PRINCIPLE REF 475136. ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE



Similar documents
PRINCIPLE. REF (200 tests/cartridge) REF (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

For In Vitro Diagnostic Use

Laboratory Procedure Manual

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

How To Test For Creatinine

Laboratory Procedure Manual

Hepatitis and Retrovirus. LIAISON XL Accurate detection of HIV infection. HIV Ab/Ag FOR OUTSIDE THE US AND CANADA ONLY

Toxoplasma gondii IgM ELISA Kit Protocol

Intact PTH. system. Intact PTH /R5. Key to symbols used 8K25

Carcinoembryonic Antigen (CEA)

Validation and Calibration. Definitions and Terminology

LIAISON XL HBsAg Quant

Biochemistry Validation Form

LIAISON XL HCV Ab Accurate diagnosis of the early stage of HCV infection

Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.

This package insert contains information to run the Creatine Kinase assay on the ARCHITECT c Systems and the AEROSET System.

Cold Agglutination Titer detecting Cold Reacting Antibodies

Vitamin B12. ADVIA Centaur System

Measles (Rubeola) IgM ELISA Catalog No. CB (96 Tests)

Cancer Antigen CA 125

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma

IgE (Human) ELISA Kit

ALKALINE PHOSPHATASE

This package insert contains information to run the Lactic Acid assay on the ARCHITECT c Systems and the AEROSET System.

CDC-Lipid Standardization Program. Laboratory Data Collection System (LDCS) Instruction Manual

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

ABORhCard. ABORhCard Package Insert ABO and Rh Blood Grouping Device

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems and the AEROSET System.

Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA

HiPer RA Test Teaching Kit

Malondialdehyde (MDA) ELISA

This package insert contains information to run the Uric Acid assay on the ARCHITECT c Systems and the AEROSET System.

Lab 2. Spectrophotometric Measurement of Glucose

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

RealLine HCV PCR Qualitative - Uni-Format

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE

1.5 Function of analyte For albumin, see separate entry. The immunoglobulins are components of the humoral arm of the immune system.

STEP-BY-STEP INSTRUCTIONS FOR INVESTIGATIONAL USE. Rapid HCV Antibody Test FOR ORAQUICK RAPID HCV ANTIBODY TEST

Rat Fibronectin ELISA Kit

This package insert contains information to run the Glucose assay on the ARCHITECT c Systems and the AEROSET System.

DHEA-S. system DHEA-S /R4. Key to symbols used 8K27

Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2

XV.Quality Assurance in the Blood Bank

Anti-ss-DNA Screen (IgA, IgG, IgM) ELISA

Development and Validation of In Vitro Diagnostic Tests. YC Lee, Ph.D. CEO

Epinephrine/Norepinephrine ELISA Kit

510(K) SUMMARY FEB Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

Anti-Zona Pellucida Antibody Latex Agglutination Test

Human Free Testosterone(F-TESTO) ELISA Kit

TOTAL PROTEIN FIBRINOGEN

Blood, Plasma, and Cellular Blood Components INTRODUCTION

How Does a Doctor Test for AIDS?

REF. Intended Use. Summary and Explanation

UNIT: Total and Direct Bilirubin

HIV Rapid Test Kits USAID Approval and Technical Guidance June 17, 2013 Revision

Human Luteinizing Hormone (LH) Custom Kit

SMF Awareness Seminar 2014

ab Complement C1q Human ELISA Kit

Rheumatoid Factor (RF)

JOINT COMMISSION INTERNATIONAL ACCREDITATION STANDARDS FOR. 2nd Edition

ab Albumin Human ELISA Kit

Dengue IgM ELISA. For the quantitative determination of IgM-class antibodies to Dengue Virus in serum.

Direct Antiglobulin Test (DAT)

ASSURING THE QUALITY OF TEST RESULTS

TRI-LEVEL CARDIAC CONTROL (CRD CONTROL 1, 2, 3)

Teriflunomide is the active metabolite of Leflunomide, a drug employed since 1994 for the treatment of rheumatoid arthritis (Baselt, 2011).

