EXPERIMENT 6. Molecular Fluorescence Spectroscopy: Quinine Assay

Similar documents
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Determination of the Mass Percentage of Copper in a Penny. Introduction

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

Chem 131A: Absorbance of Riboflavin

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy

Reaction Stoichiometry and the Formation of a Metal Ion Complex

The Determination of an Equilibrium Constant

Measuring Manganese Concentration Using Spectrophotometry

Calibration of Volumetric Glassware

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

COMMON LABORATORY APPARATUS

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

A Beer s Law Experiment

Factors Affecting Enzyme Activity

experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Solubility Product Constants

Determination of a Chemical Formula

Using the Spectrophotometer

Colorimetric Determination of Iron in Vitamin Tablets

FAJANS DETERMINATION OF CHLORIDE

Physical Properties of a Pure Substance, Water

ENZYME KINETICS ENZYME-SUBSTRATE PRODUCTS

Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow

CHEM 161: Beer s Law and Analysis of a Sports Drink

Phenolphthalein-NaOH Kinetics

SYNTHESIS AND ANALYSIS OF A COORDINATION COMPOUND OF COPPER

Evaluation copy. Titration of a Diprotic Acid: Identifying an Unknown. Computer

Chemistry 2351: Inorganic Chemistry I Laboratory Manual

INTRODUCTION TO SPECTROMETRY

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

Upon completion of this lab, the student will be able to:

Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

Calcium Analysis by EDTA Titration

Calibration of Glassware

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: )

Chem 100 Lab Experiment #9 - ACID/BASE INDICATORS

Acetic Acid Content of Vinegar: An Acid-Base Titration E10-1

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

Measurement and Calibration

PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE

Lab #11: Determination of a Chemical Equilibrium Constant

Laboratory 5: Properties of Enzymes

Quantifying Bacterial Concentration using a Calibrated Growth Curve

Analysis of Vitamin C Using Iodine. Introduction

Experiment 17: Potentiometric Titration

To see how this data can be used, follow the titration of hydrofluoric acid against sodium hydroxide below. HF (aq) + NaOH (aq) H2O (l) + NaF (aq)

Esterification Method 1 Method to 2800 mg/l (as acetic acid) Reagent Solution

Paper Chromatography: Separation and Identification of Five Metal Cations

Volumetric Analysis. Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115

Acid Base Titrations

Quaternary Ammonium Compounds DOC

Pre-Lab Notebook Content: Your notebook should include the title, date, purpose, procedure; data tables.

EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet. by Professor David Cash. September, 2008

Ion Exchange Determination of Na+ by Displacement and Zn 2+ Using Preconcentration. Reading: Harris pp , 699

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Colorimetric Determination of Iron in Vitamin Tablets

Coordination Compounds with Copper (II) Prelab (Week 2)

Experiment 1: Measurement and Density

2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY

To determine the mass of iron in one adult dose of either a ferrous sulfate or. ferrous gluconate iron supplement using a colorimetric technique.

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

18 Conductometric Titration

The Analytical Balance

Determination of Aspirin using Back Titration

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

PROTEINS (LOWRY) PROTOCOL

METHOD DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY

Synthesis of Aspirin and Oil of Wintergreen

The Measurement of Sensitivity in Fluorescence Spectroscopy

Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater

GA/7 Potentiometric Titration

Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation

Carolina s Solution Preparation Manual

Amino Acids, Peptides, and Proteins

Concentrations and Dilutions of Food Dyes

EXPERIMENT 7 Electrochemical Cells: A Discovery Exercise 1. Introduction. Discussion

Separation by Solvent Extraction

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

SOLID STATE CHEMISTRY - SURFACE ADSORPTION

8.9 - Flash Column Chromatography Guide

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Determination of calcium by Standardized EDTA Solution

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid

SODIUM CARBOXYMETHYL CELLULOSE

Phosphates (ortho- and total)

Solutions and Dilutions

Transcription:

EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay UNKNOWN Submit a clean, labeled 500-mL volumetric flask to the instructor so that your unknown quinine solution may be issued. Your name, section number, and your locker number should be written legibly on this flask. The flask does not need to be dry on the inside, but needs to have been rinsed with distilled water after it has been cleaned. The flask must be turned in at least 1 lab period before you plan to do the experiment so that the Teaching Assistants will have time to prepare the unknown. Each student will have his or her own unknown to analyze even if you are working in pairs. BACKGROUND Quinine (C 20 H 24 N 2 O 2, 324.43 g/mol) is an alkaloid extracted from the bark of the cinchona tree. It has been used for many years as an antimalarial agent. Although it does not cure malaria, it is effective in alleviating the symptoms of malarial attacks. The usual medicinal form is quinine dihydrochloride or quinine sulfate dihydrate, (C 20 H 24 N 2 O 2 ) 2 H 2SO 4 2H 2O, 782.97 g/mol. Quinine is a very strongly fluorescing compound, especially in dilute acid solution, and thus can be detected in very trace amounts. In 0.05 M H 2 SO 4, quinine has two analytically useful excitation wavelengths: λ ex = 250 and 350 nm. Regardless of which excitation wavelength is used, the wavelength of maximum fluorescence emission intensity, λ em or λ fl, is 450 nm. The basis for quantitation is that the intensity of fluorescence emission in very dilute solutions is directly proportional to the concentration of quinine if the intensity of the excitation source and other experimental factors are kept constant. Because the absolute emission intensity can vary considerably with small differences in experimental conditions, a calibration curve is prepared by measuring the fluorescence-emission intensity of accurately known quinine standard solutions. INSTRUMENTATION Turner Quantech Digital Filter Fluorometer, Model FM 109525 The instrument used in this experiment uses glass filters to select wavelength ranges appropriate for quinine, rather than some type of expensive grating or prism monochromator to isolate narrow excitation and fluorescence wavelengths. The latter type may typically have a bandpass of 0.1 or 1 nm. A bandpass is defined as the width at half-height of the maximum intensity of light that passes through the filter. It is also called the bandwidth. CHE 226 Analytical Chemistry Laboratory 34 Quinine Fluorescence

