How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer



Similar documents
What is the difference between an equivalent time sampling oscilloscope and a real-time oscilloscope?

802.11ac Power Measurement and Timing Analysis

U7248A High Speed Inter-Chip (HSIC) Electrical Test Software Infiniium Oscilloscopes

Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide

Make Better RMS Measurements with Your DMM

Variable Frequency Drive Troubleshooting with U1610A/U1620A

Agilent E363xA Series Programmable DC Power Supplies. Data Sheet

Keysight Technologies Using Fine Resolution to Improve Thermal Images. Application Note

DDR Memory Overview, Development Cycle, and Challenges

Agilent U2000 Series USB Power Sensors

Evaluating Oscilloscope Bandwidths for Your Application

Agilent E5063A ENA Series Network Analyzer

Agilent MATLAB Data Analysis Software Packages for Agilent Oscilloscopes

Agilent U2000 Series USB Power Sensors

Agilent Compatibility of the U2000 Series USB Power Sensors with Agilent Instruments. Application Note

Agilent U2000 Series USB Power Sensors

N5416A and N5417A USB Compliance Test Software for Infiniium Oscilloscopes

The Next Generation in Automated Oscilloscope Test

Agilent Infoline Web Services. Quick Reference Guide. Scan and use your phone to go to Infoline at

Keysight M9485A PXIe Multiport Vector Network Analyzer. Configuration Guide

Agilent Technologies. Troubleshooting Three-Phase AC Motors with U1210 Series Handheld Clamp Meters. Application Note

Time-Correlated Multi-domain RF Analysis with the MSO70000 Series Oscilloscope and SignalVu Software

Agilent BenchVue Software (34840B) Data capture simplified. Click, capture, done. Data Sheet

X-Series Signal Analysis. Future-ready instruments Consistent measurement framework Broadest set of applications and software

Agilent E6832A W-CDMA Calibration Application

Agilent 87421A/87422A Power Supply

Agilent Television Power Consumption Testing. Application Note

PXI and AXIe Modular Instrumentation

Keysight Technologies 8 Hints for Better Spectrum Analysis. Application Note

PCI Express Probes for Agilent E2960B PCI Express Analysis Systems

N6171A MATLAB Data Analysis Software for Agilent Instruments

Agilent Automotive Power Window Regulator Testing. Application Note

An Overview of the Electrical Validation of 10BASE-T, 100BASE-TX, and 1000BASE-T Devices

Real-Time Spectrum Analysis for Troubleshooting n/ac WLAN Devices

Agilent U1230 Series Handheld Digital Multimeters (DMMs)

Agilent FieldFox Remote Viewer

Keysight Technologies Long-Term, Remote Monitoring of Satellite Performance

Keysight Technologies Connecting Data Acquisition Systems to a Wireless Network

PXI and AXIe Modular Instrumentation

Agilent E5061B Network Analyzer. 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz

Keysight Technologies Understanding and Improving Network Analyzer Dynamic Range. Application Note

Keysight E2688A, N5384A High-Speed Serial Data Analysis and Clock Recovery Software

Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet

PCI Express 1.0 (2.5 GT/s) Protocol Test

Testing WiMAX receiver performance in a multipath propagation environment using Agilent s E6651A with an EB Propsim C8 radio channel emulator

Keysight N6854A Geolocation System. Configuration Guide

PCI Express Transmitter Electrical Validation and Compliance Testing with Agilent Infiniium Oscilloscopes

81110A Pulse Pattern Generator Simulating Distorted Signals for Tolerance Testing

Agilent Compatibility of USB Power Sensors with Agilent Instruments. Application Note

Agilent 8762/3/4A,B,C Coaxial Switches

The Real Total Cost of Ownership of Your Test Equipment

Keysight Technologies The Advantages Of Remote Labs In Engineering Education

Agilent Network Monitoring Content Intercept Manager

Agilent VEE Pro Data Sheet

Application Note. Double click System tray icon after service started. Click start/continue button