Human Parathyroid Hormone (PTH) ELISA Kit

HBV Quantitative Real Time PCR Kit

QUANTA Lite TM CCP 3.1 IgG/IgA ELISA For In Vitro Diagnostic Use CLIA Complexity: High

CANADIAN RULE BASED CLASSIFICATION SYSTEM (IVDD) Life Sciences British Columbia NRC-Industry Research Assistance Program

Guidance for Industry and FDA Staff

Cancer Antigen CA125 Human ELISA Kit

In vitro diagnostic reagent, calibrator and control material stability

This package insert contains information to run the Creatinine assay on the ARCHITECT c Systems and the AEROSET System.

MEDICAL DEVICE GUIDANCE

ab83369 Alkaline Phosphatase Assay kit (Colorimetric)

Thrombin Generation Assay

OneStep Fecal Occult Blood RapiDip InstaTest. Cat #

TRIIODOTHYRONINE (T3) ELISA Kit Protocol

EXERCISE 5: ERYTHROCYTES SEDIMENTATION RATE - ESR, SED RATE

SYSTEM. For In Vitro Diagnostic Use. Store at 2-8 C

Standardization, Calibration and Quality Control

Chapter. Guaiac Screening CHAPTER 4: GUAIAC TESTING SCREENING FOR OCCULT BLOOD. Page 1 of 5 Guaiac doc 6/24/2005

WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS

BY UV SPECTROPHOTOMETRY

Calcium. Table 1: Difference between method means in percent

Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians

Appendix B: Provincial Case Definitions for Reportable Diseases

Measuring Protein Concentration through Absorption Spectrophotometry

Transcription:

SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Rheumatoid Factor REF 475136 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Date Reviewed by: Date PRINCIPLE INTENDED USE reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems CAL 5 Plus, is intended for quantitative determination of rheumatoid factor concentration in human serum or plasma. CLINICAL SIGNIFICANCE Rheumatoid Factors are antibodies that react with the individual s own immunoglobulin. The antibodies are usually directed against the Fc region of the IgG molecule. 1 Rheumatoid Factor can be detected in the serum of the majority of patients with rheumatoid arthritis and is important for the diagnosis and prognosis of those patients with higher concentrations. These patients tend to suffer a more severe form of the illness and develop extra-joint complications more readily. 2,3 Rheumatoid factors are not disease specific and can occur in lower frequencies in several other autoimmune disorders, chronic inflammation and normal individuals. 4,5 METHODOLOGY reagent is used to measure rheumatoid factor concentration by a turbidimetric method. 6,7 In the reaction, rheumatoid factor in the patient sample combines with specific antibody to form insoluble antigen-antibody complexes. The SYNCHRON System(s) automatically proportions the appropriate sample and reagent volumes into a cuvette. The ratio used is one part sample to 50 parts reagent. The system monitors the change in absorbance at 380 nanometers. This change in absorbance is proportional to the concentration of rheumatoid factor in the sample and is used by the System to calculate and express concentration of rheumatoid factor. A multi-point, nonlinear calibrator standard (in accordance with the World Health Organization ) was used to establish lot-specific calibration parameters. 8 CHEMICAL REACTION SCHEME AUGUST 2010 Page 1 of 11

SPECIMEN TYPE OF SPECIMEN Biological fluid samples should be collected in the same manner routinely used for any laboratory test. 9 Freshly drawn serum or plasma are the preferred specimens. Acceptable anticoagulants are listed in the PROCEDURAL NOTES section of this chemistry information sheet. Whole blood or urine are not recommended for use as a sample. SPECIMEN STORAGE AND STABILITY 1. Tubes of blood are to be kept closed at all times and in a vertical position. It is recommended that the serum or plasma be physically separated from contact with cells within two hours from the time of collection. 10 2. Separated serum or plasma should not remain at room temperature longer than 8 hours. If assays are not completed within 8 hours, serum or plasma may be stored at +2 C to +8 C for up to 7 days. Frozen samples are not recommended. 9 Additional specimen storage and stability conditions as designated by this laboratory: SAMPLE VOLUME The optimum volume, when using a 0.5 ml sample cup, is 0.3 ml of sample. For optimum primary sample tube volumes and minimum volumes, refer to the Primary Tube Sample Template for your system. CRITERIA FOR UNACCEPTABLE SPECIMENS Refer to the PROCEDURAL NOTES section of this chemistry information sheet for information on unacceptable specimens. Criteria for sample rejection as designated by this laboratory: PATIENT PREPARATION Special instructions for patient preparation as designated by this laboratory: Page 2 of 11 AUGUST 2010