A monochromator-based instrument is much better for obtaining spectra, particularly those with fine structure, but a filter-based unit will often be better for routine quantitative analysis. The much larger bandpasses permit greater fractions of the excitation and fluorescent light to excite the sample and to reach the detector, respectively. This greatly increases the sensitivity of the instrument and can thus usually lower typical detection by an order of magnitude or more. Filter fluorometers are also much less expensive. Excitation Source: A 5-watt quartz-halogen lamp, which emits intense broadband radiation from 340 nm to 750 nm. Excitation Wavelength Filter: A narrow-band 360-nm filter with a bandpass of 40 nm. Emission Wavelength Filter: A sharp cut-in, long-wavelength-pass filter which transmits essentially all light with λ > 415 nm. Detector: Photomultiplier tube, model 931B PMT. Detection Limit. Stated in the manufacturer s literature as 30 ppt quinine sulfate, which is 30 ng/l or about 9 x 10-11 M. Turn the instrument on at least 15 minutes before using to allow it to warm up and stabilize. The ON-OFF switch is on the back panel near the power cord. When turned on, the instrument runs a countdown timer during which it undergoes self tests. PREPARATION OF SOLUTIONS Preparation of Stock Sulfuric Acid Solutions The two stock H 2 SO 4 solutions should have already been prepared for you by the Teaching Assistants. However, if you use up all of the 0.05 M solution while doing your experiment, prepare another 2-L batch for those that follow. It is very straightforward. 1 M H 2 SO 4. Slowly and carefully add 56 ml conc. H 2 SO 4 to about 500 ml distilled water in a 1-L beaker with stirring. This solution should have been already prepared for your use. 0.05 M H 2 SO 4. With a graduated cylinder, add 100 ml of 1 M H 2 SO 4 with a graduated cylinder to about 500 ml of distilled water in the screw-capped acid reagent jug labeled for this solution. Mix and dilute to the 2.0-L mark on the bottle and mix thoroughly. This solution should already have been prepared for your use. If you use all of or most of what is there, prepare a new batch for the students who follow you. Preparation of Quinine Stock Solution, 1000 ppm 1. Carefully weigh exactly 0.1207 g of quinine sulfate dihydrate onto a folded glassine weighing paper or into a small plastic weighing boat, and transfer this quantitatively into a 100-mL volumetric flask. A few squirts of distilled water from a wash bottle should help to wash the solid material from the weighing boat and the neck of the flask. CHE 226 Analytical Chemistry Laboratory 35 Quinine Fluorescence

2. Pipet 5.00 ml of 1 M H 2 SO 4 (located in hood #2) into the flask. Carefully dissolve all the quinine in this sulfuric acid solution by swirling before diluting to volume. This is critically important. 3. Carefully dilute to volume with distilled water and mix thoroughly. Preparation of Intermediate Quinine Stock Solution, 10.0 ppm 1. Pipet 5.00 ml of the 1000-ppm solution into a 500-mL volumetric flask. 2. Add 25.0 ml of 1 M H 2 SO 4, dilute carefully to volume with distilled water, and mix thoroughly. Preparation of Quinine Standard Solutions 1. Using volumetric transfer pipets and/or a 10-mL graduated pipet, add 1.00, 3.00, 5.00, 7.50, and 10.00 ml of the 10-ppm intermediate stock solution into five properly labeled 100-mL volumetric flasks. This will result in standard solutions of 0.1, 0.3, 0.5, 0.75, and 1.0 ppm. 2. Carefully dilute to volume with 0.05 M H 2 SO 4. 3. 0.05 M H 2 SO 4 is used as the blank. Preparation of Quinine Unknown 1. Your unknown solution is obtained from the teaching assistants in a 500-mL volumetric flask. Add 25.0 ml of 1 M H 2 SO 4 to the flask. 2. Dilute to volume with distilled water and mix thoroughly. PROCEDURE Turn on the Quantec fluorometer at least 15-20 minutes before making measurements in order to let it warm up and stabilize. Measurement of Emission Intensities 1. Carefully fill separate, clean, plastic fluorescence cuvettes (these have 4 clear sides) about ¾ full with the blank (0.05 M H 2 SO 4 ), each of the five standards, and each unknown sample. Do not touch the optical surfaces with your fingers. Instead, handle with KimWipes. Be sure to wipe any smudges off the optical surfaces of the cuvettes. 2. On the main menu of the fluorometer, press ENTER. CHE 226 Analytical Chemistry Laboratory 36 Quinine Fluorescence

