Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques



Similar documents
Introduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection

Elements of required physical infrastructures: space, schielding, and patient flow..

Required RS Training Info

The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.

Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):

Staff Doses & Practical Radiation Protection in DEXA

Radiation Safety Issues for Radiologic Technologists

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Radiologic Technology Program. Radiation Safety and Protection Program

Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health

PERSONNEL MONITORING AND DOSIMETRY POLICIES

portable x-ray survey report

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

Checklist for submission of application form for Registration

Radiation Safety Characteristics of the NOMAD Portable X-ray System

INFORMATION NOTICE: USE OF XRF DEVICES CONTAINING RADIOACTIVE MATERIAL North Dakota Department of Health Radiation Control Program (2/06)

The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.

PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS

RADIATION SHIELDING DESIGN 2010

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Analytical X-ray Safety Refresh Training. Radiation Safety Officer Department of Environmental Health and Safety

ACCELERATORS AND MEDICAL PHYSICS 2

SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration

GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

notes MERRTT IS-302 Radiological Survey Instruments and Dosimetry Devices

Agency for Health Care Administration

Shielding Solutions Who Is Nelco? NELCO Locations. Five Market Segments. Shielding Products Diagnostic. Professional Affiliations

Comparison of Medical and DOE Health Physics Programs. Kevin Lee Radiation Safety Officer Palmetto Health

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

Precise Treatment System Clinically Flexible Digital Linear Accelerator. Personalized radiotherapy solutions for everyday treatment care

X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Intensity-Modulated Radiation Therapy (IMRT)

General requirements. 1. Administrative Controls.

INTRODUCTION. A. Purpose

Assessing Radiation Dose: How to Do It Right

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

X-ray Safety Manual 2013

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

MEDICAL DOSIMETRY. COLLEGE OF APPLIED SCIENCES AND ARTS Graduate Faculty: therapy.

Page: 1 of 6 Page: 1 of 6

GE Healthcare. pet/ct for simulation. Precision in motion.

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

Prepublication Requirements

Diagnostic Exposure Tracking in the Medical Record

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

TITLE 32: ENERGY CHAPTER II: ILLINOIS EMERGENCY MANAGEMENT AGENCY SUBCHAPTER b: RADIATION PROTECTION

Safety and Radiation Protection Office Working with X-Ray Equipment

Radiological Imaging Risk: Joint Commission Sentinel Alert #47

Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

DEPARTMENT OF ENVIRONMENT, HEALTH & SAFETY RADIATION SAFETY DIVISION. University of California, Los Angeles RADIATION SAFETY GUIDANCE

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

Understanding Digital Modalities: System Integration and Use

NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

C-10 Radiation Monitoring Network

Physics testing of image detectors

X-ray (Radiography), Chest

Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards

The Design of Diagnostic Medical Facilities where Ionising Radiation is used

The Challenge of CT Dose Records

APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMMAND: PERMIT NUMBER: TYPE OF ACTION:

What is Radiology and Radiologic Technology?

Patient Dose Tracking for Imaging Studies. David E. Hintenlang, Ph.D., DABR University of Florida

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

Regulatory Guide. Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry

Radiation Safety Committee College of William and Mary

Selection, use and maintenance of portable monitoring instruments

Radiation Detection and Measurement

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

Status of Radiation Safety System at

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

Rouch, Jean. Cine-Ethnography. Minneapolis, MN, USA: University of Minnesota Press, p 238

Specific local rules for governing the use of the Niton XLt Environmental Analyser (portable XRF): Laboratory

Quality Control and Maintenance Programs

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

U.S. Bureau of Labor Statistics. Radiology Tech

COMMUNICATING RADIATION RISKS IN PAEDIATRIC IMAGING

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Personal Dosimetry in Dental Radiology

X-RAY REGULATORY GUIDE

Quality Assurance Testing of AtomLab Dose Calibrators

Transcription:

Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S. Towsley, M.S., DABR Minneapolis Oncology Regions Hospital St. Paul MN Frank Zink, Ph D Bruce Gerbi, Ph D Susanne Quarfoth Yerich, MS Patrick J. Silgen, MS Part 1 Documentation Source Objectives Detectors Techniques Objectives Part 2 Project Flow Sources of Derailment Case Studies Objectives and Environmental Measurements are Part of a Safety Program. Meet DOH, NRC, ALARA Regulations. Integrity. 20 February 2006 Radiology Manager Medical Center 2450 Riverside Avenue Minneapolis, MN 55454 Dear Radiology Manager, On January 23, 2006, I visited Medical Center Riverside Avenue Minneapolis, MN to perform a shielding integrity survey in Radiology Room 1 M-271 Installation. A survey was performed on walls surrounding this room. Conclusion: (continued) Materials and Procedure: A pressurized Ionization type survey meter was used to measure the radiation levels. The survey meter was calibrated on 3/05. The x-ray tube was used as the radiation source. All measurements are in mr/hr (milliroentgens per hour). The source was placed in the usual position at the level of the patient (i.e. table top) for typical exams and aimed at a bucket of water as a scattering medium. Measurements were made at each of the walls as labeled in the diagram below. Data The following drawings contain the data collected during my visit. Measurements below are the peak mr/hr s detected along the indicated wall. Actual data measurements are as follows: All tests on room M-271 showed that the radiation protection was at a safe level as designed. The room was design to a level of protection that conformed to the Minnesota DOH protection levels as stated in the shielding design report.

(continued) (continued) Data The following assumptions were made for the workload in this room. Exposure Exposure Table Chest Rate Wall Area Type Top Stand kvp mas (s) mr/hr A-B Control Scatter x 121 100 2.5 0.027 A-B Control Scatter x 90 100 3.5 0.071 B-C Hallwy/ Xray Rm Scatter x 121 100 2.5 0.007 C-D-E Toilet Scatter x 121 100 2.5 0.007 E-F Corridor Scatter x 121 100 2.5 0.035 E-F Corridor Primary x 121 100 2.5 0.017 F-G Hallway Scatter x 90 100 3.5 59 F-G Hallway Primary x 90 100 3.5 129 F-G Hallway Scatter x 121 100 2.5 3 F-G Hallway Scatter x 121 100 2.5 185 Technique X-Table Hip PA Chest Lat Chest X-Table PA Lat Procedures Procedures Procedures Hip Chest Chest per per per kvp mas mas mas Week Week Week 121 12 30 50 50 90 350 6 (continued) The resulting expected maximum exposure would be: Hip Chest Total Exposure Exposure Exposure Exposure per Week per Week per Week Wall Area Type mr/mas mr mr mr A-B Control Scatter 1.88E-07 0.00 A-B Control Scatter 6.90E-07 0.00 0.00 B-C Hallwy/ Xray Rm Scatter 4.86E-08 0.00 0.00 C-D-E Toilet Scatter 4.86E-08 0.00 0.00 E-F Corridor Scatter 2.43E-07 0.00 E-F Corridor Primary 1.18E-07 0.00 0.00 F-G Hallway Scatter 5.74E-04 1.20 F-G Hallway Primary 1.25E-03 2.63 F-G Hallway Scatter 2.08E-05 0.04 F-G Hallway Scatter 1.28E-03 2.70 6.58 (continued) The control room and walls A through F receive little measurable radiation. The doors labeled A-F can expect at most about 7 mr/week. Other radiographs are expected to add very little to this area. Because this area is a hallway, any person in the area would be expected to receive 1/16 th of this, or about 0.5 mr/week. This is below the Minnesota State DOH protection ALARA level of 10 mr/week. It has been a pleasure working with you and being able to supply this service. If there is any question about this report or anything else I can help with please call me at ((Area Code) Telephone Number) Sincerely, Mark Towsley, M.S., DABR Certified Diagnostic Medical Physicist Diagnostic Radiology C B E D A A F X-Ray Tube Chest Room, CT, Radiographic Room Flouroscopy Flouroscopy Room, Portable C-arm Tc-99m Source

