Introduction to Fourier Transform Infrared Spectroscopy

Similar documents
Introduction to Fourier Transform Infrared Spectrometry

FTIR Instrumentation

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data

Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs

Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID

Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping

Thermo Scientific LIFECYCLE Asset & Service Management for Healthcare

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

Thermo Scientific ConFlo IV Universal Interface

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

Thermo Scientific Qtegra Intelligent Scientific Data Solution Software. Delivering quality. Driving productivity

Thermo Scientific ExactFinder Software

LIFECYCLE. Thermo Scientific. Enterprise Solutions

Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA

UV-Vis spectrophotometers

Thermo Scientific Dionex Chromeleon 7 Chromatography Data System Software

Experiment 5. Lasers and laser mode structure

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

m/z

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

FT-NIR for Online Analysis in Polyol Production

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

FT-IR Microspectroscopy in Forensic and Crime Lab Analysis

WATERS QUANTITATIVE ANALYSIS solutions

MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC

Measurement & Analytics Measurement made easy

Signal to Noise Instrumental Excel Assignment

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

How To Control A Record System

Keysight Technologies Using Fine Resolution to Improve Thermal Images. Application Note

Back to Basics Fundamentals of Polymer Analysis

DETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY

Getting Answers at the Source Requires a New Way of Thinking About Software

Challenges in DWDM System Spectral Analysis By Laurent Begin and Jim Nerschook

Preview of Period 3: Electromagnetic Waves Radiant Energy II

13C NMR Spectroscopy

Synthetic Sensing: Proximity / Distance Sensors

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

Pesticide Analysis by Mass Spectrometry

Third Party Instrument Control with Chromeleon Chromatography Data System (CDS) Software

Title: Standard Operating Procedure for Thermo-Miran SapphIRe portable gas analyzer

Experiment #5: Qualitative Absorption Spectroscopy

Raman spectroscopy Lecture

Module 13 : Measurements on Fiber Optic Systems

White Paper. 10 Reasons to Choose Image-based Barcode Readers

DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Symmetric Stretch: allows molecule to move through space

Chemical Measurement Solutions

Simplicity. Reliability. Performance

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Preface Light Microscopy X-ray Diffraction Methods

Infrared Spectroscopy 紅 外 線 光 譜 儀

Agilent 4338B Milliohm Meter

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

Smart. Scalable. Simple.

Agilent Automotive Power Window Regulator Testing. Application Note

An Introduction to the MTG-IRS Mission

Using the Spectrophotometer

Agilent 4339B/4349B High Resistance Meters

How does a venture capitalist appraise investment opportunities?

Thermo Scientific SIEVE Software for Differential Expression Analysis

Acousto-optic modulator

Realization of a UV fisheye hyperspectral camera

Keysight Technologies How to Choose your MAC Lever. Technical Overview

Targus Wireless Remote Presenter USER S GUIDE. Making Your Mobile Life Easier. Making Your Mobile Life Easier.

How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer

It has long been a goal to achieve higher spatial resolution in optical imaging and

INFRARED PARTS MANUAL

QUANTITATIVE DETERMINATION OF COMMON TYPES OF ASBESTOS BY DIFFUSE REFLECTANCE FTIR

Time-Correlated Multi-domain RF Analysis with the MSO70000 Series Oscilloscope and SignalVu Software

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

10 Reasons to Choose Image-based ID Readers. White Paper

Product Range Overview

Make the invisible visible! SENSORS WITH EXCELLENT BACKGROUND SUPPRESSION

Ultraviolet Spectroscopy

ORIEL. FT-IR Spectroscopy SECTION FOUR FEATURES. Glossary of Terms Introduction to FT-IR Spectroscopy

h e l p s y o u C O N T R O L

[ Care and Use Manual ]

The photoionization detector (PID) utilizes ultraviolet

What is Eluent Generation?