SPECIMEN HANDLING Special instructions for specimen handling as designated by this laboratory: REAGENTS CONTENTS Each kit contains the following items: Two Reagent Cartridges (2 x 100 tests) One lot-specific Parameter Card VOLUMES PER TEST Sample Volume 9 µl ORDAC Sample Volume 4 µl Total Reagent Volume 300 µl Cartridge Volumes A 250 µl B 50 µl C REACTIVE INGREDIENTS REAGENT CONSTITUENTS Rheumatoid Factor Particle Reagent Rheumatoid Factor Reagent Buffer 36.3 ml 7.4 ml Also non-reactive chemicals necessary for optimal system performance. CAUTION Sodium azide preservative may form explosive compounds in metal drain lines. See National Institute for Occupational Safety and Health Bulletin: Explosive Azide Hazards (8/16/76). AUGUST 2010 Page 3 of 11

CAUTION Because this product is of human origin, it should be handled as though capable of transmitting infectious diseases. Each serum or plasma donor unit used in the preparation of this material was tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for antibodies to HIV and HCV and nonreactive for HbsAg. Because no test method can offer complete assurance that HIV, hepatitis B virus, and hepatitis C virus or other infectious agents are absent, this material should be handled as though capable of transmitting infectious diseases. This product may also contain other human source material for which there is no approved test. The FDA recommends such samples to be handled as specified in Centers for Disease Control s Biosafety Level 2 guidelines. 11 MATERIALS NEEDED BUT NOT SUPPLIED WITH REAGENT KIT SYNCHRON Systems CAL 5 Plus At least two levels of control material Saline REAGENT PREPARATION NOTICE Failure to mix the reagent prior to use will result in erroneous values. 1. Gently invert the cartridge several times prior to loading onto the system. 2. Check for bubbles or foam in compartments; break any bubbles. ACCEPTABLE REAGENT PEORMANCE The acceptability of a reagent is determined by successful calibration and by ensuring that quality control results are within your facility s acceptance criteria. REAGENT STORAGE AND STABILITY reagent when stored unopened at +2 C to +8 C, will obtain the shelf-life indicated on the cartridge label. Once opened, the reagent is stable for 60 days at +2 C to +8 C unless the expiration date is exceeded. DO NOT FREEZE. Reagent storage location: CALIBRATION CALIBRATOR REQUIRED SYNCHRON Systems CAL 5 Plus CALIBRATOR PREPARATION No preparation is required. Page 4 of 11 AUGUST 2010

CALIBRATOR STORAGE AND STABILITY SYNCHRON Systems CAL 5 Plus is stable until the expiration date printed on the calibrator bottle if stored capped in the original container at +2 C to +8 C. Calibrator storage location: CAUTION Because this product is of human origin, it should be handled as though capable of transmitting infectious diseases. Each serum or plasma donor unit used in the preparation of this material was tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for antibodies to HIV and HCV and nonreactive for HbsAg. Because no test method can offer complete assurance that HIV, hepatitis B virus, and hepatitis C virus or other infectious agents are absent, this material should be handled as though capable of transmitting infectious diseases. This product may also contain other human source material for which there is no approved test. The FDA recommends such samples to be handled as specified in Centers for Disease Control s Biosafety Level 2 guidelines. 11 CALIBRATION INFORMATION 1. The system must have a lot-specific parameter card and a valid calibration adjustment in memory before controls or patient samples can be run. 2. Under typical operating conditions the reagent cartridge must be calibrated every 30 days and also with certain parts replacements or maintenance procedures, as defined in the SYNCHRON LX Maintenance Manual and Instrument Log, or the UniCel DxC 600/800 System Instructions For Use (IFU) manual. This assay has within-lot calibration available. Refer to the SYNCHRON LX Operations Manual, or the UniCel DxC 600/800 System Instructions For Use (IFU) manual for information on this feature. 3. For detailed calibration instructions, refer to the SYNCHRON LX Operations Manual, or the UniCel DxC 600/800 System Instructions For Use (IFU) manual. 4. The system will automatically perform checks on the calibration and produce data at the end of calibration. In the event of a failed calibration, the data will be printed with error codes and the system will alert the operator of the failure. For information on error codes, refer to the SYNCHRON LX Diagnostics and Troubleshooting Manual, or the UniCel DxC 600/800 System Instructions For Use (IFU) manual. TRACEABILITY For Traceability information refer to the Calibrator instructions for use. QUALITY CONTROL At least two levels of control material should be analyzed daily. In addition, these controls should be run with each new calibration, with each new reagent cartridge, and after specific maintenance or troubleshooting procedures as detailed in the appropriate system manual. More frequent use of controls or the use of additional controls is left to the discretion of the user based on good laboratory practices or laboratory accreditation requirements and applicable laws. AUGUST 2010 Page 5 of 11

