Installation and Operating Instructions for. HP2 Fully Automatic High Purity Manifold

Similar documents
AUTOMATIC CHANGEOVER MANIFOLD Series 915

INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES

COMPRESSED GASES. 1.2 The contents of each cylinder and container must be clearly identified (by tag or stamp) on the cylinder.

MANI-FLEX AutoSwitch

Compressed Gas Cylinder Program

American Fireglass Outdoor Fire pit kit information. General instructions/warnings for outdoor gas pan and burner systems

Compressed Gas Cylinder Safety

VANGUARD A OPERATOR S MANUAL PORTER. FM-952 Rev. E 7/07

Guidance for Those Who Use and Store Gas Cylinders

UCLA - Compressed Gas Cylinder Storage and Handling

Product Manual. CVS Type 67AFR Filter Regulator. Introduction. Installation. Description

Your safety and the safety of others are very important.

ALVERNIA UNIVERSITY OSHA REGULATION: 29 CFR WELDING, CUTTING, AND BRAZING ( HOT WORK ) SECTION: 3600

User s Guide CARBON DIOXIDE CYLINDERS AND REGULATORS

Single- and Two- Stage Regulators

Compressed Gas Cylinder Storage and Handling Guide

tire inflator with pressure gauge

University of Tennessee Safety Guidelines

Certain specific properties of compressed gases make them highly useful in various research activities.

ENGINEERING SPECIFICATIONS FOR CO2 EXTINGUISHING AGENT PRE-ENGINEERED FIRE SUPPRESSION SYSTEMS

PREPIPED LIQUID FUEL MANIFOLDS PLPM

BERMAD Fire Protection

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

NITROUS TRANSFER PUMP INSTRUCTIONS

Compressed Gas Cylinders Safety Policy

Propane Conversion Kit PROPANE CONVERSION KIT INSTALLATION MANUAL

36G22, 36G23, 36G24 & 36G52 36J22, 36J23, 36J24 & 36J52 DSI and HSI Single Stage Combination Gas Valve

San josé OWNER S MANUAL

INSTALLATION AND OPERATIONS GUIDE FOR GRAND CANYON GAS LOG SYSTEMS

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

Pressure Vessels (Air Compressors) and LPG Tanks

Indirect-Fired Storage Water Heater Models WH-30 through WH-80 INSTALLATION AND OPERATING INSTRUCTIONS

INSTANT HOT WATER RECIRCULATING SYSTEM

Recommended Product Specifications Fuel Day Tank System Tramont UTRS Fuel Day Tank

Element D Services Plumbing

Utility Distribution Systems

Take care with oxygen Fire and explosion hazards in the use of oxygen

NFPA31FuelOilPiping 1 Revised

Compressed Gas Cylinder Safety

Model F822 thru F834 Mulsifyre Directional Spray Nozzles, Open, High Velocity General Description

This leaflet provides information on

Morrison Bros. Co. General Product Specifications

Ontario Fire Code SECTION 5.13 DIP TANKS. Illustrated Commentary. Office of the Ontario Fire Marshal

Cleaning Instructions for Burner, Pilot Assembly, and Emitter Screen. Series: LP Gas

OWNER S MANUAL FORCE 10 MARINE COMPANY HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) FAX: (604)

Oxy-Fuel Gas Welding. Given a functional oxy-fuel gas unit, instruction and demonstration of use, each student will be able to:

Technical Data. Installation

Pump Specifications 250 Series Submersible Sump / Effluent Pump 2 Solids handling

GAS CYLINDER SAFETY. 2. GRADES OF GASES BOC have divided the gases that they supply into 3 grades.

TABLE OF CONTENTS. SECTION 1 General Information: Pg 2 Description of Operation

PBX Series Quick Fit Connector Bimetallic Steam Traps

INSTRUCTION MANUAL IM042. Slide Rail Systems - Series A10 INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

Advantium 2 Plus Alarm

WHAT YOU DON T KNOW ABOUT ACCUMULATORS CAN KILL YOU!

Model: 400E-2M. Bermad Electrically Controlled Deluge Valve with Easy Lock Manual Reset. Installation Operation Maintenance. Application Engineering

Natural Gas Emergencies

1805 Series Relief Valves

Graymills. Air Operated Double Diaphragm Pumps. Operations and Maintenance Instructions

Installation, Maintenance, & Repair Series 4000B/LF4000B

BERMAD Fire Protection

G670 Intermittent Pilot Ignition Controls

HYDRAULIC LIFT TABLE CART 2200-LB.

