Visual function after bilateral implantation of a new zonal refractive aspheric multifocal intraocular lens

Size: px
Start display at page:

Download "Visual function after bilateral implantation of a new zonal refractive aspheric multifocal intraocular lens"

Transcription

1 ARTICLE Visual function after bilateral implantation of a new zonal refractive aspheric multifocal intraocular lens Gonzalo Mu~noz, MD, PhD, FEBO, Cesar Albarran-Diego, MSc, Teresa Ferrer-Blasco, PhD, Hani F. Sakla, MD, PhD, Santiago García-Lazaro, PhD PURPOSE: To evaluate visual function after bilateral implantation of a zonal refractive aspheric multifocal intraocular lens (IOL). SETTING: Private practice surgery center, Valencia, Spain. DESIGN: Cohort study. METHODS: Consecutive eyes with cataract had bilateral implantation of Lentis Mplus LS-312 multifocal IOLs. Distance, intermediate, and near visual acuities; contrast sensitivity; defocus curves; and a quality-of-vision questionnaire, including presence of halos or dysphotopsia, were evaluated 6 months postoperatively. A control group of age-matched monofocal pseudophakic patients was included to compare contrast sensitivity function. RESULTS: In the multifocal group, the mean binocular corrected distance visual acuity (logmar) was 0.04 G 0.07 at 6 m, 0.11 G 0.10 at 1 m, and 0.06 G 0.07 at 40 cm. The defocus curve showed little intermediate vision drop off. Photopic contrast sensitivity for distance was similar to the monofocal IOL contrast sensitivity function, while photopic contrast sensitivity for near and mesopic contrast sensitivity for distance with or without glare was reduced at high frequencies. The mean patient satisfaction was 8.09 G 1.30 (scale 0 to 10); 84.4% of patients were completely independent of spectacles. Moderate halos, glare, and night-vision problems were reported by 6.2%, 12.5%, and 15.6% of patients, respectively. CONCLUSION: The new-generation multifocal IOL provided adequate distance, intermediate, and, to a lesser extent, near vision with high rates of spectacle freedom. Halos occurred, and other photic phenomena should be expected in a small percentage of patients. Financial Disclosure: No author has a financial or proprietary interest in any material or method mentioned. J Cataract Refract Surg 2011; 37: Q 2011 ASCRS and ESCRS Implantation of intraocular lenses (IOLs) to obtain refractive correction for all distances is the goal of lens-based refractive surgery. Several multifocal IOL designs have been introduced; these include diffractive, zonal refractive, hybrid refractive diffractive, and pseudoaccommodating IOLs Previous studies 11,12 report better uncorrected near and intermediate vision with multifocal IOLs than with monofocal IOLs and comparable distance acuity. However, a lack of satisfaction with the outcomes of multifocal IOL implantation is a relatively frequent report by patients who do not achieve visual goals, have limited quality of vision, or develop new optical aberrations. 5,13 The Lentis Mplus (Oculentis GmbH) is a new singlepiece refractive multifocal IOL with an aspheric posterior surface. The multifocality is achieved with a near vision sector of C3.00 diopters (D) in the lower IOL segment, as in a bifocal ophthalmic lens, yielding C2.50 D at the spectacle plane. The new IOL is based on the concept of rotational asymmetry; the IOL is divided into 2 radial sectors, 1 for distance vision and 1 for near vision, both on the optical axis of the lens. The light is refracted to the near focus in a specific sector only, and the rest of the lens acts as a monofocal IOL. The potential advantages of this approach include more light to the distance focus with improved contrast sensitivity, less halo and glare, and better Q 2011 ASCRS and ESCRS Published by Elsevier Inc /$ - see front matter 2043 doi: /j.jcrs

2 2044 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL image quality, avoiding the effect of light scatter due to diffraction. 14 Comparison of this new-generation rotational asymmetric multifocal IOL with a monofocal IOL has been performed 14 ; however, the study did not include subjective assessment of photic phenomena postoperatively. The aim of the present study was to evaluate the clinical performance of the Lentis Mplus LS-312 multifocal IOL, including photic phenomena assessment. PATIENTS AND METHODS Enrollment and Baseline This prospective consecutive nonrandomized interventional clinical study comprised patients having bilateral cataract surgery with implantation of a rotationally asymmetric multifocal IOL. All patients provided written informed consent before surgery in accordance with the Declaration of Helsinki, and institutional review board approval was obtained from the hospital ethics committee. Inclusion criteria were cataract causing a significant reduction in visual quality, potential visual acuity determined by dilated near-pinhole test 15 of 0.3 logmar or better, and corneal astigmatism of 1.00 D or less in both eyes. Exclusion criteria were ocular disease other than cataract, including uveitis, amblyopia, glaucoma, retinal detachment, diabetic retinopathy, macular degeneration, corneal opacities, previous corneal refractive surgery, and neuro-ophthalmic disease. Preoperatively, all patients had a full ophthalmologic examination including refractive status; distance, intermediate, and near monocular visual acuities; slitlamp evaluation; tonometry; and fundoscopy. The uncorrected distance visual acuity (UDVA) and corrected distance visual acuity (CDVA) at 6 m and the uncorrected intermediate visual acuity (UIVA) and distance-corrected intermediate visual acuity (DCIVA) at 2 m, 1 m, and 70 cm were assessed using a calibrated Snellen chart. Uncorrected near visual acuity (UNVA), distance-corrected near visual acuity (DCNVA), and corrected near visual acuity (CNVA) using a C2.50 D addition (add) over distance correction were tested at 50 cm, 40 cm, and 33 cm with an Early Treatment Diabetic Retinopathy Study chart (Precision Vision). Corneal Submitted: March 16, Final revision submitted: May 17, Accepted: May 19, From the Refractive Surgery Department (Mu~noz, Albarran-Diego), Clınica Oftalmologica Marques de Sotelo, and Hospital NISA Valencia al Mar and the Optometry Research Group (GIO) (Ferrer-Blasco, Garcıa-Lazaro), University of Valencia, Valencia, Spain; Ebsaar Eye Surgery Center (Mu~noz, Sakla) Dubai, United Arab Emirates. Supported in part by Ministerio de Ciencia e Innovacion Research Grant (SAF ). Corresponding author: Gonzalo Mu~noz, MD, PhD, FEBO, Clınica Oftalmologica Marques de Sotelo, Avenida Marques de Sotelo 5, Planta 2 a, Pta. 5, Valencia, Spain. gon.munoz@ ono.com. refractive power was measured with an automated keratorefractometer (KR-7000, Topcon), and axial length (AL) was measured using noncontact optical biometry (IOLMaster, Carl Zeiss Meditec AG). The power of the IOL was selected for a target of emmetropia applying the SRK/T formula for eyes with an AL longer than 22.0 mm and the Hoffer-Q formula for eyes with an AL shorter than 22.0 mm An A constant of was used. The data-screening criteria proposed by Holladay et al. 19 were used to avoid refractive surprises. Multifocal Intraocular Lens Lentis Mplus (model LS-312MF) is a biconvex 1-piece multifocal acrylic IOL made with an aspheric posterior surface design. The LS-312 MF model has a 12.0 mm overall length, a 6.0 mm optic, and a C-loop haptic design with 0-degree angulation. The nonrotational symmetric multifocal IOL with a refractive design combines an aspheric asymmetric distance vision zone with a sector-shaped near vision zone with a C3.00 D add. Surgical Technique The same surgeon (G.M.) performed all operations; the second-eye surgery was performed within 1 week of the first. The technique included phacoemulsification through a 2.75 mm clear corneal incision placed on the steepest meridian, a 5.0 mm capsulorhexis, and symmetric implantation of the IOL in the capsular bag. The IOLs were implanted according to the manufacturer s instructions using a disposable 2.2 mm injector (Viscoject 2.2 Cartridge-Set, Medicel AG); that is, with the horizontal reference marks in the optic of the IOL placed on the 0- to 180-degree meridian (haptics at 90 to 270 degrees) and thus with the near sector placed inferiorly. Postoperative topical therapy included diclofenac 0.1% (Dicloabak) and tobramycin 0.3% dexamethasone 0.1% (Tobradex) for 3 weeks. Postoperative Assessments Evaluations were performed 1 day, 1 week, and 1, 3, and 6 months after the second-eye surgery and included refractive status, slitlamp evaluation, tonometry, and fundoscopy. Monocular and binocular UDVA, CDVA, UIVA, and DCIVA at 2 m, 1 m, and 70 cm and UNVA, DCNVA, and CNVA at 50 cm, 40 cm, and 33 cm were recorded. To generate defocus curves, monocular visual acuity was measured with the Snellen chart at 6 m using a variety of lens powers in a phoropter. Patients were defocused to 5.00 D spherical from the manifest refraction values, after which logmar acuity was recorded. Added negative spherical power was progressively reduced in 0.50 D steps; logmar acuity was recorded after every step until the original manifest refraction values were reached. Patients were then defocused to C3.00 D spherical from the manifest refraction, and logmar acuity was again recorded. Positive spherical power was progressively reduced in 0.50 D steps and logmar acuity recorded after every step until manifest refraction values were reached. Contrast sensitivity was tested monocularly for distance and near vision with best correction for distance under photopic conditions (90 candelas [cd]/m 2 ) and for distance with and without glare under mesopic conditions (5 cd/m 2 ) using a Functional Acuity Contrast Test (Stereo Optical Co., Inc.). 20 This test allows the presentation of sine-wave gratings at

