Operating Instructions & Maintenance Guidelines for Precision Vaporizers



Similar documents
Scope: Applies to the various anesthesia vaporizers used to deliver a known concentration of vaporized liquid. Sample. Type Interval Time Required

Understanding Anesthetic Delivery Systems Dean Knoll, CVT, VTS (Anes.) Anesthesia Technician Supervisor University of Wisconsin Madison, WI May 2003

NITROUS TRANSFER PUMP INSTRUCTIONS

THERMOSTATIC MIXING VALVE TRIM K-T9493, K-T9494

Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)

Veterinary Clinics and Hospitals

Rexroth Hydraulic Pump A10VO Series User Manual

Ozone Generator Manual

OHIO MEDICAL CORPORATION. Medical Gas Panel. Installation and Maintenance Manual

Oxy-Fuel Gas Welding. Given a functional oxy-fuel gas unit, instruction and demonstration of use, each student will be able to:

Tech. Services: (800) Fax: (800) Order Entry: (800) Fax: (800) Internet Address:

1.8 CRANKSHAFT OIL SEALS

FASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL

EXPRESS STILL WARNING It is legal in New Zealand to own and operate a still for the purpose of producing alcohol for your own consumption.

MODEL G300 BRAKE BLEEDER

Dissolved Oxygen Sensors

THE NEWCASTLE UPON TYNE HOSPITALS NHS FOUNDATION TRUST SAFE USE AND STORAGE OF LIQUID NITROGEN AND SOLID CARBON DIOXIDE (DRY ICE)

Assembly and Installation Procedures

WHAT YOU DON T KNOW ABOUT ACCUMULATORS CAN KILL YOU!

FJ2. 2 Ton Trolley Floor Jack Assembly & Operating Instructions

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE

Chemistry 112 Laboratory Experiment 6: The Reaction of Aluminum and Zinc with Hydrochloric Acid

SALTER AIRE Plus COMPRESSOR Model 8350 / 8352 / Instruction Manual. Page 1 of 9

In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..

el-720 Carbon Dioxide Gas Leak Detector User s Guide

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series

Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR İ-COLD Rev. 00

UV100A Ultraviolet Air Treatment System

BACKPACK SPRAYERS. MODEL NOS: KSP16 & KSP20 Part Nos: & OPERATING & MAINTENANCE INSTRUCTIONS GC04/12

National- Spencer Inc.

Product Manual. CVS Type 67AFR Filter Regulator. Introduction. Installation. Description

Operator Quick Guide EC SENSOR

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

SunMaxx Solar Filling Station Operating Instructions

Model 43AP Pneumatic Controller, Style B

Round Housing with Side Ports

Compressed Gas Cylinder Program

WIKA INSTRUMENT CORPORATION

Fundamentals of Mass Flow Control

Hydraulic control unit series

Peasouper Operator s Manual

SAFETY & OPERATING INSTRUCTIONS

SureSite Magnetic Liquid Level Indicator

DISSOLVED OXYGEN PROBE

Instructions for Use. Components of the GENOTROPIN PEN 12

FLUSHOMETER - TANK SYSTEM

Volkswagen Golf > VW Rabbit GTI 2006->

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA. REF Rev. L

15GAL STEEL OIL DRAIN WITH 110V PUMP

STAINLESS STEEL TANK TABLE OF CONTENTS TROUBLESHOOTING... 1 TESTING WATER TEMPERATURE... 3 TANK & JACKET ASSEMBLY... 4.

Approval of this procedure is provided by signatory indicated in the approval block at the footer of Page #1, for and on behalf of Jabiru Pty. Ltd.

Druck PV 62x pneumatic/hydraulic pressure stations

This instruction is valid for all ACD pump models shown on page 2

Installation and User Instructions Unvented Electric Storage Water Heater Models: S10UNV, S15UNV.