The following controls should be prepared and used in accordance with the package inserts. Discrepant quality control results should be evaluated by your facility. Table 1.0 Quality Control Material CONTROL NAME STORAGE TESTING PROCEDURE(S) 1. If necessary, load the reagent onto the system. A lot-specific parameter card must be loaded one time for each lot. 2. After reagent load is completed, calibration may be required. 3. Program samples and controls for analysis. 4. After loading samples and controls onto the system, follow the protocols for system operations. For detailed testing procedures, refer to the SYNCHRON LX Operations Manual, or the UniCel DxC 600/800 System Instructions For Use (IFU) manual. CALCULATIONS The SYNCHRON System(s) performs all calculations internally to produce the final reported result. The system will calculate the final result for sample dilutions made by the operator when the dilution factor is entered into the system during sample programming. REPORTING RESULTS Equivalency between the SYNCHRON LX and UniCel DxC 600/800 Systems has been established. Chemistry results between these systems are in agreement and data from representative systems may be shown. REFERENCE INTERVALS Greater than 95% of a group of 125 apparently healthy adults in Southern California showed no detectable (<20 IU/mL) with this assay. Published "upper limits of normal" vary widely among different sources. 10,12,13 Each laboratory must establish its own reference interval based on its patient population. The presence of in patient serum should be considered as one criterion of rheumatoid disease, and not as a definitive diagnosis. 14 Table 2.0 Reference intervals INTERVALS CONVENTIONAL UNITS SYNCHRON Serum or Plasma < 20 IU/mL Laboratory INTERVALS CONVENTIONAL UNITS Page 6 of 11 AUGUST 2010

Additional reporting information as designated by this laboratory: PROCEDURAL NOTES ANTICOAGULANT TEST RESULTS 1. If plasma is the sample of choice, the following anticoagulants were found to be compatible with this method based on a study of 20 healthy volunteers: Table 3.0 Acceptable Anticoagulants a ANTICOAGULANT LEVEL TESTED FOR IN VITRO INTEERENCE AVERAGE PLASMA-SERUM BIAS (IU/mL) Lithium Heparin 14 Units/mL NSI b Sodium Heparin 14 Units/mL NSI a Data shown was collected using SYNCHRON CX Systems. Equivalency between SYNCHRON LX Systems has been established by Deming regression analysis to SYNCHRON CX Systems. b NSI = No Significant Interference (within ± 20 IU/mL or 8%). 2. The following anticoagulants were found to be incompatible based on the same study: Table 4.0 Incompatible Anticoagulants a ANTICOAGULANT LEVEL TESTED FOR IN VITRO INTEERENCE PLASMA-SERUM BIAS (IU/mL) a EDTA 1.5 mg/dl +180 LIMITATIONS None identified. Bias is based on worst case instead of average. Plus (+) or minus (-) signs in this column signify positive or negative bias. INTEERENCES 1. The following substances were tested for interference with this methodology: Table 5.0 Interferences a SUBSTANCE SOURCE LEVEL OBSERVED EFFECT Hemoglobin RBC hemolysate 500 mg/dl NSI b Bilirubin Bovine 30 mg/dl NSI Lipemia Intralipid c 400 mg/dl NSI a Data shown was collected using SYNCHRON CX Systems. Equivalency between SYNCHRON LX Systems has been established by Deming regression analysis to SYNCHRON CX Systems. b NSI = No Significant Interference (within ±20 IU/mL or 8%). c Intralipid is a registered trademark of KabiVitrum, Inc., Clayton, NC 27250. AUGUST 2010 Page 7 of 11