Medical Air Systems: Scroll

FLAMMABLE AND COMBUSTIBLE LIQUIDS. n OSHA Changes

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

OHIO MEDICAL CORPORATION. Medical Gas Panel. Installation and Maintenance Manual

Hydraulic Transmission Jacks Operating Instructions & Parts Manual

INSTALLATION INSTRUCTIONS Fan Coil Replacement Coil Kit EBX & EBXX

MODEL G300 BRAKE BLEEDER

Cylinder Leak- Down Tester

Requirements for Hydrogen Storage and Use

Fluid Regulator. Instructions Parts List L LOW PRESSURE

USER S, MAINTENANCE and SERVICE INFORMATION MANUAL

INSPECTION, MAINTENANCE AND RECHARGE SERVICE MANUAL P/N 16303

KNC Model 3666 Automatic Pressure Calibration System

TWO BURNER PROPANE CAMPING STOVE

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

SECTION PACKAGED ROOFTOP AIR CONDITIONING UNITS NON-CUSTOM

OPERATIONS & MAINTENANCE (O&M) FOR PROPANE STORAGE FACILITIES

Compressed Gas Cylinder Safety Guidelines

ecomax Instructions for use Wall hung room sealed fan assisted condensing boilers For the user

A practical guide to restrictive flow orifices

All Safe, Inc. Sherwood Master Distributor Toll Free: Fax:

Mini-led spotlight with magnetic base

ASK THE EXPERT: Burner Troubleshooting Information & Maintenance

OPERATING MANUAL CONTRACTOR STYLE AXIAL FANS

orlando OWNER S MANUAL

Hydraulic control unit series

Important Ontario Fire Code Information for Building Owners in the City of Windsor

CONTROL PANEL INSTALLATION INSTRUCTIONS. Single Phase Simplex Page Phase Simplex Page 8-13

Installation Manual DIAPHRAGM WELL TANK

Installation Instructions

3 WATT LED SPOTLIGHT Model No. SLM

Fisher 377 Trip Valve

Type HSR Pressure Reducing Regulator for Residential, Commercial or Industrial Applications

OBSOLETE. VS8510,VS8520 Millivolt Gas Valve

Extrusion Flo-Valve INSTRUCTIONS-PARTS LIST psi (210 bar) Maximum Working Pressure. Model , Series K. Rev.

GUIDELINES - Gas Inspection Regulations 1. Recommendations for camping site owners

PLAN SUBMITTAL REQUIREMENTS FOR SPRAY BOOTHS AND SPRAYING ROOMS EXTINGUISHING SYSTEMS

Transcription:

Installation and Operating Instructions for HP2 Fully Automatic High Purity Manifold w w w. a m i c o. c o m

Table of Contents Introduction 3 Safety Precautions 3-4 Description 5 General Instructions 5 Specifications 5 Flexible Hose Length 6 Material of Construction 6 Standard Configurations 6-7 Dimensions 7 Oxygen Service Equipment 8 Flow Performance 8 General Maintenance 9 Shutdown 10 Warning 10 Design Changes 10 Warranty 11 2 Amico Labs

Important These instructions are for experienced operators who know the general principles and safety precautions that must be observed when handling compressed gasses. If you are not certain you fully understand the safety precautions for handling gases, we urge you to obtain and read the Material Safety Data Sheet (MSDS) for each gas being used. Do not permit untrained persons to install, operate or maintain these manifolds. Do not attempt to install or operate these manifolds until you read and fully understand these instructions. If you do not fully understand these instructions, contact Amico. Introduction Amico manifold systems are cleaned, tested and prepared for indicated gas service and are built following National Fire Protection Association and Compressed Gas Association guidelines. The manifold consists of a control panel and two supply bank headers, one service and one reserve supply, to provide an uninterrupted supply of gas for the specific gas application. The control panel is designed and built with features providing automatic changeover from the depleted Service supply bank to the Reserve supply. Pressure gauges, optional alarm signal connections and lights show system status and alert when depleted cylinders need to be replaced. Features of the automatic system include an adjustable line regulator, hoses with check valves, rigid wall-mounted headers and complete mounting hardware. Disclaimer Amico shall be not liable for errors contained herein, incidental damages nor consequential damages in connection with providing this manual or the use of material in this manual. Manufacturer s Statement The information contained in this instruction booklet has been compiled by Amico, from what are believed to be authoritative sources and is offered solely as a convenience for customers. While Amico believes that this information is accurate and factual as of the date printed, the information including design specifications is subject to change without prior notice. Safety Precautions Protect yourself and others. Read and understand the following instructions before attempting to use this equipment. Failure to understand and follow these instructions could result in serious personal injury and/or damage to equipment. Because of the many potential hazards associated with gases, read the Material Safety Data Sheet for each gas you will be using. Know and understand the physical and chemical properties of the gas being used. Observe general precautions for the use of gases. Observe safety precautions for the gas being used. Read and follow precautions on cylinder labels. Never use these manifolds with gases not compatible with the materials of construction. The use of gases not compatible with the materials of construction may cause damage to equipment or injury to personnel. If flammable gases are used with this equipment, do not locate it near open flames nor any other source of ignition. www.amico.com 3

