The effect of electron beam sterilization on transparent polymers used in medical devices

Similar documents
You have our undivided attention.

Freedom to Create In-Store Fixtures and Point-of-Purchase Displays

Injection molding Troubleshooting guide. Eastman copolyesters

Eastar copolyester technical information

Cellulose esters and Eastman Solus 2100 performance additive in ink jet applications

CYROLITE. Sterilization Methods & Considerations. Technical information

Materials. Powering Ideas.

Makrolon Rx2530. Polycarbonate (PC) Medium viscosity Injection molding Medical devices** Short description. Delivery form.

cellulosic plastics From trees to plastic

PARALOID EXL-2691A/3691A Impact Modifier For polycarbonate and polycarbonate/abs blends

Effect of Sterilization Techniques on Polymers

Resistance of Plastics to Gamma Irradiation

Eastman plasticizers. Bulk storage guidelines. Storage equipment. Tank components and materials. Piping. Pumps. Valves. Gaskets

Information on the product

Emilie Gautriaud, M.S., Keith T. Stafford, Jennifer Adamchuk, Ph.D., Mark W. Simon, Ph.D., and Duan Li Ou, Ph.D.*

Sterile Cleanroom Management

Polyolefin Heat Shrink Tubing for Tight-Tolerance Medical Applications

SERVICES FOR. Devices and Combination Products

Double Coated Urethane Foam Tapes

Araldite AV 170. Advanced Materials. One component epoxy adhesive. Structural Adhesives. Key properties. Description. Product data.

Plexiglas Altuglas Luctor ACRYLIC RESINS. Medical Grades

DuPont Packaging & Industrial Polymers

Eastman products for food and beverage applications

Bormed TM. Solutions for Healthcare Moulding and Film Applications

Eastar copolyester Thermoforming

Electron Beam Technology for Pressure Sensitive Adhesive Applications

Araldite LY 3505* / Hardeners XB 3403* / Hardener XB * / Aradur 3405*

INNOVATIVE PLASTICS RESISTANCE+ DURABILITY Chemical Resistance Performance Testing for Healthcare Materials

Temperature C Cure time to reach hours LSS > 1MPa minutes

INNOVATIVE PLASTICS HIGH & LOW SHEAR RATE RHEOLOGY

INNOVATIVE PLASTICS ULTRASONIC WELDING

Inspired By Challenge. Technical information polymer characteristics

Corning HPFS 7979, 7980, 8655 Fused Silica. Optical Materials Product Information Specialty Materials Division

Part and tooling design. Eastman Tritan copolyester

1.1 Fill fluids. Rosemount DP Level Fill Fluid Specifications July Technical Note , Rev DB

Low Pressure Molding Solutions

AMPLITUDE AND FORCE PROFILING: STUDIES IN ULTRASONIC WELDING OF THERMOPLASTICS

Two component epoxy paste adhesive

MICRO PLASTICS MOLDED NYLON 6/6 FASTENERS MECHANICAL TEST DATA - UNC - UNF SECTION

Welding of Plastics. Amit Mukund Joshi. (B.E Mechanical, A.M.I.Prod.E)

MATERIAL SAFETY DATA SHEET Page 1 of 5

Environmental Stress Crack Resistance Of Polyethylene

Naue GmbH&Co.KG. Quality Control and. Quality Assurance. Manual. For Geomembranes

Sterilization methods and equipment Lab 1-2

9130 Kallo (Kieldrecht) Houston, TX Phone: Phone : info@monumentchemical.com Fax :

Solving Healthcare Challenges. Increasing patient safety, reducing errors and improving efficiencies with Intermec Healthcare Solutions.

Sekusept Cleaner Liquid Cleaning Agent Cleaning, Desinfecting and Drying

TRANSPARENT HIGH-BARRIER RETORTABLE FILM

Two component epoxy paste adhesive. Ideal for bonding GRP, SMC and dissimilar substrates

Radiation Curable Components and Their use in Hard, Scratch Resistant Coating Applications

July 21, Dentca, Inc. c/o Mr. Claude Berthoin Denterprise International, Inc. 110 East Granada Blvd, Suite 207 Ormond Beach, Florida 32176

Overview. Eastman Aviation Solutions academy. excellence through training.