Agilent Mobile WiMAX R&D Test Set Solutions: Software and Technical Support Contract

Keysight Technologies Solutions for WLAN ac Manufacturing Test

Agilent 4339B/4349B High Resistance Meters

Agilent 4338B Milliohm Meter

Foreign Taxes Paid and Foreign Source Income INTECH Global Income Managed Volatility Fund

Agilent N5970A Interactive Functional Test Software: Installation and Getting Started

TDS5000B, TDS6000B, TDS/CSA7000B Series Acquisition Modes

Keysight Technologies N9320B RF Spectrum Analyzer

Keysight N9320B RF Spectrum Analyzer

Agilent E5063A ENA Series Network Analyzer. 100 khz to 4.5/ 8.5/18 GHz

Digital Audio Broadcasting Receiver Testing Solutions

Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity

PCI Express Transmitter PLL Testing A Comparison of Methods. Primer

How does a venture capitalist appraise investment opportunities?

Agilent EEsof EDA.

Agilent OSS access7 Signaling Meter

Image Lab Software for the GS-900 Densitometer

Installation Guide E Dielectric Probe Kit 85071E Materials Measurement Software

APPLICATION NOTE. Easily Create Custom Waveform Plug-Ins With Waveform Creator Application Software

RF Measurements Using a Modular Digitizer

Think SAFETY When Selecting a Handheld Multimeter

Agilent P940xA/C Solid State PIN Diode Switches

Agilent EEsof EDA.

Keysight Technologies How to Choose your MAC Lever. Technical Overview

Measuring ACLR Performance in LTE Transmitters. Application Note

AN3353 Application note

Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer

Agilent Modular Products M9392A PXI Vector Signal Analyzer System with Streaming Capability

Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note

TekConnect Adapters. TCA75 TCA-BNC TCA-SMA TCA-N TCA-292MM Data Sheet. Applications. Features & Benefits

High-voltage Differential Probes TMDP THDP THDP P5200A - P5202A - P5205A - P5210A

MERCER S COMPENSATION ANALYSIS AND REVIEW SYSTEM AN ONLINE TOOL DESIGNED TO TAKE THE WORK OUT OF YOUR COMPENSATION REVIEW PROCESS

1.5 GHz Active Probe TAP1500 Datasheet

Global Investing 2013 Morningstar. All Rights Reserved. 3/1/2013

Agilent Technologies N5393B PCI Express 2.0 (Gen2) Electrical Performance Validation and Compliance Software for Infiniium Oscilloscopes

Get the benefits of Norgren s unique range of Online services

Agilent OSS Customer-Centric Service Manager

Agilent U4301A PCI Express 3.0 Analyzer Module. Data Sheet. Agilent solutions for PCIe Gen1 to Gen3 analysis and emulation.

Keysight Technologies Quickly Identify and Characterize Thermal Measurement Points

The Role of Banks in Global Mergers and Acquisitions by James R. Barth, Triphon Phumiwasana, and Keven Yost *

Agilent Model Quality Assurance (MQA) Industry Standard SPICE Model Signoff and Acceptance Software

Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview

Accuracy counts! SENSORS WITH ANALOG OUTPUT

Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform

Agilent Evolution of Test Automation Using the Built-In VBA with the ENA Series RF Network Analyzers

Transcription:

How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer Application Note Introduction In a pulsed radar system, one of the key transmitter-side components is the pulse power amplifier (see Figure 1). Power amplifiers, whether solid state or vacuum tube, used in pulse radar systems need to output up to several kilowatts of power. Since WWII, engineers have been trying to perfect the design of pulsed power amplifiers to support wider bandwidth, thus allowing more data to be transferred, and to increase efficiency to enhance power consumption efficiency. RF IN RF OUT Waveform Generator Pulse Power Amplifier Modulator Power Supply Figure 1. Transmitter block diagram

