

Assessment of Microsoft Kinect Technology (Kinect for Xbox and Kinect for Windows) for Patient Monitoring during External Beam Radiotherapy

Assessment of Microsoft Kinect Technology (Kinect for Xbox and Kinect for Windows) for Patient Monitoring during External Beam Radiotherapy Assessment of Microsoft Kinect Technology (Kinect for Xbox and Kinect for Windows) for Patient Monitoring during External Beam Radiotherapy F. Tahavori, M. Alnowami, J. Jones, P. Elangovan, E. Donovan

More information

MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE

MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE K. Kohm ORBIMAGE, 1835 Lackland Hill Parkway, St. Louis, MO 63146, USA kohm.kevin@orbimage.com

More information

Comparison of Threshold-Based Segmentation Methods on Pre- and Post- Therapy PET Scans

Comparison of Threshold-Based Segmentation Methods on Pre- and Post- Therapy PET Scans M. Phillips, S.F. Barrington, D.L.G. Hill, P.K. Marsden 1 Comparison of Threshold-Based Segmentation Methods on Pre- and Post- Therapy PET Scans Michael Phillips 1 Michael.Phillips.1@city.ac.uk Sally F.

More information

What is the CyberKnife System?

What is the CyberKnife System? CYBERKNIFE Robotic Radiosurgery System PATIENT BROChURE 1 What is the CyberKnife System? have stabilizing frames bolted to their head or limit their breathing during treatment to minimize movement of the

More information

Prediction of Respiratory Motion Using A Statistical 4D Mean Motion Model

Prediction of Respiratory Motion Using A Statistical 4D Mean Motion Model Second International Workshop on -3- Prediction of Respiratory Motion Using A Statistical 4D Mean Motion Model Jan Ehrhardt 1, René Werner 1, Alexander Schmidt Richberg 1, and Heinz Handels 1 Department

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

02 CyberKnife: Treatment Delivery

02 CyberKnife: Treatment Delivery TREATMENT DELIVERY CyberKnife Treatment Delivery System The CyberKnife System is the first and only robotic radiosurgery system to offer highly precise and customizable, non-surgical treatment options

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

Palmprint Recognition. By Sree Rama Murthy kora Praveen Verma Yashwant Kashyap

Palmprint Recognition. By Sree Rama Murthy kora Praveen Verma Yashwant Kashyap Palmprint Recognition By Sree Rama Murthy kora Praveen Verma Yashwant Kashyap Palm print Palm Patterns are utilized in many applications: 1. To correlate palm patterns with medical disorders, e.g. genetic

More information

Quantitative Imaging In Clinical Trials Using PET/CT: Update

Quantitative Imaging In Clinical Trials Using PET/CT: Update Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative

More information

P R E S E N T S Dr. Mufa T. Ghadiali is skilled in all aspects of General Surgery. His General Surgery Services include: General Surgery Advanced Laparoscopic Surgery Surgical Oncology Gastrointestinal

More information

Whitepaper. Image stabilization improving camera usability

Whitepaper. Image stabilization improving camera usability Whitepaper Image stabilization improving camera usability Table of contents 1. Introduction 3 2. Vibration Impact on Video Output 3 3. Image Stabilization Techniques 3 3.1 Optical Image Stabilization 3

More information

RADIOTHERAPY Giovanna Mantello

RADIOTHERAPY Giovanna Mantello F The impact on RADIOTHERAPY Giovanna Mantello AOU Osp. Riuniti Ancona gio@mobilia.it RADIOTHERAPY OF PULMONARY LESIONS INTENSITY MODULATION HIGH GRADIENT DOSE DOSE ESCALATION IPOFRACTIONATION STEREO BODY

More information

GE Healthcare. pet/ct for simulation. Precision in motion.