Diagnostic Radiology Data X-Ray Tube Exposure Exposure Table Chest Rate Wall Area Type Top Stand kvp mas (s) mr/hr A-B Control Scatter x 121 100 2.5 0.027 A-B Control Scatter x 90 100 3.5 0.071 B-C Hallwy/ Xray Rm Scatter x 121 100 2.5 0.007 C-D-E Toilet Scatter x 121 100 2.5 0.007 E-F Corridor Scatter x 121 100 2.5 0.035 E-F Corridor Primary x 121 100 2.5 0.017 F-G Hallway Scatter x 90 100 3.5 59 F-G Hallway Primary x 90 100 3.5 129 F-G Hallway Scatter x 121 100 2.5 3 F-G Hallway Scatter x 121 100 2.5 185 Diagnostic Radiology Tc-99 Reading Reading Normalized Reading Normalized Reading Calculated Calculated Location Inside room Outside room Inside room Outside room Transmission Transmission mr/hr mr/hr mr/hr/mci mr/hr/mci Tc-99m 120 kvp Wall A-B 0.075 0.005 0.00201 0.00002 1.0% 0.1% Wall B-C 0.088 0.012 0.00237 0.00016 6.9% 0.9% Wall C-D 0.055 0.008 0.00144 0.00008 5.6% 0.7% Wall D-E 0.077 0.02 0.00206 0.00033 15.8% 2.0% Door A-E 0.095 0.021 0.00257 0.00035 13.4% 1.7% Calculated Transmission of lead at 120 kvp is the ratio of µ for Tc-99m to effective energy of a 120 kvp spectrum 40 kev Source: 10 mci of Tc-99m used Detector: Pressurized Ion Chamber Nuclear Medicine Tc-99m F-18 Or other source for a dedicated room Nuclear Medicine Net Rdg (mr/hr) Location 0 Control console 0 entrance to scanner room (door closed) 0.18 entrance to scanner room (door open) 0 Floor outside scanner room (door closed) 0.14 Entrance to uptake room 4 feet above floor (door closed) 0.13 Entrance to uptake room @ floor (door closed) 0.03 Corridor outside uptake room 4 feet above floor 0.1 Max reading corridor outside of uptake room 4 feet above floor Area outside of uptake room and scanner room (5 feet from each 0.1 door) 0.03 Downstairs storage room @ end of hallway (6 feet above floor) 0.18 Downstairs Radiologist breakroom (5 feet above floor) Max reading corridor outside Radiologist conference room (4 feet 0.08 above floor) 0.1 Max reading downstairs utility room 0.23 Max reading in computer room Compliments of Pat Silgen MS Methodist Hospital, Mpls MN Oncology Oncology Gamma Knife Gamma Knife Co-6o Source Linear Accelerator # Location mr/hr 1 Edge of couch 600 2 Foot of couch 47 3 Office 0.15 4 Office 0.10 5 Control console - work position BKG 6 Control console knee holes BKG 7 Treatment room doors @ 12 0.05 8 Treatment room doors @ 36 0.03 9 Exam/Recovery BKG 10 Corridor BKG 11 LINAC treatment room BKG 12 LINAC treatment room BKG 13 LINAC treatment room BKG 14 *2 nd floor above treatment room BKG Compliments of B. Gerbi, Ph. D. University of Minnesota

Oncology Linac Room for Elekta Precise Room 1 Hospital Name Address City, State A radiation survey on the Elekta Precise Accelerator Room 1 was performed to assure the shielding design integrity. This survey was performed on 9 February 2004. Accelerator: Elekta Precise Duel energy 6, 10 MV photons instrument: Inovision 451P ur meter Calibrated March 2003 Setup conditions: 10 MV photons. 30x30 field size 30 cm attenuating material for scatter for scatter Dose rate 450 cgy/min. Results: See Table Oncology Sample Calculations: Elekta dose rate of 4.5 Gy/hr Workload of 500 Gy/wk Beam on per week = 500 Gy/wk = 111 min/wk 4.5 Gy/min Assume occupancy of 1 Use factor of 100% for scatter Exposure to personnel at Elekta Precise control area Exposure = 530 ur/hr *10-5 msv/ur* 111 min/wk * 1 * 1 = 9.8 x 10-3 msv/wk 60 hr/min Allowable limit: Occupational exposure = 0.1 msv/wk General public exposure = 0.02 msv/wk Linac Oncology Dose Rate Gantry Reading Workload Occupancy Use Exposure Limit Gy/min factor factor msv/wk msv/wk Location Orientation Type Area ur/hr msv/hr Gy/wk Results 30x30 Field Size Coll=0 30 cm Attenuating Material Control Down Scatter Controlled 530 (0) 0.00530 500 4.5 1 1 0.0098 0.1 Pass Outside Secondary Lat (270) Secondary Uncontrolled 1000 0.01 500 4.5.0625 1 0.001 0.02 Pass 30x30 Field Size Coll=45 no Attenuating Material Outside Primary Lat (270) Primary Uncontrolled 29000 0.29 500 4.5 0.0625 0.25 0.0084 0.02 Pass Oncology Environmental Exposure Report Date Placed Monitors: 9/12/2006 Date Removed Monitors: 12/7/2006 Beam on hrs Placement of Monitors msv/ 3 Months readings msv per week Outside Secondary 0.13 0.011 Outside Primary 0.18 0.015 Oncology Environmental Exposure Report Placement of Measured msv per week Adjusted msv per week msv per week Monitors using OSL Occupanc using OSL using ion badges y Factor badges rate meter Outside Secondary 0.011 0.0625 0.0007 0.001 Outside Primary 0.015 0.0625 0.0009 0.008 Detectors Objective Practical Detector Function Detector Type Energy Response Response Handling Background Readings

