Articaine & Epinephrine Injection according to USP method



Similar documents
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

ERDOSTEINE - MONOGRAPH.

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PARACETAMOL ORAL SUSPENSION (September 2010)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Tamsulosin Hydrochloride Capsules

Analysis of Various Vitamins in Multivitamin Tablets

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

Drospirenone and Ethinyl Estradiol Tablets. Type of Posting. Revision Bulletin Posting Date. 31 July 2015 Official Date

Unique Bonding Chemistry

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

International Journal of Pharma and Bio Sciences V1(2)2010

ETHYL LAUROYL ARGINATE

Minor editorial changes have been made to update the monograph to the current USP style.

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Implementing New USP Chapters for Analytical Method Validation

I. ACID-BASE NEUTRALIZATION, TITRATION

A Complete Solution for Method Linearity in HPLC and UHPLC

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Guidance for Industry

High sensitivity assays using online SPE-LC-MS/MS -How low can you go? Mohammed Abrar Unilabs York Bioanalytical solutions, York, UK

Determination of Calcium and Magnesium in Brine

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

How To Test For Contamination In Large Volume Water

Analytical Test Method Validation Report Template

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

SUCRALOSE. White to off-white, practically odourless crystalline powder

Preparation of frequently used solutions

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: )

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

ß-CYCLODEXTRIN SYNONYMS

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Vitamin C quantification using reversed-phase ion-pairing HPLC

ph Alkalinity of Water

一 Development of microchip integrated with electrochemical sensor in conjunction with indium tin oxide electrode

Chapter 28: High-Performance Liquid Chromatography (HPLC)

Amino Acid Analyzer L-8900

Introduction. Methods. Sample Processing

VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI

Determination of Caffeine in Beverages HPLC-1

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

Overview. Purpose. Methods. Results

All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &

S olid Ph ase Extraction (SP E) Method:

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

LACHAT METHOD NUMBER D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately g CN/L g CN/L. 09e

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

The Essential CHROMacademy Guide. Mobile Phase Optimization Strategies for Reversed Phase HPLC

Analytical Methods for Cleaning Validation

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

HPLC Column Comparison Screening Study for Reversed Phase Columns

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

Journal of Chemical and Pharmaceutical Research

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

PVV Satyanaryana et al., IJSID, 2012, 2 (1), International Journal of Science Innovations and Discoveries

Anion chromatography using on-line recycled eluents

POROS CaptureSelect affinity columns for highspeed quantification of IgG Fc fusion proteins

Thermo Scientific Syncronis HPLC Columns. Technical Manual

Appendix 5 Overview of requirements in English

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

Tips for practical HPLC analysis

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

Carolina s Solution Preparation Manual

Transcription:

Application Note Biotech & pharmaceutical The most reliable LC-EC applications for Pharmaceutical & Biotech analysis ever formulated Aminoglycosides Amikacin Framycetin Sulphate Gentamicin Sulphate Kanamycin Sulphate Lincomycin Neomycin Spectinomycin Tobramycin PET imaging tracer FDG Macrolide antibiotics Azithromycin Azaerythromycin Clarithromycin Erythromycin Roxithromycin Bioanalysis of pharmaceutics Acetaminophen Artemisinin Dihydro-artemisinin Artemether Etoposide 8-OH-DPAT mesna BNP7787 Vincristine Articaine & Epinephrine Injection according to USP method n U.S. Pharmacopeia 37-NF32 (2014) n Determination of the Epinephrine contents n Analysis of organic and epinephrine-related impurities n Reproducible & Robust Introduction Articaine in combination with epinephrine is used as an anesthetic for dental procedures in a number of European countries, US and Canada. Like other local anesthetic drugs, articaine causes a transient and completely reversible state of anesthesia (loss of sensation). This drug was first synthesized by Rusching in 1969 and brought to the market in Germany by Hoechst AG under the brand name Ultracain [1]. It was approved by the FDA in April 2000 and became available two months later in the United States under the brand name Septocaine [2]. The U.S. Pharmacopoeia monograph for Articaine Hydrochloride and Epinephrine injections describes a method for the analysis of the Epinephrine contents and organic impurity analysis [3]. This method is based on HPLC in combination with electrochemical detection in the DC mode on a glassy carbon working electrode [4]. ROXY Application Note # 217_035_02 Robust Applications, Fluidly Running

