Retinal straylight measurements in patients with implanted bioanalogic polyfocal lens

Similar documents
Seeing Beyond the Symptoms

What are your options for correcting astigmatism?

AcrySof IQ Toric IOL (SN6ATT) Surgeon Keys for Success & Acknowledgement

Premium IOL Implantation Calculations in Post-LASIK Cataract Eyes Using ASCRS IOL Calculator

What is the main target for all phaco surgeons?

Pre-Op to Post-Op: A Complete Optometrist s Guide to Advanced Technology IOLs.

COMPARISON OF TWO FOLDABLE HYDROPHOBIC SINGLE PIECE MICROINCISIONAL IOLS ACRYSOF IQ AND ZARACCOM ULTRAFLEKS IN RABBIT MODEL

Welcome to the Verisyse Seminar

OMNI EYE SPECIALISTS. The Intraocular Lens

Timothy D. McGarity, M.D.

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

Cataracts & Astigmatism? Quality Distance Vision in Just One Step

Cataract Surgery after Myopic Refractive Procedures. Ray Guard Eye Center Huang Wei-Jen, MD

Standard Single-Focus Lens Implant

Multifocal Intraocular Lenses

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life

If You Receive a Monofocal Lens During your Cataract Operation You Can Get Rid of Astigmatism at the Same Time

Ophthalmic Consultants of Long Island

LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted

Clinical Evaluation. SeeLens MF Intraocular Lens

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical,

Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation

THE GUIDE TO REFRACTIVE LENS EXCHANGE SEE CLEARLY.

Incision along Steep Axis

Explanation of the Procedure

Tucson Eye Care, PC. Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens

CORRECT YOUR CATARACTS AND YOUR VISION

ClearVision. Pacific ClearVision Institute. Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens

University of Applied Sciences (former SHFA) Graduation, Fed. Dipl. Optometrist

BACKGROUND INFORMATION AND INFORMED CONSENT FOR CATARACT SURGERY AND IMPLANTATION OF AN INTRAOCULAR LENS

Alberta Police Recruit Selection Standards

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant

Cataract Surgery Patient Information

VISION PATTERN. What Can you Expect from the ReSTOR procedure?

The LenSx Laser System. Discover the assurance of bladeless cataract surgery

Faster recovery of visual acuity at all distances

SARAH PHYLLIPS. Now, a cataract procedure as unique as your eye itself. TABLE 1. Age: Depth (ACD) Axial LengTH

Presbyopia modern surgical treatments. Johan Lønsmann Poulsen, Euroeyes

Cataracts. Cataract and Primary Eye Care Service Main Number Physician Referral AT-WILLS

Discover the assurance of bladeless cataract surgery. The LenSx Laser System. Important Product Information for the LenSx Laser

PATIENT INFORMATION BOOKLET

Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens

COS Statement on Values for Uninsured Services in Canada

EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH

The Oculentis Toolbox for Lens Surgery

Glenn B. Cook, M.D., Ph.D. DIPLOMATE AMERICAN BOARD OF OPHTHALMOLOGY Coronado Eye Associates

Cataract Testing. What a Patient undergoes prior to surgery

Increasing cost of health care 2010 Prevalence of Cataract Patients: 24.4 M (NEI/NIH) of patients age 40+

WHAT IS A CATARACT, AND HOW IS IT TREATED?

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

Informed Consent for Cataract Surgery and/or Implantation of an Intraocular Lens (IOL)

Astigmatism and vision: Should all astigmatism always be corrected? 1 Ophthalmic Research Group, Aston University, Birmingham, UK

Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia

An aspheric intraocular telescope for age-related macular degeneration patients

Current Trends in Modern Cataract Surgery. Jose Ivan Quiceno, MD Division of Ophthalmology Scripps Clinic Medical Group

VA high quality, complications low with phakic IOL

The IOL Counselor Operator Manual Patient Education Concepts, Inc. Version 1.4 (2/08/2010)

CONSENT FOR CATARACT SURGERY

Table of Contents. Page. Introducing Crystalens 2. What is a cataract 3. How the eye works 4. Vision restoration options 5

Oregon Eye Specialists, PC YOUR GUIDE CATARACT SURGERY. Improving VISION. Improving LIFE.

True Solutions for Presbyopia With Laser Technology Capturing the growing presbyopia market sector.