Safety Precautions If toxic or flammable gases are used with this equipment, emergency equipment applicable to the gases in use should be available in the operating area. Many gases can cause the asphyxiation by displacing oxygen in the atmosphere. Ensure that the area where these manifolds are operated is well ventilated. Provide a device to warn personnel of oxygen depletion in the work area. Do not release toxic or flammable gases in the vicinity of personnel. Use this equipment only in well ventilated areas. Vent gases to the outside atmosphere and in an area away from personnel. Be sure that venting and disposal methods are in accordance with Federal, State, Provincial and local requirements. Locate and construct vent lines to prevent condensation or gas accumulation. Be sure the vent outlet cannot be obstructed by rain, snow, ice, insects, birds, etc. Do not interconnect vent lines; if more than one vent is needed, use separate lines. Relief devices should be installed and properly vented in all gas handling systems to protect against equipment failure and over-pressurization. Never connect this equipment to a supply source having a pressure greater than the maximum rated pressure. Refer to the Product Specifications for maximum inlet pressure. Never permit oil, grease or other combustible materials to come in contact with cylinders, manifolds and connections. Oil and grease may react and ignite when in contact with some gases particularly oxygen and nitrous oxide. Cylinder, header and master valves should always be opened very s-l-o-w-l-y. Heat of recompression may ignite combustible materials. Flexible hoses should never be kinked, twisted or bent into a radius smaller than 3 inches. Mistreatment may cause the flexible hoses to burst. Do not apply heat. Some materials may react and ignite while in contact with some gases particularly oxygen and nitrous oxide. Cylinders should always be secured with racks, chains or straps. Unrestrained cylinders may fall over and damage or break off the cylinder valve which may propel the cylinder with great force. Oxygen manifolds and cylinders should be grounded. Static discharges and lightning may ignite materials in an oxygen atmosphere, creating a fire or explosive force. Welding should not be performed near nitrous oxide piping. Excessive heat may cause the gas to dissociate, creating an explosive force. Do not use leak test solution that contains ammonia. Solutions containing ammonia may cause brass tubing to crack. Always use oxygen compatible leak test solution on oxygen or nitrous oxide service equipment. 4 Amico Labs

Description The HP2 Series fully automatic switch over manifolds accommodate multiple portable liquid cylinders or high pressure cylinders divided into two banks. The HP2 Series manifolds provide uninterrupted and continuous gas flow. The system consists of a control box and one cylinder bank per side. The manifold is cleaned and tested for the indicated gas service and built to meet NFPA 99, CSA B51, ASME B31.1, CSAZ351 and CGA requirements. Programmable Logic Controller (PLC) monitors all functions and controls the switch over process. 22mm LEDs indicate the status of each bank: IN SERVICE, READY FOR USE or BANK DEPLETED. Fully automatic switchover ensuring uninterrupted gas flow no knobs or handle to reset. Mechanical gauges display pressure in psig and BAR. Optional remote audio/visual alarm and floor stand. Optional special header configurations available. General Instructions Manifolds should be installed in accordance with guidelines stated by the National Fire Protection Association, the Compressed Gas Association, OSHA and all applicable local codes. The carbon dioxide and nitrous oxide manifolds should not be placed in a location where the temperature will exceed 120 F (49 C) or fall below 20 F (-7 C). Manifolds should be placed in a location where the temperature will not exceed 120 F (49 C) or fall below 20 F (-29 C). A manifold placed in an open location should be protected against weather conditions. During the winter, protect the manifold from ice and snow. In the summer, shade the manifold and cylinders from continuous exposure to direct rays of the sun. The manifold should be located in a clean, well ventilated area which is free of oil and combustible materials. Leave all protective covers in place until their removal is required for installation. This precaution will keep moisture and debris out of the piping interior, avoiding operational problems. All safety relief valves shall be piped/vented outside. Specifications Description Model Numbers Maximum Inlet pressure 3000 psig Outlet Pressure Range 0-110 psig Low alarm set point and switch over set point 200 psig Inlet gauges 0-4000 psig Operating temperature 5 to 50 C Power requirement 110 VAC 50-60 Hz Enclosure rating Nema 1-12 Gauge size 2½ dial Pilot lights 22mm, LED, 110 VAC Outlet pressure relief valve 150 psig www.amico.com 5