3 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL 2045 different spatial frequencies (1.5, 3, 6, 12, and 18 cycles per degree [cpd]), with increasing contrast values. Absolute values of log 10 contrast sensitivity were obtained for each combination of eye and spatial frequency, and means and standard deviations (SD) were calculated. An HD 9221 photometer (Delta OHM S.r.L.) was used to measure the lighting conditions in the room in which contrast sensitivity was measured to ensure equal lighting levels for all patients at all testing sessions. Under photopic conditions, luminance was 90 cd/m 2 and illumination was 450 lux. Mesopic conditions were 5 cd/m 2 and 4 lux. 21 Patients remained 10 minutes inside the testing room under mesopic conditions for an appropriate adaptation before mesopic contrast sensitivity was measured. Contrast sensitivity with appropriate correction for distance and near was measured in a control group of age-matched pseudophakic patients with the AR-40e monofocal IOL (Abbott Medical Optics, Inc.). Posterior capsule opacification (PCO), IOL centration, and tilt were assessed at 6 months. Posterior capsule opacification was graded as follows: 1 Z none, 2 Z mild (early development of PCO), 3 Z moderate (increased PCO with early visual acuity changes not requiring secondary capsulotomy), and 4 Z severe (PCO affecting vision and requiring neodymium:yag laser capsulotomy). Quality-of-Vision Questionnaire At the final follow-up visit after 6 months, all patients completed a 14-item questionnaire that included the following: 1. The presence of visual disturbances or photic phenomena (4 items): glare (trouble seeing street signs due to bright light on oncoming headlights), halos (rings around lights), double vision, and night-vision difficulties. 2. Visual lifestyle activities (6 items): watching television, driving at night, using a computer, reading, using a cell phone, and shaving/putting on makeup. 3. Spectacle independence for distance, intermediate, and near vision (3 items). 4. Overall satisfaction (1 item). The response rating scale for visual disturbance and lifestyle activity items was 0 Z no difficulty, 1 Z minimal difficulty, 2 and 3 Z moderate difficulty, and 4 and 5 Z severe difficulty. The response rating scale for spectacle use items was 0 Z never, 1Z rarely or occasionally, 2 Z often, and 3 Z always. The response rating scale for overall satisfaction was from 0 (least satisfied) to 10 (most satisfied). This questionnaire is a summarized version of one used by Kohnen et al., 22 which is in turn based on the survey distributed in the U.S. Food and Drug Administration clinical trials for the Alcon Acrysof IQ Restor in the National Eye Institute Visual Functioning Questionnaire-25, 23 and the Cataract TYPE questionnaire. 24 Data Management and Statistical Analysis Visual acuity values were converted to logmar notation for statistical purposes. For intermediate visual acuities at 2 m, 1 m, and 70 cm and for near visual acuities at 50 cm and 33 cm, scales were corrected accordingly as previously described. 25 Logarithmic scales were used for statistical and graphical purposes with regard to contrast sensitivity data. Manifest refractions were converted into power vector coordinates as proposed by Thibos and Horner. 26 The revised abbreviations for reporting visual acuity values proposed by Kohnen 27 were used. Statistical analysis was performed using SigmaPlot software (version 11, Systat Software, Inc.). The normality of all data samples was evaluated with the Kolmogorov-Smirnov test. The Student t test for paired data was used for all parameter comparisons between preoperative and postoperative examinations. The Student t test for unpaired data was used for comparisons of contrast sensitivity between the multifocal IOL group and the monofocal IOL group. Results are presented as the mean G SD; a P value of 0.05 was considered statistically significant. RESULTS Data were collected from 64 eyes of 32 patients (18 men, 14 women) with a mean age of 63.6 G 9.1 years (range 48 to 71 years). The mean pupil size was 3.4 G 0.8 mm under photopic conditions and 5.3 G 0.7 mm under mesopic conditions. The mean preoperative spherical equivalent (SE) was 1.30 G 2.56 D; 47 eyes (73%) were hyperopic, 12 (19%) were myopic, and 5 (8%) were emmetropic. The age-matched control group (with monofocal IOL) comprised 64 eyes of 32 patients. Adverse Events and Posterior Capsule Assessments There were no intraoperative complications. There were no cases of iris trauma, IOL decentration, or IOL tilt. Two patients had minor adverse events; 1 had bilateral dry eye requiring punctual plug insertion and 1 reported bilateral dysphotopsia in the form of monocular diplopia. At 6 months, 59 eyes (92.2%) had no or mild PCO, 4 eyes (6.3%) had grade 3 PCO, and 1 eye (1.5%) had grade 4 or severe PCO requiring capsulotomy. Refraction The postoperative refraction was close to emmetropia (mean SE [M] 0.30 G 0.33 D). There was no induced astigmatism as a consequence of cataract surgery and no significant changes in J0 or J45 components of astigmatism after vector decomposition (Table 1). Fifty-eight eyes (90.6%) were within G0.50 D of the target SE, and all the eyes were within G1.00 D. The 6 eyes (9.4%) with an SE higher than G0.50 D were all myopic with a mean residual SE of 0.75 G 0.15 D. Distance, Intermediate, and Near Vision Table 1 compares the monocular logmar visual acuity values before surgery and 6 months after surgery in eyes with the multifocal IOL, and Table 2 shows the monocular and binocular logmar visual acuity values at all tested distances. Postoperatively, there was a statistically significant improvement in monocular logmar UDVA, monocular logmar

4 2046 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL Table 1. Comparison of preoperative and 6-month postoperative data after bilateral multifocal IOL implantation. Parameter Mean G SD Preoperative Postoperative P Value Refraction (D)* M 1.30 G G 0.33!.001 J G G J G G LogMAR UDVA 0.49 G G 0.10!.0001 LogMAR CDVA 0.39 G G 0.05!.0001 LogMAR UNVA 0.53 G G 0.13!.0001 LogMAR DCNVA 0.62 G G 0.07!.0001 LogMAR CNVA 0.33 G G 0.06!.0001 LogMAR UIVA (1 m) 0.56 G G 0.12!.0001 LogMAR DCIVA (1 m) 0.53 G G 0.08!.0001 IOL power (D) 20.5 G 2.1 d d IOP (mm Hg) 16.2 G G CDVA Z corrected distance visual acuity; CNVA Z corrected near visual acuity; DCIVA Z distance corrected intermediate visual acuity; DCNVA Z distance corrected near visual acuity; IOL Z intraocular lens; IOP Z intraocular pressure; UDVA Z uncorrected distance visual acuity; UIVA Z uncorrected intermediate visual acuity; UNVA Z uncorrected near visual acuity *Manifest refractions in conventional script, sphere (S), cylinder (C) and axis (f) were converted into power vector coordinates (M, J0, J45) by the following formulas: M Z S C C/2; J0 Z ( C/2) cos (2f); J45 Z ( C/2) sin (2f). Paired Student t test CDVA, monocular logmar UNVA, monocular logmar DCNVA, and monocular logmar CNVA (all P!.01, paired Student t test). There were also statistically significant improvements in monocular log- MAR UIVA and DCIVA at 1 m (both P!.01, paired Student t test). Uncorrected visual acuity in decimal notation was at least 0.5 at near and distance in more than 98% of eyes and at all 3 intermediate distances in 76% of eyes. In the 6 eyes with a residual SE higher than 0.50 D, the mean logmar UDVA was 0.08 G 0.06 (0.83 G 0.59 lines decimal), the mean logmar UIVA at 1 m was 0.04 G 0.05 (0.91 G 0.52 lines decimal), and the mean logmar UNVA at 40 cm was 0.13 G 0.07 (0.74 G 0.65 lines decimal). This indicates that the tolerance to myopic defocus after multifocal IOL implantation was excellent. The correlation between logmar CDVA and photopic pupil size was low (r Z 0.15) and not statistically significant (PZ.23). The same was true for logmar DCNVA and photopic pupil size (r Z 0.09; PZ.52). Defocus Curve The defocus curve in Figure 1 has 2 peaks of maximum vision located at the far and near foci corresponding to 0.00 D and 2.50 D defocus, respectively, with reduced visual acuity for intermediate distances. The defocus curve has a peak of optimum near vision (mean 0.07 G 0.07 logmar) at the vergence of 2.50 D, the equivalent of 40 cm from the eye. The minimum intermediate vision occurs at a vergence of 1.00 D, or 1 m from the eye. Even so, the mean intermediate visual acuity at this point on the defocus curve (0.14 G 0.08 logmar) remains good. Contrast Sensitivity Photopic distance contrast sensitivity in the multifocal IOL group was similar to that in the monofocal IOL control group (Figure 2, top), with no statistically significant differences at any tested frequency Table 2. Monocular and binocular visual acuity at far, intermediate, and near 6 months after multifocal IOL implantation. Intermediate Near Parameter Far (6 m) (2 m) (1 m) (70 cm) (50 cm) (40 cm) (33 cm) Monocular uncorrected VA LogMAR 0.05 G G G G G G G 0.11 Decimal 0.89 G G G G G ,79 G 1, G 1.09 Monocular distance-corrected VA LogMAR 0.02 G G G G G G G 0.11 Decimal 1.05 G G G G G G G 1.09 Binocular uncorrected VA LogMAR 0.03 G G G G G G G 0.11 Decimal 0.93 G G G G G G G 1.07 Binocular distance-corrected VA LogMAR 0.04 G G G G G G G 0.06 Decimal 1.10 G G G G G G G 0.63 VA Z visual acuity