GE Healthcare. pet/ct for simulation. Precision in motion. GE Healthcare pet/ct for simulation. Precision in motion. Patient profile H: 5 10 W: 370 lbs BMI: 53 Each cancer patient is just as unique as his or her cancer. Approaches to cancer treatment are changing

More information

Image Registration and Fusion. Professor Michael Brady FRS FREng Department of Engineering Science Oxford University

Image Registration and Fusion. Professor Michael Brady FRS FREng Department of Engineering Science Oxford University Image Registration and Fusion Professor Michael Brady FRS FREng Department of Engineering Science Oxford University Image registration & information fusion Image Registration: Geometric (and Photometric)

More information

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

1. Provide clinical training in radiation oncology physics within a structured clinical environment. Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible

More information

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University

More information

Measuring Line Edge Roughness: Fluctuations in Uncertainty

Measuring Line Edge Roughness: Fluctuations in Uncertainty Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as

More information

Computerized Tomography, X-rays, and the Radon Transform

Computerized Tomography, X-rays, and the Radon Transform 1 Computerized Tomography, X-rays, and the Radon Transform 1.1 Introduction The purpose of this chapter is to give an informal introduction to the subject of tomography. There are very few mathematical

More information

Dental Radiography Core Subject. Digital Radiography

Dental Radiography Core Subject. Digital Radiography Dental Radiography Core Subject Digital Radiography Aims: To develop an understanding of the history of digital radiography, the different types of digital x-rays and the advantages and disadvantages of

More information

Purchasing a cardiac CT scanner: What the radiologist needs to know

Purchasing a cardiac CT scanner: What the radiologist needs to know Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage

More information

Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system

Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system

More information

Oftentimes, as implant surgeons, we are

Oftentimes, as implant surgeons, we are CLINICAL AVOIDING INJURY TO THE INFERIOR ALVEOLAR NERVE BY ROUTINE USE OF INTRAOPERATIVE RADIOGRAPHS DURING IMPLANT PLACEMENT Jeffrey Burstein, DDS, MD; Chris Mastin, DMD; Bach Le, DDS, MD Injury to the

More information

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology

A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology A Three-Dimensional Correlation Method for Registration of Medical Images in Radiology Michalakis F. Georgiou 1, Joachim H. Nagel 2, George N. Sfakianakis 3 1,3 Department of Radiology, University of Miami

More information

Reflection and Refraction

Reflection and Refraction Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,

More information

Comparison of Epson scanner quality for radiochromic film evaluation

Comparison of Epson scanner quality for radiochromic film evaluation JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Comparison of Epson scanner quality for radiochromic film evaluation Hani Alnawaf, 1,3 Peter K.N. Yu, 2 Martin Butson 1,2,3a South

More information

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3

GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures

More information

Pulmonary Ventilation

Pulmonary Ventilation Pulmonary Ventilation Graphics are used with permission of: Pearson Education Inc., publishing as Benjamin Cummings (http://www.aw-bc.com) Page 1. Introduction Pulmonary ventilation, or breathing, is the

More information

Modelling physiological features of Human body behavior in car crash simulations

Modelling physiological features of Human body behavior in car crash simulations Modelling physiological features of Human body behavior in car crash simulations Michel BEHR, Yves GODIO, Maxime LLARI, Christian BRUNET Laboratoire de Biomécanique Appliquée (Marseille, France) Presented

More information

Under 20 of rotation the accuracy of data collected by 2D software is maintained?

Under 20 of rotation the accuracy of data collected by 2D software is maintained? Under 20 of rotation the accuracy of data collected by 2D software is maintained? Introduction This study was undertaken in order to validate the Quintic two-dimensional software. Unlike threedimensional

More information

How To Improve Your Ct Image Quality

How To Improve Your Ct Image Quality Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

SDX and radiotherapy

SDX and radiotherapy SDX and radiotherapy ADVANTAGE OF DEEP INSPIRATION BREATH HOLD Research on the improvement of the quality of the radiotherapy of cancerous tumours that are influenced by breathing and diaphragm movements.