Detectors Objective Detectors Practical integrity ed protection limits Federal/State/Local protection limits Readily available Satisfies and meets the needs regulations Demonstrates reproducibility Demonstrates accuracy Demonstrates linearity Detectors Detector Function Detectors Detector Type Rate Meter Integrating GM Ion Chamber/Pressurized Ion Chamber Thermoluminescent Dosimeter (TLD) Optical Stimulated Luminescent Dosimeter(OSL) Detectors Response Energy Response Detectors Handling Background Readings 2σ For low level radiation measurements you can assume a poisson distribution where the SD = mean. Therefore in order to be statistically significant to the 95% level, the readings must be 2 SD apart.

Project Line Idealized Inception Finance Need Equipment Building Detraction Estimate Physicist Input Project Line Idealized Project Line Idealized Physics Approval Construction Construction Certificate of Occupation Installation -Therapy - Diagnostic State DOH - Nuclear Medicine Approval Visual inspection Visual inspection Project Line Idealized Project Line Idealized Acceptance Installation Commission -Therapy - Diagnostic Annual - Nuclear Medicine

What Really Happens Where do things break down. Stuff Happens. Inception Need Finance Equipment Building Detraction Estimate Physicist Input Project Line Stuff Happens Project Line Stuff Happens Physics Approval State DOH Approval Construction Visual inspection Project Line Stuff Happens Project Line Stuff Happens Acceptance Installation Commission -Therapy - Diagnostic Annual - Nuclear Medicine

Inception Need Equipment Finance Building Detraction Estimate Physicist Input Over Shielded Poor Both the primary walls and the ceiling contain 6 6 of concrete. Since 1992, adjacent occupancy around and above has been zero. Below Grade Below Grade Poor Ceiling Project Line Examples HAVC and Electrical Conduit Run Maze Lead Concrete L e a d D o o r Outside door 3 mr/hr 0.7 mr/hr

Iterative Process Iterative Process Iterative Process B C D Monitoring Period: 5 Jan 07 6 April 07 Low Energy machine 6x photons only Total Beam on : 7.83 hrs Total Isocenter Dose: 93,960 cgy Average # patients: 8 patients per day Integrating Dosimeter: Luxel Badge Meter: Inovision 451 P Project Line Examples Acceptance Commission Installation -Therapy - Diagnostic - Nuclear Medicine Missed Project Line Examples Flouroscopy Mammography Physician Office Annual

Changes All This department opened n 1999. In 2007 the unit was replaced and a radiation survey was performed. A construction worker was able to look down from the top of the wall and see the over time the lead seam had separated. 4 mr/hr Environmental Type s using a dose rate meter will readily identify areas of peak dose rates and alert you to potential problems. These measurements tend to be quit sensitive at low radiation levels. However, they do not represent the inherent use factor. s using long term integrating dosimeters contain the use factors but are limited to placement and sensitivity in low radiation levels. Compliments of Susanne Quarfoth Yerich, MS Environmental Measurements Conclusion Be Involved. Have a goal and state it in the report. Select the dosimeter and radiation source to match that goal.