Practical guide for quantitative 1D NMR integration Eugenio Alvarado, University of Michigan, 05/10/10



Similar documents
Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

Longitudinal Relaxation Time (T 1 ) Measurement for 1 H on VnmrJ2.2D (use on Ra, Mut, and Isis)

Nuclear Magnetic Resonance Spectroscopy

13C NMR Spectroscopy

Spin-Lattice Relaxation Times

How To Understand The Physics Of Ft Nmr

Electronic Supplementary Information

Proton Nuclear Magnetic Resonance Spectroscopy

Signal to Noise Instrumental Excel Assignment

Nuclear Magnetic Resonance Spectroscopy

Proton Nuclear Magnetic Resonance Spectroscopy

Basic NMR Concepts: A Guide for the Modern Laboratory

Signal Manipulation. time domain NMR signal in MHz range is converted to khz (audio) range by mixing with the reference ( carrier ) frequency

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

How To Understand The Measurement Process

Nuclear Magnetic Resonance notes

The Hydrogen Atom Is a Magnet.

Lecture #7 (2D NMR) Utility of Resonance Assignments

NMR - Basic principles

Proton Nuclear Magnetic Resonance ( 1 H-NMR) Spectroscopy

Nuclear Magnetic Resonance (NMR) Spectroscopy

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR)

Used to determine relative location of atoms within a molecule Most helpful spectroscopic technique in organic chemistry Related to MRI in medicine

Module 13 : Measurements on Fiber Optic Systems

NMR SPECTROSCOPY. Basic Principles, Concepts, and Applications in Chemistry. Harald Günther University of Siegen, Siegen, Germany.

Trans Fats. What is a trans fat? Trans fatty acids, or trans fats as they are known, are certain

For example: (Example is from page 50 of the Thinkbook)

Proton NMR. One Dimensional H-NMR. Cl S. Common types of NMR experiments: 1-H NMR

Nuclear Magnetic Resonance

Organic Chemistry Tenth Edition

Determination of Molecular Structure by MOLECULAR SPECTROSCOPY

NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE δ PUBLISHING

Background A nucleus with an odd atomic number or an odd mass number has a nuclear spin that can be observed by NMR spectrometers.

Solving Spectroscopy Problems

Massachusetts Institute of Technology Department of Chemistry 5.33 Advanced Chemical Instrumentation FALL SEMESTER 2005

Experiment 5. Lasers and laser mode structure

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

Nuclear Magnetic Resonance (NMR) Wade Textbook

Chapter 11 Structure Determination: Nuclear Magnetic Resonance Spectroscopy. Nuclear Magnetic Resonance Spectroscopy Nuclear Magnetic Resonance

Bruker AMX 500 MHz NMR

INFRARED SPECTROSCOPY (IR)

The Four Questions to Ask While Interpreting Spectra. 1. How many different environments are there?

NMR Spectroscopy in Notre Dame

1 Introduction to NMR Spectroscopy

NMR Spectroscopy of Aromatic Compounds (#1e)

MestRe-C User Guide Megan Bragg 04/14/05

INTRODUCTION TO CHEMICAL EXCHANGE WITH THE MEXICO PROGRAM

TOPSPIN INSTRUCTIONS

NMR and IR spectra & vibrational analysis

NMR Signal Properties & Data Processing

1 st day Basic Training Course

Examination of Proton NMR Spectra

Introduction to Nuclear Magnetic Resonance Spectroscopy

Generation and Detection of NMR Signals

Carbon, Deuterium and Heteronuclear NMR using Topspin

A Tutorial for Chemists: Using Mnova to Process, Analyze and Report 1D and 2D NMR on Your Desktop

Atomic and Nuclear Physics

Chemistry 307 Chapter 10 Nuclear Magnetic Resonance

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

How to Quickly Solve Spectrometry Problems

Nuclear Magnetic Resonance and Its Application in Condensed Matter Physics

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

Nuclear Magnetic Resonance and the Measurement of Relaxation Times of Acetone with Gadolinium

Nuclear Magnetic Resonance

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

Introduction to Nuclear Magnetic Resonance (NMR) And. NMR Metabolomics

Supporting Information

Introduction to Fourier Transform Infrared Spectrometry

HSQC and HMBC for Topspin

NMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum.

Nuclear Shielding and 1. H Chemical Shifts. 1 H NMR Spectroscopy Nuclear Magnetic Resonance

Lab 1: The Digital Oscilloscope

Running Deuterium (H2) NMR at SCS NMR facility

Nuclear Magnetic Resonance Spectroscopy

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

The Unified Scale for Referencing in NMR: New IUPAC Recommendations revised (cgf): 26 July 2010

Performance and advantages of qnmr measurements

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)

Relaxation Can T, Be Longer Than T,?