Introduction One of the key measurements required for evaluating a wide bandwidth (greater than 100 MHz) using a pulse power amplifier is pulse rise time. Currently there are several instruments capable of measuring the rise time of the RF pulsed power amplifier, including a high sampling oscilloscope with a diode detector, an RF peak power meter, and an RF spectrum analyzer using zero-span mode. This application note focuses on a new measurement instrument: the Agilent Technologies 8990B peak power analyzer (PPA). It is the highest performance peak power meter in the power meter test and measurement industry. The 8990B is easy-touse and provides highly accurate fast rise/fall times measurements. This application note illustrates how using the 8990B peak power analyzer allows you to confidently measure sub-5 nanosecond rise time on radar pulses (Figure 2). The document also explains how equivalent time sampling (ETS) is used to measure rise time. Some tips are provided to show you how to ensure you obtain the correct rise/fall time measurement results. Pulse rise and fall time measurement function and results Figure 2. Screen shot of the 8900B PPA measuring a fast RF pulse rise time 2

Pulse Shaping (Profiling) When developing a wideband power amplifier or transmitter device, signal processing and modeling techniques such as the pulse profiling or shaping have been used to make sure RF pulses transmitted out meet the bandwidth requirement (see Figure 3). Filtering design also plays an important role in determining the RF pulse profile. For example, a sloping leading edge (rise time) affects the minimum range as well as range accuracy or resolution. This means that the range resolution of a pulsed radar system is fundamentally limited by the bandwidth of the transmitted pulse; the wider the bandwidth, the better the range resolution. Amplitude 12 10 8 6 4 2 0-2 0 400.e 3 800.e 3 1.2 1.6 2 Various Pulse Profiling 0 400.e 3 800.e 3 1.2 1.6 2 Time Figure 3. Pulse profiling illustration 3

The Challenge in Measuring Fast Rise Time When trying to measure a very fast pulse rise/fall time on an RF pulse, minimizing measurement uncertainty is important. Consequently, obtaining an accurate rise time on the device under test (DUT) requires the use of a measurement device with a rise time that is faster than the expected rise/ fall time. To further reduce measurement uncertainty, it is desirable to directly measure the RF pulse from the DUT, eliminating the use of adaptors or convertors. As mentioned earlier, the 8990B PPA has a sub-5 nanosecond rise time capability using equivalent time sampling (ETS). (ETS is explained in the next section.) Using the 8990B with two wideband and fast power sensors (the N1923A and N1924A), the DUT s RF pulse signals can be measured directly. This setup is ideal, particularly when measuring a fast rise/fall time, since measurement uncertainty is kept to minimum because the signal correction is done inside the power meter. Figure 4 shows the rise time measurement error (in percentage) versus the DUT s actual rise time. The graph is constructed based on the following simple error equation: 2 2 RisetimeInstruments + RisetimeDUT RisetimeDUT Risetime Error = 100% Risetime DUT 60.0% 50.0% Agilent 8990B PPA Rise Time Measurement Error (%) Measurement Error (%) 40.0% 30.0% 20.0% 10.0% 0.0% 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 User DUT rise time ( nsec ) Figure 4. 8990B rise time measurement error vs. DUT rise time 4

Equivalent Time Sampling Equivalent time sampling (ETS) works by constructing a picture or waveform of the input signal by accumulating samples over many waves of triggering cycles. As illustrated in Figure 5, when the ETS feature of the 8990B PPA is turned on, an accurate timeto-digital converter (DTC) subdivides 10 more bins (points) over multiple triggers samples, meaning the DTC can accurately locate 10 more smaller time samples in each trigger. The sampling rate of the power meter channels (channels 1 and 4) are effectively boosted up from the 100 MSa/s real-time sampling to 1 GSa/s sampling when ETS is used. This is a common sampling technique and is used in some of the conventional digital oscilloscopes to effectively increase the sampling rate to capture fast and repetitive signals. Subsequent samples from multiple triggers and filled into the 10 sub division bins Trigger level Bin 0 Bin 1 Bin 2 Bin 3 Bin 9 Reconstruction of the repetitive signals over multiple trigger events, with accurate delta time interval Figure 5. 8990B equivalent time sampling illustration 5