Summary The Epinephrine analysis was evaluated on an Antec ALEXYS LC-EC analyzer, using the method and conditions described in the official 2014 USP monograph 37-NF32 for Articaine Hydrochloride and Epinephrine injection [3]. In this application note typical results obtained with the ALEXYS system are reported, demonstrating its performance for the analysis of the Epinephrine contents, organic impurities and Epinephrinerelated impurities in anesthetic products based on Articaine with Epinephrine. Method and results Separation Separation of Epinephrine and its related impurities is achieved using a reverse phase C8 column in combination with an acidic buffered solution with 1-Heptanesulfonate as ion-pairing agent and methanol as organic modifier (isocratic elution). In the monographs the use of the following column type is described for the separation of Epinephrine: size 250 mm, ID 4.0 mm, 5 μm packing L7. The USP packing L7 is described as: Octylsilane chemically bonded to totally porous or superficially porous silica particles 1.5 to 10 μm in diameter, or a monolithic rod. A Phenomenex Luna 5μ C8(2), 250 x 4.6 mm column was chosen for the method evaluation. This specific stationary phase is listed in the USP L7 packing list. Note that the ID of the column is slightly larger (4.6 mm), such variation in internal diameter is allowed by the USP [6]. Table 1 LC-EC Conditions Figure 1: 2 μl injection of a 22 μg/ml Epinephrine Bitartrate RS and 20 μg/ ml Norepinephrine Bitartrate RS solution in 0.5 mg/ml Potassium Metabisulfite (System suitability solution as described in the USP monograph). HPLC ALEXYS Epinephrine Analyzer. Column 4.6 mm ID x 25 cm, 5µm, packing L7 Mobile phase 50 ml of glacial acetic acid and 930 ml water. Adjust with 2N sodium hydroxide to a ph of 3.4. In this solution, dissolve 1.2g of sodium 1-heptanesulfonate and add 1 ml of 0.1 M edetate disodium and 0.298g of potassium chloride. Add 150 ml methanol Diluent 0.5 mg/ml potassium metabisulfite in water Flow rate 1.0 ml/min V injection 2 µl Temperature 30 C for separation and detection Flow cell Sencell TM with 2mm Glassy Carbon working electrode, Ag/AgCl (salt bridge) reference electrode and stainless steel auxiliary electrode, AST setting 2 Potential E= +0.65 V I-cell ca. 3 na ADF 0.5 Hz Range 1 µa and 1 na (for LOD measurements) 2