INSTALLATION AND OPERATION INSTRUCTIONS NOT FOR OXYGEN OR OTHER OXIDIZING GASES

BOWIE PUMPS OPERATION - MAINTENANCE

Professional Precharged Hunting Rifle

CALIBRATION FOR LAL20X & LAL24X

Dive Rite 200 & 300 Bar Isolator Manifold Service Manual

USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102

INSTRUCTIONS AND PARTS LIST FOR MODEL 70H & 75H HAND-OPERATED HYDRAULIC PRESS

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)

Instructions For SF 6 Refill Kit

BELLMAN CXE-25/27 INSTRUCTIONS

Class 5 to 7 Truck and Bus Hydraulic Brake System

Voltmaster Trash Pumps Model TSP2, TSP3 and TSP4

Submersion/Insertion Toroidal Sensor

Volkswagen New Beetle 2.0 Liter 4-cyl General, Engine (Engine Code AEG) 17 Engine-Lubrication system (Page GR-17)

Installation Instructions S

Failure to comply with the following cautions and warnings could cause equipment damage and personal injury.

Char-Lynn Spool Valve Hydraulic Motors. Repair Information. W Series Geroler Motors

TIPPMANN. A-5 Owner s Manual CO2 POWERED PAINTBALL MARKER

Pet hair clipper. Model Diagrams within this manual may not be drawn proportionally.

SELECTION, APPLICATION & MAINTENANCE

Use & Care. of your Iron. All about the TA B L E O F C O N T E N T S. USA and Canada

SHOT BLAST CABINET MODEL NO: CSB20B

Stainless Steel Single and Dual Circulation Kits

A/C refrigerant system, overview

IMPORTANT INFORMATION - PLEASE READ. Table of Contents. Introduction. General Information

Quick guide: using CTE security straws for cryopreservation of pronuclear stages, embryos and sperm

Tri-Homo Style Operation and Maintenance Instructions

Oil and Coolant Circulating Heating System. Model - OCSM

123 Industrial Loop Road Paynesville, MN Phone: MASTER MANUFACTURING MASTER GARDNER

ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

CDS TROUBLESHOOTING SECTION I. VACUUM Weak vacuum at wand. Gauge reads normal (10hg to 14hg)

Porter Instrument Division

Softail Oil Change. Tools

MG1532 & MG2032 MIXER-GRINDERS

Patient Manual. Oxygen Concentrator

Electric Meat Grinder

OPL BASIC. Dosing System for Professional Laundry machines. Contents

Physical Chemistry Laboratory I CHEM 445 Experiment 6 Vapor Pressure of a Pure Liquid (Revised, 01/09/06)

MODEL J SERIES CONVERTER

Extraction: Separation of Acidic Substances

TECHNICAL INSTRUCTIONS REPAIR KITS Hydroguard Series 420 Thermostatic Valve All Models (1 9)

CHAPTER 65 TAIL ROTOR DRIVE SYSTEM. Section Title Page

CARING FOR YOUR WATER HEATER

Transcription:

Operating Instructions & Maintenance Guidelines for Precision Vaporizers

This page intentionally blank.

Examination and Preparation for Use We want you to be completely satisfied with your vaporizer. Please look now to see if it is as you expected. Should anything not be exactly right, please call your sales representative right away for help. 1. Examine shipping carton for signs of external damage. 2. Inspect vaporizer for visible damage or missing parts. 3. Check that control dial operates freely. This vaporizer is designed to give you many years of trouble-free use. It is both flow rate and temperature compensated to give precise delivery of anesthetic gas. This is done by simply setting the control dial to the percentage you want to deliver. (See Performance section of manual for clinical ranges.) This vaporizer is to be operated under the normal surveillance and control of a veterinarian trained in its use. However, you need to know more about this vaporizer than just how to operate it. Please read this manual in its entirety. Abbreviated operating instructions can be found on page 14. If you have any questions or comments, we would welcome the opportunity to address them. Please call your local sales representative. Thank you! Model #: Serial #: Purchased: / / For Sales or Service support, call: 800-466-6463 or 925-463-1828 1