5 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL 2047 Figure 1. Defocus curve for the Lentis Mplus IOL 6 months after implantation. The error bars represent G SD values for each vergence. (1.5 cpd, PZ.23; 3 cpd, PZ.49; 6 cpd, PZ.71; 12 cpd, PZ.20, 18 cpd, PZ.10). Photopic near contrast sensitivity (Figure 2, bottom) was similar at low frequencies (1.5 cpd, PZ.20; 3 cpd, PZ.71; 6 cpd, PZ.34) but was significantly lower than contrast sensitivity function with the monofocal IOL at 12 cpd and 18 cpd (both P!.01). Under mesopic conditions without glare (Figure 3, top), distance contrast sensitivity with the multifocal IOL was similar to that of the monofocal IOL at low frequencies (1.5 cpd, PZ.12; 3 cpd, PZ.12; 6 cpd, PZ.09) but was significantly lower at high spatial frequencies (12 cpd and 18 cpd; both P!.01). The same was true for distance contrast sensitivity under mesopic conditions with glare (Figure 3, bottom), with no significant differences at 1.5 cpd (PZ0.14) and 3 cpd (PZ.14) but with statistically significant lower values at higher frequencies (6 cpd, 12 cpd, and 18 cpd; all P!.01). Visual Disturbances and Photic Phenomena The mean score for glare, halos, double vision, and night-vision disturbances indicated minimum to no difficulty at the 6-month follow-up (Table 3). Five (15.6%) patients reported moderate night-vision problems, 4 (12.5%) reported moderate glare, 2 (6.2%) reported moderate halos, and 1 (3.1%) had monocular double vision in both eyes that caused moderate impairment. No patient reported severe difficulty for any of the 4 visual disturbance items evaluated. Visual Lifestyle Activities At 6 months, no patient reported severe difficulty with any of the 6 lifestyle activities evaluated. None reported moderate problems for watching television (distance vision) or using a computer (intermediate vision). Moderate difficulties were reported by 3 (12.5%) of 24 patients for driving at night, 7 (21.9%) of 32 patients for reading or near work, 4 (12.5%) of 32 patients for shaving/applying makeup, and 6 (20.0%) of 30 patients for using a cell phone. Spectacle Independence and Overall Patient Satisfaction Of the 32 patients, 27 (84.4%) were completely independent of spectacles for distance, intermediate, and near vision. Thirty-one (96.9%) patients never used glasses for distance vision, 31 (96.9%) never used them for intermediate vision, and 27 (84.4%) never used them for near vision. One (3.1%) patient needed glasses often or always for distance vision, 1 (3.1%) for intermediate vision, and 4 (12.5%) for near vision. Twenty-five (78.1%) rated satisfaction as 8 or higher and 2 (6.2%) as lower than 6. DISCUSSION Visual complaints after multifocal IOL implantation fall into 2 categories. The first is decreased vision for distance, intermediate, or near as a consequence of diminished contrast sensitivity. 5 The second is photic phenomena, including glare, halos, and positive dysphotopsia, or perception of lights, sparks, lines, or colors in some parts of the visual field. 2,11 Contrast sensitivity can decrease up to 50% with some models of multifocal IOLs, 5 and it has been reported that photic phenomena are 3.5 times more likely with multifocal IOLs than with monofocal IOLs. 11 The Lentis Mplus is an acrylic nonrotational symmetric multifocal IOL designed to overcome the drawbacks of multifocal IOLs by providing high contrast sensitivity and minimizing halos and glare. The lens has a refractive design combining an aspheric asymmetric distance vision zone with a sector-shaped

6 2048 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL Figure 2. Photopic log contrast sensitivity function at far and near distances in the multifocal IOL group and the monofocal IOL control group. The asterisk represents statistically significant differences between IOLs. The error bars represent G SD values at each spatial frequency (cpd Z cycles per degree). Figure 3. Mesopic log contrast sensitivity function at far and near distances in the multifocal IOL group and the monofocal IOL control group. The asterisk represents statistically significant differences between IOLs. The error bars represent G SD values at each spatial frequency (cpd Z cycles per degree). near vision zone with a C3.00 D add. In the present study, cataract surgery using this new IOL was very predictable. The refractive outcomes were close to emmetropia in all cases, with a mean residual SE near 0.25 D and with only 6% of eyes having a defocus higher than 0.75 D. The distance, intermediate, and, to a lesser extent, near vision results were excellent in our study. The uncorrected visual acuity in decimal notation was at least 0.50 for near and distance in more than 98% of eyes, with a mean binocular logmar UDVA of 0.03 (0.93 decimal). The IOL also performed well for near vision, with a mean binocular logmar UNVA of 0.09 (0.82 decimal) at 40 cm, which was the best distance for near. After distance correction, the mean binocular logmar CDVA and DCNVA improved to 0.04 and 0.06, respectively (1.10 and 0.87 decimal, respectively). The far vision achieved with the Lentis Mplus IOL was comparable to that with other multifocal IOLs, and near vision was somewhat worse. 2,22,28,29 Several studies report that diffractive and hybrid refractive diffractive multifocal IOLs provide better near vision. de Vries et al. 6 and Kohnen et al. 22 report binocular logmar UNVA and DCNVA near 0.00 (1.00 decimal) using the Acrysof Restor SA60D3 IOL and Acrysof Restor SN6AD1 IOL (both Alcon Laboratories, Inc.), respectively. Similar results (binocular decimal DCNVA approximately 1.0) were reported by Cillino et al. 2 using the multifocal diffractive pupil-independent Tecnis ZM900 IOL (Abbott Medical Optics, Inc.) and by Alfonso et al. 28 and Kaymak and Mester 29 using the multifocal diffractive Acri.Lisa 376D IOL (Carl Zeiss Meditec AG). The most likely cause of the worse near vision with the Lentis Mplus IOL is its C3.00 D add, which is lower than the addition of the aforementioned diffractive models except the new C3.00 D Acrysof Restor. Intermediate vision was excellent with the Lentis Mplus IOL. The uncorrected decimal visual acuity was at least 0.5 at all 3 intermediate distances tested

7 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL 2049 Table 3. Mean scores for visual disturbances, visual lifestyle activities, spectacle use and overall satisfaction scores 6 months multifocal IOL implantation. Item Mean Score G SD* Visual disturbances (from 0 to 5) Glare 0.69 G 0.86 Halos 0.38 G 0.71 Double vision 0.25 G 0.62 Night vision 0.84 G 0.92 Visual lifestyle activities (from 0 to 5) Watching television 0.28 G 0.46 Driving at night 0.75 G 0.68 Using a computer 0.31 G 0.47 Reading or sewing 1.28 G 0.99 Using a cell phone 1.10 G 1.06 Shaving or putting on makeup 0.88 G 0.87 Spectacle use (from 0 to 3) For distance vision 0.06 G 0.35 For intermediate vision 0.06 G 0.35 For near vision 0.28 G 0.81 Overall satisfaction (from 0 to 10) 8.09 G 1.30 *The response rating scales: visual disturbance and lifestyle activities, 0 Z no difficulty, 1 Z minimal difficulty, 2 and 3 Z moderate difficulty, 4 and 5 Z severe difficulty; spectacle use, 0 Z never, 1 Z rarely or occasionally, 2 Z often and 3 Z always; overall satisfaction, range from 0 (least satisfied) to 10 (most satisfied) (70 cm, 1 m, and 2 m) in 76% of eyes. The mean decimal DCIVA was higher than 0.7 at all intermediate distances (0.74 at 70 cm, 0.72 at 1 m, 0.87 at 2 m). The intermediate vision performance of this multifocal IOL seems better than that reported for diffractive multifocal models, which usually provide excellent distance and near vision with a drop in intermediate vision, causing intermediate blur in some cases. 30,31 Recently, there has been a trend toward reducing the power of the near add (from C 4.00 D to C3.00 D) in some models of diffractive IOLs to improve intermediate vision or expand reading distance 22 ; therefore, the C3.00 D Acrysof Restor IOL models were developed with only 9 diffractive steps that are more widely spaced than the 12 steps of the C4.00 D IOL models. By reducing the addition for near, the near and the far points become closer and the gradual decline in vision between the 2 principal foci is minimized. Intermediate vision with diffractive models has been reported for several distances ranging from 50 to 70 cm. 22,32 34 The C3.00 D diffractive models provide a mean logmar intermediate vision at 50 cm of approximately 0.05 (0.9 decimal), 22 which is much better than the mean logmar intermediate vision at 50 cm provided by the C4.00 D models, which is approximately 0.25 (0.56 decimal). 33 However, even the C3.00 D diffractive models showed a decrease with more extended intermediate vision; at 60 cm and 70 cm, the mean logmar visual acuity fell to approximately 0.12 and 0.20 respectively (0.75 and 0.63 decimal, respectively). 22 Intermediate vision at 1 m was approximately 0.60 decimal, as can be deduced from the defocus curve of the C3.00 D diffractive models. 22 The mean decimal intermediate vision with the Lentis Mplus IOL was 0.74 at 70 cm, 0.72 at 1 m, and 0.87 at 2 m. Thus, the minimum intermediate vision at 1 m was approximately 0.75 decimal, which is better than the intermediate vision reported for C3.00 D diffractive IOLs. We believe it is important to report visual results for intermediate distances of 70 cm and 1 m because they are used in common activities, such as computer work. Refractive multifocal IOL models such as the Rezoom (Abbott Medical Optics, Inc.) are reported to provide intermediate vision similar to that of C3.00 D diffractive IOLs 35 and thus worse than the intermediate vision provided by the Lentis Mplus IOL. One study of C3.50 D refractive IOLs 35 reports a mean UIVA at the preferred intermediate distance (range 60 to 80 cm) of 0.23 logmar (0.59 decimal). Near vision with the Rezoom IOL is reported to be worse than that with IOLs with diffractive designs, as shown in numerous studies 9,10,35,36 ; therefore, this type of IOL has fallen into disuse in multifocal IOL technology. The defocus curve for the Lentis Mplus IOL is consistent with the above findings. The IOL performance for distance was similar to that with a monofocal IOL, with a decimal visual acuity of approximately The IOL provided decimal intermediate acuity of approximately 0.70 to 0.80 between 50 cm and 2 m and had the best decimal near vision at 40 cm (mean approximately 0.85). Decimal near vision at 33 cm dropped to approximately 0.75 and dropped further at 25 cm to With the multifocal IOL, there was little intermediate drop off at intermediate distances, with the lowest visual acuity threshold at 1 m ( 1.00 D of defocus). Photopic distance contrast sensitivity with the Lentis Mplus multifocal IOL was similar to that with the monofocal IOL, meaning that the multifocal IOL behaves very similarly to a monofocal IOL at distance under high illumination conditions. The photopic near contrast sensitivity was similar between the multifocal IOL and monofocal IOL except at higher frequencies. The mesopic distance contrast sensitivity values also showed differences at high frequencies and at medium and high frequencies in the case of mesopic vision with glare. Alio et al. 14 found similar photopic contrast sensitivity values in a comparison of the Lentis Mplus IOL and a monofocal IOL; they also found a trend (although not significant) toward better contrast sensitivity outcomes

