Computer Aided Engineering (CAE) Techniques Applied To Hip Implant



Similar documents
Custom Design of Knee Joint Prosthesis By Using Computerized Tomography (CT) Images and 3D Modelling

COMPUTATIONAL ENGINEERING OF FINITE ELEMENT MODELLING FOR AUTOMOTIVE APPLICATION USING ABAQUS

Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope

The mechanical behavior of the bone microstructure around dental implants

Construction of human knee bone joint model using FDM based 3D Printer from CT scan images.

Creation of an Unlimited Database of Virtual Bone. Validation and Exploitation for Orthopedic Devices

Image-based simulation of the human thorax for cardio-pulmonary applications

Lap Fillet Weld Calculations and FEA Techniques

CRASH ANALYSIS OF AN IMPACT ATTENUATOR FOR RACING CAR IN SANDWICH MATERIAL

ANALYTICAL AND EXPERIMENTAL EVALUATION OF SPRING BACK EFFECTS IN A TYPICAL COLD ROLLED SHEET

Wear-contact analysis of hip implant coated with functionally graded material

A COMPARATIVE STUDY OF TWO METHODOLOGIES FOR NON LINEAR FINITE ELEMENT ANALYSIS OF KNIFE EDGE GATE VALVE SLEEVE

METHODS FOR ACHIEVEMENT UNIFORM STRESSES DISTRIBUTION UNDER THE FOUNDATION

MASTER DEGREE PROJECT

Research Letter Biomimetic Composite-Metal Hip Resurfacing Implant

Tensile fracture analysis of blunt notched PMMA specimens by means of the Strain Energy Density

Effect of Sleeve Shrink-fit on Bearing Preload of a Machine Tool Spindle: Analysis using Finite Element Method

MATERIALS AND METHODS

Structural Integrity Analysis

Image Based-Mesh Generation for realistic Simulation of thetranscranial Current Stimulation

The Right Choice. Exeter. total hip system

Introduction to Mechanical Behavior of Biological Materials

STATIC STRUCTURAL ANALYSIS OF SUSPENSION ARM USING FINITE ELEMENT METHOD

8.2 Elastic Strain Energy

Bildgebende Verfahren als. Control 2011 Alexander Nolte CADFEM GmbH

Finite Element Formulation for Plates - Handout 3 -

Introduction to the Bertram Hip Spacer

ANALYSIS OF A LAP JOINT FRICTION CONNECTION USING HIGH STRENGTH BOLTS

Cylinder Head Gasket Contact Pressure Simulation for a Hermetic Compressor

Study on Structural Behaviour of Human Vertebral Column Using Staad.Pro

Institute of Structural Engineering Division of Computer Aided Design Laboratory of Biomechanics

DESIGN AND ANALYSIS OF BRIDGE WITH TWO ENDS FIXED ON VERTICAL WALL USING FINITE ELEMENT ANALYSIS

Why an Exactech Hip is Right for You

Multiphysics Software Applications in Reverse Engineering

A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS

COMPARISON OF STRESS BETWEEN WINKLER-BACH THEORY AND ANSYS FINITE ELEMENT METHOD FOR CRANE HOOK WITH A TRAPEZOIDAL CROSS-SECTION

BONE PRESERVATION STEM

Constrained Tetrahedral Mesh Generation of Human Organs on Segmented Volume *

An Overview of the Finite Element Analysis

EBM Inside. Additive Manufacturing of orthopedic implants

USAGE OF SELECTIVE LASER MELTING TECHNOLOGY CAPABILITIES IMPLEMENTATION FOR TEETH IMPLANTS METAL CONSTRUCTION IN TU-SOFIA

In-situ Load Testing to Evaluate New Repair Techniques

INJECTION MOLDING COOLING TIME REDUCTION AND THERMAL STRESS ANALYSIS

Biomechanics of Joints, Ligaments and Tendons.

