Performance of a portable, Digital-Signal-Processing MCA with Safeguards Germanium Detectors



Similar documents
An Innovative, portable MCA based on Digital Signal Processing. R.D. Bingham, R.M. Keyser, T.R. Twomey

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

MAESTRO 7. Multichannel Analyzer (MCA) Application Software

IDM-200 -V. Interchangeable HPGe Detector Module (IDM) for Systems Integrators

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

GammaVision 8. Gamma Spectroscopy Software. Compatible, Efficient, and Defendable Results for Gamma Spectroscopy Applications.

Networks and Connections in the Counting Room

ORTEC. Micro-trans-SPEC. Ultra-Light, Battery-Powered, Portable, LN 2 -Free, HPGe Gamma Spectrometer


A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials

AlphaVision. Alpha Spectrometry Management Software

Reconfigurable System-on-Chip Design

RADIATION DETECTION & ANALYSIS SYSTEMS FOR WATER

Automated Laboratory Quality Assurance Program: Using the ORTEC GammaVision -32 Software

WM2012 Conference, February 26 March 1, 2012, Phoenix, Arizona, USA

Global Value 7. Productivity Suite for GammaVision. Optimizing Gamma Spectrometry Processes through Secure Data Management and Measurement Automation.

Description: Multiparameter System (4 or 8 channels)

Nuclear applications at the Hellenic Centre for Marine Research (HCMR): Radioactivity in the marine environment

Tutorial 4.6 Gamma Spectrum Analysis

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

Application Note. Monitoring the Release of Radioactive Noble Gases Through the Stack of a Nuclear Power Plant (NPP): Stack Monitor System

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

Measurement solutions for nuclear safety and security. software customization services. Nuclear MeasureMeNts BusiNess unit of areva

thyroid uptake systems

ORTEC. High Resolution Detection Systems for Interdiction of Nuclear Material Trafficking. Abstract. Introduction

A Digital Spectrometer Approach to Obtaining Multiple Time-Resolved Gamma-Ray. Spectra for Pulsed Spectroscopy. W. K. Warburton a

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

RF Measurements Using a Modular Digitizer

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

FREQUENCY RESPONSE ANALYZERS

S2000 Spectrometer Data Sheet

POCKET SCOPE 2. The idea 2. Design criteria 3

Measuring Laser Power and Energy Output

Designing the Solution for Electrochemistry Potentiostat/Galvanostat І Battery Cycler І Fuel Cell Test Station І

Appendix 5 Overview of requirements in English

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

WM2012 Conference, February 26 -March 1, 2012, Phoenix, Arizona, USA

Application Note. Line Card Redundancy Design With the XRT83SL38 T1/E1 SH/LH LIU ICs

Figure 1.Block diagram of inventory management system using Proximity sensors.

DEVELOPMENT OF DEVICES AND METHODS FOR PHASE AND AC LINEARITY MEASUREMENTS IN DIGITIZERS

WIRELESS INSTRUMENTATION TECHNOLOGY

Scalar Network Analyzer

The Universal DAQ Device. Connect and measure immediately!

W a d i a D i g i t a l

Optimizing VCO PLL Evaluations & PLL Synthesizer Designs

Hideo Okawara s Mixed Signal Lecture Series. DSP-Based Testing Fundamentals 46 Per-pin Signal Generator

LLRF. Digital RF Stabilization System

Impedance 50 (75 connectors via adapters)

ORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Remote RadEye Product Family

EDM Cloud Remote Monitoring Solutions. Crystal Instruments, January 2014

Captus 3000 Thyroid Uptake And Well System

Design and Certification of ASH Radio Systems for Japan

INTREPID Perimeter Intrusion Detection System

Laboratory Technologies, Inc. Multi-Wiper. Time Saving Multiple Detector Wipe Counter

Perimeter Security System

MicroMag3 3-Axis Magnetic Sensor Module

Electronics GRETINA Project

Hello and welcome to this training module for the STM32L4 Liquid Crystal Display (LCD) controller. This controller can be used in a wide range of