Flexible Hose Length When connected to Hose length High pressure cylinders 36 Materials of Construction Parts Enclosure Headers Tubing Standard Configurations Hydrogen & Helium Flexible Hoses Steel Brass CDA 360 ASTM B16 Stainless steel TP304 copper ASTM B280/B819 Fittings Brass ASTM B16/B453 stainless steel 316 Flexible hoses (selection according to service gas) Nylon hose (Thermoplastic) Teflon hose All stainless steel hose Relief valve Regulators Pressure switches Inner core : Nylon fitting : steel Specification SAE J517 100R8 Inner core : Virgin PTFE smooth bore tubing over braid : stainless steel fitting brass Inner core : 321 stainless steel over braid : 304 stainless steel fitting : 316L stainless steel Body, seat retainer: brass ASTM B16 adjusting screw, springs : stainless steel 316 Disc : PTFE Body: forged brass diaphragm : stainless steel housing cap: forged brass Piston : stainless steel O -ring : Buna-N connection : brass Helium and hydrogen are very small molecules that permeate through Teflon and Nylon. All Amico gas cylinder stationary discharging stations are mounted with stainless steel hoses on helium and hydrogen service. Steel Panel Our equipment is mounted in an aluminum cabinet with a Lexan sticker: Hides and protects pressure switches, tubing and alarm wires No more rust or paint chipping Easy wall and stand mounting Name Tag Each equipment bears a name tag telling you important information about: Gas service Alarm set points Pressure settings Year of manufacture Model number Switchover set point 6 Amico Labs

IN SERVICE READY FOR USE BANK DEPLETED IN SERVICE READY FOR USE BANK DEPLETED Flexible Hoses Flexible hose selection is critical to get the best from your manifold. We offer three types: Thermoplastics Teflon Stainless Steel For additional safety, stainless steel hoses are available with: Armor guard As standard features, each hose has: A check valve in the CGA nipple (except 680) An elbow to avoid kinking of the hose (installed when necessary) An adaptor to the outlet to ease connection to the manifold Finally, we offer three cylinder hose connections (all CGA / gas related): Standard CGA nut & nipple Quick connects (zip nuts) Hand tight nuts Dimensions HP2 Series A 24 B 30 C 50 D 70 E 24 F* 36 *HOSES FOR BULK PACKS (CYLINDER CLUSTERS) ARE 96 LONG. CONTROL BOX DEPTH: 11 www.amico.com 7

Oxygen Service Equipment All oxygen service equipment made by Amico is cleaned as per the requirements of CGA G-4.1-1996. The flexible hoses installed on oxygen service discharging stations are Teflon inner core, stainless steel over braid with brass end connections. CAUTION: REMOVE ALL PROTECTIVE CAPS PRIOR TO ASSEMBLY. THE PROTECTIVE CAP MAY IGNITE DUE TO HEAT OF RECOMPRESSION IN AN OXYGEN SYSTEM. FLOW PERFORMANCE Flow Performance HP2 Series FLOW DIAGRAM Flow Diagram 8 Amico Labs

General Maintenance CAUTION: DO NOT USE LEAK TEST SOLUTION THAT CONTAINS AMMONIA. SOLUTIONS CONTAINING AMMONIA MAY CAUSE BRASS TUBING TO CRACK 1. Main section a. Daily b. Monthly. c. Annually 2. Manifold header 1. Record line pressure 1. Check regulators and valves for external leakage 2. Check valves for closure ability 1. Check relief valve pressures 2. Check regulator seats a. Daily: observe nitrous oxide and carbon dioxide systems for cylinder frosting and surface condensation. Should excessive condensation or frosting occur it may be necessary to increase manifold capacity. b. Monthly 1. Inspect valves for proper closure. 2. Check cylinder flexible hoses for cleanliness, flexibility, wear, leakage and thread damage. Replace damaged flexible hoses immediately. 3. Inspect flexible hoses and check valves for closure ability. c. Every 4 years 1. Replace all flexible hoses. www.amico.com 9