8 2050 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL with the monofocal IOL at high spatial frequencies under mesopic conditions. Zhao et al. 37 report similar findings for the C3.00 D Acrysof Restor SA60D3 IOL, which provided reduced contrast sensitivity compared with a monofocal IOL. The reduction in mesopic visual performance at high spatial frequencies may result in a reduced capability to resolve small objects under such conditions; for example, seeing road indications while driving at night. 3,5 Patients with Lentis Mplus IOLs reported a low incidence of significant photic phenomena. Of the patients, 6.2% reported moderate halos, 12.5% reported moderate glare, and 15.6% reported moderate nightvision problems; none of the patients rated these photic phenomena as severe. The subjective scores for halos (0.38 G 0.71), glare (0.69 G 0.86), and night-vision problems (0.84 G 0.92) were low. Using the same scale to subjectively evaluate visual disturbances caused by a diffractive multifocal IOL, Kohnen et al. 22 found a mean score of 1.3 G 1.3 for halos, 1.7 G 1.5 for glare, and 1.1 G 1.4 for night-vision problems, higher than the results in the present study. It therefore seems that the classic disadvantages of multifocal IOL technology, such as halos and glare, are minimized with the Lentis Mplus IOL. Alio et al. 14 found statistically significantly larger amounts of primary coma and tilt with the Lentis Mplus IOL than with a monofocal IOL. In the present study, aberrometry was not performed; therefore, we could not determine whether there was a correlation between postoperative photic phenomena and residual coma or other higher-order aberrations. We did not find a correlation between photic phenomena and residual astigmatism, IOL tilt, or capsulorhexis size. We also did not find a significant correlation between photopic pupil size and CDVA or DCNVA, meaning that the Lentis Mplus IOL showed pupillary independence with the pupil sizes measured in our study. Eighty-four percent of our patients with bilateral Lentis Mplus IOLs achieved full spectacle independence and were satisfied with their distance, intermediate, and near vision. For far vision, 96.9% reported never using glasses. One patient (3.1%) indicated frequent or continued use of glasses for far vision. The outcome in this case was 1.00 D in both eyes. The tolerance of the Lentis Mplus IOL to myopic defocus was excellent. The 6 eyes with a residual SE higher than 0.50 D had excellent UIVA at 1 m and maintained adequate UDVA and UNVA at 40 cm. On the defocus curve, low levels of residual myopia do not translate into a significant decrease in visual performance, which may represent another advantage of this new multifocal IOL design. In the present study, secondary interventions to improve postoperative refractive results were not considered. A secondary excimer laser procedure might have led to a lower percentage of patients needing glasses for distance vision. Similar percentages of spectacle independence for distance vision have been reported for diffractive multifocal models. 22 The percentage of patients free from glasses for intermediate vision activities was 96.9%; 1 patient reported a frequent need for glasses for intermediate distance. This can be easily explained by the defocus curve of the Lentis Mplus IOL in which the typical drop off at intermediate distances seen in other multifocal IOL models is minimized, which we believe is a main advantage of the new IOL. The percentage of patients who never used glasses for near (84.4%) was high, although lower than that reported for diffractive multifocal IOLs. 22 This may be related to the near add of C3.00 D, which yields a C2.50 D add at the spectacle plane; this may be insufficient for some near activities. Regarding the 6 lifestyle activities evaluated in the present study, none of the patients reported moderate or significant problems for activities related to distance vision (eg, watching television) or intermediate vision (eg, using a computer). However, between 12.5% and 21.8% of patients reported moderate difficulties performing activities that require very accurate near vision, such as reading/sewing, text messaging using a cell phone, or shaving/applying makeup. The mean patient satisfaction score was 8.09 G 1.30, with 78% of patients rating satisfaction as 8 or higher. Two patients (6.2%) reported a satisfaction rating lower than 6. Using the same scale, a mean subjective score of 8.3 G 1.6 was reported by Kohnen et al. 22 using a diffractive IOL model. Improvement in patient satisfaction and quality of life after bilateral implantation of multifocal IOLs in cataract patients or presbyopic refractive lens exchange in patients has also been reported in other recent studies. 38,39 In conclusion, the Lentis Mplus multifocal IOL provided excellent uncorrected distance, intermediate, and, to a lesser extent, near vision. Defocus curve showed very little intermediate drop off. Halos, glare, and other photic phenomena as a consequence of multifocal IOL technology were minimized. REFERENCES 1. Javitt JC, Steinert RF. Cataract extraction with multifocal intraocular lens implantation; a multinational clinical trial evaluating clinical, functional, and quality-of-life outcomes. Ophthalmology 2000; 107: Cillino S, Casuccio A, Di Pace F, Morreale R, Pillitteri F, Cillino G, Lodato G. One-year outcomes with new-generation multifocal intraocular lenses. Ophthalmology 2008; 115:

9 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL Montes-Mico R, Espa~na E, Bueno I, Charman WN, Menezo JL. Visual performance with multifocal intraocular lenses; mesopic contrast sensitivity under distance and near conditions. Ophthalmology 2004; 111: Jacobi FK, Kammann J, Jacobi KW, Großkopf U, Walden K. Bilateral implantation of asymmetrical diffractive multifocal intraocular lenses. Arch Ophthalmol 1999; 117: Available at: Accessed July 7, Pieh S, Weghaupt H, Skorpik C. Contrast sensitivity and glare disability with diffractive and refractive multifocal intraocular lenses. J Cataract Refract Surg 1998; 24: de Vries NE, Webers CAB, Montes-Mico R, Tahzib NG, Cheng YYY, de Brabander J, Hendrikse F, Nuijts RMMA. Long-term follow-up of a multifocal apodized diffractive intraocular lens after cataract surgery. J Cataract Refract Surg 2008; 34: Cervi~no A, Hosking SL, Montes-Mico R, Alio JL. Retinal straylight in patients with monofocal and multifocal intraocular lenses. J Cataract Refract Surg 2008; 34: Montes-Mico R, Alio JL. Distance and near contrast sensitivity function after multifocal intraocular lens implantation. J Cataract Refract Surg 2003; 29: Chang DF. Prospective functional and clinical comparison of bilateral ReZoom and ReSTOR intraocular lenses in patients 70 years or younger. J Cataract Refract Surg 2008; 34: Pepose JS, Qazi MA, Davies J, Doane JF, Loden JC, Sivalingham V, Mahmoud AM. Visual performance of patients with bilateral vs combination Crystalens, ReZoom, and Re- STOR intraocular lens implants. Am J Ophthalmol 2007; 144: Leyland M, Zinicola E. Multifocal versus monofocal intraocular lenses in cataract surgery; a systematic review. Ophthalmology 2003; 110: Ortiz D, Alio JL, Bernabeu G, Pongo V. Optical performance of monofocal and multifocal intraocular lenses in the human eye. J Cataract Refract Surg 2008; 34: Woodward MA, Randleman JB, Stulting RD. Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg 2009; 35: Alio JL, Pi~nero DP, Plaza-Puche AB, Rodriguez Chan MJ. Visual outcomes and optical performance of a monofocal intraocular lens and a new-generation multifocal intraocular lens. J Cataract Refract Surg 2011; 37: Melki SA, Safar A, Martin J, et al. Potential acuity pinhole; a simple method to measure potential visual acuity in patients with cataracts, comparison to potential acuity meter. Ophthalmology 1999; 106: ; discussion by DJ McIntyre, Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg 1990; 16: ; correction, Gavin EA, Hammond CJ. Intraocular lens power calculation in short eyes. Eye 2008; 22: Available at: nature.com/eye/journal/v22/n7/pdf/ a.pdf. Accessed July 7, Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg 1993; 19: ; errata 1994; 20: Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg 1988; 14: Pesudovs K, Hazel CA, Doran RML, Elliott DB. The usefulness of Vistech and FACT contrast sensitivity charts for cataract and refractive surgery outcomes research. Br J Ophthalmol 2004; 88: Available at: nlm.nih.gov/pmc/articles/pmc /pdf/bjo pdf. Accessed July 7, Montes-Mico R, Albarran-Diego C, Mu~noz G, Alio JL. Use of illuminance and luminance units [letter]. J Refract Surg 2004; 20:490; reply by SE Wilson, Kohnen T, Nuijts R, Levy P, Haefliger E, Alfonso JF. Visual function after bilateral implantation of apodized diffractive aspheric multifocal intraocular lenses with a C3.0 D addition. J Cataract Refract Surg 2009; 35: Mangione CM, Lee PP, Gutierrez PR, Spritzer K, Berry S, Hays RD; for the National Eye Institute Visual Function Questionnaire Field Test Investigators. Development of the 25-item National Eye Institute Visual Function Questionnaire. Arch Ophthalmol 2001; 119: Available at: ama-assn.org/cgi/reprint/119/7/1050.pdf. Accessed July 7, Javitt JC, Wang F, Trentacost DJ, Rowe M, Tarantino N. Outcomes of cataract extraction with multifocal intraocular lens implantation; functional status and quality of life. Ophthalmology 1997; 104: Cuq C, Spera C, Laurendeau C, Lafuma A, Berdeaux G. Intermediate visual acuity without spectacles following bilateral ReSTOR implantation. Eur J Ophthalmol 2008; 18: Thibos LN, Horner D. Power vector analysis of the optical outcome of refractive surgery. J Cataract Refract Surg 2001; 27: Kohnen T. New abbreviations for visual acuity values [editorial]. J Cataract Refract Surg 2009; 35: Alfonso JF, Fernandez-Vega L, Se~naris A, Montes-Mico R. Prospective study of the Acri.LISA bifocal intraocular lens. J Cataract Refract Surg 2007; 33: Kaymak H, Mester U. Erste Ergebnisse mite inter neuen aberrationskorrigierten intraokularen Bifokallinse [First results with a new aberration correcting bifocal intraocular lens]. Ophthalmologe 2007; 104: Blaylock JF, Si Z, Vickers C. Visual and refractive status at different focal distances after implantation of the ReSTOR multifocal intraocular lens. J Cataract Refract Surg 2006; 32: Chiam PJT, Chan JH, Aggarwal RK, Kasaby S. ReSTOR intraocular lens implantation in cataract surgery: quality of vision. J Cataract Refract Surg 2006; 32: ; errata, Alfonso JF, Fernandez-Vega L, Baamonde MB, Montes-Mico R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg 2007; 33: Alfonso JF, Fernandez-Vega L, Amhaz H, Montes-Mico R, Valcarcel B, Ferrer-Blasco T. Visual function after implantation of an aspheric bifocal intraocular lens. J Cataract Refract Surg 2009; 35: Alfonso JF, Fernandez-Vega L, Valcarcel B, Montes-Mico R. Visual performance after AcrySof ReSTOR aspheric intraocular lens implantation. J Optom 2008; 1: Available at: Original%20Article.pdf. Accessed July 7, Chiam PJT, Chan JH, Haider SI, Karia N, Kasaby H, Aggarwal RK. Functional vision with bilateral ReZoom and ReSTOR intraocular lenses 6 months after cataract surgery. J Cataract Refract Surg 2007; 33: Martınez Palmer A, Gomez Fai~na P, Espa~na Albelda A, Comas Serrano M, Nahra Saad D, Castilla Cespedes M. Visual function with bilateral implantation of monofocal and multifocal intraocular lenses: a prospective, randomized, controlled clinical trial. J Refract Surg 2008; 24:

10 2052 VISUAL PERFORMANCE WITH A NEW BILATERAL ZONAL REFRACTIVE IOL 37. Zhao G, Zhang J, Zhou Y, Hu L, Che C, Jiang N. Visual function after monocular implantation of apodized diffractive multifocal or single-piece monofocal intraocular lens; randomized prospective comparison. J Cataract Refract Surg 2010; 36: Lane SS, Javitt JC, Nethery DA, Waycaster C. Improvements in patient-reported outcomes and visual acuity after bilateral implantation of multifocal intraocular lenses with C3.0 diopter addition: multicenter clinical trial. J Cataract Refract Surg 2010; 36: Alfonso JF, Fernandez-Vega L, Valcarcel B, Ferrer-Blasco T, Montes-Mico R. Outcomes and patient satisfaction after presbyopic bilateral lens exchange with the ReSTOR IOL in emmetropic patients. J Refract Surg 2010; 26: First author: Gonzalo Mu~noz, MD, PhD, FEBO Refractive Surgery Department, Clínica Oftalmologica Marques de Sotelo, Valencia, Spain

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life

Providing Optimal Optics For Your Astigmatic Cataract Patients. While the cornea remains relatively stable and prolate throughout life Providing Optimal Optics For Your Astigmatic Cataract Patients David I. Geffen, OD, FAAO Why keep the crystalline lens? While the cornea remains relatively stable and prolate throughout life Unless we

More information

EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com

EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com ejpmr, 2015,2(3), 436-440 EUROPEAN JOURNAL OF PHARMACEUTICAL AND MEDICAL RESEARCH www.ejpmr.com Tumram et al. SJIF Impact Factor 2.026 Research Article ISSN 3294-3211 EJPMR CLINICAL OUTCOME OF TORIC IOL

More information

Faster recovery of visual acuity at all distances

Faster recovery of visual acuity at all distances news Special issue PresbyMAX Faster recovery of visual acuity at all distances PresbyMAX μ-monovision: The latest technique for treating presbyopia SCHWIND now offers PresbyMAX µ-monovision as a new option

More information

The Oculentis Toolbox for Lens Surgery

The Oculentis Toolbox for Lens Surgery Supplement to January 2015 Sponsored by Oculentis The Oculentis Toolbox for Lens Surgery LENTIS Comfort for cataract surgery, LENTIS Mplus Family for refractive cataract surgery, FEMTIS for laser-assisted

More information

Clinical Evaluation. SeeLens MF Intraocular Lens

Clinical Evaluation. SeeLens MF Intraocular Lens 1. HANITA LENSES Clinical Evaluation Phase 1 Report (9-12 month follow up) SeeLens MF Intraocular Lens 4 August 2011 Version 2 Page 1 of 21 CONTENTS 1. OBJECTIVES 2 2. EFFICACY AND SAFETY ASSESSMENTS 2

More information

AcrySof IQ Toric IOL (SN6ATT) Surgeon Keys for Success & Acknowledgement

AcrySof IQ Toric IOL (SN6ATT) Surgeon Keys for Success & Acknowledgement AcrySof IQ Toric IOL (SN6ATT) Surgeon Keys for Success & Acknowledgement Alcon Laboratories, Inc is pleased to announce the availability of the AcrySof IQ Toric intraocular lens, and delighted that you

More information

Premium IOL Implantation Calculations in Post-LASIK Cataract Eyes Using ASCRS IOL Calculator

Premium IOL Implantation Calculations in Post-LASIK Cataract Eyes Using ASCRS IOL Calculator Premium IOL Implantation Calculations in Post-LASIK Cataract Eyes Using ASCRS IOL Calculator Sahiba K Chailertborisuth, Saneha K. C. Borisuth, Navaneet S.C. Borisuth, MD, PhD Virdi Eye Clinic & Laser Vision

More information

OMNI EYE SPECIALISTS. The Intraocular Lens

OMNI EYE SPECIALISTS. The Intraocular Lens OMNI EYE SPECIALISTS Intraocular Lens Options In Cataract Surgery Or Refractive Lens Exchange Surgery Your decision to have vision surgery is important and the doctors at Omni Eye Specialists will explain

More information

Informed Consent for Cataract Surgery and/or Implantation of an Intraocular Lens (IOL)

Informed Consent for Cataract Surgery and/or Implantation of an Intraocular Lens (IOL) Bruce H. Brumm, MD, PC 6751 North 72 nd Street, Ste 105 Omaha, NE 68122 (402) 572-2020 800-775-5909 www.brummeye.com Informed Consent for Cataract Surgery and/or Implantation of an Intraocular Lens (IOL)

More information

Tucson Eye Care, PC. Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens

Tucson Eye Care, PC. Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens Tucson Eye Care, PC Informed Consent for Cataract Surgery And/Or Implantation of an Intraocular Lens INTRODUCTION This information is provided so that you may make an informed decision about having eye

More information

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant

ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant ADDENDUM to the Informed Consent for Cataract Surgery with Intraocular Lens Implant INTRODUCTION Except for unusual situations, a cataract operation is indicated only when you cannot function satisfactorily

More information

Informed Consent for Refractive Lens Exchange (Clear Lens Extraction)

Informed Consent for Refractive Lens Exchange (Clear Lens Extraction) Informed Consent for Refractive Lens Exchange (Clear Lens Extraction) This form is designed to ensure that you have all the information you need to make a decision about whether or not you wish to undergo

More information

Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation

Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation Wavefront-guided Excimer Laser Vision Correction After Multifocal IOL Implantation Bettina B. Jendritza, MD; Michael C. Knorz, MD; Steve Morton ABSTRACT PURPOSE: To investigate the use of wavefront-guided

More information

Astigmatism and vision: Should all astigmatism always be corrected? 1 Ophthalmic Research Group, Aston University, Birmingham, UK

Astigmatism and vision: Should all astigmatism always be corrected? 1 Ophthalmic Research Group, Aston University, Birmingham, UK Astigmatism and vision: Should all astigmatism always be corrected? James S Wolffsohn 1, Gurpreet Bhogal 1 and Sunil Shah 1,2 1 Ophthalmic Research Group, Aston University, Birmingham, UK 2 Midland Eye,

More information

LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted

LASIK. Complications. Customized Ablations. Photorefractive Keratectomy. Femtosecond Keratome for LASIK. Cornea Resculpted Refractive Surgery: Which Procedure for Which Patient? David R. Hardten, M.D. Minneapolis, Minnesota Have done research, consulting, or speaking for: Alcon, Allergan, AMO, Bausch & Lomb, Inspire, Medtronic,

More information

What are your options for correcting astigmatism?