Algorithms for digital γ-ray spectroscopy

Alpha/Beta Counting Manual Systems

NETWORK ENABLED EQUIPMENT MONITOR

AN_6521_035 APRIL 2009

ACR880 GPRS Portable Smart Card Terminal

Reducing EMI and Improving Signal Integrity Using Spread Spectrum Clocking

AGRI-ALERT 128 TOUCH AGRI-ALERT 800EZE ALARM SYSTEMS

Section 3. Sensor to ADC Design Example

Spectrum Analyzer Two models available: OGR-24 (24 GHz) and OGR-8 (8 GHz)

Realtime FFT processing in Rohde & Schwarz receivers

1 Multi-channel frequency division multiplex frequency modulation (FDM-FM) emissions

Synchronization of sampling in distributed signal processing systems

High Power Programmable DC Power Supplies PVS Series

Vector Signal Analyzer FSQ-K70

Measuring Temperature withthermistors a Tutorial David Potter

MH - Gesellschaft für Hardware/Software mbh

Multi-Range Programmable DC Power Supplies 9115 Series

BUILD VERSUS BUY. Understanding the Total Cost of Embedded Design.

Simple SDR Receiver. Looking for some hardware to learn about SDR? This project may be just what you need to explore this hot topic!

STM32L. Ultra-low-power Cortex -M3 devices

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

The Heartbeat behind Portable Medical Devices: Ultra-Low-Power Mixed-Signal Microcontrollers

Rapid Audio Prototyping

NEW. EVEN MORE data acquisition and test stand automation

Silicon Drift Detector Product Brochure Update 2013

Lecture N -1- PHYS Microcontrollers

Kit Highlights and Key Features

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

Chlorophyll Content Meter. CCM-200 plus

Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)


Application Note Noise Frequently Asked Questions

Thyroid Uptake Systems The Ultimate... User Friendly... Intuitive Captus 3000 Thyroid Uptake And Well System

CANFI: Cheap Automatic Noise Figure Indicator. Frank Schmäling DL2ALF Wolf-Henning Rech DF9IC Alexander Kurpiers DL8AAU

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

Transcription:

Performance of a portable, Digital-Signal-Processing MCA with Safeguards Germanium Detectors Bingham, R. D., Keyser, R. M., Twomey, T. R. ORTEC, PerkinElmer Instruments, Inc 801 South Illinois Avenue Oak Ridge, TN 37831 Abstract The use of digital signal processing (DSP) has become the standard for high-performance laboratory instruments for measurements in the field of nuclear spectroscopy. In all applications it offers greater stability, improved resolution and greater throughput at high count rates compared to analog or traditional designs. All of these improve the quality of the spectrum and the analysis results. Safeguards applications, especially the isotopic-ratio calculation, depend on the high resolution and stability of the MCA and detector for accurate results. Recently, advances in low-power DSP modules have made it possible to incorporate DSP in useable, portable instruments. A portable multichannel analyzer system consisting of a HPGe detector, with integral bias supply, and a MCA has been developed. This system includes all the benefits of digital processing in a complete instrument for spectroscopy. In addition, the unit includes a local display and keypad, to eliminate or significantly reduce the need for a PC in many field applications. When used with the associated "intelligent" HPGe detector system, the instrument monitors 18 operating parameters of the system, including detector temperature, to derive the overall system state of health. In addition, the unit provides authentication of the entire system and the collected data. The system has been tested using both liquid-nitrogen and electrically cooled detectors. The peak position and resolution have been measured using the commonly available mixed nuclide sources. Performance data for peak resolution and position vs countrate, peak resolution and position vs time, using Safeguards-quality HPGe detectors will be presented. Introduction Since their introduction, DSP-based Gamma-ray spectrometers have become the industry standard and are now widely used both standalone and within larger measurement systems. It has rapidly become clear that the DSP technology had no performance disadvantages relating to spectroscopy; indeed, in every case, they have shown to offer improved resolution, throughput and stability, both over long time periods and wide temperature changes. The wide choice of operating parameters coupled with a high degree of setup automation available, in some cases, means that the instrument performance is easily tuned to make the optimum use of this performance in a specific application situation. Not surprisingly, a wide variety of systems have been developed incorporating these digital spectrometers, in counting laboratories, remote monitoring, industrial on-line applications and non-destructive assay applications. DSP systems are proving capable of meeting the most demanding of applications. 1