Detection For the detection of Epinephrine and its related impurities, am-perometric detection in Direct Current (DC) mode is mandatory using a Glassy Carbon (GC) working electrode and Ag/ AgCl reference electrode. The Antec SenCell matches these requirements and was used in this evaluation. The cell was set to a static DC cell potential of +650 mv, the cell current was typical 3 na under the measurement conditions listed in table 1. The temperature for separation and detection was 30 C. Note that for optimal temperature control of the electrochemical detector the ambient temperature in the laboratory does not exceed 23 C. System Suitability A chromatogram of an 2 μl injection of a 22 μg/ml Epinephrine Bitartrate RS and 20 μg/ml Norepinephrine Bitartrate RS solution in 0.5 mg/ml Potassium Metabisulfite is show in figure 1 (system suitability solution as described in the USP monograph). Besides the USP system suitability solution also solutions were analysed containing two known epinephrinerelated substances, Adrenalone and Epinephrine sulfonic acid, see figure 2. The retention times of the substances are listed in table 2. The relative retention times for Epinephrine and Norepinephrine are in correspondence with the ones indicated in the USP monograph. Table 2 Retention Time Component Retention time (min) Epinephrine sulfonic acid 2.60 0.40 Norepinephrine 5.86 0.91 Epinephrine 6.47 1.0 Adrenalone 9.08 1.40 Relative Retention time (RRT)* *) Relative retention time (RRT) with reference to Epinephrine (6.47 min). It is evident from figure 2 that the response of Epinephrine Sulfonic Acid (ESA) is significantly lower than that of the other components. This is most likely due to the fact that the optimal oxidation potential for Epinephrine Sulfonic Acid is at a higher potential. The USP monograph demands a potential setting of E= +0.65 V for the analysis, which is not necessarily the most optimal potential for all compounds. In the USP monograph for Articaine and Epinephrine the following system suitability requirements are specified: n Resolution: not less than 1.5 between the Norepinephrine and Epinephrine peak. n Tailing factor: not more than 2.0 for the Epinephrine peak. n Relative standard deviation: not more than 1% for the Epinephrine peak from 6 injections (n=6). The system suitability is evaluated using the chromatograms obtained with the standard solution of 22 μg/ml Epinephrine Bitartrate RS and 20 μg/ml Norepinephrine Bitartrate RS solution in 0.5 mg/ml Potassium Metabisulfite (system suitability solution). The results are listed in table 3, it is evident that the system suitability requirements are met for all performance parameters. Figure 2: Identification of Epinephrine and related compounds: Chromatograms of 2 μl injections of 2.5 ppm (2.5 μg/ml) solutions of: (1) USP Epinephrine bitartrate RS (black), (2) USP Norepinephrine bitartrate RS (red), (3) Epinephrine sulfonic acid (green) and (4) Adrenalone HCl (blue) in 0.5 mg/ml Potassium metabisulfite. 3

Table 3 USP System Suitability Requirement Parameter USP criteria Measured Resolution between Norepinephrine and > 1.5 2.9 Epinephrine Tailing factor (Epinephrine) < 2.0 1.1 RDS n=6 ( Peak area Epinephrine) < 1% 0.4 Linearity, repeatability and LOD The linearity of Epinephrine and Norepinephrine were investigated in the concentration range of 4 22 μg/ml (20 μg/ ml for Norepinephrine). For both components the correlation coefficients were better than 0.999 for peak areas. The relative standard deviation (% RSD) in peak area was determined for 6 replicate injections of the system suitability solution. The RSDs in peak area were 0.4% for both components. A 2.5 ppb (2.5 ng/ml) standard mix of Epinephrine, Norepinephrine, Adrenalone and Epinephrine sulfonic acid was injected to assess the Limit Of Detection (LOD) of the compounds. See figure 3. The calculated concentration LODs are listed in table 4. The LOD here is based on a 2 μl injection and defined as the concentration that gives a signal that is three times the peak-to-peak noise. Table 4 LOD Component LOD (nm) LOD (ppb) Epinephrine sulfonic acid 14.4 3.8 Norepinephrine 2.4 0.8 Epinephrine 2.4 0.8 Adrenalone 4.1 0.9 Figure 3: Chromatograms of a 2.5 ppb standard mix of epinephrine and related compounds. Injection volume 2µL (red curve) and 10 µl (blue curve). As can be seen from table 4 and figure 3 all compounds have a detection limit below 1 ppb, with the exception of epinephrine sulfonic acid. The USP monograph demands an injection volume of 2 µl. However by increasing the injection volume to for example 10 µl the concentration LOD can be improved effectively by more than a factor of 3 as demonstrated in figure 3 when required. Under the USP conditions the Limit of Quantitation is approximately 2.5 ppb (except Epinephrine sulfonic acid). Sample Analysis To evaluate the Epinephrine assay and Epinephrinerelated impurity analysis described in the USP monograph, two stressed Epinephrine samples in metabisulfite were analyzed. One sample was kept at acidic ph the other at mild alkaline conditions: (1) 0.01% (100 ppm) epinephrine sample, acidic (ph 4) (2) 0.01% (100 ppm) epinephrine sample, basic (ph 8-9) Epinephrine assay To determine the actual content of Epinephrine in the samples, 2µl of a 40x dilution of both 0.01% Epinephrine samples were injected. Based on the declared contents of 100 ppm, this corresponds to a final concentration of 2.5 ppm (2.5 µg/ ml Epinephrine). The chromatograms of both diluted samples are shown in figure 4. 4