SMALL INCISION CATARACT SURGERY

Informed Consent for Cataract Surgery or Clear Lens Extraction with Implantation of an Intraocular Lens

LASIK and Refractive Surgery. Laser and Lens Vision Correction Options

Premium Lenses in Problematic Patients

5/24/2013 ESOIRS Moderator: Alaa Ghaith, MD. Faculty: Ahmed El Masri, MD Mohamed Shafik, MD Mohamed El Kateb, MD

Corniche Road Bld. #4WB - Office # (2) (4) EYE SURGERY

Cataract Removal and LifeStyle Implants


A comparison of higher order aberrations in eyes implanted with AcrySof IQ SN60WF and AcrySof SN60AT intraocular lenses

Visual function after bilateral implantation of a new zonal refractive aspheric multifocal intraocular lens

Ocularis Pharma, LLC. The Next Wave in Ophthalmic Drugs: Managing the Pupil to Improve Night Vision Performance Ocularis Pharma, LLC November, 2013

Straylight values 1 month after laser in situ keratomileusis and photorefractive keratectomy

How To Treat Eye Sight Problems With Eye Care

refractive surgery a closer look

Comparison of Higher Order Aberrations and Contrast Sensitivity After LASIK, Verisyse Phakic IOL, and Array Multifocal IOL

Laser cataract surgery

Integrated Eye Care. Improving. Collaborative Patient Management for Premium IOLs. Contributors: Robert E. Prouty, OD, FAAO. Stephen S.

Call today at

Cataract Surgical Packages

LASER VISION C ORRECTION REFRACTIVE SURGERY CENTER

IBRA Professional. Leading Technology for Refractive Outcome Analysis. Brochure 2015

Clouds FREE. things up Shedding light on facts about cataracts. Glass. Looking. Clearing. Parting the. Through the. A journey to clearer vision

Uncorrected astigmatism can cause visual

Frequently Asked Questions. Understanding Cataracts and Cataract Surgery

Alcon. January 2016 Page 1 of 5 Novartis AG

Comparison of Retinal Image Quality between SBK and PRK. Allen Boghossian, D.O. Durrie Vision Overland Park, KS

Verisyse Phakic IOL. Facts You Need to Know About Implantation of the Verisyse Phakic IOL (-5 to -20 D) for the Correction of Myopia (Nearsightedness)

Non-surgical Presbyopia correction

Trulign Toric Accommodating Intraocular Lenses PATIENT INFORMATION BROCHURE

Vision Glossary of Terms

POSITION PAPER November 2011 IMPROVING ACCESS TO MODERN CATARACT SURGERY IN EUROPE THROUGH EXPANDING PATIENT CHOICE. Page 1 of 8

OCT-guided Femtosecond Laser for LASIK and Presbyopia Treatment

Transcription:

Retinal straylight measurements in patients with implanted bioanalogic polyfocal lens Kateřina Liehneová 1, Sylvie Ovesná 2, Eva Adamusová 3, Martin Hložánek 1 1 Ophthalmology Department, University Hospital Královské Vinohrady and 3rd Medical Faculty, Prague, Czech Republic 2 GEMINI Eye Clinic, Zlín, Czech Republic 3 Medicem Institute, Kamenne Zehrovice, Czech Repulic Financial interests: E. Adamusová is employed by a for-profit company with an interest in the subject of the presentation.

Study Purpose The purpose of the study is to compare intraocular light scattering in pseudophakic eyes of patients with implanted bioanalogic, polyfocal lens WIOL-CF and healthy eyes of corresponding age. Materials and Methods Twelve patients (23 eyes) from 2 centers in the Czech Republic (University Hospital Královské Vinohrady and 3rd Medical Faculty and GEMINI Zlín) were examined. Mean patient age was 65 ± 6 years. The cohort included 4 males and 8 females. The follow-up time was from 3 to 12 months. All patients underwent binocular routine cataract surgery and WIOL-CF bioanalogic IOL implantation. The patients didn t have any other ocular surgery or disease. The average postoperative UDVA was 0,92 ± 0,16 Dec. All patients were measured by straylight meter C-Quant (abbreviation for Cataract-Quantifier; Oculus). The data were analyzed using statistical software QC.Expert version 3.3 (TriloByte Statistical Software). WIOL-CF data were compared to recently published retinal straylight values in patients with monofocal and multifocal intraocular lens (IOL).

Straylight The main sources of retinal straylight are: corneal scatter, iris and sclera transparency, lens, and fundus scatter. The light scattering reduces the contrast of the retinal image, and consequently decreases the quality of image projected on the retina. Visualization of retinal straylight. The optical components of the eye form an image of the outside world (left picture) on the retina (right picture). In the case of such a street scene, the picture on the retina is much degraded. Street objects are much less visible compared to the original picture. This is caused by the fact that part of the light coming from the car headlight is scattered in all forward directions (represented by the white arrows in the figure), projecting a veil of light over the retinal image which causes a decrease in the contrast of this image. This veil of light is called straylight. The lower the retinal straylight (straylight ratio s measured by C-Quant), the better the vision. Texts and figures on this slide are from freely accessible document Introduction into retinal straylight on http://www.herseninstituut.knaw.nl/