Contents DO NOT ATTEMPT INITIAL OPERATION OF THIS CALIBRATED VAPORIZER UNTIL THIS MANUAL HAS BEEN READ IN ITS ENTIRETY. Warnings... 3 Cautions... 4 Description... 5 Specification... 7 Performance... 8 Principle of Operation... 9 Installation... 10 Operating Instructions Turning On... 11 Filling and Draining General... 11 Funnel Filler, Cyprane and Ohio style... 12 Keyed Filler, Ohio style vaporizer... 13 Keyed Filler, Cyprane style vaporizer... 14 Maintenance... 16 2

Warnings DO NOT FILL THE VAPORIZER WITH ANY ANESTHETIC AGENT OTHER THAN THE ONE SPECIFIED ON THE FRONT LABEL. THE VAPORIZER IS DESIGNED FOR THAT AGENT ONLY. ANY OTHER AGENT THAN THAT SPECIFIED CAN PROVE TO BE DANGEROUS TO A PATIENT. IF A VAPORIZER IS FILLED WITH THE WRONG AGENT, DRAINING WILL NOT ELIMINATE THE AGENT SINCE SOME OF THE AGENT WILL HAVE BEEN ABSORBED INTO THE WICK. THE WICK MUST BE THOROUGHLY CLEANED AND DRIED. THIS PROCEDURE MUST BE PERFORMED AT AN AUTHORIZED SERVICE CENTER. DO NOT CARRY THE VAPORIZER BY THE CONTROL DIAL. DO NOT MODIFY, TAMPER WITH, OR DISASSEMBLE THE VAPORIZER. THERE IS A POTENTIAL DANGER OF DAMAGING THE VAPORIZER AND ALTERING THE CALIBRATION ACCURACY. DO NOT IMMERSE THE VAPORIZER IN ANY LIQUID, INCLUDING WATER. DO NOT STERILIZE THE VAPORIZER. DO NOT DRAIN THE ANESTHETIC AGENT INTO ANY CONTAINER OTHER THAN A PROPERLY MARKED CONTAINER. INCOMPLETE SEALING OF THE FILL CAP WILL RESULT IN LOSS OF GASES. DO NOT FILL THE VAPORIZER UNLESS THE CONTROL DIAL IS IN THE 'OFF' POSITION. DO NOT TURN THE DIAL 'ON' DURING FILLING OR ATTEMPT TO FILL BEYOND THE 'FULL' MARK. KEEP THE VAPORIZER UPRIGHT AT ALL TIMES. NEVER TILT A CALIBRATED VAPORIZER BEYOND 45 DEGREES WHEN CHARGED. DANGEROUS OVERDOSE MAY RESULT DURING SUBSEQUENT USE. THIS CONDITION CAN BE DETECTED BY CALIBRATION VERIFICATION ONLY. DO NOT CONNECT THE VAPORIZER DIRECTLY INTO THE PATIENT BREATHING CIRCUIT. DO NOT TURN ON TWO VAPORIZERS AT THE SAME TIME. DO NOT PUT WATER OR ANY OTHER SOLVENT IN A VAPORIZER DUE TO THE WICK S CAPACITY TO ABSORB AGENT, THE VAPORIZER WILL CONTINUE TO DELIVER THE SET CONCENTRATION FOR A CONSIDERABLE PERIOD OF TIME EVEN IF NO LIQUID LEVEL CAN BE VIEWED THROUGH THE SIGHT WINDOW. HOWEVER, DO NOT USE A VAPORIZER IN THIS CONDITION ON A PATIENT. FLUSH FLOWS THROUGH AN OPEN VAPORIZER MAY RESULT IN UNWANTED CONCENTRATIONS OF ANESTHETIC AGENT. DO NOT IGNORE MAINTENANCE OF YOUR VAPORIZER. AN ANNUAL CALIBRATION RECERTIFICATION IS HIGHLY RECOMMENDED BY THE MANUFACTURER. 3

