Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control



Similar documents
A Complete Solution for Method Linearity in HPLC and UHPLC

INTRODUCTION. The principles of which are to:

Optimize Your Process Operations by Improving Process Monitoring

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Separation of Peptides from Enzymatic Digestion on Different Acclaim Columns: A Comparative Study

Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

HPLC Column Comparison Screening Study for Reversed Phase Columns

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

The Theory of HPLC. Gradient HPLC

WATERS QUANTITATIVE ANALYSIS solutions

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

What is Eluent Generation?

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

IC-MS and LC-MS Determination of Ionic Liquids, Counterions, and Impurities

Rapid Screening Method for Illicit Drugs, Using an Advanced Solid Core UHPLC Column and UHPLC System with MS/MS Detection

How To Test For Contamination In Large Volume Water

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Comparison of the Reversed-Phase Selectivity of Solid Core HPLC Columns

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Transforming a ChromLab Software 2-D Purification Template into an Automated Multidimensional (Multi-D) Purification Workflow An Instructional Guide

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Thermo Scientific Syncronis HPLC Columns. Technical Manual

Thermo Scientific Dionex UltiMate 3000 Variable Wavelength Detectors

Protocol: HPLC (amino acids)

Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

ChromQuest 5.0 Chromatography Data System

Development of an Automated Method for Monoclonal Antibody Purification and Analysis. Introduction. Gurmil Gendeh, 1 Wim Decrop, 2 and Remco Swart 2 1

How To Use An Acquity Qda Detector

Thermo Scientific Vanquish UHPLC System. Revolutionize Your. UHPLC Experience

Chemical analysis service, Turner s Green Technology Group

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Application Note. Separation of three monoclonal antibody variants using MCSGP. Summary

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Henry Shion, Robert Birdsall, Steve Cubbedge, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Robert Birdsall, Eoin Cosgrave, Henry Shion, and Weibin Chen Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Accelerate your Quantitative LC/MS Workflows with a Fully Integrated UHPLC-QQQ Platform

columns IonPac CS18 Cation-Exchange Column

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Application Note. Purifying common light-chain bispecific antibodies using MCSGP. Summary

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Overview. Purpose. Methods. Results

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Determination of Calcium and Magnesium in Brine

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Advantages of Polar, Reversed- Phase HPLC Columns for the Analysis of Drug Metabolites

Journal of Chemical and Pharmaceutical Research

Purification of reaction mixtures using flash chromatography.

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Reversed phase C-18 chromatography: From analytical scale to preparative scale

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Analytical Test Report

Strength in Technology HPLC COLUMNS. Product Guide

Agilent Television Power Consumption Testing. Application Note

IFC. A new perspective i l ifi ti. p p g p. Single softw FractPAL software plug-in for Agilent ChemStation. Medicinal Chemistry.

Application Note. Determination of Amino acids by UHPLC with automated OPA- Derivatization by the Autosampler. Summary. Fig. 1.

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Environmental Analysis Using the ICS-3000 Ion Chromatography System

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Dionex IonPac AS19-4µm

Aspects of industrial purification of peptides using large-scale chromatography. Lars Andersson and Jonas Persson

SUCRALOSE. White to off-white, practically odourless crystalline powder

Smartline HPLC by design

Ultra High Performance Liquid Chromatograph. Nexera XR C196-E082B

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

AppsLab Library. A new online platform for Thermo Fisher Scientific applications and methods. Susanne Kramer, Ph.D. Product Manager AppsLab Library

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Appendix 5 Overview of requirements in English

Delivering a. New Era. of Technical. Support Services

Monoclonal Antibody Characterization Achieving Higher Throughput and Productivity

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

ERDOSTEINE - MONOGRAPH.