Shutdown WARNING: HAZARDOUS GASES MUST BE DISCHARGED INTO A SAFETY VENT. BE SURE TO USE A VENTING PROCEDURE THAT IS ENVIRONMENTALLY ACCEPTABLE AND COMPLIES WITH FEDERAL, STATE, PROVINCIAL AND LOCAL REQUIREMENTS. 1. Close all cylinder valves. 2. Vent the system pressure to 0 psig. If a hazardous gas is used, purge the entire system with clean, dry nitrogen gas. Continue purging until the hazardous gas level in the system is below the TLV for the gas. 3. Close all system valves by turning the knobs fully clockwise. 4. On the HP2 Series, close the line regulator by turning the T-handle counterclockwise until it reaches the stop. Warning Our equipment is primarily intended for use in compressed gas systems. Amico products are designed for use by persons technically trained in the proper use and safe handling of gas delivery systems. Due to the high pressure and hazardous gases employed in these processes, misapplication could result in injury or death. Amico expressly warns against the sale to, or use of our products by anyone other than professionally trained personnel. Do not use this equipment where pressure or temperature can exceed those listed under specifications. Through misuse, age or malfunction, components used with inert, combustible, corrosive, toxic or oxidizing gases can fail in various modes. The system designer is warned to consider the failure modes of all component parts used with the above mentioned gases and to provide adequate safeguards to prevent personal injury or damage to equipment in the event of such failure modes. Adequate safeguards can be, but are not limited to: Pressure relief devices adequately piped to a safe location. Gas detection devices connected to a proper warning audible and visual alarm. Automatic shutoff valves and/or manual shutoff valves with an emergency stop push button. Self-contained breathing apparatus. Pipeline purge system with inert gas. Fire extinguishers and/or automatic sprinklers. System designers must provide a warning to end users in the system s instructional manual if protection against a failure mode cannot be adequately provided. It should be recognized that warnings are valid for any equipment, regardless of manufacturer, and are not restricted to equipment manufactured by Amico. Amico s reputation for equipment with quality performance is well established. We feel we have the additional obligation to provide information or warnings to customers to assist them in using our equipment in a reasonable and safe manner. Design Changes In line with our commitment to continuous improvement, Amico reserves the right to make design modifications or discontinue manufacture of any equipment without prior notice. 10 Amico Labs

Lab Equipment Warranty Policy Amico Corporation warrants its Lab Equipment to be free from defects in material and workmanship for a period of twelve (12) months from the date of shipment. Within this period Amico will repair or replace any part on site, or at the factory, which is proven to be defective at Amico s cost. All hose assemblies and high purity equipment are excluded from this warranty. Any defective hoses and high purity equipment shall have a maximum warranty of one (1) year from the date of purchase. Furthermore, Amico will warrant its material to be free from defect for an additional period of four (4) years (five (5) years from the date of shipment). Within this period, Amico will replace any part, at no charge, which is proven to be defective. Shipping and Installation costs after the first twelve (12) months will be borne by the Customer. This warranty is valid only when the product has been properly installed according to Amico specifications, used in a normal manner and serviced according to factory recommendations. It does not cover failures due to damage which occurs in shipments or failures which resulted from accidents, misuse, abuse, neglect, mishandling, alteration, misapplication or damage that may be attributable to acts of God. Amico shall not be liable for incidental or consequential damages resulting from the use of the equipment. All claims for warranty must first be approved by Amico s Service Department (service@amico.com or 1-877-462-6426). A valid Return Goods Authorization (RGA) number must be obtained from Amico prior to commencement of any service work. Warranty work, which has not been pre-authorized by Amico, will not be reimbursed. www.amico.com 11

www.amico.com Amico Labs www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada 71 East Industry Court, Deer Park NY 11729, U.S.A Toll Free Tel: 1.877.264.2697 Toll Free Fax: 1.866.440.4986 Tel: 905.763.7778 Fax: 905.763.8587 Email: info@amico.com AL-SP-HGH-PUR-HP2-MNFLD MANUAL 10.4.2013