FTIR Instrumentation

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Direct and Reflected: Understanding the Truth with Y-S 3

Nuclear Magnetic Resonance Spectroscopy

Appendix A: Acquiring EPR Spectra and Optimizing Parameters. Measure the Spectrum with Nominal Settings

NMR Nuclear Magnetic Resonance

CHE334 Identification of an Unknown Compound By NMR/IR/MS

Alignment and Preprocessing for Data Analysis

Pesticide Analysis by Mass Spectrometry

NMR practice times. Mo pm Jim 2-4:30 Ning 4:30-7. Tues pm Jianing 2-4:30 Ting 4:30-7. Fri Donia 10-12:00 Ilya 2-4

Infrared Spectroscopy 紅 外 線 光 譜 儀

Single-scan longitudinal relaxation measurements in high-resolution NMR spectroscopy

Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note

right click Open Terminal mnova mnova & Using a terminal window cd /home/your-user-name pwd ls Cp cd /MNova cd /data pwd ls Cp

Ph 3504 Nuclear Magnetic Resonance and Electron Spin Resonance

Introduction to ACD version 12 NMR Software

Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory

Transcription:

Practical guide for quantitative 1D NMR integration Eugenio Alvarado, University of Michigan, 05/10/10 The purpose of this manuscript is not to present a discussion about quantitative NMR, but to offer practical advise on how to set up appropriate parameters for acquisition and processing of 1D spectra to accurately measure integrals. Detailed information can be found in the references and in standard NMR textbooks. These notes refer mainly to proton spectra; quantitative analysis of spectra of other nuclei may require different settings, but these notes can be used as a guide. It is important to keep in mind that the protocol used depends on the level of accuracy that you want to achieve. The following steps are aimed at getting about a 1% accuracy. If more is needed, you will need a more rigorous control of the parameters. Acquisition 1. Shim the sample carefully and turn spinning off. Accurate shimming is important for integration and may be crucial if signals overlap and deconvolution is needed. If spinning side bands are visible, it is advisable to turn spinning off. 2. If measuring 1 H, use 13 C decoupling during acquisition. Carbon decoupling helps clean up the region around the base of the peaks. The 13 C satellites around each hydrogen peak account for 1.07% of the integration and should be taken into account in each integral region if an accuracy of 1% or less is required. The figure below shows the aromatic region of chloroquine; the lower trace is plotted at an usual intensity and the 13 C satellites cannot be seen; in the middle trace the intensity is 100 times higher and the satellites (and impurities) can now be clearly seen; the upper spectrum was acquired with 13 C decoupling (the small signals that remain are impurities). Notice that the satellites of a signal, when present, may overlap the integral regions of other signals and interfere with the measurements. Also, by using 13 C decoupling, the

integral regions do not have to be so wide to include the satellites and they will be easier to set up in crowded regions. It will also be easier to exclude impurities or determine their presence in the integrals. To enable 13 C decoupling, go to the Acquire, Channels parameter panel and in the Dec On/Off field of the Decouple channel type nny, and in Dec Modulation type g. This turns on GARP decoupling during acquisition. On the other hand, if you are measuring any other nucleus (not 1 H or 19 F) and need proton decoupling, NOE effects from neighboring hydrogen atoms should be avoided as these will affect your integrals. If you don't need proton decoupling, set Dec On/Off in the Decouple channel (H1) to nnn. If you need proton decoupling, use inverse gated decoupling by setting Dec On/Off to nny and Dec Modulation to w (waltz decoupling). If you are measuring fluorine, proton decoupling is off by default (there is no way of doing it except in our new vnmrs-500 spectrometer with two high frequency channels). 3. Choose the spectral window so that all peaks of interest fall within 60% of its center. Take a spectrum and choose a spectral window with about 20% of empty space on both ends of the spectrum. This is recommended in order to compensate for the signal attenuation that occurs at the spectral edges due to the receiver filters. 4. Choose an appropriate digital resolution. Each peak should be described by at least 5 points above its half width in order to have a reliable integration. The digital resolution (the spacing between data points) is defined by the spectral width (sw) divided by half the number of acquired points (np). Having previously defined an appropriate spectral width, the digital resolution is then determined by the number of points which in turn is assigned by the acquisition time (at). Assuming a half width of 1 Hz (typical for a medium size organic molecule), that means a digital resolution of 0.2 Hz. At 500 MHz with a 10 ppm spectra width, an acquisition time of about 5 seconds gives a digital resolution of 0.2 Hz. In general, for proton spectra, use an acquisition time of 4 to 5 seconds. When in doubt, take a survey spectrum, put the cursor in one of the peaks of interest and type dres to get the digital resolution and linewidth of the peak; if necessary, adjust the acquisition time and repeat the procedure. Keep in mind that highly symmetric compounds and some nuclei (fluorine comes to mind) may have very narrow peaks and long relaxation times. 5. Use a 90º pulse with a long relaxation delay (d1). All signals of interest must have relaxed