Prototyping Process Choosing the best process for your project

Elastomeric Components for Pharmaceutical Applications - Actual Quality Trends - Claudia Petersen. -Senior Manager Biotechnology-

HOSTAFORM C 9021 POM Unfilled

Product manufacturing from zirconium oxide

Product Information Guide CALIBRE. Polycarbonate Resins. Impact Strength. Heat Resistance. Optical Clarity

RTV159. RTV157 and RTV159 High Strength Silicone Adhesive Sealants

Plasma Cleaner: Physics of Plasma

Color Coded Labeling System for Storing Chemicals in your Laboratory

Gamma Sterilisation Validation according to ISO Sterilising dose -

SAFETY DATA SHEET West Skyhawk Drive West Jordan, UT U.S.A. Phone

Optimizing Hair Care Products with ANGUS Specialty Neutralizer Technology from Dow Performance Additives

Alcoa Aluminum Powder. Your Link to Success

Medical Device Packaging and Sterilization: What you need to know to protect your product

Chlorinated polyethylene ELASLEN TM. Application for Wire and Cable

Materials for the best connections

Kemet DIAMOND PRODUCTS

Material Safety Data Sheet

The information contained in this document is intended for healthcare professionals only.

Technical Data Sheet February 2014

Lab PPE Guide Examples of Personal Protective Equipment

DUPONT PERFORMANCE POLYMERS Joint Design: A Critical Factor in Strong Bonds GENERAL GUIDELINES FOR ULTRASONIC, VIBRATION AND SPIN WELDING

How compact discs are made

Accurate Convenient Affordable Results in 24 Hours. Mail-to-Lab Spore Testing. Mail-to-Lab Spore Testing

Specialty Engineered Materials for Wire & Cable

Industrial Attachment Tapes X-Series. Value, Performance. & Quality. Simply X.

Scotch-Weld TM. Epoxy Adhesive 1838 B/A Green 1838 B/A Tan 1838-L B/A Translucent. Technical Data February, Product Description

D.E.H. 20. Product Information. Epoxy Curing Agent. Description. D.E.H. 20 Epoxy Curing Agent is a liquid aliphatic polyamine curing agent.

Thermoplastic Olefin Polymer of Amorphous Structure (COC) Cyclic Olefin Copolymer (COC)

Design for Manufacturability Rapid Prototyping Mold Making Production Injection Molding

Eastman Provista copolymers. Extrusion of tubes and profiles

Amphenol Sincere. Electronic Integration & Flexible Printed Circuits. Aerospace. Health Care. Heavy Equipment HEV. Industrial

DEMTROL High Performance Crude Oil Demulsifier Bases FPO. Setting the Standard for Emulsion Breaking in Production Operations

Fused Deposition Modeling: A Technology Evaluation

1.4 Metallic Impurities Iron must not exceed 150 ppm and copper shall not exceed 6 ppm when tested according to TAPPI T 266.

FM FILM ADHESIVE

Wire and Cable Product Guide

Scotch-Weld. Low-Odor Acrylic Adhesives DP810 DP810 NS. Technical Data June, 2004

THE BASICS OF STERILIZATION

BRILLIANT TONER SERIES TONERS FOR PRINTING INKS BSTS SERIES

EB Technology For Fast, Safe Curing of Inks

INJECTION MOLDING PROCESSING GUIDE Polymer

Ethylene Oxide (EtO or EO) The Future Of Gas Sterilization

Safety Data Sheet (SDS)

Ningbo Yinzhou Keao Prototyping & Mould Factory Services include : CNC machining prototypes,

3 Single-Coated Foam Tapes

STERILIZATION AND DISINFECTION

Amodel AS-1133 HS. polyphthalamide. Technical Data Sheet

Transcription:

The effect of electron beam sterilization on transparent s used in medical devices The Material Difference in medical applications