Cautions Vaporizer should be serviced by an authorized service center. Turn the vaporizer OFF when not in use. A vaporizer MUST be in a vertical upright position to indicate liquid agent levels accurately. If a screw-cap filler vaporizer is pressurized, turn the screw cap slowly before filling. If a keyed-filler vaporizer is pressurized, turn the screw slowly before filling. The delivered concentration may be different from the set concentration if the vaporizer is operated in a tilted position. Avoid kinks in the rubber tubing that will restrict flow. Eliminate leaks that will effect total flow to the patient. To fill the vaporizer when in use, it will be necessary to turn the vaporizer OFF. No repair should ever be undertaken or attempted by anyone not having the training and qualifications to perform these tasks. 4

Description General This vaporizer is designed for 'out of circuit' use in continuous flow techniques of inhalation anesthesia. This vaporizer is temperature and flow compensated so that the output remains relatively constant despite cooling due to vaporization and variations in inlet flow (see - Performance section). This vaporizer is clearly labeled with the name of the anesthetic agent for which it is designed. A single control dial with a concentration scale calibrated in % of anesthetic agent vapor per total volume (v/v) is employed to set the desired concentration of the anesthetic agent. To prevent accidental displacement of the control dial from the 'OFF' position to 'ON', a release button is incorporated in the dial assembly. Simultaneous depression of the release button and counter-clockwise rotation of the dial is necessary to achieve an 'ON' setting. An inlet filter protects the vaporizer against particulate contamination. Calibrated vaporizers are adaptable for use on many types of anesthesia machines and are available with either a funnel or key-fill system. 5

Release b tt Inlet Filler cap Sight glass Funnel filler Cyprane style vaporizer Funnel filler system Control di l Release button Upper retaining screw Outlet Filling tunnel Upper valve knob Lower valve knob Draining tunnel Sight glass Lower retaining screw Ohio style vaporizer Keyed filler system 6

Specification Calibration Vaporizers are calibrated at 21 C (70 F) and at elevated temperatures. The variation in output with temperature, flow rate and duration of use is small, and the variation in output when used with Intermittent Positive Pressure Respiration is negligible. Agent Usage The rate of consumption of anesthetic agent depends primarily on flow rate and vapor output concentration. The rate of evaporation of anesthetic agent may, with caution, be used as an approximate method of checking that the delivered output is not grossly in error, and as a means of estimating how often the vaporizer is likely to need refilling. The approximate hourly consumption of anesthetic agents can be expressed as follows: 3 x % x F Where % represents the setting of the vaporizer output percentage, F represents the input flow rate in liter/min. Example: If a vaporizer is set to deliver 2% at 6-liter/min. total input gas flow rate: Approximate rate of agent consumption = 3 x 2 x 6 = 36 ml/hour. The figures are intended only for clinical guidance and are approximate. They will vary depending upon the type of anesthetic agent employed, accuracy of graduation of flowmeters, etc., and will be grossly in error if the vaporizer drain port is not fully closed. 7

Performance Effects of Variables Temperature The effects of variation of temperature are normally negligible at commonly used combinations of dial setting and ambient temperature. The vaporizer responds very slowly to changes in ambient temperature and (to prevent the valve from closing completely) as a safety feature the temperature sensitive valve does not respond to temperatures below the range of approximately 12-15 C (54-59 F). Should the vaporizer temperature be lower than this, the output can be expected to be lower than that indicated on the dial. At temperatures above 32 C (90 F), the vaporizer output may be unpredictably high - particularly if the temperature approaches the boiling point of the specific anesthetic agent. To avoid inaccuracies due to extreme temperatures the vaporizer should be allowed to attain a temperature of 20-32 C (68-90 F). Flow The vaporizer compensates for variations in inlet flow from 300 ml to 10 liters per minute with minimal change in output concentration. Resistance to flow within the vaporizer is minimal: 40 cm/h 2 O (30 mm/hg) at 10 liters per minute. Carrier Gas Composition Small effects can occur when the carrier gas composition is changed from oxygen to air or nitrous oxide/oxygen mixture. As a general rule variation of output with carrier gas composition can be considered of negligible clinical significance since the effects, if any, are normally less than 10% of setting. Where changes occur, the usual effect is that the output is slightly depressed when nitrous oxide is employed compared to the output when oxygen is the carrier gas. The presence of nitrous oxide reduces the required inspired concentration of volatile agent and this mitigates this small depression in output from the vaporizer. Time Out of Service If the anesthetic machine on which the vaporizer is fitted is left for a period of time with no gases flowing, a concentration of agent may be observable at the machine outlet when the gas flow is turned 'on' and the vaporizer is turned 'off'. This concentration can be expected to fall rapidly to zero (e.g. within about 15 seconds at 5 liter/min.). This phenomenon is a normal characteristic of anesthetic vaporizers and machines. Clinically this is normally considered to be insignificant because of the small volume of vapor involved. Other Variables Ambient temperature, input flow rate and duration often can affect delivered concentrations, particularly when the vaporizer is used at extremes of the usual clinical range. The valve design and temperature compensation system of this vaporizer reduces these effects, under most clinical conditions, to levels that are clinically not significant. 8