Comparison of measured WIOL-CF data to other published IOLs IOL Optics Material of IOL Straylight value [log(s)] WIOL-CF (MEDICEM) Polyfocal WIGEL (unique hydrogel) 1.09±0.25 NCL (65±8 years) - Natural crystalline lens 1.19±0.07 1 ReSTOR SN6AD1 IOL (Alcon) Multifocal diffractive Hydrophobic Acrylic 1.23±0.21 2 AcrySof IQ IOL (Alcon) Monofocal aspheric Hydrophobic Acrylic 1.16±0.23 2 AR40e (AMO) Monofocal spherical Hydrophobic Acrylic 1.47±0.22 3 HQ201hep (Hexavision) Monofocal spherical Hydrophilic Acrylic 1.37±0.24 3 PC156C55 (Henan universe IOL R&M Co) Monofocal spherical PMMA 1.45±0.23 3 1 The value is derived from statistical curve from C-Quant software. 2 Cheng Peng et al.: Optical performance after bilateral implantation of apodized aspheric diffractive multifocal intraocular lenses with +3.00-D addition power; Acta Ophthalmologica 2012 3 Ya-wen Guo et al.: Comparison of the Retinal Straylight in Pseudophakic Eyes with PMMA, Hydrophobic Acrylic, and Hydrophilic Acrylic Spherical Intraocular Lens; Journal of Opthalmology 2014

Results WIOL-CF vs NCL Polyfocal WIOL-CF vs Multifocal IOL The retinal straylight in eyes with WIOL-CF was statistically significantly better than healthy population standard of corresponding age (P<0.05). The mean value of retinal straylight in WIOL-CF eyes was 1.09±0.25. Recently published retinal straylight values in pseudophakic eyes implanted with multifocal IOLs from hydrophilic acrylic were 1.23±0.21 (age: 66±8 years) 1. The difference in retinal straylight between eyes implanted with multifocal IOLs and polyfocal IOLs (WIOL-CF) was statistically significant (P<0.01). Polyfocal WIOL-CF from WIGEL vs monofocal IOLs from different materials Recently published retinal straylight values in pseudophakic eyes implanted with monofocal IOLs made from hydrophobic acrylic (Alcon and AMO) were 1.16±0.23 (age: 67±9 years) 1 and 1.47±0.22 (age: 67±7 years) 2, respectively; from hydrophilic acrylic and PMMA 1.37±0.24 (age: 63±10 years) and 1.45±0.23 (age: 65±8 years), respectively 2. The difference in retinal straylight between eyes implanted with monofocal IOLs made from hydrophilic acrylic, PMMA and hydrophobic acrylic from AMO and polyfocal IOLs made from WIGEL (WIOL-CF) was statistically significant (P<0.01). 1 Cheng Peng et al.: Optical performance after bilateral implantation of apodized aspheric diffractive multifocal intraocular lenses with +3.00-D addition power; Acta Ophthalmologica 2012 2 Ya-wen Guo et al.: Comparison of the Retinal Straylight in Pseudophakic Eyes with PMMA, Hydrophobic Acrylic, and Hydrophilic Acrylic Spherical Intraocular Lens; Journal of Opthalmology 2014

Conclusions The retinal straylight in the eyes implanted with WIOL-CF was statistically significantly better compared to healthy population standard of corresponding age. Regarding the retinal straylight WIOL-CF had better results in retinal straylight than monofocal IOLs made from hydrophobic acrylic (AMO), hydrophilic acrylic (Hexavision) and PMMA (Henan universe IOL R&M Co). Average value of straylight ratio in case of hydrophobic acrylic IOL (AcrySof IQ IOL, Alcon) is higher than WIOL-CF results, however the difference is not statistically significant. The retinal straylight in WIOL-CF patients is substantially better than in the case of multifocal hydrophobic acrylic IOLs (ReSTOR, Alcon). If we assume that four of five sources of straylight were identical in pseudophakic and phakic eyes, it could be concluded that the difference would lay mainly in the lens properties and consequently, the retinal straylight induced by WIOL-CF would correspond to the straylight induced by natural crystalline lens of lower age. This suggests that polyfocal WIOL-CF should be less susceptible to disturbing optical phenomena at low light under glare conditions than multifocal IOLs and than IOLs made from hydrophobic acrylic, hydrophilic acrylic and PMMA. This finding is consistent with better contrast sensitivity of WIOL-CF patients and could be due to the lower reflectivity and higher resistance of WIGEL (WIOL-CF material) to deposits. Moreover, smooth continuous optical surface (without transition between optical zones) may also contribute.