2. Serum samples may yield different concentration values for rheumatoid factor when analyzed at various dilutions. This phenomenon may be caused by the presence of rheumatoid factor complexes with inhibitors or interfering substances present in the serum. Sample dilution will reduce the effective concentration of the interfering substances so that the final result obtained for rheumatoid factor in the diluted sample will be increased over the result obtained for the undiluted (neat) or lesser diluted sample. 3. Refer to References (15,16,17) for other interferences caused by drugs, disease and preanalytical variables. PEORMANCE CHARACTERISTICS ANALYTIC RANGE The SYNCHRON System(s) method for the determination of this analyte provides the following analytical range: Table 6.0 Analytical Range Serum or Plasma Serum or Plasma (ORDAC) CONVENTIONAL UNITS 20 800 IU/mL 650 1770 IU/mL Samples with concentrations outside of the analytical range will be reported as "<20 IU/mL" or ">800 IU/mL" (ORDAC "<650 IU/mL" or ">1770 IU/mL"). Samples reported out as greater than the analytical range should be confirmed by enabling ORDAC, or diluting with saline, and reanalyzing. If manual dilution is used the appropriate dilution factor should be applied to the reported result. Samples reported out as "SUPPRESSED" due to RXN ERROR should be reanalyzed. REPORTABLE RANGE (AS DETERMINED ON SITE): Table 7.0 Reportable Range CONVENTIONAL UNITS S.I. UNITS SENSITIVITY Sensitivity is defined as the lowest measurable concentration which can be distinguished from zero with 95% confidence. Sensitivity for determination is 20 IU/mL. EQUIVALENCY Equivalency was assessed by Deming regression analysis of patient samples to accepted clinical methods. Serum or Plasma (in the range of 21 to 590 IU/mL): Y (SYNCHRON LX Systems) = 1.104X - 7.56 N = 86 MEAN (SYNCHRON LX Systems) = 142.9 MEAN (IMMAGE ) = 136.3 CORRELATION COEFFICIENT (r) = 0.973 Page 8 of 11 AUGUST 2010

Refer to References (18) for guidelines on performing equivalency testing. PRECISION A properly operating SYNCHRON System(s) should exhibit precision values less than or equal to the following: Table 8.0 Precision Values a b TYPE OF PRECISION Within-run Total 1 SD IU/mL CHANGEOVER VALUE a IU/mL % CV Serum/Plasma 10.0 250 4.0 Serum/Plasma (ORDAC) NA b NA 10.0 Serum/Plasma 15.0 250 6.0 Serum/Plasma (ORDAC) NA NA 15.0 When the mean of the test precision data is less than or equal to the changeover value, compare the test SD to the SD guideline given above to determine the acceptability of the precision testing. When the mean of the test precision data is greater than the changeover value, compare the test % CV to the guideline given above to determine acceptability. Changeover value = (SD guideline/cv guideline) x 100. NA = Not applicable. Comparative performance data for a SYNCHRON LX System evaluated using the NCCLS Proposed Guideline EP5-T2 appears in the table below. 19 Each laboratory should characterize their own instrument performance for comparison purposes. Table 9.0 NCCLS EP5-T2 Precision Estimate Method a TYPE OF IMPRECISION Within-run Total No. Systems No. Data Points a Test Mean Value (IU/mL) EP5-T2 Calculated Point Estimates SD %CV Serum Control 1 1 80 66.9 2.56 3.83 Serum Control 2 1 80 220.8 3.15 1.43 Serum Control 1 1 80 66.9 2.79 4.17 Serum Control 2 1 80 220.8 3.72 1.68 The point estimate is based on the pooled data from one system, run for twenty days, two runs per day, two observations per run on an instrument operated and maintained according to the manufacturer s instructions. NOTICE These degrees of precision and equivalency were obtained in typical testing procedures on a SYNCHRON LX System and are not intended to represent the performance specifications for this reagent. ADDITIONAL INFORMATION For more detailed information on SYNCHRON LX Systems or UniCel DxC Systems, refer to the appropriate system manual. SHIPPING DAMAGE If damaged product is received, notify your Beckman Coulter Clinical Support Center. AUGUST 2010 Page 9 of 11