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System

Qualitative NIR Analysis for Ingredients in the Baking Industry

INTRODUCTION. The principles of which are to:

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher)

IR READY TOGO. Spectrum Two FT-IR Spectroscopy

ProteinChip Energy Absorbing Molecules (EAM)

Spectrum 10 Spectroscopy Software

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Transcription:

Introduction to Fourier Transform Infrared Spectroscopy

Introduction What is FT-IR? FT-IR stands for Fourier transform infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR radiation is passed through a sample. Some of the infrared radiation is absorbed by the sample and some of it is passed through (transmitted). The resulting spectrum represents the molecular absorption and transmission, creating a molecular fingerprint of the sample. Like a fingerprint no two unique molecular structures produce the same infrared spectrum. This makes infrared spectroscopy useful for several types of analysis. So, what Information can FT-IR Provide? It can identify unknown materials It can determine the quality or consistency of a sample It can determine the amount of components in a mixture This booklet is an introduction to the concepts behind FT-IR spectroscopy. It covers both the basic theory of FT-IR and how it works as well as discussing some the practical aspects of FT-IR use. We hope that it gives you a good understanding of the importance and usefulness of this powerful technique.

Theory of FT-IR Why Infrared Spectroscopy? Infrared spectroscopy has been a workhorse technique for materials analysis in the laboratory for over seventy years. An infrared spectrum represents a fingerprint of a sample with absorption peaks which correspond to the frequencies of vibrations between the bonds of the atoms making up the material. Because each different material is a unique combination of atoms, no two compounds produce the exact same infrared spectrum. Therefore, infrared spectroscopy can result in a positive identification (qualitative analysis) of every different kind of material. In addition, the size of the peaks in the spectrum is a direct indication of the amount of material present. With modern software algorithms, infrared is an excellent tool for quantitative analysis. Older Technology The original infrared instruments were of the dispersive type. These instruments separated the individual frequencies of energy emitted from the infrared source. This was accomplished by the use of a prism or grating. An infrared prism works exactly the same as a visible prism which separates visible light into its colors (frequencies). A grating is a more modern dispersive element which better separates the frequencies of infrared energy. The detector measures the amount of energy at each frequency which has passed through the sample. This results in a spectrum which is a plot of intensity vs. frequency. Fourier transform infrared spectroscopy is preferred over dispersive or filter methods of infrared spectral analysis for several reasons: It is a non-destructive technique It provides a precise measurement method which requires no external calibration It can increase speed, collecting a scan every second It can increase sensitivity one second scans can be co-added together to ratio out random noise It has greater optical throughput It is mechanically simple with only one moving part

Why FT-IR? Fourier transform infrared (FT-IR) spectrometry was developed in order to overcome the limitations encountered with dispersive instruments. The main difficulty was the slow scanning process. A method for measuring all of the infrared frequencies simultaneously, rather than individually, was needed. A solution was developed which employed a very simple optical device called an interferometer. The interferometer produces a unique type of signal which has all of the infrared frequencies encoded into it. The signal can be measured very quickly, usually on the order of one second or so. Thus, the time element per sample is reduced to a matter of a few seconds rather than several minutes. Most interferometers employ a beamsplitter which takes the incoming infrared beam and divides it into two optical beams. One beam reflects off of a flat mirror which is fixed in place. The other beam reflects off of a flat mirror which is on a mechanism which allows this mirror to move a very short distance (typically a few millimeters) away from the beamsplitter. The two beams reflect off of their respective mirrors and are recombined when they meet back at the beamsplitter. Because the path that one beam travels is a fixed length and the other is constantly changing as its mirror moves, the signal which exits the interferometer is the result of these two beams interfering with each other. The resulting signal is called an interferogram which has the unique property that every data point (a function of the moving mirror position) which makes up the signal has information about every infrared frequency which comes from the source. This means that as the interferogram is measured, all frequencies are being measured simultaneously. Thus, the use of the interferometer results in extremely fast measurements. Because the analyst requires a frequency spectrum (a plot of the intensity at each individual frequency) in order to make an identification, the measured interferogram signal can not be interpreted directly. A means of decoding the individual frequencies is required. This can be accomplished via a well-known mathematical technique called the Fourier transformation. This transformation is performed by the computer which then presents the user with the desired spectral information for analysis.