The recent arrival of low-power, digital-signal processing technology has now made it feasible to incorporate the high performance expected of a digital spectrometer into a small, battery-operated package. The portable DSP systems can show equivalent performance to the laboratory based instruments, as we shall see. From a historical perspective, introduced almost 20 years ago, the LANL-developed Davidson PMCA 2 with on-board cassette tape storage, local display and some limited smart capabilities has been the instrument of choice for safeguards inspectors across the world. It has been used for routine measurements such as uranium enrichment for many years; it still provides a unique combination of features in a single integrated and portable package. Many hundreds are currently in use. Its attraction is in its combination of useful features: Battery Operated On Board Display and spectral data storage Full work shift operation (more than 10 hours) without change of battery Good quality spectroscopy (for a 1982 vintage portable system) Ability to do some calculations without an attached PC. Until now, the recent-generation of small portable MCAs, have no on-board bulk storage, display, keypad or computing power and are blind peripherals, they require the use of an associated small computer. The new instrument reported here, with the trade-name digidart TM is being produced by PerkinElmer Instruments (ORTEC). It combines the high performance attributes of the digital signal processing in a small, low-power, battery-operated package. The instrument includes an on-board display and the ability to store multiple spectra without the need for an accompanying PC. It can be viewed as a state-of-the-art-technology successor to the PMCA. In addition, the instrument includes new SMART-1 HPGe detector technology, which can monitor the system state of health and provide authentication of spectral data, very valuable in remote or unattended monitoring in a single instrument. Therefore, it can address a variety of existing and developing use needs. The wide range of applicability means that this instrument can help reduce the inventory of different MCA types required to support the spectrum of safeguards applications. Circuit Description The MCA hardware is shown schematically in Figure 1. The high voltage supply and detector monitoring are provided in a separate detector interface module (DIM). Following the front-end amplifier, a 14-bit, 10 MHz flash ADC samples the incoming pulse stream and converts it into a string of digital numbers. This is then filtered directly by a proprietary digital filter algorithm, implemented in a field-programmable gate array (FPGA). This also provides the functions of digital baseline restorer, fine gain, peak qualification, conversion gain, digital upper and lower discriminators, and spectrum stabilization. System control, keypad communications, spectral display and control of USB and/or RS232 communication is provided by the microprocessor. The microprocessor also controls the detector high voltage and provides the detector monitoring

function. The battery-backed histogram memory provides up to 16,000 channels of non-volatile data storage with optional 32,000 channels. A flash memory provides for non-volatile storage of multiple spectra. Over 600 512-channel spectra may be stored. A digital version of the Gedcke-Hale 3 livetime clock is implemented. The digital trapezoidal filter provides 99 rise time (RT) and 16 flat top (FT) choices in the ranges 0.8 S to 20 S and 0.3 S to 2 S respectively. This range of choices is wider than other DSP units to allow precise optimization of detector settings. A digital Automatic Pole Zero circuit 4 is provided, along with an optimize function to automatically optimize the flat top tilt as well as Pole Zero across the range of rise time/flat top combinations. Figure 1 Portable MCA Schematic Block Diagram Battery and Power Management The instrument is powered by a single Sony NP-F960 camcorder battery, which will operate for up to 12 hours including supplying the detector power. The battery may be replaced without loss of spectral data and the instrument restarted in seconds. The battery management circuitry is built into the instrument. Any 12V DC external supply will charge the battery and operate the instrument. Physical The instrument is packaged in an impact-resistant, ABS plastic enclosure, with rubberized overmolding, as shown in Figure 2. The outside overmold is a shock-absorbing, easy grip, stay clean coating. The display and sealed membrane keypad are recessed to protect them if the instrument is dropped. The easy-to-grip enclosure will fit in a box 20 cm x 12.7 cm x 7.7 cm and the weight including battery is <850 gm. Display and Built-in Analysis functions The 240 x 160 pixel backlit LCD (Figure 3) provides live spectral display, status information and analysis results. It features full display, display of multiple ROIs and Zoom modes. Figure 2 The portable MCA in handheld use