The effect of electron beam sterilization on transparent s used in medical devices The objective of sterilizing medical devices is to reduce the bioburden to a safe level with minimal effects on physical and optical properties. The most widely used sterilization methods in the medical industry today are electron beam (e-beam), gamma radiation, ethylene oxide (EtO), autoclave, and low-temperature hydrogen peroxide gas plasma. Recent technology advances have improved operating efficiency, making e-beam radiation a safe, efficient, reliable source of energy that is gaining in popularity. E-beam radiation typically costs less compared to gamma radiation mainly due to higher dose rates which reduce the time of exposure at the same target dose. The shorter exposure time to e-beam radiation minimizes the oxidation reactions that can occur at the surface of s and consequently has less effect on the physical and optical properties of most resins when compared to gamma radiation. Eastman conducted studies to determine the effects of e-beam radiation on the physical and optical properties of Eastman specialty plastics and various other transparent s used in the medical industry. Specific products tested are shown in Table 1. Optical properties Color measurements were made 3, 7, 14, and 42 days after e-beam radiation using a HunterLab UltraScan Sphere 8000 and the CIE L*, a*, b* color scale. Samples were concealed in dark enclosures throughout testing and were only exposed to light for measurement of color. The b* value measures the blue to yellow scale which Figure 1 shows as the most prevalent color change after exposure to ionization energy. Table 1 Formula and resin type used in e-beam testing Formula MN006 Eastman Tritan Tenite propionate 360A40012 Lipid resistant PC Medical grade acrylic Resin type cellulose acetate propionate (CAP) polycarbonate transparent ABS acrylic (PMMA) 2

The Material Difference in medical applications Figure 1 Photographs of molded resins before and after e-beam radiation at 50 kgy Time after sterilization Eastman Tritan MN006 a Lipid resistant PC Medical grade acrylic 3 days 7 days 14 days 42 days Unexposed a Next-generation DN004 3

Figure 2 Eastman specialty plastics and competitive resins after e-beam radiation at 50 kgy b* color measurements 20 15 10 b* v alue 5 0 5 10 Eastman Tritan MN006 a Lipid resistant PC Medical grade acrylic 3 7 14 42 days after e-beam radiation Unexposed a Next generation DN004 Of the resins tested, s and MN006 showed the smallest change in b* values immediately after sterilization compared to the competitive resins. 4

Figure 3 Change in b* 42 days after e-beam radiation 12 10 Change in b* 8 6 4 2 0 Eastman Tritan MN006 Tenite propionate 360A40012 Lipid resistant PC Medical grade acrylic Change in b* at 25 kgy Change in b* at 50 kgy Figure 3 shows the difference in values between the unexposed samples and the sterilized samples 42 days after exposure at 25 and 50 kgy. resins show the smallest color shift 42 days after exposure compared to the other resins tested and were very close to their initial color, especially at the lower dosage. Physical properties Physical properties were measured before and after e-beam sterilization at 25 and 50 kgy. No physical property degradation was noted for any resin tested. Tables 2 and 3 show the test results. 5

Table 2 Physical properties before and after e-beam radiation of Eastman specialty plastics and competitive resins Tensile yield strength, MPa Tensile break strain, % Flatwise impact @ 23 C; total energy, J E-beam energy, Mrad Eastman Tritan MN006 Tenite propionate 360A40012 Lipid resistant PC Unexposed 43.9 59 52 47 51 28 64 49 44 25 NA 61 52 47 51 28 63 48 44 50 42.9 62 54 47 52 28 62 48 45 Unexposed 196.2 117 71 312 271 37 146 14 6.2 25 NA 73 64 330 272 31 134 15 7.8 50 184.3 99 38 340 288 23 147 17 8.7 Unexposed NA 66 59 62 65 36 89 4.5 14 25 NA 63 57 62 66 36 95 3.8 11 50 NA 63 57 64 61 36 85 3.2 17 Medical grade acrylic Molecular weight was determined by Gel Permeation Chromatography for Eastman specialty plastics before and after e-beam radiation. Table 3 shows molecular weight of the sterilized resins as a percentage of the original molecular weight of unexposed resins. Table 3 Percent retained molecular weight after e-beam radiation of Eastman specialty plastics E-beam, energy, kgy Eastman Tritan MN006 Tenite propionate 360A40012 25 NA 99 100 100 99 88 50 93 99 99 100 99 77 6

The s and s showed no statistical change in molecular weight after exposure to e-beam radiation. Tenite propionate 360 showed significant change in molecular weight. Aliphatic s such as Tenite 360 are more prone to chain scission when exposed to ionization energy than aromatic s. Eastman s specialty medical grades for devices deliver a wide range of processing and performance attributes, including: Brilliant clarity before and after e-beam radiation, gamma radiation, EtO, and low-temperature hydrogen peroxide gas plasma sterilization Chemical resistance to most medical solvents and most medical and household cleaners Good balance of mechanical properties, including excellent toughness Flexibility to use in various processes, including injection molding, injection blow molding, extrusion blow molding, and extrusion Ability to undergo numerous secondary operations, including solvent bonding, swaging (cold forming), laser welding, ultrasonic welding, adhesive bonding, and hot-plate welding If e-beam sterilization is a key material selection criterion, Eastman specialty plastics can provide The Material Difference in your application. Contact Eastman today for more information on compatibility of Eastman resins with e-beam sterilization.