Principle of Operation Vaporizer Valve Assembly In the 'OFF' position, there is a direct link between the 'inlet' and 'outlet' within the vaporizer. When the dial is turned 'ON' the carrier gas is split into two streams; bypass and vaporizing chamber flow. Carrier gas entering the vaporizing chamber flows through agent saturated wicks. As the carrier gas passes through the wicks it becomes saturated with vapor. The amount of agent picked up in the vaporizing stream will vary due either to variation in room temperature or to the cooling which takes place when the agent is vaporized. Each causes changes in the effective vapor pressures of the anesthetic agents. Unless some form of compensating device was used, the output of the vaporizer for a given flow and dial setting would change with temperature. NOTE: This principle is one of many reasons explaining why it is recklessly dangerous to use a jar and wick style vaporizer with extremely volatile anesthetic agents such as Halothane and Isoflurane. This vaporizer has a temperature-compensating device (thermostat) to control the proportion of carrier gas entering the vaporizing chamber. If the temperature of the vaporizer falls, the thermostat closes and more carrier gas is allowed into the vaporizing chamber. If the temperature of the vaporizer increases, the thermostat opens and less carrier gas is allowed into the vaporizing chamber. In this way the output of the vaporizer remains constant under conditions of changing temperature. 9

Installation The vaporizer may be mounted onto a back bar assembly, interlock, a tabletop stand or a clamp fitted to the frame of the anesthesia apparatus. The vaporizer should always be mounted between the flow-metering unit and the patient breathing circuit-- but upstream of any flush valve, absorber or humidifier. Ensure that liquid which may accumulate in the breathing circuit or the CO 2 absorber cannot enter the vaporizer while in use or during disassembly of the circuit or when the machine is not in use. Unless otherwise specified, all vaporizers are supplied standard with 23mm inlet and outlet tapers. The standard mounting system requires bolting of the vaporizer directly to a rigid back bar of an anesthesia gas machine. Viewed from back of vaporizer IMPORTANT The direction of gas flow must be from inlet to outlet in the direction of the arrow, i.e. from left to right when viewing the vaporizer from the front. 10