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

[ Care and Use Manual ]

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

for IonSwift TM MONOLITH ANION CONCENTRATOR (MAC)

Oasis HLB Cartridges and 96-Well Plates

Transcription:

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control Dr. Frank Steiner, 1 Andreas Brunner, 1 Fraser McLeod, 1 Dr. Sergey Galushko 2 1 Dionex Corporation, Germering, Germany; 2 ChromSword, Mühltal, Germany Introduction HPLC method development is still considered to be one of the crucial bottlenecks that impede productivity in analytical laboratories. Due to the variety of parameters such as stationary phase, ph and composition of mobile phase, and column temperature, finding the optimum conditions for the analysis normally requires substantial efforts both in experimental work and result evaluation. This presentation describes an integrated solution based on ChromSwordAuto and UltiMate 3000 RSLC instrumentation. This system provides fully automatic method development and optimization, starting with a small number of initial experiments but exploring the entire design space through software intelligence to find the best analysis conditions. Screening of columns, solvents, buffers, and instrument parameters as well as fine tuning, robustness studies, and documentation are implemented in one platform. The systems are used effectively for all stages of drug discovery and drug development projects where rapid or sophisticated method development and optimization are key requirements. Thanks to the very fast separation of UHPLC, all this can be accomplished in the shortest time possible. Instrument and Method Setup A set of UHPLC columns is screened for the analysis of pharmaceutical samples on an integrated HPLC system designed for ultrafast automated method scouting. The system comprises: A pump with quaternary gradient capabilities at pressures up to 80 MPa, extended with an additional 10-position, 11-port solvent selection valve. A split-loop (flow-through) autosampler with very short cycle times. A powerful diode array detector with data collection rates up to 100 Hz, providing compound identification based on UV-vis spectra. Two high-pressure 6-position, 7-port column selection valves integrated in the thermostatted column compartment for maximum scouting flexibility. Intelligent software for easy parameter permutation, fully automated system control, and automated identification of the optimum chromatograms from the large data set.

The following method parameters were used during the method scouting study: Flow rate 1.0 ml/min Gradient 15% 95%B in 15 min, 2.5 min equilibration (for screening) UV detection at 254 nm, 3D field from 210 nm to 360 nm Injection volume 10 µl Column compartment at 30 C (temperature optimization not considered) A1 A6 A2 A3 A4 A5 A7 A8 A9 A10 Column Compartment Column 1 Column 2 Column 3 DAD Detector B C D Pump Autosampler Column 4 Column 5 Column 6 25882 Figure 1. System configuration for automated method scouting with optional 10-position solvent selection valve. Aqueous Mobile Phases (A) Organic Modifiers (B) Columns (100 x 4.6 mm) 1% v/v TFA, ph 2.0 acetonitrile Luna C18 10 mm potassium phosphate, ph 3.0 methanol Luna Phenyl-Hexyl 10 mm ammonium acetate, ph 5.0 Inertsil ODS 3 5 mm potassium phosphate, ph 6.5 Aqua C18 5 mm sodium borate, ph 9.5 Luna CN Software Setup The UltiMate 3000 Method Scouting System is controlled using the Chromeleon Chromatography Data System (CDS) software. In this solution, this setup is combined with the ChromSwordAuto chromatographic method development software. The package supports rapid development of separation methods in HPLC with a minimum number of experiments. The combination of ChromSwordAuto, Chromeleon, and the UltiMate 3000 Method Scouting System provides a system capable of developing completely new methods or optimizing existing ones quickly and in a fully automated way. 2 Automated Optimization Method of the Development Separation Power Utilizing of 1-D Software-Based Nano LC Analysis Optimization of Proteomics and Samples Direct Instrument Control