What are your options for correcting astigmatism? What are your options for correcting astigmatism? If you depend upon eyeglasses, you may have experienced the inconvenience of not being able to find your glasses when you really need them. Eyeglasses

More information

Explanation of the Procedure

Explanation of the Procedure Informed Consent Cataract Surgery with Intraocular Lens Implant Please initial below indicating that you have read and understand each section Introduction The internal lens of the eye can become cloudy

More information

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement) Mark Packer, M.D. Informed Consent for Refractive Lens Exchange (Clear Lens Replacement) This surgery involves the removal of the natural lens of my eye, even though it is not a cataract. The natural lens

More information

Multifocal Intraocular Lenses

Multifocal Intraocular Lenses Ophthalmol Clin N Am 19 (26) 89 15 Multifocal Intraocular Lenses Stephen S. Lane, MDT, Mike Morris, PhD, Lee Nordan, MD, Mark Packer, MD, FACS, Nicholas Tarantino, OD, R. Bruce Wallace III, MD, FACS Ophthalmology

More information

Patient-Reported Outcomes with LASIK (PROWL-1) Results

Patient-Reported Outcomes with LASIK (PROWL-1) Results Patient-Reported Outcomes with LASIK (PROWL-1) Results Elizabeth M. Hofmeister, MD CAPT, MC, USN Naval Medical Center San Diego Refractive Surgery Advisor for Navy Ophthalmology Assistant Professor of

More information

Supplement to January 2012. and. Advanced multifocal IOL technology for the treatment of presbyopia, astigmatism, and cataract.

Supplement to January 2012. and. Advanced multifocal IOL technology for the treatment of presbyopia, astigmatism, and cataract. Supplement to January 2012 and Advanced multifocal IOL technology for the treatment of presbyopia, astigmatism, and cataract. LENTIS Mplus and LENTIS Mplus Toric LENTIS Mplus and LENTIS Mplus Toric Advanced

More information

Seeing Beyond the Symptoms

Seeing Beyond the Symptoms Seeing Beyond the Symptoms Cataracts are one of the leading causes of vision impairment in the United States. 1 However, because cataracts form slowly and over a long period of time, many people suffer

More information

Current Trends in Modern Cataract Surgery. Jose Ivan Quiceno, MD Division of Ophthalmology Scripps Clinic Medical Group

Current Trends in Modern Cataract Surgery. Jose Ivan Quiceno, MD Division of Ophthalmology Scripps Clinic Medical Group Current Trends in Modern Cataract Surgery Jose Ivan Quiceno, MD Division of Ophthalmology Scripps Clinic Medical Group Current Trends in Modern Cataract Surgery Modern cataract surgery is not only a rehabilitative

More information

Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens

Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens INTRODUCTION This information is given to you so that you can make an informed decision about having eye surgery. Take as

More information

Common visual problems in older LASIK patients

Common visual problems in older LASIK patients 丘 子 宏 LASIK 手 術 後 的 視 覺 Visual acuity:the measurement of high contrast Snellen acuity but not other functions under different condition Quality of vision: measure the visual functions in variable condition

More information

BACKGROUND INFORMATION AND INFORMED CONSENT FOR CATARACT SURGERY AND IMPLANTATION OF AN INTRAOCULAR LENS

BACKGROUND INFORMATION AND INFORMED CONSENT FOR CATARACT SURGERY AND IMPLANTATION OF AN INTRAOCULAR LENS OMNI EYE SPECIALISTS A Madison Street Company Proudly Owned by Employees Specializing in Medical and Surgical Care of the Eye 55 Madison St, Suite 355 Denver CO 80206 303-377-2020 800-GO-2-OMNI www.omnieye.com

More information

CONSENT FOR CATARACT SURGERY

CONSENT FOR CATARACT SURGERY CONSENT FOR CATARACT SURGERY INTRODUCTION The natural crystalline lens can become cloudy due to aging, trauma, disease or medications. This condition is called a cataract. When cataracts become visually

More information

Incision along Steep Axis

Incision along Steep Axis Toric IOL An option or a must? ~ 15% cataract surgical patients >1.5 D Options: spectacles, CLs, Incision along steep axis, LRI, AK, toric IOL, Excimer Laser or a combination Walter J. Stark, MD Professor

More information

Pre-Op to Post-Op: A Complete Optometrist s Guide to Advanced Technology IOLs.

Pre-Op to Post-Op: A Complete Optometrist s Guide to Advanced Technology IOLs. Pre-Op to Post-Op: A Complete Optometrist s Guide to Advanced Technology IOLs. Your Role in Guiding Patients Through Cataract Surgery As an optometrist, you ve been guiding your patients through their

More information

The pinnacle of refractive performance.

The pinnacle of refractive performance. Introducing! The pinnacle of refractive performance. REFRACTIVE SURGERY sets a new standard in LASIK outcomes More than 98% of patients would choose it again. 1 It even outperformed glasses and contacts

More information

ClearVision. Pacific ClearVision Institute. Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens

ClearVision. Pacific ClearVision Institute. Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens ClearVision Pacific ClearVision Institute Informed Consent For Cataract Surgery And/Or Implantation of an Intraocular Lens INTRODUCTION This information is given to you so that you can make an informed

More information

Comparison of Higher Order Aberrations and Contrast Sensitivity After LASIK, Verisyse Phakic IOL, and Array Multifocal IOL

Comparison of Higher Order Aberrations and Contrast Sensitivity After LASIK, Verisyse Phakic IOL, and Array Multifocal IOL ORIGINAL ARTICLES Comparison of Higher Order Aberrations and Contrast Sensitivity After LASIK, Verisyse Phakic IOL, and Array Multifocal IOL Sasivimon Chandhrasri, MD; Michael C. Knorz, MD ABSTRACT PURPOSE:

More information

Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia

Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia Pseudo-accommodative Cornea (PAC) for the Correction of Presbyopia Alaa El Danasoury, FRCS Magrabi Hospitals & Centers Surgical options for the Correction of Presbyopia Monovision Reversal of Presbyopia:

More information

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay

Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay Comparison Combined LASIK Procedure for Ametropic Presbyopes and Planned Dual Interface for Post-LASIK Presbyopes Using Small Aperture Corneal Inlay Minoru Tomita, MD, PhD 1,2 1) Shinagawa LASIK, Tokyo,

More information

Custom Cataract Surgery. Laser and Lens Options

Custom Cataract Surgery. Laser and Lens Options Custom Cataract Surgery Laser and Lens Options What is a Cataract? More than fifty percent of people over the age of 60, and quite a few younger than that, suffer from cataracts. In fact, cataracts are

More information

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University.

IOL Power Calculation After Myopic LASIK. Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University. IOL Power Calculation After Myopic LASIK Hany Helaly, Lecturer of Ophthalmology, Faculty of Medicine, Alexandria University. SUPERVISORS Prof. Dr. Mohammad El-Hifnawy Professor of Ophthalmology Faculty

More information

Helping Patients to See the Maximum: New Treatment Options in Presbyopia and Astigmatic Correction

Helping Patients to See the Maximum: New Treatment Options in Presbyopia and Astigmatic Correction Supplement Dec 2015/Jan 2016 Helping Patients to See the Maximum: New Treatment Options in Presbyopia and Astigmatic Correction The Extended Range of Vision IOL A proven concept for presbyopia correction

More information

Optimizing refractive outcomes

Optimizing refractive outcomes APACRS The news magazine of the Asia-Pacific Association of Cataract & Refractive Surgeons Supplement to EyeWorld Asia-Pacific Winter 2014 Asia-Pacific Surgical management of astigmatism Sponsored by Abbott

More information

Insert to October 2013 THE COMPLETE PICTURE. Experiences with the ALADDIN system. Sponsored by Topcon

Insert to October 2013 THE COMPLETE PICTURE. Experiences with the ALADDIN system. Sponsored by Topcon Insert to October 2013 THE COMPLETE PICTURE Experiences with the ALADDIN system Sponsored by Topcon THE COMPLETE PICTURE Advantages of The Aladdin How this all-in-one optical biometer edged out the IOLMaster

More information

Cataract Surgery after Myopic Refractive Procedures. Ray Guard Eye Center Huang Wei-Jen, MD

Cataract Surgery after Myopic Refractive Procedures. Ray Guard Eye Center Huang Wei-Jen, MD Cataract Surgery after Myopic Refractive Procedures Ray Guard Eye Center Huang Wei-Jen, MD Financial Disclosures : * No financial interest on products mentioned Cataract Surgery after Myopic Refractive

More information

KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC,

KERATOCONUS IS A BILATERAL, ASYMMETRIC, CHRONIC, Comparison of and Intacs for Keratoconus and Post-LASIK Ectasia MUNISH SHARMA, MD, AND BRIAN S. BOXER WACHLER, MD PURPOSE: To evaluate the efficacy of single-segment Intacs and compare with double-segment