Operating Instructions OBSERVE ALL WARNINGS AND INSTRUCTIONS IN THIS MANUAL. Turning On 1. To turn the vaporizer ON, depress the control dial release button, and simultaneously turn control dial counter-clockwise to desired concentration. NOTE: To avoid the inadvertent delivery of small concentrations the control dial should be turned to OFF when the vaporizer is not in use. Filling and Draining WARNING - DO NOT FILL VAPORIZER WITH ANY AGENT OTHER THAN THE ONE SPECIFIED ON THE FRONT LABEL. THE VAPORIZER IS DESIGNED FOR THAT AGENT ONLY. ANY OTHER AGENT THAN THAT SPECIFIED CAN PROVE TO BE DANGEROUS TO A PATIENT. WARNING - DO NOT FILL VAPORIZER UNLESS THE CONTROL DIAL IS IN THE OFF POSITION. WARNING - DO NOT TURN THE DIAL ON DURING FILLING OR ATTEMPT TO FILL BEYOND THE FULL MARK. WARNING - DO NOT DRAIN THE AGENT INTO ANY CONTAINER OTHER THAN A PROPERLY MARKED CONTAINER. Periodically check the agent level. The vaporizer should be refilled at appropriate intervals. The vaporizer will function satisfactorily as long as there is agent visible in the sight glass. The vaporizer should be filled and used in an upright position. Small deviations from the upright position will not affect the output or the safety of the vaporizer. Because the agent depth is shallow in relation to the diameter of the vaporizing chamber, more frequent checks of the agent levels should be carried out when small deviations from the upright position occur. This will avoid obtaining a misleading impression of the amount of agent in the vaporizer. At intervals - ideally not exceeding two weeks - the vaporizer should be drained into an appropriately marked container when the agent level is low and the agent discarded. Less frequent intervals may be used when the anesthetic agent (e.g. Isoflurane) does not contain additives or stabilizing agents. 11

Funnel Filler To fill: Dial should be in OFF position. If the vaporizer wicks are dry, the level will fall as the wicks absorb the liquid agent. The agent is not lost. 1) Remove filler cap by turning cap counter-clockwise. Turn cap slowly if vaporizer is pressurized. Be sure drain plug is closed. 2) Verify that the agent is the same as that specified on the vaporizer. 3) Pour agent slowly into opening. Observe proper agent level through sight glass. If the vaporizer wick is dry, the level will fall slightly as the wick absorbs the agent. 4) Replace cap by turning cap clockwise. Cap should be tight to prevent leaks. Sight glass To Drain: 1) Remove the filler cap to reveal the drain plug. 2) Unscrew the plug, but do not remove. (Some vaporizers have a slot in the filler cap to unscrew plug.) 3) Drain only into a properly marked container. Drain 12

Keyed Filler System The filling system consists of three elements - the bottle collar, the adapter and the filling / draining unit fitted to the vaporizer. Key fill and drain systems are designed to prevent the wrong agent from being used in the vaporizer. For example, an Isoflurane vaporizer cannot be filled using a Halothane bottle adapter. And, the Halothane bottle adapter will not fit on either the Isoflurane bottle or into the Isoflurane vaporizer. Ohio Style Vaporizer Only To Fill: If the vaporizer wicks are dry, the level will fall as the wicks absorb the liquid agent. The agent is not lost. 1) Attach the bottle adapter to the appropriate bottle. 2) Ensure the vaporizer control is set in the OFF position. Insert the keyed block of the adapter fully into the filling (upper) tunnel, with the two adapter holes DOWN. Tighten the retaining screw located above the filling tunnel. 3) Open the upper valve knob to allow the liquid to flow gradually into the vaporizer. The bottle can be elevated to increase the fill rate. 4) Fill the vaporizer with the liquid agent until the FULL mark is reached as viewed through the sight window. 5) When the vaporizer is full, lower the bottle, close the valve knob, and remove the block from the tunnel. WARNING: - INCOMPLETE SEALING OF THE VAPORIZER VALVES ON THE KEYED FILL AND DRAIN MODELS WILL RESULT IN LOSS OF GASES. TO HELP PREVENT SUCH LOSS, MAKE SURE THAT THE FILL AND DRAIN VALVES ARE COMPLETELY CLOSED WHEN THE VAPORIZER IS IN USE. WARNING: - FAILURE TO OBTAIN A GAS-TIGHT SEAL AT BOTH ENDS OF THE BOTTLE ADAPTER MAY PERMIT LEAKAGE OF AIR INTO THE BOTTLE AND CONSEQUENT OVERFILLING OF THE VAPORIZER. To Drain: 1) Insert the keyed block of the adapter (use a properly marked empty bottle) fully into the draining (lower) tunnel, with the two adapter holes UP. Tighten the retaining screw located below the draining tunnel. 2) Open the lower valve knob to allow the liquid to flow into the bottle. The bottle must be held lower than the draining tunnel. 3) Close the valve knob when the vaporizer is emptied and remove the block from the tunnel. WARNING: - TO AVOID DEFEAT OF THE KEYED FILLING SYSTEM, DRAIN THE AGENT INTO THE PROPERLY MARKED AND KEYED BOTTLE ONLY. To avoid wasting liquid agent and to minimize the concentration of agent in room air, always cap the bottle after filling or draining the vaporizer. The bottle adapter attached to a bottle does not prevent spilling or evaporation. 13