More information

LUNG VOLUMES AND CAPACITIES

LUNG VOLUMES AND CAPACITIES LUNG VOLUMES AND CAPACITIES STANDARDS 3.1.10A, 3.1.12A Identify the function of subsystems within a larger system; analyze and describe the function, interaction and relationship Westminster College among

More information

INTRODUCTION. A. Purpose

INTRODUCTION. A. Purpose New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes

More information

Bioinstrumentation. Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University

Bioinstrumentation. Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University Bioinstrumentation Medical Instrumentation: Application and Design Third Edition John G. Webster, Editor Kuo-Sheng Cheng, Ph.D. Department of Biomedical Engineering National Cheng Kung University The Importance

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

Back to Basics: Introduction to Industrial Barcode Reading

Back to Basics: Introduction to Industrial Barcode Reading Back to Basics: Introduction to Industrial Barcode Reading 1 Agenda What is a barcode? History 1 D codes Types and terminology 2 D codes Types and terminology Marking Methods Laser Scanning Image Based

More information

IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS

IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS IN-LINE PARTICLE SIZE MEASUREMENTS FOR CEMENT AND OTHER ABRASIVE PROCESS ENVIRONMENTS 1998 A.P. Malcolmson Malvern, Inc. 10 Southville Road Southborough, MA D. J. Holve Malvern/Insitec, Inc. 2110 Omega

More information

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma): Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 47 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL Refresher Course

More information

Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration.

Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration. SOP Spirometry 1. General considerations Spirometry serves as a physiological test to quantify pulmonary disease severity and to assess clinical change in respiratory function over time. Standard spirometric

More information

BRACHYTHERAPY FOR TREATMENT OF BREAST CANCER

BRACHYTHERAPY FOR TREATMENT OF BREAST CANCER BRACHYTHERAPY FOR TREATMENT OF BREAST CANCER Coverage for services, procedures, medical devices and drugs are dependent upon benefit eligibility as outlined in the member's specific benefit plan. This

More information

Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group

Acknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group Cone Beam CT guided Radiotherapy Jan-Jakob Sonke Acknowledgements NKI-AVL: Marcel van Herk, Jose Belderbos, Suzanne van Beek, Anja Betgen, Josien de Bois, ~Rianne de Jong, Michel Frenay, Danny Minkema,

More information

Physics and Dosimetry of SBRT: Simulation and Localization. Moyed Miften, PhD University of Colorado

Physics and Dosimetry of SBRT: Simulation and Localization. Moyed Miften, PhD University of Colorado Physics and Dosimetry of SBRT: Simulation and Localization Moyed Miften, PhD University of Colorado Team Kelly Stuhr Bodo Reitz Quentin Diot Tracey Schefter Laurie Gaspar Brian Kavanagh Rad Onc Staff at

More information

CLINICAL IMPLEMENTATION OF 4-DIMENSIONAL RADIOTHERAPY FOR TREATMENT OF LUNG CANCER

CLINICAL IMPLEMENTATION OF 4-DIMENSIONAL RADIOTHERAPY FOR TREATMENT OF LUNG CANCER CLINICAL IMPLEMENTATION OF 4-DIMENSIONAL RADIOTHERAPY FOR TREATMENT OF LUNG CANCER Klinische implementatie van 4-dimensionale radiotherapie voor de behandeling van longkanker René Widjai Marcel Underberg

More information

Esperienza di stereotassia polmonare al Campus Bio-Medico: tecnica e risultati Rolando M. D Angelillo

Esperienza di stereotassia polmonare al Campus Bio-Medico: tecnica e risultati Rolando M. D Angelillo Esperienza di stereotassia polmonare al Campus Bio-Medico: tecnica e risultati Rolando M. D Angelillo Università Campus Bio-Medico di Roma - Via Álvaro del Portillo, 21-00128 Roma Italia BED 10 > 100 Gy

More information