3. Use the right cursor ( ) to select Quinine. Press RETURN. 4. The instrument asks if you wish to change the name. Select NO. 5. The instrument asks if filters are correct. Select YES. 6. The instrument asks if you would like to you as standard curve from memory. Select NO. 7. The instrument asks for the number of points for the calibration curve. Use the up-arrow ( ) to increase the number to 5 in order to be able to use your five standards. Press RETURN. 8. Beginning with the highest concentration standard, enter the concentration using the uparrow to change the values. Use ( ) to move the cursor to the appropriate position. Use the up-arrow to set the appropriate units (ppm). 9. The instrument asks to insert the sample. Do so and press RETURN. 10. Repeat for other standards. 11. Instrument asks the operator to insert the BLANK. Do so and press ZERO. 12. The instrument gives a Coefficient of Determination value. If this value is less than 0.90, you must pour out your standard solutions and repipet them from the stock quinine solution. If this value is greater than 0.90, continue. 13. Insert the unknown and Press RETURN. 14. Readout shows READ and provides the value. 15. Repeat the entire process at least two more times to obtain a set of three good replicate measurements: If the values for a set of standards and the unknown do not drift over time, this would mean that the raw fluorescence data for your unknown are within a range of about 3-4% relative. If the values do appear to drift, then the ratios of the fluorescence intensities of the standard to those of a nearby standard are within a range of about 3-4% relative. Comparison of Calibration Procedures This procedure will compare values that are (a) calculated by the instrument based on its internal calibration procedure with (b) those obtained based on the calibration curve(s) that you prepare from the raw fluorescence-emission data obtained above. 1. At main menu, select RAW FLUORESCENCE as the units of measure. 2. The instrument requests a value for the standard. Enter the value and RETURN. CHE 226 Analytical Chemistry Laboratory 37 Quinine Fluorescence

3. The instrument instructs you to insert the standard. USE the 1.0 ppm STANDARD. Do so and press RETURN. The instrument adjusts its internal gain. 4. The instrument asks Insert Blank Sample? Select YES, followed by RETURN. 5. Insert the blank and the instrument will take a reading. 6. Now insert each of the standard solutions and record the emission value for each. 7. Insert the unknown, and record the emission value. 8. Repeat the entire procedure at least twice more in order to obtain at least 3 good measurements for each sample. 9. Prepare calibration curves and calculate the average concentration of the unknown sample. When done with the entire experiment, rinse out all the plastic cuvettes with distilled water, shake off excess water, and return them to the experiment drawer so they can be reused. DATA ANALYSIS Prepare appropriate calibration curve(s) from your replicate sets of data. Depending on the nature of the data and any drift in the instrument, it may be best to (a) average all the net emission intensities for a particular solution and to prepare one calibration curve or (b) prepare several calibration curves from the separate sets of data, obtain several values for the concentration of quinine in your unknown, then average these values. Determine which one is better or if both give equivalent results. Use linear least squares to fit each of the three sets of data, and the average of all three, in your analysis. Report the best value for the concentration of quinine (in ppm) in your unknown and the estimated standard deviation of this best value. Note: By definition, parts-per-million (ppm) is a mass-ratio unit g analyte /g sample x 10 6 or µg/g. In dilute aqueous solutions, 1 ppm is equivalent to 1 µg/ml or 1 mg/l because the density of the solution is 1.00 g/ml. A 1000-ppm solution of quinine sulfate (1.000 g of quinine sulfate per liter of solution) would only be an 828.7 ppm of quinine itself. If you cannot weigh out exactly 0.1207 g of quinine sulfate, try to get it close, and simply calculate the correct values of the concentrations of your standards. HAZARDOUS WASTE DISPOSAL Dispose of ALL waste quinine and sulfuric acid solutions in the proper, labeled Hazardous Waste Container for this experiment, which is located in the hoods. 0.05 M H 2 SO 4 has a ph CHE 226 Analytical Chemistry Laboratory 38 Quinine Fluorescence

of about 1 and thus may NOT simply be run down the drain, even if you run large amounts of water with it. If you are unsure about the container, ASK. TEXT REFERENCE D. A. Skoog, D. M. West, F. J. Holler, and S. R. Crouch, Analytical Chemistry: An Introduction, 7th ed., Chapter 23, pp. 594-631. Revised January 24, 2005 Copyright by The Department of Chemistry,, 2005 CHE 226 Analytical Chemistry Laboratory 39 Quinine Fluorescence