Cyprane Style Vaporizer Only To Fill: If the vaporizer wicks are dry, the level will fall as the wicks absorb the liquid agent. The agent is not lost. 1) Attach the bottle adapter to the appropriate bottle. 2) Ensure the vaporizer control is set in the OFF position. Turn top retaining screw on vaporizer filler unit counterclockwise and withdraw filler plug. Turn cap slowly if vaporizer is pressurized. 3) Insert the keyed block of the adapter fully into the filling tunnel, with the two holes DOWN. Take care to bend tube slightly so bottle is below inlet level to prevent spillage. Turn the retaining screw above filling tunnel clockwise and tighten to seal fill adapter in the filling tunnel. 4) Raise bottle above level of filling tunnel, without kinking tube. A steady stream of air bubbles should emerge from the inner tube within 2 seconds. 5) When vaporizer is filled to the FULL level, lower the bottle. Remove fill adapter. If any excess liquid drains from the filler tunnel, allow this to escape completely before inserting and tightening filler plug. Failure to replace and tighten retaining screw will cause gas to escape through filler. The vaporizer is now ready for use. 14

To Drain: 1) Insert the fill adapter in drain (lower) tunnel, with the two holes UP, and tighten retaining screw. 2) Remove filler plug to let air vent from bottle and for draining to occur. Hold bottle slightly downward without kinking tube. CAUTION: Turn cap slowly if vaporizer is pressurized. 3) Open drain valve on front by turning counter-clockwise and vaporizer will drain. 4) At completion of draining, close drain valve (clockwise). 5) Remove the bottle adapter and replace the filler plug in the filler tunnel. Tighten retaining screw (clockwise). Failure to replace and tighten retaining screw will cause gas to escape through filler. 15

Maintenance WARNING - DO NOT MODIFY, TAMPER WITH OR DISASSEMBLE THE VAPORIZER BECAUSE OF THE DANGERS OF DAMAGING THE UNIT AND ALTERING THE ACCURACY OF GRADUATION. Observation of the instructions given earlier, regular servicing and normal professional vigilance is normally all that is required to maintain the vaporizer in a safe working condition. Schedule Every Two Weeks: The vaporizer should be drained into an appropriately marked container when the agent level is low and the agent discarded. Less frequent intervals may be used when the anesthetic agent does not contain additives or stabilizing agents. Annually: The vaporizer should be serviced at an authorized service center. This service should include: 1) Complete disassembly of components. 2) Inspection of all parts for damage and wear. 3) Thorough cleaning of all metal parts. 4) Replacement of wicks, seals and damaged, worn or outdated items. 5) Lubrication where necessary. 6) Re-assembly of vaporizer and testing for and correction of any leaks. 7) Verification of the delivered vapor concentrations under closely defined conditions at different temperatures in order to test the temperature compensating mechanism. Any re-graduation or adjustment where necessary. 8) Maintaining continuous service record. WARNING - DO NOT PUT WATER OR ANY OTHER SOLVENT IN A VAPORIZER. A VAPORIZER SHOULD BE FILLED WITH THE SPECIFIED ANESTHETIC AGENT ONLY. Clean the exterior of the vaporizer with a damp cloth. Never allow cleaning agents to accumulate in the filler, gas inlet and outlet ports, or around the control dial. Contamination If a contaminant is put into the vaporizer (e.g. inappropriate anesthetic agent, alcohol, water, etc.), contact an authorized service center immediately. Repairs Repairs should only be carried out by an authorized service center. Note: It is not possible to change the graduation range or anesthetic agent. 16