The Method Development Process Figure 2 shows the steps usually involved in method development. ChromSwordAuto can utilize information about the physico-chemical characteristics of the analytes to preselect a suitable subset of mobile and stationary phases. In an initial screening step, the software runs a set of experiments using different combinations of mobile and stationary phases to find a column with suitable selectivity and initial elution composition. In the subsequent fine method optimization, method conditions are varied further in order to find the best conditions possible. Finally, robustness testing can be performed to verify the method conditions to be robust against typical parameter changes in everyday use. Application Example: Separation of Six Analogs In this pharmaceutical application, six analogs with different substituents (see core structure in Figure 3) had to be separated by either isocratic or gradient elution. Each of them contained both primary amine and carboxylic acid functional groups, as well as degradation impurities in the form of cyclic amides. pka values varied from 5 9 depending on substituents. The application and the automated optimization have been published formerly. 1 N + H 3 COOH N + H 3 COO - collect information of physico-chemical characteristics (literature research) screening of initial conditions (columns/solvents) columns with various selectivity (separation) organic modifiers with different solvent strengths (elution) off-line computer simulation with known structure aqueous buffers with different ph range (peak shape) R 2 NH 2 COOH R 2 Degradation NH O selection of conditions for fine method development/optimization N + H 3 COO - + Water fine method development - optimization robustness testing evaluation of results evaluation of results documentation: method development report documentation: method development report R 2 Figure 3. Core structure of pharmaceutical compounds, acid-base equilibria, and ring formation (analogs differ in and R 2 ). 27474 27473 Figure 2. The method development process. 3

Results and Discussion Based on the combination of the conditions described in the Instrument and Method Setup section, a fully automated experimental screening is accomplished first. The chromatographic results for all five columns combined with either 10 mm phosphate buffer at ph 3.0 or 5 mm phosphate buffer at ph 6.5, running in either a CH 3 CN or a CH OH gradient are compared (20 different conditions) in Figure 4. From these results, the following conclusions can be drawn: 1. Although the bonding chemistry of the columns used differs widely, the influence of the mobile phase on selectivity and peak shape clearly prevails. 2. The difference between CH 3 CN and CH 3 OH is much more pronounced at ph 3.0 than at ph 6.5. 3. Peak shapes are generally better with CH 3 CN, but the difference is insignificant on some columns at ph 6.5. 4. Column Z (Luna Phenyl-Hexyl) shows the best peak shapes relative to its peers. 5. Based on the best compromise between overall peak resolution and peak shape, the combination of column Z with ph 6.5 and CH 3 OH is the most promising approach for the given separation challenge. In Figure 5, the applied workflow is depicted and the most promising conditions are highlighted. Based on this intermediate result, the setup automatically runs a fine optimization under consideration of frame conditions set by the operator. The fine optimization is performed for both an isocratic and a gradient method. It is based on Column Z running at ph 6.5 with methanol as organic modifier. screening of initial conditions (columns/solvents) s) columns: various selectivity (100 mm 4.6 mm, 5 µm) Luna C18 Luna Phenyl-Hexyl Inertsil ODS 3 Aqua C18 Luna CN fine method development - optimization columns with various selectivity (separation) organic modifiers: different solvent strength acetonitrile methanol evaluation of results organic modifiers with different solvent strengths (elution) documentation: method development report aqueous buffers with different ph range (peak shape) aqueous buffers: range of different ph 1% v/v TEA, ph 2.0 10 mm potassium phosphate, ph 3.0 10 mm ammonium acetate, ph 5.0 5 mm potassium phosphate, ph 6.5 selection of initial conditions for fine method development/optimization 27476 Figure 5. Graphic representation of the applied workflow and the intermediate results from the initial screening phase (best conditions marked in red with asterisk). Figure 4. Chromatographic results of screening phase with generic gradients. Combination of five different columns with two different buffers and two different organic modifiers (time axes are not to scale). 27475 4 Automated Optimization Method of the Development Separation Power Utilizing of 1-D Software-Based Nano LC Analysis Optimization of Proteomics and Samples Direct Instrument Control