completely before each pulse. If this condition is not met, the next pulse will produce signals that will not show their full intensities and their relative areas will be incorrect. After a 90º pulse, the magnetization relaxes with a rate that is a function of the longitudinal relaxation time, T 1, of the nucleus. Thus, after 3T 1 the magnetization has recovered to 95.0% of its original size; after 5T 1 the recovery is 99.3% and after 7T 1 it is 99.91%. Therefore, to obtain an accuracy of 1% or less, the recycle time (the time between pulses) must be at least five times the T 1 of the slowest relaxing signal of interest in the spectrum. The major problem here is that usually the relaxation times of the nuclei in the sample are not known (although they can be quickly estimated with an inversion recovery experiment) and therefore some guessing may be required. Relaxation times for hydrogen nuclei in medium size molecules are usually from 0.5 to 3 or 4 seconds; while those for 13 C can go from 0.1 to tens of seconds for quaternary carbons. For example, for atropine (MW=289, C 17H 23NO 3) in non degassed chloroform, proton T 1s go from 0.6 to 3.0 seconds; and 13 C T 1s go from 0.6 to 11.7 s. So, assuming the longest 1 H T 1 in the sample is 3 seconds, the recycle time should be 15 s. As the recycle time in a 1D spectrum is the sum of the relaxation time d1 and the acquisition time at, if the acquisition time is 5 s, the relaxation delay d1 should be set to 10 s. If you added an integration standard to you solution, keep in mind that the signals from the small molecules usually chosen as standards may relax much more slowly than typical organic molecules and can be ten seconds or more. If you use an standard and it doesn't relax completely, your calibration will be wrong. Also, carbon-13 and other nuclei can relax extremely slowly (tens of seconds). Use of relaxing agents in these cases may be very useful to speed up the measurements. 6. Verify that the spectral width is 30% or less of the inverse of the pulse width: sw 0.3/pw. If it isn't, decrease the pulse width. This is not an issue with proton spectra because the spectral widths and pulse widths are short, but it may be important for other nuclei where spectral widths and pulses are large. A radio-frequency pulse doesn't excite all regions equally but has an excitation profile that has a shape similar to the one shown below for a pw=100 μs. For quantitative integration, it is important that all signals be located within the middle 15% of the profile (shaded area), where their intensity variation will be less than 1% of the maximum intensity. The parameter pw is in units of microseconds; convert it to seconds to calculate the maximum sw. If you must know, the excitation profile function is: ƒ(υ) = sin(pw π υ)/(pw π υ).

For example, fluorine spectra have default pulse widths of about 5 μs and spectral widths of 230 ppm, or more than 90 khz. With those values, peaks near the left and right edges of the spectrum will not show their maximum intensity. The solution would be to change the pulse width to 3 μs or to make sure that all the peaks of interest are located near the center of the spectrum in an area not larger than 60 khz (0.3/0.000005). 7. Collect enough scans to get a good signal to noise ratio. The signal to noise should be higher than 1000:1 for integration errors of less than 1%. This is easy to do with proton spectra on spectrometers with indirect probes, but may not be so easy for other nuclei. Use solvent suppression if the sample contains non deuterated solvents. To calculate the signal to noise ratio of a peak, expand a section of the spectrum containing both a region with only noise and the peak of interest; then use both cursors to select a flat region containing with only noise, and type the command dsnmax. That command calculates the s/n of the tallest peak displayed relative to the average noise in the selected region. Processing 1. Use zero-filling and exponential weighting. Zero filling means increasing the number of data points before Fourier transformation. This has been found to incorporate information from the dispersive part of the transformed spectrum into the absorptive part resulting in increased integral precision. To do a one time zero filling, type on the command line fn=np*2 or to do a double zero filling type fn=np*4, then do the Fourier transformation as usual. If the spectrum is noisy, you can also increase the amount of line broadening that is usually applied to most spectra to increase the signal to noise. Normally, an exponential weighting function is applied to the FID before transformation to increase signal to noise. The parameter lb controls the amount of s/n enhancement or line broadening applied; the higher the number, the higher the noise will be suppressed, but the more the peaks will be broadened. Usually, values from 0.2 to 1.0 are adequate for 1 H and 19 F. Other nuclei may benefit from much higher values (1-5 for 13 C, 10-100 for others). In theory, the best value is that which approximates the linewidth of the peaks in Hz. Enter a value for lb and Fourier transform with wft. 2. Accurately correct the phase of the spectrum and the bias and slope of the integral. You will obtain better results if you do it manually instead of relying on the macros that try to do it automatically for you. 3. Use baseline correction to ensure a very flat baseline. It is critical to have a very flat baseline in order to obtain accurate integrals. In order to do it in vnmrj, you must first indicate what points in the spectrum belong only to the baseline and should be flattened, and what points belong to peaks. This is done with integration regions. The software will know that the points outside those regions belong to the baseline and will try to synthesize a polynomial function that reproduces the shape of all those points. The function will then be subtracted form the spectrum to give a flat baseline. Display the integral trace and cut it in pieces; not too narrow, allow at least 0.1or 0.3 ppm on either side of the peaks. Don't worry if the regions are not selective; these are only to define the baseline correction and we will define accurate regions later. All peaks must be inside a region, even the solvent and those peaks you don't want to measure. Then, apply a baseline correction with the command bc(5). If for any reason, you