A Status line provides parameter display; with the user able to choose any two from: cursor energy, location, live time, live time remaining, real time remaining, battery life remaining, Count rate, Count rate in ROI, or counts. A real time Peak Information function reports centroid, FWHM, Net and Gross area for the ROI where the cursor is positioned. A two-point energy calibration of channel and FWHM can be done at any time from the keypad. The user can define from 1 to 9 regions for the Nuclide activity display. The activity with uncertainty is calculated and reported in real time. Multiple acquisition presets are provided: live time, real time, integral peak count, peak area, MDA and uncertainty. SMART-1 HPGe Detector Support Figure 3 The portable MCA spectral display The MCA supports HPGe detectors with the SMART-1 technology, as well as traditional HPGe detector systems. The SMART-1 HPGe detector supports many advanced features, as described in another paper 5. Associated Software The unit is supported by the ORTEC standard MAESTRO TM MCA Emulator. The easy setup has been extended to support all the instrument functions via the USB hardware interface. All of the setup can be done by MAESTRO or by the standalone digidart. The philosophy is that it is easier to setup the instrument for use back at base using a large-screen PC, but all the settings may be changed in the field, where there is no PC available. The settings can be protected by password, so the settings can not be inadvertently changed. Two levels of passwords are used to allow different levels of access control. If either level of password is set, the password must be entered when the MCA is powered on. The instrument is compatible with all of the ORTEC analysis packages, including versions of MGA++ and PC/FRAM Isotopic ratio packages, and products for the ISOTOPIC analysis of soils and containers as well as general purpose gamma-analysis packages. To assist the developers of custom or in-house systems, programmer toolkits are available 6. Performance data In a recent LANL evaluation 7, the latest ORTEC digital laboratory spectrometer, the DSPEC Plus was evaluated and considered to be highly recommended. It may therefore be considered as a suitable digital benchmark for comparison purposes. The ORTEC 672 Spectorscopy Amplifier has long been considered as a performance benchmark for use in analog systems. In figure 4, the digidart performance was compared to the DSPEC Plus and an analog system incorporating the 672 and a Canberra 8076 450 MHz Wilkinson ADC under conditions which are similar to those pertaining to many safeguards measurements. The data were taken with an ORTEC 500 mm 2

safeguards planar detector 8. The data in Figure 4 show clearly that the DigiDART is capable of equivalent performance to the highest quality lab systems, both analog and digital. Even at 100 khz input rate, the peak FWHM from the digidart is 600 ev, which meets the stringent requirements of the isotopic codes such as PC/FRAM and MGA. At 100 khz, the Digidart shows almost 10 and 51% higher throughput than the DSPEC + and NIM Analog systems respectivley, with almost equal resolution (only 1.5% or 9 ev worse than the NIM system). Figure 4 Measurements were also made with an ORTEC Safeguards GEM 25, 25% relative efficiency, coaxial detector with the following warranted specifications: Model Active Diamete r (mm) Thicknes s (mm) Energ y (kev) FWHM @ 1 kcps (6 s shaping) FWHM @ 1 kcps (2 S shaping) FWHM @ 30 kcps (2 S shaping) SGD-GEM-25175 50 50 122 750eV 870 ev 880 ev The SGD-GEM detector type was developed specifically to meet the needs of PC/FRAM in high-energy mode. A smaller version of this detector has recently been tested for use with MGA in so-called 2-detector mode, with successful results 9. This detector and a larger nominally 50% relative efficiency type have been specified for use in the Los Alamos Tomographic Gamma Scanner (TGS) 10. Figure 5 shows total throughput curves as a function of input count rate. All except the upper curve were measured with an 0.8uS filter flat top The upper 1332 1.75Kev 1.95 kev 2.0 kev Figure 5 Total system throughput performance: coaxial detector

curve was measured with an 0.3uS flat top to demonstrate the ultimate throughput of the system to exceed 100KHz. This produced a 122keV peak with a resolution of 3KeV (not shown), which was very stable in FWHM and in position with increasing count rate. While of academic interest here, such a performance trade-off may be valuable elsewhere, where throughput is more important than resolution. Figure 6 shows that the system resolution performance meets the coaxial detector warranted 1KHz count rate performance specifications of 870 ev at 2uS shaping (4uS risetime) even at 30 khz. The resolution is seen to degrade towards shorter shaping times due to the charge collection time in the coaxial detector crystal being longer than the integration time. Figure 7 shows the variation in 122 kev resolution with count rate and rise time. Figure 6 In general, the shorter shaping times give poorer resolution and better stability as would be expected. Figure 7 For the 4 S data, even at the point of maximum throughput of 28 khz at 75 khz input rate (Figure 5), Figure 7 shows that the resolution at 900 ev is only 20 ev worse than the detector warranted value at 30 khz input rate. Figure 8 shows peak shift versus count rate. The stability for short shaping times is better, the tradeoff being resolution. Figure 8