Figure 6 shows the best buffer/methanol composition as an outcome of the fine optimization as well as the optimal gradient profile together with the resulting chromatograms for comparison. It can be seen that general conditions reveal excellent peak shapes for all six analytes. Both the isocratic and the gradient method fulfill the general requirements for routine analysis. This can be further consolidated in a fully automated robustness testing (not described in this poster). The gradient method separates the six compounds faster, but given the required re-equilibration, run times are similar. Overall the gradient method is superior with respect to the equal distribution of the peaks along the chromatogram (resolution). 0.8 0.7 0.6 0.5 AU 0.4 0.3 0.2 0.1 0.0 A) FINAL ISOCRATIC METHOD Time ( min) %A %B Flow (ml/min ) 0.0 46 54 1.0 14.0 46 54 1.0 0 2 4 6 8 10 12 14 16 Minutes A ChromSwordAuto-based, fully automated optimization has also been run for an impurity profiling method with an active pharmaceutical ingredient (API) of the described compound class. The resulting method had to be MS-compatible which limited the selection of applicable buffers. Figure 7A) shows the result with the fine optimized gradient profile run in an LC-MS method. The same sample was given to an experienced chromatographer for manual method development. The result of the manual development is depicted in Figure 7B). One can see that the impurity eluting closely in front of the API peak is not resolved with the manually developed method. The automated development required 24 h for the screening and 24 h for the fine optimization, while the manual development lasted close to eight working days and resulted in an inferior method. Abundance * 1E6 2.75 2.50 2.00 1.50 1.00 0.50 0.00 A) Impurity 0 2 4 6 8 10 12 14 16 18 100 90 80 70 60 50 40 30 20 0 Conc B,% 0.8 0.7 0.6 AU 0.5 0.4 0.3 0.2 0.1 B) FINAL GRADIENT METHOD Time (min) %A %B Flow (ml/min) 0.0 92 8 1.0 0.0 0 1 2 3 4 5 6 7 8 9 10 11 Minutes 7.7 30 70 1.0 12.0 30 70 1.0 27477 Figure 6. Results of automated fine optimization for A) isocratic, and B) gradient mode. Abundance * 1E6 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 0.0 B) 0 2 4 6 8 10 12 14 Minutes 100 90 80 70 60 50 40 30 20 10 0 Figure 7. Comparison of A) ChromSwordAuto-based, and B) manually developed method for real life sample impurity profiling. Conc B,% 27478 5

Conclusions The automated solution based on a combination of ChromSwordAuto software, Chromeleon software, and the UltiMate 3000 Method Scouting system can significantly reduce the labor and time spent for method development and fine optimization. References 1. Hewitt, E.F.; Lukulay, P.; Galushko, S. Implementation of a Rapid and Automated High Performance Liquid Chromatography Method Development Strategy for Pharmaceutical Drug Candidates. J. Chromatogr., A 2006, 1107, 79 87. Even for challenging applications, the results of the automated fine optimization are available within 48 h, and the resulting method is superior compared to methods developed manually in close to eight working days. If required, robustness testing and kinetical method optimization (flow/gradient volume) are also available with the described setup. ChromSwordAuto is a registered trademark of Dr. Sergey Galushko Software Entwicklung. Aqua is a trademark and Luna is a registered trademark of Phenomenex, Inc. Inertsil is a registered trademark of GL Sciences, Inc. Chromeleon and UltiMate are registered trademarks of Dionex Corporation. Passion. Power. Productivity. Dionex Corporation 1228 Titan Way P.O. Box 3603 north America U.S./Canada (847) 295-7500 Europe Austria (43) 1 616 51 25 Benelux (31) 20 683 9768; (32) 3 353 4294 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 6 Automated Optimization Method of the Development Separation Power Utilizing of 1-D Software-Based Nano LC Analysis Optimization of Proteomics and Samples Sunnyvale, CA Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Taiwan (886) 2 8751 6655 south America 94088-3603 Direct Instrument Control Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Brazil (55) 11 3731 5140 (408) 737-0700 www.dionex.com Asia Pacific Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190 LPN 2553-01 7/10 2010 Dionex Corporation