More information

The Bladeless Laser Cataract Surgery

The Bladeless Laser Cataract Surgery I HAVE SUDDENLY STARTED SEEING BLURRY. WHY WOULD THAT BE? THIS COULD BE DUE TO CATARACT A cataract is clouding of the lens in the eye that affects vision. Most cataracts are related to aging. A cataract

More information

Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens

Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens Standardized Analyses of Correction of Astigmatism With the Visian Toric Phakic Implantable Collamer Lens Donald R. Sanders, MD, PhD; Edwin J. Sarver, PhD ABSTRACT PURPOSE: To demonstrate the methodology

More information

Cataracts & Astigmatism? Quality Distance Vision in Just One Step

Cataracts & Astigmatism? Quality Distance Vision in Just One Step Cataracts & Astigmatism? Quality Distance Vision in Just One Step What would it be like to say goodbye to both cataracts and astigmatism at the same time? Imagine being able to see and enjoy quality distance

More information

VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment)

VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment) CustomVue Advantage Patient Information Sheet VISX Wavefront-Guided LASIK for Correction of Myopic Astigmatism, Hyperopic Astigmatism and Mixed Astigmatism (CustomVue LASIK Laser Treatment) Statements

More information

Clinical Study Visual and Refractive Outcomes of a Toric Presbyopia-Correcting Intraocular Lens

Clinical Study Visual and Refractive Outcomes of a Toric Presbyopia-Correcting Intraocular Lens Ophthalmology Volume 16, Article ID 745821, 6 pages http://dx.doi.org/1.1155/16/745821 Clinical Study Visual and Refractive Outcomes of a Toric Presbyopia-Correcting Intraocular Lens Alice T. Epitropoulos

More information

Everything New with IOL and Cataract Surgery

Everything New with IOL and Cataract Surgery Everything New with IOL and Cataract Surgery What an Optometrists Needs to Know Jim Owen, OD, MBA, FAAO Cataract Surgery Eye Care Still the most common eye surgery performed Aging Population makes it even

More information

No-history method of intraocular lens power calculation for cataract surgery after myopic laser in situ keratomileusis

No-history method of intraocular lens power calculation for cataract surgery after myopic laser in situ keratomileusis J CATARACT REFRACT SURG - VOL 33, JANUARY 2007 No-history method of intraocular lens power calculation for cataract surgery after myopic laser in situ keratomileusis H. John Shammas, MD, Maya C. Shammas,

More information

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening

THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening THE BEST OF BOTH WORLDS Dual-Scheimpflug and Placido Reaching a new level in refractive screening GALILEI G4 Clinical Applications Corneal Implant Planning The comes with a licensable corneal inlay software

More information

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement)

Informed Consent for Refractive Lens Exchange (Clear Lens Replacement) Drs. Fine, Hoffman and Packer, LLC PHYSICIANS AND SURGEONS, EyeMDs OPHTHALMOLOGY I. Howard Fine, M.D. Richard S. Hoffman, M.D. Mark Packer, M.D. 1550 Oak Street, Suite 5 www.finemd.com Eugene, OR 97401-7701

More information

Advanced Eyecare of Orange County Kim Doan, M.D. Eye Physician & Surgeon

Advanced Eyecare of Orange County Kim Doan, M.D. Eye Physician & Surgeon Advanced Eyecare of Orange County Kim Doan, M.D. Eye Physician & Surgeon Patient Information Sheet: Cataract Surgery And/Or Implantation of an Intraocular Lens This information is given to you so that

More information

Phacoemulsification: Considerations for Astigmatism Management Jason P. Brinton, MD and Thomas A. Oetting, MS, MD June 10, 2011

Phacoemulsification: Considerations for Astigmatism Management Jason P. Brinton, MD and Thomas A. Oetting, MS, MD June 10, 2011 Phacoemulsification: Considerations for Astigmatism Management Jason P. Brinton, MD and Thomas A. Oetting, MS, MD June 10, 2011 Introduction Patient expectations for cataract surgery have increased significantly

More information

The ASCRS Clinical Survey revealed

The ASCRS Clinical Survey revealed Supplement to EyeWorld September 2015 Supplement to EyeWorld September 2015 Modernizing cataract surgery: Navigating success with toric multifocal IOLs Supported by an educational grant from Abbott Medical

More information

What is the main target for all phaco surgeons?

What is the main target for all phaco surgeons? CORRECTION C O OF ASTIGMATISM DURING CATARACT SURGERY Abdallah dllh K. Hassouna, M.D. Sherein S. Wahba, M.D. Ain Shams University 2009 Main target What is the main target for all phaco surgeons? Main Target

More information

Premium Lenses in Problematic Patients

Premium Lenses in Problematic Patients Premium Lenses in Problematic Patients How to be successful with presbyopic lenses in eyes that have already had LASIK surgery. Walter Bethke, Managing Editor If implanting a monofocal intraocular lens

More information

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the

INTRACOR. An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the INTRACOR An excerpt from the presentations by Dr Luis Ruiz and Dr Mike Holzer and the Round Table discussion moderated by Dr Wing-Kwong Chan in the 1 Dr Luis Ruiz Presbyopia treatment with INTRACOR Luis

More information

INFORMED CONSENT FOR CATARACT AND LENS IMPLANT SURGERY

INFORMED CONSENT FOR CATARACT AND LENS IMPLANT SURGERY INFORMED CONSENT FOR CATARACT AND LENS IMPLANT SURGERY 1. GENERAL INFORMATION This information is given to you to help you make an informed decision about having cataract and/or lens implant surgery. Once

More information

What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director Shinagawa LASIK Center, Tokyo, JAPAN

What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director Shinagawa LASIK Center, Tokyo, JAPAN What now, when Presbyopia sets in? Minoru Tomita, MD, PhD Executive Director, Tokyo, JAPAN Financial disclosure Travel expenses for the ASCRS meeting were paid by Ziemer and AcuFocus, Inc. Presbyopia:

More information

Wavefront technology has been used in our

Wavefront technology has been used in our Wavefront Customized Ablations With the WASCA Asclepion Workstation Sophia I. Panagopoulou, BSc; Ioannis G. Pallikaris, MD ABSTRACT PURPOSE: WASCA (Wavefront Aberration Supported Cornea Ablation) is a

More information

Presbyopia modern surgical treatments. Johan Lønsmann Poulsen, Euroeyes

Presbyopia modern surgical treatments. Johan Lønsmann Poulsen, Euroeyes Presbyopia modern surgical treatments Johan Lønsmann Poulsen, Euroeyes Presbyopia the earliest treatments Reading stones Emperor Nero (37-68 AD) Presbyopia design trends c. 1400s c. 1680-1750 c. 1800 1850

More information

WHAT IS A CATARACT, AND HOW IS IT TREATED?

WHAT IS A CATARACT, AND HOW IS IT TREATED? 4089 TAMIAMI TRAIL NORTH SUITE A103 NAPLES, FL 34103 TELEPHONE (239) 262-2020 FAX (239) 435-1084 DOES THE PATIENT NEED OR WANT A TRANSLATOR, INTERPRETOR OR READER? YES NO TO THE PATIENT: You have the right,

More information

Cataract Surgery in Small Eyes Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Oregon Health & Science University No Financial Interests Anatomic Classification Short AC depth with

More information

Uncorrected astigmatism can cause visual

Uncorrected astigmatism can cause visual Spatial Vision and Astigmatism Correction What is the minimum amount of astigmatism that should be corrected? By Eloy A. Villegas, PhD; and Pablo Artal, PhD Uncorrected astigmatism can cause visual quality

More information

Glenn B. Cook, M.D., Ph.D. DIPLOMATE AMERICAN BOARD OF OPHTHALMOLOGY Coronado Eye Associates

Glenn B. Cook, M.D., Ph.D. DIPLOMATE AMERICAN BOARD OF OPHTHALMOLOGY Coronado Eye Associates Glenn B. Cook, M.D., Ph.D. INFORMED CONSENT FOR CATARACT OPERATION AND/OR IMPLANTATION OF INTRAOCULAR LENS INTRODUCTION: RIGHT / LEFT This information is given to you so that you can make an informed decision

More information

Increasing cost of health care 2010 Prevalence of Cataract Patients: 24.4 M (NEI/NIH) of patients age 40+

Increasing cost of health care 2010 Prevalence of Cataract Patients: 24.4 M (NEI/NIH) of patients age 40+ 7/27/15 Comanaging Modern Cataract Surgery: Guiding patients through Advanced Technology and Optimizing Surgical Outcomes Disclosures: Robert S. Stutman, OD, MBA, FAAO Alcon Speaker s Alliance Select Eye

More information

Cataract Testing. What a Patient undergoes prior to surgery

Cataract Testing. What a Patient undergoes prior to surgery Cataract Testing What a Patient undergoes prior to surgery FINANCIAL DISCLOSURE I have no financial interest or relationships to disclose What do most Technicians find to be the most mundane yet very important

More information

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State

Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State Curtin G. Kelley, M.D. Director of Vision Correction Surgery Arena Eye Surgeons Associate Clinical Professor of Ophthalmology The Ohio State University Columbus, Ohio Refractive Errors Myopia (nearsightedness)

More information

Cataract Surgery Patient Information

Cataract Surgery Patient Information Cataract Patient Information 1. Within the human eye, there is a normal structure called the lens. In youth, this lens is clear, and light rays pass through and are focused by this lens as well as the

More information

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US

FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US FIRST EXPERIENCE WITH THE ZEISS FEMTOSECOND SYSTEM IN CONJUNC- TION WITH THE MEL 80 IN THE US JON DISHLER, MD DENVER, COLORADO, USA INTRODUCTION AND STUDY OBJECTIVES This article summarizes the first US

More information

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical,

Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical, sight see OLYMPIA EYE & LASER CENTRE Your one stop vision centre Our ophthalmic centre offers comprehensive eye management, which includes medical, At the Olympia Eye & Laser Centre, our vision is to improve

More information

When To Laser, When To Implant, When To Do Both

When To Laser, When To Implant, When To Do Both When To Laser, When To Implant, When To Do Both Scott MacRae, MD Professor of Ophthalmology Professor of Visual Sciences StrongVision Refractive Surgery Center University of Rochester Eye Institute Refractive

More information

Timothy D. McGarity, M.D.