Table 5 Epinephrine content Sample USP criteria (%) Measured (%) 0.01% Epinephrine, ph 4 90.0 115.0 90.6 0.01% Epinephrine, ph 8-9 90.0 115.0 48.1 It is evident that the more instable basic sample (ph 8-9), has a significant lower contents of Epinephrine (almost half less) than the acid sample, due to oxidation/degradation of Epinephrine. Figure 4: Chromatogram of (1) Blue curve: 0.01% Epinephrine sample ph 4, diluted 40x (2.5 ppm). Red curve: 0.01% Epinephrine sample ph 8-9, diluted 40x (2.5 ppm). Injection volume 2µL The actual content of epinephrine in both samples was calculated using the response of a standard solution of 2.5 μg/ ml Epineph-rine bitartrate RS in diluent, using the following calculation: Result= (Ru/Rs) x (Cs/Cu) x 100% Organic impurities, Limit of Epinephrine Related Compounds To determine the contents of organic impurities and epinephrine related compounds, 2μL undiluted sample solution was injected and analyzed. The chromatograms of the undiluted acidic and basic 0.01% Epinephrine samples are shown in figure 5 and 6, respectively. The figures show a zoom-in on the baseline to visualize the impurities present in the samples. In the top-right corners of the figures the full chromatograms are shown. Where: Ru = Epinephrine peak area from the of sample solution Rs = Epinephrine peak area from the of standard solution Cs = Concentration of Epinephrine in the std. solution (mmol/l) Cu = Nominal Concentration of Epinephrine in the sample solution (mmol/l) Due to the fact that the standard and sample solutions originate from Epinephrine bitartrate (M = 333.29 g/mol) and Epinephrine base (M = 183.21 g/mol), respectively, it was necessary to correct for the molar mass. So instead of the concentration in mg/ml the molar concentration was used in the calculation. The calculated actual contents (%) of epinephrine in the stressed samples are listed in table 5 below. 5

Figure 5: Blue curve: zoom-in on baseline of the chromatogram of the undilut-ed 0.01% Epinephrine sample ph 4. Top-right insert: full chromatogram. Red curve: blank injection of diluent. Injection volume 2 μl. It is evident from figure 5 that besides the epinephrine related compounds there are 3 more unknown impurities with a significant response which can be quantified in the acidic sample. The blank injection shows that the peak next to Epinephrine sulfonic acid (right side) is a system peak and not a relevant impurity. The calculated contents (%) of impurities and epinephrine related compounds in the acidic Epinephrine sample are listed in table 6 below. The calculation used is the same as described in the sec-tion above for the determination of the content of epinephrine in the samples (USP Epinephrine assay). The response of the 2.5 ppm USP epinephrine standard was used for the calculation of the percentage of the impurities. The USP acceptance criteria for the amount of impurities are: n Epinephrine sulfonate: not more than 7.5% (Relative retention time approximately 0.46). n Specified impurity: not more than 8% (Relative retention time approximately 0.52). n Any other individual impurity: not more than 1%. n Total impurities: not more than 10%. Table 6 0.01% Epinephrine sample ph 4, Impurity analysis, Limit of Epinephrine related compounds Figure 6: Blue curve: zoom-in on baseline of the chromatogram of the undilut-ed 0.01% Epinephrine sample ph 8-9. Top-right insert: full chromatogram. Red curve: blank injection of diluent. Injection volume 2 μl. Impurity RRT* Measured (%) USP criteria (%) Epinephrine sulfonate** 0.40 0.003 7.5 Unknown 1 0.49 0.02 1 Unknown 2 0.78 0.19 1 Unknown 3 0.87 0.07 1 Norepinephrine 0.90 0.05 1 Adrenalone 1.40 0.11 1 Total 0.45 10 *) Relative retention time (RRT) with reference to Epinephrine (6.47 min). **) Epinephrine sulfonic acid. 6