The Sample Analysis Process The normal instrumental process is as follows: 1. The Source: Infrared energy is emitted from a glowing black-body source. This beam passes through an aperture which controls the amount of energy presented to the sample (and, ultimately, to the detector). 2. The Interferometer: The beam enters the interferometer where the spectral encoding takes place. The resulting interferogram signal then exits the interferometer. The interferometer uses a reference laser for precise wavelength calibration, mirror position control and data acquisition timing. 3. The Sample: The beam enters the sample compartment where it is transmitted through or reflected off of the surface of the sample, depending on the type of analysis being accomplished. This is where specific frequencies of energy, which are uniquely characteristic of the sample, are absorbed. 4. The Detector: The beam finally passes to the detector for final measurement. The detectors used are specially designed to measure the special interferogram signal. 5. The Computer: The measured signal is digitized and sent to the computer where the Fourier transformation takes place. The final infrared spectrum is then presented to the user for interpretation and any further manipulation. Because there needs to be a relative scale for the absorption intensity, a background spectrum must also be measured. This is normally a measurement with no sample in the beam. This can be compared to the measurement with the sample in the beam to determine the percent transmittance. This technique results in a spectrum which has all of the instrumental characteristics removed. Thus, all spectral features which are present are strictly due to the sample. A single background measurement can be used for many sample measurements because this spectrum is characteristic of the instrument itself.

A Simple Spectrometer Layout

Advantages of FT-IR Some of the major advantages of FT-IR over the dispersive technique include: Speed: Because all of the frequencies are measured simultaneously, most measurements by FT-IR are made in a matter of seconds rather than several minutes. This is sometimes referred to as the Felgett Advantage. Sensitivity: Sensitivity is dramatically improved with FT-IR for many reasons. The detectors employed are much more sensitive, the optical throughput is much higher (referred to as the Jacquinot Advantage) which results in much lower noise levels, and the fast scans enable the coaddition of several scans in order to reduce the random measurement noise to any desired level (referred to as signal averaging). Mechanical Simplicity: The moving mirror in the interferometer is the only continuously moving part in the instrument. Thus, there is very little possibility of mechanical breakdown. Internally Calibrated: These instruments employ a HeNe reference laser as an internal wavelength calibration standard (referred to as the Connes Advantage). These instruments are self-calibrating and never need to be calibrated by the user. These advantages, along with several others, make measurements made by FT-IR extremely accurate and reproducible. Thus, it a very reliable technique for positive identification of virtually any sample. The sensitivity benefits enable identification of even the smallest of contaminants. This makes FT-IR an invaluable tool for quality control or quality assurance applications whether it be batch-to-batch comparisons to quality standards or analysis of an unknown contaminant. In addition, the sensitivity and accuracy of FT-IR detectors, along with a wide variety of software algorithms, have dramatically increased the practical use of infrared for quantitative analysis. Quantitative methods can be easily developed and calibrated and can be incorporated into simple procedures for routine analysis. Thus, the Fourier transform infrared (FT-IR) technique has brought significant practical advantages to infrared spectroscopy. It has made possible the development of many new sampling techniques which were designed to tackle challenging problems which were impossible by older technology. It has made the use of infrared analysis virtually limitless.

The Thermo Scientific Family of FT-IR Technology We are the worldwide leader in FT-IR instrumentation, delivering powerful solutions to scientists since the 1970 s. Nicolet is 50 Easily upgradeable Superior optical throughput High resolution Multiple spectral ranges Rapid scanning/fast kinetics Nicolet is10 Highest confidence in laboratory based Pass/Fail product qualification Comprehensive set of regulatory compliance and performance verification tools for passing audits Fast implementation, low cost of ownership and configurability for multiple QC verifications Nicolet is5 Easy sampling Easy purge Easy interface Easy identification Easy presentation Easy to use compact system Nicolet in 10 Infrared Microscope The original integrated FT-IR microscope Fully automated microscope controls Flexible micro-atr options Microscope-specific Software Ultra-fast imaging options Nicolet Continuµm Infrared Microscope Highest spatial resolving power Simultaneous view and collect Compensating IR Objectives Extended range detectors Digital video and binocular viewing Wide range of visible light techniques Africa +27 11 822 4120 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland/Norway/Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS +43 1 333 50 34 0 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 www.thermoscientific.com 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. BR50555_E 05/13M Sample Compartment Accessories Samples as small as 50 µm using ATR Fibers, pellets, paper, powdered samples Plug and play design for quick analysis