Timothy D. McGarity, M.D. Timothy D. McGarity, M.D. Dr. McGarity received his B.S. in Microbiology, honors, at the University of Arkansas, Fayetteville, and his M.D. at the University of Arkansas for Medical Sciences, Little Rock.

More information

Management of Astigmatism in Cataract Surgery

Management of Astigmatism in Cataract Surgery Management of Astigmatism in Cataract Surgery Jonathan B. Rubenstein, M Vice- Chairman and eutsch Family Professor of Ophthalmology irector of Refractive Surgery Rush University Medical Center Financial

More information

Information For Consent For Cataract Surgery

Information For Consent For Cataract Surgery Information For Consent For Cataract Surgery Your Ophthalmologist has diagnosed you with a visually significant cataract. The following handout will explain your condition and give you the information

More information

IBRA Professional. Leading Technology for Refractive Outcome Analysis. Brochure 2015

IBRA Professional. Leading Technology for Refractive Outcome Analysis. Brochure 2015 IBRA Professional Leading Technology for Refractive Outcome Analysis Brochure 2015 Introduction IBRA is the leading all-in-one system for refractive laser and lens surgery outcomes, allowing eye surgeons

More information

Informed Consent for Cataract Surgery or Clear Lens Extraction with Implantation of an Intraocular Lens

Informed Consent for Cataract Surgery or Clear Lens Extraction with Implantation of an Intraocular Lens Informed Consent for Cataract Surgery or Clear Lens Extraction with Implantation of an Intraocular Lens Please read the following pages carefully, and initial and sign where indicated. Please do not sign

More information

VA high quality, complications low with phakic IOL

VA high quality, complications low with phakic IOL Page 1 of 5 VA high quality, complications low with phakic IOL Use in keratoconus will continue, one surgeon predicts; another ponders long-term safety Nov 1, 2007 By:Nancy Groves Ophthalmology Times Several

More information

Table of Contents. Page. Introducing Crystalens 2. What is a cataract 3. How the eye works 4. Vision restoration options 5

Table of Contents. Page. Introducing Crystalens 2. What is a cataract 3. How the eye works 4. Vision restoration options 5 P/N 50-0065, Revision B, p.1 PATIENT LABELING Table of Contents Introducing Crystalens 2 Page What is a cataract 3 How the eye works 4 Vision restoration options 5 Are you a candidate for the Crystalens?

More information

Clinical Study Minimizing Surgically Induced Astigmatism at the Time of Cataract Surgery Using a Square Posterior Limbal Incision

Clinical Study Minimizing Surgically Induced Astigmatism at the Time of Cataract Surgery Using a Square Posterior Limbal Incision Ophthalmology Volume 2011, Article ID 243170, 4 pages doi:10.1155/2011/243170 Clinical Study Minimizing Surgically Induced Astigmatism at the Time of Cataract Surgery Using a Square Posterior Limbal Incision

More information

Non-surgical Presbyopia correction

Non-surgical Presbyopia correction Non-surgical Presbyopia correction Presbyopia is the decreasing age-related ability of the eye to focus on objects at close distances! The rate of decline in near visual work is faster compared to 100

More information

Physics 1230: Light and Color

Physics 1230: Light and Color Physics 1230: Light and Color The Eye: Vision variants and Correction http://www.colorado.edu/physics/phys1230 What does 20/20 vision mean? Visual acuity is usually measured with a Snellen chart Snellen

More information

Quality Control in Refractive Surgery

Quality Control in Refractive Surgery Quality Control in Refractive Surgery Stefan Pieger*, M.Sc. Ingenieurbüro Pieger GmbH, Wendelstein, Germany Introduction Personal experience as application specialist for refractive excimer lasers since

More information

Techniques for Enhancing Cataract Surgery Patients with Residual Refractive Error. Director of Cornea Center For Excellence In Eye Care Miami, FL

Techniques for Enhancing Cataract Surgery Patients with Residual Refractive Error. Director of Cornea Center For Excellence In Eye Care Miami, FL Techniques for Enhancing Cataract Surgery Patients with Residual Refractive Error William Trattler, MD Director of Cornea Center For Excellence In Eye Care Miami, FL 1 Financial Disclosures Consulting

More information

CORRECT YOUR CATARACTS AND YOUR VISION

CORRECT YOUR CATARACTS AND YOUR VISION CORRECT YOUR CATARACTS AND YOUR VISION CATARACT INFORMATION GUIDE The day you have cataract surgery marks the beginning of a whole new world of possibilities from reduced dependence on glasses to having

More information

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy

Keratorefractive Surgery for Post-Cataract Refractive Surprise. Moataz El Sawy Keratorefractive Surgery for Post-Cataract Refractive Surprise Moataz El Sawy Departmentof Ophthalmology, Faculty of Medicine,MenoufiyaUniversity, Egypt [email protected] Abstract: Purpose: To evaluate

More information

VISION PATTERN. What Can you Expect from the ReSTOR procedure?

VISION PATTERN. What Can you Expect from the ReSTOR procedure? RESTOR Presbyopia is the normal aging process that everyone experiences over the age of 40 which affects near vision. This usually results in the need for reading glasses or bifocals. The crystalline lens

More information

How To Treat Eye Sight Problems With Eye Care

How To Treat Eye Sight Problems With Eye Care Dealing With The Refractive Surprise Richard S. Hoffman, MD Clinical Associate Professor of Ophthalmology Casey Eye Institute Oregon Health and Science University No Financial Interest Why The Surprise?

More information

THE GUIDE TO REFRACTIVE LENS EXCHANGE SEE CLEARLY.

THE GUIDE TO REFRACTIVE LENS EXCHANGE SEE CLEARLY. THE GUIDE TO REFRACTIVE LENS EXCHANGE SEE CLEARLY. EVERYBODY WANTS TO SEE CLEARLY Many of us take our sight for granted, whether it s forgetting how often we rely on it to guide us through our day-to-day

More information

Ocularis Pharma, LLC. The Next Wave in Ophthalmic Drugs: Managing the Pupil to Improve Night Vision Performance Ocularis Pharma, LLC November, 2013

Ocularis Pharma, LLC. The Next Wave in Ophthalmic Drugs: Managing the Pupil to Improve Night Vision Performance Ocularis Pharma, LLC November, 2013 Ocularis Pharma, LLC The Next Wave in Ophthalmic Drugs: Managing the Pupil to Improve Night Vision Performance Ocularis Pharma, LLC November, 2013 Ocularis Pharma, LLC Page 2 The Next Wave in Ophthalmic

More information

Benefits and Harms of Femtosecond Laser Assisted Cataract Surgery: A Systematic Review

Benefits and Harms of Femtosecond Laser Assisted Cataract Surgery: A Systematic Review 4 Department of Veterans Affairs Health Services Research & Development Service Benefits and Harms of Femtosecond Laser Assisted Cataract Surgery: A Systematic Review December 2013 Prepared for: Department

More information

True Solutions for Presbyopia With Laser Technology Capturing the growing presbyopia market sector.

True Solutions for Presbyopia With Laser Technology Capturing the growing presbyopia market sector. Supplement to June 2010 True Solutions for Presbyopia With Laser Technology Capturing the growing presbyopia market sector. Sponsored by an educational grant from Technolas Perfect Vision Technolas Perfect

More information

Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser

Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser Active Cyclotorsion Error Correction During LASIK for Myopia and Myopic Astigmatism With the NIDEK EC-5000 CX III Laser Sudhank Bharti, MD; Harkaran S. Bains ABSTRACT PURPOSE: To investigate the predictability

More information

Presbyopia Treatment by Monocular Peripheral PresbyLASIK

Presbyopia Treatment by Monocular Peripheral PresbyLASIK Presbyopia Treatment by Monocular Peripheral PresbyLASIK Robert Leonard Epstein, MD, MSEE; Mark Andrew Gurgos, COA ABSTRACT PURPOSE: To investigate monocular peripheral presby- LASIK on the non-dominant

More information

Oregon Eye Specialists, PC YOUR GUIDE CATARACT SURGERY. Improving VISION. Improving LIFE.

Oregon Eye Specialists, PC YOUR GUIDE CATARACT SURGERY. Improving VISION. Improving LIFE. Oregon Eye Specialists, PC YOUR GUIDE TO CATARACT SURGERY Dinelli M. Monson, M.D. Comprehensive Ophthalmology Physician and Surgeon Tualatin Clinic: 19250 SW 65 th Ave, Ste 215 503.692.3630 Newberg Clinic:

More information

Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda

Customized corneal ablation can be designed. Slit Skiascopic-guided Ablation Using the Nidek Laser. Scott MacRae, MD; Masanao Fujieda Slit Skiascopic-guided Ablation Using the Nidek Laser Scott MacRae, MD; Masanao Fujieda ABSTRACT PURPOSE: To present the approach of using a scanning slit refractometer (the ARK 10000) in conjunction with

More information