Shell Elements in ABAQUS/Explicit

Reconstruction of Images into 3D Models using CAD Techniques

ELASTO-PLASTIC ANALYSIS OF A HEAVY DUTY PRESS USING F.E.M AND NEUBER S APPROXIMATION METHODS

A Case Study Comparing Two Approaches for Applying Area Loads: Tributary Area Loads vs Shell Pressure Loads

Design & Drafting Services

4.3 Results Drained Conditions Undrained Conditions References Data Files Undrained Analysis of

Linear Static Analysis of a Cantilever Beam Using Beam Library (SI Units)

Simulation for the Collapse of WTC after Aeroplane Impact

Technical What s New. Autodesk Alias Product Line

Local buckling of plates made of high strength steel

New approaches in Eurocode 3 efficient global structural design

This week. CENG 732 Computer Animation. Challenges in Human Modeling. Basic Arm Model

EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.

The patellofemoral joint and the total knee replacement

X-Plain Hip Replacement Surgery - Preventing Post Op Complications Reference Summary

ENGINEERING SCIENCE H1 OUTCOME 1 - TUTORIAL 3 BENDING MOMENTS EDEXCEL HNC/D ENGINEERING SCIENCE LEVEL 4 H1 FORMERLY UNIT 21718P

Mimics Innovation Suite. Software and Services for Engineering on Anatomy TM

Stress and deformation of offshore piles under structural and wave loading

PERIPROSTHETIC IMPLANTS

STATIC AND DYNAMIC ANALYSIS OF CENTER CRACKED FINITE PLATE SUBJECTED TO UNIFORM TENSILE STRESS USING FINITE ELEMENT METHOD

Bending, Forming and Flexing Printed Circuits

An introduction to 3D draughting & solid modelling using AutoCAD

MECHANICS OF SOLIDS - BEAMS TUTORIAL 2 SHEAR FORCE AND BENDING MOMENTS IN BEAMS

Liquid Hydrogen Pressure Vessel Analysis

NOTCHES AND THEIR EFFECTS. Ali Fatemi - University of Toledo All Rights Reserved Chapter 7 Notches and Their Effects 1

Introduction to Solid Modeling Using SolidWorks 2012 SolidWorks Simulation Tutorial Page 1

GOM Optical Measuring Techniques. Deformation Systems and Applications

Solved with COMSOL Multiphysics 4.3

3D Visualization of Pre-operative Planning for Orthopedic Surgery

Design Optimization of a Customized Dental Implant Manufactured via Electron Beam Melting

Posterior Referencing. Surgical Technique

The Basics of FEA Procedure

Express Introductory Training in ANSYS Fluent Lecture 1 Introduction to the CFD Methodology

ABSTRACT INTRODUCTION

Technical Notes 3B - Brick Masonry Section Properties May 1993

NCB Periprosthetic Femur Plate System. Surgical Technique

Materials. Testing Software Data Infrastructure

BEHAVIOR OF WELDED T-STUBS SUBJECTED TO TENSILE LOADS

MECHANICS OF SOLIDS - BEAMS TUTORIAL 1 STRESSES IN BEAMS DUE TO BENDING. On completion of this tutorial you should be able to do the following.

NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE. Health Technology Appraisal

Plates and Shells: Theory and Computation - 4D9 - Dr Fehmi Cirak (fc286@) Office: Inglis building mezzanine level (INO 31)

Keywords: CFD, heat turbomachinery, Compound Lean Nozzle, Controlled Flow Nozzle, efficiency.

The elements used in commercial codes can be classified in two basic categories:

Integrative Optimization of injection-molded plastic parts. Multidisciplinary Shape Optimization including process induced properties

Analysis of Wing Leading Edge Data. Upender K. Kaul Intelligent Systems Division NASA Ames Research Center Moffett Field, CA 94035

Simulation of Residual Stresses in an Induction Hardened Roll

CHAPTER 4 4 NUMERICAL ANALYSIS

THE MODELING OF SOME TYPES OF IMPLANTS AND PROSTHESES USED FOR OSTEOINTEGRATION

Computational Modeling of Hip Replacement Surgery: Total Hip Replacement vs. Hip Resurfacing

Computer Aided Design and Drafting (CAD)

Avizo Inspect New software for industrial inspection and materials R&D

METU DEPARTMENT OF METALLURGICAL AND MATERIALS ENGINEERING

frequently asked questions Knee and Hip Joint Replacement Technology

Fluid structure interaction of a vibrating circular plate in a bounded fluid volume: simulation and experiment

How To Simulate A Kidney Blunt Impact

Transcription:

International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor, 3, A. E. Radi 1, Assoc. Prof, MTC, Cairo, Egypt 2,3, Professor of orthopedics, Faculty of medicine Ain-Shams Univ. Cairo, Egypt Abstract: Methods to predict contact stresses in femoral prostheses can provide an improved understanding of load distribution. The objectives of this study is to apply advanced computer aided engineering techniques (CAE) for predicting stress transfer to femur after implantation of standard and short femoral stems (Proxima stem). A perspective three-dimensional geometry model for the femoral bone using subject-specific geometry from X-ray computed tomography image data (CTI) is used to create 3D CAD model. Forces acting on the femur in different gait regimes were reviewed and compared while loaded with standard and then with Proxima stem. And finally a developed finite element analysis model based on the created CAD model with forces acting on both femurs loaded with standard and Proxima stem is analyzed. Keywords: Computer aided engineering, Biomechanics, Femur modeling, implant, Hip joint I. Introduction: Anatomic short femoral prostheses with a prominent lateral flare have the potential to reduce stressshielding in the femur through a more physiological stress distribution to the proximal femur. Femur bone crack and fracture is one of the most common forms of injuries during accidents. Body weight is transferred through pelvis to femur by hip joint. Bone is a living tissue, which continuously rebuilds its structure according to the direction of loads exerted on it. After insertion of a metal prosthesis into the medullary canal, the load equilibrium and the remodel tissues in the bone are disturbed. Stress-shielding activity of the implant (Fig. 1), which is stiffer, makes the bone atrophies [1]. Reduction of stresses in bone with respect to the natural state causes its adaptation to the new conditions manifested by mass changing (external remodeling) or bone density changing (internal remodeling).the latter is especially dangerous because, as already indicated, it can cause aseptic loosening of the implant. Studying different factors affecting stress-shielding activity required appropriate modelling and simulation of the hip stem and femoral bone along with cemented media used in hip replacement surgery. Femur Fig. 1, Femur bone in human skeleton The proper modeling of femur hip replacement can be achieved through the following processes: [1] Acquisition of computer tomography (CT) data of the femur. [2] Geometrical modeling of the femur and design of prosthesis stem in a computer-aided design (CAD) system. [3] Calculation of forces acting on hip during different gait regimes. [4] Finite element analysis of the selected implant prosthesis and bones. 2. Materials and Methods 2.1 Acquisition and processing of CT data: Novel techniques have been developed to convert 3D image data, obtained from Computer Tomography, automatically into numerical meshes suitable for Finite Element (FE) analysis. CT measurements [2] of the diseased hip joint were performed in order to acquire coordinates of the points defining the femur 219

shape. The CT images obtained were then processed to filter the required data and detect the femur shape (Fig. 2). Fig. 2, CT images implementation to create 3D model of femur bone MIMICS is a software suite that performs the segmentation of the anatomy through sophisticated threedimensional selection and editing tools [3]. The Mimics software is an image-processing package with 3D visualization functions that interfaces with all common scanner formats. 2.2 Femur Geometrical modeling and prosthesis stem design The processed CT data obtained by Mimics is then used either as a meshed model for rapid Finite Element Analysis of the femur or to create a 3D cad model, using AutoDesk Inventor CAD software, for prosthesis stem design as shown in figure (3). Slices for further CAD uses 220

Mesh model for Finite Element Analysis Fig. 3, 2D slices and 3D mesh results of the processed CT data Fig. 4, Processing of the 3D model (cutting) in order to obtain the interested part of the femur The proposed custom implant design was done using Autodesk Inventor CAD software (Fig. 4). A CAD model of the patient's distal femur was imported and used as the base for the design. The articulating surface was kept intact to avoid altering the gait and to preserve the correct patellar grove so as to prevent resurfacing and patellar dislocation.a set of spline curves were created along the interface surface in a radial pattern (Fig. 5), and a single spline was created to connect all curves in a central plane. A swept-blend command was used to create the smooth articulating surface. The bone-implant interface was created using a new set of curves in the same planes as the original curves. The inner curves were offset from the original curves and then manually edited to avoid any undercuts. The offset distance was kept uniform to create an implant with constant wall thickness. A new center plane curve was created as well to prevent the need for undercuts. A swept-blend-cut command was used to create the bone-implant interface. Additional cuts and fillets were added to provide an implant with smooth surfaces and edges. 2.3 Forces acting on hip during different gait regimes: Muscle forces exerted during human movement provide insight in tissue load and muscle function and/or malfunction. The use of musculoskeletal models in the estimation of muscle force has been widely reported in the literature [4]. 221