REFERENCES 1. Johnson, P. M., Faulk, W. P., "Rheumatoid Factor: Its Nature, Specificity and Production in Rheumatoid Arthritis", Clin. Immunol. Immunopathol., 6:414 440 (1976). 2. Zrein M., De Marcillac, G., Van Regenmortel M. H. V., "Quantitation of Rheumatoid Factor by Enzyme Immunoassay Using Biotinylated IgG", J. Immunol. Methods, 87:229 237 (1986). 3. Msai, A. T., Maldonado-Cocco, J. A., Kaplan, S. B., Frigenbaum, S. L., Chandler, R. W., "Prospective Study of the Early Course of Rheumatoid Arthritis in Young Adults: Comparison of Patients With and Without Rheumatoid Factor Positively at Entry and Identification Variables Correlating with Outcome", Semin. Arthritis Rheum., 5:229 326 (1976). 4. Wagner, O., Ann Clin Res., 7:168 182 (1975). 5. Boyden, A., Button, E., Germerog, D., "Precipitin Testing With Special Reference to the Measurement of Turbidity", J. Immunol., 57:211 (1947). 6. Hellsing, K., "The Effects of Different Polymers for Enhancement of the Antigen-Antibody Reaction as Measured with Nephelometry", Protides of the Biological Fluids, 23:579 (1973). 7. Spaun, J., Bentson, M. W., Olsen-Larsen, S., Hewitt, L. F., International Standard for Antistreptolysin-O, WHO Bulletin, 24:241 279 (1961). 8. Tietz, N. W., "Specimen Collection and Processing; Sources of Biological Variation", Textbook of Clinical Chemistry, 2nd Edition, W. B. Saunders, Philadelphia, PA (1994). 9. National Committee for Clinical Laboratory Standards, Procedures for the Handling and Processing of Blood Specimens, Approved Guideline, NCCLS publication H18-A, Villanova, PA (1990). 10. Painter, P., Cope, J., Smith, J., "Reference Information for the Clinical Laboratory", Chapter 50, Tietz, Textbook of Clinical Chemistry, Burtis, C., Ashwood, E., eds.) W.B. Saunders, New York (1999). 11. CDC-NIH manual, Biosafety in Microbiological and Biomedical Laboratories, U.S. Government Printing Office, Washington, D.C. (1984). 12. Tietz, N. W., ed.,clinical Guide to Laboratory Tests, W.B. Saunders, New York (1995). 13. Mayo Medical Laboratories, Interpretive Handbook: Interpretive Data for Diagnostic Laboratory Tests, Mayo Clinic, Rochester, MN (1997). 14. Arnett, F., Edworthy, S., Bloch, D., McShane, D., Fries, J., Cooper, N., et al., The American Rheumatism Association 1987, " Revised Criteria for the Classification of Rheumatoid Arthritis", Arthritis Rheum., 31:315 324 (1998). 15. Young, D. S., Effects of Drugs on Clinical Laboratory Tests, 4th Edition, AACC Press, Washington, D. C. (1995). 16. Friedman, R. B., Young, D. S.,Effects of Disease on Clinical Laboratory Tests, 3rd Edition, AACC Press, Washington, D.C. (1997). 17. Young, D. S., Effects of Preanalytical Variables on Clinical Laboratory Tests, 2nd Edition, AACC Press, Washington, D. C. (1997). 18. National Committee for Clinical Laboratory Standards, Method Comparison and Bias Estimation Using Patient Samples, Approved Guideline, NCCLS publication EP9-A, Villanova, PA (1995). Page 10 of 11 AUGUST 2010

19. National Committee for Clinical Laboratory Standards, Precision Performance of Clinical Chemistry Devices, Tentative Guideline, 2nd Edition, NCCLS publication EP5-T2, Villanova, PA (1992). Beckman Coulter Ireland Inc., Mervue Business Park, Mervue, Galway, Ireland (353 91 774068) Beckman Coulter, Inc., 250 South Kraemer Blvd., Brea, CA 92821 AUGUST 2010 Page 11 of 11