later need to re-transform the spectrum, this step must be repeated. Refer to the Commands and Parameters Reference Manual (Help menu in vnmrj) for more information about the bc command. The MestreNova software has a very impressive baseline correction routine that you can try. 4. Define appropriate integration regions. Now that the spectrum is flat, we can reset the previous integration regions and define new ones with better precision. At this point, it is very important to realize that the peaks do not start to grow from the baseline at a defined distance from the tip like a pyramid, but they grow slowly from the infinite. The shape of an NMR peak is given by the Lorentzian function: L(x)=h/(1+(x/w) 2 ), where h is the peak's height and w is its line width at half height. Obviously, we can't integrate every peak from -infinite to +infinite, so we must set a practical limit. Griffiths and Irving have shown that in order to cover 99.0% of the area of any peak, the region must spread over a range that is at least 25 times the line width of the peak in both directions. For example, with a line width of 1Hz, the region should be 50 Hz wide (0.125 ppm at 400 MHz or 0.1 ppm at 500 MHz). To get an accuracy of 0.1% the regions must cover about 75 times the linewidth in both directions. Therefore, get a good estimation of your peaks' widths (command dres) and select your integral regions accordingly. 5. Use deconvolution if the peaks of interest overlap. Integral regions should not overlap. For example, two singlets 1 Hz wide should be separated by at least 50 Hz (0.1 ppm at 500 MHz) to be able to be integrated individually with an accuracy of at least 1%. If they are not, a different method called Line Fitting or Deconvolution can be used. Essentially, this method attempts to reproduce the individual peaks in your spectrum thereby knowing their areas. Although vnmrj can do deconvolution, it is not easy to use. MestreNova on the other hand has a very nice implementation that you can try. Advanced Chemistry Development's ACD/NMR Processor Academic Edition can also do deconvolution and it is free for academic users. It can be downloaded from: http://www.acdlabs.com/resources/freeware/nmr_proc/index.php As an example, the spectrum of chloroquine diphosphate in 90% H 2O is shown below. The spectrum

was acquired at 400 MHz using WET water suppresion, with a relaxation delay of 20 s, acquisition time of 3.7 s, and pw=6.6 μs. The second signal from left to right, at 7.9 ppm, shows a lower than expected integration due to saturation exchange with the solvent (it is an NH). The signal at 3.9 ppm also shows a reduced integration due to partial saturation because of its proximity to the water signal being saturated. The expected normalized integrations are shown on red for each multiplet. The compound was used from the manufacturer's bottle without further purification (small amounts of impurities are clearly seen when the spectrum's intensity is increased). Except for the signals at 7.9 and 3.9 ppm, all integrals are within ±1% of the expected values even though the 3 regions between 7.8 and 8.1 ppm partially overlap each other. References: L. Griffiths and A. Irving, Assay by nuclear magnetic resonance spectroscopy: quantification limits. Analyst 123, 1061-1068, (1998). T. Saito, S Nakaie, M. Kinoshita, T. Ihara, S. Kinugasa, A. Nomura and T. Maeda, Practical guide to accurate quantitative solution state NMR analysis. Metrologia, 41, 213-218 (2004). P. Soininen. Quantitative 1H NMR Spectroscopy. Doctoral dissertation, Department of Biosciences, University of Kuopio, October 10, 2008. Online at: http://epublications.uef.fi/pub/urn_isbn_978-951-27-1093-5/urn_isbn_978-951-27-1093-5.pdf C. Cobas, NMR Analysis Processing and Prediction blog in http://nmr-analysis.blogspot.com/. Nov. 30, 2009; Jan. 11, 2010; Jan. 14, 2010 and others. These are very interesting discussions about this topic from the main author of MestreNova. G. F. Pauli, B. U. Jaki and D. C. Lankin, A routine experimental protocol for qhnmr illustrated with taxol, J. Nat. Prod. 70, 589-595 (2007).