Fig. 5, Femur bone and prosthesis stem modeling Fig. 6, Beam and cross-section views of applied forces and strain distributions in compression (A), bending (B), twisting (C), and shearing (D). Bones are loaded in four possible ways [5], depicted in figure (6), they can be compressed axially, bent, sheared, and twisted. Often these loads occur in combination (most long bones are mostly bent, but also compressed and twisted to varying extents).based on terminologies described in figure (7), a literature review of several analytical and experimental results [6] are plotted in figure (8) and figure (9). These results are used to simulate forces acting on femur hip for the finite element analysis. Fig. 7, femur degree of freedom terminologey 222

Fig. 8, The net muscle moments applied about the axes of the hip joint Fig. 9, Joint contact force in the hip 3. Finite Element Analysis A finite element model of spacer and femur was developed to analyze stresses [7]. The analysis was performed using ANSYS, a FEA software (ANSYS Inc., Canonsburg, United States). CAD model was obtained as described before and transferred to ANSYS for FEA. Figure (10) shows both boundary condition and mesh distribution. Effect of knee-femur connection is modeled as fixed boundary for simplicity. The cancellous bone within the femur head has a Young s modulus of 1300 MPa and Poisson s ratio 0.28 while bone cement were analysed with a Young s modulus of 2200 MPa. A 4 kn load was applied to the femoral head. femur force and boundary condition mesh distribution Fig. 10, Initial and boundary conditions for the FEA 223

4. Results and Discussion Finite element results reveal the capability of the proposed computer aided design process to handle such an analysis requirements for orthopedic implants evaluation. As shown in figure (11) and figure (12), the homogenous distribution of the strain and stress fields across the contact layer between hip stem and cemented material and between cemented material and femur bone is an accepted proof of contact establishment of the cemented fill and both hip and femur. Extensive analysis of hip stem design, material and effect of using cement or cement-less installation of the hip stem along with studying the effect of cement material can be carried out using proposed model. Also the effect of different loads in gait regimes on stress-shielded zone will be analyzed latter on. Fig. 11, Elastic strain distribution in section view (left) and 3D view (right) Fig. 12, (von-mises) Stress distribution in section view (left) and 3D view (right) 5. Conclusions Rapid custom modeling of human skeleton using advanced computed tomography (CT) techniques is believed to be the most effective way of selection of the right implant design for surgery. Advanced computer aided engineering (CAE) techniques such as CAD and FEA can provide better examination and analysis of the designed implant before surgery operation to insure high durability of the designed implant. In the research field, factors affecting stress distribution to human bones via the implant can be done using the CAE facilities for implant design optimization. 224

References [1] Mihai Ovidiu GHIBA, Lucian RUSU, Geometrical design of custom-made femoral stem prostheses, Fascicle of Management and Technological Engineering, Vol. VII, No. XVII, 2008. [2] MIMICS (10.0) tutorial software package. [3] K. L. Chelule, Dr. T. Coole, D.G. Cheshire, Fabrication of medical models from scan data via rapid prototyping techniques [4] Erdemir, A., McLean, Model-based estimation of muscle forces exerted during movements, Clinical Biomechanics Vol. 22, PP. 131-54, 2007. [5] Osbjorn M. Pearson, Daniel E. Lieberman, The Aging of Wolff s Law : Ontogeny and Responses to Mechanical Loading in Cortical Bone, YEARBOOK OF PHYSICAL ANTHROPOLOGY, Vol. 47, PP. 63 99, 2004. [6] Frank C. Anderson, Marcus G. Pandy, " Static and dynamic optimization solutions for gait are practically equivalent, Journal of Biomechanics, Vol. 34, PP. 153-161, 2001. [7] T. Thielen, S. Maas, A. Zuerbes, et al, Mechanical behaviour of standardized endoskeleton-including hip spacers implanted into composite femurs, International Journal of Medical Sciences, Vol. 6, No.5, PP.280-286, 2009. 225