Equivalent Time Sampling In the 8990B PPA, ETS is turned on automatically only when the time base setting is set below 500 ns/ div. At 500 ns/div and above, ETS is turned off; in other words it is in real time sampling of 100 MSa/s. To check whether ETS is turned on or off, simply click on the Status tab at the bottom of the 8990B measurement panel (shown in Figure 6 below). The panel also shows if the ETS acquisition status is acquiring or completed. Figure 6. 8990B ETS status bar The limitation of this measurement method is that ETS works best on repetitive signals. ETS cannot be used for single-shot or non-repetitive signals because there are not be enough data samples to reconstruct the signal waveform. For these instances, there is an option to manually turn off the ETS even if the time base is set to below 500 ns/div. This allows you to analyze non-repetitive RF pulses, provided the rise/fall time is greater than 50 ns. The other limitation of ETS is the triggering time. The longer the waiting time for one trigger to be armed and completed, the longer it takes for ETS to fill the 10 acquisition bins and subsequently build back the RF pulse waveform. 6

Tips For Obtaining Correct Rise/Fall Time Measurement The flowchart shown in Figure 7 is a guideline to optimize the rise/ fall time measurement result in the 8990B PPA. As it shows, there are two pre-determined factors that you need to know in order to obtain the best rise/all time measurement accuracy. The first is for you to know the expected rise time. This data can be found in the DUT s data specification documents. The second is for you to know whether the signal is repetitive. In most testing conditions, DUT pulses are repetitive since they can be generated from external RF signal sources or when the DUT is operating in service or self test mode. Rise/Fall time Excepted? YES Less than 50 nsec NO Repetitive Signals? YES NO Change time base to get the rising/falling edge of the pulse to occupy ~ 50% the display Set time base less than 500 ns/div, make sure ETS is turned on Manually turn off the ETS. Rise/fall time measurement uncertainty will be high Change time base to get the rising/falling edge of the pulse to occupy ~ 50% the display Figure 7. Flow Chart to Get Optimized Rise/Fall Time Measurement 7

Conclusion Accurately measuring RF pulse nanosecond range rise/fall time requires the use of a suitable measurement instrument that is properly set up. In a wide band pulse power amplifier, accuracy is especially important since the rise/fall time measurement becomes a critical performance measurement parameter. The new Agilent 8990B peak power analyzer is an ideal test solution, ensuring measurement accuracy with its 5 nano-seconds rise time performance and easy-to-use graphic interface. Reference 1) Principles Of Modern Radar: Basic Principles, Mark A. Richards, James A. Scheer, William A. Holm, Scitech Publishing, ISBN: 978-1-891121-52-4, 2010 2) Agilent 8990B Peak Power Analyzer, User s Guide, part number 08990-90005 3) Baseband Pulse Shaping for Improved Spectral Efficiency Technical Note, part number 5989-9492EN 8

www.agilent.com www.agilent.com/find/peakpoweranalyzer Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent s measurement expertise and product breadth, combined with channel partner convenience. www.axiestandard.org AdvancedTCA Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Agilent is a founding member of the AXIe consortium. www.lxistandard.org LAN extensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. www.pxisa.org PCI extensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based highperformance measurement and automation system. Agilent Advantage Services is committed to your success throughout your equipment s lifetime. To keep you competitive, we continually invest in tools and processes that speed up calibration and repair and reduce your cost of ownership. You can also use Infoline Web Services to manage equipment and services more effectively. By sharing our measurement and service expertise, we help you create the products that change our world. www.agilent.com/find/advantageservices www.agilent.com/quality For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas Canada (877) 894 4414 Brazil (11) 4197 3500 Mexico 01800 5064 800 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 375 8100 Europe & Middle East Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 /minute Germany 49 (0) 7031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 United Kingdom 44 (0) 131 452 0200 For other unlisted countries: www.agilent.com/find/contactus Revised: June 8, 2011 Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2011 Published in USA, December 23, 2011 5990-9661EN