Note that the relative retention time of Epinephrine sulfonic acid, 0.40 is slightly lower as indicated in the USP monograph (0.46). The contents of Epinephrine sulfonic acid was < 0.005%, well below the specified limit of 7.5%. All other quantified unknown impurities, as well as the total amount are below the specified limits and within the USP acceptance criteria. The calculated contents (%) of impurities and epinephrine related compounds in the alkaline Epinephrine sample are listed in table 7. In this sample no detectable level of Epinephrine sulfonate is present. At a relative retention time of 0.44 a significant impurity peak is present. It is assumed (based on its relative position to Epinephrine sulfonate) that the peak at this retention time corresponds to the specified impurity as mentioned in the USP monograph. All quantified impurities or related compounds are below 1%, which is within the USP acceptance criteria. The total amount of quantified impurities was 0.58%, well within the USP limit of 10%. Table7 0.01% Epinephrine sample ph 8-9, Impurity analysis, Limit of Epinephrine related compounds Impurity RRT* Measured (%) USP criteria (%) Unknown 1** 0.44 0.21 8 Unknown 2 0.49 0.05 1 Unknown 3 0.50 0.09 1 Unknown 4 0.78 0.07 1 Unknown 5 0.87 0.01 1 Norepinephrine 0.91 0.12 1 Unknown 6 1.16 0.01 1 Unknown 7 1.31 0.01 1 Adrenalone 1.40 0.02 1 Total 0.59 10 *) Relative retention time (RRT) with reference to Epinephrine (6.47 min). **) It is assumed (based on its relative position to Epinephrine sulfonate) that the peak at this retention time corresponds to the specified impurity as mentioned in the USP mono-graph. Conclusion The ALEXYS Epinephrine Analyzer provides a suitable solution based on the official method of the USP for the analysis of the composition and Epinephrine-related impurities in commercial Articaine with Epinephrine injectable anesthetics. 7

References 1. S.F. Malamed, Handbook of local anaesthesia, St. Louis, Mosby, 5ed, (2004), 71 2. FDA website: http://www.accessdata.fda.gov/drugsatfda_docs/ nda/2006/022010_septocaine_toc.cfm 3. Articaine Hydrochloride & Epinephrine injection, United States Pharmacopoeia (USP), USP37-NF32 (2014), 1834 1836 4. R.J. Flanagan, D. Perret, R. Whelton, Electrochemical detection in HPLC: Analysis of Drugs and Poisons, Royal Society of Chemistry, 1ed, (2005) 5. <621> Chromatography general chapter, United States Pharmacopoeia (USP), USP37-NF32 (2014), 6376 6386 Ordering number 180.0055C ALEXYS Epinephrine analyzer, including column & flow cell 250.1070B Phenomenex Luna 5µ C8(2), 250 x 4.6 mm column Antec (USA) info@myantec.com www.myantec.com phone (888) 572-0012 (toll free) Antec (worldwide) info@myantec.com www.myantec.com phone +31 71 5813333 For research purpose only. The information shown in this communication is solely to demonstrate the applicability of the ALEXYS system. The actual performance may be affected by factors beyond Antec s control. Specifications mentioned in this application note are subject to change without further notice. 8