The implication of figures 5-8 is the following: The resolution requirement in a spectroscopy problem is defined by the complexity of the spectrum, especially true in isotopic ratio measurements. This defines what shaping time must be employed for a give detector. The question may then arise as to how fast data may be taken, (how quickly can the measurement be made), or over what range of count ratescan the system operate without excessive peak shift or resolution degradation. (The optimum count rate to take data is the point of maximum throughput). What is shown here is that the digidart makes it possible to fine tune filter settings to the best effect, and that once done, the count rate stability is excellent.the system continues to produce viable data from low rates to beyond the point of maximum throughput, with peak shift of less than 40 ev (0.03%) across the range. Figure 9 Figures 9 and 10 show peak position and resolution at low- and high- energy Figure 10 over time. A mixed isotope source was acquired over a period of 5 days: 120 spectra. The count-rate was approximately 1 khz. The filter setting was 12 S risetime, 0.8 S flat top. The 59 kev and 1332.5 kev peaks were monitored. At 1332 kev the peak position and resolution drift standard deviations were 0.002% and 2% respectively, while at 59 kev the corresponding results were 0.005% and 0.9%. Conclusion A new and innovative digital high performance portable MCA has been developed and tested. In spite of the demands of designing for low power, it has been shown to provide performance at least equivalent to the best laboratory grade, digital spectrometer and the best analog systems. In practical terms there is no longer any trade off in performance for a field system. Indeed this system can be effectively used in fixed installations, where small size/low power, battery backup, local multiple storage, networkability, detector self-diagnostics and data authentication can be assets to the measurement. Its packaging and display provide superior ergonomic features, whereas multi-spectral

storage and analysis capabilities extend the usefulness of handheld MCAs. Acknowledgements: The authors wish to acknowledge Duc Vo of LANL for permission to use his unpublished data, and to Nancy A. Slater of ORTEC for taking the remainder of the data. References 1. Fuel INtegrity Evaluation and Surveillance System (FINESS), L Sihver, D. Upp and B Bjurman,.Paper presented at IEEE Nuclear Science Symposium in 1999 2. J.K. Halbig and S. Klosterbuer The Mini-MCA: An Intelligent Inspection Instrument. Presented at the International Symposium on Recent Advances in Nuclear Material Safeguards, Vienna, Austria 1982 3. R. Jenkins, R.W Gould and D.Gedcke, Quantitative X-ray Spectrometry (New York: Marcel Dekker Inc.), 1981 pp.243-267 4. US Patent No. 5,872,363 5. Twomey, T. R., Bingham, R. D., Keyser, R. M. Authentication and data quality monitoring with Safeguards HPGe detector systems, to be published 6. R.M Keyser and T.R Twomey, Networks and Connections in the Counting Room, ESARDA 21 st Annual Meeting Sevilla, Spain 1999 7. D.T Vo, Comparions of the DSPEC and DSPEC plus Spectrometer Systems, Los Alamos report LA-13671-MS, 1999 8. Data taken by D.T. Vo, Los Alamos Technical Report to be published 9. T-F Wang, K Raschke, W.D Ruhter, T.R.Twomey, Two Detector Mode MGA Analysis of Plutonium Gamma-Ray Standards using a Single Ge Detector. ESARDA 23 rd Annual Meeting, Brugge, Belgium 2001 10. For example, see S. Dittrich, J.A. Mason, A Towner, M.I Thornton, A. Ravazzani, and P Schillebeeckx, Implementation and Testing of a universal Gamma Scanner, ESARDA 23 rd Annual Meeting, Brugge, Belgium 2001