Eastman Chemical Company Corporate Headquarters P.O. Box 431 Kingsport, TN 37662-5280 U.S.A. Telephone: U.S.A. and Canada, 800-EASTMAN (800-327-8626) Other Locations, (1) 423-229-2000 Fax: (1) 423-229-1193 Eastman Chemical Latin America 9155 South Dadeland Blvd. Suite 1116 Miami, FL 33156 U.S.A. Telephone: (1) 305-671-2800 Fax: (1) 305-671-2805 Eastman Chemical B.V. Fascinatio Boulevard 602-614 2909 VA Capelle aan den IJssel The Netherlands Telephone: (31) 10 2402 111 Fax: (31) 10 2402 100 Eastman (Shanghai) Chemical Commercial Company, Ltd. Jingan Branch 1206, CITIC Square No. 1168 Nanjing Road (W) Shanghai 200041, P.R. China Telephone: (86) 21 6120-8700 Fax: (86) 21 5213-5255 Eastman Chemical Japan Ltd. MetLife Aoyama Building 5F 2-11-16 Minami Aoyama Minato-ku, Tokyo 107-0062 Japan Telephone: (81) 3-3475-9510 Fax: (81) 3-3475-9515 Eastman Chemical Asia Pacific Pte. Ltd. #05-04 Winsland House 3 Killiney Road Singapore 239519 Telephone: (65) 6831-3100 Fax: (65) 6732-4930 www.eastman.com Material Safety Data Sheets providing safety precautions that should be observed when handling and storing Eastman products are available online or by request. You should obtain and review the available material safety information before handling any of these products. If any materials mentioned are not Eastman products, appropriate industrial hygiene and other safety precautions recommended by their manufacturers should be observed. It is the responsibility of the medical device manufacturer ( Manufacturer ) to determine the suitability of all component parts and raw materials, including any Eastman product, used in its final product to ensure safety and compliance with requirements of the United States Food and Drug Administration (FDA) or other international regulatory agencies. Eastman products have not been designed for nor are they promoted for end uses that would be categorized either by the United States FDA or by the International Standards Organization (ISO) as implant devices. Eastman products are not intended for use in the following applications: (1) in any bodily implant applications for greater than 30 days, based on FDA-Modified ISO-10993, Part 1, Biological Evaluation of Medical Devices tests (including any cosmetic, reconstructive, or reproductive implant applications); (2) in any cardiac prosthetic device application, regardless of the length of time involved, including, without limitation, pacemaker leads and devices, artificial hearts, heart valves, intra-aortic balloons and control systems, and ventricular bypass assisted devices; or (3) as any critical component in any medical device that supports or sustains human life. For manufacturers of medical devices, biological evaluation of medical devices is performed to determine the potential toxicity resulting from contact of the component materials of the device with the body. The ranges of tests under FDA-Modified ISO- 10993, Part 1, Biological Evaluation of Medical Devices include cytotoxicity, sensitization, irritation or intracutaneous reactivity, systemic toxicity (acute), subchronic toxicity (subacute), implantation, and hemocompatibility. For Eastman products offered for the medical market, limited testing information is available on request. The Manufacturer of the medical device is responsible for the biological evaluation of the finished medical device. The suitability of an Eastman product in a given end-use environment is dependent on various conditions including, without limitation, chemical compatibility, temperature, part design, sterilization method, residual stresses, and external loads. It is the responsibility of the Manufacturer to evaluate its final product under actual end-use requirements and to adequately advise and warn purchasers and users thereof. DuraStar, Eastar, Eastman, Tenite, The Material Difference, and Tritan are trademarks of Eastman Chemical Company. UltraScan is a trademark of Hunter Associates Laboratory, Inc. Eastman Chemical Company, 2011. PPM-218A 10/11